You are on page 1of 4

Calibration Report

Job number: Customer IND201100148 Report number: IND200900105/4 Date: 06/07/2011 Pg 1 of 2

Mr. Concrete Kodigamam road, Puloly South Point Pedro. Digital Caliper Make Serial no. /10 no. Range Division

Unit under Test

Precise Not Available / MT031 0-150 mm 0.01 mm

Method of test Test was carried out refered to the international Refered to EA-4/02, NABL 141

standard ISO 6906 : 1984. Uncertainty

calculations

were done

Standard Laboratory working standard Grade 1 gauge blocks set no 090013 traceable to NIST via 821/276375 PTB Via 4937 PTB 06 Temperature and Humidity meter: SN:1 0745936 Place of test performed : SGS Lanka Calibration Lab

- 08,

Results of Calibration

Temperature

during test

Note: The calliper reading was checked at three different positions of the measuring and the base.

jaw,viz. tip, mid portion,

Gauge setting mm Zero setting 20.00 40.00 50.00 70.00 100.00 120.00 150.00 20.00 40.00 50.00 Tip

Caliper reading, mm Mid portion 0.00 20.00 40.00 50.01 20.01 40.00 50.00 70.00 100.00 120.00 150.00 Base

69.99
99.99 119.99 150.00

69.99
99.99 120.00 150.00

(Company

No. PV 19226)

SGS Lanka 1M Ltd

Industrial Services, AEC Building, 140, Vauxhall Street, Colombo, Head Office, 3< Floor, 140, Vauxhall Street, Colombo,

2, Sri Lanka, t + 94 (0) 11 5395118 f + 94 (0) 11 4717878 e sgs.lanka@sgs.com


t + 94 (0) 11 5376280 f + 94 (0) 11 5376281

www.sgs.com

2, Sri Lanka,

e sgs.lanka@sgs.com

www.sgs.com
Member of the SGS Group

Measurement Report
Job number: IND201100148 Report number: IND2011000143/3 Date:
06/0712011

Pg 1 of 2

Customer

Mr. Concrete Kodigamam road, Puloly South Point Pedro.

Unit under Test

Steel Rule Make - Elora Serial no./ID no. : -/MT037

Range Division

0- 1000 m
1 mm

Method of test The accuracy of this instrument was checked in accordance with Linear measurements were made by Linear Scale. Standard This certificate provides traceability of measurement to national standards through Industrial Technology Institute Sri Lanka Place of test performed: Results of Calibration Temperature during test Zero Error: Condition room at the SGS Lanka Pvt. Ltd.

NIST Japan and

20 C

Present or Zero line missing Measured Value 1 (mm)

Range 1 (mm)

Measured Value 1 (mm)

Range 1 (mm)

0-50 0-100 0-150 0-200 0-250 0-300 0-350 0-400 0-450 0-500

50 100 150 200 250 300 350 400 450 499

0-550 0-600 0-650 0-700 0-750 0-800 0-850 0-900 0-950 0-1000

549 599 649 699 749 799 849 899 949 998
(Company No. PV 19226)

SCS Lanka IPvtI Uti

Industrial Services, AEC Building, 140, Vauxhall Street, Colombo - 2, Sri Lanka, t + 94 (0) 11 5395118 f + 94 (0) 114717878
Head Office, 3" Floor, 140, Vauxhall Street, Colombo - 2, Sri Lanka, t + 94 (0) 11 5376280 f + 94 (0) 115376281

e sgs.lanka@sgs.com

WWW.sgs.com
Member of the SGS Group

e sgs.lanka@sgs.com

www.sgs.com

Measurement Report
Job number: IND2011 00148 Report number: IND2011000143/2 Date: 06/07/2011 Pg 1 of 2 Customer Mr. Concrete Kodigamam road, Puloly South Point Pedro. Steel Tape Make -Jason Serial no.llD no. : -/MT035

Unit under Test

Range Division

0-7.5 m 1 mm

Method of test The accuracy of this instrument was checked in accordance with Linear measurements were made by Linear Scale. Standard This certificate provides traceability of measurement to national standards through Industrial Technology Institute Sri Lanka Place of test performed: Results of Calibration Temperature during test Zero Error: Measured Value 1 (mm) 100 200 300 400 500 600 700 800 900 1000 none 20 Condition room at the SGS Lanka Pvt. Ltd.

NIST Japan and

DC

Range 1 (mm) 0-100 0-200 0-300 0-400 0-500 0-600 0-700 0-800 0-900 0-1000

Rangel (m) 1 - 1.100


1 - 1.200 1 - 1.300

Measured Value I (mm) 100 200 300 400 500 600 700 800 900 1000

Range I (m) 2-3 3-4


4-5

Measured Value 1 (m) 1.000 1.000 1.000 1.000 1.000 0.500

1 - 1.400 1 - 1.500 1 - 1.600 1 - 1.700 1 -1.800 1 -1.900 1 - 2.000

5-6
6-7 7 - 7.5

Width ofthe Tape: 24 mm


(Company No. PV 19226)

SGS Lanka [Pvtl Ltd

Industrial Services, AEC Building, 140, Vauxhall Street, Colombo - 2, Sri Lanka, Head Office, 3" Floor, 140, Vauxhall Street, Colombo - 2, Sri Lanka,

t + 94 (0) 11 5395118

f + 94 (0) 11 4717878

e sgs,lanka@sgs,com

www.sgs.com

t + 94 (0) 11 5376280

f + 94 (0) 11 5376281

e sgs.lanka@sgs,com

www.sgs.com
Member

of the SGS Group

Measurement Report
Job number: IND201100148 Report number: IND2011000143/1 Date: 06/07/2011 Pg 1 of 3

Customer

Mr. Concrete Kodigamam road, Puloly South Point Pedro.

Unit under Test

Fiberglass Tape Make - TSB - 50 M Serial no./ID no. : -/MT034

Range Division

0- 50

2mm

Method of test I The accuracy of this irhl strument was checked in accordance with Linear measurements were made by Linear Scale. Weight is applide only to remove wrinkles Standard This certificate provides traceability of measurement to national standards through Industrial Technology Institute Sri Lanka Place of test performed: Results of Calibration Temperature during test Zero Error: Range 1 (mm) Value 1 (mm) none Range 1 (m)
1 - 1.100 1 - 1.200 1 - 1.300 1 - 1.400 1 - 1.500 1 - 1600 1 - 1.700 1 - 1.800 1 - 1.900 1 - 2.000

NIST Japan and

Condition room at the SGS Lanka Pvt. Ltd.

20 C

Measured Value 1 (mm)

0-10 0-20 0-30 0-40 0-50 0-60 0-70 0-80 0-90 0-100

100 200 300 400 499 599 699 799 899 998

100 200 300 400 500 600 700 800 900 1000
(Company No. PV 19226)

SGS Lanka 1M) Ltd

2 S . L nka t + 94 (0) 11 5395118 I + 94 (0) 11 4717878 e sgs.lanka@sgs.com www.sgs.com I I n t' IS' d es AEC Building 140 Vauxhall Street, Coom b0 - , n a , H uds rOlaffiCee~~c 140 vau~all s;reet, Colombo _2. Sri Lanka, t + 94 (0) 11 5376280 f + 94 (0) 11 5376281 e sgs.lanka@s9s.comwww.sgs.com. FI~or ea , " Member of Ire SGS

Group