Professional Documents
Culture Documents
ELECTRICAL
Quantity Measure Range Best Equipment
Uncertainty Used
DC VOLTAGE- 0 to 200 mV ± 0.6 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC VOLTAGE- 0.22 to 2 V ± 0.9 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC VOLTAGE- 2.2 to 20 V ± 7 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC VOLTAGE- 100 to 200 V ± 330 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC VOLTAGE- 220 to 1000 V ± 0.8 mV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 0 to 20 mV @ 40 Hz ± 18 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 20 to 200 mV @ 10 Hz to 500kHz ± 34 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 0.210 to 2 V @ 10 Hz to 1 MHz ± 49 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 0.210 to 2 V @ 10 Hz to 1 MHz ± 49 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 2.1000 to 20.000 V @ 40 Hz to 20 kHz ± 260 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 21.000 to 200.000 V @ 10 Hz to 1 MHz ± 2.6 mV TRANSMILLE
SOURCE MULTI-PRODUCT
TEMPERATURE
Quantity Range Best Uncertainty Equipment Used
Temperature - -45 to 25 C ±0.02 C AMETEK RTC-
SOURCE 157B
Temperature - 30 to 75 C ±0.03 C AMETEK RTC-
SOURCE 157B
Temperature - 80 to 155 C ±0.12 C AMETEK RTC-
SOURCE 157B
Temperature - 165 to 550 C ±0.21 C GE DRUCK DRY
SOURCE TC-650
Temperature - 600 TO 800 C ±0.54 C AMETEK CTC-
SOURCE 1200A
Temperature - 800 TO 1205 C ±0.71 C AMETEK CTC-
SOURCE 1200A
Temperature - -200 C ± 0.6 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - -80 C ± 1.5 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - 0 C to 100 C ± 1.5 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - 100 C to 300 C ± 2.4 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
LIGHT
Quantity Range Best Uncertainty Equipment Used
Irradiance -SOURCE 300 – 400 nm ±2.7 % RS-10 Series
Spectral
Irradiance-
GAMMA
SCIENTIFIC
Irradiance -SOURCE 400 – 930 nm ±1.8 % RS-10 Series
Spectral