You are on page 1of 14

MEGR 6181

ENGINEERING METROLOGY

ANGULAR MEASUREMENTS

RAJESH PATEL
1. INTRODUCTION:

The angle is the arithmetic difference between two directions. The term angle is used to refer the
angular separation of intersecting lines, or to the angular spacing of points on the circumference
of a circle. It is also refer as a deviation from a reference line or plane or other datum, or to the
twist of a shaft subjected to torque. Angular spacing is frequently applied in engineering to
specify conditions, which result from the division of circle. Frequently the features which have to
be located lie on the concentric circles so that the dual parameters of radial distance and spacing
angles.

These different interpretations of angle imply different geometrical situations, which will govern
the method of measurement, both regards to the reference datum and choice of instrument.
Angular measurement requires no absolute reference standard since the 360 circle may be
accurately subdivided by testing the equality of subdivisions. Any unit of angle is thus a
geometrical concept and is not derived from material concept. Certain angle standards are
required for practical applications.

2.UNITS:

Radian is the basic unit of angle and is derived from geometrical consideration. Radian is
defined as the angle subtended at the center of a circle by an arc equal in length to the
radius. The round angle is there fore 2 π radian. The radian unit is actually unsuited to
metrological, since 2 π is not a rational number, and therefore any whole number divisor of a full
circle will necessarily also not be rational. For technical purposes, the 360 th division of the full
circle is used instead, which bears the unit degree ( °). There is no primary artifact standard for
the angle: it is defined in terms of the full circle. Angle metrology therefore is reduced to the
quest of dividing the round angle as equally as possible. . There are two alternative systems are
used for angular measurement; (a) Sexagesimal system and (b) Centesimal system.

In Sexagesimal system the right is divided into 90 degrees and the each degree is further
subdivided into 60 minutes and each minute into 60 seconds. Where as in Centesimal the right
angle is divided into 100 grades with a further subdivision of the grade into 100 minutes and the
minute into 100 seconds. Other than these two systems some trigonometric functions such as
tangent, sine or cosine are used to define angular magnitude.

3.ANGLE MEASURING TECHNIQUES:

3.1.FIXED TAPER GAGES:

Fixed taper gages are the direct checking tools. In this method the part are rubbed with gages for
displaying the condition of compliance or deviation. This method does not provide any numerical
information regarding the angular deviation between the part and the master gage.
3.2. ANGLE GAGE BLOCKS:

Fig.1 Angle gage Block


source : Geneva Gage .Inc.

A reference angle of specific angular dimension can be created by combining appropriate


individual members. The purpose of the gage block is to create a positive linear distance
represented by stack’s two end faces. Angle gage blocks are assemble to produce a specific
angular separation between two free faces.

Angle blocks are usually supplied in sets. In set each individual member has particular angle and
by combining different members it is possible to produce different angle. A small number of
blocks set can produce a large variety of different angles. With use of different free edge of the
each block it is possible to create two different angles. ( as shown in fig.2 )Angle gage blocks are
come in different accuracy.

+5 -5

+ 45
50 + 45 40

Fig. 2 a. Addition & b. Subtraction of combination of gages.

Normally gage blocks are made of hardened and ground steel. They are lightly magnetized for
better holding.

Applications of angle gage block.


a. Checking the circular dividing accuracy of the rotary table and dividing heads.
b. Setting a revolving work holding table or a magnetic chuck into required tilt position
c. Inspecting and refining the setting accuracy of tilt table
d. Providing a reference angle for inspecting features on work piece.

3.3. ADJUSTABLE ANGULAR BODIES:

In order to make an angle adjustable, one of its bounding elements must be movable. To qualify
an angle as a reference body, the movement of the boundary element must be precisely
controlled, further more, any selected mutual position of the boundary element must result in an
angular separation which is positive and of accurately known magnitude.

3.3.1.SINE BAR:

These conditions are satisfied by the Sine bar, when it is used in conjunction with flat plate and
gage block. Sine bar works on basic principle of trigonometry. The sine bar consists of a
hardened and ground bar to which two rollers are attached. To function satisfactorily the sine bar
requires to be made accurate. Important features of the sine bar are the parallelism of the top face
with two rollers, the spacing of the rollers, the equality of diameter of the rollers, and the
roundness of the rollers.

L
θ H

Fig.3 Sine bar


Source : Geneva Gages.Inc.

Sin θ = H / L

L = spacing between centers of two rollers


H = height of flat plate

Sensitivity of sine bar is known as the ability to produce a very small increase in angle by
increasing the height H by a certain small amount. The sensitivity decreases as the angle increases
from 1 to 90.

Applications:

The sine bar is an instrument for measuring how much a vertical surface is off from plumb. It can
also be used for setting surfaces to be exactly plumb such as saws, anvils, line bars, face plates,
press rolls. The sine bar is also used to measure small angles of plumb. Sine bar is used to set up
angles on a layout or in a milling machine vice.

The accuracy of the Sine bar is governed by

a. The straightness of the base generator.


b. The sensitivity and accuracy of the dial gage
c. The magnitude of the angle being measured
d. The degree of alignment of the test piece with longitude axis of sine bar.
3.3.2 SINE BLOCKS:

Fig. 4 Sine Blocks


Source : Taft- Peirce Inc.
In measuring process, the sloping platform of the sine block surface is used to support the tapered
part whose vertex-actual or virtual-must point toward the elevated end of the sine plate, in
position where the parts axis is contained in a plate precisely to the hinge of the block.

3.3.2.1 SINE BLOCK TAPER TESTING FIXTURES ( 9122 series ):

These fixtures are designed specially for precise inspection of taper parts machined on centers.
Tapered work is mounted between the centers, the sine block base elevated by gage blocks is one
half the included angle, and the work then indicated at both ends to determine the accuracy of
taper.

Taper testing fixtures combine a special 20” sine block with a pair of identical center heads that
may be positioned anywhere along a T slot running the length of the block and locked in place or
removed entirely to free the Sine Block for general purpose angle checking. Used in the
horizontal plane, a taper Testing Fixture becomes an extremely accurate Bench centers.

3.3.2.2 SINE BLOCK : ( 9118 series ):

They are heavy-duty fixtures for holding work that must be set to precise angles for layout, light
machining or inspection. Each block is equipped with a precision ground end plate that acts a
work positioning stop, and tapped holes in the top surface and sides are provided to allow a wide
variety of work clamping set ups.

Sine blocks are made of high quality alloy steel, scientifically heat treated for maximum stability
and wear resiatance, and precision ground on all working surfaces. Sine rolls are manufactured of
tool steel ground and lapped alike for diameter, roundness and straightness and are mounted
parallel to each other at their nominal distance within the specified limit.

3.4. SPIRIT LEVEL:

Precisely calibrated and graduated spirit levels are used to measure the angular tilt. They are
widely used because they are easy to understand, simple to use and reliable. The sensitivity of
spirit level is defined as the angle of tilt, which causes the bubble in the vial to move through one
scale division. Spirit level contains a glass tube of arcuate form almost completely filled with low
viscous fluid such as alcohol or benzol, except for small portion to create an air bubble. When the
spirit level is placed on a perfectly flat surface the air bubble will occupy the highest point in
middle of tube due to fluid’s greater specific mass. When the spirit level is placed on a tilted
surface the bubble will moves through a distance along the arc depending upon the tilt constancy
of radius R is essential to ensure equality of correspondence of the indicated value l to the angle
of inclination θ of the level over the length of scale. The value I = R* θ

Fig. 5 Spirit Levels


Source : Wyler Products. Inc.

The following sensitivities are available in the precision levels:

Sensitivity in mm/m Sensitivity in Arcsec Range of Measurement in


Arcsec
0.020 4.13 +/- 12.4
0.040 8.25 +/- 24.8
0.050 10.31 +/- 30.9
0.100 20.62 +/- 63.9
0.200 41.25 +/- 123.8
0.300 61.86 +/-185.6
3.5 THE ELECTRONIC LEVELS:

The electronic levels uses electronics to sense and display the angular values. A Pendulum is
freely suspended into the body of the instrument whose base simulates true horizontality
corresponding to the zero central position of the pendulum. The slight tilt of the base can cause an
electrical inequality between pendulum and body, which can be measured and interpreted in form
of angular measurement.

3.5.1 NIVELTRONIC INCLINOMETERS:

The Nileltronic inclinometers are available with the built in display system. The basic elements of
the instrument are the extremely sensitive pendulum system, the analogue pointer and a sturdy
housing of cast iron.

The measuring principle is based on the mechanical pendulum which is friction free suspended
and has therefore tendency to swing back the vertical position and at the lower end of the
pendulum a ferrite core is fastened which is penetrating a double winded coil fed by alternative
current (Principle of an inductive probe). The vibration of the pendulum will be dampened by
means of a magnetic brake. The position of the ferrite core influencing the inductance in the coil,
which is responsible for the changing of the voltage produced by the inclination. This voltage will
be treated and displayed by the galvanometer. The system is friction free and sensitive to the
slightest change of inclination.

The instrument is well liked for measurement of machine tool geometry. The analogue display
allows the recognition of measuring trends. When setting of large machine bed is to be done the
excellent zero point stability is highly appreciated. A disadvantage is the possible interference of
the other magnetic fields and the sensitivity to mechanical shocks.

Specifications:

Traveling distance between two lines corresponds to 10 to 50 micro m /m


Measuring range :+/- 0.150 to 0.750 mm/m
Analogue output : 0.24 V
Cast iron scraped prismatic bases.

3.5.2 LEVELMATIC INCLINOMETER:

Due to the use of high precise sensors this instrument is used for high precision measurements.
Principle:
The pendulum suspended by the Archimedes helical spring, is mounted between two electrodes.
Depending upon the inclined position of the system, the pendulum will swing out of the zero
position and by that, changing the capacity between the pendulum and the two electrodes. These
capacities will be transformed into different frequencies through the RC-oscillator. The ratio of
the two frequencies available will be used as the primary signal for determining the required
angle.

Specifications:

Measuring range : 2 mm/m to 20 mm/m


Sensitivity : 0.001 mm/m
Limits of errors: max. 1% of measured value
Inclination signal;2000 mV
Excitation: +/- 5 V dc

3.6 OPTICAL INSTRUMENTS:

Optical instruments are used for precise angular measurements. These instruments work on the
principle of the collimation of light.

3.6.1 AUTOCOLLIMATOR:

An autocollimator is an optical instrument that is used to measure small angles with very high
sensitivity. As such, they have a wide variety of applications including precision alignment,
detection of angular movement, verification of angle standards and angular monitoring over long
periods.

Principles of operation :

As the name suggests, the autocollimator projects a beam of collimated light. As external
reflector returns all or part of the beam back into insrument, where the beam is focused and
detected by a photo detector. Essentially the autocollimator measure the deviation between the
beam and the return beam. Because the autocollimator uses light to measure angles, it never
comes into contact with the test surface. Also because the autocollimator projects a collimated
beam, the working distance is limited only by the power of the light source.

X Vertical deviation between two reflected beams


δ Angular deflection of the mirror

Fig. 6 Autocollimator

The relation between vertical deviation and angular deflection can be given by

X = 2* f*δ

Where f= focal length of the lens.


3.6.1.1 VISUAL AUTOCOLLIMATORS :

Visual autocollimators rely on the operator’s eye as the photodetector. Visual autocollimators use
pinholes as the “light source”. The operator views the pinhole image through an eyepiece or on a
video monitor. Because the human eye acts as the photodetector, resolution will vary by user.
Typically, people can resolve from 3 to 5 seconds.

Because the human eye is able to discern multiple images simultaneously, visual autocollimator
can accommodate several return images (i.e. They can look at several reflective surfaces
simultsneously). This makes them ideal alignment instruments in applications like alignment laser
rod ends. Also, visual autocollimator are typically focused on infinity, makig them equally useful
for short or long distance measurements.

Visual pinhole autocollimators are designed for aligning multiple surfaces. Ideal
applications include:

• Easy to operate go /no-go gauge for surface alignment.


• Detecting non-parallelism in windows, laser rod ends, and optical wedges.
• Checking square ness of outside corners.
• Angle comparisons of reflecting surfaces.
• Verifying right angle prisms for angular and pyramid errors.
• Field use as a compact, lightweight, and robust autocollimator.
• Estimating angular measurements to 3 seconds.

3.6.1.2 DIGITAL AUTOCOLLIMATORS:

Digital autocollimators use an electronic photo detector to detect the reflected beam. The digital
autocollimator uses an advanced silicon photo detector and the latest DSP technology to detect
and process the reflected beam. This processing allows the instruments to compensate for the
non-linearity of the detector and the objective lens as well as the stray/scattered light that would
otherwise affect measurements. As a result the instrument s have a resolution capacity of 0.01
arc-second ( 0.0485 micro radian ), a linearity of 0.1% and a measurement rate as high as 1000
HZ. Digital autocollimators have the added benefit of direct readout of measurement data. The
data can be acquired via the LCD to display the analog out put or the instruments have direct plug
in compatibility with a computer or other electronic system. This makes high-speed data
acquisition and transfer capability a characteristic of these instruments.

Digital autocollimators are designed to measure angles in static and dynamic systems to a
very high degree of resolution. As such, application vary widely, Past application includes :

• Remote monitoring of alignment in large mechanical systems.


• Connection to servo system for creation of a stable platform in airborne system.
• Verification of angle standards such as rotary table or angle blocks.
• Measurement of carriage pitch and yaw along a rail.
• Alignment and monitoring of robotic arms.
• Operation in vacuumed chambers and space.
• Angular measurements to 0.01 arc second, with very high linearity and repeatability.

3.6.1.3 LASER AUTOCOLLIMATORS:


An autocollimator that emits a laser beam has significant advantages over conventional
autocollimator. For instance, the high intensity of the laser beam creates ultra-law noise
measurements, increasing the accuracy and repeatability of the instrument. The high intensity of
the beam also increases the autocollimator’s working distance ( to more than 15 meters with TL
160 Laser autocollimator) and permits angle measurements off of non-mirror quality surfaces.
The use of a laser source also allows the TL40 Laser autocollimator to make precision angular
measurements off of surfaces that are as small as 1 millimeter in diameter.

Laser autocollimators represent the future of precision angle measurement technology. The
superior beam intensity and collimation provided by laser light sources make them ideal
for:

Angle measurement of miniature angles, down to 1 mm diameter.


Long range angle measurement and alignment monitoring, with distances to more than 15 meters.
Measurement of non-mirror quality surfaces, including silicon wafers and bare metal.
Ultra low noise measurements in metrology and calibration laboratories.

An autocollimator is used to detect and measure small angular tilts of a reflecting surfaces placed
in front of the objective lens of the autocollimator. Ideally, the area of the reflecting surface
should be at least equal to the area of the objective lens. However, this is not generally the case
when the autocollimator is used with conjunction with angle gauges or a polygon. Therefore,
since the objective lenses fitted to most the autocollimator so that its optical axis passes through
the center of the reflecting face of the angle gauge or polygon, reducing the effect of waveform
errors to a minimum.

An Autocollimator should ideally be used in an environment where air currents in the optical path
between the autocollimator and the reflecting surface are minimal. Such air current, by
introducing changes in density and therefore of refractive index, producing random movements of
the observed image, impairing the accuracy of the autocollimator setting. For this reason, the
distance between the objective lens and the reflecting surface should be kept to a minimum and,
where practicable, the path should be shielded from the surrounding air.
The autocollimator should be rotated about its optical axis, if such provision exists, until a
movement of the reflected image perpendicular to the direction of measurement produces no
change of reading. For photoelectric autocollimators, this condition should be achieved using
photoelectric detector.

1. VISUAL AUTOCOLLIMATORS

MODEL 50 ALIGNMENT COLLIMATORS CCD –100 ALIGNMENT AUTOCOLLIMATORS


Fig 7. Visual Autocollimators , Source: Micro Radian Inc.
Specifications :
Specifications :
o Resolution of 5 arc second.
o Resolution of 3 arc-seconds.
o 1-degree field of view.
o Pinhole diameter of 100 microns. o 1-degree field of view.
o Image diameter subtends 100 arc seconds. o Pinhole diameter of 200 microns.
o Variable intensity incandescent light source. o Image diameter subtends 200 arc seconds.
o Optional 2-axis adjustable reference mirror mounted o Variable intensity incandescent light source
on front of body.

2. Digital Autocollimators

T 160 Digital Autocollimator T 100 Digital Autocollimator

Fig. 8 Digital Autocollimators


Source : Micro Radian.Inc.

Specifications :
o Maximum range of +/-600 arc second, each Specifications :
axis. o Maximum range of +/- 1200 arc second, each
o Maximum working distance of 5 meters. axis.
o Maximum resolution of 0.01 arc second. o Maximum working distance of 3 meters.
o Linearity of 0.2 %. o Maximum resolution of 0.01 arc second.
o Light source : single, visible red ultra o Linearity of 0.2 %.
bright LED. o Light source : single ,visible red ultra bright LED.
o Weight of 1.2 lbs. o Weight of 1.1 lbs.
3.Laser Autocollimator

TL 160 Laser autocollimator TL 40 laser autocollimator :


Fig.9 Laser Autocollimators
Source : Micro Radian.Inc

Specification : Specifications :
o Maximum measuring range of +/- 600 arc- o Minimum target size of 1.0 mm diameter.
seconds in each axis o Maximum measuring range of +/- 3600 arc
o Maximum resolution of 0.1 arc second. seconds in each axis.
o Measurement linearity of 0.5 %. o Maximum working distance is 1 meter.
o Simultaneous two-axis measurement. o Maximum resolution of 0.1 arc second.
o Maximum data bandwidth of 1000 Hz. o Measurement linearity of 0.5 %.
o Diode laser light source, 670nm ,class II o Maximum bandwidth of 1000 HZ.
o Maximum working distance 15 meters. o Diode laser source ,670 nm,classII

TL 160 has a built in laser diode light source, which emits a 5 mm diameter beam. The use
of a laser source gives the TL 160 a more powerful, better-collimated beam than is possible
with other light sources. The superior beam quality results in ultra-low noise, low drift
measurements. The intensity of the beam allows measurements to be taken off of surfaces
that are not mirror quality and off of surfaces that are as far as 50 feet away. The compact,
light weighty 160 is simple top operate and is ideal for use in remote monitoring
applications or as a calibration standard with tractability to the NIST.

3.6.2 ANGLE DEKKOR :

This is also a type of autocollimator. It contains a small illuminated scale in the collimating
objective lens. In normal position this scale is outside of the view of the microscope eyepiece.
The illuminated scale is projected as a parallel beam by the collimating lens. After reflected from
the work piece it will refocus by the lens in field view of eyepiece. There is another datum scale
is fixed across the center of screen and the reflected image of the illuminated scale is received at
right angle to this fixed scale. Thus the reading on the illuminated scale measures angular
deviations from one axis at 90 °to the optical axis and the reading on the fixed datum scale
measures the deviation about an axis mutually perpendicular to the other two.
Microscope
eye piece
Light
Source Prism

Glass screen

Datum scale
Illuminated scale
engraved on glass screen

Collimating
objective Lens

Work piece

Fig. 10 Angle Dekkor

The whole optical system is enclosed in a tube, which is mounted on an adjustable bracket. This
instrument is mostly used as a comparator. It is not precise as an autocollimators. It has a wide
range of application as angular variations are read directly without the operation of micrometer.

3.7 Angular Measurement using MR Technology :

Philips Semiconductors, has launched the world’s first Silicon integrated sensor system for
contact less angular measurement using Magneto resistive ( MR ) technology. Particularly
relevant in automotive and industrial applications, where angle measurement is frequently
required , the contact less MR approach provides a wear-free solution and removes the need for a
discrete solution on a hybrid. It is highly accurate and insensitive to temperature effects, magnet
ageing and displacement in a wide range.

This new contact less angle measurement system consists of a magneto resistive sensor and
sensor signal conditioning IC s. With the sensors and the signal conditioning electronics the
system is robust against dirt, dust and liquid as well as high temperature and mechanical
destruction. MR based systems tolerate variations in field strength caused by ageing to
temperature-sensitivity of the magnet as well as mechanical tolerances.

Automotive applications of the MR based system include chassis and seat positioning, throttle,
variable ventile timing, suspension and positioning for drive wire.

You might also like