BeneLink Module

Operator’s Manual
© Copyright 2016-2019 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All
rights reserved.
For this Operator’s Manual, the issue date is 2019-1.

WARNING

 Federal Law (USA) restricts this device to sale by or on the order of a


physician.

I
Intellectual Property Statement
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called
Mindray) owns the intellectual property rights to this Mindray product and this
manual. This manual may refer to information protected by copyrights or patents and
does not convey any license under the patent rights of Mindray, nor the rights of
others.

Mindray intends to maintain the contents of this manual as confidential information.


Disclosure of the information in this manual in any manner whatsoever without the
written permission of Mindray is strictly forbidden.

Release, amendment, reproduction, distribution, rental, adaption and translation of


this manual in any manner whatsoever without the written permission of Mindray is
strictly forbidden.

and are the registered trademarks or trademarks

owned by Mindray in China and other countries. All other trademarks that appear in
this manual are used only for editorial purposes without the intention of improperly
using them. They are the property of their respective owners.

Responsibility on the Manufacturer Party


Contents of this manual are subject to changes without prior notice.

All information contained in this manual is believed to be correct. Mindray shall not
be liable for errors contained herein nor for incidental or consequential damages in
connection with the furnishing, performance, or use of this manual.

Mindray is responsible for the effects on safety, reliability and performance of this
product, only if:

II
 all installation operations, expansions, changes, modifications and repairs of
this product are conducted by Mindray authorized personnel;

 the electrical installation of the relevant room complies with the applicable
national and local requirements;

 the product is used in accordance with the instructions for use.

WARNING

 This equipment must be operated by skilled/trained clinical


professionals.

 It is important for the hospital or organization that employs this


equipment to carry out a reasonable service/maintenance plan.
Neglect of this may result in machine breakdown or personal injury.

Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS
FOR ANY PARTICULAR PURPOSE.

Exemptions
Mindray's obligation or liability under this warranty does not include any
transportation or other charges or liability for direct, indirect or consequential
damages or delay resulting from the improper use or application of the product or
the use of parts or accessories not approved by Mindray or repairs by people other
than Mindray authorized personnel.

III
This warranty shall not extend to

 Malfunction or damage caused by improper use or man-made failure.

 Malfunction or damage caused by unstable or out-of-range power input.

 Malfunction or damage caused by force majeure such as fire and


earthquake.

 Malfunction or damage caused by improper operation or repair by


unqualified or unauthorized service people.

 Malfunction of the instrument or part whose serial number is not legible


enough.

 Others not caused by instrument or part itself.

Company Contact
Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Mindray Building, Keji 12th Road South, Hi-tech industrial park,
Address:
Nanshan,Shenzhen 518057,P.R.China
Website: www.mindray.com
E-mail Address: service@mindray.com
Tel: +86 755 81888998
Fax: +86 755 26582680

EC-Representative: Shanghai International Holding Corp. GmbH(Europe)


Address: Eiffestraβe 80, 20537 Hamburg, Germany
Tel: 0049-40-2513175
Fax: 0049-40-255726

IV
Contents
1 Safety .........................................................................................................................1-1

1.1 Safety Information ..................................................................................................................... 1-1


1.2 Warnings ....................................................................................................................................... 1-2
1.3 Cautions ........................................................................................................................................ 1-2
1.4 Notes.............................................................................................................................................. 1-3

2 Interfacing with External Devices ..............................................................................2-1

2.1 BeneLink Introduction .............................................................................................................. 2-1


2.2 Physical Specifications .............................................................................................................. 2-1
2.3 Data transmission ....................................................................................................................... 2-1
2.4 Supported Devices ..................................................................................................................... 2-2
2.4.1 Supported Anesthesia Machines .............................................................................. 2-2
2.4.2 Supported Ventilators .................................................................................................. 2-3
2.4.3 Supported tcGas Monitoring Devices ...................................................................... 2-4
2.4.4 Supported Infusion System (for BeneVision N Series Monitors) ....................... 2-5
2.4.5 Supported Other Devices............................................................................................ 2-5
2.5 Connecting External Devices .................................................................................................. 2-7
2.6 Setting the ID Adapter .............................................................................................................. 2-8
2.7 Accessories ................................................................................................................................... 2-9

3 Integrating the Anesthesia Machine ..........................................................................3-1

3.1 Draeger Apollo ............................................................................................................................ 3-1


3.1.1 Output Signals—Parameters...................................................................................... 3-1
3.1.2 Output Signals—Alarms .............................................................................................. 3-5
3.1.3 Output Signals – Waveforms (For BeneVision N series Monitors) ...................3-10
3.2 Draeger Fabius GS /Fabius plus/Fabius Tiro.......................................................................3-10
3.2.1 Output Signals - Parameters .....................................................................................3-10
3.2.2 Output Signals - Alarms .............................................................................................3-13
3.3 Draeger Perseus A500 .............................................................................................................3-14

1
3.3.1 Output Signals—Parameters ................................................................................... 3-14
3.3.2 Output Signals—Alarms ........................................................................................... 3-18
3.3.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................... 3-22
3.4 Draeger Primus ......................................................................................................................... 3-23
3.4.1 Output Signals - Parameters .................................................................................... 3-23
3.4.2 Output Signals - Alarms............................................................................................. 3-27
3.4.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................... 3-31
3.5 Draeger Zeus/ZeusIE ............................................................................................................... 3-32
3.5.1 Output Signals—Parameters ................................................................................... 3-32
3.5.2 Output Signals—Alarms ........................................................................................... 3-36
3.5.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................... 3-39
3.6 GE Aespire 7900&7100/ Aestiva 7900&7100..................................................................... 3-40
3.6.1 Output Signals - Parameters .................................................................................... 3-40
3.6.2 Output Signals - Alarms............................................................................................. 3-41
3.6.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................... 3-44
3.7 GE AespireView ......................................................................................................................... 3-44
3.7.1 Output Signals - Parameters .................................................................................... 3-44
3.7.2 Output Signals - Alarms............................................................................................. 3-46
3.7.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................... 3-49
3.7.4 Output Signals – Respiratory Loops (For BeneVision N series Monitors) ..... 3-49
3.8 GE Aisys/Avance/Aisys CS2 .................................................................................................... 3-49
3.8.1 Output Signals - Parameters .................................................................................... 3-49
3.8.2 Output Signals - Alarms............................................................................................. 3-55
3.8.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................... 3-60
3.8.4 Output Signals – Respiratory Loops (For BeneVision N series Monitors) ..... 3-61
3.9 HuL Leon .................................................................................................................................... 3-61
3.9.1 Output Signals—Parameters ................................................................................... 3-61
3.9.2 Output Signals—Alarms ........................................................................................... 3-64
3.10 HuL Leon Plus ......................................................................................................................... 3-70
3.10.1 Output Signals—Parameters ................................................................................. 3-70
3.10.2 Output Signals—Alarms ......................................................................................... 3-73

2
3.11 MAQUET FLOW-i ..................................................................................................................... 3-79
3.11.1 Output Signals - Parameters .................................................................................. 3-79
3.11.2 Output Signals - Alarms........................................................................................... 3-82
3.12 Mindray A7/A5/A4/A3/Mindray WATO 25/60/50/30/20/35/55/65/55/65 ............... 3-85
3.12.1 Output Signals - Parameters .................................................................................. 3-85
3.12.2 Output Signals - Alarms........................................................................................... 3-89
3.12.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................. 3-96
3.12.4 Output Signals – Respiratory Loops (For BeneVision N series Monitors) ... 3-96

4 Integrating the Ventilator ..........................................................................................4-1

4.1 AirLiquide ALMS Monnal T75 .................................................................................................. 4-1


4.1.1 Output Signals—Parameters...................................................................................... 4-1
4.1.2 Output Signals—Alarms .............................................................................................. 4-2
4.2 Carefusion Avea .......................................................................................................................... 4-5
4.2.1 Output Signals - Parameters ....................................................................................... 4-5
4.2.2 Output Signals - Alarms ............................................................................................... 4-8
4.2.3 Output Signals - Waveforms (For BeneVision N series Monitors).................... 4-10
4.3 Carefusion Vela.......................................................................................................................... 4-10
4.3.1 Output Signals - Parameters ..................................................................................... 4-10
4.3.2 Output Signals - Alarms ............................................................................................. 4-13
4.4 Draeger Babylog 8000 plus/Babylog 8000 ........................................................................ 4-15
4.4.1 Output Signals—Parameters.................................................................................... 4-15
4.4.2 Output Signals—Technical Alarms ......................................................................... 4-16
4.5 Draeger Babylog VN500 ......................................................................................................... 4-17
4.5.1 Output Signals—Parameters.................................................................................... 4-17
4.5.2 Output Signals - Alarms ............................................................................................. 4-20
4.6 Draeger Evita 2 .......................................................................................................................... 4-23
4.6.1 Output Signals - Parameters ..................................................................................... 4-23
4.6.2 Output Signals - Alarms ............................................................................................. 4-26
4.7 Draeger Evita2 dura/Evita 4/Evita XL................................................................................... 4-28
4.7.1 Output Signals - Parameters ..................................................................................... 4-28

3
4.7.2 Output Signals - Alarms............................................................................................. 4-32
4.7.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................... 4-35
4.8 Draeger Evita V300 .................................................................................................................. 4-36
4.8.1 Output Signals—Parameters ................................................................................... 4-36
4.8.2 Output Signals—Alarms ........................................................................................... 4-41
4.9 Draeger Infinity V500 .............................................................................................................. 4-44
4.9.1 Output Signals—Parameters ................................................................................... 4-44
4.9.2 Output Signals—Alarms ........................................................................................... 4-48
4.10 Draeger Savina 300 ............................................................................................................... 4-51
4.10.1 Output Signals—Parameters ................................................................................. 4-51
4.10.2 Output Signals—Alarms ......................................................................................... 4-54
4.11 GE Carescape R860 ................................................................................................................ 4-57
4.11.1 Output Signals—Parameters ................................................................................. 4-57
4.11.2 Output Signals—Alarms ......................................................................................... 4-61
4.11.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................. 4-66
4.11.4 Output Signals – Respiratory Loops (For BeneVision N series Monitors) ... 4-66
4.12 GE Engström Carestation ..................................................................................................... 4-66
4.12.1 Output Signals - Parameters .................................................................................. 4-66
4.12.2 Output Signals - Alarms .......................................................................................... 4-71
4.13 Hamilton C2/C3/C1 /C1 NEO/C6/C6S/T1 ......................................................................... 4-75
4.13.1 Output Signals - Parameters .................................................................................. 4-75
4.13.2 Output Signals - Alarms .......................................................................................... 4-78
4.14 Hamilton G5 (Block Protocol) ............................................................................................. 4-79
4.14.1 Output Signals - Parameters .................................................................................. 4-79
4.14.2 Output Signals - Alarms .......................................................................................... 4-82
4.14.3 Output Signals - Waveforms (For BeneVision N series Monitors) ................. 4-87
4.15 Hamilton G5 (Polling Protocol)........................................................................................... 4-88
4.15.1 Output Signals - Parameters .................................................................................. 4-88
4.15.2 Output Signals - Alarms .......................................................................................... 4-90
4.16 Hamilton Galileo .................................................................................................................... 4-91
4.16.1 Output Signals - Parameters .................................................................................. 4-91

4
4.16.2 Output Signals - Alarms........................................................................................... 4-94
4.17 Hamilton S1 ............................................................................................................................. 4-95
4.17.1 Output Signals—Parameters ................................................................................. 4-95
4.17.2 Output Signals—Alarms ......................................................................................... 4-99
4.17.3 Output Signals - Waveforms (For BeneVision N series Monitors) .............. 4-104
4.18 HuL Leoni Plus...................................................................................................................... 4-105
4.18.1 Output Signals—Parameters .............................................................................. 4-105
4.18.2 Output Signals—Alarms ...................................................................................... 4-107
4.19 MAQUET SERVO-i/SERVO-s............................................................................................... 4-111
4.19.1 Output Signals - Parameters ............................................................................... 4-111
4.19.2 Output Signals - Alarms........................................................................................ 4-114
4.19.3 Output Signals - WaveForms (for BeneVision N Series Monitors) .............. 4-117
4.19.4 Output Signals -Respiratory Loops (for BeneVision N Series Monitors)... 4-117
4.20 MAQUET SERVO-U/N.......................................................................................................... 4-118
4.20.1 Output Signals—Parameters .............................................................................. 4-118
4.20.2 Output Signals—Alarms ...................................................................................... 4-121
4.20.3 Output Signals—Waveforms (for BeneVision N Series Monitors) ............. 4-123
4.20.4 Output Signals—Respiratory Loops (for BeneVision N Series Monitors) 4-124
4.21 Mindray SV300/350/600/650/800/850 .......................................................................... 4-124
4.21.1 Output Signals - Parameters ............................................................................... 4-124
4.21.2 Output Signals - Alarms........................................................................................ 4-128
4.21.3 Output Signals - Waveforms (For BeneVision N series Monitors) .............. 4-136
4.21.4 Output Signals – Respiratory Loops (For BeneVision N series Monitors) 4-136
4.22 Mindray SynoVent E3/E5 ................................................................................................... 4-136
4.22.1 Output Signals - Parameters ............................................................................... 4-136
4.22.2 Output Signals - Alarms........................................................................................ 4-140
4.23 Newport E360 ...................................................................................................................... 4-147
4.23.1 Output Signals - Parameters ............................................................................... 4-147
4.23.2 Output Signals - Alarms........................................................................................ 4-149
4.24 Philips Respironics V60 (SNDA Protocol)....................................................................... 4-153
4.24.1 Output Signals - Parameters ............................................................................... 4-153

5
4.24.2 Output Signals - Alarms ........................................................................................4-153
4.25 Philips Respironics V60 (VRPT Protocol) ........................................................................4-154
4.25.1 Output Signals - Parameters ................................................................................4-154
4.25.2 Output Signals - Alarms ........................................................................................4-155
4.26 Puritan Bennett PB840/PB980 (SNDF Protocol) ...........................................................4-157
4.26.1 Output Signals - Parameters ................................................................................4-157
4.26.2 Output Signals - Alarms ........................................................................................4-159
4.27 Puritan Bennett PB840/PB980 (SNDA Protocol) ..........................................................4-161
4.27.1 Output Signals - Parameters ................................................................................4-161
4.27.2 Output Signals - Alarms ........................................................................................4-163
4.28 ResMed VSIII ..........................................................................................................................4-164
4.28.1 Output Signals—Parameters ...............................................................................4-164
4.28.2 Output Signals—Alarms .......................................................................................4-165

4.29 Salvia Elisa 800/600/500/300............................................................................................4-166


4.29.1 Output Signals - Parameters ................................................................................4-166
4.29.2 Output Signals - Alarms ........................................................................................4-170

4.29.3 Output Signals - Waveforms (For BeneVision N series Monitors) ............4-173


5 Integrating the tcGas Monitoring Device ...................................................................5-1

5.1 Biometer TCM40 ..........................................................................................................................5-1


5.1.1 Output Signals - Parameters .......................................................................................5-1
5.1.2 Output Signals - Alarms................................................................................................5-1
5.2 Biometer TCM4/TCM 5/TCM CombiM ....................................................................................5-2
5.2.1 Output Signals - Parameters .......................................................................................5-2
5.2.2 Output Signals - Alarms................................................................................................5-2
5.3 Biometer TCM Tosca ...................................................................................................................5-3
5.3.1 Output Signals - Parameters .......................................................................................5-3
5.3.2 Output Signals - Alarms................................................................................................5-3
5.4 SenTec Digital Monitor ..............................................................................................................5-4
5.4.1 Output Signals - Parameters .......................................................................................5-4
5.4.2 Output Signals - Alarms................................................................................................5-4

6
6 Integrating Infusion System (for BeneVision N Series Monitors) ...............................6-1

6.1 B.Braun Infusion Space/Perfusor Space Pumps.................................................................. 6-1


6.1.1 Output Signals - Parameters ....................................................................................... 6-1
6.1.2 Output Signals – Technical Alarms............................................................................ 6-1
6.2 MedimaNet Server Infusing Monitoring and Archiving Network System ................... 6-2
6.2.1 Output Signals - Parameters ....................................................................................... 6-2
6.3 Fresenius Agilia Injectomat Infusion Pumps ....................................................................... 6-2
6.3.1 Output Signals - Parameters ....................................................................................... 6-2
6.3.2 Output Signals – Technical Alarms............................................................................ 6-3
6.4 Fresenius Agilia Volumat Infusion Pumps ............................................................................ 6-4
6.4.1 Output Signals - Parameters ....................................................................................... 6-4
6.4.2 Output Signals – Technical Alarms............................................................................ 6-4
6.5 Mindray BeneFusion DS5 Infusion Supervision System (For BeneVision N series
Monitors) ............................................................................................................................................. 6-5

6.5.1 Output Signals - Parameters ....................................................................................... 6-5


6.5.2 Output Signals – Technical Alarms............................................................................ 6-6

7 Integrating Other Devices ..........................................................................................7-1

7.1 MAQUET IABP .............................................................................................................................. 7-1


7.1.1 Output Signals - Parameters ....................................................................................... 7-1
7.1.2 Output Signals - Alarms ............................................................................................... 7-1
7.2 MAQUET PulsioFlex Cardiac Monitoring Device ................................................................ 7-3
7.2.1 Output Signals - Parameters ....................................................................................... 7-3
7.2.2 Output Signals - Alarms ............................................................................................... 7-5
7.3 Narcotrend Compact M EEG Monitoring Devices .............................................................. 7-6
7.3.1 Output Signals - Parameters ....................................................................................... 7-6
7.3.2 Output Signals -Technical Alarms ............................................................................. 7-7
7.3.3 Output Signals - Waveforms (For BeneVision N series Monitors)...................... 7-8
7.4 Organon TOF-Watch SX NMT Monitoring Device .............................................................. 7-8
7.4.1 Output Signals - Parameters ....................................................................................... 7-8
7.4.2 Output Signals - Alarms ............................................................................................... 7-9

7
8 BeneLink Function Test ..............................................................................................8-1

8.1 Preparing the Tools .....................................................................................................................8-1


8.2 Checking the Function ..............................................................................................................8-1
8.3 Setting the Virtual Serial Port (For BeneVision N series Monitors) .................................8-2
8.3.1 Setting the Virtual Serial Port Properties .................................................................8-2
8.3.2 Setting the ID of the ID Adapter .................................................................................8-3
8.3.3 Parsing the Monitor Data .............................................................................................8-3
8.4 Setting the Serial Ports ..............................................................................................................8-4
8.4.1 Setting the Serial Ports of the Anesthesia Machines ............................................8-4
8.4.2 Setting the Serial Ports of the Ventilator ..................................................................8-6
8.4.3 Setting the Serial Ports of the Pumps .................................................................... 8-10
8.4.4 Setting the Serial Ports of the TcGas Monitoring Devices ................................ 8-12
8.4.5 Setting the Serial Ports of Other External Devices ............................................. 8-13
8.4.6 Installation and Test Report...................................................................................... 8-14

9 Alarm ID of External Devices ......................................................................................9-1

9.1 Alarm IDs of Anesthesia Machines .........................................................................................9-1


9.2 Alarm IDs of Ventilators .............................................................................................................9-5
9.3 Alarm IDs of tcGas monitoring devices .............................................................................. 9-10
9.4 Alarm IDs of NMT monitoring device ................................................................................. 9-11
9.5 Alarm IDs of Infusion Systems (for BeneVision N series Monitors).............................. 9-11

8
1 Safety

1.1 Safety Information

WARNING

 Indicates a potential hazard or unsafe practice that, if not avoided,


could result in death or serious injury.

CAUTION

 Indicates a potential hazard or unsafe practice that, if not avoided,


could result in minor personal injury or product/property damage.

NOTE

 Provides application tips or other useful information to ensure that you


get the most from your product.

1-1
1.2 Warnings
WARNING

 Devices of the same category can not be connected to the BeneLink


module simultaneously.

 A patient monitor supports up to two BeneLink module sonly.

 The BeneLink module receives parameter, alarm, and waveform data


from external devices but will not modify this data. For example, the
external device may display "---", but send "0" to the patient monitor.

 The signal labels used on the patient monitor may be different from
those given on the external device.

 The alarms from external devices may be delayed before transmission


to the patient monitor.

 There can be differences between the alarm priorities displayed on


your monitors and the priorities displayed on external devices
interfaced through BeneLink.

 When the monitor connects external devices via the BeneLink module,
the alarm delay for disconnection is less than 10 seconds.

1-2
1.3 Cautions
CAUTION

 First installation and debugging should be executed by Mindray


service personnel or authorized technician.

 Please check the compatibility of the external device and the ID


adapter before connection. Otherwise, unpredictable system failure
may be resulted.

 Ports on the BeneLink module are not conventional network


connectors. They are intended for connecting with the serial port of
designated devices only. Do not connect them to public network
interfaces.

1.4 Notes
NOTE

 The alarm messages from external devices are derived from the open
protocol of corresponding external device. For more information about
these alarms, please see the operator’s manual of corresponding
devices.

 For the intended use and contraindication of the external devices, refer
to their operator’s manuals.

 When displayed in the monitor main screen, parameter labels of


external devices are prefixed with the plug sign “+”. For example, if
SpO2 is from an external device, its label is displayed as “+SpO2”, and
its waveform label is displayed as “+Pleth”.

 A plus sign “+” is added before each alarm message from external
devices.

1-3
FOR YOUR NOTES

1-4
2 Interfacing with External Devices

2.1 BeneLink Introduction


BeneLink module is intended for connecting external devices, such as ventilators and
anesthesia machines, to the BeneVision N series monitor or BeneView T series monitor.
It allows information (patient data, alarms, etc.) from external devices to be displayed,
saved, recorded, or printed through the monitor. If the monitor is connected with the
central monitoring system (CMS) or eGateway, information from external devices can
also be transmitted to the CMS or eGateway.

2.2 Physical Specifications


The physical specifications of the BeneLink module are as follows:

 Size:136.5×40×102 mm

 weight:0.35 kg

2.3 Data transmission


The Benelink module is connected with external devices via the serial ports (physical
interfaces are RJ45 connectors). You can simultaneously connect up to four external
devices via one BeneLink module. Parameters, respiratory waveforms, loops, alarm
data from the external devices can be output.

2-1
2.4 Supported Devices
Please contact us or our service personnel for the most recent information on the
supported devices.

2.4.1 Supported Anesthesia Machines


Brand Model ID for ID Type of serial port
adapter adapting cable

Draeger Apollo 444FBBB1 Type C


Fabius GS 4446BBBA Type C
Fabius plus/Tiro 4446BBBA Type C
Perseus A500 4435BBCB No need
Primus 4450BBB0 Type C
Zeus/ZeusIE 445ABBA6 No need
GE Aespire 7900/7100 4F37B0C9 Type D
Aespire View 4145bebb Type D
Aestiva 7900/7100 4F37B0C9 Type D
Aisys 4F41B0BF Type D
Aisys CS2 4F41B0BF Type D
Avance 4F41B0BF Type D
HuL Leon 484CB7B4 Type C
Leon Plus 4850B7B0 Type C
MAQUET FLOW-i 4D46B2BA Type B
Mindray A7/A5/ A4 /A3 4D52B2AE No need
(A4 is for BeneVision
N Series monitors
only)
WATO 4D52B2AE No need
EX-65/55/35/25/60/5
0/30/20/65 Pro/55
Pro

2-2
2.4.2 Supported Ventilators
Brand Model ID for ID Type of serial
adapter port adapting
cable

AirLiquide ALMS Monnal T75 4154BEAC No need


Carefusion Avea 5645a9bb Type E
Vela 564CA9B4 Type E
Draeger Babylog8000/ Babylog8000 4442BBBE Type B
plus
Evita 2/Evita 2 dura/Evita XL 4434BBCC Type B
Evita 4 4434BBCC Type B
Evita V300 4433BBCD No need
Infinity V500/Babylog VN500 4456BBAA No need
Savina 300 4441BBBF Type B
GE Carescape R860 4F52B0AE Type B
Engström(Carestation) 4F45B0BB Type B
Hamilton C2/C3/C1/T1/C1 3270CD90 Type B
NEO/C6/C6S(C1/T1/C1
NEO/C6/C6S is for BeneVision
N Series monitors only)
G5 (Block Protocol) 3542CABE Type B
G5 (Polling Protocol) 3550CAB0 Type B
Galileo (polling) 4750B8B0 Type B
S1 (Block Protocol) 5331ACCF Type B
S1 (Polling Protocol) 3550CAB0 Type B
HuL Leon i Plus 4849B7B7 Type C
MAQUET SERVO-i/SERVO-s 4D53B2AD Type B
SERVO-U/N 4d55B2AB Type B
(SERVO-N is for BeneVision
N Series monitors only)

2-3
Brand Model ID for ID Type of serial
adapter port adapting
cable

Mindray SV 300/350/600/650/800/850 4D56B2AA No need


SynoVent E3/ SynoVent E5 4D56B2AA No need
Newport E360 4E50B1B0 Type B
Philips Respironics V60 5637A9C9 25-pin (male) to
(VRPT) 9-pin (female)+
5636A9CA Type B
(SDNA)
Puritan PB840 5042AFBE No need
Bennett (SNDF protocol)
PB840 5031AFCF No need
(SNDA protocol)
PB980 5042AFBE Type C
(SNDF protocol)
PB980 5031AFCF Type C
(SNDA protocol)
ResMed VSIII 5653A9AD Type C
Salvia Elisa 800/600/500/300 5341acbf No need

2.4.3 Supported tcGas Monitoring Devices


Brand Model ID for ID Type of serial port
adapter adapting cable
Biometer TCM40/TCM Tosca 5443ABBD Type C
TCM4/TCM CombiM
TCM5 5443ABBD Type B
SenTec SenTec Digital Monitor 5354ACAC Type C

2-4
2.4.4 Supported Infusion System (for BeneVision N Series
Monitors)
Brand Model ID for ID Type of serial port
adapter adapting cable

B.Braun Infusomat Space 4250BDB0 Type F


Perfusor Space 4250BDB0 Type F
Fresenius Agilia Injectomat 4650b9b0 Type G
Agilia Injectomat MC 4650b9b0 Type G
Agilia Injectomat TIVA 4650b9b0 Type G
Agilia Volumat 4650b9b0 Type G
Agilia Volumat MC 4650b9b0 Type G
Medima MedimaNet Server 4D50B2B0 Directly connects
the port of
TCP/RS232
adapting module
Mindary BeneFusion DS5 534bacb5 Type H

2.4.5 Supported Other Devices


Brand Model ID for ID Type of serial
adapter port adapting
cable

MAQUET/cardi PulsioFlex (for 5055afab USB to serial


ac monitoring BeneVision N series adapter(ATEN
device monitors) USB to Serial
Adapter UC-232A)
MAQUET/IABP IABP(CS300,CS100) (for 4941b6bf Mail to mail,
BeneVision N series directly
monitors) connected.
2-5
Brand Model ID for ID Type of serial
adapter port adapting
cable

Maximum size of
the connector is
15 mm × 33 mm.
Narcotrend/EE Compact M (for 4e43b1bd No Need
G monitoring BeneVision N series
device monitors)
Organon/ NMT TOF Watch SX 5457ABA9 Type C
monitoring
device

2-6
2.5 Connecting External Devices
The external devices are connected with the BeneLink module through an ID adapter,
which supports only its matching device. Please refer to the following procedure to
connect an external device:
(5)

(3) (4)
(1)
(2)

(6)

(1) BeneLink module (2) Label


(3) RJ45 connecting cable (4) ID adapter
(5) External device (6) Serial port adapting cable (optional)

1. Make sure that BeneLink contain the device drivers which match the devices you want to
connect. If it doesn’t, you could install driver to BeneLink by using PC upgrade program.
After upgrade, stick a label on BeneLink module to indicate installed device drives.

2. Insert the BeneLink module into the SMR or module rack.

3. Connect the ID adapter that matches the external device to the BeneLink
module with an RJ45 connecting cable.

4. Plug the ID adapter into the RS232 port on the external device. Some external
devices may have ports incompatible with the ID adapter. In this case, a serial
port adapting cable is required.

2-7
5. Stick a device name label to the RJ45 connecting cable at the end close to the
BeneLink module. When the BeneLink module is connected to several external
devices, you can tell devices easily with these labels.

6. Switch on the external device.

After the external device is connected to the monitor, the indicators on both the ID
adapter and the BeneLink module illuminate to show that the monitor successfully
communicates with the external device.

2.6 Setting the ID Adapter


The ID adapter has already been correctly configured before leaving the factory. If
you want to re-configure the ID adapter, follow this procedure:

1. Select the [Main Menu] quick key → turn to the third page → from the
[System] column select [Maintenance] → input the factory maintenance
password → select [Ok] → select the [Factory Maintenance] tab → select
the [Setup] tab→ select the [Update ID module] button.

2. Set [Benelink Module Port] and [ID].

 The setting of[Benelink Module Port] shall be consistent with the port
the RJ45 connecting cable is connected to.

 You must connect the RJ45 connecting cable to the selected port when
re-configuring the ID adapter. Otherwise, ID adapter re-configuration
will fail.

 Set [ID] to configure a new ID to the ID adapter. Refer to Column “ID for
ID Adapter” for the ID information for each external device.

2-8
2.7 Accessories
PN Descripiton

115-
115-00727
15- adapter

Serial port adapting cable, Type A, male to female,


009-001767-00
direct-connected

Serial port adapting cable, Type B, male to male,


009-001768-00
cross-over

009-001769-00 Serial port adapting cable, Type C,male to male

009-002943-00 Serial port adapting cable, Type D, 9-pin to 15-pin

009-004613-00 Serial port adapting cable, Type E, 9-pin to RJ45 connector

009-008485-00 Serial port adapting cable, Type F, 9-pin

009-008624-00 Serial port adapting cable, Type G, 9-pin

009-009488-00 Serial port adapting cable, Type J

009-004286-00 Serial cable assembly

047-004857-00 ID adapter label

047-004859-00 Network line label

009-001770-00 RJ45 connecting cable

2-9
FOR YOUR NOTES

2-10
3 Integrating the Anesthesia Machine

3.1 Draeger Apollo


3.1.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Plateau
Pplat cmH2O, hPa, mbar Yes
pressure
Pmean cmH2O, hPa, mbar Mean pressure Yes
Inspired tidal
VTi ml Yes
volume
MV L/min Minute volume Yes
Spontaneous
fspn bpm Yes
respiratory rate
Compl ml/cmH2O,ml/hPa,ml/mbar Compliance Yes
Respiratory rate
RRCO2 bpm Yes
of CO2
End-tidal
EtCO2 %,kPa,mmHg Yes
carbon dioxide
Fraction of
FiCO2 %,kPa,mmHg inspired carbon Yes
dioxide

3-1
Label Units Description Trend, record, print
Fractional
concentration
FiO2 %,kPa,mmHg Yes
of O2 in inspired
gas
EtO2 %,kPa,mmHg End-tidal O2 Yes
Difference
between
ΔO2 %,kPa,mmHg No
inspiratory and
expiratory O2
Tapnea sec Apnea time No
FiN2O % Yes
FiIso %
Inspired
FiDes %
anesthetic
FiEnf %
agent
FiSev %
FiHal %
EtN2O % Yes
EtEnf %
End-tidal
EtDes %
anesthetic
EtIso %
agent
EtSev %
EtHal %
Inspired Yes
FiAA % anesthetic
agent
End-tidal Yes
EtAA % anesthetic
agent
FiAA 2nd % 2nd Insp. Agent Yes

3-2
Label Units Description Trend, record, print
EtAA 2nd % 2nd Exp. Agent Yes
Inspired
Insp. minimum
\ No
MAC alveolar
concentration
Expired
minimum
Exp. MAC \ No
alveolar
concentration
HALLev ml
ENFLev ml Anesthetic
ISOLev ml agent No
DESLev ml consupmtion
SEVLev ml
N2O Flow L/min N2O flow No
Air Flow L/min Air flow No
O2 Flow L/min O2 flow No
% Arterial oxygen
SpO2 saturation from Yes
pulse oximetry
PR bpm Pulse rate Yes
Oxygen
VO2 ml/min Yes
consumption
Expiratory tidal
VTe ml Yes
volume
EtO2% % End-tidal O2 Yes

3-3
Label Units Description Trend, record, print
Fractional
concentration
FiO2% % Yes
of O2 in inspired
gas
Fraction of
EtCO2% % inspired carbon Yes
dioxide
End-tidal
FiCO2% % Yes
carbon dioxide
Difference
between
ΔO2 % % No
inspiratory and
expiratory O2
Setting Parameters
ml Inspired tidal No
VTi
volume
f bpm Breath rate No
bpm Minimum No
FreqMIN breath
frequency
% Percentage of No
inspiratory
TIP:TI
plateau time in
inspiratory time
sec Time for the No
pressure to rise
Tslope
to target
pressure

3-4
Label Units Description Trend, record, print
sec Time of No
Tinsp
inspiration
cmH2O, hPa, mbar Pressure control No
Pinsp level of
inspiration
Pressure No
Psupp cmH2O, hPa, mbar
support level
Maximal No
Pmax cmH2O, hPa, mbar breathing
pressure
Inspiratory No
F-Trigger L/min trigger level
(flow trigger)

3.1.2 Output Signals—Alarms


Mindray Patient Monitor Draeger Apollo Alarm

Priority Label Label


Physiological alarms
Apnea Apnea/Apnea Vent
Volume Apnea > 2
Apnea Vol
min
Pressure Apnea Apnea Pres
PAW High Paw High
High
PAW Low Paw Negtive
FiO2 Low % O2 LOW
CONT PRES PEEP > Pressure Threshold for 15 sec
CO2 Apnea Apnea CO2
No Pulse SPO2 NO PULS

3-5
Mindray Patient Monitor Draeger Apollo Alarm

Priority Label Label


PR Low SPO2 PULS LO
SpO2 Low SPO2 LOW
FiO2 High FiO2 High
VTe Low TIDAL VOL?
MV High MIN Vol HIGH
MV Low MIN Vol Low
PEEP High Peep High
EtCO2 High EtCO2 High
EtCO2 Low EtCO2 Low
FiCO2 High INSP CO2 HIGH
FiN2O High FI N2O HIGH
EtHAL High EXP. HAL HIGH
FiHal High %HAL HIGH
FiHal Low %HAL Low
Medium EtENF High EXP. ENF HIGH
FiENF High %ENF HIGH
FiENF Low %ENF Low
EtISO High EXP.ISO HIGH
FiISO High %ISO HIGH
FiISO Low %ISO Low
EtSEV High EXP.SEV HIGH
FiSEV High %SEV HIGH
FiSEV Low %SEV Low
EtDES High EXP.DES HIGH
FiDES High %DES HIGH
FiDES Low %DES Low
MAC Low MAC Low?

3-6
Mindray Patient Monitor Draeger Apollo Alarm

Priority Label Label


PR High SPO2 PULS HI
SpO2 High SPO2 HIGH
Technical alarms
O2 Supply Failure O2 Supply?
No Fresh Gas NO Fressh gas
Circuit Occluded CIRCLE OCCL
VENT DISC VENT DISC
If the monitor shows the alarm [High
Technical alarms], the external device
may have the alarms:
VENT ERR
INT.TMP.HIGH
High
O2CYL.DISCON
CHK N2O CYL
High Technical Alarm
NO N2O DELIV
NO AIR DELIV
FG-OVER?
VENT. UNLOCK
AW-TEMP HIGH
NO N2O
NO O2 DELIV
Patient Circuit Leak LEAKAGE
Fresh Gas? FRESH GAS?
If the monitor shows the alarm [AG
Medium Module abnormal], the external device
AG Module abnormal may have the following alarms:
MIXED AGENT
CO2/AGT ERR

3-7
Mindray Patient Monitor Draeger Apollo Alarm

Priority Label Label


N2O ERR
AGT ERR
2ND AGENT
If the monitor shows the alarm [CO2
Module abnormal], the external device
CO2 Module abnormal
may have the following alarms:
CO2 LINE BLK
If the monitor shows the alarm
[Medium Technical alarms], the
external device may have the following
alarms:
POWER FAIL
BATTERY LOW
N2O SUPPLY?
PRESSURE LIM
MIXER INOP
Medium Technical P MAX?
Alarm SAFETY O2 ON
FG FLOW LIM
LOSS OF DATA
% O2 ERR
SET CANCELED
FG TOO HIGH
FG ACTIVE
FG AIR SENS?
FG O2 SENS?
FG N2O SENS?
Low NO AIR NO AIR

3-8
Mindray Patient Monitor Draeger Apollo Alarm

Priority Label Label


NO O2 SUPPLY NO O2 SUPPLY
If the monitor shows the alarm [SpO2
Module abnormal], the external device
SpO2 Module may have the following alarms:
abnormal SPO2SEN DISC
SPO2 ALRM OF
SPO2 ERR
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
FAN ERR
PWR SPLY ERR
PRESS ERR
VOL ERR
LO O2 SUPPLY
CHK O2 CYL
O2 CYL OPEN
Low Technical Alarm N2O CYLOPEN
AIR CYL OPEN
COM VENT ERR
APOLLO COM1?
APOLLO COM2?
N2O CYL .SENS?
AIR CYL SENS?
O2 CYL .SENS?
AIR CYL?
PRESS RELIEF
INSP VOL ERR

3-9
3.1.3 Output Signals – Waveforms (For BeneVision N series
Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
CO2 mmHg, kPa, %

3.2 Draeger Fabius GS /Fabius plus/Fabius Tiro


3.2.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
cmH2O, hPa, Positive end-expiratory
PEEP Yes
mbar pressure
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
cmH2O, hPa,
Pplat Plateau pressure Yes
mbar
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
VTe ml Expiratory tidal volume Yes
MV L/min Minute volume Yes
Spontaneous
fspn bpm Yes
respiratory rate
RRCO2 bpm Respiratory rate of CO2 Yes
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Fraction of inspired
FiCO2 %,kPa,mmHg Yes
carbon dioxide

3-10
Label Units Description Trend, record, print
Fractional
FiO2 %,kPa,mmHg concentration of O2 in Yes
inspired gas
Fraction of inspired
FiN2O % Yes
nitrous oxide
EtN2O % End-tidal N2O Yes
FiDes % Yes
FiSev % Yes
Inspired anesthetic
FiEnf % Yes
agent
FiIso % Yes
FiHal % Yes
EtEnf % Yes
EtDes % Yes
End-tidal anesthetic
EtIso % Yes
agent
EtSev % Yes
EtHal % Yes
Inspired anesthetic
FiAA % Yes
agent
End-tidal anesthetic
EtAA % Yes
agent
N2O Flow L/min N2O flow No
Air Flow L/min Air flow No
O2 Flow L/min O2 flow No
Fractional
FiO2% % concentration of O2 in Yes
inspired gas
Fraction of inspired
FiCO2% % Yes
carbon dioxide

3-11
Label Units Description Trend, record, print
End-tidal carbon
EtCO2% % Yes
dioxide
Setting Parameters
VTi ml Inspired tidal volume No
f bpm Breath rate No
Inspiratory
I:E \ time:Expiratory time No
ratio
Percentage of
TIP:TI % inspiratory plateau No
time in inspiratory time
Tinsp sec Time of inspiration No
cmH2O, hPa, Pressure control level
Pinsp No
mbar of inspiration
cmH2O, hPa,
Psupp Pressure support level No
mbar
cmH2O, hPa, Maximal breathing
Pmax No
mbar pressure
Inspiratory trigger
F-Trigger L/min No
level (flow trigger)
Insp Flow L/min Inspiration flow No

3-12
3.2.2 Output Signals - Alarms
Mindray Patient Monitor Draeger Fabius Alarm

Priority Label Label

Physiological alarms
Apnea APNEA VENT
Volume Apnea > 2
APNEA VOL
min
Pressure Apnea APNEA PRES
High
FiO2 Low % O2 LOW
Paw High PAW HIGH
Paw Low PAW NEGATIVE
CONT PRES CONT PRES
FiO2 High % O2 HIGH
MV High MIN VOL HIGH
Medium MV Low MIN VOL LOW
PEEP High PEEP HIGH
PRESS EXP High PRESS EXP High
Low PRESSURE LIM PRESSURE LIM
Technical alarms
O2 Supply Failure LO O2 SUPPLY
APL VALVE? APL VALVE ?
No Fresh Gas NO FRESHGAS
High
If the monitor shows the alarm [High
High Technical
Technical alarms], the external device
alarms
may have the alarms: VENT ERR.
EXP-VALVE? EXP-VALVE ?
FRESH GAS? FRESH GAS ?
Medium
Medium Technical If the monitor shows the alarm [Medium
alarms Technical alarms], the external device

3-13
Mindray Patient Monitor Draeger Fabius Alarm

Priority Label Label


may have the following alarms:
BATTERY LOW
PRESS ERR
VOL ERR
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
SPEAKER FAIL
POWER FAIL
Low Technical CAL % O2 ?
Low
alarms % O2 ERR
TIME LIMITED
RS232COM ERR
PORT 1 ERROR
PORT 2 ERROR
THRESHOLD LO

3.3 Draeger Perseus A500


3.3.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
bpm Total respiratory
ftot Yes
rate
cmH2O/L/s,hPa/L/s,mbar/L
RAW Airway resistance Yes
/s
VCO2 ml/min CO2 production No
HALLev ml Anesthetic agent No

3-14
Label Units Description Trend, record, print
ENFLev ml consupmtion
ISOLev ml
DESLev ml
SEVLev ml
ml/min Oxygen
VO2 Yes
consumption
Pmean cmH2O, hPa, mbar Mean pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
cmH2O, hPa, mbar Positive
PEEP end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
\ Inspired
minimum
Insp. MAC No
alveolar
concentration
\ Expired minimum
Exp. MAC alveolar No
concentration
FiN2O %
FiIso %
FiDes % Inspired
Yes
FiEnf % anesthetic agent
FiSev %
FiHal %
EtN2O %
EtEnf % End-tidal
Yes
EtDes % anesthetic agent
EtIso %

3-15
Label Units Description Trend, record, print
EtSev %
EtHal %
Spontaneous
MVspn L/min breathed minute Yes
volume
MV L/min Minute volume Yes
Tapnea sec Apnea time No
Difference
between
ΔO2 %,kPa,mmHg No
inspiratory and
expiratory O2
Respiratory rate
RRCO2 bpm Yes
of CO2
Fraction of
FiCO2 %,kPa,mmHg inspired carbon Yes
dioxide
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
N2O Flow L/min N2O flow No
Air Flow L/min Air flow No
O2 Flow L/min O2 flow No
Inspired
FiAA % Yes
anesthetic agent
End-tidal
EtAA % Yes
anesthetic agent
FiAA 2nd % 2nd Insp. Agent Yes
EtAA 2nd % 2nd Exp. Agent Yes

3-16
Label Units Description Trend, record, print
Fractional
FiO2 %,kPa,mmHg concentration of Yes
O2 in inspired gas
EtO2 %,kPa,mmHg End-tidal O2 Yes
Inspired tidal
VTi ml Yes
volume
Fraction of
FiCO2% % inspired carbon Yes
dioxide
End-tidal carbon
EtCO2% % Yes
dioxide
Difference
between
ΔO2 % % No
inspiratory and
expiratory O2
EtO2% % End-tidal O2 Yes
Fractional
FiO2% % concentration of Yes
O2 in inspired gas
Setting Parameters
Inspired tidal No
VTi ml
volume
f bpm Breath rate No
Time of No
Tinsp sec
inspiration
PS above No
cmH2O, hPa, mbar PS above PEEP
PEEP
Maximum airway No
Pmax cmH2O, hPa, mbar
pressure

3-17
Label Units Description Trend, record, print
Inspiratory No
F-Trigger L/min trigger level (flow
trigger)
Percentage of No
inspiratory
TIP:TI %
plateau time in
inspiratory time
Time for the No
Tslope sec pressure to rise to
target pressure
FG ml/min Fresh gas flow No
Pressure control No
Pinsp cmH2O, hPa, mbar level of
inspiration
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure

3.3.2 Output Signals—Alarms


Mindray Patient Monitor Draeger A500 Alarm
Priority Label Label
Physiological alarms
Apnea APNEA or APNEA VOL
FiO2 Low %O2 LOW
CO2 Apnea APNEA CO2
High
Pressure Apnea APNEA PRES
Paw High PAW HIGH
Paw Low PAW LOW / PAW NEGATIVE

3-18
Mindray Patient Monitor Draeger A500 Alarm
Priority Label Label
CONT PRES CONT PRES
FiHal High % HAL HIGH
FiEnf High % ENF HIGH
FiIso High % ISO HIGH
MV Low MIN VOL LOW
FiSev High % SEV HIGH
FiDes High % DES HIGH
EtCO2 Low ET CO2 LOW
EtCO2 High ET CO2 HIGH
FiHal Low % HAL LOW
Medium FiEnf Low % ENF LOW
FiIso Low % ISO LOW
FiDes Low % DES LOW
FiSev Low % SEV LOW
FiCO2 High INSP CO2 HI
MV High MIN VOL HIGH
PEEP High PEEP HIGH
VTe High TIDAL VOL HI
MAC Low MAC LOW?
FiN2O High % N2O HIGH
Technical alarms
O2 Supply Failure O2 SUPPLY?
High NO Fresh Gas NO FRESHGAS
VENT DISC VENT ASSEMBL

3-19
Mindray Patient Monitor Draeger A500 Alarm
Priority Label Label
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
VENT ERR
MIXER INOP
INT.TMP.HIGH
High Technical AIR PRESS HI
Alarm HI O2 SUPPLY
SYSTEM FAULT
N2O CYL.?
NO N2O
NO OXYGEN
NO AIR
FG EXTERN?
FRESH GAS? FRESH GAS?
EXP VALVE? EXP-VALVE?
If the monitor shows the alarm [Medium
Technical alarms], the external device
may have the following alarms:
BATTERY LOW
% O2 ERR
Medium N2O SUPPLY ?
Medium Technical
POWER FAIL
Alarm
SAFETY O2 ON
FG LIMITED
LOSS OF DATA
SET.CANCELED
FG TOO HIGH
FG ACTIVE
3-20
Mindray Patient Monitor Draeger A500 Alarm
Priority Label Label
ABS.PRESENT?
HOSES MIXED?
WRONG HOSES?
AIR ENTRAIN
VENT PAUSE?
If the monitor shows the alarm [CO2
CO2 Module
Module abnormal], the external device
abnormal
may have the alarm: CO2 LINE BLK
If the monitor shows the alarm [AG
Module abnormal], the external device
may have the following alarms:
AG Module
MIXED AGENT
abnormal
CO2 AGT ERR
2nd AGENT
WATERTR.OLD?
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
RS232COM ERR
PRESS ERR
WATER TRAP ?
Low Technical VENT TEMP HI
Low
Alarm VOL ERR
FAN ERR
N2O PRESS HI
O2 CYL. ?
VOLAT SUPPLY
CO2-LINE ?
PWR SPLY ERR
3-21
Mindray Patient Monitor Draeger A500 Alarm
Priority Label Label
TIDAL VOL.?
INSP VOL ERR
N2OCYL.SENS?
AIRCYL.SENS?
O2 CYL.SENS?
AIR CYL.?
PMIN REACHED
PRESS RELIEF
ABSORB. OLD?
ID-FUNC-INOP
HOSE OLD?

3.3.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
CO2 mmHg, kPa, %
O2 %

3-22
3.4 Draeger Primus
3.4.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume
Inspired tidal
VTi ml Yes
volume
MV L/min Minute volume Yes
Spontaneous
fspn bpm Yes
respiratory rate
ml/cmH2O,ml/hPa,ml/
Compl Compliance Yes
mbar
Respiratory rate of
RRCO2 bpm Yes
CO2
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Fraction of inspired
FiCO2 %,kPa,mmHg Yes
carbon dioxide
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
EtO2 %,kPa,mmHg End-tidal O2 Yes

3-23
Label Units Description Trend, record, print
Difference between
ΔO2 %,kPa,mmHg inspiratory and No
expiratory O2
Tapnea sec Apnea time No
FiN2O % Yes
FiIso % Yes
FiDes % Inspired anesthetic Yes
FiEnf % agent Yes
FiSev % Yes
FiHal % Yes
EtN2O % Yes
EtEnf % Yes
EtDes % End-tidal Yes
EtIso % anesthetic agent Yes
EtSev % Yes
EtHal % Yes
Inspired anesthetic
FiAA % Yes
agent
End-tidal
EtAA % Yes
anesthetic agent
FiAA 2nd % 2nd Insp. Agent Yes
EtAA
% 2nd Exp. Agent
2nd Yes
Inspired minimum
Insp.
\ alveolar No
MAC
concentration
Expired minimum
Exp.
\ alveolar No
MAC
concentration

3-24
Label Units Description Trend, record, print
HALLev ml
ENFLev ml
Anesthetic agent
ISOLev ml No
consupmtion
DESLev ml
SEVLev ml
Oxygen
VO2 ml/min Yes
consumption
N2O
N2O flow No
Flow L/min
Air Flow L/min Air flow No
O2 Flow L/min O2 flow No
Arterial oxygen
SpO2 % saturation from Yes
pulse oximetry
PR bpm Pulse rate Yes
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
EtO2% % End-tidal O2 Yes
Difference between
ΔO2 % % inspiratory and No
expiratory O2
Fraction of inspired
FiCO2% % Yes
carbon dioxide
End-tidal carbon
EtCO2% % Yes
dioxide
Setting Parameters
f bpm Breath rate No

3-25
Label Units Description Trend, record, print
Minimum breath No
FreqMIN bpm
frequency
Percentage of No
inspiratory plateau
TIP:TI %
time in inspiratory
time
Time for the No
Tslope sec pressure to rise to
target pressure
Tinsp sec Time of inspiration No
Pressure control No
Pinsp cmH2O, hPa, mbar
level of inspiration
Pressure support No
Psupp cmH2O, hPa, mbar
level
Maximal breathing No
Pmax cmH2O, hPa, mbar
pressure
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Inspired tidal No
VTi ml
volume
FG ml/min Fresh gas flow No

3-26
3.4.2 Output Signals - Alarms
Mindray Patient Monitor Draeger Primus Alarm

Priority Label Label

Physiological alarms
Apnea APNEA/APNEA VENT
Volume Apnea > 2
APNEA VOL
min
Pressure Apnea APNEA PRES
Paw High PAW HIGH
Paw Low PAW NEGATIVE
High
FiO2 Low %O2 LOW
CONT PRES CONT PRES
CO2 Apnea APNEA CO2
No Pulse NO SPO2 PULS
PR Low SPO2 PULS LO
SPO2 Low SPO2 LOW
FiO2 High %O2 HIGH
VTe Low VT LOW
MV High MIN VOL HIGH
MV Low MIN VOL LOW
PEEP High PEEP HIGH
EtCO2 High ET CO2 HIGH
Medium EtCO2 Low ET CO2 Low
FiCO2 High INSP CO2 HIGH
FiN2O High FI N2O HIGH
EtHal High EXP. HAL HIGH
FiHal High % HAL HIGH
FiHal Low % HAL LOW
EtEnf High EXP. ENF HIGH

3-27
Mindray Patient Monitor Draeger Primus Alarm

Priority Label Label


FiEnf High % ENF HIGH
FiEnf Low % ENF LOW
EtIso High EXP. ISO HIGH
FiIso High % ISO HIGH
FiIso Low % ISO LOW
EtSev High EXP. SEV HIGH
FiSev High % SEV HIGH
FiSev Low % SEV LOW
EtDes High EXP. DES HIGH
FiDes High % DES HIGH
FiDes Low % DES LOW
MAC Low MAC LOW?
PR High SPO2 PULS HI
SpO2 High SPO2 HIGH
Technical alarms
O2 Supply Failure O2 SUPPLY ?
No Fresh Gas NO FRESHGAS
Circuit Occluded CIRCLE OCCL
VENT DISC VENT DISC
If the monitor shows the alarm [High
Technical alarms], the external device
High
may have the following alarms:
High Technical VENT ERR
alarms INT.TMP.HIGH
O2CYL.DISCON
CHK N2O CYL
NO N2O DELIV

3-28
Mindray Patient Monitor Draeger Primus Alarm

Priority Label Label


NO O2 DELIV.
NO AIR DELIV
FG X-OVER ?
VENT.UNLOCKD
AW-TEMP HIGH
NO N2O
Patient Circuit Leak LEAKAGE
FRESH GAS? FRESH GAS?
If the monitor shows the alarm [Medium
Technical alarms], the external device
may have the following alarms:
POWER FAIL
BATTERY LOW
N2O SUPPLY ?
PRESSURE LIM
MIXER INOP
P MAX
Medium
Medium Technical SAFETY O2 ON
alarms FG.FLOW LIM.
LOSS OF DATA
HOSES MIXED?
WRONG HOSES?
% O2 ERR
SET.CANCELED
FG TOO HIGH
FG ACTIVE
FG AIR SENS?
FG O2 SENS?

3-29
Mindray Patient Monitor Draeger Primus Alarm

Priority Label Label


FG N2O SENS?
ABS. PRESENT?
If the monitor shows the alarm [AG
Module abnormal], the external device
may have the following alarms:
WATERTR. OLD?
AG Module
MIXED AGENT
abnormal
CO2/AGT ERR
N2O ERR
AGT ERR
2nd AGENT
If the monitor shows the alarm [CO2
CO2 Module Module abnormal], the external device
Medium
abnormal may have the following alarm:
CO2 LINE BLK
NO AIR NO AIR
NO O2 SUPPLY NO O2 SUPPLY
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
FAN ERR
Low PWR SPLY ERR
Low Technical
PRESS ERR
alarms
VOL ERR
LO O2 SUPPLY
CHK O2 CYL
ID-FUNC-INOP
HOSE OLD?

3-30
Mindray Patient Monitor Draeger Primus Alarm

Priority Label Label


HOSE MISSING
COM VENT ERR
APOLLO COM1?
APOLLO COM2?
N2OCYL.SENS?
AIRCYL.SENS?
O2 CYL.SENS?
AIR CYL.?
PRESS RELIEF
ABSORB. OLD?
INSP VOL ERR
If the monitor shows the alarm [SpO2
Module abnormal], the external device
SpO2 Module may have the following alarms:
abnormal SPO2SEN DISC
SPO2 ALRM OF
SPO2 ERR

3.4.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
CO2 mmHg, kPa, %

3-31
3.5 Draeger Zeus/ZeusIE
3.5.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
EtO2 %,kPa,mmHg End-tidal O2 Yes
FiN2O %
FiIso %
FiDes % Inspired anesthetic
Yes
FiEnf % agent
FiSev %
FiHal %
EtN2O %
EtEnf %
EtDes % End-tidal
Yes
EtIso % anesthetic agent
EtSev %
EtHal %
Inspired anesthetic
FiAA % Yes
agent
End-tidal
EtAA % Yes
anesthetic agent
FiAA 2nd % 2nd Insp. Agent Yes
EtAA 2nd % 2nd Exp. Agent Yes
Expired minimum
Exp. MAC \ alveolar No
concentration
Pmean cmH2O, hPa, mbar Mean pressure Yes

3-32
Label Units Description Trend, record, print
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Spontaneous
MVspn L/min breathed minute Yes
volume
Ppeak cmH2O, hPa, mbar Peak pressure Yes
ml Expiratory tidal
VTe Yes
volume
ml Inspired tidal
VTi Yes
volume
L/min Leakage minute
MVLEAK No
volume
MV L/min Minute volume Yes
bpm Respiratory rate of
RRCO2 Yes
CO2
bpm Spontaneous
fspn Yes
respiratory rate
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Fraction of inspired
FiCO2 %,kPa,mmHg Yes
carbon dioxide
cmH2O/L/s,hPa/L/s,mb
RAW Airway resistance Yes
ar/L/s
ISOLev ml
Anesthetic agent
DESLev ml No
consupmtion
SEVLev ml

3-33
Label Units Description Trend, record, print
Arterial oxygen
SpO2 % saturation from Yes
pulse oximetry
BIS \ Bispectral index Yes
EMG dB Electromyograph Yes
SQI % Signal quality index Yes
SR % Suppression ratio Yes
Spectral edge
SEF Hz Yes
frequency
TP dB Total power Yes
BC /min Burst count Yes
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
EtO2% % End-tidal O2 Yes
Fraction of inspired
FiCO2% % Yes
carbon dioxide
End-tidal carbon
EtCO2% % Yes
dioxide
oxygen supply
PO2 kPa
pressure No

N2O Flow L/min N2O flow No

Air Flow L/min Air flow No


O2 Flow L/min O2 flow No
Setting Parameters
Tinsp sec Time of inspiration No
Texp sec Expiratory time No

3-34
Label Units Description Trend, record, print
Maximal breathing No
Pmax cmH2O, hPa, mbar
pressure
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Percentage of No
inspiratory plateau
TIP:TI %
time in inspiratory
time
Pressure control No
Pinsp cmH2O, hPa, mbar
level of inspiration
f bpm Breath rate No
FG ml/min Fresh gas flow No
Time for the No
Tslope sec pressure to rise to
target pressure
Pressure support No
Psupp cmH2O, hPa, mbar
level
Percentage of No
inspiratory plateau
I:E \
time in inspiratory
time
PS above No
cmH2O, hPa, mbar PS above PEEP
PEEP
Inspired tidal No
VTi ml
volume

3-35
3.5.2 Output Signals—Alarms
Mindray Patient Monitor Draeger Zeus/Zeus IE Alarm

Priority Label Label


Physiological alarms
Apnea APNEA or APNEA VOL
Pressure Apnea APNEA PRES
Paw Low PAW LOW or PAW NEGATIVE
Paw High PAW HIGH
High CONT PRES CONT PRES
CO2 Apnea APNEA CO2
FiO2 Low %O2 LOW
No Pulse SPO2 NO PULS
SpO2 Low SPO2 LOW
FiHal High % HAL HIGH
FiEnf High % ENF HIGH
Filso High % ISO HIGH
FiSev High % FiSEV HIGH
FiDes High % FiDES HIGH
FiHal Low % FiHAL LOW
FiEnf Low %FiENF LOW
Filso Low %FiISO LOW
Medium
FiSev Low % FiSEV LOW
FiDes Low % FiDES LOW
FiN2O High % N2O HIGH
FiN2O Low % N2O LOW
FiAA High INSP AGT.HI.
MV Low MIN VOL LOW
MV High MIN VOL HIGH
PEEP High PEEP HIGH

3-36
Mindray Patient Monitor Draeger Zeus/Zeus IE Alarm

Priority Label Label


VTe High TIDAL VOL HI
SpO2 High SPO2 HIGH
BIS High BISX HIGH
BIS Low BISX LOW
EtCO2 High ET CO2 HIGH
EtCO2 Low ET CO2 LOW
FiCO2 High INSP CO2 HI
FiO2 High %O2 HIGH
Low PRESSURE LIM PRESS LIM
Technical alarms
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
INT COM ERR
COM VENT ERR
INT.TMP.HIGH
High Technical Alarm
POWER FAIL
High VENT ERR
VA+MIX ERR
MIXER ERR
O2 CYL. ?
N2O CYL.?

No Fresh Gas NO FRESHGAS

O2 Supply Failure O2 SUPPLY FAIL

If the monitor shows the alarm


Medium Technical
Medium [Medium Technical alarms], the
Alarm
external device may have the following

3-37
Mindray Patient Monitor Draeger Zeus/Zeus IE Alarm

Priority Label Label


alarms:
WRONG AGENT
FLOW SENSOR?
PRESS ERR
VOL ERR
BATTERY LOW
% O2 ERR
PS LIMITED
MIXER INOP
N2O SUPPLY?
SAFETY O2 ON
NO AIR
FG-FLOW HIGH
If the monitor presents the alarm [CO2
Module abnormal], the external device
CO2 Module abnormal
may present the alarm:
CO2 LINE BLK
Fresh Gas LEAK_GAS_LOW
If the monitor shows the alarm [AG
Module abnormal], the external device
may have the following alarms:
AG Module abnormal MIX AGENT
N2O ERR
AGT ERR
TOW AGENT
Low No O2 Supply NO O2 DELIV

3-38
Mindray Patient Monitor Draeger Zeus/Zeus IE Alarm

Priority Label Label


If the monitor shows the alarm [SpO2
Module abnormal], the external device
SpO2 Module
may have the following alarms:
abnormal
SpO2 SEN DISC
SpO2 ERR
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
AIR TRAPPING
CO2-LINE?
Low Technical Alarm
NO N2O
PBAG INOP
BIS INOP
BIS SENS?
BISX OUT RANGE

3.5.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
CO2 mmHg, kPa, %
O2 %

3-39
3.6 GE Aespire 7900&7100/ Aestiva 7900&7100
3.6.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Expiratory tidal
VTe ml Yes
volume
Expiratory minute
MVe L/min Yes
volume
O2% % Oxygen concentration Yes
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
cmH2O, hPa,
Pplat Plateau pressure Yes
mbar
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
cmH2O, hPa, Minimum airway
Pmin No
mbar pressure
ftot bpm Total respiratory rate Yes
Setting Parameters
VT ml Tidal volume No
f bpm Breath rate No
\ Percentage of No
inspiratory plateau
I:E
time in inspiratory
time
% Percentage of No
inspiratory plateau
TIP:TI
time in inspiratory
time

3-40
Label Units Description Trend, record, print
cmH2O, hPa, Positive No
PEEP mbar end-expiratory
pressure
cmH2O, hPa, No
Plimit Pressure limit level
mbar
cmH2O, hPa, Pressure control level No
Pinsp
mbar of inspiration

3.6.2 Output Signals - Alarms


Mindray Patient Monitor GE7900/7100 Alarm

Priority Label Label

Physiological alarms
FiO2 Low Low O2
Paw High High Paw
Paw Low Low Paw
High
High Paw Sustained Sustained Paw (shutdown)
Volume Apnea > 2
Volume Apnea > 2 min
min
FiO2 High High O2
Sub-Atmospheric
Sub-Atmospheric Paw
Paw
MV Low Low VE
Medium
MV High High VE
VTe Low Low Vte
VTe High High Vte
Volume Apnea Volume Apnea
Low Pressure limiting Sustained Paw

3-41
Mindray Patient Monitor GE7900/7100 Alarm

Priority Label Label

Technical alarms
No Fresh Gas No Fresh Gas Flow
O2 Supply Failure No O2 Pressure
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
Pinspired Not Achieved
Inspiration Stopped
+15V SIB Out-of-Range
+15V Manifold Out-of-Range
Display Voltage Out-of-Range
Vaux_ref Out-of-Range
Vext_ref Out-of-Range
High A/D Converter Failure
High Technical
CPU Failure
alarms
Memory (EEPROM) Failure
Memory (flash) Failure
Memory (RAM) Failure
Memory (video) Failure
Bootup Memory Failure
Software Watchdog Failure
Hardware Watchdog Failure
Internal Clock Too Fast
Internal Clock Too Slow
CPU Internal Error
Control Settings Input Has Failed
Medium Technical If the monitor shows the alarm
Medium
alarms [Medium Technical alarms], the

3-42
Mindray Patient Monitor GE7900/7100 Alarm

Priority Label Label


external device may have the following
alarms:
No Pressure Mode/PEEP
Inspiratory Overshoot
Manifold Pressure Sensor Failure
High Pressure Limit Reached (min sys)
Inspiratory Reverse Flow
Expiratory Reverse Flow
Check Flow Sensors
Flow Valve Failure
Gas Inlet Valve Failure
Bootup Gas Inlet Valve Failure
Memory (redundant storage) Fail
No Battery
Low Battery Charge
Low VE Limit Set.
Battery in Use On Battery
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
Check O2 Sensor
O2 Calibration Error
Low
Low Technical alarms PEEP Not Achieved
Vt Not Achieved
No Inspiratory Flow Sensor
No Expiratory Flow Sensor
Insp Vt/Vte Mismatch
Vdel Mismatch

3-43
Mindray Patient Monitor GE7900/7100 Alarm

Priority Label Label


Bellows Empty
+Vanalog Failure
-Vanalog Failure
Flow Sensor Cal Data Corrupt
Low Battery
Low Battery (shutdown)
Battery Voltage Out Of Range
Battery Current Out Of Range
Circuit Auxiliary
Auxiliary Breathing Circuit
Service Calibrations Due

3.6.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min

3.7 GE AespireView
3.7.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
VTe ml Expiratory tidal volume Yes
L/min Expiratory minute
MVe Yes
volume
ftot bpm Total respiratory rate Yes

3-44
Label Units Description Trend, record, print
O2% % Oxygen concentration Yes
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
cmH2O, hPa,
Pplat Plateau pressure Yes
mbar
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
L/min Minimum airway
Pmin No
pressure
bpm Spontaneous breathed
Yes
MVspn minute volume
ml Spontaneous respiratory
Yes
fspn rate
VTi L/min Inspired tidal volume Yes
cmH2O, hPa, Inspiratory minute
Yes
MVi mbar volume
PEEPtot mmHg Total PEEP No
ATMP L/min Barometric pressure No
O2 Flow L/min O2 flow No
N2O Flow ml N2O flow No
Air Flow L/min Air flow No
Setting Parameters
VT ml Tidal volume No
f bpm Breath rate No
\ Inspiratory
I:E time:Expiratory time No
ratio

3-45
Label Units Description Trend, record, print
% Percentage of inspiratory
TIP:TI plateau time in No
inspiratory time
cmH2O, hPa, Positive end-expiratory
PEEP No
mbar pressure
cmH2O, hPa,
Plimit Pressure limit level No
mbar
cmH2O, hPa, Pressure control level of
Pinsp No
mbar inspiration
cmH2O, hPa,
Psupp mbar Pressure support level No

L/min Inspiratory trigger


No
F-Trigger level (flow trigger)
Tinsp sec Time of inspiration No

3.7.2 Output Signals - Alarms


Mindray Patient Monitor GE AespireView Alarm

Priority Label Label

Physiological alarms
Paw High pPeak High
Paw Low pPeak Low
High
Volume Apnea > 2
Volume Apnea > 2 min
min
Sub-Atmospheric
Sub-Atmospheric Paw
Paw
Medium MV Low Low VE
MV High High VE
VTe Low Low Vte

3-46
Mindray Patient Monitor GE AespireView Alarm

Priority Label Label


VTe High High Vte
Volume Apnea Volume Apnea
Low Pressure Limiting Sustained Paw
Technical alarms
No Fresh Gas No Fresh Gas Flow
O2 Supply Failure No O2 Pressure
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
Other Priority Alarms
Ventilator Has No Diver Gas
Battery Charging
High Low Battery Voltage
High Technical
Manifold Pressure Sensor Failure
alarms
Pressure Limit Switch Failure
Flow Valve Failure
Valve Power Failure
No Insp flow sensor
No Exp flow Sensor
Replace exp flow
Replace Insp Flow Sensor
Patient Circuit Leak Patient Circuit Leak
If the monitor shows the alarm [Medium
Technical alarms], the external device
Medium Medium Technical may have the following alarms:
alarms High Circuit O2
Low Circuit O2
No O2 Cell Sensor

3-47
Mindray Patient Monitor GE AespireView Alarm

Priority Label Label


No Pressure Cntrl/PEEP
Inspiration Stopped
Inspiratory Reverse Flow
Expiratory Reverse Flow
Check Flow Sensors
No Air Pressure
Battery Failure
Battery Failure High
Battery Current High
ACGO On
Increase Low MV Limit
Display Failure
Loss of Backup Audio
Battery in Use Running On Battery (No AC)
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
Replace O2 Cell
O2 Cell Calibration Error
Low Low Technical Vt Not Achieved
alarms No Inspiratory Flow Sensor
No Expiratory Flow Sensor
Dry or Replace Sensors
System Leak?
Unable to drive bellows
Memory (EEPROM) Failure

3-48
3.7.3 Output Signals - Waveforms (For BeneVision N series
Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL

3.7.4 Output Signals – Respiratory Loops (For BeneVision N series


Monitors)
Label Units
PV Loop cmH2O/ml, hPa/ml, mbar/ml
FV Loop L/min/ml
PF Loop cmH2O/L/min, hPa/L/min, mbar/L/min

3.8 GE Aisys/Avance/Aisys CS2


3.8.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Expiratory tidal
VTe ml Yes
volume
Expiratory minute
MVe L/min Yes
volume
Total respiratory
ftot bpm Yes
rate
Oxygen
O2% % Yes
concentration
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes

3-49
Label Units Description Trend, record, print
Minimum airway
Pmin cmH2O, hPa, mbar No
pressure
Spontaneous
MVspn L/min breathed minute Yes
volume
Spontaneous
fspn bpm Yes
respiratory rate
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
ml/cmH2O,ml/hPa,ml/mb
Compl Compliance Yes
ar
cmH2O/L/s,hPa/L/s,mbar/
RAW Airway resistance Yes
L/s
Inspired tidal
VTi ml Yes
volume
Inspiratory mimute
MVi L/min Yes
volume
Paux Peak auxiliary
cmH2O, hPa, mbar No
Peak pressure
Paux Mean auxiliary
cmH2O, hPa, mbar No
Mean pressure
Minimum auxiliary
Paux Min cmH2O, hPa, mbar No
pressure
Extrinsic positive
PEEPe cmH2O, hPa, mbar end-expiratory No
pressure
PEEPtot cmH2O, hPa, mbar Total PEEP No
PEEPi min Intrinsic PEEP age No

3-50
Label Units Description Trend, record, print
time (elapsed time since
last maneuver)
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
P0.1 age (elapsed
P0.1 time min time since last No
maneuver)
Barometric
ATMP mmHg No
pressure
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
EtO2 %,kPa,mmHg End-tidal O2 Yes
%,kPa,mmHg Difference between
ΔO2 inspiratory and No
expiratory O2
Fraction of inspired
FiCO2 %,kPa,mmHg Yes
carbon dioxide
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Respiratory rate of
RRCO2 bpm Yes
CO2
Inspired anesthetic
FiAA % Yes
agent
End-tidal
EtAA % Yes
anesthetic agent
FiAA 2nd % 2nd Insp. Agent Yes
EtAA 2nd % 2nd Exp. Agent Yes
FiN2O % Fraction of inspired Yes

3-51
Label Units Description Trend, record, print
nitrous oxide
EtN2O % End-tidal N2O Yes
Minimum alveolar
MAC \ Yes
concentration
Oxygen
VO2 ml/min Yes
consumption
Oxygen
VO2/m2 ml/min/m2 consumption per No
body surface area
Oxygen
VO2/kg ml/min/kg consumption per No
body weight
VCO2 ml/min CO2 production No
EE kcal/day Energy expenditure No
Respiratory
RQ \ No
quotient
oxygen supply
PO2 kPa No
pressure
N2O supply
PN2O kPa No
pressure

Pair kPa air supply pressure No

Oxygen cylinder
O2 cyl. kPa No
pressure
O2 Secondary oxygen
kPa No
cyl.2nd cylinder pressure
N2O cylinder
N2O cyl. kPa No
pressure
Air cylinder
air cyl. kPa No
pressure

3-52
Label Units Description Trend, record, print
Des flow ml/h
Enf flow ml/h
Anesthetic agent
Iso flow ml/h No
flow
Hal flow ml/h
Sev flow ml/h
O2 Flow L/min O2 flow No
N2O Flow L/min N2O flow No
Air Flow L/min Air flow No
Inspiratory
I:E \ time:Expiratory No
time ratio
Fractional residual
FRC ml No
capacity
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
EtO2% % End-tidal O2 Yes
Difference between
ΔO2 % % inspiratory and No
expiratory O2
Fraction of inspired
FiCO2% % Yes
carbon dioxide
End-tidal carbon
EtCO2% % Yes
dioxide
FiHal %
FiIso %
Inspired anesthetic
FiEnf % Yes
agent
FiDes %
FiSev %

3-53
Label Units Description Trend, record, print
EtSev %
EtHal %
End-tidal
EtIso % Yes
anesthetic agent
EtEnf %
EtDes %
Setting Parameters
VT ml Tidal volume No
f bpm Breath rate No
Inspiratory
I:E \ time:Expiratory No
time ratio
Percentage of
inspiratory plateau
TIP:TI % No
time in inspiratory
time
Positive
PEEP cmH2O, hPa, mbar end-expiratory No
pressure
Plimit cmH2O, hPa, mbar Pressure limit level No
Pressure control
Pinsp cmH2O, hPa, mbar No
level of inspiration
Pressure support
Psupp cmH2O, hPa, mbar No
level
Maximal breathing
Pmax cmH2O, hPa, mbar No
pressure
IBW kg Ideal body weight No
BSA m2 Body surface area No
Rise
% rise time% No
Time%

3-54
Label Units Description Trend, record, print
Inspiratory trigger
F-Trigger L/min No
level (flow trigger)
Inspiratory trigger
P-Trigger cmH2O, hPa, mbar level(pressure No
trigger)
Tinsp sec Time of inspiration No

Tpause% % Pause Time No

3.8.2 Output Signals - Alarms


Mindray Patient Monitor GE Avance&Aisys Alarm

Priority Label Label

Physiological alarms
Paw High High Paw
Paw Low Low Paw
High Paw Sustained High Paw Sustained
Volume Apnea > 2
Volume Apnea > 2 min
High min
EtO2 Low Low etO2
EtO2 High High etO2
FiO2 Low Low FiO2
CO2 Apnea CO2 Apnea
Sub-Atmospheric
Sub-Atmospheric Paw
Paw
MV Low Low VE
MV High High VE
Medium
VTe Low Low Vte
VTe High High Vte
Volume Apnea Volume Apnea
RR High High RR

3-55
Mindray Patient Monitor GE Avance&Aisys Alarm

Priority Label Label


RR Low Low RR
EtCO2 Low Low etCO2
EtCO2 High High etCO2
FiCO2 High High FiCO2
EtAA Low Low etAA
EtAA High High etAA
FiAA Low Low FiAA
FiAA High High FiAA
FiO2 High High FiO2
PRESSURE LIM Plimit Reached
Low
Pressure Limiting Sustained Paw
Technical alarms
No Fresh Gas No Fresh Gas Flow
O2 Supply Failure No O2 Pressure
Circuit Occluded Circuit Occluded
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
Pmax Reached
Pinspired Not Achieved
High
Other Priority Alarms (for high priority
High Technical
alarms not assigned a unique bit)
alarms
No VO2
High FiN2O
Low Drive Gas Pressure
Low Battery Charge
Low Battery (No AC)
Control Settings Failure
3-56
Mindray Patient Monitor GE Avance&Aisys Alarm

Priority Label Label


Standby ON (set when anesthesia system
is not in therapy mode or when
respiratory care ventilator is in standby)
Therapy Computer Failure
Monitoring Computer Failure
Display Computer Failure
System Error
Mixer Failure
Mixer Leak
Mixer Control Failure
Vent Failure
Mechanical Ventilation Disabled
Patient Detected (while in standby)
High O2 Supply Pressure
High Air Supply Pressure
Patient Circuit Leak Patient Circuit Leak
If the monitor shows the alarm [AG
Module abnormal], the external device
may have the following alarms:
MGAS ANE_WARMING_UP (5-minute
warming up)
Medium AG Module MGAS WARMING_UP (2-minute warming
abnormal up)
No VO2
FiO2 > 85%
Alternate O2 ON
Air Only Mode
MGAS Failure

3-57
Mindray Patient Monitor GE Avance&Aisys Alarm

Priority Label Label


Module Not Compatible
Vaporizer Cassette Failure
AA Delivery Disabled
MGAS Outlet Occluded
MGAS Filter Blocked
MGAS Sample Line Blocked
MGAS No Sample Line
MGAS Replace Water Trap
Vaporizer Cassette Agent Level Low
No Vaporizer Cassette
Vaporizer Failure
Vaporizer Leak
AA Control Failure
If the monitor shows the alarm [Medium
Technical alarms], the external device
may have the following alarms:
High Circuit O2
Low Circuit O2
No O2 Cell Sensor
No Pressure Cntrl/PEEP
Medium Technical
Inspiration Stopped
alarms
Inspiratory Reverse Flow
Expiratory Reverse Flow
Check Flow Sensors
No Air Pressure
No VO2
No VO2, Artifact
No VO2, High Bypass Flow

3-58
Mindray Patient Monitor GE Avance&Aisys Alarm

Priority Label Label


No Battery
Battery Failure
Battery Charger Failure
Non Circle Circuit Selected
Expiratory Flow Sensed with Non Circle
Circuit
Verify Low VE Limit
Fan Failure
Heater Failure
Power Supply Failure
Display Failure
Breathing System Failure
Sensor Interface Board Failure
ACGO Failure
SCGO Failure
Primary Audio Failure
Backup Audio Failure
Nebulizer Failure
No Nebulizer
Battery in Use Running On Battery (No AC)
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
Low Low Technical ASR on
alarms Replace O2 Cell
O2 Cell Calibration Error
PEEP Not Achieved
Vt Not Achieved

3-59
Mindray Patient Monitor GE Avance&Aisys Alarm

Priority Label Label


No Inspiratory Flow Sensor
No Expiratory Flow Sensor
Insp Vt/Vte Mismatch (VTE > Insp VT)
Vdel Mismatch (System Leak)
Bellows Empty
No N2O Pressure
Memory (EEPROM) Failure
Flow Sensor Cal Data Corrupt
Service Calibrations Due.

3.8.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL
CO2 mmHg, kPa, %
O2 %
N2O %
AA %
Hal %
Des %
Sev %
Enf %
Iso %

3-60
3.8.4 Output Signals – Respiratory Loops (For BeneVision N series
Monitors)
Label Units
PV Loop cmH2O/ml, hPa/ml, mbar/ml
FV Loop L/min/ml
PF Loop cmH2O/L/min, hPa/L/min, mbar/L/min

3.9 HuL Leon


3.9.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Oxygen
O2% % Yes
concentration
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume
Inspired tidal
VTi ml Yes
volume
MV L/min Minute volume Yes
Total respiratory
ftot bpm Yes
rate
Spontaneous
fspn bpm Yes
respiratory rate

3-61
Label Units Description Trend, record, print
ml/cmH2O,ml/hPa,ml/m
Compl Compliance Yes
bar
cmH2O/L/s,hPa/L/s,mbar
RAW Airway resistance Yes
/L/s
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Fraction of inspired
FiCO2 %,kPa,mmHg Yes
carbon dioxide
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
EtO2 %,kPa,mmHg End-tidal O2 Yes
FiN2O %
FiIso %
FiDes % Inspired anesthetic
Yes
FiEnf % agent
FiSev %
FiHal %
EtN2O %
EtEnf %
EtDes % End-tidal
Yes
EtIso % anesthetic agent
EtSev %
EtHal %
Minimum alveolar
MAC \ Yes
concentration
Fractional
FiO2% % concentration of O2 Yes
in inspired gas

3-62
Label Units Description Trend, record, print
EtO2% % End-tidal O2 Yes
End-tidal carbon
EtCO2% % Yes
dioxide
Fraction of inspired
FiCO2% % Yes
carbon dioxide
Inspired anesthetic
FiAA % Yes
agent
FiAA_2nd % 2nd Insp. Agent Yes
End-tidal
EtAA % Yes
anesthetic agent
EtAA_2nd % 2nd Exp. Agent Yes
Setting Parameters
VT ml Tidal volume No
Inspired tidal No
VTi ml
volume
f bpm Breath rate No
Inspiratory No
I:E \ time:Expiratory
time ratio
Insp Flow L/min Inspiration flow No
Tinsp sec Time of inspiration No
Pressure control No
Pinsp cmH2O, hPa, mbar
level of inspiration
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Maximal breathing No
Pmax cmH2O, hPa, mbar
pressure
Tpause % Apnea Time No

3-63
Label Units Description Trend, record, print
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure

3.9.2 Output Signals—Alarms


Mindray Patient Monitor HuL Leon Alarm

Priority Label Label


Physiological alarms
Apnea: Backup Breath was triggered
Apnea
/Apnea
CO2 Apnea Apnea CO2
High
High Paw
Patient pressure continuously too high
Sustained
Paw High Ppeak high
EtCO2 High Expiratory CO2 high
EtCO2 Low Expiratory CO2 low
FiEnf High ENF insp. too high
FiCO2 High Inspiratory CO2 high
FiDes High DES insp. too high
FiDes Low DES insp. too low
FiEnf High ENF insp. too low
Medium
FiHal High HAL insp. too high
FiHal Low HAL insp. too low
FiIso High ISO insp. too high
FiIso Low ISO insp. too low
FiO2 High Inspiratory O2 high
FiO2 Low Inspiratory O2 low
FiSev High SEVO insp. too high
3-64
Mindray Patient Monitor HuL Leon Alarm

Priority Label Label


FiSev Low SEVO insp. too low
MV High MV high
MV Low MV low
PRESS EXP High No release of pressure during expiration
RR Low FreqCO2 low
RR High FreqCO2 high
VTe Low Vte low
Low PRESSURE LIM PMax setting reached too early
Technical alarms
Gasmeasurement Occlusion (Artema
Circuit Occluded
AION)
1. No driving gas. Mechanical
Drive Gas Pressure ventilation stopped. Only Man/Spont
Low possible.
2. Freshgas shortage
O2 and Air supply failed. Dosing fresh gas
No Fresh Gas
stopped
1. O2 cell fresh gas failed. Please
High change. Fresh gas is 100% O2
2.Deviation O2 Concentration fresh gas.
O2 Supply Failure
Fresh gas is 100% O2
3.O2 cell of fresh gas not calibrated. Fresh
gas is 100% O2
Patient module unlocked. Ventilation
VENT DISC
stopped /Disconnection
If the monitor shows the alarm [High
High Technical
Technical alarms], the external device
Alarm
may have the following alarms:
3-65
Mindray Patient Monitor HuL Leon Alarm

Priority Label Label


Battery empty. Mechanical ventilation
stopped. Only Man/Spont possib
Battery empty. Supply voltage Low
Calib. needed: Remove O2-Cell short-time.
Checksum Error
Checksum Fail PIC Conductor
CFB Timeout
Checksum Fail PIC Monitor
CO2 absorber removed. Circle system
short-circuited
Communication Fail CFB
Communication Fail Conductor PIC
Communication Fail Power PIC
Communication Fail Monitor PIC
(Busy Timeout)
Communication Fail Monitor PIC
(Read Timeout)
Communication Fail Monitor PIC
(Write Timeout)
Controllerboard EEPROM checksum failed
Controllerboard EEPROM not write
protected
Driving gas blender failed.
Encoder without function
Ext. fresh gas outlet active
Expiratory flow measurement failed. No
expiratory volume measurement.
Failsafe

3-66
Mindray Patient Monitor HuL Leon Alarm

Priority Label Label


Failure O2 Measurement. Please calibrate
O2 Cell
FiO2 Cell badly calibrated
Flowsensor contaminated. No
measurement of expiratory flow.
Flowsensor contaminated. No
measurement of inspiratory flow.
Flowsensor disconnected. No volume
measurement.
Fresh gas blender failed (flow High). Turn
on emergency dosing
Fresh gas blender failed (flow Low). Turn
on emergency dosing!
Fresh gas blender failed (N2O). Turn on
emergency dosing!
Fresh gas blender failed (no flow
measurement) Turn on emergency
dosing!
Fresh gas blender failed (O2). Turn on
emergency dosing!
Fresh gas blender failed (valves). Fresh gas
is 100% O2
Gas Measurement failed (Artema AION)
Inspiratory flow measurement failed. No
inspiratory volume measurement.
Mains Fail Conductor PIC
Mains Fail Monitor PIC
No water trap

3-67
Mindray Patient Monitor HuL Leon Alarm

Priority Label Label


Read settings differ from written setting
Security relay broken
Sensor fail O2 measuremnet fresh gas.
Switched to 100% O2 fresh gas flow
Sensor Fail Patient Pressure
System Fail Monitor Artema AION
Systemtest skipped too many times
Technical Failure CFB (see error log)
Technical Failure NetDCU (see error log)
Technical failure. Only Man/Spont
possible
Versions not compatible
Zero flow. Flow sensor not calibrated

Patient Circuit Leak Leak high

If the monitor shows the alarm [Medium


Technical alarms], the external device
may have the following alarms:
Battery almost empty
Battery Check/Charge Fail
Batteries deep discharged. Please
Medium Medium Technical calibrate
Alarm Battery Fail
Battery falsely connected or damaged
Change Water Trap (Artema AION)
Gas Measurement unreliable (Artema
AION)
Set pressure not reachable
Set volume not reachable

3-68
Mindray Patient Monitor HuL Leon Alarm

Priority Label Label


Broken loudspeaker. Audible alarming not
possible
Broken microphone. No checking of
audible alarming
Battery in Use Device running on batteries
Air supply failed, Fresh gas with 100% O2
NO AIR
/Air supply failed
O2 supply failed. Dosing fresh gas withair
NO O2 SUPPLY (=21% O2). /O2 supply failed /Piped O2
supply too low
O2 Sensor
FiO2 cell failed. Please change.
Unconnected
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
Air and N2O supply failed, Fresh gas with
Low
100% O2
Air supply failed. Driving gas is O2
Air supply pressure High
Low Technical Check external O2 measurement
Alarm CO2 absorber removed. Circle system
short-circuited
Emergency dosing active
Emergency dosing still active. Please turn
off emergency dosing
Exhalation Condition not reached
Ext. fresh gas outlet active
Failure during communication with

3-69
Mindray Patient Monitor HuL Leon Alarm

Priority Label Label


VueLink
Fan Fail
HL7 server not available
MemoryStick Fail
No Primary Agent detected
N2O supply failed
N2O supply failed, Fresh gas with 100% O2
N2O supply from reserve
O2 ZGA supply failed. Driving gas is air
Patient module open
Piped N2O supply pressure High
Piped N2O supply Low
Piped O2 supply pressure High
VueLink not connected

3.10 HuL Leon Plus


3.10.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Oxygen
O2% % Yes
concentration
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes

3-70
Label Units Description Trend, record, print
Expiratory tidal
VTe ml Yes
volume
VTi ml Inspired tidal volume Yes
MV L/min Minute volume Yes
ftot bpm Total respiratory rate Yes
Spontaneous
fspn bpm Yes
respiratory rate
ml/cmH2O,ml/hPa,m
Compl Compliance Yes
l/mbar
cmH2O/L/s,hPa/L/s,
RAW Airway resistance Yes
mbar/L/s
Fraction of inspired
FiCO2 %.kPa,mmHg Yes
carbon dioxide
End-tidal carbon
EtCO2 %.kPa,mmHg Yes
dioxide
Fractional
FiO2 %.kPa,mmHg concentration of O2 Yes
in inspired gas
EtO2 %.kPa,mmHg End-tidal O2 Yes
FiN2O %
FiIso %
FiDes % Inspired anesthetic
Yes
FiEnf % agent
FiSev %
FiHal %
EtN2O %
EtEnf % End-tidal anesthetic
Yes
EtDes % agent
EtIso %

3-71
Label Units Description Trend, record, print
EtSev %
EtHal %
Minimum alveolar
MAC \ Yes
concentration
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
EtO2% % End-tidal O2 Yes
Fraction of inspired
FiCO2% % Yes
carbon dioxide
End-tidal carbon
EtCO2% % Yes
dioxide
Inspired anesthetic
FiAA % Yes
agent
FiAA_2nd % 2nd Insp. Agent Yes
End-tidal anesthetic
EtAA % Yes
agent
EtAA_2nd % 2nd Exp. Agent Yes
Setting Parameters
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure
VT ml Tidal volume No
VTi ml Inspired tidal volume No
f bpm Breath rate No
Insp Flow L/min Inspiration flow No
Inspiratory No
I:E \ time:Expiratory time
ratio

3-72
Label Units Description Trend, record, print
Tinsp sec Time of inspiration No
Pressure control level No
Pinsp cmH2O, hPa, mbar
of inspiration
Maximal breathing No
Pmax cmH2O, hPa, mbar
pressure
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Tpause % Apnea Time No

3.10.2 Output Signals—Alarms


Mindray Patient Monitor HuL LeonPlus Alarm

Priority Label Label


Physiological alarms
Apnea Apnea: Backup Breath was triggered /Apnea
CO2 Apnea Apnea CO2
FiO2 Low Inspiratory O2 low
High
Paw High Ppeak high
High Paw
Patient pressure continuously too high
Sustained
EtCO2 High Expiratory CO2 high
EtCO2 Low Expiratory CO2 low
FiCO2 High Inspiratory CO2 high
FiDes Low DES insp. too low
Medium FiDes High DES insp. too high
FiEnf Low ENF insp. too low
FiEnf High ENF insp. too high
FiHal High HAL insp. too high
FiHal Low HAL insp. too low
3-73
Mindray Patient Monitor HuL LeonPlus Alarm

Priority Label Label


FiIso High ISO insp. too high
FiIso Low ISO insp. too low
FiO2 High Inspiratory O2 high
FiSev High SEVO insp. too high
FiSev Low SEVO insp. too low
MV High MV high
MV Low MV low
PRESS EXP High No release of pressure during expiration
RR Low FreqCO2 low
RR High FreqCO2 high
VTe Low Vte low
Low PRESSURE LIM PMax setting reached too early
Technical alarms
Circuit Occluded Gasmeasurement Occlusion (Artema AION)
Drive Gas No driving gas. Mechanical ventilation
Pressure Low stopped.Only Man/Spont possible
O2 and Air supply failed. Dosing fresh gas
No Fresh Gas
stopped.
O2 cell fresh gas failed. Please change. Fresh
O2 Supply Failure
gas is 100% O2
High
Patient module unlocked. Ventilation
VENT DISC
stopped. Disconnection
If the monitor shows the alarm [High
Technical alarms], the external device may
High Technical
have the following alarms:
Alarm
Battery empty. Mechanical ventilation
stopped. Only Man/Spont possib

3-74
Mindray Patient Monitor HuL LeonPlus Alarm

Priority Label Label


Battery empty. Supply voltage Low
Calib. needed: Remove O2-Cell short-time.
CFB Timeout
Checksum Error
Checksum Fail PIC Conductor
Checksum Fail PIC Monitor
Communication Fail CFB
Communication Fail Conductor PIC
Communication Fail Monitor PIC (Busy
Timeout)
Communication Fail Monitor PIC (Read
Timeout)
Communication Fail Monitor PIC (Write
Timeout)
Communication Fail Power PIC
Controllerboard EEPROM checksum failed
Controllerboard EEPROM not write
protected
CO2 absorber removed. Circle system
short-circuited
Driving gas blender failed.
Encoder without function
Ext. fresh gas outlet active
Expiratory flow measurement failed. No
expiratory volume measurement.
Failsafe
Failure O2 Measurement. Please calibrate O2
Cell

3-75
Mindray Patient Monitor HuL LeonPlus Alarm

Priority Label Label


FiO2 Cell badly calibrated
Flowsensor contaminated. No
measurement of expiratory flow.
Flowsensor contaminated. No
measurement of inspiratory flow.
Flowsensor disconnected. No volume
measurement.
Fresh gas blender failed (flow High). Turn on
emergency dosing
Fresh gas blender failed (flow Low). Turn on
emergency dosing!
Fresh gas blender failed (N2O). Turn on
emergency dosing!
Fresh gas blender failed (no flow
measurement) Turn on emergency dosing!
Fresh gas blender failed (O2). Turn on
emergency dosing!
Fresh gas blender failed (valves). Fresh gas is
100% O2
Gas Measurement failed (Artema AION)
Inspiratory flow measurement failed. No
inspiratory volume measurement.
Mains Fail Conductor PIC
Mains Fail Monitor PIC
No water trap
Read settings differ from written setting
Security relay broken
Sensor fail O2 measuremnet fresh gas.

3-76
Mindray Patient Monitor HuL LeonPlus Alarm

Priority Label Label


Switched to 100% O2 fresh gas flow
Sensor Fail Patient Pressure
System Fail Monitor Artema AION
Systemtest skipped too many times
Technical Failure CFB (see error log)
Technical Failure NetDCU (see error log)
Technical failure. Only Man/Spont possible
Versions not compatible
Zero flow. Flow sensor not calibrated
Patient Circuit
Leak high
Leak
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
Battery almost empty
Battery Check/Charge Fail
Batteries deep discharged. Please calibrate
Battery Fail
Medium
Medium Technical Battery falsely connected or damaged
Alarm Broken loudspeaker. Audible alarming not
possible
Change Water Trap (Artema AION)
Gas Measurement unreliable (Artema AION)
Set pressure not reachable
Set volume not reachable
Broken microphone. No checking of audible
alarming
Low Battery in Use Device running on batteries

3-77
Mindray Patient Monitor HuL LeonPlus Alarm

Priority Label Label


Air supply failed, Fresh gas with 100% O2
NO AIR
/Air supply failed
O2 supply failed. Dosing fresh gas with air
NO O2 SUPPLY (=21% O2). /O2 supply failed /Piped O2
supply too low
O2 Sensor
FiO2 cell failed. Please change
Unconnected
If the monitor shows the alarm [Low
Technical alarms], the external device may
have the following alarms:
Air and N2O supply failed, Fresh gas with
100% O2
Air supply failed. Driving gas is O2
Air supply pressure High
Check external O2 measurement
CO2 absorber removed. Circle system
short-circuited
Low Technical
Emergency dosing active
Alarm
Emergency dosing still active. Please turn
off emergency dosing
Exhalation Condition not reached
Ext. fresh gas outlet active
Failure during communication with VueLink
Fan Fail
HL7 server not available
MemoryStick Fail
N2O supply failed
N2O supply failed, Fresh gas with 100% O2

3-78
Mindray Patient Monitor HuL LeonPlus Alarm

Priority Label Label


N2O supply from reserve
No Primary Agent detected
Patient module open
Piped N2O supply pressure High
Piped N2O supply Low
Piped O2 supply pressure High
O2 ZGA supply failed. Driving gas is air
VueLink not connected

3.11 MAQUET FLOW-i


3.11.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Inspired tidal
VTi ml Yes
volume
Expiratory minute
MVe L/min Yes
volume
Inspiratory mimute
MVi L/min Yes
volume
Total respiratory
ftot bpm Yes
rate

3-79
Label Units Description Trend, record, print
Inspiratory
I:E \ time:Expiratory No
time ratio
Exp Flow L/min Expiratory flow No
ml/cmH2O,ml/hPa,ml
Compl Compliance Yes
/mbar
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Fraction of inspired
FiCO2 %,kPa,mmHg Yes
carbon dioxide
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
EtO2 %,kPa,mmHg End-tidal O2 Yes
Fraction of inspired
FiN2O % Yes
nitrous oxide
EtN2O % End-tidal N2O Yes
Inspired anesthetic
FiAA % Yes
agent
End-tidal
EtAA % Yes
anesthetic agent
FiAA 2nd % 2nd Insp. Agent Yes
EtAA 2nd % 2nd Exp. Agent Yes
Minimum alveolar
MAC \ Yes
concentration
Oxygen supply
PO2 kPa No
pressure
N2O supply
PN2O kPa
pressure No

3-80
Label Units Description Trend, record, print
Pair kPa Air supply pressure No
Duty cycle or ratio
of inspiration time
to total breathing
Ti/Ttot \ cycle time (only No
during
spontaneous
breathing)
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
EtO2% % End-tidal O2 Yes
Fraction of inspired
FiCO2% % Yes
carbon dioxide
End-tidal carbon
EtCO2% % Yes
dioxide
Expiratory tidal
VTe ml Yes
volume
Setting Parameters
VT ml Tidal volume No
MV L/min Minute volume No
f bpm Breath rate No
Inspiratory No
I:E \ time:Expiratory
time ratio
Time for the No
Tslope sec pressure to rise to
target pressure
Tinsp sec Time of inspiration No

3-81
Label Units Description Trend, record, print
PC above No
cmH2O, hPa, mbar PC above PEEP
PEEP
PS above No
cmH2O, hPa, mbar PS above PEEP
PEEP
Inspiratory trigger No
P-Trigger cmH2O, hPa, mbar level(pressure
trigger)
Inspiratory trigger No
F-Trigger \
level (flow trigger)
Insp Flow L/min Inspiratory flow No
FG ml/min Fresh gas flow No
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure
Rise Time% % rise time% No
Tpause% % Pause Time No
Tinsp% % Time of inspiration No
Tpause s sec Pause Time No

3.11.2 Output Signals - Alarms


Mindray Patient Monitor MAQUET Alarm
Priority Label Label
Physiological alarms
Apnea Apnea
Paw High Paw High
High High Paw Sustained High continuous pressure
EtO2 High EtO2 High
EtO2 Low EtO2 Low

3-82
Mindray Patient Monitor MAQUET Alarm
Priority Label Label
FiO2 Low FiO2 Low
MV High MV too high
MV Low MV too Low
PEEP High PEEP High
PEEP Low PEEP Low
EtCO2 High EtCO2 High
EtCO2 Low EtCO2 Low
FiCO2 High FiCO2 High
FiN2O High FiN2O High
EtIso High EtISO High
EtIso Low EtIso Low
Medium FiIso High FiIso High
FiIso Low FiIso Low
EtSev High EtSev High
EtSev Low EtSev Low
FiSev High FiSev High
FiSev Low FiSev Low
EtDes High EtDes High
EtDes Low EtDes Low
FiDes High FiDes High
FiDes Low FiDes Low
FiO2 High FiO2 High
RR High frequency high
Low
RR Low frequency low
Technical alarms
Circuit Occluded Gas sampling tube Occlusion
High If the monitor shows the alarm [High
High Technical alarms
Technical alarms], the external device

3-83
Mindray Patient Monitor MAQUET Alarm
Priority Label Label
may have the following alarms:
Mixture of Anesthesia agents
Gas Supply
Cross contamination of anesthesic
Agents
Vaporizer liquid level
battery alarm
patient Cassette remove
patient Cassette exchange
check tubing alarm
If the monitor shows the alarm
[Medium Technical alarms], the
external device may have the following
alarms:
Medium Technical Gas Analyzer water trap
Medium
alarms Gas Analyzer water trap missing
internal communicaiton failture
High continuous APL
Pressure alarm
Leakage alarm
Low Battery in Use Battery operation

3-84
3.12 Mindray A7/A5/A4/A3/Mindray WATO
25/60/50/30/20/35/55/65/55/65
NOTE

 Only the Benevision N Sereis monitor supports integration with the


Mindray A4 anesthesia machine.

3.12.1 Output Signals - Parameters


Label Units Description Trend, record, print
Monitoring Parameters
Inspiratory
I:E \ time:Expiratory No
time ratio
Mandatory
fmand bpm breathing Yes
frequency
Spontaneous
fspn bpm Yes
respiratory rate
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory minute
MVe L/min Yes
volume
Expiratory tidal
VTe ml Yes
volume
cmH2O/L/s,hPa/L/s,mbar/
RAW Airway resistance Yes
L/s

3-85
Label Units Description Trend, record, print
ml/cmH2O,ml/hPa,ml/mb
Compl Compliance Yes
ar
N2O
L/min N2O flow No
Flow
Air Flow L/min Air flow No
O2 Flow L/min O2 flow No
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Fraction of inspired
FiCO2 %,kPa,mmHg Yes
carbon dioxide
Fraction of inspired
FiN2O % Yes
nitrous oxide
EtN2O % End-tidal N2O Yes
EtDes % Yes
EtSev % Yes
End-tidal anesthetic
EtEnf % Yes
agent
EtIso % Yes
EtHal % Yes
FiDes % Yes
FiSev % Yes
Inspired anesthetic
FiEnf % Yes
agent
FiIso % Yes
FiHal % Yes
Inspired anesthetic
FiAA % Yes
agent
End-tidal anesthetic
EtAA % Yes
agent
Minimum alveolar
MAC \ Yes
concentration

3-86
Label Units Description Trend, record, print
BIS \ Bispectral index Yes
SQI % Signal quality index Yes
SR % Suppression ratio Yes
EMG dB Electromyograph Yes
Spectral edge
SEF Hz Yes
frequency
TP dB Total power Yes
BC /min Burst count Yes
HALLev ml
ENFLev ml
Anesthetic agent
ISOLev ml No
consupmtion
DESLev ml
SEVLev ml
ftot bpm Total breath rate Yes
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas

EtO2 %,kPa,mmHg End-tidal O2

Fractional Yes
FiO2% % concentration of O2
in inspired gas
Yes
EtO2% % End-tidal O2

Spontaneous Yes
MV Spn L/min
minute volume
Difference between
the inspiratory and
MV Leak L/min No
expiratory minute
volume

3-87
Label Units Description Trend, record, print
Setting Parameters
L/min Inspiratory trigger No
F-Trigger
level (flow trigger)
VT ml Tidal volume No
f bpm Breath rate No
cmH2O, hPa, mbar Positive No
PEEP end-expiratory
pressure
Plimit cmH2O, hPa, mbar Pressure limit level No
cmH2O, hPa, mbar Pressure control No
Pinsp
level of inspiration
cmH2O, hPa, mbar Pressure support No
Psupp
level
Tinsp sec Time of inspiration No
sec Time for the No
Tslope pressure to rise to
target pressure
Minimum breath No
FreqMin bpm
frequency
Percentage of No
inspiratory plateau
TIP:TI %
time in inspiratory
time

3-88
3.12.2 Output Signals - Alarms
Mindray Patient Monitor Mindray Anesthesia Alarm

Priority Label Label

Physiological alarms
Apnea Apnea Alarm
Volume
Volume Apnea>2min
Apnea>2min
Paw High Paw Too High
Paw Low Paw Too Low
High FiO2 Low FiO2 Too Low
EtO2 High EtO2 Too High
EtO2 Low EtO2 Too Low
High Paw
Sustained Airway Pressure too High
Sustained
CO2 Apnea Apnea CO2
MV High MV Too High
MV Low MV Too Low
BIS High BIS Too High
BIS Low BIS Too Low
EtCO2 High EtCO2 Too High
EtCO2 Low EtCO2 Too Low
FiCO2 High FiCO2 Too High
Medium
EtN2O Low EtN2O Too Low
EtN2O High EtN2O Too High
FiN2O Low FiN2O Too Low
FiN2O High FiN2O Too High
EtHal Low EtHal Too Low
EtHal High EtHal Too High
FiHal Low FiHal Too Low

3-89
Mindray Patient Monitor Mindray Anesthesia Alarm

Priority Label Label


FiHal High FiHal Too High
EtEnf Low EtEnf Too Low
EtEnf High EtEnf Too High
FiEnf Low FiEnf Too Low
FiEnf High FiEnf Too High
EtIso Low EtIso Too Low
EtIso High EtIso Too High
FiIso Low FiIso Too Low
FiIso High FiIso Too High
EtSev Low EtSev Too Low
EtSev High EtSev Too High
FiSev Low FiSev Too Low
FiSev High FiSev Too High
EtDes Low EtDes Too Low
EtDes High EtDes Too High
FiDes Low FiDes Too Low
FiDes High FiDes Too High
FiO2 High FiO2 Too High
FiCO2 Low FiCO2 Too Low
Low Pressure limiting Pressure Limiting
Technical alarms
Drive Gas Pressure
Drive Gas Pressure Low
Low
No Fresh Gas No Fresh Gas
High
O2 Supply Failure O2 Supply Failure
High Technical If the monitor shows the alarm [High
alarms Technical alarms], the external device

3-90
Mindray Patient Monitor Mindray Anesthesia Alarm

Priority Label Label


may have the following alarms:
Negative Pressure
Check Flow Sensors
O2-N2O Ratio Error
Flowmeter Comm Stop
Aux Control Module Comm Stop
Power System Comm Stop
System DOWN for battery
depletion!Power Board High Temp
Breathing Circuit Not MountedFlowmeter
Cal. Data Error 01/02
Flowmeter Hardware Malfunction
01/02/03/04/05/06/07
Ventilator Hardware Malfunction
01/02/03/04/05/06/07/08/09/10/11/12
Aux Control Module Comm Error
Power Supply Voltage Error
Low Battery Voltage! Backup Flow Control
Deployment Failure
Flowmeter Comm ErrorAuxi Ctrl Module
Hardware Error 01/02/03/04/05
Ventilator Comm Error
Ventilator Selftest Error
Power System Comm Error
Keyboard Init Error
RT Clock Need Reset
RT Clock Not Exist
Mechanical Ventilation Failure

3-91
Mindray Patient Monitor Mindray Anesthesia Alarm

Priority Label Label


Paw < -10cmH2O
Device Fault, Ventilate Manually
Patient Circuit Leak Patient Circuit Leak
If the monitor shows the alarm [Medium
Technical alarms], the external device
may have the following alarms:
PEEP Valve Failure
Insp Valve Failure
ACGO 3-way Valve Failure
Replace O2 Sensor
Battery Undetected
IP Address Conflict
Pressure Monitoring Channel Failure
Ventilator Comm Stop
CO2 Canister Not MountedElectronic Flow
Medium Medium Technical
Control Error
alarms
Backup Flow Control Retraction Failure
Air Supply Failure
N2O not detected
Backup Flow Control Valves Open
Total Flow Sensor Self Test Time Out
Backup Flow Control Error
PEEP Safety Valve Failure
Insp Reverse Flow
Exp Reverse Flow
ACGO On
TVe Below Control Range
Key Error

3-92
Mindray Patient Monitor Mindray Anesthesia Alarm

Priority Label Label


O2 Flush Failure
If the monitor shows the alarm [AG
Module Abnormal], the external device
may have the following alarms:
AG Hardware Error
O2 Sensor Error
External AG Self Test Error
AG Hardware Malfunction
AG Init Error
AG No Watertrap
AG Watertrap Type Wrong
AG Change Watertrap
AG Comm Stop
AG Airway Occluded
AG Module
AG Comm Error
Abnormal
AG Data Limit Error
AG Zero Failed
AG Cal. Failed
AG Accuracy Error
O2/N2O/CO2/Enf/Iso/Sev/Hal/Des Accuracy
Unspecified
Mixed anesthetic gas and MAC < 3
Mixed anesthetic gas and MAC >=3
EtCO2/FiCO2/EtN2O/FiN2O/EtHal/FiHal/EtE
nf/FiEnf/EtIso/FiIso/EtSev/FiSev/EtDes/FiD
es/EtO2/FiO2 Over Range
Internal AG Error 01 02 03 04 05 06 07 08
09 10 11 12

3-93
Mindray Patient Monitor Mindray Anesthesia Alarm

Priority Label Label


If the monitor shows the alarm [BIS
Module Abnormal], the external device
may have the following alarms:
BIS Init Error
BISx Disconnected
BIS Comm Error
BIS Over Range
BIS High Imped.
BIS Sensor Off
BIS DSC Error
BIS Module
BIS DSC Malf
abnormal
BIS No Cable
BIS No Sensor
BIS Wrong Sensor Type
BIS SQI<50%
BIS SQI<15%
BIS Sensor Expired
BIS Sensor Fault
BIS Sensor Too Many Uses
Disconnect/Reconnect BIS
BIS Self Test Error
Battery in Use Battery in Use
O2 Sensor
O2 Sensor Disconnected
Unconnected
Low If the monitor shows the alarm [Low
Low Technical Technical alarms], the external device
alarms may have the following alarms:
Flow Sensor Failure

3-94
Mindray Patient Monitor Mindray Anesthesia Alarm

Priority Label Label


Pinsp Not Achieved
TV Not Achieved
Calibrate O2 Sensor
Heating Module Failure
Auto Vent UNUSEBIS Electrode
Unconnected
BIS Electrode 1/2/3/4 Lead Off
BIS Electrode 1/2/3/4 High Imped.
N2O Flow High
O2 Flow High
Air Flow High
Flowmeter zero cal errACGO Not found
Branch Flow Not Achieved
Balance Gas Branch Flow Not Achieved
Backup Flow Control is enabled
Flowmeter Comm Stop
Calibrate Flow Sensor
Calibrate PEEP Valve
TV Comp Disabled
TVe > TVi
N2O Sensor Error
O2 Sensor Error
Air Sensor Error
3-way Valve Failure
Sensor Zero Failed
TV Delivery High
Flowmeter Zero Failed
Ventilator Init Error

3-95
3.12.3 Output Signals - Waveforms (For BeneVision N series
Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL
CO2 mmHg, kPa, %
O2 %
N2O %
AA %
Hal %
Des %
Sev %
Enf %
Iso %

3.12.4 Output Signals – Respiratory Loops (For BeneVision N


series Monitors)
Label Units
PV Loop cmH2O/ml, hPa/ml, mbar/ml
FV Loop L/min/ml
PF Loop cmH2O/L/min, hPa/L/min, mbar/L/min

3-96
4 Integrating the Ventilator

4.1 AirLiquide ALMS Monnal T75


4.1.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
cmH2O, hPa, Positive end-expiratory
PEEP Yes
mbar pressure
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
cmH2O, hPa,
Pplat Plateau pressure Yes
mbar
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
VTe ml Expiratory tidal volume Yes
VTi ml Inspiratory tidal volume Yes
MVe L/min Expiratory minute volume Yes
ftot bpm Total respiratory rate Yes
Fractional concentration
FiO2 %,kPa,mmHg Yes
of O2 in inspired gas
Fractional concentration
FiO2% % Yes
of O2 in inspired gas
EtCO2 %,kPa,mmHg End-tidal carbon dioxide Yes
Setting Parameters
VT ml Tidal volume No
f bpm Breath rate No
cmH2O, hPa, Positive end-expiratory No
PEEP
mbar pressure

4-1
Label Units Description Trend, record, print
Inspiratory No
I:E \
time:Expiratory time ratio
cmH2O, hPa, No
Psupp Pressure support level
mbar
Tinsp sec Time of inspiration No
4.1.2 Output Signals—Alarms
Mindray Patient Monitor ALMS Monnal T75 Alarm

Priority Label Label


Physiological alarms
Apnea Ventilation Apnea ventilation
VT Not Achieved VT not reached
CO2 Apnea CO2 apnea
Apnea Apnea
FiO2 High High FiO2
FiO2 Low Low FiO2
High FiO2 Alarm 100% FiO2 during 2 min
MV High High MVe
MV Low Low MVe
Paw High High pressure
Ppeak High High Ppeak
Patient demand higher than set peak
Ppeak Low
flow
EtCO2 High High etCO2
EtCO2 Low Low etCO2
FiCO2 High High FiCO2
Medium
PEEPe High PEEP greater than set PEEP + 5 cmH2O
RR High High RR
RR Low Low RR

4-2
Mindray Patient Monitor ALMS Monnal T75 Alarm

Priority Label Label


VTe High High VTe

VTe Low Low VTe

Low Plimit Reached PI limit reached


Technical alarms
Airway Obstructed? Expiration blocked
Expiratory flow sensor failure
Check Flow Sensors
Inspiratory flow sensor failure
O2 Supply Pressure
O2 supply failure
Low
Patient Disconnected Patient disconnection
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
Internal battery low
O2 sensor failure
Restart self-tests
High
Technical failure detected
Ventilator inoperative
Minor failure detected
High Technical Alarm
Adjust ventilation parameters
Nurse Call
Internal battery failure
Air supply failure
N2O supply failure
Xe supply failure
N2O sensor failure
Xe sensor failure
AA sensor failure
4-3
Mindray Patient Monitor ALMS Monnal T75 Alarm

Priority Label Label


Gas analyzer failure
Hypoxic mix
PS limit reached
Slow Inflation in progress
Sustained exhalation in progress
Maximum FiO2 during 2 min
O2 flush in progress
O2 monitoring disabled
Expiratory monitoring disabled
High AA
Low AA
Low FiCO2
If the monitor shows the alarm
[Medium Technical alarms], the
Medium Technical
Medium external device may have the following
Alarm
alarms:
Internal battery discharged
Airway Leak? Important leak
Ventilator operates from external
Battery in Use battery
Ventilator operates from internal battery
Low If the monitor shows the alarm [Low
Technical alarms], the external device
Low Technical Alarm may have the following alarms:
Maintenance required
CO2 sensor failure

4-4
4.2 Carefusion Avea
4.2.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
C20/C \ Compliance Ratio Yes
ml/cmH2O,ml/hPa,ml
Cdyn Dynamic compliance Yes
/mbar
ml/cmH2O,ml/hPa,ml
Cstat Static compliance Yes
/mbar
Oxygen
O2% % Yes
concentration
Inspiratory time:
I:E \ No
Expiratory time ratio
Leak Comp % Leak compensation No
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Exp.Flow L/min Expiratory flow No
Insp.Flow L/min Inspiratory flow No
Pmean cmH2O, hPa, mbar Mean pressure Yes
Oxygen supply
PO2 kPa No
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
ftot bpm Total breath rate Yes
Spontaneous
fspn bpm Yes
respiratory rate
Total volume of gas
Vdel ml No
delivered
MVspn L/min Spontaneous Yes

4-5
Label Units Description Trend, record, print
breathed minute
volume
WOB J/L Work of breathing Yes
VTi ml Inspired tidal volume Yes
Spontaneous
VTe spn ml expiratory tidal Yes
volume
VT/kg ml/kg TVe/IBW No
Mandatory Expiratory
VTe Mand ml No
tidal volume
Expiratory tidal
VTe ml Yes
volume
Expiratory minute
MVe L/min Yes
volume
Tinsp sec Time of inspiration No
Texp sec Expiratory time No
cmH2O/L/s,hPa/L/s,m
Ri Inspiratory resistance Yes
bar/L/s
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
Pair kPa Air supply pressure No
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
Extrinsic positive
PEEPe cmH2O, hPa, mbar end-expiratory No
pressure

4-6
Label Units Description Trend, record, print
cmH2O/L/s,hPa/L/s,m
Re Expiratory resistance Yes
bar/L/s
Mandatory breathing
fmand bpm Yes
frequency
Negative inspiratory
NIF cmH2O, hPa, mbar No
force
Setting Parameters
Fractional
O2% % concentration of O2 Yes
in inspired gas
Base Flow L/min Base Flow No
Inspiration
Exp% % No
termination level
Peak Flow L/min Peak flow No
Tpause sec Apnea Time No
Phigh cmH2O, hPa, mbar Upper pressure level No
Plow cmH2O, hPa, mbar Lower pressure level No
Positive
PEEP cmH2O, hPa, mbar end-expiratory No
pressure
Pressure support
Psupp cmH2O, hPa, mbar No
level
f bpm Breath rate No
Trise sec Rise time No
Tinsp sec Time of inspiration No
Time for the lower
Tlow sec No
pressure level
Inspiratory trigger
F-Trigger L/min No
level (flow trigger)

4-7
Label Units Description Trend, record, print
Inspiratory trigger
P-Trigger cmH2O, hPa, mbar level (pressure No
trigger)
VT ml Tidal volume No
Pressure control level
Pinsp cmH2O, hPa, mbar No
of inspiration
Time for the upper
Thigh sec No
pressure level
Maximum inspiration
Ti Max sec No
time

4.2.2 Output Signals - Alarms


Mindray Patient Monitor Carefusion Avea

Priority Label Label

Physiological alarms
Ppeak High HIGH Ppeak
PEEP Low LOW PEEP
MV Low LOW Ve
High MV High HIGH Ve
FiO2 Low LOW FiO2
FiO2 High HIGH FiO2
Ppeak Low LOW Ppeak
RR High HIGH RATE
Medium VTe Low LOW Vte
VTe High HIGH Vte
Technical alarms
High High Technical If the monitor shows the alarm [High

4-8
Mindray Patient Monitor Carefusion Avea

Priority Label Label


alarms Technical alarms], the external device
may have the following alarms:
Vent Inop
Loss of Gas
Circuit Disconnect
Ext High Ppeak
Safety Valve
CIRCUIT OCCLUSION
Apnea Interval
Loss of O2
Loss of Air
Loss of Heliox
Low Battery
Loss of A/C
ILV Disconnect
If the monitor shows the alarm [Medium
Technical alarms], the external device
Medium Technical
Medium may have the following alarms:
alarms
Invalid Gas ID
Fan Failure
Battery in Use
If the monitor shows the alarm [Low
Technical alarms], the external device
Low Low Technical may have the following alarms:
alarms Insp Time Limit
I:E Limit
Volume Limit

4-9
4.2.3 Output Signals - Waveforms (For BeneVision N series
Monitors)
Label Units
Paw cmH2O, mbar, hPa
Flow L/min
Volume mL

4.3 Carefusion Vela


4.3.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Positive
cmH2O, hPa,
PEEP end-expiratory Yes
mbar
pressure
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
Expiratory tidal
VTe ml Yes
volume
VTi ml Inspired tidal volume Yes
Spontaneous
VTe spn ml expiratory tidal Yes
volume
Spontaneous
MVspn L/min breathed minute Yes
volume
Expiratory minute
MVe L/min Yes
volume

4-10
Label Units Description Trend, record, print
ftot bpm Total breath rate Yes
Spontaneous
fspn bpm Yes
respiratory rate
Inspiratory
I:E \ time:Expiratory time No
ratio
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
Texp sec Expiratory time No
Oxygen supply
PO2 kPa No
pressure
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
Setting Parameters
Positive
cmH2O, hPa,
PEEP end-expiratory No
mbar
pressure
VT ml Tidal volume No
f bpm Breath rate No
Base Flow L/min Base Flow No
Inspiratory trigger No
F-trigger L/min
level (flow trigger)

4-11
Label Units Description Trend, record, print
cmH2O, hPa, Pressure support No
Psupp
mbar level
Tinsp sec Time of inspiration No
cmH2O, hPa, Pressure control level No
Pinsp
mbar of inspiration
cmH2O, hPa, No
Phigh Upper pressure level
mbar
cmH2O, hPa, No
Plow Lower pressure level
mbar
Time for the upper No
Thigh sec
pressure level
Time for the lower No
Tlow sec
pressure level
Inspiration No
Exp% %
termination level
Peak Flow L/min Peak flow No
Maximum inspiration No
Ti max sec
time
Oxygen No
O2% %
concentration
Pressure control level No
Tpause sec
of inspiration

4-12
4.3.2 Output Signals - Alarms
Mindray Patient Monitor Carefusion Vela Alarm

Priority Label Label

Physiological alarms
MV Low Low Ve
Apnea APNEA INTERVAL
PEEP High HIGH PEEP
High
Ppeak High HIGH PIP or HIGH PIP SUST
Ppeak Low LOW PIP
FiO2 Alarm %O2 RANGE ERROR
EtCO2 High High EtCO2
Medium EtCO2 Low Low EtCO2
RR High HIGH RATE
Technical alarms
O2 cell cal. Needed CHECK O2 CAL
O2 Supply Pressure Low O2 Inlet LOW
Circuit Disconnect Circuit FAULT
If the monitor shows the alarm [High
Technical alarms], the external device
High
may have the following alarms:
High Technical alarms LOW BATTERY
MOTOR FAULT
VENT INOP
H/W FAULT
If the monitor shows the alarm
[Medium Technical alarms], the
Medium Technical
Medium external device may have the
alarms
following alarms:
CHECK EVENTS

4-13
Mindray Patient Monitor Carefusion Vela Alarm

Priority Label Label


LOW CLOCK BATTERY
CO2 COMMS ERROR
CO2 Out Of Range
CO2 Sensor Fault
CO2 Sensor Temp
CO2 Zero Reqd
FAN FAILURE
CO2 Check Adapter
DEFAULTS
O2 SENSOR FAILURE
Invalid EtCO2
O2 INLET HIGH
MED BATTERY
XDCR FAULT
Battery in Use ON BATTERY POWER
If the monitor shows the alarm [Low
Technical alarms], the external device
Low
Low Technical alarms may have the following alarms:
INVALID SERIAL NUMBER
NO CAL DATA

4-14
4.4 Draeger Babylog 8000 plus/Babylog 8000
4.4.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
ml/cmH2O,ml/hPa,ml Dynamic
Cdyn Yes
/mbar compliance
cmH2O/L/s,hPa/L/s,m Dynamic lung Yes
Rdyn
bar/L/s resistance
Pmean cmH2O, hPa, mbar Mean pressure Yes
Positive Yes
PEEP cmH2O, hPa, mbar end-expiratory
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Inspired tidal
VTi ml Yes
volume
Leak
Leak Comp % No
compensation
MV L/min Minute volume Yes
Total
ftot bpm Yes
respiratory rate
Fractional
concentration
FiO2 %,kPa,mmHg Yes
of O2 in inspired
gas
Fractional
concentration
FiO2% % Yes
of O2 in inspired
gas
Setting Parameters

4-15
Label Units Description Trend, record, print
Oxygen
O2% % No
concentration
Time of No
Tinsp sec
inspiration
Frequency of No
fSIMV bpm
SIMV
PEEP/CPAP cmH2O, hPa, mbar PEEP/CPAP No
Apnea tidal No
TApnea sec
volume
Maximum No
Pmax cmH2O, hPa, mbar
airway pressure
f bpm Breath rate No
VT ml Tidal volume No
Inspiratory No
I:E \ time:Expiratory
time ratio

4.4.2 Output Signals—Technical Alarms


Mindray Patient Monitor Draeger Babylog Alarm

Priority Label Label


If the monitor shows the alarm [Medium
Technical alarms], the external device
Medium Technical
Medium may have the following alarms:
Alarm
VOL ERR
% O2 ERR

4-16
4.5 Draeger Babylog VN500
4.5.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Cdyn ml/cmH2O,ml/hPa,ml/ Dynamic
Yes
mbar compliance
Rdyn cmH2O/L/s,hPa/L/s,mb Dynamic lung
Yes
ar/L/s resistance
VCO2 ml/min CO2 production No
Pmin cmH2O, hPa, mbar Minimum airway
No
pressure
Pmean cmH2O, hPa, mbar Mean pressure Yes
PEEP cmH2O, hPa, mbar Positive
end-expiratory Yes
pressure
fmand bpm Mandatory
breathing Yes
frequency
Ppeak cmH2O, hPa, mbar Peak pressure Yes
VTe spn ml Spontaneous
expiratory tidal Yes
volume
MVLEAK L/min Leakage minute
No
volume
Leak Comp % Leak compensation No
fspn bpm Spontaneous
Yes
respiratory rate
MV L/min Minute volume Yes
RSBI 1/(min.L) Rapid shallow
Yes
breathing index

4-17
Label Units Description Trend, record, print
ftot bpm Total respiratory
Yes
rate
EtCO2% % End-tidal carbon
Yes
dioxide
EtCO2 %,kPa,mmHg End-tidal carbon
Yes
dioxide
I:E \ Inspiratory time:
No
Expiratory time ratio
FiO2% % Fractional
concentration of O2 Yes
in inspired gas
FiO2 %,kPa,mmHg Fractional
concentration of O2 Yes
in inspired gas
MVe L/min Expiratory minute
Yes
volume
MVi L/min Inspiratory mimute
Yes
volume
VTe ml Expiratory tidal
Yes
volume
VTi ml Inspired tidal
Yes
volume
VT/kg ml TVe/IBW No
Setting Parameters
O2% % Fractional
concentration of O2 Yes
in inspired gas
Flow L/min Flow No
Tinsp sec Time of inspiration No

4-18
Label Units Description Trend, record, print
f bpm Breath rate No
PEEP cmH2O, hPa, mbar Positive
end-expiratory No
pressure
△int.PEEP cmH2O, hPa, mbar Intermittent PEEP No
Plow cmH2O, hPa, mbar Lower pressure
No
level
Phigh cmH2O, hPa, mbar Upper pressure
No
level
Tlow sec Time for the lower
No
pressure level
Thigh sec Time for the upper
No
pressure level
TApnea sec Apnea Time No
Psupp cmH2O, hPa, mbar Pressure support
No
level
Pmax cmH2O, hPa, mbar Maximal breathing
No
pressure
F-Trigger L/min Inspiratory trigger
No
level (flow trigger)
Trise sec Rise time No
FlowAssist cmH2O.s/L,hPa.s/L,mb
Flow assist No
ar.s/L
Vol Assist cmH2O/L,hPa/L,mbar/
Volume assist No
L
fapnea bpm Breath rate for
No
apnea ventilation
Pinsp cmH2O, hPa, mbar Pressure control
No
level of inspiration

4-19
Label Units Description Trend, record, print
ATC % Automatic Tube
No
Compensation
Tube ID mm Tube ID No
Tdisconnect sec Delay time for
"Airway pressure No
lower alarm limit"
Ti max sec Maximum
No
inspiration time
Exp% % Inspiration
No
termination level
VT ml Tidal volume No
VTapnea ml Apnea tidal volume No
I:E \ Inspiratory time:
No
Expiratory time ratio

4.5.2 Output Signals - Alarms


Mindray Patient Monitor Draeger BabyLog VN500 Alarm

Priority Label Label

Physiological alarms
Paw High Airway pressure high
Paw Low Airway pressure low
MV High MV high
MV Low MV low
High
FiO2 High FiO2 high
FiO2 low FiO2 low
PEEP High PEEP high
PEEP Low PEEP low

4-20
Mindray Patient Monitor Draeger BabyLog VN500 Alarm

Priority Label Label

Physiological alarms
Apnea Ventilation Apnea ventilation
Apnea APNEA RESP
EtCO2 High etCO2 high
EtCO2 Low etCO2 low
Medium VTe Low Tidal volume low
RR High High respiratory rate
VOL INCONST VT not reached
Technical alarms
Air Supply
Pressure Low Air supply down
O2 Supply
Pressure Low O2 supply down
CLEAN CO2 Clean CO2 cuvette
Power Failure Internal power supply failure
Check Flow
Exp Time Err
Sensors
EXP-VALVE? Expiratory valve malfunction
High
Negative Airway
Airway pressure negative
Pressure
Neo Flow Sensor
Neo. flow sensor changed ?
Error
Circuit Disconnect Disconnection Ventilator
No O2 Pressure O2 supply down
Airway
Tube Obstruct
Obstructed?
High Technical If the monitor shows the alarm [High

4-21
Mindray Patient Monitor Draeger BabyLog VN500 Alarm

Priority Label Label

Physiological alarms
Alarms Technical alarms], the external device may
have the following alarms:
CO2 NOT CAL
BATERY ERR
SPEAKER FAIL
VENT ERR
VOL CAL
VOL ERR
AW TEMP INOP
AIR PRESS HI
HI O2 SUPPLY
SYSTEM FAULT
BATTER ERR
COOLING
LOSS OF DATA
HOSE ERR
EJECTOR INOP
MAP LOW
O2 Err
PRESS_ERR
AW_TEM_SENS
INT_TEMP_HI
Evita Err
Medium If the monitor shows the alarm [Medium
Medium Technical Technical alarms], the external device may
Alarms have the following alarms:
GAS FAILURE

4-22
Mindray Patient Monitor Draeger BabyLog VN500 Alarm

Priority Label Label

Physiological alarms
NO AIR
AMB_PRESS
BATT LOW
CHECK_EVITA
NEBULIZ OF
VT LIMITED
Low Battery In use Internal battery activated
If the monitor shows the alarm [Low
Technical alarms], the external device may
Low Technical have the following alarms:
Alarms CO2 ERR CO2 sensor ?
Plow>high limit
Plow<low limit

4.6 Draeger Evita 2


4.6.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume

4-23
Label Units Description Trend, record, print
MV L/min Minute volume Yes
Spontaneous
MVspn L/min breathed minute Yes
volume
Total respiratory
ftot bpm Yes
rate
Spontaneous
fspn bpm breathing Yes
frequency
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
cmH2O/L/s,hPa/L/s,mbar/ Dynamic lung
Rdyn Yes
L/s resistance
ml/cmH2O,ml/hPa,ml/mb Dynamic
Cdyn Yes
ar compliance
Minimum airway
Pmin cmH2O, hPa, mbar No
pressure
Vtrap ml Trapped volume No
Inspiratory
T ℃,℉ breathing gas No
temperature
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide

4-24
Label Units Description Trend, record, print
Vds ml Dead space No
VCO2 ml/min CO2 production No
End-tidal carbon
EtCO2% % Yes
dioxide
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
Setting Parameters
Oxygen
O2% % No
concentration
Positive
PEEP cmH2O, hPa, mbar end-expiratory No
pressure
VT Tidal volume No
fSIMV bpm Frequency of SIMV No
Inspiratory
I:E \ time:Expiratory No
time ratio
△int.PEEP cmH2O, hPa, mbar Intermittent PEEP No
Inspiratory trigger
F-Trigger L/min No
level (flow trigger)
Inspiratory trigger
P-Trigger cmH2O, hPa, mbar level (pressure No
trigger)
Upper pressure
Phigh cmH2O, hPa, mbar No
level
Lower pressure
Plow cmH2O, hPa, mbar No
level

4-25
Label Units Description Trend, record, print
Time for the upper
Thigh sec No
pressure level
Time for the lower
Tlow sec No
pressure level
Maximum airway
Pmax cmH2O, hPa, mbar No
pressure
Flow L/min Flow No
TApnea sec Apnea Time No
ASB ramp sec ASB ramp No
Assisted
PASB cmH2O, hPa, mbar spontaneous No
breathing

4.6.2 Output Signals - Alarms


Mindray Patient Monitor Draeger Evita 2 Alarm

Priority Label Label

Physiological alarms
Paw High PAW HIGH
Paw Low PAW LOW
MV High MIN VOL HIGH
MV Low MIN VOL LOW
Apnea APNEA EVITA
High
FiO2 High % O2 HIGH
FiO2 Low % O2 LOW
AW-TEMP HIGH AW-TEMP HIGH
PEEP High PEEP HIGH
ASB>4s ASB > 4 SEC

4-26
Mindray Patient Monitor Draeger Evita 2 Alarm

Priority Label Label

Physiological alarms
EtCO2 High ET CO2 HIGH
EtCO2 Low ET CO2 LOW
Medium
VOL INCONST VOL INCONST
RR High RESP RATE HI
Technical alarms
Air Supply Pressure
Low AIR SUPPLY ?
Check Flow Sensors FLOW SENSOR?
CLEAN CO2 CLEAN CO2
EXP-VALVE? EXP-VALVE ?
O2 Supply Pressure
LO O2 SUPPLY
Low
If the monitor shows the alarm [High
Technical alarms], the external device may
have the following alarms:
High
VOL ERR
PRESS ERR
AW-TEMP INOP
High Technical
AW-TEMP SENS
alarms
CO2 NOT CAL
% O2 ERR
EVITA ERR
COOLING INOP
CYCLE FAILED
COOLING INOP ADVISORY
Low Low Technical If the monitor shows the alarm [Low

4-27
Mindray Patient Monitor Draeger Evita 2 Alarm

Priority Label Label

Physiological alarms
alarms Technical alarms], the external device may
have the following alarms:
CO2 ERR
CO2 SENS?
MIXER INOP
SYNCHRO INOP
MIXER INOP ADV

4.7 Draeger Evita2 dura/Evita 4


4.7.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume
MV L/min Minute volume Yes
Spontaneous
MVspn L/min breathed minute Yes
volume
Total respiratory
ftot bpm Yes
rate
4-28
Label Units Description Trend, record, print
Spontaneous
fspn bpm breathing Yes
frequency
Fractional
FiO2 %,kPa,mmHg concentration of Yes
O2 in inspired gas
cmH2O/L/s,hPa/L/s, Dynamic lung
Rdyn Yes
mbar/L/s resistance
ml/cmH2O,ml/hPa, Dynamic
Cdyn Yes
ml/mbar compliance
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
Minimum airway
Pmin cmH2O, hPa, mbar No
pressure
Vtrap ml Trapped Volume No
Inspiratory
T ℃,℉ breathing gas No
temperature
Negative
NIF cmH2O, hPa, mbar No
inspiratory force
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Vds ml Dead space No
VCO2 ml/min CO2 production No

4-29
Label Units Description Trend, record, print
PR bpm Pulse rate Yes
Arterial oxygen
SpO2 % saturation from Yes
pulse oximetry
End-tidal carbon
EtCO2% % Yes
dioxide
Fractional
FiO2% % concentration of Yes
O2 in inspired gas
Setting Parameters
Oxygen
O2% % No
concentration
Positive
PEEP cmH2O, hPa, mbar end-expiratory No
pressure
VT ml Tidal volume No
f bpm Breath rate No
Inspiratory time:
I:E \ Expiratory time No
ratio
△int.PEEP cmH2O, hPa, mbar Intermittent PEEP No
Inspiratory
trigger
F-Trigger L/min No
level (flow
trigger)
Upper pressure
Phigh cmH2O, hPa, mbar No
level
Lower pressure
Plow cmH2O, hPa, mbar No
level

4-30
Label Units Description Trend, record, print
Time for the
Thigh sec upper pressure No
level
Time for the
Tlow sec lower pressure No
level
Maximum airway
Pmax cmH2O, hPa, mbar No
pressure
Time of
Tinsp sec No
inspiration
Flow L/min Flow No
TApnea sec Apnea time No
ASB ramp sec ASB ramp No
Assisted
PASB cmH2O, hPa, mbar spontaneous No
breathing
cmH2O.s/L,hPa.s/L, No
FlowAssist Flow assist
mbar.s/L
cmH2O/L,hPa/L,mb No
Vol.Assist Volume assist
ar/L
Delay time for No
Tdisconnect sec "Airway pressure
lower alarm limit"
Tube ID mm Tube ID No
Automatic Tube No
ATC %
Compensation
Pressure control No
Pinsp cmH2O, hPa, mbar level of
inspiration

4-31
Label Units Description Trend, record, print
ml Apnea tidal
VTapnea No
volume
Papnea cmH2O, hPa, mbar Apnea pressure No
bpm Breath rate for
fapnea No
apnea ventilation

4.7.2 Output Signals - Alarms


Mindray Patient Monitor Draeger Evita 4/ Evita2 dura Alarm

Priority Label Label

Physiological alarms
Paw High PAW HIGH
Paw Low PAW LOW
MV High MIN VOL HIGH
MV Low MIN VOL LOW
Apnea APNEA EVITA
FiO2 High % O2 HIGH
FiO2 Low % O2 LOW
High AW-TEMP HIGH AW-TEMP HI
PEEP High PEEP HIGH
ASB>4s ASB > 4 SEC
No Pulse NO SPO2 PULS
PR Low SPO2 PULS LO
SpO2 Low SPO2 LOW
PR High SPO2 PULS HI
SpO2 High SPO2 HIGH
VTe High TIDVOL HI
Medium
EtCO2 High ET CO2 HIGH

4-32
Mindray Patient Monitor Draeger Evita 4/ Evita2 dura Alarm

Priority Label Label


EtCO2 Low ET CO2 LOW
VOL INCONST VOL INCONST
RR High RESP RATE HI
ASB > 1.5s ASB > 1,5 SEC
Low PPS-TI > 1.5s PPS-TI > 1,5S
ASB > Tinsp ASB > TINSP
Technical alarms
Air Supply Pressure
AIR SUPPLY ?
Low
O2 Supply Pressure
LO O2 SUPPLY
Low
Airway Obstructed? TUBE OBSTRUC
Check Flow Sensors FLOW SENSOR?
CLEAN CO2 CLEAN CO2
EXP VALVE EXP-VALVE ?
If the monitor shows the alarm [High
Technical alarms], the external device may
High have the following alarms:
VOL ERR
PRESS ERR
AW-TEMP INOP
High Technical
EVITA ERR
alarms
CYCLE FAILED
N-VOL ERR
NEO FLOW ?
CO2 ZERO CAL
SPO2 SEN DISC
SPO2 ERR
4-33
Mindray Patient Monitor Draeger Evita 4/ Evita2 dura Alarm

Priority Label Label


BATTERY ERR

FAN ERR

AIR PRESS HI
HI O2 SUPPLY
LOSS OF DATA
REM.PAD-ERR
PEEP V ERR
CO2 NOT CAL

If the monitor shows the alarm [Medium


Technical alarms], the external device may
have the following alarms:
BATT. < 2MIN
Medium Technical
Medium CHECK EVITA
alarms
EVITA STDBY
AMB PRESS ?
NEBULIZ OFF
ERR MULTIPCB
Battery in Use BATTERY ON
If the monitor shows the alarm [Low
Technical alarms], the external device may
have the following alarms:
CO2 ERR
Low Low Technical
CO2 SENSOR ?
alarms
MIXER INOP
SYNCHRO INOP
INSPHOLD END
EXSPHOLD END
4-34
4.7.3 Output Signals - Waveforms (For BeneVision N series
Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL

NOTE

 The patient monitorin does not output waveforms from the Draeger
Evita XL ventilator.

4-35
4.8 Draeger Evita V300
4.8.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
ml/cmH2O,ml/hPa, Dynamic
Cdyn Yes
ml/mbar compliance
cmH2O/L/s,hPa/L/ Dynamic lung
Rdyn Yes
s,mbar/L/s resistance
VCO2 ml/min CO2 production No
cmH2O, hPa, mbar Minimum airway No
Pmin
pressure
cmH2O, hPa, mbar 100 ms occlusion No
P0.1
pressure
Pmean cmH2O, hPa, mbar Mean pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
cmH2O, hPa, mbar Positive Yes
PEEP end-expiratory
pressure
cmH2O, hPa, mbar Intrinsic positive No
PEEPi end-expiratory
pressure
bpm Mandatory Yes
fmand breathing
frequency
Ppeak cmH2O, hPa, mbar Peak pressure Yes

4-36
Label Units Description Trend, record, print
ml Volume trapped No
in the lung by
intrinsic PEEP,
Vtrap and not exhaled
during
subsequent
expiration
ml Spontaneous Yes
VTe spn expiratory tidal
volume
Vds ml Dead space No
cmH2O, hPa, mbar Negative No
NIF
inspiratory force
L/min Leakage minute
MVLEAK No
volume
% Leak
Leak Comp No
compensation
bpm Spontaneous Yes
fspn
respiratory rate
MV L/min Minute volume Yes
1/(min.L) Rapid shallow Yes
RSBI
breathing index
bpm Total respiratory Yes
ftot
rate
End-tidal carbon Yes
EtCO2 %,kPa,mmHg
dioxide
Inspiratory
I:E \ time:Expiratory No
time ratio

4-37
Label Units Description Trend, record, print
Fractional Yes
FiO2 %,kPa,mmHg concentration of
O2 in inspired gas
L/min Expiratory Yes
MVe
minute volume
L/min Inspiratory Yes
MVi
mimute volume
ml Expiratory tidal Yes
VTe
volume
ml Inspired tidal Yes
VTi
volume
VT/kg ml/kg TVe/IBW No
ml CO2 tidal No
VTCO2
elimination
End-tidal carbon Yes
EtCO2% %
dioxide
Fractional Yes
FiO2% % concentration of
O2 in inspired gas
Setting Parameters
% Oxygen No
O2 %
concentration
Flow L/min Flow No
sec Time of No
Tinsp
inspiration
f bpm Breath rate No
cmH2O, hPa, mbar Positive No
PEEP end-expiratory
pressure

4-38
Label Units Description Trend, record, print
Δint PEEP cmH2O, hPa, mbar Intermittent PEEP No
cmH2O, hPa, mbar Lower pressure No
Plow
level
cmH2O, hPa, mbar Upper pressure No
Phigh
level
sec Time for the No
Tlow lower pressure
level
sec Time for the No
Thigh upper pressure
level
TApnea sec Apnea time No
cmH2O, hPa, mbar Pressure support No
Psupp
level
cmH2O, hPa, mbar Maximum airway No
Pmax
pressure
L/min Inspiratory No
F- Triger r trigger level (flow
trigger)
Trise sec Rise time No
cmH2O.s/L,hPa.s/L, No
Flow Assist Flow assist
mbar.s/L
cmH2O/L,hPa/L,m No
Vol Assist Volume assist
bar/L
bpm Breath rate for No
fapnea
apnea ventilation
cmH2O, hPa, mbar Pressure control No
Pinsp level of
inspiration

4-39
Label Units Description Trend, record, print
% Automatic Tube No
ATC
Compensation
Tube ID mm Tube ID No
sec Delay time for No
Tdisconnect "Airway pressure
lower alarm limit"
sec Maximum No
Ti max
inspiration time
VT ml Tidal volume No
ml Apnea tidal No
VTapnea
volume
Inspiration No
Exp% %
termination level
Trigger % Trigger No
Inspiratory time:
I:E \ Expiratory time No
ratio

4-40
4.8.2 Output Signals—Alarms
Mindray Patient Monitor Draeger Evita V300 Alarm

Priority Label Label


Physiological alarms
FiO2 Low FiO2 LOW
FiO2 High O2 HIGH
Paw High Airway pressure high
Paw Low PAW LOW
MV High MIN VOL HIGH
High MV Low MIN VOL LOW
Apnea APNEA RESP
PEEP High PEEP HIGH
PEEP Low PEEP LOW
Apnea Ventilation APNEA VENT
VT Not Achieved TIDAL VOL LO
EtCO2 Low ETCO2 LOW
EtCO2 High ETCO2 HIGH
Medium VOL INCONST VOL INCONST
RR High RESP RATE HI
VTe High TIDAL VOL HI

Technical alarms

Air Supply Pressure Low AIR SUPPLY


O2 Supply Pressure Low LOW O2 SUPPLY
FLOW SENSOR
Check Flow Sensors
EXP TIME ERR
High
CLEAN CO2 CLEAN CO2
Negative Airway Pressure PAW NEGATIVE
EXP-VALVE? EXP-VALVE
No O2 Pressure NO OXYGEN

4-41
Mindray Patient Monitor Draeger Evita V300 Alarm

Priority Label Label


Circuit Disconnect DISCONNECT
Neo Flow Sensor Error NEO FLOW
Airway Obstructed? TUBE OBSTRUC
Power Failure POWER ERR
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
CO2 NOT CAL
BATTERY ERR
SPEAKER FAIL
EVITA ERR
% O2 ERR
VOL CAL
VOL ERR
PRESS ERR
AW-TEMP INOP
High Technical Alarm
COOLING
INT.TMP.HIGH
CO2 SENSOR
AIR PRESS HI
HI O2 SUPPLY
SYSTEM FAUL
LOSS OF DATA
HOSE ERROR
SC ABORTED
SC INOP
CENTRAL HYPO
PERS TACHYP

4-42
Mindray Patient Monitor Draeger Evita V300 Alarm

Priority Label Label


UNEXPL HYPER
If the monitor shows the alarm
[Medium Technical alarms], the
external device may have the
following alarms:
GAS FAILURE
Medium Medium Technical Alarm
NO AIR
AMB PRESS
CHECK EVITA
NEBULIZ. OFF
BATT. LOW
Battery in Use BATTERY ON
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
Low EXPHOLD END
Low Technical Alarm
PMIN REACHED
PLOW LOW
PLOW HIGH
CO2 ERR

4-43
4.9 Draeger Infinity V500
4.9.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
ml/cmH2O,ml/hP Dynamic
Cdyn Yes
a,ml/mbar compliance
cmH2O/L/s,hPa/L/ Dynamic lung
Rdyn Yes
s,mbar/L/s resistance
VCO2 ml/min CO2 production No
cmH2O, hPa, Minimum airway
Pmin No
mbar pressure
cmH2O, hPa, 100 ms occlusion
P0.1 No
mbar pressure
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
cmH2O, hPa,
Pplat Plateau pressure Yes
mbar
Positive
cmH2O, hPa,
PEEP end-expiratory Yes
mbar
pressure
Intrinsic positive
cmH2O, hPa,
PEEPi end-expiratory No
mbar
pressure
Mandatory
fmand bpm breathing Yes
frequency
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar

4-44
Label Units Description Trend, record, print
Volume trapped in
the lung by
intrinsic PEEP, and
Vtrap ml No
not exhaled during
subsequent
expiration
Spontaneous
VTe spn ml expiratory tidal Yes
volume
Vds ml Dead space No
cmH2O, hPa, Negative
NIF No
mbar inspiratory force
Leakage minute
MVLEAK L/min Yes
volume
Leak Comp % Leak compensation No
bpm Spontaneous
fspn Yes
respiratory rate
MV L/min Minute volume Yes
1/(min.L) Rapid shallow
RSBI Yes
breathing index
bpm Total respiratory
ftot Yes
rate
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
End-tidal carbon
EtCO2% % Yes
dioxide
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas

4-45
Label Units Description Trend, record, print
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
Expiratory tidal
VTe ml Yes
volume
Inspired tidal
VTi ml Yes
volume
Expiratory minute
MVe L/min Yes
volume
Inspiratory mimute
MVi L/min Yes
volume
CO2 tidal
VTCO2 ml No
elimination
Inspiratory
I:E \ time:Expiratory No
time ratio
ml/cmH2O,ml/hP
Cstat Static compliance Yes
a,ml/mbar
VT/kg ml/kg TVe/IBW No
Setting Parameters
Oxygen
O2% % No
concentration
Flow sec Flow No
Tinsp sec Time of inspiration No
Inspiratory
I:E \ time:Expiratory No
time ratio
f bpm Breath rate No

4-46
Label Units Description Trend, record, print
Positive
cmH2O, hPa,
PEEP end-expiratory No
mbar
pressure
cmH2O, hPa,
Δint PEEP Intermittent PEEP No
mbar
cmH2O, hPa, Lower pressure
Plow No
mbar level
cmH2O, hPa, Upper pressure
Phigh No
mbar level
Time for the lower
Tlow sec No
pressure level
Time for the upper
Thigh sec No
pressure level
TApnea sec Apnea time No
cmH2O, hPa, Pressure support
Psupp No
mbar level
cmH2O, hPa, Maximum airway
Pmax No
mbar pressure
Inspiratory trigger
F-triger L/min No
level (flow trigger)
Trise sec Rise time No
cmH2O.s/L,hPa.s/
Flow Assist Flow assist No
L,mbar.s/L
cmH2O/L,hPa/L,m
Vol Assist Volume assist No
bar/L
VT ml Tidal volume No
Breath rate for
fapnea bpm No
apnea ventilation

4-47
Label Units Description Trend, record, print
cmH2O, hPa, Pressure control
Pinsp No
mbar level of inspiration
Automatic tube
ATC % No
compensation
Tube ID mm Tube ID No
Delay time for
Tdisconnect sec "Airway pressure No
lower alarm limit"
Maximum
Ti max sec No
inspiration time
VTapnea ml Apnea tidal volume No
Tpause % Apnea Time No
Inspiration
Exp% % No
termination level

4.9.2 Output Signals—Alarms


Mindray Patient Monitor Draeger Evita V500 Alarm

Priority Label Label


Physiological alarms
FiO2 Low FiO2 low
Paw High Airway pressure high
MV Low MV low
MV High MV high
High FiO2 High FiO2 high
Apnea APNEA RESP
Paw Low Airway pressure low
PEEP High PEEP high
PEEP Low PEEP low
4-48
Mindray Patient Monitor Draeger Evita V500 Alarm

Priority Label Label


Apnea ventilation Apnea ventilation
EtCO2 Low etCO2 low
EtCO2 High etCO2 high
RR High High respiratory rate
Medium
VTe High Tidal volume high
VTe Low Tidal volume low
VOL INCONST VT not reached
Technical alarms
Air Supply Pressure
Air supply down
Low
O2 Supply Pressure
O2 supply down
Low
CLEAN CO2 Clean CO2 cuvette
Power Failure Internal power supply failure
Check Flow Sensors Flow sensor failure
EXP-VALVE? Expiratory valve malfunction
Negative Airway
Airway pressure negative
Pressure
High
Neo Flow Sensor
Neo. flow sensor changed ?
Error
Circuit Disconnect Disconnection Ventilator
No O2 Pressure O2 supply down
AirWay Obstructed Tube Obstruct
If the monitor shows the alarm [High
Technical alarms], the external device may
High Technical
have the following alarms:
Alarm
CO2 NOT CAL
BATERY ERR
4-49
Mindray Patient Monitor Draeger Evita V500 Alarm

Priority Label Label


SPEAKER FAIL
VENT ERR
VOL CAL
VOL ERR
AW TEMP INOP
AIR PRESS HI
HI O2 SUPPLY
SYSTEM FAULT
BATTER ERR
COOLING
LOSS OF DATA
HOSE ERR
O2 Error
Press Err
AW TEMP Sens
INT TEM HIGH
Evita Err
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
Air supply down
Medium Technical
Medium Nebulization finished
Alarm
Ambient pressure sensor?
Device check incomplete
GAS FAILURE
BATT LOW
Battery in Use Internal battery activated
Low
Low Technical If the monitor shows the alarm [Low

4-50
Mindray Patient Monitor Draeger Evita V500 Alarm

Priority Label Label


Alarm Technical alarms], the external device may
have the following alarms:
Exp. hold interrupted
CO2 measurement failed
CO2 sensor ?
Plow>high limit
Plow<low limit
PMIN REACHED
CO2 Err

4.10 Draeger Savina 300


4.10.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
ml/cmH2O,ml/hP Dynamic
Cdyn Yes
a,ml/mbar compliance
cmH2O/L/s,hPa/L/ Dynamic lung
Rdyn Yes
s,mbar/L/s resistance
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
Positive
cmH2O, hPa,
PEEP end-expiratory Yes
mbar
pressure
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
Expiratory tidal
VTe ml Yes
volume

4-51
Label Units Description Trend, record, print
Inspired tidal
VTi ml Yes
volume
MV L/min Minute volume Yes
Total respiratory
ftot bpm Yes
rate
Spontaneous
MVspn L/min breathed minute Yes
volume
Spontaneous
fspn bpm Yes
respiratory rate
Fractional
FiO2 %,kPa,mmHg concentration of Yes
O2 in inspired gas
Inspiratory
T ℃, ℉ breathing gas No
temperature
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
cmH2O, hPa,
Pplat Plateau pressure Yes
mbar
Inspiratory time:
I:E \ Expiratory time No
ratio
End-tidal carbon
EtCO2% % Yes
dioxide

4-52
Label Units Description Trend, record, print
Fractional
FiO2% % concentration of Yes
O2 in inspired gas
Setting Parameters
Oxygen
O2 % % No
concentration
VT ml Tidal volume No
Apnea tidal No
VTapnea ml
volume
Breath rate for No
fapnea bpm
apnea ventilation
f bpm Breath rate No
Inspiratory time: No
I:E \ Expiratory time
ratio
cmH2O, hPa, No
△int.PEEP Intermittent PEEP
mbar
Inspiratory No
F-Trigger L/min trigger level (flow
trigger)
Tapnea sec Apnea Time No
Assisted No
cmH2O, hPa,
PASB spontaneous
mbar
breathing
Pressure control No
cmH2O, hPa,
Pinsp level of
mbar
inspiration
cmH2O, hPa, Maximum airway No
Pmax
mbar pressure

4-53
Label Units Description Trend, record, print
Delay time for No
Tdisconnect sec "Airway pressure
lower alarm limit"
cmH2O/s,hPa/s,m No
FlowACC Flow acceleration
bar/s
cmH2O, hPa, Lower pressure No
Plow
mbar level
cmH2O, hPa, Upper pressure No
Phigh
mbar level
Time for the No
Thigh sec upper pressure
level
Time for the No
Tlow sec lower pressure
level
Time of No
Tinsp sec
inspiration
Positive No
cmH2O, hPa,
PEEP end-expiratory
mbar
pressure

4.10.2 Output Signals—Alarms


Mindray Patient Monitor Draeger Savina 300 Alarm

Priority Label Label

Physiological alarms
Paw High Air pressure high
High Paw Low Air pressure low
MV High MV high
4-54
Mindray Patient Monitor Draeger Savina 300 Alarm

Priority Label Label


MV Low MV low
FiO2 High O2 HIGH
FiO2 Low O2 low
PEEP High PEEP HIGH
ASB>4s Assist Spontaneous Breathing>4s
Apnea Ventilation Apnea ventilation
Apnea APNEA SAVINA
AW-TEMP HIGH Airway temperature high
VTe High Tidal volume high
VTe Low Tidal volume low
Medium RR High RESP RATE HI
EtCO2 Low EtCO2 low
EtCO2 High EtCO2 high
Technical alarms
O2 Supply Pressure
O2 supply pressure low
Low
Check Flow Sensors Check flow sensor
EXP-VALVE? Check expiratory valve?
Clean CO2 CleanCO2
If the monitor shows the alarm [High
Technical alarms], the external device may
High
have the following alarms:
High O2 supply pressure
High Technical
insp.O2 measurement in operation
Alarm
Failen to cycle
Problem with fan
Problem with PEEP control
volume measurement inoperable
4-55
Mindray Patient Monitor Draeger Savina 300 Alarm

Priority Label Label


Pressure measurement inoperable
CO2 not calibrated
COOLING
SAVINA ERR
BATTERY ERR
AW TEMP INOP
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
Check cooling
Medium Technical
Medium CHECK SAVINA 300
Alarm
NO nubelizer
SAVINA STDBY
BATTERY LOW
MICROFILTER
If the monitor shows the alarm [Low
Technical alarms], the external device may
have the following alarms:
Low Technical
Low Inspiration hold aborted
Alarm
CO2 device failure
CO2 sensor disconnected
Expiration hold aborted

4-56
4.11 GE Carescape R860
4.11.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume
VTi ml Inspired tidal volume Yes
VT/kg ml/kg TVe/IBW No
Spontaneous
VTe spn ml expiratory tidal Yes
volume
Spontaneous
MVspn L/min breathed minute Yes
volume
Expiratory minute
MVe L/min Yes
volume
Inspiratory mimute
MVi L/min Yes
volume
Mandatory breathing
fmand bpm Yes
frequency
Spontaneous
fspn bpm Yes
respiratory rate
Inspiratory
I:E \ time:Expiratory time No
ratio

4-57
Label Units Description Trend, record, print
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
EtO2 %,kPa,mmHg End-tidal O2 Yes
Difference between
ΔO2 %,kPa,mmHg inspiratory and No
expiratory O2
cmH2O/L/s,hPa/L/s Dynamic lung
Rdyn Yes
,mbar/L/s resistance
ml/cmH2O,ml/hPa,
Cstat Static compliance Yes
ml/mbar
ml/cmH2O,ml/hPa,
Cdyn Dynamic compliance Yes
ml/mbar
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
Texp sec Expiratory time No
Peak auxiliary
Paux Peak cmH2O, hPa, mbar No
pressure
Mean auxiliary
Paux Mean cmH2O, hPa, mbar No
pressure
Minimum auxiliary
Paux Min cmH2O, hPa, mbar No
pressure
oxygen supply
PO2 kPa No
pressure
Pair kPa air supply pressure No
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure

4-58
Label Units Description Trend, record, print
Extrinsic positive
PEEPe cmH2O, hPa, mbar end-expiratory No
pressure
PEEPtot cmH2O, hPa, mbar Total PEEP No
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Respiratory rate of
RRCO2 bpm Yes
CO2
EE kcal/day Energy expenditure No
RQ \ Respiratory quotient No
VO2 ml/min Oxygen consumption Yes
VCO2 ml/min CO2 production No
Oxygen consumption
VO2/m2 ml/min/m2 No
per body surface area
CO2 consumption per
VCO2/ m2 ml/min/m2 No
body surface area
Oxygen consumption
VO2/kg ml/min/kg No
per body weight
CO2 consumption per
VCO2/kg ml/min/kg No
body weight
ftot bpm Total respiratory rate Yes
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
EtO2% %, End-tidal O2 Yes
End-tidal carbon
EtCO2% % Yes
dioxide

4-59
Label Units Description Trend, record, print
Difference between
ΔO2 % % inspiratory and No
expiratory O2
Setting Parameters
Positive
PEEP cmH2O, hPa, mbar end-expiratory No
pressure
VT ml Tidal volume No
Base Flow L/min Base Flow No
Tsupp sec Support time No
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Inspiratory trigger No
P-Trigger cmH2O, hPa, mbar level (pressure
trigger)
Pressure support No
Psupp cmH2O, hPa, mbar
level
Plimit cmH2O, hPa, mbar Pressure limit level No
Tinsp sec Time of inspiration No
Pressure control level
Pinsp cmH2O, hPa, mbar No
of inspiration
Trise sec Rise time No
Tpause sec Apnea Time No
Phigh cmH2O, hPa, mbar Upper pressure level No
Plow cmH2O, hPa, mbar Lower pressure level No
Time for the upper No
Thigh sec
pressure level
Time for the lower No
Tlow sec
pressure level

4-60
Label Units Description Trend, record, print
Maximum airway No
Pmax cmH2O, hPa, mbar
pressure
Flow L/min Flow No
Tapnea sec Apnea Time No
Delay time for No
Tdisconnect sec "Airway pressure
lower alarm limit"
Tube ID mm Tube ID No
Oxygen No
O2% %
concentration
Inspiratory No
I:E \ time:Expiratory time
ratio

4.11.2 Output Signals—Alarms


Mindray Patient Monitor GE Carescape R860 Alarm

Priority Label Label


Physiological alarms
Apnea Apnea
AW-TEMP HIGH Air Temp High
FiO2 High FiO2 high
FiO2 Low FiO2 low
MV High MVexp high
High
MV Low MVexp low
Ppeak High Ppeak High
Ppeak Low Ppeak Low
Pressure Sustained Sustained Paw
VT Not Achieved Tidal volume not delivered(VT Not
4-61
Mindray Patient Monitor GE Carescape R860 Alarm

Priority Label Label


Achieved)
EtCO2 High EtCO2 High
EtCO2 Low EtCO2 Low
EtO2 High EtO2 High
EtO2 Low EtO2 Low
Paux High Paux High
PEEPe High PEEPe High
Medium
PEEPe Low PEEPe Low
PEEPi High PEEPi High
RR High RR High
RR Low RR Low
VTe High VTexp high
VTe Low VTexp Low
Low Plimit Reached Plimit Reached
Technical alarms
Air Supply Pressure
Air Supply Pressure Low
Low
Airway Obstructed? Breathing Circuit Occlusion
Circuit Leak Circuit Leak
Expiratory Flow Sensor Error
Check Flow Sensors
Exp Flow Sensor Failure
High Negative Airway
Negative Airway Pressure
Pressure
Neo Flow Sensor
Neo Flow Sensor Error
Error
No Gas Supply
No Gas Supply Pressure
Pressure
O2 Supply Pressure O2 Supply Pressure Low
4-62
Mindray Patient Monitor GE Carescape R860 Alarm

Priority Label Label


Low
Patient Connection
Patient Connection Leak
Leak
Patient detected
Patient Connected?
(Patient Connected?)
Patient Disconnected Patient Disconnected
Power Failure Power Supply Fail
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
Air Supply Pressure High
Air Temp Sensor Error
Clean Neo Flow Sensor
FiO2 Control Error
Relief Valve Opened
Low Internal Battery 1 Min
Low Internal Battery 5 Min
High Technical Alarm Low Internal Battery 10 Min
Mixed Gas Temp Sensor Error
Neo Flow Sensor Reversed
Neo Flow Sensor Off
No D-Lite Sensor?
No Expiratory Flow Sensor
No Neo Flow Sensor
O2 Supply Pressure High
O2 Temp Sensor Error
Pressure Sensor Failure
Primary Audio Failure

4-63
Mindray Patient Monitor GE Carescape R860 Alarm

Priority Label Label


Relief Valve Failure
Replace Neo Flow Sensor
Temp High Shutdown Possible
Total Flow Sensor
Communication Failure
Volume Delivery Error
FiO2 Sensor
O2 Sensor Failure
Disconnected
If the monitor shows the alarm [Medium
Technical alarms], the external device
may have the following alarms:
Backup Ventilation on
Check D-fend
Check Sample Gas Out
Circuit Leak Alarm Off
Medium Fans Require Service
Medium Technical
Low Internal Battery 20 Min
Alarm
Module Fail No CO2, O2 Data
No Battery Backup
Replace D-fend
Sample Flow Deviation
Sample Line Blocked
SBT Ended
No Battery Backup
Backup Audio Failure
Low Battery in Use Battery in use
If the monitor shows the alarm [Low
Low Low Technical Alarm
Technical alarms], the external device

4-64
Mindray Patient Monitor GE Carescape R860 Alarm

Priority Label Label


may have the following alarms:Air Supply
Pressure Sensor Out of Range
Alarm Light Failure
Alarms Silenced
Apnea Alarm Off
Cannot Calculate FRC
Carrier Board Overheat
Case Fan Speed Fail
Connect Nebulizer
Controls Frozen Need Service
CPU Fan Speed Fail
CPU Overheat
Missed Scheduled FRC
Module Not Compatible
Module Warming Up 2 Min
Module Warming Up 5 Min
No Patient Effort
MVexp Low Alarm Off
O2 Supply Pressure Sensor Out of Range
Pinsp Sensor Out of Range
Pexp Sensor Out of Range
Paux Sensor Out of Range
SBT Completed successfully
Touchscreen Failure

4-65
4.11.3 Output Signals - Waveforms (For BeneVision N series
Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL
CO2 %,kPa,mmHg

4.11.4 Output Signals – Respiratory Loops (For BeneVision N


series Monitors)
Label Units
PV Loop cmH2O/ml, hPa/ml, mbar/ml
FV Loop L/min/ml
PF Loop cmH2O/L/min, hPa/L/min, mbar/L/min

4.12 GE Engström Carestation


4.12.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume
VTi ml Inspired tidal volume Yes
VT/kg ml/kg TVe/IBW No
Spontaneous
VTe spn ml expiratory tidal Yes
volume

4-66
Label Units Description Trend, record, print
Spontaneous
MVspn L/min breathed minute Yes
volume
Expiratory minute
MVe L/min Yes
volume
Inspiratory mimute
MVi L/min Yes
volume
ftot bpm Total respiratory rate Yes
Mandatory breathing
fmand bpm No
frequency
Spontaneous
fspn bpm Yes
breathing frequency
Inspiratory time:
I:E \ Yes
Expiratory time ratio
Fractional
FiO2 %,kPa,mmHg concentration of O2 Yes
in inspired gas
EtO2 %,kPa,mmHg End-tidal O2 Yes
Difference between
ΔO2 %,kPa,mmHg inspiratory and No
expiratory O2
cmH2O/L/s,hPa/L/s Dynamic lung
Rdyn Yes
,mbar/L/s resistance
ml/cmH2O,ml/hPa,
Cstat Static compliance Yes
ml/mbar
ml/cmH2O,ml/hPa,
Cdyn Dynamic compliance Yes
ml/mbar
Rapid shallow
RSBI 1/(min.L) Yes
breathing index

4-67
Label Units Description Trend, record, print
Texp sec Expiratory time No
Peak auxiliary
Paux Peak cmH2O, hPa, mbar No
pressure
Mean auxiliary
Paux Mean cmH2O, hPa, mbar No
pressure
Minimum auxiliary
Paux Min cmH2O, hPa, mbar No
pressure
oxygen supply
PO2 kPa No
pressure
Pair kPa air supply pressure No
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
Extrinsic positive
PEEPe cmH2O, hPa, mbar end-expiratory No
pressure
PEEPtot cmH2O, hPa, mbar Total PEEP No
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Respiratory rate of
RRCO2 bpm Yes
CO2
EE kcal/day Energy expenditure No
RQ \ Respiratory quotient No
VO2 ml/min Oxygen consumption Yes
VCO2 ml/min CO2 production No
Oxygen consumption
VO2/m2 ml/min/m2 No
per body surface area
CO2 consumption per
VCO2/m2 ml/min/m2 No
body surface area

4-68
Label Units Description Trend, record, print
Oxygen consumption
VO2/kg ml/min/kg No
per body weight
CO2 consumption per
VCO2/kg ml/min/kg No
body weight
Fractional
FiO2% % concentration of O2 Yes
in inspired gas
EtO2% % End-tidal O2 Yes
End-tidal carbon
EtCO2% % Yes
dioxide
Difference between
ΔO2 % % inspiratory and No
expiratory O2
Setting Parameters
Oxygen
O2% % No
concentration
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure
VT ml Tidal volume No
Base Flow L/min Base Flow No
Tsupp sec Support time No
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Trigger % Trigger No
Inspiratory trigger No
P-Trigger cmH2O, hPa, mbar level (pressure
trigger)

4-69
Label Units Description Trend, record, print
Pressure support No
Psupp cmH2O, hPa, mbar
level
Plimit cmH2O, hPa, mbar Pressure limit level No
Tinsp sec Time of inspiration No
Pressure control level No
Pinsp cmH2O, hPa, mbar
of inspiration
Trise sec Rise time No
Phigh cmH2O, hPa, mbar Upper pressure level No
Plow cmH2O, hPa, mbar Lower pressure level No
Time for the upper No
Thigh sec
pressure level
Time for the lower No
Tlow sec
pressure level
Maximum airway No
Pmax cmH2O, hPa, mbar
pressure
Flow L/min Flow No
Tapnea sec Apnea time No
Delay time for No
Tdisconnect sec "Airway pressure
lower alarm limit"
Tube ID mm Tube ID No
Tpause sec Apnea Time No
Inspiratory time:
I:E \
Expiratory time ratio

4-70
4.12.2 Output Signals - Alarms
Mindray Patient Monitor GE Engstrom Carestation Alarm

Priority Label Label

Physiological alarms
MV High MVexp High

MV Low MVexp Low

Apnea Apnea
FiO2 High FiO2 High

High FiO2 Low FiO2 Low


Pressure Sustained Sustained Paw
VT not Achieved TV Not Achieved
AW TEMP High Air Temp High
Ppeak High Ppeak High
Ppeak Low Ppeak Low
VTe Low TVexp Low
RR Low RR Low
EtCO2 High EtCO2 High
EtCO2 Low EtCO2 Low
RR High RR High
EtO2 High EtO2 High
Medium
EtO2 Low EtO2 Low
PEEPe High PEEPe High
PEEPe Low PEEPe Low
PEEPi High PEEPi High
VTe High TVexp High
Paux High Paux High
Plimit Reached Plimit Reached
Low
Base Flow High Base Flow High
Technical alarms
4-71
Mindray Patient Monitor GE Engstrom Carestation Alarm

Priority Label Label


Air Supply Pressure
Air Supply Pressure Low
Low
O2 Supply Pressure
O2 Supply Pressure Low
Low
No Gas Supply
No Gas Supply Pressure
Pressure
Airway Obstructed? Breathing Circuit Occlusion
Check Flow Sensors Exp Flow Sensor Error
Neo Flow Sensor
Neo Flow Sensor Error
Error
Patient Connected? Patient Connected?
Negative Airway
Negative Airway Pressure
Pressure
Circuit Leak Circuit Leak
High
Patient Connection
Patient Connection Leak
Leak
Patient Disconnected Patient Disconnected
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms: temp high
shutdown possible
primary audio failure
High Technical Alarm Neo Flow Sendor Reversed
Clean Neo Flow Sensor
Replace Neo Flow Sensor
Neo Flow Sensor Off
No Expiratory Flow Sensor
No Neo Flow Sensor

4-72
Mindray Patient Monitor GE Engstrom Carestation Alarm

Priority Label Label


Air Supply Pressure High
O2 Supply Pressure High
FiO2 Control Error
Volume Delivery Error
Air Temp Sensor Error
O2 Temp Sensor Error
Mixed Gas Temp Sensor Error
Total Flow Sensor Communication Failure
No D-Lite Sensor?
Pressure Sensor Failure
Low Internal Battery 20 Min
Low Internal Battery 10 Min
Low Internal Battery 5 Min
Low Internal Battery 1 Min
Relief Valve Failure
No Patient Effort
FiO2 Sensor
O2 Sensor Failure
Disconnected
If the monitor shows the alarm [Medium
Technical alarms], the external device
may have the following alarms:
Check Sample Gas Out
Medium
Medium Technical Replace D-fend
Alarm Sample Line Blocked
Check D-fend
Sample Flow Deviation
Module Fail No CO2, O2 Data
No Battery Backup

4-73
Mindray Patient Monitor GE Engstrom Carestation Alarm

Priority Label Label


Fans Require Service
Backup Mode Active
Backcup audio failure
Display Fans Fail
Battery in Use On Battery
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
Air Supply Pressure Sensor Out of Range
O2 Supply Pressure Sensor Out of Range
Pinsp Sensor Out of Range
Pexp Sensor Out of Range
Paux Sensor Out of Range
Module Not Compatible
Controls Frozen Need Service
Connect Nebulizer
Low
Low Technical Alarm SBT Ends < 2 Miniutes
SBT Completed
SBT Ended
Cannot Calculate FRC
Missed Scheduled FRC
FRC Series Stopped
Relief Valve Opened
Circuit Leak Alarm Off
Apnea Alarm Off
MVexp Low Alarm Off
Unable to Deliver TV
VO2/VCO2/CO2/O2 Out of Range

4-74
Mindray Patient Monitor GE Engstrom Carestation Alarm

Priority Label Label


No VO2
FiO2 > 85%
Artifact
Module Warming Up 2 Min
Module Warming Up 5 Min

Alarms Silenced.

4.13 Hamilton C2/C3/C1 /C1 NEO/C6/C6S/T1

NOTE

 Only the Benevision N Sereis monitor supports integration with the


Hamilton C1/C1 NEO/C6/C6S/T1 ventilator.

4.13.1 Output Signals - Parameters


Label Units Description Trend, record, print
Monitoring Parameters
Oxygen
O2% % Yes
concentration
MV L/min Minute volume Yes
Inspiratory
I: E \ time:Expiratory time No
ratio
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
VTi ml Inspired tidal volume Yes
Expiratory tidal
VTe ml Yes
volume
4-75
Label Units Description Trend, record, print
ftot bpm Total respiratory rate Yes
Spontaneous
Yes
fspn bpm respiratory rate
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Texp sec Expiratory time No
cmH2O/L/s,hPa/L/s,m
Ri Inspiratory resistance Yes
bar/L/s
ml/cmH2O,ml/hPa,ml
Cstat Static compliance Yes
/mbar
Insp.Flow L/min Inspiration flow No
Spontaneous
VTe spn ml expiratory tidal Yes
volume
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Exp. Flow L/min Expiratory flow No
Expiratory time
RCexp sec No
constant
Pressure time
PTP cmH2O.s,hPa.s,mbar.s No
product
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Setting Parameters

4-76
Label Units Description Trend, record, print
fCMV bpm CMV frequency No
fSIMV bpm Frequency of SIMV No
VT ml Tidal volume No
PEEP/CPAP cmH2O, hPa, mbar PEEP/CPAP No
Pressure support No
Psupp cmH2O, hPa, mbar
level
Oxygen No
O2% %
concentration
MV L/min Minute volume No
Phigh cmH2O, hPa, mbar Upper pressure level No
Inspiratory No
I: E \ time:Expiratory time
ratio
Inspiration No
Exp% %
termination level
Ramp msec Ramp No
IBW kg Ideal body weight No
Percentage of minute No
%MinVol % volume to be
delivered

4-77
4.13.2 Output Signals - Alarms
Mindray Patient Monitor Hamilton Alarm

Priority Label Label

Physiological alarms
Paw High High Pressue
Apnea Apnea
High Loss of PEEP Loss of PEEP
MV Low Low Min Vol
MV High High Min Vol
Medium RR High High Rate
Technical alarms
Disconnection
Disconnection Ventilator
ventilator side
Patient
Disconnection Patient
High Disconnected
If the monitor shows the alarm [High
High Technical
Technical alarms], the external device may
alarms
have the alarm: Oxygen Concentration.
If the monitor shows the alarm [Medium
Technical alarms], the external device may
Medium Technical
Medium have the alarm:
alarms
Gas Supply
Spez Alarm
If the monitor shows the alarm [Low
Technical alarms], the external device may
Low Technical
Low have the following alarms:
alarms
Operator
General Alarm

4-78
4.14 Hamilton G5 (Block Protocol)
4.14.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
MV L/min Minute volume Yes
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Minimum airway
Pmin cmH2O, hPa, mbar No
pressure
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Pressure time
PTP cmH2O.s,hPa.s,mba.s No
product
Insp.Flow L/min Inspiration flow No
Exp. Flow L/min Expiratory flow No
Inspired tidal
VTi ml Yes
volume
Expiratory tidal
VTe ml Yes
volume
Spontaneous
VTe spn ml expiratory tidal Yes
volume

4-79
Label Units Description Trend, record, print
Spontaneous
MVspn L/min breathed minute Yes
volume
Total respiratory
ftot bpm Yes
rate
Spontaneous
fspn bpm Yes
respiratory rate
Texp sec Expiratory time No
Inspiratory
I:E \ time:Expiratory time No
ratio
cmH2O/L/s,hPa/L/s,mb Inspiratory
Ri Yes
ar/L/s resistance
cmH2O/L/s,hPa/L/s,mb Expiratory
Re Yes
ar/L/s resistance
ml/cmH2O,ml/hPa,ml/
Cstat Static compliance Yes
mbar
Expiratory time
RCexp sec No
constant
Inspiratory time
RCinsp sec No
constant
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
Oxygen
O2% % Yes
concentration
WOB J/L Work of breathing Yes
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
VCO2 ml/min CO2 production No

4-80
Label Units Description Trend, record, print
PR bpm Pulse rate Yes
Arterial oxygen
SpO2 % saturation from Yes
pulse oximetry
VT/Kg ml/kg TVe/IBW No
Setting Parameters
f bpm Breath rate No
VT ml Tidal volume No
TPause % Apnea Time No
cmH2O, hPa, mbar Inspiratory trigger No
P-Trigger level(pressure
trigger)
PEEP/CPAP cmH2O, hPa, mbar PEEP/CPAP No
cmH2O, hPa, mbar Lower pressure No
Plow
level
cmH2O, hPa, mbar Pressure support No
Psupp
level
MV L/min Minute volume No
Plimit cmH2O, hPa, mbar Pressure limit level No
Pressure control
Pinsp cmH2O, hPa, mbar No
level of inspiration
Upper pressure No
Phigh cmH2O, hPa, mbar
level
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Inspiratory No
I:E \ time:Expiratory time
ratio
Peak Flow L/min Peak flow No

4-81
Label Units Description Trend, record, print
Inspiration No
Exp% %
termination level
Ramp msec Ramp No
IBW kg Ideal body weight No
Percentage of No
%MinVol % minute volume to
be delivered
Time for the lower No
Tlow sec
pressure level
Time for the upper No
Thigh sec
pressure level
Maximum No
Ti max sec
inspiration time
Inspiratory pause No
Tip sec
time
Tube ID mm Tube ID No
Base flow L/min Base Flow No
Oxygen No
O2% %
concentration

4.14.2 Output Signals - Alarms


Mindray Patient Monitor Hamilton G5 (Block) Alarm

Priority Label Label

Physiological alarms
Paw High High pressure
Paw Low Low pressure
High
FiO2 High High oxygen
FiO2 Low Low oxygen
4-82
Mindray Patient Monitor Hamilton G5 (Block) Alarm

Priority Label Label


Apnea Apnea
Apnea Ventilation Apnea ventilation
SpO2 Low SpO2 too low
SpO2 High SpO2 too high
MV Low Low minute volume
MV High High minute volume
Loss of PEEP Loss of PEEP
RR High High frequency
RR Low Low frequency
Medium
EtCO2 High High PetCO2
EtCO2 Low Low PetCO2
Technical alarms
Patient Disconnection Patient
Disconnected Disconnection on patient side
Air Supply Pressure
Air supply failed
Low
O2 Supply Pressure
Oxygen supply failed
Low
O2 cell disconnect O2 cell missing
High O2 cell cal. Needed O2 cell calibration needed
Power Failure Loss of mains power
Check Flow Sensors Check Flow Sensor type
No Gas Supply
All gas supplies failed
Pressure
Disconnection Disconnection Ventilator
ventilator side Disconnection on ventilator side
High Technical If the monitor shows the alarm [High

4-83
Mindray Patient Monitor Hamilton G5 (Block) Alarm

Priority Label Label


alarms Technical alarms], the external device may
have the following alarms:
Wrong Flow Sensor type
O2 cell defective
Disconncetion
Low internal pressure
High pressure during sigh
Pressure not released
Exhalation obstructed
TF5514:Check loudspeaker
Internal battery empty
Ventilator unit connection lost
Check internal battery
O2 and air supply Oxygen + air supplies failed
O2 and heliox
Oxygen + heliox supplies failed
supply
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
High leak
Medium Low tidal volume
Medium Technical High tidal volume
alarms Turn the Flow Sensor
APV init failed
Internal battery low
Panel connection lost
Heliox supply failed
SpO2: sensor error(left slot)

4-84
Mindray Patient Monitor Hamilton G5 (Block) Alarm

Priority Label Label


SpO2: sensor error(right slot)
SpO2: no sensor (left slot)
SpO2: no sensor (right slot)
SpO2: patient disconnected (left slot)
SpO2: patient disconnected (right slot)
SpO2: light interference (left slot)
SpO2: light interference (right slot)
SpO2: poor signal (left slot)
SpO2: poor signal (right slot)
Large change in FiO2
Recruitment maneuver in process
Brightness test alarm
AERONEB disconnected
Cuff disconnection
Air + heliox supplies failed
Oxygenation adjustment OFF (no SpO2)
Ventilation adjustment OFF (no PetCO2)
No hemodynamic status available
High HLI
MV oszillation
FiO2 oszillation
PEEP oszillation
Cuff high pressure
FiO2 set to 100% due to low saturation
Check P ramp
If the monitor shows the alarm [Low
Low Technical
Low Technical alarms], the external device may
alarms
have the following alarms:

4-85
Mindray Patient Monitor Hamilton G5 (Block) Alarm

Priority Label Label


Volume Low for nebulizer
ASV: Check high pressure limit
APV: Check high pressure limit
Pressure low limit reached
Check %MinVol
Check Body Wt
ASV: Cannot meet target
Check PEEP/high pressure limit
Check PEEP/Pcontrol
Check PEEP/Psupport

Check trigger
Check %TI
Check pause
Check I:E
Check Vt
Check rate
Check peak flow
Check TI
Check Flow Pattern
Flow sensor calibration needed
Expiratory valve calibration needed
Apnea ventilation ended
Maximum leak compensation
Low ExpMinVol alarm off
CO2 sensor calibration needed
Check CO2 airway adapter
CO2 sensor disconnected

4-86
Mindray Patient Monitor Hamilton G5 (Block) Alarm

Priority Label Label


CO2 sensor over temperature
CO2 sensor faulty
External battery empty
Sensor simulation active
IRV
Cuff leak
IntelliCuff not found
Check VThigh limit
AERONEB module disconnected
Oxygenation adjustment OFF (no SpO2)
Ventilation adjustment OFF (no PetCO2)
Check CO2 sampling line
Check INTELLiVENT PEEP limit setting
Set low limit of ExpMinVol alarm
Recruitment in Progress
Oxygenation Controller on Limit
Ventilation Controller on Limit
SBT conditions fulfilled
SBT in progress

4.14.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL
CO2 mmHg, kPa, %

4-87
4.15 Hamilton G5 (Polling Protocol)
4.15.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
O2% % Oxygen concentration Yes
MV L/min Minute volume Yes
Inspiratory
I: E \ time:Expiratory time No
ratio
End-tidal carbon
EtCO2 %,mmHg,kPa Yes
dioxide
Arterial oxygen
SpO2 % saturation from pulse Yes
oximetry
PR bpm Pulse rate Yes
VTi ml Inspired tidal volume Yes
VTe ml Expiratory tidal volume Yes
ftot bpm Total respiratory rate Yes
Spontaneous
Yes
fspn bpm respiratory rate
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Texp sec Expiratory time No
cmH2O/L/s,hPa/L/s,
Ri Inspiratory resistance Yes
mbar/L/s
cmH2O/L/s,hPa/L/s,
Re Expiratory resistance Yes
mbar/L/s
ml/cmH2O,ml/hPa,
Cstat Static compliance Yes
ml/mbar

4-88
Label Units Description Trend, record, print
Insp.Flow L/min Inspiration flow No
Spontaneous
VTe spn ml Yes
expiratory tidal volume
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
Minimum airway
Pmin cnH2O,hPa,mbar No
pressure
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Exp. Flow L/min Expiratory flow No
Expiratory time
RCexp sec No
constant
Inspiratory time
RCinsp sec No
constant
WOB J/L Work of breathing Yes
cmH2O.s,hPa.s,mba
PTP Pressure time product No
r.s
Positive end-expiratory
PEEP cmH2O, hPa, mbar Yes
pressure
Setting Parameters
Phigh cmH2O,hPa,mba Upper pressure level No
fCMV bpm CMV frequency No
fSIMV bpm Frequency of SIMV No
VT ml Tidal volume No
Tpause % Apnea Time No
Inspiratory trigger No
P-Trigger cmH2O,hPa,mba
level(pressure trigger)
PEEP/CPAP cmH2O, hPa, mbar PEEP/CPAP No

4-89
Label Units Description Trend, record, print
Psupp cmH2O, hPa, mbar Pressure support level No
O2% % Oxygen concentration No
MV L/min Minute volume No
Inspiratory
I: E \ time:Expiratory time No
ratio
Peak Flow L/min Peak flow No
Inspiration termination No
Exp% %
level
Ramp msec Ramp No
IBW kg Ideal body weight No
Percentage of minute No
%MinVol %
volume to be delivered

4.15.2 Output Signals - Alarms


Mindray Patient Monitor Hamilton G5 (Polling) Alarm

Priority Label Label

Physiological alarms
Paw High High pressure
Apnea Apnea
High Loss of PEEP Loss of PEEP
MV Low Low minute volume
MV High High minute volume
Medium RR High High frequency
Technical alarms
Disconnection Disconnection Ventilator or,
High
ventilator side Disconnection on ventilator side

4-90
Mindray Patient Monitor Hamilton G5 (Polling) Alarm

Priority Label Label


Patient Disconnection Patient or,
Disconnected Disconnection on patient side
If the monitor shows the alarm [High
High Technical Technical alarms], the external device may
alarms have the following alarms: OXYGEN
CONCENTRATION.
If the monitor shows the alarm [Medium
Technical alarms], the external device may
Medium Technical
Medium have the following alarms:
alarms
Gas Supply
SPEZ ALARM
If the monitor shows the alarm [Low
Technical alarms], the external device may
Low Technical
Low have the following alarms:
alarms
Operator
General Alarm

4.16 Hamilton Galileo


4.16.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Oxygen
O2% % Yes
concentration
MV L/min Minute volume Yes
Inspiratory
I: E \ time:Expiratory time No
ratio

4-91
Label Units Description Trend, record, print
VTi ml Inspired tidal volume Yes
Expiratory tidal
VTe ml Yes
volume
ftot bpm Total respiratory rate Yes
Spontaneous
Yes
fspn bpm respiratory rate
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Texp sec Expiratory time No
cmH2O/L/s,hPa/L/s,m
Ri Inspiratory resistance Yes
bar/L/s
cmH2O/L/s,hPa/L/s,m
Re Expiratory resistance Yes
bar/L/s
ml/cmH2O,ml/hPa,ml
Cstat Static compliance Yes
/mbar
Insp.Flow L/min Inspiration flow No
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
Minimum airway
Pmin cmH2O, hPa, mbar No
pressure
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Exp. Flow L/min Expiratory flow No
Expiratory time
RCexp sec No
constant
Inspiratory time
RCinsp sec No
constant

4-92
Label Units Description Trend, record, print
WOB J/L Work of breathing Yes
Pressure time
PTP cmH2O.s,hPa.s,mbar.s No
product
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Setting Parameters
fCMV bpm CMV frequency No
fSIMV bpm Frequency of SIMV No
VT ml Tidal volume No
Tpause % Apnea Time No
Inspiratory trigger No
P-Trigger cmH2O, hPa, mbar level(pressure
trigger)
PEEP/CPAP cmH2O, hPa, mbar PEEP/CPAP No
Pressure support No
Psupp cmH2O, hPa, mbar
level
Oxygen No
O2% %
concentration
Phigh cmH2O, hPa, mbar Upper pressure level No
Inspiratory No
I: E \ time:Expiratory time
ratio
Peak Flow L/min Peak flow No
Inspiration No
Exp% %
termination level
Ramp msec Ramp No
IBW kg Ideal body weight No

4-93
Label Units Description Trend, record, print
Percentage of minute No
%MinVol % volume to be
delivered

4.16.2 Output Signals - Alarms


Mindray Patient Monitor Hamilton Galileo Alarm

Priority Label Label

Physiological alarms
Paw High High Pressue
Apnea Apnea
High Loss of PEEP Loss of PEEP
MV Low Low Min Vol
MV High High Min Vol
Medium RR High High Rate
Technical alarms
Disconnection
Disconnection Ventilator
ventilator side
Patient
Disconnection Patient
Disconnected
High
If the monitor shows the alarm [High
High Technical Technical alarms], the external device may
alarms have the alarm:
Oxygen Concentration
If the monitor shows the alarm [Medium
Medium Technical alarms], the external device may
Medium
Technical alarms have the alarm:
Gas Supply

4-94
Mindray Patient Monitor Hamilton Galileo Alarm

Priority Label Label


If the monitor shows the alarm [Low
Technical alarms], the external device may
Low Technical
Low have the following alarms:
alarms
Operator
General Alarm

4.17 Hamilton S1
4.17.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Oxygen
O2% % Yes
concentration
Positive
cmH2O, hPa,
PEEP end-expiratory Yes
mbar
pressure
cmH2O, hPa, Yes
Ppeak Peak pressure
mbar
cmH2O, hPa, Yes
Pplat Plateau pressure
mbar
cmH2O, hPa, Yes
Pmean Mean pressure
mbar
Expiratory tidal Yes
VTe ml
volume
Inspired tidal Yes
VTi ml
volume
VT/kg ml/kg TVe/IBW No

4-95
Label Units Description Trend, record, print
Spontaneous Yes
VTe spn ml expiratory tidal
volume
MV L/min Minute volume Yes
Spontaneous Yes
MVspn L/min breathed minute
volume
Total respiratory Yes
ftot bpm
rate
Spontaneous Yes
fspn bpm
respiratory rate
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
Inspiratory time:
I:E \ Expiratory time No
ratio
ml/cmH2O,ml/
Cstat Static compliance Yes
hPa,ml/mbar
WOB J/L Work of breathing Yes
Insp.Flow L/min Inspiration flow No
Exp. Flow L/min Expiratory flow No
VCO2 ml/min CO2 production No
PR bpm Pulse rate Yes
Texp sec Expiratory time No
cmH2O/L/s,hPa Inspiratory
Ri Yes
/L/s,mbar/L/s resistance
cmH2O/L/s,hPa Expiratory
Re Yes
/L/s,mbar/L/s resistance

4-96
Label Units Description Trend, record, print
Expiratory time
RCexp sec No
constant
Inspiratory time No
RCinsp sec
constant
cmH2O.s,hPa.s, Pressure time No
PTP
mbar.s product
cmH2O, hPa, Minimum airway No
Pmin
mbar pressure
cmH2O, hPa, 100 ms occlusion No
P0.1
mbar pressure
Intrinsic positive
cmH2O, hPa,
PEEPi end-expiratory No
mbar
pressure
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Arterial oxygen
SpO2 % saturation from Yes
pulse oximetry
Setting Parameters
VT ml Tidal volume No
f bpm Breath rate No
Inspiratory time: No
I:E \ Expiratory time
ratio
Base Flow L/min Base Flow No
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)

4-97
Label Units Description Trend, record, print
Inspiratory trigger No
cmH2O, hPa,
P-Trigger level (pressure
mbar
trigger)
cmH2O, hPa, Pressure support No
Psupp
mbar level
cmH2O, hPa, Pressure control No
Pinsp
mbar level of inspiration
Tpause % Apnea Time No
cmH2O, hPa, Upper pressure No
Phigh
mbar level
cmH2O, hPa, Lower pressure No
Plow
mbar level
Time for the upper No
Thigh sec
pressure level
Time for the lower No
Tlow sec
pressure level
Inspiration No
Exp% %
termination level
cmH2O, hPa, No
Plimit Pressure limit level
mbar
cmH2O, hPa, No
PEEP/CPAP PEEP/CPAP
mbar
Peak Flow L/min Peak flow No
IBW kg Ideal body weight No
Maximum No
Ti max sec
inspiration time
Inspiratory pause No
Tip sec
time
Ramp msec Ramp No

4-98
Label Units Description Trend, record, print
Percentage of No
%MinVol % minute volume to
be delivered
Tube ID mm Tube ID No
Oxygen No
O2% %
concentration
MV L/min Minute volume No

4.17.2 Output Signals—Alarms


Mindray Patient Monitor Hamilton S1(Block) Alarm

Priority Label Label


Physiological alarms
FiO2 High high Oxygen
FiO2 Low low Oxygen
PawLow Low pressure
MV High High minite volume
MV Low Low minite volume
Apnea Apnea
High
Paw High high pressure
Loss of PEEP Loss of PEEP
Apnea
Apnea ventilation
Ventilation
SpO2 Low SpO2 too low
SpO2 High SpO2 too high
RR Low Low frequency
RR High High frequency
Medium
EtCO2 High High PetCO2
EtCO2 Low Low PetCO2
4-99
Mindray Patient Monitor Hamilton S1(Block) Alarm

Priority Label Label


Technical alarms
Patient
Disconnection Patient
Disconnected
Air Supply
Air supply
Pressure Low
O2 Supply
Oxygen supply
Pressure Low
O2 cell disconnect O2 cell missing
O2 cell cal.
O2 cell cal. needed
Needed
Disconnection
Disconnection ventilator
ventilator side
Power Failure Loss of mains power
Check Flow
Check Flow Sensor type
High Sensors
No Gas Supply
All gas supplies failed
Pressure
If the monitor shows the alarm [High
Technical alarms], the external device may
have the following alarms:
Wrong Flow Sensor type
O2 cell defective
High Technical
Disconnection
Alarm
Low internal pressure
High pressure during sigh
Pressure not released
Exhalation obstructed
TF 5514: Check loudspeaker

4-100
Mindray Patient Monitor Hamilton S1(Block) Alarm

Priority Label Label


Internal battery empty
Ventilator unit connection lost
Check internal battery
O2 and air supply Oxygen and air supply
O2 and heliox
Oxygen and heliox supply
supply
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
High leak
Low tidal volume
High tidal volume
Turn the Flow Sensor
APV init Failed
Check P-ramp
Medium Internal battery Low
Medium Panel connection lost
Technical Alarm Heliox supply failed
SPO2:sensor error(left slot)
SPO2:sensor error(right slot)
SPO2:no sensor(left slot)
SPO2:no sensor(right slot)
SPO2:patient disconnected(left slot)
SPO2:patient disconnected(right slot)
SPO2:light interference(left slot)
SPO2:light interference(right slot)
SPO2:poor signal (left slot)
SPO2:poor signal (right slot)

4-101
Mindray Patient Monitor Hamilton S1(Block) Alarm

Priority Label Label


Large change in FiO2
Recruitment meaneuver in progress
Brightness test alarm
AERONEB disconnected
Cuff disconnection
Air +heliox supplies failed
Oxygenation adjustment OFF(no SpO2)
Ventilation adjustment OFF(no PetO2)
No hemodynamic staus avaliable
High HLI
MV oszillation
FiO2 oszillation
PEEP oszillation
Cuff high pressure
FiO2 set to 100% due to low saturation
If the monitor shows the alarm [Low
Technical alarms], the external device may
have the following alarms:
Volume too low for nebulizerASV: Check high
pressure limit
APV: Check high pressure limit
Low Technical
Low pressure low limit reached
Alarm
Check %MinVol
Check Body Wt
ASV:Cannot meet target
Check PEEP/high pressure limit
Check PEEP/Pcontrol
Check PEEP/Psupport

4-102
Mindray Patient Monitor Hamilton S1(Block) Alarm

Priority Label Label


Check trigger
Check %TI
Check pause
Check I:E
Check Vt
Check rate
Check peak flow
Check TI
Check FlowPattern
Flow Sensor calibration needed
Expiratory valve calibration needed
Apnea ventilation ended
Maximum leak compensation
Low ExpMinVol alarm off
CO2 Sensor calibration needed
Check CO2 airway adapter
CO2 sensor disconnected
CO2 sensor over temperature
CO2 sensor faulty
External battery empty
Sensor simulation active
IRV
Cuff leak
IntelliCuff not found
Check VThigh limit
AERONEB modle disconnected
Oxygenation adjustment OFF(no SpO2)
Ventilation adjustment OFF(no PetO2)

4-103
Mindray Patient Monitor Hamilton S1(Block) Alarm

Priority Label Label


Check CO2 sampling line
Check INTELLIVENT PEEP limit setting
Set low limit for ExpMinVol alram
Recuritment in progress
Oxygenation controller on limit
Vetilation controller on limit
SBT conditions fulfilled
SBT in progress

4.17.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL
CO2 mmHg, kPa, %

4-104
4.18 HuL Leoni Plus
4.18.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Positive
PEEP cmH2O, hPa, mbar end-expiratory No
pressure
Fractional
FiO2 %.kPa,mmHg concentration of Yes
O2 in inspired gas
Fractional
FiO2% % concentration of Yes
O2 in inspired gas
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume
Inspiratory tidal
VTi ml Yes
volume
MV L/min Minute volume Yes
Total respiratory
ftot bpm Yes
rate
Spontaneous
fspn bpm Yes
respiratory rate
ml/cmH2O,ml/hPa, Static lung
Rstat Yes
ml/mbar resistance
ml/cmH2O,ml/hPa, Dynamic
Cdyn Yes
ml/mbar compliance

4-105
Label Units Description Trend, record, print
Arterial oxygen
SpO2 % saturation from Yes
pulse oximetry
Setting Parameters
Oxygen No
O2% %
concentration
VT ml Tidal volume No
Breath rate for No
fapnea bpm
apnea ventilation
f bpm Breath rate No
Inspiratory time: No
I:E \ Expiratory time
ratio
Inspiratory No
F-Trigger L/min trigger level (flow
trigger)
Trigger % Trigger No
Time of No
Tinsp sec
inspiration
Pressure control No
Pinsp cmH2O, hPa, mbar level of
inspiration
Maximum airway No
Pmax cmH2O, hPa, mbar
pressure
Flow L/min Flow No
TApnea sec Apnea time No
Pressure support No
Psupp cmH2O, hPa, mbar
level

4-106
Label Units Description Trend, record, print
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure

4.18.2 Output Signals—Alarms


Mindray Patient Monitor HuL Leoni Plus Alarm

Priority Label Label


Physiological alarms
Apnea Apnea
FiO2 High O2 high
FiO2 Low P: O2 too low
MV High MV high
MV Low MV low
PEEP Not Achieved PEEP pressure not reached
High PEEP High PEEP too high
Pinsp Not Achieved Set pressure not reachable
Ppeak High Ppeak high
Ppeak Low Ppeak low
SpO2 High SpO2 too high
SpO2 Low SpO2 too low
VT Not Achieved Volume not reached
RR High Frequency too high
Medium VTe High VTe too high
VTe Low Vte too low
Technical alarms
Airway Obstructed? Tube occluded
High If the monitor shows the alarm [Check Folw
Check Flow Sensors
Sensors], the external device may have the
4-107
Mindray Patient Monitor HuL Leoni Plus Alarm

Priority Label Label


following alarms:
Flow sensor broken
Flowsensor fail
Calibrate Flowsensor
Flowsensor contaminated.
No O2 Pressure O2 supply failed. Freshgas is Air
Patient
Disconnection
Disconnected
If the monitor shows the alarm [High
Technical alarms], the external device may
have the following alarms:
Air supply
Air supply failed. Freshgas is O2
Battery empty. Mechanical ventilation
stopped.
Battery empty. Supply voltage Low
Excess pressure Exsp-Tube
High Technical Excess pressure Insp-Tube
High
Alarm O2 and Air supply failed. Dosing fresh gas
stopped.
O2 supply
Deviation pressure sensors
Technical Failure
Current consumption High
Technical Failure
3.3V supply on NetDCU High
3.3V supply on NetDCU Low
5V supply on NetDCU High

4-108
Mindray Patient Monitor HuL Leoni Plus Alarm

Priority Label Label


5V supply on NetDCU Low
12V supply on NetDCU High
12V supply on NetDCU Low
24V supply on NetDCU High
24V supply on NetDCU Low
Versions not compatible
Failsafe
Controllerboard EEPROM checksum failed
Patient safe: Reboot the device
Sensor Fail Patient Pressure
Driving gas blender failed.
Checksum error
Encoder without function
FiO2 Sensor
Oxy Measurement Fail
Disconnected
Patient Circuit Leak Leak too high
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
Battery almost empty
Medium Check O2 concentration
Medium Technical Low Perfusion
Alarm O2 control aborted
Oximetry cable failure
Oximetry cable not connected
SpO2: communication error
SpO2: No adhesive sensor connected
SpO2: No cable connected

4-109
Mindray Patient Monitor HuL Leoni Plus Alarm

Priority Label Label


SpO2: Sensor failure
SpO2: Sensor not connected
SpO2: Sensor off patient
Battery Fail
Battery not connected
Broken loudspeaker. Audible alarming not
possible.
Batteries deep discharged. Please change.
O2 Calibration failure
Broken microphone. No checking of audible
alarming
Battery in Use Device running on batteries
If the monitor shows the alarm [Low
Technical alarms], the external device may
have the following alarms:
Low Low Technical
Observe battery runtime
Alarm
Fan Fail
Controllerboard EEPROM not write
protected

4-110
4.19 MAQUET SERVO-i/SERVO-s
4.19.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Oxygen
O2% % Yes
concentration
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume
Inspired tidal
VTi ml Yes
volume
Spontaneous
MVspn L/min breathed minute Yes
volume
Expiratory minute
MVe L/min Yes
volume
Inspiratory mimute
MVi L/min Yes
volume
Total respiratory
ftot bpm Yes
rate
Spontaneous
fspn bpm Yes
respiratory rate
Inspiratory
I:E \ time:Expiratory No
time ratio

4-111
Label Units Description Trend, record, print
Leak Comp % Leak compensation No
ml/cmH2O,ml/hPa,
Cstat Static compliance Yes
ml/mbar
ml/cmH2O,ml/hPa, Dynamic
Cdyn Yes
ml/mbar compliance
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
WOB J/L Work of breathing Yes
Exp. Flow L/min Expiratory flow No
cmH2O/L/s,hPa/L/s, Inspiratory
Ri Yes
mbar/L/s resistance
cmH2O/L/s,hPa/L/s, Expiratory
Re Yes
mbar/L/s resistance
oxygen supply
PO2 kPa
pressure No
Pair kPa air supply pressure No
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
PEEPtot cmH2O, hPa, mbar Total PEEP No
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
End-tidal carbon
EtCO2% % Yes
dioxide
VCO2 ml/min CO2 production No
CO2 tidal No
VTCO2 ml
elimination

4-112
Label Units Description Trend, record, print
Duty cycle or ratio No
of inspiration time
to total breathing
TI/TTOT \ cycle time (only
during
spontaneous
breathing)
Setting Parameters
VT ml Tidal volume No
MV L/min Minute volume No
fCMV bpm CMV frequency No
fSIMV bpm Frequency of SIMV No
Inspiratory No
I:E \ time:Expiratory
time ratio
Inspiratory trigger No
F-Trigger \
level (flow trigger)
Inspiratory trigger No
P-Trigger cmH2O, hPa, mbar level (pressure
trigger)
Tinsp sec Time of inspiration No
Tpause % Apnea Time No
Rise Time% % rise time% No
Upper pressure No
Phigh cmH2O, hPa, mbar
level
Time for the upper No
Thigh sec
pressure level
Time at PEEP level No
Tpeep sec
in Bi-Vent

4-113
Label Units Description Trend, record, print
PC above PEEP cmH2O, hPa, mbar PC above PEEP No
PS above PEEP cmH2O, hPa, mbar PS above PEEP No
PEEP/CPAP cmH2O, hPa, mbar PEEP/CPAP No
Plimit cmH2O, hPa, mbar Pressure limit level No
Breath rate for No
fapnea bpm
apnea ventilation
Tpause sec Pause Time % No
Trise sec Rise time No
TApnea sec Apnea time No
VTapnea ml Apnea tidal volume No
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure
Oxygen No
O2% %
concentration

4.19.2 Output Signals - Alarms


Mindray Patient Monitor MAQUET Alarm

Priority Label Label

Physiological alarms
Airway pressure alarm Upper pressure limit
Paw High
exceeded
MV High Exp.Minute volume too high
High MV Low Exp.Minute volume too low
Apnea Apnea alarm
FiO2 High O2 conc.too high
FiO2 Low O2 conc.too low

4-114
Mindray Patient Monitor MAQUET Alarm

Priority Label Label


PEEP High PEEP High
PEEP Low PEEP Low
EtCO2 High EtCO2 conc.too high
EtCO2 Low EtCO2 conc.too low
Medium
RR Low Breath frequency Low
RR High Breath frequency High
Technical alarms
No Gas Supply
Gas supply alarm
Pressure
O2 cell disconnect O2 cell disconnect
If the monitor shows the alarm [High
Technical alarms], the external device may
have the following alarms:
Breathing system uP Module error
Inspiratory control uP Module error
Monitoring System uP Module error
Battery alarm
High Power Failure
High Technical Mains Failure
alarms O2 potentiometer error
CMV potentiometer error
Range Switch error
Mode Switch error
Barometer error
High continuous pressure
Overrange
Computer Interface Emulator hardware
error

4-115
Mindray Patient Monitor MAQUET Alarm

Priority Label Label


NIV,Leakage out of range
NIV,Time in waiting position exceeds 2 min
regulation pressure limited
NIV,No patient effort detected
Unreliable Edi signal
Check catheterposition/Edi invalid
No consistent patient effort
Check catheter position/RR and HR coupling
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
Panel Interface uP Module error
Exp.flow &CO2 linearization uP Module error
Medium Technical Alarm buff
Medium
alarms CI Battery Voltage
Pneumatic-Edi out of synch
Edi activity low
No Edi signal detected
Unsuccessful manual gas change alarm
Check Heliox adapter
Low Check tubing Check tubing
If the monitor shows the alarm [Low
Low Technical Technical alarms], the external device may
alarms have the following alarms:
Leakage fraction too high

4-116
4.19.3 Output Signals - WaveForms (for BeneVision N Series
Monitors)
WaveForm Unit
Paw cmH2O, hPa, mbar
Flow L/min
Volume ml
CO2 %,kPa,mmHg

4.19.4 Output Signals -Respiratory Loops (for BeneVision N Series


Monitors)
Loops Unit
PV Loop cmH2O/ml, hPa/ml, mbar/ml
FV Loop L/min/ml
cmH2O/L/min, hPa/L/min,
PF Loop mbar/L/min

4-117
4.20 MAQUET SERVO-U/N
NOTE

 Only the Benevision N Sereis monitor supports integration with the


MAQUET SERVO-N ventilator.

4.20.1 Output Signals—Parameters


Label Units Description Trend, record, print
Monitoring Parameters
Oxygen
O2% % Yes
concentration
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Expiratory tidal
VTe ml Yes
volume
Inspired tidal
VTi ml Yes
volume
VT/kg ml/kg TVe/IBW No
Expiratory minute
MVe L/min Yes
volume
Inspiratory mimute
MVi L/min Yes
volume
Spontaneous
MVspn L/min breathed minute Yes
volume

4-118
Label Units Description Trend, record, print
Total respiratory
ftot bpm Yes
rate
Spontaneous
fspn bpm Yes
respiratory rate
Inspiratory
I:E \ time:Expiratory No
time ratio
Duty cycle or ratio
of inspiration time
to total breathing
TI/TTOT \ cycle time (only No
during
spontaneous
breathing)
Leak Comp % Leak compensation No
ml/cmH2O,ml/hPa,
Cstat Static compliance Yes
ml/mbar
ml/cmH2O,ml/hPa, Dynamic
Cdyn Yes
ml/mbar compliance
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
WOB J/L Work of breathing Yes
Exp. Flow L/min Expiratory flow No
cmH2O/L/s,hPa/L/s, Inspiratory
Ri Yes
mbar/L/s resistance
cmH2O/L/s,hPa/L/s, Expiratory
Re Yes
mbar/L/s resistance
oxygen supply
PO2 kPa No
pressure

4-119
Label Units Description Trend, record, print
Pair kPa air supply pressure No
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
PEEPtot cmH2O, hPa, mbar Total PEEP No
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
VCO2 ml/min CO2 production No
CO2 tidal
VTCO2 ml No
elimination
End-tidal carbon
EtCO2% % Yes
dioxide
Setting Parameters
VT ml Tidal volume No
VTapnea ml Apnea tidal volume No
MV L/min Minute volume No
Breath rate for No
fapnea bpm
apnea ventilation
f bpm Breath rate No
fSIMV bpm Frequency of SIMV No
Inspiratory time: No
I:E \ Expiratory time
ratio
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Inspiratory trigger No
P-Trigger cmH2O, hPa, mbar level (pressure
trigger)
Plimit cmH2O, hPa, mbar Pressure limit level No
Tinsp sec Time of inspiration No

4-120
Label Units Description Trend, record, print
Tpause % Apnea Time No
Trise sec Rise time No
Rise Time% % rise time% No
Upper pressure No
Phigh cmH2O, hPa, mbar
level
Time for the upper No
Thigh sec
pressure level
Time at PEEP level No
Tpeep sec
in Bi-Vent
PC above PEEP cmH2O, hPa, mbar PC above PEEP No
PS above PEEP cmH2O, hPa, mbar PS above PEEP No
PEEP/CPAP cmH2O, hPa, mbar PEEP/CPAP No
Flow L/min Flow No
TApnea sec Apnea time No
Oxygen No
O2% %
concentration
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure

4.20.2 Output Signals—Alarms


Mindray Patient Monitor MAQUET SERVO-U/N Alarm

Priority Label Label


Physiological alarms
Apnea Apnea
FiO2 High O2 concentration high
High
FiO2 Low O2 concentration low
High Paw Sustained Airway pressure continuously high
4-121
Mindray Patient Monitor MAQUET SERVO-U/N Alarm

Priority Label Label


MV High Expiratory minute volume high
MV Low Expiratory minute volume low
Paw High Airway pressure high
PEEP High PEEP high
PEEP Low PEEP low
EtCO2 High EtCO2 high
EtCO2 Low EtCO2 low
Medium
RR High Respiratory rate High
RR Low Respiratory rate low
Technical alarms
If the monitor shows the alarm [High
Technical alarms], the external device
may have the following alarms:
Battery alarm
High Technical Alarm
Overrange alarm
Patient disconnected > 1 min
High Time in waiting position > 2 min
Leakage too high
No Gas Supply
Gas supply alarm
Pressure
O2 cell disconnect O2 cell/sensor failure

Patient Disconnected Patient circuit disconnected

If the monitor shows the alarm [Medium


Technical alarms], the external device
Medium Technical
Medium may have the following alarms:
Alarm
CPAP high
CPAP low

4-122
Mindray Patient Monitor MAQUET SERVO-U/N Alarm

Priority Label Label


Edi signal interference from ECG
Edi signal invalid
Expiratory cassette disconnected
Leakage High
No consistent patient effort
No patient effort
The nebulizer cannot be run on one
battery
Battery in Use Battery operation
If the monitor shows the alarm [Low
Technical alarms], the external device
may have the following alarms:
Low Expiratory cassette replaced
Low Technical Alarm
Inconsistent Edi signal
Low Edi signal
No Edi signal detected
Volume delivery restricted

4.20.3 Output Signals—Waveforms (for BeneVision N Series


Monitors)
WaveForm Unit
Paw cmH2O, hPa, mbar
Flow L/min
Volume ml
CO2 %,kPa,mmHg

4-123
4.20.4 Output Signals—Respiratory Loops (for BeneVision N
Series Monitors)
Loops Unit
PV Loop cmH2O/ml, hPa/ml, mbar/ml
FV Loop L/min/ml
cmH2O/L/min, hPa/L/min,
PF Loop mbar/L/min

4.21 Mindray SV300/350/600/650/800/850


NOTE

 Only SV300 with software version 04.00.00 or later can be connected to


the BeneLink module.

4.21.1 Output Signals - Parameters


Label Units Description Trend, record, print
Monitoring Parameters
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Oxygen
O2% % Yes
concentration
Pmean cmH2O, hPa, mbar Mean pressure Yes
Inspired tidal
VTi ml Yes
volume

4-124
Label Units Description Trend, record, print
Expiratory tidal
VTe ml Yes
volume
VT/kg ml/kg TVe/IBW No
Spontaneous
VTe spn ml expiratory tidal Yes
volume
Expiratory minute
MVe L/min Yes
volume
Spontaneous
MVspn L/min breathed minute Yes
volume
Total respiratory
ftot bpm Yes
rate
Mandatory
fmand bpm breathing Yes
frequency
Spontaneous
fspn bpm Yes
respiratory rate
Inspiratory time:
I:E \ Expiratory time No
ratio
Leakage minute
MV Leak L/min No
volume
Fractional
FiO2 %,Kpa,mmHg concentration of O2 Yes
in inspired gas
ml/cmH2O,ml/hPa,
Cstat Static compliance Yes
ml/mbar

4-125
Label Units Description Trend, record, print
ml/cmH2O,ml/hPa, Dynamic
Cdyn Yes
ml/mbar compliance
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
Imposed work of
WOBimp J/min Yes
breathing
cmH2O/L/s,hPa/L/s, Inspiratory
Ri Yes
mbar/L/s resistance
cmH2O/L/s,hPa/L/s, Expiratory
Re Yes
mbar/L/s resistance
Expiratory time
RCexp sec No
constant
Negative
NIF cmH2O, hPa, mbar No
inspiratory force
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
End-tidal carbon
EtCO2 %,Kpa,mmHg Yes
dioxide
VCO2 ml/min CO2 production No
PR bpm Pulse rate Yes
Arterial oxygen
SpO2 % saturation from Yes
pulse oximetry
Insp.Flow L/min Inspiration flow No
Fraction of inspired
FiCO2 %,Kpa,mmHg Yes
carbon dioxide

4-126
Label Units Description Trend, record, print
Setting Parameters
Oxygen No
O2% %
concentration
VT ml Tidal volume No
VTApnea ml Apnea tidal volume No
Breath rate for No
fapnea bpm
apnea ventilation
fSIMV bpm Frequency of SIMV No
f bpm Breath rate No
fsigh bpm Sigh rate No
△int. PEEP cmH2O, hPa, mbar Intermittent PEEP No
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Inspiratory trigger No
P-Trigger cmH2O, hPa, mbar level (pressure
trigger)
Pressure support No
Psupp cmH2O, hPa, mbar
level
Tinsp sec Time of inspiration No
Pressure control No
Pinsp cmH2O, hPa, mbar
level of inspiration
Papnea cmH2O, hPa, mbar Apnea pressure No
Tpause % Apnea Time No
Trise sec Rise time No
Upper pressure No
Phigh cmH2O, hPa, mbar
level
Lower pressure No
Plow cmH2O, hPa, mbar
level

4-127
Label Units Description Trend, record, print
Time for the upper No
Thigh sec
pressure level
Time for the lower No
Tlow sec
pressure level
Inspiration No
Exp% %
termination level
Flow L/min Flow No
IBW kg Ideal body weight No
Tube ID mm Tube ID No
Automatic Tube No
ATC %
Compensation
Positive No
PEEP cmH2O, hPa, mbar end-expiratory
pressure
Plimit cmH2O, hPa, mbar Pressure limit level

4.21.2 Output Signals - Alarms


Mindray Patient Monitor Mindray SV Alarm

Priority Label Label

Physiological alarms
Paw High Paw Too High
Paw Low Paw Too Low
MV High MV Too High
High MV Low MV Too Low
FiO2 High FiO2 Too High
FiO2 Low FiO2 Too Low
Apnea Apnea Ventilation

4-128
Mindray Patient Monitor Mindray SV Alarm

Priority Label Label


Ventilation
PEEP High PEEP Too High
Apnea Apnea
Pinsp Not
Achieved Pinsp Not Achieved
PEEP Low PEEP Too Low
SpO2 High SpO2 Too High
SpO2 Low SpO2 Too Low
PR High PR Too High
PR Low PR Too Low
VT Not Achieved Tidal Volume Not Achieved
High Circuit O2 O2% Too High
Low Circuit O2 O2% Too Low
CO2 Apnea Apnea CO2
VTe High TVe Too High
VTe Low TVe Too Low
Medium EtCO2 High EtCO2 Too High
EtCO2 Low EtCO2 Too Low
FiCO2 High FiCO2 Too High
RR High ftot Too High
Low Plimit Reached Pressure Limited
Technical alarms
Air Supply
Air Supply Pressure Low
Pressure Low
High O2 Supply O2 Supply Pressure Low\
Pressure Low O2 Supply Failure
No Gas Supply No Gas Supply Pressure

4-129
Mindray Patient Monitor Mindray SV Alarm

Priority Label Label


Pressure
Airway
Obstructed? Airway Obstructed?
Tube
Disconnected? Tube Disconnected?
Sustained Airway
Pressure Sustained Airway Pressure
Neo Flow Sensor
Error Neo Flow Sensor Failure
Insp gas
temperature
High Insp. Gas Temp Too High
If the monitor shows the alarm [High
Technical alarms], the external device may
have the following alarms:
RT Clock Not Exist
Keyboard Comm Stop
Keyboard Selftest Error
Ventilator Reset Error
High Technical Battery Exhaust! Syst. Down!
alarms Low Battery Voltage
Ctrl Module Comm Error
Ctrl Module Comm Stop
Ctrl Module Selftest Error
Protection Module Comm Error
Protection Module Selftest Error
Protection Module Comm Stop
Pressure Sensor Failure

4-130
Mindray Patient Monitor Mindray SV Alarm

Priority Label Label


Insp. Limb Failure
Ventilator Selftest Error
Please perform pressure calibration.
Please perform flow calibration.
CO2 Comm Stop
CO2 Comm Error
CO2 Hardware Error
CO2 Init Error
CO2 Selftest Error
Safety Valve Failure
Exp. Valve Failure
Device Failure 01/02/03/04/05
/06/07/08/09/12/14/15/16/17/18
19/20/21
Insp. Temp. Sensor Failure
Internal Power Error
Insp Valve Off
Turbine Failed
Turbine Temperature too high
Inspiratory Module Comm Stop
Expiratory Module Comm Stop
Wrong Flow Sensor Type
Inspiratory Module Self Test Error
Expiratory Module Self Test Error
Pressure Sensor Zero Error
Battery 1 Temperature Abnormal. Charging
Failed
Battery 2 Temperature Abnormal. Charging

4-131
Mindray Patient Monitor Mindray SV Alarm

Priority Label Label


Failed
Battery 1 Charging Error
Battery 2 Charging Error
Battery 1 Aged
Battery 2 Aged
Battery 1 Comm Err
Battery 2 Comm Err
Battery 1 Error
Battery 2 Error
The machine may be powered off due to high
battery temperature.
Power Board Comm Stop
Protective Module Initialization Error
SpO2 Comm Stop
SpO2 Desat
No Pulse
The machine will be powered off soon.
Connect to an external power supply.
Battery Undetected
Monitoring Module Initialization Error
Machine Internal Voltage Error
Memory Error (Machine Error 02)
CO2 Module Zero Error
Neo. Flow Sensor Reversed
No Neo.Flow Sensor
Wrong Neo.Flow Sensor Type
Neo.Flow Sensor Failure
Neo.Flow Sensor Overrange

4-132
Mindray Patient Monitor Mindray SV Alarm

Priority Label Label


Blower Temperature High
Blower Failure 03/04
Flow Sensor Type Error
Blower Battery Failure 1/2/3/4/5
FiO2 Sensor
FiO2 Sensor Disconnected
Disconnected
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
Storage Card Error
Key Error
Fan Failure
Internal Temperature High
Exp. Flow Sensor Failure
O2 Sensor Failure
O2 Insp. Limb Failure
Medium
Medium Nebulizer Valve Failure
Technical alarms 3-way Valve Failure
Please Replace CO2 Sensor
Turbine Rotational Speed Abnormal
Turbine Temp Sensor Failure
Technical Error 01
Ambient Atmospheric Pressure Sensor Failure
HEPA Filter Pressure Sensor Failure
High battery temperature. Connect to the AC
power supply.
Battery too low. Connect to the AC power
supply.

4-133
Mindray Patient Monitor Mindray SV Alarm

Priority Label Label


Replace SpO2 Sensor
SpO2 Init Err
Inspiratory Branch Airway Obstructed
Keyboard Comm Stop
Keyboard Self Test Error
Buzzer Error
CO2 Hardware Error
CO2 Init Err
CO2 Self Test Error
Inspiratory Temp Sensor Failure
Neo Flow Sensor Disabled
Blower Failure 05
Blower Fan Failure
Airway Leak? Airway Leak?
Battery in Use Battery in Use
CO2 No Water
trap CO2 No Watertrap
Tinsp Long Tinsp Too Long
If the monitor shows the alarm [Low
Technical alarms], the external device may
Low have the following alarms:
RT Clock Need Reset
Low Technical IP Address Conflict
alarms Loading Default Config. Failed
Restoring Last Config. Failed
Insp. Hold Interrupted
Exp. Hold Interrupted
Heating Module Failure

4-134
Mindray Patient Monitor Mindray SV Alarm

Priority Label Label


Please calibrate O2 sensor.
Buzzer Failure
Volume Limited
CO2 Sensor High Temp
CO2 Sensor Low Temp
CO2 High Airway Pressure
CO2 Low Airway Pressure
CO2 High Barometric
CO2 Low Barometric
CO2 Sampleline Occluded
CO2 System Error
EtCO2 Overrange
FiCO2 Overrange
CO2 Zero Failed
Please calibrate flow sensor.
CO2 No Sensor
Replace HEPA Filter
Sign Cycle Pressure Limit
SpO2 Sensor Off
No SpO2 Sensor
SpO2 Too Much Light
SpO2 Non-Pulsatile
SpO2 Overrange
PR Overrange
Key Error
AMV: Cannot Meet Target

4-135
4.21.3 Output Signals - Waveforms (For BeneVision N series
Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL
CO2 mmHg, kPa, %
SpO2 /

4.21.4 Output Signals – Respiratory Loops (For BeneVision N


series Monitors)
Label Units
PV Loop cmH2O/ml, hPa/ml, mbar/ml
FV Loop L/min/ml
PF Loop cmH2O/L/min, hPa/L/min, mbar/L/min

4.22 Mindray SynoVent E3/E5


4.22.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
cmH2O, hPa, Positive end-expiratory
PEEP Yes
mbar pressure
Inspiratory time:
I:E \ Yes
Expiratory time ratio
PR bpm Pulse rate Yes
Arterial oxygen
SpO2 % saturation from pulse Yes
oximetry

4-136
Label Units Description Trend, record, print
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
cmH2O, hPa,
Pplat Plateau pressure Yes
mbar
O2% % Oxygen concentration Yes
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
VTe ml Expiratory tidal volume Yes
Spontaneous
VTe spn ml Yes
expiratory tidal volume
Expiratory minute
MVe L/min Yes
volume
Spontaneous breathed
MVspn L/min Yes
minute volume
ftot bpm Total respiratory rate Yes
Mandatory breathing
fmand bpm Yes
frequency
Spontaneous
fspn bpm Yes
respiratory rate
Leakage minute
MV Leak L/min No
volume
Fractional
FiO2 %,kPa,mmHg concentration of O2 in Yes
inspired gas
ml/cmH2O,ml/hP
Cstat Static compliance Yes
a,ml/mbar
ml/cmH2O,ml/hP
Cdyn Dynamic compliance Yes
a,ml/mbar

4-137
Label Units Description Trend, record, print
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
cmH2O/L/s,hPa/L
Ri Inspiratory resistance Yes
/s,mbar/L/s
cmH2O/L/s,hPa/L
Re Expiratory resistance Yes
/s,mbar/L/s
cmH2O, hPa, Negative inspiratory
NIF No
mbar force
End-tidal carbon
EtCO2 %,kPa,mmHg Yes
dioxide
Fraction of inspired
FiCO2 %,kPa,mmHg Yes
carbon dioxide
VTi ml Inspired tidal volume Yes
Expiratory time
RCexp sec No
constant
VT/kg ml/kg TVe/IBW No
VCO2 ml/min CO2 production No
Insp.Flow L/min Inspiration flow No
cmH2O, hPa, 100 ms occlusion
P0.1 No
mbar pressure
Intrinsic positive
cmH2O, hPa,
PEEPi end-expiratory No
mbar
pressure
Imposed work of
WOBimp J/min Yes
breathing
Setting Parameters
cmH2O, hPa, Positive end-expiratory Yes
PEEP
mbar pressure
O2% % Oxygen concentration Yes

4-138
Label Units Description Trend, record, print
VT ml Tidal volume No
Breath rate for apnea No
fapnea bpm
ventilation
fSIMV bpm Frequency of SIMV No
cmH2O, hPa, No
△int. PEEP Intermittent PEEP
mbar
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
cmH2O, hPa, Inspiratory trigger No
P-Trigger
mbar level (pressure trigger)
cmH2O, hPa, No
Psupp Pressure support level
mbar
Tinsp sec Time of inspiration No
cmH2O, hPa, Pressure control level No
Pinsp
mbar of inspiration
cmH2O, hPa, No
Papnea Apnea pressure
mbar
Trise sec Rise time No
cmH2O, hPa, No
Phigh Upper pressure level
mbar
cmH2O, hPa, No
Plow Lower pressure level
mbar
Time for the upper No
Thigh sec
pressure level
Time for the lower No
Tlow sec
pressure level
Inspiration termination No
Exp% %
level
Flow L/min Flow No

4-139
Label Units Description Trend, record, print
IBW kg Ideal body weight No
f bpm Breath rate No
cmH2O, hPa, No
Plimit Pressure limit level
mbar
Tube ID mm Tube ID No
Automatic Tube No
ATC %
Compensation
VTApnea ml Apnea tidal volume No
Tpause % Apnea Time No
fsigh bpm Sigh rate No

4.22.2 Output Signals - Alarms


Mindray Patient Monitor Mindray E3/E5 Alarm

Priority Label Label

Physiological alarms
Paw High Paw Too High
Paw Low Paw Too Low
MV High MV Too High
MV Low MV Too Low
Apnea Apnea
Apnea Ventilation Apnea Ventilation
High
FiO2 High FiO2 Too High
FiO2 Low FiO2 Too Low
PR High PR Too High
PR Low PR Too Low
SPO2 High SpO2 Too High
SPO2 Low SpO2 Too Low

4-140
Mindray Patient Monitor Mindray E3/E5 Alarm

Priority Label Label


PEEP Low PEEP Too Low
PEEP High PEEP Too High
Pinsp Not Achieved Pinsp Not Achieved
CO2 Apnea Apnea CO2
High Circuit O2 O2% Too High
Low Circuit O2 O2% Too Low
VTNot Achieved Tidal Volume Not Achieved
VTe High TVe Too High
RR High ftot Too High
EtCO2 High EtCO2 Too High
Medium
EtCO2 Low EtCO2 Too Low
VTe Low TVe Too Low
FiCO2 High FiCO2 Too High
Low Plimit Reached Pressure Limited
Technical alarms
Air Supply Pressure
Air Supply Pressure Low
Low
O2 Supply Pressure O2 Supply Pressure Low
Low O2 Supply Failure
No Gas Supply
Pressure No Gas Supply Pressure
High
Airway Obstructed? Airway Obstructed?
Tube
Disconnected? Tube Disconnected?
Sustained Airway
Pressure Sustained Airway Pressure
Neo. Flow Sensor Neo Flow Sensor Failure

4-141
Mindray Patient Monitor Mindray E3/E5 Alarm

Priority Label Label


Failure
Insp gas
temperature High Insp. Gas Temp Too High
If the monitor shows the alarm [High
Technical alarms], the external device may
have the following alarms:
RT Clock Not Exist
Keyboard Comm Stop
Keyboard Selftest Error
Ventilator Reset Error
Battery Exhaust! Syst. Down!
Low Battery Voltage
Ctrl Module Comm Error
Ctrl Module Comm Stop
Ctrl Module Selftest Error
High Technical
Protection Module Comm Error
alarms
Protection Module Comm Stop
Protection Module Selftest Err
Pressure Sensor Failure
Air Insp. Limb Failure
Ventilator Reset Error
Please perform pressure cal.
Please perform flow cal.
CO2 Comm Stop
CO2 Comm Error
CO2 Hardware Error
CO2 Init Error
CO2 Selftest Error

4-142
Mindray Patient Monitor Mindray E3/E5 Alarm

Priority Label Label


Safety Valve Failure
Exp. Valve Failure
Device Failure 01/02/03/04/05
/06/07/08/09/12/14/15/16/17/18
19/20/21
Insp. Temp. Sensor Failure
Internal Power Error
Insp Valve Off
Turbine Failed
Turbine Temperature too high
Inspiratory Module Comm Stop
Expiratory Module Comm Stop
Wrong Flow Sensor Type
Inspiratory Module Self Test Error
Expiratory Module Self Test Error
Pressure Sensor Zero Error
Battery 1 Temperature Abnormal. Charging
Failed
Battery 2 Temperature Abnormal. Charging
Failed
Battery 1 Charging Error
Battery 2 Charging Error
Battery 1 Aged
Battery 2 Aged
Battery 1 Comm Err
Battery 2 Comm Err
Battery 1 Error
Battery 2 Error

4-143
Mindray Patient Monitor Mindray E3/E5 Alarm

Priority Label Label


The machine may be powered off due to
high battery temperature.
Power Board Comm Stop
Protective Module Initialization Error
SpO2 Comm Stop
SpO2 Desat
No Pulse
The machine will be powered off soon.
Connect to an external power supply.
Battery Undetected
Monitoring Module Initialization Error
Machine Internal Voltage Error
Memory Error (Machine Error 02)
CO2 Module Zero Error
Neo. Flow Sensor Reversed
No Neo.Flow Sensor
Wrong Neo.Flow Sensor Type
Neo.Flow Sensor Failure
Neo.Flow Sensor Overrange
Blower Temperature High
Blower Failure 03/04/05
Flow Sensor Type Error
Blower Battery Failure 1/2/3/4/5
FiO2 Sensor
FiO2 Sensor Disconnected
Disconnected
Medium If the monitor shows the alarm [Medium
Medium Technical
Technical alarms], the external device may
alarms
have the following alarms:

4-144
Mindray Patient Monitor Mindray E3/E5 Alarm

Priority Label Label


Storage Card Err
Key Error
Battery Undetected
Fan Failure
Internal Temperature High
Exp. Flow Sensor Failure
O2 Sensor Failure
O2 Insp. Limb Failure
Nebulizer Valve Failure
3-way Valve Failure
Replace CO2 Sensor
Turbine Rotational Speed Abnormal
Turbine Temp Sensor Failure
Ambient Atmospheric Pressure Sensor
Failure
Technical Error 01
HEPA Filter Pressure Sensor Failure
High battery temperature. Connect to the
AC power supply.
Battery too low. Connect to the AC power
supply.
Replace SpO2 Sensor
SpO2 Init Err
Inspiratory Branch Airway Obstructed
Keyboard Comm Stop
Keyboard Self Test Error
Buzzer Error
CO2 Hardware Error

4-145
Mindray Patient Monitor Mindray E3/E5 Alarm

Priority Label Label


CO2 Init Err
CO2 Self Test Error
Inspiratory Temp Sensor Failure
Neo Flow Sensor Disabled
Blower Failure 05
Blower Fan Failure
Airway Leak? Airway Leak?
Battery in Use Battery in Use
Tinsp Long Tinsp Too Long
CO2 No Water trap CO2 No Watertrap
If the monitor shows the alarm [Low
Technical alarms], the external device may
have the following alarms:
RT Clock Need Reset
IP Address Conflict
Loading Default Config. Failed
Restoring Last Config. Failed
Low
Insp. Hold Interrupted
Low Technical Exp. Hold Interrupted
alarms Heating Module Failure
Please calibrate O2 sensor.
Buzzer Failure
Volume Limited
CO2 Sensor High Temp
CO2 Sensor Low Temp
CO2 High Airway Pressure
CO2 Low Airway Pressure
CO2 High Barometric

4-146
Mindray Patient Monitor Mindray E3/E5 Alarm

Priority Label Label


CO2 Low Barometric
CO2 Sampleline Occluded
CO2 System Error
EtCO2 Overrange
FiCO2 Overrange
CO2 Zero Failed
Calibrate Flow Sensor.
CO2 No Sensor
Replace HEPA Filter
Sign Cycle Pressure Limit
SpO2 Sensor Off
No SpO2 Sensor
SpO2 Too Much Light
SpO2 Non-Pulsatile
SpO2 Overrange
PR Overrange
Key Error
AMV: Cannot Meet Target

4.23 Newport E360


4.23.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Positive
cmH2O, hPa,
PEEP end-expiratory Yes
mbar
pressure

4-147
Label Units Description Trend, record, print
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
Expiratory tidal
VTe ml Yes
volume
Inspired tidal
VTi ml Yes
volume
Spontaneous
MVspn L/min breathed minute Yes
volume
Expiratory minute
MVe L/min Yes
volume
Inspiratory
MVi L/min Yes
mimute volume
Total respiratory
ftot bpm Yes
rate
Spontaneous
fspn bpm Yes
respiratory rate
Inspiratory time:
I:E \ Expiratory time No
ratio
Fractional
FiO2 %,kPa,mmHg concentration of Yes
O2 in inspired gas
Fractional
FiO2% % concentration of Yes
O2 in inspired gas
Insp.Flow L/min Inspiration flow No

4-148
Label Units Description Trend, record, print
Exp. Flow L/min Expiratory flow No
cmH2O, hPa,
PEEPtot Total PEEP No
mbar
Setting Parameters
VT ml Tidal volume No
f bpm Breath rate No
Inspiratory trigger
F-Trigger L/min No
level (flow trigger)
Inspiratory trigger
cmH2O, hPa,
P-Trigger level (pressure No
mbar
trigger)
cmH2O, hPa, Pressure support
Psupp No
mbar level
cmH2O, hPa, Pressure limit
Plimit No
mbar level
Time of
Tinsp sec No
inspiration
cmH2O, hPa,
PEEP/CPAP PEEP/CPAP No
mbar

4.23.2 Output Signals - Alarms


Mindray Patient Monitor Newport E360 Alarm

Priority Label Label

Physiological alarms
Paw High High Paw
Paw Low Low Paw
High
MV High High Exhale Mv
MV Low Low Exhale Mv
4-149
Mindray Patient Monitor Newport E360 Alarm

Priority Label Label


Apnea Apnea Alarm
FiO2 High FiO2 High
FiO2 Low FiO2 Low
VT Not Achieved Volume Target Not Met
High Baseline High Baseline
Low Baseline Low Baseline
Sustained Hbline Sustained Hbline
Medium RR High Resp. Rate Alarm
Technical alarms
Air Supply Pressure
Air Supply Loss
Low
O2 Supply Pressure
Low O2 Supply Loss
Check Flow Sensors Flow Sensor Error
Power Failure Power Failure
Tinsp too Short Insp Time too Short
If the monitor shows the alarm [High
Technical alarms], the external device may
High have the following alarms:
Device Alert
No O2 Power-Up
High Technical Control EEPROM Failure
alarms Low Battery
Transducer Error
Mon Internal System Failed
Ctrol Internal System Failed
Air Servo Valve Leak
O2 Servo Valve Leak
4-150
Mindray Patient Monitor Newport E360 Alarm

Priority Label Label


Control RAM Failed
Control ROM Failed
Control CPU Failed
Monitor RAM Failed
Monitor ROM Failed
Monitor CPU Failed
Dual RAM Failed
Monitor Tasks Failed
Control Processor Failed
Control Tasks Failed
Monitor Processor Failed
Fan Failure
Air Flow Sensor EEPROM Failure
O2 Flow Sensor EEPROM Failure
O2 and air supply Air & O2 Supply Loss
FiO2 Sensor
Disconnected FiO2 Sensor Disconnected
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
Medium
Flow Sensor Cal Failed
Medium Technical
FiO2 Sensor Bad
alarms
O2 Sensor Cal Failed
External Battery
Check Flow Sensor Board
NO TEST
Low Technical If the monitor shows the alarm [Low
Low
alarms Technical alarms], the external device may

4-151
Mindray Patient Monitor Newport E360 Alarm

Priority Label Label


have the following alarms:
I:E Ratio Inverse violation
Plimit<Pbase
Psupport+Pbase>60cmH2O
Pbase>Low Paw
Tidal Volume Out of Range
Flow Out of Range
Ti Out of Range
Rate Out of Range
Psupport Out of Range
Plimit Out of Range
PEEP/CPAP Out of Range
Flow Trigger Out of Range
CPM Blinking
EXH. VALVE CAL. Failed: Prox < 1
EXH. VALVE CAL. Failed: Prox > 0.5
EXH. VALVE CAL. Failed: Prox Low
EXH. VALVE CAL. Failed: Flow < 1
Leak Test Failed
Tinsp too Long Insp Time too Long
Battery in Use Battery in Use

4-152
4.24 Philips Respironics V60 (SNDA Protocol)
4.24.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
ftot bpm Total breath rate Yes
ml Expiratory tidal
VTe Yes
volume
MV L/min Minute volume Yes
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
Setting Parameters
f bpm Breath rate No
Oxygen
O2% % Yes
concentration
Positive
cmH2O, hPa,
PEEP end-expiratory No
mbar
pressure
cmH2O, hPa, Pressure support
Psupp No
mbar level
cmH2O, hPa, Pressure control
Pinsp No
mbar level of inspiration
Tinsp sec Time of inspiration No

4.24.2 Output Signals - Alarms


Mindray Patient Monitor Philips V60(SNDA) Alarm

Priority Label Label

Physiological alarms
Paw Low High inhalation pressure
High
Paw High Low inhalation pressure
4-153
Mindray Patient Monitor Philips V60(SNDA) Alarm

Priority Label Label


Low inspiratory pressure
Low exhaled minute volume
MV Low
Low minute volume
RR High High respiratory rate
Medium Low exhaled mandatory spontaneous tidal
VTe Low
volume
Technical alarms
O2 Supply Pressure
Low oxygen supply pressure
Low
Patient Disconnect Occlusion or I-time too long
High If the monitor shows the alarm [High
High Technical Technical alarms], the external device may
alarms have the following alarms:
Low battery

4.25 Philips Respironics V60 (VRPT Protocol)


4.25.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
ftot bpm Total breath rate Yes
Expiratory tidal
VTe ml Yes
volume
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
Expiratory minute
MVe L/min Yes
volume

4-154
Label Units Description Trend, record, print
Leakage minute
MVLEAK L/min No
volume
Oxygen
TI/TTOT \ No
concentration
Setting Parameters
f bpm Breath rate No
Oxygen
O2% % No
concentration
Positive
cmH2O, hPa,
PEEP end-expiratory No
mbar
pressure
cmH2O, hPa, Pressure control
Pinsp No
mbar level of inspiration
Tinsp sec Time of inspiration No

4.25.2 Output Signals - Alarms


Mindray Patient Monitor Philips V60(VRPT) Alarm

Priority Label Label

Physiological alarms
Low inhalation pressure,
Paw Low
Low inspiratory pressure
Paw High High inhalation pressure
High
Low exhaled minute volume
MV Low
Low minute volume
PEEP Low Low PEEP
RR High High respiratory rate
Medium RR Low Apnea
VTe Low Low exhaled mandatory spontaneous tidal
4-155
Mindray Patient Monitor Philips V60(VRPT) Alarm

Priority Label Label


volume
Technical alarms
O2 Supply
Low O2 supply
Pressure Low
If the monitor shows the alarm [High
Technical alarms], the external device may
have the following alarms:
High Technical
Low internal battery alarm status
High alarms
Primary alarm failure
Air source fault alarm status
O2 valve stuck closed alarm status
Airway
Occlusion
Obstructed?
EXP-VALVE? Safety valve
If the monitor shows the alarm [Medium
Medium Technical Technical alarms], the external device may
Medium
alarms have the alarm:
High enclosure temperature.
Airway Leak? High leak
If the monitor shows the alarm [Low
Low Low Technical Technical alarms], the external device may
alarms have the alarm:
Nonvolatile memory failure.

4-156
4.26 Puritan Bennett PB840/PB980 (SNDF Protocol)
4.26.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
Oxygen
O2% % Yes
concentration
Positive
PEEP cmH2O, hPa, mbar end-expiratory Yes
pressure
Ppeak cmH2O, hPa, mbar Peak pressure Yes
Pplat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
ml Expiratory tidal
VTe Yes
volume
VTi ml Inspired tidal volume Yes
ml Spontaneous
VTe spn expiratory tidal Yes
volume
L/min Spontaneous
MVspn breathed minute Yes
volume
L/min Expiratory minute
MVe Yes
volume
ftot bpm Total respiratory rate Yes
Inspiratory time:
I:E \ No
Expiratory time ratio
Leakage minute
MVLEAK L/min No
volume
Leak Comp % Leak compensation No

4-157
Label Units Description Trend, record, print
cmH2O/L/s,hPa/L/s,
Rstat Static lung resistance Yes
mbar/L/s
cmH2O/L/s,hPa/L/s, Dynamic lung
Rdyn Yes
mbar/L/s resistance
ml/cmH2O,ml/hPa,
Cstat Static compliance Yes
ml/mbar
ml/cmH2O,ml/hPa,
Cdyn Dynamic compliance Yes
ml/mbar
Rapid shallow
RSBI 1/(min.L) Yes
breathing index
WOB J/L Work of breathing Yes
Negative inspiratory
NIF cmH2O, hPa, mbar No
force
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure
Intrinsic positive
PEEPi cmH2O, hPa, mbar end-expiratory No
pressure
PEEPtot cmH2O, hPa, mbar Total PEEP No
Setting Parameters
VT ml Tidal volume No
VTapnea ml Apnea tidal volume No
Breath rate for apnea
fapnea bpm No
ventilation
f bpm Breath rate No
Peak Flow L/min Peak flow No
TApnea sec Apnea interval No
IBW kg Ideal body weight No

4-158
Label Units Description Trend, record, print
Maximum inspiration
Ti max sec No
time
Tube ID mm Tube ID No
Phigh cmH2O, hPa, mbar Upper pressure level No
Plow cmH2O, hPa, mbar Lower pressure level No
Time for the upper
Thigh sec No
pressure level
Time for the lower
Tlow sec No
pressure level
Base Flow L/min Base Flow No
Inspiratory time:
I:E \ No
Expiratory time ratio
Inspiratory trigger
F-Trigger L/min No
level (flow trigger)
Inspiratory trigger
P-Trigger cmH2O, hPa, mbar level (pressure No
trigger)
Pressure support
Psupp cmH2O, hPa, mbar No
level
Tplat sec Plateau time No
Rise Time% % Rise time% No
PEEP/CPAP cmH2O, hPa, mbar PEEP/CPAP No

4.26.2 Output Signals - Alarms


Mindray Patient Monitor Puritan Bennett 840/980 Alarm

Priority Label Label

Physiological alarms
High MV High High Exhaled minute Volume
4-159
Mindray Patient Monitor Puritan Bennett 840/980 Alarm

Priority Label Label


MV Low Low Exhaled Minute Volume
Apnea Apnea
FiO2 Low Low O2 %
Ppeak High High Ppeak
Ppeak Low Low Ppeak
VTe High High Exhaled Tidal Volume
RR High High ftot
Medium
VTe Low Low Exhaled Mandatory Tidal Volume Alarm
EtO2 High High O2 Percent
Technical alarms
Air Supply
No Air Supply
Pressure Low
O2 Supply
Pressure Low No O2 Supply
Airway
Obstructed? Severe Occlusion
Circuit
Disconnect Circuit Disconnect
Power Failure Loss of Power
High
If the monitor shows the alarm [High Technical
alarms], the external device may have the
following alarms:
Compressor Inoperative
High Technical
Compliance Limited VT
alarms
Procedure Error
PAV Startup Too Long
PAV R& C Not Assessed
Volume Not Delivered(VC+)
4-160
Mindray Patient Monitor Puritan Bennett 840/980 Alarm

Priority Label Label


Volume Not Delivered(VS)
If the monitor shows the alarm [Low Technical
alarms], the external device may have the
Low Technical
following alarms: Inoperative Battery
Low alarms
Low Battery
Ac Power Loss
Tinsp too Long Inspiration Too Long

4.27 Puritan Bennett PB840/PB980 (SNDA Protocol)


4.27.1 Output Signals - Parameters
Label Units Description Trend, record, print
Monitoring Parameters
cmH2O, hPa,
Ppeak Peak pressure Yes
mbar
cmH2O, hPa,
Pplat Plateau pressure Yes
mbar
cmH2O, hPa,
Pmean Mean pressure Yes
mbar
ml Expiratory tidal
VTe Yes
volume
L/min Spontaneous
MVspn breathed minute Yes
volume
L/min Expiratory minute
MVe Yes
volume
bpm Total respiratory
ftot Yes
rate

4-161
Label Units Description Trend, record, print
\ Inspiratory time:
I:E Expiratory time No
ratio
cmH2O/L/s,hPa/L/ Static lung
Rstat Yes
s,mbar/L/s resistance
cmH2O/L/s,hPa/L/ Dynamic lung
Rdyn Yes
s,mbar/L/s resistance
ml/cmH2O,ml/hP
Cstat Static compliance Yes
a,ml/mbar
ml/cmH2O,ml/hP Dynamic
Cdyn Yes
a,ml/mbar compliance
cmH2O, hPa, Negative
NIF No
mbar inspiratory force
PR bpm Pulse rate Yes
% Arterial oxygen
SpO2 saturation from Yes
pulse oximetry
Setting Parameters
VT ml Tidal volume No
VTapnea ml Apnea tidal volume No
bpm Breath rate for
fapnea No
apnea ventilation
f bpm Breath rate No
Inspiratory time:
I:E \ Expiratory time No
ratio
Base Flow L/min Base Flow No
L/min Inspiratory trigger
F-Trigger No
level (flow trigger)

4-162
Label Units Description Trend, record, print
cmH2O, hPa, Inspiratory trigger
P-Trigger mbar level (pressure No
trigger)
cmH2O, hPa, Pressure support
Psupp No
mbar level
Tplat sec Plateau time No
cmH2O, hPa,
PEEP/CPAP PEEP/CPAP No
mbar
Peak Flow L/min Peak flow No
TApnea sec Apnea interval No
cmH2O, hPa, Upper pressure
Phigh No
mbar level
cmH2O, hPa, Lower pressure
Plow No
mbar level
fsigh bpm Sigh Rate No
VTsigh ml Sigh Tidal Volume No

4.27.2 Output Signals - Alarms


Mindray Patient Monitor Puritan Bennett 840/980 Alarm

Priority Label Label

Physiological alarms
SpO2 Low Low Saturation Alarm
SpO2 High High Saturation Alarm
MV Low Low Exhaled minute Volume
Apnea Apnea
PR Low Low Pulse Rate Alarm
High PR High High Pulse Rate Alarm

4-163
Mindray Patient Monitor Puritan Bennett 840/980 Alarm

Priority Label Label


Ppeak Low Low Ppeak
Ppeak High High Ppeak
VTe Low Low Exhaled Mandatory Tidal Volume Alarm
Medium
RR High High ftot
Technical alarms
Air Supply
No Air Supply
Pressure Low
O2 Supply
High
Pressure Low No O2 Supply
EXP VALVE Exhalation Valve Leak
Circuit Disconnect Circuit Disconnect
If the monitor shows the alarm [Low
Low Technical
Low Technical alarms], the external device may
alarms
have the following alarms: Low Battery

4.28 ResMed VSIII


4.28.1 Output Signals—Parameters
Label Units Description Trend, record, print
Monitoring Parameters
ftot bpm Total respiratory rate Yes
L/min Leakage minute
MVLEAK No
volume
Setting Parameters
Positive No
cmH2O, hPa,
PEEP end-expiratory
mbar
pressure
VT ml Tidal volume No
4-164
Label Units Description Trend, record, print
f bpm Breath rate No
Inspiratory trigger No
F-Trigger L/min
level (flow trigger)
Tinsp sec Time of inspiration No
cmH2O, hPa, Pressure control level No
Pinsp
mbar of inspiration
Inspiration
Exp% % No
termination level
Maximum inspiration
Ti max sec No
time

4.28.2 Output Signals—Alarms


Mindray Patient Monitor Resmed VSIII Alarm

Priority Label Label

Physiological alarms
Paw High High pressure
High Paw Low Low Pressure
Apnea Apnea alarm
Medium RR High High Frequency alarm
Technical alarms
Power Failure Main disconnect
Tube Disconnected? Turbine alarm
O2 cell cal.Needed FiO2 Cell Defective
High O2 cell disconnect Fi O2 Cell Missing
If the monitor shows the alarm [High
High Technical
Technical alarms], the external device may
Alarm
have the following alarms:

4-165
Mindray Patient Monitor Resmed VSIII Alarm

Priority Label Label

No power supply
Without external DC power supply
Technical Alarm
Internal battery temperature out of range
Patient Circuit Leak Patient circuit disconnected
If the monitor shows the alarm [Medium
Technical alarms], the external device may
Medium Medium Technical have the following alarms:
Alarm With power supply
With mains power
No power supply

4.29 Salvia Elisa 800/600/500/300


NOTE

 Only the Benevision N Sereis monitor supports integration with the


Salvia Elisa 500/300 ventilator.

4.29.1 Output Signals - Parameters


Label Units Description Trend, record, print
Monitoring Parameters
Total respiratory
Vte ml Yes
rate
Leakage minute
Mve L/min No
volume
Total respiratory
ftot bpm Yes
rate

4-166
Label Units Description Trend, record, print
Oxygen
O2% % Yes
concentration
Ppeak cmH2O, hPa, mbar Peak pressure Yes
pPlat cmH2O, hPa, mbar Plateau pressure Yes
Pmean cmH2O, hPa, mbar Mean pressure Yes
Minimum airway
Pmin cmH2O, hPa, mbar No
pressure
Spontaneous
MVspn L/min breathed minute Yes
volume
Spontaneous
fspn bpm Yes
respiratory rate
Intrinsic positive
Peepi cmH2O, hPa, mbar end-expiratory No
pressure
ml/cmH2O,ml/hPa,ml Dynamic
Cdyn Yes
/mbar compliance
cmH2O/L/s,hPa/L/s,m Inspiratory
Ri Yes
bar/L/s resistance
Inspired tidal
Vti ml Yes
volume
Inspiratory
MVi L/min Yes
mimute volume
Extrinsic positive
PEEPe cmH2O, hPa, mbar end-expiratory No
pressure
PEEPtot cmH2O, hPa, mbar Total PEEP No
100 ms occlusion
P0.1 cmH2O, hPa, mbar No
pressure

4-167
Label Units Description Trend, record, print
Fractional
FiO2% % concentration of Yes
O2 in inspired gas
End-tidal carbon
EtCO2% % Yes
dioxide
Inspiratory time:
I:E \ Expiratory time No
ratio
Leak
Leak Comp % No
compensation
oxygen supply
PO2 kPa No
pressure
Work of
WOB J/L Yes
breathing
Spontaneous
Vte Spn ml expiratory tidal Yes
volume
Vt/Kg ml/kg TVe/IBW No
Time of
Tinsp Sec No
inspiration
Texp Sec Expiratory time No
Rapid shallow
RSBI 1/min·L Yes
breathing index
air supply
Pair kPa
pressure No
Expiratory time
RCexp Sec No
constant
Vtrap ml Trapped volume No

4-168
Label Units Description Trend, record, print
Respiratory
RQ \ No
quotient
O2 Flow L/min O2 flow No
Air Flow L/min Air flow No
Setting Parameters
Positive No
VT Ml end-expiratory
pressure
f bpm Tidal volume No
I:E \ Breath rate No
Inspiratory No
PEEP cmH2O, hPa, mbar trigger level (flow
trigger)
Time of No
Plimit cmH2O, hPa, mbar
inspiration
Pressure control No
Pinsp cmH2O, hPa, mbar level of
inspiration
Inspiratory time:
Psupp cmH2O, hPa, mbar Expiratory time No
ratio
Pressure limit
Pmax cmH2O, hPa, mbar No
level
Pressure support
IBW kg No
level
Maximum airway
Ramp ms No
pressure
Ideal body
ASB Rmap Sec No
weight

4-169
Label Units Description Trend, record, print
Trise Sec Ramp No
F-Trigger L/min ASB ramp No
P-Trigger cmH2O, hPa, mbar Rise time No
Inspiratory
trigger
Tinsp sec No
level (pressure
trigger)
Tpause % Apnea Time No
Assisted
PASB cmH2O, hPa, mbar spontaneous
breathing No
Base Flow L/min Base Flow No
Oxygen
O2% % No
concentration

4.29.2 Output Signals - Alarms


Mindray Patient Monitor Draeger Evita 4/ Evita2 dura Alarm

Priority Label Label

Physiological alarms
Paw High High Paw
Paw Low Low Paw
High Paw
High Paw Sustained
Sustained
High SubAtmospheric
Sub-Atmospheric Paw
Paw
Pmax Reached Pmax Reached
Pinspired Not
Pinspired Not Achieved
Achieved
4-170
Mindray Patient Monitor Draeger Evita 4/ Evita2 dura Alarm

Priority Label Label


PEEP Not Achieved PEEP Not Achieved
No Peep CPAP No Pressure Cntrl/PEEP
MV Too Low Low Mve
MV Too High High Mve
Vt Not Achieved Vt Not Achieved
Volume Apnea 2
Volume Apnea > 2 min
min
FiO2 Too Low Low FiO2
FiO2 Too High High FiO2
VTe Too Low Low Vte
VTe Too High High Vte
Volume Apnea Volume Apnea
RR Too Low Low RR
Medium
RR Too High High RR
EtCo2 Too Low Low etCO2
EtCo2 Too High High etCO2
FiCo2 Too High High FiCO2
Low Plimit Reached Plimit Reached
Technical alarms
Check flow Sensors Check Flow Sensor
No Fresh Gas Flow No Fresh Gas Flow
No VO2 High FiN2O No VO2, High FiN2O
No O2 Pressure No O2 Pressure
High
Negative airway
Sustained Paw
pressure
High Technical If the monitor shows the alarm [High
alarms Technical alarms], the external device may

4-171
Mindray Patient Monitor Draeger Evita 4/ Evita2 dura Alarm

Priority Label Label


have the following alarms:
Circuit Occluded
Inspiration Stopped
No Battery
Low Battery Charge
Low Battery (No AC)
Battery deeply discharged
Standby ON
High O2 Supply Pressure
High Air Supply Pressure
If the monitor shows the alarm [Medium
Technical alarms], the external device may
have the following alarms:
High Circuit O2
Low Circuit O2
No O2 Cell Sensor
No Air Pressure
Running On Battery
Medium Technical Battery Failure
Medium
alarms Battery Charger Failure
Fan Failure
Heater Failure
Power Supply Failure
Display Failure
Breathing System Failure
Primary Audio Failure
Backup Audio Failure
Nebulizer Failure

4-172
Mindray Patient Monitor Draeger Evita 4/ Evita2 dura Alarm

Priority Label Label


Medium Patient Circuit Leak Patient Circuit Leak
12 Hour Test 12 Hour Test
If the monitor shows the alarm [Low
Technical alarms], the external device may
have the following alarms:

ASR on
ARC On
Low Low Technical
Replace O2 Cell
alarms
O2 Cell Calibration Error
No Inspiratory Flow Sensor
No Expiratory Flow Sensor
Patient Circuit Leak Silenced
Memory (EEPROM) Failure
Service Calibrations Due

4.29.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
Paw cmH2O, mbar, hpa
Flow L/min
Volume mL
CO2 mmHg, kPa, %

4-173
FOR YOUR NOTES

4-174
5 Integrating the tcGas Monitoring Device
5.1 Biometer TCM40
5.1.1 Output Signals - Parameters
Label Units Description Trend, record, print
Transcutaneous carbon
tcpCO2 mmHg,kPa dioxide partial Yes
pressures
Transcutaneous
tcpO2 mmHg,kPa oxygen partial Yes
pressures
Power mW Sensor heating power Yes
Tsensor ℃,℉ Sensor temperature Yes
PR bpm Pulse rate Yes
Arterial oxygen
SpO2 % saturation from pulse Yes
oximetry

5.1.2 Output Signals - Alarms


Mindray Patient Monitor TCM40 Alarm

Priority Label Label

Physiological alarms
High SpO2 Alarm SpO2 Alarm high/low
tcpCO2 Alarm tcpCO2 Alarm high/low
Medium tcpO2 Alarm tcpO2 Alarm high/low
PR Alarm Pulse Alarm high/low
Technical alarms
tcGas Battery Depleted Battery level critically low
High
TCM Temperature High Monitor temperature too high

5-1
Mindray Patient Monitor TCM40 Alarm

Priority Label Label


Medium tcGas Low Battery Battery level low
Low TCM Alert TCM Alert

5.2 Biometer TCM4/TCM 5/TCM CombiM


5.2.1 Output Signals - Parameters
Label Units Description Trend, record, print
mmHg,kPa Transcutaneous carbon
tcpCO2 Yes
dioxide partial pressures
mmHg,kPa Transcutaneous oxygen Yes
tcpO2
partial pressures
Power mW Sensor heating power Yes
Tsensor ℃,℉ Sensor temperature Yes

5.2.2 Output Signals - Alarms


Mindray Patient Monitor Tcm Alarm

Priority Label Label

Physiological alarms
tcpCO2 Alarm pCO2 alarm high/low
Medium
tcpO2 Alarm pO2 alarm high/low
Technical alarms
tcGas Battery Depleted Battery level critically low
High
TCM Temperature High Monitor temperature too high
Medium tcGas Low Battery Battery level low
Low TCM Alert TCM Alert

5-2
5.3 Biometer TCM Tosca
5.3.1 Output Signals - Parameters
Label Units Description Trend, record, print
mmHg,kPa Transcutaneous carbon
tcpCO2 Yes
dioxide partial pressures
Power mW Sensor heating power Yes
Tsensor ℃,℉ Sensor temperature Yes
PR bpm Pulse rate Yes
% Arterial oxygen saturation Yes
SpO2
from pulse oximetry

5.3.2 Output Signals - Alarms


Mindray Patient Monitor Tcm Tosca Alarm

Priority Label Label

Physiological alarms
SpO2 Alarm SpO2 Alarm high/low
Medium tcpCO2 Alarm pCO2 Alarm high/low
PR Alarm Pulse Alarm high/low
Technical alarms
tcGas Battery Depleted Battery level critically low
High
TCM Temperature High Monitor temperature too high
Medium tcGas Low Battery Battery level low
Low TCM Alert TCM Alert

5-3
5.4 SenTec Digital Monitor
5.4.1 Output Signals - Parameters
Label Units Description Trend, record, print
mmHg,kPa Transcutaneous carbon
tcpCO2 Yes
dioxide partial pressures
mmHg,kPa Transcutaneous oxygen partial Yes
tcpO2
pressures
Power mW Sensor heating power Yes
Tsensor ℃,℉ Sensor temperature Yes
PR bpm Pulse rate Yes
% Arterial oxygen saturation Yes
SpO2
from pulse oximetry

5.4.2 Output Signals - Alarms


Mindray Patient Monitor SenTec Digital Monitor Alarm

Priority Label Label

Physiological alarms
High SpO2 Alarm SpO2 Alarm high/low
tcpCO2 Alarm tcpCO2 Alarm high/low
Medium tcpO2 Alarm tcpO2 Alarm high/low
PR Alarm Pulse Alarm high/low
Technical alarms
High tcGas Battery Depleted SDMS Battery Critical
Medium tcGas Low Battery SDMS Battery Low

5-4
6 Integrating Infusion System (for BeneVision N
Series Monitors)

6.1 B.Braun Infusion Space/Perfusor Space Pumps


6.1.1 Output Signals - Parameters
Label Units Description Trend, record, print
Rate ml/hr Flow Rate No
V-infused ml Infused Volume No
T-Remain h:m Time Remain No
T-infused h:m Time Infused No
Drug Name \ Drug Name No

6.1.2 Output Signals – Technical Alarms


Mindray Patient Monitor B.Braun Alarm

Priority Label Label


Low Battery Battery alarm
Syringe Empty Syringe empty
Occlusion Pressure too high
Volume infused
end of infusion Infusion time expired x hour limit is reached.
High Only for PCA operation
Door Open Syringe holder open
Drop alarm, summary of all drop alarms
Drop alarm
Drop alarm: no drops
Air alarm, summary of all air alarms
Air bubble
Air alarm: Air bubble too large > 0.3 ml
6-1
Mindray Patient Monitor B.Braun Alarm

Priority Label Label


Air alarm: Air rate exceeded > 4 ml/h
Air alarm: sensor test faulty
Syringe Near Empty syringe empty pre-alarm

6.2 MedimaNet Server Infusing Monitoring and Archiving


Network System
6.2.1 Output Signals - Parameters
Label Units Description Trend, record, print
Rate ml/hr Flow rate Yes
T-infused h:m Acculated infusion time Yes
DrugName \ Drug name No
Status \ Pump connection status No

6.3 Fresenius Agilia Injectomat Infusion Pumps


6.3.1 Output Signals - Parameters
Label Unit Description Trend, record, print
Rate ml/hr Flow Rate No
V-infused ml Infused Volume No
Drug Name \ Drug Name No
VTBI ml Left Volume No

6-2
6.3.2 Output Signals – Technical Alarms
Mindray Patient Monitor Fresenius Injectomat Alarm

Priority Label Label


Low Battery Battery/Pre-Alarm Battery
end of limit volume
end of infusion end of volume/time
end of infusion
occlusion
Occlusion
rise of pressure
High syringe clamp
Mechanical Abnormal flange
plunger head
Syringe disengagement disengagment
Drop too less infusion flow rate low
Drop too many infusion flow rate high
Syringe Empty syringe empty
Pre-Occlusion occlusion(pre-alarm)
Medium
System Abnormal power disconnection
end of limit volume(pre-alarm)
Low Infusion Near Done end of volume/time(pre-alarm)
end of infusion(pre-alarm)

6-3
6.4 Fresenius Agilia Volumat Infusion Pumps
6.4.1 Output Signals - Parameters
Label Units Description Trend, record, print
Rate ml/h Flow Rate No
V-infused ml Infused Volume No
Drug Name \ Drug Name No
VTBI ml Left Volume No

6.4.2 Output Signals – Technical Alarms


Mindray Patient Monitor Fresenius Volumat Alarm

Priority Label Label


battery alarm
Low Battery
battery pre-alarm
end of infusion alarm
end of infusion
end of secondary infusion alarm
downstream occlusion alarm
upstream occlusion alarm
Occlusion occlusion detected by drop sensor
occlusion check system failed
pressure increase
High
Mechanical Abnormal clamp missing
absence of drop sensor
Sensor alarm
presence of drop sensor
under flow alarm detected by drop
Drop too less sensor
low flow rate or low dose
over flow alarm detected by drop
Drop too many sensor
high flow rate or high dose

6-4
Mindray Patient Monitor Fresenius Volumat Alarm

Priority Label Label


Door Open door opened alarm
Drop alarm uncontrolled flow alarm
Air alarm (air presence or air detection
Air bubble failure)
Air alarm (air volume exceeded limit)
Pre-Occlusion downstream occlusion pre-alarm
set not detected by upstream pressure
sensor
Medium set not detected by downstream
System Abnormal
pressure sensor
set not installed correctly
main disconnection
Low Infusion Near Done end of infusion pre-alarm

6.5 Mindray BeneFusion DS5 Series Infusion Supervision


System (For BeneVision N series Monitors)
6.5.1 Output Signals - Parameters
Label Units Description Trend, record, print
Rate ml/h Flow Rate No
V-infused ml Infused Volume No
Drug Name \ Drug name No
VTBI ml Volume To Be Infused No
T-Infused h:m:s Infused Time No
T-Remain h:m:s Left Time No
Status \ Infused status No
Mode \ Infused Mode No

6-5
6.5.2 Output Signals – Technical Alarms
Mindray Patient Monitor Mindray Pump Alarm

Priority Label Label

battery Exhausted
Battery Low
Battery Low
end of infusion Over
Occlusion
Occlusion
Top Occlusion
Dislocated
System Disengagement
Not Install
High
Empty
Syringe Empty
Empty Bottle
Syringe Near Empty Near Empty
Door Open Door Open
Drop alarm Drip UnExptional
Air bubble Air Bubble
KVO Done KVO Done
System Failure
Med System Abnormal
System Exceptional
Low Infusion Near Done Near over

6-6
7 Integrating Other Devices

7.1 MAQUET IABP


7.1.1 Output Signals - Parameters
Label Units Description Trend, record, print
HR bpm Heart Rate Yes
Art-S mmHg,kPa Systolic Blood Pressure Yes
Art-D mmHg,kPa Diastolic Blood Pressure Yes
Art-M mmHg,kPa Mean Blood Pressure Yes
Aug. mmHg,kPa Augmented Blood Yes

7.1.2 Output Signals - Alarms


Mindray Patient Monitor Maquet IABP Alarm

Priority Label Label

Physiological alarms
Augmentation Below
Medium Augmentation below limit set
limit set
Tchnical alarms
NO Trigger NO Trigger
NO Pressure Trigger NO Pressure Trigger
Check Pacer Timing Check Pacer Timing
High Trigger Interference Trigger Interference
Rapid gas loss Rapid gas loss
Leak in IAB circuit Leak in IAB circuit
IAB Disconnected IAB Disconnected
Blood detected Blood detected

7-1
Mindray Patient Monitor Maquet IABP Alarm

Priority Label Label


Check IAB Catheter Check IAB catheter
High driver pressure High drive pressure
Low Vacuum Low Vacuum
System Failure SYSTEM FAILURE
Electrical Test Failure ELECTRICAL TEST FAILURE
Autofill Failure Autofill Failure
Autofill Failure - No
helium Autofill Failure - NoHelium
Safety disk test failure Safety Disk test failure
No patient status
available No Patient Status Available
Autofill Required Autofill Required
Poor Signal Persist Poor Signals Persist
ECG Detected ECG Detected
Irregular Trigger Irregular Trigger
Low helium Low Helium
Low Battery Low Battery
Low Battery(EXT) Low Battery (EXT)
Maintenance required MAINTENANCE REQUIRED
Prolonged Time in
Prolonged Time in Standby
Medium Standby
Irregular pressure trigger Irregular Pressure Trigger
No pressure source
No Pressure Source Available
available
Poor Signal Quality Poor Signal Quality
Unable to update timing Unable To Update Timing
Unable to Calibrate
Unable to Calibrate Sensor IAB
Sensor IAB
7-2
Mindray Patient Monitor Maquet IABP Alarm

Priority Label Label


Sensor IAB Calibration
Sensor IAB Calibration Expired
Expired
Medium Sensor IAB Failure Sensor IAB Failure
Sensor IAB Module
Medium
Failure Sensor IAB Module Failure

7.2 MAQUET PulsioFlex Cardiac Monitoring Device


7.2.1 Output Signals - Parameters
Label Units Description Trend, record, print
CCI L/min/m2 Continuous Cardiac Index Yes
SVV % Stroke volume variation Yes
SVRI DS.m2/cm5 Systemic Vascular
Yes
Resistance Index
GEF % Global Ejection Fraction Yes
GEDI ml/min Global End-Diastolic
Yes
Volume Index
ELWI ml/kg Extra Vascular Lung Water
Yes
Index
PDR %/min Plasma Disappearance
No
Rate (of ICG)
dpMx mmHg/s Index of left ventricular
Yes
contractility
PVPI \ Pulmonary Vascular
Yes
Permeability Index
CFI L/min Cardiac Function Index Yes
CPI W/m2 Cardiac Power Index Yes
SVI ml/m2 Stroke Volume Index Yes

7-3
Label Units Description Trend, record, print
CCO L/min Continuous Cardiac
Yes
Output
SVR DS/cm5 Systemic Vascular
Yes
Resistance
PPV % Pulse pressure variation Yes
C.I. L/min/m2 Discontinuous Cardiac
Yes
Index from thermodilution
C.O. L/min Discontinuous Cardiac
Output from Yes
thermodilution
CPO W Cardiac Power Yes
GEDV ml Global End-Diastolic
Yes
Volume
EVLW ml Extra Vascular Lung Water Yes
R15 % Retention Rate (of ICG) No
SV ml Stroke Volume Yes
TB ℃,℉ Blood temperature Yes
pArt-S mmHg,kPa Systolic arterial pressure Yes
pArt-D mmHg,kPa Diastolic arterial pressure Yes
pArt-M mmHg,kPa Mean arterial pressure Yes
ScvO2 % Central Venous Oxygen
Yes
Saturation
DO2 ml/min Oxygen Delivery No
DO2I ml/min/m2 Oxygen Delivery Index No
VO2 ml/min Oxygen Consumption No
VO2I ml/min/m2 Oxygen Consumption
No
Index
O2ER % Oxygen Extraction Ratio No

7-4
7.2.2 Output Signals - Alarms
Mindray Patient Monitor Maquet PulsioFlex Alarm

Priority Label Label

Physiological alarms
pArt-S Low ApSys Low
pArt-S High ApSys High
pArt-M Low ApMap Low
pArt-M High ApMap High
CCO Low CO Low
High
CCO High CO High
CCI Low CI Low
CCI High CI High
ScvO2 High ScvO2 High
ScvO2 Low ScvO2 Low
Technical Alarms
Map Drop-check catheter Map drop-Check catheter
High AP/HR Not Available-please
HR/AP not available-Please check
check

7-5
7.3 Narcotrend Compact M EEG Monitoring Devices
7.3.1 Output Signals - Parameters
Label Units Description Trend, record, print
CFI1 \ NI Yes
EMG1 % EMG Yes
Delta1 % DeltaRel1 No
Delta2 % DeltaRel2 No
Theta1 % ThetaRel1 No
Theta2 % ThetaRel2 No
Alpha1 % AlphaRel1 No
Alpha2 % AlphaRel2 No
Beta1 % BetaRel1 No
Beta2 % BetaRel2 No
TP1 µV2 Power1 Yes
TP2 µV2 Power2 Yes
MF1 Hz Median1 No
MF2 Hz Median2 No
SEF1 Hz EdgeFreq1 Yes
SEF2 Hz EdgeFreq2 Yes
BSR1 % BSRShort1 Yes
STI1 % STI1 Yes
STI2 % STI2 Yes
BSR2 % BSRShort2 Yes
Imp1a kΩ Imp1a No
Imp1b kΩ Imp1b No
ImpRef kΩ ImpRef No
Imp2a kΩ Imp2a No
Imp2b kΩ Imp2b No

7-6
7.3.2 Output Signals -Technical Alarms
Mindray Patient Monitor Narcotrend-Compact M Alarm

Priority Label Label


CH1 Impedance Impedance
CH1 Offset Offset
CH1 EMG EMG
CH1 50 Hz / 60 Hz 50 Hz / 60 Hz
CH1 Interferences Interferences
CH1 Sharp Transients Sharp Transients
CH1 Artifact Artifact
CH1 Start Start
CH1 Inhomogeneous Inhomogeneous
CH1 Undifferentiated EEG Undifferentiated EEG
CH2 Impedance Impedance
CH2 Offset Offset
CH2 EMG EMG
Prompt
CH2 50 Hz / 60 Hz 50 Hz / 60 Hz
CH2 Interferences Interferences
CH2 Sharp Transients Sharp Transients
CH2 Artifact Artifact
CH2 Start Start
CH2 Inhomogeneous Inhomogeneous
CH2 Undifferentiated EEG Undifferentiated EEG
No patient lead is connected No patient lead is connected
1a electrode bad (impedance) 1a electrode bad
1b electrode bad (impedance) 1b electrode bad
Ref electrode bad (impedance) Ref electrode bad
2a electrode bad (impedance) 2a electrode bad
2b electrode bad (impedance) 2b electrode bad

7-7
Mindray Patient Monitor Narcotrend-Compact M Alarm

Priority Label Label


Electrode check (the raw EEG
Electrode Check
may not valid)

7.3.3 Output Signals - Waveforms (For BeneVision N series


Monitors)
Label Units
EEG1 µV
EEG2 µV

7.4 Organon TOF-Watch SX NMT Monitoring Device


7.4.1 Output Signals - Parameters
Label Units Description Trend, record, print
TOF-Ratio % Train-of-four ratio Yes
TOF-Count \ Train-of-four count Yes
PTC \ post tetanic count Yes
Single \ Single twitch stimulation Yes
Tskin ℃, ℉ Skin temperature No
T1 % Size of twitch 1 in %/ No
T2 % Size of twitch 2 in %/ No
T3 % / Size of twitch 3 in % No
T4 % / Size of twitch 4 in % No

7-8
7.4.2 Output Signals - Alarms
Mindray Patient Monitor TOF-Watch SX Alarm

Priority Label Label

Physiological alarms
Medium TOF Alarm TOF Alarm Enabled
Technical alarms
High TWSX Battery Depleted Battery empty
Medium TWSX Low Battery Battery Low
TWSX No Acceleration
AccelleCable Not Mounted
Sensor
TWSX No Temp Sensor TempCable Not Mounted
TWSX No Stimulation
SurfaceCable Not Mounted
Cable
SkinResist Too high
Low Calibration Error
BadAccSignal
BadTempSignal
TWSX Technical Alarm
StimMissing
AccMissing
BadElectrode
Internal Error

7-9
FOR YOUR NOTES

7-10
8 BeneLink Function Test

8.1 Preparing the Tools


Prepare the tools for the function test according to the type of the external device you
install. Please see the Instructions for Use of the corresponding external device for
guidance. For the function test of ventilator and anesthesia machine, at least the
following tools are needed:

 BeneVision patient monitor with BeneLink module properly installed

 Anesthesia machine or ventilator

 Gas source (tube or gas cylinder), including air or O2 at least, and with N2O or
other anesthesia gases as options

 Tube that connects the patient(or test lung)

 Test lung and a matching Y-pipe, or other accessories

8.2 Checking the Function


To check the function of the BeneLink module, follow this procedure:

1. Connect the BeneLink module to the ventilator or the anesthesia machine. See
Device Connection and Setup for more details.

2. Connect the gas supply and test lungs to the ventilator or anesthesia machine, turn
on the device, and configure as follows:

 Set the serial port of the external device by referring to Serial Port
Configuration List.

 Set the pressure control mode and check if the ventilator or anesthesia
machine works normally.

8-1
3. Make sure the ID adapter is correctly configured, and the green indicator of
corresponding port on the BeneLink module illuminates constantly.

4. Access the [Integrated Devices] screen on the patient monitor. Check that the device
type (ventilator or anesthesia machine) and ventilation mode are correctly displayed.

5. Select parameters PEEP, Pmean, VTe, MV, I:E, and f(RR) respectively on the patient
monitor and check if the parameter values displayed on the patient monitor are
consistent with those displayed on the ventilator or anesthesia machine.

6. Re-configure the above parameters on the ventilator or the anesthesia machine and
check if the parameter values displayed on the patient monitor change accordingly.

7. Trigger alarms [MV Low], [Airway Pressure Too High], [PAW High], [Peak High] and [No
Gas Supply] (no Air or O2) on the ventilator or the anesthesia machine. Check
whether these alarm messages are correctly recorded in the alarm list of the patient
monitor.

8. Switch the ventilator or anesthesia machine to volume control ventilation mode.


Check if the ventilation mode displayed on the patient monitor changes accordingly,
and if the parameter values of PEEP, Pmean, VTe, MV, I:E, and f(RR) are correctly
displayed.

8.3 Setting the Virtual Serial Port (For BeneVision N series


Monitors)
The ID adapter provides the virtual serial port. When the ID adapter is connected to
the BeneLink module, the third party device can obtain the realtime monitor data
through the virtual serial port.

8.3.1 Setting the Virtual Serial Port Properties


You should set the virtual serial port to realize the communication between the
monitor and the third party device. To set the serial port, proceed as follows:

8-2
1. Select the [Main Menu] quick key → from the [System] column select
[Maintenance]→ input the required password → select .

2. Select the [Others] tab →[Parameter Output Setup].

3. Set the [Baud Rate], [Parity Mode], [Data Bits] or [Stop Bits] as desired.

8.3.2 Setting the ID of the ID Adapter


The ID of the the virtual serial port is 6000a000. For the method of setting the ID, refer to
2.6 Setting the ID Adapter.

8.3.3 Parsing the Monitor Data


To parse the monitor data, refer to “DIAP Communication Protocol Service Manual (PN:
046-012201-00)”.

8-3
8.4 Setting the Serial Ports
8.4.1 Setting the Serial Ports of the Anesthesia Machines
Brand Model Protocol Baud Data Stop Parity Remark
Rate Bits Bits

Draeger Apollo Medibus 9600 8 1 Even Push knob/Switch screen/power together →


System Service Screen → Serial Port
Fabius GS Medibus 9600 7 1 Even Push knob/Switch screen/power together →
System Service Screen → Serial Port
Fabius plus/Tiro Medibus 9600 7 1 Even Push knob/Switch screen/power together
→ System Service Screen → Serial Port
Perseus A500 Medibus 9600 8 1 Even System Setup → System → interface
Primus Medibus 9600 8 1 Even Push knob/Switch screen/power together →
System Service Screen → Serial Port
ZeusIE Medibus 9600 / 1 Even Biomed Service → Basic Setup → Medibus
Setup → Com1/Com2
Commands to Medibus device: ON
Medibus version: SW4.n
Port : ON
GE Aespire 7900/7100 / 19200 7 1 Odd No need to set

8-4
Brand Model Protocol Baud Data Stop Parity Remark
Rate Bits Bits

Aestiva 7900/7100 / 19200 7 1 Odd No need to set


Aisys/Aisys CS2 / 19200 7 1 Odd No need to set
Avance / 19200 7 1 Odd No need to set
Hul Leon / 19200 8 1 None Standby → switch screen → Login →
Leon Plus / 19200 8 1 None Service Engineer(PW:0002) → Configuration
→ Page 2 → Serial Protocol: On, Mode:
request driven
MAQUET FLOW-i / 38400 8 1 Even No need to set
Mindray WATO / / / / / No need to set
EX-65/60/55/50/3
5/30/25/20
A7/A5/A4/A3/EX-5 / 57600 8 1 None No need to set
5 Pro/EX-65 Pro

8-5
8.4.2 Setting the Serial Ports of the Ventilator
Brand Model Protocol Baud Data Stop Parity Remark
Rate Bits Bits

AirLiquide ALMS T75 / 38400 8 1 None No need to set


Carefusion Vela / 115200 8 1 None /
Draeger Babylog / 9600 / 1 None Cal. Config → config → com→ Start
8000/8000plus bit: 1
Evita 2 Medibus 19200 / 1 Even Channel A: No need to set
Channel B: Menu Mode → Other
Evita 2 Dura Medibus 19200 / 1 Even Cal. Conf. → Device → Interface
Evita 4 Medibus 19200 / 1 Even Configuration → Ventilation → System
Defaults → Interface
Evita XL Medibus 19200 / 1 Even /
Evita V300 Medibus 19200 / 1 Even System setup → system → interface
Medibus X → COM
Infinity Medibus 19200 / 1 Even System setup → system → interface
V500/Babylog Medibus X → COM →Start bit: 1
VN500
Savina 300 Medibus 19200 / 1 Even System Setup → Interface

8-6
Brand Model Protocol Baud Data Stop Parity Remark
Rate Bits Bits

Medibus X
GE Carescape R860 / 19200 7 1 Odd No need to set
Engström / 19200 7 1 Odd No need to set
(Carestation)
Hamilton C2/C3/C1/T1/C1 Polling 9600 7 2 Even System → General → More → RS232
NEO/C6/C6S Protocol: GALILEO compatible
G5 Block 38400 8 1 None Enter the Configuration menu.
G5/S1 Polling 9600 7 2 Even Enter the Configuration menu.
Galileo Polling 9600 7 2 Even 1. Enter the Standby mode.
2. Press and hold hard keys “O2 ICON”
and “Lung ICON”, “Lock/Unlock ICON”
and “On/Off ICON”.
3. Select Configuration → Interface.
4. Set com1/2 to Hamilton C5/Polling.
S1 Block 38400 8 1 None Enter standby mode→ hold ”O2 ICON” &
“Lung ICON” hard button → hold
“Lock/Unlock ICON” & “On/Off ICON” hard

8-7
Brand Model Protocol Baud Data Stop Parity Remark
Rate Bits Bits

button → configuration → interface


→ com1/2: Hamilton G5/Block
Hul Leon i Plus / 19200 8 1 None Standby → switch screen → Login →
Service Engineer (PW:0002)
→ ”Connections → Serial Protocol: On,
Mode: request driven
MAQUET SERVO-i/SERVO-s / 9600 8 1 Even No need to set
SERVO-U/N / 38400 8 1 Even No need to set
Mindray SV / / / / / No need to set
300/350/600/650
/800/850
SynoVent E3/E5
Newport E360 Newport 38400 8 1 None Setup & Calibration → Technical →
Comm. Protocol: Newport
Philips Respironics V60 SNDA VRPT 19200 / / / Turn off machine → Hold confirm
button and then push power button →
Enter Diagnosis mode

8-8
Brand Model Protocol Baud Data Stop Parity Remark
Rate Bits Bits

Need an extra 25 pin to 9 pin adapter


Puritan Bennett PB840 SNDF 38400 8 / None Ventilator Setting → Other Screens →
SNDA Communications Setup
PB980 SNDF / / / / COM1: 840 DCI
SNDA
ResMed VSIII / 9600 8 1 / No need to set
Salvia Elisa800/600/500 / 19200 8 1 / No need to set
/300

8-9
8.4.3 Setting the Serial Ports of the Pumps
Brand Model Protocol Baud Data Stop Parity Remark
Rate Bits Bits

B.Braun Infusomat Space / 57600 8 1 None Using the Type F cable to connect the
Perfusor Space spaceCom of the spaceStation (if cascading
is needed. Refer to the official documents
on how to cascade several spaceStation.
Use an PC to connect the docking station
through Ethernet → Open IE and input
192.168.100.41 → Configuration → BCC
configuration panel : BCC 3.3x , interface:
com1, Baudrate:57600, Stop bits: 1,
Databits: 8, Character Stuffing: On,
Operation Mode: Request
Fresenius Fresenius Pump / 115200 8 1 None No need to set
Agilia Injectomat,
Agilia Injectomat
MC,
Agilia Injectomat

8-10
Brand Model Protocol Baud Data Stop Parity Remark
Rate Bits Bits

TIVA, Agilia
Volumat,
Agilia Volumat MC
Medima MedimaNet Server HL7 115200 8 1 None Use a serial to Ethernet module connect
with BeneLink ID module and MedimaNet
Server, setup serial to Ethernet module base
on serial parameters and network
parameters.
Mindray BeneFusion DS5 / 115200 8 1 None No need to set

8-11
8.4.4 Setting the Serial Ports of the TcGas Monitoring Devices
Brand Model Protocol Baud Rate Data Stop Parity Remark
Bits Bits

Biometer TCM MonLink 9600 8 1 Even Setup(PW: 19100) → Tech.settings →


Tosca/CombiM Continuous data output: MonLink
/TCM4/TCM40 You need also the serial interface adapter
with cable from Radiometer. (SN:
636-649)
TCM5 MonLink / / / / Service (PW: TCM5PW1)→ Serial →
protocol: Monlink for TCM4 Series
SenTec SenTec Digital SenTecLink 115200 / / / Interfaces → Serial Interface →
Monitor) Protocol: SenTecLink

8-12
8.4.5 Setting the Serial Ports of Other External Devices
Brand Model Protocol Baud Data Bits Stop Parity Remark
Rate Bits

MAQUET/ IABP IABP(CS300,CS1 / / / / / Adapting cable: mail to mail, directly


00) connected. Maximum size of the
connector is 15 mm × 33 mm.
MAQUET/cardiac PulsioFlex / / / / / Cable: ATEN USB to Serial Adapter
monitoring device UC-232A
Main Menu → System → Service →
LANCONF (Password) → Data Output
→ tick ”Data output” → set protocol
as” PulsioFlex V1.0”
Narcotrend/EEG Compact M / 115200 8 1 None No Need to set
monitoring device
Organon/ NMT TOF-Watch SX / 19200 8 1 None No Need to set.
monitoring device The serial interface adapter with cable
from Organon is required.

8-13
8.4.6 Installation and Test Report
Basic Information

Hospital Department
Serial number of ID Name of external
adapter device
ID of the external Type of serial port
device adapting cable
Software version and other information of
the external device

Test Result
Checking the connection
(Yes/No)

Can the patient monitor and the external device be assembled


together using designated accessories?
Does the green indicator of corresponding port on the
BeneLink module illuminate while the other indicators are not?
Are there numerics or characters displayed on the [Integrated
Devices] screen of the patient monitor?
Are the device type and parameter values displayed correctly
on the [Integrated Devices] screen of the patient monitor
when the external device just enters working mode?
Is the ventilation mode correctly displayed on the patient
monitor? Does it change correctly when the ventilation mode
on the external device is changed?

Checking the parameters

Parameters on
Parameters on the
Value the patient Value
external device
monitor
PEEP PEEP

8-14
Pmean Pmean

Ppeak Ppeak

f (RR) f (RR)

VTe VTe

Checking the alarms


Alarm Alarm messages displayed on the Alarm messages
external device displayed in the
alarm list of the
patient monitor
RR High

Apnea

Patient Disconnected

Other information

8-15
FOR YOUR NOTES

8-16
9 Alarm ID of External Devices

If the storage, display, or audio settings of a specific external device alarm are different
from its category or priority, you can set them individually by adding the Alarm ID to the
alarm list. Refer to “Interfacing with External Devices” of the N series Operator’s Manual for
the method.
This chapter lists the alarm IDs of external devices.

9.1 Alarm IDs of Anesthesia Machines


Alarm ID Alarm message Category Alarm priority Source

0xA5A1 Apnea Phys High Anes

Volume Apnea > 2


0xA5A2 Phys High Anes
min

0xA5A3 Pressure Apnea Phys High Anes

0xA5A4 Paw High Phys High Anes

0xA5A5 Paw Low Phys High Anes

0xA5A6 High Paw Sustained Phys High Anes

0xA5A7 Pressure Limiting Phys Low Anes

0xA5A8 Sub-Atmospheric Paw Phys Medium Anes

0xA5A9 FiO2 High Phys Medium Anes

0xA5AA FiO2 Low Phys High Anes

0xA5AB VTe High Phys Medium Anes

0xA5AC VTe Low Phys Medium Anes

0xA5AD MV High Phys Medium Anes

9-1
Alarm ID Alarm message Category Alarm priority Source

0xA5AE MV Low Phys Medium Anes

0xA5AF RR High Phys Low Anes

0xA5B0 RR Low Phys Low Anes

0xA5B1 PEEP High Phys Medium Anes

0xA5B2 PRESS EXP High Phys Medium Anes

0xA5B3 PRESSURE LIM Phys Low Anes

0xA5B4 CONT PRES Phys High Anes

0xA5B5 Volume Apnea Phys Medium Anes

0xA5B6 CO2 Apnea Phys High Anes

0xA5B7 RR Low Phys Medium Anes

0xA5B8 RR High Phys Medium Anes

0xA5B9 EtCO2 High Phys Medium Anes

0xA5BA EtCO2 Low Phys Medium Anes

0xA5BB FiCO2 High Phys Medium Anes

0xA5BC FiCO2 Low Phys Medium Anes

0xA5BD EtN2O High Phys Medium Anes

0xA5BE EtN2O Low Phys Medium Anes

0xA5BF FiN2O High Phys Medium Anes

0xA5C0 FiN2O Low Phys Medium Anes

0xA5C1 EtHal High Phys Medium Anes

0xA5C2 EtHal Low Phys Medium Anes

0xA5C3 FiHal High Phys Medium Anes

0xA5C4 FiHal Low Phys Medium Anes

0xA5C5 EtEnf High Phys Medium Anes


9-2
Alarm ID Alarm message Category Alarm priority Source

0xA5C6 EtEnf Low Phys Medium Anes

0xA5C7 FiEnf High Phys Medium Anes

0xA5C8 FiEnf Low Phys Medium Anes

0xA5C9 EtIso High Phys Medium Anes

0xA5CA EtIso Low Phys Medium Anes

0xA5CB FiIso High Phys Medium Anes

0xA5CC FiIso Low Phys Medium Anes

0xA5CD EtSev High Phys Medium Anes

0xA5CE EtSev Low Phys Medium Anes

0xA5CF FiSev High Phys Medium Anes

0xA5D0 FiSev Low Phys Medium Anes

0xA5D1 EtDes High Phys Medium Anes

0xA5D2 EtDes Low Phys Medium Anes

0xA5D3 FiDes High Phys Medium Anes

0xA5D4 FiDes Low Phys Medium Anes

0xA5D5 MAC Low Phys Medium Anes

0xA5D6 EtAA Low Phys Medium Anes

0xA5D7 EtAA High Phys Medium Anes

0xA5D8 FiAA Low Phys Medium Anes

0xA5D9 FiAA High Phys Medium Anes

0xA5DA EtO2 High Phys High Anes

0xA5DB EtO2 Low Phys High Anes

0xA5DC FiO2 High Phys High Anes

0xA5DD FiO2 Low Phys High Anes


9-3
Alarm ID Alarm message Category Alarm priority Source

0xA5DE BIS High Phys Medium Anes

0xA5DF BIS Low Phys Medium Anes

0xA5E0 No Pulse Phys High Anes

0xA5E1 PR Low Phys High Anes

0xA5E2 PR High Phys Medium Anes

0xA5E3 SpO2 High Phys Medium Anes

0xA5E4 SpO2 Low Phys High Anes

0xA5E5 Patient Circuit Leak Tech Medium Anes

O2 Sensor
0xA5E6 Tech Low Anes
Unconnected

Drive Gas Pressure


0xA5E7 Tech High Anes
Low

0xA5E8 O2 Supply Failure Tech High Anes

0xA5E9 Battery in Use Tech Low Anes

0xA5EA APL VALVE? Tech High Anes

0xA5EB EXP-VALVE? Tech Medium Anes

0xA5EC FRESH GAS? Tech Medium Anes

0xA5ED No Fresh Gas Tech High Anes

0xA5EE Circuit Occluded Tech High Anes

0xA5EF VENT DISC Tech High Anes

0xA5F0 NO AIR Tech Low Anes

0xA5F1 NO O2 SUPPLY Tech Low Anes

0xA5F2 High Technical Alarm Tech High Anes

9-4
Alarm ID Alarm message Category Alarm priority Source

Medium Technical
0xA5F3 Tech Medium Anes
Alarm

0xA5F4 Low Technical Alarm Tech Low Anes

0xA5F5 Prompt Tech Prompt Anes

CO2 Module
0xA5F6 Tech Medium Anes
Abnormal

0xA5F7 AG Module Abnormal Tech Medium Anes

0xA5F8 BIS Module Abnormal Tech Medium Anes

SpO2 Module
0xA5F9 Tech Low Anes
Abnormal

0xA5FA PEEP Low Phys Medium Anes

9.2 Alarm IDs of Ventilators


Alarm ID Alarm message Category Alarm priority Source

0xA66D FiCO2 High Phys Medium Vent

0xA66E VOL INCONST Phys Medium Vent

0xA66F RR High Phys Medium Vent

0xA670 AW-TEMP HIGH Phys High Vent

0xA671 PEEP High Phys High Vent

0xA672 ASB > 4s Phys High Vent

0xA673 ASB > 1.5s Phys Low Vent

9-5
Alarm ID Alarm message Category Alarm priority Source

0xA674 PPS-TI > 1.5s Phys Low Vent

0xA675 ASB > Tinsp Phys Low Vent

0xA676 No Pulse Phys High Vent

0xA677 PR Low Phys High Vent

0xA678 SpO2 Low Phys High Vent

0xA679 PR High Phys High Vent

0xA67A SpO2 High Phys High Vent

0xA67B VTe Low Phys Medium Vent

0xA67C High Paw Sustained Phys High Vent

0xA67D Pressure Sustained Phys High Vent

0xA67E Sub-Atmospheric Paw Phys Medium Vent

0xA67F Pmax Reached Phys High Vent

0xA680 Pinsp Not Achieved Phys High Vent

0xA681 PEEP Not Achieved Phys High Vent

No Pressure
0xA682 Phys High Vent
PEEP/CPAP

0xA683 VT Not Achieved Phys High Vent

0xA684 Volume Apnea Phys Medium Vent

Volume Apnea > 2


0xA685 Phys High Vent
min

0xA686 High Circuit O2 Phys High Vent

0xA687 Low Circuit O2 Phys High Vent

0xA688 CO2 Apnea Phys High Vent

0xA689 EtO2 High Phys Medium Vent

9-6
Alarm ID Alarm message Category Alarm priority Source

0xA68A EtO2 Low Phys Medium Vent

0xA68B Ppeak High Phys High Vent

0xA68C Ppeak Low Phys High Vent

0xA68D PEEPe High Phys Medium Vent

0xA68E PEEPe Low Phys Medium Vent

0xA68F PEEPi High Phys Medium Vent

0xA690 Paux High Phys Medium Vent

0xA691 Base Flow High Phys Low Vent

0xA692 Loss of PEEP Phys High Vent

0xA693 Low Baseline Phys High Vent

0xA694 High Baseline Phys High Vent

0xA695 Sustained Hbline Phys High Vent

Air Supply Pressure


0xA696 Tech High Vent
Low

O2 Supply Pressure
0xA697 Tech High Vent
Low

No Gas Supply
0xA698 Tech High Vent
Pressure

0xA699 Airway Obstructed? Tech High Vent

0xA69A Tube Disconnected? Tech High Vent

0xA69B Airway Leak? Tech Low Vent

0xA69C Battery in Use Tech Low Vent

0xA69D Check Flow Sensors Tech High Vent

0xA69E EXP-VALVE? Tech High Vent

9-7
Alarm ID Alarm message Category Alarm priority Source

0xA69F CLEAN CO2 Tech High Vent

Drive Gas Pressure


0xA6A0 Tech High Vent
Low

0xA6A1 Patient Circuit Leak Tech Medium Vent

0xA6A2 Neo Flow Sensor Error Tech High Vent

0xA6A3 O2 and air supply Tech Medium Vent

0xA6A4 O2 and heliox supply Tech Medium Vent

Sustained Airway
0xA6A5 Tech High Vent
Pressure

Insp gas temperature


0xA6A6 Tech High Vent
high

0xA6A7 Tinsp Long Tech Low Vent

0xA6A8 CO2 No Watertrap Tech Low Vent

0xA6A9 No VO2, High FiN2O Tech High Vent

0xA6AA No O2 Pressure Tech High Vent

0xA6AB No Fresh Gas Flow Tech High Vent

0xA6AC No VO2, FiO2 > 85% Tech High Vent

MGAS Replace Water


0xA6AD Tech High Vent
Trap

0xA6AE 12 Hour Test Tech Low Vent

0xA6AF Patient Connected? Tech High Vent

Negative Airway
0xA6B0 Tech High Vent
Pressure

0xA6B1 Circuit Leak Tech High Vent

9-8
Alarm ID Alarm message Category Alarm priority Source

Patient Connection
0xA6B2 Tech High Vent
Leak

0xA6B3 Patient Disconnected Tech High Vent

0xA6B4 O2 cell disconnect Tech High Vent

0xA6B5 Check tubing Tech Low Vent

Disconnection
0xA6B6 Tech High Vent
ventilator side

0xA6B7 O2 cell cal Needed Tech High Vent

0xA6B8 Circuit Disconnect Tech High Vent

0xA6B9 Power Failure Tech High Vent

0xA6BA Tinsp Short Tech High Vent

FiO2 Sensor
0xA6BB Tech Medium Vent
Disconnected

0xA6BC High Technical Alarm Tech High Vent

Medium Technical
0xA6BD Tech Medium Vent
Alarm

0xA6BE Low Technical Alarm Tech Low Vent

0xA6BF Prompt Tech Prompt Vent

0xA6C0 PEEP Low Phys High Vent

0xA6C1 FiO2 Alarm Phys High Vent

0xA96E High DCO2 Tech Medium Vent

0xA96F Low DCO2 Tech Medium Vent

0xA970 SPO2 Desat Phys High Vent

9-9
9.3 Alarm IDs of tcGas monitoring devices
Alarm ID Alarm message Category Alarm priority Source

0xA6CB PR Alarm Phys Medium TcGas

0xA6CC SpO2 Alarm Phys Medium TcGas

0xA6CD tcpCO2 Alarm Phys Medium TcGas

0xA6CE tcpO2 Alarm Phys Medium TcGas

0xA6CF tcGas Low Battery Tech Medium TcGas

tcGas Battery Tech


0xA6D0 High TcGas
Depleted

TCM Temperature Tech


0xA6D1 High TcGas
High

0xA6D2 TCM Alert Tech Low TcGas

9-10
9.4 Alarm IDs of NMT monitoring device
Alarm ID Alarm message Category Alarm priority Source

0xA6C2 TWSX Low Battery Tech Medium NMT

TWSX Battery Tech


0xA6C3 High NMT
Depleted

TWSX No Acceleration Tech


0xA6C4 Low NMT
Sensor

TWSX No Temp Tech


0xA6C5 Low NMT
Sensor

TWSX No Stimulation Tech


0xA6C6 Low NMT
Cable

0xA6C7 TWSX Technical Alarm Tech Low NMT

0xA6C8 TWSX Technical Alarm Tech Low NMT

0xA6C9 TWSX Technical Alarm Tech Low NMT

0xA6CA TOF Alarm Phys Medium NMT

0xA6D3 TWSX Technical Alarm Tech Low NMT

0xA6D4 TWSX Technical Alarm Tech Low NMT

0xA6D5 TWSX Not Cal Tech Prompt NMT

9.5 Alarm IDs of Infusion Systems (for BeneVision N series


Monitors)
Alarm ID Alarm message Category Alarm priority Source

0xA73F Pre-Occlusion Tech Medium Pump

0xA740 Pre-Occlusion Tech Medium Pump

0xA741 Pre-Occlusion Tech Medium Pump

9-11
Alarm ID Alarm message Category Alarm priority Source

0xA742 Pre-Occlusion Tech Medium Pump

0xA743 Pre-Occlusion Tech Medium Pump

0xA744 Pre-Occlusion Tech Medium Pump

0xA745 Pre-Occlusion Tech Medium Pump

0xA746 Pre-Occlusion Tech Medium Pump

0xA747 Pre-Occlusion Tech Medium Pump

0xA748 Pre-Occlusion Tech Medium Pump

0xA749 Pre-Occlusion Tech Medium Pump

0xA74A Pre-Occlusion Tech Medium Pump

0xA74B Pre-Occlusion Tech Medium Pump

0xA74C Pre-Occlusion Tech Medium Pump

0xA74D Pre-Occlusion Tech Medium Pump

0xA74E Pre-Occlusion Tech Medium Pump

0xA74F Pre-Occlusion Tech Medium Pump

0xA750 Pre-Occlusion Tech Medium Pump

0xA751 Pre-Occlusion Tech Medium Pump

0xA752 Pre-Occlusion Tech Medium Pump

0xA753 Pre-Occlusion Tech Medium Pump

0xA754 Pre-Occlusion Tech Medium Pump

0xA755 Pre-Occlusion Tech Medium Pump

0xA756 Pre-Occlusion Tech Medium Pump

0xA757 Occlusion Tech High Pump

0xA758 Occlusion Tech High Pump

0xA759 Occlusion Tech High Pump


9-12
Alarm ID Alarm message Category Alarm priority Source

0xA75A Occlusion Tech High Pump

0xA75B Occlusion Tech High Pump

0xA75C Occlusion Tech High Pump

0xA75D Occlusion Tech High Pump

0xA75E Occlusion Tech High Pump

0xA75F Occlusion Tech High Pump

0xA760 Occlusion Tech High Pump

0xA761 Occlusion Tech High Pump

0xA762 Occlusion Tech High Pump

0xA763 Occlusion Tech High Pump

0xA764 Occlusion Tech High Pump

0xA765 Occlusion Tech High Pump

0xA766 Occlusion Tech High Pump

0xA767 Occlusion Tech High Pump

0xA768 Occlusion Tech High Pump

0xA769 Occlusion Tech High Pump

0xA76A Occlusion Tech High Pump

0xA76B Occlusion Tech High Pump

0xA76C Occlusion Tech High Pump

0xA76D Occlusion Tech High Pump

0xA76E Occlusion Tech High Pump

0xA76F Air bubble Tech High Pump

0xA770 Air bubble Tech High Pump

0xA771 Air bubble Tech High Pump


9-13
Alarm ID Alarm message Category Alarm priority Source

0xA772 Air bubble Tech High Pump

0xA773 Air bubble Tech High Pump

0xA774 Air bubble Tech High Pump

0xA775 Air bubble Tech High Pump

0xA776 Air bubble Tech High Pump

0xA777 Air bubble Tech High Pump

0xA778 Air bubble Tech High Pump

0xA779 Air bubble Tech High Pump

0xA77A Air bubble Tech High Pump

0xA77B Air bubble Tech High Pump

0xA77C Air bubble Tech High Pump

0xA77D Air bubble Tech High Pump

0xA77E Air bubble Tech High Pump

0xA77F Air bubble Tech High Pump

0xA780 Air bubble Tech High Pump

0xA781 Air bubble Tech High Pump

0xA782 Air bubble Tech High Pump

0xA783 Air bubble Tech High Pump

0xA784 Air bubble Tech High Pump

0xA785 Air bubble Tech High Pump

0xA786 Air bubble Tech High Pump

0xA787 Drop too less Tech High Pump

0xA788 Drop too less Tech High Pump

0xA789 Drop too less Tech High Pump


9-14
Alarm ID Alarm message Category Alarm priority Source

0xA78A Drop too less Tech High Pump

0xA78B Drop too less Tech High Pump

0xA78C Drop too less Tech High Pump

0xA78D Drop too less Tech High Pump

0xA78E Drop too less Tech High Pump

0xA78F Drop too less Tech High Pump

0xA790 Drop too less Tech High Pump

0xA791 Drop too less Tech High Pump

0xA792 Drop too less Tech High Pump

0xA793 Drop too less Tech High Pump

0xA794 Drop too less Tech High Pump

0xA795 Drop too less Tech High Pump

0xA796 Drop too less Tech High Pump

0xA797 Drop too less Tech High Pump

0xA798 Drop too less Tech High Pump

0xA799 Drop too less Tech High Pump

0xA79A Drop too less Tech High Pump

0xA79B Drop too less Tech High Pump

0xA79C Drop too less Tech High Pump

0xA79D Drop too less Tech High Pump

0xA79E Drop too less Tech High Pump

0xA79F Drop too many Tech High Pump

0xA7A0 Drop too many Tech High Pump

0xA7A1 Drop too many Tech High Pump


9-15
Alarm ID Alarm message Category Alarm priority Source

0xA7A2 Drop too many Tech High Pump

0xA7A3 Drop too many Tech High Pump

0xA7A4 Drop too many Tech High Pump

0xA7A5 Drop too many Tech High Pump

0xA7A6 Drop too many Tech High Pump

0xA7A7 Drop too many Tech High Pump

0xA7A8 Drop too many Tech High Pump

0xA7A9 Drop too many Tech High Pump

0xA7AA Drop too many Tech High Pump

0xA7AB Drop too many Tech High Pump

0xA7AC Drop too many Tech High Pump

0xA7AD Drop too many Tech High Pump

0xA7AE Drop too many Tech High Pump

0xA7AF Drop too many Tech High Pump

0xA7B0 Drop too many Tech High Pump

0xA7B1 Drop too many Tech High Pump

0xA7B2 Drop too many Tech High Pump

0xA7B3 Drop too many Tech High Pump

0xA7B4 Drop too many Tech High Pump

0xA7B5 Drop too many Tech High Pump

0xA7B6 Drop too many Tech High Pump

0xA7DE Drop alarm Tech High Pump

0xA7DF Drop alarm Tech High Pump

0xA7E0 Drop alarm Tech High Pump


9-16
Alarm ID Alarm message Category Alarm priority Source

0xA7E1 Drop alarm Tech High Pump

0xA7E2 Drop alarm Tech High Pump

0xA7E3 Drop alarm Tech High Pump

0xA7E4 Drop alarm Tech High Pump

0xA7E5 Drop alarm Tech High Pump

0xA7E6 Drop alarm Tech High Pump

0xA7E7 Drop alarm Tech High Pump

0xA7E8 Drop alarm Tech High Pump

0xA7E9 Drop alarm Tech High Pump

0xA7EA Drop alarm Tech High Pump

0xA7EB Drop alarm Tech High Pump

0xA7EC Drop alarm Tech High Pump

0xA7ED Drop alarm Tech High Pump

0xA7EE Drop alarm Tech High Pump

0xA7EF Drop alarm Tech High Pump

0xA7F0 Drop alarm Tech High Pump

0xA7F1 Drop alarm Tech High Pump

0xA7F2 Drop alarm Tech High Pump

0xA7F3 Drop alarm Tech High Pump

0xA7F4 Drop alarm Tech High Pump

0xA7F5 Drop alarm Tech High Pump

0xA7F6 Syringe disengagement Tech High Pump

0xA7F7 Syringe disengagement Tech High Pump

0xA7F8 Syringe disengagement Tech High Pump


9-17
Alarm ID Alarm message Category Alarm priority Source

0xA7F9 Syringe disengagement Tech High Pump

0xA7FA Syringe disengagement Tech High Pump

0xA7FB Syringe disengagement Tech High Pump

0xA7FC Syringe disengagement Tech High Pump

0xA7FD Syringe disengagement Tech High Pump

0xA7FE Syringe disengagement Tech High Pump

0xA7FF Syringe disengagement Tech High Pump

0xA800 Syringe disengagement Tech High Pump

0xA801 Syringe disengagement Tech High Pump

0xA802 Syringe disengagement Tech High Pump

0xA803 Syringe disengagement Tech High Pump

0xA804 Syringe disengagement Tech High Pump

0xA805 Syringe disengagement Tech High Pump

0xA806 Syringe disengagement Tech High Pump

0xA807 Syringe disengagement Tech High Pump

0xA808 Syringe disengagement Tech High Pump

0xA809 Syringe disengagement Tech High Pump

0xA80A Syringe disengagement Tech High Pump

0xA80B Syringe disengagement Tech High Pump

0xA80C Syringe disengagement Tech High Pump

0xA80D Syringe disengagement Tech High Pump

0xA80E Mechanical Abnormal Tech High Pump

0xA80F Mechanical Abnormal Tech High Pump

0xA810 Mechanical Abnormal Tech High Pump


9-18
Alarm ID Alarm message Category Alarm priority Source

0xA811 Mechanical Abnormal Tech High Pump

0xA812 Mechanical Abnormal Tech High Pump

0xA813 Mechanical Abnormal Tech High Pump

0xA814 Mechanical Abnormal Tech High Pump

0xA815 Mechanical Abnormal Tech High Pump

0xA816 Mechanical Abnormal Tech High Pump

0xA817 Mechanical Abnormal Tech High Pump

0xA818 Mechanical Abnormal Tech High Pump

0xA819 Mechanical Abnormal Tech High Pump

0xA81A Mechanical Abnormal Tech High Pump

0xA81B Mechanical Abnormal Tech High Pump

0xA81C Mechanical Abnormal Tech High Pump

0xA81D Mechanical Abnormal Tech High Pump

0xA81E Mechanical Abnormal Tech High Pump

0xA81F Mechanical Abnormal Tech High Pump

0xA820 Mechanical Abnormal Tech High Pump

0xA821 Mechanical Abnormal Tech High Pump

0xA822 Mechanical Abnormal Tech High Pump

0xA823 Mechanical Abnormal Tech High Pump

0xA824 Mechanical Abnormal Tech High Pump

0xA825 Mechanical Abnormal Tech High Pump

0xA826 Sensor alarm Tech High Pump

0xA827 Sensor alarm Tech High Pump

0xA828 Sensor alarm Tech High Pump


9-19
Alarm ID Alarm message Category Alarm priority Source

0xA829 Sensor alarm Tech High Pump

0xA82A Sensor alarm Tech High Pump

0xA82B Sensor alarm Tech High Pump

0xA82C Sensor alarm Tech High Pump

0xA82D Sensor alarm Tech High Pump

0xA82E Sensor alarm Tech High Pump

0xA82F Sensor alarm Tech High Pump

0xA830 Sensor alarm Tech High Pump

0xA831 Sensor alarm Tech High Pump

0xA832 Sensor alarm Tech High Pump

0xA833 Sensor alarm Tech High Pump

0xA834 Sensor alarm Tech High Pump

0xA835 Sensor alarm Tech High Pump

0xA836 Sensor alarm Tech High Pump

0xA837 Sensor alarm Tech High Pump

0xA838 Sensor alarm Tech High Pump

0xA839 Sensor alarm Tech High Pump

0xA83A Sensor alarm Tech High Pump

0xA83B Sensor alarm Tech High Pump

0xA83C Sensor alarm Tech High Pump

0xA83D Sensor alarm Tech High Pump

0xA83E Door Open Tech High Pump

0xA83F Door Open Tech High Pump

0xA840 Door Open Tech High Pump


9-20
Alarm ID Alarm message Category Alarm priority Source

0xA841 Door Open Tech High Pump

0xA842 Door Open Tech High Pump

0xA843 Door Open Tech High Pump

0xA844 Door Open Tech High Pump

0xA845 Door Open Tech High Pump

0xA846 Door Open Tech High Pump

0xA847 Door Open Tech High Pump

0xA848 Door Open Tech High Pump

0xA849 Door Open Tech High Pump

0xA84A Door Open Tech High Pump

0xA84B Door Open Tech High Pump

0xA84C Door Open Tech High Pump

0xA84D Door Open Tech High Pump

0xA84E Door Open Tech High Pump

0xA84F Door Open Tech High Pump

0xA850 Door Open Tech High Pump

0xA851 Door Open Tech High Pump

0xA852 Door Open Tech High Pump

0xA853 Door Open Tech High Pump

0xA854 Door Open Tech High Pump

0xA855 Door Open Tech High Pump

0xA856 end of infusion Tech High Pump

0xA857 end of infusion Tech High Pump

0xA858 end of infusion Tech High Pump


9-21
Alarm ID Alarm message Category Alarm priority Source

0xA859 end of infusion Tech High Pump

0xA85A end of infusion Tech High Pump

0xA85B end of infusion Tech High Pump

0xA85C end of infusion Tech High Pump

0xA85D end of infusion Tech High Pump

0xA85E end of infusion Tech High Pump

0xA85F end of infusion Tech High Pump

0xA860 end of infusion Tech High Pump

0xA861 end of infusion Tech High Pump

0xA862 end of infusion Tech High Pump

0xA863 end of infusion Tech High Pump

0xA864 end of infusion Tech High Pump

0xA865 end of infusion Tech High Pump

0xA866 end of infusion Tech High Pump

0xA867 end of infusion Tech High Pump

0xA868 end of infusion Tech High Pump

0xA869 end of infusion Tech High Pump

0xA86A end of infusion Tech High Pump

0xA86B end of infusion Tech High Pump

0xA86C end of infusion Tech High Pump

0xA86D end of infusion Tech High Pump

0xA86E Syringe Empty Tech High Pump

0xA86F Syringe Empty Tech High Pump

0xA870 Syringe Empty Tech High Pump


9-22
Alarm ID Alarm message Category Alarm priority Source

0xA871 Syringe Empty Tech High Pump

0xA872 Syringe Empty Tech High Pump

0xA873 Syringe Empty Tech High Pump

0xA874 Syringe Empty Tech High Pump

0xA875 Syringe Empty Tech High Pump

0xA876 Syringe Empty Tech High Pump

0xA877 Syringe Empty Tech High Pump

0xA878 Syringe Empty Tech High Pump

0xA879 Syringe Empty Tech High Pump

0xA87A Syringe Empty Tech High Pump

0xA87B Syringe Empty Tech High Pump

0xA87C Syringe Empty Tech High Pump

0xA87D Syringe Empty Tech High Pump

0xA87E Syringe Empty Tech High Pump

0xA87F Syringe Empty Tech High Pump

0xA880 Syringe Empty Tech High Pump

0xA881 Syringe Empty Tech High Pump

0xA882 Syringe Empty Tech High Pump

0xA883 Syringe Empty Tech High Pump

0xA884 Syringe Empty Tech High Pump

0xA885 Syringe Empty Tech High Pump

0xA886 Syringe Near Empty Tech Low Pump

0xA887 Syringe Near Empty Tech Low Pump

0xA888 Syringe Near Empty Tech Low Pump


9-23
Alarm ID Alarm message Category Alarm priority Source

0xA889 Syringe Near Empty Tech Low Pump

0xA88A Syringe Near Empty Tech Low Pump

0xA88B Syringe Near Empty Tech Low Pump

0xA88C Syringe Near Empty Tech Low Pump

0xA88D Syringe Near Empty Tech Low Pump

0xA88E Syringe Near Empty Tech Low Pump

0xA88F Syringe Near Empty Tech Low Pump

0xA890 Syringe Near Empty Tech Low Pump

0xA891 Syringe Near Empty Tech Low Pump

0xA892 Syringe Near Empty Tech Low Pump

0xA893 Syringe Near Empty Tech Low Pump

0xA894 Syringe Near Empty Tech Low Pump

0xA895 Syringe Near Empty Tech Low Pump

0xA896 Syringe Near Empty Tech Low Pump

0xA897 Syringe Near Empty Tech Low Pump

0xA898 Syringe Near Empty Tech Low Pump

0xA899 Syringe Near Empty Tech Low Pump

0xA89A Syringe Near Empty Tech Low Pump

0xA89B Syringe Near Empty Tech Low Pump

0xA89C Syringe Near Empty Tech Low Pump

0xA89D Syringe Near Empty Tech Low Pump

0xA89E Infusion Near Done Tech Low Pump

0xA89F Infusion Near Done Tech Low Pump

0xA8A0 Infusion Near Done Tech Low Pump


9-24
Alarm ID Alarm message Category Alarm priority Source

0xA8A1 Infusion Near Done Tech Low Pump

0xA8A2 Infusion Near Done Tech Low Pump

0xA8A3 Infusion Near Done Tech Low Pump

0xA8A4 Infusion Near Done Tech Low Pump

0xA8A5 Infusion Near Done Tech Low Pump

0xA8A6 Infusion Near Done Tech Low Pump

0xA8A7 Infusion Near Done Tech Low Pump

0xA8A8 Infusion Near Done Tech Low Pump

0xA8A9 Infusion Near Done Tech Low Pump

0xA8AA Infusion Near Done Tech Low Pump

0xA8AB Infusion Near Done Tech Low Pump

0xA8AC Infusion Near Done Tech Low Pump

0xA8AD Infusion Near Done Tech Low Pump

0xA8AE Infusion Near Done Tech Low Pump

0xA8AF Infusion Near Done Tech Low Pump

0xA8B0 Infusion Near Done Tech Low Pump

0xA8B1 Infusion Near Done Tech Low Pump

0xA8B2 Infusion Near Done Tech Low Pump

0xA8B3 Infusion Near Done Tech Low Pump

0xA8B4 Infusion Near Done Tech Low Pump

0xA8B5 Infusion Near Done Tech Low Pump

0xA8B6 System Abnormal Tech Medium Pump

0xA8B7 System Abnormal Tech Medium Pump

0xA8B8 System Abnormal Tech Medium Pump


9-25
Alarm ID Alarm message Category Alarm priority Source

0xA8B9 System Abnormal Tech Medium Pump

0xA8BA System Abnormal Tech Medium Pump

0xA8BB System Abnormal Tech Medium Pump

0xA8BC System Abnormal Tech Medium Pump

0xA8BD System Abnormal Tech Medium Pump

0xA8BE System Abnormal Tech Medium Pump

0xA8BF System Abnormal Tech Medium Pump

0xA8C0 System Abnormal Tech Medium Pump

0xA8C1 System Abnormal Tech Medium Pump

0xA8C2 System Abnormal Tech Medium Pump

0xA8C3 System Abnormal Tech Medium Pump

0xA8C4 System Abnormal Tech Medium Pump

0xA8C5 System Abnormal Tech Medium Pump

0xA8C6 System Abnormal Tech Medium Pump

0xA8C7 System Abnormal Tech Medium Pump

0xA8C8 System Abnormal Tech Medium Pump

0xA8C9 System Abnormal Tech Medium Pump

0xA8CA System Abnormal Tech Medium Pump

0xA8CB System Abnormal Tech Medium Pump

0xA8CC System Abnormal Tech Medium Pump

0xA8CD System Abnormal Tech Medium Pump

0xA8CE Low Battery Tech High Pump

0xA8CF Low Battery Tech High Pump

0xA8D0 Low Battery Tech High Pump


9-26
Alarm ID Alarm message Category Alarm priority Source

0xA8D1 Low Battery Tech High Pump

0xA8D2 Low Battery Tech High Pump

0xA8D3 Low Battery Tech High Pump

0xA8D4 Low Battery Tech High Pump

0xA8D5 Low Battery Tech High Pump

0xA8D6 Low Battery Tech High Pump

0xA8D7 Low Battery Tech High Pump

0xA8D8 Low Battery Tech High Pump

0xA8D9 Low Battery Tech High Pump

0xA8DA Low Battery Tech High Pump

0xA8DB Low Battery Tech High Pump

0xA8DC Low Battery Tech High Pump

0xA8DD Low Battery Tech High Pump

0xA8DE Low Battery Tech High Pump

0xA8DF Low Battery Tech High Pump

0xA8E0 Low Battery Tech High Pump

0xA8E1 Low Battery Tech High Pump

0xA8E2 Low Battery Tech High Pump

0xA8E3 Low Battery Tech High Pump

0xA8E4 Low Battery Tech High Pump

0xA8E5 Low Battery Tech High Pump

0xA971 KVO Done Tech High Pump

0xA972 KVO Done Tech High Pump

0xA973 KVO Done Tech High Pump


9-27
Alarm ID Alarm message Category Alarm priority Source

0xA974 KVO Done Tech High Pump

0xA975 KVO Done Tech High Pump

0xA976 KVO Done Tech High Pump

0xA977 KVO Done Tech High Pump

0xA978 KVO Done Tech High Pump

0xA979 KVO Done Tech High Pump

0xA97A KVO Done Tech High Pump

0xA97B KVO Done Tech High Pump

0xA97C KVO Done Tech High Pump

0xA97D KVO Done Tech High Pump

0xA97E KVO Done Tech High Pump

0xA97F KVO Done Tech High Pump

0xA980 KVO Done Tech High Pump

0xA981 KVO Done Tech High Pump

0xA982 KVO Done Tech High Pump

0xA983 KVO Done Tech High Pump

0xA984 KVO Done Tech High Pump

0xA985 KVO Done Tech High Pump

0xA986 KVO Done Tech High Pump

0xA987 KVO Done Tech High Pump

0xA988 KVO Done Tech High Pump

0xA7C3 Electrode check Tech Prompt EEG

0xA7C4 CH1 Impedance Tech Prompt EEG

0xA7C5 CH1 Offset Tech Prompt EEG


9-28
Alarm ID Alarm message Category Alarm priority Source

0xA7C6 CH1 EMG Tech Prompt EEG

0xA7C7 CH1 50 Hz / 60 Hz Tech Prompt EEG

0xA7C8 CH1 Interferences Tech Prompt EEG

0xA7C9 CH1 Sharp Transients Tech Prompt EEG

0xA7CA CH1 Artifact Tech Prompt EEG

0xA7CB CH1 Start Tech Prompt EEG

0xA7CC CH1 Inhomogeneous Tech Prompt EEG

CH1 Undifferentiated
0xA7CD Tech Prompt EEG
EEG

0xA7CE CH2 Impedance Tech Prompt EEG

0xA7CF CH2 Offset Tech Prompt EEG

0xA7D0 CH2 EMG Tech Prompt EEG

0xA7D1 CH2 50 Hz / 60 Hz Tech Prompt EEG

0xA7D2 CH2 Interferences Tech Prompt EEG

0xA7D3 CH2 Sharp Transients Tech Prompt EEG

0xA7D4 CH2 Artifact Tech Prompt EEG

0xA7D5 CH2 Start Tech Prompt EEG

0xA7D6 CH2 Inhomogeneous Tech Prompt EEG

CH2 Undifferentiated
0xA7D7 Tech Prompt EEG
EEG

No patient lead is
0xA7D8 Tech Prompt EEG
connected

0xA7D9 1a electrode bad Tech Prompt EEG

0xA7DA 1b electrode bad Tech Prompt EEG

9-29
Alarm ID Alarm message Category Alarm priority Source

0xA7DB Ref electrode bad Tech Prompt EEG

0xA7DC 2a electrode bad Tech Prompt EEG

0xA7DD 2b electrode bad Tech Prompt EEG

0xA962 CH1 High SR Phys High EEG

0xA963 CH1 Low CFI Phys High EEG

0xA964 CH1 High CFI Phys High EEG

CH1 Perform Electrode


0xA965 Tech Low EEG
Check

CH1 Re-prepocess
0xA966 Tech Low EEG
Electrodes

CH1 Check Patient


0xA967 Tech Low EEG
Cable and Leads

0xA968 CH2 High SR Phys High EEG

0xA969 CH2 Low CFI Phys High EEG

0xA96A CH2 High CFI Phys High EEG

CH2 Perform Electrode


0xA96B Tech Low EEG
Check

CH2 Re-prepocess
0xA96C Tech Low EEG
Electrodes

CH2 Check Patient


0xA96D Tech Low EEG
Cable and Leads

0xA8E6 NO Trigger Tech High IABP

0xA8E7 NO Pressure Trigger Tech High IABP

NO Pressure Trigger
0xA8E8 Tech High IABP
Zero Transducer

9-30
Alarm ID Alarm message Category Alarm priority Source

0xA8E9 HR Low Phys Medium IABP

0xA8EA Check Pacer Timing Tech High IABP

0xA8EB ECG Detected Tech Medium IABP

0xA8EC Irregular Trigger Tech Medium IABP

0xA8ED Trigger Interference Tech High IABP

0xA8EE Rapid gas loss Tech High IABP

0xA8EF Leak in IAB circuit Tech High IABP

0xA8F0 IAB Disconnected Tech High IABP

0xA8F1 Blood detected Tech High IABP

0xA8F2 Check IAB Catheter Tech High IABP

0xA8F3 High driver pressure Tech High IABP

0xA8F4 Low Vacuum Tech High IABP

Augmentation Below
0xA8F5 Phys Medium IABP
limit set

0xA8F6 Low helium Tech Medium IABP

0xA8F7 Low Battery Tech Medium IABP

0xA8F8 System Failure Tech High IABP

0xA8F9 Electrical Test Failure Tech High IABP

0xA8FA Autofill Failure Tech High IABP

Autofill Failure - No
0xA8FB Tech High IABP
helium

0xA8FC Safety disk test failure Tech High IABP

No patient status
0xA8FD Tech High IABP
available

9-31
Alarm ID Alarm message Category Alarm priority Source

0xA8FE Autofill Required Tech High IABP

0xA8FF Low Battery(EXT) Tech Medium IABP

0xA900 Maintenance required Tech Medium IABP

Prolonged Time in
0xA901 Tech Medium IABP
Standby

0xA902 Poor Signal Persist Tech High IABP

Irregular pressure
0xA903 Tech Medium IABP
trigger

No pressure source
0xA904 Tech Medium IABP
available

0xA905 Poor Signal Quality Tech Medium IABP

Unable to update
0xA906 Tech Medium IABP
timing

0xA907 Sensor IAB Failure Tech Medium IABP

Sensor IAB Module


0xA908 Tech Medium IABP
Failure

Unable to Calibrate
0xA909 Tech Medium IABP
Sensor IAB

Sensor IAB Calibration


0xA90A Tech Medium IABP
Expired

0xA7B7 pArt-S Low Phys High PICCO

0xA7B8 pArt-S High Phys High PICCO

0xA7B9 pArt-M Low Phys High PICCO

0xA7BA pArt-M High Phys High PICCO

0xA7BB CCO Low Phys High PICCO

9-32
Alarm ID Alarm message Category Alarm priority Source

0xA7BC CCO High Phys High PICCO

0xA7BD CCI Low Phys High PICCO

0xA7BE CCI High Phys High PICCO

0xA7BF ScvO2 Low Phys High PICCO

0xA7C0 ScvO2 High Phys High PICCO

Map Drop-check
0xA7C1 Tech High PICCO
catheter

AP/HR Not
0xA7C2 Tech High PICCO
Available-please check

9-33
FOR YOUR NOTES

9-34
046-009023-00 (8.0)