You are on page 1of 17

DPharp_DP Flow

YSAPE

A90-TCS-421
Copyright © Yokogawa Electric Corporation -1-
July. 2011
Continuous R&D in Sensor Technology

1970’s 1980’s 1990’s 2000’s


Pneumatic

Analog

Analog
P10 E10 Digital
UNI⊿

YOKOGAWA YA11
Force Balance Piezoresistive
Capacitance DPharp DPharp DPharp
Silicone Resonant
EJ EJA EJX

DPharp = DP high accuracy resonance pressure

A90-TCS-421
Copyright © Yokogawa Electric Corporation -2-
July. 2011 Proprietary info goes here…
2
DPharp Products Solution

Basic Solution Premium Value

EJA
Best-in-class Accuracy
MV (Mass Flow)
SIL2
World Fastest Response Time
Dual Output (Analog + Pulse)
Advanced Diagnostics Advanced Diagnostics
Basic Features Best-in-class MTBF Best-in-class MTBF
of DPharp Sensors
Long term stability 5 years Long term stability 10 years
under all conditions under all conditions
No hysteresis, no drift No hysteresis, no drift
Multi-sensing (DP, SP, T) Multi-sensing (DP, SP, T)
High Accuracy, High Resolution High Accuracy, High Resolution
Full Digital Sensor Full Digital Sensor
A90-TCS-421
Copyright © Yokogawa Electric Corporation -3-
July. 2011 Proprietary info goes here…
3
Reverse Flow

Forward Flow
– Operating as normal
Flow Reversed Forward Reverse

– Output will go into low


saturation condition
– High & low side pressure
measurement swapped
in electronics
– Operating as normal

A90-TCS-421
Copyright © Yokogawa Electric Corporation -4-
July. 2011
Improving Process Variability

Liquid, Gas & Steam Flow Measurement


– DPharp Sensor Technology
• Best in class accuracy & long term stability
Gas & Steam Flow Measurement
– Multivariable Mass Flow
• Improved flow measurement for gases & steam
• Multivariable measurement & transmission

A90-TCS-421
Copyright © Yokogawa Electric Corporation -5-
July. 2011 Proprietary info goes here…
5
What is Process Variability?

BEFORE AFTER
Offset due to Variation Offset due to Variation

Product Operating Product


Spec. Set Point Spec.

Operating
Set Point

A reduction means less waste of end product, plus


less raw material and utilities consumption
A90-TCS-421
Copyright © Yokogawa Electric Corporation -6-
July. 2011 Proprietary info goes here…
6
Total Accuracy

E1 = Reference Accuracy
E2 = Ambient Temperature Effect
E3 = Static Span Effect
E4 = Static Zero Effect
E5 = Overpressure Effect
Root Mean Square of all the errors

E1  E2  ( E3  E4 )  E5
2 2 2 2

A90-TCS-421
Copyright © Yokogawa Electric Corporation -7-
July. 2011 Proprietary info goes here…
7
Calibration Frequency

Better Installed Accuracy & Long Term Stability

0.60

Requested
0.50 Performance
Total accuracy (% of span)

0.5% of span

0.40

0.30

0.20

0.10

0.00
0 12 24 36 48 60 72 84 96 108 120 132 144

Months in service

A90-TCS-421
Copyright © Yokogawa Electric Corporation -8-
July. 2011 Proprietary info goes here…
8
Introduction to Multivariable Mass Flow

Measurement & Compensation in One Device


Measurement
– DP, static pressure & process temperature
Compensation
– Primary element factors
– Fluid property factors based on fixed composition

A90-TCS-421
Copyright © Yokogawa Electric Corporation -9-
July. 2011 Proprietary info goes here…
9
Traditional System

Flow Computer
(MW100)

Differential Pressure Static Pressure Temperature


Transmitter Transmitter Transmitter

A90-TCS-421
Copyright © Yokogawa Electric Corporation - 10 -
July. 2011 Proprietary info goes here…
10
Multi-Sensing System

Flow Computer
(Stardom)

Multi-Sensing (Differential & Static) Temperature


Pressure Transmitter Transmitter

A90-TCS-421
Copyright © Yokogawa Electric Corporation - 11 -
July. 2011 Proprietary info goes here…
11
EJX900 Series Multivariable Mass Flow

Dynamically Compensated Mass Flow


– Primary elements factors (orifice plates, venturis
& nozzles)
– Fluid properties (natural gas, steam, other gases &
liquids)
Other Process Variables
– Differential Pressure
– Pressure
– Temperature

A90-TCS-421
Copyright © Yokogawa Electric Corporation - 12 -
July. 2011 Proprietary info goes here…
12
Dynamic Compensation

Beta Ratio
Pressure
Gas Expansion Factor
Isoentropic Exponent Orifice Bore
Differential Pressure
Units Conversion Factor
Reference Orifice Bore

Qmass = N Cd E Y1 d2 DP(ρ) Temperature

Density
Discharge Coefficient Velocity of Approach Factor

Beta Ratio Reynolds Number


Beta Ratio
Temperature Compressibility Factor
Pipe ID Pipe ID Gas Constant
Orifice Bore Velocity Molecular Weight
Differential Pressure
Temperature Viscosity Fluid Constants
Temperature Pressure
Density
Temperature

Compressibility Factor
Gas Constant
Pressure Key
Temperature

Data Entered Measured Calculated

A90-TCS-421
Copyright © Yokogawa Electric Corporation - 13 -
July. 2011 Proprietary info goes here…
13
Flow Errors from Density

Impact of 3.6% Density Variation on Flow Measurement


Impact on Repeatability & Accuracy

4
3.6
3.5
3
2.5
2 1.8
1.5
1
0.5 <0.1
0
Uncompensated Uncompensated Compensated
Velocity Meter DP Flow Meter Meters
A90-TCS-421
Copyright © Yokogawa Electric Corporation - 14 -
July. 2011 Proprietary info goes here…
14
EJX900 Series - Retransmission

Primary Variable (4 – 20 mA output)


– Normally assigned to flow
– Pulsed output for totalised flow
– HART compliant system
• Digital output for DP, SP & PT
– Non HART compliant system
• HART to analogue converter generates additional 4 – 20
mA outputs
All Digital Foundation Fieldbus (H1)

A90-TCS-421
Copyright © Yokogawa Electric Corporation - 15 -
July. 2011 Proprietary info goes here…
15
EJX900 Series - Total Cost of Ownership

Reduced Installed Cost


– Less wiring
– Fewer power supplies (& barriers)
– Less terminations
Reduced Maintenance
– Less frequent calibrations
– Less instruments to maintain
– Reduced spares

A90-TCS-421
Copyright © Yokogawa Electric Corporation - 16 -
July. 2011 Proprietary info goes here…
16
Thank you for your attention!

A90-TCS-421
Copyright © Yokogawa Electric Corporation - 17 -
July. 2011

You might also like