SoundPro DLX

Hand Held Sound Level Meter & Real-Time Frequency Analyzer

Owners Manual

SOUNDPRO DLX REAL TIME ANALYZER


OWNERS MANUAL

Part number 053-273B August 2005

CHAPTER

Overview chapter

Congratulations for choosing SoundPro DLX for your all-around sound level measurement needs. This instrument is an advanced sound level meter capable of producing a wide variety of point, averaged and integrated measurements of sound. Depending upon the SoundPro DLX model selected, many measurements can be broken into octave or third-octave components for an even finer analysis. Measurements can be viewed while they are occurring and in a summary display between measurement sessions. You can print from the summary display and transmit measurements results to a Quest application running on a personal computer for offline data analysis.

Instr ument char acteristics


Microphones
SoundPro DLX operates with interchangeable condenser or electret microphones designed for free field, random incidence and pressure measurements. To reduce distortions due to sound reflections, you can install an optional extender cable between the instrument and the preamplifier.

Multiple meters
SoundPro DLX is actually three meters in one. Each meter is separately configured, which enables you to make simultaneous measurements under different conditions. One of the meters can be set to perform octave and third-octave measurements of the sound signal.

Display
Both setup and measurements information is displayed on a touch-sensitive touchscreen. The displays are rich in detail and color, with full-featured capabilities that allow you to dynamically control what youre viewing in multiple ways. As a setup tool, you can quickly make selections from contextsensitive menus and view the settings in multiple displays. When viewing measurement results, you have a choice of viewing in dynamic, tabulated or historical formats, in some cases using panning and zooming options to obtain better resolution or observe long time trends.

Running studies
Each session consists of one or more studies performed with the same measurement settings. During a session, you can monitor accumulating results for that session. At your option, you can add additional information such as voice notes and drawings to a study when the study ends. Between sessions, you can view the final results in a Session Manager for sessions and studies residing in memory. Between sessions you can perform a wide variety of session editing, including renaming the session or any of its studies, edit much of the non-measurement information in the session and add additional voice notes and drawings.

Part number 053-273B

ii COPYRIGHT

Copyright
Subject to the inclusion of this Copyright Page, permission is granted to copy and distribute this manual. Permission to copy and distribute does not include permission to modify the text or illustrations in the manual in any way. Translations of this manual are prohibited without the express written permission of Quest Technologies, Inc.

Trademarks

The following are declared or US registered trademarks of Quest Technologies, Inc.: Quest, QuestSuite Professional, SoundPro. Windows is a US registered trademark of Microsoft, Inc. Hewlett-Packard is a US registered trademark of the Hewlett-Packard Development Company, L.P.

Updates
For the latest updates and corrections to this manual, contact your dealer or visit Quests website at http://www.quest-technologies.com. Feedback on the manual is welcomed and should be sent to service@questtechnologies.com.

Copyright 2004, 2005 Quest Technologies, Inc. Part number 053-273B

CONTENTS

Contents SoundPro DLX User Manual


Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii

Chapter 1: Overview chapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1


Instrument characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 About this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

Chapter 2: Getting started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3


Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ports and connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The keypad. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Powering-up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Customizing for your use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Display controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The SLM and RTA display groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Information and control area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Setup displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Display utilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sound Level Meter (SLM) display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SLM History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SLM Tabular display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Additional meter settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Real Time Analyzer (RTA) display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . RTA History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . RTA Tabular display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Sound curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Part number 053-273B

3 3 4 6 7 9 11 13 14 16 19 20 25 30 33 35 41 44 46 47

Chapter 3: Display fundamentals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11

Chapter 4: SLM display group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25

Chapter 5: RTA display group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41

ii CONTENTS

Chapter 6: Logging measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51


Logging Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51 Logging display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54

Chapter 7: Using the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .57


Positioning the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Running studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calibrations and verifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Adding information to a study . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preset triggers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Viewing current session results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Session Manager display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Common features and functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Session folders, displays and fields . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Study displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Deleting in the Session Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exporting sessions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Printing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The Global Setup display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Meter Profile Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Date and Time Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Security Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Battery, Sound, Display Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Meter Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Network Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Web Server Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacement and optional parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The back cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Microphone polarization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Pinouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . QuestSuite Professional II application. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Software system diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 58 59 63 67 71 73 74 76 79 83 83 84 87 89 93 94 96 97 98 99 101 101 105 107 108 109

Chapter 8: Managing session data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73

Chapter 9: Global setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87

Chapter 10: Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .101

Part number 053-273B

CONTENTS iii

Appendix A: Printout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111 Appendix B: Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113 Appendix C: Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123


Time-weighted measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Peaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Time-averaged measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Integrated measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Timing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Other formula parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Standards and approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Logging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Special functions (Class 1 models only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Environmental characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Mechanical characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 124 124 125 127 127 129 131 131 135 135 135 136 137 138 140 141

Appendix D: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129

Appendix E: Quest policies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143


Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .145

Part number 053-273B

iv CONTENTS

This page intentionally left blank

Part number 053-273B

LISTS

Figures
Chapter 1: Chapter 2:
21 22 23 24 31 32 33 34 35 36 37 38 39 41 42 43 44 45 46 47 48 49 410 411 412 51 52 53 54

Overview chapter Getting started

SoundPro DLX physical components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 Ports in the base of the instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 The five navigation keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 Security Logon display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8

Chapter 3:

Display fundamentals

Main Menu and display keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 Three types of scrolling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 Measurement display groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15 Status panels in the SLM display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 The Navigation Legend in the SLM display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17 Opening the Input Panel (SLM display) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 Input Panel (unshifted) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 Zooming and panning in the Log History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22 Color dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23

Chapter 4:

SLM display group

Sound Level Meter (SLM) display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25 Shared meter panel characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26 Bandpass analysis controls and indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29 Analysis Type menu (Class/Type 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29 SLM History display for Meter 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31 Parameter panel in the SLM History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31 Interval Menu for the Back Erase function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 SLM Tabular display for Meter 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34 SLM Settings page in Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35 One of the SLM Settings pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36 All four SLM Settings pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37 Leq/Lavg advisory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38

Chapter 5:

RTA display group

Real Time Analysis (RTA) display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41 RTA Parameter Panel features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42 Analysis Area in the RTA display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43 RTA History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44

Part number 053-273B

vi LISTS

55 56 57 58 61 62 63 71 72 73 74 75 76 77

Analysis Area in the RTA History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45 RTA Tabular display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46 RTA display with a sound curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47 RTA Settings page in Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48

Chapter 6:

Logging measurements

Logging Settings page with the seconds list displayed . . . . . . . . . . . . . . . . . . . . . . . . . . .52 Log History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54 Composite Menu in the Log History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56

Chapter 7:

Using the instrument

Illustration of a typical session . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59 Calibration Settings page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60 Study Information display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .63 Drawing Pad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65 Voice Note Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66 Trigger Manager in threshold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68 Summary display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72

Chapter 8:

Managing session data

81 Expanded session with one of its studies also expanded . . . . . . . . . . . . . . . . . . . . . . . . .73 82 Session Manager showing required and optional components . . . . . . . . . . . . . . . . . . . . .74 83 Contents of the Sessions General List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .76 84 Opening the Meter Settings display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77 85 A sessions Statistics display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77 86 A sessions Exceedance display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78 87 Session Summary display for TWA measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79 88 Expansion of a Studys General List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80 89 Opening the Logged Measurements display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81 810 Calibration display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82 811 Print display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84

Chapter 9:
91 92 93 94 95 96 97 98 99 910

Global setup

Features in the Global Setup display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87 Touchscreen calibration displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88 Save Profile dialog box and Input Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90 Date and Time Settings page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93 Security Settings page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94 Battery, Sound, Display Settings page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .96 Meter Information page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97 Network Settings page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98 Web Server Settings page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99 Web viewing home page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100

Part number 053-273B

LISTS vii

Chapter 10:
101 102 103 104 105 108 A1 A2 B1 B2 B3 B4 B5 B6 B7

Miscellaneous

Removing/replacing the back cover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .102 Remove/install a NiMH battery pack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103 Positioning NiMH batteries during installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .103 Installing alkaline batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104 Reset switch location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105 General Purpose Accessory connector (Ribbon: 26-MINI-D) . . . . . . . . . . . . . . . . . . . .107 Page 1 of a SoundPro DLX report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111 Page 2 of a SoundPro DLX report . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112

Appendix A: Printout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .111

Appendix B:

Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113

Slow response to a tone burst . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .118 Fast response to a tone burst . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .118 Impulse response to a tone burst . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119 Plot of A-weighted corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121 Plot of C-weighted corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121 Plot of Z-weighted corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122 C-A weighting response curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .122

Appendix C: Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123 Appendix D: Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129 Appendix E: Quest policies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .143

Tables
21 31 32 33 71 81 D1 D2 Keys and their functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 Navigation icons in Online Help system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 Function Icons and their meanings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 Status Bar icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18 Calibration History data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61 Manager Function icons. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75 Frequency ranges, by microphone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131 Characteristics of microphones alone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139

Part number 053-273B

viii LISTS

This page intentionally left blank

Part number 053-273B

2 OVERVIEW CHAPTER
Communications

Communications
SoundPro DLX is able to communicate with external devices in a variety of ways.

Ethernet ~ With an optional adapter, exchange information locally or through a computer network. USB ~ Connect by means of a USB adapter supplied with SoundPro DLX. Infrared ~ Print to a standalone or computer-supported printer via an infrared link. Wireless ~ With an optional adapter, exchange information point-to-point with your computer or through a wireless LAN. Web viewing ~ Remotely view SoundPro DLX displays live in a web browser.

Additional capabilities
You can periodically log data during studies. Data logging fills the accumulators only for the logging interval, then resets the accumulator to zero and repeats for the next logging interval. If your model is capable of obtaining filtered measurement results, you can log separately for each filter. You can secure your instrument against unauthorized use. If you shut down with security enabled, only a person who knows the security code will be able to use it. You can run studies automatically, either by triggering operation from a threshold or according to the calendar. Studies can be started, paused or the session in which they are running stopped. When stopping a session, you turn off the automatic triggering capability as well. You can compare filtered measurement results online with sound curves. Each sound curve shows the values previously measured over the audio spectrum. Sound curves help you detect changes during a study or differences between measurement results between studies.

About this manual


For best results, become familiar with this owners manual. If you have any uncertainty about terms, see the Glossary for definitions. If you need quick access to a particular topic, figure or table, see the Index at the back or the listings of contents, figures and tables at the front of the manual.

Notes in the text


Useful additional information or information of special importance is often shown in a special note format. The note below illustrates and explains the note format. Example: NOTE: The note format has three elementsa standard symbol at the left, a keyword such as NOTE or CAUTION that identifies the note type and two parallel lines separating note text from adjacent text.

Part number 053-273B

CHAPTER

Getting started

This chapter starts you off with the things you need to know to assemble the instrument and get it operating.

Unpacking
SoundPro DLX is shipped disassembled and unpowered. Your first task is to perform the minor assembly. To operate it, you must either install batteries or connect the external power supply.

Assembl y
Connect the preamplifier to the instrument case as illustrated in Figure 21, which also identifies some of the case features. If you have more than one microphone, you also need to assemble the microphone you will be using to the preamplifier. Measurement Microphone Preamplifier Knurled Nut

Infrared Port

Touchscreen

Keypad

Figure 21

SoundPro DLX physical components

Part number 053-273B

4 GETTING STARTED
Preamplifier

Preamplifier
The 6-pin connector on the preamplifier fits the mating connector on the instrument case in only one way. When assembling, rotate the preamplifier until it slips into the case connector. While gently applying downward pressure on the preamplifier with one hand, twist the black knurled nut with the other hand clockwise to tighten. CAUTION: Do not touch the gold-plated center pin of the preamplifiers microphone connector.

Measurement microphone
If installing a microphone, align it with the preamplifier connector, press together and twist the microphone to tighten.

Por ts and connector s


All wired connections to the case are made to ports located in the base of the instrument (Figure 22). Power Headset CompactFlash slot General Purpose Accessory connector Stylus Figure 22 Ports in the base of the instrument

Infrared window
You can print from SoundPro DLX to a standalone printer that has infrared capability. The information for the report is transmitted from the instrument through its infrared port (Figure 21). See Printing on page 84 for information about setting up a printing session.

Power jack
An external power supply is furnished with SoundPro DLX. Two cables are connected to the power supplyone for input AC power (standard AC plug) and the other for output DC power (pin plug). To provide external power to the instrument, connect the pin plug to the power jack on the base of the instrument, then connect the AC plug to an available power source. If you need to know about the power jack pinout, see External power on page 107. When external power is applied with the standard Nickel Metal Hydride batteries installed, a built-in charger operates as needed to ensure that the batteries remain charged. For information about installing

Part number 053-273B

GETTING STARTED 5
Headset port

batteries, see Installing or changing batteries on page 102. Battery charging is signalled by the Charging indicator on the instruments front panel.

Headset port
[OPTIONAL FEATURE] If your SoundPro DLX model contains the voice vote feature, you can plug a standard headset into this port to record and listen to voice notes. For information about adding a voice note during a session, see Voice note on page 65. For information about editing and adding voice notes between sessions, see Voice notes on page 82. The pinout for the headset port is given under Headset port on page 107.

CompactFlash slot
Standard CompactFlash port that has multiple uses.

CompactFlash ~ Plug in a CompactFlash card to provide access to external storage. Ethernet ~ With an optional adapter, operate SoundPro DLX on a wired network. Wireless ~ With an optional adapter, operate SoundPro DLX on a wireless network.

For wired or wireless operation, see Network Settings on page 98 to set the communications parameters.

General Purpose Accessory connector


A 26-pin connector for supplied cables or for customer-wired cables for the following purposes. The pinout for the connector is given under General Purpose Accessory connector on page 107.

USB interface ~ Cable with a USB connector is furnished with the instrument. Analog (AC Output) ~ Cable is optional (Cables on page 142). Logic (TTL) In ~ Not yet implemented. Logic (TTL) Out ~ Zero and 5 VDC levels available in response to threshold triggering. See Setting triggers on page 68.

When operating SoundPro DLX with a USB connection, see Network Settings on page 98 to set the communications parameters.

Part number 053-273B

6 GETTING STARTED
The keypad

T he keypad
All but the SoundPro DLX Navigation keys are labeled on its keypad panel. These labeled keys are explained in Table 21, below. Table 21 Keys and their functions Keys Key names and functions Main Menu ~ Press to display the Main Menu, from which you can make selections that will take you to other displays. Alternate functions ~ Press this key to toggle the Navigation Keys. See Alternate functions on page 17 for more information. Session stop ~ Press to stop a session. If a study is running when you press , the study is ended as well. See Sessions on page 58 for more information. Run/Pause ~ Press to start or stop a study within a session (pausing the session). See Studies on page 58 for more information. On/Off Backlight ~ When the instrument is switched off, press to switch it on. When the instrument is on, press briefly to turn backlighting on or off. Press and hold for about 5 seconds to turn the instrument off.

Navigation keys
The five keys in the center that of the keypad, pictured in Figure 23, are called the Navigation keys You can use them to move between displays and menus, but they often have other functions as well.

Figure 23

The five navigation keys

Part number 053-273B

GETTING STARTED 7
Standby indicator

Execute key ~ The circular center key is called the Execute key because its often used to complete an action initiated by other keys or controls. Up/Down keys ~ The keys above and below the Execute key, called the Up/Down keys in this manual, are commonly used to navigate in a vertical direction in a display. An example is to move from option to option in a vertical menu. Left/Right keys ~ The keys on either side of the Execute key, called the Left/Right keys in this manual, are commonly used to navigate in a horizontal direction. An example is to move between displays that are alternatives in a group of displays.

Standby indicator
Not implemented at this time.

Charging indicator
With rechargeable battery packs installed and SoundPro DLX connected to an AC source, battery recharging is indicated by a blinking green light.

Powering-up
A power-up start occurs when power is provided to the instrument after it was completely unpowered. The unpowered condition exists for a new instrument, when the user removes all sources of power (batteries and AC) or after a reset operation. After completing a power-up start, you can switch the instrument into hibernation for quick startups. See Starting from hibernation on page 8. NOTE: When powering up a new instrument, install the rechargeable batteries packs provided, connect AC power and allow the batteries to fully charge before switching the instrument on. See Installing or changing batteries on page 102 for battery installation.
To perform a power-up start:

1. Provide battery power, external power or both. After a few seconds, the Quest globe logo appears and powering up proceeds automatically.

Battery power ~ See Installing or changing batteries on page 102 for installation instructions. When both batteries are installed and external power supply is provided, the instrument runs on external power. External power ~ Connect the external power supply. For information, see Power jack on page 4.

2. The powering up sequence takes at least a minute. When it is finished, the Security Logon display appears (Figure 24).

Part number 053-273B

8 GETTING STARTED
Starting from hibernation

Figure 24

Security Logon display

NOTE: Logon security exists by default. You can disable it and change the logon code after successfully logging on (Security Settings on page 94). 3. Tap the security code (factory default code is 1-2-3-4) 4. Tap Enter in the display. If you entered the correct code, the Sound Level Meter (SLM) display appears. See The SLM and RTA display groups on page 14 for an overview of the six monitoring displays, including the Sound Level Meter (SLM) display.

Starting from hibernation


When the instrument is on, press and hold for about five seconds to put the instrument into hibernation. The hibernation state is also called the power off state. While you are holding the key, a message appears in the display telling you how much time is left before the instrument powers off. If you release the key before the counter reaches zero, the instrument remains on. NOTE: If you attempt to power off during a study, a message with a counter appears telling you that the session will be paused (study ended) in two seconds. If you hold the key until that time expires, the session is paused and the standard power off message appears giving you three more seconds to decide if you want the power off to occur. In hibernation, the display is turned off to conserve power consumption, and all but (the On/Off/Backlight key) is inactive. Sufficient power is provided by the main batteries or external power source, however, to preserve all the software initializations that were performed during the powering on operation. A restart from hibernation is therefore much faster than a restart from an unpowered condition.

Part number 053-273B

GETTING STARTED 9
Software reset

Software reset
If your instrument ever appears to be operating abnormally, you should remove all operating power to return it to an unpowered condition. You can do that with either a software or a hardware reset. Try the software reset first. If that doesnt correct the problem, see Hardware reset on page 104.
To do a soft reset:

1. Using three fingers, simultaneously press the Up/Down/Right navigation keys on the front panel. 2. Hold the keys depressed for several seconds until the display begins to break up. 3. Release the keys and wait. Powering up should begin after a short delay.

Customizing for your use


Once you have the instrument operating, youll probably want to customize it for your conditions and individual use. For example, you can set the time and date (Date and Time Settings on page 93), change speaker volume Battery, Sound, Display Settings on page 96) and specify a language (Meter Information on page 97). To get busy with measurements, you may want to choose a meter profile (Meter Profile Settings on page 89) to establish a favored measurement setup.

Part number 053-273B

10 GETTING STARTED
Customizing for your use

This page intentionally left blank

Part number 053-273B

CHAPTER

Display fundamentals
T he Main Menu

This chapter introduces some basic features and functions in SoundPro DLX displays.

You can get to any display in the instrument by first showing the Main Menu, then choosing from that. Most selections on the Main Menu will take you to groups of displays, where you will use the navigation keys to move from display to display within the group. To show the Main Menu, press . Figure 31 illustrates the use of this key to show the Main Menu and the appearance of the menu itself. The menu contains a stack of panels, each of which contains an icon representing the display or group of displays available with that choice. You can select from the Main Menu using the navigation keys, or you can select by tapping the SoundPro DLX stylus on the touchscreen. The procedure below introduces the navigation key method. See Navigation Panel on page 16 for information about selecting directly from the touchscreen.
Using navigation keys to select from the Main Menu:

1. After showing the Main Menu, press the Up/Down keys to make a selection. A selection is indicated by a red box around the selected panel. In Figure 31, the top panel is selected. 2. Press to execute the selection.

NOTE: You can accomplish both steps of this procedure by tapping in the Main Menu using the touchscreen stylus. See User input on page 13.

Part number 053-273B

12 DISPLAY FUNDAMENTALS
Online help system

Chapter 4, SLM display group Chapter 5, RTA display group Chapter 6, Logging measurements Viewing current session results on page 71 Chapter 8, Managing session data Setup displays on page 19 Chapter 9, Global setup Online help system, below Figure 31 Main Menu and display keys

Online help system


The bottom panel on the Main Menu is a link to SoundPro DLXs Online Help system. The home page for Online Help contains a touch-sensitive representation of the Main Menu, shown above. Tap in that graphic for information about the indicated display. The home page, as well as every other page in the help system, contains navigation icons at the bottom. Table 31 Navigation icons in Online Help system Icon Home page Keyworded topics ~ Tap to jump to the keyword index, then tap a letter in the index to jump to a list of keywords beginning with that letter. Definitions ~ Tap for an index of definition types. Definitions are organized into three categories: General, Measurements and Measurement Parameters. Procedures ~ Tap to jump to a list of procedures, alphabetically arranged. Tap in the list for information about that type of procedure. Back ~ When moving from topic to topic in Online Help, tap this icon to go back one page. Forward ~ After tapping the Back arrow, tap this forward arrow to reverse direction and jump forward in the navigation sequence. Purpose

Part number 053-273B

DISPLAY FUNDAMENTALS 13
Display controls

Display contr ols


User input
SoundPro DLX has two types of user input deviceskeys and the touchscreen. When controlling with a key, press the key. When working directly on the touchscreen, use the SoundPro DLX stylus that is stored in a slot at the base of the instrument. The stylus is illustrated at the right..

Tap ~ A brief touch with the tip of the stylus. Double-tap ~ Two brief touches with the stylus in quick succession. Touch and drag ~ Touch the screen and hold in contact while moving the stylus across the screen. Touch and hold ~ When selections are available from a scroll list, touch the scroll arrow and hold to scroll continuously through the list.

Scrolling
Sometimes you can change the information on display or set a parameter by scrolling. SoundPro DLX has three types of scrolling: menu scrolling, list box scrolling and screen scrolling. Examples of these three types are given in Figure 32. Top of list (dimmed arrow) More menu choices (dark arrow) Scroll Arrow

Scroll Bar

Menu Figure 32

List box Three types of scrolling

Vertical Scroll Bar

Menu scrolling ~ Menus show your choices in a long list, but not necessarily all at once. You may have to scroll the menu to see other choices. When scrolling is available, the menu will have arrows at the top and bottom of the list, as shown in the example above. Tap a dark arrow to scroll in that direction. A dimmed arrow means youre reached the end of the list in that direction. Either tap to select from the list or use the keys. When using keys, press the Up/Down keys to scroll, then press to select. Scrolling with keys wraps around, which means you can quickly jump from one end of the list to the other. This can come in handy for long lists.

Part number 053-273B

14 DISPLAY FUNDAMENTALS
The SLM and RTA display groups

List box scrolling ~ List boxes show you only the current choice, but you can change that current choice by scrolling. Tap arrows in the list box field to scroll through the choices. You cant use keys to scroll in a list box (not a wraparound list). Screen scrolling ~ When the amount of information in a display exceeds screen size, a scroll bar appears depending upon the excess. When the excess is left to right, a horizontal scroll bar appears. When the excess is top to bottom, a vertical scroll bar appears. Tap a scroll bar arrow to move the display in that direction. Tap and hold for continuous scrolling. You can also tap, hold and drag the Scroll Bar.

T he SLM and RTA display g r oups


After successfully completing the power-on procedure, the first display you will see is the Sound Level Meter display (SLM display). This display is one of three displays in the SLM Group. This group, and the closely corresponding Real Time Analysis Group (RTA Group) are shown in Figure 33.

SLM Group
The sound detected by the microphone can contain sound frequency components that vary over a relatively wide spectrum, depending upon the microphone type (Microphones on page 139). In general, displays in the SLM Group present measurements of that sound signal for the entire audio spectrum (broadband) without regard to variations among the frequency components. The SLM display in the SLM Group shows results for three measurements on all three meters at once. The SLM Tabular display shows you real-time tabulations of a large set of measurements, meter by meter. The SLM History display charts the last 30 seconds of measurements you select, meter by meter. Detailed information about the features and functions in these displays is given in Chapter 4, SLM display group.

RTA Group
The three displays of the RTA Group show results for one of the three Meter 1 measurements. If youre running broadband studies youll see broadband results in the RTA displays. If youre running filtered studies, however, youll see both broadband and filtered results in the RTA displays. The RTA Group contains the RTA display, the RTA Tabular display and the RTA History display that in many ways each correspond to the similarly named displays in the SLM Group. Detailed information about the features and functions in these displays is given in Chapter 5, RTA display group.

Navigational similarities
Both groups are available as separate selections from the Main Menu. After selecting, you can move between the displays of each of the groups in analogous ways. These navigational similarities are illustrated in Figure 33.

Part number 053-273B

DISPLAY FUNDAMENTALS 15
Navigational similarities

Sound Level Meter (SLM)

SLM History

SLM Tabular

Real Time Analyzer (RTA)

Main Menu RTA History Figure 33 RTA Tabular Measurement display groups

Part number 053-273B

16 DISPLAY FUNDAMENTALS
Information and control area

Infor mation and contr ol ar ea


The bottom portion of all displays in the SLM and RTA groups is an information and control area that contains a Navigation Panel, a Status Bar and the displays Title Bar (Figure 34). The Title Bar is a feature of all SoundPro DLX displays. Most displays contain the Navigation Panel, but only those displays that are directly associated with controlling measurements contain the Status Bar.

Navigation Panel Status Bar Title Bar Figure 34 Status panels in the SLM display

Navigation Panel
The icons in this panel are arranged in three clusters that are separated by short vertical bars. Within each cluster are touch-sensitive icons that look like the physical navigation keys. Associated with the key icons in each cluster are function icons. The key icons in a cluster are all touch-sensitive. Tapping a key icon executes the associated function in that cluster. In addition, pressing the actual keys that look like the key icons executes the same function. Table 32 explains most of those functions. Table 32 Function Icons and their meanings Function Icon Meaning Change to a different Display. Change to a different meter. Go to the corresponding Instrument setup display. Select one of the display Parameters. Change the Range Base and Range Ceiling. Press the Execute key. Change the selected filter band. Back Erase: Remove data for a selected time period. Capture a sound curve (Sound curves on page 47).

Part number 053-273B

DISPLAY FUNDAMENTALS 17
Status Bar

Alternate functions
The functions icons that appear in the Navigation Panel frequently differ between displays. Some displays lack function icons for some navigation keys. Others have multiple functions for those keys. When multiple functions exist, you need to press (Alt f key) to change to the alternate functions for the display. Pressing again changes back to the original functions, which are known as the standard or default functions for the display. Thus, acts like a toggle between the standard and alternate functions and function assignments. Figure 35 identifies elements in the Navigation Legend of the SLM display for the displays standard and alternate functions. Standard Functions

Function Icons

Alternate Functions

Left/Right keys Figure 35

Up/Down keys

Execute key

The Navigation Legend in the SLM display

Zoom functions
Not all function icons are shown in Table 32. In addition to functions varying with the display and with , a Zoom command in the display can introduce zoom and pan functions. For an explanation of zooming and panning in SoundPro DLX, as well as when and how you can do it, see Zoom and pan on page 21.

Status Bar
The Status Bar gives you critical information about the status of the instrument and the running of studies, including a display of the Run Time Clock. Status information is signalled by icons that change appearance when certain conditions change. Table 33 explains the meanings of the Status Bar icons.

Part number 053-273B

18 DISPLAY FUNDAMENTALS
Status Bar

Table 33 Status Bar icons Status Icons Meanings Logging MemoryShaded area drops as memory fills with data. Battery ConditionAppears in the second icon position when operating under battery power. Green area shrinks as batteries discharge. Combined power ~ Appears in the second icon position when both batteries and external power are connected. The instrument operates under external power but may be charging the batteries. Stopped SessionAppears when no session is open (alternates in place with the Running Study and Paused Session icons). Running StudyAppears when a study is running (alternates in place with the Paused Session and Stopped Session icons). Paused SessionAppears when the session is paused (alternates in place with the Paused Session and Stopped Session icons). Overload ConditionBlackened during an input overload; dimmed otherwise.a Under Range ConditionBlackened during an under range condition; dimmed otherwise. Triggers enabledAppears and blinks or is steady when you have one or both trigger types enabled. See Preset triggers on page 67. Preamp disconnected ~ Appears and blinks if the preamp is electrically disconnected from the body or from an optional extension cable.
a) The overload status bar icon will turn black indicating the top of the linear operating range has been reached before displayed signal level linearity errors exceed the class 1 tolerance for level linearity error.

Part number 053-273B

DISPLAY FUNDAMENTALS 19
Title Bar

Run Time Clock


The Run Time Clock shows the elapsed time for a study. When you press to start a study, the Run Time Clock immediately resets to zero, the default name for the study appears, then the clock appears in place of the studys name and shows for the duration of that study. During a study, the Run Time Clock steadily increments in hours, minutes and seconds. When no study is running, the Run Time Clock shows the final run time for the last study that ran.

Title Bar
All SoundPro DLX displays are named in a Title Bar that appears at the bottom of the display. The Title Bar also holds the Input Panel icon. For information about this icon and using the Input Panel, see Input Panel on page 20.

Setup displays
Most SoundPro DLX setup procedures are contained in two groups of displays Instrument Setup and Global Setupand each consists of many setup pages. Both displays are available as selections from the Main menu. Select the top icon shown at the right to enter Instrument Setup. Select the bottom icon shown at the right to enter Global Setup. The selection always takes you to the page in the display that you last visited. These setup displays have many common features. See Chapter 9, Global setup, for information about all Global Setup displays. Information about the five Instrument Setup displays is given in the following sections.

Sound Level Meter Settings ~ See Additional meter settings on page 35 Real Time Analyzer Settings ~ See Sound curves on page 47 Trigger Manager ~ See Preset triggers on page 67. Logging Settings ~ Logging Setup on page 51 Calibration Settings ~ See Calibrations and verifications on page 59.

Part number 053-273B

20 DISPLAY FUNDAMENTALS
Display utilities

Display utilities
This section covers a number of software utilities that can appear at your command in many displays.

Input Panel
The Input Panel icon is located in the displays title area at the bottom of the touchscreen. Display the Input Panel by tapping that icon. Figure 36 illustrates the Input Panel in the SLM display.

Displays Title Area

Input Panel Icon

Input Panel Title Bar Figure 36 Opening the Input Panel (SLM display)

Moving ~ To move the Input Panel around the display, touch its title bar and drag. Closing ~ To close the Input Panel, tap the Input Panel icon again.

Using the Input Panel


To add or edit text in an editable text field, open the Input Panel. Tap in the text field. A blinking cursor appears where you tapped.

Backspace Return Arrows Accent Figure 37 Space Input Panel (unshifted)

Keyboard characters change when you tap a special function key (CAP, Shift, Ctrl and ).

Replace all text ~ Tap any key while all text is highlighted to begin entering new text. Position the cursor ~ Tap in the field where you want to start. After establishing a starting position, you can tap arrow keys to move from character to character.

Part number 053-273B

DISPLAY FUNDAMENTALS 21
Zoom and pan

Remove a character (non-shifted keyboard) ~ Position the cursor to the right of an unwanted character, then tap the Backspace key. Remove one or more characters (shifted keyboard) ~ Position the cursor to the right of the unwanted character(s), then tap the left arrow to sequentially highlight characters. Alternatively, tap and drag over the characters to highlight them. Tap Del to remove. Add a character ~ Position the cursor to the right of wanted characters, then tap keys to add new text. One capital/symbol ~ Tap Shift before entering the character. After you tap, the text characters revert to lower case letters. String of capitals ~ Tap CAP before entering text. Tap again to return to lower case letters. Accented letters/symbols ~ Tap the Accent key (). After you tap, the text characters revert to unaccented letters. Escape ~ Tap Esc to remove settings (Ctrl, Shift, )

Zoom and pan


When you enter a display in which measurements are graphed, the scale on the y-axis is always set to the current Measurement Range. In some displays, however, you can alter this range by zooming and panning. Zooming means that you can magnify or shrink the range and display the result over the entire display. Panning means that you can shift the range up or down in the display, changing what remains visible.

Result of zooming and panning


Figure 38 illustrates zooming and panning in the Log History display. The display at the right in this figure is the zoomed and panned version of the original display at the left for the same data. Relative to the original display, the zoomed and panned display shows the following features.

The full-scale range has been decreased from 100 dB to 50 dB, which means that the zoomed display shows a magnified version of the original display. While keeping the Range Ceiling nearly constant (small change from 90 dB to 85 dB), the Range Base has been moved up from 10 dB to 35 dB so that all of the measurement data remains in view. The Execute key has changed its function, as indicated by the red X in the zoomed display. Press Execute to end zooming and panning.

Part number 053-273B

22 DISPLAY FUNDAMENTALS
Zoom and pan

dB Range

Original display Figure 38

Zoomed and panned display

Zooming and panning in the Log History display

Operating zoom and pan controls


The zoom and pan controls become available when you tap the Zoom command in the display. In the Logging display example given above, the Zoom command is at the right edge of the Navigation Panel. Not all displays have the Zoom command in the same location. See the sections covering other displays for information about the availability of zooming and panning and the location of the Zoom command.
To zoom and pan:

1. Press Zoom to make the zoom and pan controls available in the Navigation Panel. 2. Press the Left/Right keys on the front panel to zoom in or out. You can also zoom by tapping the icons associated with the Left/Right keys in the Navigation Panel. Zoom changes affect the top and the bottom of the display range by the same amount but in opposite directions. For example, if the bottom of the range increases by 5 dB, the top of the range decreases by 5 dB to net a range reduction of 10 dB (magnification). 3. Press the Up/Down keys on the front panel to pan up or down. You can also pan by tapping the icons associated with the Up/Down keys in the Navigation Panel. Pan changes affect the top and the bottom of the display range by the same amount and in the same direction. For example, if the bottom of the range increases by 5 dB, the top of the range also increases by 5 dB for a range shift upward of 5 dB. 4. When done viewing under zoom and pan conditions, press the Execute key to restore the original settings.

Part number 053-273B

DISPLAY FUNDAMENTALS 23
Changing colors

Changing colors
If you can change a color, follow the instructions for displaying the Color dialog box. When this dialog box appears, a black rectangle surrounds the current color selection. NOTE: Only Basic colors are available. Please disregard the Custom Colors section
To change the color

1. Tap any color in the Color dialog box. The black rectangle moves to the new color. 2. Tap . The Color dialog box closes and the new color is applied. You can cancel out of the Color dialog box without changing the color by tapping . Figure 39 Color dialog box

Part number 053-273B

24 DISPLAY FUNDAMENTALS
Changing colors

This page intentionally left blank

Part number 053-273B

CHAPTER

SLM display group

SoundPro DLX operates with three sound level meters, all making independent measurements at once. This chapter explains how to set up the meters and how to use the displays in the SLM group to view measurement results.

Sound Level Meter (SLM) displ ay


As explained under The SLM and RTA display groups on page 14, you can launch the SLM displayone of the displays of the SLM display groupfrom the Main Menu. Press to display the Main Menu, scroll to the icon shown at the right, then press . The SLM display appears (Figure 41). You can also select directly from the Main Menu by tapping the icon.

Meter 1 Panel

Meter 2 Panel

Meter 3 Panel

Display status area

Figure 41

Sound Level Meter (SLM) display

Meter panels
The distinguishing feature of the SLM display is the presence of three meter panels. Except for the Measurement Range parameter, each meter functions entirely separately from the other two and has its own setup controls and measurement displays. Meter 2 and Meter 3 are identical in appearance and function, but Meter 1 has one feature that the other two dont havethe filtering controls. These filtering controls are located on a tab just above Meter 1 in the SLM display. To learn about using setting the filtering parameters, see Filtered analysis on page 28.

Part number 053-273B

26 SLM DISPLAY GROUP


Common meter panel characteristics

Common meter panel characteristics


The characteristics that all three SoundPro DLX meters have in common are identified in Figure 42. Although the characteristics are common to all three meters, you can customize meter settings so that each meter makes measurements in different ways. Frequency Weighting Time Response Measurement 1 Meter Number

Measurement 2 Measurement 3

Range Base Figure 42

Maximum SPL SPL Activity Bar Shared meter panel characteristics

Range Ceiling

Parameters
The information shown in red in each meter panel are the current parametric settings for the meter. These settings control the operation of the meter and are user-selectable within limits. See Changing settings in SLM displays on page 27 for information about making changes. The setup parameters and the options available are specified below. For a detailed explanation of these parameters and their settings, see Appendix C, Measurements.

Frequency Weighting ~ Each meter can filter the incoming signal according to a frequency weighting curve that you select. The standardized weighting curves available are A, C, Z and C-A. A-weighting and C-weighting are industry standards that give greater weight to mid-range frequencies; Z-weighting means no weighting across the sound spectrum; C-A weighting means that both C-weighting and A-weighting are applied and the differences reported. Time Response ~ Each meter can react to the time-varying signal with a selectable time constant. The standardized time constants available are Fast, Impulse and Slow. Measurement Range ~ The Measurement Range is defined by the sum of the Range Base and the instruments dynamic range. To change the Range Base, see Changing meter parameters and measurements on page 28. The dynamic range is a fixed characteristic but depends upon the Analysis Type setting. If Analysis Type is set to Off, the dynamic range is 100 dB. If set to either 1/1 or 1/3, the dynamic range is reduced to 80 dB. For information about changing the Analysis Type, see Analysis Type menu on page 29.
Part number 053-273B

SLM DISPLAY GROUP 27


Changing settings in SLM displays

Measurements
You can select from a list of many measurements for each of three positions in the meter panel. These measurements are identified in Figure 42 as Measurement 1, Measurement 2 and Measurement 3. The names are shown in light blue and the values are shown in black. For Measurement 1, the unit of measurement for Measurement 1 is also shown and colored blue. Any measurement can be selected for any of the three measurements of a meter. To select a different measurement, see Changing settings in SLM displays on page 27. For Meter 1, all three measurements report in accordance with the filter control settings (Filtered analysis on page 28). If the Analysis Type is set to OFF, all three Meter 1 measurements are broadband values. If filtering is chosen, all three measurements of Meter 1 are the filtered results for the selected filter.

SPL Activity Scale


In addition to showing the values of three selected measurements, each meter provides a visual indication of the current sound pressure level (SPL) on its SPL Activity Scale. All three SPL Activity Scales are calibrated to the same range, but the SPL activity indicated by each meter depends upon the meters setup. The numbers at the ends of the SPL Activity Scale specify the minimum (Range Base) and maximum (Range Ceiling) values of the calibrated range.

Activity Bar ~ The dark blue portion of the SPL Activity Scale represents the current SPL activity within the indicated range. If you do not see the Activity Bar or if it runs all the way across the Activity Scale, try changing the range setting. See To change the range setting: on page 28. Maximum SPL ~ When you see a light blue band at the right of the Activity Bar, that represents the maximum SPL in the preceding one-second interval.

Changing settings in SLM displays


The following instructions apply to all displays in the SLM display group. When you change something in one display, for example a meter or a meter parameter, the change is effective in all other SoundPro DLX displays in which that setting applies.

Selecting meters
Change a meters setting by first selecting the meter. A selected meter has orange borders to set it apart from the other two meters.
To select a meter with navigation keys:

1. The Navigation Panel must be showing the standard functions (Navigation Panel on page 16). 2. Press the Up/Down keys to select the meter.
To select a meter using the touchscreen:

Tap anywhere in the meter panel.

Part number 053-273B

28 SLM DISPLAY GROUP


Filtered analysis

Changing meter parameters and measurements


Each meter has adjustable parameters. The collection of settings of those parameters for a meter is the meters current configuration or setup. All but the range setting are selected from menu options. Some menus are long and contain Scroll Arrows (Scrolling on page 13).
To make changes on the touchscreen:

1. Tap the setting. A menu appears (also selects the meter if not previously selected). 2. Tap to select from the menu. (You may have to tap Scroll Arrows to reach all options.) The new setting is applied and the menu closes.
To make changes with keys:

1. Select the meter as explained above. 2. If the Navigational Panel is not showing the Standard functions, press . 3. Look on the Navigation Panel for the keys associated with the selection icon ( ). Press those keys to step through the settings and highlight the parameter you want to change. 4. Press 6. Press . A menu appears. . The new setting is applied and the menu closes. 5. Press the Up/Down keys to select from the menu.

To change the range setting:

The three meters share the same Measurement Range, but the range can be translated up or down in 10 dB increments. For example, the Range Base can be increased from 30 dB to 50 dB, and when you do, the Range Ceiling for broadband measurements increases from 130 dB to 150 dB 1. Navigate to the SLM display. 2. The Navigation Bar must be showing the alternate functions (Alternate functions on page 17). 3. Press the Up/Down keys to translate the range. NOTE: The optimum range has been selected when the SPL activity bar (SPL Activity Scale on page 27) is operating near the center of its range. If the overload status bar icon turns black, increase the level range. If the under range status bar icon turns black, decrease the level-range (these icons are discussed under Status Bar on page 17.

Filtered analysis
SoundPro DLX always makes measurements over the applicable frequency range of the instrument (broadband analysis), but you can also filter measurements for display and accumulation purposes. Filtered analysis only applies to Meter 1. Depending upon your model, you can analyze the broadband signal using octave or third-octave filter bands.

Part number 053-273B

SLM DISPLAY GROUP 29


Filtered analysis

When doing a filtered analysis, all measurements shown in the Meter 1 panel are either broadband or filtered values according to your choice. The tab at the top of the Meter 1 panel in the SPL display indicates the type of analysis and the selected filter, if any (Figure 43). Analysis Type setting Filter setting

Filter menu

Figure 43

Bandpass analysis controls and indicators

Analysis Type menu


Choose between broadband and filtered analysis from the Analysis Type menu (Figure 44). This menu is available in the SLM display and in other displays of the SLM and RTA groups.

OFF ~ No filtering (broadband analysis). If the Analysis Type is set to OFF, you cannot make a selection from the Filter menu.

Figure 44 Analysis 1/1 ~ Full octave analysis. If the Analysis Type is set to 1/1, you Type menu (Class/Type 1) can select from among 9 or 11 filter options from the Filter menu depending upon your SoundPro DLX model. 1/3 ~ Third-octave analysis. If the Analysis Type is set to 1/3, you can select from among 27 or 33 filter options from the Filter menu depending upon your SoundPro DLX model.

To set the analysis type:

1. Navigate to the SLM display. 2. If a session is open, press


Part number 053-273B

to stop it.

30 SLM DISPLAY GROUP


SLM History display

3. Tap the Analysis Type setting on the Meter 1 tab. The Analysis Type menu appears. 4. Tap to select one of the available selections. The menu closes, your selection is conformed on the Meter 1 tab and in all other displays where the selected filter is shown.

Filter menu
Except when the Analysis Type is set OFF (broadband), you can display either filtered or broadband results for all measurements selected in the Meter 1 panel. Select the filter by selecting its center frequency from the Filter menu. Select a broadband display by selecting OFF from the Filter menu.
To select from the filter menu:

1. Navigate to the SLM display. 2. If a session is open, press to stop it. 3. Select the analysis type, as explained above. 4. Tap the Filter setting on the Meter 1 tab. The Filter menu appears. An example of the Filter menu is shown in Figure 43, but the menu might look different depending upon whether your model provides third-octave analysis and your Analysis Type selection. 5. Tap to select either broadband (OFF) or one of the filters. The menu closes, your selection is conformed on the Meter 1 tab and in all other displays where the selected filter is shown.

SLM Histor y display


The SLM History display plots graphs of selected measurements for the last thirty seconds. The measurement curves update on a second-by-second basis, adding current data at the right, losing old data at the left, so that the curves appear to scroll automatically to the left. Unlike the logging history, this historical record is not saved. Also unlike logging, the results shown in the SLM History display pertain to the entire study, not just to incremental logging intervals. A particular advantage of the SLM History display is that you can use it to monitor the last thirty seconds of SPL values, whether you are running a study or not.

Navigating to the display


In general, you reach the SLM History display after first selecting the SLM display group from the Main Menu (Sound Level Meter (SLM) display on page 25). In the SLM display, press the Left key to go to the SLM History display. An example of this display is shown in Figure 45.

Part number 053-273B

SLM DISPLAY GROUP 31


Parameter Panel

Parameter Panel Measurement Legend

Display Range

Tracking Area

Figure 45

SLM History display for Meter 1

Parameter Panel
The Parameter Panel identifies the selected meter and its configuration (Figure 46). Frequency weighting Time Response Analysis Type Selected Filter Meter Number

Figure 46

Parameter panel in the SLM History display

You can change the meter and all but the range parameter directly in this display. See Changing settings in SLM displays on page 27.

Measurements
The Measurement Legend identifies the four measurements tracked in this display by name and by curve color. You can change any of these measurement directly in this display, and the change updates in all other displays where this selection applies. See Changing settings in SLM displays on page 27.
To change curve color

General information about working with the Color dialog box is given under Changing colors on page 23.

Part number 053-273B

32 SLM DISPLAY GROUP


Tracking Area

1. Tap the color in the Measurement Legend. 2. Tap any color in the Color dialog box. 3. Tap to complete the operation.

Tracking Area
Measurement results are shown in the Tracking Area of the display, where the values at each second of time are graphed against the Display Range. The display opens with the Display Range set to the Measurement Range (Figure 45).

Changing the Display Range


You can scale and shift the Display Range by zooming and panning. To make the zoom functions available, tap Zoom in the Navigation Panel. For information about using zoom and pan in SoundPro DLX displays, see Zoom and pan on page 21. NOTE: The zoom and pan changes are valid only in this display. You must cancel them before you can leave this display

Back Erase function


NOTE: This feature available only in Class/Type 1 models. While a study is running, you can erase newly-acquired data at any time using the Back Erase function in the SLM History display. The data are erased for a selectable time interval that immediately precedes the moment you executed the command. For example, if you select a 10-second interval, the last 10 seconds of newly-acquired data are erased from memory. The time interval options change during the study, with five seconds being the first available option and two hours being the longest available option.
To selectively erase data

1. If the Navigation Panel is set to the standard functions, the Back Erase icon identified in Figure 47 will be associated with the Execute key. If not, press . 2. Press Execute. An Back Erase Interval menu appears. 3. Tap to select from the menu. The Back Erase Interval menu closes, the data is erased and the Run Time Clock setting is corrected.

Part number 053-273B

SLM DISPLAY GROUP 33


SLM Tabular display

Back Erase Interval menu

Back Erase icon Figure 47 Interval Menu for the Back Erase function

SLM Ta bular display


The SLM History display shows the current values for measurements for one meter at a time. You can zoom measurements to change the number of measurements that are displayed. The SPL measurement updates in real time as the sound level changes, whether you are running a study or not. All other measurements update as a result of accumulations and comparisons made during a study. If the study is completed, the display shows the studys final values.

Navigating to the display


In general, you reach the SLM Tabular display after first selecting the SLM display group from the Main Menu (Sound Level Meter (SLM) display on page 25). In the SLM display, press the Right key to go to the SLM Tabular display. An example of the display is shown in Figure 45.

Part number 053-273B

34 SLM DISPLAY GROUP


Parameter Panel

Parameter Panel

Tabulation Area

Figure 48

SLM Tabular display for Meter 1

Parameter Panel
This panel is identical in features and function to the corresponding panel in the SLM History display. See Parameter Panel on page 31 for more information.

Tabulation Area
The Tabulation Area identifies and gives the current values of measurements for the selected meter in accordance with a measurement zoom setting. If viewing while a study is ongoing, the results update as you watch. If viewing for a completed study, the results are the last values recorded for the study. Only results for an ongoing study or for the last completed study can be viewed in this display.

Controlling the display


You can select which meter youre viewing by pressing the Up/Down keys or tapping the corresponding key icons in the Navigation Panel. You can also zoom measurements to change how many are displayed and their sizes.
To zoom measurements:

1. Set the Navigation Panel to its alternate functions (Navigation Panel on page 16). A magnifying glass icon will be associated with the Left/Right keys 2. Press the Left/Right keys to zoom measurements.

Part number 053-273B

SLM DISPLAY GROUP 35


Additional meter settings

Ad di tional meter settings


Some meter parameters can changed on Sound Level Meter Settings (SLM Settings) pages of Instrument Setup. You can change these settings at any time unless a session is open. When a session is open, the Instrument Setup icon on the Main Menu is unavailable.

Working in SLM Settings pages


To display a SLM Settings page, press to display the Main Menu, then tap on the touchscreen to enter Instrument Setup. In Instrument Setup, you may need to navigate to the SLM Settings pages.

Navigating in Instrument Setup


All sections and pages of Instrument Setup have the same features at the top of the display. Figure 49 is example of an SLM Settings page showing the common Instrument Setup features. You will know if youre in the SLM Settings section by the name on the displays Selection Bar and by the position of the bars red pointer. For the SLM Settings section, that pointer should be indicating the icon shown at the right. Navigate in Instrument Setup using any of the techniques below.

Tap an icon in the displays Page Navigator. Tap the Left/Right key icons in the displays Navigation Panel. Press the physical keys represented by the key icons in the displays Navigation Panel. Page Navigator Selection Bar

Variable Page Area

Navigation Panel Figure 49 SLM Settings page in Instrument Setup

Part number 053-273B

36 SLM DISPLAY GROUP


Features of SLM Settings pages

Navigation shortcut
You can jump directly to the RTA Settings page and avoid the Main menu and navigation in Instrument Setup, and it works in reverse too. In the RTA display (set to the standard Navigation Panel functions), tap or press . In the RTA Settings page of Instrument Setup, tap or press . This shortcut icon is shown in Figure 49.

Features of SLM Settings pages


SLM Settings occupies four pages in Instrument Setup. Every page contains Meter radio buttons and paging buttons. The general characteristics of these pages is shown in Figure 410. Work areas, minus the radio buttons, for the pages are shown in Figure 411.

Meter buttons

Page back button

Page forward button

Figure 410

One of the SLM Settings pages

Part number 053-273B

SLM DISPLAY GROUP 37


Making changes in SLM Settings

Page 1

Page 4

Page 2

Page 3 Figure 411 All four SLM Settings pages

Making changes in SLM Settings


Move between the SLM Settings pages by tapping the Page Forward and Page Back buttons. Before making changes in parameter settings, select a meter by tapping its radio button. Most parameters are set when you select from a menu, but Integration Threshold and Exchange Rate are a little different. Also distinctive is the checkbox on page 2 that allows you to add additional study information during a session.

Integration Threshold
You can set an integration threshold but choose not to apply this parameter. Set the threshold by selecting from a menu just like all other SLM Settings parameters. To apply the threshold, select Enable. Clear Enable if you want the meter to inactivate this setting.

LN Source Meter
SoundPro DLX continually updates exceedance measurements at all 100 percentage points during studies, but you can choose four exceedance percentages for online viewing. These values, identified as Ln1, Ln2, Ln3 and Ln4 are derived from SPL measurements made by one of the three meters. Follow the instructions under Selecting settings from menus: on page 39 to set these percentage points. Tap one of the three radio buttons to specify the LN Source Meter for all exceedance measurements.

Part number 053-273B

38 SLM DISPLAY GROUP


Making changes in SLM Settings

Exchange Rate
The Exchange Rate parameter and Integration Threshold settings determine the meaning of the time averaged SPL measurement. For an Exchange Rate of 3 dB and an inactivated Integration Threshold, the measurement is called the equivalent continuous SPL and is written as Leq. If, however, the Exchange Rate is set to some other value or if an Integration Threshold is used, the determination is a different average measurement and is written as Lavg. For information about setting these parameters, see Making changes in SLM Settings on page 37. SoundPro DLX tells you in the displays and in selection menus whether the time averaged SPL measurement is Leq or Lavg according to the Exchange Rate and Integration Threshold settings. If you change either Exchange Rate or Integration Threshold and that change results in a change in the meaning of the averaged SPL measurement, displays and menus are automatically updated with the new notation. Also, an advisory appears to inform you of the change. An example of the Leq/Lavg advisory is shown in Figure 412. In this example, the advisory tells you that the meaning of the time averaged SPL has changed from Leq to Lavg for Meter 1.

Figure 412

Leq/Lavg advisory

NOTE: The Leq/Lavg advisory remains on display until all references to Leq/Lavg for that meter are changed. If you had selected Leq or Lavg in Logging Settings for multiple filters, the advisory takes longer to clear.

Study information
Studies may be run with, or without, adding additional information to them during the session. Additional information includes voice notes and drawings, but you can also change the study name, identify the user and specify the location where the study was conducted. All of this information can be added or changed in the Session Manager after a session is stopped. See Study displays on page 79 for information about modifying studies in the Session Manager. If you would like to add or change study information during the session, select the Show study info screen in Instrument Setup. For additional information about this feature, see Adding information to a study on page 63. That check box is located on Page 2 of the SLM Settings pages.

Part number 053-273B

SLM DISPLAY GROUP 39


Making changes in SLM Settings

Changing settings
The meter parameters that can be changed in SLM Settings are distributed over four pages in Instrument Setup. All of these settings are included in a profile that can be established as explained under Meter Profile Settings on page 89.

Page 1 ~ Criterion Level, Criterion Time and Projected Time. Page 2 ~ Upper Level Limit (ULL) and Peak Count Threshold (PCT). Page 3 ~ Point exceedance levels (L1, L2, L3, L4). Page 4 ~ Integration Threshold, Exchange Rate and Taktm Interval.

Selecting settings from menus:

1. Select the SLM Settings page in Instrument Setup by tapping the Page Back/Page Forward icons. 2. Tap the radio button to select a meter. 3. Tap in the field for the settings you want to change. A cursor appears in the field. 4. Tap, or touch and hold, the scroll buttons at the right of the field to select. 5. Repeat for any other settings you wish to change. NOTE: Projection time only enters into calculations for Projected Dose (Appendix B, Glossary). When setting Projection Time, set hours and minutes separately. Tap to select one or the other, then use the scroll buttons to change the value.

Part number 053-273B

40 SLM DISPLAY GROUP


Making changes in SLM Settings

This page intentionally left blank

Part number 053-273B

CHAPTER

RTA display group

This chapter explains the features and functions of the RTA display group, which consists of three related displays. The primary purpose of these displays is to show measurement results for one measurement at a time for all available filters. For information about setting up for filtered analysis, see Filtered analysis on page 28.

Real Time Anal yzer (RTA) di spl ay


As explained under The SLM and RTA display groups on page 14, you can launch the RTA displayone of the displays of the RTA display groupfrom the Main Menu. Press to open the Main Menu, scroll to the icon shown at the right and press . The RTA display appears (Figure 51). You can also select directly from the Main Menu by tapping the icon. Parameter Panel

Display Range

Analysis Area

Figure 51

Real Time Analysis (RTA) display

Part number 053-273B

42 RTA DISPLAY GROUP


RTA Parameter Panels

RTA Parameter Panels


All three displays in the RTA display group contain the same Parameter Panel at the top (Figure 52). The panel displays current parameter setup for Meter 1. It also identifies and gives information about the RTA Measurement.

Frequency Weighting

Analysis Type indicator

Parameters

Zoom Command

Time Response

Selected Filter Figure 52

RTA Measurement RTA Parameter Panel features

RTA Measurement
Each Parameter Panel in the RTA Group identifies a Selected Filter and provides information about the RTA Measurement. The RTA Measurement is for the selected measurement in the selected filter band. The RTA Measurement is the same measurement selected as Measurement 1 for Meter 1 in the SLM Group displays, but you can also select it in any display of the RTA Group. You can also select the filter and the Analysis Type in any display of the RTA Group.

Changing settings
You can change the settings of any parameter that appears in the Parameter Panel using either the touchscreen or keys. All changes are selected from menus, sometimes after scrolling to view more options (Scrolling on page 13). The changes you make to Meter 1 parameters in RTA displays update in all displays of the RTA Group and the SLM Group.
To change settings using the touchscreen:

1. Tap any parameter. An appropriate menu appears. 2. Tap to select from the menu. The new setting is applied and the menu closes. (You may have to tap scroll icons to reach all options.)
To make changes with keys:

1. If the Navigation Panel is not showing the standard functions, press 3. Press . A menu appears.

2. Press the Up/Down keys to highlight the measurement in the Parameter Panel.

Part number 053-273B

RTA DISPLAY GROUP 43


Analysis Area, RTA display

4. Press the Up/Down keys to select from the menu. If you see dark scroll icons, keep pressing to display more menu options. 5. Press . The new setting is applied and the menu closes.
Alternate way to change the Selected Filter:

1. If the Navigation Panel is not showing the alternate functions, press

2. Press the Left/Right keys. The filter indicated in the Parameter Panel changes. NOTE: You can also tap the icons in the Navigation Panel associated with the Left/Right key icons.

Analysis Area, RTA display


The RTA display contains an Analysis Area that shows both filtered and broadband values for the RTA Measurement. That Analysis Area of the RTA display is shown in Figure 53. Red markers at the top and bottom of the Analysis Area identify the Selected Filter. Selected Filter Meter 1, 2, 3 Broadband values Filter values Y (dB)

X (frequency)

Selected Filter Figure 53 Analysis Area in the RTA display

Part number 053-273B

44 RTA DISPLAY GROUP


Measurement range

Measurement range
You can scale and shift the Display Range by zooming and panning. To make the these functions available, tap Zoom in the Parameter Panel. The zoom and pan controls appear in the Navigation Panel. For information about using them, see Zoom and pan on page 21.

RTA Hi stor y display


During a study, the RTA History display shows the last 15 seconds of results for one measurement plotted by time, frequency and level. Level is color-coded to provide a third dimension in the 2dimensional display. You can track SPL in the RTA History display because SPL represents a current measurement. But until you run a study you cannot track any other measurement in the display because those measurements depend upon what happened before.

Navigating to the display


In general, you reach the RTA History display after first selecting the RTA display group from the Main Menu (The SLM and RTA display groups on page 14). When viewing the RTA display, press the Left key to go to the RTA History display (Figure 54). Parameter Panel Level Legend

Y (seconds) Analysis Area X (frequency) Color coded level

Figure 54

RTA History display

Part number 053-273B

RTA DISPLAY GROUP 45


Analysis Area, RTA History display

Analysis Area, RTA History display


The RTA History display shows a 3-dimensional display of information for the last 15 seconds for the RTA Measurement. Just as in the RTA display, the filter frequency range and three meters are represented along the x-axis. The y-axis shows the progression of time. Values as time progresses in the filter and broadband categories is represented by a change of color, and you can choose the color scheme to represent values over the measurement range. Level Legend

Elapsed Time

Filter markers

Filters Figure 55

Meters 1, 2, 3 Broadband

Analysis Area in the RTA History display

Changing the color scheme


The Level Legend in the RTA History display shows a color scheme that relates a particular set of colors to the Measurement Range. The extents of the Measurement Range are shown at the extremes of the color scheme. You can choose from among several color schemes.
To change the color scheme:

1. Tap the Level Legend to display the options. The current selection has a red box around it. 2. Tap one of the options. The menu closes and new results entering at the bottom of the Analysis Area are colored according to the new scheme.

Part number 053-273B

46 RTA DISPLAY GROUP


RTA Tabular display

RTA Ta bular display


The RTA History display shows the current values in all passbands of the Analysis Type for the RTA Measurement. If monitoring for SPL, the values update as the sound level changes, whether you are running a study or not. For all other measurements, the values update in response to accumulations and comparisons during a study. If the study is completed, the display shows the studys final values.

Navigating to the display


In general, you reach the RTA Tabular display after first selecting the RTA display group from the Main Menu (Real Time Analyzer (RTA) display on page 41). In the RTA display, press the Right key to go to the RTA History display (Figure 56). Parameter Panel

Tabulated values

Figure 56

RTA Tabular display

Tabulated values
Values for each filter are represented at the right of the band frequencies over the frequency range.

Part number 053-273B

RTA DISPLAY GROUP 47


Sound curves

Sound cur ves


A sound curve is a line graph showing the value in each band at the moment it was captured. If you recorded a line curve during a session, the sound curve is saved for later use and immediately displayed in the RTA display. Sound curves are handy to show differences between measurements made at different times or under different conditions. Figure 57 shows a saved sound curve superimposed on an RTA display. With a sound curve present, the display exhibits several types of coloring.

Higher sound curve values ~ If sound curve values exceed current values, the area between the sound curve and current values is shown in white. Lower sound curve values ~ If current values exceed sound curve values, the sound curve values are shown in blue and the excess above sound curves are shown in red.

Sound Curve

Figure 57

RTA display with a sound curve

Working with sound curves


Sound curves are controlled on the Real Time Analyzer Settings (RTA Settings) page of Instrument Setup. If you want to record sound curves, you must enable this feature before you begin recording in a session. Once enabled, you can record as many sound curves as you want in the session. You can also stop sound curve recording during the session. To display the RTA Settings page, press to display the Main Menu, then tap on the touchscreen to enter Instrument Setup. If you need help navigating to the RTA Settings page in Instrument Setup, see Navigating in Instrument Setup on page 35. The navigation icon in Instrument Setup for that page is shown at the right.

Part number 053-273B

48 RTA DISPLAY GROUP


Features of the RTA Settings page

Navigation shortcut
You can jump directly to the RTA Settings page and avoid the Main menu and navigation in Instrument Setup, and it works in reverse too. In the RTA display (set to the standard Navigation Panel functions), tap or press . In the RTA Settings page of Instrument Setup, tap or press . This shortcut icon is shown in Figure 58.

Features of the RTA Settings page


The RTA Settings page is shown in Figure 58. Note that it contains the Enable Sound Curve Capture check box and an Available Sound Curves list of sound curves. Preceding the name of each sound curve in the list is the notation 1/1 or 1/3, which indicate whether the sound curve was recorded during octave or third-octave filtering, respectively.

Figure 58

RTA Settings page in Instrument Setup

Recording sound curves


Use these procedures to add new sound curves to the Available Sound Curves list in the RTA Settings page of Instrument Setup. You can perform these procedures whenever no session is open.
To prepare for sound curve recording:

1. Go to the RTA Settings page. 2. Tap to check Enable Sound Curve Capture. 3. Go to the display you want to use to monitor the study. 4. Start the first study in the session (Running studies on page 58).
To record a sound curve:

Part number 053-273B

RTA DISPLAY GROUP 49


Managing sound curves

1. In the RTA display, press to select the alternate key functions. Note in the Navigation Panel that the camera icon, , is associated with the Execute key. 2. When ready to record a sound curve, press Panel appear. . The Sound Curve Name dialog box and Input

3. Use the Input Panel to modify the name or enter a new name for the sound curve. 4. [Optional] If you are done capturing sound curves during the session, clear Enable Sound Curve Capture in the Sound Curve Name dialog box. The camera icon will no longer be available in m the Navigation Panel. 5. Tap Save. The sound curve is added to the Available Sound Curves list in RTA Settings.

Managing sound curves


Use these procedures to manage existing sound curves. You can perform these procedures whenever no session is open.
To apply a sound curve:

Use this procedure to apply a sound curve in the RTA display. 1. Go to RTA Settings page. 2. In the Available Sound Curves list, tap to select the check box next to the sound curve that you want to view in the RTA display.
To remove an applied sound curve:

Use this procedure to remove the sound curve that appears in the RTA display. 1. Go to RTA Settings page. 2. In the Available Sound Curves list, tap to clear the checkbox next to the applied sound curve.
To delete a sound curve:

Use this procedure to remove a sound curve from the Available Sound Curves list in the RTA Settings page. 1. Go to the RTA Settings page. 2. Tap to select a sound curve in the Available Sound Curves list. 3. Tap Delete. A confirmation dialog box appears. 4. Tap Yes.

Part number 053-273B

50 RTA DISPLAY GROUP


Managing sound curves

This page intentionally left blank

Part number 053-273B

CHAPTER

Logging measurements

This chapter explains how to set the Logging Rate, select measurements to be logged and view logging results in real time.

Log ging Setup


The logging setup functions are provided on the Logging Settings page of Instrument Setup. You can change these settings at any time unless a session is open. When a session is open, the Instrument Setup icon on the Main Menu is unavailable. To display the Logging Settings page, press to display the Main Menu, then tap on the touchscreen to enter Instrument Setup. If you need help navigating to the Logging Settings page in Instrument Setup, see Navigating in Instrument Setup on page 35. The navigation icon in Instrument Setup for that page is shown at the right.

External factors that affect logging settings


In the Logging Settings display, you will have an opportunity to choose measurements to be logged. Those choices will, in some cases, depend upon what happens outside the Logging Settings display.

Before working in Logging settings. Set Analysis Type (Meter 1 only) ~ Set the Analysis Type before choosing logged measurements
in the Logged Settings display (Analysis Type menu on page 29). This is because changes you make to Analysis Type do not automatically update in the Logging Settings display. If, for example, you chose a filtered analysis, then chose measurements to log according to that analysis, then subsequently changed to a different Analysis Type, some of those previous logging selections might no longer be valid. You can always determine which measurements were actually logged by opening the Logged Measurements display in the Session Manager (Logged Measurements on page 80).

Meter Profile ~ A meter profile determines what measurements each of the three meters can make. You must select the meter profile before displaying Logging Setup to be able to select measurements available in that profile. For information about working with meter profiles, see Meter Profile Settings on page 89. Leq/Lavg measurements ~ You cannot log both Leq and Lavg on a meter at the same time, but you can log them simultaneously on different meters. These two measurements are alternatives for each meter based on your Exchange Rate and Integration Threshold settings in SLM Settings (Making changes in SLM Settings on page 37). Whenever you changes those settings, the Logging Settings display updates in accordance with the new settings.

Part number 053-273B

52 LOGGING MEASUREMENTS
Features of the Logging Settings page

Features of the Logging Settings page


The RTA Settings page is shown in Figure 61. Note that it contains meter radio buttons and a Logging List. The Logging List is for the selected meter, which is Meter 1 in this illustration

Meter radio buttons

Logging List

Figure 61

Logging Settings page with the seconds list displayed

Logging rate
The Logging Rate area has hour, minute, second and tenths of a second fields where you can set the logging rate by setting the logging interval (the rate is the inverse of the interval). Set the logging rate by selecting in those fields. The longest logging interval you can select is, essentially, 24 hours.
To set the logging rate:

1. Tap the arrow next to any field in the Log Rate panel. A menu appears. In the Logging Settings page in Figure 61, the seconds list is displayed. 2. Tap to select from the menu. The selection is displayed in the field. 3. Repeat steps 12 until you have the interval you want. NOTE: You may get an alert if you attempt to set the rate too high in relation to the total number of measurements previously selected (see Limits on page 53). When you tap OK to close the alert, your new logging rate will be ignored. Choose a lower logging rate or fewer measurements accordingly.

Part number 053-273B

LOGGING MEASUREMENTS 53
Logging Measurements

Logging Measurements
Each meter has its own Logging List containing selected measurements to be logged. The Meter 1 Logging List can also contain filter specifications if the logging selections were made after a fulloctave or third-octave Analysis Type was previously selected.

Limits
You are not allowed to log a total of more than 104 measurements at a time. Also, you are not allowed to log a total of more than 12 measurements if you have the logging interval set to less than one second. These limits apply to measurements for all three meters combined, not just to measurements for each meter. If you attempt to exceed the limits, you will be alerted when attempting to add measurements to a logging list and your selection(s) will be ignored.

Adding measurements
Although you can log broadband measurements for any of the meters, you can also log filtered measurements for Meter 1 when its Analysis Type is set to either full or third-octave analysis. You can add as many filtered measurements to Meter 1s Logging List as you wish (up to the limits previously stated), including all of them at once with a single convenient selection.
To add a broadband measurement:

This procedure applies to Meter 1 with Analysis Type set to Off, to Meter 2 and to Meter 3. 1. Tap to select a meter radio button. The Logging List for that meter appears below. 2. Tap the button above the Logging List. A menu appears. 3. Tap to select a measurement from the menu. The measurement is added to the Logging List. If adding for Meter 1, the entry in the Filter column is set to OFF. 4. Repeat for other meters or measurements.
To add a filtered measurement:

This procedure applies only to Meter 1 with Analysis Type set to 1/1 or 1/3. 1. Tap to select the Meter 1 radio button. The Meter 1 Logging List appears below. 2. Tap the button above the Logging List. A menu appears. 3. Tap to select a measurement from the menu. A filter list appears. 4. Tap to select a filter, OFF or ALL from the list.

Tap OFF to add the broadband measurement only (top of the list). Tap ALL to add all filtered measurements plus the broadband measurement (end of the list).

NOTE: The list wraps around. Sometimes, for example when the list first appears, you can move quickly to the opposite end of the list by tapping the Up/Down keys or corresponding list scroll arrows.

Part number 053-273B

54 LOGGING MEASUREMENTS
Logging display

5. Repeat for other measurements or meters.

Removing measurements
You can remove one measurement at a time from a meters Logging List or all measurements at once for that meter.
To remove measurement:

1. Tap to select a Meter radio button. The Logging List for that meter appears below. 2. Tap to highlight a measurement in the meters Logging List. 3. Tap the list. button above the Logging List (single dash). The measurement is removed from the

To remove all measurements at once:

1. Tap to select a Meter radio button. The Logging List for that meter appears below. 2. Tap the button above the Logging List (double-dash). A dialog appears. 3. Tap Yes in the dialog. The dialog closes and all measurements for the meter are deleted at once.

Log ging display


Up to four color-coded measurements can be viewed at once in the Log History display. To open the Log History display, press to display the Main Menu, scroll to the icon shown at the right and press .

Selection Panel Selected Measurement

Display Range

Data Display

Zoom Command

Figure 62

Log History display


Part number 053-273B

LOGGING MEASUREMENTS 55
Display characteristics

Display characteristics
At the moment you start a study, scaled data for selected measurements begin appearing at the right in the display. The values move to the left as time advances. When you stop the study, the final curves for those measurements are frozen in place. The X-axis in the Log History display is a count of the most recent logging events and is therefore indexed to time. If, for example, you have the logging rate set at five logging events per second (logging interval = 0.20 seconds), 60 on the x-axis represents a time interval of 12 seconds.

Selection Panel
The Selection Panel identifies which measurements are currently selected for display and shows the colors associated with each measurement curve. Because the measurements can be taken from any of the three meter Logging Lists, the Selection Panel identifies both measurement and the corresponding meter. For Meter 1 selections, in addition, the filter list selection(s) are shown.

Measurement scaling
Each logged measurement is scaled in the display, and you can read its amplitude from the scale at the left edge of the Display Range (Figure 62). This scale defaults to the existing Measurement Scale, but you can change it temporarily with the zoom and pan controls. To make the zoom and controls available, tap Zoom in the Navigation Panel. For information about these controls and how to use them, see Zoom and pan on page 21. You must cancel the zoom and pan controls before you can leave this display.

Selecting measurements to display


You can view a graph for up to four measurements. For each of the four choices, you can select any measurement from a Composite menu that contains all logging selections for all meters. The selections available on the Composite Menu are those measurements you selected for logging on the Logging Settings page of Instrument Setup (Logging Measurements on page 53).

Part number 053-273B

56 LOGGING MEASUREMENTS
Selecting measurements to display

Logging Interval

Color swatches

Composite Menu

Figure 63

Composite Menu in the Log History display

To select a measurement to display:

1. Tap in the Selection Panel where one of the measurements is described. The Composite Menu appears (Figure 63). 2. Tap to select a measurement from the menu. The Selection Panel changes to describe the new selection.
To change a curves color:

1. Tap a color swatch in the Selection Panel. The Color dialog box opens with a box surrounding the currently-selected color. 2. Tap a color. The box moves to that color. 3. Tap to complete the selection. The color swatch and curve color change to match the selection.

Part number 053-273B

CHAPTER

Using the instrument

This chapter provides an overview of the use of the instrument to collect data, from the initial setup of the microphone and hardware to the real time viewing of measurement results. You can also view results after theyre acquired, and for that information, see Chapter 8, Managing session data.

Positioning the instr ument


Tripod mounting
Whenever possible, the meter should be mounted on a tripod in a relatively open area to minimize reflections from your body or nearby large objects. Avoid placing against a wall or in a corner. If holding the instrument in your hand, keep your arm fully extended.

Preamplifier cable
If body reflections are a concern, connect the microphone to the preamplifier, then connect a preamplifier cable from the preamplifier to the instrument. Quest makes preamplifier cables in several lengths.

Microphone orientation
When using a random incidence microphone or pressure microphone, point it approximately 70 degrees away from the sound source. When using a free field microphone, point it directly toward the noise source.

Avoidance of contaminating fields


The instrument The SoundPro DLX is fully shielded and tested to work in fields up to 10 V/m, but contaminating fields can raise the SoundPro DLX noise floor. CAUTION: Handheld transceivers and cell phones in close proximity to the instrument can easily exceed 10 V/m. Working near high voltage lines or other transmitting antennae can lead to higher, erroneous measurements.

Part number 053-273B

58 USING THE INSTRUMENT


Running studies

Running studies
Studies are run within sessions. You can record as many sessions, and as many studies within sessions, as there is room available in data memory. You can repetitively save measurements during studies by logging ones that you choose. You can do post-study verifications during session pauses that relate to the last calibration performed on the instrument.

Sessions
Collected data is organized by sessions. You cannot change measurement parameters during any part of a session. SoundPro DLX does not have a session open when you see on the Status Bar. All accumulators are reset to zero when a session is opened.
To run a session

1. Unplug all external cables and the charger. 2. Point the microphone toward the noise source.

Press

to stop the session. The first study starts.

To stop a session

Press

. If a study is running at that time, it is stopped as well.

Studies
A study is a recording period during a session. SoundPro DLX is running a study when you see the Status Bar.
To run a study

on

1. Press 2. Press

to start the study within a session. If no session is open at the time, the session opens. to stop a running study. A stopped study starts a session pause.

Pauses
A pause is an interval during a session when no study is running (no data being accumulated or recorded). The session is paused when you see on the Status Bar.

Logging
You can capture measurement values along the way in studies by logging. Logging occurs at a Logging Rate that you can set and for measurements you select (Chapter 6, Logging measurements). At each logging event, the values for that measurement are saved, and then the initial conditions for each logged measurement are reset to zero to prepare for measurements in the next logging interval.

Part number 053-273B

USING THE INSTRUMENT 59


Session time line

Session time line


Figure 71 illustrates a measurement session showing one measurement in relation to the session status. Session status is indicated by the status icons that appear on the Status Bar representing the run, pause and session stopped states (Status Bar on page 17). Session Level

Study 1

Study 2

I Logging Interval

Pause 1

Pause 2

Session status Figure 71 Illustration of a typical session

In this example, icons that appear on the status bar during a session are used to indicate session status. Assuming that no session was open when Study 1 starts, pressing starts Study 1 and the session at once. When you press to stop Study 1, the session remains open but is in a pause state. You can continue to press to start and stop studies without stopping the session. During all the pauses, no data is either accumulated or logged, but SPL is continuously measured and displayed by the instrument during those intervals. You can stop the session by pressing during a session pause, as shown here, or you can stop it by pressing to stop the last study and session at once, thereby ignoring the terminating pause.

Calibr ations and veri fications


Use a Quest calibrator to calibrate the instrument outside sessions and to verify a calibration during pause periods within a session (post-study verification). If you dont have a Quest calibrator, see Miscellaneous on page 142 for specifications and ordering information. Perform these calibration and verification procedures in the Calibration Settings page, which is part of Instrument Setup. To display Calibration Settings, press to display the Main Menu, then tap on the touchscreen to enter Instrument Setup. Navigate in the displays Page Manager to the icon shown at the right. If you need help with this navigation, see Navigating in Instrument Setup on page 35.

Part number 053-273B

60 USING THE INSTRUMENT


Calibration Settings page

Calibration Settings page


The Calibration Settings page has data fields where you enter information about the calibrator and can view the instruments calibration history (Figure 72). You can also view the currently measured SPL. CAUTION: You should not attempt to change the setting for 200 V microphone polarization on this page except to match the needs of a different microphone (Microphone polarization on page 105). If you do try, a message will appear asking you to confirm.

Calibrator information Polarization setting Calibration results Calibrate button Measured Value Figure 72 Calibration Settings page

Calibrator information
Quest makes two types of calibration kits: QC-10 and QC-20. The QC-10 has fixed output settings. The QC-20 has optional selections for both output level and frequency. Information on the output settings can be obtained from the calibrator label.

Level ~ The calibrator output level is in decibels. If you have a choice of output level, Quest recommends 114 dB for ordinary calibrations. Frequency ~ If you have a choice of output frequency, Quest recommends 1000 Hz for ordinary calibrations. Calibrator Serial Number ~ Read from the label.

Part number 053-273B

USING THE INSTRUMENT 61


Preparation

Calibration History list


Date/time entries identify both calibrations and post-study verifications.

For each successful calibration, an entry line appears in the Calibration History list that begins with the icon, , and is followed by the date and time of the calibration. For each successful post-study verification, an entry appears in the Calibration History list that begins with the icon, , and is followed by the date and time of the calibration. Each new entry is added at the top of the Calibration History list, and the rest are arranged in descending time order after that.

Double-tap a date/time entry in the Calibration History list to see the data retained for the event. Table 71 explains each of these data items. Table 71: Calibration History data Frequency ~ Value you entered into the Frequency field. Level ~ Value you entered into the Level field. Change ~ Difference in decibels between this calibration, or post-study verification, and the preceding one. Substantial deviations from zero when calibrating the same microphone might mean microphone degradation. Calibrator Serial Number. Polarization ~ Specifies whether the 200 VDC microphone polarization was turned on or off. Reference Range ~ The measurement range chosen by the instrument in response to the calibration signal.

Preparation
This procedure relates to calibration. When performing a post-study verification, you cannot change any of the calibrator settings.
To prepare and assemble:

1. Switch the calibrator to ON. If the LOBAT indicator is on, replace the calibrator batteries before using the calibrator. 2. If the calibrator has more than one output frequency, select 1000 Hz. 3. Press the calibrator adapter over the SoundPro DLX microphone until it fits evenly and snugly against the adapters retaining ridge. If you have more than one microphone, make sure you use the calibration adapter that is properly sized for it. 4. Slowly place the calibrator over the adapter, pressing the two together for an even and snug fit.
Part number 053-273B

62 USING THE INSTRUMENT


Performing a calibration

Performing a calibration
The instrument must be in a session stop state (session stop icon appearing in the Status Bar). Prepare the calibrator before starting the calibration procedure.
Calibration procedure:

1. Verify or set the value in the Level data box so that it agrees with the calibrators output level. 2. Verify or set the value in the Frequency data box so that it agrees with the calibrators output frequency. 3. When ready to calibrate, verify again that the calibrator is emitting sound. If not, switch the calibrator off and on again to restore the output, or replace calibrator batteries. 4. Press the Calibrate button. You will be notified on the touchscreen if the calibration was successful. If successful, the event is entered in the Calibration History list.

Performing a post-study verification


Prepare just as you did for a calibration. You must use the same calibrator, both to ensure consistency and because the calibrator information is taken from the last calibration for the instrument and you will not be allowed to change any of it. Also, the Calibrate button will be relabeled Verify.
Post-study verification procedure:

SoundPro DLX must be in a pause state (icon

showing on the Status Bar).

1. When ready to calibrate, verify that the calibrator is emitting sound. If not, switch the calibrator off and on again to restore the output, or replace calibrator batteries. 2. Press the Verify button. You will be notified on the touchscreen if the post-study verification was successful. If successful, the event is entered in the Calibration History list. 3. Disassemble to prepare for the next study.

Viewing the measured value


You will always be able to view the current SPL measurement from the Measured Value at the bottom of the Calibration Settings page. If the Measured Value exceeds the Range Ceiling, the Range Ceiling is displayed instead of the Measured Value and the up arrow at the right it is darkened. You can use the Measured Value to verify the measurement status during a calibration (or post-study verification) procedure. To do that, first verify that the measurement range includes the calibrators output level (Changing settings in SLM displays on page 27).

Calibrator detached ~ The ambient SPL is displayed as the Measured Value. Calibrator attached, ON, but prior to calibration ~ The instruments current accuracy in relation to an external calibrator is indicated as the difference between the Measured Value and the

Part number 053-273B

USING THE INSTRUMENT 63


Adding information to a study

calibrators output level. For example, if calibrators output is set to 114 dB and the Measured Value is 113.8 dB, the instrument is currently reading 0.2 dB low.

Calibrator attached, ON, after calibration ~ View the result of the calibration or post-study verification. The Measured Value should be within 0.1 dB of the calibrators output level.

Ad di ng infor mation to a study


SoundPro DLX automatically names sessions and studies for you. You can view and change these names in the Session Manager whenever no session is open. Optionally, you can provide identifying information for a study without waiting until the session is stopped. You can name the study, name the user, identify a location and provide comments. You can add a drawing, a voice note, or both, to the study at that time as well (Class/Type 1 models only). To add this information to a study during a session, select the Show study info screen check box in Instrument Setup before starting the session. See Making changes in SLM Settings on page 37.

Pointing the Page Navigator


If you selected Show study info screen in Instrument Setup, the Study Information display appears when you stop a study (Figure 73). At the top of this display is a Page Navigator which allows you to move to related displays in this display group. Select any page of this group by pointing the Page Navigator to it. Information about selecting pages in display that contains a Page Navigator is available at Setup displays on page 19.

Page Navigator

Figure 73

Study Information display

Part number 053-273B

64 USING THE INSTRUMENT


Adding text to study information

Adding text to study information


After selecting the Study Information display in Instrument Setup, it appears the next time you press pause to stop the study. The display opens with the Page Navigator pointing to the icon at the right and the Text Information page displayed (Figure 73). This page is one of three work areas available in the Study Information display. The other two pages are for drawings and voice notes (Class/Type 1 models only). For information about changing pages, see Setup displays on page 19.
To add text information:

1. You can enter characters using the Input Panel. Alternatively for the User Name and Location fields, you can select from among previous entries. 2. Tap the Input Panel icon in the lower right corner. The Input Panel appears. For information about using it, see Input Panel on page 20. 3. Tap in a field to enter characters. If applicable, tap the arrow at the right of the box and select from an existing entry. 4. Repeat step 2 for other fields, as desired. If entering in the Comments box, you can scroll through a long entry by tapping the Scroll Arrows at the edge of the box. 5. When finished, tap the Input Panel Icon to remove it from the display. 6. If you wish, you can also add a voice note or drawing to the study before leaving the Study Information display. See those sections in this chapter. Otherwise, finish as explained under Exiting the display on page 67.

Drawing
NOTE: This feature available only in Class/Type 1 models. Point the Page Navigator to the icon shown at the right to open the Drawing Pad (Figure 74).

Part number 053-273B

USING THE INSTRUMENT 65


Voice note

Line Width

Line Color

Figure 74
To create a drawing:

Drawing Pad

1. Create a drawing by simply drawing on the touchscreen with the stylus. To control your drawing, tap either of the options below before drawing.

Line Width ~ Tap one of the Line Width selections. Line color ~ Tap the Line Color icon to display the Color dialog box. See Changing colors on page 23 for information about working in the Color dialog box.

2. If you change your mind, tap Clear to remove the drawing completely. 3. If you wish, you can also add text information or a voice note to the study before leaving the Study Information display. See those sections in this chapter. Otherwise, finish as explained under Exiting the display on page 67.

Voice note
NOTE: This feature available only in Class/Type 1 models. You can add a 10-second sound recording, called a voice note, to the study. When preparing a voice note, you can play and re-record the voice note before accepting it. You can record voice notes either through the measurement microphone or through a headset microphone. For information about connecting a headset microphone, see Headset port on page 5.
To prepare a voice note:

1. Select the icon shown at the right in the Page Navigator. The Voice Note Recorder appears (Figure 74).
Part number 053-273B

66 USING THE INSTRUMENT


Voice note

Stop

Play

Start

Recording Indicator

Figure 75

Voice Note Recorder

2. Tap the Start button to begin the recording. 3. Talk into the microphone. During recording, the start button is dimmed and a bar moves across the Recording Indicator, second by second, to indicate elapsed time.

Tap the Stop button if you want to record for less than ten seconds. Tap the Play button to play back the recording. This button is dimmed except when a recording exists for the study. If you dont hear the voice note, check the speaker volume setting in Global Setup (Battery, Sound, Display Settings on page 96). Tap the Start button again to re-record over the previous recording.

4. If you wish, you can also add text information or a drawing to the study before leaving the Study Information display. See those sections in this chapter. Otherwise, finish as explained under Exiting the display, below.

Part number 053-273B

USING THE INSTRUMENT 67


Exiting the display

Exiting the display


You can escape this display without saving or save first, then return to the former display. Before exiting, you can prevent the Study Information display from appearing after the next study. NOTE: Exiting with saving study information does not affect the measurement results, which are always saved.

Disabling the display


You can disable this display by clearing the checkbox in Instrument Setup (Making changes in SLM Settings on page 37). But you can also stop its reappearance in a session by clearing the Show after each study check box in the Study Information display. This also clears the corresponding checkbox in Instrument Setup.

Cancel or save
When you exit by saving, all of the information is saved with the study. If you entered a user name and location, those entries are added to the respective selection lists. When no session is open, you can edit saved study information in the Session Manager (Study displays on page 79). NOTE: If you change your mind during a cancel or save operation, point the Page Navigator to a different selection.
To exit without saving:

1. Tap the X icon (shown at the right) in the displays Navigation Panel. button appears in the middle of the display. 2. Tap Yes. Your display at the time you stopped the study reappears.
To save, then exit:

1. Tap the disk icon (shown at the right) in the displays Navigation Panel. A button appears in the middle of the display. 2. Tap Yes. Your display at the time you stopped the study reappears.

Pr eset trig ger s


You can also set triggers to control study runs or set output logic levels. You can trigger in relation to SPL or timed settings in a single session. You can also combine triggered control with manual control in the session. Timed studies can be run from instrument hibernation.

Part number 053-273B

68 USING THE INSTRUMENT


Trigger Manager

Trigger Manager
Configure and enable triggering in the Trigger Manager, which is part of Instrument Setup. To display the Trigger Manager, press to display the Main Menu, then tap on the touchscreen to enter Instrument Setup. Navigate in the Page Manager to the icon shown at the right. If you need help with this navigation, see Navigating in Instrument Setup on page 35. The Trigger Manager page is shown in Figure 76.

Page Navigator

Threshold setup area

Figure 76

Trigger Manager in threshold mode

Setting triggers
Click one of the radio buttons at the top of the Trigger Manager to select and configure either threshold or timed triggers. The threshold and timed triggers operate separately and require separate settings.

Threshold trigger mode


When you tap the threshold radio button, the threshold setup area shown in Figure 76 appears. This setup area provides for independent settings for two triggers. You can use triggers to start and stop studies in relation to measurements within a specified range. That range is determined by the two level settings for triggers 1 and 2. Both levels are referenced to the broadband sound pressure level measured by Meter 1.

Part number 053-273B

USING THE INSTRUMENT 69


Setting triggers

To configure each threshold trigger, tap the arrows at the edge of the selection boxes to select when a trigger event can occur and what action it causes. Other conditions must also be met for that action to take place, as explained under Putting triggered actions to work, below.

Condition ~ Select whether the SPL must be greater than (>) or less than (<) the threshold level set for this trigger. Action ~ Two action types are available: controlling a study or setting TTL output levels. To control a study, select a pause, start or stop action. If the control action is impossible, for example, trying to start a study when one is already running, the trigger event is ignored. There are two TTL channels available at the General Purpose Accessory connector, and the triggers can be used to set either of them high or low. Level ~ Select a value for the threshold level. You can choose any value in the range 0200 dB in 0.1 dB steps.

Timed trigger mode


The timed trigger starts and pauses studies repeatedly in the session according to date, time and repetition settings. For each setting, tap the arrow in the setting field to display options. For information about selecting dates and times, see Date and Time Settings on page 93.

Date ~ Select the date for the first timed study to occur. Tap the down arrow in the data box to display a calendar. In addition to selecting from the calendar, you can select the instruments current date by tapping the Today: line at the bottom of the calendar. Time ~ Select the time for the first timed study to occur. Tap the current setting for hours, minutes or seconds, then tap the up or down arrows to change (24-hour clock). For all but Occurs = Hourly, the time will be the same for every occurrence. Duration ~ Tap the current setting for hours, minutes or seconds, then tap the up or down arrows to change (Range is 0:0:0 up to 24:59:59). Occurs (Repetition Rate) ~ Tap the down arrows to display a list, then tap to select from the list. If you choose Weekly, the study will run on the same day each subsequent week. If you choose Weekdays, checkboxes for the days of the week will appear. Select those days of the week when you want studies to run.

Notes on timed trigger setup Starting times ~ For all settings of Occurs except Hourly, all timed studies will start at the same
time on subsequent days. For Occurs = Hourly, all timed studies will be started exactly one or more hours after the first study was started.

Long Duration ~ For Occurs = Hourly, setting Duration to longer than an hour will overlap one or more subsequent runs and cause them to be skipped.

Part number 053-273B

70 USING THE INSTRUMENT


Putting triggered actions to work

Putting triggered actions to work


You can do a trigger setup, but triggered control cant occur unless several conditions are met.

The trigger must be enabled. The instrument must be in a measurement status (threshold triggering only). Trigger conditions must be valid. The trigger action must be possible.

Enabling triggers
Enable threshold and timed triggers by selecting the Enable check boxes in the Trigger Manager. You must enable threshold and timed triggers separately. Threshold and timed triggers can be enabled at the same time if you wish. When you enable threshold triggering, both threshold triggers are enabled at the same time.

Disabling triggers
There are three ways to disable triggers.

Enable check box ~ Clear to disable threshold triggers and the time trigger separately. Triggered Stop (threshold trigger only) ~ If a threshold trigger requires a stop action and is valid, the study and session not only end but all triggering is disabled as well. Manual Stop ~ If you manually stop a session by pressing disable triggering at the same time. while triggering is enabled, you

Measurement status
Threshold triggering requires an SPL measurement, which only occurs when the instrument is processing a sound signal. Its not necessary for the instrument to be running a study or in a session for SPL measurement to occur. Timed triggering does not require the instrument to be actively measuring a sound signal. The instrument could be in a hibernation state when the timed trigger even occurs and still respond properly. If in hibernation, the instrument starts as usual, runs a study, and then shuts back down to a hibernation state when the triggered study ends.

Disabling triggers
You can disable threshold or timed triggers by clearing the Enable check box in the Trigger Manager. Clearing one type has no effect on the enable status of the other type, but disabling threshold triggers disables both threshold triggers at once.

Valid conditions
Triggering cant occur unless the conditions you established in Trigger Manager are valid. That means, for threshold triggering, that the action must be possible. For Timed Triggering, that means that no study is already underway when the time trigger occurs.
Part number 053-273B

USING THE INSTRUMENT 71


Viewing current session results

Trigger mode established


Even when triggering is enabled, it cant occur until the instrument is armed. Arming capability is available in any of display of the SPL or RTA display groups. To arm triggers after leaving the Trigger Manager, select the display and press . Arming wont start a study but will prepare the instrument to respond to the next trigger event. When either trigger is armed, the Trigger Icon shown at the right appears in the Status Bar. The icon blinks when a trigger is enabled but the instrument is unarmed. The Trigger Icon is steady when the instrument is armed. You can run start and pause studies in the usual way when the instrument is armed, but you wont be able to change settings in Instrument Setup or Global Setup. Nor will you be able to view sessions in Session Manager. You end arming by pressing .

Triggering interactions
Triggers can effect your studies as you run them manually. If a triggered pause or stop occurs while you are manually running a study, SoundPro DLX will comply as if you did it manually. Likewise, you can manually override a study that was triggered to start.

V i ewing cur r ent session r esults


You can view measurement results for the current session in the Summary display. You can select the Summary display at any time. To select the Summary display, press to display the Main Menu, scroll to the Summary display icon (shown at right), press .

Summary display
The Summary display tabulates accumulated values during a session for a selected meter. If a study is running while you view this display, the values update every 10 seconds. To view results for a different meter, press the Up/Down keys. Figure 77 shows measurement results for Meter 1.

Part number 053-273B

72 USING THE INSTRUMENT


Summary display

Meter Number

Figure 77

Summary display

Part number 053-273B

CHAPTER

Managing session data

This chapter explains the features and functions of the Session Manager. You can view and print information from any session when no session is open. You can add information such as drawings and voice notes and edit information added previously (Adding information to a study on page 63). Use the Summary display to view accumulating measurement results during a session (Viewing current session results on page 71). To select the Session Manager display, press the icon shown at the right and press . to display the Main Menu, scroll to

Session Mana ger display


Information is presented in the Session Manager in a tree structure of folders, books, lists, displays and fields. The top level structure is a list of named folders, each of which represents a session in data memory. If you double-tap a folder name, it expands to reveal a list of named books, each of which represents a study in the session. An expanded session or an expanded study also reveals the lists, displays and fields in that session or study. Figure 81 shows the Session Manager with an expanded study within an expanded session. Double-tap an expanded session or study to collapse it again. NOTE: Vertical or horizontal scroll bars appear in the Session Manager when there is too much information in that direction to show at one time in the display.

Expanded session Expanded study

Manager Functions Bar Figure 81


Part number 053-273B

Expanded session with one of its studies also expanded

74 MANAGING SESSION DATA


Block diagram

Block diagram
Not all sessions and studies contain the same types of information. For example, the Logged Measurements and Calibration elements shown for the expanded studies in Figure 82 are options under the control of the person conducting the study. In addition, voice notes and drawings are optional elements that you may wish to add to a study. The block diagram in Figure 82 indicates the required and optional elements in the Session Manager. If an element is optional, it is shown grayed in this illustration. Session1 Session General Session Settings Session Statistics Session4 Session Exceedance Session Summary ... Figure 82 Study1 Study2 Study3 .. .

Session2

Session3

Study General Statistics Exceedance Logged Measurements Study Summary Calibration Drawing Pad Voice Note Session Manager showing required and optional components

Common featur es and functions


Displays in the Session Manager share many of the following features and functions.

Manager Functions Bar


A Manager Functions Bar lies along the bottom of every Session Manager display. The icons on this bar represent functions that may be available depending upon the context (Table 81). When a function is not available in the current Session Manager display, its icon on the Manager Functions bar is dimmed.

Part number 053-273B

MANAGING SESSION DATA 75


Scroll bars

NOTE: The voice note and drawing functions are available only in Class/Type 1 models. Table 81 Manager Function icons Icon Manager Function Add or edit text. Delete selected item. Print session results and information. Collapse everything in the display. Return to previous display. Add or edit voice note. Add or edit drawing. Delete all sessions. Export a session.

Scroll bars
When there is too much information to be viewed at once in the display, one or more scroll bars appear. A vertical scroll bar appears when the number of sessions or expansions of sessions or books is too large for the display; a horizontal scroll bar appears when one or more data fields are too long for the display.

Reorganizing columns
Many Session Manager displays contain columns of information. Examples are the Settings display and measurement results in the Session Summary and Study Summary displays. You can sort and size all of them similarly.

Sorting ~ Sort on a column by tapping the column heading. Sort in reverse order by tapping again. Column width ~ Touch and drag a column divider in the heading to resize the column. If you make the total column width too wide for the display, a horizontal scroll bar appears.

Part number 053-273B

76 MANAGING SESSION DATA


Renaming, entering text and editing

Renaming, entering text and editing


You have many opportunities to add or change text information inn the Session Manager. For example, you can change the names of the sessions and the studies. You can also add or change text information in the General List of sessions and studies. For all of these, the manipulation of text follows the same general procedure.
To edit a text field in Session Manager:

1. Display the text field. 2. Tap in the text field. A cursor appears. 3. Tap on the Manager Functions Bar. The Input Panel appears (Input Panel on page 20). 4. Edit using the Input Panel (Input Panel on page 20). 5. Tap the Input Panels return key. 6. If finished editing text fields, tap the Input Panel icon to close the Input Panel.

Session folder s, displays and fields


Each session folder contains three displays: the General List, the Settings display and the Session Summary display.

Session name
The name of a session is automatically generated in the form, SessionX with sequential numbers for X, but you can rename it in the Session Manager if you wish (Renaming, entering text and editing, above).

Sessions General List


Figure 83 shows an example of a Sessions General List. This list always contains three fields.

Figure 83

Contents of the Sessions General List

Comments field ~ The text in this field defaults to Comments. If you want to edit it in the Session Manager, see Renaming, entering text and editing, above. Start field ~ Specifies the sessions starting date and time (same as the date and time that the sessions first study started). Stop field ~ Specifies the sessions ending date and time (same as the date and time that the sessions last study ended).
Part number 053-273B

MANAGING SESSION DATA 77


Settings display

Settings display
This display contains parameter settings for all meters. Double-tap Settings or its associated icon to show the Settings display (Figure 84). To return to the full Session Manager display, tap .

Column Headings

Opened session folder Settings display Figure 84 Opening the Meter Settings display

Statistics
You can view a graph of the percentage distribution of sound during the session, called a Statistics display. For a definition of statistics, see Appendix B, Glossary. Statistics displays are available for every session and for every study in a session. An example of a Statistics display is shown in Figure 85.

Opened session folder Statistics display Figure 85 A sessions Statistics display

Part number 053-273B

78 MANAGING SESSION DATA


Exceedance

Exceedance
You can view a graph of exceedances over the entire percentage range, called an Exceedance display. For a definition of Exceedance, see Appendix B, Glossary. An Exceedance graph decreases from left to right as increasing percentages of sound are included for lower levels of exceedance. This display is available for every session and for every study in a session. An example of an Exceedance display is shown in Figure 86.

Opened session folder Exceedance display Figure 86 A sessions Exceedance display

Sessions Summary display


The Sessions Summary display contains a list of measurements performed in the session. Double-tap a measurement to see the results for that measurement in a Summary display. If octave or third-octave analysis was performed during that session, results for all the filters are given. Figure 87 illustrates the two-step process you use to get to a Summary display from an open Session folder. To back out of a Summary display to the, tap Manager, tap again. . To return to the Measurement List in the Session

Part number 053-273B

MANAGING SESSION DATA 79


Study displays

Opened session folder

Measurement List TWA results Figure 87 Session Summary display for TWA measurement

Study displ ays


Each study book contains two required displays: the General display and the Summary display. Optionally, it also can contain Logging Measurements, Calibration, voice note or drawing displays.

Naming studies
The name of a study is automatically generated in the form, StudyX with sequential numbers for X, but you can rename it in the Session Manager if you wish (Renaming, entering text and editing, above).

Studys General List


Expand to see a Studys General List by tapping the expansion icon in front it. This General List always contains five fields. Figure 88 illustrates how to get to a Studys General List from the Study book.

Part number 053-273B

80 MANAGING SESSION DATA


Logged Measurements

Figure 88

Expansion of a Studys General List

Fields in the display Comments ~ The text in this field defaults to Comments. If you want to edit it in the Session
Manager, see Renaming, entering text and editing on page 76 above.

User Name ~ The text in this field defaults to User Name. If you want to edit it in the Session Manager, see Renaming, entering text and editing on page 76 above. Location ~ The text in this field defaults to Location. If you want to edit it in the Session Manager, see Renaming, entering text and editing on page 76 above. Start ~ Specifies the studys starting date and time. Stop ~ Specifies the studys ending date and time.

Logged Measurements
If you had selected measurements for logging before starting the session, a Logged Measurements link exists for the study (Chapter 6, Logging measurements). Double-tap the Logged Measurements link to open the Logged Measurements display (Figure 89). To return to the full Session Manager display, tap .

Part number 053-273B

MANAGING SESSION DATA 81


Study statistics and exceedance

Logged Measurements display Figure 89 Opening the Logged Measurements display

Study statistics and exceedance


Statistics and Exceedance displays are available for studies, just as they are for sessions. See the corresponding sections under Session folders, displays and fields on page 76for additional information.

Study Summary
The Study Summary display is analogous to the Session Summary display, except that it contains the results for one study. See Statistics on page 77 for information about this display.

Calibration
The Calibration display in the Session Manager shows calibration data and post-study verification data, if any, for the study. The calibration information is for the last time a calibration was done prior to the session in which the study was conducted. The post-study verification, if done for the study, is shown after the calibration data. To return to the full Session Manager display, tap .

Part number 053-273B

82 MANAGING SESSION DATA


Voice notes

Calibration data

Location of Post-study verification data

Calibration display Figure 810 Calibration display

Voice notes
If you had selected Show study info screen and recorded a voice note during the session, you can replay and edit it in the Study Manager. You can also rename the previously recorded voice note, and you can add additional voice notes in the Study Manager if you wish.
To replay, edit or add a voice note:

1. Double-tap the voice note. The Voice Note Recorder appears.

Replay ~ The voice note plays automatically. Edit or add ~ Follow the instructions given under Voice note on page 65.

2. Repeat step 1 as needed to continue replaying, editing or adding. 3. When finished, tap in the Manager Functions Bar to leave the Voice Note Recorder and return to the Session Manager.
To rename an existing voice note:

1. Tap the voice note to highlight it. 2. Proceed as explained under Renaming, entering text and editing on page 76.

Drawings
If you had selected Show study info screen and entered a drawing into the session, you can view and edit it in the Study Manager. You can also rename the previously recorded drawing, and you can add additional drawings in the Study Manager if you wish.
To view, edit or add a drawing:

1. Double-tap the drawing. The Drawing Pad appears.

Review ~ The Drawing Pad contains the drawing.

Part number 053-273B

MANAGING SESSION DATA 83


Deleting in the Session Manager

Edit or add ~ Follow the instructions given under Drawing on page 64. in the Manager Functions Bar to leave the Drawing Pad and return to the

2. Repeat step 1 as needed to continue editing or adding. 3. When finished, tap Session Manager.

To rename an existing drawing:

1. Tap the drawing to highlight it. 2. Proceed as explained under Renaming, entering text and editing on page 76.

Deleting i n the Session Mana ger


You can delete individual sessions, studies, voice notes and drawings in the Session Manager. You can delete all sessions at once by tapping on the Manager Functions Bar. Before responding to your request to delete all sessions, a window appears asking you to confirm your intention.
To delete an item:

1. Make the item visible, if necessary, by expanding in the sessions and studies. 2. Tap to highlight the item. 3. Tap on the Manager Functions Bar. A dialog box opens. 4. Tap Yes. For large items, a progress bar appears while the deletion is occurring.

Expor ting sessions


You can export sessions, one at a time, to a CompactFlash card in the Session Manager.
To export a session:

1. Make sure you have a CF card in the slot and that it has sufficient memory. 2. Collapse the display or scroll, if necessary, so that you can see the name of the session you wish to export. 3. Tap the session name to highlight it. 4. Tap on the Manager Functions Bar. A confirmation dialog box appears. 5. Tap Yes to export the session.

Part number 053-273B

84 MANAGING SESSION DATA


Printing

Printing
You can print selected results for sessions and studies from the Session Manager. Only printers that operate via infrared communication are supported. See Infrared on page 136 for the printer specification. NOTE: You can also print sessions and studies that are transferred to QuestSuite Professional II. For information about using that software application, see QuestSuite Professional II application on page 108.

Print display
Use the Print display in Session Manager to design and print a report (Figure 811).

To print from a session with or without also printing from studies in the session, tap the session name in the Session Manager. The Print display appears with the name of the session and each study in the session listed in the Select Items to Print list. Alternatively, if you only want to print a study, tap the name of the study in the Session Manager and proceed directly to setting up the report.

Figure 811 Print display

Contents of the report


For each session and study selected, the name of the item and general information heads the report for that item. The general information for an item is the same information that you would see in the General display for that item in the Session Manager. For studies that have associated calibrations, the calibration information is printed next. Any other information printed for the item depends upon your selections in the Print Options section of the Print display. See Appendix A, Printout, for an example of a printed report for a study.
Part number 053-273B

MANAGING SESSION DATA 85


Print display

There are six categories of print options distributed across six, named tabs in the Print display. Tap one of the tabs to make it active. The active tab stands out in front of all others. You can cycle through all tabs by tapping the arrows next to the tabs, but that doesnt change which tab is currently active.

Summary ~ Select Enable on the Summary tab to include all broadband measurements made by all three meters for the selected item. If filtered measurements are available for Meter 1, you can also select Filter Values or Filter Charts to include these measurements in either a tabulated or graphed format, or both formats. Clear Enable to exclude this summary information from the report. Statistics ~ Select Enable on the Statistics tab and also select Charts, Values or both of them to add statistics information to the report. Clear Enable to exclude the statistics information from the report. Statistics information is derived from the LN Source Meter (LN Source Meter on page 37). Exceedance ~ Select Enable on the Statistics tab and also select Charts, Values or both of them to add this information to the report. Clear Enable to exclude the exceedance information from the report. Statistics information is derived from the LN Source Meter (LN Source Meter on page 37).

To set up a report:

1. In Select Items to Print, tap to select what you want in the report. You can make more than one selection. 2. In Print Options, tap a tab. Tap the arrows next to the tabs to move other tabs into view. 3. If you want to include that type of information in the report, select Enable. Clear the Enable checkbox is you dont want that in the report. 4. If the page provides additional options, select or clear to control how the data is presented. 5. Repeat steps 2-4 above for other tabs, as appropriate. 6. If you want to edit the title, tap in the Report Title field and edit as explained under Renaming, entering text and editing on page 76.

Generating the report


SoundPro DLX is compatible with printers that accept infrared transmission. Set up the printer and have it ready to receive an infrared transmission as explained in the printers Owners Manual.
To generate a report:

1. Physically position SoundPro DLX so that its IR port faces the printer IR port and is no more than one meter away. 2. In the Print display, tap Print. The button changes immediately to Cancel. Many printers indicate when they are receiving and processing data.


Part number 053-273B

If you want to stop printing before the job is completed, press Cancel. The report is finished when the Cancel button changes back to Print.

86 MANAGING SESSION DATA


Print display

This page intentionally left blank

Part number 053-273B

CHAPTER

Global setup

The purpose of this chapter is to explain how to configure instrument parameters that dont directly relate to the meters or the measurements. In many cases, you only need to make these settings once, or very infrequently, as your circumstances or needs change.

T he Global Setup display


The Global Setup display contains many pages for setup functions. You can change global settings at any time unless a session is open. When a session is open, the Global Setup icon on the Main Menu is unavailable.

Common features
Each page of the Global Setup display contains identical navigation and calibration features.

Navigation
The navigation features, located at the top and bottom of each page, enable you to vary the information that appears in the page area of the display. When you select a page, the name of the page appears in the displays Selection Bar and a pointer on the Selection Bar moves to the icon for that page in the Page Navigator (see an example in Figure 91). You can select a page in any of several ways.

Tap an icon in the displays Page Navigator. Tap the Left/Right key icons in the displays Navigation Panel. Press the physical keys represented by the key icons in the displays Navigation Panel. Page Navigator Selection Bar

Variable Page Area

Navigation Panel Figure 91 Features in the Global Setup display

Part number 053-273B

88 GLOBAL SETUP
Common features

Touchscreen calibration
All Global Setup displays allow you to calibrate the touchscreen if you suspect that is necessary. A calibration problem might exist if you tap the touchscreen to activate a control or change data, and the action doesnt occur as expected. You can calibrate the touchscreen at any time.
To calibrate the touchscreen:

1. Tap in the Navigation Panel or press display in Figure 92).

. A Touchscreen Calibration display appears (left

NOTE: If touchscreen calibration is so far out of alignment that you cannot operate it by tapping, follow the instructions under Using only the keys, below. 2. Follow the instructions to tap on the targets in the display. 3. If your entries are acceptable, a Calibration Confirmation display appears (right display in Figure 92). 4. Follow the instructions in the second calibration display. The Calibration Confirmation display closes and youre back at the Battery, Sound, Display Settings page.

Using only the keys


You can also get to the calibration display without using the touchscreen itself.
To calibrate using only keys:

1. Display a Global Setup page by pressing and the Up/Down keys, as explained under The Global Setup display on page 87. Global Setup opens on the page you used last. 2. Press . The Touchscreen Calibration display shown at the left in Figure 92 appears. 3. Follow steps 24 of the previous procedure.

Touchscreen Calibration Figure 92

Calibration Confirmation

Touchscreen calibration displays

Part number 053-273B

GLOBAL SETUP 89
Meter Profile Settings

Meter Pr ofile Settings


Manage meter profiles in the Meter Profile Settings page of Global Setup. In Global Setup, select the page represented by the icon shown at the left.

What is a meter profile?


A SoundPro DLX meter profile completely specifies the setup of the instrument to perform measurements, and controls some other features as well. All of the following are characteristics of a meter profile.

Profile name ~ All stored profiles have names, but profiles that are factory-supplied have reserved names that you may not use for your own profiles. Profile description ~ A stored profile can have a description. To see the profiles description, double-tap the profiles name in the Available Profiles list. Enabled measurements ~ You can enable measurements in the Meter Profile Settings page that affects the choices available on measurement lists in other displays, including the SLM displays, the RTA displays and the Logging Settings page. NOTE: Enabling measurements in a profile only restricts which measurements can be selected for viewing in the SLM and RTA displays. Those not enabled are still accumulated and included in the study results. The affect on logged measurements is different. Only measurements enabled in a profile will be available for logging.

SLM settings ~ All settings in the SLM display, both parameter settings and viewing settings (see Sound Level Meter (SLM) display on page 25). SLM settings ~ All settings in the SLM Settings pages of Instrument Setup except the Show study info check box setting (see Additional meter settings on page 35). Logging settings ~ All settings in the Logging Settings pages of Instrument Setup (see Logging Setup on page 51). parameter settings in the SLM Settings pages of Instrument setup (except the Show Study check box).

Loading a stored profile


You can load a stored profile to make it the instruments active profile. When a profile is loaded, its name appears in the Profile field of the Meter Profiles Settings page.
To load a profile:

1. Tap to highlight the profile name in the Available Profiles field. You may need to scroll the vertical scroll bar to see the profile. 2. Tap . When loading is complete, its name appears in the Profile field.

Part number 053-273B

90 GLOBAL SETUP
Creating or modifying a stored profile

NOTE: You may see a notice appear during loading about Leq and Lavg. The notice announces a change in naming resulting from the loaded profile. For an explanation, see Making changes in SLM Settings on page 37.

Creating or modifying a stored profile


You can create an active profile from a manual setup or by loading one of the stored meter profiles. If you want to store the result as a new profile, save to any but an existing profile name. Except for factory-supplied profiles, you can use this procedure to modify a stored profile by saving to the same name. You are not allowed to save to any of the factory-supplied names, which are reserved.
To create/modify a profile:

In the following procedure, you will create a new profile if it has a unique name. You will overwrite an existing profile if you save with the same name as an existing profile. You are not allowed to save to the name of a factory-supplied profile. 1. Create an active profile with a manual setup or by loading one of the existing profiles. 2. If you wish, edit the active profile. 3. Tap at the left of the Available Profiles list. The Save Profile dialog box appears with a blinking cursor in the Profile Name field. The Input Panel also appears (Figure 93). 4. Type the name on the Input Panel. Alternatively, tap the arrow at the edge of the Profile Name field to display a list of previously entered names, then tap to select from the list.

Input Panel

Figure 93

Save Profile dialog box and Input Panel

5. If you want to enter a description, tap in the Description field.


Part number 053-273B

GLOBAL SETUP 91
Editing stored profiles

6. Type the description on the Input Panel. 7. When finished entering text, tap Save. If you created a new profile, its name is added to the Available Profile list and entered in the Profile field. It is now the active profile.

Editing stored profiles


You can load any profile, edit it and save the edited version is you wish. If you load and edit one of the reserved profiles (factory default), you cannot save it under the reserved name but you can use it that way. But you can save an edited factory default under a different name as a custom profile. Edit Enabled Measurements in the profile separately from editing other settings.
To change the profiles name or description:

1. Tap to highlight the name or description in the Available Profiles list. To see the description, double-tap the name. 2. Tap the button at the left edge of the Available Profiles list. The Input Panel appears. 3. Type on the Input Panel to add/edit the text. 4. Tap the Return key on the Input Panel. 5. Tap the Input Panel icon. The Input Panel disappears.
To edit the enables measurements:

1. Load the profile, as explained under Loading a stored profile on page 89. 2. Select or clear checkboxes in Enabled Measurements. Only selected measurements will appear on the measurement lists. 3. If you want to start over, tap located at the left of the Enabled Measurements list. All checkboxes will be reset to whatever state they had in the loaded profile. 4. When satisfied with your settings, tap loaded profile.
To edit all other settings:

. Your enabled measurements selections are saved to the

This procedure does not change the Enabled Measurements settings in the profile. 1. Load the profile, as explained under Loading a stored profile on page 89. 2. Manually change settings in other displays. NOTE: When you change measurement parameters, the name in the Profile field in Meter Profile Settings changes to red. 3. When finished, tap located at the left of the Available Profiles list.

Part number 053-273B

92 GLOBAL SETUP
Locked profiles

Locked profiles
You can prevent any changes to profiles with a locking option on the Meter Profile Settings page. Select Lock Profile to set the lock. Clear Lock Profile to restore normal access. Three conditions exist for locked profiles.

Users see (Prohibited icon) instead of a menu when they attempt to change measurement settings in the SLM or RTA displays. The Instrument Settings selection in the Main Menu is unavailable. Users cannot change anything in the Meter Profile Settings page until Lock Profile is cleared.

Deleting a profile
You can delete any user profile provided that its not the current profile. You are not allowed to delete factory default profiles.
To delete a profile:

1. Tap to highlight the profile name in the Available Profiles list. You may need to scroll the vertical scroll bar to see the profile. 2. Tap . A Delete dialog box appears. 3. Tap Yes.

Part number 053-273B

GLOBAL SETUP 93
Date and Time Settings

Date and Time Settings


Use the Date and Time Settings page to change clock and calendar settings. In Global Setup, select the page represented by the icon shown at the left. An example of the Data and Time Settings page is shown in Figure 94.

New day setting

Current day setting

New time setting

Current time setting

Figure 94

Date and Time Settings page

Changing settings
You can make all changes directly in the display. When you are done making changes, tap display to save all changes and update the calendar and clock settings to new settings.
To change the year:

in the

1. Tap the year in the calendar title bar. A Year box with Scroll Arrows opens. 2. Tap the Scroll Arrows to select the year (Scrolling on page 13). 3. Tap anywhere else in the display to close the Year box.
To change the month:

1. Tap the month in the calendar title bar. A Month menu appears. 2. Tap to select from the menu. The menu closes and the new month is substituted. Alternatively, you can tap the left and right Scroll Arrows in the Calendar Title Bar to step sequentially through the months (and years).
To change the day:

Tap the day in the calendar. The selected day is shown with reversed text in a small blue box.
Part number 053-273B

94 GLOBAL SETUP
Security Settings

To change the time:

Set the elements of the time individually. 1. Tap in the Time box to highlight the hour, minute or second (24-hour clock). 2. Tap the Scroll Arrows to change the setting. 3. Repeat steps 12, as necessary, for the other elements in the Time box.

Security Settings
If instrument security is enabled and you know the security code, you have full access to the instrument. Full access means the ability to log on to the instrument after shutdown, disable the security system and change the security code. Logging on with a security code was discussed under The SLM and RTA display groups on page 14. If you want to change the state of instrument security or change the security code, you must unlock the Security Settings page of Global Setup (Figure 95). In Global Setup, select the page represented by the icon shown at the left.

Figure 95

Security Settings page

Part number 053-273B

GLOBAL SETUP 95
Unlocking Security Settings

Unlocking Security Settings


Only a person who knows the security code can unlock the Security Settings page and make changes in it. Changes in Security Changes take effect when you re-lock the page. The page is automatically relocked if you power-down the instrument.
To unlock security settings:

1. Tap the Input Panel icon. The Input Panel appears (Input Panel on page 20). 2. Touch and drag the cursor over the asterisk in the Security Code dialog box. 3. Enter the security code on the Input Panel. 4. Tap . If you entered the correct code, the asterisks in Security Code are dimmed and the same code is entered in the New Security Code and Confirm dialog boxes. Skip to step 8. 5. If you entered the wrong code, a Security dialog box appears telling you that the code is invalid. 6. Tap OK in the dialog box. 7. Repeat steps 34. 8. Leave the Input Panel open if you plan to set a new security code.

Editing Security Settings


Security settings must be unlocked before you can perform any of the operations below. See the unlock procedure above.
To enable/disable instrument security:

1. To enable, select Enable Security. To disable, clear Enable Security. 2. If finished making changes, tap
To set a new security code:

to save the new setting and re-lock security settings.

1. Touch and drag the cursor over the asterisks in the New Security Code dialog box. 2. Enter the new security code on the Input Panel. 3. Touch and drag the cursor over the asterisks in the Confirm dialog box. 4. Enter the same code on the Input Panel. 5. If finished making changes, tap to save the new setting and re-lock security settings. If the settings match, the new code is saved. 6. If the settings do not match, a Security dialog box appears. 7. Tap OK and repeat steps 15. NOTE: Before leaving Security Settings, tap the Input Panel icon to close the Input Panel.

Part number 053-273B

96 GLOBAL SETUP
Battery, Sound, Display Settings

Batter y, Sound, Display Settings


Use the Battery, Sound, Display Settings page to make changes to those global properties. In Global Setup, select the page represented by the icon shown at the left. An example of this page is shown in Figure 96.

Figure 96

Battery, Sound, Display Settings page

Battery status
The Battery Status area of this display shows you the percentage of charge. With no batteries installed, the battery status registers as 0%. If you have rechargeable batteries installed, you may see the Battery Status display interrupted periodically by a charging notice.

Backlight on time
You can select from a list of options how long the backlight will stay on without pressing one or more of the keys. To select a time, touch the down arrow in the Backlight on time box. A list appears. Scroll in the list to see all the options (Scrolling on page 13). Tap to select an option.

Speaker volume
Tap on the slider bar. If the pointer is not at that point, it jumps toward the point. Tap until the pointer is at a desired location. The speaker sounds to show the result as you tap to adjust. Tap to make the change effective.

Backlight Brightness
Tap on the slider bar. If the pointer is not at that point, it jumps toward the point. Tap until the pointer is at a desired location. The backlight changes as you adjust. Tap to make the change effective.
Part number 053-273B

GLOBAL SETUP 97
Screen contrast

Screen contrast
Tap on the slider bar. If the pointer is not at that point, it jumps toward the point. Tap until the pointer is at a desired location. The contrast changes as you adjust. Tap to make the change effective.

Meter Infor mation


Use the Meter Information page to find the instruments serial number and the versions of all installed software modules. You can change SoundPro DLXs display language on this page. In Global Setup, select the page represented by the icon shown at the left. An example of this page is shown in Figure 97.

Figure 97

Meter Information page

Updating the version list


There may be circumstances when you would like to ensure that all items in the list are up to date. Tap to refresh the list with information in memory.

Selecting a language
The current display language is identified in the Language dialog box. Select from a list in the box to change the language.
To change the language:

1. Tap the arrow in the dialog box below the Language dialog box. A list of available languages appears. 2. Tap to select from that list.
Part number 053-273B

98 GLOBAL SETUP
Network Settings

3. Tap

to begin operating with the selected language.

Networ k Setti ngs


Use the Network Settings page to enter and edit setups for wired, wireless or USB communications. In Global Setup, select the page represented by the icon shown at the left. An example of this page showing a wireless network configuration is shown in Figure 98.

Figure 98

Network Settings page

General setup procedure


All three setup types, including USB, require configurations that are compatible with the network with which the instrument will be communicating. You can enter and edit all applicable settings in this display. If you are unsure about what settings to use, see the network administrator.
To prepare for network operation:

1. Tap to select USB, Wireless or Wired in the Setup Type dialog box. 2. Enter the necessary communications settings. 3. Tap to save settings. If a window appears regarding a connection, follow its instructions.

Part number 053-273B

GLOBAL SETUP 99
Web Server Settings

Web Ser ver Settings


When enabled for web viewing, the instrument acts as a web server for remote access to data and settings in the instrument. In Global Setup, select the page represented by the icon shown at the left. An example of the Web Server Settings page is shown in Figure 99.

Viewing in a browser
You can use any of the three communications setups for web viewing. If, however, there are other devices multicasting in the same network, you also need to enter appropriate settings for the multicast and port parameters. For multicast parameter options, see your network administrator.
To configure for web viewing:

1. Connect the hardware needed for wired, wireless or USB communications (Ports and connectors on page 4).

Figure 99 Web Server Settings page

2. Set up or verify the setup for the type of communications youre using (Network Settings on page 98). 3. In Web Server Settings, select Remote Web Viewing to enable it. 4. Skip to the last step unless you need to set the multicasting parameters. 5. Tap the Input Panel icon to open the Input Panel. 6. Edit in the Multicast Address dialog box. If entering the whole address, touch and drag across the contents of Multicast Address, then tap a new address on the Input Panel. 7. Repeat step 3 for Multicast Port. 8. Tap to save the settings.

Viewing
Web viewing is available at any computer in the network that has a web browser. Enter the instruments IP address in the browser in the usual way. For example, if the IP address is 192.168.50.4, enter http://192.168.50.4. The SoundPro DLX home page will appear (Figure 910). NOTE: If you are already viewing in the browser and select or clear Remote Web Viewing, the browser doesnt respond immediately. Click Refresh in the browser to manually force an immediate update.

Part number 053-273B

100 GLOBAL SETUP


Viewing in a browser

Figure 910 Web viewing home page

Web pages
In addition to the home page, four other pages are available. To display any SoundPro DLX web page, click the name of the page on the gray bar located near the top of any page. The selected page appears in a different color according to your browser preferences.

Home ~ First page that appears when you enter SoundPro DLX web pages. Measurements ~ Display of the Sound Level Meter and Real Time Analyzer displays, side-byside and showing what the instrument is showing in those displays at the time. Setup ~ Allows you to change the settings in the meter that affect the results in the Sound Level Meter and Real Time Analyzer displays, including the measurement range. Once you click Send, the meter settings change and the results can be viewed in the web view Measurements display. Information ~ Identifies the instrument and the software versions currently running in it. Data ~ Shows you the sessions in the instrument. If you click on a session in this web view, you will see a summary of results for that session. You can print a report of the summary information from this display.

Part number 053-273B

Miscellaneous
Replacement and optional par ts

10

CHAPTER

This chapter contains information on a variety of topics that may arise infrequently if at all.

Of the three component parts to a fully assembled SoundPro DLX instrument, only the microphone can be replaced by optional parts. No other preamplifier or instrument body is available. The standard microphone is an electret microphone, the Quest QE4936, but you can order the instrument with other types or obtain a different type later. The instrument is shipped with the correct microphone polarization for the microphone you ordered with it. If you change microphones after purchase, make sure youre using the correct microphone polarization setting. See the section below for that information. You can also order and install a cable between the microphone and the preamplifier for the purpose of reducing reflection errors. For information about replacement and optional parts, see Appendix D, Specifications.

T he back cover
The reset switch and main batteries are inside the back cover.
To remove the back cover:

1. Press the Button Latch firmly while sliding the cover toward the base of the instrument. The covers six edge pins will slide out of the mating slots (one pin and slot pair circled in Figure 101). 2. When the pins have cleared the slots, lift off the back cover and lay it aside.
To replace the back cover:

1. Line up the edge pins with the slots in the case. 2. Press down until the cover fits flush against the back of the instrument. You may need to squeeze the sides of the cover slightly to guide the pins into the slots. 3. Slide the cover toward the top of the instrument until the Button Latch re-engages, then release the button. The cover should fit tight all the way around.

Part number 053-273B

102 MISCELLANEOUS
Installing or changing batteries

Figure 101

Removing/replacing the back cover

Installing or changing batteries


SoundPro DLX is shipped with rechargeable Lithium-ion (Li-ion) batteries in two battery packs. Older SoundPro DLX models were shipped with rechargeable Nickel Metal Hydride batteries (NiMH) batteries in two battery packs. All models are also shipped with tubes into which standard AA alkaline cells can be placed and the tubes then installed in place of the rechargeable batteries. This section explains how to remove and install all three of these battery types.

Li-ion battery packs


Remove the back of the instrument as explained above. The batteries are installed in virtually the same way as the NiMH battery packs, explained under NiMH battery packs, below, except for battery pack orientation. First, note that there is a front and rear end of each battery pack. The cable connector is closer to the front end, and there are indentations along the ring at the front end that fit corresponding ribs in the battery well. Also, the case at the rear of the battery pack is designed so that it fits against a rear rib at the bottom of the battery well. Install the Li-ion battery packs straight down (no turning or twisting) with the front indentations sliding along the front ribs and the battery packs resting snug against the battery well. Plug in as explained under NiMH battery packs, below.

NiMH battery packs


Remove the back of the instrument as explained above. Each battery pack is connected to the instrument by a cable with a plug at the end. Insert the plug in the mating socket (Figure 102). The plug fits only one way and latches in place. Before replacing the cover, make sure the batteries are both oriented so that the battery wiring and wiring packs clear the cover (Figure 103).

Part number 053-273B

MISCELLANEOUS 103
Installing or changing batteries

Figure 102
To install a NiMH battery pack:

Remove/install a NiMH battery pack

1. With the battery pack close to its slot in the case, orient the plug with the latch toward the inside of the instrument. 2. Press the battery plug into the case socket until the plug locks in place. 3. Lay the battery in the slot, rotating it as shown in Figure 103 so that the wiring pack will not stick up and interfere with the cover.

Figure 103 4. Repeat for the other battery. 5. Replace the back cover.
To remove a NiMH battery pack:

Positioning NiMH batteries during installation

1. Remove the back cover, as explained above.

Part number 053-273B

104 MISCELLANEOUS
Hardware reset

2. Lift out one of the batteries without tugging on the battery cable. This will enable you to get a good grip on the plug. See Figure 102 for this and the following steps in the procedure. 3. Grab the plug between your forefinger and thumb, depressing the latch. 4. Pull the plug out of the socket and lift the battery out.

Alkaline batteries
Two tubes for alkaline battery cells are provided with the instrument. Install with reference to the drawing in Figure 104. NOTE: SoundPro DLXs built-in charging circuit works only through the socket provided for the rechargeable battery packs, not the end connectors used for the alkaline cells. Alkaline cells are not exposed to recharging current.
To install alkaline batteries:

1. Insert three AA alkaline cells in one of the battery tubes with all cells facing the same direction. 2. Lay the assembled tube in the cases battery slot with the positive battery end toward the instruments base. 3. Repeat these steps for the other tube. 4. Replace the back cover.

Figure 104

Installing alkaline batteries

Hardware reset
Perform a hardware reset if the instrument appears to be operating abnormally and youve already tried a software reset without success (Software reset on page 9).
To do a hard reset: Part number 053-273B

MISCELLANEOUS 105
Microphone polarization

1. Turn the instrument over and remove the back cover (see The back cover, above). 2. Locate the Reset Access hole in the case next to one of the battery slots (see Figure 105). 3. Find a nonconducting, blunt-tip rod that is small enough to fit through the hole. It needs to penetrate approximately 3/8". 4. Poke the rod straight through the Reset Access hole until it contacts the shaft of a push-button switch below. 5. Push the rod in, then release (no need to hold). You should hear a tiny click when operating the reset switch.

If successful, the SoundPro DLX display will start clearing, indicating that a reboot operation has started. If not successful, try again. Make sure that power to the instrument is available when performing a hardware reset.

6. Once the power-on sequence begins, replace the back cover and follow the procedure given under Powering-up on page 7.

Figure 105

Reset switch location

Mi cr ophone polarization
The standard SoundPro DLX microphone for all models is an pre-polarized electret deviceit does not need external polarization. Other microphones available as options from Quest are condenser type microphones that do not carry their own charge and require an external polarization. For that reason the SoundPro DLX has a selection for microphone polarization. For information about microphone options, see Microphones on page 139.

Part number 053-273B

106 MISCELLANEOUS
Checking or changing microphone polarization

Checking or changing microphone polarization


Use the procedure below to verify, and if necessary change, the microphone polarization setting. SoundPro DLX is shipped with that setting disabled because its standard microphone is an electret type. If you should install a condenser microphone in place of a functioning electret microphone, simply enable the 200V potential setting. But when planning to install an electret microphone in place of a functioning condenser microphone, you must ensure that the setting is disabled before installing the microphone. Applying the 200V potential to an electret microphone can degrade or damage it. NOTE: The procedures below assume youre already familiar with the instrument, its controls and menus. If you need help with these procedures, see Chapter 2, Getting started. For information about working in the Calibration Settings page, see Calibrations and verifications on page 59.
Disabling the 200V polarization potential

1. Do not have the electret microphone installed when performing this procedure. You can have the preamplifier attached to the body without the microphone installed, or you can remove the preamplifier from the body. See Preamplifier on page 4 for assembly information. 2. Press 3. Press to switch the instrument on. and select Instrument Setup.

4. In Instrument Setup, tap in the Page Navigator to select the page represented by the icon shown at the right. This will open the Calibration Settings page. 5. For an electret microphone, clear the 200 V microphone polarization check box. 6. Install the microphone. You should see a normal SPL reading for the Measured Value at the bottom of the Calibration Settings page.
Enabling the 200V polarization potential

1. The instrument can remain fully assembled for this procedure. 2. Press 3. Press to switch the instrument on. and select Instrument Setup.

4. In Instrument Setup, tap in the Page Navigator to select the page represented by the icon shown at the right. This will open the Calibration Settings page. 5. For a condenser microphone, select the 200 V microphone polarization check box. If the microphone is installed, you should see a normal SPL reading for the Measured Value at the bottom of the Calibration Settings page.

Part number 053-273B

MISCELLANEOUS 107
Pinouts

Pinouts
External power
This connector is located on the base of the instrument body. See Ports and connectors on page 4 for the exact location. Figure 106 External power connector (12 VDC, 2.5 A)

Headset port
This connector is located on the base of the instrument body. See Ports and connectors on page 4 for the exact location.)

Figure 107 Headset connector (Phone jack: 2.5 mm

General Purpose Accessory connector


This connector is located on the base of the instrument body. See Ports and connectors on page 4 for the exact location.

Figure 108

General Purpose Accessory connector (Ribbon: 26-MINI-D)

Part number 053-273B

108 MISCELLANEOUS
QuestSuite Professional II application

QuestSuite Pr ofessional II a pplication


You can exchange information between SoundPro DLX and a personal computer with the QuestSuite Professional II (QSPII) application. This application is designed to provide offline services for many types of Quest instruments.

Transfer session data ~ Move session data from the instrument to QSPII. Display ~ Display session data in the computer using named templates or a wide variety of presentations that you create. Print ~ Generate reports in Portable Document Format (pdf). View ~ View measurements live, while a study is being performed. Program the instrument ~ Transfer setup information to the instrument to control future studies. Profiles ~ Use QSPII to store, transmit and share profiles between instruments.

Part number 053-273B

MISCELLANEOUS 109
Software system diagram

Softwar e system dia g r am

Part number 053-273B

110 MISCELLANEOUS
Software system diagram

This page intentionally left blank

Part number 053-273B

APPENDIX

Printout

This appendix contains a sample printout from SoundPro DLX. Printouts like these can be generated in the Session Manager (Printing on page 84).

General information

Settings

Summary section with Filter Values enabled

Figure A1
Part number 053-273B

Page 1 of a SoundPro DLX report

112 PRINTOUT

Summary section with Filter Values enabled

Figure A2

Page 2 of a SoundPro DLX report

Part number 053-273B

APPENDIX

Glossary

This section contains an alphabetical listing of terms related to the analysis and measurement of sound by SoundPro DLX. Some of the terms are the measurements performed by SoundPro DLX, some terms are the user-adjustable parameters for those measurements, and other terms are names for nonnumerical settings for some of those parameters. Explanations for the SoundPro DLX measurements included in this list are all found in Appendix C, Measurements.

A
A Weighting parameter setting. See Weighting on page 120 for information about this and other Weighting settings.

Bandwidth
A range of frequencies defined by an upper band-edge frequency, f2, and a lower band-edge frequency, f1. The bandwidth is the difference between the two frequencies (f2 f1 Hz).

C
A Weighting parameter setting. See Weighting on page 120 for information about this and other Weighting settings.

C-A
A Weighting parameter setting. See Weighting on page 120 for information about this and other Weighting settings.

Center frequency
The filter identification you see in SoundPro DLX displays and menus that refer to the center frequencies of octave or third-octave passbands. Center frequencies can be exactly calculated, and are listed for the audio spectrum in ANSI and IEC specifications. The center frequencies cited in SoundPro DLX, however, are preferred values because theyre not quite exact but very close and more convenient for reference purposes.

CNEL
A SoundPro DLX measurement. See CNEL on page 125.

Criterion Level (CL)


The average sound level for a Criterion Time (generally 8 hours) that would result in a allowable dose of 100% for an individual exposed to that sound.

Criterion Time (CT)


The time over which the Criterion Level is established, generally eight hours.
Part number 053-273B

114 GLOSSARY

Decibel (dB)
A dimensionless unit, commonly called the level in sound measurements, that compares two like quantities, such as powers or amplitudes, by forming a ratio of the two and taking the logarithm of the ratio. Decibel measurements express large differences with smaller numbers.

Dose
A SoundPro DLX measurement. See Dose on page 125.

Dynamic Range
The Dynamic Range of a sound pressure meter is the difference in decibels between the maximum SPL the instrument can measure accurately and the instruments noise floor. SoundPro DLXs Dynamic Range is 100 dB for broadband measurements and 80 dB for bandpass measurements.

Exceedance
The sound level that is equalled or exceeded for N% of the time over a defined period, generally either the course of a study or a session. Exceedances are determined in 1% intervals over the entire percentage range.

Exceedance Level (Ln) A SoundPro DLX setting used to determine a measurement of Exceedance. See LnX on
page 123.

Exceedance Percentage
The percentage value for which a determination of an Exceedance Level is made. Exceedance Percentages range from 0100 in one-percent increments.

Exchange Rate (ER)


The trade-off relationship between a change in sound level and the change in time for a fixed sound dose. For a 3 dB exchange rate, if the time of exposure is doubled, the sound level would have to be reduced by 3 dB to produce the same accumulated dose.

ExpHrs
A SoundPro DLX measurement. See ExpHrs on page 125.

ExpSec
A SoundPro DLX measurement. See ExpSec on page 125.

Fast
A Response Time parameter setting. See Response time on page 117 for information about this and other Response Time settings.

Part number 053-273B

GLOSSARY 115

Filter
For electronic processing, the means by which a signal is restricted to a particular bandwidth (passband). A filter can be uniquely specified by the two band-edge frequencies that define its limits and is commonly designated by its center frequency.

Impulse
A Response Time parameter setting. See Response time on page 117 for information about this and other Response Time settings.

Integrating Threshold The A-weighted sound level at which a sound measurement device begins to integrate noise signals into the measurement. An integrating threshold is commonly used in personal dosimeters to determine an individuals exposure to environmental noise. Lavg
A SoundPro DLX measurement. See Lavg on page 124.

LEP,d
A SoundPro DLX measurement. See LEP,d on page 126.

LDN
A SoundPro DLX measurement. See LDN on page 126.

Leq
A SoundPro DLX measurement. See Leq on page 124.

Level
In sound measurements, the logarithm of the ratio of two sound pressure quantities.

Lmin
A SoundPro DLX measurement. See Lmin on page 123.

Lmax
A SoundPro DLX measurement. See Lmax on page 123.

LnX
A SoundPro DLX measurement. See LnX on page 123.

Lpk
A SoundPro DLX measurement. See Lpk on page 124.

Part number 053-273B

116 GLOSSARY

Logging
In sound studies, the process of periodically saving measurements at fixed intervals during a study where each logged measurement is determined over the period of the logging interval.

Octave band
An octave band is a range of frequencies where the upper limiting frequency is twice the value of the lower limiting frequency. These limits, called band-edges, are the frequencies at which a filter transmits half as much power as the filter transmits at its center frequency.

Measurement
A numerical characteristic of the detected sound that is reported by SoundPro DLX. Measurements are determined by algorithms implemented in SoundPro DLX using user-adjustable parameters. Example of measurement include point measurements, time averaged measurements and time integrated measurements of SPL, along with times of occurrences of salient features in sound signals.

Measurement Range The decibel range within which the measurements are valid. The Measurement Range can be calculated by adding the instruments Dynamic Range to its Range Base or subtracting the Dynamic Range from the instruments Range Ceiling. Mntime
A SoundPro DLX measurement. See Mntime on page 127.

Mxtime
A SoundPro DLX measurement. See Mxtime on page 127.

Noise Unwanted sound. OLtime


A SoundPro DLX measurement. See OLtime on page 127.

OL%
A SoundPro DLX measurement. See OL% on page 127.

Noise Floor
The noise floor is the level, in decibels, at which the signal to be measured can no longer be differentiated from the internal noise of the instrument.

Pascal (Pa)
Unit of pressure equal to 1 Newton per square meter.

Part number 053-273B

GLOSSARY 117

Peak Count Threshold (PCT)


A SoundPro DLX parameter. This is the level you set for the determination of peak count (PKcnt on page 124). Each increment in the count means that the signal went above the PCT and subsequently fell below it.

PDose
A SoundPro DLX measurement. See PDose on page 126.

PKcnt
A SoundPro DLX measurement. See PKcnt on page 124.

PKtime
A SoundPro DLX measurement. See PKtime on page 127.

Projection time (Prt)


A SoundPro DLX parameter. This is the variable amount of time used to make a projected dose calculation (PDose) from an actual dose measurement.

Range Base
The minimum value of the Measurement Range. The Range Base is an adjustable parameter that can take on values between -10 dB and 80 dB when the Dynamic Range is 100 dB (broadband range) and values between -10 dB and 100 dB when the Dynamic Range is 80 dB (bandpass analysis range).

Range Ceiling
A SoundPro DLX display parameter. This is the maximum value of the Measurement Range. SoundPro DLX can measure up to a Range Ceiling of 180 dB.

Reference Sound Pressure


This reference pressure is defined for acoustic measurements to be 20 Pa (1 Pascal) when measured in air at 1000 Hz. This is approximately the threshold of normal hearing.

Response time
A SoundPro DLX parameter. This is an averaging characteristic that reduces the contributions of higher frequency components. Response time is commonly described by a time constant, which is a measurement of the time it takes for a system exposed to an abrupt input to rise or fall to the end condition. The SoundPro DLX response time parameter has three options: Slow, Fast and Impulse. For these, the associated time constants are 1.0 seconds, 0.125 seconds and 0.35 ms, respectively. Systems with a reduced response to higher frequencies will change by different amounts depending upon the length of time theyre exposed to an input. Figure B1 through Figure B3 below illustrate how SoundPro DLX responds to varying tone bursts for each of its response time options.

Part number 053-273B

118 GLOSSARY

Relative Response (dB)

Burst duration, t (seconds) Figure B1 Slow response to a tone burst

Relative Response (dB)

Burst duration, t (milliseconds) Figure B2 Fast response to a tone burst

Part number 053-273B

GLOSSARY 119

Relative Response (dB)

Burst duration, t (milliseconds) Figure B3 Impulse response to a tone burst

SEL
A SoundPro DLX measurement. See SEL on page 126.

Slow
A Response Time parameter setting. See Response time on page 117 for information about this and other Response Time settings.

SPL
A SoundPro DLX measurement. See SPL on page 124.

Statistics
A SoundPro DLX measurement. During studies and sessions, SoundPro DLX continually determines how often sound at a particular level occurs in relation to the occurrence of sound at all levels. For this percentage distribution, continuous sound level measurements are grouped in 1 dB intervals. For example, sound measurements during a study might be spread over a 40 100 dB range, and during that measurement period the sound at 60 dB occurred 50% of the time.

Taktm
A SoundPro DLX measurement. See Taktm on page 125.

Taktm Interval
A SoundPro DLX parameter. This is the time interval over which Taktmaximal values are integrated, usually set to either three or five seconds. Taktmaximal is the maximum level (Lavg) encountered during the Taktm Interval, which is selectable in the SLM Settings display (Making changes in SLM Settings on page 37). Taktmaximal measurements are required by some countries noise regulations.

Part number 053-273B

120 GLOSSARY

A Taktmaximal measurement calculates a higher average level for highly impulsive sounds than does an Leq measurement. Taktm is affected by the Exchange Rate and should be calculated with ER = 3 dB.

Third-octave band
A range of frequencies defined in ANSI S1.11. Each octave band contains three third-octave bands.

TWA
A SoundPro DLX measurement. See TWA on page 126.

ULTime
A SoundPro DLX measurement. See ULTime on page 127.

Upper Level Limit (ULL) URTime


A SoundPro DLX measurement. See URTime on page 127.

UR%
A SoundPro DLX measurement. See UR% on page 127.

Weighting
A SoundPro DLX parameter. The human ear does not respond to all sound in the same way. Higher and lower sound frequencies are attenuated so that normal hearing is confined to sound in the range of 20 Hz to 20,000 Hz (20 kHz). In addition, the frequency response depends to some degree on the sound pressure level as well. SoundPro DLX includes a weighting parameter that allows you to choose from among several weighting curves that provide varying corrections to the frequency components of the SPL measurementsA-weighting, C-weighting, Z-weighting or C-A weighting. A-weighting and C-weighting have been found over the years to best adjust for the hearing response at moderate and low sound levels, respectively. A-weighting is commonly used for general purpose sound measurements while C-weighting is applied when evaluating hearing protectors. When no adjustment for the hearing response is needed, investigators will often choose zero weighting (Z-weighting) which applies no adjustments in the measurement spectrum. C-A weighting is often used when performing community noise monitoring because it tends to enhance low frequency noise. The four figures below show the effect of weighting across the measurement spectrum for these four types of weighting. Relative response means the amount of attenuation at a particular frequency relative to the amount of weighting at 1 kHz. For each of these plots, the center curve is the weighting curve and the curves on either side of it represent tolerances specified in IEC standard 61672-1, March 2001, Sections 5.4 and 5.16.3.

Part number 053-273B

GLOSSARY 121

Relative Response (dB)

Frequency (Hz) Figure B4 Plot of A-weighted corrections

Relative Response (dB)

Frequency (Hz) Figure B5 Plot of C-weighted corrections

Part number 053-273B

122 GLOSSARY

Relative Response (dB)

Frequency (Hz) Figure B6 Plot of Z-weighted corrections

Relative Response (dB)

Frequency (Hz) Figure B7 C-A weighting response curves

Windscreen
A covering for the measurement microphone designed to minimize distortions in the sound signal by physical contact or wind turbulence.

Z
A Weighting parameter setting. See Weighting on page 120 for information about this and other Weighting settings.

Part number 053-273B

APPENDIX

Measurements

SoundPro DLX detects and analyzes sound using a wide variety of measurements. This section describes those measurements and, where applicable, gives calculation formulae. For definitions of terms and parameters, see Appendix B, Glossary. Some parameters are defined under Other formula parameters on page 127. For the purposes of formulae below, these are definitions: LS ~ Time-varying sound level, in decibels, that exceeds the Threshold Level. The response time of the measuring instrument is indicated by the second letter (Slow or Fast). RTIME ~ Run Time during which the measurement is made. TC ~ Time Constant for the exponential time weighting.

Time-weighted measur ements


Sound level measurements are reported after the sound signal is processed through an exponential time-weighting circuit with a user-selectable time constant.

Lmin
Minimum Level ~ Lowest SPL measured over a time interval.

Lmax
Maximum Level ~ Highest SPL measured over a time interval.

LnX
Exceedance level ~ The level, in decibels, that was equaled or exceeded by the SPL a fixed percentage of time during an interval. For example, Ln60 = 40 dB means that the sound measured during an interval exceeded 40 dB more than 60% of the time. Exceedance measurements in SoundPro DLX are made at integer values of percentage. SoundPro DLX calculates exceedance levels for integral percentages throughout the 0100% range. You can select exceedance levels for online viewing at fixed percentages and at four custom percentages, which are designed as L1, L2, L3 and L4. To set custom exceedance levels, see Making changes in SLM Settings on page 37.

Part number 053-273B

124 MEASUREMENTS
Peaks

SPL
Sound Pressure Level ~ Logarithm of the ratio of two measurements of sound pressure, where the denominator of the ratio, Pr, is the Reference Sound Pressure (see Appendix B, Glossary, for an explanation of this term).

PSPL = 20 log ---- dB P r

Peaks
Lpk
Peak Level ~ Highest SPL measured over a time interval.

PKcnt
Peak Count ~ Count of the number of times that the sound level went above the Peak Count Threshold (PCT) and then returned below it (Peak Count Threshold (PCT) on page 117).

Time-aver a ged measur ements


Lavg
Average Level ~ Average sound level calculated for an arbitrary Exchange Rate.
RTIME

L avg = ER

log 2

2
0

LS ER

dt log 2 ( RTIME ) dB

NOTE: Lavg = Leq for an Exchange Rate of 3 dB and no Integrating Threshold (disabled). See Other formula parameters on page 127 for an explanation of these parameters.

Leq
Equivalent Continuous Level ~ Average sound level calculated for an Exchange Rate of 3 dB.

L eq = 3

log 2

RTIME

2
0

LS 3

dt log 2 ( RTIME ) dB

Part number 053-273B

MEASUREMENTS 125
Integrated measurements

Taktm
An exposure measurement that is used commonly when monitoring with respect to regulations in Germany, this is a time integration of measurements called Taktmaximal values. Each Taktmaximal value is the highest level during either a 3 or a 5 second interval (Taktmaximal Interval). You can set the Taktmaximal Interval on an SLM Settings page (Changing settings on page 39).

Inte g r ated measur ements


CNEL
Community Noise Exposure Level ~ The accumulated exposure to sound measured in a 24-hour study, where the sound level is artificially increased before the integration process at certain times during the study. Between the hours of 7 p.m. and 10 p.m., the sound level is increased by 5 dB; between the hours of 10 p.m. and 7 a.m., the sound level is increased by 10 dB. A 3 dB Exchange Rate is used unless otherwise specified.

Dose
The ratio of the accumulated exposure to sound to the maximum exposure allowed under a Criterion Level standard.

100 Dose = -------TC

RTIME

2
0

( LS CL ) ( ER )

dt

ExpHrs
Hours of sound exposure ~ Accumulated exposure to sound measured in hours using an Exchange Rate of 3 dB. The display will switch to ExpSec if RTIME is less than one hour.

EXP Hrs = [ 2 ExpSec

( L eq 94 ) 3.01

RTIME 2 ] ------------------- Pa H 3600

Seconds of sound exposure ~ Accumulated exposure to sound measured in seconds using an Exchange Rate of 3 dB. The display will switch to ExpHrs if RTIME is greater than 3600 seconds.

EXP Sec = [ 2
Part number 053-273B

( L eq 94 ) 3.01

] RTIME Pa S

126 MEASUREMENTS
Integrated measurements

LEP,d
A measurement of daily personal noise exposure that is commonly used in Europe.

LDN
Day/Night sound level ~ The accumulated exposure to sound measured in a 24-hour study, where the sound level is artificially increased by adding 10 dB to it during the hours of 10 p.m. and 7 a.m. Typically, LDN is measured without an Integrating Threshold and at a 3 dB Exchange Rate.

SEL
Sound Exposure Level ~ The continuous sound level lasting one-second that would deliver the same sound exposure as the accumulated exposure measured over a Run Time interval. SEL is typically determined with a 3 dB Exchange Rate, slow exponential time-weighting and without regard to an Integrating Threshold.

SEL = ER

log 2

RTIME

2
0

LS ER

dt

dB

PDose
Projected dose ~ Calculated value for sound dose over a Projection Time (Prt) determined from an actual dose measurement (Dose) made for a particular run time interval.

PRT PDose Prt = DOSE ------------------- % RTIME TWA


Time Weighted Average ~ The accumulated sound level over a run time interval, then averaged with respect to the Criterion Time interval. The TWA measurement is obtained with an Integrating Threshold, slow exponential time-weighting and frequency weighting.

TWA = ER

log 2

RTIME

2
0

LS ER

dt log 2 ( CT )

dB

Part number 053-273B

MEASUREMENTS 127
Timing

Timing
Mntime
Time into a study (or, if logged, into a logging interval) when the minimum SPL occurs (HH:MM:SS).

Mxtime
Time into a study (or, if logged, into a logging interval) when the maximum SPL occurs (HH:MM:SS)

OLtime
Overload Time ~ The period of time during which the input signal is above the linear measurement range of the instrument.

OL%
Overload Percentage ~ Fraction of Run Time during which an overload condition exists.

OLtime OL% = 100 ------------------- percent RTIME PKtime


Time into a study (or, if logged, into a logging interval) when the peak SPL occurs (HH:MM:SS).

ULTime
Upper Limit time ~ Total time that the sound level exceeds a preset level.

URTime
Under Range Time ~ The period of time during which the input signal falls below the linear measurement range of the instrument.

UR%
Under Range Percentage ~ Fraction of Run Time during which an under range condition exists.

URtime UR% = 100 ------------------- percent RTIME

Other for mul a par ameter s


LS ~ Time-varying sound level, in decibels, that exceeds the Threshold Level. The response time of the measuring instrument is indicated by the second letter (Slow or Fast). RTIME ~ Run Time during which the measurement is made. TC ~ Time Constant for the exponential time weighting. ER ~ Exchange Rate.
Part number 053-273B

128 MEASUREMENTS
Other formula parameters

This page intentionally left blank

Part number 053-273B

APPENDIX

Specifications
Standar ds and a ppr ovals
Acoustics

IEC61672-1 (2003) (2002-05) ~ Electroacoustics - Sound level meters - Part 1: Specifications. IEC 61672-2 (2003-04) ~ Electroacoustics - Sound level meters - Part 2: Pattern evaluation tests. ANSI S1.4-1983 (R2001) ~ American National Standard Specification for Sound Level Meters. IEC 61260 (1995-08) ~ Electroacoustics - Octave-band and fractional-octave-band filters. ANSI S1.11-2004 ~ American National Standard 'Specification for Octave-Band and Fractional-Octave-Band Analog and Digital Filters. ANSI S1.43-1997 (R2002) ~ American National Standard Specifications for Integrating Averaging Sound Level Meters. Earlier standards ~ Also fulfills all requirements of IEC60651 and IEC60804. CE ~ Certificate of Conformity according to the EMC directive, low-voltage directive, and applicable acoustic measurement standards. Documentation on-file at Quest Technologies, Oconomowoc, WI, USA. EN61010.1 ~ Safety requirements for electrical equipment for measurement, control and laboratory use.

European

EMC emissions and immunity


Emissions 61000 series, CISPR 11 (2003-03) ~ Industrial, scientific and medical (ISM) radio-frequency
equipment - Electromagnetic disturbance characteristics - Limits and methods of measurement.

RFI Class B (Residential, commercial, & light-industrial) ~ 150 kHz - 30 MHz Conducted, 30 MHz - 1 GHz radiated.

Part number 053-273B

130 SPECIFICATIONS
EMC emissions and immunity

Immunity (Susceptibility)
NOTES:

Level A ~ Unit operates as intended. No degradation of performance or loss of function. Level B ~ Unit operates as intended after the test. Level C ~ Temporary loss of function, with either self-recovery or by operation of controls. IEC 61000-4-02 (2001-04) (EMC) - Part 4-2 ~ Testing and measurement techniques Electrostatic discharge immunity test. 8 kV AIR, 4kV Continuous, 3kV Capacitive. Level B. IEC 61000-4-03 (2002-09) (EMC) - Part 4-3 ~ Testing and measurement techniques Radiated, radio-frequency, electromagnetic field immunity test. 30 MHz - 1 GHZ, 10 V/m, 1 kHz 80% modulation). Level A, except a reading of 55 dB was noted between 9500 MHZ and 1 GHZ. IEC 61000-4-04 (1995-01) (EMC) - Part 4 ~ Testing and measurement techniques - Section 4: Electrical fast transient/burst immunity test. Basic EMC Publication. 2 kV, on AC, DC, & Control Lines, 1 kV IO Lines. Level B. IEC 61000-4-05 (2001-04) (EMC) - Part 4-5 ~ Testing and measurement techniques - Surge immunity test. 2kV Common-mode, 1kV Differential mode. Level B.

IEC 61000-4-06 (2003-05) (EMC) - Part 4-6 ~ Testing and measurement techniques Immunity to conducted disturbances, induced by radio-frequency fields. 0.15 - 80 MHz, 10 Vrms, 1kHz, 80% modulation. Level A. NOTE: Readings above 70 dB are not affected except between 10 MHZ and 16 MHZ at 10 V/m. IEC 61000-4-08 (2001-03) (EMC) - Part 4-8 ~ Testing and measurement techniques - Power frequency magnetic field immunity test. 50 Hz, 30 A/m. Level A. IEC 61000-4-11 (2004-03) (EMC) - Part 4-11 ~ Testing and measurement techniques Voltage dips, short interruptions and voltage variations immunity tests. With internal Lithium ion batteries >25% charged. Level A.

Part number 053-273B

SPECIFICATIONS 131
General

Gener al
Warm up

Hard RESET / Cold Start from loss of power ~ Less than 2 minutes. Normal (Power-up, from Sleep) ~ 30 seconds.

User Interface

Display ~ 3.9" Diagonal, VGA resolution (240 x 320 pixels), front-lit, color reflective, TFT liquid-crystal display with touch-panel. Keypad ~ Ten translucent push-buttons with backlight. Backlight ~ User adjustable. Color ~ 65,536 colors (16-bit). Contrast ~ User adjustable. Languages ~ English, French, Spanish, Italian, Portuguese and German.

Measur ements
Types performed
Dose Lpk OL% URtime UR% Custom Ln4 Pdose Leq CNEL SEL Ln90 Mntime Lavg Lepd ULtime ExpSec Custom Ln1 Mxtime Lmin TWA Taktm LND Custom Ln2 PKtime Lmax OLtime PKcnt Ln10 CustomLn3 SPLa

a) Time History Data only

See Appendix C, Measurements, for a full listing of measurements performed by SoundPro DLX and, in some cases, the formulas used.

Part number 053-273B

132 SPECIFICATIONS
Ranges

Ranges
Overall frequency range
The ranges shown in Table D1 represent performance of the microphone and instrument combined. Table D1: Frequency ranges, by microphone Microphone Type BK4936 QE7052 QE4110 QE4130 QE4170a Description 1/2" Electret 1/2" Electret 1/4" Condenser 1/2" Condenser 1" Condenser Pressure Class/Type 1 2 1 1 1 Frequency Range 8Hz to 20 kHz 20Hz to 17 kHz 5Hz to 24.5 kHz 3Hz to 24.5 kHz 3Hz to 10 kHz

a) Pressure microphone QE4170 is not to be used for free-field or random-incidence measurements.

Display ranges
The instrument has a 100 dB range in the SLM mode and an 80 dB range in the RTA mode.

SLM (100 dB) mode ~ Ten Selectable overlapping ranges in 10 dB steps.

Lowest SLM mode range: -10 to 90. Highest SLM mode range: 80 to 180. Lowest RTA mode range: -10 to 70. Highest RTA mode range: 100 to 180.

RTA (80 dB) mode ~ Twelve selectable overlapping ranges in 10 dB steps.

Maximum PEAK level


3 dB above full-scale reading.

Maximum Crest Factor


75 dB crest factor. For example, up to 140 dB SPL with BK4936 microphone at 65 dBC. Overload indication will occur if upper range is exceeded. Other optional microphones will raise or lower the SPL limit.

Part number 053-273B

SPECIFICATIONS 133
Parameters

Parameters
Weighting A-weighting, C-weighting, C-A weighting (broadband only).
Z-weighting (previously called Linear): 3Hz (-3dB) to 24.5 kHz (-3dB). In the RTA mode, any one of the three weightings may be selected. A different weighting may be selected for each of the three meters.

Response Times Slow, Fast, Impulse, and Peak. Peak response ~ Available for all three meters with A, C, or Z-weighting settings. Peak
response has an onset time constant of < 50 microseconds.

Octave band filters


ANSI S1.11-1986, Order 3, Type 2, Sub-type C and IEC R225-1966. Center Frequencies: (1/3 Mode) 33 center frequencies from 12.5 Hz to 20 kHz. (1/1 Mode) 11 center frequencies from 16 Hz to 16 kHz.

Threshold
Off or user selectable level 0-140 dB.

Exchange Rate
3, 4, 5, 6.

Run modes
Level triggered run/pause and clock/calendar triggered power on and run for programmed run duration.

Signal processing
Detector Pulse range (C-weighted) ~ 75 dB (Crest Factor). Measuring range ~ When measuring a steady signal near the lower range limit, you get the
entire upper part of the range (up to 75 dB) for crest factor capability.

Digital Signal Processing (DSP)


24-bit samples @ 52.734 KHz.

Part number 053-273B

134 SPECIFICATIONS
Signal processing

Instrument Dynamic Range


Up to 143 dB. Electrical range: -118 dBV to 24.6 dBV (Peak). Measured at the input to the preamplifier.

Primary Indicator/Linearity Range


Tested with a sinusoidal signal input.

SLM ~ 100 dB, unless limited by noise. RTA ~ 80 dB, unless limited by noise floor.

Level Linearity
Inside the primary indicator range. It is electrically tested on the reference range (20 to 120 dB) with a sinusoidal input signal. Tolerance is 0.7 dB referenced to 400 mV at 114 dB SPL.

Attenuator Accuracy
Referred to the Reference Range and the Reference SPL. Within 0.5 dB from 3Hz to 22.4 kHz

Integration Time (Settling Time)


One minute when measuring a short impulse. Five seconds when integrating a constant input signal.

Bandwidth
3Hz to 24.5 kHz (Z-weighted, 1 dB).

Filter bands Design ~ Optimized Parks-McClellan analytical method. Frequency ratio ~ Base-ten system. Reference attenuation ~ Zero. Sampling ~ Within the system sampling rate of 52,734.375 samples per second, the sampling
rate for individual filters is an octave-based decimation of 2.

Octave passband ~ The digital octave-filters are flat within <0.3 dB in the passband. The half-power points are at 0.707 fc and 1.414 fc, where fc is the bands center frequency. approximately 20 dB down. The 10fc frequencies are approximately 95 dB down and the fc/10 frequencies are approximately 70 dB down.
Octave skirts ~ The 2fc frequencies are approximately 30 dB down and the fc/2 frequencies are

Third Octave passband ~ The 1/3 filters are flat within < 0.3 dB in the passband. The half power points are at 0.89 fc and 1.12 fc. Third Octave skirts ~ The fc/2 and 2fc frequencies are approximately 50 dB down. Real time ~ For each nominal filter bandwidth the frequency range for real-time operation and real-time spectral analyses of transient and time-varying signals is 12.5 Hz to 20kHz, center frequencies.
Part number 053-273B

SPECIFICATIONS 135
References

References

Range

SLM (100 dB) mode ~ 20 dB to 120 dB. RTA (80 dB) mode ~ 40 dB to 120 dB.

SPL ~ 114 dB (Using BK4936). 400 mVrms) electrical @ 114 dB SPL. Frequency ~ 1 kHz. Direction ~ 0o when using a Free Field Microphone. Sound is arriving from directly in front of the microphone diaphragm. (A normal line extending from the center of the microphone diaphragm.).

Resolution
0.1 dB.

Calibr ation

Initial factory calibration is traceable in conjunction with ISO9001:2000 and A2LA Registered to ISO/IEC Guide 25/EN17025 Factory calibration and initial user calibration are stored in unit to be compared with calibrations that are taken during or after measurements are taken. Calibrators ~ Quest QC-10 or QC-20.

Log ging
Standards 32 MB: 1 sec. logging. Allows logging of up to 104 measurement values for 5-6 hours. At 0.1 sec. logging rate (10X/Sec.), you can store 12 values/interval for 4-5 hours. Standard memory allows you to store over 2 million logged data points, plus additional summary information for each study and overall session.

Special functions (Class 1 models onl y)



Back Erase ~ Selectable 5 s to 24 h of data can be erased. Voice Notes ~ 10 s sound recording can be attached to measurements so that verbal comments can be stored along with the measurement, using the microphone or headset microphone. Voice notes can be played back using an earphone/headphone connected to the headset port or to the instruments built-in speaker. Drawing Pad ~ Drawings and text notes can be attached to measurements using the stylus. Sound Curve Capture ~ Captures and stores a snapshot of the spectral footprint of a noise source. Allows you to build libraries of multiple sound curves and then recall them for onscreen comparison with future spectral measurements.

Part number 053-273B

136 SPECIFICATIONS
Ports

Por ts
Infrared
Located on the side of the instrument. Conforms to the IRDA standard. Can be used as a printer port (Hewlett-Packard Photosmart, model 1215/1315, or similar.)

General Purpose Accessory


Located at the base of the instrument. Multiple purposes.

AC Output ~ The AC output is used with the optional AC/DC output cable. (4.0 V P-P maximum). The output impedance is 1000 ohms. Dynamic range is limited to 80 dB. Connected equipment should have an input impedance of > 10K ohms. The output can be shorted without damaging the meter or changing the meter reading. If full scale is set at 140dB, range will be from about 60 dB to 140 dB. DC Output ~ The output range is 0-5.0 volts, which corresponds to an SPL range of -24 dB to 180.0 dB (2.44 mV/0.1 dB). Output impedance is 1000 ohms. The output can be shorted without damaging the meter or changing the meter reading. TTL ~ Two channels are available, each of which can be set high or low by a triggered response (Trigger Manager on page 68). USB ~ Uses IP over USB protocol. Compatible with the 1.1 USB standard.

Headset
Accepts headset with 2.5 mm connector for voice in and audio out.

CompactFlash
Can be used for any of the following purposes.

CD card ~ With optional CD card, use as external memory to receive exported session information (Exporting sessions on page 83). Ethernet ~ With optional Ethernet adapter, use to connect to a network. Wireless ~ With optional wireless adapter, use to connect to a network.

Part number 053-273B

SPECIFICATIONS 137
Environmental characteristics

Envir onmental char acteristics


Temperature

Operating temperature ~ (< 0.5 dB effect) -10 C to +50 C (+14 F to +122 F). Storage temperature ~ -25 C to +70 C (-13 F to +158 F).

Humidity

(< 0.5 dB effect) 90% RH (non-condensing) @ 40C (104 F). (< 0.5 dB effect) for 10% < RH < 90% @ 40 C (104 F) and 1 kHz.

Shock and vibration



Non-operating IEC68-2-6: Vibration: 0.3mm, 20m/s (2G), 10-500 Hz. Non-operating IEC68-2-27: Shock: 1000 m/s (100G). Non-operating IEC68-2-29: Bump: 1000 bumps at 250 m/s (25G). Vibration Susceptibility: (Operating) ~ Vibrations were applied between 20 Hz and 1 KHz. in 3 axes. With a dummy microphone installed and vibration at 1 m/s (~0.1 G), the maximum deviation from the noise floor in the worst case was 8.2 dB. At 10 m/s (~1 G), the maximum deviation from the noise floor in the worst case was 18.6 dB. When the microphone extension cable is used and only the pre-amp and microphone are shaken at 10 m/s (~1 G), the maximum deviation from the noise floor in the worst case was 6.3 dB.

External fields

Magnetic field susceptibility ~ As measured per IEC61672-1 Section 6.6. A magnetic field of 80A/m at 50 or 60Hz has no effect at a reference sound level of 74 dBA. Intense sound fIelds ~There is a minimum of 25 dB separation between internal microphonics and external SPL. (20 dB is needed to meet the standard). Above 250 Hz, there is at least 60 dB of separation. Electric ~ An electric field of 10 V/m over 150kHz to 1 GHz produces a maximum reading of 55 dBZ. Electrostatic ~ Negligible interference as long as the protection grid is kept on the microphone. Up to 1 kV over DC-100 kHz. Electro-Static Discharge (ESD) ~ All externally accessible screws, pins, cables, and connectors are protected up to 4 KV Contact discharge and 8 KV air discharge. Internal accessible battery pins and terminals (when the rear cover is removed) are fully protected when the unit is in it's standby state.

Part number 053-273B

138 SPECIFICATIONS
Electrical characteristics

Electrical char acteristics


Batteries
SoundPro DLX uses two types of batteries in normal operation: a main battery set and a backup battery. The main batteries can be either rechargeable Lithium-ion or alkaline batteries. Earlier models used rechargeable Nickel Metal Hydride, but these batteries are no longer available for SoundPro DLX.

Battery performance
The following data were obtained for the instrument operating on battery power alone (without AC power), and with fresh or fully-charged batteries. During the tests, the instrument was operated continually in the run mode. The testing ended when the instrument shut itself down for lack of sufficient battery power. While operating solely under battery power, the instrument conforms to all applicable tolerance limits.

Lithium-ion ~ Without backlight: 9 hours; with backlight: 6-7 hours. Nickel Metal Hydride ~ Without backlight: 4 hours; with backlight: 3 hours. Alkaline ~ Without backlight: 1.5 hours; with backlight: 1 hour.

External power supply


Switching type. Furnishes power to operate the instrument as well as to operate the internal charger when rechargeable batteries are installed.

Input ~ 100 - 240 VAC at 50/60 Hz and 0.8A. Output: ~ +12 VDC, 2.5A. Power input jack ~ Requires 11 to 16 VDC.

Backlight OFF ~ For externally-powered, long-term measurements, the unit consumes approximately 400 mA at 12V input (4.8W). Backlight ON ~ Power consumption increases to 500 mA (6.0W).

Meters

Input ~ The input impedance is 100 Kohm in series with 11F with approximately 100 pF loading to ground. Frequency range ~ "Z" (Linear) weighting. Electrical performance, meter and pre-amp with 059-703 BNC input adapter.

0.1 dB ~ 16 Hz to 24.4 kHz. 1.0 dB ~ 2.5 Hz to 25.3 kHz. 3.0 dB ~ 2.0 Hz to 25.8 kHz.

Part number 053-273B

SPECIFICATIONS 139
Preamplifier

Preamplifier

Dimensions ~ Directly accepts 1/2 in. (0.52" or 13.2 mm) microphone. Other sizes require an adapter. Assembly ~ Preamp is removable. Cable ~ Will drive up to 30 meters of cable with negligible signal loss. Input impedance ~ Greater than 1 G with approximately 3 pF loading. Signal limit ~ 12 VAC rms (~ 21 dBV rms, ~ 24 dBV peak).

Internal noise
After calibration at 400 mV @ 114 dB (-28dBV at 94 dB-SPL, 1kHz), like a nominal QE4936 microphone, typical noise levels with a 12pF input load will be:

10.6 V-A (22.5 dBA). 50.1 V-C (36.0 dBC). 202 V-Z (48.1 dBZ, 3Hz to 24.5 kHz).

Microphones
Polarization voltage
Regulated (low-current) 200V DC ( 2%) available for unpolarized condenser microphones. The microphone polarization can be switched off by the user for pre-polarized electret microphones.

Microphones available BK4936 ~ Standard, electret-type microphone for Class/Type 1 models. Use for sound levels
in the range of 29142 dB.

QE7052 ~ Standard, electret-type microphone for Class/Type 2 models. Low cost and disposable, typically used for general purpose free-field measurements. Use for sound levels in the range of 28143 dB. QE4110 ~ A small diameter free-field condenser microphone especially suited for very high levels. Recommended for measurements between 45 and 167 dB, or for highly accurate highfrequency measurements. QE4130 ~ A good free-field condenser microphone for general electro-acoustic purposes, loudspeaker and microphone measurements. Flat response to 30 KHz for a sound levels in the range of 35150 dB. QE4170 ~ Pressure style condenser microphone used for coupler measurements, audiometer calibration, low-frequency measurements, low-level measurements (due to lower noise floor) or as a laboratory standard. Use for sound levels in the range of 18140 dB.

Part number 053-273B

140 SPECIFICATIONS
Mechanical characteristics

Table D2: Characteristics of microphones alone Property/Model Part number Nominal diameter Bias voltage Response characteristic Frequency response ( 2 dB) Sensitivity (dBV)c Sensitivity (mV) db (1 kHz, 1/3) noise dBA noise dBC noise dBZ noise Mic range dBA min (recommended) dB Peak Nominal capacitance (pF) Accuracy BK4936a 059-523 1/2 inch Electret Free-Field 8 Hz 20 kHz -28 40 11.4 22.2 33.9 45.6 29 142 12 Class 1 QE7052b 056-317 1/2 inch Electret Free-Field QE4110 059-413 1/4 inch 200V Free-Field QE4130 058-659 1/2 inch 200V Free-Field QE4170 058-488 1 inch 200V Pressure 3 Hz 10 kHz -26 50 -0.8 11.8 20.9 34.1 18 140 60 Class 1

20 Hz 17 kHz 5 Hz 24.5 kHz 3 Hz 24.5 kHz -29 35 10.2 21.3 33.5 45.7 28 143 15 Class 2 -49 3 26.6 45 56.3 68.1 45 167 6.5 Class 1 -36 15 17.5 28.1 40.1 52.9 35 150 18 Class 1

a) Standard microphone for the Class/Type 1 models. b) Standard microphone for the Class/Type 2 models. c) Relative to 1 V/Pa (mV/Pa).

Mechanical char acteri stics



Tripod mount ~ A threaded insert on back of the meter accepts a standard -20 tripod mounting screw. Housing ~ Tough ABS/Polycarbonate with Internal EMC shielding and external ergonomic handgrips. Size ~ 10.2 cm wide 33.7 cm long 4.8 cm thick (4" 13" 2"). Weight (with Lithium-ion batteries) ~ Approximately 900 g (2 lbs).

Part number 053-273B

SPECIFICATIONS 141
Parts

Par ts
Original equipment kits
053-830 053-831 053-832 053-833

SoundPro DLX-1-1/3 (Class/Type 1 instrument with 1/1 and 1/3 filtering capability, 053-840 microphone kit and 100-425 headphone /microphone). SoundPro DLX-1-1/1 (Class/Type 1 instrument with 1/1 filtering capability, 053-840 microphone kit and 100-425 headphone /microphone). SoundPro DLX-2-1/3 (Class/Type 2 instrument with 1/1 and 1/3 filtering capability and 053-841 microphone kit). SoundPro DLX-2-1/1 (Class/Type 1 instrument with 1/1 filtering capability and 053-841 microphone kit).

Original equipment replacements


053-840 053-841 053-809 053-237 057-823 053-263 053-434 053-855 053-164 058-115 059-344 100-424 100-425

SoundPro DLX Class 1 BK4936 microphone kit (BK4936 microphone, 056-990 microphone-to-calibrator adapter and 059-344 windscreen). SoundPro Class 2 QE7052 microphone kit (BK7052 microphone, 056-990 microphone-to-calibrator adapter and 059-344 windscreen). SoundPro DLX preamp. SoundPro DLX AA battery tube. Package of two. SoundPro DLX Lithium-ion batteries. Package of two. Universal power supply 12V, 2.5A, 2.1 mm plug. SoundPro DLX DLX dust boot. SoundPro DLX USB cable to pc.
Wrist strap, 1/4" - 20. WS-3 windscreen for 1.0" microphone. Package of 3. WS-7 windscreen for 0.5" microphone. Package of 3.

SoundPro DLX stylus.


Headphone /microphone with 2.5 mm plug.

Part number 053-273B

142 SPECIFICATIONS
Optional microphones

Optional microphones
053-842 053-843 053-844

SoundPro DLX QE4170 microphone kit (QE4170 microphone and machined adapter to preamp). SoundPro DLX Class 1 QE4130 microphone kit (QE4130 microphone, 056-990 microphone-to-calibrator adapter and 059-3447 windscreen). SoundPro DLX Class 1 QE4110 microphone kit (QE4110 microphone with adapter, 056-990 microphone-to-calibrator adapter and 059-344 windscreen). QE4170 pressure microphone, 1" condenser type. QE4130 free field microphone, 0.5" condenser type. QE4110 free field microphone, 0.25" condenser type.

058-488 058-659 059-413

Cables
053-851 053-852 053-853 053-870 053-882 SoundPro DLX preamp extension cable, 1 M. SoundPro DLX preamp extension cable, 3 M. SoundPro DLX preamp extension cable, 15 M. SoundPro DLX automobile cable. SoundPro DLX AC/DC cable.

Miscellaneous
053-372 059-703 016-097 016-098 016-116 056-981 056-982 056-990 059-045 059-458 053-176 053-845 059-703 Direct coupled input adapter. Input adapter with 18 pF capacitor. 10 Mbps Ethernet adapter with cable for CompactFlash slot. 11 Mbps 802.11b wireless adapter for SoundPro DLX CompactFlash slot. 11 Mbps 802.11b wireless adapter for pc. QC-10 Calibrator; 114dB at 1000 Hz Output. QC-20 Calibrator; Selectable 94dB or 114dB at 250 Hz or 1000 Hz Output. Microphone-to-calibrator adapter for QC-10 and QC-20 calibrators. TP-1 tripod (height range: 18.5 in. - 48.9 in.). 1.5V AA batteries. Six required for 053-237 tubes. SoundPro DLX NiMH battery pack. Package of two. SoundPro basic carry case. 18 pF adapter

Part number 053-273B

APPENDIX

Quest policies
Service

Congratulations! You have purchased one of the finest instruments available, manufactured by one of the most respected names in safety & industrial hygiene instrumentation. Your instrument is backed by a limited warranty that seeks complete customer satisfaction. Should your instrument require service for any reason, you can expect prompt and courteous attention. You must obtain a return authorization prior to shipment. We reserve the right to refuse any shipments forwarded without prior authorization. The following information will expedite the service process and is required when obtaining return authorization:

Model and serial number of each instrument. Description of work required and symptoms of any failures for each instrument. For non-warranty service onlyVISA, MasterCard or American Express credit card or company purchase order number. Billing and/or return shipping addresses. Use one of the methods below to obtain return authorization, service pricing and shipping instructions.

International customers
Contact your local, factory-authorized distributor from whom the product was purchased. To obtain the name of the local factory-authorized distributor, contact us via e-mail at service@quest-technologies.com,via telephone at +(1)-262-567-9157 or via fax at +(1) 262-567-4047.

United States customers only



Go to the service section of our web site at www.quest-technologies.com. Contact us via e-mail at service@quest-technologies.com. Contact us at (800) 245-0779. Office hours are 8:00 a.m. to 5:00 p.m. United States Central Time.

Warranty
Quest Technologies warrants our instruments to be free from defects in materials and workmanship for one year under normal conditions of use and service. For United States customers, we will replace or repair (our option) defective instruments at no charge, excluding batteries, abuse, misuse, alterations, physical damage, or instruments previously repaired by other than Quest Technologies. Microphones, sensors, printers, and chart recorders may have shorter or longer warranty periods. This warranty states our total obligation in place of any other warranties expressed or implied. Our warranty does not include any liability or obligation directly resulting from any defective instrument or product or any associated damages, injuries, or property loss, including loss of use or measurement data. For warranty outside the United States, a minimum of one year warranty applies subject to the same limitation and exceptions as above with service provided or arranged through the authorized Quest distributor or our Quest European Service Laboratory. Foreign purchasers should contact the local Quest authorized sales agent for details. Part number 053-273B

144 QUEST POLICIES


Warranty

This page intentionally left blank

Part number 053-273B

INDEX

Index
A
ac/dc Out, connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Alternate functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 Analysis Type menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Analyzed Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 A-weighting Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Display Backlighting setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Contrast setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 Global Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Log History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 RTA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41, 44 RTA History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 RTA Tabular . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46 Session Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 SLM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 SLM History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30, 30 SLM Tabular . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Title Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Dose, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . 125 Drawing Add during session . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64 Session Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82 Drawing Pad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64, 82 Dynamic Range, Defined . . . . . . . . . . . . . . . . . . . . . . . . . 114 Dynamic range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Measurement range . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

B
Back Erase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Backlight, Global Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Bandpass analysis, controls . . . . . . . . . . . . . . . . . . . . . . . . . 28 Bandwidth, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 Battery Global Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 Indicator icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Browser, viewing in . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99

C
C-A weighting Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Calibrate, Touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Calibration, instrument . . . . . . . . . . . . . . . . . . . . . . . . . 5963 Center frequency, Defined . . . . . . . . . . . . . . . . . . . . . . . . . 113 Clock, Run time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 CNEL, defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125 Code, security . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Color dialog box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Communications, Compact flash slot . . . . . . . . . . . . . . . . . . 5 Compact flash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Connector, Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Connectors, Case location . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Criterion Level Change setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 Criterion Time Change setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113 C-weighting Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

E
Electret microphone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Email, Quest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143 Exceedance Change settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123 Percent, defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114 Exceedance display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 Exchange Rate Change setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114 ExpHrs, defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125, 125 Export session . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74, 83 ExpSec, defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125, 125 External power, Indicator icon . . . . . . . . . . . . . . . . . . . . . . 18

F
Fast response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Filter Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114 Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Free field, microphone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57

D
Date and Time ?? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94 Decibel, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114

G
Global Setup Calibration Settings . . . . . . . . . . . . . . . . . . . . . . . . 5963

Part number 053-273B

146 INDEX

Global Setup (continued) Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Go . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

H
Headset Accessory connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Connector pinout . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Help system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

Meter Profile Settings . . . . . . . . . . . . . . . . . . . . . . . . . . 8992 Microphone Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Polarizing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Mntime, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 Mxtime, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127

N
Navigation Panel Global Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Introduced . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Network, Global Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Noise floor, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116 Notes, in text . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

I
Impulse response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Indicator Panel, introduction . . . . . . . . . . . . . . . . . . . . . . . . 17 Information Global Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 Web view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Infrared, Window . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Input Panel, operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Instrument security . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9495 Instrument Setup Logging Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . 5154 SLM Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3539 Integrating threshold, Defined . . . . . . . . . . . . . . . . . . . . . . 115 Integration threshold, Enable/disable . . . . . . . . . . . . . . . . . 37

O
Octave band, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116 OL, See Overload Omnidirectional, microphone . . . . . . . . . . . . . . . . . . . . . . . 57 Online Help system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Overload % Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 Indicator icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Time measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . 127

L
Language, select . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 Lavg, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . 124 LDN, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . 126 LEP,d, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . 126 Leq, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . . 124 Level, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116 Lmax, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . 123 Lmin, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . 123 LnX, See Exceedance Log History display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54 Logged measurements, Session Manager . . . . . . . . . . . . . . 80 Logging Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116 Memory icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Selections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Lpk, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . . 124

P
Page Navigator Global Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87 Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Study Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Parameter Panel RTA Group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42 SLM Group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Pascal, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116 Passband, Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Pause Controlling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Indicator icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 PCT, Change setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 PDose, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . 126 Phone, Quest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143 PKcnt, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124 PKtime, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 Polarization potential, Enable/disable . . . . . . . . . . . . . . . . 105 Power Connector pinout . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107 Port connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Preamplifier, Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Preamplifier cable, use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Pressure microphone, use . . . . . . . . . . . . . . . . . . . . . . . . . . 57 Part number 053-273B

M
Manager Functions Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74 Measurement, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116 Measurement range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Measurements, Introduction . . . . . . . . . . . . . . . . . . . . . . . . 27 Meter, Select in SLM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Meter characteristics . . . . . . . . . . . . . . . . . . . . . . . . . . . 2628

INDEX 147

Printing reports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8485 from Web View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Standalone setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84 Profiles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8992 Locking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92 Projection Time Change setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 116

Q
Quest, Contact information . . . . . . . . . . . . . . . . . . . . . . . . 143 QuestSuite, Print from . . . . . . . . . . . . . . . . . . . . . . . . . 84 Note

R
Random incidence, microphone . . . . . . . . . . . . . . . . . . . . . 57 Range Change in SLM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Reference Sound Pressure, Defined . . . . . . . . . . . . . . . . . 117 Response, Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Response time, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . 117 RS-422 Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 RTA Introduction to group . . . . . . . . . . . . . . . . . . . . . . . . . . 14 RTA Settings shortcut . . . . . . . . . . . . . . . . . . . . . . . . . 48 RTA display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4144 Sound curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 Zoom and pan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 RTA Group, Parameter Panels . . . . . . . . . . . . . . . . . . . . . . 42 RTA History display . . . . . . . . . . . . . . . . . . . . . . . . . . . 4445 RTA Tabular display . . . . . . . . . . . . . . . . . . . . . . . . . . . 4646 Run, Indicator icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Run time Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Session Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76

S
Screen contrast, Global Setup . . . . . . . . . . . . . . . . . . . . . . . 97 Scrolling In Session Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . 75 Introduced . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Security Logon Change code . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Disable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95 Security page . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9495 Security Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9495 SEL, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . 126 Serial, Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Service warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

Session Comments field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Controlling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Delete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Final summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 Indicator icon . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Ongoing summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71 Session Manager Collapse structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 Delete items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Exceedance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78 Export session . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Logged measurements . . . . . . . . . . . . . . . . . . . . . . . . . 80 Manager Functions Bar . . . . . . . . . . . . . . . . . . . . . . . . 74 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Statistics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Summary file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76 Settings Change in SLM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117 Session Manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Setup displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Shortcut RTA Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48 SLM Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 SLM Introduction to group . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Parameter Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 SLM Settings shortcut . . . . . . . . . . . . . . . . . . . . . . . . . 36 SLM display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2530 SLM History, Back Erase . . . . . . . . . . . . . . . . . . . . . . . . . . 32 SLM History display . . . . . . . . . . . . . . . . . . . . . . . . . . . 3033 SLM Tabular display . . . . . . . . . . . . . . . . . . . . . . . . . . . 3334 Slow response . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Sound curve procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47 SoundPro DLX Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109 Speaker, Global Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96 SPL, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . . 123 SPL activity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Statistics, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Statistics display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77 Study Comments field . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Delete . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83 Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Rename . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 Run time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Running . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58 Summary file . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81

Part number 053-273B

148 INDEX

Study (continued) User name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80 Study information display Disable during session . . . . . . . . . . . . . . . . . . . . . . . . . 67 Enable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38, 63 Using . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6367 Stylus for touchscreen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Summary display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71

T
Taktm Interval, defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119 Measurement, defined . . . . . . . . . . . . . . . . . . . . . . . . 125 Taktm Interval, Change setting . . . . . . . . . . . . . . . . . . . . . . 39 Tap, explained . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Third-octave band, Defined . . . . . . . . . . . . . . . . . . . . . . . . 120 Time response, Meter parameter . . . . . . . . . . . . . . . . . . . . . 26 Touch And drag, explained . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 And hold, explained . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Touchscreen Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Tripod, use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57 TTL I/O, Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 TWA, Measurement, defined . . . . . . . . . . . . . . . . . . . . . . 126

Web viewing, setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99 Website, Quest . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143 Weighting Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Meter parameter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Windscreen, Defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Wired Compact flash slot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Wireless Compact flash slot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98 Wireless setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

Z
Zoom and pan Detailed instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 RTA display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 SLM History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Z-weighting Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120 Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

U
UL, See Upper Limit ULL, Change setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 Under Range Percentage, defined . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 Time, defined . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127 Under range, Indictor icon . . . . . . . . . . . . . . . . . . . . . . . . . . 18 Upper Limit Time, Measurement, defined . . . . . . . . . . . . 127 UR, See Under Range USB Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

V
Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97 Voice Note Add during session . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65 Edit, add, rename . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

W
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143 Web view information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100