AEI305.82 TO 83 2
Load cell specifications

The general specification defines the minimum


requirement for

• Construction
• Performance
• Features of the Load cells

AEI305.82 TO 83 3
Functional Performance Requirements
The accuracy of the measuring system shall be
better than 0.2%

This includes
• error of hysteresis
• Linearity
• Repeatability
• Ambient temperature. Effects

AEI305.82 TO 83 4
Functional Performance Requirements…

• Should have built in load cells.

• Should be immune to Radio Frequency Interference


/ Electro Magnetic Interference.

• Ambient temperature limit for operation -200ºC to


600ºC

AEI305.82 TO 83 5
Functional Performance Requirements…

• The out put of the load cell should be more than 2mv /volt.

• Load cell impendence shall be more than 600 Ω

AEI305.82 TO 83 6
Tagging

A permanently attached Stainless tag stamped


with the following information:

• Manufactures Name / Trade Mark

• Model No.

• Serial No.

AEI305.82 TO 83 7
Tagging…

• Tag No

• Supply Voltage

• Current Output 4-20ma

• Service

• Any other which vendor may like to specify

AEI305.82 TO 83 8
Packing

• All parts & openings shall be sealed.

• The connections shall be properly covered with


wooden, plastic or metal friction plugs

• The equipment is adequately protected & packed to


meet the shipping & delivery requirements.

AEI305.82 TO 83 9
Special Test Requirements

• Ambient Temperature

• Maximum : 43 o C

• Minimum : 9oC

• Design : 50 o C

AEI305.82 TO 83 10
Special Test Requirements…

• Relative Humidity
Max / Mean: 85% at 35 o C / 75% at 30 o C

• Atmospheric Environment :Highly Corrosive

• Rainfall :1300 – 2400 mm per annum

AEI305.82 TO 83 11
Hydraulic Load cell specifications

• Accuracy : ±1% F.S.

• Thermal drift : 0.1% F.S. / °C

• Operating temperature : −20 to +60°C

• Overload : 1.2 × F.S.

AEI305.82 TO 83 12
Hydraulic Load cell specifications…

• Measurement ranges: 45 kg to 450000 kg

• Inaccuracy : ± 0.1% to ± 1% of full scale

• Cost :750 $ to 1500 $ (Dallers)

AEI305.82 TO 83 13
Hydraulic Load cell specifications…

Capacity 250 500 750 1000 1 2000


KN KN KN KN 500 KN
KN
Cell thickness 28 28 28 28 30 30
mm m mm mm mm mm
m
Distribution 30 40 40 45 45 45
plate thickness mm m mm mm mm mm
m

AEI305.82 TO 83 14
The HYDLO Hydraulic Load Cell

AEI305.82 TO 83 15
AEI305.82 TO 83 16
AEI305.82 TO 83 17
Summary

We have discussed about

• Specifications of Load cells

• Specifications of Hydraulic load cells

AEI305.82 TO 83 18
Quiz

1. The general specification of load cell defines the


minimum requirement for
a) Construction

b) Performance

c) Features of the Load cells

d) Above all

AEI305.82 TO 83 19
1. The accuracy of the measuring system (Load cell)
shall be better than
0.5%

0.7%

0.2%

0.1%

AEI305.82 TO 83 20
1. The out put of the load cell shall be more than
a) 2mv /volt

b) 1mv /volt

c) 0.5mv /volt

d) 0.2mv /volt

AEI305.82 TO 83 21
1. The Load cell impendence shall be more than
a) 200 Ω
b) 400 Ω
c) 600 Ω
d) 500 Ω

AEI305.82 TO 83 22
• The accuracy of the hydraulic load cell should be
a) ±0.1% F.S

b) ±0.5% F.S

c) ±2% F.S

d) ±1% F.S

AEI305.82 TO 83 23
1. The operating temperature of the hydraulic
load cell is
a) −20 to +60°C

b) −2 to +60°C

c) −50 to +600°C

d) None

AEI305.82 TO 83 24
1. The measurement range of the hydraulic load cell is
a) 4.5 kg to 45000 kg
b) 45 kg to 450000 kg
c) 0.45 kg to 4500 kg
d) 450 kg to 4500000 kg

AEI305.82 TO 83 25
Frequently Asked Questions

1. List the general specifications of the Load cells.

3. List the specifications of the Hydraulic Load cells.

AEI305.82 TO 83 26

You might also like