Professional Documents
Culture Documents
=
n
X X
s
i
( ) % 100
|
.
|
\
|
=
X
s
CV
Statistical Measurements
Measure of spread or dispersion of data
Range: Difference between the largest and smallest
observation
Standard deviation: is a measure of the spread of data
about the mean
For symmetric distribution 75% of the data lies between
(mean - 2s) and (mean + 2s)
Coefficient of variation: standardize the variation to
compare data measured in different scales.
( )( )
( ) ( )
=
2 2
Y Y X X
Y Y X X
r
i i
i i
Statistical Measurements
Percentile: gives the percentage of a distribution
that is less than or equal to the percentile number.
Standard error of the mean (SEM): Express the
variability to be expected among the mean in future
samples.
Correlation Coefficient r: is a measure of a linear
relationship between numerical variables x and y for
paired observations
Methods for inference about a value in a
population of subjects from a set of
observations.
Estimation and confidence of interval:
are used to estimate specific parameters such as the
mean and the variance.
Hypothesis testing and P-value:
reveals whether the sample gives enough evidence
for us to reject the null hypothesis. P-value
indicates how often the observed difference would
occur by chance alone.
Methods for measuring the accuracy of a
diagnostic procedure
Sensitivity of a test: Probability of its
yielding positive results in patients who
actually have the disease.
Specificity of a test: Probability of its
yielding negative results in patients who do
not have the disease
Prior Probability: the prevalence of the
condition prior to the test.
Characteristics of Instrument Performance
Two classes of characteristics are used to
evaluated and compare new instrument
Static Characteristics: describe the
performance for dc or very low frequency
input.
Dynamic Characteristics: describe the
performance for ac and high frequency input.
1.9 Generalized Static Characteristics
Parameters used to evaluate medical instrument:
Accuracy: The difference between the true value and
the measured value divided by the true value
Precision: The number of distinguishable alternatives
from which a given results is selected {2.434 or 2.43}
Resolution: The smallest increment quantity that can
be measured with certainty
Reproducibility: The ability to give the same output
for equal inputs applied over some period of time.
1.9 Generalized Static Characteristics
Parameters used to evaluate medical instrument:
Statistical Control: Systematic errors or bias are
tolerable or can be removed by calibration.
Statistical Sensitivity: the ratio of the incremental
output quantity to the incremental input quantity, Gd.
( )( )
( )
=
2
d
2
d
d d
x x n
y x y x n
m
( )( ) ( )( )
( )
=
2
d
2
d
d d
2
d
x x n
x y x x y
b
b mx y + =
d
Finding static sensitivity G
d
using line equation with the
minimal sum of the squared difference between data points
and the line
Figure 1.3 (b) Static sensitivity:
zero drift and sensitivity drift.
Dotted lines indicate that zero
drift and sensitivity drift can be
negative.
Zero Drift: all output values increase or decrease
by the same amount due to manufacturing
misalignment, variation in ambient temperature,
vibration,.
Sensitivity Drift: Output change in proportion to the
magnitude of the input. Change in the slope of the
calibration curve.
Figure 1.4 (a) Basic definition of linearity
for a system or element. The same linear
system or element is shown four times for
different inputs. (b) A graphical illustration
of independent nonlinearity equals A% of
the reading, or B% of full scale, whichever
is greater (that is, whichever permits the
larger error).
x
d
(Input)
B% of full scale
A% of reading
Overall tolerance band
Least-squares
straight line
(a)
(b)
Point at which
A% of reading = B% of full scale
y (Output)
x
1
(x
1
+ x
2
)
y
1
x
2
Kx
1
Ky
1 y
2
Linear
system
Linear
system
Linear
system
Linear
system
and and
(y
1
+ y
2
)
Linearity
Independent nonlinearity
- A% deviation of the reading
- B% deviation of the full scale
Input Ranges (I ):
Minimum resolvable input < I < normal linear operating range
A linear system described by the following equation
y=2x+3. Find the overall tolerance band for the system if
the input range is 0 to 10 and its independent nonlinearity is
0.5% deviation of the full scale and 1.5% deviation of the
reading.
Example
variable flow
iable effort var
d2
d1
= =
X
X
Z
x
2
d2
2
d1
d2 d1
X Z
Z
X
X X P
x
x
= = =
Input Impedance:
disturb the quantity being measured.
X
d1
: desired input (voltage, force, pressure)
X
d2
: implicit input (current, velocity, flow)
P = X
d1
.X
d2
:Power transferred across the tissue-sensor
interface
Generalized input impedance Z
x
Goal: Minimize P, when measuring effort variable X
d1
, by
maximizing Z
x
which in return will minimize the flow
variable X
d2
.
Loading effect is minimized when source impedance Z
s
is
much smaller then the Z
x
) ( ) (
0 1 0 1
t x b
dt
dx
b
dt
x d
b t y a
dt
dy
a
dt
y d
a
m
m
m
n
n
n
+ + + = + + +
( ) ( ) ) ( ) (
0 1 0 1
t x b D b D b t y a D a D a
m
m
n
n
+ + + = + + +
0 1
0 1
) (
) (
a D a D a
b D b D b
D x
D y
n
n
m
m
+ + +
+ + +
=
1.10 Generalized Dynamic Characteristics
Most medical instrument process signals that are functions
of time. The input x(t) is related to the output y(t) by
a
i
and b
i
depend on the physical and electrical parameters
of the system.
Transfer Functions
The output can be predicted for any input (transient,
periodic, or random)
0 1
0 1
) (
) (
a D a D a
b D b D b
D x
D y
n
n
m
m
+ + +
+ + +
=
0 1
0 1
) ( ) (
) ( ) (
) (
) (
) (
a j a j a
b j b j b
j X
j Y
j H
n
n
m
m
+ + +
+ + +
= = e
Frequency Transfer Function
Can be found by replacing D by je
Example:
If x(t) = A
x
sin (e t)
then y(t) = |H(e)| A
x
sin (e t + /_H(e))
Figure 1.5 (a) A linear
potentiometer, an example of a
zero-order system. (b) Linear
static characteristic for this
system. (c) Step response is
proportional to input. (d)
Sinusoidal frequency response
is constant with zero phase
shift.
Zero-Order Instrument
a
0
y(t) = b
0
x(t)
K
a
b
j X
j Y
D x
D y
= = =
0
0
) (
) (
) (
) (
=
K: static sensitivity
First-Order Instrument
) ( ) (
) (
0 0 1
t x b t y a
dt
t dy
a = +
( ) ) ( ) ( 1 D t Kx t y = +
D
K
D x
D y
+
=
1 ) (
) (
( )
( )
( )
( )
( ) 1 / arctan
1
1
2 2
K
j X
j Y
j
K
j X
j Y
=
+
=
+
=
|
0
0
0
1
a
b
K
a
a
= = t
Where t is the time constant
( ) ( )
t /
1
t
e K t y
=
t
1
(c)
(a)
C
+
+
y(t)
Output y(t)
Input x(t)
Slope = K = 1
(b)
Y (je)
X (je)
Log
scale
1.0
0.707
Log scale e
(d)
0
45
90
Log scale e
t
1
0.63
t
L
t
S
t
L
t
S
e
S
e
L
|
t
L
t
S
x(t)
x(t)
y(t)
R
First-Order Instrument
D
K
D x
D y
+
=
1 ) (
) (
( ) ( )
t /
1
t
e K t y
=
Example 1.1: High-pass filter
) ( ) (
) (
t x t y
dt
t dy
RC = +
1 ) ( 1 = = = t x K RC t
Second-Order Instrument
( ) ( )
( ) ( ) t x b t y a
dt
t dy
a
dt
t y d
a
0 0 1
2
2
2
= + +
( ) ( ) t Kx t y
D
n n
=
(
+ + 1
2
2
2
units input by defined units output y, sensitivit static
0
0
= =
a
b
K
rad/s frequency, natural undamped
2
0
= =
a
a
n
ess dimensionl ratio, damping
2
2 0
1
= =
a a
a
( )
( )
1
2
2
2
+ +
=
n n
D
D
K
D x
D y
Operational Transfer Function
( )
( ) ( ) ( )
( )
( )
( ) | |
K
j X
j Y
j j
K
j X
j Y
n n
n n
n n
/ /
2
arctan
/ 4 / 1
1 / 2 /
2 2 2
2
2
2
=
+
=
+ +
=
|
Frequency Transfer Function
Many medical instrument are 2
nd
order or higher
Figure 1.7 (a) Force-measuring
spring scale, an example of a
second-order instrument. (b)
Static sensitivity.
Output y ( t )
(b)
(d) (c)
1
K
s
x(t)
y(t)
y
n
y
n + 1
Resonance
2
Log
scale
1
2
-90
0.5
1
2
-180
1
0.5
0.5
Log scale e
Log scale e
K
1
t
t
Input x(t)
Slope K =
1
K
s
0
e
n
e
n
Y (je)
X (je)
|
Output
displacement
(a)
Input
Force x(t)
0
y(t)
2
nd
order mechanical force-measuring Instrument
( )
( )
( )
( )
2
2
dt
t y d
M t y K
dt
t dy
B t x
s
=
s
K K / 1 =
M
K
s
n
=
M K
B
s
2
=
B = viscosity constant
K
s
= spring constant
Natural freq.
Damping ratio
(c) Step response for overdamped case , = 2, critically damped case , = 1,
underdamped case , = 0.5.
(d) Sinusoidal steady-state frequency response, , = 2, , = 1, , = 0.5.
: 1 >
( ) K Ke
Ke
t y
t t
n n
+
+
=
|
.
|
\
|
|
.
|
\
|
+ 1
2
2
1
2
2
2 2
1 2
1
1 2
1
: 1 =
( ) ( ) K Ke t t y
t
n
n
+ + =
1
: 1 <
( ) ( )
2
2
2
1 arcsin
1 sin
1
K t K
e
t y
n
t
n
=
+ +
=
|
|
Overdamped
Underdamped
Critically damped
1
K
s
y(t)
0.5
t
2
1 , e e =
n d
Damped natural freq.
2
1
2 / 3
t
n
n
=
|
2
1
1
2 / 7
t
n
n
=
+
|
( )
( )
2
1
2
2 2
2
2
1
1
2
ln
1
2
exp
1
2 / 7
exp
1
1
2 / 3
exp
1
y
y
K
y
y
n
n
n
n
n
n
n
n
= A =
|
|
.
|
\
|
|
|
.
|
\
|
=
|
|
.
|
\
|
(
(
|
|
.
|
\
|
|
|
|
.
|
\
|
(
(
|
|
.
|
\
|
=
+
+
|
|
2 2
4 A +
A
=
and
Example 1.2: for underdamped second-order instruments,
find the damping ratio from the step response
Logarithmic decrement
( ) ( ) K t K
e
t y
n
t
n
+ +
=
|
2
2
1 sin
1
( ) ( )
d
t Kx t y =
d
t >
( )
( )
d
j
Ke
j X
j Y
t
=
Time Delay System
Log
scale
Log scale e
Log scale e
K
0
Y (je)
X (je)
|
d