LECTURE NOTES
ON
Mr. Y.HARIKRISHNA
ASST.PROFESSOR
1
UNIT 1: MEASUREMENTS
ClassificationofMeasuringInstruments
Theelectrical measuringinstrumentscanbeclassifiedasbelowMeasuringIn
struments
Absolute Secondary
instruments instruments
EeffectsofElectricCurrentUtilizedInMeasuringInstrume
nts
Theprincipleofoperation ofmeasuring instrumentsdepend upon the variouse
ffectsofelectriccurrentusedintheiroperation.Theyareclassifiedasinthetablesh
ownbelow.
Effects Typeofmeasur Instruments
used ements utilizingtheseeffects
MagneticEffect Current,Voltage,Power Ammeter, Voltmeter,
andEnergyonbothA.C.a Integratingmetersandother
ndD.C.System. electricalinstruments
Thermal effect Current,Voltageonboth Ammeter,Voltmeter
A.C.andD.C.system
Electromagnetic Voltage,Current,Power Voltmeter,Ammeter,Watt
InductionEffect energyinA.C.system meterenergymeter
ChemicalEffect AmperehoursinD.C.S Voltmeter,Ammeter
ystem
ElectroStatic Voltmeteron bothA.C. Voltmeter
effect andD.C.System
MainpartsofIndicatingInstruments
Indicatinginstrumentsarethosewhichindicatethevalueofthequantityund
ermeasurement.Themainpartsofindicatinginstrumentsare.,
1. Pointer.
2. PreCalibratedScale.
3. Movingsystemwithspindle,pivotedinjewelledbearings.
4. Deflectingoroperatingtorque.
3
5. Controllingorrestoringtorque.
6. Dampingtorque.
Pointerandscale
Theshapeandsizeofthepointernormallydependsonthetypeandsizeofinstru
ments,usuallyaluminiumtubeorstripisusedasamaterial forpointer.Theendofthe
pointerisflattenedtotheshapeofanarrowheadand
ismadetomoveoveramirrorandgraduatedscale.
Themirrorhelpsinavoidingparallelaxis.Thearmofthe pointerusuallyhasado
ublebendto
adjustthearrowheadinleveltothescale.Asmallcounterweightisattachedattheta
ilendforbalancing.
Fig.2.3Differenttypesofpointers
Aclearlymarkedgraduatedscaleisattachedtothecaseoverthepointer.Thel
engthofthescaledependsuponthelengthofthepointerandtheangletowhichitcanr
otateforfullscaledeflection.
Thescaleisprintedonastiffcordusuallyglazedandwhitebackgroundandbl
ockmaking.
Therearetwotypesofscales usedinindicatinginstruments,namelyuniformsca
leandnonuniformscale.Inuniformscale,eachdivisionisequallydevided.Thistype
ofscaleisusedinmovingcoilinstruments.
Thenon-
uniformscaleiscrampedatendsandbroadlydevidedinmiddlepartofthescale.We
canseethistypeofscaleinmovingironinstrumentsand
a.c.measuringinstruments.
TorquesofIndicatingInstruments
Inmostoftheindicatinginstrumentsthreedistincttorquesarerequiredforoper
ation.Theyare
a) Atorquetomovethepointerfromitszeropositioninsameproportiontothe
magnitudeofthequantitybeingmeasured.Thistorqueiscalleddeflectin
gtorque.
b) Ifthe pointer hastostopatavalueindicatingthe magnitudeofthe quantity,the
reshouldbeanequalandoppositetorqueatthatinstanttokeepthepointerin
stableposition.Thistorqueiscalledcontrollingtorque.
c) Whentwotorquesexistsinoppositedirectionoscillationresultsduetowhi
chthepointertakessometimeequaltotheoscillationalternationtimetoseti
ELECTRICAL AND ELECTRONIC MEASUREMENTS
ntostablepositiontobringthepointertorest,atthequickestpossibletime,an
othertorqueactsinthe directionsopposite to themotionofthepointeratthe
deflectedposition.Itiscalleddampingtorque.
1. DeflectingTorque(T)orDeflectingForce
d
Thedeflectingtorquecausesthemovingsystemoftheinstrumenttomovefrom
itsinitialzeroposition.Themagnitudeofthedeflectingtorquedependsupon,the
magnitudeofthemeasurablequantity.
Thetorqueisproducedbytheeffectsofelectriccurrentsuchasmagnetic,elect
ro-
magneting,heating,electrostaticetc.Themethodofproductionofdeflectingtor
queanditsrelationtothemeasurablequantitydependsuponthetypeofinstrument.
Pointer
TD Pivot
CounterBalance
(DeflectingTorque)
Fig2.4TheActionofDeflectingTorque
2. ControllingTorque(T)orControllingForce
c
Theforceactingonamovingsystem,inoppositedirectiontodeflectingtorqu
eandmakesthepointer stableatitsfinaldeflectedposition
iscalledControllingTorqueorControllingForce.Thepointerwilltrytorotateasmot
orinabsenceofthistorque.Italsobringsthepointerbacktozeroposition,afterdisco
nnectingthemeterfromthecircuit.
Fig2.5TheActionofControllingTorque
5
Thecontrollingtorqueisprovidedinthefollowingways.
1. Springcontrol 2.Gravitycontrol
a. SpringControl
Inthis,thespindleofthemovingsystemrestsontwopivots.Ahelical hairsprin
gusuallymadeofphosphorbronzeisattachedtothemovingsysteminsuchawayt
hatitsoneend(innerside)isfixedtothemovingsystem,whiletheouterendtoarigidb
ody.Thestressinthespringexerts,atorqueonthemovingmechanismoppositetoth
emotionofthemovingmechanism.Thistorqueisproportionaltotheangleofdefle
ction.
Initiallywhenthepointerisatzerodeflectedposition,the controllingtorqueisze
roasthereexistsnostressinthehelicalspring.Atthefinal deflectedposition,torquepr
oducedbythespringwouldbecomeequalandoppositetodeflectingtorque.
Graduatedscale
Pointer Jewelbearing
Spring
Counterweight
Fig.2.6TheActionofSpringControlTorque
b. GravityControl
Inthismethod,asmallweightisattachedtothemovingsysteminsuchawayth
atitgetsactivatedatthetimeofdeflectionandproducesacontrollingtorquein pro
portiontothedeflection.Themagnitudeofthecontrollingtorquecanbevariedbyad
justingthepositionofthecontrollingweightuponthearm.
Theabovefigureshowsthearrangementsandcorrespondingpositionsofthec
ontrolarm forzerodeflection.Inthezerodeflectedpositionofthepointer,thecontro
lweightstandsverticalduetowhichtheforceduetogravitydirectlyactsonthepointe
randnotorqueexistsonthepointer.Asthedeflectiontakesplacethecontrolarmals
oshiftsfromverticalpositionwithan angletovertical(position)tothatofdeflection.
Butthedisadvantagesofgravitycontrolare
(i) Itgivescrampedscale
(ii) Theinstrumenthastobekeptvertical
ELECTRICAL AND ELECTRONIC MEASUREMENTS
Fig.2.7GravityControlTorque
DampingTorque
Intheindicatinginstruments,thepointerwilloscillateaboutitsdeflectedpositi
onbeforecomingtorestduetoinertia.Tobringthemovingparttorestatthequickestp
ossibletimeatthedeflectedposition,aforceisusedwhichiscalleddampingforce.
Itoperateonthemovingsystemtobringitrestatthequickestpossibletime.The instr
ument provided witha dampingsystemis knownasdeadbeattypeinstrument.Diff
erentdevicesareusedtoprovidesuitabledamping.
Thevariousmethodsadoptedfordampingare.,
d) Airfrictiondamping
e) Fluidfrictiondamping
f) Eddycurrentdamping
The systemused varieswiththe typeofinstrument.Eachmethodis discussedin
thesucceedingparagraphs.
a. Airfrictiondamping
Twomethodsareemployedinairfrictiondamping.Thefirstmethodusesalighta
luminiumpistonattachedtothemovingsystemasshowninfigureThepistonmove
sinsidetheairchamberalongwiththepointer.
Asthepistonslidesinthechamber,theairinsidetheclosedspaceiscompresse
dandthepressuresodevelopedopposesthemotionofthepiston.
7
Theonlycarethatmustbetakensothatthepistonisnotbentorthepistondoesn
ottouchthewallsofthechamber. Itwouldleadtoseriouserrorsin
reading.
Enclosedair
chamber
Spring
Piston
Pointer
Fig.2.8AirFrictionDampingwithPistonairchamber
Thesecondmethodemploysavanemountedonthespindleofthemovingsyste
m.
Itismadeofthinsheetofaluminiumandmovesinaclosedsectorshapedbox.T
hismethodismoreadvantageous.Sincethevanerotatealongwiththespindleinthe
sectorbox.Carehastobetakenatthetimeofmountingtoseethatthevaneisfreeinsi
dethesectorbox.
Fig.2.9AirFrictionDampingwithaluminiumvane
b. Fluidfrictiondamping
Inthismethodofdamping,alightvaneisattachedtothespindlewhichisdipped
in apotofhighviscusdampingoil.Themotionofthemovingsystemisalwaysoppos
edbythefrictionofthedampingoil.Thedampingforcecreatedincreaseswiththein
creaseinvelocityofthevane.Thereexistsnodampingforcewhenthevaneisstatio
nery.
ELECTRICAL AND ELECTRONIC MEASUREMENTS
Fig.2.10FluidFrictionDamping
c. EddyCurrentDamping
Inthissystem,acirculardisc(copperoraluminium)discismountedonthespindl
eanditismadetomoveinamagneticfieldalongwiththespindle.Thedisccutsthem
agneticlinesofforceofthemagnetandeddycurrentswillbeinducedinit.Theinduc
ededdycurrentsopposesthemotionandthusadampingforceisdevelopedintheinstr
ument.
Thereisan anothertypeofeddycurrentdampingsystem,which isusedinperm
anentmagnetmovingcoilinstruments.Thecoiliswoundon thisaluminiumframe,inw
hicheddycurrentsareproducedwhenthecoilmovesinthefieldofthepermanentma
gnet.
Fig2.11EddyCurrentDampingByThinAluminiumFrame
9
Fig2.12EddyCurrentDampingByThinAluminiumDisc
AmmeterandVoltmeter
Ammeterisaninstrumentinwhichtorqueisproducedby thecurrenttobemeas
ured.Ithaslowresistancecoil,sothatitcanbeconnectedinseriestothecircuitwitho
utanyappreciablevoltagedropinit.Whereasavoltmeterhasahigh resistancecoila
nditcarriesa currentwhich isproportionaltovoltageacrossthecircuittobemeasure
dandthetorqueisproducedduetothiscurrent.
Theonlydifferencebetweenthetwo(Ammeterandvoltmeter)isinthemagn
itudeofthecurrentandmethod ofconnectingin
circuit.AmmeterisconnectedinserieswithcircuitandVoltmeterisconnectedinp
arallelwiththecircuit.
Ammeterandvoltmeteraregroupedtogetherbecausethereisnoessentialdiffe
renceintheprinciplebetweenthem.
Fig2.13Ammeter Voltmeter
ELECTRICAL AND ELECTRONIC MEASUREMENTS
Ammeterandvoltmetersareclassifiedasfollows.
a) Movingcoilinstrument
(i) Permanentmagnetictype (ii)Electrodynamicordynamictype
b) Movingironinstrument
(i) Attractiontype (ii)Repulsiontype
c) Hotwiretype
d) Inductiontype
(i) Splitphasewinding (ii)Shadedpoletype
e) Electrostatictype
PermanentMagnetTypeMovingCoilInstrument
Thisinstrumentisalsocalledmovingcoilinstrument.Itismoreaccurate,sensiti
veandhas uniformscale.Itismostcommonlyused forDCmeasurements.
Principle
Itworkson theprincipleofelectrodynamiceffectorprincipleofDCmotor.Itstat
esthat,wheneveracurrentcarryingconductorisplacedinamagneticfield,itexper
iencessomemechanical forceandmovesinaparticulardirection.
Construction
ItconsistsofaU-
shapedpermanentmagnet.Thispowerfulmagnetismadeofalnico.Arectanglecoil
ofmanyturnswoundonlightaluminiumframeiskeptinsidethepolepiecesasshown
inthefigure.Thisframeissupportedbydelicatebearingandtowhichalightpointeris
attached.Thealuminiumframeprovidessupportforthecoilandalsoactsasdampin
gdevicebyeddycurrent.
Working
Fig ConstructionofPermanentMagnetMovingCoilinstrument
11
Whenthisinstrumentisconnectedinacircuit,thecurrentispassedthroughtheco
il.Asthecoiliskeptin magnetic field,itexperiencesatorqueanddeflectedduetoelectr
odynamiceffect.The deflectingtorqueisproportionaltothequantityofelectricityor
currentwhichisundermeasurement.
ControllingTorque
Controlofthecoilmomentisbyspringcontrolmethod,byusingtwophosph
orbronzehairspring-
oneaboveandonebelow.They additionallyservesthepurposeofpassingofcurrent
inandoutthecoil.
Dampingtorque
EddycurrentdampingisprovidedbytheAluminumframe,onwhichthecopp
ercoil was made.When the coilrotatesinthemagnetic field,theAluminiumframeals
omovesalongwiththecoil.Henceeddycurrentswillproduceintheframeandthere
bydampingtorqueisdeveloped.
Applications
Thistypeofinstrumentsaremovesensitive,accurateandhas uniformscale.
TheycanbeusedformeasurementofDCvoltagesandcurrents.
MovingIronInstruments
Thistypeofinstrumentsworkontheprinciplesof magnetismi.e.,attractionorre
pulsionbetweentwomagnetizedironpieces.Therearetwotypesof movingironinstr
uments.
Theyare
1) Attractiontypemovingironinstrument
2) Repulsiontype movingironinstrument
MovingIronInstrument-AttractiontypePrinciple
Movingiron instrumentworkson the principleof magnetism.i.e.,Amagnetatt
ractstheironpieceorUnlikemagneticpolesattracteachother.
Constructionandworking
Itconsistsofafixedaircorecoilmadebyinsulatedcopperwireandanovalsha
pedsoftirondiscfittedonaspindle.Thespindleispivotedbetweentwojewelledbe
aringsneartothecoil.Apointerisattachedtothespindle,whichmovesonapre-
calibratednonuniformscale.
ELECTRICAL AND ELECTRONIC MEASUREMENTS
Fig.2.15AttractiontypeMovingIronInstrument
13
ConstructionandWorking
Inthis,afixedcoilisusedasanelectromagnetandtwosoftironrods/stripsa
rearrangedinsidethecoilasshowninfigure.Themovablestrip/rodcarriersapoin
terthatmovesoverthecalibratedscale.
Thedeflectingtorqueisbymagneticeffectofelectriccurrent.Thecontrol
lingtorqueisbyspringsorgravitysystemanddampingisbyairfrictionsystem.
SectionalView PlanView
Fig.2.16RepulsiontypeMovingIronInstrument
Whentheinstrumentisconnectedinacircuitthecurrentpassedthroughthefix
edcoilandsetsupamagneticfieldwhichmagnetisesthetworodsofsimilarpolarit
y.Thenarepulsiveforcewilldevelopbetweenthemandspindlemoved.
Thenthepointerattachedtothespindlemovesonthescaleshowingthevalue
ofelectrical quantity.Theforceofrepulsionisapproximatelyproportionaltothesqu
areof thecurrentpassingthrough thecoilandhencethescaleiscrowdedatends.
Whateverthedirectionofcurrentinthecoil,thepolarityinducedontheironst
ripsissame.Hencealwaystherewillbearepulsiveforce.Thereforeitcanbeusedf
orACandDCmeasurements.
Applications
TheMovingIronrepulsiontypeinstrumentcanbeusedforACandDCmeas
urements(VoltmeterandAmmeter)
ELECTRICAL AND ELECTRONIC MEASUREMENTS
AdvantageandDisadvantagesofMovingIronInstruments
1. TheseinstrumentscanbeusedbothonA.CandD.Csystem.
2. Theseinstrumentsarerobustandfreefrommaintenance
3. Itpossesseshighstartingtorque.
4. Itcanwithstandmomentaryoverloads.
5. Itcangivereasonableaccuracyinthereading.
6. Inthisinstrumentthescaleisnonuniform.
7. Powerconsumptionishighatlowvoltage
8. Thestiffnessofthespringdecreaseswithincreaseintemperature.
9. Hysteresisandstraymagneticfieldcauseserrorsinthereading.
10. Changeinfrequencyofoperationcausesseriouserror.
DifferencesbetweentheMovingCoilandMovingIronInstrum
ents
Movingcoilinstruments MovingIroninstruments
1. ItworksontheprincipleofD 1.Itworksontheprinciplesofma
Cmotor gnetism.
2. Deflectiontorqueisproportionalt 2. Deflectiontorqueisproportional
ocurrent tosquareofcurrentDampin
3. Dampingisprovidedbyeddy.cu 3. gisprovidedby airdamping
rrent Gravitycontrolledinstruments.Co
4. Springcontrolledinstruments 4. ntrollingtorqueisproportionaltoSi
5. Controllingtorqueisproportional5. n
toangleofdefection 6. Nonuniformscale.
6. Scaleisuniform 7. Robust,reliableaccurate.
7. Delicate,sensitiveanda
ccurate.
8. Costly. 8. Cheap.
9. Lowpowerconsumption 9. high powerconsumptionthan
movingcoil.
10. ItisusedonlyinD.C. 10.ItisusedbothinA.C.andD.C.
Circuits. Circuits.
11. Canbeusedasvoltmeter,Am-
11.CanbeusedasAmmeter,meter,Galvanometer,ohmmeter
VoltmeterandWattmeter.
15
ExtensionofRangeofMCandMIInstruments-
NeedofShuntsandMultiplier
Normallyallthemeasuringinstrumentsaredesignedfora(current)rangebetw
eenfewmicroampstomilliampsforfullscaledeflection.Iftheinstrumenthas tobe use
dforhigherranges of currentandvoltagecertain amountof resistanceisaddedinserie
sorparalleltocoil.
Ityouobserveanammeterasmallwireisconnectedtothecoilandahighvaluere
sistorisconnectedinthecaseofvoltmeter.Sometimeswecanseetheseadditional
partsoutsideoftheinstrument.
Usuallymangainandnichromeisusedasresistancematerial.Itwoundonains
ulatedformerandleadsaresolderedtotwolargecopperblocks.Toincreasethedissipa
tingsurfacemangainisemployedintheformofthinstrips.
a. Shunt(Rsh)
Shuntisasmall amountofresistanceconnectedinparallel withammeter.Itisus
edtoextendtherangeofAmmeter.
Im
Rm I
Ish
V Rsh Load
2.17Shunt-extensionofammeter
Letthecurrentforfullscaledeflection(FSD)of Ammeteris=
Im,Amps.
ThemeterresistanceofcoilresistanceRm,
Loadcurrentornewrange-I,Amps
Currentisshuntresistance,IshAmpsShuntre
sistance=Rsh,ohms.
ImRm .ohms
Shuntresistance Rsh II
m
b. Multiplier(R)
se
AMultiplier isalargeamount of resistence connectedinserieswith voltmeter.
Itisusedtoextendtherange(measuringcapacityofvoltmeter.
ELECTRICAL AND ELECTRONIC MEASUREMENTS
I=I+I
m L IL
Voltmeter
Rm
Load
V
R se
Fig.2.18Multiplier-
extensionofVoltmeterLetthefullscaledeflection(FSD)voltag
e=V,VoltsThemeterresistence =Rm,ohms,
Loadvoltageornewrange =V,Volts
VV
ThemultiplierresistanceRse .Rm.ohms
V
17
Introduction to CRO – Cathode Ray Oscilloscope
The cathode ray oscilloscope is an extremely useful and versatile laboratory instrument
used for studying wave shapes of alternating currents and voltages as well as for meas-
urement of voltage, current, power and frequency, in fact, almost any quantity that
involves amplitude and waveform.
It allows the user to see the amplitude of electrical signals as a function of time on the
screen.
It is widely used for trouble shooting radio and TV receivers as well as laboratory work
involving research and” design.
It can also be employed for studying the wave shape of a signal with respect to amplitude
distortion and deviation from the normal.
In true sense the cathode ray oscilloscope has been one of the most important tools in
the design and development of modern electronic circuits.
The CRO consists of the following:
(i) CRT
(ii) Vertical amplifier
(iii) Delay line
(iv) Horizontal amplifier
(v) Time-base generator
(vi) Triggering circuit
(vii) Power supply
CATHODE-RAY TUBE:
The electron gun or electron emitter, the deflecting system and the fluorescent
screen are the three major components of a general purpose CRT
Electron Gun
In the electron gun of the CRT, electrons are emitted, converted into a sharp beam
and focused upon the fluorescent screen. The electron beam consists of an indirectly
heated cathode, a control grid, an accelerating electrode and a focusing anode. The
electrodes are connected to the base pins. The cathode emitting the electrons is surrounded
by a control grid with a fine hole at its centre. The accelerated electron beam passes
ELECTRICAL AND ELECTRONIC MEASUREMENTS
through the fine hole. The negative voltage at the control grid controls the flow of electrons in
the electron beam, and consequently, the brightness of the spot on the CRO screen is
controlled.
Deflection Systems:
Electrostatic deflection of an electron beam is used in a general purpose oscilloscope. The
deflecting system consists of a pair of horizontal and vertical deflecting plates. Let us consider
two parallel vertical deflecting plates P1 and P2. The beam is focused at point O on the screen in
the absence of a deflecting plate voltage. If a positive voltage is applied to plate P1 with respect
to plate P2, the negatively charged electrons are attracted towards the positive plate P1, and
these electrons will come to focus at point Y1 on the fluorescent screen.
The deflection is proportional to the deflecting voltage between the plates. If the polarity of
the deflecting voltage is reversed, the spot appears at the point Y2, as shown in Fig. 14-3(a).
To deflect the beam horizontally, an alternating voltage is applied to the horizontal deflecting
plates and the spot on the screen horizontally, as shown in Fig. 14-3(b). The electrons will
focus at point X2. By changing the polarity of voltage, the beam will focus at point X1. Thus, the
horizontal movement is controlled along X1OX2 line.
Phosphor is used as screen material on the inner surface of a CRT. Phosphor absorbs
the energy of the incident electrons. The spot of light is produced on the screen where the
electron beam hits. The bombarding electrons striking the screen, release secondary emission
electrons. These electrons are collected or trapped by an aqueous solution of graphite called
“Aquadag” which is connected to the second anode. Collection of the secondary electrons is
necessary to keep the screen in a state of electrical equilibrium. The type of phosphor used,
determines the color of the light spot. The brightest available phosphor isotope, P31, produces
yellow–green light with relative luminance of 99.99%.
Figure 14-5(a) shows a sine wave applied to vertical deflecting plates and a
repetitive ramp or saw-tooth applied to the horizontal plates. The ramp waveform at
the horizontal plates causes the electron beam to be deflected horizontally across the
screen. If the waveforms are perfectly synchronized then the exact sine wave applied
to the vertical display appears on the CRO display screen
ELECTRICAL AND ELECTRONIC MEASUREMENTS
TRIANGULAR WAVEFORM:
Similarly the display of the triangular waveform is as shown in Fig
TIME-BASE GENERATORS:
The CRO is used to display a waveform that varies as a function of time. If the wave form is to
be accurately reproduced, the beam should have a constant horizontal velocity. As the beam
velocity is a function of the deflecting voltage, the deflecting voltage must increase linearly with
time. A voltage with such characteristics is called a ramp voltage. If the voltage decreases rapidly
to zero—with the waveform repeatedly produced, as shown in Fig. 14-6—we observe a pattern
which is generally called a saw-tooth waveform. The time taken to return to its initial value is
known as flyback or return time.
MEASUREMENTS USING THE CATHODE-RAY OSCILLOSCOPE:
1)Measurement of Frequency:
2) Measurement of Phase:
ELECTRICAL AND ELECTRONIC MEASUREMENTS
UNIT II
CLASSIFICATION OFRESISTANCES
Forthepurposesofmeasurements,theresistancesareclassifiedintothreemajorgroupsbased
on their numericalrange of values as under:
• Lowresistance (0 to 1 ohm)
• Medium resistance (1 to100 kilo-ohm) and
• Highresistance(>100 kilo-ohm)
1. Lowresistance(0to1ohm):AVMethod,KelvinDoubleBridge,potentiometer,doctoroh
mmeter, etc.
2. Medium resistance(1 to 100 kilo-ohm): AV method, wheatstone’s bridge,
substitution method, etc.
3. Highresistance(>100kilo-ohm):AVmethod,Fallofpotentialmethod,Megger,loss of
charge method, substitution method, bridge method, etc.
heaverage value of all the trails will give the measured valueofresistance.
ThismethodisalsoreefedasthepotentialdropmethodorVAmethod.Here,themeterrangesar
e tobe chosencarefullybasedonthecircuitconditionsandtheresistancevaluetobe obtained. This
methodsuffers from the connectionerrors.
Wheatstone bridge
Thebridgeissaidtobebalancedwhenthegalvanometercurrentiszero.Thisiscalledasthe
position of null readingin thegalvanometer,which is used here as anull detector ofthebridge
under thebalanced conditions.
(iii) Errorsduetotheeffectoftheconnectingwiresandleadresistors:theconnectingleadwireresi
stance will affect the value of theunknown resistance, especiallywhen it is a
lowresistancevalue.Thus,theconnectingleadwireresistanceshavetobeaccountedforwhile
measuringalow resistance.
(iv) Contactresistanceerrors:Thecontactresistanceoftheleadsalsoaffectsthevalueofthemeasu
randresistance,justasinpoint(iii)above.Thisresistancevaluedependsonthecleanliness of
thecontactsurfaces andthe pressure applied to the circuit.
Limitations of WS bridge
TheWSbridgemethodisusedformeasurementofresistancesthatarenumericallyintherange
ofafewohmstoseveralkilo-ohms.Theupperlimitissetbythereductioninsensitivityto
unbalancecausedbythe highresistances, as per theequation (7).
Thus,theWSbridgesensitivityishighwhen(R/S)=1.Itdecreasesastheratio(R/S)becomeseit
herlargerorsmallerthanunityvalue.Thisalsocausesareductioninaccuracywith which
thebridgeisbalanced.
theKelvin’sDoubleBridge
Ofthevariousmethodsusedformeasurementoflowresistances,theKelvin’sDoubleBridge(
KDB)isthemostwidelyusedmethod.Here,thebridgeisdesignedasanimprovedormodifiedWSB
ridgewiththeeffectofcontactandconnectingleadwireresistancesconsidered.Thecircuitarrange
mentforKDBisasshowninfigure3below.Risthelowresistancetobemeasured,Sisthestandardlo
wresistance,ristheverysmalllinkresistancebetweenRandS,calledtheyokeresistance,P,Q,p,qare
thenoninductiveresistances,onesetofwhich,PporQqisvariable.Thebridgeissuppliedwithabatte
ryEandaregulatingresistanceRC.AgalvanometerofinternalresistanceRg,connectedasshown, is
usedas a nulldetector.
Fig.3Kelvin’sDouble Bridge
Balance EquationofKDB
ThebalanceequationofKDBcanbederivedasunder.Usingstar-deltaconversionprinciple,
theKDBcircuit of figure 3 can be simplifiedas shown in figure-4 below.
ELECTRICAL AND ELECTRONIC MEASUREMENTS
It is thus observed that the KDB has now become a simplified WS bridge-
ABCD as showninfigure4(c). Itistobe noted that they oke resistancer1 is
ineffective now, since itisinth ezero current branch, in series withth
egalvanometer. Thus, under balance conditions ,wehave,
P (S+r3) =Q (R+r2) (13)
Where,
CBridges
Maxwell Bridge
capacitance.
Themaxwellbridge is limited to the measurement of medium-Qcoils(1<Q<10).
Hay Bridge
ScheringBridge:
TheScheringbridge, oneof themost importantbridges, is usedextensivelyfor
themeasurement ofcapacitors.
Schering bridge for measurement of capacitance
Wien Bridge:
Applications :
– Frequencymeasureent
– Notchfilter
– Frequency-determiningelement
AndersonBridge:
In theAndersonBridge the unknown inductanceismeasured in terms
ofaknowncapacitanceandresistance.
this method is capable of precisemeasurements ofinductance over
awiderangeofvaluesfrom a fewmicro-henrys to severalhenrysand is the best
bridge method
Capacitance Bridge:
SinglePhase(1- )InductionTypeEnergyMeter
Principle
Thesinglephaseinductiontypeenergymeterworksontheprincipleofelectr
omagneticinduction.Arotatingmagneticfieldisdevelopeddue
tointeractionof currentin pressurecoil andcurrentcoilwith
aluminiumdiscplacednearcoils.
ConstructionandWorking
Thefigureshowsessential partsofsinglephaseinductionmeter.Itconsistsofap
ressurecoilmadeofthincopperwireofmanyturns(alsocalledshuntmagnet);acu
rrentcoilmadeofthickcopperwireofoneortwoturns(alsocalledseriesmagnet),a
n aluminium discmountedonspindleandplacedbetweentwojeweledbearingsanda
recordingmechanismconsistsofwormandwheelthroughwhichtherotationofspi
ndleisrecordedoncyclometer.
ConstructionalDetailsofSinglePhaseEnergymeter
Abrakingmagnetisarrangedonadisctocontrolitsmovementandtostopthemo
vementundernoload.Itproducesadampingtorque/force.Hereeddycurrentsyste
mofdampingisused.Aphasedifferenceof900issetbetweencurrentcoilandpress
urecoilwiththehelpofcoppershadedrings.
Thetwofieldfluxesproducedby the twocoilsacton aluminiumdiscinduceeddy
currentsandhencethediscrotatesduetotheinteractionofthetwofluxesdeveloped.
Thespeedofdiscisproportionaltotheenergyconsumed/powerconsumedbythel
oad.
Thenumberofrevolutionscompletedbythediscforone‘Kwh’iscalledmeter
constant.Usuallyitisprintedonnameplateofmeter.
Fig3.4InternalpartsofEnergymeter
Errorsin1- Induction typeenergymeter
(1) PhaseerrororPowerfactorerror
Thiserrorisduetoimproperphasedifferencebetweenfluxandvoltagedueto
shuntcoil.With this,thediscrotatesfasteratlowp.f.loads.Itiscorrectedbyadjustingt
hepositionoftheshadingrings.
(2) Speederror
Sometimesthespeedofdiscisnotproportionfornon-
inductiveloads.Thisisduetoincorrectadjustmentofbrakingmagnet.Thiscan beeli
minatedbyrepositioningthebrakemagnet.
(3) FrictionCompensationorfrictionerror
Inmanycasesanunwantedbrakingtorqueondiscwilldevelopedduetofrictio
nalforcesatdiscandregistermechanism.Thiscausesforcreepingerror.Thiscanbea
djustedbyrepositioningtheshadedringsoncentrallimb/shuntmagnet.
(4) CreepingError
Sometimesenergymeterrunsslowly,evenat no
loadcondition.Thisphenomenonissaidtobecreeping.Thismayoccurduetoove
rfrictioncompensation,straymagneticfieldormechanicalvibrations.Thiscanbe
preventedbyproperadjustmentoffrictioncompensation.Usuallytwosmallhole
sarecutnearthecircumferenceofthealuminiumdisc.Sometimesasmallpieceofiro
nwireisattachedtotheedgeofthediscwhichwhenattractedbybrakingmagnet.
(5) Temperatureerror
Theerrorduetotemperatureinmeterreadingisveryless.Anincreaseintempe
ratureresultsinincreasedresistancetotheeddycurrentpath reducingthedrivingtor
queandbrakingtorque.Italsoincreasetheresistanceoftheshuntcoilcausingsmall
ercurrenttoflow,duetowhichthedrivingtorqueisreduced.Athighertemperaturesth
efluxfrombrakingmagnetisalsoless.The netresultisto
neutralizeeachotherseffectand error causeddueto
temperatureisnotappreciable.
(6) Frequencyeffect
Increaseinfrequencyrisestheimpedanceofthecoilswiththeresultlesscurrent
flowsthroughthem.Thetorqueproducedisalsoproportionatelyreduced.
Hencethemeterhastobeadjustedatfixedfrequencytoobtainminimumerror.
PolyphaseInductionTypeEnergyMeter
Polyphaseenergymetersareobtainedbycombiningtwoormoresinglephase
energymeters.Twodiscsaremountedonthesamecommonspindleandeachdischas
itsownbrakingmagnet.Thecommonspindledrivesasinglegeartrain.Eachunitispr
ovidedwithitsownshadingrings.Theworkingissimilartosinglephaseinductiontyp
eenergymeter.
ThreePhase(3 EnergyMeters
Thethreephaseenergymetersworksonthesameprincipleassinglephaseener
gymeter.Theenergyinthethreephasecircuitcanbemeasuredeitherbytwosepara
tesinglephaseenergymetersorthreeseparateenergymetersorsingleThreephase
energy meter
Thesearealsocalledpolyphaseenergymeters.Theyweretwotypes1)3-
Phase,3-wireACEnergymeter.2)3-Phase,4-wireACEnergymeter.
(1) 3-Phase,3-wireACEnergymeter.
Thethreephasethree
wireenergymeterworksontheprincipleofmeasurementofpowerby twowattme
termethod.Itconsistsoftwoseparatepressurecoilsandtwocurrentcoils.Theyprod
uceadrivingforceonacommondisc.Butinsomemodels,theremay betwodiscsattac
hedtoacommonspindlewithoneelementattopandtheotheratbottom.
Threephase,Threewireenergymeter
(2) 3-Phase,4-wireACEnergymeter.
Thismeterhasthree1-
Phaseelementsworkingonthreediscsfixedtothesamespindlepivotedonajewelle
dbottombearingandapintypetopbearing.
Thediscsarearrangedtorevolveintheairgapofthetwoelectromagnetsonesup
pliedwithcurrentcoilsandothertothevoltage(potentialcoil)Thefluxduetothetwoe
lectromagnetsproducesarotatingfield.
Thetorqueandsothespeedofdiscisproportionaltothepower.Ifweconsider
thetimethenthenumberof revelutionsofthediscisproportionaltotheenergy.Foradj
ustmentofpowerfactorandfriction,shadingringsareprovided.Thedampingisprovi
dedwithpermanentmagnets.Thespindleisgearedbyachainofwheelstothecounte
rwhichindicatestheenergy suppliedthroughthemeter.
Theprincipleofthismeterissameasofthree wattmetermethodmeasuringpow
er,asshowninthefigure,thepowerisW+W+W 1 2 3
ThreePhase,FourwireEnergymeter.
\
PowerFactorMeter
Thepowerfactormeterisusedtomeasurethecosinevalueofphaseanglebetwe
enappliedvoltageandcurrentinanACcircuitdirectly.Theworkingprincipleissi
milartodynamometerinstrument.Henceitisalsocalledasdynamometertypep.f.
meter.Ithasnocontroldevicei.e.,arrangementforcontrolingtorque.Therearetw
otypes.,
(1)Singlephase p.f.meter (2)Threephase p.f.meter.
SinglePhaseDynamometerType P.F.meter
Principle
PowerfactormeterisusedtomeasurethePowerFactorofacircuit/load.Dyna
mometertypePowerFactormeterworksontheprincipleofdynamometer.Itstatestha
t,wheneveracurrentcarryingconductorisplacedinamagneticfield,itexperiences
aforceandmovesawayfromthemagneticfield.
Construction
TheconstructionaldetailsofPowerfactormeterisasshown inthefigure.Thisi
nstrumenthasastationarycoilwhichisdividedintotwosectionsFandF 1 2
andtwomovingcoils1 a whicharerigidlyattachedtoeachotherand
C ndC 2
mountedonthesamespindle.Thefixedcoilsarecalledcurrentcoilsandthemoving
coilsarepressurecoilsorvoltagecoils.
SinglePhaseDynamometerTypeP.F.meter
Working
WhenPowerfactormeterisconnectedinthecircuit,aphasedifference
900iscreatedbetween,thecurrentI 1 a becauseoftheresistanceand
ndI 2
inductanceofthemovingcoils.BecauseofthiscurrentIandI
1 2 flowinthe
movingcoil.Hencethemovingcoilofthemeterdeflects,dependinguponthepower
factorandshowsthereadingwiththehelpofpointer.
F1
F2
R1 I1
Sup-
ply
Load
R2
ThreePhaseDynamometerTypeP.F. meter
MeasurementofPower-Watt-meters
Tillnowwehavestudiedaboutmeasurementofcurrentandvoltage(acordc)by
AmmeterandVoltmeter.NowwewilllearnaboutmeasurementofElectricPowerby
Wattmeter.
Wattmetermeasurestheelectricpowerbeingconsumedinthecircuit/equip
mentwhereitisconnected.Smallwattmetersmeasurespowerinwattswhereasth
ewattmetersinelectricsubstationsandpowerstationsmeasuresthepowerinkilowat
ts(Kw)andmega watts(Mw). Wattmetersmainlyarefollowingtypes
1.Dynamometertypeormovingcoiltypewattmeter
2.Inductiontypewattmeter.
Fig2.19Wattmeter
DynamometerTypeInstruments
Electrodynamometertypeinstrumentisalsoamovingcoilinstrumentandcanb
eusedforACandDCmeasurements.WiththiswecanmeasureCurrent,Voltageand
Power.Butmainlyitisusedaswattmeter.Itsscaleisuniformwhenthisinstrumentisu
sedaswattmeter.
Principle
Thedynamometertypeinstrumentworkson the principleofElectrodynamiceff
ectofelectriccurrent(PrincipleofMotor).Itstatesthat“whenacurrentcarryingc
onductoris placedinamagneticfield,itexperiencesaforceandmovesawayfromthe
magneticfield”.
a) Sectionalview b)SchematicRepresentation
Dynamometertypeinstrument
Construction
TheinstrumentconsistsoftwofixedcoilsandamovablecoilThefixedcoilsar
emadeofafewturnsofthickcopperwireandcalledcurrentcoil.Themovablecoilof
finecopperwireandmadeonthinaluminiumframeandcalledpressurecoil.Itisplac
edonthespindlewhichwaspivotedontwobearings.Ahairspringisattachedtospin
dletoprovidecontrollingtorque.Dampingisprovidedbyeddycurrentsinducedin
aluminiumframe.Theotherarrangementsareasshowninfigure.
Working
Whentheinstrumentisconnectedinacircuit,thecurrentpassesthroughthebo
thcoils.Thefluxproducedby currentcoilisproportionaltoloadcurrentandthecurr
entthroughmovingcoilisproportionaltothevoltageacrosstheload.Becauseoft
hesetwotypes
offluxes,differentmagneticfieldsweredeveloped.Hencethemovingcoilexperie
ncesaforceandmoves(deflects).
Thedeflectingtorqueisproportionaltothe productofvoltageand currentinthetwo
coils.
Applications
TheDynamometertypeinstrumentcanbeusedasVoltmeter,Ammeterand
Wattmeter.
InductiontypeInstruments
Theinductiontypeofinstrumentshaswideapplicationsaswattmeterandener
gymeterbutcanalsobeusedasammeterandvoltmeter.Thedeflectingtorqueispr
oducedby theeddycurrents inducedinanlightaluminiumorcopperdiscordrumby
revolvingfluxcreatedbyelectromagnets.Ithasawidescalespreadover3000
InductiontypeinstrumentsareusedonlyforA.C.Measurements.
InductionWattmeterPrinciple
ThePrincipleofinductionwattmetersisthe sameasthat
ofinductionammeter.Theoperationofallinductiontypeinstrumentdependsup
ontheproductionoftorqueduetothereactionbetweentheflux( )andtheeddy
currentin ametaldiscduetootherflux.
Constructionandworking
Thewattmeterhastwolaminatedelectromagnets,oneofwhichisexcitedbyth
ecurrentinthemain circuitandcalledaseriesmagnet.Theotherisexcitedbycurrent
whichisproportionaltothevoltageofthecircuitandcalledshuntmagnet.
Fig2.21Inductiontypewatt-meter
Athinaluminumdiscissomountedthatitscutsthefluxofbothmagnets.Hence
twoeddycurrentsareproducedinthedisc,andtorqueisproducedinthedisc.Thed
eflectiontorqueisproducedduetotheinteractionoftheseeddycurrentsandinduci
ngflux.Twoorthreecopperringsarefittedonthecentrallimbof theshuntmagneta
ndcanbealsoadjustedtomaketheresultantfluxintheshuntmagnetlagbehindtheap
pliedvoltageby900.
Inductionwattmetershownaboveisspring-
controlled.Thespringbeingfittedtothespindleofthemovingsystemwhichalsoca
rriesthepointer.Thescaleisuniformlyevenandextendsover3000.
Currentsupto100Acanbehandledbysuchwattmeterdirectly.Butforcurre
ntsgreaterthanthisvaluetheyareusedinconjuctionwithcurrenttransformer.Th
epressurecoilispurposelymadeasmuchinductiveaspossibleorderthatthefluxthr
oughitshouldlagbehindthevoltageby900.
AdvantagesandlimitationofInductionWattmeter
Thesewattmeterspossestheadvantagesoffairlylongscaleextendingover3000
.
Theyarefreeformtheeffectsofstrayfields,andhavegooddamping.They arepra
cticallyfreefromfrequencyerrors.Howevertheyaresubjectedto(sometimes)seri
oustemperatureerrors,becausethemain effectoftemperatureisontheresistanceo
ftheeddycurrentpaths.
UNIT 4
INSTRUMENT TRANSFORMERS