Professional Documents
Culture Documents
from 400 to 700 nm. It is designed for use with the sample port facing up or forward.
It can also be used port-down with an optional stand. The sample is illuminated by a
xenon flash lamp and reflected light is collected by a 15- station fiber optic ring. For
This most versatile 00/450 spectrophotometer measures reflected color the way the
eye sees it, including the effect of gloss. It measures samples in virtually any form
from solids and powders to liquids. The system uses automation extensively with
Procedure
1. Configure software to read using the desired color scale, illuminant, and
observer.
2. Install the 0.20 inch (5 mm) port plate and standardize the instrument for
the 0.125 inch (3.175mm) area of view. First standardize on the black glass,
and hold it in place. Make sure that the area of the sample to be measured
4. Take a single color reading of the sample. Rotate the sample 90 0 or move
the sample to another area containing the same color and read it at least once
more. Average the multiple color readings for a single color measurement.
allows you to better estimate the color of the designated area(s) as a whole.
sample cup viewing port can be used to measure the reflectance of translucent semi-
solids in a glass sample cup with a ring and disk set and opaque cover.
1. Configure software to read using the desired color scale, illuminant, and
observer.
2. Orient the instrument with the port facing up. Replace the regular port insert
3. Standardize the instrument for the 1.75 inch port size and 1.0 inch area of
view using the black and calibrated white standards. Measuring with a smaller
4. Insert the 10-mm black ring into the 2.5 inch glass sample cup so that it
settles flat on the bottom of the cup. The plastic ring fixes the internal path
length of light through the semi-solid sample at 10mm while excluding outside
Then fill the cup with the translucent semi-solid sample until the sample is
6. Push the white ceramic disk down through the semi-solid sample until it
rests firmly on top of the black ring. The disk provides a white background to
direct light that has traveled through the sample back to the detector.
7. Your goal is to have the sample appear smooth and opaque through the
9. Cover the sample cup with the opaque cover. The opaque cover prevents
any ambient light from outside the instrument from leaking into the detector.
10. Take a single reading of the sample through the bottom of the sample cup.
If the sample is uniform in color or the tolerances are wide, one reading may
with sample replacement for a single color measurement represent the color
of the batch. Scoop or pour the sample out of the sample cup, refill it, and
measure again several times from the same batch. Average the readings
together for a single color measurement representing the color of the batch.
11. Record the average color values for the sample batch.
The Hunter L,a,b color space is a 3-dimensional rectangular color space based on
the opponent-colors theory.
RESULTS
Colour values of extruded maize flour based snack food: L * = 70.1 ± 2.2
a * = -1 ± -0.4
b * = 17.4 ± 2