You are on page 1of 18

MMT -530M Revision 7 Jan.

2000 Page 1

Site Plannin
(CZ-70 and up)

Guide

GE 1.0T Signa Horizon Mobile MR System

@ 1999

Calumet Coach Company

All Rights Reserved

This guide contains confidential information of Calumet Coach Company. You may not copy it or any part of it without the written permission of Calumet Coach Company. This manual may be used only by you, and only for the purposes for which it was intended. You may not disclose this manual or the confidential information it contains outside your company. If you wish to copy part of this manual or to use it other than as described above, you must write to Calumet Coach Company seeking permission to do so: Keith Moore, President Calumet Coach Company 2150 E. Dolton Road Calumet City, IL 60409-1411 USA

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used. either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may Dot be reproduced or disclosed without prior consent.

MMT-530M
Revision 7 Jan. 2000 Page 2

LIST OF REVISIONS
Revisions 01 02 03 04 05 06 07 Initial Release Revised width to 98"; changed magnet centerline to 128" from rear of unit Revised width to 100" Revised magnet centerline for units 47A and up Added wireless LAN antenna location Added grounding requirements; added phone line for data link Revised generator housing

Date
04-1996 07-1996 02-1997 06-1997 05-1998 05-1999 01-2000

Notice

In accordance with our policy of continued product improvement, Calumet Coach Company reserves the right to make changes in the equipment, design, specifications and materials of the product described herein. Any problems or questions related to the components or systems covered in this booklet may be directed to: Calumet Coach Company Service Department 2150 E. Dolton Road Calumet City, Illinois 60409-1411 USA Telephone: 800-839-0630 (24/7 Service) Fax: 708-868-5101
http://www.calumetcoach.com

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

MMT-530M Revision 7 Jan. 2000 Page 3

Table of Contents
Section 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. Support Pad Requirements Customer Power Requirements Mobile Grounding Requirements Telephone Service Water Requirements Figure 1 - Medical System Figure 2 - Pad Layout Figure 3 - Curb Side Elevation Figure 4 - RussellStoll Service Outlet Figure 5 - RussellStoll Chart Figure 6 - Mobile Grounding Requirements Chart Figure 7 - Turning Requirements Figure 8 - Stair and Platform Detail Appendix: Photographic References Page 4 6 8 9 10 11 12 13 14 15 16 17 18 Al

The information contained in this booklet is property of Calumet Coach Company. The contents are considered not be used. either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. reproduced or disclosed without prior consent.

to be confidential and may The contents may not be

MMT -530M Revision 7 Jan. 2000 Page 4

Support Pad Requirements


GE 1.0T Signa Horizon Mobile MR System
The following is a list of recommendations and requirements for a concrete support pad. However, due to varying site conditions, the actual pad design should be prepared by an appropriately licensed structural or architectural engineer.

Recommended Support Pad A full pad measuring10'-0" x 40'-0", locatedas shownon Figure2 and Figure3 as the crosshatching,s the i
recommended support pad.

Recommended Service Pad A full pad measuring 21' -0" x 55'-6", located as shown on Figure 2 and Figure 3 is recommended to provide a service access.

Support Pad Depth Recommendations for the width and length of the pad are given above. Based upon the weight distribution information given in Figure 3 and existing site conditions, the depth should be determined by a local contractor. It is recommended that non-ferrous reinforcement materials be used for pad reinforcement. GE must approve plans for pad construction. Support Pad Levelness The support pad must be level to ensure proper operation of the MRI system. The pad must not exceed .125" deviation in 10'-0". If the minimum support pads are selected, rather then the recommended single pad, they must also meet this specification. Vehicle Access A firm, level surface is required around the mobile unit to provide access to the tractor/trailer, aid with patient handling, servicing the unit, and delivering of cryogens.

Steel Reinforced Concrete Pad


A maximum of 4 lbs per square foot of steel is allowed in the concrete pads beneath the mobile unit.

---

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. reproduced or disclosed without prior consent.

to be confidential and may The contents may not be

MMT-S30M Revision 7 Jan. 2000 Page 5

Support Pad Requirements


GE 1.0T Signa Horizon Mobile MR System
Minimum Support Pad Reauirements
A front pad measuring 10'-0" x 4'-0" and a rear pad measuring 10'-0" x 15'-0" located as shown on Figure 2
will provide the minimum support requirements. Recommended Attachment to Facilitv An inflatable air bag or soft seal is recommended at the point of connection from the unit to the facility. Fixed or solid connections may hinder imaging quality. Contact Calumet Coach Company or the local GE representative prior to construction if the proposed connection varies from the recommended. Exclusion Zone An area of 5'-0" x 5'-0", located directly below the magnet vent should be fenced off to prevent injury in the event of magnet quenches. Vehicle Movement The MRI is very sensitive to vibration and moving metal. Consequently, all vehicle traffic shall be kept as far away as possible from the pad. Moving ferrous materials having the listed masses should be limited to areas as described in the GE site planning publication. Contact GE to obtain the latest version. Vibration I Foundation DesiQn It is essential to minimize vibration to ensure high resolution Image quality. To achieve this, GE has produced a scope of work, specifications, and test methods for new and existing sites. GE in their Signa Horizon site planning publication provides this documentation. Contact GE to obtain the latest version.

SwinQ Clearance Note


Please verify the actual dimensions of the rearmost projections on the cab of your tractor to the centerline of tandem suspension or centerline of the fifth wheel plate on your tractor. Refer to Figure 7 for proper tractor sizing information.

The information contained in this booklet is property of Calumet Coach Company. The contents are considered not be used, either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. reproduced or disclosed without prior consent.

to be confidential and may The contents may not be

MMT-530M Revision 7 Jan. 2000 Page 6

Customer Power Requirements


GE 1.0T Signa Horizon Mobile MR System
Electrical Service
A single electrical power source is required for operation of the MRI system. 480 Volt A.C., 3 Phase, 125 KVA, fused at 150 Amps

Load Reaulation at Line Freauencv Wiresare to be sizedsuchthat the line voltagedropsfromthe powersourceto the mobileunit is less then2% of thenominalvoltageforratedloadof themobileunit. Confiauration
Three phase, five wire, wye connection, with ground and neutral (not less than AWG #1/0).

Freauencv
60 Hz j: 0.5% Phase Balance: + 2% maximum phase-to-phase line voltage difference lowest phase. Maximum voltage variation: j: 2% from nominal steady state (under the worst case conditions of line voltage) Maximum voltage range: 420 volts - 500 volts Source impedance: 480 volt line to line impedance <= 0.1 ohms

. . . . .

Connector Tvpe
The unit is supplied with a 35-foot (usable) cable and male connector. The connector is a RussellSto1l200 Amp plug #DS2504MP.

Customer Facilitv
The facility must have the matching receptacle as specified in Figure 4 and Figure 5. The receptacle is a RussellStoll #DF2504FRAB.

Voltaae Suraes
Transient voltage variations caused by external loads must not:

. .
.

Exceed + 10% exceed five cycles duration Occur more than ten times an hour.

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may Dot be used, either partiaIly or whoIly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

--

MMT-530M Revision 7 Jan. 2000 Page 7

Customer Power Requirements


GE 1.0T Signa Horizon Mobile MR System
Ground Conductor An insulated ground conductor sized in accordance with National, State, and local codes shall be installed between the facility vault and the MRI System ground bus location in the power distribution unit. Exclusion Zone The MRI unit is equipped with magnetic shielding. The exclusion zone for cardiac pacemakers, neurostimulators, and other biostimulation devices is recommended at 5 gauss (0.5mT). Signs, provided by GE, must be posted to alert all who approach the unit of this requirement. The l.OT magnet systems exclusion zone does not extend into normally accessible locations that are exterior to the mobile unit. The appropriate warning signs are permanently attached to the scan room doors. The 3 gauss line (0.3mT) is located approximately 3 feet from the unit walls. The 1 gauss line (O.lmT) is located approximately 8 feet from the van walls. The 0.5 gauss line (0.05mT) is located approximately 12 feet from the van walls.

R.F. Shieldina The R.F. shieldis includedwiththe MRI systemand willprovideminimumlevelof attenuation:10Mhz - 100 Mhz,85 db (planewave,electricfield,magneticwave).
Note: The standard connector for the unit is a RussellStoll #DS2504MP. If an existing site currently implements a different connector, connector configuration, or the available power supply varies from above specifications, please contact Calumet Coach Company to arrange for a compatible power connector before the unit leaves the facility.

WARNING:
It is the operator's responsibility to verify that the shore power receptacle is electrically compatible with the mobile system's umbilical connection prior to plugging in. Plugging into a receptacle not electrically compatible could cause injury or damage. It is also recommended to inspect the power cord, connector, and fasteners prior to connection. If during inspection you suspect internal or external damage has occurred, have a qualified electrician inspect and repair the damage before further use. Follow the maintenance schedule in the Operator and Maintenance Manual for safe operation of the mobile unit.

The information contained in this booklet is property of Calumet Coach Company. The contents are considered not be used. either panially or wholly, for any purpose inconsistent with the purpose for which it was produced. reproduced or disclosed without prior consent.

to be confidential and may The contents may not be

---

MMT -530M Revision 7 Jan. 2000 Page 8

Mobile Grounding Requirements


GE 1.0T Signa Horizon Mobile MR System
NOTE: For the associated drawinQ please refer to FiQure 6

. . . .

All work is to be done in accordance with local and national electrical codes. Information shown here is only a recommendation and must be verified local site codes and national site codes. Ground wires inside enclosures are to be taped green for the entire visual length for identification. If a separately derived secondary system transformer is used, a bonding jumper between the grounded conductor (neutral) and the equipment - grounding conductor must be used.

GroundinQ The ground for our system shall originate at the system power source (transformer) or first access point of power into the facility, and be continuous to our system power disconnect in the room. This ground can be spliced with "High Compression Fittings" and should be terminated at each distribution panel it passes through. When it is broken for a connection to a panel, it shall be connected into an approved grounding block with the incoming and outgoing ground in the same grounding block. This grounding block is then connected to the steel panel. Never use the steel or any other material of the panel as the grounding block. The connection at the power source shall be at the grounding point of the "Neutral Ground" if a "WYE" transformer is being used, or at the typical grounding points of a separately derived system. In the case of an external facility, it shall be bonded to the facility ground point at the service entrance.

Ground Wire
The ground wire shall be copper wire and the same size as the disconnect feeders. The ground wire impedance from our system disconnect, including the ground rod, shall not have an impedance greater than 2 ohms to earth as measured by one of the applicable techniques described in Section 4 of ANSIlIEEE Standard 142-1982.

Special GroundinQ Note: The unit must have an earth driven groundrod within five (5) feet of the hospitablepower receptacle. A
grounding cable of a minimum 4 GA must be connected between the grounding rod and the grounding pin of the hospital power receptacle, and another cable to be kept as short as possible, and must not exceed 8 feet in length. A separate grounding conductor must still be run with the phase conductors to the source of power from the grounding pin of the hospital power receptacle in accordance with NEe. Article 250-23.

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered not be used, either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. reproduced or disclosed without prior consent.

to be confidential and may The contents may not be

MMT-530M Revision 7 Jan. 2000 Page 9

Telephone Service Requirements


GE 1.0T Signa Horizon Mobile MR System
Telephone Service

. . . .

The unit is supplied with three telephone connections and three data line connections. The connector type that is used is a model Hubbell PH-6595 (inlet), supplied by Calumet Coach. Two Hubbell PH-6599 50 foot telephone-connecting cables are included with the unit. The data connections utilize a 50-foot UTP CAT 5 Cable with an RJ-45 connector. The customer is also required to supply three Hubbell phone connectors, model PH-6597 (weatherproof phone outlets) at the site.

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used, either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

MMT-530M Revision 7 Jan. 2000 Page 10

Water Requirements
GE 1.0T Signa Horizon Mobile MR System
During winter conditions provisions must be made to ensure that any and all water lines do not freeze a result of weather conditions.

Humidifier Water Fill


The unit contains a water storage tank for the humidifier. This tank is located in the equipment room and must always contain water to insure the specified humidity level remains constant. There are two options for filling the tank: . A %" I.P.S. male threaded hose connection is located under the equipment room.

If a water supply hose is not available, the humidifier storage tank may be filled manually via a port located on the top of the tank.

Potable Water Supplv Reauirements (option)


A cold water supply line is required, with a flow rate of 5 gallons per minute, 45-60 psi and a maximum temperature of 70 degrees Fahrenheit. The unit will be supplied with a %" diameter, 20' long hose terminated with a %" I.P.S. male threaded hose connector. The facility is to provide a W' female connector to connect to the units 20' long hose.

Waste Water Connections (sink option)


A 1Y2"IPS male connection is required to sink wastewater.

Both the water supply and the wastewater drain connections are to be contained in a recessed water connection box located as shown in Figure 1. Any pipes and drains within 20'-0" of the magnet isocenter should be of non-ferrous materials such as PVC, copper, or brass. All plumbing must comply with applicable codes. The unit is supplied with a 20' long 1W' diameter drain hose terminated with a 1W' male threaded connector for sanitary wastewater drainage. The facility must provide means of sanitary wastewater drainage from the system that comply with locally applicable codes.

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

----

~ ~~~!!
270.
128. 1&8'

MMT-530M Revision 7 Jan. 2000 Page 11


12;'~2'

OPTIONAL LA. 3"

~e'"
EXHAUST

4' 4' MY "'tN~OR I :--2~ ~G.\CYMEDIAOF'TION 18.... v ,2.....

,'-..

SEIMCE DOOR

rAN

'00'

hmT

L PATIENTLOWER EXCHANCER L COL HAT B

=
62
36"

.r_.l-~

~3
,"01:;

~~z ~d
1'\

h
I~

HVAC UNIT 72,000 BTU

208 PANEl volT

VARE DETEC~ON PANEL I rWATER


TANK

RO<.J oooR

" I

....

..
4'-

H y

HVAC UNIT 60,000 BTU

~UIi4IDIF'IE

PATIENT n wrTH u SAF'ETY HAND RAILS

1---6'
66 1/2'

VENT DEFUCTOR

~
3';"&.

~
rPAcI\PM~CER

r~
2'

LJ I'
12-

90' 9&'

123-

36.,& 1/2't---

,"

2"
1 AI. Q!,!!;L..

7"

7~

d'/4'

'11 ~

~-

r Duct

10:
~

AIR OUCT

II1UI

HVAC

UNIT

96 1 8.1-"'RI o.ERGENCY STOF

'AN '~
~~
~ Of IUGNI

U~I~
I
I 108. 128. f-2&'

.,.

*
~

;:::::j:
!-<CY

MIt>'"

'TI

I~ ~~ ~~
t=
/

16/1 ,NET

~ 5 "," i~ ~~~ !

/,,20

TMK

gl:; ~ ~~ u~

r---

'---

~
I I

~
13&'

GENERATOR

IANDCJ 208 VOlTPANEL

nRE DETEcnoN PANEL

f--tiUUIOIFlEI

KING PIN It

,.'

[F,

It. LANDING LEGS

EST. WEIGHT

2215801

(roR KINGPIN)

32'---I1-4'
48

117 3/".

48'-0"

(iWR

:tiH~ENTE~:.~l)

Fiaure 1: GE 1.0T Siana Horizon Mobile _MR System


The information contained in this booklet is property of Calumet Coach Company, The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

~ ~~~!!
55'-6'
FULL SERvICE PAD

MMT-530M Revision 7 Jan. 2000 Page 12

6'
MINIMUM

IS'
SUPPORT PAD

21'

4' MINIMUM SUPPOI PAD APPRQXIWATE LOCATION 'tt:a:CAl/TEI AND G.E, EXTERN.t.l. OPTfHS INTERfACE P.

9'-6"

4'-6"

I
I 8'-4'

10' AXL SPACING

C.l.jMAGNET 10'-0'

1'-141/2"

3'1

lAIC

I
10'

IA/C
APPRO~.
. Of' MAG. VENTING

21'

6'-6"

I~~J
P"t'RAMID STANDS

,tPPAOXIMATELY e'-o.

IN FRONTOF'

THE

Ale

~~ ~NT
CAAQE

UNITS F"QR SERVICE NTER THE UNIT IS IN PLACE.

CEN. UN[ OPENING

TO

m...

APPoo"'

,[1.Y "-0"X20'-0"CLAR rOR NELIUNFlLl ACCESS

0 SERVICE OF IHUMIDIFIER FILL

CLIP-ON

STAIRS

NOTE: PLATFORM, WHEN REQUIRED, SHALL BE A SITE SPECIFIC OPTION. WHEN USED,THE PLATFORM MUST BE MECHANICALLY ISOLATED FROM THE TRAILER,

3'-8"

7'-6"

10'-7"

.25'-4
CZ 70 AND UP

1/2"

MININUM ClAR OPENING REQUIRED FDA F'ACIUT'I' UNI(

2'-4

1/2'

10'-3"

Figure 2: Pad Layout


The information contained in this booklet is property of Calumet Coach Company, The contents are considered to be confidential and may not be used, either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

~ ~~~!!
1'-4 1/24S'

MMT-530M Revision 7 Jan. 2000 Page 13

10'-8'

j
I

C.L/w.GNET
7'-3
VENT DEFlECTOR SERVICE

1/2'

5'-71

HATCh-"""""""

J
HANDRAILS

CONTACT GE HEN..THNET SVCS FOR OPTIONAL WIRELESS LAN COt.lMUNICATION INSTALLATION

OF WAlL MOUNTED

COMPONENTS AT HOSPITAL SITE

2'-6-

..5"RECESS

FRO'"

FACE/TRAILER

13'-6"

FLOOR

LEVEL

I I I I I I
L
PATIENT Uf'T

I I
AIC

3'l

GEN.

7'-6"

I
C.l.
KINGPIN

.'

g' ,L./PYRANIO STANDS 10'


AXLE SPACIN'

4" 2 1/2" '-S"


FRAMED

S'-2"
OPENING

10'-3"
UNK

MINIMUM CLEAR

Jp~',.;~

~6~ACIUTY

9'-6 3/".

I
EST. REAR WT. 34000 LBS,

C.L/LANDING

GEAR

I TOTALWEIGHT APPROXIMATElY 62500


LBS. EST. FRONT WT. 24000 LBS.

Fiaure 3: Curbside Elevation


The infc)fJnation contained in this booklet is property of Calumet Coach Company, The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced, The contents may not be reproduced or disclosed without prior consent.

~ ~~~!!
SAFETY DISCONNECT SWITCH
ON

MMT-530M Revision 7 Jan. 2000 Page 14

150 AMPERE FUSING FRS-R-100 OR EQUAL.

OFF

RECEPTACLE: #DF2504FRAB W/SPRING DOOR

HEIGHT TO BE GOVERNED BY NATIONAL AND LOCAL CODES


GROUND ROD FINISHED GRADE

480 VOLT 30, 60Hz ONLY

CUSTOMER MUST CLEARLY LABEL

~~~~~~
3 PHASES + GROUNDAC 480V 150A SERVICE WARNING: HOSPITALRECEPTACLE WIRINGAND MOBILE UNIT PLUG WIRING MUST BE INITIALLYCHECKEDFOR PROPER CONTINUITYBY A QUALIFIEDELECTRICIAN

.lliITL.
THE INFORMATION CONTAINEDIN THIS DIAGRAMIS RECOMMENDED TO MEET THE REQUIREMENTS THE EQUIPMENTOPERATEDAND IS OF SUBJECT TO LOCAL CODE APPROVAL.

Fiaure 4: RusseliStoll Service Outlet

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used. either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

~ ~~~!!
3C/N/G AC 480V 60 Hz RECEPTACLE
MAIN DISCONNECT 3C/N/G AC 480V 150 AMPERE

MMT-530M Revision7 Jan,2000 Page 15

NEUTRAL-WHITE

PHASE 'C'-BLUE(208/240V),

BROWN(480V)

o
PHASE 'A' -BLACK(208/240V), YELLOW( 480V)

GROUND - GREEN

PHASE 'B'-RED(208/240V),

ORANGE(480V)

VIEW FROM BACK SIDE OF RECEPTACLE INSTALLATIONBY CUSTOMER

WIRING CHART AMPS/WIRE DESCRIPTION WIRES POLES 5 MAXIMUMWIRE SIZE AS REQUIREDBY LOCAL CODES ABBREVIATIONS:
3C-PHASE CONDUcroRS N - NEUTRAL G-GROUND

RECEPTACLE PART NUMBER 480 VOLT (200 AMP) 5 WIRE RUSSELLSTOLL RECEPTACLE #DF 2504 FRAB THIS RECEPTACLE MUST BE WATERPROOF

480 VOLT AC

3 WYE CONNECTION150 AMP PHASE TOTAL 3 <YNEUTRAL & GROUND

Fiaure 5: RusseliStoll Chart

The infonnation the purpose

contained

in this booklet

is property of Calumet The contents

Coach Company.

The contents without

are considered prior consent.

to be confidential

and may not be used, either partially

or wholly,

for any purpose

inconsistent

with

for which it was produced.

may not be reproduced

or disclosed

~ ~~~!!:
I
ul

MMT-530M Revision 7 Jan. 2000 Page 16

GROUND BUS BAR

r
ELECTRICAL IPING P

FEEDER
Ll

[!>
FROM FACILITY SERVICE EQUIPMENT

I
I

I I I I

l
DISCONNECT LOCATED AT VAN CONNECTON SEE ELECTRICAL REQUIREMENTS FOR AMP. RATINGS
L1

L2 L3 IN
I

I
I

f
I I
I

L2

r I

RUSSELLSTOLL DS2S04FP FEMALE

RECEPTACLE

OREQUIVALENT

I I
#1/0 AWG COPPER I MIN.

G
SERVICE EARTH MAIN GROUNDING AT FACILITY SERVICE

L ---------------------

TO GROUND ROD AT RECEPTACLE

Fiaure 6: Mobile Groundina Reauirements Chart

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used. either partially or wholly. for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

~ ~~~!!
PATH

MMT-530M Revision 7 Jan. 2000 Page 17

OF OVERHANG

PATH OF RIGHT FRONT I./HEEL./

--C.L SUSPENS ION

// /

/1 I
\1\

50'-0'
OUTER
TO CENTER RADIUS POINT OF

27'-0'

RIGHT FRONT

I./HEEL

~
PATH OF LEFT REAR SIDE

1\\
1 \"v

I "/~
.........

I I I I I I
r
1

----------------------------ISO' S'-O'

-------------------------------

S'-O'

Fiaure 7: Turnina Reauirements


A minimum "A" dimension of 92' is required from rearmost projection to centerline of tandem suspension. This provides swing clearance for generator set which is mounted on the front of the trailer. Hospital is responsible to ensure the access route is clear of obstructions when the trailer is scheduled to arrive or depart. The 50' minimum outside turning radius shown here has been calculated using an international harvester (Navistar) tractor Model COF-9670 with a 161" wheelbase. Turning radius will vary with towing tractor. Customer must confirm the turning radius on their tractor and prepare each site with adequate space to accommodate it.

The information contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.

~ eAh~~!!
3'-0a.

MMT-530M Revision 7 Jan,2000 Page 18

r
r-4'-4i!' _0.
9!' TYP.

0i!T10NAL

STAIR

LOCATlDN
'

-o...o-.-o-.-o-.-o-o o
0000000000.0'.0.0.0.0."

3'-O!'

I L

J.

~4'-O~'

1'-4i!'

TRAILER

FLOOR

6i!'

STAIR

ANO

PLATFORM

ASSEMBLY

STAIR

WITHOUT

PLATFORM

ASSEMBLY

Fiaure 8: Stair and Platform Detail

The infonnation contained in this booklet is property of Calumet Coach Company. The contents are considered to be confidential and may not be used, either partially or wholly, for any purpose inconsistent with the purpose for which it was produced. The contents may not be reproduced or disclosed without prior consent.