To start logging, select B/Data Logging in the Data Logging configuration window. Continue by selecting C/Start Logging in the Logging window. If the Trigger setting is Periodic, you can follow how the logging advances in the lower part of the window. Both the amount of samples to be logged and the remaining time is displayed. If the Trigger setting is Keyboard, accept the values by pressing B/Accept Readings. Now the lower part of the window displays the amount of samples already logged and the time since the logging was started.
20.10.2000 15:32
DATA LOGGING Abs bar Pt100 385 C K NiCr/NiAl C Posit. kantti kHz Posit. kantti kHz Posit. kantti kHz Posit. kantti kHz 38802 s
Stop Logging MENU
MC5 Multifunction Calibrator Quick Guide for the Data Logging Utility
Applies to MC5 firmware version 1.90
Count
To stop/interrupt the logging, be it periodically triggered or triggered from the keyboard, select C/Stop Logging.
Channel 2 RTD-Temp. [C] 24.18 24.18 24.17 24.18 24.17 24.17 24.17 24.17 24.16 24.17 24.16 24.16 24.16
2
Back
MENU
xx x xx xx xxx x xxx - xx xx x x xxx x xxxxx x Transferring the Results to a Personal Computer - xxxx xxx xx xxxx xxx xx x A 32-bit Windows software called MC5_Datalog.exexx shipped xxx - xxx xx x xx is xxx x together with MC5 if you bought the Data Logging option. Install and start this soft- xx xx ware just as any other Windows software. x x xxx x xxxxx x - xxxx xxx xx xxxx xxx xx x All communication between the PC and MC5 is initiated from - xxx software xx xxx x xxx MC5_Datalog.exe. More information of the xx x xxin Appendix 3. - xx xx x x xxx x xxxxx x - xxxx xxx xx xxxx xxx xx x
3 Beamex 2000, 2002
Contents
* * * * * General Configuring Starting the Data Log Viewing the Results Transferring the Results to a Personal Computer
Data_Logging_Quick_Guide / 010702
Data Logging
General
The Data Logging utility is a firmware option that allows you to collect data using MC5. The collected data may then be viewed, transferred to a personal computer (PC) and printed using a utility shipped together with MC5 if the Data Logging option is purchased. The maximum amount of measuring channels can be up to seven (valid only if your MC5 includes a sufficient amount of measuring modules/ports). Data Logging may be configured to collect data using different triggers and saving methods as described further on.
Assigning Measurement Ports to Channels Move the cursor to one of the measurement channel rows in the Data Logging configuration window. Press C/Edit to open the following channel settings window: Select the measurement port first. The other settings depend on the port setting. To clear a channel, select port None in the channel settings window. To remove all channel settings, select D/MENU, 1/ClearAllChannels in the Data Logging configuration window.
20.10.2000 15:01
DATA LOGGING SETUP Channel Port Pressure Type Unit 1 P3:B Abs bar
Configuring
To start configuring the Data Logging (starting from Basic Mode), press D/Menu, C/Others, 5/Data Logging. The upper part of the Data Logging configuration window is reserved for seven measurement channels and the lower part for general settings. The following subchapters describe how they are configured.
20.10.2000 14:57
Cancel
Edit
OK
Trigger Available options: Periodic and Keyboard. When Keyboard is selected, the fields Interval, Samples and Time are not needed and thus not shown. Save Method Options: Instant, Average, Minimum, Maximum, Min & Max and All values. Interval Accepted values: 1 to 3600 seconds. Enter the value as an integer. Samples Maximum amount of samples: 13043. Time
1 2 3 4 5 6 7
20.09.2000 15:02
DATA LOGGING SETUP P3:B bar abs ET:RTD(mea) C Pt100 385 ET:TCi(mea) C K NiCr/niAl None None None None Periodic Instant Value 60 720 12 0 0
Basic Mode
Data Logging
Edit
MENU
Basic Mode
Data Logging
Edit
MENU
Automatically calculated based on the Interval and Samples settings. Alternatively, by changing the Time settings, the Samples value is automatically changed to suit the new Time settings. Maximum value: 30 days (720 hours). Note. In addition to the previously mentioned limits, the maximum values of Samples and Time settings also depend on the available free memory. If you enter values that require more memory than is available, MC5 automatically replaces the entered value with the currently allowed maximum value.