You are on page 1of 17

QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS

ELECTRICAL
Quantity Measure Range Best Equipment
Uncertainty Used
DC VOLTAGE- 0 to 200 mV ± 0.6 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC VOLTAGE- 0.22 to 2 V ± 0.9 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC VOLTAGE- 2.2 to 20 V ± 7 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC VOLTAGE- 100 to 200 V ± 330 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC VOLTAGE- 220 to 1000 V ± 0.8 mV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 0 to 20 mV @ 40 Hz ± 18 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 20 to 200 mV @ 10 Hz to 500kHz ± 34 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 0.210 to 2 V @ 10 Hz to 1 MHz ± 49 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 0.210 to 2 V @ 10 Hz to 1 MHz ± 49 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 2.1000 to 20.000 V @ 40 Hz to 20 kHz ± 260 µV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC VOLTAGE- 21.000 to 200.000 V @ 10 Hz to 1 MHz ± 2.6 mV TRANSMILLE
SOURCE MULTI-PRODUCT

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


CALIBRATOR
3010A
AC VOLTAGE- 210.000 to 1000.000 V @ 10 Hz to 1 MHz ± 45 mV TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC CURRENT- 0.0000 µA to 200.0000 µA ± 0.2 nA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC CURRENT- 0.210000 mA to 2.000000 mA ± 2.7 nA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC CURRENT- 5.000000 mA to 20.00000 mA ± 34 nA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC CURRENT- 100.0000 mA to 200.0000 mA ± 380 nA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC CURRENT- 1.000000 A to 2.000000 A ± 8.7 µA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
DC CURRENT- 10.00000 A to 30.00000 A ± 0.48 mA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC CURRENT- 25.000 µA to 200.000 µA @ 40 Hz to 10 kHz ± 15 nA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC CURRENT- 0.210000 mA to 2.000000 mA @ 40 Hz to 10 ± 55 nA TRANSMILLE
SOURCE kHz MULTI-PRODUCT
CALIBRATOR
3010A
AC CURRENT- 2.1000 mA to 20.00000 mA @ 40 Hz to 10 ± 430 nA TRANSMILLE
SOURCE kHz MULTI-PRODUCT
CALIBRATOR
3010A
AC CURRENT- 21.000 mA to 200.000 mA @ 40 Hz to 10 kHz ± 4.2 µA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


AC CURRENT- 0.21000 A to 2.00000 A @ 40 Hz to 5 kHz ± 42 µA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
AC CURRENT- 2.10000 A to 30.00000 A @ 40 Hz to 1 kHz ± 0.48 mA TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
RESISTANCE - 100 mΩ to 800 mΩ ± 3.2 mΩ MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 1 Ω to 800 Ω ± 80 mΩ MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 1 kΩ to 10 kΩ ± 1000 mΩ MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 10 kΩ to 800 kΩ ± 2000 mΩ MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 1 MΩ to 80 MΩ ± 500 Ω MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 100 MΩ to 800 MΩ ± 130 k Ω MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 1 GΩ to 8 GΩ ± 1.3 MΩ MULTI-
SOURCE FUNCTION

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 10 GΩ to 100 GΩ ± 48 MΩ MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 10 GΩ to 100 GΩ ± 0.46 GΩ MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
RESISTANCE - 1TΩ to 9 TΩ ± 4.8 GΩ MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5320A
FREQUENCY- 0 – 10 MHz ± 20 Hz TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
FREQUENCY- 100 – 300 MHZ ± 600 Hz TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
FREQUENCY- Upto 600 MHz ± 1.2 kHz TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
INDUCTANCE- 1.031 to 29.345mH @ 1kHz ± 1.3 µH TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
INDUCTANCE- 49.410 mH to 100.78mH @ 1kHz ± 15 µH TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A
INDUCTANCE- 1.0120 H to 10.530 H @ 1kHz ± 400 µH TRANSMILLE
SOURCE MULTI-PRODUCT
CALIBRATOR
3010A

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


CAPACIATANCE- Upto 400.0 pF @ 1kHz & 5 kHz ± 0.13 pF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 1.00 nF to 10 nF @ 1kHz & 5 kHz ± 1.2 pF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 11 nF to 33nF @ 1kHz & 5 kHz ± 9.0 pF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- From 34 nF to 110 nF @ 1kHz & 5 kHz ± 20.0 pF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 200 nF to 300 nF @ 1kHz & 5 kHz ± 0.24 nF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 1.00000 µF to 10.0000 µF @ 100 Hz ± 0.70 nF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 11.0000 µF to 30.0000 µF @ 100 Hz ± 0.70 nF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 11.00000 µF to 30.0000 µF @ 100 Hz ± 28 nF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 20.00000 µF to 100.000 µF @ 100 Hz ± 0.14 µF (Upper MULTI-
SOURCE Limit) FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 200.000 µF to 330.000 µF @ 100 Hz ± 49 nF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- 1.10000 mF to 33.0000 mF ± 0.24 µF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
CAPACIATANCE- Up-to 100.00 mF ±24 µF MULTI-
SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
THERMOCOUPLE 600 C TO 1820 C ±60 µV /0.34 C MULTI-
SOURCE / MEASURE FUNCTION
– TYPE B ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
THERMOCOUPLE -250 C TO 1000 C ±60 µV /0.21 C MULTI-
SOURCE / MEASURE FUNCTION
– TYPE E ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
THERMOCOUPLE 0 C TO 2316 C ±60 µV/ 0.31 C MULTI-
SOURCE / MEASURE FUNCTION
– TYPE C ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


THERMOCOUPLE -210 C TO 1200 C ±60 µV/ 0.17 C MULTI-
SOURCE / MEASURE FUNCTION
– TYPE J ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
THERMOCOUPLE -200 C TO 1372 C ±60 µV/ 0.26 C MULTI-
SOURCE / MEASURE FUNCTION
– TYPE K ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
THERMOCOUPLE 0 C TO 1767 C ±60 µV / 0.35 C MULTI-
SOURCE / MEASURE FUNCTION
– TYPE R ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
THERMOCOUPLE 0 C TO 1767 C ±60 µV/ 0.36 C MULTI-
SOURCE / MEASURE FUNCTION
– TYPE S ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
THERMOCOUPLE -250 C TO 120 C ±60 µV/ 0.24 MULTI-
SOURCE / MEASURE FUNCTION
– TYPE T ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (385) -200 C ± 0.6 mΩ MULTI-
@ 18.521 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (385) -80 C ± 1.5 mΩ MULTI-
@ 68.326 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (385) 0C ± 1.5 mΩ MULTI-
@ 100.000 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (385) 100 C ± 2.4 mΩ MULTI-
@ 138.505 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (385) 300 C ± 2.4 mΩ MULTI-
@ 212.052 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (385) 400 C ± 3.6 mΩ MULTI-
@ 247.092 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (385) 630 C ± 3.6 mΩ MULTI-
@ 323.302 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (385) 800 C ± 3.6 mΩ MULTI-
@ 375.704 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –PT 100 (3926) 100 C ± 2.4 mΩ MULTI-
@ 375.704 Ω - FUNCTION
SOURCE ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A
RTD –Ni 120 @ 100 C ± 2.4 mΩ MULTI-
200.67 Ω - SOURCE FUNCTION
ELECTRICAL
TESTER
CALIBRATOR –
FLUKE 5520A

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


DC VOLTAGE- 0.0000000 to 200.0000 mV ± 3 µV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC VOLTAGE- 0.0000000 to 2.0000000 V ± 1.4 µV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC VOLTAGE- 2.0000000 to 20.000000 V ± 0.7 µV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC VOLTAGE- 100.00000 to 200.00000 V ± 0.9 µV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC VOLTAGE- 220.00000 to 1000.0000 V ± 1.3 µV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC VOLTAGE- 0.000000 to 200.0000 mV @ 1 to 100 kHz ± 0.03 mV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC VOLTAGE- 0.000000 to 2.00000 V @ 1 kHz to 1000kHz ± 0.010 mV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC VOLTAGE- 2.200000 to 19. 00000 V @ 1 to 1000 kHz ± 0.013 mV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC VOLTAGE- 22.00000 to 99.00000 V @ 1 to 100 kHz ± 0.018 mV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC VOLTAGE- 100.0000 to 500.0000 V @ 1 to 30 kHz ± 0.03 mV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC VOLTAGE- Up to 1000 V @ 3 kHz to 10 kHz ± 0.03 mV/V FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC CURRENT - 0.000000 – 200.0000 µA ±3 µA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC CURRENT - 0.000000 – 2.000000 mA ±3 µA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC CURRENT - 2.200000 – 10.00000 mA ±3 µA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC CURRENT - 10.20000 – 100.0000 mA ±3 µA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


DC CURRENT - Up to 200.0000 mA ±3 µA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC CURRENT - 0.00000 to 2.00000 A ±0.06 mA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
DC CURRENT - 10.0000 to 20.0000 A ±0.06 mA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC CURRENT - 0.00000 to 200.0000 µA @ 0.3 to 10 Khz ±0.026 mA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC CURRENT - 0.00000 to 200.000 mA @ 0.3 to 10 Khz ±0.05 mA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC CURRENT - 0.00000 to 2.00000 A @ 0.3 to 10 Khz ±0.11 mA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
AC CURRENT - 2.20000 to 20.0000 A @ 0.3 to 10 Khz ±0.21 mA/A FLUKE 8508A
MEASURE PRECISION
MULTIMETER
RESISTANCE (TRUE 1.000063 TO 100.00023 Ω ±18 µΩ/Ω FLUKE 8508A
OHMS) - MEASURE PRECISION
MULTIMETER
RESISTANCE (TRUE 1.000063 TO 100.00023 Ω ±18 µΩ/Ω FLUKE 8508A
OHMS) - MEASURE PRECISION
MULTIMETER
RESISTANCE (TRUE 1.000063 kΩ TO 10.000248 kΩ ±2.0 µΩ/Ω FLUKE 8508A
OHMS) - MEASURE PRECISION
MULTIMETER
RESISTANCE - Up to 99.9996 kΩ ±4.0 µΩ/Ω FLUKE 8508A
MEASURE PRECISION
MULTIMETER
RESISTANCE - 0.999997 MΩ TO 100.0080 MΩ ±4.0 µΩ/Ω FLUKE 8508A
MEASURE PRECISION
MULTIMETER
RESISTANCE - Up to 1.000467 GΩ ±20.0 µΩ/Ω FLUKE 8508A
MEASURE PRECISION
MULTIMETER
RESISTANCE (HIGH Up to 10.01591 MΩ ±4.0 µΩ/Ω FLUKE 8508A
VOLTAGE) - PRECISION
MEASURE MULTIMETER
RESISTANCE (HIGH Up to 100.0080 MΩ ±5.0 µΩ/Ω FLUKE 8508A
VOLTAGE) - PRECISION
MEASURE MULTIMETER

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


RESISTANCE (HIGH Up to 1.000467 GΩ ±10.0 µΩ/Ω FLUKE 8508A
VOLTAGE) - PRECISION
MEASURE MULTIMETER
RESISTANCE (HIGH Up to 10.0003 GΩ ±30.0 µΩ/Ω FLUKE 8508A
VOLTAGE) - PRECISION
MEASURE MULTIMETER
MASS/FORCE
Quantity Range Best Uncertainty Equipment Used
MASS-SOURCE 1 mg to 500 mg (Class E1) ± 0.0015 mg Troemner Ultra
class Mass Sets
MASS-SOURCE 1g to 100 g (Class E1) ± 0.0050 mg Troemner Ultra
class Mass Sets
MASS-SOURCE Up to 500 g (Class E1) ± 0.080 mg Troemner Ultra
class Mass Sets
MASS-SOURCE 1 kg (Class E1) ± 0.40 mg Troemner Ultra
class Mass Sets
MASS-SOURCE 2 kg (Class E1) ± 0.80 mg Troemner Ultra
class Mass Sets
MASS-SOURCE 3 kg (Class E1) ± 1.20 mg Troemner Ultra
class Mass Sets
MASS-SOURCE 5 kg (Class E1) ± 2.00 mg Troemner Ultra
class Mass Sets
MASS-SOURCE 20 kg (Class E1) ± 8.00 mg Troemner Ultra
class Mass Sets
MASS-SOURCE 20 Kg to 220 Kg (Class F) ±12.5 mg Troemner Mass
Sets
MASS-SOURCE 240 Kg to 520 Kg (Class F) ±15.0 mg Troemner Mass
Sets
MASS-SOURCE Up to 800 Kg (Class F) ±21.0 mg Troemner Mass
Sets
MASS-MEASURE 1 mg to 500 mg (Class F2) ±0.26 mg Ohaus Analytical
Balance AV264C
MASS-MEASURE 1 g to 250 g (Class F2) ±0.31 mg Ohaus Analytical
Balance AV264C
MASS-MEASURE 200 g to 1000 g (Class F2) ±0.06 g Ohaus Top
Loading Balance
EP4101C
MASS-MEASURE 1000 g to 2000 g (Class F2) ±0.06 g Ohaus Top
Loading Balance
EP4101C
MASS-MEASURE 2000 g to 4000 g (Class F2) ±0.06 g Ohaus Top
Loading Balance
EP4101C
MASS-MEASURE Upto 30,000 g (Class F2) ±0.01 g Ohaus CD-11
Digital Platform

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


DIMESNIONAL
Quantity Range Best Uncertainty Equipment Used
Dimensional – 0 up to 10 mm ± 0.08 µm Mitutoyo -
PRECISION BLOCKS Gauge Block Sets
Dimensional – 10 up to 25 mm ± 0.10 µm Mitutoyo -
PRECISION BLOCKS Gauge Block Sets
Dimensional – 25 up to 50 mm ± 0.12 µm Mitutoyo -
PRECISION BLOCKS Gauge Block Sets
Dimensional – 50 up to 75 mm ± 0.15 µm Mitutoyo -
PRECISION BLOCKS Gauge Block Sets
Dimensional – 75 up to 100 mm ± 0.18 µm Mitutoyo -
PRECISION BLOCKS Gauge Block Sets
Dimensional – Upto 300 mm ± 0.26 µm Mitutoyo -
PRECISION BLOCKS Gauge Block Sets
TORQUE
Quantity Range Best Uncertainty Equipment Used
TORQUE-SOURCE & 10.00 to 100.00 Nm Clockwise ±0.14 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE-SOURCE & 10.00 to 100.00 Nm Anti-Clockwise ±0.16 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE-SOURCE & 50.00 to 500.00 Nm Clockwise ±0.17 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE-SOURCE & 50.00 to 500.00 Anti-Clockwise ±0.13 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE-SOURCE & 2.500 to 25.00 Nm Clockwise ±0.19 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE-SOURCE & 2.500to 25.00 Nm Anti-Clockwise ±0.19 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE-SOURCE & 300.0 to 3000.0 Nm Clockwise ±0.072 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE-SOURCE & 300.0 to 3000.0 Nm Anti-Clockwise ±0.066 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE-SOURCE & 100.0 to 1000.0 Nm Clockwise ±0.095 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


TORQUE-SOURCE & 100.0 to 1000.0 Nm Anti-Clockwise ±0.098 % NORBAR T-BOX
MEASURE & STATIC
TRANSDUCER
TORQUE- MEASURE Excited Voltages @ 250 Ω to 1000 Ω ± 0.0014 % NORBAR T-BOX
PRESSURE
Quantity Range Best Uncertainty Equipment Used
PRESSURE 0 to 42.019 Bar Gauge – High Range ± 0.0010 Bar Digital Pressure
(PNEUMATIC) – Indicator PACE
SOURCE & 6000
MEASURE
PRESSURE 43.0000 to 125.0672 Bar Gauge – High ± 0.0039 Bar Digital Pressure
(PNEUMATIC) – Range Indicator PACE
SOURCE & 6000
MEASURE
PRESSURE 126.0000 TO 167.0905 Bar Gauge – High ± 0.0090 Bar Digital Pressure
(PNEUMATIC) – Range Indicator PACE
SOURCE & 6000
MEASURE
PRESSURE 168.0000 TO 210.1140 Bar Gauge – Low ± 0.0102 Bar Digital Pressure
(PNEUMATIC) – Range Indicator PACE
SOURCE & 6000
MEASURE
PRESSURE -0.900533 to 2.801791 Bar Gauge – Low ± 0.000010 Bar Digital Pressure
(PNEUMATIC) – Range Indicator PACE
SOURCE & 6000
MEASURE
PRESSURE 3.000000 to 5.603541 Bar Gauge – Low ± 0.000136 Bar Digital Pressure
(PNEUMATIC) – Range Indicator PACE
SOURCE & 6000
MEASURE
PRESSURE 5.700000 TO 7.004410 Bar Gauge – Low ± 0.000180 Bar Digital Pressure
(PNEUMATIC) – Range Indicator PACE
SOURCE & 6000
MEASURE
PRESSURE 0.750 to 1.150 Bar Absolute ± 0.000032 Bar Digital Pressure
(PNEUMATIC) – Indicator PACE
SOURCE & 6000
MEASURE
PRESSURE 0.5 to 50.0000 Mpa ±0.000031 Mpa Ruska
(HYDRAULIC) – Deadweight
SOURCE & Tester
MEASURE
PRESSURE 10.05115 to 99.8988 Mpa ± 0.00065 Mpa Ruska
(HYDRAULIC) – Deadweight
SOURCE & Tester
MEASURE

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


PRESSURE 983.457 Kpa to 4982.89 Kpa ±0.12 Kpa Ruska
(HYDRAULIC) – Deadweight
SOURCE & Tester
MEASURE
PRESSURE 0.5 to 50.0000 Mpa ±0.000031 Mpa Ruska
(HYDRAULIC) – Deadweight
SOURCE & Tester
MEASURE
PRESSURE 0.5 to 50.0000 Mpa ±0.000031 Mpa Ruska
(HYDRAULIC) – Deadweight
SOURCE & Tester
MEASURE
PRESSURE 0 to 40,000 Psi ±0.01 % Additel Digital
(HYDRAULIC) – Pressure
SOURCE &
MEASURE

TEMPERATURE
Quantity Range Best Uncertainty Equipment Used
Temperature - -45 to 25 C ±0.02 C AMETEK RTC-
SOURCE 157B
Temperature - 30 to 75 C ±0.03 C AMETEK RTC-
SOURCE 157B
Temperature - 80 to 155 C ±0.12 C AMETEK RTC-
SOURCE 157B
Temperature - 165 to 550 C ±0.21 C GE DRUCK DRY
SOURCE TC-650
Temperature - 600 TO 800 C ±0.54 C AMETEK CTC-
SOURCE 1200A
Temperature - 800 TO 1205 C ±0.71 C AMETEK CTC-
SOURCE 1200A
Temperature - -200 C ± 0.6 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - -80 C ± 1.5 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - 0 C to 100 C ± 1.5 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - 100 C to 300 C ± 2.4 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


Temperature - 300 C to 400 C ± 2.4 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - 400 C to 630 C ± 3.6 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - 630 C to 800 C ± 3.6 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
Temperature - Up to 1200 C ± 3.6 mΩ FLUKE PRECSION
MEASURE MULTIMETER
8846A
TEMPERATURE – Up to 35 C ±0.06 C Fluke 4181
SOURCE Precision
(INFRARED) Infrared
Calibrator
TEMPERATURE – Up to 100 C ±0.06 C Fluke 4181
SOURCE Precision
(INFRARED) Infrared
Calibrator
TEMPERATURE – Up to 200 C ±0.08 C Fluke 4181
SOURCE Precision
(INFRARED) Infrared
Calibrator
TEMPERATURE – Up to 350 C ±0.10 C Fluke 4181
SOURCE Precision
(INFRARED) Infrared
Calibrator
TEMPERATURE – Up to 500 C ±0.16 C Fluke 4181
SOURCE Precision
(INFRARED) Infrared
Calibrator
VOLUME
Quantity Range Best Uncertainty Equipment Used
Volume 100 Ml ±0.0007 mL Pycnometer

LIGHT
Quantity Range Best Uncertainty Equipment Used
Irradiance -SOURCE 300 – 400 nm ±2.7 % RS-10 Series
Spectral
Irradiance-
GAMMA
SCIENTIFIC
Irradiance -SOURCE 400 – 930 nm ±1.8 % RS-10 Series
Spectral

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


Irradiance-
GAMMA
SCIENTIFIC
Irradiance -SOURCE 930 – 1200 nm ±1.4 % RS-10 Series
Spectral
Irradiance-
GAMMA
SCIENTIFIC
Illuminance (Lux) - Throughout Range ±0.74 % RS-10 Series
SOURCE Spectral
Irradiance-
GAMMA
SCIENTIFIC
Color Temperature Throughout Range ±7K RS-10 Series
K) -SOURCE Spectral
Irradiance-
GAMMA
SCIENTIFIC
SOUND
Quantity Range Best Uncertainty Equipment Used
Sound - SOURCE 75, 80 85, 90, 95, 100, 105 dB @ 125 to ±0.006 THD ACS-101
20000 Hz (HF Phones) Audiometer
Calibration
System
Sound - SOURCE 75, 80 85, 90, 95, 100, 105 dB @ 125 to ±0.007 THD ACS-101
20000 Hz (Bone Level) Audiometer
Calibration
System
Sound - SOURCE 75, 80 85, 90, 95, 100, 105 dB @ 125 to ±0.004 THD ACS-101
20000 Hz (Electrical) Audiometer
Calibration
System
Sound - SOURCE 94 dB ± 0.75 dB @ A & C Filters ±2 % SC-942 Sound
Calibrator
Sound - SOURCE 114 dB ± 0.9 dB @ A & CY Filters ±5 % SC-942 Sound
Calibrator
Sound - MEASURE 0 to >155 dB re 20 μPa ±0.20 dB 377A13 Precision
Condenser
Microphone
VIBRATION
Quantity Range Best Uncertainty Equipment Used
Vibration - SOURCE 1.000 m/s2 @ 15.92 to 1280.00 Hz ± 1% VC 21 Vibration
Calibrator
Metra-Mess

Issued date: 15 Dec 18 QF-62


QF-62 Rev: 01

SCOPE AND CMC OF MASTER STANDARDS


Vibration - SOURCE 2.000 m/s2 @ 15.92 to 1280.00 Hz ± 1% VC 21 Vibration
Calibrator
Metra-Mess
Vibration - SOURCE 5.000 m/s2 @ 40.00 to 1280.00 Hz ± 1% VC 21 Vibration
Calibrator
Metra-Mess
Vibration - SOURCE 10.000 m/s2 @ 80.00 to 1280.00 Hz ± 1% VC 21 Vibration
Calibrator
Metra-Mess
Vibration - SOURCE 20.000 m/s2 @ 159.20 to 1280.00 Hz ± 1% VC 21 Vibration
Calibrator
Metra-Mess

Issued date: 15 Dec 18 QF-62