Professional Documents
Culture Documents
The loss of potential energy(P.E) when the electron moves from cathode to accelerating
anode;
P.E = e Ea
y = (½)(eεy/m)(x/Vox)²
This is the equation of a parabola. The slope at any point (x,y) is
dy/dx = eεy/mV²ox
Put x = ld in above equation we get the value of .
The Apparent origin is thus at the centre of deflection plates. The deflection D on the screen is given by,
D = L tan θ = L.(e.Ed.ld)/mdV²ox
Out of these varieties, the materials P31 is used commonly for general purpose oscilloscopes due to following
characteristics:
• It gives color to which human eye response is maximum.
• It gives short persistence required to avoid multiple image display.
• It has high burn resistance to avoid the accidental damage.
• Its illumination level is high.
• It provides high writing speed.
Note: the light output of a fluorescent screen is proportional to the number of bombarding electrons, i.e., to
the beam current.
Delay line
Lumped parameter delay line
Distributed parameter delay line
Cont.
Dual trace or Multiple trace
• Two different input signals are used i.e Channels A & B with attenuators
and preamplifiers. A delay line is used between electronic switch &
vertical amplifier that alternately connects the input to main vertical
amplifiers
Alternate mode
• A slight reduction in the amplitude of the waveform and a slight change in the
wave shape occurs with this probe. To avoid this possibility, a high impedance
compensated probe, called a low capacitance probe or a 10 : 1 probe, is used.
• Active probes are designed to provide an efficient method of coupling high
frequency, fast rise time signals to the CRO input. Usually active probes have very
high input impedance, with less attenuation than passive probes. Active devices
may be diodes, FETs, BJTs, etc.
• Active probes are more expensive and bulky than passive probes, but they are
useful for small signal measurements, because their attenuation is less.
Active probes
• Active Probes Using FETs
The FET is used as the active element to amplify the
input signal. Although the voltage gain of the FET follower
circuit shown is unity, the follower circuit provides a power
gain so that the input impedance can be increased. To be
effective the FET must be mounted directly in the voltage
probe tip, so that the capacitance of the interconnecting
cable can be eliminated. This requires that the power for the
FET be supplied from the oscilloscope to the FET in the probe
tip. The FET voltage follower drives a coaxial cable, but
instead of the cable connecting directly to the high input
impedance of the oscilloscope, it is terminated in its
characteristic impedance.
• There is no signal attenuation between the FET Amplifier
and the probe tip. The range of the signals that can be
handled by the FET probe is limited to the dynamic range
of the FET amplifier and is typically less than a few volts.
To handle a larger dynamic range,
external attenuators are added at the probe tip. Active
probes have limited use because the FET probe effectively
becomes an FET attenuator. Therefore, oscilloscopes are
typically used with a 10 to 1 attenuator probe.
Current Probe
• They provide a method of inductively coupling the signal to the CRO input, so that
a direct electrical connection to the test circuit is not required.
• The current probe consists of a sensor, a coax cable & a termination circuit a
shown in Fig.
Cont.
SPLIT-CORE
• Passive current probe, the most popular type, can be opened & clipped around a
conductor (see Fig. 17) whose current is to be measured.
• The current sensing device of this probe is a “Current Transformer” of split core
design, consisting a stationary U-piece & a Movable flat piece.
• A multi –turn coil of approximately 25 turns is wound on one leg of the ferrite core
to form the secondary turn primary.
• The input signal to the probe is the current in the conductor under test; the o/p
signal is the voltage developed across the transformer secondary.
• This current probe senses only the changes in current & hence can be used
only to measure A C Signal.
• When correctly terminated, the sensitivity of this probe is of the order of
10ma/mv.
• The transformer o/p voltage is coupled from the probe head to the termination via
a coaxial cable.
• The termination circuitry can be passive or active, depending on the kind of probe
generally the termination of the coax is in its characteristics impedance.
• Additional circuitry to improve the response characteristics of the probe is also
contained in the termination box.
Digital Storage Oscilloscope
Working
• The input signal is applied to the amplifier and attenuator section.
• The oscilloscope uses same type of amplifier and attenuator circuitry as
used in the conventional oscilloscopes.
• The attenuated signal is then applied to the vertical amplifier.
• To digitize the analog signal, analog to digital (A/D) converter is used.
• The output of the vertical amplifier is applied to the A/D converter
section.
• The successive approximation type of A/D converter is most oftenly used
in the digital storage oscilloscopes.
• The sampling rate and memory size are selected depending upon the
duration & the waveform to be recorded.
• Once the input signal is sampled, the A/D converter digitizes it.
• The signal is then captured in the memory.
• Once it is stored in the memory, many manipulations are possible as
memory can be readout without being erased.
The digital storage oscilloscope has
three modes
1. Roll mode
2. Store mode
3. Hold or save mode
Advantages
i) It is easier to operate and has more capability.
ii) The storage time is infinite.
iii) The display flexibility is available. The number of traces that can be stored
and recalled depends on the size of the memory.
iv) The cursor measurement is possible.
v) The characters can be displayed on screen along with the waveform
which can indicate waveform information such as minimum, maximum,
frequency, amplitude etc.
vi) The X-Y plots, B-H curve, P-V diagrams can be displayed.
vii) The pretrigger viewing feature allows to display the waveform before
trigger pulse.
viii) Keeping the records is possible by transmitting the data to computer
system where the further processing is possible.
ix) Signal processing is possible which includes translating the raw data into
finished information e.g. computing parameters of a captured signal like r.m.s.
value, energy stored etc.
Applications