Professional Documents
Culture Documents
Contents
Introduction to Validation Stages
of qualifications of Autoclave
Validation
Validation
Validation
Protocol of Autoclave
of Dry Heat Sterilizers And Tunnel
Validation
Validation
evidence which provides a high degree of assurance that a specific process meeting its attributes.
It
prove the safety efficacy, Purity & effectiveness of the drug product, medical devices & biologics in the marketplace & health system.
is Vital for
Safety
Fewer
interruptions of work Lower repair costs Elimination of premature replacement Less standby equipment Identification of high maintenance cost Reduction of variation in results Greater confidence in the reliability of results
PQ should always be done by the user because it is very application specific, and the vendor may not be familiar with these. As PQ should be done on a daily basis, this practically limits this task to the user.
Validation
Part 1. Overview on qualification and validation Part 2. Qualification of HVAC and water systems Part 3. Cleaning validation Part 4. Analytical method validation Part 5. Computerized system validation Part 6. Qualification of systems and equipment Part 7. Non sterile product process validation
Validation
Stages of qualification
Performance qualification
Change control
Validation
Results of calibration maintenance, verification
Defined schedule
Frequency based on Factors
Periodic
Requalification
After change
Equipment qualification
Design
Installation Operational
Performance
qualification (PQ)
Installation Qualification(IQ)
Installation qualification establishes that the instrument is received as designed and specified, that it is properly installed in the selected environment, and that this environment is suitable for the operation and use of the instrument. The qualification involves the coordinated efforts of vendor operating department project team (which provide input into the
qualification (OQ) is the process of demonstrating that an instrument will function according to its operational specification in the selected environment."
proper operation of equipment is verified by performing the test functions specified in the protocol. is drawn regarding the operation of equipment after the test functions are checked and all data has been analyzed.
The
A conclusion
Following
1.Application S.O.Ps
2.Utilization List
3.Process Description 4.Test Instrument Utilized To Conduct Test 5.Test Instrument Calibration 6.Critical Parameters 7.Test Function (List) 8.Test Function Summaries
Performance Qualification(PQ)
"Performance
Qualification (PQ) is the process of demonstrating that an instrument consistently performs according to a specification appropriate for its routine use ". PQ should always be performed under conditions that are similar to routine sample analysis. PQ should be performed on a daily basis or whenever the equipment is being used. In practice, PQ can mean system suitability testing, where critical key system performance characteristics are measured and compared with documented.
A. Introduction
Sterile products have several unique dosage form properties, such as Freedom from micro-organisms, Freedom from pyrogens, Freedom from particulates, Extremely high standards of purity and quality; However, the ultimate goal in the manufacture of a sterile product is absolute absence of microbial contamination.
Introduction(Con..)
1. To build sterility into a product 2. To demonstrate to a certain maximum level of probability that the processing and sterilization methods have established sterility to all units of a product batch
D value
It is time required for a 90% reduction in microbial population. Quantitative expression of rate of killing of micro organism. In other words, the D value will be affected by The type of microorganism used as BI, The formulation components and characteristics The surface on which the micro-organism is exposed The temperature, gas concentration, or radiation dose of sterilization process.
D value found by 2 methods, 1) Survivor curve method (log number of surviving organism versus time/gas concentration/radiation dose) 2) Fraction negative method
Z value Used exclusively in validation of heat sterilization process. Z value is reciprocal of slope of plot of log D verses T at which D value is found i.e. increase in temperature required to reduce D value of organism by 90 % (1 log reduction) F value Used exclusively in validation of heat sterilization process. It is time in min required to kill all spores in suspension at 121oC. Measures equivalent time
Moist heat (autoclave) Dry heat oven or tunnel Ethylene oxide Peracetic acid
2.Gas
3.Radiation
Gamma
Beta Ultraviolet
The
removal of air from the chamber and replacement with saturated steam.
Autoclaves
properly and all operations qualified through installation qualification and operation qualification (IQ/OQ).
Copper constantan wires coated with Teflon are a popular choice as thermocouple monitors.
Accuracy of thermocouples should be 0.5C. Temperature accuracy is especially important in steam sterilization
validation.
2.
Dry heat
B. subtilis var. niger spores B. subtilis, 5230 spores B. subtilis var. niger spores
3.
Ethylene Oxide
4.
Radiation
C. Heat-Distribution Studies
Heat-distribution
Cont..
Heat-distribution
effect of the load size and/or configuration on the cool spot location.
The
and the mean chamber temperature should be not greater than 2.5C .
Greater
equipment malfunction.
D. Heat-Penetration Studies
This
validation process.
The
main purpose is to determine the F0 value of container cold spot for containers 100 ml is
Cont..
Thermocouples
in the steam exhaust line, and in constanttemperature baths outside the chamber.
F0
temperature recorded by the thermocouple inside the container at the coolest area of the load.
F0
Heat-Penetration Studies(Con..)
Three
critical parameter associated with all wet heat sterilization Processes: 1. A minimum F value 2. A design F value 3. A sterilization process time Any changes in the load size, load configuration, or container characteristics must be accompanied; To prove that the cool spot location has not changed or, If it has, that it receives the design F0 time exposure from the sterilization cycle used.
E. Equipment Qualification
Prior
affect
requirements) are:
Accurate temperature and pressure measurement Air removal to some predefined level of vacuum distribution and uniformity in the
Temperature
chamber.
The
following :
1.Calibration of temperature and pressure sensors
provide additional necessary assurance that adequate lethality has been delivered to all parts of the load.
Calibrated
After
inoculated items or spore strips are recovered and subjected to microbiological test
procedures.
Strips
medium (soybean casein digest medium is typical) and incubated for up to seven days.
loads
are
protected
from
such
contamination by their primary containers (vials, bags) and many nonproduct loads are protected by
For
A. Introduction
Mainly
III.
Microwave Sterilizers
AND
CIRCULATION:
The
sterilizer releases large amounts of heat energy that serves to heat the items in the sterilizer.
In
Tunnel Steriliser
Positive pressure to entrance Even distribution of heat Belt speed recorder HEPA-filtered cooling air
The four main mechanism through which Heat transfer occurs are:
3. Heat-penetration studies:
located in the areas likely to present the greatest resistance to reaching the desired temperature.
Minimum studied.
and
maximum
temperatures
as
4. Mechanical repeatability:
During
all
these in
studies, of
repeatability
terms
temperature consistency, and reliability and sensitivity of all the oven and instrumental controls must be verified.
Air flow must be balanced in order to provide a gradual decrease in air temperature as items move along the conveyor.
In
2. Heat-Distribution Studies:
Thermocouples
cold spots are located as a function of the width of the belt and height of the tunnel chamber.
Peak
3. Heat-Penetration Studies:
Prior
sterilization,
4. Mechanical Repeatability:
Tunnel
sterilizers
must
demonstrate
consistency and reliability of all the tunnel controls (heat zone temperatures, belt speed) must be proved during the physical validation studies.
sterile and pyrogen-free commodities, validation studies must be done using both micro-organisms and microbial endotoxins.
The
Procedures for the validation of a tunnel sterilization: overkill approach is selected for the validation
The
study.
Select
Run
pattern.
Determine
adjacent
to
those
Several
various
temperatures
and/or
cycle
time
Z value of 54C.
Detectors
Loggers Logger
Wireless Temperature
Infrared Thermometer
Constant Temperature
Baths
Stopwatch
Voltmeter
or Ammeter
Optical Tachometer
INSTALLATION QUALIFICATION
The
checked
recommendations. A. Structural:
Check
correct leveling, proper insulation, presence of seals, and inspect for structural damage.
B. Filters: All filters used within the system must be recorded, such as those used with air (supply, recirculating) or in other utilities (e.g., steam, water). Some HEPA filters may need to be checked periodically by performing an integrity test or DOP. C. Electrical: Ensure conformance to National Electrical Code Standards D. HVAC: Ensure the system provides the RH, temperature, and pressure differential required.
E. Air Supply: Identify source (direct from the HVAC system or room air), duct size, duct material of construction, and air classification. F. Ventilation: Check that the ventilation exhaust duct exhausts to an appropriate area (not to an aseptic environment), and identify the method used to prevent back-flow. G. Door Gaskets: Check integrity of gaskets and materials of construction.
H. Heaters: Record the manufacturers model number, the number of heating elements, and the voltage, amperage, and wattage of the elements for the heaters. I. Lubricants: Make certain that any lubricants used cannot contaminate the material being sterilized or depyrogenated. J. Blowers: The blower must be mechanically sound, the volute in place and correctly balanced, and that the blades rotate in the correct direction.
OPERATIONAL QUALIFICATION
A. Temperature Monitors: The temperature controllers, recorders, and sensors on the process equipment must be calibrated before the unit can be operated reliably. B. Cycle Timer: The accuracy of the timer must be determined, so that assurance is provided for cycle length. C. Door Interlocks: If a unit is equipped with double doors, the interlocks must operate such that the door leading to the aseptic area cannot be opened.
D. Heaters: All of the heating elements must be functional. It is preferable to have them monitored continuously with ammeters in order that burnedout elements can be immediately detected. E. Cooling Coils: To enable a faster cool-down cycle, the air is often circulated across coolant coils. F. Belts: The belt speed is a critical operating parameter in both continuous hot-air tunnels and flame sterilizers. Recorders for charting the belt speed are recommended for units with adjustable speed settings.
G. Particulate Counts:
Particulate
the containers before and after sterilization to quantitate the particle load. H. Chamber Leaks:
The
QUALIFICATION TESTING
Upon
for:
Heat distribution studies, Loaded-chamber testing consisting of heat distribution and heat penetration studies.
1) Component Mapping Studies Before conducting the loaded-chamber heat penetration studies, component mapping should be conducted. The studies help to determine the coolest point within a specific load and item. 2) Empty-Chamber Testing The initial testing is performed on an empty oven or tunnel to establish the uniformity of temperature distribution. The thermodynamic characteristics of the empty unit are depicted in a temperature distribution profile.
3) Loaded-Chamber Studies
For
should be reduced. For tunnels, the temperature set point should be reduced and the belt speed increased if possible.
4) Bio-Challenge/Pyro-Challenge Studies
or endotoxin.
The challenge can be accomplished using commercial strips or suspensions of B. subtilis spores for sterilization or E. Coli endotoxin for depyrogenation.
assurance.
QUALIFICATION REPORT
After
the empty and loaded-chamber studies and biochallenge studies have been completed, the data must be analyzed to ascertain that all testing requirements have been achieved. The results of the biochallenge studies and F value computation must demonstrate the required degree of lethality according to the protocol. The following information should be provided in the process qualification validation report: 1. Protocol achievement 2. Summary of data 3. Deviations 4. Diagram
References
Berry
validation second edition, revised and expanded; Marcel Dekker series; 83-110.
Agalloco
J.A,
Carleton
F.A,
Validation
of
edition; 432
73