You are on page 1of 13

SECTION 8.

SERVICE GUIDES
C-ARMIIMS2760
IMS 2760 IMAGE SYSTEM OPERATIONAL PARAMETERS
AND SERVICE MENUS

INTRODUCTION:

This section of the image system manual was prepared to provide the service engineer the default
values, operational parameters of the image system, and to assist i n the adjustment of document
cameras, and monitors w i t h test patterns.

NOTE: THIS SECTION PROVIDES GUIDES FOR SERVICE AND IS NOT INTENDED TO BE THE
SOLE MEANS OF TROUBLESHOOTING THE C-ARM AND IMS2760 DIGITAL SYSTEM.

IMS2760 I M A G E S Y S T E M S E R V I C E M O D E :
To enter the service mode the Service Engineer must start the system and w h e n fully operational i n c l u d i n g the
C - A R M communication follow the steps below.

Press and h o l d SHIFT key, n o w press the E D G E key.

Note: The E D G E key w i l l be located i n the lower left corner of the monitor cart keyboard and w h e n
pressed w i l l respond to activate the service mode.

System option keys are shown i n dotted lines. The E D G E is filled i n this picture to highlight
the k e y .
ACQUISITION

I
SHIFT MSA
EDGE WNDW ! FRAME
Max-Op I RATE I
KEY I
L '
RSA
Road | NOISE VCR ! I MASK ] i CINE
Map I 1 1

^ A Ji. .

LIH TEXT I CINE


DSA NOISE ON/OFF | RESET I
I > L _ J
Acquisition section of the monitor keyboard control panel

Once y o u have entered the service mode the right monitor w i l l display a menu w h i c h list the options
for operational default selections and service related functions.

NOTE: The system default files are set at the time they leave the factory at IMS and should not be
changed until the operator has a chance to use the system. It is recommended to write down all
of the system default values before first operation by user as a reference.

The service mode allows a great deal of custom setting to the systems operations to meet the
needs of the operator/user, however it should be noted that miss adjustment will cause image
degradation and lead to poor image quality. Therefore please read the following information
and write down all the data on the screens after installation for future reference.

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 -1
EXIT R U N - T I M E S O F T W A R E :
It may become necessary to exit the system run-time software to load or otherwise check software files
and perform certain system operations w h i c h require the run-time software to be disabled. Therefore
follow the instructions below to exit the run-time software.

THIS S H O U L D O N L Y B E D O N E B Y S O M E O N E T H A T U N D E R S T A N D S F U L L Y T H E O P E R A T I O N O F
BOTH T H E IMAGE SYSTEM A N D THE C - A R M

1) T u r n off the C - A r m system including the image system. O p e n rear cover of the monitor
cart. Remove the two 6 m m alien bolts from the image system mounting bracket and
gently swing open the mounting bracket being careful not to damage interconnecting
cables. See Section 7 of IMS2760 service manual for location of connectors.

2) Install V G A or T T L monitor depending on system configuration.


V G A card w i l l be installed on all systems that include R D C 2000 board.
M G P or T T L card w i l l be installed on all systems that do not have a R D C 2000.

3) Connect A T style keyboard to Image System keyboard connector. See Section 7 of


PMS2760 service m a n u a l for location of connector.
Note: If system is equipped w i t h a RDC2000 board the key board connector
w i l l be a PC-2 type connector. Adapter f r o m your keyboard maybe
necessary to connect it to the Image System.

4) T u r n on the system using the power o n key of the C - A r m or the monitor cart. Wait until
the system is fully operational then press the N E W P A T I E N T K E Y w a i t for right
monitor to display the entry menu. N o w press and h o l d the SHIFT key, then press the
D S A key.

Note: The D S A w i l l be located i n the lower left corner of the monitor cart
keyboard depending on the optional equipment of the IMS2760 image
system the D S A key symbol maybe or m a y not be visible on the
keyboard overlay printing. H o w e v e r the key is active under software
operational control and if pressed w i l l respond to activate the service
mode.

5) The system w i l l exit to the C:IMS2760> directory. Further instructions w i l l be given on


a case b y case bases for use of software upgrades and file transfers if they become
necessary.

CAUTIONS AND WARNINGS:

SYSTEM FILES:
Call IMS national service before attempting to make nay changes to the
system files or otherwise copy files to or from the system hard drives.

The system uses very highly advanced read/ write routines for files and images which
do not require the DOS operating system and therefore files and images are not visible
to normal DOS commands.

Therefore any attempts to view copy or otherwise perform tasks not authorized by IMS
will result in system failure and or damage to the system hard drives and void
warranty.

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 - 2
SECURITY:
The image system is equipped with a security system . any attempts to delete files and
or add files to the system may compromised this security system and cause the system
to be inoperable. The files from one system can not be installed on a another without
causing the system to fail. Please do not attempt to do so as damage to the system will
result voiding warranty.

COPYING OTHER SOFTWARE PROGRAMS ON TO THE SYSTEM:


PLEASE DO NOT ATTEMPT TO COPY PROGRAMS ON TO THE IMAGE SYSTEM .

THE SYSTEM IS VERY SPECIFIC IN THE AREAS THAT CAN HAVE DOS FILES
AND WERE IT CAN NOT.

THE IMAGE SYSTEM FILES WILL BE OVER WHITEN IF YOU ATTEMPT TO COPY
OTHER SERVICE RELATED TYPE SOFTWARE PROGRAMS CAUSING A SYSTEM
FAILURE .

IF ANY ATTEMPTS ARE MADE TO COPY FILES OTHER THAN THOSE SO


DESIGNATED BY IMS TO THE SYSTEM WILL VOID ALL WARRANTY OF THE
SYSTEM.

SERVICE SOFTWARE SETUP DISPLAY MENUS:

The service software is used for setting the system operational parameters and user preference items. The
software is broken u p into six parts as listed below.

To enter a selection press the corresponding number on the monitor cart keyboard and then press E N T E R

1) W N D W G a i n and Offset

2) Noise Reduction, L I H

3) Default Files and System Parameters

4) Test Patterns

5) D S A , M S A , R S A , L o o k u p Tables

6) Update System Software

The following information is listed i n order of the selection number on the main service menu.

N O T E : Only train Service Engineers who understand and are trained should attempt to use the
following information. Incorrect entry of data will cause system failure and or poor image
quality.

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 - 3
DESCRIPTION OF SERVICE SOFTWARE SELECTIONS:

1) G A I N and OFFSET

The gain and offset functions of the image processor provide the necessary calibration to match the
analog to digital converter's range to the amplitude and level of the video input.

The Offset adjustment is used to set the A / D converter's lower range to the black level i n the video
signal. A n y signal below this level w i l l result i n all pixel values being converted to black until the
video signal rises above the offset value.

The G a i n is used to match the m a x i m u m signal amplitude (white level) to the A / D ' s m a x i m u m
conversion capacity (i.e. 10 bit = 1024 gray scale).

EFFECTS OF GAIN AND OFFSET:

The following steps 1-15 must be performed on the I M S 2000 / 2760 Imaging System prior to adjusting
the G A I N / O F F S E T values.

UPS 2000 Output Levels:

WARNING
Exercise extreme caution when measuring on the VIP2000 and UPS2000 boards. Make sure power is off
before attaching meter and scopes to the board sets as permanent damage will occur if components are
inadvertently shorted. The subsequent shorting of one of the boards can be detected when boards are
inspected.

The following adjustments w i l l make sure the image system input and output video is calibrated properly.

1. T u r n off power to system and open image system to gain access to the image boards. See section 7.0 of the
IMS2760 service manual for location of boards. See also the section 8 for location of test points.

2. Energize system.

3 Enter the Service M e n u by h o l d i n g d o w n the Shift Key while pressing the "Edge K e y " .

4. From the Service M e n u , select item # 3. wait then press # 0 u n i t l l M a s k Circle Disabled is selected.

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2-4
5. Press E N T E R key then press E S C A P E untill y o u return to the m a i n sevice menu. N o w press # 4 Test
Patterns, then select # 4 (Step Wedge).

6. Press the Escape K e y twice to return to normal operation mode and press the M O N / L - R key to transfer
the image to the right side monitor. The Step Wedge image should n o w appear on both monitors (left
and right).

7. Connect oscilloscope to the video output of the left monitor, leaving the 75 ohm termination switch
"on". A step wedge going u p on both sides of the oscilloscope w i l l be seen.

8. Adjust potentiometer R43 on U P 2000 board for a signal of 1.02 V p - p .

9. Connect oscilloscope to the video output of the right monitor, leaving the 75 ohm termination switch
"on". A step wedge going u p on both sides of the oscilloscope w i l l be seen.

10. Adjust potentiometer R40 on UPS2000 board for a signal of 1.02 V p - p .

11. Adjust potentiometer R45 to the m i n i m u m value ~ 1.1V at J16 terminated.


(Higher value due to A d d i t i o n a l equipment attached to this line. i.e. documentation camera, Paper
printer, D I C O M , Laser printers etc ) N o w go to step 4 and turn the M a s k Circle to Enabled.

Left & Right Video BNC output Level

V I P 2000 D . C . Offset a n d G a i n A / D Convertor Calibration For f u l l Scale Conversion.

WARNING
Exercise extreme caution when measuring on the VIP2000 and UPS2000 boards . Make sure power is off
before attaching meter and scopes to the board sets as permanent damage will occur if components are
inadvertently shorted.

Please refer to section 8.3 of the IS 2760 service manual for location of test points.

12. O n the VIP2000 board, connect D C voltmeter between W 6 (gnd) and W22 (DC gain).

13. Adjust R l 12 for a reading of 0.54 V d c

14. O n the VIP2000 board, connect D C voltmeter between W 6 (gnd) and W19 (offset).

15. Adjust R92 for a reading of 2.64 V d c

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 - 5
DIMENSIONS IN INCHES
13.075
12.750
4.700

•HFS1i t
LEVEL 1

OP
• •
! If 5 5Wc IM • F M

cm| I
4 ^
lT
t — n n , • FM
MM ?
ID
I - - o
BBS"
«gUoL
cff
W7 RU 894C42 RH7 • '

•09
r:;J am

4.50 4.10 r -i r
CWj I Q w
__pjg II M
UJD- •
| C'"_f~~1

Q mUr- 4.250
uw «4 R m • *

U<9 U47
"'7-2 in
U7»

• *

n
UN
U7
w.150fr
U»7 , ,
C1SI -[ ~[-
Hi?
1.250 j
H U73

Li: .300
.200
Dc -^IU^S wac .SHVJX.
2) LIVE NOISE REDUCTION

The system incorporates weighted frame averaging to help reduce noise i n the live and last image
hold. The system employs an averaging technique termed "exponential averaging". This type of
averaging w i l l reduce from 50 % to 87% of the noise i n the live video image depending o n the selected
value, i.e. 2,3,4,5,6,7,8 or 16 frames. Therefore the service software allows the service engineer to
select the value of noise reduction best suited for the customer as a default setting w h e n the system turns
on. The selection of the image noise reduction is accomplished by selecting " 2 " then Pressing E N T E R on
the monitor cart control keyboard.

The operator is allowed to select from the control keyboard L O W , M E D , H I G H noise reduction w h i c h
i n turn is a value set by the service engineer i n the service menu. The value for L O W , M E D , or H I G H can
be set to any value from 2 to 16 frames. The best results are given below for a new system.

LOW = 3 MED= 5 H I G H = 16 These values are set at the factory w h e n the system is shipped.

SETTING T H E N E W V A L U E :

Select the number from list on the screen b y pressing the corresponding number select function
L O W , M E D , H I G H . Then enter value for noise reduction by pressing the numerical keys for number
of video frames to be averaged, i.e. 2, 3,4,5, - 16 needed then press the E N T E R key to save the new value
for noise reduction.

N O T E : Selection 9 provides a base level of image noise reduction for system even if operator select no
N R reduction. T h i s base value s h o u l d be 2 to 3 frames but no less than 2. T h i s w i l l l i m i t a l l settings of
noise reduction N R , and L I H from going below the m i n i m u m set b y this selection.

NOTE: When all entries are complete press ENTER to save the screen and it's new values to digital
disk

LAST I M A G E H O L D NOISE REDUCTION

L I H NOISE:
Refers to the function used a the end of a fluoroscopy w h e n the radiation h a n d / f o o t s w i t c h is
released and the processing system averages images to reduce noise i n the image to provide better
image quality.

The L A S T I M A G E H O L D feature of noise reduction allows the operator to select a different value
of noise reduction w h e n the D r . releases the fluoroscopic footswitch this type of noise reduction is called
a finite impulse response filter . The system averages image together w i t h a fixed pre-set number of
video image frames.

E X A M P L E : L I H set for four (4) image. The D R . releases the footswitch, C - A r m sends a signal to the
IMS2760 digital system to start the L I H mode. The image processor sets the value of image to the 4
images preselected then accumulates them i n the image buffers and applies the L I H algorithm to the
images i n the buffer. The results are then displayed on the process monitor and image system sends
signal to the C - A r m to stop the fluoroscopic radiation . The length of h o l d on of radiation is determined
by the number of images selected for last image hold rule of thumb is 8 frames of integration takes about
300 to 400ms.

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 - 6
3. D E F A U L T FILES AND SYSTEM PARAMETERS
These Sub- menus set the system parameters and operators preferences for certain operations. The
following list shows the default setting i n the service program.
1) Screen Alignments
2) Image System Parameters
3) System Alerts
4) D S A / C I N E Parameters
5) C - A r m Parameters

1) Screen Alignments

1) Text Circle Radius "enter value" [ 238]

2) Auto Window Circle Radius "enter value" [130]

3) Screen Center X axis "enter value" [512]

4) Screen Center Y axis "enter value" [254]

5) Mask Circle Radius " enter value" [239]

6) Digital Document Rotation Radius "enter value" [ 242]


7) Digital Document Rotation " D D R " = [Enabled]
" toggle using 7 key"

8) Rotation Degree =1 "enter value" [5]


maximum 30 degrees

9) Rotation & Shift K e y =15 "enter value" [30]


maximum 90 degrees"

0) Mask Circle = Enabled/disabled " toggle using 0 key"

2) Image System Parameters


1) (Auto Clear Text During Fluoro) or (Manually Turn O f f Text)
" toggle using 1 key "
2) (Live Monitor Transfer) or (Simple Monitor Copy)
"toggle with 2 key"

3) Last Image Hold Off (LIH O F F ) on/off "toggle with 3 k e y " [Off]

4) Menu Time-out "enter value" [989]

5) Display Dose : ( k V , mA, Time) or ( C G y C m 2 ) or ( k V , m A ,


C G y C m 2 ) "toggle using 5 key"
6) Sharp L o w " enter value" [10]
7) Sharp High " enter value" [25]

8) L I H Edge Enhanced = O F F / L O W / H I G H "toggle using 1 key"


Note off / low / high to add image enhancement when foot pedal is
released by operator. [OFF]

9) Edge " enter value" [1]

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 - 7
3) System Alerts
1) Radiation Alert = on/off " toggle using 1 key" [On]

2) D S A Alert = on/off " toggle using 2 key" [On]

3) Store Alert = on/off " toggle using 3 key" [On]

4) Confirm Image Delete = on/off "toggle using 4 key" [On]

4) DSA/CINE Parameters

1) D S A Noise Filter "enter value" [16]

2) Protect Cine Run = on/off " toggle using 2 k e y " [On]

3) Default Frame Rate "enter value" [4]

4) Cine Auto
toggle Store:
using (Allowed), or (Always), or (Never)
4 key'

5) C-Arm Control Parameters


1) (Adjust Contrast A t C - A r m ) , or (Recall Images A t C - A r m ) , or
(Auto Select Contrast/Recall) " toggle using 1 k e y "

2) Pulsed fluoro off time "enter value" [200]


Sets the radiation off time between pulses to control pulse rate.

3) Pulsed Auto Store = on/off " toggle using 3 key " [On]
When Pulsed Fluoro Auto Store is on and auto store is on. The last
image held when operator releases the footswitch is stored to hard
disk. Auto Trans and Both work the same as to day in pulse.

4) 1 st D S A Delay "enter value" [3984]


Sets the time needed for image stabilization to lock in k V , m A and
mask acquire time for the first.

5) 2nd D S A Delay "enter value" [498]


Sets the delay for the next D S A acquisition. The second time D S A
comes on and the operator does turn off D S A and stays in the same
position the system only requires a small delay for mask acquisition.

6) Cine C - A r m Auto Tech L o c k = on/off [Off]


(When selected to off the image system w i l l not send signal to
lock in technique factors of generator k V / m A . The C I N E acquisition
first delay and second delay will also be deactivated).

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2-8
4. TEST PATTERNS

The test patterns are used for setting the monitor alignment and system testing.

Select the number "5 TEST P A T T E R N S " from the menu. Then one of the
following

1) G r i d
Provides a grid pattern to adjust the monitor linearly

2) Circle
Provides a circle that should be adjusted to 24.5 c m V / H for proper
image size on the monitor. The large circle size should be set w i t h
monitor adjustments for correct size and linearly then the Video Camera of the
C - A r m should be adjusted for size / linearly. The camera image circle size
should be the same as the circle of the test pattern. To check video image size
press "7" on the key board and live video w h e n footswitch is pressed w i l l is
displayed on the Left monitor. To adjust right monitor reverse video cables from
left and right monitors.

3) G r a y scale
Provides a continues B L K to W H T gradient)

4) Step wedge
Provides a uniform step wedge of a value from B L K to W H T of 11 steps.
This is used to set the video output level of the UPS2000 board and to calibration
the optional D I C O M interface.

5) Resolution
Provides a resolution test for monitor and hard copy camera

6) Test Pattern
Provides a combination of gray scale, resolution, and grid pattern for setting
monitors and hard copy camera images)

7) L i v e video input
Provides a live image under the circle and grid patterns for adjustment of the
monitors.

8) Freeze video input


Provides the service engineer the ability to freeze the image for v i e w i n g
without constant fluoroscopic radiation on. Press the " 8 " key when y o u need to
freeze an image for v i e w i n g

SPECIAL SELECTION o> -Ok ^ W * .


9) C o p y B M P file to floppy
Provides the operator w i t h the ability to save images to a floppy disk, 4 images
can be stored on empty I B M formatted floppy disk

C o p y B I N A R Y file to f l o p p y
USED O N L Y TO S E N D I M A G E S OF F U L L I M A G E G R A Y S C A L E T O
T H E F A C T O R Y FOR E V A L U A T I O N OF I M A G E Q U A L I T Y
C o p y to H A R D D I S K
USED O N L Y TO SEND IMAGES OF FULL I M A G E G R A Y SCALE TO

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 - 9
TO T H E F A C T O R Y FOR E V A L U A T I O N OF I M A G E QUALITY

NOTE: WITH SELECTION NUMBER "9" PRESS THE "9" KEY


REPEATEDLY UNTIL THE PROPER SELECTION IS DISPLAYED ON
THE MONITOR THEN PRESS ENTER TO SAVE.

ALWAYS LEAVE THE SYSTEM SET FOR "BMP" FILE COPY


FOR NORMAL OPERATION .

T o exit the test pattern mode press E S C the k e y to exit back to the service m e n u

5) DSA,MSA,RSA, Look Up Table:

Provides the necessary image processing parameters to set the 16 bit A L U in the R D C 2 0 0 0
board to adjust the L U T curve to enhance the difference in the low black levels of the
subtracted image to clearly visualize the injection contrast media over low contrast areas of
bone and dense tissue.

Each of the different Subtracted functions or modes could require based on the operators
preferences a sperate and defined L U T curve and W I N D O W L E V E L value.

1) D S A Look U p Table

2) M S A Look U p Table

3) R S A Look U p Table

CAUTION: Adjustment of curve can cause sever image distortion and poor
image quality if adjustments are performed incorrectly.
It is therefore recommended that service engineers contact IMS
national Service before any attempts are made to change the LUT
curve values. 7>vhc- Ji^h ^p ^ -u^Ju^ -gjum &-4J*-^- - i * * * - * * * ^ <L&*^^Y

To adjust the curve values press the U P / D o w n key to raise or lower a values on the curve. Use
the R I G H T / L E F T arrow keys to move the point on the curve to raise or lower.

Press the W N D W key to adjust the monitor brightness /contrast level. Use U P / D O W N arrow
keys to adjust brightness ( O F F S E T ) . Use The R I G H T / L E F T arrow keys to adjust Contrast
(GAIN).

Press Escape to exit without saving the new values.

Press Enter to save new values.

Press the fluoroscopy footswitch to view image on the left monitor for adjustment.

NOTE: Y o u must select Noise reduction and edge enhancement before you enter service
program to be able to view the process values in the D S A , M S A , R S A service
adjustment procedures.

6) Update, System Software:


Future use to allow system software upgrades and version releases.

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 - 10
C-ARM POWER TURN
CN
IMS 2760 IMAGE SYSTEM I BOOT-UP FLOW CHART
OK

BIOS TEST
C P U / V OTHER ERROR WAIT J REPAIR SYSTEM
BOAR1 D AND i . e . M E M O R Y , etc. "I PROBLEM
MEV ORY DISPLAYED ON TTL

OK

BIOS
HDD/CONTROLLER •

OK BIOS CONTROLLER OF
EPROM SOFTWARE CARD FAIL •
FCC/CONTROLLER
BOOTS PAST THE FLOPPY
AND SIMULATES THE A: DRIVE
YES^

BIOS CONFIGURATION
SCREEN DISPLAYED ON ERROR WAIT
TTL MONITOR i.e., HDD failure, etc.
DISPLAYED ON TTL MONITOR T H E SYSTEM
WILL BY-PASS IN 20 SEC OR SO AND DISPLAY
BIOS CONTROLS DOS BOOT A N ERROR MESSAGE ON THE SCREEN
SEQUENCE AND THEN TRY TO BOOT. IF IT C A N NOT FIND
T H E DOS ON A: OR C: DRIVE IT WILL DISPLAY
THEMES AGE
DRIVE NOT READY
INSTALLED FLOPPY INSERT BOOT DISKETTE IN A:
DOS WITH PRESS ANY KEY TO CONTINUE
COMMAND.COM
DOS CHECKS FOR
A: or C:
"CONFIG.SYS" FILE

REPAIR SYSTEM
SOFTWARE DISPLAYS ON MONITOR
PROBLEM
Mor>_stand =1: Lamp=l;
YES •
VER IMS2760.00, September 15,1997: Test
Fault -Line is Active!
Fluoro-Prog 0 O N
DOS CONFIGURES THE - Fluoroscopy ON
SYSTEM FOR DEVICES HalfDose ON
Left Monitor Active
Fluoro-Prog 0 Off
Fluoro- Prog 1 ON
HalfDose Off
DOS LOADS T HE COMMAND
PROC 3SSOR SYSTEM IS NOW OPERATIONAL AND VIDEO
MONITORS DISPLAY WHITE CIRCLE
TEXT & C - A R M CONTROL PANEL IS ON
SOFTWARE TEST
THE COMMAND RDC2000 BOARD
PROCESSOR LOOKS FOR
SYSTEM STOPS
A N "AUTOEXEC.BAT' FILE
DISPLAY FAILS SOFTWARE TEST SYSTEM CONTINUES TO SYSTEM CONTINUES TO RUN
1. DOS PROMPT <•
I/O BOARD RUN IMS2760.EXE IMS2760.EXE PROGRAM
2. NO MESSAGE PROGRAM UNTIL UNTIL POWER IS TURNED OFF
SYSTEM HANGS SYSTEM SHUT DOWN IS
SOFTWARE TEST PERFORMED BY USER IN
UPS200 BOARD SOFTWARE.

SYSTEM OFF
END
SOFTWARE TEST
VIP 2000 BOARD
BJS4S9 SOFTWARE
1. CLOSE A L L M G E FILES
2. CLOSE A L L PATIENT DATA
MONITOR DISPLAY FILES
CYLINDERS 1008, 3. EXIT TO DOS ENVIROMENT
T H E COMMAND PROCESSOR T H E COMMAND PROCESSOR
SECTORS 55, COMMAND LINE INPUT
EXECUTES THE DISPLAYS THE DOS TIME AND
HEADS 5,
" A U T O E X E C . B A T ' FILE D A T E PROMPTS
M B Y T E 162.3

DOS ENVIROMENT
DOS ENVIROMENT
COMMAND U N E INPUT
COMMAND U N E INPUT
~T~
T H E "AUOTEXEC.BAT FILE USER TURNS
1. CHANGES TO C: DRIVE OFF POWER TO
IF SYSTEM C A N NOT OPEN
2. OPENS THE IMS2760 C-ARM
OR FIND A FILE, OR
DIRECTORY DIRECTORY THE MONITOR
3. STARTS T H E IMS27600.EXE
WILL DISPLAY ERROR
PROGRAM ON THE C: DRIVE
MESSAGE AND EXIT TO DOS SYSTEM OFF
END

Exposcop 7000 / IMS2760 Section 8.9 09/21/97


International Medical Systems Service Guides Rev.2 -11

You might also like