Conductivity Products

REP SERV TRAINING

Copyright Yokogawa Electric Corporation

The Products - Conductivity


Analyzers:

EXA SC202
-Two-Wire Transmitter - 24 VDC - 4-20 mA Output (HART) - Fieldbus or Profibus

EXA SC402
- Four-Wire Analyzer - 110 VAC - NEMA 4X - Dual 4-20 mA Outputs - 4 Contact Relays - PI Control - USP23 Built-In

EXA DC402
- Two Inputs for Cond. - Four-Wire Analyzer - 110 VAC - NEMA 4X - Dual 4-20 mA Outputs - 4 Contact Relays - PI Control - USP23 Built-In

EXAxt SC450
- Four-Wire Analyzer - Intuitive Touch-Screen - Trend Graph Screen - 2 4-20mA / 4 Contacts - HART Communication - PID Control Function -USP23 Built-In -Class I, Div II

- Class I, Div I (II)


- FM, CSA, or ATEX - USP23 Built-In
Copyright Yokogawa Electric Corporation

The Products - Conductivity


Low-Cost 100 Series for Conductivity:

EXA SC100
- 1/4 DIN Panel M ount - Large Four Digit Display - Easy Automatic Calibration - 4-20mA / 2 Contacts - UL Rated (Pending)

Copyright Yokogawa Electric Corporation

The Products - Conductivity


Sensors:

SC42 (Small Bore: 12mm)


316SS body Insertion with NPT fitting or Triclamp Low Cost and Small Footprint Low Ranges Only

SC4A (Small Bore: 19mm)


316SS or Titanium body Insertion w/NPT fitting, Triclamp, or Retractable/w PR4A Mid-Price Quality Sensor Low Ranges Only

SC42 (Large Bore:30mm)


Variety of Materials Insertion, Immersion; Flow-Thru (with FF/FS40) Priced Higher -the BEST sensor for LOW and HIGH ranges

Copyright Yokogawa Electric Corporation

Why is Conductivity measured?


Used as a process control parameter. Used as an Alarm point indicator. Virtually maintenance free.

Copyright Yokogawa Electric Corporation

What is Electrolytic Conductivity?


Definition:
Specifi c Resista nce

Is

a Measure of the Ability of a Solution to Carry an Electric Current......(Charged Particles in Motion)


10 K 1 K 10 0 1 0 1

100 Meg.

10 Meg.

1 Meg .

100 K

OHM/ CM S/ CM

. 0 1 Specific

. 1

1 0

10 0

1,0 00

10,0 00

100, 000 1,000,000 10% H2S O4 3% NaO H

Conduct ance
Good Distil led Wate r

Ultra Pure Water

10 PPM NaCl

Oleum (nomi nal)

Sea Water (3.5% NaCl)

Sieme ns or Mhos

Natural Waters (typical)


Copyright Yokogawa Electric Corporation

What is A Standard Reference Cell?

Two Plates in a volume of Fluid. Different Volume =Different

Cell

Copyright Yokogawa Electric Corporation

Constant Same Fluid in a different volume will give a different conductivity reading.

Standard Reference Cell


1 cm

1 cm

A
1 cm

Length cm Cell Constant = ---------- = ---- -2 Area cm


FLOW

Copyright Yokogawa Electric Corporation

How is Conductivity The term Conductivity is derived from Measured? Ohms Law which is defined as:
E= I R

In other words when a voltage E is applied across an electrical conductor (water), then an electrical current I will flow which is dependent on the resistance R Alternating the conductor ( water). of

Current Conductivity is defined as the


B

reciprocal of resistance and it can be expressed as: G= I / R

A
FLOW

Copyright Yokogawa Electric Corporation

Increasing the amount of ions present in the fluid, increases the ability of the fluid to transport the electrical current. Conductivity goes up.

nductivity Cells Come In Different Siz


Small Size-12 MM OD Lowest Cost Low Conductivity Ranges Large Size-25 MM OD Higher Cost Lowest Conductivity Ranges and Highest Range
YE W

YOKOGA

Mo

Mid Size-19 MM OD Low Cost Lowest Conductivity Ranges


Copyright Yokogawa Electric Corporation

EXA Conductivity Analyzers and Available in both contacting and inductive Sensors
(electrodes) conductivity methods Accurate (0.5%) measurement Of everything from ultrapure water to highconcentration solutions in temperatures from -20C to 130C Auto-ranging Eliminates multiple sensor requirements and allows a single sensor to handle 0.055 S to 2,000,000 S Sensor Diagnostics Ensures measurement integrity and sensor quality in heavy scaling applications
Copyright Yokogawa Electric Corporation

EXA Conductivity Analyzers and Special Temperature Compensation Modes Sensors


Features non-linear correction for pure water and concentrated solutions (examples: HCl, NaOH, NH3), linear programmable coefficients and a 25-point programmable matrix
PI control (mA or pulse)

Eliminating the need for a separate controller

Copyright Yokogawa Electric Corporation

Event Logbook Function Used to record events like cell constant changes, calibrations and diagnostic messages

Inductive - Principle of Operation


Sample (Liquid Loop)
DRI VE COI L

Input Voltage To Coil

The Induced Current is Dependent on the Conductivity of the Sample

PICKU P COIL

DRIV E
PICKUP COIL

COIL

Sample (Liquid Loop)

Acts as a single turn winding


Copyright Yokogawa Electric Corporation

PROCESS FLUID

ISC40 - Key Features


ONE BORE SIZE The only sensor on the market that is able to measure from 0.6 uS to 2,000,000 uS with the same bore. The use of matched (patterned permeability) coils within the sensor permit unprecedented temperature stability and thus rangability. Since the matched coils respond in the exact manner to temperature swings, no accuracy is lost due to these temperature changes. TEMPERATURE SENSOR The temperature sensor provides a rapid temperature measurement and permits a rapid compensation for the effects of temperature on the conductivity measurement. AUTO ZEROING
Copyright Yokogawa Electric Corporation

Continuos auto zeroing (every 20 seconds), plus matched coils eliminates the need for

ISC40 - Other Benefits


STRAIN RELIEF Cable protection from twists and pulls during installation and calibration and gives longer life. 316 SS SHANK the 3166 shank goes all the way through the sensor. This makes it the strongest inductive sensor currently available a the one best able to handle water hammer, back pressure problems, surries, etc. PEEK PEEK or Poly ether ether ketone possess extremely valuable properties as a material of construction such as: the ability to handle continuos working temperatures of 250oC (482oF), being insoluble to all common solvents and

Copyright Yokogawa Electric Corporation

Inductive - Sensor Holders


Immersion Holder Flow Through Holder Sub-Assembly Holder

MNPT 316 Stainless Steel PVDF Polypropylene

Tri-Clover

Copyright Yokogawa Electric Corporation

Inductive - Sensor Holders


ISC40FD Immersion Holder

ISC40FD Immersion Holder

Copyright Yokogawa Electric Corporation

nductive - Sensors Pipe Mounting


Cabl e Nut 1 1/4NMPT 3/4G
Customer supplied Pipe 1 1/4 FNPT

3/4 G to 1 1/4MNPT

S2510046G1 PVDF S2510046F1 316 SS

Copyright Yokogawa Electric Corporation

Selection: Contacting Vs Inductive


Contacting Cells

All High Purity Applications Ranges Below 100 microsiemens All Ranges Except High Purity Above 100 microsiemens All % Concentration Ranges All Custom Ranges

Inductive Cells

Example - Ounces per Gallon

All Applications With Undissolved Solids

Match the Sensor to the Applicati on

Copyright Yokogawa Electric Corporation

Conductivity: Contacting
Full scale range 0 to 50, us/cm: No

Toroidal

Vs

Toroidal
- Lacks sensitivity

Contacting
Yes - Can measure down to high purity water

Process dirty or corrosive:

Yes - Resistant to fouling; corrosion resistant materials available

No - Very susceptible to fouling. Metal electrodes can corrode

Concentrated solutions:

Yes - Can measure highest conductivity possible

No - There can be polarization effects. Concentrated solutions tend to be fouling and corrosive

Major advantages:

Lower maintenance due to Fouling and corrosion resistance;High range capability

Sensitivity for low conductivity measurements Usually smaller and easier to mount.

Copyright Yokogawa Electric Corporation

Contacting Conductivity
YEW

YOKOGAWA
SC41/SP34

CONDUCTIVITY CELL

Measures Volume Resistivity


Required for High Purity Water Certified Cell Constant Wide Range Capability Quick Disconnect Installation Long Life Cell

Copyright Yokogawa Electric Corporation

EXA - Key Features


Autoranging = Rangeability

PURE S/cm

0.1 S

1000 mS

CONCENTRATION mS/cm

C = 10.0 C = 1.0 C = .01 & 0.1

Capacitance

Polarization
0.1 S 1000 mS

Copyright Yokogawa Electric Corporation

Selection of Wrong Cells Constants Causes problems


A B
Polarization

FLOW
Copyright Yokogawa Electric Corporation

Wrong Range (Current Density) Causes Problems


A B
Capacitance
B

A
A

FLOW
Copyright Yokogawa Electric Corporation

Auto Read Point Adjustment Alarms Upon Detection


A B

Auto Read Point

ALAR M

Copyright Yokogawa Electric Corporation

FLOW

SENSOR DIAGNOSTICS
YEW

YOKOGAWA
SC41/SP34

CONDUCTIVITY CELL

CONNECTIONS

FOULING

Same Basic Functionalit y as EXA pH


OUT OF PROCESS
Copyright Yokogawa Electric Corporation

Two Cell Measurements


YEW YOKOGAWA CONDUCTIVITY CELL SC41/SP34 YEW YOKOGAWA CONDUCTIVITY CELL SC4 1/SP34

Two cell inputs. Two measured outputs

or two calculations.
One measured output

& one calculation.

DC402 Analyzer
Copyright Yokogawa Electric Corporation

Conductivity Calibration
& Configuration Section

Copyright Yokogawa Electric Corporation

Conductivity Setup
Contacting and Inductive

Air Cal

All EXA SC202, SC402 , ISC402, IC200 and the EXAxt SC150 permit an Air Cal to be performed. With SC202,SC402 and SC150 Air Cal is done via the Access Codes. ISC402 and IC200 via front panel. the small zero current in the circuitry and needs to be performed only once during initial startup. The benefit of performing an Air Cal, is two fold.

The function of Air Cal is to compensate for

Copyright Yokogawa Electric Corporation

Special Inductive Issues


Always do a Installation Factor Adjustment When using a ISC40 FF flow through Fitting. ( Default is 1.7/cm.). The IF adjustment corrects the shape of the magnetic field so no contact is made with the ISC40FF walls.

magnetic field
Copyright Yokogawa Electric Corporation

Inductive Sensors -Installation Factor



Always ensure that the proper installation factor is used. See Access Code 02 ( ISC402) or Access Code 05 (IC200) for the proper value. With a ISC40FF- S (Stainless Steel) it is 1.55. For a ISC40P or ISC40F( Poly or Kynar) it is 1.7. Default is 1.7.

Pipe and vessel installations may however not give the 1 minimum clearance to use the default setting.

ISC40FF- S
Copyright Yokogawa Electric Corporation

Conductivity Calibration
Contacting and Inductive How To Calibrate
DOs
1. Always CLEAN and RINSE the Cell 2.Use FRESH CALIBRATION Solutions. 3.If working with low conductivity Solutions, clean the beakers with DI Water before pouring in Solutions. 4.Pick Calibration Solutions in the middle range of the for the cell constant if possible. 5.Ensure the discharge holes of cell are covered. 6. Follow proper grab sample calibration procedures.

Discharge Holes

DONTs
1.Do Not Calibrate With Extremely of Calibration The Frequency Low Conductivity Solutions Depends On: ( below 10 S). 2. Do Not wait before checking grab sample Process Solution Maintenance Copyright Yokogawa Electric Corporation values with low conductivity values.

Desired Accuracy

Conductivity Calibration
Inductive Special Issues
DONTs
1. Never allow the ISC40 Sensor to touch the bottom or the sides of the beaker that the calibration solution or grab sample is held in !! The shape of the magnetic fields (Installation Factor) is warped and a incorrect calibration will result. 2. If possible do not use the ISC40 in low conductivity solutions ( below 10 S).
DRIV E COIL PICKU P COIL

Magnetic Fields

Copyright Yokogawa Electric Corporation

APPLICATIONS Calibration of Conductivity

Correct calibration requires standard solutions. Works well on higher conductivity ranges (2000 S) Interferences are too high at low conductivity values ( below 100S) Carbon Dioxide from the air dissolves in the water and increases conductivity One Point Calibration
microsiemens 1.5 1 0.5 0 0.01
Copyright Yokogawa Electric Corporation

pH 7.5 7 6.5 6 5.5 5

pH

Conductivity 0.1 1 % of air saturation 10 100

Conductivity - Calibration
GRAB SAMPLE METHOD

Changes in Process
HOL D

0.236
NaCL

conditions can be accounted for by proper grab sampling.

26.8C

Process

POWER HOL D CELL CON ST TEM P COEF F CAL RANG E

Bring a high quality lab unit or conductivity handheld to the process floor.

ENT

Run a sample stream.

CONDUCTIVI TY METER

Sample
Copyright Yokogawa Electric Corporation

This gives the same conductivity and process temperature as the process.

Sample Calibration of Conductivity


PROBLEM: Carbon Dioxide from the air dissolves in the water and increases conductivity and decreases pH value. The higher the agitation rate or longer the sample flows, the higher the ingress. Sample Calibration values will be high. Why? Carbon Dioxide from the air dissolves and forms Carbonic Acid.pH value changes and so does conductivity. SOLUTION: Slowly fill beaker by using small flex tubing, placed in bottom of beaker. This stops agitation and entrained ingress.
Carbon Dioxide Ingress

Sample Point

Copyright Yokogawa Electric Corporation

CONDUCTIVITY APPLICATIONS

Copyright Yokogawa Electric Corporation

APPLICATION Non Homogenous Applications S


Are Solutions that contain: Undissolved solids Air bubbles

Immiscible liquids (like solvents) Have lower conductivity because the impurities take up space that normally would have conductive liquid.
0 50 100 0 50 100

50

100

Just One Phase reads 100%

Uniform Distribution reads lower

Non-Uniform Distribution is noisy

Copyright Yokogawa Electric Corporation

APPLICATIO Conductivity Percent NS Concentration Curves

Good Sensors for application

Bad Sensors for application

Copyright Yokogawa Electric Corporation

Conductivity rises linearly with concentration in dilute solutions. Ions interfere with one another in concentrated solutions so conductivity may stop rising or actually fall.

APPLICATION S BOILER BLOWDOWN


pH T RA N SM IT TER

CONDENSATE RETURN LINE MAKE UP WATER LEVEL CONTROL


LOOP 1 20. 0

USER
ST EA M H EA DER
CO N DU CTI VI T Y T RA N SMI T TER

TIC-1076 FAIL ALM % DO C

8. 19 PV: 13. 0

SV:
05.0

MV:
0.0

C
SHIFT

0
YOKOGAWA

STEA M DRU M PRETREATMENT, CLARIFIER, SOFTENER OR DEIONIZER BO IL ER FEED WA T ER

CO N DU CTI VI T Y SENSO R

Y OKOG Y E CONDUCTIVIT AWA W SC41/SP Y CELL 34

Good Sensors for application


DEAERATOR A U TO M A TI C V A LV E MA N U A L V A LV E

FEEDWATER CHEMICAL TREATMENT

CO N TIN U O U S O R SU RFA CE BLO WDO WN LIN E

Bad Sensors for application

Copyright Yokogawa Electric Corporation

BO TT O M BLO WDO WN L IN E

Typical Ultrapure Water System


CC = Cation Conductivity CR = Cation Ratio (2 Sensors) pH = pH Measurement RO = Reverse Osmosis Passage/Rejection (2 Sensors) SC = Specific Conductivity

Key

RO Inlet Stream

Cation CR

Anion pH SC pH SC Mixed Bed pH SC Deion Storage SC

RO
POWER/ BOILERS

Cation RO CR

Anion pH SC Exhaust Steam Turbine Condenser

Return Condensate

Sample Cooler SC Cation Column CC

Steam Polisher Deaerator pH SC Boiler

SC

Return Condensate

Copyright Yokogawa Electric Corporation

Ultrapure Water Cycle, cont.d...


Reverse Osmosis Filters Conductivity: Ratio or % Passage/Rejection (two sensors) pH occasionally, for membrane life Demineralizers: Cation, Anion, Mixed Bed Cation Cation Conductivity (two sensors in ratio) Charged Cation Beads remove (change to acid) positive Ions and release H+ into stream Cation Conductivity has special Temp Comp to take the H+ into account % Concentration (1 sensor) Beads are regenerated with acid (typically HCL or H2SO4)

Copyright Yokogawa Electric Corporation

Ultrapure Water Cycle, cont.d..


Anion Specific Conductivity (1 or 2 sensors) Charged Beads remove (change to water) negative ions % Concentration (1 sensor) Beads are regenerated with Caustic (NaOH) pH also, occasionally Completeness of after-regeneration Rinse Mixed Bed (Polisher) Specific Conductivity (1 sensor) Removal of both + & - Ions (fine cleaning, < 0.1 uS/cm) pH occasionally for complete monitoring

Copyright Yokogawa Electric Corporation

Ultrapure Water Cycle, cont.d..


Deion Storage Specific Conductivity (1 sensor) Ensure Water Quality after storage Second Polisher Specific Conductivity Final cleaning (Boiler applications)

Copyright Yokogawa Electric Corporation

Ultrapure Water Cycle Analyzer Summary


Boiler Steam & Water Cycle, Measurement Points
Re Sample Point 1. Boiler Blowdown Measurement Specific cond 0-1000 microS pH to purity 8-11 Specific cond 0-1 microS pH hi-purity 6-8 Cation cond 0-2 microS pH hi-purity 6-8 Cation cond 0-10 microS pH hi-purity 8-11 Reason Control TDS Scale & Corrosion Demineralizer Regeneration Demineralizer Regeneration Superheater contamination Superheater contamination HP heater contamination Anti-corrosion check Instrument ISC402 & ISC40 ISC402 & ISC40 SC402 /SC200 & SC42-SP34 PH402/PH202 & FF20/FS20 w / -SE option DC402/SC402 /SC200 & SC42-SP34 PH402/PH202 & FF20/FS20 w / -SE option DC402/SC402 /SC200 & SC42-SP34 PH402/PH202 & FF20/FS20 w / -SE option

2. Demineralizer Outlet

3. Superheater Outlet

4. Economiser Inlet

Copyright Yokogawa Electric Corporation