Professional Documents
Culture Documents
Slide 1 of 28
Validation
Introduction
l
Slide 2 of 28
Validation
Objectives
To introduce the concepts of :
l l
l
l l
Analytical procedure
Extent of validation Method transfer
Validation
Validation of analytical procedures requires:
l l l l l
Qualified and calibrated instruments Documented methods Reliable reference standards Qualified analysts Sample integrity
Slide 4 of 28
WHO EDM 1/2002
Validation
Validation protocol for analytical method
l l l l l l l
Statement of purpose and scope Responsibilities Documented test method List of materials and equipment Procedure for the experiments for each parameter Statistical analysis Acceptance criteria for each performance parameter
Slide 5 of 28
Validation
Qualification of the instrument
Make, model and makers manual Modifications
Slide 6 of 28
Validation
Characteristics of analytical procedures (1)
Accuracy
Precision
Repeatability Reproducibility
Slide 7 of 28
Validation
Relationship between accuracy and precision
Accurate but Inaccurate but imprecise precise Module 1,, Part 4: QC-related validation Slide 8 of 28
Validation
Characteristics of analytical procedures (2)
Ruggedness Robustness Variability caused by:
Day-to-day variations Analyst-to-analyst Laboratory-to-laboratory Instrument-to-instrument Chromatographic column-to-column Reagent kit-to-kit Instability of analytical reagents
Module 1,, Part 4: QC-related validation Slide 9 of 28
WHO EDM 1/2002
Validation
Characteristics of analytical procedures (3)
Linearity and range
Specificity Sensitivity Limit of detection Limit of quantitation
Slide 10 of 28
Validation
Calculated analyte in mg/mL
0.040 0.035 0.030 0.025 0.020 0.015 0.010 0.01
y
Calculated mg/ml
0.0100 0.0150
0.0101 0.0145
3
0.04
0.0200
0.0250
0.0210
0.0260
0.015
0.02
0.025
0.03
0.035
5
6
Module 1,, Part 4: QC-related validation Slide 11 of 28
0.0300
0.0400
0.0294
0.0410
Validation
Linearity Statistics
Intercept
Slope
-0.0002
1.0237
0.9978 0.9985
Correlation coefficient
Pearson Olkin and Pratt
3.4%
WHO EDM 1/2002
Validation
LOQ, LOD and SNR
Limit of Quantitation
Limit of Detection Signal to Noise Ratio
Peak A LOD
Baseline noise
Peak B LOQ
Slide 13 of 28
Validation
Different classes of analytical tests
Class A: To establish identity Class B: To detect and quantitate impurities Class C: To determine quantitatively the concentration Class D: To assess the characteristics
Slide 14 of 28
Validation
Characteristic
Accuracy Precision A
quant.
B
Limit test
C X X
D X* X
X X
X
X
X X
X X
X
X X
X X
X
X X
X
Validation
Extent of validation
New methods require complete validation
Pharmacopoeial methods require partial validation (or verification)
Slide 16 of 28
Validation
Analytical method transfer
Method transfer protocol and procedure
precision accuracy ruggedness
Validation
Chemical laboratory validation requirements (1)
Balances
Chromatography HPLC, HPTLC, GC, TLC Dissolution or disintegration apparatus Karl Fischer moisture determination Melting, softening or freezing point apparatus Ovens, refrigerators, incubators
Slide 18 of 28
Validation
Chemical laboratory validation requirements (2)
pH meter
Polarimeter - optical rotation Refractometer
Validation
Typical validation of HPCL assay (1)
System suitability (performance check)
system precision column efficiency symmetry factor capacity factor
Slide 20 of 28
Validation
Typical validation of HPLC assay (2)
Method validation
specificity accuracy precision linearity robustness
Slide 21 of 28
Validation
Biological assays
Can be difficult to "validate"
"Validity" on a case by case basis Strictly adhere to the Biological Testing monographs in pharmacopoeias
Slide 22 of 28
Validation
Microbiological testing requiring validation
Microbial limit testing
Microbial count Sterility testing Preservative effectiveness testing Environmental monitoring program
Biological testing
Module 1,, Part 4: QC-related validation Slide 23 of 28
WHO EDM 1/2002
Validation
Validation of microbial test procedures (1)
Virtually impossible to completely validate test procedures for every microorganism Neutralize /inactivate inhibitory substances, or dilute
Slide 24 of 28
Validation
Validation of microbial test procedures (2)
Incubation temperature and time
Media may not grow all microorganisms Variations in media may affect recovery Inhibitory disinfectants or preservatives Sample
procedures
handling, storage, transport
Module 1,, Part 4: QC-related validation Slide 25 of 28
WHO EDM 1/2002
Validation
Microbiological viable count method validation (1)
Methods
pour plate / spread plate membrane filtration Most Probable Number
Validation
Microbiological viable count method validation (2)
Membrane filtration conditions Incubation conditions Acceptance criteria
Slide 27 of 28
Validation
Sterility testing validation requirements
Media growth promotion, sterility, pH
Product validation Stasis testing Environmental monitoring Negative controls Challenge organisms
Module 1,, Part 4: QC-related validation Slide 28 of 28
WHO EDM 1/2002