Ventilation in Neonates Mammady Al maghayreh Date: October 17-19, 2011 Venue: Pr. Rahma Hospital/Auditourium Introduction
It is an invasive life-support procedure with many effects on the cardiopulmonary system The goal is to optimize both gas exchange and clinical status at minimum FiO2 and ventilator pressure. The ventilator strategy employed to accomplish this goal depends in part on the infant disease process. Conventional positive pressure ventilation remains the mainstay of assisted ventilation in neonates despite the development of new ventilatory techniques.
2 Physiologic functions of the lung Ventilation The movement of air between the atmosphere and the respiratory portion of the lungs Perfusion The flow of blood through the lungs Diffusion The transfer of gases between the air-filled spaces in the lungs and the blood.
Ventilation Depends on the conducting airways: Nasopharynx and oropharynx Larynx Tracheobronchial tree Open Alveoli Function: Moves air in and out of the lung, warms and humidifes. Airways do not participate in gas exchange. Lung Compliance
The ease with which lungs can be expanded Specifically, the measure of the change in lung volume that occurs with a given change in transpulmonary pressure Determined by two main factors Distensibility of the lung tissue and surrounding thoracic cage Surface tension of the alveoli Pulmonary Mechanics during Assisted Ventilation
Compliance (ml/cmH 2 O) =Distensible nature of lungs and chest wall. = Change in volume (ml) X change in pressure (cmH 2 O) Neonates have greater chest wall Compliance Premature infants with RDS have stiffer lungs (poorly compliant lungs). Neonatal lung Normal 0.003-0.006 L/cm H 2 O with RDS 0.0005-0.001 L/cm H 2 O Resistance 10 Resistance (cmH 2 O/L/sec)= Property of airways and lungs to resist gas. = Change in pressure (cmH 2 O) X Change in flow (L/sec) Time constant of the respiratory system = Resistance x Compliance Resistance in infants with normal lungs ranges from 25 to 50 cm H2O/L/sec. It is increased in intubated babies and ranges from 50 to 100 cm H2O/L/sec. Pulmonary mechanics Time constant The time taken for the airway pressure (and volume) changes to equilibrate throughout the lung is proportional to the compliance and resistance of the respiratory system
Time constant = Compliance x Resistance
12 Pulmonary mechanics
Almost full equilibration: 3-5 time constants
100 80 60 40 20 0 1 2 3 4 5 Time constants
C h a n g e
i n
p r e s s u r e
( % )
63 86 95 98 99 14 Time Constant Inspiratory time must be 3-5 X time constant 1) One time conststant = time for alveoli to discharge 63% of its volume through the airway. 2) Two time constant = 84% of the volume leaves 3) Three time constant = 95% of volume leaves. Lung Mechanics Differ in Different Disease States Disease Compliance ml/cm H 2 O Resistance (cm/H2O/ ml/s) Time Constant(s) FRC (ml/kg) V/Q Matching Work Normal Term 4-6 20-40 0.25 30 --- --- RDS Decreased Decreased Decreased Decreased Decreased Increased Meconium Aspiration Decreased Increased Increased Increased Decreased Increased BPD Increased/ Decreased Increased Increased Increased Decreased Increased Air leak Decreased Increased Increased Increased Decreased Increased VLBW apnea Decreased Decreased Decreased Decreased Decreased Increased 15 Basic Ventilator Parameters FiO 2
Fractional concentration of inspired oxygen delivered expressed as a % (21-100) Breath Rate (f) The number of times over a one minute period inspiration is initiated (bpm)
Tidal volume (V T ) The amount of gas that is delivered during inspiration expressed in mls or Liters. Inspired or exhaled. Flow The velocity of gas flow or volume of gas per minute
Phase Variables Trigger (start)- begins inspiratory flow Cycling (end)- ends inspiratory flow Limiting (continue)- places a maximum value on a control variable pressure volume flow time Breath Type Only Two (for now)! Mandatory Ventilator does the work Ventilator controls start and stop Spontaneous Patient takes on work Patient controls start and stop Trigger Variable- Start of a Breath Time - control ventilation Pressure - patient assisted Flow - patient assisted Volume - patient assisted Manual - operator control The Control Variable- Inspiratory Breath Delivery Flow (volume) controlled pressure may vary Pressure controlled flow and volume may vary Time controlled (HFOV) pressure, flow, volume may vary Inspiratory - delivery limits Maximum value that can be reached but will not end the breath- Volume Flow Pressure
Expiratory - baseline Positive End Expiratory Pressure Expiratory Retard Negative End Expiratory Pressure Expiratory Hold Time Limited Exhalation PEEP Definition Positive end expiratory pressure Application of a constant, positive pressure such that at end exhalation, airway pressure does not return to a 0 baseline
Used with other mechanical ventilation modes such as A/C, SIMV, or PCV
Referred to as CPAP when applied to spontaneous breaths PEEP Increases functional residual capacity (FRC) and improves oxygenation Recruits collapsed alveoli Splints and distends patent alveoli Redistributes lung fluid from alveoli to perivascular space 5 cm H 2 O PEEP Gas Exchange during Assisted Ventilation Carbon Dioxide (CO 2 ) Diffuses rapidly from the blood into the alveoli. Its elimination depends largely on alveolar ventilation. Minute alveolar ventilation= (Tidal volume Dead space) x Frequency. Tidal volume is determined by the pressure gradient between inspiration and expiration. 25 CPAP Definition Continuous positive airway pressure Application of constant positive pressure throughout the spontaneous ventilatory cycle
No mechanical inspiratory assistance is provided Requires active spontaneous respiratory drive
Same physiologic effects as PEEP Gas Exchange during Assisted Ventilation (cont.) Carbon Dioxide (CO 2 ) - (cont.) Inspiratory duration may partially determine the tidal volume Tidal volume can be decreased by shortening the inspiratory time. Changes in ventilator frequency have a strong effect on CO2 elimination 27 Oxygen Oxygen exchange depends on matching perfusion with ventilation. Oxygenation is determined by the mean airway pressure applied. Paw = (PIP PEEP) [Ti/ (Ti +Te)] + PEEP
28 Gas Exchange during Assisted Ventilation (cont.) Mean airway pressure will be augmented by increasing any of the following: Inspiratory flow PIP I:E ratio PEEP Frequency (or rate) by shortening Te
29 Gas Exchange during Assisted Ventilation (cont.) The effect of mean airway pressure on oxygenation is related to: Optimizing the lung volume Preventing atelectasis and consequently improving ventilation perfusion relationships.
30 Gas Exchange during Assisted Ventilation (cont.) Conventional Neonatal Ventilator
31 Volume-Cycled Ventilators
Less used to ventilate neonates Deliver a fixed volume irrespective of pressure Flow and I:E ratio determine the tidal volume Does not work well for RDS patient 32 Pressure-Limited, Time-Cycled Peak inspiratory pressure, and inspiratory timing are selected Continuous flow of fresh heated humidified gas It allows the infant to make spontaneous respiratory efforts Fighting neonates may face air leak syndrome
33 Patient Triggered Ventilation Neonate is able to initiate ventilatory breath by: 1) Abdominal motion 2) Chest wall impedance 3) Airway flow 4) Great degree of synchronacy between patient and ventilator 34 PatientTriggered Ventilators (PTV)
The patient is able to initiate ventilator breaths by 1) Abdominal motion 2) Chest wall impedance 3) Airway flow 4) Great degree of synchronacy between patient and ventilator
Triggering the ventilator setting detector The system support the patient if it didn't breath improved tidal volume and blood gases It can be Synchronized (SIMV) or Assisted/ control (A/C) modes Weaning is by reducing the PIP Cerebral blood flow is controlled Reduce the duration of ventilation and ease weaning 35 Troubleshooting Changes in Tidal Volume during Pressure Ventilation of the Neonate
Tidal Volume Change Possible Cues Solutions Increase Increased compliance, decreased resistance, decreased PEEP, increased inspiratory time, decreased leak. Reduce peak inspiratory pressure. Decrease Decreased compliance, increased resistance, decreased peak inspiratory pressure, increased PEEP, decreased inspiratory time, increased leak. Suction airway Administer surfactant. Increase inspiratory pressure, performing a transillumination to check for pneumothorax, obtaining chest radiography, and check tube positioning. 36 Indications for Mechanical Ventilation Absolute indications Severe hypoxemia with a PaO2 less than 50mm Hg despite FiO2 of 0.8 Respiratory acidosis with pH of less than 7.20 to 7:25 or PaCO2 above 60mm Hg. Severe prolonged apnea
37
USAID-Funded Health Systems Strengthening II Project 38 Relative indications Frequent intermittent apnea unresponsive to drug therapy Early treatment when use of mechanical ventilation is anticipated because of deteriorating gas exchange Relieving work of breathing in an infant with signs of respiratory difficulty Initiation of exogenous surfactant therapy in infants with RDS
39 Indications for Mechanical Ventilation (cont.) Volume vs Pressure Control Ventilation Volume Ventilation
Volume delivery constant Inspiratory pressure varies Inspiratory flow constant Inspiratory time determined by set flow and V T Pressure Ventilation
Volume delivery varies Inspiratory pressure constant Inspiratory flow varies Inspiratory time set by clinician The Effects of Ventilator Setting Changes on Blood Gases
Effects on Blood Gas Tensions Ventilator setting changes PaCO 2 PaO 2
Increase PIP Decrease Increase Increase PEEP Increase Increase Increase Frequency Decrease Increase Increase I:E Ratio ------ Increase Increase FiO 2 ------ Increase Increase Flow Decrease Increase 41 Starting Ventilator Settings
Guidelines for Endotracheal Tube Size Infant Weight (gm) Endotracheal Tube Internal Diameter < 1,000 2.5mm 1,000-2,000 3.0mm 2,000-3,000 3.5mm > 3,000 3.5 - 4.00mm 42 Initial Settings for mechanical ventilation Setting Instructions for Use Peak inspiratory pressure (PIP) As needed to provide tidal volume of 5-7ml/kg. Positive end-expiratory pressure (PEEP) 3-5cm H 2 O Rate 40- 60/minute. Inspiratory time 0.3- 0.4 seconds Fractional inspired oxygen concentration (FiO 2 ) maintain SpO
The subsequent settings for mechanical ventilation Subsequent Settings PEEP PIP Low PaO 2 , low PaCO 2 Increase Low PaO 2 , high PaCO 2 Increase High PaO 2 , high PaCO 2 Decrease High PaO 2 , low PaCO 2 Decrease 44 Starting Ventilator Settings (cont.)
Monitoring the infant during mechanical ventilation First blood gas after 15-30 mins Blood gas after 15-30 mins of every change Regularly blood gas every 6 hrs Continuous vital sings monitoring 45 Starting Ventilator Settings (cont.) Deterioration during Mechanical Ventilation Sudden clinical deterioration Mechanical or electrical ventilator failure Disconnected tube or leaking connection Endotracheal tube displacement or blockage Pneumothorax 46 Deterioration during Mechanical Ventilation (cont.) Gradual deterioration Inappropriate ventilator setting Intraventricular hemorrhage Baby fighting against the ventilator PDA Anemia Infection 47 Paralysis and Sedation
The use of neuromuscular blockade is not routinely indicated Sedation is restricted to cases when agitation interferes with ventilatory support and when infants fight the ventilator. It is necessary to increase ventilator pressure after initiation of neuromuscular blockade.
48 Weaning When the patient is stable, FiO2 and PIP are weaned first. Decrease PIP as tolerated and as chest rise diminishes. When PIP is around 20, attention is directed to FiO2 and then to the respiratory rate alternating with each other As frequency is decreased, Te should be prolonged For larger infants, endotracheal CPAP when PIP 15- 18 cm H2O and FiO2 40%
49 Weaning (cont.) The infant can be weaned to oxygen hood when PEEP is 4cm H2O For less than 1.750gm, when PIP is less than 15cm and FiO2 30% decrease respiratory rate to 15-20/ min then to nasal CPAP In most infants, when ventilator frequency of approximately 15 breaths per minute is tolerated, endotracheal CPAP may be tried for a short period of time before extubation.
50 Weaning (cont.) Atelectasis after extubation is common in preterm infants recovering from RDS. Use of nasal CPAP may prevent atelectasis. Steroids are not routine before extubation, but if there was prolonged intubation or previous failed attempts of extubation, a short course of steroids may facilitate extubation. If strider developed epinephrine aerosols and steroids may be helpful 51 Complications of Mechanical Ventilation
Suction 53 Goals of Pulmonary Hygiene Maintain a patent airway by clearing secretions Promote optimal pulmonary oxygenation and ventilation Prevent pulmonary infection from accumulated secretions Facilitate removal of pulmonary debris 54 Chest Physiotherapy Indication Intubated neonates: Chest physiotherapy should only be applied if it is clearly indicated. No CPT after surfactant administration CPT in early RDS increases incidence of intraventricular hemorrhage Post-extubation Chronic lung disease of prematurity
55 Physiotherapy and surgery After abdominal or cardiac surgery Postoperative physiotherapy should never be routine but should be used judiciously. 56 Chest Physiotherapy Indication (cont.) Chest Physiotherapy Technique Positioning
Vibration 57 When It Is Indicated During intubation and ventilation Post ex-tubation Premature with chronic lung disease After abdominal and chest surgery
58 Suction Methods of suctioning: Open Closed Catheter suction ca be: Deep Shallow Suctioning should be performed under strict sterile preparation 59 Suction Procedure Shallow suction is recommended to prevent trauma Deep suction may cause apnea and vagal stimulation Duration of suction should be 15 seconds 60 Follow-up Care
Hyper-oxygenation for at least 1 min especially for hypoxemic infants Hyperventilation should not be routinely used. The patient should be monitored for adverse reactions 61 Indications
Secretions in peripheral airways should not be directly removed by endotracheal suctioning Suctioning should be performed only when clinically indicated To remove accumulated pulmonary secretions To maintain the patency and integrity of the artificial airway 62 Signs of Increased Pulmonary Secretions Increased peak inspiratory pressure during volume- controlled mechanical ventilation Decreased tidal volume during pressure-controlled ventilation Deterioration of oxygen saturation and/or arterial blood gas values Visible secretions in the airway Acute respiratory distress Suspected aspiration of gastric or upper-airway secretions The need to obtain a sputum specimen
63 Complications
Decrease in dynamic lung compliance and functional residual capacity Atelectasis Hypoxia/hypoxemia Tissue trauma Bronchoconstriction/bronchospasm Increased microbial colonization Changes in cerebral blood flow and increased intracranial pressure Hypertension or hypotension Cardiac dysrhythmias
64 Assessment of Outcome
Improvement in appearance of ventilator graphics and breath sounds Decreased need for ventilation support Improvement in arterial blood gas values or saturation Removal of pulmonary secretions
65 The following should be monitored prior to, during, and after the procedure: Breath sounds Oxygen saturation Pulse rate Skin color Respiratory rate and pattern Sputum characteristics: color, volume, consistency, and odor Ventilator
66 Monitoring during suction (cont.)
Recommendations Endotracheal suctioning should be performed only when secretions are present and not routinely. (1C) Pre-oxygenation should be considered if the patient has a clinically important reduction in oxygen saturation with suctioning. (2B) Performing suctioning without disconnecting the patient from the ventilator (2B) Use of shallow suction (2B) 67 Routine use of normal saline instillation prior to endotracheal suction should not be performed. (2C) A suction catheter is used that occludes less than 70% of the lumen of the ETT in infants. (2C) The duration of the suctioning event be limited to less than 15 seconds. (2C)
68 Recommendations (cont.) High Frequency Ventilation (HFV)
Rescue following failure of conventional ventilation (PPHN, Meconium).2,3 Air leak syndromes (pneumothorax, pulmonary interstitial emphysema) 4 To reduce barotrauma when conventional ventilator settings are high F r e q u e n c y
High frequency ventilation rate (Hz, cycles per second) M A P
Mean airway pressure (cmH 2 O) A m p l i t u d e
delta P or power is the variation around the MAP O x y g e n a t i o n
i s
d e p e n d e n t
o n
M A P
a n d
F i O 2
MAP provides a constant distending pressure equivalent to CPAP. This inflates the lung to a constant and optimal lung volume maximising the area for gas exchange and preventing alveolar collapse in the expiratory phase. Ventilation is dependent on amplitude and to lesser degree frequency. Thus when using HFV CO 2 elimination and oxygenation are independent The clinician sets
The amplitude . Frequency of the pressure wave generated by the ventilator piston or diaphragm Mean airway pressure (MAP) Inspiratory time. Fractional inspired concentration (FiO 2 ) USAID-Funded Health Systems Strengthening II Project 71 Optimal lung volume strategy (aim to maximise recruitment of alveoli). Set MAP 2-3 cmH2O above the MAP on conventional ventilation MAP in 1-2 cmH2O steps until oxygenation improves Set frequency to 10 Hz
73 HFV Low volume strategy (aim to minimise lung trauma) Set MAP equal to the MAP on conventional ventilation Set frequency to 10 Hz Adjust amplitude to get an adequate chest wall vibration. Making adjustments once established on HFV
Poor Oxygenation Over Oxygenation Under Ventilation Over Ventilation Increase FiO 2
Decrease FiO 2
Increase Amplitude Decrease Amplitude Increase MAP (1-2cmH 2 O) Decrease MAP (1-2cmH 2 O) Decrease Frequency (1-2Hz) if Amplitude Maximal Increase Frequency (1-2Hz) if Amplitude Minimal