Overview
Purpose of measurement system analysis (MSA)
Organizational uses of MSA
Measurement system errors
Accuracy and Precision of measurements
MSA process flow
Real world application
Exercise
Measurement System
Analysis (MSA)
aka Gauge R&R
A tool used to evaluate the statistical properties of
process measurement systems.
Purpose of MSA is to statistically verify that
current measurement systems provide:
Representative values of the characteristic being
measured
Unbiased results
Minimal variability
Organizational Uses
Mandatory requirement for QS 9000
certification.
Identify potential source of process
variation.
Minimize defects.
Increase product quality.
Measurement System Errors
Accuracy: difference between the observed
measurement and the actual measurement.

Precision: variation that occurs when


measuring the same part with the same
instrument.
Measurement System Error

Precise but Accurate but Not accurate Accurate


not accurate not precise or precise and
precise
Accuracy of Measurement
Broken down into three components:
1. Stability: the consistency of
measurements over time.
2. Accuracy: a measure of the amount of
bias in the system.
3. Linearity: a measure of the bias values
through the expected range of
measurements.
Precision of Measurement
Precision, aka Measurement Variation, can
be broken down into two components:
1. Repeatability (Equipment variation): variation
in measurements under exact conditions.
2. Reproducibility (Appraiser variation): variation
in the average of measurements when different
operators measure the same part.
MSA Process Flow
1. Preparation for study
2. Evaluate stability
3. Evaluate resolution
4. Determine accuracy
5. Calibration
6. Evaluate linearity
7. Determine repeatability and reproducibility
Preparation for Study
Objective: establish process parameter for
the study.

Process:
1. Determine which measurement system will
be studied.
2. Establish test procedure.
Preparation for Study
-Cont-
3. Establish the number of sample parts, the
number of repeated readings, and the
number of operators that will be used.
4. Choose operators and sample parts.
Evaluate Stability
Objective: evaluate measurement system
to determine if the system is in statistical
control.

Procedure:
1. Choose sample standards.
2. Measure sample standards three to five
times.
Evaluate Stability
-Cont-
3. Plot data on a x-bar and R chart.

Analysis:
1. Determine if process is in control.
2. If process is unstable determine and correct
the cause.
Evaluate Resolution
Objective: determine if the measurement
system can identify and differentiate
between small changes in the given
characteristic.

Process:
1. Choose a sample standard.
Evaluate Resolution
-Cont-
2. Measure the sample standard three to five
times.

3. Repeat the process 10 to 25 times.

4. Plot data on a R chart.


Evaluate Resolution
-Cont-
Analysis
1. The resolution is inadequate if:
- There are only one, two, or three possible
values for the range, or
- There are only four possible values for the
range when n >= 3.
Determine Accuracy
Objective: determine the variation between the
observed measurement and the actual
measurement of a part.

Process:
1. Choose sample standards.
2. Measure sample standards 15 to 25 times using
the same measuring device, the same operator, and
the same setup.
Determine Accuracy
-Cont-
3. Calculate x-bar
4. Calculate bias
- Bias = Average Reference Value
5. Calculate the upper and lower 95%
confidence limit (CL).
Determine Accuracy
-Cont-
Analysis

1. If reference value is within the 95% CL


then the bias is insignificant.

2. If reference value is outside the 95% CL


then the bias is significant and measurement
system must be recalibrated.
Calibration
Objective: to ensure the instrument is
accurate, and measurement bias is
minimized.

Process: calibrate instrument IAW


manufacturers instructions.
Evaluate Linearity
Objective: determine the difference between
the obtained value and a reference value
using the same instrument over the entire
measurement space.

Process:
1. Choose three to five sample standards that
cover the measurement space.
Evaluate Linearity
-Cont-
2. Measure sample standards 15 to 25 times.
3. Calculate the average of the readings.
4. Calculate bias.
5. Plot reference values on x-y graph.
6. Calculate slope of the linear regression line.
7. Calculate linearity and percent linearity.
8. Calculate R2.
Evaluate Linearity
-Cont-
Analysis

1. The closer the slope is to zero, the better the


instrument.

2. R2 gives indication of how well the best-


fit line accounts for variability in the x-y
graph.
Determine Repeatability and
Reproducibility
Objective: determine variation in a set of
measurement using a single instrument that
can be credited to the instrument itself, and
to the entire measurement system.

Process
1. Generate random order for operators and
parts to complete the run.
Determine Repeatability and
Reproducibility -Cont-
2. Repeat process for subsequent runs.
3. Have operators take measurements.

Analysis:
1. Plot data
2. Run ANOVA (analysis of variance) on data.
Determine Repeatability and
Reproducibility -Cont-
3. Calculate total variance.
4. Calculate % Contribution and determine if
acceptable.
5. Calculate % Contribution (R&R)
6. Calculate Process to Tolerance ratio (P/T)
for repeatability.
7. Determine if P/T is acceptable.
Real World Application
Case study performed by Tirthankar
Dasgupta and S.V.S.N Murthy, Indian
Statistical Institute, 2001.
- Gauge R&R study of automobile radiator
manufacturer.
- After studying four characteristics of
radiator components the following results
were obtained:
Table 1. Results of Preliminary
Study

Least
Characteristic Gauge Count TV %EV %AV %R&R
Filler neck hole dia. Vernier calipers 0.01 0.1107 35.77 72.88 81.12
Tube crown thickness Dial vernier calipers 0.02 0.1866 49.19 38.06 62.1
Tube width Dial vernier calipers 0.02 0.0365 10.71 43.83 46.87
Inlet hole dia. Vernier calipers 0.01 0.1057 86.56 20.09 88.83

Source: Dasgupta & Murthy, Total Quality Management, Vol. 12, No. 6, 2001
Real World Application
-Cont-
Any system having greater than 30% gauge
R&R is considered inadequate. As seen in
Table 1, all four characteristics %R&R is
inadequate.
Investigation of the measurement system
led to a subsequent reduction of %R&R in
three of the four characteristics to between
12% and 23%.
Real World Application
-Cont-
Further investigation of the fourth
characteristic, inlet hole diameter, led the
examiners to a manufacturing problem. The
team discovered high ovality in the inlet
hole, which was caused by the cutting tool.
The tool was modified to reduce ovality.
Real World Application
-Cont-
Benefits of the study
1. Reduced measurement variation.
2. Increased operator confidence regarding
their aptitude for conducting gauge R&R
studies.
3. Paved the way for further studies within the
firm.
An Exercise Calculating EV,
AV, R&R, and TV
Given: EV = 5.15(s0) , AV = 5.25(s1)
R&R = (EV2 + AV2)
TV = (EV2 + AV2 + PV2)
Where: s0 = gauge standard deviation = 0.05
s1 = true appraiser standard
deviation = 0.1
PV = part-to-part variation = 0.02
An Exercise Calculating EV,
AV, R&R, and TV

Calculate R&R and TV

Is the calculated R&R acceptable?


Summary
MSA studies are required for QS 9000
certification. However, MSA can prove
beneficial to any firm that uses
measurement systems whether they are
seeking QS 9000 certification, or not.

MSA studies are a tool that aid in ensuring


quality at all levels of a process.
Summary
-Cont-

According to David C. Crosby, If you


dont know the capability of your
measurement system, you dont know if
your measurements, or your products, are
good or bad.
References
Crosby, David C. A Managers Guide to Gauge R&R, Rubber World
218 (1998): 16-17.
Goyal, Niraj. Selecting Appropriate Metrics.
www.isixsigma.com/library/content.c020930a.asp. 4 Oct 2003.
Hemanth, S. Anomoly In Normality.
www.isixsigma.com/library/content/c02820a.asp. 4 Oct 2003.
Niles, Kim. Characterizing the Measurement Process.
www.isixsigma.com/library/content/c020527a.asp. 4 Oct 2003.
Measurement Systems Analysis Overview.
http://mathstat.carleton.ca/~help/minitab/QCMEASYS.pdf.
4 OCT 2003.
Measurement System Analysis Resolution and Granularity.
www.isixsigma.com/library/content/c000903a.asp. 4 Oct 2003.
References
Measurement System Capability Manual.
www.6sigma.us/MSASymbolsandNotation.html. 3 Oct 2003.
Dasgupta, Tirthankar & Murthy, S.V.S.N. Looking beyond audit-
oriented evaluation of gauge repeatability and reproducibility: A case
study. Total Quality Management. 12 (2001): 649-655.