You are on page 1of 15

PULSE OXIMETRY 2923

I. C. Gleadhill, A. R. Schwartz, N. Schubert, R. A. Wise, S. Per- R. C. Eberhart, S. K. Dengle, and R. M. Curtis, Mathematical and
mutt, and P. L. Smith, Upper airway collapsibility in snorers experimental methods for design and evaluation of membrane
and patients with obstructive hypopnea and apnea. Am. J. Re- oxygenators. Artif. Organs 1978; 2:19–34.
spir. Crit. Care Med. 1991; 143:1300–1303. J. B. Zwischenberger, V. J. Cardenas, W. Tao, S. C. Niranjan, J. W.
I. A. Rybak, K. A. Moxon, S. Giszter, and J. K. Chapin, Compu- Clark, and A. Bidani, Intravascular membrane oxygenation and
tational modeling of integration of voluntary/behavioral and carbondioxide removal with IVOX: can improved design and per-
automatic mechanisms for breathing control. Adv. Exp. Med. missive hypercapnia achieve adequate respiratory support during
Biol. 2001; 499:425–430. severe respiratory failure? Artif. Organs 1994; 18:833–839.
S. M. Botros and E. N. Bruce, Neural network implementation of V. Mates, J. C. Anderson, J. Hildebrandt, and M. P. Hlastala,
a three-phase model of respiratory rhythm generation. Biol. Modeling diffusion limitation of gas exchange in lungs con-
Cybern. 1990; 63:143–153. taining perfluorocarbon. J. Appl. Physiol. 1999; 86:273–284.
E. N. Bruce, Chemoreflex and vagal afferent mechanisms enhance M. Oud, Lung function interpolation by means of neural-network-
breath to breath variability of breathing. Respir. Physiol. 1997; supported analysis of respiration sounds. Med. Eng. Phys.
110:237–244. 2003; 25:309–316.
G. Longobardo, C. J. Evangelisti, and N. S. Cherniack, Effects of M. Kompis, H. Pasterkamp, and G. R. Wodicka, Acoustic imaging
neural drives on breathing in the awake state in humans. Re- of the human chest. Chest 2001; 120:1309–1321.
spir. Physiol. 2002; 129:317–333. H. A. Mansy, T. J. Royston, R. A. Balk, and R. H. Sandler, Pneumo-
G. S. Longobardo, N. S. Cherniack, and B. Gothe, Factors affecting thorax detection using pulmonary acoustic transmission mea-
respiratory system stability. Ann. Biomed. Eng. 1989; 17: surements. Med. Biol. Eng. Comput. 2002; 40:520–525.
377–396. A. A. Kamal, Spectrum analysis of respiratory sound: application
D. W. Carley and D. C. Shannon, A minimal mathematical model of to smokers and non-smokers. Front Med. Biol. Eng. 1997;
human periodic breathing. J. Appl. Physiol. 1988; 65:1400–1409. 8:165–177.
A. ElHefnawy, G. M. Saidel, E. N. Bruce, and N. S. Cherniack, E. M. Williams, R. Hamilton, L. Sutton, and C. E. Hahn, Mea-
Stability analysis of CO2 control of ventilation. J. Appl. Phy- surement of respiratory parameters by using inspired oxygen
siol. 1990; 69:498–503. sinusoidal forcing signals. J. Appl. Physiol. 1996; 81:998–1006.
K. G. Hickling, The pressure-volume curve is greatly modified by P. Dhawale and E. N. Bruce, Estimation of hypoxic ventilatory
recruitment. A mathematical model of ARDS lungs. Am. J. dynamics using pseudorandom inputs. Respir. Physiol. 1995;
Respir. Crit. Care Med. 1998; 158:194–202. 101:267–278.
C. J. Joyce and K. G. Hickling, Permissive hypercapnia and gas T. B. Martonen, Y. Yang, D. Hwang, and J. S. Fleming, Computer
exchange in lungs with high Qs/Qt: a mathematical model. Br. model of human lung morphology to complement SPECT an-
J. Anaesth. 1996; 77:678–683. alyses. Int. J. Biomed. Comput. 1995; 40:5–16.
W. C. Burke, P. S. Crooke, T. W. Marcy, A. B. Adams, and J. J. Marini, E. Denny and R. C. Schroter, A mathematical model for the mor-
Comparison of mathematical and mechanical models of pressure- phology of the pulmonary acinus. J. Biomech. Eng. 1996;
controlled ventilation. J. Appl. Physiol. 1993; 74:922–933. 118:210–215.
A. B. Adams, P. L. Bliss, and J. Hotchkiss, Effects of respiratory N. S. Cherniack and G. S. Longobardo, Periodic breathing in sleep.
impedance on the performance of bi-level pressure ventilators. In: N. A. Saunders and C. E. Sullivan, eds., Sleep and Breath-
Respir. Care 2000; 45:390–400. ing, 2nd ed. New York: Marcel Dekker, 1994.
J. J. Marini, P. S. Crooke, and J. D. Truwit, Determinants and E. N. Bruce and J. A. Daubenspeck, Mechanisms and analysis of
limits of pressure-preset ventilation: a mathematical model of ventilatory stability. In: J. A. Dempsey and A. I. Pack, eds., Reg-
pressure control. J. Appl. Physiol. 1989; 67:1081–1092. ulation of Breathing, 2nd ed. New York: Marcel Dekker, 1995.
J. Bonassa, Mathematical model for a new mode of artificial ven- J. A. Dempsey, C. A. Smith, C. A. Harms, C. Chow and K. W.
tilation: volume assisted pressure supported ventilation: a Saupe, Breathing instability. Sleep 1996; 19:236–247.
comparative study. Artif. Organs 1995; 19:256–262. M. C. K. Khoo, Periodic breathing and central apnea. In: M. Altose
P. F. Niederer, R. Leuthold, E. H. Bush, D. R. Spahn, and E. R. and Y. Kawakami, eds., Control of Breathing in Health and
Schmid, High-frequency ventilation: oscillatory dynamics. Disease. New York: Marcel Dekker, 1999.
Crit. Care Med. 1994; 22:S58–S65. M. C. K. Khoo, Physiological Control Systems. Piscataway, NJ:
G. M. Saidel and S. C. Jepson, Control and evaluation of high- IEEE Press, 2000.
frequency jet ventilation: mechanical lung model. J. Biomed.
Eng. 1990; 12:496–502.
F. T. Tehrani, A microcomputer oxygen control system for venti-
latory therapy. Ann. Biomed. Eng. 1992; 20:547–558. PULSE OXIMETRY
B. Muller, A. Hasman, and J. A. Blom, Building intelligent alarm
systems by combining mathematical models and inductive ma- YITZHAK MENDELSON
chine learning techniques. Int. J. Biomed. Comput. 1996; Worcester Polytechnic Institute
41:107–124. Worcester, MA
S. C. Niranjan, A. Bidani, F. Ghorbel, J. B. Zwischenberger, and J.
W. Clark, Theoretical study of inspiratory flow waveforms dur-
ing mechanical ventilation on pulmonary blood flow and gas
exchange. Comput. Biomed. Res. 1999; 32:355–390. 1. OXYGEN TRANSPORT IN BLOOD
D. W. Kaczka, E. P. Ingenito, S. C. Body, S. E. Duffy, S. J. Mentzer,
M. M. DeCamp, and K. R. Lutchen, Inspiratory lung impedance Oxygen is vital to the function and survival of every cell in
in COPD: effects of PEEP and immediate impact of lung volume the body so the viability of many tissues is critically de-
reduction surgery. J. Appl. Physiol. 2001; 90:1833–1841. pendent on continuous delivery of an adequate oxygen
2924 PULSE OXIMETRY

supply. Without adequate oxygen supply, irreversible ODC. In clinical applications in which it is important to
cell damage can occur rapidly. Therefore, accurate mea- detect fast-changing physiological conditions that may
surement of oxygen and other gases in blood can provide lead to hypoxemia, it is preferred to measure arterial
valuable information concerning the efficiency of the SO2 directly rather than to measure pO2 and then calcu-
circulatory and respiratory system. Various diagnostic late SO2 based on the ODC. On the other hand, when
measurements are available to assist the clinician in the hyperoxia is a clinical concern, pO2 monitoring is more
determination of patient oxygenation and to protect pa- appropriate than measurement of SO2.
tients against life-threatening and often unpredictable Traditionally, blood oxygen levels have been measured
hypoxic conditions. These methods commonly assess pul- by invasive sampling, either through an indwelling arte-
monary gas exchange, the adequacy of alveolar ventila- rial catheter or by arterial puncture, and analyzed in a
tion, blood gas transport, and tissue oxygenation. clinical laboratory with a bench-top blood gas analyzer.
Oxygen is transported in the blood in two distinct Central clinical ‘‘stat’’ laboratories can provide accurate
states. Only about 2% of O2 carried in blood is physically measurements of multiple respiratory and metabolic
dissolved in the plasma because O2 is not soluble in water. variables from small blood samples. However, this prac-
This quantity is proportional to the O2 tension or partial tice presents significant drawbacks in critical care
pressure (pO2). The quantity of dissolved O2 in blood is medicine particularly because blood gas values may
relatively small and is not sufficient to sustain normal change rapidly in certain clinical situations, particularly
body activity. Therefore, to increase the oxygen-carrying in patients with unstable respiratory or cardiopulmonary
capacity of blood, under normal physiological conditions, conditions, and the prolonged delay time between sample
nearly 98% of the O2 transported in the arterial blood is acquisition and the receipt of laboratory results. In neo-
combined with hemoglobin (Hb), the respiratory pigment natal applications, for example, frequent blood sampling
present inside the red blood cells. can cause significant blood loss especially for very small
When Hb is combined with O2 in a reversible chemical premature infants even if micro-blood samples are used
reaction, it is oxidized and forms oxyhemoglobin (HbO2). for analysis. Furthermore, blood gas values are available
Oxyhemoglobin saturation (SO2) represents the relative only intermittently and, therefore, indicate the status
concentration of HbO2 that is bound to oxygen in a given of the patient only at the time the blood samples were
quantity of blood, which is expressed as a percent of drawn. The inevitable delay and lack of continuous infor-
the total concentration of reduced Hb and HbO2. This mation can lead to potential diagnostic errors particularly
quantity is commonly referred to as functional oxygen in critically ill patients in whom rapid, and often life-
saturation and is defined as threatening, cardiopulmonary changes can occur during
short periods of time.
½HbO2  The increasing cost of health care today makes the use
Oxyhemoglobin Saturationð%Þ ¼  100: of noninvasive oxygen monitoring attractive because in-
½Hb þ ½HbO2 
stantaneous changes in blood oxygenation can be recog-
After one molecule of Hb combines with one oxygen mole- nized and adequately corrected for before irreversible
cule, the affinity of the Hb molecule to combine with more tissue damage occurs. Furthermore, continuous monitor-
oxygen molecules is greater and continues to increase as ing also provides real-time trending information to assist
more oxygen molecules are combined. This unique binding the clinician in assessing the response to therapeutic in-
property results in a sigmoid-shaped oxyhemoglobin disso- terventions without the time, risk, and high cost associ-
ciation curve (ODC) that relates the pO2 and SO2 in blood. ated with clinical laboratory analysis.
Noninvasive transcutaneous measurement of pO2 has
been available since the early 1970s (1). The principle is
2. CLINICAL MOTIVATION based on O2 diffusion through the skin and the concept of
oxygen polarography using a Clark-type electrode. Be-
Arterial pO2 and SO2, which are both good indicators to cause O2 diffusion through the stratum corneum is nor-
assess the adequacy of O2 transport in blood, are related mally too low to obtain measurable pO2 at the skin
by the ODC. However, each variable has a different phys- surface, the sensor contains a small heating element to
iological meaning. Arterial pO2 determines the efficiency induce hyperemia and speed O2 diffusion through the
of alveolar ventilation. On the other hand, arterial SO2, skin. Therefore, the position of the electrode must be
which accounts for most O2 carried by the arterial blood, is changed frequently to avoid skin burns. Additionally, be-
an important clinical parameter because it helps to assess cause polarography is based on an electrochemical reac-
how well the arterial blood is oxygenated. It should be tion, before it is applied to the skin, the sensor must first
kept in mind that pO2 or SO2 are both important in the be calibrated in vitro with a known gas mixture.
assessment of the patient’s oxygenation status. Although The principal application of transcutaneous pO2 moni-
it is possible to determine SO2 from pO2 or vice versa toring has been limited to newborn infants because of
based on the ODC, large errors may result unless the pH, the relatively thick layer of the epidermis at birth. In
temperature, the partial pressure of CO2 (pCO2), and the adults, the greater skin thickness prevents accurate and
2,3-diphosphoglycerate (2,3-DPG) level in the blood are reproducible measurement of arterial pO2. Besides this dif-
also measured simultaneously. All of the above factors ference and other practical concerns as noted above, trans-
play an important role in controlling the affinity of Hb for cutaneous pO2 measurements are slower and more
O2 and thus can lead to considerable displacement of the complicated compared with pulse oximetry. For example,
PULSE OXIMETRY 2925

10 by a substance can be related to its concentration using


the basic relationship

It ¼ Io  10ecd ;
Extinction coefficient

1 MetHb
where It is the intensity of the transmitted light, Io is the
HbO2
intensity of the incident light, e is the extinction coefficient
(or molar absorptivity, which represent the fraction of
0.1 Hb light absorbed at a specific wavelength) of the sample,
c is the concentration of the sample, and d corresponds to
the light path length through the sample. This relation-
HbCO ship requires monochromatic (i.e., single wavelength) in-
cident light and collimated (i.e., parallel beams) light
0.01
600 700 800 900 1000 transmission through the sample. In other words, the
Wavelength (nm) sample does not scatter any light. In addition, it is as-
sumed that the incident light intensity is equal to the sum
Figure 1. Optical absorption spectra of Hb, HbO2, MetHb, and
of the light intensities transmitted and absorbed by the
HbCO in the visible and near-infrared regions of the spectrum.
sample. It must be noted also that if the incident light is
scatters by the sample, the effective pathlength of the
after a transcutaneous pO2 electrode is applied to the skin, transmitted light through the sample is increased, and
it takes about 5–10 minutes to establish a stable and accu- thus, the Beer–Lambert law is no longer valid.
rate reading. For these reasons, transcutaneous pO2 mea- Assuming for simplicity that a thoroughly mixed hem-
surements remained popular in neonatal applications until olyzed blood sample (i.e., blood in which the red blood cell
the early 1980s when pulse oximetry was first introduced membrane has been ruptured and destroyed) consists of a
into clinical practice. Soon after its introduction, the two-component homogeneous mixture composed of only Hb
popularity of pulse oximetry rose rapidly, whereas the pop- and HbO2, the light absorption by this homogenous mixture
ularity of transcutaneous pO2 monitoring has gradually is equal to the sum of light absorption by Hb and HbO2.
declined. Therefore, it is possible to derive two independent algebraic
equations to describe the absorption of the mixture using
two distinct wavelengths. This set of equations can be then
3. PRINCIPLE OF BLOOD OXIMETRY be solved simultaneously to find the concentration of each
component in the mixture. Using the two equations de-
3.1. Optical Spectra of Blood scribed above, SO2 can be determined from the relationship
The basic concept of oxygen saturation measurement (ox-
imetry) dates back to Robert Boyle who, in 1700, first ob- ODðl1 Þ
SO2 ¼ A  B ;
served that blood can change its color when exposed to air. ODðl2 Þ
Early methods that were developed to estimate the rela-
tive amount of O2 in blood are based on the observation where A and B are functions of the specific extinction coef-
that the characteristic color of deoxygenated Hb is dark ficients of Hb and HbO2 and OD is defined as the optical
red, whereas when Hb is chemically combined with O2 to density [i.e., log10(Io/It)] of the substance. In other words,
form HbO2 has a distinctly bright red color. SO2 can be determined by measuring the incident and
Optical oximetry relies on the significant difference in transmitted light intensities of a homogenous sample solu-
the optical absorption of Hb and HbO2 in the visible part of tion containing an unknown concentration of Hb and HbO2.
the spectrum, as illustrated in Fig. 1. Optical methods for This concept is the underlying principle used by conven-
measuring SO2 are based on light absorption by blood at tional bench-top clinical CO-oximeters to measure SO2 and
two specific wavelengths: l1, where there is a large differ- other hemoglobin derivatives in blood.
ence in light absorption between Hb and HbO2 (e.g.,
660 nm); and a second wavelength, l2, which can be an
4. HISTORY OF NONINVASIVE OXIMETRY
isobestic wavelength (a wavelength at which the absorp-
tion of Hb and HbO2 are equal, e.g., 805 nm) or a different
Since the early 1930s, considerable efforts have been made
wavelength in the near infrared region of the spectrum
to develop noninvasive optical techniques for accurate
between 880 and 940 nm, where the absorption of Hb is
measurements of arterial oxygen saturation (2). Matthes
slightly smaller compared with HbO2.
(3,4) developed the first noninvasive oximeter using two
wavelengths of light. However, the need to compensate for
3.2. Calculation of Oxygen Saturation
large changes in tissue thickness, unknown blood content,
The absorption of light at a specific wavelength by a ho- and variations in skin color among different individuals
mogeneous substance in a solution contained inside a par- was recognized early on as a practical obstacle in achiev-
allel-wall optically transparent holder (optical cuvette) ing accurate calibration.
can be accurately determined with the Beer–Lambert Early application of the Beer–Lambert law to noninva-
law. Accordingly, the fraction of radiant energy absorbed sive measurement of arterial SO2 began by Millikan et al. (5)
2926 PULSE OXIMETRY

and Wood and Geraci (6) during World War II because of the thickness and skin pigmentation. Thus, to determine ar-
need to monitor and protect pilots flying at high altitudes in terial SO2, a computer inside the oximeter had to solve
unpressurized cockpits from losing consciousness. This de- eight algebraic equations simultaneously. These equa-
vice essentially used the earlobe as an optical cuvette tions, which were originally derived based on Beer–Lam-
containing blood. However, the ear tissue contains non- bert law, are similar to the relationship describing the
hemolyzed whole blood, and hence, the transilluminated optical absorption of a two-component mixture. However,
light undergoes partial reflection and multiple scattering by because the exact optical extinction coefficients of the dif-
tissue and blood. Furthermore, besides arterial blood, the ferent light absorbers in the tissue including light scat-
ear lobe also contains venous blood and other nonblood sub- tering by the ear pinna were unknown, a brute force
stances that interact with the light. Consequently, initial approach was taken by empirically deriving a set of char-
attempts to measure SO2 noninvasively required a zero- acteristic extinction coefficients based on data obtained
point calibration by subtracting the light attenuated by the from human volunteers.
bloodless tissue and skin from the total amount of light Although the H-P ear oximeter became popular within
transmitted through the ear. The zero-point calibration was pulmonary medicine, its major disadvantages were the
accomplished by compressing the ear using an ear probe need to heat the skin to arterialize the blood, the heavy
equipped with a transparent pressure capsule that was ini- weight of the fiber-optic cable, and the large size of the ear
tially inflated to a pressure that exceeds the arterial blood probe (approximately 10  10 cm), which limited the ap-
pressure, thus rendering the ear pinna almost bloodless. plication of this oximeter to adult patients. Manufacturing
After the initial baseline transmission of light through the of the H-P ear oximeter, which was the prominent fore-
bloodless ear was measured, the pressure in the capsule was runner to the pulse oximeter, stopped in the early 1980s.
released and a second measurement through the uncom-
pressed ear was taken. SO2 was then determined based on
the difference between these two readings. Although this 5. THEORY OF PULSE OXIMETRY
approach was the first significant step toward an accurate
and absolute noninvasive measurement of SO2, it was not Noninvasive monitoring of arterial oxygen saturation by
successful mainly because of poor reproducibility and the pulse oximetry (SpO2) was first adopted by anesthesiolo-
complicated calibration process involved. The other techni- gists in the early 1980s in an effort to optimize patient
cal challenge that contributed to significant errors was the safety. The goal of this technology was to identify as early
inability to distinguish between the light absorbed by arte- as possible unrecognized episodes of hypoxemia (insuffi-
rial and venous blood because oxygen saturation of venous cient blood oxygenation) that often resulted in irreversible
blood is lower than arterial blood and depends on local phys- tissue damage and costly malpractice insurance claims.
iological conditions. During the past decade, pulse oximetry has become a rap-
The first widely used commercial ear oximeter, which idly growing practice in many fields of clinical medicine,
was considered the forerunner of pulse oximetry, was de- including subacute and long-term care. Whether used as a
veloped by the Hewlett-Packard (H-P) company in the safety guard or as a diagnostic tool, this new modality is
early 1970s (7). Briefly, a high-intensity tungsten lamp widely acknowledged to be one of the most important
was used to generate a broad spectrum of light. Eight technological advances in patient monitoring. The most
narrow-band optical interference filters were mounted on important advantage of pulse oximetry is the capability to
a rotating wheel that intercepted the light path sequen- provide continuous, safe, and cost-effective monitoring of
tially to provide a source of monochromatic wavelength blood oxygenation noninvasively at the patient bedside. It
selection. These filtered light beam pulses entered a fiber- is particularly useful when a patient is unstable and sub-
optic cable that carried it to the ear. A second fiber-optic ject to rapid or unpredictable oxygen desaturation. Pulse
cable carried the light pulses transmitted across the ear- oximeters are easy to use, require no user calibration, and
lobe back to the instrument for detection and analysis. To provide virtually maintenance-free operation.
measure arterial SO2, the ear probe was positioned across The history and principle of pulse oximetry has been
the ear pinna and a temperature-controlled heater within reviewed extensively in several books and manuscripts
the ear probe maintained the tissue at a temperature of (8–15). This breakthrough idea came about in the 1970s
411C, which caused an increase in local blood flow through when Takuo Aoyagi was researching how to measure car-
the ear that essentially ‘‘arterializes’’ the blood. diac output noninvasively based on a dye dilution washout
The H-P ear oximeter measured arterial SO2 by com- curve using light transmission through the ear based on
paring the intensity of light transmitted through the ear Wood and Geraci’s method (16–18). While performing ex-
at eight selective wavelengths within the 650–1050 nm periments on animals, he noted that arterial pulsations
spectral region. These wavelengths were selected because caused fluctuations in light transmission that interfered
the light in this spectral range is least attenuated by tis- with measurements of changes in dye concentration over
sues. Wavelengths outside of this range are heavily ab- time. These arterial pulsations made it difficult to achieve
sorbed either by blood or by water. The design of the accurate cardiac output measurements. Realizing that the
H-P ear oximeter assumed that the ear pinna acts as a minute pulsations (i.e., modulation) in the transmitted
fixed path length optical cuvette. The method then as- light are sensitive to SO2, and therefore can neatly cir-
sumed that a homogenous mixture of eight independent cumvent many problems of earlier nonpulsatile oximetry,
light absorbers can adequately represent the bulk optical Aoyagi’s discovery quickly spurred the commercialization
properties of the ear pinna, including variations in tissue of the first clinically useful noninvasive pulse oximeter.
PULSE OXIMETRY 2927

nonpulsatile (direct current, or DC) component of the


Arterial pulse PPG that is characteristic of the background light ab-
sorbed by the tissue, the residual nonpulsatile arterial
Residual arterial blood present in the vascular bed during diastole, and ve-
blood nous blood. This effective scaling yields a normalized AC/
Light absorption

Venous blood DC ratio for both the red and the infrared PPGs. The ratio
of these two normalized AC/DC ratios, which is largely
independent of the incident light intensity, is denoted by
R and is equal to
Bloodless tissue ACR =DCR
R¼ ;
ACIR =DCIR

where ACR is the pulsatile red component, ACIR is the


pulsatile infrared component, DCR is the nonpulsatile red
Time component, and DCIR is the nonpulsatile infrared compo-
nent of the corresponding PPGs. It is this adjusted R ratio
Figure 2. Variations in light attenuation by tissue illustrating
that the pulse oximeter uses to compute SpO2.
the rhythmic effect of arterial pulsation. The arterial pulse cor-
responds to the AC component of the PPG waveform. The DC
component corresponds to the total absorption by the residual 5.1. Calibration of Pulse Oximeters
arterial blood present in the vascular bed during diastole, the ve- Precalibration of a pulse oximeter by the manufacturer
nous blood and bloodless tissue components.
must be performed empirically by calculating the ratios of
the normalized AC/DC signals obtained from the red and
infrared PPGs during a stepwise hypoxic study. The cal-
Pulse oximetry exploits the time-variant photo- ibration data are normally acquired from 10–20 healthy
plethysmographic (PPG) signal that is produced by young adult volunteers breathing hypoxic gas mixtures to
changes in arterial blood volume and changes in the ori- induce hypoxemia. During the study, data measured by
entations of red blood cell associated with cardiac contrac- the pulse oximeter are compared simultaneously with
tion and relaxation, as illustrated in Fig. 2. The technique SaO2 values obtained from an indwelling radial artery
also assumes that venous blood does not pulsate and that catheter and analyzed in vitro by a standard multiwave-
the amplitude of the arterial PPG pulse measured around length CO-oximeter. The data collected are later used to
660 nm is sensitive to changes in arterial SO2. Taking determine specific calibration coefficients and create a
these assumptions into consideration, SpO2 can be derived look-up table that is permanently programmed into the
by analyzing only the changes in the amplitude of the oximeter for converting optical measurements to SpO2
pulsatile red and infrared PPGs measured in the visible values. This method of calibration can be lengthy and
and near-infrared region of the spectrum. costly, but it must be conducted by each pulse oximeter
The unique advantage of this PPG-based noninvasive manufacturer to gain U.S. Food and Drug Administration
approach is that only two wavelengths are required to de- (FDA) approval.
termine SpO2, which greatly simplifies the configuration Because of this empirical calibration, different brands of
of the optical sensor. Furthermore, the pulsatile portion of pulse oximeters may display different values depending on
the PPG signal, which is related to the inflow of arterial the internal calibration of the oximeter. A typical calibra-
blood into the cutaneous vascular tissue, is used to sepa- tion relationship between the normalized R ratio and the
rate the arterial blood from all other absorbing substances arterial SpO2 is shown in Fig. 3. This calibration can be
present in the illuminating path. Thus, light absorbed by described for example by the mathematical relationship (19)
the skin, which depends on skin color and the composition
of subcutaneous tissues, as well as on the optical absorp- k1 þ k2 R
SpO2 ¼ ;
tion by venous blood, remains constant during the cardiac k3 þ k4 R
cycle and therefore can be ignored. Additionally, rendering
the tissue bloodless during the calibration process and the or the data can be fitted by a polynomial relationship of the
need to heat the tissue to ‘‘arterialize’’ the blood, which form (9)
were essential in earlier models of nonpulsatile oximeters,
are no longer necessary. SpO2 ¼ AR2 þ BR þ C:
Practically, tissue absorption can vary widely between
individuals and probe locations. Furthermore, transmit- The values of the different coefficients in these equations
ted light intensities depend on detector sensitivity and (k1–k4 and A, B, C) are determined by empirical calibration.
incident light intensities. Therefore, a normalization (i.e., Note that at about 85% SpO2, the amount of light absorbed
scaling) process is commonly employed in which the pulsa- by the Hb and HbO2 is relatively equal and, therefore, the
tile (alternating current, or AC) component of the red and normalized amplitudes of the red and infrared signals are
infrared PPGs measured by a photodetector, which results equal (i.e., R ¼ 1:0). Because the optical properties of blood
from the influx of arterial blood and expansion of the cu- are similar among different individuals, individual instru-
taneous vascular bed, is divided by the corresponding ment calibration is not required in the field.
2928 PULSE OXIMETRY

100 was based on a bulky fiber-optic cable that was used as a


90 split optical guide to conduct light from a quartz halogen
lamp to the measurement site and to conduct the light
80
transmitted through the finger back to a photodetector
% Oxygen saturation

70 (PD). The light source and PD were both housed inside the
60 oximeter. Narrow optical band pass filters in combination
with a mechanical chopper were used to properly select
50
the red and infrared wavelengths and to synchronize the
40 detection of the transmitted light with the measurements
30 made by the PD.
20 An improved design of a noninvasive pulse oximeter
probe was introduced in the United States by BIOX and
10
Nellcor in the early 1980s based on a much simpler, more
0 compact, and significantly less expensive design that elim-
0 1.0 2.0 3.0 4.0 5.0
inated the need for a fiber-optic cable and other bulky op-
Normalized R ratio
tical components. It included readily available solid-state
Figure 3. Empirical relationship between arterial oxygen satu- components that also allowed manufacturers to signifi-
ration and the normalized R ratio. cantly reduce the overall size and weight of the probe. A
typical pulse oximeter probe comprises of a single highly
sensitive silicon PD and a pair of small, but bright, red and
The decision of what should be considered a ‘‘safe’’ low- infrared light emitting diodes (LEDs) mounted inside a
level SO2 for calibrating pulse oximeters on healthy vol- small and light-weight probe. Additionally, using LEDs
unteers remains controversial. Typically, most manufac- reduced significantly the amount of power consumption
turers perform in vivo calibrations by varying the inspired and eliminated the amount of heat generated by the in-
O2 concentrations to yield arterial SO2 values between candescent light source in previous designs.
70% and 100%. However, pulse oximeters are capable of Most manufacturers of pulse oximeters offer different
measuring SpO2 values lower than 70%. Occasionally, types of disposable and reusable sensors that are suitable
even during clinical calibration studies designed to en- for neonatal, pediatrics, and adult applications. In trans-
sure that SO2 levels remain above 70%, a few arterial SO2 mission-type sensors, the LEDs and PD are placed on op-
values can fall below 70%, although the number of data posite sides across a pulsating arterial bed. The most
points in this range is generally too small to be included in popular sensor configurations are flexible bandage-type
the calibration data set. Therefore, manufacturers typi- disposable sensors (Fig. 4) or reusable finger clip-on type
cally prefer to extrapolate the calibration curves to deter- probes (Fig. 5). Some manufacturers also offer earlobe clip
mine SpO2 values below 70%. and nose-bridge type sensors.
In general, the accuracy of most noninvasive pulse oxi- In certain clinical situations in which transmission
meters is acceptable for a wide range of clinical applica- sensors cannot be used, it is possible to use reflectance-
tions. Under normal operating conditions, the accuracy of type sensors. The feasibility to use reflected light rather
pulse oximeters compared with a CO-oximeter is typically than transillumination to measure SpO2 was first intro-
within þ /  2–3% in the arterial SO2 range between 70% duced by Mendelson et al. in the early 1980s (25,26). Re-
and 100% (20–24). Below 70%, the accuracy normally de- flectance sensors use similar optical components as
teriorates considerably. transmission sensors except that the LEDs and PD are
mounted adjacent to one another, as illustrated in Fig. 6.
5.2. Pulse Oximeter Sensors The most compelling reason for developing reflectance-
The sensor of the original pulse oximeter as described by type sensors is the ability to measure SpO2 from alterna-
Yoshiya et al. and manufactured by Minolta in Japan (18) tive body locations such as the head. For instance, a re-
flectance sensor can be used in perinatology to monitor a
fetus during delivery. Forehead reflectance sensors have
also been advocated to guard against acute changes in
hypoxemia, particularly in patients with poor peripheral
circulation, because it can be manifested in the head more
rapidly than in the extremities (27–30). Clinical studies

(a) (b)
Figure 4. A typical disposable transmission-type finger sensor of Figure 5. A typical reusable transmittance-type pulse oximeter
a pulse oximeter (a) taped to a subject’s finger (b). probe.
PULSE OXIMETRY 2929

sensors employ a single PD element, typically with an ac-


tive area of about 12–15 mm2. Normally, a relatively small
PD chip is adequate for measuring strong transmission
PPGs because most of the light emitted from the LEDs is
diffused by the skin and subcutaneous tissues predomi-
Figure 6. A typical reflectance-type pulse oximeter probe.
nantly in a forward-scattering direction. However, in re-
flection mode, only a small fraction of the incident light is
backscattered by the tissues. Additionally, the backscat-
found that the lag time to detect hypoxemia in the finger- tered light intensity reaching the skin surface is normally
tips during peripheral vasoconstriction can be longer by distributed over an area larger than 15 mm2. Typically,
about 90 seconds compared with the forehead. These stud- the backscattered light is spread concentrically around
ies also confirmed that transmission sensors mounted on the LEDs, but its intensity is inversely proportional to the
the fingertip are the slowest to respond to changes in SpO2 square distance from the location of the LEDs.
because the fingertips are more affected by thermoregula- To improve the signal-to-noise ratio (SNR) and reliabil-
tory vasoconstriction than the forehead. ity of reflection pulse oximeters, several sensor geometries
Pulse oximeters store the unique calibration curves that have been described based on a radial arrangement of a
are used to convert optical readings to SpO2 values inside larger area PD or multiple LEDs. For example, Mendelson
the monitor. This approach can limit the manufacturer’s et al. (34–36) and Konig et al. (37) designed a reflectance
ability to introduce new sensor designs that might require sensor prototype consisting of multiple discrete PDs or a
different calibration choices. To overcome this limitation, a single annular-shaped PD mounted symmetrically around
new line of pulse oximeter probes (OxiMax) that includes a a pair of red and infrared LEDs. Takatani et al. (38,39)
digital memory chip was recently introduced by Nellcor described a different sensor configuration to increase
(31). The memory chip is imbedded within each sensor and the amount of incident light based on ten LEDs that
contains all calibration and operating characteristics for were arranged symmetrically around a single PD chip.
that individual probe, which enables the monitor to oper-
ate accurately with a more diverse range of sensors. 5.3. Sensor Attachment
Pulse oximetry is typically used in hospitals, although
Several locations on the body, such as the fingertips and
several products and research efforts are focused on devel-
earlobes, are suitable for monitoring SpO2 with transmis-
oping more portable and wearable pulse oximeters with
sion sensors in adults. The most popular sites in adults are
unique sensor configurations. Some efforts are also di-
the fingertips because these locations are convenient to
rected toward the development of units that are more suit-
use and a good PPG signal can be quickly obtained. On
able for long-term ambulatory applications. Nonin Medical
neonates and pediatric patients, the toes, ankles, or wrists
(Plymouth, MN) managed to condense and integrate the
can be used as alternative monitoring sites. Because light
electronics of a pulse oximeter together with a transmis-
is highly scattered in tissues, other locations on the skin
sion-type sensor into a unique finger clip probe, creating
that are not accessible to conventional transillumination,
the smallest self-contained finger pulse oximeter called the
such as the forehead or the temple region, can be moni-
Onyx. The Onyx pulse oximeter weighs only 2 ounces,
tored using reflectance sensors. These sensors are typi-
measures 1.3  1.3  2.2 inches, and is powered by two
cally attached to the skin using a double-sided adhesive
AAA batteries providing continuous use for up to 18 hours.
tape. A headband can be used to hold the reflectance sen-
Rhee et al. (32,33) developed a miniaturized wireless
sor in place and to minimize ambient light interferences.
pulse oximeter sensor worn by the subject as a finger base
ring. The isolated ring configuration separates the sensor
5.4. Instrumentation
unit from the rest of the ring body that is much heavier
than the optical sensor unit alone. The separation is Hardware implementation of pulse oximeters varies
achieved by using two concentric rings that are mechan- widely among different manufacturers. In general, a mi-
ically decoupled. The inner ring contains a pair of red and croprocessor is used to generate a sequence of digital time-
infrared LEDs and a PD configured as a reflectance mode multiplexing pulses to synchronously drive the red and
sensor, whereas the outer ring comprises the rest of the infrared LEDs such that only one LED is turned on at a
pulse oximeter circuitry including a battery and a radio- time. Typically, this is accomplished by sequencing the red
frequency transmitter. A thin and flexible cable connects and infrared LED on and off, which allows an interval
the two rings. Forces from mechanical contacts with the when both LEDs are turned off to detect changes in am-
outer ring are therefore isolated from the optical sensor. bient background light. These digital switching pulses are
This efficient double-ring design has the potential of min- also used to demultiplex the red and infrared signals be-
imizing the influence of external forces and therefore of cause light is detected by a single PD. Narrow-width
reducing the effects of motion artifacts. switching pulses also help to increase the peak illuminat-
The design of a reflectance-mode pulse oximeter de- ing intensity of the LEDs without the risk of exceeding the
pends on the ability to fabricate a sensor that has im- maximum rated average power dissipation that can dam-
proved sensitivity and can detect sufficiently strong PPGs age the LEDs. Depending on the particular design, addi-
from various locations on the body combined with sophis- tional sample-and-hold circuits are sometimes used to
ticated digital signal algorithms to process the relatively separate the composite signals measured by the PD into
weak and often noisy signals. Commercial pulse oximeter two different analog channels corresponding to the red
2930 PULSE OXIMETRY

and infrared PPGs. The demultiplexed PPG signals can the AC and DC components of the PPG waveforms. Addi-
then be amplified and high-pass filtered individually to tionally, dedicated software algorithms are commonly
separate the AC and DC components of each PPG wave- employed to implement digital filtering, improve the
form before it is further digitized by an analog-to-digital SNR by averaging multiple measurements for AGC, dis-
converter (ADC). In some pulse oximeters, the composite play different diagnostic or warning information to keep
PPG measured by the PD is digitized directly by an ADC, the user informed, and implement various user-selected
which considerably simplifies the amount of preprocessing SpO2 and HR alarm settings.
hardware used, although this approach requires a higher
bit resolution ADC (typically 14–16 bits or more) to
5.5. Signal Processing
achieve adequate SNR. Some manufacturers use auto-
matic gain control (AGC) either to adjust the light inten- There are no standard methods for processing the PPG
sity of each LED or to change the gain of the photodiode waveforms and improving the SNR in pulse oximetry. Sig-
amplification. This process is performed to ensure that the nal processing algorithms used by manufacturers for com-
AC and DC levels of the PPGs signals always remain puting and relating the R ratios to SpO2 and for
within a predetermined range. This feature is important measuring heart rate are proprietary information and
to maximize the amplitude of the analog signal before it is therefore protected by patent laws. Review of the patent
digitized by the ADC and to compensate for variations in literature can reveal some signal processing algorithms
skin color and tissue properties. used by manufacturers. Typically, these algorithms range
Besides SpO2, most pulse oximeters also provide sev- from simple digital filtering and time-domain analysis
eral other displays. Most pulse oximeters, other than based on a weighted moving average technique to more
handheld devices typically used for spot-checking, com- sophisticated frequency-domain processing, including
monly include analog or digital bar graph-type displays nonlinear filtering, autocorrelation, adaptive filtering,
in the form of a horizontally scrolling arterial waveform and feature extractions, to name a few. Rusch et al. (40)
resembling a blood pressure pulse, or a vertical moving compared several digital signal processing algorithms
bar corresponding to the relative amplitude of the arterial based on frequency-domain analysis and found that
pulse (Fig. 7). These important features enable the clini- computation of SpO2 by fast fourier transform (FFT) or
cian to assess in real time the quality and reliability of the discrete cosine transform were as accurate as weighted
measurement. For example, the amplitude, shape, and moving average algorithms.
stability of the arterial waveform can often be used as an Because pulse oximeters use the pulsatile signal of the
indication of motion artifacts or adequate perfusion con- PPG, several time-domain approaches can be used to de-
ditions when erratic SpO2 or HR values are displayed, rive the R ratios and compute the patient’s HR. Addition-
particularly in monitoring compromised patients. Simi- ally, the response time of a pulse oximeter depends on the
larly, if the patient’s heart rate displayed by the pulse oxi- number of data points averaged before the readings are
meter differs considerably from the actual heart rate, displayed. Some methods use the peak-to-trough ampli-
which often is available as an independent measurement tudes of the pulsatile red and infrared components to de-
if the patient is also connected to an electrocardiogram termine the AC components at each wavelength. However,
(ECG) monitor, the SpO2 values displayed by the pulse this simplified approach has a deleterious effect on the
oximeter should be questioned. time response of the pulse oximeter to acute changes in
The software used in pulse oximeters typically per- SpO2 because the readings depend on the patient’s heart
forms several important functions. Depending on the rate, and more importantly, new raw signals are available
hardware features implemented in a particular pulse oxi- for averaging only once every heart beat. An improved
meter model, the software usually controls the sequencing approach is based on averaging many incremental
of the LEDs and may perform digital filtering to separate changes along the steep slope of the PPG pulse. This ap-
proach results in many more data points between succes-
sive heart beats and, consequently, leads to a shorter
instrument response time. In addition, averaging more
data points improves the accuracy and stability of the
values displayed. Because the AC component of each PPG
signal is a bipolar waveform, in the time domain, it is rel-
atively simple to measure the duration of each heart beat
using a simple zero-crossing algorithm.
An alternative approach for processing the PPG signals
is based on a frequency-domain approach. Accordingly, the
composite red and infrared PPG signals (before the DC
and AC components are separated) are first processed by
an FFT algorithm. The FFT spectra for each PPG signal
are then processed by a separate algorithm to identify
the amplitude and frequency of the dominant peaks. Typ-
Figure 7. The Nellcor Model N595 pulse oximeter (Courtesy of ically, the amplitude of the zero-frequency peak corre-
Nellcor Puritan Bennett, Inc. Pleasanton, CA). (This figure is avail- sponds to the value of the DC component. Likewise, the
able in full color at http://www.mrw.interscience.wiley.com/ebe.) amplitude of the highest spectral peak residing between
PULSE OXIMETRY 2931

1 and 3 Hz (i.e., 60 to 180 beats per minute), which coin- achieved with an additional (i.e., third) measurement
cides with the subject’s HR, corresponds to the amplitude channel with an illuminating wavelength that exhibits
of the AC component. contrasting anatomical absorption characteristics. Co-
Several methods are typically employed to improve the etzee and Elghazzawi (47) described a different noise re-
accuracy of pulse oximeters, especially in clinical situa- duction algorithm that first detects relatively clean PPG
tions involving motion artifacts. The primary motivation sections from which the HR is estimated. Then, the HR
is the reduction in the occurrence of false alarms. The most value is used to construct a synthetic reference signal that
common method to reduce the effect of motion artifacts is matches an idealized pulse signal. An adaptive filter con-
by digital signal processing and averaging of consecutive tinuously processes the sensor signals, reconstructing sig-
measurements over a fixed number of seconds before the nals in a linear subspace defined by the reference signal. A
final reading is displayed. However, this approach can projective subspace algorithm is then applied to find SpO2.
slow the response time of the pulse oximeter to alert cli-
nicians when abrupt changes in hypoxemia occur. There-
fore, many brands permit the user to select the averaging 6. PERFORMANCE LIMITATIONS OF PULSE OXIMETRY
time depending on the user needs.
Some manufacturers used the R-wave of the patient’s Certain clinical situations and confounding factors can in-
ECG to synchronize with the optical measurements, terfere with the proper acquisition of reliable data and
which thereby improved the detection of noisy pulsatile the interpretation of pulse oximeter readings. The most
signals by enhancing the SNR using multiple time-aver- common cause of interruptions in the continuous readings
aged signals (41,42). The disadvantage of this approach is of a pulse oximeter (called ‘‘dropouts’’) seems to be from
that it relies on the availability of a separate ECG monitor. motion artifacts and low peripheral blood perfusion. Cli-
Other techniques use a priori assumptions about specific nicians must understand these shortcomings so that er-
feature recognition in a typical PPG waveform and simply roneous readings can be properly identified.
suppress the readings when excessive motion artifacts
corrupt the basic template features of the detected 6.1. Low Peripheral Vascular Perfusion
PPGs. However, because even small artifacts can severely Because pulse oximeters rely on adequate arterial pulsa-
distort the temporal features of the PPG, this approach tion, significant reduction in peripheral vascular pulsa-
tends to discard potentially useful data, and timeouts tion, most notably from systemic hypotension, local
are common. vasoconstriction, hypothermia, hypovolemia, and during
In 1989, the Masimo Corporation introduced a different cardiopulmonary bypass, can produce a signal too small to
signal extraction technology based on an adaptive filtering be processed reliably by a pulse oximeter (48–51). In cer-
approach (43,44). This approach, which is used to improve tain situations, locally applied vasodilating agents or gen-
the SNR in signals corrupted with overlapping frequency tle warming of the skin by covering the sensor with a cloth
noise bands, recognizes that during patient motion, in ad- can be used to enhance the pulsatile signal. In general,
dition to a noisy arterial pulse, the venous blood compo- however, pulse oximeters become more susceptible to
nent is also susceptible to perturbations caused by motion artifacts under conditions of low peripheral perfu-
dynamic redistribution attributed to gravity or external sion. When the oximeter cannot detect adequate pulsatile
pressure and therefore represents a significant source of signals, a ‘‘low perfusion’’ or similar type of diagnostic
interdependent noise. Because a pulse oximeter assumes message is usually displayed by the oximeter to alert the
that the AC component of the PPG is caused by arterial user of possible peripheral blood perfusion problems.
pulsation, venous perturbations during movement can
cause significant errors and lead to false alarms. The
6.2. Motion Artifacts
method attempts to estimate the magnitude of this noise
and cancel its contribution from the PPG during patient Pulse oximeters must detect a pulsatile signal that is nor-
movement by an adaptive noise cancellation approach. mally a very small fraction (typically about 1–3%) of the
The Masimo approach, which is based on a proprietary nonpulsatile component in the PPG. Therefore, any tran-
statistical signal processing algorithm termed Discrete sient motion of the sensor relative to the skin that is usu-
Saturation Transform (DST), builds a reference signal ally accompanied by changes in the optical coupling
from the incoming red and infrared PPGs for each between the probe and the illuminated tissue can cause
percent of SpO2 value. This reference is passed through significant artifacts. Because these transient artifacts can
the adaptive filter to cancel the correlated frequencies be- sometimes mimic a normal heart beat, erroneous readings
tween the reference and the corrupted PPG signal. may be displayed if the instrument cannot differentiate
Additional signal processing methods to improve the between pulsations that are from motion artifacts and
performance of pulse oximeters by reducing the sensitivity normal arterial pulsations.
to motion artifacts are being developed by the manufac- Pulse oximeters were originally used by anesthesiolo-
turers and independent investigators. Hayes and Smith gists in the operating room on motionless patients. Soon
(45,46) proposed that a nonlinear PPG artifact reduction after, this technology was adopted for use in the intensive
method can be used to significantly reduce the effect of care unit and general wards, where patients can be free to
changes in the probe coupling, with an electronic process- move around or may become restless. In these circum-
ing algorithm based on inversion of a physical artifact stances, the tolerance of pulse oximeters to motion arti-
model. Deterministic removal of the modeled artifact is facts emerged as one of the most challenging technical
2932 PULSE OXIMETRY

problems to solve. In fact, motion artifacts have been cited the wavelengths common in pulse oximetry. Several re-
by clinicians as the most common cause of false alarms, ports suggest that the readings of pulse oximeters may be
loss of signal, and inaccurate readings in a pulse oximeter. affected by high-frequency radio interference from ESUs
The high frequency of false alarms associated with motion and direct exposure to high-intensity light sources such as
artifacts often leads to unsettling tendencies among clini- infrared, xenon, and fluorescent lamps (57–59), although a
cians, especially in a busy hospital ward, to ignore or even controlled study conducted by Fluck et al. (60) concluded
turn off pulse oximeter alarms and thereby compromise that the effect of ambient light on the accuracy of pulse
patient safety. oximeters is statistically and clinically insignificant. Some
Motion artifacts, such as during shivering or seizure manufacturers of pulse oximeters try to minimize the ef-
activity, are usually recognized by false or erratic heart fect of stray light by taking intermittent optical back-
rate displays or by distorted PPG waveforms. Several clin- ground readings when the LEDs in the sensor are turned
ical studies compared how different brands of pulse oxi- off and subtracting these readings from measurements
meters respond to motion artifacts under various taken by the PD when the two LEDs are switched on. Ad-
conditions (52–54). It is important to keep in mind, how- ditionally, room light flickering from fluorescent lights can
ever, that the interpretations of clinical comparison stud- be greatly reduced by sequencing the LEDs at a switching
ies to evaluate the performance of different pulse frequency that is an integer multiple of the power line
oximeters can depend on the pulse oximeter model used frequency. Wrapping the sensor with an opaque material
and the particular revision of the software algorithms. can continue to reduce these light interferences.

6.3. Venous Pulsations 6.6. Interpretation of Pulse Oximeter Readings


Another potential problem with pulse oximeter measure- When comparing pulse oximeter readings to standard in
ments are artifacts induced by venous pulsations. vitro CO-oximetry, which are widely used as a reference
As the veins become engorged with blood, the venous for calibrating pulse oximeters, it is important to under-
blood pulsates at the same frequency as the arterial blood. stand the principal differences between these two mea-
Because pulse oximeters rely on the presence of a pulsatile surements. For example, some bench-top CO-oximeters
signal, and SO2 of venous blood is lower than arterial use four distinct wavelengths between 500 and 600 nm to
blood, some pulsatile red component in the PPG signal measure total Hb, HbO2, HbCO, and methemoglobinemia
may actually contain venous blood of lower SO2 mixed (MetHb) in a given sample of blood. Hence, the value of
with the higher SO2 normally present in the arterial HbO2 reported by a CO-oximeter is commonly referred to
blood. Therefore, increased venous pulsations interfere as fractional oxyhemoglobin. Numerically, it is equal to
with the measurements, which results in artificially low the concentration of HbO2 expressed as a percentage of
SpO2 readings. the total of all active and inactive forms of Hb with respect
Sometimes, if sensors are applied too snugly in an at- to their O2 binding capability. Therefore, it is normally
tempt to reduce motion artifacts, they can lead to venous defined using the relationship
pulsation and even necrosis. Other situations leading to
venous pulsations may be attributed, for example, to right ½HbO2 
heart failure, tricuspid regurgitation, high airway pres- HbO2 ¼  100%:
½Hb þ ½HbO2  þ ½HbCO þ ½MetHb
sures, VALSALVA maneuvers, and changes in body posi-
tion such as during certain surgical procedures when the In contrast to a CO-oximeter, pulse oximeters perform op-
patient is placed in a Trendelenberg position (upper body tical measurements using only two wavelengths. Conse-
and head tilted downward by 30–40 degrees) causing ve- quently, a pulse oximeter cannot distinguish between the
nous blood pooling in the head. relative concentrations of each type of Hb derivative pres-
ent in the blood because it is calibrated to display func-
6.4. Effect of Fetal Hb tional oxyhemoglobin saturation as defined above.
At birth, fetal hemoglobin constitutes 60% to 95% of the The presence of significant amounts of dysfunctional
total Hb in blood of both preterm and term newborns. hemoglobins (i.e., hemoglobin derivatives that are not ca-
Therefore, one of the concerns clinicians may have relates pable of reversibly binding with O2), such as HbCO and
to the interpretation of pulse oximeter readings in new- MetHb, can confuse the interpretation of the readings by a
borns because pulse oximeters are calibrated on adult pa- pulse oximeter. Therefore, clinicians must recognize the
tients. However, several studies have shown that the above-mentioned differences in interpreting pulse oxi-
difference in the optical absorption of adult and fetal Hb meter readings.
is insignificantly small to affect the clinical accuracy of As noted in Fig. 1, in the red region around 660 nm, Hb
pulse oximetry (19,55,56). and MetHb have similar extinction coefficients, whereas
in the near-infrared region, MetHb absorbs considerable
more light compared with Hb and HbO2. MetHb is formed
6.5. Interference from Electrosurgical Units (ESUs) and
when the iron in Hb is oxidized from the ferrous (Fe þ þ ) to
Ambient Light
the ferric (Fe þ þ þ ) state. Normally, MetHb is present in
The energy produced by incandescent and quartz-halogen the blood in relatively low concentrations (o1%), and thus
light sources has spectral components in the visible and its effect on a pulse oximeter is virtually insignificant.
near-infrared wavelength regions, which also overlap with However, clinical studies have shown that MetHb levels
PULSE OXIMETRY 2933

may be high congenitally or various medications, most transport from the operating room to the recovery room; in
notably sulfa and sodium nitroprusside, can increase in emergency medicine; and more recently in ambulatory
the amount of MetHb in the blood and produce erroneous applications (69–88).
low readings by a pulse oximeter (61). This effect is pre-
dictable from the optical absorption property of MetHb 7.1. Neonatal and Pediatric Pulse Oximetry
shown in Fig. 1 because an increase in the concentration of
Hyperoxia in premature neonates is associated with an
MetHb tends to drive the R ratio toward unity and a cor-
increased risk of developing retinopathy of prematurity or
responding SpO2 value of 85%.
chronic lung disease because the oxyhemoglobin dissocia-
HbCO, which is a normal product of Hb catabolism in
tion curve is flattened at oxygen saturation values above
the body and is formed when carbon monoxide (CO) com-
90%, and errors of only a few percent in SpO2 measure-
bines with Hb, readily competes with Hb for O2 binding
ments could easily represent a large and significant error
sites. The quantity of HbCO in the blood of nonsmoking
in estimating the oxygen tension value. Therefore, sup-
subjects is typically below 2%, although it can be higher
plemental oxygen delivery to premature infants must be
depending on local environmental conditions. In smokers
closely controlled to obtain optimal tissue oxygenation
(or smoke inhalation victims), however, the level of HbCO
while minimizing potential serious side effects associated
may rise above 10%. As shown in Fig. 1, the absorption of
with hyper- or hypoxemia.
HbCO in the near-infrared region is very small, and for
Although several studies demonstrated that the infor-
wavelengths greater than 920 nm, it is practically negli-
mation obtained from a pulse oximeter can be used as a
gible. On the contrary, the optical absorption of HbCO in
feedback signal in a closed-loop controller algorithm to
the red region of the spectrum is similar to HbO2, which
automatically adjust the inspired oxygen fraction during
gives the blood a ‘‘cherry red’’ color. Thus, as expected from
the administration of supplemental oxygen (89–92), the
Fig. 1, a two-wavelength pulse oximeter interprets HbCO
main utility of pulse oximetry in neonatology remains the
as HbO2. Clinical studies confirmed that elevated levels of
prevention of hypoxia rather than hyperoxia.
HbCO lead to a falsely high SpO2 readings (62–64).
Pulse oximetry has also been used as an apnea monitor
Because pulse oximetry is based on the principle of
at home for infants and children at risk for sudden infant
light absorption, nail polish or radiographic dyes intro-
death syndrome and to check the adequacy of supplemen-
duced into the blood stream intravenously can adversely
tal oxygen therapy in patients with chronic obstructive
affect the readings of a pulse oximeter. For example, meth-
pulmonary disease. However, the expense and frequent
ylene blue, which is commonly used as an antidote in the
false alarms from motion artifacts have limited the wide
treatment of an elevated level of MetHb, has a high ab-
use of pulse oximeters for these specific applications.
sorbance peak around 670 nm, and this wavelength closely
coincides with the 660 nm wavelength used in pulse oxi-
7.2. Retinal Oximetry
metry. Clinical studies found that methylene blue can
cause spuriously low SpO2 readings (65–67). Attempts to measure the oxygen saturation within the ret-
inal vasculature noninvasively dates back to the 1950s
(93). This interest was motivated by the idea that measur-
7. CLINICAL APPLICATIONS ing oxygen saturation in blood supplying the retinal arter-
ies can be used to infer how much oxygen is supplied to
Since the invention of pulse oximetry in the mid-1970s, cerebral tissues because the ophthalmic artery that sup-
the technology has rapidly spread and became an invalu- plies oxygen to the retina is also a branch of the internal
able monitoring tool in many clinical settings. Monitoring carotid artery that supplies blood to the brain. A retinal
O2 concentration in the inspired air has long been a stan- oximeter has also been suggested as a noninvasive tech-
dard procedure to confirm that hypoxic gas concentrations nique that may be used to identify cerebral bleeding before
are not delivered inadvertently to the patient during an- changes in vital signs and to provide a reliable index of
esthesia well before pulse oximeters became available. oxygen delivery during shock resuscitation of trauma vic-
Shortly after pulse oximetry was introduced, in 1986, it tims. Additionally, measurement of retinal oxygen utiliza-
was recommended as a standard of care for basic intraop- tion may provide clinical information about the metabolic
erative monitoring by the American Society of Anesthesi- state of the retina. If the ocular blood flow and oxygen sat-
ologists (68). In 1988, the Society for Critical Care uration of the arterioles and venules in the retina are
Medicine adopted a similar recommendation that pulse known, the information can be used to diagnose and study
oximetry be used to monitor patients undergoing oxygen retinal metabolic function in different disease states such
therapy. Continuous monitoring of pulse oximetry is now a as diabetic retinopathy and macular degeneration.
de facto standard of care for virtually all cases that require Retinal oximetry exploits the unique transparency of
intensive or critical care management. the ocular media that enables reflectance measurements
The popularity of pulse oximetry stems from the prem- from the retinal fundus by shining light through the pupil.
ise that pulse oximetry is capable of early detection of The incident light impinging on the retina interacts with
changes in a patient’s condition, which therefore allows the blood, the vessel wall, and the pigmented epithelium
rapid intervention that could prevent adverse conse- background. Consequently, some light reflected and back-
quences. To date, pulse oximetry has been widely used in scattered from the retina does not interact with the
various clinical applications including surgery, critical blood. In addition, the relationship between the different
care, hypoxemia screening, and exercise; during patient components of the reflected light depends on path length,
2934 PULSE OXIMETRY

hematocrit, and blood flow. Moreover, the absolute amount fetal assessment of acid–base balance and, thus indirectly,
of light reflected from the retina is wavelength dependent the adequacy of oxygenation. However, fetal scalp sam-
and can vary depending on the reflectance from surface pling has drawbacks because it is an invasive procedure,
and deeper tissue components (94). requires operator’s skill, and provides information about
Numerous techniques have been employed to develop a the fetal acid–base balance only at the time of sampling.
retinal oximeter with various degrees of success. Most Therefore, this method has not been universally or rou-
methods are based on digital images acquired at two tinely adopted by obstetricians.
wavelengths (569 and 600 nm) (95), three wavelengths The potential application of reflectance pulse oximetry
(558, 569, and 586 nm) (96), and four wavelengths (488, to fetal monitoring during labor has been demonstrated by
635, 670, and 830 nm) (97). de Kock et al. (98) developed several groups since the early 1990s (103–105). The main
an in vitro system to simulate the retinal circulation and objective was to detect early signs of hypoxia and perform
ocular optics based on a dual-wavelength (660 and trend monitoring in the fetus soon after the cervix is dilated
940 nm) reflectance pulse oximeter and found that it can and the amniotic membranes are ruptured. This technology
detect changes in retinal SpO2 based on data from the in is viewed as an adjunct to electronic FHR monitoring in the
vitro model. Despite these efforts, because there are mul- presence of a nonreassuring fetal HR pattern to help ob-
tiple variables that affect the absolute intensity of the re- stetricians determine whether the fetus is receiving enough
flected light that is used to calculate oxygen saturation in oxygen so that a Caesarian section can be avoided.
the retina, the method is used as a research tool, but a Several early designs of fetal pulse oximeters were
clinical useful device that provides accurate reading with- based on attempts to affix the sensor to the presenting
out the need for in vivo calibration is not yet available. part of the fetus scalp. Different methods were proposed,
including tissue adhesive, vacuum cups, inflatable bal-
7.3. Esophageal Pulse Oximetry loons, and a spiral needle used in the attachment of a
scalp ECG electrode. However, these fixation methods
Because transmission or reflection-type sensors that mea-
failed to achieve satisfactory results. Nellcor, Inc. has de-
sure SpO2 from the skin rely on adequate peripheral per-
veloped a unique fetal pulse oximeter probe that over-
fusion at the measurement site, they can fail to provide
comes this problem. The disposable probe contains the red
accurate readings in patients with compromised periph-
and infrared LEDs and the PD assembly for measuring the
eral pulsatile circulation. Such clinical situations occur,
fetal PPG signals and two contact electrodes to determine
for example, in states of hypovolemia, hypothermia, or
through changes in impedance measurements if the optical
vasoconstriction. Several studies investigated the alterna-
components are contacting the tissue. These components
tive use of a reflectance-type sensor mounted inside a spe-
are mounted inside a smooth and pliable housing with a
cially modified esophageal probe to monitor SpO2 from
flexible curved tip acting as a fulcrum as shown in Fig. 8.
core organs within the mediastinum in patients undergo-
The sensor is attached to a handle with a removable stylet.
ing prolonged cardiothoracic bypass surgery and in the
Because the sensor cannot be observed during application,
intensive care unit. The idea is based on the contention
directional markings on the sensor’s handle are used to
that the use of a more central monitoring site such as the
orient the optical components toward the fetus face, and
esophagus increases the likelihood of obtaining more reli-
depth markings aid the obstetrician in positioning the sen-
able measurements because it is a better perfused organ
sor against the skin. The probe can be used only on fetuses
than the extremities during periods of poor peripheral
in vertex (head down position) presentation. It is inserted
perfusion (99–101).
transvaginally into the mother’s uterus once the maternal
membranes have ruptured. Normal uterine forces coupled
7.4. Fetal Pulse Oximetry
with the special design of the probe hold it in place against
Electronic fetal heart rate (FHR) monitoring is used rou- the cheek or temple of the fetus for the duration of the la-
tinely in labor as a standard for intrapartum assessment bor and delivery as illustrated in Fig. 9.
of fetal well-being. Although it is considered to be a sen- In addition to the problems associated with pulse ox-
sitive technique and can identify a fetus with a normal imetry, as was previously discussed, fetal pulse oximetry
tracing as being uncompromised, the diagnostic value of also poses unique and additional challenges. In particular,
FHR monitoring as an indirect measure of fetal oxygen- sensor design and monitoring site are important factors
ation and acid-base balance is limited and is open for in- that can compromise its performance as a diagnostic tool
terpretation. In clinical situations in which FHR tracing is of fetal hypoxia and may lead to adverse outcomes. The
reassuring, the method has a predictive value of almost SNR in fetal SpO2 monitoring is significantly smaller com-
99% on the well-being of the fetus. However, an abnormal pared with conventional pulse oximetry, which is primar-
FHR has a positive predictive value of only 50% and there- ily because of the smaller PPG amplitudes obtained from
fore exhibits relatively poor specificity in detecting fetal the sensor application site. Additionally, motion artifact
compromise (102). during contractions caused by fetal and maternal move-
Several approaches to fetal assessment have been de- ments present significant technical challenges. A multi-
veloped with various degrees of success in an attempt to center study (106) found also that there is a relatively high
resolve the lack of specificity associated with electronic incidence of signal dropout that may be attributed to fetal
FHR monitoring as an indication of fetal distress during scalp congestion (edema), vernix, site and probe applica-
labor. For example, blood sampling from the fetal scalp to tion, and movement artifacts. The study found that reli-
determine fetal blood pH has been employed to assist in able readings can only be obtained 60–70% of the time.
PULSE OXIMETRY 2935

Sensor's fulcrum tip

Contact
electrodes LEDs
Optical 735 nm and 890 nm
components
and contacts
Photodetector

(a) (b)

(c)
Figure 8. The Nellcor Model N-400 fetal pulse oximeter and the special probe developed to mon-
itor SpO2 from the fetal cheek (Courtesy of Nellcor Puritan Bennett, Inc. Pleasanton, CA). (This
figure is available in full color at http://www.mrw.interscience.wiley.com/ebe.)

Despite these limitations, the FDA approved the mar- controversial and contradicting whether this technology
keting of the Nellcor fetal oximeter in 2000. However, the is ready for routine clinical practice (108–111).
American College of Obstetricians and Gynecologists has
so far not endorsed the adoption of fetal pulse oximetry as
7.5. Combining Pulse Oximetry and Transcutaneous pCO2
a standard for routine intrapartum use citing concerns
Measurement
that ‘‘its introduction could further escalate the cost of
medical care without necessarily improving clinical out- In addition to measuring arterial oxygen saturation, it is
come’’ (107). The recommendation was based on insuffi- frequently also necessary to determine the pCO2 in arte-
cient clinical studies to show that the use of fetal pulse rial blood to determine CO2 elimination and assess more
oximetry cuts the rate of Caesarian section deliveries. De- accurately the respiratory status of the patient.
spite continued efforts to evaluate the utility of this tech- Gisiger et al. (112) and Eberhard et al. (113) developed
nology in randomized clinical trials conducted by the NIH a miniaturized combined sensor for simultaneous moni-
and the manufacturer, several recent opinions remain toring of SpO2 and transcutaneous pCO2 in pediatric and
adult patients during surgery. The combined sensor is at-
tached to the earlobe by a low-pressure clip. The transcu-
taneous CO2 sensor is based on a potentiometric
measurement by a Stow–Severinghaus-type electrode
(114,115), which determines the pH of an electrolyte layer
separated from the skin by a permeable transparent mem-
brane. A change of pH is proportional to the logarithm of a
pCO2 change. The other part of this sensor comprises the
optical components of a pulse oximeter sensor, although
the signal measured by the photodiode originates both
from transmission as well as from reflection of light from a
reflective element mounted at the opposite side of the ear-
lobe clip. A small heater is imbedded in the housing to
raise the temperature of the earlobe to 421C to achieve
local arterialization of the cutaneous tissue underneath
the sensor that is normally required for accurate transcu-
taneous pCO2 measurement. This approach can also be
advantageous for enhancing SpO2 monitoring by increas-
ing blood flow during periods of low peripheral perfusion
and thereby the strength of the PPG signals (116).

7.6. Noninvasive Blood Pressure Measurement by Pulse


Oximetry
Figure 9. Resting location of the Nellcor fetal pulse oximeter Because the PPG waveform corresponds remarkably well
probe (Courtesy of Nellcor Puritan Bennett, Inc. Pleasanton, CA). to the blood pressure waveform and is displayed on many
2936 PULSE OXIMETRY

pulse oximeters, several studies for alternative measure- oximeter performs according to the manufacturer’s func-
ment of blood pressure noninvasively using a pulse oxi- tional specifications. In the quest to produce reliable and
meter have been reported (117–120). The method requires reproducible pulse oximeter simulators, various methods
both devices to be placed on the same extremity and the have been employed based on mechanical pumping of
blood pressure cuff positioned proximal to the location of blood through an artificial finger model or nonblood de-
the pulse oximeter probe. The technique is based on noting vices that incorporate optical or mechanical means to sim-
the disappearance and/or reappearance of the PPG wave- ulate the properties of blood pulsation through a human
form during the inflation and deflation of the blood pres- finger (130).
sure cuff. The technique might be useful in patients for Several manufacturers of medical testing equipment
whom standard oscillometric blood pressure measure- (e.g., Metron, DNI Nevada, and Bio-Tek Instruments) de-
ments are not reliable or obtainable, such as those with veloped nonblood optoelectronic simulators for testing the
very low blood pressure. However, the main shortcoming performance of different pulse oximeters in vitro over a
of this approach is that only systolic blood pressure can be wide range of oxygen saturation and heart rate values.
measured and the oximetric determination is subjective. Some of these devices include simulations of different pa-
tient conditions that can be encountered in the clinical
7.7. Pulse Oximetry Waveform Interpretations setting, such as the response to weak and noisy PPGs.
These simulators were developed for field testing of pulse
In addition to the dependence of the PPG amplitude and
oximeter models from specific manufacturers and there-
frequency on oxygen saturation and HR, the morphology
fore are not universal. Typical testing may involve simu-
of the PPG contains spectral information that can be use-
lations of the electronic signals generated by the PD to test
ful for clinical interpretation of various physiological con-
the electronic processing circuitry in the pulse oximeter or
ditions and cardiovascular diseases (121,122). Therefore,
testing the optical sensor separately to identify broken
in addition to SpO2 and HR information, several manu-
wires. It is important to realize that these simulators are
facturers typically display the AC component of the infra-
only used to test the functional capability of the pulse oxi-
red PPG waveform because it does not depend on
meter relative to its original factory settings and are not
variations in SO2. Normally, the variable DC baseline of
intended to test the absolute accuracy of the pulse oxi-
the PPG is removed, and autoscaling and autocentering
meter or the validity of the algorithms used by the man-
routines are used to ensure that the amplitude of the AC
ufacturer for calibrating their pulse oximeters.
waveform is maximized and remains on the display
Several in vitro tissue phantoms using blood were pro-
screen. Some manufacturers include an option to turn off
posed for evaluating different physiological parameters or
these automatic functions.
physical factors that can affect the accuracy of a pulse oxi-
Early investigations recognized that the AC waveform
meter. These artificial tissue phantoms attempt to simu-
of the PPG correlates to changes in sympathetic tone of
late the optical absorption and scattering properties of
the peripheral blood vessels and thus can change dramat-
biological tissues using nonhemolyzed blood. Such models
ically with cardiovascular shock and sedation (123–126).
can be useful because deliberately desaturating human
For example, episodes of increased sympathetic tone, such
subjects is a contentious issue mainly because of the asso-
as vasoconstriction attributed to pharmacologic agents or
ciated risks that subjects may suffer irreversible brain
physiologic conditions elicited by cold temperatures, cause
damage breathing hypoxic gas mixtures, which can lead
the amplitude of the AC component to decrease. Likewise,
to inaccurate readings at low levels of SpO2. Moreover, it is
with vasodilation, the amplitude is increased. This infor-
sometimes unethical or impractical for example to accu-
mation can be useful for an anesthesiologist to follow sym-
rately study in vivo the effect of certain physiological pa-
pathetic tone and sympathetic response to surgical
rameters, such as elevated levels of CO, MetHb, or certain
stimulation. Other investigators showed that normal re-
intravenous dyes; the effect of variations in hematocrit; or
spiratory activity can also affect the amplitude of the PPG
changes in blood flow conditions. Mendelson and Kent
waveform. For example, it was noted that during sponta-
(131) developed a circulating blood phantom of a finger to
neous breathing, an 8–10 peripheral pulse trace shows a
investigate the effect of HbCO on the accuracy of a pulse
slow phasic respiratory wave that modulates the envelope
oximeter. Likewise, de Kock and Tarassenko (132), Vegfors
of the AC components of the PPG. These low-frequency
et al. (133), and Reynolds et al. (134) used a similar in vitro
modulations of the PPG signal are directly related to the
approach to investigate the effects of hematocrit, blood
pressure and volume of the air contained in the lungs
flow, blood volume, and degree of pulsatility on the nor-
(127,128). Nakajima et al. (129) have demonstrated ways
malized R ratio that is used in calibrating a pulse oximeter.
of isolating these low-frequency components using digital
A noncirculating tissue model that requires a smaller
filters. The resulting waveforms are representative of a
amount of blood was described by Volgyesi et al. (135).
subject’s breathing pattern.

8.1. In vitro Calibration and Verification


8. PULSE OXIMETER SIMULATORS
A novel prototype device was proposed for standardized
Although pulse oximeters are precalibrated in vivo by and universal calibration of a pulse oximeter in vitro
their manufacturer using human subjects, the many pulse based on a database of time-resolved raw optical trans-
oximeters available on the market increased the demand mission spectra that was previously recorded from human
for a simple and cost-effective way to test whether a pulse fingers during the calibration studies of pulse oximeters
PULSE OXIMETRY 2937

(136–139). The in vitro calibration process is simulated by 5. G. A. Millikan, J. R. Pappenheimer, A. J. Rawson, and J. P.
playing back the stored optical transmission spectra that Hervey, Continuous measurement of oxygen saturation in
were recorded in vivo for different arterial SO2 values by man. Am. J. Physiol. 1941; 133:390.
means of a special artificial finger calibrator that has op- 6. E. H. Wood and J. E. Geraci, Photoelectric determination of
tical scattering properties similar to real human fingers arterial oxygen saturation in man. J. Lab. Clin. Med. 1949;
and a specially designed tissue spectrometer. In contrast 34:387–401.
with in vitro simulation models of a human finger, this in 7. E. B. Merrick and T. J. Hayes, Continuous, non-invasive
vitro spectrophotometric method is based on data recorded measurements of arterial blood oxygen levels. Hewlett-Pack-
in vivo from real human subjects and therefore bear the ard J. 1976; 28(2):2–9.
basic interaction between a pulse oximeter and a human 8. J. P. Payne and J. W. Severinghaus, eds., Pulse Oximetry.
finger. These data can be played back to test the accuracy New York: Springer-Verlag, 1987.
and calibration procedure of any pulse oximeter. It can 9. J. G. Webster, ed., Design of Pulse Oximeters. Institute of
also be used to investigate how the pulse oximeter re- Physics, Boca Raton, Florida, 1997.
sponds to rapid changes in oxygen saturation and motion 10. M. Yeldennan and W. New, Jr, Evaluation of Pulse Oximetry.
interferences. Anesthesiology 1983; 59(4):349–352.
11. C. M. Alexander, L. E. Teller, and J. B. Gross, Principles of
pulse oximetry: Theoretical and practical considerations.
8.2. Future Trends Anesth. Analg. 1989; 68:368–376.
Pulse oximetry has elegantly combined the principles of 12. K. K. Tremper and S. J. Barker, Monitoring of oxygen. In:
photoplethysmography and spectrophotometry to mea- C. L. Lake, ed., Clinical Monitoring for Anesthesia and Crit-
sure noninvasively the oxygen saturation of arterial blood. ical Care, 2nd ed. Philadelphia, PA: W.B. Saunders, 1994.
The method has been widely welcomed by many practi- 13. E. H. Thilo, J. B. Curlander, and W. W. Hay, Pulse oximetry.
tioners and undoubtedly serves as an important tool used In: Y. W. Brans and W. W. Hay, eds., Physiological Monitoring
by clinicians to assess the status of patient oxygenation. and Instrument Diagnosis in Perinatal and Neonatal Medi-
cine. Cambridge, UK: Cambridge University Press, 1995.
Pressure on manufacturers to overcome several short-
comings and improve the performance of pulse oximeters 14. M. W. Wukitsch, M. T. Petterson, D. R. Tobler, and J. A.
Pologe, Pulse oximetry: Analysis of theory, technology, and
has led to improvement and expansion of the technology in
practice. J. Clin. Monit. 1988; 4:290–301.
an attempt to satisfy diverse clinical requirements. De-
15. J. Pologe, Pulse oximetry: Technical aspects. Int. Anesth.
spite almost two decades of significant technological
Clin. 1987; 25(3):142–153.
achievements, the scope of pulse oximetry is still expand-
16. T. Aoyagi and K. Miyasaka, Pulse oximetry: Its invention,
ing. Improvements in performance of pulse oximeters, es-
contribution to medicine, and future tasks. Anesth. Analg.
pecially the most troublesome need to reduce the 2002; 94:S1–S3.
occurrence of false alarm rates and provide more reliable
17. T. Aoyagi, M. Kishi, K. Yamaguchi, and S. Watanabe,
measurements under conditions of low peripheral perfu- Improvement of the earpiece oximeter. Abstracts of the Jap-
sion, continue to be made as new digital signal processing anese Society of Medical Electronics and Biological Engi-
hardware and more advanced software algorithms are neering. 1974: 90–91.
being developed by different manufacturers. 18. I. Yoshiya, Y. Shimada, and K. Tanaka, Spectrophotometric
Future trends include improvements in performance by monitoring of arterial oxygen saturation in the fingertip.
developing more advanced digital signal processing algo- Med. Biol. Eng. Comput. 1980; 18:27–32.
rithms, more flexible and intelligent alarm settings, and 19. Y. Mendelson and J. C. Kent, Variations in optical absorption
the combination of pulse oximeters with other monitoring spectra of adult and fetal hemoglobins and its effect on
signals. Additional efforts are also underway to integrate pulse oximetry. IEEE Trans. Biomed. Eng. 1989; 36(8):
wireless telemetry with pulse oximetry and expand its use 844–848.
in ambulatory monitoring, in ambulances, during helicop- 20. P. F. Wouters, H. Gehring, G. Meyfroidt, L. Ponz, et al.,
ter transport, and at home. Wearable pulse oximeters are Accuracy of pulse oximeters: The European multi-center
also being developed for civilian and military applications trial. Anesth. Analg. 2002; 94:S13–S16.
by first responders and soldiers during routine training 21. L. A. Jensen, J. E. Onyskiw, and N. G. N. Prasad, Meta
and combat applications. analysis of arterial oxygen saturation monitoring by pulse
oximetry in adults. Heart Lung 1998; 27:387–408.
22. D. Hess, M. Kochansky, L. Hassett, R. Frick, and W. O.
BIBLIOGRAPHY Rexrode, An evaluation of the Nellcor N-10 portable pulse
oximeter. Respir. Care 1986; 31:796–802.
1. R. Huch, A. Huch, and D. W. Lubbers, Transcutaneous PO2. 23. J. W. Severinghaus and K. H. Naifeh, Accuracy of response
New York: Thieme-Stratton, 1981. of six pulse oximeters to profound hypoxia. Anesthesiology
2. J. W. Severinghaus and P. B. Astrup, History of blood gas 1987; 67:551–558.
analysis. VI. Oximetry. J. Clin. Monit. 1986; 2:270–288. 24. J. W. Severinghaus, K. H. Naifeh, and S. O. Koh, Errors in 14
3. K. Matthes, Uber den einfluss der atmung auf die sauer- pulse oximeters during profound hypoxemia. J. Clin. Monit.
stoffsattingungen des arterienblutes. Arch. Exp. Pathol. 1989; 5:72–81.
Phannacol. 1934; 176:683–696. 25. Y. Mendelson, P. W. Cheung, D. G. Fleming, and M. R. Neu-
4. K. Matthes, Untersuchungen uber die sauerstoffsattingun- man, Spectrophotometric transcutaneous reflectance oxime-
gen des menschlichen arterienblutes. Arch. Exp. Pathol. try. 2nd Int. Symp. Continuous Blood Gas Monitoring,
Pharmacol. 1935; 179:698–711. Zurich, Switzerland, 1981.

You might also like