Professional Documents
Culture Documents
Mazen Kherallah, MD, FCCP Consultant Intensivist King Faisal Specialist Hospital
2. Control Variables
Pressure, volume, dual
3. Phase Variables
Trigger, Limit and Cycle
Lung Mechanics
Equation of Motion
ventilation pressure =
(to deliver tidal volume)
resistive pressure
(to make air flow through the airways)
elastic pressure
(to inflate lungs and chest wall)
Phase Variables
Trigger (start)- begins inspiratory flow (start) Cycling (end)- ends inspiratory flow (end) Limiting (continue)- places a maximum value on a control (continue)variable
pressure volume flow time
Spontaneous
Patient takes on work Patient controls start and stop
Pressure controlled
flow and volume may vary
Volume/Flow Control
Inspiration
20
Pressure Control
Inspiration
20
Expiration
Expiration
Paw
Pressure
0 20 1
Paw
2 0 0 20 1 2
Volume
0 3 0 1 2 0 3 0 1 2
Flow
Time (s)
Time (s)
-3
-3
Paw
cmH20 -20 120
SEC
6
INSP
Flow
L/min
SEC
6
EXH
120
Manufacturer; ventilator
VIASYS Healthcare; Bird 8400Sti and Tbird VIASYS Healthcare; Bear 1000 Siemens; servo 300 Cardiopulmonary corporation; Venturi Siemens; servo 300 Hamilton; Galileo Drager; Evita 4 Cardiopulmonary corporation; Venturi Hamilton; Galileo
Name
VolumeVolume-assured pressure support Pressure augmentation Volume support Variable pressure support PressurePressure-regulated volume control Adaptive pressure ventilation Autoflow Variable pressure control Adaptive support ventilation
Dual control breath to breath: PressurePressure-limited flow-cycled flowventilation Dual control breath to breath: PressurePressure-limited time-cycled timeventilation
Bear 1000
Bird 8400Sti
Tbird
40
Paw
cmH20
-20 0.6
Volume
L
Set tidal volume cycle threshold Tidal volume Tidal volume not met met
Flow cycle
Flow
L/min
60
trigger
Pressure at Pressure support
no
flow= 25% peak
yes
delivered VT set VT
yes
Cycle off inspiration
no
Insp flow > Set flow
yes
no
Switch to flow control at peak flow setting
no yes
delivered VT = set VT
no
yes
Dual Control Breath-to-Breath Breath-topressure-limited flow-cycled ventilation pressureflowVolume Support Tidal volume is used as feedback control to adjust the pressure support level All breaths are patient triggered, pressure limited, and flow-cycled. flow Automatic weaning of pressure support as long as tidal volume matches minimum required VT (VT set in a feedback loop to adjust pressure).
Servo 300
Maquet Servo-i
VS vs VAPS
How does volume support differ from VAPS ?
In volume support, we are trying to adjust pressure so that, within a few breaths, desired VT is reached. In VAPS, we are aiming for desired VT tacked on to the end of a breath if a pressurelimited breath is going to fail to achieve VT
VS (Volume Support)
Entirely a spontaneous mode Delivers a patient triggered (pressure or flow), pressure targeted, flow cycled breath
Can also be timed cycled (if TI is extended for some reason) or pressure cycled (if pressure rises too high).
Similar to pressure support except VS also targets set VT. It adjusts pressure (up or down) to achieve the set volume (the maximum pressure change is < 3 cm H2O and ranges from 0 cm H2O to 5 cm H2O below the high pressure alarm setting Used for patients ready to be weaned from the ventilator and for patients who cannot do all the WOB but who are breathing spontaneously
VS (Volume Support)
(1), VS test breath (5 cm H2O); (2), pressure is increased slowly until target volume is achieved; (3), maximum available pressure is 5 cm H2O below upper pressure limit; (4), VT higher than set VT delivered results in lower pressure; (5), patient can trigger breath; (6) if apnea alarm is detected, ventilator switches to PRVC
yes
no
calculate compliance
trigger
yes
cycle off
no
Dual Control Breath-to-Breath Breath-topressure-limited flow-cycled ventilation pressureflowVolume Support Little data to show it actually works. If pressure support level increases to maintain TV in pt with increased airways resistance, PEEPi may increase. If minimum TV set too high, weaning may be delayed.
VS (Volume Support)
Indications
Spontaneous breathing patient who require minimum E Patients who have inspiratory effort who need adaptive support Patients who are asynchronous with the ventilator Used for patient who are ready to wean
VS (Volume Support)
Advantages
Guaranteed VT and E Pressure supported breaths using the lowest required pressure Decreases the patients spontaneous respiratory rate Decreases patient WOB Allows patient control of I:E time Breath by breath analysis Variable I to meet the patients demand
VS (Volume Support)
Disadvantages
Spontaneous ventilation required VT selected may be too large or small for patient Varying mean airway pressure Auto-PEEP may affect proper functioning Auto A sudden increase in respiratory rate and demand may result in a decrease in ventilator support
Dual Control Breath-to-Breath Breath-topressure-limited time-cycled ventilation pressuretimePressure Regulated Volume Control
Servo 300
Maquet Servo-i
Dual Control Breath-to-Breath Breath-topressure-limited time-cycled ventilation pressuretimePressure Regulated Volume Control Delivers patient or timed triggered, pressure-targeted pressure(controlled) and time-cycled breaths time Ventilator measures VT delivered with VT set on the controls. If delivered VT is less or more, ventilator increases or decreases pressure delivered until set VT and delivered VT are equal
Dual Control Breath-to-Breath Breath-topressure-limited time-cycled ventilation pressuretimePressure Regulated Volume Control
The ventilator will not allow delivered pressure to rise higher than 5 cm H2O below set upper pressure limit
Example: If upper pressure limit is set to 35 cm H2O and
the ventilator requires more than 30 cm H2O to deliver a targeted VT of 500 mL, an alarm will sound alerting the clinician that too much pressure is being required to deliver set volume (may be due to bronchospasm, secretions, changes in CL, etc.)
PRVC. (1), Test breath (5 cm H2O); (2) pressure is increased to deliver set volume; (3), maximum available pressure; (4), breath delivered at preset E, at preset f, and during preset TI; (5), when VT corresponds to set value, pressure remains constant; (6), if preset volume increases, pressure decreases; the ventilator continually monitors and adapts to the patients needs
yes
no
calculate compliance
trigger
yes
cycle off
no
Paw
cmH20 -20 120
SEC
6
INSP
Flow
L/min
SEC
6
EXH
120
Paw
cmH20 -20 120
SEC
6
INSP
Flow
L/min
SEC
6
EXH
120
Paw
cmH20 -20 120
SEC
6
INSP
Flow
L/min
SEC
6
EXH
120
Pressure then raises to assure that the set tidal volume is delivered
Automode
The ventilator switch between mandatory and spontaneous breathing modes Combines volume support (VS) and pressure-regulated pressurevolume control (PRVC) If patient is paralyzed; the ventilator will provide PRVC. All breaths are mandatory that are ventilator triggered, pressure controlled and time cycled; the pressure is adjusted to maintain the set tidal volume. If the patient breathes spontaneously for two consecutive breaths, the ventilator switches to VS. All breaths are patient triggered, pressure limited, and flow cycled. If the patient becomes apneic for 12 seconds; the ventilator switches back to PRVC
Advantages
Full to partial ventilatory support Allows spontaneous ventilation with safety net Patients E remains stable Prevents hypoventilation
Drager Evita 4
BiLevels
BiLevel Ventilation
Spontaneous Breaths 60 Spontaneous Breaths Synchronized Transitions
Paw
cmH20
1
-20
BiLevel Ventilation
PEEPHigh + PS Pressure Support
60
PEEPH
Paw
cmH20
PEEPL
1
-20
APRV always implies an inverse I:E ratio All spontaneous breathing is done at upper pressure level
Spontaneous Breaths 60
Paw
cmH20
Releases
1
-20
Modes of Ventilation
Questions?