You are on page 1of 3

24/01/2011

012345
^0

www.gmecorp-usa.com/A/CP/SW/TM/S

GE Medical Systems
X-Ray Service Engineering
Senographe 600T Senix HF Help System
____________________________________
1
2
3
4
5

Seno Term Info


Error code detail
Board location map
Chassis location
Board LED Status

^$
^1

SENO TERM FUNCTIONS


___________________

ALT+h -- Gives you access SENO HELP


Ctrl+C -- Transfers control to Seno diagnostics
Ctrl+D -- Transfers control to Seno front panel
ALT+q -- Terminates SENO TERM

^$
^2

E10
E20
E30
E40
E50
E80
E90
E100
E110
E120
E130
E140
E150
E160
E170
E171
E190
^$
^3

ERROR CODES
___________
---------------------------------------------------------------------

GENERATOR CONTROL FAULT


INTERFACE BOARD FAULT
CPU FAULT
PM BOARD FAULT
BUCKY TIMEOUT
CPU FAULT
GENERATOR EXPOSURE INHIBIT
FILAMENT BOOST FAULT
ANODE STARTER FAULT
TUBE ARCING FAULT
POWER FAULT - INVERTER VOLTAGE ABSENT
CONVERTER INTERLOCK FAULT
FILAMENT CURRENT FAULT - .1 FS
FILAMENT CURRENT FAULT - .3 FS
REALTIME CLOCK ERROR
EXPOSURE TIMEOUT
POWER LINE ERROR

(400-PL7)
(400-PL14)
(400-PL1)
(400-PL14)

(400-PL3) & CONVERTER


(400-PL4)
(400-PL14)

BOARD LOCATIONS
_______________

gmecorp-usa.com/A/CP/SW//HELP.TXT

1/3

24/01/2011

www.gmecorp-usa.com/A/CP/SW/TM/S

LOC
DESCRIPTION
PART #
-------------------------------------PL1 --- HEATER CURRENT
45203227
400 CHASSIS FRONT VIEW
PL2 --- HEATER SUPPLY
45202164
PL3 --- AC SUPPLY CONTROL
C800047
_________________________
PL4 --- CONVERTER CONTROL
45203229
|PPPP
PP
P
P |
PL5 --|LLLL
LL
L
L |
PL6 --|1234
78
1
1 |
PL7 --- GENERATOR INTERFACE
C800038
|||
||
1
4 |
PL8 --- X-RAY ROOM INTERFACE 45203670
|||
||
|
| |
PL9 --|||
||
|
| |
PL10 -|||
||
|
| |
PL11 -- PHOTOMULTIPLIER
45070685
|||
||
|
| |
PL12 -|_________________________|
PL13 -PL14 -- CPU BOARD
C809727
^$
^4
___REAR_____
CHASSIS LOCATIONS
____FRONT____
|_____
|
_________________
| __________ |
||
|
|
||
| |
||500 |
|
||
500
| |
||____|
|
100 -- CONTROL PANEL INT.
||__________| |
|___________ |
200 -- POWER SUPPLY CHASSIS
|
|
||
702
||
300 -- CONVERTER
|
|
||
||
400 -- MAIN CARD RACK
|
|
||__
||
500 -- ELECTROMECHANICAL
|
|
|
|_______||
600 -- ROTOR CONTROL
|
|
|____________|
| ___________ |
||
||
||
||
||
400
||
||
400
||
||__________||
||___________||
| __________ |
| ___________ |
||
300
||
||
300
||
||__________||
||___________||
||
200
||
||
200
||
||__________||
||___________||
||600 |
|
|
|
||____|______|
|_____________|
^$
^5
Board Led States
________________
System in 2 Pt. (Non Photo timed mode)
Ready for first trigger (prep)
Power Supply Chassis: The 4 neon indicators and 13 green LED's are lit.
Converter Chassis

: The red front panel indicator must be lit. This


indicates converter capacitor bank charged.

Filament Board (Pl1): Green De2 LED


Yellow De3 LED
Yellow De4 LED
Green De5 LED
Heater Supply

:Lit
:Lit
:off
:off

(Pl2): L1 neon indicator :off


Green De1 LED
:off
Green De2 LED
:off

gmecorp-usa.com/A/CP/SW//HELP.TXT

2/3

24/01/2011

www.gmecorp-usa.com/A/CP/SW/TM/S
Green De3 LED
:lit

^PAUSE
Line Supply

(PL3): Yellow De1 LED


Yellow De2 LED
Red De3 LED
Yellow De4 LED

:lit
:lit
:off
:lit

Inverter Comm. (PL4): Green De1 LED


Yellow De2 LED
Yellow De3 LED
Yellow De4 LED
Yellow De5 LED
Red De6 LED
Red De7 LED
Red De8 LED
Green De9 LED
Green De10 LED

:lit
:off
:off
:off
:lit
:off
:off
:off
:lit
:lit

Generator

(PL7): No LED's

^PAUSE
Room Interface (PL8): Green De1 LED
Green De2 LED
Green De3 LED
Green De4 LED
Green De5 LED
Yellow De6 LED
Yellow De7 LED
Green De8 LED
Green De9 LED
Green De10 LED

:off
:off
:off
:off
:lit(Al/Mo filter,diaphram present)
:off
:off
:off
:off
:off

Kv Ma Board (on HT ): LED's D51,53 - 58 :lit


Control Pan.(Pl6)

^$

: Green D8 LED
Green 24v LED
Yellow De1 LED
Yellow Drx LED
Yellow De2 TX LED

:lit
:lit
:off
:lit
:lit

^$

gmecorp-usa.com/A/CP/SW//HELP.TXT