You are on page 1of 262

GE Healthcare

Definium™ 6000
System Schematics and Drawings

Direction 5135611-100
Revision 4
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Page 2
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

LEGAL NOTES
TRADEMARKS
All other products and their name brands are trademarks of their respective holders.

COPYRIGHTS
All Material Copyright © 2005 ~ 2007 by General Electric Company, Inc. All rights reserved.

Page 3
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

This page is intentionally left blank.

Page 4
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Revision History

Revision Date Reason for change


1 Nov. 1, 2005 Initial Release.
2 Jun. 20, 2006 Update Product Name.
3 Nov. 30, 2006 Update Drawings
4 May 10, 2007 Update Drawings

Page 5
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

This page is intentionally left blank.

Page 6
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

IMPORTANT PRECAUTIONS

Page 7
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

LANGUAGE

警告 • 本维修手册仅提供英文版本。

(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。

VÝSTRAHA • TENTO PROVOZNÍ NÁVOD EXISTUJE POUZE V ANGLICKÉM JAZYCE.


(CS) • V PŘÍPADĚ, ŽE EXTERNÍ SLUŽBA ZÁKAZNÍKŮM POTŘEBUJE NÁVOD V JINÉM JAZYCE, JE ZAJIŠTĚNÍ PŘEKLADU DO ODPOVÍDAJÍCÍHO JAZYKA ÚKOLEM ZÁKAZNÍKA.
• NESNAŽTE SE O ÚDRŽBU TOHOTO ZAŘÍZENÍ, ANIŽ BYSTE SI PŘEČETLI TENTO PROVOZNÍ NÁVOD A POCHOPILI JEHO OBSAH.
• V PŘÍPADĚ NEDODRŽOVÁNÍ TÉTO VÝSTRAHY MŮŽE DOJÍT K PORANĚNÍ PRACOVNÍKA PRODEJNÍHO SERVISU, OBSLUŽNÉHO PERSONÁLU NEBO PACIENTŮ VLIVEM ELEKTRICKÉHOP
PROUDU, RESPEKTIVE VLIVEM MECHANICKÝCH ČI JINÝCH RIZIK.

ADVARSEL • DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK.


(DA) • HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE.
• FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE DENNE SERVICEMANUAL HAR VÆRET KONSULTERET OG ER FORSTÅET.
• MANGLENDE OVERHOLDELSE AF DENNE ADVARSEL KAN MEDFØRE SKADE PÅ GRUND AF ELEKTRISK, MEKANISK ELLER ANDEN FARE FOR TEKNIKEREN, OPERATØREN ELLER
PATIENTEN.

WAARSCHUWING • DEZE ONDERHOUDSHANDLEIDING IS ENKEL IN HET ENGELS VERKRIJGBAAR.


(NL) • ALS HET ONDERHOUDSPERSONEEL EEN ANDERE TAAL VEREIST, DAN IS DE KLANT VERANTWOORDELIJK VOOR DE VERTALING ERVAN.
• PROBEER DE APPARATUUR NIET TE ONDERHOUDEN VOORDAT DEZE ONDERHOUDSHANDLEIDING WERD GERAADPLEEGD EN BEGREPEN IS.
• INDIEN DEZE WAARSCHUWING NIET WORDT OPGEVOLGD, ZOU HET ONDERHOUDSPERSONEEL, DE OPERATOR OF EEN PATIËNT GEWOND KUNNEN RAKEN ALS GEVOLG VAN EEN
ELEKTRISCHE SCHOK, MECHANISCHE OF ANDERE GEVAREN.

WARNING • THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY.


(EN) • IF A CUSTOMER'S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMER'S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES.
• DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
• FAILURE TO HEED THIS WARNING MAY RESULT IN INJURY TO THE SERVICE PROVIDER, OPERATOR, OR PATIENT FROM ELECTRIC SHOCK, OR FROM MECHANICAL OR OTHER HAZARDS.

HOIATUS • KÄESOLEV TEENINDUSJUHEND ON SAADAVAL AINULT INGLISE KEELES.


(ET) • KUI KLIENDITEENINDUSE OSUTAJA NÕUAB JUHENDIT INGLISE KEELEST ERINEVAS KEELES, VASTUTAB KLIENT TÕLKETEENUSE OSUTAMISE EEST.
• ÄRGE ÜRITAGE SEADMEID TEENINDADA ENNE EELNEVALT KÄESOLEVA TEENINDUSJUHENDIGA TUTVUMIST JA SELLEST ARU SAAMIST.
• KÄESOLEVA HOIATUSE EIRAMINE VÕIB PÕHJUSTADA TEENUSEOSUTAJA, OPERAATORI VÕI PATSIENDI VIGASTAMIST ELEKTRILÖÖGI, MEHAANILISE VÕI MUU OHU TAGAJÄRJEL.

Page 8
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

VAROITUS • TÄMÄ HUOLTO-OHJE ON SAATAVILLA VAIN ENGLANNIKSI.


(FI) • JOS ASIAKKAAN HUOLTOHENKILÖSTÖ VAATII MUUTA KUIN ENGLANNINKIELISTÄ MATERIAALIA, TARVITTAVAN KÄÄNNÖKSEN HANKKIMINEN ON ASIAKKAAN VASTUULLA.
• ÄLÄ YRITÄ KORJATA LAITTEISTOA ENNEN KUIN OLET VARMASTI LUKENUT JA YMMÄRTÄNYT TÄMÄN HUOLTO-OHJEEN.
• MIKÄLI TÄTÄ VAROITUSTA EI NOUDATETA, SEURAUKSENA VOI OLLA HUOLTOHENKILÖSTÖN, LAITTEISTON KÄYTTÄJÄN TAI POTILAAN VAHINGOITTUMINEN SÄHKÖISKUN, MEKAANISEN
VIAN TAI MUUN VAARATILANTEEN VUOKSI.

ATTENTION • CE MANUEL DE SERVICE N'EST DISPONIBLE QU'EN ANGLAIS.


(FR) • SI LE TECHNICIEN DU CLIENT A BESOIN DE CE MANUEL DANS UNE AUTRE LANGUE QUE L'ANGLAIS, C'EST AU CLIENT QU'IL INCOMBE DE LE FAIRE TRADUIRE.
• NE PAS TENTER D'INTERVENIR SUR LES EQUIPEMENTS TANT QUE LE MANUEL SERVICE N'A PAS ETE CONSULTE ET COMPRIS
• LE NON-RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN, L'OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES, MÉCANIQUES OU AUTRES.

WARNUNG • DIESE SERVICEANLEITUNG EXISTIERT NUR IN ENGLISCHER SPRACHE.


(DE) • FALLS EIN FREMDER KUNDENDIENST EINE ANDERE SPRACHE BENÖTIGT, IST ES AUFGABE DES KUNDEN FÜR EINE ENTSPRECHENDE ÜBERSETZUNG ZU SORGEN.
• VERSUCHEN SIE NICHT DIESE ANLAGE ZU WARTEN, OHNE DIESE SERVICEANLEITUNG GELESEN UND VERSTANDEN ZU HABEN.
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH STROMSCHLÄGE,
MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.

ΠΡΟΕΙΔΟΠΟΙΗΣΗ • ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ ΔΙΑΤΙΘΕΤΑΙ ΣΤΑ ΑΓΓΛΙΚΑ ΜΟΝΟ.


(EL) • ΕΑΝ ΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ ΕΝΟΣ ΠΕΛΑΤΗ ΑΠΑΙΤΕΙ ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕ ΓΛΩΣΣΑ ΕΚΤΟΣ ΤΩΝ ΑΓΓΛΙΚΩΝ, ΑΠΟΤΕΛΕΙ ΕΥΘΥΝΗ ΤΟΥ ΠΕΛΑΤΗ ΝΑ ΠΑΡΕΧΕΙ ΥΠΗΡΕΣΙΕΣ
ΜΕΤΑΦΡΑΣΗΣ.
• ΜΗΝ ΕΠΙΧΕΙΡΗΣΕΤΕ ΤΗΝ ΕΚΤΕΛΕΣΗ ΕΡΓΑΣΙΩΝ ΣΕΡΒΙΣ ΣΤΟΝ ΕΞΟΠΛΙΣΜΟ ΕΚΤΟΣ ΕΑΝ ΕΧΕΤΕ ΣΥΜΒΟΥΛΕΥΤΕΙ ΚΑΙ ΕΧΕΤΕ ΚΑΤΑΝΟΗΣΕΙ ΤΟ ΠΑΡΟΝ ΕΓΧΕΙΡΙΔΙΟ ΣΕΡΒΙΣ.
• ΕΑΝ ΔΕ ΛΑΒΕΤΕ ΥΠΟΨΗ ΤΗΝ ΠΡΟΕΙΔΟΠΟΙΗΣΗ ΑΥΤΗ, ΕΝΔΕΧΕΤΑΙ ΝΑ ΠΡΟΚΛΗΘΕΙ ΤΡΑΥΜΑΤΙΣΜΟΣ ΣΤΟ ΑΤΟΜΟ ΠΑΡΟΧΗΣ ΣΕΡΒΙΣ, ΣΤΟ ΧΕΙΡΙΣΤΗ Ή ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ,
ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ.

FIGYELMEZTETÉS • EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL.


(HU) • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE.
• NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
• EZEN FIGYELMEZTETÉS FIGYELMEN KÍVÜL HAGYÁSA A SZOLGÁLTATÓ, MŰKÖDTETŐ VAGY A BETEG ÁRAMÜTÉS, MECHANIKAI VAGY EGYÉB VESZÉLYHELYZET MIATTI SÉRÜLÉSÉT
EREDMÉNYEZHETI.

AÐVÖRUN • ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU.


(IS) • EF AÐ ÞJÓNUSTUVEITANDI VIÐSKIPTAMANNS ÞARFNAST ANNAS TUNGUMÁLS EN ENSKU, ER ÞAÐ SKYLDA VIÐSKIPTAMANNS AÐ SKAFFA TUNGUMÁLAÞJÓNUSTU.
• REYNIÐ EKKI AÐ AFGREIÐA TÆKIÐ NEMA AÐ ÞESSI ÞJÓNUSTUHANDBÓK HEFUR VERIÐ SKOÐUÐ OG SKILIN.
• BROT Á SINNA ÞESSARI AÐVÖRUN GETUR LEITT TIL MEIÐSLA Á ÞJÓNUSTUVEITANDA, STJÓRNANDA EÐA SJÚKLINGS FRÁ RAFLOSTI, VÉLRÆNU EÐA ÖÐRUM ÁHÆTTUM.

Page 9
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

AVVERTENZA • IL PRESENTE MANUALE DI MANUTENZIONE E DISPONIBILE SOLTANTO IN INGLESE.


(IT) • SE UN ADDETTO ALLA MANUTENZIONE ESTERNO ALLA GEMS RICHIEDE IL MANUALE IN UNA LINGUA DIVERSA, IL CLIENTE E TENUTO A PROVVEDERE DIRETTAMENTE ALLA TRADUZIONE.
• SI PROCEDA ALLA MANUTENZIONE DELL'APPARECCHIATURA SOLO DOPO AVER CONSULTATO IL PRESENTE MANUALE ED AVERNE COMPRESO IL CONTENUTO
• IL NON RISPETTO DELLA PRESENTE AVVERTENZA POTREBBE FAR COMPIERE OPERAZIONI DA CUI DERIVINO LESIONI ALL'ADDETTO ALLA MANUTENZIONE, ALL'UTILIZZATORE ED AL PAZIENTE PER FOLGORAZIONE
ELETTRICA, PER URTI MECCANICI OD ALTRI RISCHI.

• このサービスマニュアルには英語版しかありません。

(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業者の責任で行うものとさせていただきます。

• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでください。

• この警告に従わない場合、サービスを担当される方、操作員あるいは患者さんが、感電や機械的又はその他の危険により負傷する可能性があります。

경고 • 본 서비스 지침서는 영어로만 이용하실 수 있습니다.

(KO) • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우, 번역 서비스를 제공하는 것은 고객의 책임입니다.

• 본 서비스 지침서를 참고했고 이해하지 않는 한은 해당 장비를 수리하려고 시도하지 마십시오.

• 이 경고에 유의하지 않으면 전기 쇼크, 기계상의 혹은 다른 위험으로부터 서비스 제공자, 운영자 혹은 환자에게 위해를 가할 수 있습니다.

BRĪDINĀJUMS • ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ.


(LV) • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU.
• NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS IZLASĪŠANAS UN SAPRAŠANAS.
• ŠĪ BRĪDINĀJUMA NEIEVĒROŠANA VAR RADĪT ELEKTRISKĀS STRĀVAS TRIECIENA, MEHĀNISKU VAI CITU RISKU IZRAISĪTU TRAUMU APKALPES SNIEDZĒJAM, OPERATORAM VAI
PACIENTAM.

ĮSPĖJIMAS • ŠIS EKSPLOATAVIMO VADOVAS YRA PRIEINAMAS TIK ANGLŲ KALBA.


(LT) • JEI KLIENTO PASLAUGŲ TIEKĖJAS REIKALAUJA VADOVO KITA KALBA – NE ANGLŲ, NUMATYTI VERTIMO PASLAUGAS YRA KLIENTO ATSAKOMYBĖ.
• NEMĖGINKITE ATLIKTI ĮRANGOS TECHNINĖS PRIEŽIŪROS, NEBENT ATSIŽVELGĖTE Į ŠĮ EKSPLOATAVIMO VADOVĄ IR JĮ SUPRATOTE.
• JEI NEATKREIPSITE DĖMESIO Į ŠĮ PERSPĖJIMĄ, GALIMI SUŽALOJIMAI DĖL ELEKTROS ŠOKO, MECHANINIŲ AR KITŲ PAVOJŲ PASLAUGŲ TIEKĖJUI, OPERATORIUI AR PACIENTUI.

Page 10
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

ADVARSEL • DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK.


(NO) • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE.
• IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
• MANGLENDE HENSYN TIL DENNE ADVARSELEN KAN FØRE TIL AT SERVICELEVERANDØREN, OPERATØREN ELLER PASIENTEN SKADES PÅ GRUNN AV ELEKTRISK STØT, MEKANISKE
ELLER ANDRE FARER.

OSTRZEŻENIE • NINIEJSZY PODRĘCZNIK SERWISOWY DOSTĘPNY JEST JEDYNIE W JĘZYKU ANGIELSKIM.


(PL) • JEŚLI DOSTAWCA USŁUG KLIENTA WYMAGA JĘZYKA INNEGO NIŻ ANGIELSKI, ZAPEWNIENIE USŁUGI TŁUMACZENIA JEST OBOWIĄZKIEM KLIENTA.
• NIE PRÓBOWAĆ SERWISOWAĆ WYPOSAŻENIA BEZ ZAPOZNANIA SIĘ I ZROZUMIENIA NINIEJSZEGO PODRĘCZNIKA SERWISOWEGO.
• NIEZASTOSOWANIE SIĘ DO TEGO OSTRZEŻENIA MOŻE SPOWODOWAĆ URAZY DOSTAWCY USŁUG, OPERATORA LUB PACJENTA W WYNIKU PORAŻENIA ELEKTRYCZNEGO, ZAGROŻENIA MECHANICZNEGO BĄDŹ
INNEGO.

ATENÇÃO • ESTE MANUAL DE ASSISTÊNCIA TÉCNICA SÓ SE ENCONTRA DISPONÍVEL EM INGLÊS.


(PT) • SE QUALQUER OUTRO SERVIÇO DE ASSISTÊNCIA TÉCNICA, QUE NÃO A GEMS, SOLICITAR ESTES MANUAIS NOUTRO IDIOMA, É DA RESPONSABILIDADE DO CLIENTE FORNECER OS
SERVIÇOS DE TRADUÇÃO.
• NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTÉNCIA TÉCNICA
• O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA DO TÉCNICO, OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS, MECÂNICOS OU OUTROS.

ATENŢIE • ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ.


(RO) • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
• NU ÎNCERCAŢI SĂ REPARAŢI ECHIPAMENTUL DECÂT ULTERIOR CONSULTĂRII ŞI ÎNŢELEGERII ACESTUI MANUAL DE SERVICE.
• IGNORAREA ACESTUI AVERTISMENT AR PUTEA DUCE LA RĂNIREA DEPANATORULUI, OPERATORULUI SAU PACIENTULUI ÎN URMA PERICOLELOR DE ELECTROCUTARE, MECANICE SAU
DE ALTĂ NATURĂ.

ОСТОРОЖНО! • ДАННОЕ РУКОВОДСТВО ПО ОБСЛУЖИВАНИЮ ПРЕДЛАГАЕТСЯ ТОЛЬКО НА АНГЛИЙСКОМ ЯЗЫКЕ.


(RU) • ЕСЛИ СЕРВИСНОМУ ПЕРСОНАЛУ КЛИЕНТА НЕОБХОДИМО РУКОВОДСТВО НЕ НА АНГЛИЙСКОМ, А НА КАКОМ-ТО ДРУГОМ ЯЗЫКЕ, КЛИЕНТУ СЛЕДУЕТ САМОСТОЯТЕЛЬНО
ОБЕСПЕЧИТЬ ПЕРЕВОД.
• ПЕРЕД ОБСЛУЖИВАНИЕМ ОБОРУДОВАНИЯ ОБЯЗАТЕЛЬНО ОБРАТИТЕСЬ К ДАННОМУ РУКОВОДСТВУ И ПОЙМИТЕ ИЗЛОЖЕННЫЕ В НЕМ СВЕДЕНИЯ.
• НЕСОБЛЮДЕНИЕ ТРЕБОВАНИЙ ДАННОГО ПРЕДУПРЕЖДЕНИЯ МОЖЕТ ПРИВЕСТИ К ТОМУ, ЧТО СПЕЦИАЛИСТ ПО ОБСЛУЖИВАНИЮ, ОПЕРАТОР ИЛИ ПАЦИЕНТ ПОЛУЧАТ УДАР
ЭЛЕКТРИЧЕСКИМ ТОКОМ, МЕХАНИЧЕСКУЮ ТРАВМУ ИЛИ ДРУГОЕ ПОВРЕЖДЕНИЕ.

UPOZORNENIE • TENTO NÁVOD NA OBSLUHU JE K DISPOZÍCII LEN V ANGLIČTINE.


(SK) • AK ZÁKAZNÍKOV POSKYTOVATEĽ SLUŽIEB VYŽADUJE INÝ JAZYK AKO ANGLIČTINU, POSKYTNUTIE PREKLADATEĽSKÝCH SLUŽIEB JE ZODPOVEDNOSŤOU ZÁKAZNÍKA.
• NEPOKÚŠAJTE SA O OBSLUHU ZARIADENIA SKÔR, AKO SI NEPREČÍTATE NÁVOD NA OBLUHU A NEPOROZUMIETE MU.
• ZANEDBANIE TOHTO UPOZORNENIA MÔŽE VYÚSTIŤ DO ZRANENIA POSKYTOVATEĽA SLUŽIEB, OBSLUHUJÚCEJ OSOBY ALEBO PACIENTA ELEKTRICKÝM PRÚDOM, DO MECHANICKÉHO
ALEBO INÉHO NEBEZPEČENSTVA.

Page 11
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

ATENCION • ESTE MANUAL DE SERVICIO SOLO EXISTE EN INGLES.


(ES) • SI ALGUN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLES, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCION.
• NO SE DEBERA DAR SERVICIO TECNICO AL EQUIPO, SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
• LA NO OBSERVANCIA DEL PRESENTE AVISO PUEDE DAR LUGAR A QUE EL PROVEEDOR DE SERVICIOS, EL OPERADOR O EL PACIENTE SUFRAN LESIONES PROVOCADAS POR CAUSAS
ELÉCTRICAS, MECÁNICAS O DE OTRA NATURALEZA.

VARNING • DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA.


(SV) • OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA ÖVERSÄTTNINGSTJÄNSTER.
• FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
• OM DU INTE TAR HÄNSYN TILL DEN HÄR VARNINGEN KAN DET RESULTERA I SKADOR PÅ SERVICETEKNIKERN, OPERATÖREN ELLER PATIENTEN TILL FÖLJD AV ELEKTRISKA STÖTAR,
MEKANISKA FAROR ELLER ANDRA FAROR.

DİKKAT • BU SERVİS KILAVUZUNUN SADECE İNGİLİZCESİ MEVCUTTUR.


(TR) • EĞER MÜŞTERİ TEKNİSYENİ BU KILAVUZU İNGİLİZCE DIŞINDA BİR BAŞKA LİSANDAN TALEP EDERSE, BUNU TERCÜME ETTİRMEK MÜŞTERİYE DÜŞER.
• SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE ETMEYİNİZ.
• BU UYARIYA UYULMAMASI, ELEKTRİK, MEKANİK VEYA DİĞER TEHLİKELERDEN DOLAYI TEKNİSYEN, OPERATÖR VEYA HASTANIN YARALANMASINA YOL AÇABİLİR.

Page 12
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

DAMAGE IN TRANSPORTATION
All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving
agent. Whether noted or concealed, damage MUST be reported to the carrier immediately upon discovery, or in any event, within 14 days after receipt, and the contents and containers held for inspection by the carrier. A transportation company
will not pay a claim for damage if an inspection is not requested within this 14 day period.
Call GEHC Global Parts 1-800-548-3366 and select option 8, immediately after damage is found. At this time be ready to supply name of carrier, delivery date, consignee name, freight or express bill number, item damaged and extent of damage.
Complete instructions regarding claim procedure are found in Section S of the Policy And Procedures Bulletins.
14 July 1993

CERTIFIED ELECTRICAL CONTRACTOR STATEMENT


All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors. In addition, electrical feeds into the Power Distribution Unit shall be performed
by licensed electrical contractors. Other connections between pieces of electrical equipment, calibrations and testing shall be performed by qualified GE Healthcare personnel. The products involved (and the accompanying electrical installations)
are highly sophisticated, and special engineering competence is required. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the
requirements of the applicable electrical codes.

The purchaser of GE equipment shall only utilize qualified personnel (i.e., GE’s field engineers, personnel of third-party service companies with equivalent training, or licensed electricians) to perform electrical servicing on the equipment.

IMPORTANT...X-RAY PROTECTION
X-ray equipment, if not properly used, may cause injury. Accordingly, the instructions herein contained should be thoroughly read and understood by everyone who will use the equipment before you attempt to place this equipment in operation.
The General Electric Company, Healthcare Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the useful beam, no practical design of equipment can provide complete protection. Nor can any practical design compel the operator to take adequate
precautions to prevent the possibility of any persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully acquainted with the recommendations of the National Council on Radiation Protection and Measurements as published in NCRP Reports available from
NCRP Publications, 7910 Woodmont Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on Radiation Protection, and of any other local authorities, and take adequate steps to protect against injury.
The equipment is sold with the understanding that the General Electric Company, Healthcare Group, its agents, and representatives have no responsibility for injury or damage which may result from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.

OMISSIONS & ERRORS


Customers, please contact your GE Sales or Service representatives.
GE personnel, please use the GEHC PQR Process to report all omissions, errors, and defects in this publication.

Page 13
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

This page is intentionally left blank.

Page 14
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Preface - Publication Conventions


Purpose: Standardized conventions for representing information is a uniform way of communicating information to a reader in a consistent manner. Conventions are used so that the reader can easily recognize the actions or decisions that must
be made. There are a number of character and paragraph styles used in this publication to accomplish this task. Please become familiar with them before proceeding forward.

Section 1.0
Safety & Hazard Information
Proper product safety labeling allows a person to safely use or service a product. The format and style for safety communications reflected in this publication represents the harmonization of IEC/ISO 3864 and ANSI Z535 standards.
Within this publication, different paragraph and character styles are used to indicated potential hazards. Paragraph prefixes, such as hazard, caution, danger and warning, are used to identify important safety information. Text (Hazard) styles are
applied to the paragraph contents that are applicable to each specific safety statement.

1.1 Hazard Messages


Any action that will, could or potentially cause personal injury will be preceded by the safety alert symbol and an appropriate signal word. The safety alert symbol is the triangle with an exclamation mark within it. It’s always used next to the signal
word to indicate the severity of the hazard. Together, they are used to indicate a hazard exists.
Signal words describe the severity of possible human injuries that may be encountered. The alert symbol and signal word are placed immediately before any paragraph they affect. Safety information includes:
1.) Signal Word - The seriousness level of the hazard.
2.) Symbol or Pictorial - The consequence of interaction with the hazard.
3.) Word Message:
a.) The nature of the hazard (i.e. the type of hazard)
b.) How to avoid the hazard.
The safety alert symbol is not used when an action can only cause equipment damage.

1.2 Text Format of Signal Words


DANGER - INDICATES AN IMMINENTLY HAZARDOUS SITUATION WHICH, IF NOT AVOIDED, WILL RESULT IN DEATH OR SERIOUS INJURY. THIS SIGNAL WORD IS TO BE LIMITED TO THE
MOST EXTREME SITUATIONS.
WARNING - INDICATES A POTENTIALLY HAZARDOUS SITUATION WHICH, IF NOT AVOIDED, COULD RESULT IN DEATH OR SERIOUS INJURY.
Caution - Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices.
Notice - Indicates information or a company policy that relates directly or indirectly to the safety of personnel or protection of property. This signal word is associated directly with a hazard or hazardous situation and is used in
place of 'DANGER,' 'WARNING,' or 'CAUTION.' It can include:
• Destruction of a disk drive
• Potential for internal mechanical damage, such as to a X-ray tube

Page 15
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1.3 Symbols and Pictorials Used


The following Symbols and Pictorials are be used in this publication. These graphical icons (symbols) may be used to make you aware of specific types of hazards that could possibly cause harm.

keep_up magnetic biohazard compressgas ppe-hearing

fragile impact corrosive heavyobject ppe-2people

static_elec heat general laser ppe-respitory

keep_dry pinch radiation poisongas ppe-loto

general explosive electrical flammable ppe-eye

torque crush/mechanical tipping Read Manual ppe-gloves

ce instuction poisonmatl entanglement instuction

Section 1.0 Safety & Hazard Information Page 16


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 2.0
Publication Conventions
2.1 General Paragraph and Character Styles
Prefixes are used to highlight important non-safety related information. Paragraph prefixes (such as Purpose, Example, Comment or Note) are used to identify important but non-safety related information. Text styles are also applied to text within
each paragraph modified by the specific prefix.

EXAMPLES OF PREFIXES USED FOR GENERAL INFORMATION:


Purpose: Introduces and provides meaning as to the information contained within the chapter, section or subsection (such as used at the beginning this chapter, for example).
Note: Conveys information that should be considered important to the reader.
Example: Used to make the reader aware that the paragraph(s) that follow are examples of information possibly stated previously.
Comment: Represents “additional” information that may or may not be relevant to your situation.

2.2 Page Layout


Publication Part Number & Revision Number Publication Title

The current section and its title The current chapter and its title
are always shown in the footer of are always shown in the footer of
the left (even) page. the right (odd) page.
An exclamation point in a triangle is used Paragraphs preceeded by a symbol
to indicate important information to the user. (e.g. bullets) contain information that
Paragraphs preceeded by Alphanumeric has no specific order.
characters (e.g. numbers) contain infor-
mation that must be followed in a specific order.

Headers and footers in this publication are designed to allow you to quickly identify your location. The document part number and revision number appears in every header on every page. Odd numbered page footers indicate the current chapter,
its title and current page number. Even page footers show the current section and its title, as well as the current page number.

2.3 Computer Screen Output/Input Text Character Styles


Within this publication, mono-spaced character styles (fonts) are used to indicate computer text that’s either screen input and output. Mono-spaced fonts, such courier, are used to indicated text direction. When you type at your keyboard, you are
generating computer input. Occasionally you will see the math operator “greater-than” and “less-than” symbols used to indicate the start and finish of variable output. When reading text generated by the computer, you are reading it as computer
generated output. In addition to direction, characters are italicized (e.g. italics) to indicate information specific to your system or site.

Page 17
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Example: This paragraph’s font represents computer generated screen “fixed” output. Its output is fixed from the sense that it does not vary from application to application. It’s the most
Fixed Output commonly used style used to indicate filenames, paths and text that do not change from system to system. The character style used is a fixed width such as courier.
Example: This paragraph’s font represents computer screen output that is “variable”. It’s used to represent output that varies from application to application or system to system. Variable
Variable Output output is sometimes found placed between greater-than and less-than operators for clarification. For example: <variable_ouput> or <3.45.120.3>. In both cases, the < and > operators
are not part of the actual input.
Example: This paragraph’s font represents fixed input. It’s computer input that is typed-in via the keyboard. Typed input that does not vary from application to application or system to
Fixed Input system. Fixed text the user is required to supply as input. For example: cd /usr/3p
Example: This paragraph’s font represents computer input that can vary from application to application or system to system. With variable text, the user is required to supply system dependent
Variable Input input or information. Variable input sometimes is placed between greater-than and less-than operators. For example: <variable_input>. In these cases, the (<>) operators would be
dropped prior to input. For example: ypcat hosts | grep <3.45.120.3> would be typed into the computer as
ypcat hosts | grep 3.45.120.3
without the greater-than and less-than operators.

2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys)
Different character styles are used to indicate actions requiring the reader to press either a hard or soft button, switch or key. Physical hardware, such as buttons and switches, are called hard keys because they are hard wired or mechanical in
nature. A keyboard or on/off switch would be a hard key. Software or computer generated buttons are called soft keys because they are software generated. Software driven menu buttons are an example of such keys. Soft and hard keys are
represented differently in this publication.
Example: A power switch ON/OFF or a keyboard key like ENTER is indicated by applying a character style that uses both over and under-lined bold text that is bold. This is a hard key.
Hard Keys
Example: Whereas the computer MENU button that you would click with your mouse or touch with your hand uses over and under-lined regular text. This is a soft key.
Soft Keys

Section 2.0 Publication Conventions Page 18


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Table of Contents
Preface - Publication Conventions........................................................................................................................................................................................................................................ 15
Section 1.0 —
Safety & Hazard Information......................................................................................................................................................................................................................................................................... 15
1.1 Hazard Messages........................................................................................................................................................................................................................................................................................................................................ 15
1.2 Text Format of Signal Words ....................................................................................................................................................................................................................................................................................................................... 15
1.3 Symbols and Pictorials Used ....................................................................................................................................................................................................................................................................................................................... 16
Section 2.0 —
Publication Conventions............................................................................................................................................................................................................................................................................... 17
2.1 General Paragraph and Character Styles.................................................................................................................................................................................................................................................................................................... 17
2.2 Page Layout................................................................................................................................................................................................................................................................................................................................................. 17
2.3 Computer Screen Output/Input Text Character Styles ................................................................................................................................................................................................................................................................................ 17
2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys) ....................................................................................................................................................................................................................................................................... 18

Page 19
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Chapter 1 - Introduction to Schematics .............................................................................................................................................................................................................................. 23


Section 1.0 —
Content and Organization ............................................................................................................................................................................................................................................................................ 23
Section 2.0 —
Drawing Conventions ................................................................................................................................................................................................................................................................................... 23

Chapter 2 - System................................................................................................................................................................................................................................................................ 25
Section 1.0 —
System Interconnect ..................................................................................................................................................................................................................................................................................... 25
Section 2.0 —
System Architecture (CANopen based) ...................................................................................................................................................................................................................................................... 26
Section 3.0 —
System Jumpers, Switch Settings............................................................................................................................................................................................................................................................... 27

Chapter 3 - Over-Head Tube Support (OTS) ....................................................................................................................................................................................................................... 29


Section 1.0 —
FeiTian OTS, 5139720SCH............................................................................................................................................................................................................................................................................ 29
Section 2.0 —
FeiTian Arm Logic Module, 6000000SCH.................................................................................................................................................................................................................................................... 31
Section 3.0 —
FeiTian I/F Board, 5151217SCH.................................................................................................................................................................................................................................................................... 44
Section 4.0 —
OTS Console interconnection...................................................................................................................................................................................................................................................................... 60
Section 5.0 —
Cable Routing ................................................................................................................................................................................................................................................................................................ 61

Chapter 4 - Wall Stand .......................................................................................................................................................................................................................................................... 67


Section 1.0 —
FeiTian Wall Stand, 5136848SCH................................................................................................................................................................................................................................................................. 67
Section 2.0 —
FT WS Motor Control, 1234567FAB ............................................................................................................................................................................................................................................................. 70
2.1 FT WS Motor Control Top Silk ..................................................................................................................................................................................................................................................................................................................... 70
2.2 FT WS Motor Control Bottom Silk................................................................................................................................................................................................................................................................................................................ 71

Page 20
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 3.0 —
FT WS Transfer Top Silk............................................................................................................................................................................................................................................................................... 72
Section 4.0 —
WS Motor Control Board SCH...................................................................................................................................................................................................................................................................... 73
Section 5.0 —
Remote Control Transfer .............................................................................................................................................................................................................................................................................. 80
Section 6.0 —
Grid Sensor Board, 5145484 ........................................................................................................................................................................................................................................................................ 86
Section 7.0 —
Display Board, 5140868 ................................................................................................................................................................................................................................................................................ 88
Section 8.0 —
Column Button, 5140871 .............................................................................................................................................................................................................................................................................. 91
Section 9.0 —
Wall Stand Schematics, 5136848 ................................................................................................................................................................................................................................................................. 95
Section 10.0 —
Ion Chamber, 2210104SCH......................................................................................................................................................................................................................................................................... 101
Section 11.0 —
Cable Routing .............................................................................................................................................................................................................................................................................................. 102

Chapter 5 - System Cabinet ............................................................................................................................................................................................................................................... 107


Section 1.0 — 5KV Power Distribution Unit, 2293487.............................................................................................................................................................................................................................. 108
Section 2.0 — "Halo" Relay Board, 2407245............................................................................................................................................................................................................................................. 110
Section 3.0 — Corona 2 Board, 5138201................................................................................................................................................................................................................................................... 116
Section 4.0 — Jedi X-ray Generator, 2374870 .......................................................................................................................................................................................................................................... 157
4.1 Overview .................................................................................................................................................................................................................................................................................................................................................... 157
4.2 Jedi Power Module, 2375932 .................................................................................................................................................................................................................................................................................................................... 158
4.2.1 Generic IF Board, 2295467-3 ........................................................................................................................................................................................................................................................................................................... 158
4.2.2 AEC Board, 2222253-5 .................................................................................................................................................................................................................................................................................................................... 167
4.2.3 KV Control Board, 2382791-3 .......................................................................................................................................................................................................................................................................................................... 173
4.2.4 Inverter Power Assembly, 2268972.................................................................................................................................................................................................................................................................................................. 196
4.2.4.1 Quad Snubber Board, 2307323 ....................................................................................................................................................................................................................................................................................... 198
4.2.4.2 Gate Cmd Board, 2209547-3 ........................................................................................................................................................................................................................................................................................... 202
4.3 HV Tank Assembly, 2186730 .................................................................................................................................................................................................................................................................................................................... 209
4.3.1 KV Measure Board, 2184565 ........................................................................................................................................................................................................................................................................................................... 211
4.4 EMC and AC/DC Module, 2372292 ........................................................................................................................................................................................................................................................................................................... 213
4.4.1 EMC Filter Board, 2209840-2........................................................................................................................................................................................................................................................................................................... 214
4.4.2 AC/DC Board, 2300401-2 (schematic not available)........................................................................................................................................................................................................................................................................ 217
4.5 Auxiliary Module, 2381787......................................................................................................................................................................................................................................................................................................................... 218
4.5.1 Heater Board, 2271116 .................................................................................................................................................................................................................................................................................................................... 218
4.5.2 LVPS Power Supply, 2212166-2 ...................................................................................................................................................................................................................................................................................................... 226
4.5.3 Rotation Board, 2242438-3 .............................................................................................................................................................................................................................................................................................................. 233

Page 21
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 5.0 — System Cabinet Bulkhead, 5134189.................................................................................................................................................................................................................................. 246

Chapter 6 - Master Interconnect Schematic (MIS) and Charts........................................................................................................................................................................................ 247


Section 1.0 —
Grounding & Wire Gauge Selection in Master Interconnect Schematics .............................................................................................................................................................................................. 247
1.1 Before You Begin....................................................................................................................................................................................................................................................................................................................................... 247
1.2 Equal-potential Grounding “General Rules” ............................................................................................................................................................................................................................................................................................... 247
1.3 equal potential Grounding - “Recommendations” ...................................................................................................................................................................................................................................................................................... 247
1.4 Vendor “Add-on Components” - Ground Requirements ............................................................................................................................................................................................................................................................................ 247
1.5 Grounding using the MIS Map ................................................................................................................................................................................................................................................................................................................... 247
1.5.1 Purpose ............................................................................................................................................................................................................................................................................................................................................ 247
1.5.2 MIS Map Symbols ............................................................................................................................................................................................................................................................................................................................ 247
1.5.3 Power Distribution ............................................................................................................................................................................................................................................................................................................................ 249
1.5.4 Wire Gauge Guidelines For Field Supplied Power Runs.................................................................................................................................................................................................................................................................. 249
1.5.4.1 NEC Minimum .................................................................................................................................................................................................................................................................................................................. 249
1.5.4.2 Voltage Regulation........................................................................................................................................................................................................................................................................................................... 249
Section 2.0 —
The Master Interconnect Schematic (MIS) Map........................................................................................................................................................................................................................................ 253
2.1 Introduction ................................................................................................................................................................................................................................................................................................................................................ 253
2.1.1 Using the MIS Map ........................................................................................................................................................................................................................................................................................................................... 253
2.1.2 Symbols/Conventions....................................................................................................................................................................................................................................................................................................................... 253
2.2 System MIS Maps (Full Sized)................................................................................................................................................................................................................................................................................................................... 254
Section 3.0 —
MIS CHARTS ................................................................................................................................................................................................................................................................................................ 255
3.1 mis chart 11629a ....................................................................................................................................................................................................................................................................................................................................... 255
3.2 MIS Chart 11644A ..................................................................................................................................................................................................................................................................................................................................... 255
3.3 mis chart 11690a ....................................................................................................................................................................................................................................................................................................................................... 255
3.4 mis chart 11691a ....................................................................................................................................................................................................................................................................................................................................... 255
3.5 MIS Chart 11708A ..................................................................................................................................................................................................................................................................................................................................... 255
3.6 MIS Chart 11710A ..................................................................................................................................................................................................................................................................................................................................... 255
3.7 MIS Chart 11711A ..................................................................................................................................................................................................................................................................................................................................... 255
3.8 MIS Chart 11755A ..................................................................................................................................................................................................................................................................................................................................... 255
3.9 MIS Chart 11756A ..................................................................................................................................................................................................................................................................................................................................... 255
3.10 MIS Chart 11757A ..................................................................................................................................................................................................................................................................................................................................... 256
3.11 MIS Chart 11759A ..................................................................................................................................................................................................................................................................................................................................... 256
3.12 mis chart 11760a ....................................................................................................................................................................................................................................................................................................................................... 256
3.13 mis chart 11761a ....................................................................................................................................................................................................................................................................................................................................... 256
3.14 mis chart 11762a ....................................................................................................................................................................................................................................................................................................................................... 257
3.15 mis chart 11763a ....................................................................................................................................................................................................................................................................................................................................... 257
3.16 mis chart 11764a ....................................................................................................................................................................................................................................................................................................................................... 257
3.17 mis chart 11767a ....................................................................................................................................................................................................................................................................................................................................... 258
3.18 mis chart 11768a ....................................................................................................................................................................................................................................................................................................................................... 258
3.19 mis chart 11769a ....................................................................................................................................................................................................................................................................................................................................... 259
3.20 mis chart 11770a ....................................................................................................................................................................................................................................................................................................................................... 259
3.21 mis chart 11774a ....................................................................................................................................................................................................................................................................................................................................... 259
3.22 mis chart 11776a ....................................................................................................................................................................................................................................................................................................................................... 259
3.23 MIS Chart 20101A ..................................................................................................................................................................................................................................................................................................................................... 260
3.24 MIS Chart 20102A ..................................................................................................................................................................................................................................................................................................................................... 260

Page 22
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Chapter 1 - Introduction to Schematics


Section 1.0
Content and Organization
This service manual contains schematics for the assemblies that make up the Definium 6000 system. Schematics are arranged in alpha-numeric order, using the reference designators assigned each assembly. Reference designators are brief
codes used to identify and locate the assembly within a system.
Schematics are preceded by a Title Page that contains the schematic number, a reference designator, and the schematic’s name. Signal Name Lists are provided for interfaces such as: backplanes, wire harnesses, or transition panels, that do
not have many active components. These interfaces electrically link the different assemblies together and must be included in the schematics document for proper signal tracing.

Section 2.0
Drawing Conventions
Some schematics contained in this document use a hierarchical design philosophy. This hierarchical philosophy allows for easier schematic representation of complex circuitry. To a typical schematic user however, the hierarchical schematics
may seem difficult to follow. So to address this issue, there are listed below, some informative highlights which should help in the understanding of hierarchical design schematics.
• Hierarchical schematics are structured from a high level (i.e.,complete system) to a low level (i.e., discrete components). In circuit board schematic applications this means that there will typically be a block diagram on the first schematic
sheets, which represents the entire circuit board. Then from this diagram, each of the blocks is broken down into the discrete component circuitry.
• For individual signal connections, a similar schematic structuring technique is used. At the block diagram schematic level, signals are grouped together into buses, which travel between the blocks. The individual signal connections are first
shown separately when the blocks are broken down into the discrete component circuitry.
• Buses are drawn thicker than the individual signal lines and they also have two numbers shown in parenthesis, immediately following the bus name. These numbers indicate how many individual signals are contained in the bus. For example;
(7:0) indicates that there are eight total signals in the bus, ranging from 0 through 7. If the buses happen to be data or address buses, the numbers in parenthesis also indicate the most significant bit and least significant bit as follows,
(MSB:LSB).
• When an individual signal line enters or leaves a bus, it is drawn using a diagonal line called a ripper. Next to the ripper, a number is also drawn. This ripper number is unique to the signal line that it represents. In other words, if a ripper
number is 3 and the signal line name that leaves the bus is XAC, every time XAC leaves or enters that bus, the ripper number, will always be 3. Examples of ripper symbols are shown in Figure 1-1.
• There are some special symbols, called ports and pages, used to link buses and signal lines from one sheet of the schematic to others. Port symbols link signals between different functional blocks, while the page symbols link signals within
the same functional block. Examples of these symbols are shown in Figure 1-1.
This was just a brief summary of some of the more significant differences associated with hierarchical design schematics. In general, most other schematic guidelines used before the hierarchical applications, still apply. For example; schematic
inputs are still on the left and outputs on the right, and schematic flow is from left-to-right and top-to-bottom. Also, signal names are still consistent throughout an entire schematic design, whenever possible.

Rippe r Example

BUS
RIPPER
ADDR( 3 : 0 ) SIGNAL

0 ADDR(0)
1 ADDR(1)
2 ADDR(2)
3 ADDR(3)

Off Sheet Examp les

Port Out Port In Port In/Out On Page Off Page Page In/Out

Figure 1-1 Example

Chapter 1 - Introduction to Schematics Page 23


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

This page is intentionally left blank.

Section 2.0 Drawing Conventions Page 24


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Chapter 2 - System
Section 1.0
System Interconnect

FeiTian System Interconnect Cable Diagram Wallstand


A A
Magic PC

RJ45 PLUG
RJ45 REC
WallBox
Ethernet Interface
PC Embedded

PCI GbE card


Hospital Passthrough Blukhead inside wallstand
M RJ45

F RJ45

Ethernet
M RJ45
F RJ45

M SUB-D 9P

M SUB-D 9P
Detector Ethernet

F SUB-D 9P

F SUB-D 9P
Network System CANopen

B B
M SUB-D 9P

M SUB-D 9P

M SUB-D 9P
F SUB-D 9P

F SUB-D 9P

F SUB-D 9P
CAN1

Jedi CAN Jedi CAN


M USB A

F USB A

Mouse USB

M MATE-N-LOCK 3P

M MATE-N-LOCK 3P
F MATE-N-LOCK 3P

F MATE-N-LOCK 3P
PCI CAN card

120VAC (for WS)


Keyboard
M SUB-D 9P

M SUB-D 9P

M SUB-D 9P
F SUB-D 9P

F SUB-D 9P

F SUB-D 9P
System CANopen System CANopen
CAN2
M USB A

F USB A

USB
Image Monitor

M MATE-N-LOCK 3P

M MATE-N-LOCK 3P
F MATE-N-LOCK 3P

F MATE-N-LOCK 3P
C 120VAC (for Detector PS)
C
PCI RS232X2 card

BARCODE
M USB A

M MATE-N-LOCK 3P
F USB A

F MATE-N-LOCK 3P
USB
READER

M IEC320

F IEC320
120VAC 120VAC
RS232 Interface

M SUB-D 9P

M SUB-D 9P
F SUB-D 9P

F SUB-D 9P
PC Embedded

Ion Chamber
M SUB-D 9P
F SUB-D 9P

M DVI-A
F DVI-A

DVI-I

M3 STUD

M3 STUD
D D

M3 RING

M3 RING
GROUNDING
M VGA 15P

F VGA 15P

M IEC320

M IEC320
F IEC320

F IEC320

VGA 120VAC

System
OTS
M3 STUD

M3 STUD
M3 RING

M3 RING
GROUNDING
Cabinet

M SUB-HD 15P

F SUB-HD 15P
UIF

M3 STUD
M3 RING
GROUNDING
E Handswitch
E

M SUB-D 9P

M SUB-D 9P
F SUB-D 9P

F SUB-D 9P
RS232
RCIM
M SUB-D 9P

M SUB-D 50P

M SUB-D 50P
F SUB-D 9P

F SUB-D 50P
F SUB-D 50P
M SCSI 50P

USER I/O USER I/O


F SCSI 50P

M SUB-D 9P

M SUB-D 9P
F SUB-D 9P

F SUB-D 9P
RCIM Power RCIM Power System CANopen

F Passthrough
F

M MATE-N-LOCK 3P

M MATE-N-LOCK 3P
F MATE-N-LOCK 3P

F MATE-N-LOCK 3P
120VAC
OTS
Legend:
Hospital 320-
M SUB-D 9P

M SUB-D 9P
F SUB-D 9P

F SUB-D 9P
= Ethernet
GROUNDING
= CANopen bus
480VAC 3 phase 3 phase power cable To Cabinet
Circuit Breaker

= Jedi CAN bus


supply
G G
M MATE-N-

F MATE-N-
LOCK 12P

LOCK 12P

= RS232/RS422 serial communication GROUNDING


TUBE&FAN
= USB serial communication Connect to tube bulkhead
XRAY ON LIGHT
= VGA/DVI-I video HV+
To Jedi Tank Connect to tube directly
= AC Power line HV-
DOOR To PDU Terminals
= DC Power line
INTERLOCK
= Grounding line
H = Others
TITLE

FeiTian System Architecture


REV H
FIRST MADE FOR: FeiTian
3
GEHL MEDICAL SYSTEMS 51364101 BLK SHT/CONT ON
1/2
PDF created with pdfFactory
1 Pro trial version 2
www.pdffactory.com 3 4 5 6 7 8
Figure 2-1 System Intercoonnect

Chapter 2 - System Page 25


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 2.0
System Architecture (CANopen based)

A FeiTian System CANopen, Serial line & Real-Time Topology A

KEYBOARD

MOUSE
BARCODE
READER
System Cabinet (Reuse) DVI-I

Jedi Interface USB USB


USB

TERM Hospital
B Network
Ethernet
Thunder Magic Image Monitor B
TERM
(reuse) Gigabit Ethernet
Jedi CAN TERM

System CANopen Wallstand IO Detector Wallstand


Jedi RT board Detector Power Supply
ARM-CORE
C C
With Terms

NODE ID
Vertical
CAN HUB

WS CANopen 0x34
AqBiSS/SSI
(reserve 0x35) Encoder
Fixed Table CANopen
(Jumper Connectors) OTS CANopen Tilt
Handswitch AqBiSS/SSI
Encoder
Detector PS
TERM NODE ID
NODE ID=0x30
Reserved 31 RCIM 0x50
D D
USER I/O (reuse)
Corona 2
ION
Chamber
Ion Chamber

E PDU control PDU status


E

XRAY-ON Light

OTS IO board OTS


AqBiSS/SSI
Longitude
Encoder
ARM-CORE
NODE ID
Collimator
Room Door 0x38 AqBiSS/SSI
Lateral Reserved ID
F PDU (reserve 0x39) Encoder 0x26,27,28,29,41,61 F
SW Vertical
Vertical Motor AqBiSS/SSI
AMP. Encoder TERM
NODE ID Angulation
0x05 AqBiSS/SSI
Encoder

Buttons&LEDs signal
G (RT or RS232)
G
VGA OTS Console
RS232 for touchscreen
RS232

Rotor, Tube Fan, Pressure Switch TUBE

H TITLE

FeiTian System Architecture


REV H
FIRST MADE FOR: FeiTian
3
GEHL MEDICAL SYSTEMS 51364101 BLK SHT/CONT ON
2/2
DF created with pdfFactory
1 Pro trial version 2
www.pdffactory.com 3 4 5 6 7 8
Figure 2-2 Sys Architecture (CANopen based)

Section 2.0 System Architecture (CANopen based) Page 26


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 3.0
System Jumpers, Switch Settings

System Cabinet Wall Stand OTS


Switch Settings Switch & Jumper Settings Switch & Jumper Settings
Corona Bd. WS Detector Power Supply ALB Bd. (5161473)
DIP switch settings DIP Switch Settings (Top of P.S.) J5 OPEN J6 OPEN
S1 S2 S3 J17 - 2 CLOSE J17 - 3 CLOSE
S8 S7 S6 S5 S4 S3 S2 S1
1 - On 1 - Off 1 - On J18 - 2 CLOSE J18 - 3 CLOSE
ON OFF ON OFF ON ON ON ON
2 - On 2 - Off 2 - Off J27 CLOSE J28 CLOSE
3 - Off 3 - Off 3 - On DIP Switch Settings (Front of P.S.)
4 - Off 4 - On 4 - Off
S3 S2 S1
OFF OFF ON

ALB Bd (5161473)
J5 OPEN J6 OPEN
J17 OPEN J18 OPEN
J27 OPEN J28 OPEN

Figure 2-3 System Jumpers & Switch Settings

Chapter 2 - System Page 27


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

This page is intentionally left blank.

Section 3.0 System Jumpers, Switch Settings Page 28


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Chapter 3 - Over-Head Tube Support (OTS)


Section 1.0
FeiTian OTS, 5139720SCH
1 2 3 4 5 6 7 8

Area Designations GE Company Confidential


A A1 - OTS ALB Assembly
A1 Blk 1 OTS ALB Assembly A1 A1 Blk2
A
A2 - OTS Power Assembly 5148479
A3 - OTS Top Assembly
A4 - OTS Lateral Rail Assembly
1
A5 - OTS Angulation Assembly 5V
9
A6 - OTS Console 2
GND Lateral Load
CLK- Encoder
R4 10
CLK+ A3 E3
3 DATA- 5161024
11 DATA+
B B
1 1
5V 9 5V
Longitudinal GND
9 GND
2 2
Load Encoder A4 E1 CLK- L1 10
CLK-
10 CLK+
5161024 CLK+
3 Amplifier 3 Vertical Load
Data- DATA-
11 11 DATA+ Encoder
Data+ A1B2 5161083 R5 5 A3 E2
J22 J22 12 ENA+ 5161024
ENA-
6
ENB+
C 5V
1 13
ENB- C
Angulation 9 7 ENC+
GND 14
2 ENC-
Load Encoder CLK-
10 L2
5150464 CLK+
A5 E4 3 1
Data- 5V
Data+ 11 6
7 GND
ENA+
2
ENA-
R2 8
ENB+ EN Vertical
3
RXD ENB- A3 Driving Motor
9
D To A6 J1
TXD
12VDC L4 J10 4
ENC+
ENC-
M1 5150166 D
GND
1
Motor V+

J19
R1 2 MTR
Motor V-
To System Cabinet

CAN_H 7
CAN_L 2
9 Advanced Logic Board
CAN_V+
3 L5 J5 A1B1 5161473
CAN_GND 8 7
CAN_H_lbpk 24V Vertical Lock
E CAN_L_Lbpk
4 2
Long Lock
A3L2 5150462 E
8 Lateral Lock
24V
3 A3L3 5140106
Long Lock

J8
R3
6 24V Vertical Clutch
Angulation Lock 8 A3C1
A5L4 24V 1 5150460
5150465 Ang Lock
3 Long Lock

Column Rotation 7 L3 9 Longitudinal Lock


24V 24V
F A5S1 2 4 A4 L1 F
J9

Switch 5117344-44 Zxis SW Long Lock


5140106

Collimator Rotation 3
24V
A5S2 Coll SW
11
Switch 5117344-45
1
1

From Power Box


24V
2 GND
GND 2

A2blk2 Out
5 24VDC
CAN_H
4
L6 3
CAN_L J J1 R6 0V
Collimatior 7 4 37V
G CAN_V+
CAN_GND
10
4
5 Shield
G
6,8
37VAC
0VAC 9,12

J6

NOTES
1. All sheets of this schematic will be at the same revision. The
H revision of this sheet will be the revision of the entire document.
REVISIONS
MADE
APPROVALS YY/MM/DD TITLE REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian OTS Schematics
2. Indicates signals are part of a cable
1 Gao Shuo 26 Oct 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 2
APPR
GEHL MEDICAL 5139720SCH SHT/CONT ON
- - OTHER SYSTEMS 1 /2

1 2 3 4 5 6 7 8

Figure 3-1 5139720SCH_s1_r1_1


Chapter 3 - Over-Head Tube Support (OTS) Page 29
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

GE Company Confidential
A A

GND

A2 Blk2 A2 Blk1
0VDC 6

)* GND

B 24VDC 6
Switch Power B
Supply (SP)
$&/ $&1
5
1
2 GND *1'
Out

AC-L
C 4
C

From System Cabinet


37VAC Transformer
-

-
3 (T1) AC-N
0VAC 3 GND
- 2

In
Filter - 0VAC
(FT1) 1 110VAC
Fuse (F1)
AC-L
250V 10A
AC-L
D D

Power Box Assembly A2


Fuse (F2)
5148483
250V 10A

E E
RXD0

To System Cabinet
TXD0 HV Anode
GND From A1Blk1 L6
12VDC HV Cathode
J1 A5 Xray Tube Anode Stator
Fan/Pressure Switch
RXD1 GND
F TXD1 F
To System Console

OTS Console GND


5146497 A2
Red
Red GND
Green
Green GND
Area Designations
J2 Blue
Blue GND A1 - OTS ALB Assembly
HSYNC A2 - OTS Power Assembly NOTES
G VSYNC A3 - OTS Top Assembly 1. All sheets of this schematic will be at the same revision. The
G
SGND
A4 - OTS Lateral Rail Assembly revision of this sheet will be the revision of the entire document.
A5 - OTS Angulation Assembly
2. Indicates signals are part of a cable
A6 - OTS Console

H REVISIONS
MADE
APPROVALS YY/MM/DD TITLE REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian OTS Schematics
1 Gao Shuo 26 Oct 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 2
APPR
GEHL MEDICAL 5139720SCH SHT/CONT ON
- - OTHER SYSTEMS 2 /2

1 2 3 4 5 6 7 8

Figure 3-2 5139720SCH_s1_r1_2

Section 1.0 FeiTian OTS, 5139720SCH Page 30


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 2.0
FeiTian Arm Logic Module, 6000000SCH
1 2 3 4 5 6 7 8

FEITIAN ARM LOGIC MODULE


A

LPC2294 I2 MEMORY I3 PAGE DESCRIPTION OF CHANGE BY


1 ORIGINAL DESIGN WANG ZHUYING
ARM_JTAG<5..0> DATA<31..0> DATA<31..0>
2 DESIGN UPDATE WANG ZHUYING
CONFIG_FPGA<4..0> ADDR<23..0> ADDR<23..0>

ARM_CLK ARM_CON<9..0> ARM_CON<9..0>

ARM_RESET READY_BUSY READY_BUSY

EXT_RESET ARM_IO<30..0> ARM_RESET NOTES:


1. THE VARIATION AND VERSION FOR THIS
PWA IS 6000000 A.
2. THIS COVER SHEET CONTROLS THE RE-
VISION STATUS OF THE ENTIRE DOCUMENT.
THE REVISION OF ALL SHEETS OF THIS
DOCUMENT WILL BE AT THE SAME LEVEL.
3. THIS SCHEMATIC MEETS THE SPECIFICA-
B TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
4. ELECTRONIC SIGNATURES FOR BOTH THE
FPGA I4 SCHEMATIC REVIEWER(S) AND APPROVER(S)
CAN BE FOUND WITHIN E-PDM.
INTERFACE I6
ARM_RESET FPGA_CLK 5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS.
ARM_CLK ARM_CON<9..0> ARM_RESET EXT_RESET
RESISTORS ARE IN OHMS.
CONFIG_FPGA<4..0> ADDR<23..0> ARM_CLK FPGA_CLK INDUCTORS ARE IN HENRIES
FPGA_JTAG<4..0> DATA<31..0> ARM_IO<30..0>

ADC_CON<7..1> FPGA_IO<131..0> FPGA_IO<131..0>

ARM_IO<30..0> FPGA_CAN<2..1> FPGA_CAN<2..1> ANALOG_INPUT<8..1>

POWER I5

BSCAN I9 ADC I8

FPGA_JTAG<4..0> ARM_RESET ADC_CON<7..1> ANALOG_INPUT<8..1>

ARM_JTAG<5..0>

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Wed Nov 30 13:59:51 2005 armlogic 1 of 1 6000000SCH 1 1 of 13


1 2 3 4 5 6 7 8

Figure 3-3 6000000SCH_s1_r2_1

Chapter 3 - Over-Head Tube Support (OTS) Page 31


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

AS CONFIG CABLE
FPGA 3V3
J5

A 2 2 1 1 DCLK 2
R32 4 4 3 3 CONF_DONE 4
N_CE 6 6 5 5 N_CONFIG 0
N_CSO 8 8 7 7 CFG_DATA 1
10K 10 9 ASDIO
10 9
0.1W
1%
PIN U14
STRAIGHT EPCS4
DUAL_ROW_HEADER_BREAK
GND GND
1 CFG_DATA 2 DATA CS_N 1 N_CSO

DCLK 6 DCLK 2

ASDI 5 ASDIO
1V5
FL1
BLM31A601S VCC=3V3;GND=GND
1 2 VCCA_PLL1
600 OHM
C26 C30 C33 C94
1U 2.2U 0.1U 10N
10V 10V 16V 50V
10% 10% 10% 10% FPGA_JTAG<4..0>
IO
B
3V3 1V5

GND U13
1V5
EP1C12F324C8
FL2
BLM31A601S
1 2 VCCA_PLL2 VCCINT<11-0> CEO_N K2
VCCIO1<3-0> STATUS_N L12 N_STATUS 3
600 OHM
VCCIO2<3-0>
C27 C31 C93 C95 VCCIO3<3-0> TDO K13 FPGA_TDO 0
1U 2.2U 0.1U 10N VCCIO4<3-0> IO_D3_CLKUSR D3
10V 10V 16V 50V VCCA_PLL1 J5 VCCA_PLL1 IO_C3_INIT_DONE C3
10% 10% 10% 10% VCCA_PLL2 J12 VCCA_PLL2 CONF_DONE K17 CONF_DONE 4
IO_C4_DEV_CLRN C4 ARM_RESET
IN
N_CE J7 CE_N IO_B3_DEV_OE B3
0 N_CONFIG J2 CONFIG_N IO_J1_NCSO J1 N_CSO
3V3 K6 ASDIO
IO_K6_ASDO
SOLDER FOR 1 FPGA_TCK K18 TCK
C 2 FPGA_TMS K14 TMS IO_K4_PLL1_OUT0P K4 3V3
PS CONFIG R95 3 FPGA_TDI J17 TDI IO_K5_PLL1_OUT0N K5
GND 10K IO_K15_PLL2_OUT1P K15 R89
Y2 MSEL0:1 = 1:0 0.1W 1 CFG_DATA H7 DATA0 IO_K16_PLL2_OUT1N K16 0 N_CONFIG
3V3 1% 2 DCLK L1
S1803B-16.0000 DCLK
FL4 K3 K1 10K
MSEL0 GNDA_PLL1
BLM31A601S K7 J6 0.1W 1%
R87 MSEL1 GNDG_PLL1
1 2 4 VCC VOUT 3 FPGA_CLK J3 CLK0 GNDG_PLL2 J18
ARM_CLK J4 CLK1 GNDA_PLL2 K12 R90
600 OHM 1 2 33.2 J15 3 N_STATUS
CLK2 GND<27-0>
ENABLE GND 0.1W J16 CLK3
C28 C29 C32 1% 10K
R31 R96
1K 0 SYM 1 OF 3 0.1W 1%
1U 0.1U 10N 16MHZ
10V 16V 50V 0.1W SOLDER FOR 0.1W ACTIVE SERIAL
10% 10% 10% 1% 5% CONFIG MODE R91
AS CONFIG MESL0 = 0 4 CONF_DONE
MESL1 = 0
10K
MSEL0:1 = 0:0
GND 0.1W 1%

GND GND GND

GND
FPGA_CLK CONFIG_FPGA<4..0>
OUT IO

D IN ARM_CLK

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:26 2005 fpga 1 of 3 6000000SCH 1 2 of 13


1 2 3 4 5 6 7 8

Figure 3-4 6000000SCH_s1_r2_2

Section 2.0 FeiTian Arm Logic Module, 6000000SCH Page 32


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

FPGA IO
FPGA_IO<131..0>
IO

ARM_IO<30..0>
IO

A DATA<31..0>
IO

3V3 3V3 3V3 3V3

U13 U13
EP1C12F324C8 EP1C12F324C8

D4 IO_D4_VREF0B1 IO_C14_VREF0B2 C14 E14 IO_E14_VREF0B3 IO_P13_VREF0B4 P13


H6 IO_H6_VREF1B1 IO_E10_VREF1B2 E10 J14 IO_J14_VREF1B3 IO_P9_VREF1B4 P9
R1 IO_R1_VREF2B1 IO_A6_VREF2B2 A6 P16 IO_P16_VREF2B3 IO_R5_VREF2B4 R5

0 FPGA_IO<0> C2 IO_C2 IO_C16 C16 FPGA_IO<44> 44 2 P02_SCL T16 IO_T16 IO_U3 U3 ADDR<17> 17
1 FPGA_IO<1> D2 IO_D2 IO_B16 B16 FPGA_IO<45> 45 0 DATA<0> T17 IO_T17 IO_V4 V4 FPGA_IO<102> 102
2 FPGA_IO<2> D1 IO_D1 IO_G11 G11 FPGA_IO<46> 46 3 P03_SDA R17 IO_R17 IO_M8 M8 ADDR<16> 16
3 FPGA_IO<3> E3 IO_E3 IO_F11 F11 FPGA_IO<47> 47 1 DATA<1> R18 IO_R18 IO_N8 N8 ADDR<15> 15
4 FPGA_IO<4> E2 IO_E2 IO_B15 B15 FPGA_IO<48> 48 2 DATA<2> R15 IO_R15 IO_T4 T4 ADDR<14> 14
5 FPGA_IO<5> F1 IO_F1 IO_A15 A15 FPGA_IO<49> 49 3 DATA<3> R16 IO_R16 IO_U4 U4 FPGA_IO<103> 103
6 FPGA_IO<6> E4 IO_E4 IO_C15 C15 FPGA_IO<50> 50 4 P04_SCK0 P17 IO_P17 IO_T5 T5 FPGA_IO<104> 104
7 FPGA_IO<7> E5 IO_E5 IO_D14 D14 FPGA_IO<51> 51 5 P05_MISO0 P15 IO_P15 IO_U5 U5 FPGA_IO<105> 105
8 FPGA_IO<8> F2 IO_F2 IO_B14 B14 FPGA_IO<52> 52 6 P06_MOSI0 P14 IO_P14 IO_R4 R4 FPGA_IO<106> 106
9 FPGA_IO<9> F3 IO_F3 IO_E13 E13 FPGA_IO<53> 53 7 P07_SSEL0 N14 IO_N14 IO_V6 V6 FPGA_IO<107> 107
10 FPGA_IO<10> F4 IO_F4 IO_G10 G10 FPGA_IO<54> 54 4 DATA<4> N18 IO_N18 IO_U6 U6 FPGA_IO<108> 108
11 FPGA_IO<11> F5 IO_F5 IO_F10 F10 FPGA_IO<55> 55 5 DATA<5> N17 IO_N17 IO_P6 P6 ADDR<13> 13
12 FPGA_IO<12> G1 IO_G1 IO_B13 B13 FPGA_IO<56> 56 8 P08_TXD1 N13 IO_N13 IO_P7 P7 ADDR<12> 12
B 13 FPGA_IO<13> G2 IO_G2 IO_A13 A13 FPGA_IO<57> 57 9 P09_RXD1 N12 IO_N12 IO_T6 T6 ADDR<11> 11
14 FPGA_IO<14> F6 IO_F6 IO_D13 D13 FPGA_IO<58> 58 6 DATA<6> N16 IO_N16 IO_R6 R6 FPGA_IO<109> 109
15 FPGA_IO<15> F7 IO_F7 IO_C13 C13 FPGA_IO<59> 59 7 DATA<7> N15 IO_N15 IO_U7 U7 FPGA_IO<110> 110
1 SWITCH<1> G3 IO_G3 IO_D12 D12 FPGA_IO<60> 60 10 P010_RTS1 M18 IO_M18 IO_V7 V7 ADDR<10> 10
16 FPGA_IO<16> G4 IO_G4 IO_C12 C12 FPGA_IO<61> 61 11 P011_CTS1 M17 IO_M17 IO_T7 T7 ADDR<9> 9
17 FPGA_IO<17> G5 IO_G5 IO_B12 B12 FPGA_IO<62> 62 12 P012_DSR1 M14 IO_M14 IO_R7 R7 FPGA_IO<111> 111
2 SWITCH<2> G6 IO_G6 IO_A12 A12 FPGA_IO<63> 63 8 DATA<8> M15 IO_M15 IO_U8 U8 FPGA_IO<112> 112
18 FPGA_IO<18> H1 IO_H1 IO_C11 C11 FPGA_IO<64> 64 9 DATA<9> M16 IO_M16 IO_V8 V8 FPGA_IO<113> 113
19 FPGA_IO<19> H2 IO_H2 IO_D11 D11 FPGA_IO<65> 65 13 P013_DTR1 L18 IO_L18 IO_T8 T8 FPGA_IO<114> 114
20 FPGA_IO<20> H3 IO_H3 IO_B11 B11 FPGA_IO<66> 66 1 FPGA_CAN_TX L17 IO_L17 IO_R8 R8 FPGA_IO<115> 115
21 FPGA_IO<21> H4 IO_H4 IO_A11 A11 FPGA_IO<67> 67 10 DATA<10> M13 IO_M13 IO_U9 U9 ADDR<8> 8
3 SWITCH<3> H5 IO_H5 IO_E11 E11 FPGA_IO<68> 68 11 DATA<11> L13 IO_L13 IO_V9 V9 ADDR<7> 7
22 FPGA_IO<22> L7 IO_L7 IO_C10 C10 FPGA_IO<69> 69 2 FPGA_CAN_RX L16 IO_L16 IO_R9 R9 FPGA_IO<116> 116
23 FPGA_IO<23> L6 IO_L6 IO_D10 D10 FPGA_IO<70> 70 12 DATA<12> L15 IO_L15 IO_T9 T9 FPGA_IO<117> 117
24 FPGA_IO<24> L2 IO_L2 IO_B10 B10 FPGA_IO<71> 71 13 DATA<13> L14 IO_L14 IO_M9 M9 ADDR<6> 6
4 SWITCH<4> L3 IO_L3 IO_A10 A10 FPGA_IO<72> 72 14 DATA<14> J13 IO_J13 IO_N9 N9 ADDR<5> 5
25 FPGA_IO<25> L5 IO_L5 IO_G9 G9 FPGA_IO<73> 73 14 P014_DCD1 H13 IO_H13 IO_U10 U10 FPGA_IO<118> 118
26 FPGA_IO<26> L4 IO_L4 IO_F9 F9 FPGA_IO<74> 74 15 P015_EINT2 H14 IO_H14 IO_V10 V10 ADDR<4> 4
5 SWITCH<5> M1 IO_M1 IO_D9 D9 FPGA_IO<75> 75 15 DATA<15> H15 IO_H15 IO_T10 T10 FPGA_IO<119> 119
6 SWITCH<6> M3 IO_M3 IO_C9 C9 FPGA_IO<76> 76 26 CAN1_TX H16 IO_H16 IO_R10 R10 FPGA_IO<120> 120
27 FPGA_IO<27> M2 IO_M2 IO_A9 A9 FPGA_IO<77> 77 25 CAN1_RX H17 IO_H17 IO_P10 P10 FPGA_IO<121> 121
28 FPGA_IO<28> M5 IO_M5 IO_B9 B9 FPGA_IO<78> 78 24 CAN2_TX H18 IO_H18 IO_R11 R11 FPGA_IO<122> 122
29 FPGA_IO<29> M4 IO_M4 IO_D8 D8 FPGA_IO<79> 79 23 CAN2_RX G18 IO_G18 IO_T11 T11 FPGA_IO<123> 123
C 30 FPGA_IO<30> N1 IO_N1 IO_C8 C8 FPGA_IO<80> 80 21 P021_PWM5 G17 IO_G17 IO_U11 U11 ADDR<3> 3
31 FPGA_IO<31> N2 IO_N2 IO_A8 A8 ADDR<18> 18 87 FPGA_IO<87> G13 IO_G13 IO_V11 V11 ADDR<2> 2
32 FPGA_IO<32> M6 IO_M6 IO_B8 B8 ADDR<19> 19 88 FPGA_IO<88> G14 IO_G14 IO_V12 V12 ADDR<1> 1
7 SWITCH<7> N7 IO_N7 IO_E8 E8 FPGA_IO<81> 81 89 FPGA_IO<89> G15 IO_G15 IO_U12 U12 FPGA_IO<124> 124
8 SWITCH<8> N5 IO_N5 IO_E7 E7 FPGA_IO<82> 82 90 FPGA_IO<90> G16 IO_G16 IO_T12 T12 ADDR<0> 0
33 FPGA_IO<33> N6 IO_N6 IO_A7 A7 ADDR<20> 20 91 FPGA_IO<91> G12 IO_G12 IO_R12 R12 FPGA_IO<125> 125
34 FPGA_IO<34> N3 IO_N3 IO_B7 B7 ADDR<21> 21 22 P022 F12 IO_F12 IO_V13 V13 FPGA_IO<126> 126
35 FPGA_IO<35> N4 IO_N4 IO_D7 D7 FPGA_IO<83> 83 92 FPGA_IO<92> F18 IO_F18 IO_U13 U13 FPGA_IO<127> 127
36 FPGA_IO<36> P5 IO_P5 IO_C7 C7 FPGA_IO<84> 84 93 FPGA_IO<93> F17 IO_F17 IO_T13 T13 FPGA_IO<128> 128 LED<4..1> 4B8<
37 FPGA_IO<37> P2 IO_P2 IO_E6 E6 FPGA_IO<85> 85 94 FPGA_IO<94> F13 IO_F13 IO_R13 R13 FPGA_IO<129> 129
38 FPGA_IO<38> P3 IO_P3 IO_D6 D6 FPGA_IO<86> 86 95 FPGA_IO<95> F14 IO_F14 IO_N10 N10 FPGA_IO<130> 130
39 FPGA_IO<39> P4 IO_P4 IO_B6 B6 ADDR<22> 22 96 FPGA_IO<96> F16 IO_F16 IO_M10 M10 FPGA_IO<131> 131
40 FPGA_IO<40> R2 IO_R2 IO_C6 C6 ADDR<23> 23 97 FPGA_IO<97> F15 IO_F15 IO_P12 P12 LED<1> 1
41 FPGA_IO<41> R3 IO_R3 IO_B5 B5 AD_CLKIN 1 98 FPGA_IO<98> E17 IO_E17 IO_U14 U14 LED<2> 2
42 FPGA_IO<42> T2 IO_T2 IO_C5 C5 AD_CONVST 2 99 FPGA_IO<99> E16 IO_E16 IO_T14 T14 LED<3> 3
43 FPGA_IO<43> T3 IO_T3 IO_D5 D5 AD_SYNC 3 100 FPGA_IO<100> E15 IO_E15 IO_R14 R14 LED<4> 4
IO_A4 A4 AD_SCLK 4 101 FPGA_IO<101> D18 IO_D18 IO_V15 V15 BLS<3> 8
IO_B4 B4 AD_DOUT 5 0 CS0 D16 IO_D16 IO_U15 U15 BLS<2> 7
IO_F8 F8 AD_DIN 6 1 CS1 D15 IO_D15 IO_N11 N11 BLS<1> 6
IO_G8 G8 AD_BUSY 7 2 CS2 C17 IO_C17 IO_M11 M11 BLS<0> 5
4A6<> 3 CS3 D17 IO_D17 IO_U16 U16 OE 9
4A6<> SYM 2 OF 3 T15 WE 4
IO_T15

SWITCH<8..1> SYM 3 OF 3
ARM_CON<9..0>
IN
D FPGA_CAN<2..1>
IO
ADC_CON<7..1> ADDR<23..0>
IO IN

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:27 2005 fpga 2 of 3 6000000SCH 1 3 of 13


1 2 3 4 5 6 7 8

Figure 3-5 6000000SCH_s1_r2_3

Chapter 3 - Over-Head Tube Support (OTS) Page 33


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

SWITCHS & LEDS


A
3V3

R36
1 SWITCH<8..1> 3D1<
10K
0.1W 1%
R37 S2
2 CHS_04TB
10K 3V3
0.1W 1% 1 SWITCH<1> 1 8
I1 O1
DS1
R38
3 2 SWITCH<2> 2 7 HSMG-C650
I2 O2
R44 GREEN
10K 3 SWITCH<3> 3 6 2 1
I3 O3
0.1W 1%
4 SWITCH<4> 4 5 332
R39 I4 O4 GREEN
4 0.1W LED<4..1> 3C8>
1%
U15
10K DS2
74LVC_LCX244
0.1W 1% HSMG-C650
B R40 R45 GREEN
5 S3 2 1 OE_N 1
CHS_04TB
10K 332 18 2 LED<1> 1
GREEN Y1 A1
0.1W 1% 0.1W 16 4 LED<2> 2
Y2 A2
5 SWITCH<5> 1 8 1% 14 6 LED<3> 3
R41 I1 O1 Y3 A3
6 DS3 12 8 LED<4> 4
Y4 A4
6 SWITCH<6> 2 7 HSMS-C650
I2 O2
10K
R46 RED
0.1W 1% 7 SWITCH<7> 3 6 2 1
I3 O3 VCC=3V3;GND=GND
R42
7 8 SWITCH<8> 4 5 332
I4 O4 RED
0.1W
10K 1%
0.1W 1% DS4
HSMY-C650
R43
8 GND R47 YELLOW
2 1
10K
0.1W 1% 332
YELLOW
0.1W GND
1%

3V3 1V5

C35 C36 C37 C38 C39 C40 C41 C42 C43 C44 C45 C46 C47 C48 C49 C50
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V
10% 10% 10% 10% 10% 10% 10% 10% 10% 10% 10% 10% 10% 10% 10% 10%

GND GND

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:27 2005 fpga 3 of 3 6000000SCH 1 4 of 13


1 2 3 4 5 6 7 8

Figure 3-6 6000000SCH_s1_r2_4

Section 2.0 FeiTian Arm Logic Module, 6000000SCH Page 34


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

AD7856, 14BIT, 8CHANNEL


VCC

C76 R75 C79


A VCC

10N 10 22U
50V 0.1W 6.3V
10% 1% 10%

C77
R77 R78 R79 R80 R81
10K 10K 10K 10K 10K
22U 0.1W 0.1W 0.1W 0.1W 0.1W
6.3V 1% 1% 1% 1% 1%
10%

6
1
C78 C80

0.1U 0.1U
16V 16V
10% 10%
U21
A
AD7856
GND_ANALOG GND ADC_INTERFACE<7..1>

5 AVDD DVDD 18 U5
VCC 74LVC_LCX244
1 ANALOG_INPUT<1> 9 AIN1 CLKIN 22 CLKIN 1
2 ANALOG_INPUT<2> 10 AIN2 SCLK 23 SCLK 4
B 3 ANALOG_INPUT<3> 11 AIN3 1 OE_N<1>
4 ANALOG_INPUT<4> 12 AIN4 CONVST_N 1 CONVST 2 R76
5 ANALOG_INPUT<5> 13 AIN5 1K 1 AD_CLKIN 2 A1<1> Y1<1> 18 CLKIN 1
6 ANALOG_INPUT<6> 14 AIN6 BUSY 2 BUSY 7 0.1W 2 AD_CONVST 4 A2<1> Y2<1> 16 CONVST 2
7 ANALOG_INPUT<7> 15 AIN7 1% 3 AD_SYNC 6 A3<1> Y3<1> 14 SYNC 3
8 ANALOG_INPUT<8> 16 AIN8 SLEEP_N 3 SLEEP 4 AD_SCLK 8 A4<1> Y4<1> 12 SCLK 4

4 REFIN_REFOUT 19 OE_N<0>
C81
7 CREF1 SYNC_N 24 SYNC 3 6 AD_DIN 11 A1<0> Y1<0> 9 DIN 6
8 CREF2 13 A2<0> Y2<0> 7
0.1U DIN 21 DIN 6 5 DOUT 15 A3<0> Y3<0> 5 AD_DOUT 5
16V CAL 17 CAL_N DOUT 20 DOUT 5 7 BUSY 17 A4<0> Y4<0> 3 AD_BUSY 7
C75 10%
0.1U 6 AGND DGND 19
16V C83 VCC=VCC;GND=GND
10% VCC=3V3;GND=GND
C84
0.47U 10N GND
16V 50V
10% 10% R74
ADC_CON<7..1>
C82 IO
C 0
0.1W
10N 5%
50V
A 10% A
GND_ANALOG GND GND_ANALOG GND

ANALOG_INPUT<8..1>
IN
2

8
7

7
1

1
D4 D6 D8 D10 VCC
BAS40_04 BAS40_04 BAS40_04 BAS40_04
3

3
AK

AK

AK

AK
1 A K2 1 A K2 1 A K2 1 A K2
C85 C86 C87 C88 C89 C90 C91 C92
40V 40V 40V 40V
10N 10N 10N 10N 10N 10N 10N 10N
50V 50V 50V 50V 50V 50V 50V 50V
10% 10% 10% 10% 10% 10% 10% 10%
D5 D7 D9 D11
BAS40_04 BAS40_04 BAS40_04 BAS40_04
3

3
AK

AK

AK

AK
1 A K2 1 A K2 1 A K2 1 A K2
40V 40V 40V 40V

D A
GND_ANALOG
A
GND_ANALOG

5 of 13 GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:55:10 2005 adc 1 of 1 6000000SCH 1 5 of 13


1 2 3 4 5 6 7 8

Figure 3-7 6000000SCH_s1_r2_5

Chapter 3 - Over-Head Tube Support (OTS) Page 35


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

LPC2294 8B8> ARM_RESET

ARM_IO<30..0>
IO

A U6
74LVC-LCX14
3V3 1V8
R13
ARM_CLK 10 B A 11 U1
OUT
LPC2292FBD144
33.2
0.1W
VCC=3V3;GND=GND
3V3 1%
V18(1..0)
143 V18A

V3(9..0)
Y1 U6 14 V3A
SM7745HV-10.000 74LVC-LCX14
FL3 ARM_RESET 135 RESET_N TD1 22 CAN1_TX 26
BLM31A601S C34
1 2 4 VCC VOUT 3 13 A B 12 CPU_CLK 142 XTAL1
141 XTAL2
600 OHM 1 2 100P
ENABLE GND 50V 0 P00_TXD0 42 91 CS0 0
VCC=3V3;GND=GND P0_0_TXD0_PWM1 P1_0_CS0
C23 C24 C25 5% 1 P01_RXD0 49 90 OE 9
R12 P0_1_RXD0_PWM3_EINT0 P1_1_OE
1U 10N 0.1U 1K 10MHZ 2 P02_SCL 50 P0_2_SCL_CAP0_0
10V 50V 16V 0.1W 3 P03_SDA 58 P0_3_SDA_MAT0_0_EINT1 P1_16_TRACEPKT0 34
10% 10% 10% 1% 4 P04_SCK0 59 P0_4_SCK0_CAP0_1 P1_17_TRACEPKT1 24 ARM_CON<9..0> 7C8>
OUT
5 P05_MISO0 61 P0_5_MISO0_MAT0_1 P1_18_TRACEPKT2 15
6 P06_MOSI0 68 P0_6_MOSI0_CAP0_2 P1_19_TRACEPKT3 7
B 7 P07_SSEL 69 P0_7_SSEL0_PWM2_EINT2 P1_20_TRACESYNC 102
8 P08_TXD1 75 P0_8_TXD1_PWM4 P1_21_PIPESTAT0 95
9 P09_RXD1 76 P0_9_RXD1_PWM6_EINT3 P1_22_PIPESTAT1 86
10 P010_RTS1 78 P0_10_RTS1_CAP1_0 P1_23_PIPESTAT2 82 R92
11 P011_CTS1 83 P0_11_CTS1_CAP1_1 P1_24_TRACECLK 70 4.7K
12 P012_DSR1 84 P0_12_DSR1_MAT1_0 P1_25_EXTIN0 60 0.0625W
GND 13 P013_DTR1 85 P0_13_DTR1_MAT1_1 P1_26_RTCK 52 ARM_RTCK 0 0.1%
GND 14 P014_DCD1 92 P0_14_DCD1_EINT1 P1_27_TDO 144 ARM_TDO 1
15 P015_EINT2 99 P0_15_RI1_EINT2 P1_28_TDI 140 ARM_TDI 2
16 P016_EINT0 100 P0_16_EINT0_MAT0_2_CAP0_2 P1_29_TCK 126 ARM_TCK 3
17 P017_SCK1 101 P0_17_CAP1_2_SCK1_MAT1_2 P1_30_TMS 113 ARM_TMS 4 GND
18 P018_MISO1 121 P0_18_CAP1_3_MISO1_MAT1_3 P1_31_TRST 43 ARM_TRST 5
19 P019_MOSI1 122 P0_19_MAT1_2_MOSI1_CAP1_2
NO SOLDER 20 P020_SSEL1 123 P0_20_MAT1_3_SSEL1_EINT3
21 P021_PWM5 4 P0_21_PWM5_CAP1_3 ARM_JTAG<5..0>
IO
R28 22 P022 5 P0_22_CAP0_0_MAT0_0
READY_BUSY P015_EINT2 15 23 CAN2_RX 6 P0_23_RD2
IN
24 CAN2_TX 8 P0_24_TD2
0 25 CAN1_RX 21 P0_25_RD1
5% 0.1W 27 AIN0 23 P0_27_AIN0_CAP0_1_MAT0_1
28 AIN1 25 P0_28_AIN1_CAP0_2_MAT0_2 VSS(11..0)
R33 29 AIN2 32 P0_29_AIN2_CAP0_3_MAT0_3 VSSA_PLL 138
FEED_DOG P022 22 30 AIN3 33 P0_30_AIN3_EINT3_CAP0_0 VSSA 139
C
0 SYM 1 OF 2
5% 0.1W
R35
AGND

3V3 0
NO SOLDER FOR ISP MODE 5% 0.1W
R11 R14
P014_DCD1 14 16 P016_EINT0 N_CONFIG 0 A
AGND GND
10K 0
0.1W 5% 0.1W
R17
3V3 1% 4.7K R15
0.0625W 17 P017_SCK1 CFG_DATA 1
0.1%
0
5% 0.1W
R16
18 P018_MISO1 DCLK 2
GND
0
R34 LED 5% 0.1W
332
0.1W HEARTBEAT R18
1% 19 P019_MOSI1 N_STATUS 3
U7
DS8 0
74LVC-LCX08
HSMS-C650 5% 0.1W
D
A 12 R19
2 1 11 Y P021_PWM5 21 20 P020_SSEL1 CONF_DONE 4 CONFIG_FPGA<4..0>
IO
B 13
0
RED
5% 0.1W
VCC=3V3;GND=GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:24 2005 lpc2294 1 of 3 6000000SCH 1 6 of 13


1 2 3 4 5 6 7 8

Figure 3-8 6000000SCH_s1_r2_6

Section 2.0 FeiTian Arm Logic Module, 6000000SCH Page 36


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

LPC2294
DATA<31..0> ADDR<23..0>
IO OUT
A U1
LPC2292FBD144

0 DATA<0> 98 P2_0_D0 P3_0_A0 89 ADDR<0> 0


1 DATA<1> 105 P2_1_D1 P3_1_A1 88 ADDR<1> 1
2 DATA<2> 106 P2_2_D2 P3_2_A2 87 ADDR<2> 2
3 DATA<3> 108 P2_3_D3 P3_3_A3 81 ADDR<3> 3

VOLTAGE MONITOR 4
5
6
DATA<4>
DATA<5>
DATA<6>
109
114
115
P2_4_D4
P2_5_D5
P2_6_D6
P3_4_A4
P3_5_A5
P3_6_A6
80
74
73
ADDR<4>
ADDR<5>
ADDR<6>
4
5
6
7 DATA<7> 116 P2_7_D7 P3_7_A7 72 ADDR<7> 7
8 DATA<8> 117 P2_8_D8 P3_8_A8 71 ADDR<8> 8
9 DATA<9> 118 P2_9_D9 P3_9_A9 66 ADDR<9> 9
VCC R1 10 DATA<10> 120 P2_10_D10 P3_10_A10 65 ADDR<10> 10
27 11 DATA<11> 124 P2_11_D11 P3_11_A11 64 ADDR<11> 11
12 DATA<12> 125 P2_12_D12 P3_12_A12 63 ADDR<12> 12
10K 13 DATA<13> 127 62 ADDR<13> 13
P2_13_D13 P3_13_A13
0.0625W C1 14 DATA<14> 129 56 ADDR<14> 14
P2_14_D14 P3_14_A14
0.1% 15 DATA<15> 130 55 ADDR<15> 15
R5 0.1U P2_15_D15 P3_15_A15
10K 16V 16 DATA<16> 131 P2_16_D16 P3_16_A16 53 ADDR<16> 16
0.0625W 10% 17 DATA<17> 132 P2_17_D17 P3_17_A17 48 ADDR<17> 17
0.1% 18 DATA<18> 133 P2_18_D18 P3_18_A18 47 ADDR<18> 18
19 DATA<19> 134 P2_19_D19 P3_19_A19 46 ADDR<19> 19
20 DATA<20> 136 P2_20_D20 P3_20_A20 45 ADDR<20> 20
21 DATA<21> 137 P2_21_D21 P3_21_A21 44 ADDR<21> 21
B 22 DATA<22> 1 P2_22_D22 P3_22_A22 41 ADDR<22> 22
23 DATA<23> 10 P2_23_D23 P3_23_A23_XCLK 40 ADDR<23> 23
A 24 DATA<24> 11 36 CS3 3
P2_24_D24 P3_24_CS3
AGND 25 DATA<25> 12 P2_25_D25 P3_25_CS2 35 CS2 2
3V3 R2 26 DATA<26> 13 P2_26_D26_BOOT0 P3_26_CS1 30 CS1 1
28 27 DATA<27> 16 P2_27_D27_BOOT1 P3_27_WE 29 WE 4
28 DATA<28> 17 P2_28_D28 P3_28_BLS3_AIN7 28 BLS<3> 8
10K 29 DATA<29> 18 27 BLS<2> 7
P2_29_D29 P3_29_BLS2_AIN6
0.0625W 30 DATA<30> 19 97 BLS<1> 6
P2_30_D30_AIN4 P3_30_BLS1
0.1% C2 31 DATA<31> 20 96 BLS<0> 5
R6 P2_31_D31_AIN5 P3_31_BLS0
10K 0.1U
0.0625W 16V SYM 2 OF 2 ARM_CON<8..0> 6B8>
0.1% OUT
10%

3V3

3V3
A
R97
AGND 10K
1V8 R3 0.1W R98 3V3
29 1% 10K
0.1W
C 1K
BOOT1 DATA<27> 27 1%
0.0625W
0.1% C3
R7
10K 0.1U R9 26 DATA<26> BOOT0
0.0625W 16V NO SOLDER 10K C5 C6 C7 C9 C11 C13 C15 C17
0.1% 10% 0.1W 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
1% R10 16V 16V 16V 16V 16V 16V 16V 16V
NO SOLDER 10K 10% 10% 10% 10% 10% 10% 10% 10%
0.1W
1%

A GND
AGND
1V5 R4 GND
30 1V8
1K
BOOT 1:0 = 11 INTERNAL BOOTUP C8 C10 C12 C14 C16 C18
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
0.0625W
0.1%
10 EXTERNAL 32BIT 16V 16V 16V 16V 16V 16V
R8 C4 10% 10% 10% 10% 10% 10%
10K 0.1U 01 EXTERNAL 16BIT
0.0625W 16V
0.1% 10% 00 EXTERNAL 8BIT GND

D A
AGND ARM_IO<30..0> 6D5<>

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:24 2005 lpc2294 2 of 3 6000000SCH 1 7 of 13


1 2 3 4 5 6 7 8

Figure 3-9 6000000SCH_s1_r2_7

Chapter 3 - Over-Head Tube Support (OTS) Page 37


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

ARM RESET & WATCH DOG


A
VCC

R24
10K
0.1W
1%

EXT_RESET
IN

J2
B VCC VCC 3V3
ARM_CLK 1 2

JUMPER

R26
6D6> FEED_DOG R20 R22 C19 R25 10K
2.49K 10K 0.1U 10K 0.1W
0.1W 0.1W 16V 0.1W 1%
1% 1% U24 10% 1%
MAX6301
S1
KT11B1JM U7
8 VCC 74LVC-LCX08

1 1 3 3 1 RESET_IN 4 A
6 WDI Y 6 ARM_RESET 6A3<
OUT
2 2 4 4 5 WDS RESET_N 7 5 B
C
3 SRT VCC=3V3;GND=GND
4 SWT GND 2

C20 C21 C22 5V


R21
1K 2.2U 47N 4.7N
0.1W 10V 16V 50V
1% 10% 10% 10%
500 X TWD = 6.3 S
TRP = 125 MS

GND GND GND

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:24 2005 lpc2294 3 of 3 6000000SCH 1 8 of 13


1 2 3 4 5 6 7 8

Figure 3-10 6000000SCH_s1_r2_8

Section 2.0 FeiTian Arm Logic Module, 6000000SCH Page 38


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

MEMORY
ADDR<23..0> 9C8<
IN
DATA<31..0>
IO
I293 I294 I295
K6F1616U6C K6F1616U6C K6F1616U6C K6F1616U6C
I292
A U2 U8 U10 U12
k6f1616u6c k6f1616u6c k6f1616u6c k6f1616u6c

2 ADDR<2> A3 A0 I/O1 B6 DATA<0> 0 2 ADDR<2> A3 A0 I/O1 B6 DATA<16> 16 2 ADDR<2> A3 A0 I/O1 B6 DATA<0> 0 2 ADDR<2> A3 A0 I/O1 B6 DATA<16> 16
3 ADDR<3> A4 A1 I/O2 C5 DATA<1> 1 3 ADDR<3> A4 A1 I/O2 C5 DATA<17> 17 3 ADDR<3> A4 A1 I/O2 C5 DATA<1> 1 3 ADDR<3> A4 A1 I/O2 C5 DATA<17> 17
4 ADDR<4> A5 A2 I/O3 C6 DATA<2> 2 4 ADDR<4> A5 A2 I/O3 C6 DATA<18> 18 4 ADDR<4> A5 A2 I/O3 C6 DATA<2> 2 4 ADDR<4> A5 A2 I/O3 C6 DATA<18> 18
5 ADDR<5> B3 A3 I/O4 D5 DATA<3> 3 5 ADDR<5> B3 A3 I/O4 D5 DATA<19> 19 5 ADDR<5> B3 A3 I/O4 D5 DATA<3> 3 5 ADDR<5> B3 A3 I/O4 D5 DATA<19> 19
6 ADDR<6> B4 A4 I/O5 E5 DATA<4> 4 6 ADDR<6> B4 A4 I/O5 E5 DATA<20> 20 6 ADDR<6> B4 A4 I/O5 E5 DATA<4> 4 6 ADDR<6> B4 A4 I/O5 E5 DATA<20> 20
7 ADDR<7> C3 A5 I/O6 F5 DATA<5> 5 7 ADDR<7> C3 A5 I/O6 F5 DATA<21> 21 7 ADDR<7> C3 A5 I/O6 F5 DATA<5> 5 7 ADDR<7> C3 A5 I/O6 F5 DATA<21> 21
8 ADDR<8> C4 A6 I/O7 F6 DATA<6> 6 8 ADDR<8> C4 A6 I/O7 F6 DATA<22> 22 8 ADDR<8> C4 A6 I/O7 F6 DATA<6> 6 8 ADDR<8> C4 A6 I/O7 F6 DATA<22> 22
9 ADDR<9> D4 A7 I/O8 G6 DATA<7> 7 9 ADDR<9> D4 A7 I/O8 G6 DATA<23> 23 9 ADDR<9> D4 A7 I/O8 G6 DATA<7> 7 9 ADDR<9> D4 A7 I/O8 G6 DATA<23> 23
10 ADDR<10> H2 A8 I/O9 B1 DATA<8> 8 10 ADDR<10> H2 A8 I/O9 B1 DATA<24> 24 10 ADDR<10> H2 A8 I/O9 B1 DATA<8> 8 10 ADDR<10> H2 A8 I/O9 B1 DATA<24> 24
11 ADDR<11> H3 A9 I/O10 C1 DATA<9> 9 11 ADDR<11> H3 A9 I/O10 C1 DATA<25> 25 11 ADDR<11> H3 A9 I/O10 C1 DATA<9> 9 11 ADDR<11> H3 A9 I/O10 C1 DATA<25> 25
12 ADDR<12> H4 A10 I/O11 C2 DATA<10> 10 12 ADDR<12> H4 A10 I/O11 C2 DATA<26> 26 12 ADDR<12> H4 A10 I/O11 C2 DATA<10> 10 12 ADDR<12> H4 A10 I/O11 C2 DATA<26> 26
13 ADDR<13> H5 A11 I/O12 D2 DATA<11> 11 13 ADDR<13> H5 A11 I/O12 D2 DATA<27> 27 13 ADDR<13> H5 A11 I/O12 D2 DATA<11> 11 13 ADDR<13> H5 A11 I/O12 D2 DATA<27> 27
14 ADDR<14> G3 A12 I/O13 E2 DATA<12> 12 14 ADDR<14> G3 A12 I/O13 E2 DATA<28> 28 14 ADDR<14> G3 A12 I/O13 E2 DATA<12> 12 14 ADDR<14> G3 A12 I/O13 E2 DATA<28> 28
15 ADDR<15> G4 A13 I/O14 F2 DATA<13> 13 15 ADDR<15> G4 A13 I/O14 F2 DATA<29> 29 15 ADDR<15> G4 A13 I/O14 F2 DATA<13> 13 15 ADDR<15> G4 A13 I/O14 F2 DATA<29> 29
16 ADDR<16> F3 A14 I/O15 F1 DATA<14> 14 16 ADDR<16> F3 A14 I/O15 F1 DATA<30> 30 16 ADDR<16> F3 A14 I/O15 F1 DATA<14> 14 16 ADDR<16> F3 A14 I/O15 F1 DATA<30> 30
17 ADDR<17> F4 A15 I/O16 G1 DATA<15> 15 17 ADDR<17> F4 A15 I/O16 G1 DATA<31> 31 17 ADDR<17> F4 A15 I/O16 G1 DATA<15> 15 17 ADDR<17> F4 A15 I/O16 G1 DATA<31> 31
18 ADDR<18> E4 A16 18 ADDR<18> E4 A16 18 ADDR<18> E4 A16 18 ADDR<18> E4 A16
19 ADDR<19> D3 A17 19 ADDR<19> D3 A17 19 ADDR<19> D3 A17 19 ADDR<19> D3 A17
20 ADDR<20> H1 A18 20 ADDR<20> H1 A18 20 ADDR<20> H1 A18 20 ADDR<20> H1 A18
21 ADDR<21> G2 A19 21 ADDR<21> G2 A19 21 ADDR<21> G2 A19 21 ADDR<21> G2 A19

5 BLS<0> A1 LB 7 BLS<2> A1 LB 5 BLS<0> A1 LB 7 BLS<2> A1 LB


6 BLS<1> B2 UB 8 BLS<3> B2 UB 6 BLS<1> B2 UB 8 BLS<3> B2 UB
B OE A2 OE OE A2 OE OE A2 OE OE A2 OE
WE G5 WE WE G5 WE WE G5 WE WE G5 WE

4MB B5 CS1 4MB B5 CS1 8MB B5 CS1 8MB B5 CS1


A6 CS2 A6 CS2 A6 CS2 A6 CS2

VCC=3V3;GND=GND VCC=3V3;GND=GND VCC=3V3;GND=GND VCC=3V3;GND=GND

ARM_RESET
IN
OE 9
WE 4

ARM_CON<9..0>
IN
I291
U3 I290 U9
CS0 0
AM29LV640D AM29LV640D

2 ADDR<2> E2 A0 D0 E3 DATA<0> 0 2 ADDR<2> E2 A0 D0 E3 DATA<16> 16 3V3


3 ADDR<3> D2 A1 D1 H3 DATA<1> 1 3 ADDR<3> D2 A1 D1 H3 DATA<17> 17
C 4 ADDR<4> C2 A2 D2 E4 DATA<2> 2 4 ADDR<4> C2 A2 D2 E4 DATA<18> 18
5 ADDR<5> A2 A3 D3 H4 DATA<3> 3 5 ADDR<5> A2 A3 D3 H4 DATA<19> 19 R29
6 ADDR<6> B2 A4 D4 H5 DATA<4> 4 6 ADDR<6> B2 A4 D4 H5 DATA<20> 20 10K U23 U11
7 ADDR<7> D3 A5 D5 E5 DATA<5> 5 7 ADDR<7> D3 A5 D5 E5 DATA<21> 21 0.1W 74LVC-LCX32 74LVC-LCX04
8 ADDR<8> C3 A6 D6 H6 DATA<6> 6 8 ADDR<8> C3 A6 D6 H6 DATA<22> 22 1%
9 ADDR<9> A3 A7 D7 E6 DATA<7> 7 9 ADDR<9> A3 A7 D7 E6 DATA<23> 23 A 1 2 1 ADDR<22> 22
Y A
10 ADDR<10> B6 A8 D8 F3 DATA<8> 8 10 ADDR<10> B6 A8 D8 F3 DATA<24> 24 8MB 3
Y
11 ADDR<11> A6 A9 D9 G3 DATA<9> 9 11 ADDR<11> A6 A9 D9 G3 DATA<25> 25 2
12 ADDR<12> C6 F4 DATA<10> 10 12 ADDR<12> C6 F4 DATA<26> 26 B
A10 D10 A10 D10 VCC=3V3;GND=GND
13 ADDR<13> D6 A11 D11 G4 DATA<11> 11 13 ADDR<13> D6 A11 D11 G4 DATA<27> 27 VCC=3V3;GND=GND
14 ADDR<14> B7 A12 D12 F5 DATA<12> 12 14 ADDR<14> B7 A12 D12 F5 DATA<28> 28 3V3 CS1 1 ADDR<23..0> 9A8<
IN
15 ADDR<15> A7 A13 D13 G6 DATA<13> 13 15 ADDR<15> A7 A13 D13 G6 DATA<29> 29
16 ADDR<16> C7 A14 D14 F6 DATA<14> 14 16 ADDR<16> C7 A14 D14 F6 DATA<30> 30
17 ADDR<17> D7 A15 D15 G7 DATA<15> 15 17 ADDR<17> D7 A15 D15 G7 DATA<31> 31 R30
18 ADDR<18> E7 A16 18 ADDR<18> E7 A16 10K U23
19 ADDR<19> B3 A17 19 ADDR<19> B3 A17 0.1W 74LVC-LCX32
20 ADDR<20> C4 A18 RY_BY A4 READY_BUSY 20 ADDR<20> C4 A18 RY_BY A4 READY_BUSY 1%
21 ADDR<21> D5 A19 21 ADDR<21> D5 A19 A 4
22 ADDR<22> D4 A20 22 ADDR<22> D4 A20 3V3 4MB 6
Y
23 ADDR<23> C5 A21 WP_ACC B4 ACC 23 ADDR<23> C5 A21 WP_ACC B4 ACC 5 CS1 1
B
WE A5 WE 3V3 WE A5 WE 3V3 VCC=3V3;GND=GND
OE G2 OE VCCIO[1..0] OE G2 OE VCCIO[1..0]
F2 CE F2 CE
R88
B5 RESET B5 RESET R27 10K
D 10K 0.1W
1%
0.1W
VCC=3V3;GND=GND VCC=3V3;GND=GND 1%
READY_BUSY
OUT

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:25 2005 memory 1 of 1 6000000SCH 1 9 of 13


1 2 3 4 5 6 7 8

Figure 3-11 6000000SCH_s1_r2_9

Chapter 3 - Over-Head Tube Support (OTS) Page 39


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

INTERFACE
A
FPGA_IO<131..0>
IO

VCC
C96
I265 I266
VCC J11

B
A

A
100U J6
16V NCA NCB
GND NCA NCB
GND
10% 1 1 2 2 1 1 2 2
3 3 4 4 68 3 3 4 4 102
5 5 6 6 69 5 5 6 6 103
FPGA_CLK 7 7 8 8 ARM_CLK 70 7 7 8 8 104
IN IN
11C1< 11A8< ARM_RESET 9 9 10 10 EXT_RESET 71 9 9 10 10 105
IN OUT
11 11 12 12 72 11 11 12 12 106
0 13 13 14 14 34 13 13 14 14
1 15 15 16 16 35 73 15 15 16 16 107
2 17 17 18 18 36 74 17 17 18 18 108
3 19 19 20 20 37 75 19 19 20 20 109
DUMMY_MH3_50 DUMMY_MH3_50 4 21 21 22 22 38 76 21 21 22 22 110
1PIN 1PIN 23 23 24 24 77 23 23 24 24 111
MH2 MH1 5 25 25 26 26 39 25 25 26 26
B 1 1 6 27 27 28 28 40 78 27 27 28 28 112
7 29 29 30 30 41 79 29 29 30 30 113
1

8 31 31 32 32 42 80 31 31 32 32 114
9 33 33 34 34 43 81 33 33 34 34 115
35 35 36 36 82 35 35 36 36 116
10 37 37 38 38 44 37 37 38 38
11 39 39 40 40 45 83 39 39 40 40 117
12 41 41 42 42 46 84 41 41 42 42 118
13 43 43 44 44 47 85 43 43 44 44 119
14 45 45 46 46 48 86 45 45 46 46 120
47 47 48 48 87 47 47 48 48 121
15 49 49 50 50 49 49 49 50 50
DUMMY_MH3_50 DUMMY_MH3_50 16 51 51 52 52 50 88 51 51 52 52 122
1PIN 1PIN 17 53 53 54 54 51 89 53 53 54 54 123
MH3 MH4 18 55 55 56 56 52 90 55 55 56 56 124
1 1 19 57 57 58 58 53 91 57 57 58 58 125
59 59 60 60 92 59 59 60 60 126
1

20 61 61 62 62 54 61 61 62 62
21 63 63 64 64 55 93 63 63 64 64 127
22 65 65 66 66 56 94 65 65 66 66 128
23 67 67 68 68 57 95 67 67 68 68 129
24 69 69 70 70 58 96 69 69 70 70 130
GND 71 71 72 72 97 71 71 72 72 131
C 25 73 73 74 74 59 73 73 74 74
26 75 75 76 76 60 98 75 75 76 76
27 77 77 78 78 61 99 77 77 78 78
28
29
79
81
79 80 80
82
62
63
100
101
79
81
79 80 80
82
RESERVED FOR FUTURE USE
81 82 81 82
83 83 84 84 83 83 84 84
30 85 85 86 86 64 1 ANALOG_INPUT<1> 85 85 86 86 ANALOG_INPUT<2> 2
31 87 87 88 88 65 87 87 88 88
32 89 89 90 90 66 3 ANALOG_INPUT<3> 89 89 90 90 ANALOG_INPUT<4> 4
33 91 91 92 92 67 91 91 92 92
11B8<> RS232<3..0> 1 ISP_RXD 93 93 94 94 ISP_TXD 0 5 ANALOG_INPUT<5> 93 93 94 94 ANALOG_INPUT<6> 6
3 232_RXD 95 95 96 96 232_TXD 2 95 95 96 96
11C8<> CAN<2..1> 1 CAN_H 97 97 98 98 CAN_L 2 7 ANALOG_INPUT<7> 97 97 98 98 ANALOG_INPUT<8> 8
99 99 100 100 99 99 100 100
SFC_150_T2 SFC_150_T2

A A
GND_ANALOG GND_ANALOG
GND GND

ANALOG_INPUT<8..1>
OUT
D

Made By Date Title


GE Healthcare
Wang Zhuying 30-NOV-2005 ARM LOGIC MODULE Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:30 2005 interface 1 of 2 6000000SCH 1 10 of 13


1 2 3 4 5 6 7 8

Figure 3-12 6000000SCH_s1_r2_10

Section 2.0 FeiTian Arm Logic Module, 6000000SCH Page 40


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8
3V3
3V3 3V3

CAN & RS-232 C66 C68


U17
MAX3232C
R67

20K
C74
U4
74VHC123A
0.1U 0.1U 1U R73
16V 16V C71 0.1W 10V 332
10% 10% 16 1% 10% 14 0.1W
VCC 1U CEXT
2 10V 15 13 DS7 1%
VP REXT_CEXT Q
6 10% HSMY-C650
VN
A 1 A_N
1 C1P 2 B Q_N 4 1 2
3 C1M ISP_TXD 0 3 RD
4 C2P YELLOW
5 C2M R94
C63 C65 232_TXD 2 R68 VCC=3V3;GND=GND
0.1U 0.1U 0 P00_TXD0 11 T1IN T1OUT 14 ISP_TXD 10K R71 VCC=VCC;GND=GND
16V 16V 8 P08_TXD1 10 7 0 0.1W 0
T2IN T2OUT
10% 10% 1 ISP_RXD 13 12 P01_RXD0 0.1W 5% 1 1% 0.1W ARM_RESET
R1IN R1OUT IN 10B2<
8 R2IN R2OUT 9 5% 11C1<
1 P01_RXD0
GND 15 R93
R23 P09_RXD1 9 R70
3 232_RXD 10K
0 0.1W 0
0 0.1W 5% 1% P00_TXD0
0.1W 5%

GND GND
TP7
RS232<3..0> 10C1<>
TIN
VCC ARM_IO<30..0>
IO
B R63 R64
U16 10K 1K
0.1W 0.1W CAN_VCC
DCR010505U
1% 1% CAN_VCC

1 VS1 VOUT 14
2 VS2
ERROR_N 17
28 SYNCIN VREC 12
C60 26 0VIN1 ENABLE 16 C73
0.47U 27 0VIN2 0VOUT 13 C64 C67 0.1U
16V 47U 0.47U 16V
10% C62 10V 16V 10%
5V 1U 10% 10% U20
10V U18 TJA1050
10% VCC HCPL0720-060
TIN

5 VREF
TP5

1 VDD1 VDD2 8 8 S CANH 7 CAN_H 1

1 FPGA_CAN_TX 2 VI VO 6 1 TXD
GND CAN_GND
3V3 3V3 C69 4 5 4 6 CAN_L 2
C 0.1U GND1 GND2
RXD CANL

R59 C61 16V


U4 10% VCC=CAN_VCC;GND=CAN_GND
74VHC123A CAN<2..1> 10C1<>
20K 1U CAN_GND
R65
0.1W 10V 332
1% 10% 6 0.1W
CEXT R72
7 5 DS6 1% 1
REXT_CEXT Q
HSMY-C650 GND
9 121
A_N U19
10 12 1 2 VCC 0.1W 1%
B Q_N HCPL0720-060
0B2<
1A8< ARM_RESET 11 RD
IN
YELLOW
8 VDD2 VDD1 1 J7
R62
0 VCC=3V3;GND=GND R66
R60 0.1W 2 FPGA_CAN_RX 6 VO VI 2 1 2 2
10K 5%
0.1W C70 5 4 C72 392
1% FPGA_CAN_TX 1 0.1W
0.1U GND2 GND1 0.1U JUMPER
16V 16V 1%
10% 10%
FPGA_CAN_RX 2

R61
10K
D 0.1W
1% GND CAN_GND

GND

IO FPGA_CAN<2..1> GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:30 2005 interface 2 of 2 6000000SCH 1 11 of 13


1 2 3 4 5 6 7 8

Figure 3-13 6000000SCH_s1_r2_11

Chapter 3 - Over-Head Tube Support (OTS) Page 41


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

POWER SUPPLYS
TP1

TIN

A
R50
5.0V TO 3.3V
1K
VCC 0.1W
1% D2 TP9
BAS116
TIN

75V
2 DS5 R55 3V3
HSMG-C650 1K
POWER GREEN VR2 0.1W
1%
LT1529-3.3
INDICATOR
1

5 VIN VOUT 1
4 SHDN_N SENSE 2

R48 R49 C53 C54 GND1 3


1K 1K 10U 0.1U GND2 6 C58
0.1W 0.1W 16V 16V 100U
1% 1% +80%-20% 10% 16V
3.3V 10%

B
GND GND GND

1V5 1V8

5.0V TO 1.5V 5.0V TO 1.8V


TIN

TIN
R51 R56
1V5
TP2

TP3
1K 1K
0.125W 0.125W
VCC 1% VCC 1%
D29 D3
BAS116 BAS116

C 75V 75V
R52 R57
VR1 681 VR3 1.4K
LT1963EQ 0.0625W LT1963EQ 0.1W
0.1% 1%

R69 2 VIN VOUT 4 C55 R54 2 VIN VOUT 4 C59


1 SHDN_N SENSE_ADJ 5 10U 1 SHDN_N SENSE_ADJ 5 10U
16V 16V
1K +80%-20% 1K +80%-20%
GND<1-0> GND<1-0>
0.125W 0.125W
1% 1%
1.21V TO 20V R53 1.21V TO 20V R58
2.87K 2.87K
0.1W 0.1W
C51 C52 1% C56 C57 1%
0.1U 10U TP6 0.1U 10U TP4
16V 16V 16V 16V
10% +80%-20% TIN 10% +80%-20% TIN

GND GND
D

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:29 2005 power 1 of 1 6000000SCH 1 12 of 13


1 2 3 4 5 6 7 8

Figure 3-14 6000000SCH_s1_r2_12

Section 2.0 FeiTian Arm Logic Module, 6000000SCH Page 42


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

BOUNDRY SCAN & JTAG HEADER 3V3

5
6
8
7
R83

0.0625W
10K

R85

10K
5%
0.1W
ARM_JTAG<5..0> 1% J10
IO

4
3
2
1
1 1 2 2
5 ARM_TRST 3 3 4 4
2 ARM_TDI 5 5 6 6
4 ARM_TMS 7 7 8 8
3 ARM_TCK 9 9 10 10
0 ARM_RTCK 11 11 12 12
1 ARM_TDO 13 13 14 14
U22 ARM_RESET 15 15 16 16
IN
74LVC_LCX244 17 17 18 18
19 19 20 20
R84
1 OE_N<1> 4.7K
0.0625W PIN
FPGA_JTAG<4..0> 1 FPGA_TCK 2 A1<1> Y1<1> 18 ARM_TCK 3 0.1% STRAIGHT
IO
2 FPGA_TMS 4 A2<1> Y2<1> 16 ARM_TMS 4 DUAL_ROW_HEADER_SHROUD
0 FPGA_TDO 6 A3<1> Y3<1> 14 ARM_TDI 2
1 ARM_TDO 8 A4<1> Y4<1> 12
GND GND
B 19 OE_N<0>

11 A1<0> Y1<0> 9 JTAG_TDO


13 A2<0> Y2<0> 7
15 A3<0> Y3<0> 5
17 A4<0> Y4<0> 3

VCC=VCC;GND=GND
VCC=3V3;GND=GND
GND
3V3

3V3
U11

5
6
8
7
74LVC-LCX04
R82 GND

0.0625W
4 3 BSCAN_EN

R86

10K
5%
Y A
C 10K
0.1W
VCC=3V3;GND=GND
1%

4
3
2
1
4

J8 3V3
J9
1 2

2 2 1 1 FPGA_TCK 1
JUMPER 4 4 3 3 JTAG_TDO
ARM_TRST 6 6 5 5 FPGA_TMS 2
8 8 7 7
10 10 9 9 FPGA_TDI 3
GND

PIN
STRAIGHT
DUAL_ROW_HEADER_BREAK
GND

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet

Fri Dec 09 09:25:32 2005 bscan 1 of 1 6000000SCH 1 13 of 13


1 2 3 4 5 6 7 8

Figure 3-15 6000000SCH_s1_r2_13

Chapter 3 - Over-Head Tube Support (OTS) Page 43


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 3.0
FeiTian I/F Board, 5151217SCH
1 2 3 4 5 6 7 8

A NOTES:
1. THE VARIATION AND VERSION FOR THIS
REVISION ECR/ECO DATE SIGN OFF PWA IS 5151217 REV1.
A 18-FEBR-05 WANG ZY 2. THIS COVER SHEET CONTROLS THE RE-
C 16-SEP-05 WANG ZY VISION STATUS OF THE ENTIRE DOCUMENT.
THE REVISION OF ALL SHEETS OF THIS
1 12-DEC-05 WANG ZY DOCUMENT WILL BE AT THE SAME LEVEL.
3. THIS SCHEMATIC MEETS THE SPECIFICA-
TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
4. ELECTRONIC SIGNATURES FOR BOTH THE
SCHEMATIC REVIEWER(S) AND APPROVER(S)
CAN BE FOUND WITHIN E-PDM.
5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS.
RESISTORS ARE IN OHMS.
INDUCTORS ARE IN HENRIES
B

Made By Date Title


GE Healthcare
Wang Zhuying 24-Mar-2006 FEITIAN I/F BOARD Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Mon Dec 12 17:20:57 2005 otslb 1 of 2 5151217 1 1 of 16
1 2 3 4 5 6 7 8
Figure 3-16 5151217SCH_s1_r3_1

Section 3.0 FeiTian I/F Board, 5151217SCH Page 44


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

I16

A CANOPEN_BD_IF
SWITCH<14..1> ENCODER_IO<14..1>
LOCK_FDBK<6..1> LOCK_CTRL<8..1>
IO_BUS<63..0> VOL_MON<3..1>
CAN_ID<11..0> 24V_MTR
V_FDBK_DRIVE<1..0>
V_FDBK<1..0>

B
I3
I6 I8

DRIVE INTERFACE AMP_ENCODER


MOT_U_RELAY AMP_RELAY
VOL_MON<3..1>
LOCK_FDBK<6..1> LOCK_CTRL<8..1> CAN_ID<11..0> AMP_CAN<2..1> ENCODER_IO<14..1>
RS232<4..1>
EXT_LED<3..1> IO_BUS<63..0> IO_BUS<63..0> LOCK_CTRL<8..1> AMP_CANID<5..0>
ARM_RESET
MOTOR_24V_FDBK LOCK<6..1> EXTEND_IO<15..1> AMP_FAULT MOTOR_24V_FDBK
EN_RESET
DRIVE_24V AMP_FAULT CAN<9..1>
MTR_CONTROL AMP_RELAY MOT_U_RELAY
UIF_12V AMP_CANID<5..0>
MTR_CONTROL V_FDBK<1..0>
AMP_CAN<2..1> V_FDBK_DRIVE<1..0>

I14
I15
CONNECTORS I13 EXTEND_IO
DRIVE_24V UIF_12V
EXT_LED<3..1> CAN<9..1> POWER EN_RESET
24V_MTR
RS232<4..1> LOCK<6..1>
EXTEND_IO<15..1>
SWITCH<14..1> LAMP_PWR<1..0> LAMP_PWR<1..0> ARM_RESET
EXT_LED<3..1>

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Thu Nov 17 10:36:08 2005 otslb 2 of 2 5151217 1 2 of 16
1 2 3 4 5 6 7 8
Figure 3-17 5151217SCH_s1_r3_2

Chapter 3 - Over-Head Tube Support (OTS) Page 45


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

LOCK<6..1> OUT OUT LOCK_FDBK<6..1>

A 24V VCC
GND
24V

R18
24V_GND

0.1W
C1

392
LOCK_CTRL<8..1>

1%
IN
D10

11
0.125W

0.125W
MURS120T3

R3

R9
10K

10K
470P

GND
R24

1%

1%
50V 2 VOUT
5% 200V 1 LONG_LOCK_FDBK1
D 2.21K 3
R12

R21

0.1W
Q1 0.1W LM324

10%
25V

100
1 G

VCC
C4

1U

1%

R33

0.1W
IRFP254 1% AR1

392
FL9

1%
10K BL03RN2R1

4
0.125W S
1 LONG_LOCK
2 1

4.99K
VCC

0.1W
0.1%
1%

R6
NA

MURS120T3
Q7 R25 R30
ULN2803 LONG_LOCKED

MURS320T3
D7
GND

+80%-20%
200V
R15
200V

0.1W
2.21K 301K

392
GND

1%
0.1W 0.1W

50V
C65
1 LONG_CTL 1 18 LONG_DRV 1 D4

1U
1% 1%

D1
2 LAT_CTL 2 17 LAT_DRV 2 24V_GND MURS320T3
3 VERT_CTL 3 16 VERTLOCK_DRV 3 LONG_LOCKED
4 VERT_CLH_CTL 4 15 VERTCLH_DRV 4 VCC
5 ANG_CTL 5 14 ANG_DRV 5 24V 200V GND
6 ROT_CTL 6 13 ROT_DRV 6
B DRIVE_RELAY 7 12 RALY_IN 24V GND

R19

0.1W
392
8 11 24V 24V_GND

1%
24V_GND
ULN2803 10 C2 D11

11
0.125W

0.125W
MURS120T3

R10
R4
10K

10K
9

GND
R26

1%

1%
470P 6 VOUT
50V 200V 7 LAT_LOCK_FDBK 2
5% D 2.21K 5
R13

R22

0.1W
Q2 0.1W LM324

10%
25V

100
24V_GND 2 G

VCC
C5

1U

1%

R34

0.1W
IRFP254 1% AR1

392
1%
10K

4
0.125W S
FL10
4.99K
VCC
0.1W
0.1%
1%
R7

BL03RN2R1

MURS120T3
1 2
LAT_LOCK 2 R27 R31
LAT_LOCKED
NA

MURS320T3
D8
GND

+80%-20%
200V
R16
200V

0.1W
2.21K 301K

392
GND

1%
0.1W 0.1W

50V
C66

1U
D2
24V_GND 1% 1%
D5
MURS320T3
LAT_LOCKED
C 200V GND

24V_GND
24V VCC
GND
24V

R20
24V_GND

0.1W
C3

392
1%
D12

11
0.125W

0.125W

MURS120T3
R11
R5
10K

10K

470P

GND
R28
1%

1%

50V 9 VOUT
5% 200V 8 VERT_LOCK_FDBK 3
D 2.21K 10
R14

R23

0.1W
Q3 0.1W LM324

10%
25V

100
3 G

VCC
C6

1U

1%

R35

0.1W
IRFP254 1% AR1

392
FL11

1%
10K BL03RN2R1

4
4.99K

S
0.1W
0.1%

0.125W
R8

1 VERT_LOCK
2 3 VCC
1%
NA
MURS120T3

R29 R32
VERT_LOCKED
MURS320T3
D9

GND

+80%-20%
200V
R17
200V

0.1W

2.21K 301K
392

GND
1%

0.1W 0.1W

50V
C67
24V_GND D6

1U
D3

MURS320T3 1% 1%
D LOCK_DRV<1..6> VERT_LOCKED

200V
24V_GND
GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Fri Mar 24 12:57:57 2006 drive 1 of 3 5151217 1 3 of 16
1 2 3 4 5 6 7 8
Figure 3-18 5151217SCH_s1_r3_3

Section 3.0 FeiTian I/F Board, 5151217SCH Page 46


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

LOCK<1..6> LOCK_FDBK<6..1>
LOCK_DRV<1..6>

24V
VCC
A
C7

R51

0.1W
392
1%
0.125W

0.125W
R36

R42

11
470P

10K

10K
24V

1%

1%
50V

GND
5% R57
D22 13 VOUT
D 14 VERT_CLH_FDBK 4
R45 Q4 MURS120T3 2.21K
4 G 12

R54

0.1W
IRFP254 0.1W LM324

10%
25V

100
C10

VCC
1U

1%

R66

0.1W
10K 1% AR1

392
200V

1%
0.125W S

4
4.99K
R39

0.1W
0.1%
1% FL12 VCC

MURS120T3
BL03RN2R1
1 2 VERT_CLH 4 R58 R63

MURS320T3
D19
VERT_CLHED

200V
R48
200V

0.1W
NA

392

+80%-20%
1%
2.21K 301K
D13
D16 0.1W 0.1W

50V
C68
MURS320T3 VERT_CLHED

1U
1% 1%

200V

B
24V
VCC

C8

R52

0.1W
392
1%
0.125W

0.125W
R37

R43

11
470P
10K

10K

24V
1%

1%

50V

GND
5% R59
D23 2 VOUT
D 1 ANG_LOCK_FDBK 5
R46 Q5 MURS120T3 2.21K
5 G 3

R55

0.1W
IRFP254 0.1W LM324

10%
25V

100
C11

VCC
1U

1%

R67

0.1W
10K 1% AR2

392
200V

1%
0.125W S

4
1% FL13
4.99K
R40

VCC
0.1W
0.1%

MURS120T3

BL03RN2R1
1 2 AUG_LOCK 5 R60 R64
MURS320T3
D20

AUG_LOCKED
200V
R49
200V

0.1W

NA
392

+80%-20%
1%

2.21K 301K
D14

D17
C 0.1W 0.1W

50V
C69
MURS320T3

1U
1% 1%

200V AUG_LOCKED

24V VCC

C9

R53

0.1W
392
1%
0.125W

0.125W
R38

R44

11
10K

10K

470P 24V
1%

1%

50V

GND
5% R61
D24 6 VOUT
D 7 ROT_LOCK_FDBK 6
R47 Q6 MURS120T3 2.21K
6 G 5

R56

0.1W
IRFP254 0.1W LM324

10%
25V

100
C12

VCC
1U

1%

R68

0.1W
10K 1% AR2

392
200V

1%
0.125W S

4
4.99K
R41

0.1W
0.1%

1% FL14 VCC
MURS120T3

BL03RN2R1
1 2 ROT_LOCK 6 R62 R65
MURS320T3
D21

ROT_LOCKED
200V
R50
200V

0.1W

NA
D
392

+80%-20%
1%

2.21K 301K
D15

D18 0.1W 0.1W

50V
C70
MURS320T3

1U
1% 1%
ROT_LOCKED
200V
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:35 2005 drive 2 of 3 5151217 1 4 of 16
1 2 3 4 5 6 7 8

Figure 3-19 5151217SCH_s1_r3_4

Chapter 3 - Over-Head Tube Support (OTS) Page 47


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

K1
G2R_1_24VDC
A LOCK, BREAK 24V FEEDBACK
DS15 RALY_IN 1 COIL1
24V CB2 HSMS-C650 5 COIL2
1410-L210-L2F1-S02-5A R105 CLOSED 2
DRIVE_24V_FDBK 2 1 4 COM OPN 3 MOT_U_RELAY IO
4.99K RED D31
0.1W MURS120T3
5A FL15 0.1% 200V
240VAC,28VDC BL03RN2R1 24V
1 2 DRIVE_24V OUT R106 GND
NA AMP_RELAY
4.99K IO
0.1W
0.1%

R104

0.1W
20K
1%
U7
MC1489A
C38
1 3
IN O
B 0.1U
50V RESP_CTRL

2
10% VCC=VCC;GND=GND U2
74LVC_LCX244
GND

MOTOR 24V FEEDBACK U7


MC1489A
19 OE_N
R102 11 A1 Y1 9 DRIVER_24V_3V3 45
IN MOTOR_24V_FDBK 4 6 13 A2 Y2 7 MOTOR_24V_3V3 46
IN O 15 A3 Y3 5 UIF_12V_3V3 47
20K 0.1U
17 A4 Y4 3 AMP_FAULT_3V3 53
0.1W 10%
50V RESP_CTRL
C39

5
12V 1% VCC=VCC;GND=GND
VCC=3V3;GND=GND IO_BUS<63..0> IO

U7
CB3 GND MC1489A AMP_FAULT IN
1410-L210-L2F1-S02-2A R103
10 8
20K IN O
R100

R101
3.32K

3.32K
0.1W
0.1%

0.1W
0.1%

0.1W RESP_CTRL VCC


9

C
0.1U

2A 1% VCC=VCC;GND=GND
10%
50V
C40

240VAC,28VDC
GND

R199

R198

R217
0.1W

0.1W

0.1W
20K

20K

20K
1%

1%

1%
2
DS14
HSMS-C650

GND
RED

3
1
U2
74LVC_LCX244
1

IO IO_BUS<63..0>
1 OE_N
GND 48 EXT_LED_DRV1 2 A1 Y1 18 EXT_LED1 1
49 EXT_LED_DRV2 4 A2 Y2 16 EXT_LED2 2
50 EXT_LED_DRV3 6 A3 Y3 14 EXT_LED3 3 VCC
IN MTR_CONTROL 8 A4 Y4 12 DRIVE_RELAY
UIF_12V OUT
VCC=3V3;GND=GND EXT_LED<3..1> OUT

0.125W
R152
10K
1%
D
DRIVE_RELAY

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:36 2005 drive 3 of 3 5151217 1 5 of 16
1 2 3 4 5 6 7 8
Figure 3-20 5151217SCH_s1_r3_5

Section 3.0 FeiTian I/F Board, 5151217SCH Page 48


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8
IO MOT_U_RELAY
FL42 J19
IO LOCK_CTRL<8..1> BMB1812-P2A-121
1 2 1 1
IO AMP_CAN<2..1>
J21
J23 120 OHM
2 2 VERTICAL MOTOR
2 2 1 1 AMP_RELAY IO PIN
AMP_TXD 2 2 1 1 AMP_RXD 4 4 3 3 STRAIGHT
4 4 3 3 6 6 5 5 MATE_N_LOK
1 AMP_CAN_TX 6 6 5 5 AMP_CAN_RX 2 8 8 7 7 FL43 DS16
HSMS-C650
A 8
10
8 7 7
9 AMP_FAULT MOT_V
10
12
10 9 9
11 MOT_V 1
BMB1812-P2A-121
2
10 9 OUT 12 11 J24
12 12 11 11 IN10 5 14 14 13 13 2 1
3 IN9 14 14 13 13 IN8 1 16 16 15 15 120 OHM
0 IN7 16 15 IN6 4 18 17 FL18 AMP_TXD 1

R182
16 15 18 17 RED 1

R181
4.99K

4.99K
0.1W
0.1%

0.1W
0.1%
7 IN2 18 18 17 17 IN4 0 20 20 19 19 BMB1812-P2A-121 AMP_RXD 2 2
8 IN3 20 20 19 19 IN1 2 MOT_W 22 22 21 21 MOT_W HV_24V_GND 1 2 GND 3 3
22 22 21 21 24 24 23 23
1 IN5 24 24 23 23 26 26 25 25 120 OHM
26 26 25 25 28 28 27 27 MOTOR_24V_FDBK OUT PIN
EN_/X 28 28 27 27 EN_X 30 30 29 29 GND STRAIGHT
EN_/B 30 30 29 29 EN_B 32 32 31 31 GND FRICTIONLOCK_HEADER
EN_/A 32 32 31 31 EN_A 34 34 33 33
36 36 35 35
HV_24V_GND 38 38 37 37 HV_24V D32 CB4 FL44 24V
GND SOCKET
STRAIGHT
GND
HV_24V
40
42
40
42
39
41
39
41 HV_24V
MR760
1
BMB1812-P2A-121
2 VCC
AMPLIFIER RS232
DUAL_ROW_SOCKET_STRIP 44 44 43 43
46 46 45 45 1KV 120 OHM
48 48 47 47 6.3A U23

1000U
COPLEY AMPLIFIER 240VAC,28VDC

0.125W
63V
20%
50 49 AHV

R185
C91
50 49 MAX488E

4.99K
1410-L210-L2F1-S02-6.3A

1%
ACM-055-18 1
MA_CLK+

R193
SOCKET VCC

0.1W
0.1%
110
STRAIGHT Y 5
DUAL_ROW_SOCKET_STRIP 9 MA_CLK 3 DI
4
2
0

5
3
1

B VERT_MOTOR_ENCODER<5..0>
SL_DATA+ 8 A
Z 6 MA_CLK-
HV_24V_GND RO 2 SL_DATA 10

R188

0.1W
0.1%
110
IN AMP_CANID<5..0> 7 B
GND 4
V_FDBK_DRIVE<1..0> OUT SL_DATA-
GND
VCC

ENCODERS VCC VCC


0.1W 0.1%

0.1W 0.1%

0.125W
R175

0.1%
10K
R172

R173

J22 U20
110

110

VCC DS34C86T U21 U24

0.125W

0.125W
R177

R186
MAX488E MAX488E

4.99K

4.99K
VCC 1 1 26 26 MOT_CHI_INV 4

1%

1%
GND 2 2 27 27 MOT_CHI 5 4 ENA_C V_CLK+ SSI_CLK1+
0 MOT_CHA_INV 3 28 V_CHA 2 3 VERT_LD_CHA 0 1 1

R183
3 28 INAP OUTA VCC VCC

R191
0.1W
0.1%

0.1W
0.1%
110

110
1 MOT_CHA 4 4 29 29 V_CHA_INV 1 INAN Y 5 Y 5
2 MOT_CHB_INV 5 5 30 30 V_CHB 6 INCP OUTC 5 VERT_LD_CHB 1 1 V_CLK 3 DI 3 SSI_CLK1 3 DI
3 MOT_CHB 6 6 31 31 V_CHB_INV 7 INCN Z 6 V_CLK- Z SSI_CLK1-
6
VCC 7 7 32 32 GND V_DATA+ 8 A SSI_DATA1+
8 A
C MA_CLK+ 8 33 SL_DATA+ 12 RO 2 V_DATA 2 RO 2 SSI_DATA1 4

R179

R189
8 33 ENB_D

0.1W
0.1%

0.1W
0.1%
110

110
MA_CLK- 9 9 34 34 SL_DATA- 14 INBP OUTB 13 7 B 7 B
VCC 10 10 35 35 GND 15 INBN GND 4 GND 4
SSI_DATA1- 11 11 36 36 SSI_CLK1- 10 INDP OUTD 11
SSI_DATA1+ 12 12 37 37 SSI_CLK1+ 9 INDN V_DATA- SSI_DATA1-
12 SET_ZERO1 13 13 38 38 SCREEN
VCC 14 14 39 39 GND GND GND
SSI_DATA2- 15 15 40 40 SSI_CLK2- VCC=VCC;GND=GND
SSI_DATA2+ 16 16 41 41 SSI_CLK2+ VCC VCC
13 SET_ZERO2 17 17 42 42 SCREEN
VCC 18 18 43 43 GND
SSI_DATA3- 19 44 SSI_CLK3-
R174

R176

19 44 U22 U25
0.125W 1%

0.125W 1%

0.125W

0.125W
SSI_DATA3+ 20 45 SSI_CLK3+

R178

R187
4.99K

4.99K
20 45 MAX488E MAX488E
1K

1K

14 SET_ZERO3 21 21 46 46 SCREEN

1%

1%
VCC 22 22 47 47 GND SSI_CLK2+ SSI_CLK3+
V_DATA- 23 48 V_CLK- 1 1

R184

R192
23 48 VCC VCC

0.1W
0.1%

0.1W
0.1%
110

110
V_DATA+ 24 24 49 49 V_CLK+ Y 5 Y 5
11 V_SET_ZERO 25 25 50 50 SCREEN 5 SSI_CLK2 3 DI 7 SSI_CLK3 3 DI
Z 6 SSI_CLK2- Z 6 SSI_CLK3-
G1 G1 SHIELD 8
SSI_DATA2+ A 8
SSI_DATA3+ A
G2 R180 RO 2 SSI_DATA2 6 RO 2 SSI_DATA3 8

R190
G2

0.1W
0.1%

0.1W
0.1%
110

110
GND GND 7 B 7 B
GND GND 4 GND 4
SOCKET
STRAIGHT SSI_DATA2- SSI_DATA3-
SCSI SHIELD GND GND
D GND GND

ENCODER_IO<14..1> IO

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:42 2005 amp_encoder 1 of 1 5151217 1 6 of 16
1 2 3 4 5 6 7 8
Figure 3-21 5151217SCH_s1_r3_6

Chapter 3 - Over-Head Tube Support (OTS) Page 49


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A VCC
24V TO 5.0V

TIN
R93
24V & LAMP POWER

TP7
5.11K

R143
PS1

3.16K
0.1W
0.1%
0.125W

2
1% LM2676-ADJ

HSMG-C650
TP3

DS1
FL38

20%
33U
ZJYS51R5-2P

GREEN
5A
L1
TIN 24V_IN 1 4 2 INPUT SWIT_OUT 1 C20
CR1 6 FEEDBACK BOOST 3
SMCJ26A 7 ON_OFF

SM6T27A

1
TP1 24V_GND GND 4 10000P
50V

27V
2 3 TAB_GND 8

LSM340J

SMBJ5.0
10%

CR2

40V
R144
33V

6.7V
TIN

0.1U

0.1W
100U

100U
10%
50V

16V
10%
C16

50V
20%

C18

C22
200 OHM

1K
1%

R71

0.1W
D26

CR4

499
24V_GND

1%
0.125W
5.11K
R70

TP5 J1 24V_GND TP10


1%

D25 TIN TIN


MR760 F1 V 1 1
24V_IN 2 2
Vi 3 3
1KV 4 4
24V TO 12V

TIN
5 5 R94 12V GND
LAMP_PWR<1..0>

TP8
B OUT 0 LAMP_GND
5.11K

R148
PIN PS2

9.09K
0.1W
0.1%
TP4 STRAIGHT 0.125W LM2676-ADJ
MATE_N_LOK_WHITE FL39 1%
TIN

20%
33U

2
CB1 ZJYS51R5-2P

L2

5A

HSMG-C650
C21

DS2
1410-L210-L2F1-S02-6.3A 24V_IN 1 4 2 INPUT SWIT_OUT 1
6 FEEDBACK BOOST 3

GREEN
1 LAMP_37VAC SHIELD 7 ON_OFF
GND 4 10000P
0.1W 0.1%

0.1W 0.1%

P6SMB18A
24V_GND 2 3 TAB_GND 8 50V

LSM340J

1
6.3A 10%

40V
R206
0.1U

0.1W
100U

100U
240VAC,28VDC

10%
50V

18V
C17

50V
20%

C19

C23

50V
20%
200 OHM
4.99K

4.99K

1K
1%
R78

R79

D27

CR5

R72

0.1W
1K
1%
DS3 TP11
HSMS-C650 D28 SHIELD
R77 BAS116 TIN
0 1 2

RED 10K 75V


0.125W
1% GND
C 24V TO 3.3V 3V3

TIN
R96

TP9
5.11K

R207
1.74K
PS3

0.1W
24V 0.125W

2
FL41 LM2676-ADJ

1%
24V 1%

DS10
HSMG-C650
ZJYS51R5-2P FL40

20%
33U
24V_IN 1 4 ZJYS51R5-2P

GREEN
L3

5A
24V_IN 1 4 2 INPUT SWIT_OUT 1 C36
6 FEEDBACK BOOST 3
7 ON_OFF

1
24V_GND 2 3 GND 4 10000P
4.7U

0.1U

50V
10%
50V
10%
50V

C89

24V_GND 2 3 8
C71

TAB_GND

LSM340J

SMBJ5.0
10%

40V
R237
200 OHM

6.7V
0.1U

0.1W
100U

100U
10%
50V

16V
10%
C34

50V
20%

C35

C37
200 OHM

1K
1%

D30

CR7

R95

0.1W
499
1%
TP12

TIN

GND
GND

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Mar 28 16:24:12 2006 power 1 of 1 5151217 1 7 of 16
1 2 3 4 5 6 7 8

Figure 3-22 5151217SCH_s1_r3_7

Section 3.0 FeiTian I/F Board, 5151217SCH Page 50


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

CAN<9..1> IO

CAN_L_COLL
J28
COLLIMATOR LAMP & CAN

2
J17
CANOPEN TO SYSTEM 2 CAN_L 2 1 CAN_L_COLL
CANOPEN DEBUG 4 CAN_L_LPBK 1 1 2 3 3
A 24V
J5 J6 JUMPER J4
JUMPER
CAN_H_COLL
1 RESERVED 1 6 CAN_OPT 6 1 6 24V_GND 2 1

2
1 6 1 6 J18 2 1
2 CAN_L 2 2 7 7 CAN_H 7 2 CAN_L 2 2 7 7 CAN_H 7 3 3
3 CAN_V- 3 3 8 8 CAN_H_LPBK 8 3 CAN_V- 3 3 8 8 CAN_H_COLL 8 CAN_H_LPBK 1 1 2 3 3 COLL_CAN_H 5 5 4 4 COLL_CAN_L
4 CAN_L_LPBK 4 4 9 9 CAN_V+ 9 CAN_L_COLL 4 4 9 9 CAN_V+ 9 6 6
5 SHIELD 5 5 5 SHIELD 5 5 JUMPER 1 LAMP_37VAC 8 8 7 7 CAN_V+ 9
G1 G1 G1 G1 0 LAMP_GND 9 9
G2 G2 G2 G2 SHIELD 11 11 10 10 CAN_V- 3
0 LAMP_GND 12 12
J15
SOCKET SOCKET
STRAIGHT 2 CAN_L 1 2 STRAIGHT PIN
STD_DENSITY_NONMAGNETIC STD_DENSITY_NONMAGNETIC STRAIGHT
MATE_N_LOK
JUMPER

0.1W
R1
121
1%
J16 SHIELD J27

7 CAN_H 1 2 7 CAN_H 2 1
LAMP_PWR<1..0> IN
B JUMPER JUMPER

SWITCH<14..1> OUT

LOCK<6..1> IN
DIGITAL INPUT
VERTICAL LIMIT SWITCH LONG/LAT/VERT BRAKE/CLUTCH
J9
24V
J3 J8 5 AUG_LOCK 1 1
7 ZAXIS_SW 2 2 9 9
12V 8 COLL_SW 3 3 10 10 RS232<4..1> IO
1 1 LONG_SW 1 4 VERT_CLH 1 1 6 6 DRIVE_24V IN 6 ROT_LOCK 4 4 11 11
14 14 2 2 LAT_SW 2 3 VERT_LOCK 2 2 7 7 DRIVE_24V 11 5 5 12 12
15 15 3 3 DETENT_SW 3 2 LAT_LOCK 3 3 8 8 DRIVE_24V 12 6 6 13 13
16 16 4 4 VERT_SW 4 1 LONG_LOCK 4 4 9 9 DRIVE_24V 13 7 7 14 14 GND UIF_12V
C 17 17 5 5 ANG_SW 5 5 5 14 8 8 15 15 IN
18 18 6 6 ROT_SW 6 G1 G1 J10
19 19 7 7 ZAXIS_SW 7 G2 G2 SHIELD G1 G1
20 20 8 8 COLL_SW 8 G2 G2 SHIELD
21 21 9 9 POS_LIM_SWITCH 9 1 1 6 6
GND 22 22 10 10 NEG_LIM_SWITCH 10 SOCKET 1 RXD0_EXT 2 2 7 7 RXD1_EXT 3
23 23 11 11 EXT_LED<1> 1 STRAIGHT SOCKET 2 TXD0_EXT 3 3 8 8 TXD1_EXT 4
24 24 12 12 EXT_LED<2> 2 STD_DENSITY_NONMAGNETIC STRAIGHT 4 4 9 9
25 25 13 13 EXT_LED<3> 3 STD_DENSITY 5 5
G1 G1
G1 G1 SHIELD G2 G2 SHIELD
G2 G2

PIN
COLLIMATOR ANGLE & SWITCH GND SOCKET
STRAIGHT
GND

STRAIGHT STD_DENSITY_NONMAGNETIC
STD_DENSITY_NONMAGNETIC MH1 MH3
EXT_LED<3..1> IN MH5_00_M4WASHER MH5_00_M4WASHER
SHIELD CONSOLE RS232 & POWER
1

SHIELD
EXTERNAL LEDS MH2
MH5_00_M4WASHER
MH5
MH5_00_M4WASHER MH7 MH9

0.25W
250V

R2
20%
C15

22N
1

ALB_HOLE ALB_HOLE

1M
1%
MOUNT HOLE

1
MH4 MH6 ARM MODULE HOLE
MH5_00_M4WASHER MH5_00_M4WASHER
D 2 2
1

1
0.125W

0.125W

0.125W

VCC M1 M2 GND GND


R97

R98

R99
511

511

511

AMP_HOLE AMP_HOLE MH8 MH10


1%

1%

1%

SHIELD
2

2 1 2 1
ALB_HOLE ALB_HOLE
GE Healthcare
AMPLIFIER HOLE Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Fri Mar 24 11:29:55 2006 connectors 1 of 1 5151217 1 8 of 16
1 2 3 4 5 6 7 8
Figure 3-23 5151217SCH_s1_r3_8

Chapter 3 - Over-Head Tube Support (OTS) Page 51


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

INTERFACE FOR WALLSTAND


IO EXTEND_IO<15..1> U17 WS_INTERFACE<18..1>
74LVC_LCX244

A 19 OE_N
1 11 A1 Y1 9 1 VCC
2 13 A2 Y2 7 2 R156
3
4
15
17
A3
A4
Y3
Y4
5
3
3
4
1 WS_H_BRIDGE
10K
0.125W
VCC=3V3;GND=GND 0.1% J13
R158
2 1 IN1_1 1 1
2 D1 2 2 9 9 IN1_2 4
U15 10K 14 FB1 3 3 10 10 FS_N1 16
74LVC_LCX244 0.125W 3 IN2_2 4 4 11 11 IN2_1 5
0.1% 15 FS_N2 5 5 12 12 D2 6
R164 6 6 13 13 FB2 17
1 OE_N 3 7 7 14 14
8 8 15 15 MOTOR_CON_RST
5 2 A1 Y1 18 5 10K
6 4 A2 Y2 16 6 0.125W G1 G1 SHIELD
7 6 A3 Y3 14 7 0.1% G2 G2
8 8 A4 Y4 12 8 R166
4
10K SOCKET
VCC=3V3;GND=GND 0.125W STRAIGHT
0.1% STD_DENSITY
B GND 24V_MTR
R168 OUT
U15 5
74LVC_LCX244 HV_24V_GND
10K
0.125W
OE_N 19 0.1%
R197 HV_24V_GND
9 9 Y1 A1 11 14 6
10 7 Y2 A2 13 15
11 5 Y3 A3 15 16 10K
12 3 Y4 A4 17 17 0.125W
0.1%

VCC=3V3;GND=GND 7
R210 WS_TRANSCEIVER
VCC 10K
0.125W J14
0.1%
R211 GND
U17 8 1 1
0.125W

7 COL_EN1 2 14
R240

74LVC_LCX244 2 14
0.1%

10K
10K

8 COL_CLK 3 3 15 15 COL_EN2 12
C 0.125W 9 COL_SO 4 4 16 16 COL_SI 18
OE_N 1 0.1% 10 DIS_EN 5 5 17 17
R212 11 DIS_SO 6 6 18 18 DIS_CLK 13
13 18 Y1 A1 2 18 9 VCC 7 7 19 19 VCC
MOTOR_CON_RST 16 Y2 A2 4 ARM_RESET IN 8 8 20 20
9 14 Y3 A3 6 14 10K 9 9 21 21
10 12 Y4 A4 8 15 0.125W 12V 10 10 22 22 12V
0.1% 11 11 23 23
R213 24V 12 12 24 24 24V
VCC=3V3;GND=GND 10 13 13 25 25
10K G1 G1 SHIELD
GND 0.125W G2 G2
0.1%

SOCKET
STRAIGHT
STD_DENSITY
SHIELD

12

13
11

0.125W

0.125W

0.125W
R214

R215

R216
0%

0%

0%
0

0
MULTIPLEX WITH OTS
2

3
1

D EXTERNAL LED IN EXT_LED<3..1>

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:48 2005 extend_io 1 of 2 5151217 1 9 of 16
1 2 3 4 5 6 7 8

Figure 3-24 5151217SCH_s1_r3_9

Section 3.0 FeiTian I/F Board, 5151217SCH Page 52


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

VCC

C14
1 2
0.1U
50V
10%
R92

GND 392

2
0.1W

HSMS-C650
DS4
1%

0.25W

RED
R91
U5

1M
1%
NE555

1
B 8 VCC U6 U6
74HCT04 74HCT04
6 THRESHOLD DISCHARGE 7
5 CNTRLVOLTAGE OP 3 13 A Y 12 11 A Y 10 MOTOR_CON_RST
U6 U6 2 TRIGGER
74HCT04 74HCT04 4 RST GND 1
C13 C 3
R69 VCC=VCC;GND=GND VCC=VCC;GND=GND
IN EN_RESET 9 A Y 8 5 A Y 6 1 2 1 2 B
1 Q8

1
2.2U
10K 2.2U MMBT4401

10%
C33

50V
0.1W R90 E 2

2
VCC=VCC;GND=GND VCC=VCC;GND=GND 50V 1 2

0.1U
0.1%

10%
50V
10%

C43
2
10K
0.1W

1
0.1%
D29
BAS116
3 1

75V

C
GND

Made By Date Title


GE Healthcare
Wang Zhuying 24-Mar-2006 FEITIAN I/F BOARD Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:48 2005 extend_io 2 of 2 5151217 1 10 of 16
1 2 3 4 5 6 7 8
Figure 3-25 5151217SCH_s1_r3_10

Chapter 3 - Over-Head Tube Support (OTS) Page 53


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

IN SWITCH<14..1> U4
74LVC_LCX244 IO_BUS<63..0> IO
U1 U3
MC1489A MC1489A
R73 R80 1 OE_N<1>
1 LONG_SW 1 3 LONG_ACT 0 2 LAT_SW 1 3 LAT_ACT 1
IN O IN O 0 LONG_ACT 2 A1<1> Y1<1> 18 LONG_ACT_3V3 0
15K 15K 1 LAT_ACT 4 A2<1> Y2<1> 16 LAT_ACT_3V3 1

0.1U

0.1U
0.1W RESP_CTRL 0.1W RESP_CTRL

10%

10%
50V

50V
0.125W

0.125W
C25

C29
2 DETENT_ACT 6 14 DETENT_ACT_3V3 2
R218

R222
2

2
1% VCC=VCC;GND=GND 1% VCC=VCC;GND=GND A3<1> Y3<1>
A 2K 3 VERT_ACT 8 A4<1> Y4<1> 12 VERT_ACT_3V3 3
1%

2K
1%
12V
19 OE_N<0>
4 ANG_ACT 11 A1<0> Y1<0> 9 ANG_ACT_3V3 4
GND U1 GND U3 5 ROT_ACT 13 A2<0> Y2<0> 7 ROT_ACT_3V3 5
MC1489A MC1489A 6 ZAIXS_ACT 15 A3<0> Y3<0> 5 ZAIXS_ACT_3V3 6
R74 R81 7 COLL_ACT 17 A4<0> Y4<0> 3 COLL_ACT_3V3 7
3 DETENT_SW 4 6 DETENT_ACT 2 4 VERT_SW 4 6 VERT_ACT 3
15K IN O 15K IN O
VCC=3V3;GND=GND
0.1U

0.1U
0.1W RESP_CTRL 0.1W RESP_CTRL
10%

10%
0.125W

50V

0.125W

50V
C26

C30
R219

R223
5

5
1% VCC=VCC;GND=GND 1% VCC=VCC;GND=GND
2K
1%

2K
1%
GND

GND U1 GND U3
MC1489A MC1489A LOCK_FDBK<6..1> IN
R75 R82
5 ANG_SW 10 8 ANG_ACT 4 6 ROT_SW 10 8 ROT_ACT 5
15K IN O 15K IN O
0.1U

0.1U
0.1W RESP_CTRL 0.1W RESP_CTRL
10%

10%
0.125W

50V

0.125W

50V
C27
R220

R224

C31
9

9
1% VCC=VCC;GND=GND 1% VCC=VCC;GND=GND U10
GND 74LVC_LCX244
2K
1%

2K
1%
B 1 OE_N<1>
GND U1 GND U3 1 LONG_LOCK_FDBK 2 A1<1> Y1<1> 18 LONG_LKBK_3V3 8
MC1489A MC1489A 2 LAT_LOCK_FDBK 4 A2<1> Y2<1> 16 LAT_LKBK_3V3 9
R76 R83 3 VERT_LOCK_FDBK 6 A3<1> Y3<1> 14 VERT_LKBK_3V3 10
7 ZAXIS_SW 13 11 ZAIXS_ACT 6 8 COLL_SW 13 11 COLL_ACT 7 4 VERT_CLH_FDBK 8 A4<1> Y4<1> 12 VERT_CLBK_3V3 11
15K IN O 15K IN O
19 OE_N<0>
0.1U

0.1U
0.1W RESP_CTRL 0.1W RESP_CTRL
10%

10%
0.125W

50V

0.125W

50V
C28

C32
R225
12

12
R221

1% VCC=VCC;GND=GND 1% VCC=VCC;GND=GND 5 ANG_LOCK_FDBK 11 A1<0> Y1<0> 9 ANG_LKBK_3V3 12


2K
1%

2K
1%
12V 6 ROT_LOCK_FDBK 13 A2<0> Y2<0> 7 ROT_LKBK_3V3 13
2 V_DATA 15 A3<0> Y3<0> 5 V_DATA_3V3 14
15 V_CLK_3V3
17 A4<0> Y4<0> 3 V_CLK 1
GND GND
VCC=3V3;GND=GND
SWITCH_IO<9..0>

R153
11 V_SET_ZERO
C
0
0.125W 0%
LOCK_CTRL<8..1> OUT R154
12 SET_ZERO1
U8 0 U9 GND U11
74LVC_LCX244 0.125W 0% 74LVC_LCX244 74LVC_LCX244
R157
13 SET_ZERO2
OE_N<1> 1 OE_N<1> 1 24 1 OE_N<1>
0 25
1 LONG_CTL 18 Y1<1> A1<1> 2 LONG_CTL_3V3 28 0.125W 0% 18 Y1<1> A1<1> 2 VZERO_3V3 16 SCLK1_3V3 2 A1<1> Y1<1> 18 SSI_CLK1 3
2 LAT_CTL 16 Y2<1> A2<1> 4 LAT_CTL_3V3 29 R159 16 Y2<1> A2<1> 4 ZERO1_3V3 4 SSI_DATA1 4 A2<1> Y2<1> 16 SDATA1_3V3 17
3 VERT_CTL 14 Y3<1> A3<1> 6 VERT_CTL_3V3 30 14 SET_ZERO3 14 Y3<1> A3<1> 6 ZERO2_3V3 18 SCLK2_3V3 6 A3<1> Y3<1> 14 SSI_CLK2 5
4 VERT_CLH_CTL 12 Y4<1> A4<1> 8 VERT_CLH_CTL_3V331 12 Y4<1> A4<1> 8 ZERO3_3V3 6 SSI_DATA2 8 A4<1> Y4<1> 12 SDATA2_3V3 19
0 26
OE_N<0> 19 0.125W 0% OE_N<0> 19 27 19 OE_N<0>
GND
5 ANG_CTL 9 Y1<0> A1<0> 11 ANG_CTL_3V3 32 9 Y1<0> A1<0> 11 20 SCLK3_3V3 11 A1<0> Y1<0> 9 SSI_CLK3 7
6 ROT_CTL 7 Y2<0> A2<0> 13 ROT_CTL_3V3 33 7 Y2<0> A2<0> 13 8 SSI_DATA3 13 A2<0> Y2<0> 7 SDATA3_3V3 21
7 AMP_POS_LIM 5 Y3<0> A3<0> 15 AMP_POS_3V3 34 51 POS_LIM_3V3 5 Y3<0> A3<0> 15POS_LIM_ACT8 22 MA_CLK_3V315 A3<0> Y3<0> 5 MA_CLK 9
8 AMP_REG_LIM 3 Y4<0> A4<0> 17 AMP_REG_3V3 35 52 NEG_LIM_3V3 3 Y4<0> A4<0> 17NEG_LIM_ACT9 10 SL_DATA 17 A4<0> Y4<0> 3 SL_DATA_3V3 23

VCC=3V3;GND=GND VCC=3V3;GND=GND VCC=3V3;GND=GND


D GND GND

IO ENCODER_IO<14..1>

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:51 2005 canopen_bd_if 1 of 4 5151217 1 11 of 16
1 2 3 4 5 6 7 8
Figure 3-26 5151217SCH_s1_r3_11

Section 3.0 FeiTian I/F Board, 5151217SCH Page 54


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

12V

IN SWITCH<14..1> U12
74LVC_LCX244 IO_BUS<63..0> IO
U13
MC1489A GND 1 OE_N
R145
11 1 3 ADD<1> 2 A1 Y1 18 54
IN O 4 A2 Y2 16 55
15K 6 A3 Y3 14 56

0.1U
0.1W RESP_CTRL

10%
50V
0.125W

C45
8 12 57

R135

2
1% VCC=VCC;GND=GND A4 Y4

2K
1%
VCC=3V3;GND=GND

GND U13
MC1489A
R146
12 13 11 ADD<2>
B 15K IN O

0.1U
0.1W RESP_CTRL

10%
0.125W

50V
C44
R136

12
1% VCC=VCC;GND=GND
2K
1%

GND U13
MC1489A U12
R147 74LVC_LCX244
13 4 6 ADD<3> GND
15K IN O
19 OE_N
0.1U

0.1W RESP_CTRL
10%
0.125W

50V
C62

5
R141

1% VCC=VCC;GND=GND V_FDBK_DRIVE<1..0> 0 11 A1 Y1 9 0 V_FDBK<1..0>


2K
1%

IN 13 7 OUT
1 A2 Y2 1
15 A3 Y3 5
17 A4 Y4 3
GND U13
MC1489A VCC=3V3;GND=GND
R149
C 14 10 8 ADD<4>
15K IN O
0.1U

0.1W RESP_CTRL
10%
0.125W

50V
C63
R142

1% VCC=VCC;GND=GND
2K
1%

GND

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:50 2005 canopen_bd_if 2 of 4 5151217 1 12 of 16
1 2 3 4 5 6 7 8

Figure 3-27 5151217SCH_s1_r3_12

Chapter 3 - Over-Head Tube Support (OTS) Page 55


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

IN SWITCH<14..1> SWITCH_IO<9..0>
U27
R200
MC1489A VOLTAGE MONITOR
9 POS_LIM_SWITCH 1 3 POS_LIM_ACT 8
A 12V 15K IN O 24V R131
1 VOL_MON<3..1> OUT

0.1U
0.1W RESP_CTRL

10%
50V
C92

2
1% VCC=VCC;GND=GND 20K

R133
R226

3.32K
0.1U

0.1W
0.1%
0.125W

10%
50V
C41
0.1%
2K
0.125W
1% GND
U27 A
MC1489A
R201 12V R132 GND_ANALOG
10 NEG_LIM_SWITCH 13 11 NEG_LIM_ACT 9 2
15K IN O 20K

R134
9.09K
0.1U

0.1U

0.1W
0.1%
0.1W RESP_CTRL 0.125W

10%

10%
50V

50V
C93

C42
12
R227 1% VCC=VCC;GND=GND 0.1%

2K
0.125W
1% GND A
GND_ANALOG
R150
IN 24V_MTR 3
20K
B

R151
3.32K
0.1U

0.1W
0.1%
0.125W

10%
50V
C64
ENCODER_IO<14..1> VCC 0.1%
IO
R118
1 V_CLK
10K
VCC 0.125W A
LOCK_CTRL<8..1> 0.1% GND_ANALOG
OUT
R109 R119
1 LONG_CTL 3 SSI_CLK1 3V3 CANOPEN ID
10K 10K
0.125W 0.125W R127
0.1% 0.1% 0 CAN_ID<11..0> OUT
R110 R120 10K
2 LAT_CTL 5 SSI_CLK2
0.125W 0.1% S1
10K 10K R128 219-04MST
0.125W 0.125W 1
0.1% 0.1% 3V3 R137 S3
R111 R121 10K 0 CAN_ID<0> 1 I1 O1 8 8 219-04MST
C 3 VERT_CTL 7 SSI_CLK3 0.125W 0.1%
R129 1 CAN_ID<1> 2 I2 O2 7 10K
10K 10K 2 0.125W 0.1% 8 CAN_ID<8> 1 I1 O1 8
0.125W 0.125W 2 CAN_ID<2> 3 I3 O3 6 R138
0.1% 0.1% 10K 9 9 CAN_ID<9> 2 I2 O2 7
R112 R122 0.125W 0.1% 3 CAN_ID<3> 4 I4 O4 5
4 VERT_CLH_CTL 9 MA_CLK R130 10K 10CAN_ID<10> 3 I3 O3 6
3 0.125W 0.1%
10K 10K R139 11CAN_ID<11> 4 I4 O4 5
0.125W 0.125W 10K S2 10
0.1% 0.1% 0.125W 0.1% 219-04MST
R113 R123 R202 10K
5 ANG_CTL 11 V_SET_ZERO 4 0.125W 0.1%
4 CAN_ID<4> 1 I1 O1 8 R140
10K 10K 10K 11 GND
0.125W 0.125W 0.125W 0.1% 5 CAN_ID<5> 2 I2 O2 7
0.1% 0.1% R203 10K
R114 R124 5 6 CAN_ID<6> 3 I3 O3 6 0.125W 0.1%
6 ROT_CTL 12 SET_ZERO1
10K 7 CAN_ID<7> 4 I4 O4 5
10K 10K 0.125W 0.1%
0.125W 0.125W R204
0.1% 0.1% 6
R115 R125 10K
7 AMP_POS_LIM 13 SET_ZERO2
0.125W 0.1% GND
10K 10K
D 0.125W 0.125W R205
7
0.1% 0.1%
R116 R126 10K
8 AMP_NEG_LIM 14 SET_ZERO3 0.125W 0.1%
10K 10K
0.125W
0.1%
0.125W
0.1%
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:51 2005 canopen_bd_if 3 of 4 5151217 1 13 of 16
1 2 3 4 5 6 7 8

Figure 3-28 5151217SCH_s1_r3_13

Section 3.0 FeiTian I/F Board, 5151217SCH Page 56


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

DIAG_LED<7..0>
3V3

R84
DS5
HSMG-C650 DIAG LED DRIVE
2 1 0
332 GREEN
0.125W U26
1% DS6 74LVC_LCX244
HSMG-C650 IO_BUS<63..0> IO
R85
2 1 1 OE_N<1> 1
332 GREEN 0 DIAG_LED<0> 18 Y1<1> A1<1> 2 LED0 37
0.125W 1 DIAG_LED<1> 16 Y2<1> A2<1> 4 LED1 38
1% DS7 2 DIAG_LED<2> 14 Y3<1> A3<1> 6 LED2 39
HSMG-C650
B 3 DIAG_LED<3> 12 Y4<1> A4<1> 8 LED3 40
R86
2 1 2 OE_N<0> 19
332 GREEN 4 DIAG_LED<4> 9 Y1<0> A1<0> 11 LED4 41
0.125W 5 DIAG_LED<5> 7 Y2<0> A2<0> 13 LED5 42
1% DS8 6 DIAG_LED<6> 5 Y3<0> A3<0> 15 LED6 43
HSMG-C650 7 DIAG_LED<7> 3 Y4<0> A4<0> 17 LED7 44
R87
2 1 3
332 VCC=3V3;GND=GND
0.125W GREEN
1%
DS9
HSMG-C650
R88 GND
2 1 4
332 GREEN
0.125W
1% DS11
HSMG-C650
C R89
2 1 5
332 GREEN
0.125W
1% DS12
HSMG-C650
R107
2 1 6
332 GREEN
0.125W
1% DS13
HSMG-C650
R108
2 1 7
332 GREEN
0.125W
1%

Made By Date Title


GE Healthcare
Wang Zhuying 24-Mar-2006 FEITIAN I/F BOARD Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:51 2005 canopen_bd_if 4 of 4 5151217 1 14 of 16
1 2 3 4 5 6 7 8
Figure 3-29 5151217SCH_s1_r3_14

Chapter 3 - Over-Head Tube Support (OTS) Page 57


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

IO IO_BUS<63..0>

EXTEND_IO<15..1> IO

AMP_CANID_3V3<5..0>
VCC I394 VCC
A

B
A
I395
NCA NCB

B
A
1 1 2 2
3 3 4 4 NCA NCB
5 5 6 6 GND 1 1 2 2 GND
7 7 8 8 54 3 3 4 4
OUT ARM_RESET 9 9 10 10 55 5 5 6 6
11 11 12 12 56 7 7 8 8
39 LED2 13 13 14 14 EXTEND_IO<1> 1 0 AMP_CANID_3V3<0> 9 9 10 10 AMP_CANID_3V3<4> 4
37 LED0 15 15 16 16 EXTEND_IO<2> 2 1 AMP_CANID_3V3<1> 11 11 12 12 AMP_CANID_3V3<5> 5
40 LED3 17 17 18 18 EXTEND_IO<3> 3 13 13 14 14
41 LED4 19 19 20 20 2 AMP_CANID_3V3<2> 15 15 16 16 EXTEND_IO<7> 7
42 LED5 21 21 22 22 EXTEND_IO<4> 4 3 AMP_CANID_3V3<3> 17 17 18 18 EXTEND_IO<8> 8
23 23 24 24 57 19 19 20 20
38 LED1 25 25 26 26 EXTEND_IO<5> 5 21 21 22 22 EXTEND_IO<9> 9
43 LED6 27 27 28 28 23 23 24 24
44 LED7 29 29 30 30 EXTEND_IO<6> 6 25 25 26 26
1 LAT_ACT_3V3 31 31 32 32 27 27 28 28 EXTEND_IO<10> 10
2 DETENT_ACT_3V3 33 33 34 34 SL_DATA_3V3 23 IN V_FDBK<1..0> 0 29 29 30 30 EXTEND_IO<11> 11
35 35 36 36 1 31 31 32 32 EXTEND_IO<12> 12
3 VERT_ACT_3V3 37 37 38 38 VZERO_3V3 24 33 33 34 34
0 LONG_ACT_3V3 39 39 40 40 35 35 36 36 EXTEND_IO<13> 13
4 ANG_ACT_3V3 41 41 42 42 AMP_FAULT_3V3 53 37 37 38 38
5 ROT_ACT_3V3 43 43 44 44 39 39 40 40 EXTEND_IO<14> 14
6 ZAIXS_ACT_3V3 45 45 46 46 POS_LIM_3V3 51 0 CAN_ID<0> 41 41 42 42 EXTEND_IO<15> 15
B 7 COLL_ACT_3V3
47
49
47 48 48
50 NEG_LIM_3V3 52 1 CAN_ID<1>
43
45
43 44 44
46 MTR_CONTROL
49 50 45 46 OUT
51 51 52 52 2 CAN_ID<2> 47 47 48 48
8 LONG_LKBK_3V3 53 53 54 54 EXT_LED_DRV1 48 49 49 50 50
55 55 56 56 DRIVE_24V_3V3 45 51 51 52 52
9 LAT_LKBK_3V3 57 57 58 58 EXT_LED_DRV2 49 3 CAN_ID<3> 53 53 54 54 EN_RESET OUT
59 59 60 60 55 55 56 56
10 VERT_LKBK_3V3 61 61 62 62 EXT_LED_DRV3 50 4 CAN_ID<4> 57 57 58 58
63 63 64 64 MOTOR_24V_3V3 46 5 CAN_ID<5> 59 59 60 60
11 VERT_CLBK_3V3 65 65 66 66 UIF_12V_3V3 47 61 61 62 62
12 ANG_LKBK_3V3 67 67 68 68 ZERO1_3V3 25 63 63 64 64 3V3
13 ROT_LKBK_3V3 69 69 70 70 ZERO2_3V3 26 6 CAN_ID<6> 65 65 66 66
71 71 72 72 67 67 68 68
15 V_CLK_3V3 73 73 74 74 LONG_CTL_3V3 28 7 CAN_ID<7> 69 69 70 70 AMP_CAN_TX 1
14 V_DATA_3V3 75 75 76 76 ZERO3_3V3 27 8 CAN_ID<8> 71 71 72 72 AMP_CAN_RX 2
16 SCLK1_3V3 77 77 78 78 LAT_CTL_3V3 29 73 73 74 74
17 SDATA1_3V3 79 79 80 80 VERT_CTL_3V3 30 75 75 76 76 R160 R161
21 SDATA3_3V3 81 81 82 82 VERT_CLH_CTL_3V3 31 9 CAN_ID<9> 77 77 78 78 20K 20K
83 83 84 84 10 CAN_ID<10> 79 79 80 80 0.125W 0.125W
18 SCLK2_3V3 85 85 86 86 ROT_CTL_3V3 33 11 CAN_ID<11> 81 81 82 82 1% 1%
22 MA_CLK_3V3 87 87 88 88 ANG_CTL_3V3 32 83 83 84 84
19 SDATA2_3V3 89 89 90 90 AMP_POS_3V3 34 85 85 86 86 1 AMP_CAN_TX
20 SCLK3_3V3 91 91 92 92 AMP_NEG_3V3 35 87 87 88 88
C 1 93 93 94 94 2 89 89 90 90
3 95 95 96 96 4 91 91 92 92
7 97 97 98 98 2 93 93 94 94 2 AMP_CAN_RX
99 99 100 100 95 95 96 96
SFC_150_T2 97 97 98 98
99 99 100 100
GND GND SFC_150_T2
IO RS232<4..1> AMP_CAN<2..1> IO
A
CAN<9..1> A GND_ANALOG
IO GND_ANALOG

2
3
1
IN CAN_ID<11..0> IN VOL_MON<3..1>
VCC
3V3
12V

0.1U

C113

0.1U

C115

0.1U
C111

10%

10%

10%
50V

50V

50V
0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

C102

0.1U
10%

10%

10%

10%

10%

10%

10%

10%

10%
50V

50V

50V

50V

50V

50V

50V

50V

50V
C46

C47

C48

C49

C50

C52

C53
C51

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U

0.1U
10%

10%

10%

10%

10%

10%

10%

10%
50V

50V

50V

50V

50V

50V

50V

50V
C54

C55

C56

C57

C58

C59

C60
GND

C61

C112

0.1U

C114

0.1U

C116

0.1U
VCC

10%

10%

10%
50V

50V

50V
0.1U

0.1U

0.1U

0.1U

0.1U

C100

0.1U

0.1U

C103

0.1U

24V
C101

D
10%

10%

10%

10%

10%

10%

10%

10%
50V

50V

50V

50V

50V

50V

50V

50V
C95

C96

C97

C98

C99

C104

0.1U

C105

0.1U

C106

0.1U

C107

0.1U

C108

0.1U

C109

0.1U

C110

0.1U
10%

10%

10%

10%

10%

10%

10%
50V

50V

50V

50V

50V

50V

50V
GND

GND
GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:39 2005 interface 1 of 2 5151217 1 15 of 16
1 2 3 4 5 6 7 8

Figure 3-30 5151217SCH_s1_r3_15

Section 3.0 FeiTian I/F Board, 5151217SCH Page 58


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A CAN ADDRESS CONFIGURATION

VCC
VCC

R234 AMP_CANID<5..0> OUT


0
10K
B 0.125W U19
74LVC_LCX244
0.1U

0.1U

0.1U

C124

0.1U

C126

0.1U

C128

0.1U

C130

0.1U

C132

0.1U

C134

0.1U
0.1%
10%

10%

10%

10%

10%

10%

10%

10%

10%
50V

50V

50V

50V

50V

50V

50V

50V

50V
C24

C79

C84

R117
1
OE_N 1
10K
0.125W 0 AMP_CANID<0> 18 Y1 A1 2 AMP_CANID_3V3<0> 0
0.1% R235 1 AMP_CANID<1> 16 Y2 A2 4 AMP_CANID_3V3<1> 1
2 2 AMP_CANID<2> 14 Y3 A3 6 AMP_CANID_3V3<2> 2
3 AMP_CANID<3> 12 Y4 A4 8 AMP_CANID_3V3<3> 3
10K
3V3 0.125W
GND R194 0.1% VCC=3V3;GND=GND
3
10K
0.125W
0.1% R236
4 U19
0.1U

0.1U

0.1U

C125

0.1U

C127

0.1U

C129

0.1U

0.1U

C133

0.1U

C135

0.1U
C131
10%

10%

10%

10%

10%

10%

10%

10%

10%
50V

50V

50V

50V

50V

50V

50V

50V

50V
C76

C85

GND
C81

10K 74LVC_LCX244
0.125W
R233 0.1% OE_N 19
C 5
4 AMP_CANID<4> 9 Y1 A1 11 AMP_CANID_3V3<4> 4 AMP_CANID_3V3<5..0>
10K 5 AMP_CANID<5> 7 Y2 A2 13 AMP_CANID_3V3<5> 5
0.125W 5 Y3 A3 15 R155
0.1% 3 Y4 A4 17
GND 10K
VCC=3V3;GND=GND 0.125W
0.1%
GND

GE Healthcare
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue Dec 13 14:29:39 2005 interface 2 of 2 5151217 1 16 of 16
1 2 3 4 5 6 7 8

Figure 3-31 5151217SCH_s1_r3_16

Chapter 3 - Over-Head Tube Support (OTS) Page 59


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 4.0
OTS Console interconnection

1 2 3 4 5 6 7 8

A Inverter A
CON3 2PIN

Touch
CON1 5PIN LCD
screen
5117344-49
B CON2 2PIN CN1 31PIN B
The DB9F and DB9M
connectors of 5117344-37 Back 5117344-48
must be fasten with 4 screws. Plate
The part number of screws P3 5PIN
CN4 5PIN CN1 32PIN
are 5161171 and 5165211
C 5117344-37 P2 10PIN
C
CN3 9PIN CN4 6PIN
The DB15F and DB15M CN2 5117344-34 P4 2PIN
connectors of 5117344-35 9PIN
5117344-35 T/S control
must be fasten with 4 screws. CN5 15PIN
CN1 VGA board board
The part number of screws
15PIN
D are 5161171 and 5165211 5117344-38 D
CN9 2PIN
Lock
5151066 Photoelectric
CN7 3PIN switch
5117344-42 5117344-43
Lock on Back Plate Lock Grounding CN8 26PIN
Grounding Grounding pole
3 pole pole
E 5117344-36 E
8 1

CN1
J 5PIN

CN2 LED board


F Grounding F
pole J1 5PIN J2 8PIN
Button board
Mounting Plate

ᡔᴃ㽕∖ ˖
G G
&DEOH⛞᥹ࠄ/RFNⱘᐌᓔ㾺⚍Ϟ
ˈ 咥
ˉ 
ˈ 㑶
ˉ 
DŽ
GE customized cable
ϸϾ⛞⚍ࡴ⛁㓽ㅵPP DŽ
Provided with device by supplier

H REV MADE BY DATE


REVISIONS
PCN NO. DESCRIPTION
APPROVALS
MADE
YY/MM/DD TITLE

OTS Console interconnection 1 REV H


1 Yin Na 27 Sept 2005
CHECK
FIRST MADE FOR: Fei Tian
1
APPR
GEHL MEDICAL 1/1SHT/CONT ON
OTHER SYSTEMS 1/1

1 2 3 4 5 6 7 8

Figure 3-32 OTS Console interconnection

Section 4.0 OTS Console interconnection Page 60


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 5.0
Cable Routing
Mark
Number Part No. Description Quantity Screw Screw (new)
OTS
ASSEMBLY 1 5147719 FeiTian OTS Cable in Cable hose
1 5117344-8 Feitian OTS Collimator Cable 1
2 5117344-13 OTS Angulation Ground Cable 1
3 5117344-14 OTS Angulation Encoder Cable 1
4 5117344-15 OTS Angulation Control Cable 1
5 5117344-16 OTS Long Encoder Cable 1 2
6 5117344-17 OTS Long Control Cable 1 2
7 5117344-20 OTS Console Cable 1
8 5146500-7 FeiTian OTS CAN and CANopen Cable 1
9 5146500-8 FeiTian OTS Ground Cable 1
10 5146500-9 FeiTian Tube Stator,Fan&Pressure Switch Cable 1
11 5160469-1 FeiTian Tube Anode cable 1
12 5160469-2 FeiTian Tube Cathode Cable 1
13 5146500-12 FeiTian OTS Power Cable 1
14 5146500-13 FeiTian OTS Console Touchscreen Cable 1
15 5146500-14 FeiTianOTS Console VGA Cable 1
16 5117344-47 OTS Console GND Cable
17 5161016 Upper Cable Hose Connector
18 5161022 Middle Cable Hose Connector
19 5161187 Lower Cable Hose Connector
20 5140441 Cable Hose 1
21 5160788 Cable Hose 2 1
22 5160789 Cable Hose 3
23 5161171 Hex Nut AMP745563-1 4 4
ASSEMBLY 2 5147721 FeiTian OTS Cable in ALB Box
1 5117344-5 OTS CANopen Inbox Cable 1 2 2
2 5117344-6 OTS Encoder Inbox Cable 1 10 2
3 5117344-28 OTS Collimator angle and switch Inbox Cable 1 2
4 5117344-29 OTS Console RS232 Inbox Cable 1 2 2
5 5117344-30 OTS Collimator Lamp Inbox Cable 1
6 5117344-31 OTS power Inbox Cable 1
7 5117344-32 OTS Lock Inbox Cable 1 2 2
8 5117344-33 OTS Motor Power Inbox Cable 1
9 5161171 Hex Nut AMP745563-1 18
10 5165211 Feitian cable screw AN03321 8
ASSEMBLY 3 5147723 FeiTian OTS Cable in Power Box
2 5117344-41 Feitian OTS Power Box SP-L Cable 1
1 5117344-2 Feitian OTS ALB Power Cable 1
ASSEMBLY 4 5147727 FeiTian OTS Cable out of Cable hose and Box
1 5117344-4 OTS Motor Power Cable 1
2 5117344-10 OTS Lock Cable 1
3 5117344-50 OTS Lateral Encoder Connector Cable 1
4 5117344-51 OTS Vertical Encoder Connector Cable 1
5 5117344-40 Feitian OTS ALB Power Connecter Cable 1
Total 31pecs 5161171 Hex Nut AMP745563-1 15 0

Chapter 3 - Over-Head Tube Support (OTS) Page 61


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

ASSEMBLY 5 5151560 FeiTian Program OTS Cable in Console


5117344-34 1
5117344-35 OTS Console VGA Inbox Cable 1 2 2
5117344-36 OTS Console LED and Botton Cable 1
5117344-37 OTS Console RS232 and Touchscreen Cable 1 2 2
5117344-38 OTS Console Lock Cable 1
5117344-42 Feitian OTS Console VGA BD to mounting plate GND Cable 1
5117344-43 Feitian OTS Console mounting plate to back plate GND Cable 1
5117344-47
5117344-48
8pecs 5151066 FeiTian OTS Console--photoelectric switch 1
Total 8 5161171 Hex Nut AMP745563-1 4
5165211 Feitian cable screw AN03321 4
5150166 OTS Vertical Servo Motor 1
5150466 OTS Collimator Lamp Transfomer 1
5159613 Feitian OTS Power Filter 1
5117344-44 OTS Xaxis Switch Cable 1
5117344-45 OTS Collimator Rotation Switch Cable 1
5161024 Feitian Encoder 1

Section 5.0 Cable Routing Page 62


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A A

NOTE: The length of cable 3 out of


B Carriage cable duct is 300 mm
B
NOTE:The length of cable 3 out (include connector)
of cable hose 1 is 700mm
300+10mm

3 700+10mm 300+10mm
$QJXODWLRQHQFRUGHU Lateral Ctrl
Angulation lock
4 750+10mm 300+10mm (5117344-) Vertical Lock Ctrl
C Zaxis switch
1100+15mm 300+10mm Vertical Clutch Ctrl C
7
Long Lock Ctrl
300+10mm 8
1 650+10mm 300+10mm
collimator

Collimator (5117344-45) 700+10mm Long Lock


Switch
D NOTE:The length of
cable 15 between
(5161024)
Long Encoder D
Carriage Cable Duct
6 23(2)
and Orbit is 300mm 8 SKL OTS CAN
5 23(2) A25 J41
15 1100+15mm 300+10mm
PC VGA
12 400+10mm 300+10mm
SKL HV Transformer Cathode
11130+10mm 100mm 300+10mm
E Console 2 SKL HV Transformer Anode E
GND GND
10 450+10mm 300+10mm
SKL A25 J42
PC RS232
2 350+10mm 300+10mm
16 1100+15mm 300+10mm 300+10mm
GND 9 GND
810mm 610mm 710mm 13 SKL
See related Drawing Print for complete definition.

2
DRAWING SHEET - FOR REFERENCE ONLY.

F 4 Power F
NOTE: The length of cable 13
Out of cable chain is 610mm

In
G Power Box G

H REV MADE BY DATE


REVISIONS
ECR/ECO NO. DESCRIPTION
MADE
APPROVALS YY/MM/DD TITLE REV H
2 Li Dantong 10 Jan 2006 ECR 2023875 Change length
CHECK
FIRST MADE FOR: Feitian
4
APPR
3 Li Dantong 18 Apr 2006 ECR 2028228 Add description
SHT/CONT ON
OTHER
GEHL MEDICAL SYSTEMS
4 Li Dantong 07 Aug 2006 ECR 2031908 Change length

1 2 3 4 5 6 7 8

Figure 3-33 Cable Routing in Cable Hose

Chapter 3 - Over-Head Tube Support (OTS) Page 63


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

ALB Box
7 9(2) 10(2)
2 9(10) 10(2)

L1
J22 J8 J19 8 R1
L2
DB50-F DB9-F 2P-M
R2

3 9(2) R3
L3 J9
10(2)
9(2) 4
ALB DB15-F R4
L4 J10
DB9-F R5
10(2)
L5 9(2) 1 J6 12P-F
5P-M
DB9-F 6
J4 J1 R6
L6
5

Figure 3-34 ALB Box

Section 5.0 Cable Routing Page 64


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A A

B B

C C
2

1
D D
In Out

Transformer
Switch Power

E E
See related Drawing Print for complete definition.
DRAWING SHEET - FOR REFERENCE ONLY.

F F

G G

H REV MADE BY DATE


REVISIONS
ECR/ECO NO. DESCRIPTION
MADE
APPROVALS YY/MM/DD TITLE REV H
2 Li Dantong 16 Jan 2006 ECR 2023875 Initial Released
CHECK
FIRST MADE FOR: Feitian 2
APPR
SHT/CONT ON
OTHER
GEHL MEDICAL SYSTEMS

1 2 3 4 5 6 7 8

Figure 3-35 Cable Routing in Power Box

Chapter 3 - Over-Head Tube Support (OTS) Page 65


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A A

B B

ALB Box
C L1 (5151390) C
R1
1 Motor Lateral Lock
L2 Lateral Lock Ctrl
(5150166)
R2 Vertical Lock
Vertical Lock Ctrl
ALB 2
R3 Vertical Clutch
L3
D Vertical Clutch Ctrl D
R4
3 5161024 Lateral
Encoder
L4 Long Lock Ctrl
4 Vertical
(5117344-17)
R5 5161024
L5 Encoder
Long Lock
L6 R6

E E

Power box Out


5
In
F F

G G

H REV MADE BY DATE


REVISIONS
ECR/ECO NO. DESCRIPTION
MADE
APPROVALS YY/MM/DD TITLE REV H
1 Li Dantong 30 Oct 2005 ECO 2021188 Initial Released
CHECK
FIRST MADE FOR: Feitian 1
APPR

GEHL MEDICAL SYSTEMS SHT/CONT ON


OTHER

1 2 3 4 5 6 7 8

Figure 3-36 Cable Routing Out Of Cable Hose and Box

Section 5.0 Cable Routing Page 66


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Chapter 4 - Wall Stand


Section 1.0
FeiTian Wall Stand, 5136848SCH
1 2 3 4 5 6 7 8
A2 Internal A4 Blk 1 GE Company Confidential Internal A4 Blk2 To Shuttl e A 2
To Shuttle A2

1 1 1 7
5V 1 Ti l t i n g
15 15 15 35
GND IN1-1 Drive+ A2 A

J12
J4
3 3 3 9 2 Dr i v i n g
A CLK- D1

Motor Control Board


11 11 11 8 3 Drive- M2
CLK+ Advanced Logic Board FB1 Motor

eiTian all tand


Tilting 2 34 4

J22
2 2 Data- IN2-2
A2 5161462
J3

Load 10 10 10 33 5
Data+ FS-N2
A1 B2 Ver t i c a l

51360
E2 4 4 4 3 DC24GND 6

A4 B1
Encoder A-
12 4 7 Drive+ A5 Dr i v i n g

J11
12 12 A+ DC24GND

J3
5 5 5 5 8 M1
Drive- Motor

J13
DC24-MTR

J2
B- 9
13 13 13 6 IN1-2
B+
10

Buckhead
To Main
A1F2 A1J4 A7E1
FS-N1
11
IN2-1
1 1 1 12
AC-L
B
B Vertical Up D2
Handle Assembly

13 DC2 24V

J10
2 2 2

J1
Vertical Down FB2 DC2 24GND 5146712 AC-N
3 3 3 14
Tilting Up DC24GND Earth A4 P1
4 4 4 DC24-MTR 15
Tilting Down
J1

5 5 18
GND Earth

port
To Earth
6 6 14 Motor
DC12V
7 7 5

J3
9
Tilting Senor
19
Power Su p p l y
9 GND 22 5V 1 1 1 1
32 GND 15 15 15 15
48 CLK- 3 3 3 3
1 1 1 7 C
To Main Buckhead

Foot Up 49 11 11 11 11
C J3 3 2 2
Foot Down
8
23
CLK+
DATA- 2 2 2 2
J7

5 5 5 20 Encoder

J22
GND 24 10 10 10 10 A5

J2
DATA+
29 ENA- 5 5 5 5 E1
28 ENA+ 13 13 13 13 Ver t i c a l
2 2 2 31 6 6 6 6
3 3 CAN_L 3 ENB- Load
30 14 14 14 14

Sys CAN
CAN_GND ENB+ En c o d e r
4 4 4 J5 25 4 4 4 4
CAN_L_Lbpk
J5

J2 7 7 7 VZero
CAN_H
8 8 CAN_H_lbpk 8
5 5 CAN Shield 5 8 DC1 24V
3 3 Ver t i c a l D
D 6 Clutch
1 1 A5 C1 Clutch
7 DC1 24V 2 2 Ver t i c a l

J9
2 2 2

J8
CAN_L 3 Break1 8 8 A5 B1
3 3 3 Brake
Det e c t o r 6 6
Sys CAN

CAN_GND 1 DC1 24V Ver t i c a l


4 4 4 7 7 A5 B2
2
J6

Power CAN_L_Lbpk Break2


J4

J4 7 7
CAN_H
7 Brake2
Supply 8 8 CAN_H_lbpk 8 (res e r v e )
A6 P3 5 5 CAN Shield 5
Buckhead
To Main

5 1 1
A1F1 A1J4 A7E1

FAN1+ 6 A2 F1 FAN1
1 6 E

J13
J9
FAN1-
2
E AC-L 2 2 GND 1 Power
6
4
FAN2+ 2,3
7,8 7 A2 F2 FAN2
1 1 2 J1 FAN2-
3 3
DC1 24V
5
Input
AC-N Shield Earth
A4 P1
5161462 A1 B3
Transfer I/F Board
J8
port
To Earth

Earth 2
COL-EN1
15
COL-EN2
COL-CLK 3 Area Designations
Co n t r o l COL-SO 4
Power Su p p l y COL-SI 16 A1 - WS Main Bulkhead
F DIS-EN 5
A2 - WS Shuttle Assembly F
DIS-CLK 18
J1

6 A3 - WS Detector Housing Assembly


To Detector Housing

1 1 DIS-SO
DC12V DC5V 8,9,20,21 A4 - WS EEBox Assembly
2 2 1,7,10,19
SI- GND A5 - WS Top Assembly
3 SI+ 3 11,23
6 6 DC12V A6 - WS Detector Bulkhead
DC5V GND 12,22
7 7 13,25 A7 - WS GND port
EN-LED- DC24V
8 8 14,24
A3B5

EN-LED+ GND
J3

9 9
J6

SO-
10 10
G SO+ G

A5 D1
11 11 1 1
CLK- DC5V
12 CLK+ 12 2 2
13 13 DATA- 3
EN-KEY- 3
14 14 DATA+
EN-KEY+ 4 4
CLK-

J14
15 15 5 5

J1
GND CLK+
J2

EN- 6 6
EN+ 7 7
GND 8 8
NOTES GND 9 9
1. All sheets of this schematic will be at the same revision. The
H revision of this sheet will be the revision of the entire document.
REVISIONS
MADE
APPROVALS YY/MM/DD TITLE
REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian Wallstand Schematics
2. = Indicates signals are part of a cable
1 Ye Bin 10 Oct. 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 1
APPR
GEHL MEDICAL 5136848SCH SHT/CONT ON
- - OTHER SYSTEMS 1 /5

1 2 3 4 5 6 7 8

Figure 4-1 5136848SCH_r1_1

Chapter 4 - Wall Stand Page 67


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A1 Blk GE Company Confidential


Area Designations

To Detector Housing

To Shuttle
A A1 - WS Main Bulkhead 1 GND
1 A
A2 - WS Shuttle Assembly 2 Middle Area Select 2

A3 IonCham
IonChamber
3 Left Area Select 3
A3 - WS Detector Housing Assembly
4 AEC1 4
A4 - WS EEBox Assembly I oC h a m b e r 5 V_Out 5
A5 - WS Top Assembly J1
6 Right_Area Select 6
7

FG
A6 - WS Detector Bulkhead 7 -15 VDC Rail
8 +15 VDC Rail 8
A7 - WS GND port
9 GND 9

FG
B B

To EEBox Assembly
FG
2 2

A4 Blk1 J1
3 3 CAN_L
CAN_GND
4 4
CA N p o r t J2 7 7
CAN_L_Lbpk FG
CAN_H -S Switch Power
8 8 CAN_H_lbpk
5 5 - Supply
CAN Shield
+ A4 P2

A4 Blk1 J5
1 +S
Foot Up
C Fo ot p e d a l J3
3
Foot Down AC- AC- C
5 L N
GND

AC-N

Earth port
AC-L
AC-L
Filter Fuse A1 F1
WS P o w e r J4
A1 FT1 250V 5A FG
-S Switch Power
D AC-L D

A1 E1
- Supply
Fuse A1 F2
AC-L + A4 P1
WS 250V 5A
Filter +S
D et e c t o r J5
A1 FT2 AC- AC-
Power
Fuse A1 F3 L N
250V 5A
AC-L
AC-N
E E
Earth

Earth

FG
F F
Ethernet Port J2 Can FG
Detector

DC Port
A3 D1 Detector Power

output
J5 DC
Supply
A6 P3

AC- AC-
L N
G AC-L G
AC-N
NOTES
1. All sheets of this schematic will be at the same revision. The
revision of this sheet will be the revision of the entire document.

2. = Indicates signals are part of a cable

H REVISIONS
MADE
APPROVALS YY/MM/DD TITLE
REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian Wallstand Schematics
1 Ye Bin 10 Oct. 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 1
APPR
GEHL MEDICAL 5136848SCH SHT/CONT ON
- - OTHER SYSTEMS 2 /5

1 2 3 4 5 6 7 8

Figure 4-2 5136848SCH_r1_2

Section 1.0 FeiTian Wall Stand, 5136848SCH Page 68


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

Area Designations

A A1 - WS Main Bulkhead A

A3B5
Collimator Remote Control Board 1
A2 - WS Shuttle Assembly 1

A3 B6
Collimator Remote Control Board 2
To EEBo
DC12V
A3 - WS Detector Housing Assembly 2
SI-
3 SI+
A4 - WS EEBox Assembly 6 DC5V
A5 - WS Top Assembly 7
EN-LED-

A4 Blk1 J6
A6 - WS Detector Bulkhead 8

Assembly
EN-LED+
A7 - WS GND port 9 SO-
J1 10 J1
SO+
11
CLK-
12
B 13
CLK+ B
EN-KEY-
14 EN-KEY+
15
GND

C C
1 1
DC12V
2 SI- 4
3 SI+ 5
4 SI- 2
5 SI+ 3
6 DC5V 6
7 7
EN-LED-
8 EN-LED+ 8 J2
J2 9 SO- 9
10
D 11
SO+ 10
11
D
CLK-
12 CLK+ 12
13 EN-KEY- 13
14 EN-KEY+ 14
15 GND 15

J3 J4 J3 J4
E E

S 2 S 3 S 4 GND
S1

GND D1-1
F D1-2 F

J4
NC
C
NO

Detector
G GRID Sensor Board
in G
Housing
Sensor
A3 B7 A3 D1

NOTES
1. All sheets of this schematic will be at the same revision. The
revision of this sheet will be the revision of the entire document.

2. = Indicates signals are part of a cable


H REVISIONS
MADE
APPROVALS YY/MM/DD TITLE
REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian Wallstand Schematics
1 Ye Bin 10 Oct. 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 1
APPR
GEHL MEDICAL 5136848SCH SHT/CONT ON
- - OTHER SYSTEMS 3 /5

1 2 3 4 5 6 7 8

Figure 4-3 5136848SCH_r1_3

Chapter 4 - Wall Stand Page 69


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 2.0
FT WS Motor Control, 1234567FAB
2.1 FT WS Motor Control Top Silk

Figure 4-4 1234567FAB

Section 2.0 FT WS Motor Control, 1234567FAB Page 70


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

2.2 FT WS Motor Control Bottom Silk

Figure 4-5 1234567FAB

Chapter 4 - Wall Stand Page 71


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 3.0
FT WS Transfer Top Silk

Figure 4-6 1234567FAB

Section 3.0 FT WS Transfer Top Silk Page 72


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 4.0
WS Motor Control Board SCH
1 2 3 4 5 6 7 8

A A

NOTES:
REVISION ECR/ECO DATE SIGN OFF 1. THE VARIATION AND VERSION FOR THIS
A 02-MAR-05 YE BIN PWA IS 2000000 A.
B 2. THIS COVER SHEET CONTROLS THE RE- B
B 01-MAY-05 YE BIN VISION STATUS OF THE ENTIRE DOCUMENT.
C 18-AUG-05 YE BIN THE REVISION OF ALL SHEETS OF THIS
DOCUMENT WILL BE AT THE SAME LEVEL.
3. THIS SCHEMATIC MEETS THE SPECIFICA-
TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
4. ELECTRONIC SIGNATURES FOR BOTH THE
SCHEMATIC REVIEWER(S) AND APPROVER(S)
CAN BE FOUND WITHIN E-PDM.
5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS.
RESISTORS ARE IN OHMS.
INDUCTORS ARE IN HENRIES
C C

D D

REV 2 01July04
Made By Date Title
GE Healthcare
John Q Public DD-MMM-YYYY Design Name

B Size
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Sun Aug 21 16:41:31 2005 otslb 1 of 2 2xxxxxxSCH A 1 of 7
1 2 3 4 5 6 7 8
Figure 4-7 2xxxxxxSCH-1

Chapter 4 - Wall Stand Page 73


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A A

I3

DRIVE

DRIVE_MOTOR<10..0>

MOTOR_OUT<3..0>
B B

I14
I13
CONNECTORS
POWER
MOTOR_OUT<3..0>

C C
DRIVE_MOTOR<10..0>

D D

REV 2 01July04
GE Healthcare

B Size
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Tue May 03 17:22:28 2005 otslb 2 of 2 2xxxxxxSCH A 2 of 7
1 2 3 4 5 6 7 8
Figure 4-8 2xxxxxxSCH-2

Section 4.0 WS Motor Control Board SCH Page 74


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

IO DRIVE_MOTOR<10..0>

U1
74AHCT14
FL14
R34 BLM31A601S R216 VCC
A 4 8 B A 9 1 2 A
100 600 OHM 100

4
VCC=VCC;GND=GND 0.125W 1% C96 0.125W 1%
10N CR9 CR9 R222
50V MG052S14A300 MG052S14A300 1K
10% 17.5V 17.5V 0.125W
1%

2
R220
10K VCC

2
DS7

1
0.125W HSMH-C190
GND 1%

2
RED
AR1 C26
GND GND

8
R215 R219

1
LM358

VP
VCC 6
7 0.1U FL19
R24 10K 10K 5 VOUT 50V GND BLM31P500S
0.125W 0.125W 10%

VN
1K C97 50 OHM
0.125W 1% 1%

4
22N
1%

2
U1 50V
74AHCT14 DS6 10% VCC_SAMPLE1

8 1
AR1
HSMH-C190 LM358

VP
R23 RED 3
13 A B 12 GND R221 C101 1
4.99K 0.1U 2 VOUT

1
1K

VN
0.125W 50V
0.125W 1% 10%

4
1% R224
B VCC=VCC;GND=GND
10K
0.125W GND
B
1%
R226
GND

MG052S14A300 10K
0.125W
1%
3

C95 C17 GND VCC R214


U1 CR7 10N CR8 2.2N C103
74AHCT14 50V MG052S14A300 50V
17.5V 10% 17.5V 5% 10K FL16
R30 0.125W C98 BL03RN2R1 +24V_MOTOR1 0.1U
3 10 11 FL13 1% 1 2 R37 C102 R225 50V
1

B A 10%
BLM31A601S R210 1K 1U 10K
100 1 2 0.1U NA 0.25W 35V 0.125W
VCC=VCC;GND=GND 0.125W 1% 50V 1% 10% 1%
600 OHM 100 10%
U1 0.125W 1%
74AHCT14 C99 HV_BRIDGE1 GND
FL10
R32 BLM31A601S R211 GND MOTOR_OUT<3..0>
C 0 1 A B 2 1 2 VALUE U4
OUT C
V MC33486A
100 600 OHM 100 T
U1 VCC=VCC;GND=GND 0.125W 1% 0.125W 1% D5 D7
74AHCT14 21 VBAT OUT2A 13 MURS320T3 MURS320T3
FL11 24V_GND OUT2B 14 200V 200V
R33 BLM31A601S R212 19 ST OUT2C 15 FL17
1 3 A B 4 1 2 OUT2D 16 BL03RN2R1
3 IN1 H_OUT11 1 2 MOTOR_DRIVE_OUT11 0
100 600 OHM 100 18 IN2 GLS2 17
VCC=VCC;GND=GND 0.125W 1% 0.125W 1% NA
U1 20 WAKE OUT1A 5
74AHCT14 OUT1B 6
FL12 2 CURR OUT1C 7
R31 BLM31A601S R213 OUT1D 8
2 5 A B 6 1 2 9 NC1
10 NC2 GLS1 4
100 600 OHM 100 11 NC3
0.125W 1% 0.125W 1% 12 1
4

NC4 GND
VCC=VCC;GND=GND C14 FL18
2.2N CR7 CR8 C19 BL03RN2R1
50V MG052S14A300 MG052S14A300 2.2N H_OUT12 1 2 MOTOR_DRIVE_OUT12 1
5% 17.5V 17.5V 50V
5% NA
2

VCC D6 D8
C13 MURS320T3 MURS320T3
D D
D G Q3 G Q5 200V 200V D
0.1U GND GND IRFP254 IRFP254

REV 2 01July04
50V GND
10% S S

24V_GND
GE Healthcare

B Size
24V_GND Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Sun Aug 21 18:49:02 2005 drive 1 of 3 2xxxxxxSCH A 3 of 7
1 2 3 4 5 6 7 8
Figure 4-9 2xxxxxxSCH-3

Chapter 4 - Wall Stand Page 75


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

IO DRIVE_MOTOR<10..0>

U28
74AHCT14
FL29
R29 BLM31A601S R15 VCC
A 9 10 B A 11 1 2 A
100 600 OHM 100

4
VCC=VCC;GND=GND 0.125W 1% C2 0.125W 1%
10N CR6 CR6 R19
50V MG052S14A300 MG052S14A300 1K
10% 17.5V 17.5V 0.125W
1%

2
R236
10K VCC

2
DS5

1
0.125W

2
HSMH-C190
GND 1% RED C25
AR2
GND GND

8
R14 R18

1
LM358

VP
VCC 6 0.1U FL33
7 50V GND BLM31P500S
R8 10K 10K 5 VOUT 10% 50 OHM
0.125W 0.125W

VN
1K C3
0.125W 1% 1%

4
22N
1% VCC_SAMPLE2

2
U28 50V

1
74AHCT14 DS4 10% AR2

8
HSMH-C190 LM358

VP
R7 RED 3
9 A B 8 GND R238 C7 1
4.99K 0.1U 2 VOUT
1
1K

VN
0.125W 50V
0.125W 1% 10%

4
1% R242
B VCC=VCC;GND=GND
10K
0.125W
B
1% GND
R22
GND
MG052S14A300 10K
0.125W
1%
3

C1 GND VCC
3

C11 R13
U28 CR3 10N 2.2N C9
74AHCT14 50V CR5 50V
17.5V 10% MG052S14A300 5% 10K FL31
R25 17.5V 0.125W C4 BL03RN2R1 +24V_MOTOR2 0.1U
8 12 13 FL27 1% 1 2 R20 C8 R21 50V
1

B A 10%
BLM31A601S R9 1K 1U 10K
1

100 1 2 0.1U NA 0.25W 35V 0.125W


VCC=VCC;GND=GND 0.125W 1% 50V 1% 10% 1%
600 OHM 100 10%
U28 0.125W 1%
74AHCT14 C5 GND
FL24 HV_BRIDGE2
R28 BLM31A601S R10 GND
C 5 1 A B 2 1 2 VALUE C
V
VCC=VCC;GND=GND 100 600 OHM 100 T MOTOR_OUT<3..0>
0.125W 1% 0.125W 1% OUT
D1 D3
U28 U3 MURS320T3 MURS320T3
74AHCT14 FL25 24V_GND MC33486A 200V 200V
R27 BLM31A601S R11 FL28
6 3 A B 4 1 2 BL03RN2R1
21 VBAT OUT2A 13 1 2 MOTOR_DRIVE_OUT21 2
100 600 OHM 100 OUT2B 14
VCC=VCC;GND=GND 0.125W 1% 0.125W 1% 19 ST OUT2C 15 NA
U28 OUT2D 16
74AHCT14 3 IN1
FL26 18 IN2 GLS2 17
R26 BLM31A601S R12
7 5 A B 6 1 2 20 WAKE OUT1A 5
OUT1B 6
100 600 OHM 100 2 CURR OUT1C 7
0.125W 1% 0.125W 1% 8
4

VCC=VCC;GND=GND OUT1D
C10 C12 9 NC1 FL32
2.2N CR3 CR5 2.2N 10 NC2 GLS1 4 BL03RN2R1
50V MG052S14A300 MG052S14A300 50V 11 NC3 1 2 MOTOR_DRIVE_OUT22 3
5% 17.5V 17.5V 5% 12 NC4 GND 1
NA
VCC C21
2

D2 D4
MURS320T3 MURS320T3
D 0.1U
200V 200V D
50V GND GND GND

REV 2 01July04
10% D D
G Q2 G Q4
IRFP254 IRFP254
S S
24V_GND GE Healthcare

B Size
24V_GND Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Sun Aug 21 18:49:02 2005 drive 2 of 3 2xxxxxxSCH A 4 of 7
1 2 3 4 5 6 7 8
Figure 4-10 2xxxxxxSCH-4

Section 4.0 WS Motor Control Board SCH Page 76


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

IO DRIVE_MOTOR<10..0>
VCC C27
A A
0.1U
50V GND
10%

K1
G6B_2114P_US24VDC
+24V_MOTOR

U2 U2 8 COILP
74VHC_AHC00 74VHC_AHC00 1 COILN
+24V_MOTOR1
4 A 9 A 3 OPEN1 OPEN2 4
Y 6 Y 8 6 CLOSED1 CLOSED2 5
10 5 B 10 B

VCC=VCC;GND=GND VCC=VCC;GND=GND
Q1
ULN2803

1 I1 O1 18
B 2
3
I2
I3
O2
O3
17
16
B
4 I4 O4 15
5 I5 O5 14
6 I6 O6 13
7 I7 O7 12
8 I8 O8 11 +24V_MOTOR
10
COMMON
GND 9

U2 U2
74VHC_AHC00 74VHC_AHC00 K2
+24V_MOTOR G6B_2114P_US24VDC
1 A 12 A
Y 3 Y 11
10 2 B 13 B 8 COILP
1 COILN
+24V_MOTOR2
VCC=VCC;GND=GND VCC=VCC;GND=GND 24V_GND 3 OPEN1 OPEN2 4
6 CLOSED1 CLOSED2 5
C C
24V_GND

D D

REV 2 01July04
GE Healthcare

B Size
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Sun Aug 21 18:49:02 2005 drive 3 of 3 2xxxxxxSCH A 5 of 7
1 2 3 4 5 6 7 8
Figure 4-11 2xxxxxxSCH-5

Chapter 4 - Wall Stand Page 77


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

24V & LAMP POWER

A A

24V TO 5.0V

TIN
R93

TP7
5.11K VCC
0.125W R4
TP3 1% 3.16K
0.1W PS1
TIN 24V_IN 24V 1% LM2676
D25 FL1

20%
33U
MR760 ZJYS51R5-2P VCC_O

5A
L1
24V_IN1 24V_IN 1 4 2 INPUT SWIT_OUT 1 C20 R36
J1 6 FEEDBACK BOOST 3 499
1KV 7 ON_OFF 0.125W

SM6T27A
GND 4 10000P 1%
1 2 3 8 50V

27V
1 TAB_GND

LSM340J
SMCJ26A

SMBJ5.0
100U

0.1U
10%

50V
B B

10%
C16

50V
20%

C18
2

CR2

40V
2
1000U

6.7V
100U

100U
C15

63V
20%

C22

C24
33V

16V
10%

16V
10%
3 3 200 OHM R6
4
CR1

D26

CR4
1K

2
4
0.1W DS1
24V_GND 1% HSMG-C190

TIN
PIN GREEN

TP10
STRAIGHT
MATE_N_LOK

1
24V_GND DS3 GND
GND HSMH-C190

2 1

RED
R3 R5
4.99K 4.99K
R1 R2 0.125W 0.125W 24V_GND
4.99K 4.99K CB1 1% 1%
0.125W 0.125W 1410-L210-L2F1-S02-10A
1% 1% +24V_MOTOR
C C
2

DS2 10A
HSMG-C190 240VAC,28VDC
GREEN
1

24V_GND

D D

REV 2 01July04
GE Healthcare

B Size
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Sun Aug 21 18:49:06 2005 power 1 of 1 2xxxxxxSCH A 6 of 7
1 2 3 4 5 6 7 8
Figure 4-12 2xxxxxxSCH-6

Section 4.0 WS Motor Control Board SCH Page 78


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

IO DRIVE_MOTOR<10..0>

A A

MOTOR_OUT<3..0> IN

FL2
BLM31P601SG FL8
1 1 2 BLM31P601SG J3
1 2 0 0 MOTOR_DRIVE_OUT11
600 OHM
FL3 600 OHM 1 1
BLM31P601SG FL9 2 2
3 1 2 BLM31P601SG 3 3
1 2 2 1 MOTOR_DRIVE_OUT12
600 OHM J2
FL4 600 OHM PIN
BLM31P601SG FL20 STRAIGHT
5 1 2 1 1 BLM31P601SG MATE_N_LOK
9 9 2 2 1 2 4
600 OHM
B FL5
10
11
10
11
3
4
3
4 600 OHM B
BLM31P601SG 12 12 5 5 FL21
7 1 2 13 13 6 6 BLM31P601SG
14 14 7 7 1 2 6
600 OHM 15 15 8 8
FL6 600 OHM
BLM31P601SG G1 G1 FL22
9 1 2 G2 G2 BLM31P601SG
1 2 8
600 OHM
PIN 600 OHM
FL15 STRAIGHT
BLM31P601SG STD_DENSITY
10 1 2 FL23 J4
BLM31P500S
600 OHM 1 2 2 MOTOR_DRIVE_OUT21
1 1
50 OHM 2 2
3 3
FL7
24V_IN BLM31P500S 3 MOTOR_DRIVE_OUT22
1 2 PIN
C 50 OHM STRAIGHT C
MATE_N_LOK

MH1 MH3 MH5


MH5_00_M4WASHER MH5_00_M4WASHER MH5_00_M4WASHER
1

R35 C23
1MEG 22N
0.5W 250V
1% 20%

MH2 MH4
MH5_00_M4WASHER MH5_00_M4WASHER
1

D D

REV 2 01July04
GE Healthcare

B Size
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Sun Aug 21 18:49:10 2005 connectors 1 of 1 2xxxxxxSCH A 7 of 7
1 2 3 4 5 6 7 8
Figure 4-13 2xxxxxxSCH-7

Chapter 4 - Wall Stand Page 79


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 5.0
Remote Control Transfer
1 2 3 4 5 6 7 8

A A

NOTES:
REVISION ECR/ECO DATE SIGN OFF 1. THE VARIATION AND VERSION FOR THIS
A 02-MAR-05 YE BIN PWA IS 2000000 A.
B 2. THIS COVER SHEET CONTROLS THE RE- B
DD-MM-YY VISION STATUS OF THE ENTIRE DOCUMENT.
DD-MM-YY THE REVISION OF ALL SHEETS OF THIS
DOCUMENT WILL BE AT THE SAME LEVEL.
3. THIS SCHEMATIC MEETS THE SPECIFICA-
TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
4. ELECTRONIC SIGNATURES FOR BOTH THE
SCHEMATIC REVIEWER(S) AND APPROVER(S)
CAN BE FOUND WITHIN E-PDM.
5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS.
RESISTORS ARE IN OHMS.
INDUCTORS ARE IN HENRIES
C C

D D

REV 2 01July04
Made By Date Title
GE Healthcare
John Q Public DD-MMM-YYYY Design Name

B Size
Company Restricted
Block Path Block Sheet Dwg No Rev Sheet
Sun Aug 21 16:41:31 2005 otslb 1 of 2 2xxxxxxSCH A 1 of 7
1 2 3 4 5 6 7 8
Figure 4-14 2xxxxxxSCH-1

Section 5.0 Remote Control Transfer Page 80


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

BUS_FPGA<7..0>

A A

FL1
BLM31P601SG
0 COL_ENDIS 1 2
FL9
600 OHM BLM31P601SG
1 2 COL_ENLED 1
FL2
BLM31P601SG 600 OHM VCC
2 COL_CLK 1 2 R2
J1 FL10
600 OHM BLM31P601SG

2
1 2 COL_SI 3 C1 C4 1K
DS1
FL3 1 100U 0.125W
1 0.1U HSMG-C190
BLM31P601SG 2 14 600 OHM 50V 50V 1%
2 14 GREEN
B 4 COL_SO 1 2 3 3 15 15 20% 10% B
4 4 16 16 FL11

1
600 OHM 5 17 BLM31P601SG
5 17
6 6 18 18 1 2 DIS_CLK 6
FL4 7 7 19 19
BLM31P601SG 8 20 600 OHM GND
8 20
5 DIS_EN 1 2 9 9 21 21
10 10 22 22
600 OHM 11 23 12V
11 23 J4
12 12 24 24
FL5 13 13 25 25 R3
BLM31P601SG VCC_FPGA 1 1
7 DIS_SO 1 2 G1 G1 2 2

2
G2 C2 C5 2K
G2 DS2
600 OHM 100U 0.125W
0.1U HSMG-C190
50V 50V 1%
PIN GREEN
SOCKET STRAIGHT 20% 10%
STRAIGHT MATE_N_LOK

1
STD_DENSITY

GND_FPGA
VCC_FPGA J5 GND
12V_FPGA
C 24V_FPGA C
12V_FPGA 1 1
2 2
2

2
24V
PIN R1
FL6 FL7 FL8 FL12 STRAIGHT

2
BL03RN2R1 BL03RN2R1 BL03RN2R1 BL03RN2R1 MATE_N_LOK DS3
C3 C6 4.99K
NA NA NA NA HSMG-C190
100U 0.125W
0.1U GREEN
50V 50V 1%
J6
VCC 12V 24V 20% 10%
1

1
24V_FPGA 1 1
GND GND_FPGA 2 2
GND

PIN
MH1 MH3 STRAIGHT
MH5_00_M4WASHER MH5_00_M4WASHER MATE_N_LOK
1

MH5
MH5_00_M4WASHER
1

D MH2 MH4 C19 R17 D


MH5_00_M4WASHER MH5_00_M4WASHER 22N 1MEG
250V 0.5W
1

20% 1%

REV 2 01July04
GE Healthcare
GND Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Tue Mar 08 17:10:30 2005 ftwsremotecontrol 2 of 6 2xxxxxxSCH A 2 of 6


1 2 3 4 5 6 7 8

Figure 4-15 2xxxxxxSCH-2

Chapter 4 - Wall Stand Page 81


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A A

DISPLAY_OUT<5..0>

FL13
BLM31P601SG FL19
B 4 DIS_DATAN 1 2 J2 BLM31P601SG B
1 2 DIS_ENN 0
600 OHM
FL14 VCC_DIS 1 6 600 OHM
1 6
BLM31P601SG 2 2 7 7 FL20
5 DIS_DATAP 1 2 3 3 8 8 BLM31P601SG
4 4 9 9 1 2 DIS_ENP 1
600 OHM 5 5
FL15 G1 GND_DIS 600 OHM
G1
BLM31P601SG G2 G2
2 DIS_CLKN 1 2
600 OHM
SOCKET
FL16 STRAIGHT
BLM31P601SG STD_DENSITY_NONMAGNETIC

2
3 DIS_CLKP 1 2
600 OHM
FL17 FL18
BLM31P500S BLM31P500S
50 OHM 50 OHM

C VCC C
1

1
C7 C8
100U 0.1U
50V 50V
20% 10%

GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Tue Mar 08 17:10:30 2005 ftwsremotecontrol 3 of 6 2xxxxxxSCH A 3 of 6


1 2 3 4 5 6 7 8
Figure 4-16 2xxxxxxSCH-3

Section 5.0 Remote Control Transfer Page 82


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

COL_REMOTE_OUT<11..0>

A A
12V

C11 C12
47U 0.1U
35V 50V

2
20% 10%

FL27
BL03RN2R1
FL21 NA FL30
BLM31P601SG BLM31P601SG
6 COL_SIN 1 2 1 2 COL_SO2N 10

1
600 OHM 12V_COL 600 OHM
FL22 FL31
BLM31P601SG BLM31P601SG
7 COL_SIP 1 2 1 2 COL_SO2P 11
J3
600 OHM 600 OHM
FL23 1 1 FL32
B BLM31P601SG 2 2 9 9 BLM31P601SG B
8 COL_SON 1 2 3 3 10 10 1 2 COL_CLKN 4
4 4 11 11
600 OHM 5 12 600 OHM
5 12
FL24 6 6 13 13 FL33
BLM31P601SG 7 7 14 14 BLM31P601SG
9 COL_SOP 1 2 8 8 15 15 1 2 COL_CLKP 5
600 OHM G1 600 OHM
FL34
G1
FL25 G2 G2 BLM31P601SG
BLM31P601SG 1 2 COL_ENKEYN 2
0 COL_ENDISN 1 2
600 OHM
SOCKET
600 OHM FL35
STRAIGHT
FL26 STD_DENSITY GND_COL BLM31P601SG
BLM31P601SG 1 2 COL_ENKEYP 3
1 COL_ENDISP 1 2
VCC_COL 600 OHM
600 OHM

2
C FL28 FL29 C
BL03RN2R1 BL03RN2R1
NA NA
VCC
1

1
C9 C10
47U 0.1U
35V 50V
20% 10%

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Tue Mar 08 17:10:30 2005 ftwsremotecontrol 4 of 6 2xxxxxxSCH A 4 of 6


1 2 3 4 5 6 7 8
Figure 4-17 2xxxxxxSCH-4

Chapter 4 - Wall Stand Page 83


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

DISPLAY_OUT<5..0>
BUS_FPGA<7..0>

A A

VCC

R4
U1 U1 1K
74HCT04 74HCT04 0.125W
1% DIS_ENN 0
5 1 2 9 8
A Y A Y
R6
121
VCC=VCC;GND=GND VCC=VCC;GND=GND 0.125W
1%
DIS_ENP 1

U2
DS34C87T DIS_CLKN 2
U1 U1
74HCT04 74HCT04
4 A_B_CNTRL R7
B 6 3
A Y
4 11
A Y
10 1 INA INVA 2 121 B
NON_INVA 3 0.125W
7 INB INVB 6 1%
VCC=VCC;GND=GND VCC=VCC;GND=GND NON_INVB 5 DIS_CLKP 3
12 C_D_CNTRL
9 INC INVC 10
NON_INVC 11
15 IND INVD 14 DIS_DATAN 4
U1 U1 NON_INVD 13
74HCT04 74HCT04 R8
121
7 5 6 13 12 VCC=VCC;GND=GND 0.125W
A Y A Y 1%

VCC=VCC;GND=GND VCC=VCC;GND=GND DIS_DATAP 5

R5
1K
0.125W
1%

C C

VCC

C13 C14
0.1U 0.1U
50V 50V
10% 10%

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Tue Mar 08 17:10:30 2005 ftwsremotecontrol 5 of 6 2xxxxxxSCH A 5 of 6


1 2 3 4 5 6 7 8
Figure 4-18 2xxxxxxSCH-5

Section 5.0 Remote Control Transfer Page 84


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

BUS_FPGA<7..0>

COL_REMOTE_OUT<11..0>

A VCC
A

R10
U3 U3 1K
74HCT04 74HCT04 0.125W
1% VCC C17 COL_ENDISN 0
0 COL_ENDIS 1 2 3 4
A Y A Y
R11
0.1U 200
VCC=VCC;GND=GND VCC=VCC;GND=GND 50V 0.125W
10% 1%
COL_ENDISP 1

U5
DS34C87T COL_ENKEYN 2
U3 U3
74HCT04 74HCT04
4 A_B_CNTRL R12
1 COL_ENKEY 5 6 9 8 1 INA INVA 2 200
A Y A Y
NON_INVA 3 0.125W
7 INB INVB 6 1%
VCC=VCC;GND=GND VCC=VCC;GND=GND NON_INVB 5 COL_ENKEYP 3
12 C_D_CNTRL
9 INC INVC 10
B NON_INVC 11 B
15 IND INVD 14 COL_CLKN 4
U3 U3 NON_INVD 13
74HCT04 74HCT04 R13
200
2 COL_CLK 13 12 11 10 VCC=VCC;GND=GND 0.125W
A Y A Y 1%
COL_CLKP 5
VCC=VCC;GND=GND VCC=VCC;GND=GND

COL_SON 8
R14
200
VCC 0.125W
U4 U4
74HCT04 74HCT04 C18 1% COL_SOP 9

4 COL_SO 1 2 3 4
A Y A Y 0.1U COL_SO2N 10
50V
10%
VCC=VCC;GND=GND VCC=VCC;GND=GND
R15
U6 200
MAX488E 0.125W
C 1% C
U4 U4 COL_SO2P 11
74HCT04 74HCT04 1 VCC
Y 5
3 COL_SI 6 5 8 9 3 DI COL_SIN 6
Y A Y A
Z 6
8 A
VCC=VCC;GND=GND VCC=VCC;GND=GND RO 2 R16
7 B 200
GND 4 0.125W
1%

U4 U4 COL_SIP 7
74HCT04 74HCT04

13 12 11 10
A Y A Y

VCC=VCC;GND=GND VCC=VCC;GND=GND
VCC

C15 C16
0.1U 0.1U
50V 50V
10% 10% R9
1K
D 0.125W
1%
D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Tue Mar 08 17:10:30 2005 ftwsremotecontrol 6 of 6 2xxxxxxSCH A 6 of 6


1 2 3 4 5 6 7 8
Figure 4-19 2xxxxxxSCH-6

Chapter 4 - Wall Stand Page 85


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 6.0
Grid Sensor Board, 5145484
1 2 3 4 5 6 7 8

A A

REVISION ECR/ECO DATE SIGN OFF


1 12-OCT-05 JIANG MING

MH2 MH1
1

B S1 B
J

S2
1
2
3
4
S3 5
6

S4
NOTES:
1. THE VARIATION AND VERSION FOR THIS
PWA IS 5145484 .
2. THIS COVER SHEET CONTROLS THE RE-
VISION STATUS OF THE ENTIRE DOCUMENT.
THE REVISION OF ALL SHEETS OF THIS
DOCUMENT WILL BE AT THE SAME LEVEL.
C 3. THIS SCHEMATIC MEETS THE SPECIFICA- C
TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
4. ELECTRONIC SIGNATURES FOR BOTH THE
SCHEMATIC REVIEWER(S) AND APPROVER(S)
CAN BE FOUND WITHIN E-PDM.
5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS.
RESISTORS ARE IN OHMS.
INDUCTORS ARE IN HENRIES

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Fri Oct 14 18:35:50 2005 grid_sensor 1 of 1 5145484SCH 1 1 of 1

1 2 3 4 5 6 7 8

Figure 4-20 5145484SCH

Section 6.0 Grid Sensor Board, 5145484 Page 86


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 4-21 5145484ADW

Chapter 4 - Wall Stand Page 87


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 7.0
Display Board, 5140868
1 2 3 4 5 6 7 8

A2

B CA N3 N4 N5
A N
M C
K P A1 A A1 A A1 A
A H U F B F B F B A
G D G C G C G C
T R E DP E DP E DP
VCC U2 F S D A2 D A2 D A2
3 VDD E DP
VCC
12 DATA_IN DATA_OUT 18

11 CLOCK A 7
C1

C2

C3

C4
B 6
10 ENABLE_N C 5
2

D 4
2
BL03RN2R1

BL03RN2R1

E
F 1
G 20
H 19
MC14489B

BANK1 9

2
FL1
FL2 BANK2 13 LED6 LED5 LED4
BANK3 15
1

BANK4 16
BANK5 17

1
RX 8
VSS 14

R2
B B
A1
12 EN_N
FL3 4 EN
BLM31P601SG U3 B CA N1 N2
1 2 1 INN A N
2 INP O 3 M C
FL4 R3 K P A1 A A1 A
1 2 H U F B F B
G D G C G C
BLM31P601SG T R E DP E DP
VCC U1 F S D A2 D A2
J1 1 1 12 EN_N 3 VDD E DP
6 6 2 2 FL5 4 EN
7 7 3 3 BLM31P601SG U3 12 DATA_IN DATA_OUT 18
8 8 4 4 1 2 15 INN
9 9 5 5 14 INP O 13 11 CLOCK A 7
FL6 R4 B 6
G1 G1 1 2 10 ENABLE_N C 5
G2 G2 D 4
BLM31P601SG E 2
1
C 12 EN_N
F
G 20 C
FL7 4 EN H 19
MC14489B
BLM31P601SG U3

2
1 2 7 INN BANK1 9 LED1 LED2 LED3
6 INP O 5 BANK2 13
FL8 R5 BANK3 15
1 2 BANK4 16
BANK5 17

1
BLM31P601SG
RX 8
VSS 14
NOTES:
R1

1. THE VARIATION AND VERSION FOR THIS


PWA IS 5140868.
2. THIS COVER SHEET CONTROLS THE RE-
VISION STATUS OF THE ENTIRE DOCUMENT.
MH1 MH2 MH3 MH4
THE REVISION OF ALL SHEETS OF THIS
REVISION ECR/ECO DATE SIGN OFF
DOCUMENT WILL BE AT THE SAME LEVEL.

1
3. THIS SCHEMATIC MEETS THE SPECIFICA- 1 12-OCT-05 JIANG MING
TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
D 4. ELECTRONIC SIGNATURES FOR BOTH THE
D
SCHEMATIC REVIEWER(S) AND APPROVER(S)

REV 2 01July04
CAN BE FOUND WITHIN E-PDM.
5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS. GE Healthcare
RESISTORS ARE IN OHMS. Company Restricted

B Size
INDUCTORS ARE IN HENRIES Block Path Block Sheet Dwg No Rev Sheet

Tue Oct 18 14:05:08 2005 display 1 of 1 5140868SCH 1 1 of 1

1 2 3 4 5 6 7 8

Figure 4-22 5140868SCH

Section 7.0 Display Board, 5140868 Page 88


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 4-23 5140868ADW1

Chapter 4 - Wall Stand Page 89


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 4-24 5140868ADW2

Section 7.0 Display Board, 5140868 Page 90


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 8.0
Column Button, 5140871

1 2 3 4 5 6 7 8

A A

NOTES:
1. THE VARIATION AND VERSION FOR THIS
PWA IS 5140871. REVISION ECR/ECO DATE SIGN OFF
2. THIS COVER SHEET CONTROLS THE RE-
1 12-OCT-05 JIANG MING
VISION STATUS OF THE ENTIRE DOCUMENT.
B THE REVISION OF ALL SHEETS OF THIS B
DOCUMENT WILL BE AT THE SAME LEVEL.
3. THIS SCHEMATIC MEETS THE SPECIFICA-
TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
4. ELECTRONIC SIGNATURES FOR BOTH THE
SCHEMATIC REVIEWER(S) AND APPROVER(S)
CAN BE FOUND WITHIN E-PDM.
5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS.
RESISTORS ARE IN OHMS.
INDUCTORS ARE IN HENRIES

C C

D D

REV 2 01July04
Made By Date Title
GE Healthcare
Jiang Ming 12-Oct-2005 col_button Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Tue Oct 18 16:05:13 2005 col_button 1 of 2 5140871SCH 1 1 of 1

1 2 3 4 5 6 7 8

Figure 4-25 5140871SCH1

Chapter 4 - Wall Stand Page 91


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

MH1 MH2 MH3

VCC
U3
MH4 MH5 MH6 VCC
Y
DI R21
Z VCC
A A A
RO
B
GND
VCC

C2

C3

C4

C5
C1
O2

O3

O4
O1

3
4

4
VCC
I1

I2

I3

I4

12V S1 S2 S3 S4 S5
U2
VDD C11 C12 C13 C14 C15

2
1

1
VPWR

12V FL1 SI SB1


J1 J2 SO SB2
1 1 SCLK
9 2 2 9 INTB SP1
10 3 3 10 CSB SP2
11 4 4 11 VBG SP3
12 5 5 12 R2 SP4
13 6 6 13 VCC
B 14 7 7 14 RSTB SG1 B
15 8 8 15 SG2
MASL SG3
G1 VCC FL2 G1 SYNC SG4 C16 C17 C18 C19
G2 G2 SG5
VSS SG6
MC33884

2
1

1
VCC S6 S7 S8 S9
FL3

J3 1

3
4

4
R3 R4
1 2
2 3
3 4 J4
C31 C32 C33 5

C C
BLM31P601SG

VCC U1
VDD

DATA_IN DATA_OUT
FL4 FL5 FL6 FL7 FL8 FL9 FL10 FL11
CLOCK A
B
ENABLE_N C
D
E
F
R23 R24 R25 R22 G DS11 DS12 DS13 DS14
MC14489B H DS21 DS22 DS23 DS24
DS31 DS32 DS33
VCC BANK1
BANK2
BANK3
BANK4
BANK5
EN

EN

EN

EN
INN

INN

INN

INN
EN_N

EN_N

EN_N

EN_N
INP

INP

INP

INP

RX
VSS

D U4 U4 U4 U4
D

R1
O

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Tue Oct 18 16:05:08 2005 col_button 2 of 2 5140871SCH 1 2 of 2

1 2 3 4 5 6 7 8

Figure 4-26 5140871SCH2

Section 8.0 Column Button, 5140871 Page 92


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 4-27 5140871ADW1

Chapter 4 - Wall Stand Page 93


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 4-28 5140871ADW2

Section 8.0 Column Button, 5140871 Page 94


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 9.0
Wall Stand Schematics, 5136848
1 2 3 4 5 6 7 8
A2 Internal A4 Blk 1 GE Company Confidential Internal A4 Blk2 To Shuttle A2
To Shuttle A2

1 1 1 7
5V 1 Tilting
15 15 15 35
GND IN1-1 Drive+ A2 A

J12
J4
3 3 3 9 2 Driving
A CLK- D1

Motor Control Board


11 11 11 8 3 Drive- M2
CLK+ Advanced Logic Board FB1 Motor

eiTian all tand


Tilting 2 34 4

J22
2 2 Data- IN2-2
A2 5161462

J3
Load 10 10 10 33 5
Data+ FS-N2
A1 B2 Vertical

51360
E2 4 4 4 3 DC24GND 6

A4 B1
Encoder A-
12 4 7 Drive+ A5 Driving

J11
12 12 A+ DC24GND

J3
5 5 5 5 8 M1
Drive- Motor

J13
DC24-MTR

J2
B- 9
13 13 13 6 IN1-2
B+
10

Buckhead
To Main
A1F2 A1J4 A7E1
FS-N1
11
IN2-1
1 1 1 12
AC-L
B
B Vertical Up D2
Handle Assembly

13 DC2 24V

J10
2 2

J1
2 Vertical Down FB2 5146712 AC-N
3 3 14 DC2 24GND
3 Tilting Up DC24GND A4 P1
15 Earth
4 4 Tilting Down 4 DC24-MTR
J1

5 5 18
GND Earth

port
To Earth
6 6 14 Motor
DC12V
7 7 5

J3
9
Tilting Senor
19
Power Supply
9 GND 22 5V 1 1 1 1
32 GND 15 15 15 15
48 3 3 3 3
1 1 1 7 CLK- C
To Main Buckhead

Foot Up 49 11 11 11 11
C J3 3 2 2
Foot Down
8
23
CLK+
DATA- 2 2 2 2
J7

5 5 5 20 Encoder

J22
GND 24 10 10 10 10 A5

J2
DATA+
29 ENA- 5 5 5 5 E1
28 ENA+ 13 13 13 13 Vertical
2 2 2 31 6 6 6 6
3 3 CAN_L 3 ENB- Load
30 14 14 14 14

Sys CAN
CAN_GND ENB+ Encoder
4 4 4 25 4 4 4 4

J5
CAN_L_Lbpk
J5

J2 7 7 7 VZero
CAN_H
8 8 CAN_H_lbpk 8
5 5 CAN Shield 5 8 3 3 Vertical D
D 6
DC1 24V
Clutch
1 1 A5 C1 Clutch
7 DC1 24V 2 2 Vertical

J9
2 2 2

J8
CAN_L 3 Break1 8 8 A5 B1
3 3 3 Brake
Detector 6 6

Sys CAN
CAN_GND 1 DC1 24V Vertical
4 4 4 7 7 A5 B2
Power CAN_L_Lbpk J6 2 Break2
J4

J4 7 7
CAN_H
7 Brake2
Supply 8 8 CAN_H_lbpk 8 (reserve)
A6 P3 5 5 CAN Shield 5
Buckhead
To Main

5 1 1
A1F1 A1J4 A7E1

FAN1+ 6 A2 F1 FAN1
1 6 E

J13
J9
FAN1-
2
E AC-L 2 2 GND 1 Power
6
4
FAN2+ 2,3
7,8 7 A2 F2 FAN2
1 1 2 J1 FAN2-
3 3
DC1 24V
5
Input
AC-N Shield Earth
A4 P1
5161462 A1 B3
Transfer I/F Board
J8
port
To Earth

Earth 2
COL-EN1
15
COL-EN2
COL-CLK 3 Area Designations
Control COL-SO 4
Power Supply COL-SI 1 6 A1 - WS Main Bulkhead
F DIS-EN 5
A2 - WS Shuttle Assembly F
DIS-CLK 18
J1

6 A3 - WS Detector Housing Assembly


To Detector Housing

1 1 DIS-SO
DC12V DC5V 8,9,20,21 A4 - WS EEBox Assembly
2 2 1,7,10,19
SI- GND A5 - WS Top Assembly
3 SI+ 3 11,23
6 6 DC12V A6 - WS Detector Bulkhead
DC5V GND 12,22
7 7 13,25 A7 - WS GND port
EN-LED- DC24V
8 8 14,24
A3B5

EN-LED+ GND
J3

9 9
J6

SO-
10 10
G SO+ G

A5 D1
11 11 1 1
CLK- DC5V
12 CLK+ 12 2 2
13 13 DATA- 3
EN-KEY- 3
14 14 DATA+
EN-KEY+ 4 4
CLK-

J14
15 15 5 5

J1
GND CLK+
J2

EN- 6 6
EN+ 7 7
GND 8 8
NOTES GND 9 9
1. All sheets of this schematic will be at the same revision. The
H revision of this sheet will be the revision of the entire document.
REVISIONS
MADE
APPROVALS YY/MM/DD TITLE
REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian Wallstand Schematics
2. = Indicates signals are part of a cable
1 Ye Bin 10 Oct. 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 1
APPR
GEHL MEDICAL 5136848SCH SHT/CONT ON
- - OTHER SYSTEMS 1 /5

1 2 3 4 5 6 7 8
Figure 4-29 5136848SCH_rev1 (1)

Chapter 4 - Wall Stand Page 95


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A1 Blk GE Company Confidential


Area Designations

To Detector Housing

To Shuttle
A A1 - WS Main Bulkhead 1 GND
1 A
A2 - WS Shuttle Assembly 2 Middle Area Select 2

A3 IonCham
IonChamber
3 Left Area Select 3
A3 - WS Detector Housing Assembly
4 AEC1 4
A4 - WS EEBox Assembly I oC h a m b e r 5 V_Out 5
A5 - WS Top Assembly J1
6 Right_Area Select 6
7

FG
A6 - WS Detector Bulkhead 7 -15 VDC Rail
8 +15 VDC Rail 8
A7 - WS GND port
9 GND 9

FG
B B

To EEBox Assembly
FG
2 2

A4 Blk1 J1
3 3 CAN_L
CAN_GND
4 4
CA N p o r t J2 7 7
CAN_L_Lbpk FG
CAN_H -S Switch Power
8 8 CAN_H_lbpk
5 5 - Supply
CAN Shield
+ A4 P2

A4 Blk1 J5
1 +S
Foot Up
C Fo ot p e d a l J3
3
Foot Down AC- AC- C
5 L N
GND

AC-N

Earth port
AC-L
AC-L
Filter Fuse A1 F1
WS P o w e r J4
A1 FT1 250V 5A FG
-S Switch Power
D AC-L D

A1 E1
- Supply
Fuse A1 F2
AC-L + A4 P1
WS 250V 5A
Filter +S
D et e c t o r J5
A1 FT2 AC- AC-
Power
Fuse A1 F3 L N
250V 5A
AC-L
AC-N
E E
Earth

Earth

FG
F Ethernet Port F
J2 Can FG
Detector

DC Port
A3 D1 Detector Power

output
J5 DC
Supply
A6 P3

AC- AC-
L N
G AC-L G
AC-N
NOTES
1. All sheets of this schematic will be at the same revision. The
revision of this sheet will be the revision of the entire document.

2. = Indicates signals are part of a cable

H REVISIONS
MADE
APPROVALS YY/MM/DD TITLE
REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian Wallstand Schematics
1 Ye Bin 10 Oct. 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 1
APPR
GEHL MEDICAL 5136848SCH SHT/CONT ON
- - OTHER SYSTEMS 2 /5

1 2 3 4 5 6 7 8

Figure 4-30 5136848SCH_rev1 (2)

Section 9.0 Wall Stand Schematics, 5136848 Page 96


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

Area Designations

A A1 - WS Main Bulkhead A

A3B5
Collimator Remote Control Board 1
A2 - WS Shuttle Assembly 1

A3 B6
Collimator Remote Control Board 2
To EEBo
DC12V
A3 - WS Detector Housing Assembly 2
SI-
3 SI+
A4 - WS EEBox Assembly 6 DC5V
A5 - WS Top Assembly 7
EN-LED-

A4 Blk1 J6
A6 - WS Detector Bulkhead 8

Assembly
EN-LED+
A7 - WS GND port 9 SO-
J1 10 J1
SO+
11
CLK-
12
B 13
CLK+ B
EN-KEY-
14 EN-KEY+
15
GND

C C
1 1
DC12V
2 SI- 4
3 SI+ 5
4 SI- 2
5 SI+ 3
6 DC5V 6
7 7
EN-LED-
8 EN-LED+ 8 J2
J2 9 SO- 9
10
D 11
SO+ 10
11
D
CLK-
12 CLK+ 12
13 EN-KEY- 13
14 EN-KEY+ 14
15 GND 15

J3 J4 J3 J4
E E

S 2 S 3 S 4 GND
S1

GND D1-1
F D1-2 F

J4
NC
C
NO

Detector
G in G
GRID Sensor Board Housing
Sensor
A3 B7 A3 D1

NOTES
1. All sheets of this schematic will be at the same revision. The
revision of this sheet will be the revision of the entire document.

2. = Indicates signals are part of a cable


H REVISIONS
MADE
APPROVALS YY/MM/DD TITLE REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian Wallstand Schematics
1 Ye Bin 10 Oct. 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 1
APPR
GEHL MEDICAL 5136848SCH SHT/CONT ON
- - OTHER SYSTEMS 3 /5

1 2 3 4 5 6 7 8
Figure 4-31 5136848SCH_rev1 (3)

Chapter 4 - Wall Stand Page 97


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

GE Company Confidential

A WS Grounded A
B
Fuse3
Detector
Filter
110Vac Detector
Power Supply

B WS Collum B
Cover not
A Grounded
DC24V Vertical
Fuse2 Motor DC Motor
Filter

Motor
110Vac
Power
DC24V Control
Tilting
Board DC24V
DC Motor
C C

Fuse1 DC24V Vertical


Control
Encoder
Power

Tilting

D C Encoder
D
ALB
Detector in House Sensor Vertical Brake
Board
Vertical Clutch

Collimator Remote FAN1


E Control Board 2 FAN2
E
Lamp and FOV Button

Vert Clutch

GRID Sensor Fei Tian WS


F Board WS
WS Insulation diagram F
Foot
Handle
Pedal

Lamp and FOV Button

Collimator Remote Transfer I/F Board Display Board


Control Board 1
G G
Detector Housing Cover
not Grounded NOTES
1. All sheets of this schematic will be at the same revision. The
revision of this sheet will be the revision of the entire document.

2. = Indicates signals are part of a cable

H REVISIONS
MADE
APPROVALS YY/MM/DD TITLE
REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian WS Schematics
1 Ye Bin 6 Sept 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 2
APPR

OTHER
GEHL MEDICAL SYSTEMS 5136848SCH SHT/CONT ON
- - 4 /5

1 2 3 4 5 6 7 8

Figure 4-32 5136848SCH_rev1 (4)

Section 9.0 Wall Stand Schematics, 5136848 Page 98


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A GE Company Confidential A

B B

C C

D D

晻  
E E
晻  0 .1晻  

Any accessible dead mental part


on WS, which is protectively
F earthed
F

End of the grounding


cable for WS

G G
NOTES
1. All sheets of this schematic will be at the same revision. The
revision of this sheet will be the revi sion of the entire document.

2. = Indicates signals are part of a cable

T IT L
H REVISIONS
MADE
APPROVALS YY/MM/DD E
REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian WS Schematics
1 Ye Bin 6 Sept 2005 Initial Design
CHECK
Fei Tian
FIRST MADE FOR: 2
APPR

OTHER
GEHL MEDICAL SYSTEMS 5136848SCH SHT/CONT
ON
- - 5 /5

1 2 3 4 5 6 7 8

Figure 4-33 5136848SCH_rev1 (5)

Chapter 4 - Wall Stand Page 99


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

GE Company Confidential
A Reset A
FPGA Logic Clock builder CLK-16M
CLK-10M
Display board
Display Board LED control Module
(SPI)

Digital Input Register Array


B B

DI
DI sample Module
Tilting
positioning
arithmetic
C Module C

Collimator
Remote
Control
Module
(SPI) ARM Bus Interface
D D

Vertical
positioning
arithmetic
DO Module
DO output Module

E E

Digital Output Register Array

Voltage Diagnostic A/D Control Module Analog Input Register Array


F F
Tilting_CLK BISS Encoder Control
Tilting_DATA Module
Encoder Input Register Array
Vertical_CLK SSI Encoder Control
Vertical_DATA Module

G TX
G
CAN Hub RX
Reckon speed based on
Encoder data

NOTES
1. All sheets of this schematic will be at the same revision. The
H revision of this sheet will be the revision of the entire document.
REVISIONS
MADE
APPROVALS YY/MM/DD TITLE
REV H
REV MADE BY DATE ECR NO. DESCRIPTION Fei Tian WS Schematics
2. = Indicates signals are part of a cable
1 Ye Bin Oct. 20 2005 Initial Design
CHECK
FIRST MADE FOR: Fei Tian 1
APPR
GEHL MEDICAL 5161462SCH SHT/CONT ON
- - OTHER SYSTEMS 1 /1

1 2 3 4 5 6 7 8

Figure 4-34 5136848SCH_rev1 (6)

Section 9.0 Wall Stand Schematics, 5136848 Page 100


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 10.0
Ion Chamber, 2210104SCH

Figure 4-35 2210104SCH_r0

Chapter 4 - Wall Stand Page 101


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 11.0
Cable Routing
Wallstand cable list
NO. Part No. Description Quantity Screw Screw (new)
(include assembly 5146709,5147169,5147171.This is
5136368 Feitian Wallstand Cable Assembly all wallstand cable assembly)
1 5136791 Feitian Wallstand EEBox DisplayBD Cable 1 2 2
2 5136792 Feitian Wallstand EEBox Collimater Remote Control Output Cablel 1 2 2
3 5136794 Feitian Wallstand EEBox Interface To RemoteControlTransmit BD Cable 1 0 4
4 5136800 Feitian Wallstand EEBox CANOpen Cable 1 2 2
Feitian Wallstand EEBox DetectorPowerSupply CANOpen AND
5 5136801 1 3 2
Handle_TiltingSensor AND Footswitch AND FAN Cable
6 5136810 Feitian Wallstand EEBox 24VDC1 Cable 1
7 5136811 Feitian Wallstand EEBox 24VDC2 Cable 1
8 5136812 Feitian Wallstand EEBox Vertical AND Tilting Position Encoder Cable 1 3 2
9 5136814 Feitian Wallstand InterfaceToMotorControl Cable 1 2 4
10 5136815 Feitian Wallstand EEBox Vertical Motor cable 1
11 5136816 Feitian Wallstand EEBox Tilting Motor Cable 1
12 5136817 Feitian Wallstand EEBox Vertical Break_Clutch Cable 1 2 2
13 5141880 Feitian Wallstand Fan Control Cable in Detector Housing inside EEBox 1 1
14 5141874 Feitian Wallstand Detector Power Supply CANOpen Cable inside EEBox 1 2 2
15 5136369 Feitian Wallstand PowerSupply Input Cable 1
16 5136370 Feitian Wallstand 24VDC1 Cable1 1
17 5136371 Feitian Wallstand 24VDC2 Cable 1
18 5136372 Feitian Wallstand CANOpen Cable 1 1
19 5136373 Feitian Wallstand Detector CANOpen Cable 1
20 5136374 Feitian Wallstand Display BD Cable 1
21 5136375 Feitian Wallstand Collimater Control_1 Cable 1
22 5136376 Feitian Wallstand Collimater Control_2 Cable 1
23 5136380 Feitian Wallstand Footswitch Cable 1 1
24 5136381 Feitian Wallstand Handle_TiltingSensor Cable 1
25 5136382 Feitian Wallstand Handle_TiltingSensor_2 Cable 1
26 5136383 Feitian Wallstand Tilting OptSensor Cable 1 sensor supplied by supplier
27 5136385 Feitian Wallstand Tilting Position Encoder Cable 1
28 5136386 Feitian Wallstand Vertical Motor Cable 1
29 5136387 Feitian Wallstand Vertical Break_Clutch Cable 1
30 5136388 Feitian Wallstand Tilting Motor Cable 1
31 5136389 Feitian Wallstand Fan Control Cable 1
32 5136390 Feitian Wallstand Detector PowerSupply DCConvert Cable 1
33 5136391 Feitian Wallstand Ion Chanber Cable 1 1
34 5136427 Feitian Wallstand GND_ARM_House Cable 1
35 5136428 Feitian Wallstand GND_Buckhead_EEBox Cable 1
36 5136429 Feitian Wallstand GND_Buckhead_ARM Cable 1
37 5136503 Feitian Wallstand Motor Drive Socket Cable 1
38 5136503-1 Feitian Wallstand motor socket cable 1
39 5136895 Feitian Wallstand GRID Sensor Cable 1
40 5136919 Feitian Wallstand Detector Power Supply Input Cable 1

Section 11.0 Cable Routing Page 102


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

41 5137111 Feitian Wallstand Tilting Encoder Socket Cable 1 Encoder Pin supplied by supplier
42 5147939 FeiTian Wallstand Detector Power supply GND Cable 1
43 5147940 FeiTian Wallstand Detector Power supply Buck head Cable 1
44 5161387 Wallstand Vertical Encoder Connecter Cable
45 5161171 Cable screw 26
46 5165211 Feitian cable screw AN03321 22
ASSEMBLY 1 5146709 FeiTian Wallstand EEBox Cable
5136791 Feitian Wallstand EEBox DisplayBD Cable 1 2
ASSEMBLY 2 5147169 FeiTian Wallstand Cable in orbit
5136388 Feitian Wallstand Tilting Motor Cable 1
5136389 Feitian Wallstand Fan Control Cable 1
5136381 Feitian Wallstand Handle_TiltingSensor Cable 1
5136385 Feitian Wallstand Tilting Position Encoder Cable 1
5136375 Feitian Wallstand Collimater Control_1 Cable 1
5136429 Feitian Wallstand GND_Buckhead_ARM Cable 1
5136391 Feitian Wallstand Ion Chanber Cable 1 1
5161171 Cable screw 1 27
ASSEMBLY 3 5147171 Feitian Wallstand Inside cable except EEBox and ARM

Chapter 4 - Wall Stand Page 103


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A5

Vertical Motor Display Board Hight Encoder


A Rated Power: 25 Watt
Brake
Two line LED
A
18 Watt (5161024)
Rated Voltage: 24VDC Three LED Light
Rated Speed: 3000r/min Clutch
10 Watt

2
15
A4
Power 110V
Supply 2 24V
JSW300A-24

GND
B B
30 28 29 31 20 43 17
35

11 10 12 13 1 8 7
A4 44(2) 44(2) 44(2) 44(2)

(51
J4 J1
368
J3 10) WS MC BD 2 J2
45(4) 9
Tilt Up
C Tilt Down J1
C
J9 45(2) 3
Vert U p Handle Switch
Vert Down VS10N051C2 J8 45(2) 2
Handle
Handle J22 45(2) WS ALB BD
Box1 9DB-M 2 2 45(2)
2
25 J13 J5
J3
45(2)
J14 2 45(2) J6
Tilt Up
Tilt Down 3 45(4)
Handle Switch 2 45(2) J3 2
Vert U p 2
VS10N051C2 J1 WS Transfer BD J2
45(2)
D Vert Down D
Handle
Handle
Box2
44(2) 44(2) 44(2) 44(2) 44(2) 44(2)
5 8 14 4 2 5 6
Extend DI
26 GND
24 27 19 18 21 23 16
44(2) 44(2)
Tilting Limit Switch Power Supply 1
A4
120V 0V HK150A-24
EE-SX771 0V24V 24V
110V

E A2
E
15
A3
34 Shuttle 36
36
GND
Detector 2 Fan *2
Housing 10 Watt
GND

Collimator Control 15DB-F 15DB-F Collimator Control


Panel (left) 9 Buttons Panel(right) 9 Buttons 44(2) J1
10 LED Light 9DB-M 22 10 LED Light 9DB-M Tilting Motor 37 A1 J3
Buck Head 2 J2
Rated Power: 25 Watt
5P 5P Rated Voltage: 24VDC 15 J4
38
40 39
Tilting 27 J5
6P
F Encoder FUSE1
FUSE2
F
GRID Sensor Detector In House FUSE3
Sensor

33
Ion Chamber
A7 GND
Detector
A6

5145500-6 2
Buck Head
G GND
41 G
32
25DB-M 9DB-F

Load 0V GND 42
Detector
120V 0V
Power Supply

9DB-M 40

H REVISIONS APPROVALS YY/MM/


DD
TITLE
REV H
REV MADE BY DATE ECR/ECO NO. DESCRIPTION
MADE
Feitian Wallstand cable Routing
2 6 Dec Add cable screw5165211
CHECK 3
Li Dantong 2005 ECR 2023874 FIRST MADE FOR: Feitian
3 18 Apr Add cable screw5165211
APPR
Li Dantong ECR 2028228
2006 GEHL MEDICAL 5136368ADW
SHT/CONT
ON
OTHER SYSTEMS

1 2 3 4 5 6 7 8

Figure 4-36 FeiTian Wallstand Cable Routing (1)

Section 11.0 Cable Routing Page 104


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A A

B B
Wallstand
Wallstand Detector Power
Bulkhead supply Bulkhead

J1

5146500-4 Ion Chamber 5136381


9DB-F 9DB-M

C C
J3 3

Foot 5136380
Switch 9DB-F
9DB-M

J2 3

5146500-1 CANOpen 5136372


9DB-M
9DB-F
D D
J4

5146500-2 EMC 5136369


Filter

L-COM
J5 RJ45 RJ45 5146500-6

5146500-3 5136919
E EMC
Filter
E

AMP14PIN

Detector Power Cable DB25-M

1 FUSE1 2
1 FUSE2 2
F 1 FUSE3 2 F
GND

G G

H REVISIONS
MADE
APPROVALS YY/MM/
DD
TITLE
REV H
REV MADE BY DATE ECR/ECO NO. DESCRIPTION Feitian Wallstand cable Routing
2 Li Dantong 31 Oct 2005 ECR 2023874
CHECK
FIRST MADE FOR: Feitian
3
Change the position APPR
3 Li Dantong 18 Apr 2006 ECR 2028228
of J2 and J3 GEHL MEDICAL SHT/CONT
5136368ADW ON
OTHER SYSTEMS

1 2 3 4 5 6 7 8

Figure 4-37 FeiTian Wallstand Cable Routing (2)

Chapter 4 - Wall Stand Page 105


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

This page is intentionally left blank.

Section 11.0 Cable Routing Page 106


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Chapter 5 - System Cabinet


A25 BULKHEAD
Thunder System Cabinet SCHEMATICS – 5128145SCH
A8 A2
+24VDC Power 50DB-F J109
Supply RCIM

CABINET FANS
9DB-M J110
TABLE
50SCS1 RT BUS

9DB-F J108
JEDI CAN
3MnL-P
'%)
-
276&$1

'%)
- /223%$&.
:6&$1 &211(&725

'%) -
7$%/( /223%$&.&211(&725
&$1 )25',$*1267,&621/<

'%)
-
&$10$*,&

9DB-F

9DB-M

9DB-M SERIAL A
TABLE CHILLER
15DB-F
9DB-M J103
CONDITIONER STATUS TBL
9DB-M

9DB-F SERIAL B
WALL STAND CHILLER
9DB-M 9DB-M 9DB-M 9DB-M
9DB-F J203
CONDITIONER STATUS WS

25DB-F 3MnL-P -
&21752/522032:(5

3MnL-P -
:632:(5
J1
A7 PDU CORONA
J1 PWR CTL
J8 6MnL-P 3MnL-P -
Positioner J29
120VAC Out IDC2 PWR 7$%/('(732:(5

J12 J2 3MnL-P 5MnL-P J16 J28 -


Wall Stand MAGIC 5MnL-P 3MnL-P
120VAC In (Future) :6'(7
PC PWR
(switched) 5MnL-P 32:(5
J13 J23
J3 3MnL-P WS PWR 5MnL-P
J14 IDC1 -
J25 3MnL-P
Cab. Fans TABLE DET 27632:(5
5MnL-P
J4 PWR
5MnL-P J15 J26
9DB-F J13 WS DET 5MnL-P 6MnL-P -
Logic Bd. PWR 7$%/(32:(5
J5
J27 5MnL-P
3MnL-P J9
5MnL-P J12 J22 A9 JEDI
Pwr. Supply
J6 OTS PWR 5MnL-P 9DB-M 9DB-F J81
AEC1
TABLE ION CHAMBER
5MnL-P J11 J21
3MnL-P J10 J7 TABLE PWR 8MnL-P
Magic PC / 3MnL-P COMM 9DB-M 9DB-F
Jedi Aux 1 AEC2 J14
ICD2 5MnL-P J14 J24 15DB-F WALL STAND ION CHAMBER
IDC1 PWR
J11
Jedi Aux 2 TB2 A10 HALO RELAY J?
BOARD

230VAC J?
Service Outlet
J2
120VAC A11 FUSE
Service Outlet BLOCK ?PNX-S J42
J?
STATOR1
TB1 Input Power
Stator Tube 1
?PNX-S 12MnL-C
J?
Mains 125A / 600V
3 Phase Power
Cathode From
Hospital Mains Switch
OTS
125A / 600V Anode

5128145SCH Term
M.MEJIA Block
125A / 600V
Rev 1 Protective Earth
20JUN2005
Check if IEC requires a secondary restraint on the input mains

Figure 5-1 5128145SCH_s1_r1

Chapter 5 - System Cabinet Page 107


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 1.0 — 5KV Power Distribution Unit, 2293487

Figure 5-2 2293487SCH_s1_r2

Section 1.0 — 5KV Power Distribution Unit, 2293487 Page 108


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-3 2293487SCH_s2_r2

Chapter 5 - System Cabinet Page 109


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 2.0 — "Halo" Relay Board, 2407245

Figure 5-4 2407245ADW_s0001_r2

Section 2.0 — "Halo" Relay Board, 2407245 Page 110


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

REVISION ECR/ECO DATE SIGN OFF


A 20-FEB-04 SWQ
A B 10-MAR-04 SWQ A
C 24-MAR-04 SWQ NOTES:
1. THE VARIATION AND VERSION FOR THIS
1 13-MAY-04 BMW PWA IS 2407245 2.
2 2005760 23-AUG-04 BMW 2. THIS COVER SHEET CONTROLS THE RE-
VISION STATUS OF THE ENTIRE DOCUMENT.
THE REVISION OF ALL SHEETS OF THIS
DOCUMENT WILL BE AT THE SAME LEVEL.
3. THIS SCHEMATIC MEETS THE SPECIFICA-
TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
4. ELECTRONIC SIGNATURES FOR BOTH THE
SCHEMATIC REVIEWER(S) AND APPROVER(S)
CAN BE FOUND WITHIN E-PDM.
5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS.
RESISTORS ARE IN OHMS.
B INDUCTORS ARE IN HENRIES B

C C

D D

REV 2 1July04
Made By Date Title
GE Healthcare
SONG WEN QING 24-Mar-04 HALO RELAY BOARD Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Fri Aug 27 11:51:22 2004 pdu_relay_board 1 of 5 2407245SCH 2 1 of 5

1 2 3 4 5 6 7 8

Figure 5-5 2407245SCH_s0001_r2.

Chapter 5 - System Cabinet Page 111


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

4A2<

4A2< 3A5> TABLE_PWR PDU_J1_OUT230V_P 3B5< 4D2< 3D5> TABLE_DETECTOR_PWR


4A5< 4B5< 4D2<
J25
J21 J11
A A
1 1 1 1 1 1
2 2 2 2 2 2
TABLE PWR 3 3 3 3 3 3
4 4 PDU_J1_OUT230V_N 4 4 PDU J1 OUT230V TABLE DETECTOR PWR 4 4
5 5 5 5 5 5
6 6 6 6
7 7 PDU_J1_OUT230V_G 7 7
8 8 8 8 PIN
STRAIGHT
MATE_N_LOK_WHITE PDU_J8_OUT_P 3B8< 3D5<
PIN PIN
STRAIGHT STRAIGHT 4A5< 3A8> WS_DETECTOR_PWR J15
MATE_N_LOK MATE_N_LOK
J26
4B2< 1 1
2 2
1 1 3 3
4B2< 3B5> OTS_PWR PDU_J1_OUT120V_P 3B5< 2 2 4 4
3 3 PDU_J8_OUT_N 5 5 PDU J8 OUT
J22 J12 WS DETECTOR PWR 4 4
5 5 PDU_J8_OUT_G
PIN
1 1 1 1 STRAIGHT
2 2 2 2 PIN MATE_N_LOK_WHITE
OTS PWR 3 3 PDU_J1_OUT120V_N 3 3 PDU J1 OUT120V STRAIGHT
B 4 4 4 4 MATE_N_LOK_WHITE B
5 5 PDU_J1_OUT120V_G 5 5
4B5< 3B8> PDU_K7_PWR

PIN PIN J27


STRAIGHT STRAIGHT
MATE_N_LOK_WHITE MATE_N_LOK_WHITE
1 1
2 2
3 3
PDU K7 PWR 4 4
4B2< 5 5

4B2< 3C5> WS_PWR PDU_J2_OUT_P 3C5< PIN


STRAIGHT
J23 J13 MATE_N_LOK_WHITE
4B5< 4C5<

1 1 1 1
2 2 2 2 4B5< 3C8> MAGIC_PC_PWR PDU_J7_OUT_P 3C8<
3 3 PDU_J2_OUT_N 3 3 PDU J2 OUT
C WS PWR 4 4 4 4 J28 J16
C
5 5 PDU_J2_OUT_G 5 5

MAGIC PC PWR 1 1 1 1
PIN PIN 2 2 2 2
STRAIGHT STRAIGHT 3 3 3 3
MATE_N_LOK_WHITE MATE_N_LOK_WHITE 4 4 4 4 PDU J7 OUT
5 5 5 5

PIN PIN
STRAIGHT STRAIGHT
4C2< MATE_N_LOK_WHITE MATE_N_LOK_WHITE

4C2< 3C5> 2ND_PC_PWR PDU_J3_OUT_P 3C5< 4C5< 3C8> 3RD_PC_PWR

J24 J14 J29

1 1 1 1 1 1
2 2 2 2 2 2
3 3 PDU_J3_OUT_N 3 3 PDU J3 OUT 3 3 PDU_J7_OUT_N
2ND PC PWR 4 4 4 4 3RD PC PWR 4 4
5 5 PDU_J3_OUT_G 5 5 5 5 PDU_J7_OUT_G

D PIN PIN PIN D


STRAIGHT STRAIGHT STRAIGHT
MATE_N_LOK_WHITE MATE_N_LOK_WHITE MATE_N_LOK_WHITE

REV 2 1July04
Made By Date Title
GE Healthcare
SONG WEN QING 24-Mar-04 HALO RELAY BOARD Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Wed May 05 09:05:18 2004 pdu_relay_board 2 of 5 2407245SCH 2 2 of 5

1 2 3 4 5 6 7 8

Figure 5-6 2407245SCH_s0002_r2

Section 2.0 — "Halo" Relay Board, 2407245 Page 112


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A 24VDC TEST POINT A


RED
4A1< 5B3> 24V0 4PIN
4PIN K16
K11 ALE15B24
ALE15B24

1N4148
C1 COILP CLOSED1 COM WS_DETECTOR_PWR 2A5<

75V
1N4148
C1 COILP CLOSED1 COM TABLE_PWR 2A2<

D16
75V
K16_RTN 4A5<

D11
K11_RTN 4A2<
BLACK C2 COILN OPEN1 NO 3D5<
BLACK 5 WS_DETECTOR_PWR_CONTROL_N PDU_J8_OUT_P 2A8>
C2 COILN OPEN1 NO
0 TABLE_PWR_CONTROL_N PDU_J1_OUT230V_P 2A4> 3B8<

4PIN 4PIN
K12 K17
ALE15B24 ALE15B24

1N4148

1N4148
C1 COILP CLOSED1 COM OTS_PWR 2B2< C1 COILP CLOSED1 COM PDU_K7_PWR 2B5<

75V

75V
D12

D17
K12_RTN K17_RTN 4B5<
4B2<
B BLACK C2 COILN OPEN1 NO BLACK C2 COILN OPEN1 NO B
1 OTS_PWR_CONTROL_N PDU_J1_OUT120V_P 2B4> 6 PDU_K7_PWR_CONTROL_N PDU_J8_OUT_P 2A8> 3B8<
3D5<

4PIN 4PIN
K13 K18
ALE15B24 ALE15B24
1N4148

1N4148
C1 COILP CLOSED1 COM WS_PWR C1 COILP CLOSED1 COM MAGIC_PC_PWR 2C5<
75V

75V
2C2<
D13

D18
K13_RTN K18_RTN 4B5<
4B2<
BLACK C2 COILN OPEN1 NO BLACK C2 COILN OPEN1 NO
2 WS_PWR_CONTROL_N PDU_J2_OUT_P 2C4> 7 MAGIC_PC_PWR_CONTROL_N PDU_J7_OUT_P 2C8> 3C8<

4PIN 4PIN
K14 K19
ALE15B24 ALE15B24

1N4148
C1 COILP CLOSED1 COM 2ND_PC_PWR C1 COILP CLOSED1 COM 3RD_PC_PWR 2D5<
C C

75V
2D2<
1N4148

D19
75V

K14_RTN K19_RTN 4C5<


D14

4C2<
BLACK C2 COILN OPEN1 NO BLACK C2 COILN OPEN1 NO
3 2ND_PC_PWR_CONTROL_N PDU_J3_OUT_P 2D4> 8 3RD_PC_PWR_CONTROL_N PDU_J7_OUT_P 2C8> 3C8<

4PIN
K15
ALE15B24

C1 COILP CLOSED1 COM TABLE_DETECTOR_PWR 2A5<


1N4148

75V

K15_RTN 4D2<
D15

BLACK C2 COILN OPEN1 NO


4 TABLE_DETECTOR_PWR_CONTROL_N PDU_J8_OUT_P 2A8>
3B8<

5C2>
4A1<
RELAY_BOARD_CONNECTOR<17..0>
D D

REV 2 1July04
Made By Date Title
GE Healthcare
SONG WEN QING 24-Mar-04 HALO RELAY BOARD Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Wed May 05 09:05:19 2004 pdu_relay_board 3 of 5 2407245SCH 2 3 of 5

1 2 3 4 5 6 7 8

Figure 5-7 2407245SCH_s0003_r2

Chapter 5 - System Cabinet Page 113


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

3D1< 5C2> RELAY_BOARD_CONNECTOR<17..0>

A A
K21 K26
RT424730 RT424615

2A4> PDU_J1_OUT230V_N 4D2< 4B5< 2A7> PDU_J8_OUT_N


1 COIL1 K21_RTN 1 COIL1 K26_RTN
2A2< 3A5> TABLE_PWR 2 COIL2 2A5< 3A8> WS_DETECTOR_PWR 2 COIL2
BLACK BLACK
3A2<
5B3> 24V0 4 COM CLOSED 3 4 COM CLOSED 3
OPEN 5 TABLE_PWR_STATUS 9 OPEN 5 WS_DETECTOR_PWR_STATUS 14

K22 K27
RT424615 RT424615

2B4> PDU_J1_OUT120V_N 4A5< 2A7> PDU_J8_OUT_N


1 COIL1 K22_RTN 4D2< 1 COIL1 K27_RTN
2B2< 3B5> OTS_PWR 2 COIL2 2B5< 3B8> PDU_K7_PWR 2 COIL2
BLACK BLACK
4 COM CLOSED 3 4 COM CLOSED 3
OPEN 5 OTS_PWR_STATUS 10 OPEN 5 PDU_K7_PWR_STATUS 15

B B

K23 K28
RT424615 RT424615

2C4> PDU_J2_OUT_N 4C5< 2C8> PDU_J7_OUT_N


1 COIL1 K23_RTN 1 COIL1 K28_RTN
2C2< 3C5> WS_PWR 2 COIL2 2C5< 3C8> MAGIC_PC_PWR 2 COIL2
BLACK BLACK
4 COM CLOSED 3 4 COM CLOSED 3
OPEN 5 WS_PWR_STATUS 11 OPEN 5 MAGIC_PC_PWR_STATUS 16

K24 K29
RT424615 RT424615
C 2C4> PDU_J3_OUT_N 4B5< 2C8> PDU_J7_OUT_N C
1 COIL1 K24_RTN 1 COIL1 K29_RTN
2D2< 3C5> 2ND_PC_PWR 2 COIL2 2D5< 3C8> 3RD_PC_PWR 2 COIL2
BLACK BLACK
4 COM CLOSED 3 4 COM CLOSED 3
OPEN 5 2ND_PC_PWR_STATUS 12 OPEN 5 3RD_PC_PWR_STATUS 17

K25
RT424615

4B5< 4A5< 2A7> PDU_J8_OUT_N


1 COIL1 K25_RTN
2A5< 3D5> TABLE_DETECTOR_PWR 2 COIL2
BLACK
4 COM CLOSED 3
OPEN 5 TABLE_DETECTOR_PWR_STATUS 13

D D

REV 2 1July04
Made By Date Title
GE Healthcare
SONG WEN QING 24-Mar-04 HALO RELAY BOARD Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Wed May 05 09:05:20 2004 pdu_relay_board 4 of 5 2407245SCH 2 4 of 5

1 2 3 4 5 6 7 8

Figure 5-8 2407245SCH_s0004_r2

Section 2.0 — "Halo" Relay Board, 2407245 Page 114


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A A

CTRL SIGNAL CONNECTOR


J1

1 1
14 14 2 2
4A1< 24V0 15 15 3 3 TABLE_PWR_CONTROL_N 0
3A2< 16 16 4 4 OTS_PWR_CONTROL_N 1
9 TABLE_PWR_STATUS 17 17 5 5 WS_PWR_CONTROL_N 2
10 OTS_PWR_STATUS 18 18 6 6 2ND_PC_PWR_CONTROL_N 3
11 WS_PWR_STATUS 19 19 7 7 TABLE_DETECTOR_PWR_CONTROL_N 4
12 2ND_PC_PWR_STATUS 20 20 8 8 WS_DETECTOR_PWR_CONTROL_N 5
13 TABLE_DETECTOR_PWR_STATUS 21 21 9 9 PDU_K7_PWR_CONTROL_N 6
14 WS_DETECTOR_PWR_STATUS 22 22 10 10 MAGIC_PC_PWR_CONTROL_N 7
B 15 PDU_K7_PWR_STATUS 23 23 11 11 3RD_PC_PWR_CONTROL_N 8 B
16 MAGIC_PC_PWR_STATUS 24 24 12 12
17 3RD_PC_PWR_STATUS 25 25 13 13

G1 G1
G2 G2

PIN
STRAIGHT
STD_DENSITY

4A1< 3D1< RELAY_BOARD_CONNECTOR<17..0>

C C

=== MOUNTING HOLES ===


MH1 MH2 MH3 MH4 MH5 MH6 MH7 MH8
MH4_50 MH4_50 MH4_50 MH4_50 MH4_50 MH4_50 MH4_50 MH4_50
1

D D

REV 2 1July04
Made By Date Title
GE Healthcare
SONG WEN QING 24-Mar-04 HALO RELAY BOARD Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet

Wed May 05 09:05:21 2004 pdu_relay_board 5 of 5 2407245SCH 2 5 of 5

1 2 3 4 5 6 7 8

Figure 5-9 2407245SCH_s0005_r2

Chapter 5 - System Cabinet Page 115


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 3.0 — Corona 2 Board, 5138201

Figure 5-10 5138201ADW_s1_r2

Section 3.0 — Corona 2 Board, 5138201 Page 116


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

NOTES:
------
1. THE VARIATION AND VERSION FOR THIS
PWA IS 5138201 A.
REVISION ECR/ECO DATE SIGN OFF 2. THIS COVER SHEET CONTROLS THE RE-
1 30-APR-05 K. COOMBS VISION STATUS OF THE ENTIRE DOCUMENT.
A THE REVISION OF ALL SHEETS OF THIS A
2 31-AUG-05 K. COOMBS
DOCUMENT WILL BE AT THE SAME LEVEL.
3. THIS SCHEMATIC MEETS THE SPECIFICA-
TIONS OF GLOBAL SCHEMATIC STANDARDS
2299773GSP.
4. ELECTRONIC SIGNATURES FOR BOTH THE
SCHEMATIC REVIEWER(S) AND APPROVER(S)
CAN BE FOUND WITHIN E-PDM.
5. UNLESS OTHERWISE SPECIFIED:
CAPACITORS ARE IN FARADS.
RESISTORS ARE IN OHMS.
INDUCTORS ARE IN HENRIES
6. THIS SCHEMATIC WAS CAPTURED WITH CADENCE
CONCEPT HDL
7. ACTIVE LOW SIGNALS ARE INCIDATED WITH
_N AFTER THE SIGNAL NAME
8. DIFFERENTIAL SIGNAL PAIRS ARE INDICATED WITH A _H
B AFTER THE SIGNAL NAME FOR THE ACTIVE HIGH SIGNAL B
SIGNAL AND A _L FOR THE ACTIVE LOW SIGNAL
9. UNLESS OTHERWISE SPECIFIED, ALL RESISTORS
ARE 1% TOLERANCE, 0.0625W, 0402 PACKAGE SIZE

C C

D D

REV 2 01July04
Made By Date Title
GE Healthcare
Kevin Coombs 30-Apr-2005 Corona 2 Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Wed Aug 31 07:46:50 2005 corona2 1 of 8 5138201SCH 2 1 of 40


1 2 3 4 5 6 7 8

Figure 5-11 5138201SCH_s1_r2

Chapter 5 - System Cabinet Page 117


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

I127

I62
FPGA
CC2_HIERARCHY PAGES 22-24
I152
PAGES 27-37 CAN_HUB
A A
MEM_INT_OE
MEM_INT_OE CONFIG<5..0>
CONFIG[5..0]
CONFIG<5..0> PAGES 19-21
NC FLASH_DATA[15..0] OTS_CAN_BUS[9..1]
DSP_XINT_BUS<2..0> FLASH_DATA<15..0> FLASH_DATA<15..0> CAN_HUB_LEDS<20..0> CAN_HUB_LEDS<20..0> OTS_CAN_BUS<9..1>
NC FLASH_CTL[6..0] WS_CAN_BUS[9..1]
DSP_GPIO_A<15..0> FLASH_CTL<6..0> FLASH_CTL<6..0> CAN_TX_RX_BUS<9..0> CAN_TX_RX_BUS<9..0> WS_CAN_BUS<9..1>
NC FLASH_ADDR[24..0] TABLE_CAN_BUS[9..1]
ADC_IN<9..0> FLASH_ADDR<24..0> FLASH_ADDR<24..0> ADC_CONTROL<4..0> ADC_CONTROL<4..0> TABLE_CAN_BUS<9..1>
CAN_RX_TO_DSP CAN_TX_FROM_DSP MAGIC_CAN_BUS[9..1]
CAN_RX_TO_DSP CAN_TX_FROM_DSP CAN_TX_FROM_DSP HUB_CTRL<4..0> HUB_CTRL<4..0> MAGIC_CAN_BUS<9..1>
NC NC FPGA_WD SPARE_CAN_BUS[9..1]
CAN_RX_FROM_XCVR CAN_TX_TO_XCVR FPGA_WD SPARE_CAN_BUS<9..1>
CC2_TDI FPGA_CLK
CC2_TDI FPGA_CLK FPGA_CLK
FPGA_RS_REQ CC2_TDO I128
FPGA_RS_REQ CC2_TDO CC2_TDO
FPGA_WD TCK[7..0] RTB
FPGA_WD TCK<7..0> TCK<7..0>
PAGES 12-15
X_DATA[15..0] TMS[7..0]
USER_PERIPH_DATA<15..0> TMS<7..0> TMS<7..0>
X_ADR[18..0] TRST_N[3..0] FPGA_TDO
USER_PERIPH_ADR<18..0> TRST_N<3..0> FPGA_TDO JEDI_RX_TX<1..0> JEDI_RX_TX<1..0>
X_CTL[8..0] USER_RST_N IDC_ATLAS_RESET[1..0]
USER_PERIPH_CTL<8..0> USER_RST_N USER_RST_N AEC_RX_TX<1..0> AEC_RX_TX<1..0> IDC_ATLAS_RESET<1..0>
NC NC X_CTL[8..0] JEDI_CAN[1..0]
CPLD2_GPIO<16..0> DIN<3..0> USER_PERIPH_CTL<8..0> IDC_RTB_RESET_RESULT IDC_RTB_RESET_RESULT JEDI_CAN<1..0>
B B
EXT_DIN[3..0] EXT_DOUT[3..0] X_ADR[18..0] JEDI_RT_BUS[7..0]
EXT_DIN<3..0> EXT_DOUT<3..0> USER_PERIPH_ADR<18..0> IDC_ATLAS_CTRL<34..0> IDC_ATLAS_CTRL<34..0> JEDI_RT_BUS<7..0>
EXT_JTAG[5..0] TI_JTAG[8..0] X_DATA[15..0] CAN_AEC_BRT[1..0]
EXT_JTAG<5..0> TI_JTAG<8..0> USER_PERIPH_DATA<15..0> IDC_RTB_RESET_CMD IDC_RTB_RESET_CMD CAN_AEC_BRT<1..0>
CF_JTAG[3..0] NC MEM_INT_OE IDC_ATLAS_RTB[15..0]
CF_JTAG<3..0> CAN_BUS<9..1> MEM_INT_OE JEDI_RTB_CTRL<12..0> JEDI_RTB_CTRL<12..0> IDC_ATLAS_RTB<15..0>
NC CLOCK_FAIL_N
DSP_UART<1..0> CLOCK_FAIL_N CLOCK_FAIL_N
ALT_CLK FPGA_RS_REQ
ALT_CLK ALT_CLK FPGA_RS_REQ
CAN_RX_TO_DSP
CAN_RX_TO_DSP
RCIM1_TX[15..0]
RCIM1_TX<15..0>
LVDS_RX_3V3[19..0]
LVDS_RX_3V3<19..0> RCIM_ENERGIZE
LVDS_TX_3V3[11..0] I140
LVDS_TX_3V3<11..0> FPGA<39..0>
PDU_IF
SPARE_SRTB<5..0>
PAGES 16-18
I122
C RCIM_WARM_24V C
RCIM_GATING RCIM_WARM_24V
PAGES 9-11 POWER_ON_FROM_RCIM2
POWER_ON_FROM_RCIM2

RCIM_ENERGIZE POWER_ON_FROM_RCIM
RCIM_ENERGIZE POWER_ON_FROM_RCIM
RCIM1_TX<15..0>
FPGA[39..0] PDU_CONNECTOR<6..0>
FPGA<39..0> PDU_CONNECTOR<6..0>
LVDS_RX_3V3<19..0>
FPGA_TDO HALO_CONNECTOR<17..0>
FPGA_TDO HALO_CONNECTOR<17..0>
LVDS_TX_3V3<11..0>
TMS[7..0] ESTOP1_TABLE
TMS<7..0> ESTOP1_TABLE
SPARE_SRTB[5..0]
SPARE_SRTB<5..0>
TRST_N[3..0] ESTOP2_TABLE
TRST_N<3..0> ESTOP2_TABLE
RCIM_RTB[35..1]
RCIM1_RTB<35..1>
TCK[7..0] ESTOP1_EXTRA
TCK<7..0> ESTOP1_EXTRA
RCIM2_SRTB[7..0]
RCIM2_SRTB<7..0>
ESTOP2_EXTRA
ESTOP2_EXTRA
SPARE_SRTB_RS485[3..0] CB_TRIP_12V0
SPARE_SRTB_RS485<3..0> CB_TRIP_12V0
ESTOP_RCIM
ESTOP_RCIM
CC2_TDI CB_TRIP_24V0
CC2_TDI CB_TRIP_24V0
ESTOP_ATLAS
ESTOP_ATLAS
POWER_ON_FROM_ATLAS
POWER_ON_FROM_ATLAS

I61 I60
D D
DECOUPLING POWER_SUPPLY
PAGES 38-40 PAGES 25-26

REV 2 01July04
CB_TRIP_12V0

CB_TRIP_24V0
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Wed Aug 31 08:45:31 2005 corona2 2 of 8 5138201SCH 2 2 of 40


1 2 3 4 5 6 7 8

Figure 5-12 5138201SCH_s2_r2

Section 3.0 — Corona 2 Board, 5138201 Page 118


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= IDC CONNECTORS =======

IDC_ATLAS_RTB[15..0]

J11 IDC1
A A
1 RECEPTOR_READY_L 1 1 26 26 RECEPTOR_READY_H 0
7 PRE_TRIGGER_L 2 2 27 27 PRE_TRIGGER_H 6
5 XRAY_ON_L 3 3 28 28 XRAY_ON_H 4
3 EXP_CMD_L 4 4 29 29 EXP_CMD_H 2
9 EXP_GATE_L 5 5 30 30 EXP_GATE_H 8
6 6 31 31
7 7 32 32
8 8 33 33
9 9 34 34
10 10 35 35
11 11 36 36 SILK SCREEN NOTE:
12 12 37 37 THE IDC CONNECTORS ARE CONNECTED
13 13 38 38 TOGETHER INTERNALLY. IT DOES NOT
14 14 39 39 MATTER WHICH IDC PLUGS INTO WHICH
15 15 40 40 CONNECTOR.
16 16 41 41
0 RESET_L 17 17 42 42 RESET_H 1
18 18 43 43
19 19 44 44
20 20 45 45
21 21 46 46
22 22 47 47
23 23 48 48
24 24 49 49
B 25 25 50 50 MH8 B
MH3_20
G1 G1

1
G2 G2 MH1
MH3_20
MH9

1
SOCKET MH3_20
RIGHT_ANGLE

1
SCSI
GND

MH10
MH4_00

1
J12 IDC2

1 RECEPTOR_READY_L 1 1 26 26 RECEPTOR_READY_H 0
7 PRE_TRIGGER_L 2 2 27 27 PRE_TRIGGER_H 6
5 XRAY_ON_L 3 3 28 28 XRAY_ON_H 4
3 EXP_CMD_L 4 4 29 29 EXP_CMD_H 2 GND
9 EXP_GATE_L 5 5 30 30 EXP_GATE_H 8
C 6 6 31 31 C
7 7 32 32
8 8 33 33
9 9 34 34
10 10 35 35
11 11 36 36
12 12 37 37
13 13 38 38
14 14 39 39
15 15 40 40
16 16 41 41
0 RESET_L 17 17 42 42 RESET_H 1
18 18 43 43
19 19 44 44
20 20 45 45
21 21 46 46
22 22 47 47
23 23 48 48
24 24 49 49
25 25 50 50

G1 G1
G2 G2 MH2
MH3_20

SOCKET 1 IDC_ATLAS_RESET[1..0]
D RIGHT_ANGLE D
SCSI
GND

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:00 2005 corona2 3 of 8 5138201SCH 2 3 of 40


1 2 3 4 5 6 7 8

Figure 5-13 5138201SCH_s3_r2

Chapter 5 - System Cabinet Page 119


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= RCIM1 INTERFACE =======


RCIM_RTB[35..1]

J14 RCIM1
A A
1 RESET_SW_H 1 1 26 26 RESET_SW_L 26
2 AUTO_POS_SW_H 2 2 27 27 AUTO_POS_SW_L 27
3 PREP_SW_H 3 3 28 28 PREP_SW_L 28
4 EXP_SW_H 4 4 29 29 EXP_SW_L 29
5 TRIAL_SW_H 5 5 30 30 TRIAL_SW_L 30
6 XRAY_ON_H 6 6 31 31 XRAY_ON_L 31
7 TUBE_OVERHEAT_H 7 7 32 32 TUBE_OVERHEAT_L 32
8 EXP_INHIBIT_H 8 8 33 33 EXP_INHIBIT_L 33
9 ERR_MESSAGE_H 9 9 34 34 ERR_MESSAGE_L 34
10 LIMIT_INDICATION_H 10 10 35 35 LIMIT_INDICATION_L 35
RCIM_WARM_24V 11 11 36 36
12 12 37 37
13 13 38 38
24V0_FUSED 14 14 39 39
SPARE_1_H 15 15 40 40 SPARE_1_L
SPARE_2_H 16 16 41 41 SPARE_2_L
RCIM_COLD_POWER 17 17 42 42
POWER_ON_FROM_RCIM 18 18 43 43
19 19 44 44
20 20 45 45
ESTOP_RCIM 21 21 46 46
SPARE_3_H 22 22 47 47 SPARE_3_L
23 23 48 48
24 24 49 49
B 25 25 50 50 B
G1 G1
G2 G2
MH3
GND MH3_20
SOCKET GND

1
STRAIGHT
SCSI

======= JEDI INTERFACE =======


C C
CAN_AEC_BRT[1..0]

JEDI_CAN[1..0] 12V0_FUSED
J3 JEDI

1 1
JEDI_RT_BUS[7..0] 9 9 2 2 EXT_CAN1_H 1
0 EXT_CAN1_L 10 10 3 3 CAN_AEC_BRT_H 0
1 CAN_AEC_BRT_L 11 11 4 4 RESET_H 0
1 RESET_L 12 12 5 5 XRAY_CMD_H 2
3 XRAY_CMD_L 13 13 6 6 XRAY_ON_H 4
5 XRAY_ON_L 14 14 7 7 EXP_DMD_H 6
7 EXP_DMD_L 15 15 8 8

G1 G1
G2 G2 MH4
MH3_20
1
PIN
RIGHT_ANGLE
STD_DENSITY
D D
GND

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:00 2005 corona2 4 of 8 5138201SCH 2 4 of 40


1 2 3 4 5 6 7 8

Figure 5-14 5138201SCH_s4_r2

Section 3.0 — Corona 2 Board, 5138201 Page 120


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

PDU_CONNECTOR[6..0] ======= PDU CONNECTOR =======


A 0 PDU_K1_CONTROL_N A
J16 PDU ESTOP_SW_PWR_OUT 5
24V0_FUSED

1 1 6 6 24V0_FUSED
2 2 7 7
1 PDU_K2_CONTROL_N 3 3 8 8 DOOR_INLK_STATUS 4
4 4 9 9 ROOM_EMERGENCY_POWER_OFF 6
5 5
G1 G1
G2 G2
2 PDU_K4_CONTROL_N

SOCKET MH6
STRAIGHT MH3_20 5
STD_DENSITY

1
3 PDU_K5_CONTROL_N

GND

B B

======= EXTRA E-STOPS =======


HALO_CONNECTOR[17..0]
EXTRA E-STOP J13
======= HALO RELAY CONNECTOR ======= 24V0_FUSED

ESTOP1_EXTRA 1 1
2 2 6 6
ESTOP2_EXTRA 3 3 7 7
4 4 8 8
24V0_FUSED 24V0_FUSED 5 5 9 9

J15 HALO G1 G1
G2 G2

1 1
2 2 14 14 PIN
0 TABLE_PWR_CONTROL_N 3 3 15 15 STRAIGHT
C 1 OTS_PWR_CONTROL_N 4 4 16 16 STD_DENSITY
C
2 WS_PWR_CONTROL_N 5 5 17 17 TABLE_PWR_STATUS 9
3 2ND_PC_PWR_CONTROL_N 6 6 18 18 OTS_PWR_STATUS 10
4 TABLE_DETECTOR_PWR_CONTROL_N 7 7 19 19 WS_PWR_STATUS 11 GND
5 WS_DETECTOR_PWR_CONTROL_N 8 8 20 20 2ND_PC_PWR_STATUS 12
6 PDU_K7_PWR_CONTROL_N 9 9 21 21 TABLE_DETECTOR_PWR_STATUS 13
7 MAGIC_PC_PWR_CONTROL_N 10 10 22 22 WS_DETECTOR_PWR_STATUS 14
8 3RD_PC_PWR_CONTROL_N 11 11 23 23 PDU_K7_PWR_STATUS 15
12 12 24 24 MAGIC_PC_PWR_STATUS 16
13 13 25 25 3RD_PC_PWR_STATUS 17 ======= TABLE E-STOPS =======
G1 G1 MH5
G2 G2 MH4_00 TABLE E-STOP J17
24V0_FUSED
1

SOCKET ESTOP1_TABLE 1 1
STRAIGHT 2 2 6 6
STD_DENSITY ESTOP2_TABLE 3 3 7 7
4 4 8 8
GND 5 5 9 9

G1 G1
G2 G2
MH7
MH3_20
PIN

1
D STRAIGHT D
STD_DENSITY

REV 2 01July04
GND

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:00 2005 corona2 5 of 8 5138201SCH 2 5 of 40


1 2 3 4 5 6 7 8

Figure 5-15 5138201SCH_s5_r2

Chapter 5 - System Cabinet Page 121


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= SPARE / LIGHTNING CAN PORT ======= ======= CAN HUB CONNECTORS =======
LIGHTNING
CAN PORT J4 OTS CAN J5

1 RESERVED 1 1 1 RESERVED 1 1 6 6 GND_OPT 6


A 2 CAN_L 2 2 6 6 GND_OPT 6 2 CAN_L 2 2 7 7 CAN_H 7 A
3 CAN_GND_CON 3 3 7 7 CAN_H 7 3 CAN_GND_CON 3 3 8 8 CAN_H_LPBK 8
4 CAN_L_LPBK 4 4 8 8 CAN_H_LPBK 8 4 CAN_L_LPBK 4 4 9 9 CAN_V 9
5 CAN_SHIELD 5 5 9 9 CAN_V 9 5 CAN_SHIELD 5 5
G1 G1
G1 G1 G2 G2
G2 G2

SOCKET
PIN RIGHT_ANGLE
GND RIGHT_ANGLE STD_DENSITY GND
STD_DENSITY SILKSCREEN NOTE:
USE THIS CAN PORT IN LIGHTNING SYSTEMS (IDC
SPARE_CAN_BUS[9..1] PRESENT). SWITCH 1 & 2 SHOULD BE ON, AND OTS_CAN_BUS[9..1]
SWITCH 3 SHOULD BE OFF.

S1
219-04MST W/S CAN J6

7 8 O1 I1 1 8 1 RESERVED 1 1 6 6 GND_OPT 6
2 CAN_L 2 2 7 7 CAN_H 7
2 7 O2 I2 2 4 3 CAN_GND_CON 3 3 8 8 CAN_H_LPBK 8
R14 4 CAN_L_LPBK 4 4 9 9 CAN_V 9
2 6 O3 I3 3 7 5 CAN_SHIELD 5 5
G1 G1
121
B 0.1W
5 O4 I4 4 G2 G2 B
1%
SOCKET
RIGHT_ANGLE
STD_DENSITY GND

WS_CAN_BUS[9..1]
======= ATLAS RTB =======
TABLE CAN J7
IDC_ATLAS_RESET[1..0]

IDC_ATLAS_RTB[15..0] 1 RESERVED 1 1 6 6 GND_OPT 6


J2 2 CAN_L 2 2 7 7 CAN_H 7
J2 ATLAS RTB 24V0_FUSED 3 CAN_GND_CON 3 3 8 8 CAN_H_LPBK 8
4 CAN_L_LPBK 4 4 9 9 CAN_V 9
19 19 RESET_H 1 5 CAN_SHIELD 5 5
0 EXPOSE_EN_H 1 1 10 10 MOTION_EN_H 8 20 20 RESET_L 0 G1 G1
1 EXPOSE_EN_L 2 2 11 11 MOTION_EN_L 9 21 21 COLD(SW)_24V G2 G2
2 EXPOSE_CMD_H 3 3 12 12 SPARE5_H 10 22 22
3 EXPOSE_CMD_L 4 4 13 13 SPARE5_L 11 23 23 COLD(SW)_GND
C 4 XRAY_ON_H 5 5 14 14 SPARE6_H 12 24 24 SOCKET
C
5 XRAY_ON_L 6 6 15 15 SPARE6_L 13 25 25 RIGHT_ANGLE
6 BRIGHT_SIG_H 7 7 16 16 SPARE7_H 14 26 26 STD_DENSITY GND
7 BRIGHT_SIG_L 8 8 17 17 SPARE7_L 15
9 9 18 18 G1 G1
G2 G2 GND TABLE_CAN_BUS[9..1]

SOCKET
RIGHT_ANGLE SOCKET
HIGH_DENSITY RIGHT_ANGLE MAGIC PC CAN J8
HIGH_DENSITY GND
ESTOP_ATLAS
1 RESERVED 1 1 6 6 GND_OPT 6
POWER_ON_FROM_ATLAS 2 CAN_L 2 2 7 7 CAN_H 7
3 CAN_GND_CON 3 3 8 8 CAN_H_LPBK 8
4 CAN_L_LPBK 4 4 9 9 CAN_V 9
5 CAN_SHIELD 5 5
G1 G1
G2 G2

======= 24VDC POWER INPUTS ======= SOCKET


RIGHT_ANGLE
GND
24V0_IN1 24V0_IN2 STD_DENSITY

D J18 24VDC POWER J22 24VDC POWER 2 MAGIC_CAN_BUS[9..1] D

REV 2 01July04
1 1 1 1
2 2 2 2

PIN PIN GE Healthcare


STRAIGHT STRAIGHT Company Restricted

B Size
MATE_N_LOK MATE_N_LOK
GND GND Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:44:59 2005 corona2 6 of 8 5138201SCH 2 6 of 40


1 2 3 4 5 6 7 8

Figure 5-16 5138201SCH_s6_r2

Section 3.0 — Corona 2 Board, 5138201 Page 122


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= SYSTEMBIST DIGITAL I/O =======


SYSTEMBIST I/O J20

A A
0 1 1 2 2 0
1 3 3 4 4 1
2 5 5 6 6 2
3 7 7 8 8 3
9 9 10 10
EXT_DOUT[3..0] EXT_DIN[3..0]

PIN R600 R601 R602 R603


STRAIGHT 10K 10K 1K 1K
DUAL_ROW_HEADER_BREAK 0.0625W 0.0625W
1% 1% 1% 1%

3V3

GND

======= SYSTEMBIST JTAG HEADER =======


R7 EXT_TRST_N 5
R8 EXT_TCK 4
3V3 R9 EXT_TDI 3
R10 EXT_TMS 2
B 0.0625W 10K 1%
B

SYSTEMBIST JTAG
J19

1 1 2 2
3 3 4 4
5 5 6 6
7 7 8 8 EXT_TDO 1
9 9 10 10

PIN
GND STRAIGHT
DUAL_ROW_HEADER_BREAK
======= TI JTAG CONNECTOR ======= EXT_PT_N 0

EXT_JTAG[5..0]
C 3V3 C

R1 R2 R3 R4 TI JTAG
======= CPLD PROGRAMMING HEADER =======
4.7K J1
0.0625W DUAL_ROW_HEADER_BREAK 3V3 R11 TCK 3
1% R12 TDI 2
R13 TMS 1
TMS 1 1 2 2 TRSTN
TDI 3 4 0.0625W 10K 1%
3 4
PD 5 5 6 6 R6
TDO 7 7 8 8 R5 4.7K
TCK_RET 9 9 10 10 0.0625W
TCK 11 11 12 12 1%
EMU0 13 13 14 14 EMU1 CPLD JTAG J21

PIN 1 1 2 2
STRAIGHT 3 3 4 4
5 5 6 6
7 7 8 8 TDO 0
9 9 10 10
GND CF_JTAG[3..0]
D D
3
8
7
6
5

0
1

TI_JTAG[8..0] PIN
GND STRAIGHT

REV 2 01July04
DUAL_ROW_HEADER_BREAK

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:01 2005 corona2 7 of 8 5138201SCH 2 7 of 40


1 2 3 4 5 6 7 8

Figure 5-17 5138201SCH_s7_r2

Chapter 5 - System Cabinet Page 123


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= SERIAL REAL TIME BUS TO RCIM2 / MIF =======


A A

RCIM2 SRTB
RCIM2_SRTB[7..0] 24V0_FUSED
J9

0 CLOCK_H 1 1 POWER_ON_FROM_RCIM2
1 CLOCK_L 2 2 9 9
3 3 10 10
2 CORONA2_TX_H 4 4 11 11 ESTOP_RCIM
3 CORONA2_TX_L 5 5 12 12
6 6 13 13 RCIM_WARM_24V
4 CORONA2_RX_H 7 7 14 14 RESET_H 6
5 CORONA2_RX_L 8 8 15 15 RESET_L 7

G1 G1
G2 G2

SOCKET
RIGHT_ANGLE
GND STD_DENSITY GND

B B

======= SPARE SERIAL REAL TIME BUS FOR GATING =======

GATING SRTB
SPARE_SRTB[5..0] 24V0_FUSED
J10 SPARE_SRTB_RS485[3..0]

0 CLOCK_H 1 1
C 1 CLOCK_L 2 2 9 9 C
3 3 10 10 RS485_1_H 0
2 CORONA2_TX_H 4 4 11 11 RS485_1_L 1
3 CORONA2_TX_L 5 5 12 12 RS485_2_H 2
6 6 13 13 RS485_2_L 3
4 CORONA2_RX_H 7 7 14 14
5 CORONA2_RX_L 8 8 15 15

G1 G1
G2 G2

SOCKET
RIGHT_ANGLE
GND STD_DENSITY GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:01 2005 corona2 8 of 8 5138201SCH 2 8 of 40


1 2 3 4 5 6 7 8

Figure 5-18 5138201SCH_s8_r2

Section 3.0 — Corona 2 Board, 5138201 Page 124


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= RCIM1, RCIM2, AND GATING LVDS/RS-485 RECEIVERS =======


R41 R64 RCIM1_RTB[35..1]
IO
1
560 40.2
R34 R55
3V3 20K 0.0625W 20 0.1W
0.0625W 1% 0.1W 1%
1% R42 1% R65
A 26 A
2 THE RCIM1 RS-485 RX LINES ARE CONVERTED
R19 R21 R23 R25 R27 R29 R31 TO LVDS WITH THIS CIRCUIT. VIN MAX=5V;
7.5K 7.5K 7.5K 7.5K 7.5K 7.5K 7.5K R35 R43 R56 R66 VIN MIN=1.5V.
0.0625W 20K 20
1% 1% 1% 1% 1% 1% 1% 0.0625W 0.1W THE TERMINATION SEEN BY THE DRIVER IS
1% R44 1% R67 1220 || 20 + 80.4 = 100.1 OHMS
27
3 5V / 100.1 OHMS = 50MA
R15 3V3 50MA * 20.20 OHM = 1.01 V
1K U1 R36 R45 R57 R68
0.0625W SCAN92LV090 20K 20 1.5V / 100.1 OHMS = 15MA
1% 0.0625W 0.1W 15MA * 19.80 OHMS = 297 MV
1% R46 1% R69
VCC<4..0> 28
LVDS_RX_3V3[19..0] AVCC<4..0> 4
IO
GND
18 G2 DE R37 R47 R58 R70
20K 20
0 A4 DIN<0> DOP_RIP<0> D5 RESET_SW_H 0.0625W 0.1W
1 D4 RO<0> DON_RIN<0> B5 RESET_SW_L 1% R48 1% R71
29
2 C4 DIN<1> DOP_RIP<1> A7 AUTO_POS_SW_H 5
3 A3 RO<1> DON_RIN<1> B6 AUTO_POS_SW_L
R38 R49 R59 R72
4 A2 DIN<2> DOP_RIP<2> C6 PREP_SW_H 20K 20
B 5 B3 RO<2> DON_RIN<2> C7 PREP_SW_L 0.0625W
1%
0.1W
1%
B
R50 R73
6 C3 DIN<3> DOP_RIP<3> B8 EXP_SW_H 30
7 C2 RO<3> DON_RIN<3> D6 EXP_SW_L SPARE_SRTB_RS485[3..0]
IO
8 D2 DIN<4> DOP_RIP<4> D8 TRIAL_SW_H 0
9 C1 RO<4> DON_RIN<4> E8 TRIAL_SW_L THE SPARE (GATING) RS-485 RX LINES
R39 R51 R62 R74 ARE CONVERTED TO LVDS WITH THIS
10 E3 DIN<5> DOP_RIP<5> E6 GATING_RS485_1_H 20K 20 CIRCUIT. SEE ABOVE FOR MATH.
11 G1 RO<5> DON_RIN<5> G8 GATING_RS485_1_L 0.0625W 0.1W
1% R52 1% R75
12 G3 DIN<6> DOP_RIP<6> F7 GATING_RS485_2_H 1
13 H2 RO<6> DON_RIN<6> F6 GATING_RS485_2_L 2

14 H3 DIN<7> DOP_RIP<7> G6 RCIM2_RX_H R40 R53 R63 R76


15 R16 F4 RO<7> DON_RIN<7> H7 RCIM2_RX_L 20K 20
0.0625W 0.1W
16 G4 DIN<8> DOP_RIP<8> G5 RCIM2_RESET_H 1% R54 1% R77
17 R17 E4 RO<8> DON_RIN<8> E5 RCIM2_RESET_L 3

19 1% 56 0.0625W H1 4 RCIM2_SRTB[7..0]
RE_N IO
FPGA_TDO D1 TDI TDO E2 R60
C IN
1 B2 TCK 100 C
1 E7 TMS AGND<4..0> 0.0625W THESE TWO PAIRS ARE ACTUAL LVDS.
0 F8 TRST_N GND<5..0> 1%
5
TCK[7..0]
IN
6
R18 3V3
TMS[7..0] 10K GND R20 R22 R24 R26 R28 R30 R32 R61
IN
0.0625W 4.7K 4.7K 4.7K 4.7K 4.7K 4.7K 4.7K 100
1% 0.0625W 0.0625W
TRST_N[3..0] 1% 1% 1% 1% 1% 1% 1% 1%
IN
7

GND R33
TDO_LVDS_INT
56
0.0625W
1%
PLACE THE TERMINATION RESISTORS FOR THE
LVDS PAIRS (7,8) FIRST AND ACAP TO THEIR PINS.

PLACE THE RS-485 TERMINATION RESISTORS (0-6)


ACAP TO THEIR PINS SECOND.
D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:18 2005 rcim_gating 1 of 3 5138201SCH 2 9 of 40


1 2 3 4 5 6 7 8

Figure 5-19 5138201SCH_s9_r2

Chapter 5 - System Cabinet Page 125


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= RCIM2 AND GATING LVDS DRIVERS =======

A A

R78 3V3
1K U2
0.0625W SCAN92LV090
1% RCIM2_SRTB[7..0]

VCC<4..0>
LVDS_TX_3V3[11..0] AVCC<4..0> CLK_H 0
IO
GND
10 G2 DE R82
100
0 A4 DIN<0> DOP_RIP<0> D5 0.0625W
1 D4 RO<0> DON_RIN<0> B5 1%
CLK_L 1
2 C4 DIN<1> DOP_RIP<1> A7 TX_H 2
3 A3 RO<1> DON_RIN<1> B6
B R83 B
4 A2 DIN<2> DOP_RIP<2> C6 100
5 B3 RO<2> DON_RIN<2> C7 0.0625W
1%
6 C3 DIN<3> DOP_RIP<3> B8 TX_L 3
7 C2 RO<3> DON_RIN<3> D6
SPARE_SRTB[5..0]
IO
8 D2 DIN<4> DOP_RIP<4> D8
9 C1 RO<4> DON_RIN<4> E8 CLK_H 0

E3 DIN<5> DOP_RIP<5> E6 R84


G1 RO<5> DON_RIN<5> G8 100
0.0625W
G3 DIN<6> DOP_RIP<6> F7 1%
H2 RO<6> DON_RIN<6> F6 CLK_L 1
TX_H 2
H3 DIN<7> DOP_RIP<7> G6
F4 RO<7> DON_RIN<7> H7 R85
100
G4 DIN<8> DOP_RIP<8> G5 0.0625W
E4 RO<8> DON_RIN<8> E5 1%
TX_L 3
11 H1 RE_N
C R81
C
TDO_LVDS_INT D1 TDI TDO E2 CC2_TDI
OUT
TCK[7..0] 2 B2 TCK
TMS[7..0] 2 E7 56
TMS AGND<4..0>
TRST_N[3..0] 0 F8 0.0625W
TRST_N GND<5..0>
1%

R79 R80 GND


10K 10K 3V3

1% 1%

PLACE THE TERMINATION RESISTORS FOR THE


LVDS PAIRS (0-3) ACAP TO THEIR PINS.

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:19 2005 rcim_gating 2 of 3 5138201SCH 2 10 of 40


1 2 3 4 5 6 7 8

Figure 5-20 5138201SCH_s10_r2

Section 3.0 — Corona 2 Board, 5138201 Page 126


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= RCIM1 RS-422 TRANSMIT CHANNELS =======

A RCIM1_RTB[35..1] U7 9 A
MAX3485E

ERROR_MESSAGE
9 RO 1
R86 6 RE_N 2 R93
100 7 DE 3 11 1K
0.1W DI 4 10 0.0625W
1% 1%
34
VCC=3V3;GND=GND

U3 12 GND
MAX3485E

LIMIT_INDICATION
10 RO 1
R87 6 RE_N 2 R94
100 7 DE 3 14 1K
0.1W DI 4 13 0.0625W
1% 1%
35
VCC=3V3;GND=GND

GND
B U4 0 B
MAX3485E
X_RAY_ON

6 RO 1
R88 6 RE_N 2 R95
100 7 DE 3 2 1K
0.1W DI 4 1 0.0625W
1% 1%
31
VCC=3V3;GND=GND

GND
U5 3
MAX3485E
TUBE_OVERHEAT

7 RO 1
R89 6 RE_N 2
100 7 DE 3 5
0.1W DI 4 4 R96
1% 1K
32 0.0625W
C VCC=3V3;GND=GND 1% C

U6 6
MAX3485E GND
EXP_INHIBIT

8 RO 1
R90 6 RE_N 2
100 7 DE 3 8
0.1W DI 4 7
1%
33
VCC=3V3;GND=GND R91 R92
1K 1K
0.0625W 0.0625W
1% 1% RCIM1_TX[15..0]
IO

GND GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:19 2005 rcim_gating 3 of 3 5138201SCH 2 11 of 40


1 2 3 4 5 6 7 8

Figure 5-21 5138201SCH_s11_r2

Chapter 5 - System Cabinet Page 127


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= JEDI RTB CHANNELS =======

A
======= JEDI CAN XCVR ======= A
RESET JEDI_RX_TX[1..0]
IO
U8
JEDI_RT_BUS<7..0> MAX1487E 5 U50
IO
SN65HVD230D

RO 1 RO00
0 6 RE_N 2 VREF 5 JEDI_CAN[1..0]
IO
1 7 DE 3 0 1 4 R
R
DI 4 1 CANH 7 1
0 1 D CANL 6 0
D
VCC=5V0;GND=GND 8 RS

VCC=3V3;GND=GND

PERFORM A 5-3.3V CONVERSION


ON THE RECEIVER OUTPUT.
GND
XRAY_CMD
U9
MAX1487E 2

RO 1 RO01 U48
B 2 6 RE_N 2 74LVC_LCX16244 B
3 7 DE 3 3
DI 4 4
12 48 OE_N

VCC=5V0;GND=GND RO00 41 A1 Y1 8
======= JEDI CAN AEC BRT XCVR =======
RO01 40 A2 Y2 9
RO02 38 A3 Y3 11 AEC_RX_TX[1..0]
IO
RO03 37 A4 Y4 12
U51
SN65HVD230D
VCC=3V3;GND=GND

VREF 5 CAN_AEC_BRT[1..0]
IO
1 4 R
R
11 CANH 7 0
XRAY_ON 0 1 D CANL 6 1
D
U10
MAX1487E R182 8 RS
1K
0.0625W
RO 1 RO02 1% VCC=3V3;GND=GND
4 6 RE_N 2
C 5 7 DE 3 6 C
DI 4 7
GND
GND
VCC=5V0;GND=GND

8
EXP_DMD
U11
MAX1487E

RO 1 RO03
6 6 RE_N 2
7 7 DE 3 9
DI 4 10

VCC=5V0;GND=GND R181
1K
D 0.0625W
1%
D
JEDI_RTB_CTRL<12..0>
IO

REV 2 01July04
GND

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:24 2005 rtb 1 of 4 5138201SCH 2 12 of 40


1 2 3 4 5 6 7 8

Figure 5-22 5138201SCH_s12_r2

Section 3.0 — Corona 2 Board, 5138201 Page 128


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= IDC / ATLAS RTB CHANNELS =======

A A

IDC RTL NAMES ARE IDC_ATLAS_CTRL[34..0]


IO
IN PARENTHESIS.

U12 U16
MAX1487E MAX1487E
32
(RECEPTOR_READY_H) (EXP_GATE_H)
RO 1 RO0 RO 1 RO4
0 EXPOSE_EN_H 6 RE_N 2 1 8 MOTION_EN_H 6 RE_N 2 17 U48
1 EXPOSE_EN_L 7 DE 3 2 9 MOTION_EN_L 7 DE 3 18 74LVC_LCX16244
DI 4 3 DI 4 19
(RECEPTOR_READY_L) (EXP_GATE_L)
25 OE_N
VCC=5V0;GND=GND VCC=5V0;GND=GND
RO0 36 A1 Y1 13 0
U13 U17 RO1 35 A2 Y2 14 4
MAX1487E MAX1487E RO2 33 A3 Y3 16 8
RO3 32 A4 Y4 17 12
(EXP_CMD_H)
RO 1 RO1 RO 1 RO5
2 EXPOSE_CMD_H 6 RE_N 2 5 10 SPARE5_H 6 RE_N 2 21 VCC=3V3;GND=GND
3 EXPOSE_CMD_L 7 DE 3 6 11 SPARE5_L 7 DE 3 22
B DI 4 7 DI 4 23 33 B
(EXP_CMD_L)
U48
VCC=5V0;GND=GND VCC=5V0;GND=GND 74LVC_LCX16244

U14 U18
MAX1487E MAX1487E 24 OE_N

(XRAY_ON_H) RO4 30 A1 Y1 19 16
RO 1 RO2 RO 1 RO6 RO5 29 A2 Y2 20 20
4 XRAY_ON_H 6 RE_N 2 9 12 SPARE6_H 6 RE_N 2 25 RO6 27 A3 Y3 22 24
5 XRAY_ON_L 7 DE 3 10 13 SPARE6_L 7 DE 3 26 RO7 26 A4 Y4 23 28
DI 4 11 DI 4 27
(XRAY_ON_L)
VCC=3V3;GND=GND
VCC=5V0;GND=GND VCC=5V0;GND=GND

U15 U19
MAX1487E MAX1487E
R187 R189
(PRE_TRIGGER_H) 1K 1K
RO 1 RO3 RO 1 RO7 0.0625W 0.0625W
6 BRIGHT_SIG_H 6 RE_N 2 13 14 SPARE7_H 6 RE_N 2 29 1% 1%
C 7 BRIGHT_SIG_L 7 DE 3 14 15 SPARE7_L 7 DE 3 30 C
DI 4 15 DI 4 31
(PRE_TRIGGER_L)

VCC=5V0;GND=GND VCC=5V0;GND=GND GND GND


IDC_ATLAS_RTB[15..0]
IO

2
6
10
NOTE: THE IDC CONNECTOR AND THE ATLAS RTB CONNECTOR 14
SHARE SOME RS485 TRANSCEIVERS. THEIR USE IS MUTALLY EXCLUSIVE. 18
22
26
30

R183 R184 R185 R186 R188 R190 R191 R192


1K 1K 1K 1K 1K 1K 1K 1K
0.0625W
1% 1% 1% 1% 1% 1% 1% 1%

D D
GND

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:25 2005 rtb 2 of 4 5138201SCH 2 13 of 40


1 2 3 4 5 6 7 8

Figure 5-23 5138201SCH_s13_r2

Chapter 5 - System Cabinet Page 129


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= TERMINATION FOR THE ATLAS RTB =======

A A
IDC_ATLAS_RTB[15..0]

5V0 5V0 5V0 5V0

0
1 R205 R211 R217 R223
470 470 470 470
2 0.0625W 0.0625W 0.0625W 0.0625W
3 1% 1% 1% 1%

4
5 R206 R212 R218 R224
100 100 100 100
6 0.25W 0.25W 0.25W 0.25W
B 7 1% 1% 1% 1% B
THE TEAM DECIDED AT THE CORONA2
SCH REVIEW ON 5/17/2005 TO MAKE THE
TERMINATION ON CORONA PERMANENT. R207 R213 R219 R225 PLACE THE TERMINATION RESISTORS
470 470 470 470 NEAR THE ASSOCIATED TRANSCEIVER.
0.0625W 0.0625W 0.0625W 0.0625W
1% 1% 1% 1%

GND GND GND GND

5V0 5V0 5V0 5V0


C C
8
9 R208 R214 R220 R226
470 470 470 470
10 0.0625W 0.0625W 0.0625W 0.0625W
11 1% 1% 1% 1%

12
13 R209 R215 R221 R227
100 100 100 100
14 0.25W 0.25W 0.25W 0.25W
15 1% 1% 1% 1%

R210 R216 R222 R228


470 470 470 470
0.0625W 0.0625W 0.0625W 0.0625W
1% 1% 1% 1%

GND GND GND GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:25 2005 rtb 3 of 4 5138201SCH 2 14 of 40


1 2 3 4 5 6 7 8

Figure 5-24 5138201SCH_s14_r2

Section 3.0 — Corona 2 Board, 5138201 Page 130


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= RT BUS RESET CIRCUIT GENERATION =======

A 5V0 A
3V3

R194
10K U52
0.0625W 74LVC-LCX14 R201
1% 221 N12V0 12V0_FUSED
0.1W
IDC_RTB_RESET_CMD 1 2 1%
IN A B

RTB RESET

1
VCC=3V3;GND=GND DS9
HSMH-C190
RED D8
BAS116 K1
TQ2SA-5V

2
C
R198 75V
B 1 COILP
Q1 10 COILN
150 MMBT3904
E D10
0.0625W 3 2
R200 D9 COM<1> CLOSED<1> BAS116
1% 4 RESET_LOW
BAS116 OPEN<1>
RESET_HIGH 8 COM<0> CLOSED<0> 9
1K 7
OPEN<0> 75V
0.0625W
B 1%
75V B
GND

0
1
IDC_ATLAS_RESET[1..0]
IO

======= RT BUS RESET RECEIVE CIRCUIT =======

C C
U53
CNY17-4S
5V0
R197 R199
RESET_HIGH 1 A B 6
U52
2K 2K C 5 74LVC-LCX14
0.1W 0.1W
R202
1% 1% 10K 2 E 4
R193 R195 R196 D7 R204
10K 10K 10K 0.1U 0.1W K 3 4 IDC_RTB_RESET_RESULT
BAS116 A B OUT
C1

0.1W 0.1W 0.1W 50V 0.1%


10% 75V
0.1% 0.1% 0.1% 10K
0.0625W C2
R203 VCC=3V3;GND=GND
1K 1%
0.1U
RESET_LOW 0.0625W 16V
1% +80%-20%

GND GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:25 2005 rtb 4 of 4 5138201SCH 2 15 of 40


1 2 3 4 5 6 7 8

Figure 5-25 5138201SCH_s15_r2

Chapter 5 - System Cabinet Page 131


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= RELAY ACTIVATION FOR THE HALO AND PDU =======


FPGA[39..0]
IO
U55
74VHC-AHC08

A 0 1 A A
U54 Y 3 TABLE_PWR_CONTROL
74LVC-LCX04 2 B 11 3RD_PC_PWR_CONTROL
R243
1 1 2 VCC=3V3;GND=GND 1K
A Y
0.0625W
R230 R237 1%
1K VCC=3V3;GND=GND 1K
0.0625W 0.0625W
1% 1%
U55 12 GND PDU_K1_CONTROL
74VHC-AHC08 R240
GND GND 1K
2 9 A 0.0625W
U54 Y 8 OTS_PWR_CONTROL 1% 3RD_PC_PWR_CONTROL_N 8
74LVC-LCX04 10 B
ACTIVE LOW CONTROL SIGNALS
3 5 6 VCC=3V3;GND=GND FOR THE RELAYS INSIDE PDU.
A Y
GND
13 PDU_K2_CONTROL PDU_CONNECTOR<6..0>
IO
R231 VCC=3V3;GND=GND R238
1K 1K R244
0.0625W 0.0625W 1K Q3
1% 1% 0.0625W ULN2803
U55 1%
74VHC-AHC08
B GND GND 1 I1 O1 18 B
4 4 A 2 I2 O2 17 PDU_K1_CONTROL_N 0
U54 Y 6 WS_POWER_CONTROL GND 3 I3 O3 16 PDU_K2_CONTROL_N 1
74LVC-LCX04 5 B 14 PDU_K4_CONTROL 4 I4 O4 15 PDU_K4_CONTROL_N 2
5 I5 O5 14 PDU_K5_CONTROL_N 3
5 3 4 VCC=3V3;GND=GND R241 6 I6 O6 13
A Y
1K 7 I7 O7 12
R229 R239 Q2 0.0625W 8 I8 O8 11
1K VCC=3V3;GND=GND 1K ULN2803 1% 10
0.0625W 0.0625W COMMON 24V0
1% 1% GND 9
1 I1 O1 18 TABLE_PWR_CONTROL_N 0
2 I2 O2 17 OTS_PWR_CONTROL_N 1 GND
GND GND 3 I3 O3 16 WS_PWR_CONTROL_N 2
6 2ND_PC_PWR_CONTROL 4 I4 O4 15 2ND_PC_PWR_CONTROL_N 3
5 I5 O5 14 TABLE_DETECTOR_PWR_CONTROL_N 4
R234 6 I6 O6 13 WS_DETECTOR_PWR_CONTROL_N 5
1K 7 I7 O7 12 PDU_K7_PWR_CONTROL_N 6
0.0625W 8 I8 O8 11 MAGIC_PC_PWR_CONTROL_N 7
1% 10
COMMON
GND 9
24V0
C GND C

7 TABLE_DETECTOR_PWR_CONTROL 15 PDU_K5_CONTROL

R232 R242
1K 1K
0.0625W 0.0625W
1% 1%

GND
8 WS_DETECTOR_PWR_CONTROL GND
R235
1K
0.0625W
1%

GND
9 PDU_K7_PWR_CONTROL
R233
1K ACTIVE LOW CONTROL
0.0625W SIGNALS FOR THE
1% HALO RELAY BOARD.

D D
GND
10 MAGIC_PC_PWR_CONTROL HALO_CONNECTOR<17..0>
IO

REV 2 01July04
R236
1K
0.0625W
1% GE Healthcare
Company Restricted

B Size
GND Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:27 2005 pdu_if 1 of 3 5138201SCH 2 16 of 40


1 2 3 4 5 6 7 8

Figure 5-26 5138201SCH_s16_r2

Section 3.0 — Corona 2 Board, 5138201 Page 132


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= 24V TO 3.3V CONVERSION (HALO) =======


HALO_CONNECTOR<17..0>
U52 U56
74LVC-LCX14 74LVC-LCX14

R245 R260
A 9 TABLE_PWR_STATUS 5 A B 6 16 14 WS_DETECTOR_PWR_STATUS 9 A B 8 21 A
47K C3 47K C11
R252 R265
0.0625W 10K 0.0625W 10K
1U VCC=3V3;GND=GND 1U VCC=3V3;GND=GND
1% 10V 0.0625W 1% 10V 0.0625W
10% 1% 10% 1%
U52
74LVC-LCX14

R246
10 OTS_PWR_STATUS 9 A B 8 17
47K C4 R253
0.0625W 10K
1U VCC=3V3;GND=GND
1% 10V 0.0625W U56
10% 1% 74LVC-LCX14

U52 R261
74LVC-LCX14 15 PDU_K7_PWR_STATUS 13 A B 12 22
47K C12
R247 R266
11 WS_PWR_STATUS 11 10 18 0.0625W 10K
A B 1U VCC=3V3;GND=GND
1% 10V 0.0625W
47K C5 10% 1%
R254
0.0625W 10K
1U VCC=3V3;GND=GND
1% 10V 0.0625W
B 10% 1% B
U52
74LVC-LCX14

R248
12 2ND_PC_PWR_STATUS 13 A B 12 19
U57
47K C6 R255 74LVC-LCX14
0.0625W 10K
1U VCC=3V3;GND=GND
1% 10V 0.0625W R262
10% 1% 16 MAGIC_PC_PWR_STATUS 1 2 23
A B
47K C13 R267
0.0625W 10K
U56 1U VCC=3V3;GND=GND
1% 10V 0.0625W
74LVC-LCX14
10% 1%
R249
13 TABLE_DETECTOR_PWR_STATUS 3 A B 4 20
47K C7 R256
0.0625W 10K
1U VCC=3V3;GND=GND
1% 10V 0.0625W
C 10% 1% C

U56
74LVC-LCX14
U56
74LVC-LCX14 R259
17 3RD_PC_PWR_STATUS 11 A B 10 24
R250
CB_TRIP_12V0 1 2 38 47K C10
IN A B R264
0.0625W 10K
1U VCC=3V3;GND=GND
20K C8 1% 10V 0.0625W
R257
0.0625W 10K 10% 1%
1U VCC=3V3;GND=GND
1% 10V 0.0625W U57
10% 1% 24VDC DOOR INLK FEEDBACK FROM PDU. 74LVC-LCX14

R263
U56 4 DOOR_INLK_STATUS 3 A B 4 25
74LVC-LCX14
47K C14 R268
0.0625W 10K
R251 1U VCC=3V3;GND=GND
CB_TRIP_24V0 5 6 39 1% 10V 0.0625W
IN A B
10% 1%
47K C9 R258
0.0625W 10K PDU_CONNECTOR<6..0>
1U VCC=3V3;GND=GND
1% 10V 0.0625W
D 10% 1% D
FPGA[39..0]

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:28 2005 pdu_if 2 of 3 5138201SCH 2 17 of 40


1 2 3 4 5 6 7 8

Figure 5-27 5138201SCH_s17_r2

Chapter 5 - System Cabinet Page 133


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= 24V TO 3.3V CONVERSION (E-STOPS) =======


U58
74LVC-LCX14

U57 R291
74LVC-LCX14 36 8 B A 9 POWER_ON_FROM_ATLAS
IN
47K
A R274 R285 C21
0.0625W
A
ESTOP1_TABLE 9 A B 8 26 VCC=3V3;GND=GND 10K 1U
IN 1%
0.0625W 10V
47K C15 1% 10%
R279 U58
0.0625W 10K
1U VCC=3V3;GND=GND 74LVC-LCX14
1% 10V 0.0625W
10% 1% GND GND R292
U57 33 10 B A 11 POWER_ON_FROM_RCIM2
IN
74LVC-LCX14
C22 47K
R286
GND GND 10K 0.0625W
R272 VCC=3V3;GND=GND 1U
ESTOP_RCIM 11 10 27 0.0625W 10V 1%
IN A B
1% 10%
47K C16 R280
0.0625W 10K
1U VCC=3V3;GND=GND U58
1% 10V 0.0625W GND GND
74LVC-LCX14
10% 1%
U58 R293
74LVC-LCX14 30 12 B A 13 ESTOP2_TABLE
IN
GND GND C23 47K
R271 R287
POWER_ON_FROM_RCIM 3 4 28 10K 0.0625W
IN A B VCC=3V3;GND=GND 1U
0.0625W 10V 1%
47K C17 1% 10%
R281
0.0625W 10K
1U VCC=3V3;GND=GND
1% 10V 0.0625W U59
B 10% 1% 74LVC-LCX14 B
U58 GND GND
74LVC-LCX14 R294
PDU_CONNECTOR[6..0] 31 2 B A 1 ESTOP1_EXTRA
IN
R273 GND GND
6 ROOM_E_POWER_OFF 5 6 29 C24 47K
A B R288
10K 0.0625W
VCC=3V3;GND=GND 1U
47K C18 0.0625W 10V 1%
R282
0.0625W 10K 1% 10%
1U VCC=3V3;GND=GND
1% 10V 0.0625W
10% 1% U59
U57 74LVC-LCX14
74LVC-LCX14 GND GND
R295
R275 GND GND 37 4 B A 3
ESTOP_ATLAS 13 A B 12 35
IN 47K
R289 C25 R297
47K C19 10K 0.0625W 4.02K
R283 VCC=3V3;GND=GND 1U
0.0625W 10K 0.0625W 10V 1% 0.25W
1U VCC=3V3;GND=GND
1% 10V 0.0625W 1% 10% 1%
10% 1% U59

1
74LVC-LCX14 DS11
U58 HSMH-C190
C 74LVC-LCX14 GND GND R296 RED
C
GND GND 34 6 B A 5
R276 RCIM BREAKER TRIP

2
ESTOP2_EXTRA 1 2 32 C26 47K
IN A B R290
10K 0.0625W
VCC=3V3;GND=GND 1U
47K C20 0.0625W 10V 1% GND
R284
0.0625W 10K 1% 10%
1U VCC=3V3;GND=GND
1% 10V 0.0625W
10% 1%
GND GND
FPGA[39..0]
GND GND
5V0

3V3 R278
221 K2
R269 0.1W TQ2SA-5V
10K 1% D11
0.0625W RCIM_ENERGIZE BAS116
2

1% DS10 75V 1 COILP CB1


HSMG-C190 10 COILN 1410-L210-L2F1-S02-2A
GREEN 3
U57 3 COM CLOSED 2 1 RCIM_24V_TRIP
74LVC-LCX14 OPEN 4 RCIM_WARM_24V
1

OUT
C
R270 2
2A
D IN
RCIM_ENERGIZE 5 A B 6 B
Q4 240VAC,28VDC
D
24V0
150 MMBT3904
0.0625W E

REV 2 01July04
VCC=3V3;GND=GND R277
1%

1K
0.0625W GE Healthcare
1% GND Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:28 2005 pdu_if 3 of 3 5138201SCH 2 18 of 40


1 2 3 4 5 6 7 8

Figure 5-28 5138201SCH_s18_r2

Section 3.0 — Corona 2 Board, 5138201 Page 134


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= CAN HUB HARDWARE 1 OF 2 =======

A A
HUB_CTRL[4..0]
IN

U22 CAN_TX_RX_BUS[9..0]
IO
TLP3115 3V3
U24
R302 SN65HVD230D
HUB_CTRL[0] 1 A D1 4
330 5 VREF
12V0_FUSED 0.0625W R 4 OTS_CAN_RX 0
1% R
2 3 7 CANH
OTS_CAN_BUS[9..1] K D2 6 CANL D 1 OTS_CAN_TX 1
IO D
9
7 CAN_H RS 8 ADC_CONTROL[4..0]
IO
GND
R298 R306
121 VCC=3V3;GND=GND 0 3V3
0.1W 0.0625W 3V3_CLEAN
1%
2 CAN_L U20
3 TLP3115 C28 C29
B ADC_INPUT[0] 0.1U 0.1U B
GND 16V 16V
1 A D1 4 R304 U26
GND 121 TLV1570CPW
R300 0.1W
HUB_CTRL[0] 1%
2 3 GND 15 AVDD
330 K D2 GND 6 DVDD
0.0625W
1% 0 9 SCLK R310
GND GND 1 10 SDIN SDOUT 11 4
2 12 CS_N 3V3
3 8 56
FS
0.0625W

2
AIN 20 1%
AIN
3V3_CLEAN 13 REF

U23 0 5 CH0 MO 19 AIN FL1


TLP3115 3V3 C27 1 4 CH1 BLM21PG221SN1
U25 0.1U 2 3 CH2 220 OHM
R303 SN65HVD230D 16V 3 2 CH3
HUB_CTRL[1] 1 A D1 4 4 1 CH4
18 CH5 3V3_CLEAN
C C

1
330 5 3V3 17 14
VREF CH6 AGND
12V0_FUSED 0.0625W R 4 WS_CAN_RX 2 16 7 C30
R CH7 DGND
1% 2 3 7 CANH GND R308 1U
WS_CAN_BUS[9..1] K D2 6 CANL D 1 WS_CAN_TX 3 1K 10V
IO D
9 10%
7 CAN_H RS 8 1%
GND GND
R299 R307
121 VCC=3V3;GND=GND 0 R309 GND
0.1W 0.0625W 1K PLACE THE FILTER ACAP
1% TO THE TLV1570.
2 CAN_L U21 1%
3 TLP3115
ADC_INPUT[1]
GND GND
1 A D1 4 R305
GND 121
R301 0.1W
HUB_CTRL[1] 1% ADC_INPUT[4..0]
2 3
330 K D2
0.0625W
1%
GND GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:30 2005 can_hub 1 of 3 5138201SCH 2 19 of 40


1 2 3 4 5 6 7 8

Figure 5-29 5138201SCH_s19_r2

Chapter 5 - System Cabinet Page 135


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= CAN HUB HARDWARE 2 OF 2 =======

A A
HUB_CTRL[4..0]

U29 CAN_TX_RX_BUS[9..0]
TLP3115 3V3
U31
R315 SN65HVD230D
HUB_CTRL[2] 1 A D1 4
330 5 VREF
12V0_FUSED 0.0625W R 4 TABLE_CAN_RX 4
1% R
2 3 7 CANH
TABLE_CAN_BUS[9..1] K D2 6 CANL D 1 TABLE_CAN_TX 5
IO D
9
7 CAN_H RS 8 8
GND
R311 R319 9
121 VCC=3V3;GND=GND 0
0.1W 0.0625W U34
1% TLP3115 3V3
2 CAN_L U27 U35
3 TLP3115 R322 SN65HVD230D
B ADC_INPUT[2] HUB_CTRL[4] 1 A D1 4 B
GND
1 A D1 4 330 5
R317 VREF
GND 121 12V0_FUSED 0.0625W R 4 SPARE_CAN_RX
1% R
R313 0.1W 2 3 7 CANH
HUB_CTRL[2] 1% K D2 6 CANL D 1 SPARE_CAN_TX
D
2 3 9
330 K D2 7 CAN_H RS 8
0.0625W GND
1%
R325
GND GND VCC=3V3;GND=GND 0
0.0625W

2 CAN_L U33
3 TLP3115
ADC_INPUT[4]
GND
U30 1 A D1 4 R324
TLP3115 3V3 GND 121
U32 0.1W
R316 SN65HVD230D 1%
HUB_CTRL[3] 1 A D1 4 2 3
K D2
C 330 5 VREF
C
12V0_FUSED 0.0625W R 4 MAGIC_CAN_RX 6
1% R
2 3 7 CANH GND GND
MAGIC_CAN_BUS[9..1] K D2 6 CANL D 1 MAGIC_CAN_TX 7
IO D
9
7 CAN_H RS 8
GND R323
R312 R320 HUB_CTRL[4]
121 VCC=3V3;GND=GND 0
0.1W 0.0625W 330
1% 0.0625W
2 CAN_L 1%
U28
3 TLP3115
ADC_INPUT[3]
GND
1 A D1 4 R318
GND 121
R314 0.1W
HUB_CTRL[3] 1% ADC_INPUT[4..0]
2 3
330 K D2
0.0625W
1%
GND GND

D D

REV 2 01July04
SPARE_CAN_BUS[9..1]
IO

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:30 2005 can_hub 2 of 3 5138201SCH 2 20 of 40


1 2 3 4 5 6 7 8

Figure 5-30 5138201SCH_s20_r2

Section 3.0 — Corona 2 Board, 5138201 Page 136


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= CAN HUB STATUS LEDS =======

CAN_HUB_LEDS[20..0]
IN
DS12 DS22
HSMG-C190 HSMG-C190
A R326 R336 A
0 2 1 10 2 1
150 150
GREEN GREEN
0.0625W 0.0625W
DRAW A BOX AROUND THIS LED
1% 1%
AND LABEL IT "CAN HUB ACTIVE".
DS13 3V3 DS23 3V3
HSMH-C190 HSMH-C190
DS32
R327 R337
1 2 1 11 2 1 HSMG-C190

150 150 20 2 1
RED RED
0.0625W 0.0625W
1% 1%
GREEN
DS14 DS24
HSMG-C190
DRAW A BOX AROUND THESE 5 LEDS. HSMG-C190
DRAW A BOX AROUND THESE 5 LEDS.
LABEL THE BOX "OTS CAN PORT". LABEL LABEL THE BOX "TABLE CAN PORT". LABEL R346
R328 EACH OF THEM ACCORDING TO THIS LIST: R338 EACH OF THEM ACCORDING TO THIS LIST: 150
2 2 1 12 2 1 0.0625W
GREEN LED 1 - "ACTIVE" GREEN LED 1 - "ACTIVE" 1%
150 150
GREEN
0.0625W
RED LED 2 - "BAD TERM." GREEN
0.0625W
RED LED 2 - "BAD TERM."
1%
GREEN LED 3 - "TRANSMIT" 1%
GREEN LED 3 - "TRANSMIT"
DS15
YELLOW LED - "ERROR FRAME" DS25
YELLOW LED - "ERROR FRAME"
RED LED - "PORT HUNG" RED LED - "PORT HUNG" GND
HSMY-C190 HSMY-C190
R329 R339
3 2 1 13 2 1
B 150 150
B
YELLOW YELLOW
0.0625W 0.0625W
1% 1%
DS16 DS26
HSMH-C190 HSMH-C190
R330 R340
4 1 2 14 1 2
150 150
RED RED
0.0625W 0.0625W
1% 1%
GND GND
DS17 DS27
HSMG-C190 HSMG-C190
R331 R341
5 2 1 15 2 1
150 150
GREEN GREEN
0.0625W 0.0625W
1% 1%
DS18 3V3 DS28 3V3
C HSMH-C190 HSMH-C190 C
R332 R342
6 2 1 16 2 1
150 150
RED RED
0.0625W 0.0625W
1%
DRAW A BOX AROUND THESE 5 LEDS. LABEL 1%
DS19
THE BOX "WALL STAND CAN PORT". LABEL DS29
DRAW A BOX AROUND THESE 5 LEDS.
HSMG-C190
EACH OF THEM ACCORDING TO THIS LIST: HSMG-C190
LABEL THE BOX "MAGIC PC CAN PORT". LABEL
EACH OF THEM ACCORDING TO THIS LIST:
R333 GREEN LED 1 - "ACTIVE" R343
7 2 1 RED LED 2 - "BAD TERM." 17 2 1 GREEN LED 1 - "ACTIVE"
150
GREEN LED 3 - "TRANSMIT" 150
RED LED 2 - "BAD TERM."
GREEN
0.0625W
YELLOW LED - "ERROR FRAME" GREEN
0.0625W
GREEN LED 3 - "TRANSMIT"
1%
RED LED - "PORT HUNG" 1%
YELLOW LED - "ERROR FRAME"
DS20 DS30
RED LED - "PORT HUNG"
HSMY-C190 HSMY-C190
R334 R344
8 2 1 18 2 1
150 150
YELLOW YELLOW
0.0625W 0.0625W
1% 1%
DS21 DS31
HSMH-C190 HSMH-C190
R335 R345
D 9 1 2 19 1 2 D
150 150
RED RED
0.0625W 0.0625W

REV 2 01July04
1% 1%
GND GND

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:47:20 2005 can_hub 3 of 3 5138201SCH 2 21 of 40


1 2 3 4 5 6 7 8

Figure 5-31 5138201SCH_s21_r2

Chapter 5 - System Cabinet Page 137


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= CYCLONE FPGA BLOCK 1 OF 3 =======

DS2
HSMG-C190
R103
A 2 1 A
150
GREEN
CONFIG[5..0] 0 CONF_DONE 0.0625W
IO FPGA HB
1 INIT_DONE 1%
2 STATUS_N DS3
HSMY-C190
U49 R104
1V5_PLL1 3V3 EP1C20F400C7 2 1

1V5 150
YELLOW
1V5_PLL2 L2 0.0625W
VCCINT<19-0> CEO_N FPGA SAFE
L17 1%
VCCIO1<4-0> STATUS_N
R98 VCCIO2<4-0>
GND 10K L20 FPGA_TDO DS4
VCCIO3<4-0> TDO R100
1% D3 HEARTBEAT HSMG-C190
VCCIO4<4-0> IO_D3_CLKUSR
3 K4 VCCA_PLL1 IO_C3_INIT_DONE C3 INIT_DONE R105
4 K17 VCCA_PLL2 CONF_DONE L18 CONF_DONE 2 1
IO_C4_DEV_CLRN C4 NORMALM_LED
L5 B3 ERROR_LED0 150
CE_N IO_B3_DEV_OE GREEN
TCK[7..0] 5 CONFIG_N K1 K2 ERROR_LED1 0.0625W
IN CONFIG_N IO_K2_NCSO FPGA NORMAL
L4 ERROR_LED2 1%
IO_L4_ASDO
TMS[7..0] 0 L19 TCK
IN DS5
0 L16 TMS IO_L8_PLL1_OUTP L8 R101 LVDS_TX_3V3[0] EXPOSURE
K18 M8 ERROR_LED3 HSMH-C190
TDI IO_M8_PLL1_OUTN ERROR
IO_L13_PLL2_OUTP L13 R102 LVDS_TX_3V3[4] R106
B CONFIG[4] K3 DATA0 IO_M13_PLL2_OUTN M13 1 2 B
3V3 CONFIG[3] L3 DCLK
L1 K7 0.0625W 150
MSEL0 GNDA_PLL1 RED
L6 K20 1% 0.0625W
MSEL1 GNDA_PLL2
CC2_TDO K5 L7 56 1%
IN R97 CLK0 GNDG_PLL1
10K K6 L15 DS6
R99 CLK1 GNDG_PLL2 HALO
1% 10K L14 HSMH-C190
CLK2 GND<39-0> ERROR
1% K14 CLK3 R107
1 2
SYM 1 OF 3
GND 150
PASSIVE CONFIGURATION: RED
GND 0.0625W
MSEL0 = 1; MSEL1 = 0
1%
USER_RST_N DS7
IN HSMH-C190
E2 R108
FPGA_CLK 1 2
IN
150
RED
ALT_CLK 0.0625W
IN 1%
DS8
HSMH-C190
C E3 R109
C
1 2
150
RED
0.0625W
1%
LAYOUT NOTES:
1. THE FPGA_CLOCK LINE SHOULD SPLIT BETWEEN CLK1 AND CLK2
AS CLOSE TO THE FPGA AS POSSIBLE (MINIMUM SPANNING TREE.
2. THE LEDS SHOULD BE PLACED IN A BANK (ON THE EDGE OF THE
BOARD IF POSSIBLE). LABEL THEM PER THE SCHEMATIC. GND

MEM_INT_OE
OUT

FPGA_TDO
OUT

LVDS_TX_3V3[11..0]

D D

REV 2 01July04
Made By Date Title
GE Healthcare
Kevin Coombs 30-Apr-2005 Corona 2 Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:21 2005 fpga 1 of 3 5138201SCH 2 22 of 40


1 2 3 4 5 6 7 8

Figure 5-32 5138201SCH_s22_r2

Section 3.0 — Corona 2 Board, 5138201 Page 138


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= CYCLONE FPGA BLOCK 2 OF 3 =======

CAN_HUB_LEDS[20..0]
OUT
U49
EP1C20F400C7
A A
4 D4 IO_D4_VREF0B1 IO_D15_VREF0B2 D15 0
5 J8 IO_J8_VREF1B1 IO_F10_VREF1B2 F10 1
6 U1 IO_U1_VREF2B1 IO_E6_VREF2B2 E6 2

7 T4 IO_T4_DPCLK0 IO_C6_DPCLK2 C6 3
8 F3 IO_F3_DPCLK1 IO_C15_DPCLK3 C15 33 FPGA<39..0>
IO
9 C2 IO_C2 IO_C17 C17 32
CLOCK_FAIL_N D2 IO_D2 IO_B18 B18 31
IN
RCIM_ENERGIZE D1 IO_D1 IO_B17 B17 30
OUT
IDC_RTB_RESET_RESULT E4 IO_E4 IO_A17 A17 29
IN
IDC_RTB_RESET_CMD E5 IO_E5 IO_C16 C16 28
OUT
FPGA_WD E3 IO_E3 IO_B16 B16 27
OUT
FPGA_RS_REQ E2 IO_E2 IO_D16 D16 26
OUT
4 SPARE_RX_H F4 IO_F4 IO_E16 E16 25
R110 SPARE_RX_L F5 IO_F5 IO_B15 B15 24
SPARE_SRTB[5..0] 100 F2 IO_F2 IO_A15 A15 23
IO
0.1W ON_BOARD_PWR_SW F1 IO_F1 IO_E15 E15 22
5 1% F6 IO_F6 IO_F14 F14 21
FPGA[39] G5 IO_G5 IO_A14 A14 20
FPGA[38] G1 IO_G1 IO_B14 B14 19 S5
PLACE THE ESD SUPRESSORS CR1 CR2 FPGA[37] G2 IO_G2 IO_E14 E14 18 3V3 KT11B1JM
NEAR THE CONNECTOR. TVSF0603 TVSF0603 FPGA[36] G6 IO_G6 IO_E13 E13 17
8KV 8KV FPGA[35] G7 IO_G7 IO_C14 C14 16
B FPGA[34] G3 IO_G3 IO_D14 D14 15 1 1 3 3 ON_BOARD_PWR_SW B
0 G4 IO_G4 IO_A13 A13 14
1 H7 IO_H7 IO_B13 B13 13 2 2 4 4
GND 2 H1 IO_H1 IO_C13 C13 12
3 H2 IO_H2 IO_D13 D13 11
HUB_CTRL[4..0] 4 H3 IO_H3 IO_E12 E12 10 R604 C136
OUT
8 H4 IO_H4 IO_F12 F12 9 1K 1U
7 J1 IO_J1 IO_A12 A12 8 DRAW A BOX AROUND THIS 0.0625W 10V
6 J2 IO_J2 IO_B12 B12 7 SWITCH AND LABEL IT: 1% 10%
5 H5 IO_H5 IO_D12 D12 6 "POWER ON SYSTEM"
4 H6 IO_H6 IO_C12 C12 5
3 J3 IO_J3 IO_E11 E11 4
2 J4 IO_J4 IO_A11 A11 3
1 J5 IO_J5 IO_B11 B11 2
0 J6 IO_J6 IO_D11 D11 1
18 J7 IO_J7 IO_C11 C11 0 GND
USER_PERIPH_CTL[8..0] 17 M2 IO_M2 IO_D10 D10 1
IO
16 M1 IO_M1 IO_C10 C10 R127 2
15 M5 A10 3 56
IO_M5 IO_A10
14 M6 B10 5 0.0625W
IO_M6 IO_B10
13 M4 E10 6 1%
IO_M4 IO_E10 R128
12 M3 IO_M3 IO_E9 E9 7
11 M7 IO_M7 IO_C9 C9 8
C 10 N6 IO_N6 IO_D9 D9 9 C
9 N1 IO_N1 IO_A9 A9 10
8 N2 IO_N2 IO_B9 B9 11
7 N4 IO_N4 IO_C8 C8 0
6 N3 IO_N3 IO_D8 D8 1
5 N5 IO_N5 IO_A8 A8 2
4 N7 IO_N7 IO_B8 B8 3
3 P7 IO_P7 IO_E8 E8 4 LVDS_TX_3V3[11..0]
IO
2 P2 IO_P2 IO_F8 F8 5
1 P1 IO_P1 IO_C7 C7 6
0 P6 IO_P6 IO_D7 D7 7
USER_PERIPH_ADR[18..0] 15 R111 P5 IO_P5 IO_A7 A7 8
IN
14 R112 P3 IO_P3 IO_B7 B7 9
13 R113 P4 IO_P4 IO_E7 E7 10
12 R114 R1 IO_R1 IO_F7 F7 11
11 R115 R2 IO_R2 IO_A6 A6 12
10 R116 R6 IO_R6 IO_B6 B6 13
9 R117 R5 IO_R5 IO_B5 B5 14
8 R118 R3 IO_R3 IO_C5 C5 15
7 R119 R4 IO_R4 IO_D6 D6 16
6 R120 T2 IO_T2 IO_D5 D5 17
5 R121 T3 IO_T3 IO_A4 A4 18
4 R122 U4 IO_U4 IO_B4 B4 19
3 R123 U2 IO_U2
2 R124 U3 IO_U3 LVDS_RX_3V3[19..0]
IO
1 R125 V2 IO_V2
D 0 R126 V3 IO_V3 D
USER_PERIPH_DATA[15..0]
IO 56 SYM 2 OF 3
0.0625W

REV 2 01July04
1%

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:21 2005 fpga 2 of 3 5138201SCH 2 23 of 40


1 2 3 4 5 6 7 8

Figure 5-33 5138201SCH_s23_r2

Chapter 5 - System Cabinet Page 139


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= CYCLONE FPGA BLOCK 3 OF 3 =======


CAN_HUB_LEDS[20..0]
OUT

EP1C20F400C7
U49

A 15 D17 IO_D17_VREF0B3 IO_Y15_VREF0B4 Y15 10 A


16 K19 IO_K19_VREF1B3 IO_V10_VREF1B4 V10 11
17 U17 IO_U17_VREF2B3 IO_T7_VREF2B4 T7 12

18 F18 IO_F18_DPCLK4 IO_W15_DPCLK6 W15 13


19 T19 IO_T19_DPCLK5 IO_T6_DPCLK7 T6 14

RCIM1_TX[15..0] 20 W18 IO_W18 IO_W3 W3 CAN_TX_FROM_DSP


IO IN
0 V18 IO_V18 IO_Y4 Y4 CAN_RX_TO_DSP
OUT
1 V19 IO_V19 IO_V4 V4 12
2 U20 IO_U20 IO_W4 W4 11
3 U18 IO_U18 IO_T5 T5 10
4 U19 IO_U19 IO_U5 U5 9
5 T18 IO_T18 IO_V5 V5 8
6 T17 IO_T17 IO_W5 W5 7
7 R16 IO_R16 IO_W6 W6 6
8 R19 IO_R19 IO_Y6 Y6 5
9 R20 IO_R20 IO_U6 U6 4
10 R17 IO_R17 IO_V6 V6 3
11 R18 IO_R18 IO_W7 W7 2
12 R15 IO_R15 IO_Y7 Y7 1
13 P14 IO_P14 IO_R7 R7 0
CAN_TX_RX_BUS[9..0] 14 P18 IO_P18 IO_T8 T8 R133 15 JEDI_RTB_CTRL[12..0]
IO IO
15 P17 IO_P17 IO_V7 V7 R134 14
0 OTS_RX P16 IO_P16 IO_U7 U7 R135 13
1 OTS_TX P15 IO_P15 IO_V8 V8 R136 12
B 2 WS_RX P19 IO_P19 IO_U8 U8 R137 11 B
3 WS_TX P20 IO_P20 IO_W8 W8 R138 10
4 TABLE_RX N14 IO_N14 IO_Y8 Y8 R139 9
5 TABLE_TX N18 IO_N18 IO_U9 U9 R140 8
6 MAGIC_RX N17 IO_N17 IO_V9 V9 R141 7
7 MAGIC_TX N19 IO_N19 IO_T9 T9 R142 6
ADC_CONTROL[4..0] 8 SPARE_TX N20 IO_N20 IO_R9 R9 R143 5
IO
9 SPARE_RX N16 IO_N16 IO_Y9 Y9 R144 4
0 CLK R129 N15 IO_N15 IO_W9 W9 R145 3
1 MO R130 M18 IO_M18 IO_T10 T10 R146 2
2 CS_N R131 M17 IO_M17 IO_U10 U10 R147 1
3 FS R132 M15 IO_M15 IO_W10 W10 R148 0 FLASH_DATA[15..0]
IO
4 SO M16 IO_M16 IO_Y10 Y10 R149 24
1 M20 IO_M20 IO_V11 V11 R150 23
0 M19 IO_M19 IO_U11 U11 R151 22
M14 IO_M14 IO_W11 W11 R152 21
AEC_RX_TX[1..0] 0.0625W J13 Y11 20
IO IO_J13 IO_Y11 R153
34 56 1% K16 R11 19
IO_K16 IO_R11 R154
33 K15 IO_K15 IO_Y12 Y12 R155 18
32 J18 IO_J18 IO_W12 W12 R156 17
31 J17 IO_J17 IO_T11 T11 R157 16
30 J14 IO_J14 IO_T12 T12 R158 15
29 H14 IO_H14 IO_U12 U12 R159 14
C 28 J20 IO_J20 IO_V12 V12 R160 13 C
27 J19 IO_J19 IO_T13 T13 R161 12
26 J15 IO_J15 IO_R13 R13 R162 11
25 J16 IO_J16 IO_Y13 Y13 R163 10
24 H20 IO_H20 IO_W13 W13 R164 9
23 H19 IO_H19 IO_U13 U13 R165 8
22 H17 IO_H17 IO_V13 V13 R166 7
21 H18 IO_H18 IO_R14 R14 R167 6
20 H16 IO_H16 IO_T14 T14 R168 5
19 G17 IO_G17 IO_W14 W14 R169 4
18 G18 IO_G18 IO_Y14 Y14 R170 3
17 H15 IO_H15 IO_U14 U14 R171 2
16 G14 IO_G14 IO_V14 V14 R172 1
15 G19 IO_G19 IO_V15 V15 R173 0 FLASH_ADDR[24..0]
OUT
14 G20 IO_G20 IO_U15 U15 R174 MEM_BYTE_N 6
13 G15 IO_G15 IO_T15 T15 R175 MEM_RESET_N 5
12 G16 IO_G16 IO_T16 T16 R176 MEM_WP_N 4
11 F20 IO_F20 IO_W16 W16 R177 MEM_WE_N 3
10 F19 IO_F19 IO_V16 V16 R178 MEM_OE_N 2
9 F15 IO_F15 IO_V17 V17 R179 MEM_CE_N 1
8 F16 IO_F16 IO_U16 U16 R180 MEM_BUSY_N 0
7 E19 IO_E19 IO_Y17 Y17 1
6 E18 IO_E18 IO_W17 W17 0 FLASH_CTL[6..0]
IO
5 F17 IO_F17
4 E17 56 JEDI_RX_TX[1..0]
IO_E17 IO
3 D20 0.0625W
IO_D20
1%
D 2 D19 IO_D19 D
1 D18 IO_D18
0 C19 IO_C19

REV 2 01July04
C18 IO_C18

IDC_ATLAS_CTRL[34..0] SYM 3 0F 3
IO
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:22 2005 fpga 3 of 3 5138201SCH 2 24 of 40


1 2 3 4 5 6 7 8

Figure 5-34 5138201SCH_s24_r2

Section 3.0 — Corona 2 Board, 5138201 Page 140


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= SWITCHING POWER SUPPLIES =======


LAYOUT NOTES:
1. PLACE COMPONENTS ACAP TO THE SWITCHER, WITH THE FOLLOWING PRIORITY:
1. INPUT DECOUPLING CAP TP3
2. SCHOTTKY DIODE 12V
A 3. INDUCTOR TIN A
4. BULK INPUT CAPS
2. ROUTE FEEDBACK TRACES AND COMPONENTS FAR AWAY FROM THE OUTPUTS
3. ABSOLUTELY NOTHING SHOULD BE UNDER THE INDUCTOR OR DIODE, OR THE 24V0
PS3
LM2676
R361
1K
0.1W
12V0
TRACE CONNECTING THEM, UNLESS SEPARATED BY A GROUND PLANE. 24.0000
4. ROUTE THE SW OUTPUT (BETWEEN INDUCTOR/DIODE/LM2676) UNDER A GROUND PLANE. 3A L3 1%
5. SEE NATIONAL APP NOTE AN-1229 FOR ADDITIONAL LAYOUT GUIDELINES. 2 INPUT SWIT_OUT 1
6 FEEDBACK BOOST 3 68U
12V0 7 ON_OFF 20% R363
DESIGN NOTES: C41 C42 GND 4 3A 1.65K
1. 12V SUPPLY DESIGNED FOR 2.5A OUTPUT. 1000U 0.1U TAB_GND 8 C46 C48 C49 0.1W
2. 3.3V SUPPLY DESIGNED FOR 3A OUTPUT. 50V 50V R355 10N D15 100U 100U 1%
20% 10% 9.09K 50V B340A 16V 16V
3. 5V SUPPLY DESIGNED FOR 2A OUTPUT.
24V0

2
4. -12V SUPPLY DESIGNED FOR 0.5A OUTPUT. TP13 0.1W 1.2V TO 37V 10% 40V 20% 20% DS35
24V0 1% HSMG-C190
TIN GREEN
24V0_IN2 D17 12V
SS36
R356

1
1.02K
0.1W
60V 1%
GND
24V0_IN1 D18 GND
SS36
GND
TP1 TP4
60V
B 5V0 3V3 B
TIN TIN

12V0
PS1
LM2676
R351
1K
0.1W
5V0 12V0
PS4
LM2676
R362
1K
0.1W
3V3
L1 1% L4 1%
2 INPUT SWIT_OUT 1 2 INPUT SWIT_OUT 1
6 FEEDBACK BOOST 3 33U 6 FEEDBACK BOOST 3 33U
7 ON_OFF 20% R353 7 ON_OFF 20% R366
C31 C33 5V0 GND 4 5A 332 C43 C44 C45 3V3 GND 4 5A 150
47U 0.1U TAB_GND 8 C36 C38 C39 0.1W 47U 47U 0.1U TAB_GND 8 C47 C50 C51 0.1W
20V 50V 10N D12 100U 100U 1% 20V 20V 50V 10N D16 100U 100U 1%
10% 10% R347 50V B340A 16V 16V 10% 10% 10% R359 50V B340A 16V 16V

2
3.16K 1.2V TO 37V 10% 40V 20% 20% DS33 1.74K 1.2V TO 37V 10% 40V 20% 20% DS38
0.1W HSMG-C190 0.1W HSMG-C190
1% GREEN 1% GREEN
5V0 3V3

1
R348 R360
GND 1K 1K
0.1W 0.1W
1% GND 1%
C GND GND C

GND GND CB2


1410-L210-L2F1-S02-1A CB_TRIP_12V0
OUT
12V0 3 R364
PS2 1
12V0 LM2676

1
4.02K
2 DS36
0.25W
L2 HSMH-C190
2 INPUT SWIT_OUT 1 1A
240VAC,28VDC
12V0_FUSED 1%
RED
6 FEEDBACK BOOST 3 68U 12V TRIP
7 ON_OFF 20% R354

2
C32 C34 GND 4 3A 1.65K R357
47U 0.1U TAB_GND 8 C37 C40 0.1W 100
20V 50V GND 10N D13 100U D14 1% GND 1W GND
10% 10% R349 50V B340A 16V B340A 5% GND
2

C35 9.09K 1.2V TO 37V 10% 40V 20% 40V DS34 CB3
22U 0.1W HSMG-C190 R358 1410-L210-L2F1-S02-1A CB_TRIP_24V0
35V 1% OUT
GREEN 100 24V0 3 R365
10% N12V 1W 1
5%

1
1

1.82K
R350 2 DS37
GND 1.02K TP2 0.1W
HSMH-C190
0.1W N12V N12V0 1A
240VAC,28VDC
24V0_FUSED 1%
RED
1%
D
TIN
N12V0 PLACE THIS LOAD AWAY
MISC 24V TRIP
D

2
R352 FROM THE SWITCHER
1K

REV 2 01July04
N12V0 0.1W GND
1%

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:03 2005 power_supply 1 of 2 5138201SCH 2 25 of 40


1 2 3 4 5 6 7 8

Figure 5-35 5138201SCH_s25_r2

Chapter 5 - System Cabinet Page 141


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= 1V5 LINEAR POWER SUPPLY ======= ======= 1V5 INDICATOR CIRCUIT =======
TP5

A 1V5 TIN 5V0 A


3V3
VR2
LT1963EQ
R367
1K
0.1W
1V5
1%
2 VIN VOUT 4 R375
1 SHDN_N SENSE_ADJ 5 221
R368 1V5 0.1W
R605 GND<1-0> 681 1%
C52 C54 0.1W
10U 0.1U 1% C58 1V5

2
10V 16V 1K 100U
1.21V TO 20V DS40
+80%-20% +80%-20% 0.0625W 16V HSMG-C190
1% 20%
R369 GREEN
2.87K
0.1W

1
1%
C
R371
B
GND Q6
150 MMBT3904
0.0625W E
LAYOUT NOTES: GND R373
1%
1. PLACE THE INPUT CAPS AS CLOSE AS POSSIBLE TO THE REGULATOR.
2. TRACES BETWEEN THE INPUT CAPS AND THE DEVICE SHOULD BE RATED FOR 1.5A.
1K
0.1W
B 1%
B
GND

======= PLL SUPPLY FILTERS ======= ======= 1V9 INDICATOR CIRCUIT =======
5V0

1V5
FL2
1V5_PLL1 1V5
FL3
1V5_PLL2
GND
BLM11A121S BLM11A121S R374
1 2 1 2 221 TP6 TP7 TP8
C 1V9 0.1W C
120 OHM 120 OHM 1%
TIN TIN TIN
C53 C55 C56 C57 C59 C60
10U 0.1U 1N 10U 0.1U 1N 1V9

2
10V 16V 50V 10V 16V 50V DS39
+80%-20% +80%-20% 10% +80%-20% +80%-20% 10% HSMG-C190
GREEN

1
C
R370
B GND
Q5
GND GND 150 MMBT3904
E PLACE THESE AROUND THE BOARD
0.0625W
R372
1%
LAYOUT NOTES:
1K
1. PLACE THE SMALL VALUE CAPS AS CLOSE AS POSSIBLE TO THEIR POWER PINS.
0.1W
2. IF NOT ON A PLANE, THE TRACE BETWEEN THE BULK AND SMALL CAPS SHOULD BE 20 MIL MIN.
1%

GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:04 2005 power_supply 2 of 2 5138201SCH 2 26 of 40


1 2 3 4 5 6 7 8

Figure 5-36 5138201SCH_s26_r2

Section 3.0 — Corona 2 Board, 5138201 Page 142


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

USER_PERIPH_DATA[15..0]
IO
USER_PERIPH_ADR[18..0]
OUT
USER_PERIPH_CTL[8..0]
IO
CC2_TDO
OUT
I38 I55
CC2_DSP CC2_PWR_ISO_CLK
FPGA_WD
IN
A USER_PERIPH_DATA<15..0> A
FPGA_RS_REQ
USER_PERIPH_ADR<18..0> IN
DSP_UART[1..0] FPGA_CLK
IO DSP_UART<1..0> USER_PERIPH_CTL<8..0> FPGA_WD FPGA_CLK OUT
CC2_TDO FPGA_RS_REQ
DSP_WD CAN_RX_FROM_XCVR
DSP_WD DSP_WD CAN_RX_FROM_XCVR OUT
ADC_IN[9..0] DSP_RS_REQ
IN ADC_IN<9..0> DSP_RS_REQ DSP_RS_REQ
TDI_DSP
TDI_DSP TDI_DSP
DSP_GPIO_A[15..0]
IO DSP_GPIO_A<15..0>
DSP_XINT_BUS[2..0] CPLD2_GPIO[16..0]
IO DSP_XINT_BUS<2..0> CPLD2_GPIO<16..0> IO
CAN_SETUP[7..0]
CAN_SETUP<7..0> CAN_SETUP<7..0>
CAN_RX_TO_DSP
IN CAN_RX_TO_DSP
CAN_TX_FROM_DSP CAN_TX_TO_XCVR
CLK_RST<5..0> CAN_TX_FROM_DSP OUT IN CAN_TX_TO_XCVR
TMS<7..0>
SB_STATUS_CTRL<6..0> TCK<7..0>

TI_JTAG_SEL
TMS<7..0>
TCK<7..0>
TXA_EN_N
TRST_N<3..0> TXA_EN_N TXA_EN_N
RXA_EN_N CAN_BUS[9..1]
RXA_EN_N RXA_EN_N CAN_BUS<9..1> IO
CAN3V3_OFF_N
CAN3V3_OFF_N CAN3V3_OFF_N CLK_RST<5..0>
EN_3V3_OFF_N
TI_JTAG<8..0> EN_3V3_OFF_N EN_3V3_OFF_N
B DSP_CLK B
CAN3V3_TST_N ALT_CLK ALT_CLK DSP_CLK
TDI_CPLD2 CAN_3V3TST_N
CAN3V3_TST_N
DSP_CLK

TMS[7..0]
OUT
TCK[7..0]
OUT
TRST_N[3..0]
OUT
TI_JTAG[8..0]
IO

0
EXT_JTAG[5..0]
IO
EXT_DIN[3..0]
IN
EXT_DOUT[3..0]
OUT
CF_JTAG[3..0] 2 CLOCK_FAIL_N
IO OUT
C C
ALT_CLK 1 USER_RST_N
OUT OUT
I65
CC2_SYSTEMBIST
CLK_RST[5..0]
CF_JTAG<3..0> CLK_RST<5..0>
CONFIG[5..0]
EXT_DOUT<3..0> CONFIG<5..0> IO
FLASH_DATA[15..0]
EXT_DIN<3..0> FLASH_DATA<15..0> IO
FLASH_CTL[6..0]
EXT_JTAG<5..0> FLASH_CTL<6..0> IO
FLASH_ADDR[24..0]
FLASH_ADDR<24..0> IO

TDI_CPLD2
CC2_TDI
TRST_N<3..0> CC2_TDI IN
TCK<7..0>
TMS<7..0> EMU1_FROM_SB
TI_JTAG_SEL
TI_JTAG_SEL
SB_STATUS_CTRL[6..0]
SB_STATUS_CTRL<6..0>
DIN[3..0] CLK_RST[5..0]
IN DIN<3..0>
MEM_INT_OE
IN MEM_INT_OE
CC2_DECOUPLING CAN CORE 2 REVISION LOG

I35 REVISION DATE ENGINEER DESCRIPTION OF CHANGES


D D
1 5/13/2005 PAUL MONROE INITIAL RELEASE.

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Wed Aug 31 08:53:26 2005 cc2_hierarchy 1 of 1 5138201SCH 2 27 of 40


1 2 3 4 5 6 7 8

Figure 5-37 5138201SCH_s27_r2

Chapter 5 - System Cabinet Page 143


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= TMS320F2812 DSP (1 OF 2) =======

DIAG_VOLTAGES[6..0]

A A

3
0
2
4
5

0
1

1
3V3 3V3 3V3 3V3 3V3 3V3 3V3

2
ADC_IN[9..0] K K K K K K K
IN
TO USER CIRCUITRY
D19 D6

3
8
9

2
4
5
6
7
1
3 3 3 3 3 3 3
3V3 AK AK D1 AK D2 AK D3 AK D4 AK D5 AK BAT54S

D4
C4

D5

C2
C3

D3
D2
A3

B4
A4
E5
B5

F5
C1

D1
30V

B1
1V9 A A A A A A A

1
ADCINA0
ADCINA2
ADCINA3
ADCINA4
ADCINA5
ADCINA6
ADCINA7
ADCINB0
ADCINB2
ADCINB3
ADCINB4
ADCINB5
ADCINB6
ADCINB7
ADCINA1

ADCINB1
FL5
2

BLM21PG221SN1 C244 C245


220 OHM R422 10U 10U GND
3V3_A E1 AVDDREFBG ADCREFP E2 20K 10V 10V
FL4 VDDA<1-0> ADCREFM E4 0.0625W +80%-20% +80%-20%
BLM21PG221SN1 B2 VDDAIO ADCRESEXT F2 1%
1

220 OHM N8 VDD3VFL


VDD<9-0>
VDDIO<5-0> GND_ANALOG
A6 A
VDD1
1

R426
DSP_CLK K9 X1_XCLKIN XCLKOUT F11 ALT_CLK
OUT
B M9 X2 R395
56
B
A13 TESTSEL TDO D12 DSP_TDO
C238 C239 C240 C241 C242 C243 DSP_RESET_N D6 0.0625W
XRS_N
1U 0.1U 10N 1U 0.1U 10N TEST ONLY M7 TEST1 XA0 G5 56 1% 0.0625W
R396 0 1%
10V 16V 16V 10V 16V 16V TEST ONLY N7 TEST2 XA1 M2 R397 1
10% +80%-20% 10% 10% +80%-20% 10% TEST ONLY E6 ADCBGREFIN XA2 N10 R398 2
GND XA3 M11 R399 3 DSP_JTAG[6..0]
B3 ADCLO XA4 J12 R400 4
U36 XA5 H13 R401 5
R377 DSP_TMS D13 TMS TMS320F2812GHHA XA6 G14 R402 6
DSP_TDI C13 TDI XA7 F12 R403 7 DSP_TDO 6
DSP_TCK A12 TCK XA8 E12 R404 8 DSP_TMS 2
0 DSP_TRSTN B12 D14 9 DSP_TDI 5
TRST_N XA9 R405
0.125W DSP_EMU0 D11 C12 10 DSP_TCK 3
EMU0 XA10 R406
0% DSP_EMU1 C9 A14 11 DSP_TRSTN 4
EMU1 XA11 R407
A C11 12 DSP_EMU0 1
XA12 R408
GND GND_ANALOG 0 R379 G3 XD0 XA13 E10 R409 13 DSP_EMU1 0
JOIN GNDS AT ONE POINT 1 R380 H3 XD1 XA14 A10 R410 14
2 R381 H5 XD2 XA15 B9 R411 15
PLACE R AT CORNER OF BGA 3 R382 J3 XD3 XA16 A8 R412 16
DO NOT ROUTE DIGITAL SIGNALS 4 R383 K3 XD4 XA17 B7 R413 17
UNDER/OVER THE AGND ISLAND 5 R384 J5 XD5 XA18 D7 R414 18
6 R385 L3 XD6
C 7 R386 L5 XD7 XHOLDA_N K10 R415 0 USER_PERIPH_ADR[18..0] C
OUT
DSP_CLK 8 R387 P7 XD8 XZCS0AND1_N P1 R416 1 TO USER CIRCUITRY
IN
9 R388 M8 XD9 XZCS2_N P13 R417 2
10 R389 L9 XD10 XZCS6AND7_N B13 R418 3
11 R390 N9 XD11 XWE_N N11 R419 4
12 R391 L14 XD12 XRD_N M3 R420 5
CLK_RST[5..0] 13 R392 J10 XD13 XR_NW N4 R421 6
IO
R376 14 R393 B11 XD14
1 DSP_RESET_N 15 A9 E3 56
R394 XD15 AVSSREFBG R423
0.0625W 1
VSSA<1-0>
1K 56 1% 0.0625W F1 A2 1%
XMP_NMC VSSAIO
0.0625W 8 E7 C6
XHOLD_N VSS1 R424
1% 7 B6 3V3 2
XREADY VSS<15-0>
RESISTOR TO PREVENT CONTENTION R378
10K SYM 1 OF 2
IN CASE THE DSP WATCHDOG DRIVES R425
0.0625W A 3
THE RESET LINE LOW WHEN SYSTEM
BIST IS DRIVING IT INACTIVE. 1% GND GND_ANALOG
10K
0.0625W
1%

GND

USER_PERIPH_CTL[8..0]
IO
TO USER CIRCUITRY
D D
USER_PERIPH_DATA[15..0]
IO
TO USER CIRCUITRY

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:10 2005 cc2_dsp 1 of 3 5138201SCH 2 28 of 40


1 2 3 4 5 6 7 8

Figure 5-38 5138201SCH_s28_r2

Section 3.0 — Corona 2 Board, 5138201 Page 144


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= TMS320F32812 DSP (2 OF 2) ======= ======= VOLT / TEMP MONITOR =======


U38 DIAG_VOLTAGES[6..0]
A LM60 A
3V3
1V5
DSP_XINT_BUS[2..0] 1 VSP VOUT 2 0
IN
2 GND 3
1 C246 C247 R449
CAN3V3_TST_N 0 R443 0.1U 1U 1K
IN
R439 R441 1K 16V 10V
1K 1K 0.0625W +80%-20% 10% 1%
1% 1
1% 1%
C251
GND 1U
U36 GND 10V 1V9
TMS320F2812GHHA 10%

DSP_GPIO_A[15..0] GND 5V0 R452


IO
0 M12 GPIOA0_PWM1 GPIOD0_NT1CTRIP_PDPINTA H14 3V3 1K
1 M14 GPIOA1_PWM2 GPIOD1_NT2CTRIP_NEVASOC G10 GND
2 L12 GPIOA2_PWM3 R445 1%
3 L13 GPIOA3_PWM4 GPIOD5_NT3CTRIP_PDPINTB P10 1.5K 2
4 K11 GPIOA4_PWM5 GPIOD6_NT4CTRIP_NEVBSOC P11 0 R437 R440 R442
5 K14 GPIOA5_PWM6 1 10K 10K 10K 1% C252
6 J11 GPIOA6_T1PWM_T1CMP GPIOE0_XINT1_NXBIO D9 2 0.0625W 3 1U
7 J13 GPIOA7_T2PWM_T2CMP GPIOE1_XINT2_ADCSOC D8 1% 10V
8 H10 E8 U37 M25P40 10%
GPIOA8_CAP1_QEP1 GPIOE2_XNMI_XINT13
B 9 H11 GPIOA9_CAP2_QEP2 R446 C248 12V0 B
10 H12 GPIOA10_CAP3_QEPI1 R435 1K 1U
11 F14 M1 56 5 10V
GPIOA11_TDIRA GPIOF0_SPISIMOA D R450
12 F13 GPIOA12_TCLKINA GPIOF1_SPISOMIA N1 R436 6 C 1% 10% 4.7K GND
13 E13 K2 56
GPIOA13_NC1TRIP GPIOF2_SPICLKA R444
14 E11 GPIOA14_NC2TRIP GPIOF3_SPISTEA K4 1 S_N Q 2 1%
15 F10 GPIOA15_NC3TRIP 3 W_N 4
CAN_SETUP[7..0] C7 7 33 GND GND
IN GPIOF4_SCITXDA HOLD_N
0 N2 A7 0.0625W
GPIOB0_PWM7 GPIOF5_SCIRXDA
1 P2 1%
GPIOB1_PWM8
2 N3 GPIOB2_PWM9 GPIOF6_CANTXA N12 VCC=3V3;GND=GND C250 R451
3 P3 GPIOB3_PWM10 GPIOF7_CANRXA N13 24V0_IN1 1U 1K
4 L4 GPIOB4_PWM11 4 MBIT SERIAL FLASH 10V
5 M4 GPIOB5_PWM12 GPIOF8_MCLKXA J1 FOR DSP USAGE 10% 1%
6 K5 GPIOB6_T3PWM_T3CMP GPIOF9_MCLKRA H2 R447
7 N5 GPIOB7_T4PWM_T4CMP GPIOF10_MFSXA H4 10K
6 TST_DONE M5 GPIOB8_CAP4_QEP3 GPIOF11_MFSRA J2 24V0_IN2
0 TEST_SEL1_N M6 GPIOB9_CAP5_QEP4 GPIOF12_MDXA G1 1% GND GND
1 TEST_SEL2_N P6 GPIOB10_CAP6_QEPI2 GPIOF13_MDRA G2 5
2 TEST_SEL4_N L8 GPIOB11_TDIRB R607
3 TEST_SEL8_N K8 GPIOB12_TCLKINB GPIOF14_XF_NXPLLDIS A11 PLL_EN_AND_LED_FB 3V3 10K
4 START_N N6 GPIOB13_NC4TRIP R434 R448 C249
5 PASS_N L6 P14 56 1K 1%
GPIOB14_NC5TRIP GPIOG4_SCITXDB 1U
C K7 GPIOB15_NC6TRIP GPIOG5_SCIRXDB M13 R438 10V 6 C
SYM 2 OF 2 10K 1% 10%

R427 1% C141 R608


0 1U 1K
GND GND 10V
1K 10% 1%
R428
0.0625W 1 3V3
1%
DSP HB
R429
2 U60 R433 GND GND
GND 3V3 74LVC125A DS1 150
HSMG-C190 0.0625W TX 1
R430 R606
3 1K 1 1% RX 0
0.0625W OE_N 3 1 2
R431 1% 2 Y CAN_TX_FROM_DSP
OUT
4 A GREEN
VCC=3V3;GND=GND CAN_RX_TO_DSP
IN
R432
5 PLL_EN_AND_LED_FB TXA_EN_N
OUT
SB_STATUS_CTRL[6..0] 10K RXA_EN_N
IO 0.0625W OUT
1% EN_3V3_OFF_N
OUT
SYSTEM BIST CONTROL
AND STATUS PULL-UPS CAN3V3_OFF_N
OUT
D DSP_UART[1..0] D
IO DSP_RS_REQ
OUT

REV 2 01July04
DSP_WD
OUT

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:10 2005 cc2_dsp 2 of 3 5138201SCH 2 29 of 40


1 2 3 4 5 6 7 8

Figure 5-39 5138201SCH_s29_r2

Chapter 5 - System Cabinet Page 145


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= DSP JTAG MUX =======

TI_JTAG[8..0] DSP_JTAG[8..0]
IO
0 EMU1 0
A A
8 EMU0 1

TDI 5

TMS 2

TCK 3

TRSTN 4
U39
74LVC_LCX16244

TI_SEL_N 1 OE_N<3>

1 TI_TRSTN 47 A1<3> Y1<3> 2 R455


6 TI_TCK 46 A2<3> Y2<3> 3 R456
7 TI_TMS 44 A3<3> Y3<3> 5 R457
5 TI_TDI 43 A4<3> Y4<3> 6 R458

SB_SEL_N 48 OE_N<2>

3 SB_TRSTN 41 A1<2> Y1<2> 8 R459


6 SB_TCK 40 A2<2> Y2<2> 9 R460
6 SB_TMS 38 A3<2> Y3<2> 11 R461
B TDI_DSP 37 A4<2> Y4<2> 12 R462 B
TI_SEL_N 25 OE_N<1>

36 A1<1> Y1<1> 13 R463


UNUSED_PD 35 A2<1> Y2<1> 14
UNUSED_PD 33 A3<1> Y3<1> 16
32 A4<1> Y4<1> 17 R464

SB_SEL_N 24 OE_N<0>

UNUSED_PD 30 A1<0> Y1<0> 19


UNUSED_PD 29 A2<0> Y2<0> 20
UNUSED_PD 27 A3<0> Y3<0> 22 R465
26 A4<0> Y4<0> 23 CC2_TDO
OUT
56
0.0625W
VCC=3V3;GND=GND
1%
TDO 6 UNUSED_PD

R466
10K
C 0.0625W C
1%

3 TI_TDO

GND

TRST_N[3..0] 3V3
IN
TCK[7..0]
IN
TI JTAG ENABLE 3.3V TP
TMS[7..0]
IN
TP9 TP10 3V3 C253
TDI_DSP 0.1U
IN
TIN TIN 16V U40
+80%-20% 74LVC1G00

5 VCC
R454 GND
TI_JTAG_SEL 1 A
IN Y 4 TI_SEL_N
1K 2 B
R453 R609
1K 0.0625W 1K 3
GND
0.0625W 1% SB_SEL_N 0.0625W
1% 1%
D D
GND

REV 2 01July04
GND

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:11 2005 cc2_dsp 3 of 3 5138201SCH 2 30 of 40


1 2 3 4 5 6 7 8

Figure 5-40 5138201SCH_s30_r2

Section 3.0 — Corona 2 Board, 5138201 Page 146


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

3V3
======= SYSTEM BIST CPLD PAGE 1 =======
R498
10K
0.0625W
1%
MEM_INT_OE CLK_RST[5..0]
IN IO
A 3V3 A
U46
R505 EPM2210F256C4 1
EXT_DIN[3..0] 10K
IN
0.0625W
1% VCCINT<7..0> R519
VCCIO1<3..0> 1K EXT_DOUT[3..0]
OUT
VCCIO2<3..0> 0.0625W
VCCIO3<3..0> 1%
GND VCCIO4<3..0>

0 CPLD_CLK H5 IO_H5_GCLK0
J5 IO_J5_GCLK1 DOUT[3..0]
J12 IO_J12_GCLK2
5 CC2_RST_N H12 IO_H12_GCLK3 GND
M8 IO_M8_DEV_OE
CF_JTAG[3..0] M9 IO_M9_DEV_NCLR IO_B16 B16
IO
IO_C13 C13 SB_STATUS_CTRL[6..0]
0 IO
2 CF_TDI R506 L6 TDI IO_A15 A15
EXT_JTAG[5..0] 1 CF_TMS 0 N4 C12
IO R507 TMS IO_C12
3 CF_TCK 0 P3 B14
R508 TCK IO_B14
0 CF_TD0 R504 M5 TDO IO_D12 D12
CONFIG[5..0] IO_B13 B13 TST_DONE 6
IO 56 D3 IO_D3 IO_C11 C11
0.0625W 1%
B 1 INIT_DONE C2 IO_C2 IO_A13 A13 B
C3 IO_C3 IO_D11 D11 2
E3 IO_E3 IO_B12 B12 MEM_INT_OE
0 CONF_DONE D2 IO_D2 IO_E11 E11 3
E4 IO_E4 IO_A12 A12
3V3 3V3 3V3 2 STATUS_N D1 IO_D1 IO_C10 C10 USER_RST_N
E5 IO_E5 IO_B11 B11
E2 IO_E2 IO_D10 D10 1
R495 R496 R497 3V3 F3 IO_F3 IO_A11 A11
10K 10K 10K 4 EXT_TCK E1 IO_E1 IO_E10 E10 0
EXT_DOUT[0] F4 IO_F4 IO_B10 B10 START_N 4 TRST_N[3..0]
1% 1% 1% OUT
R499 2 EXT_TMS F2 IO_F2 IO_C9 C9 R520 R521 R522 R523
10K EXT_DOUT[1] F5 A10 0 CC2_TDI 1K
IO_F5 IO_A10 R509
2 0.0625W 3V3 F1 IO_F1 IO_D9 D9 0.0625W
1% F6 IO_F6 IO_B9 B9 R510 0 1%
0 G2 IO_G2 IO_E9 E9
R500 G3 IO_G3 IO_A9 A9 R511 2
5 10K G1 IO_G1 IO_A8 A8 R512 3
0.0625W EXT_DOUT[3] G4 IO_G4 IO_B8 B8 R513 1 GND GND GND GND
3 1% H2 IO_H2 IO_E8 E8 TMS[7..0]
OUT
EXT_DOUT[2] G5 IO_G5 IO_A7 A7 R514 2
0 EXT_PT_N H1 IO_H1 IO_D8 D8
5 EXT_TRST_N H3 IO_H3 IO_B7 B7 R515 3
C J1 IO_J1 IO_C8 C8 C
H4 IO_H4 IO_A6 A6 R516 0
J2 IO_J2 IO_E7 E7 DIN[0]
J4 IO_J4 IO_B6 B6 R517 1
K1 IO_K1 IO_D7 D7 DIN[1] TCK[7..0]
56 0.0625W 1% OUT
J3 IO_J3 IO_A5 A5 TEST_SEL1_N 1
3 EXT_TDI K2 IO_K2 IO_C7 C7
EXT_DIN[1] K5 IO_K5 IO_B5 B5 TEST_SEL3_N 3
1 EXT_TDO R501 L1 IO_L1 IO_E6 E6 DOUT[0]
EXT_DIN[0] K4 IO_K4 IO_A4 A4 TEST_SEL0_N 0
L2 IO_L2 IO_D6 D6 DOUT[1]
K3 IO_K3 IO_B4 B4 TEST_SEL2_N 2
5 CONFIG_N M1 IO_M1 IO_C6 C6
EXT_DIN[2] L5 IO_L5 IO_C4 C4
4 DATA0 R502 M2 IO_M2 IO_C5 C5
EXT_DIN[3] L4 IO_L4 IO_B3 B3
L3 IO_L3 IO_D5 D5
3 DCLK R503 N1 IO_N1 IO_A2 A2
M4 IO_M4 IO_B1 B1 FAIL_PASS_N 5
56 N2 D4
IO_N2 IO_D4
0.0625W M3 IO_M3
DIN[3..0] 1% N3 DS43
IN IO_N3 GNDINT<7..0>
P2 HSMH-C190
IO_P2 GNDIO<15..0>
R518
SYM 1 OF 2 1 2
150
D GND RED
0.0625W
D
BIST FAIL
CC2_TDI 1%
IN

REV 2 01July04
GND

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:15 2005 cc2_systembist 1 of 3 5138201SCH 2 31 of 40


1 2 3 4 5 6 7 8

Figure 5-41 5138201SCH_s31_r2

Chapter 5 - System Cabinet Page 147


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= SYSTEM BIST CPLD PAGE 2 =======

DIN[3..0]

A A
FLASH_DATA[15..0]
IO
11 R524
6 R525
4 R526
3 R527
1 R528

U46
EPM2210F256C4 TMS[7..0]

13 R529 P14 IO_P14 IO_R1 R1


P15 IO_P15 IO_P4 P4
N13 IO_N13 IO_T2 T2 R569 4
8 R530 N14 IO_N14 IO_P5 P5 TCK[7..0]
10 R531 M14 IO_M14 IO_R3 R3 R570 5
14 R532 N15 IO_N15 IO_N5 N5
9 R533 M13 IO_M13 IO_R4 R4
15 R534 N16 IO_N16 IO_P6 P6 R571 6
2 R535 L14 IO_L14 IO_T4 T4
12 R536 M15 IO_M15 IO_N6 N6 DOUT[3]
2 MEM_OE_N R537 L13 IO_L13 IO_R5 R5 R572 4
7 R538 M16 IO_M16 IO_M6 M6 DOUT[2]
L12 IO_L12 IO_T5 T5
B 5 R539 L15 IO_L15 IO_P7 P7 B
L11 IO_L11 IO_R6 R6 R573 5
24 R540 L16 IO_L16 IO_N7 N7 DIN[3]
0 R541 K14 IO_K14 IO_T6 T6
0 R542 K15 IO_K15 IO_M7 M7 DIN[2]
1 MEM_CE_N R543 K13 IO_K13 IO_R7 R7
16 R544 K16 IO_K16 IO_P8 P8 R574 7
K12 IO_K12 IO_T7 T7 R575 6
J15 IO_J15 IO_N8 N8
11 R545 J14 IO_J14 IO_R8 R8
15 R546 J16 IO_J16 IO_N9 N9
19 R547 J13 IO_J13 IO_T8 T8 R576 7
23 R548 H16 IO_H16 IO_T9 T9
1 H13 R9 56
R549 IO_H13 IO_R9
20 H15 P9 0.0625W
R550 IO_H15 IO_P9
5 H14 T10 1%
R551 IO_H14 IO_T10
14 R552 G16 IO_G16 IO_M10 M10 MEM_BYTE_N 6
G12 IO_G12 IO_R10 R10
10 R553 G15 IO_G15 IO_N10 N10
G13 IO_G13 IO_T11 T11 R577 TDI_CPLD2
OUT
5 MEM_RST_N R554 F16 IO_F16 IO_P10 P10
21 G14 R11 56
R555 IO_G14 IO_R11
6 F15 M11 0.0625W DOUT[3..0]
R556 IO_F15 IO_M11
C F11 IO_F11 IO_T12 T12 1% C
8 R557 E16 IO_E16 IO_N11 N11 R611
F12 IO_F12 IO_R12 R12 MEM_WP_N 4 3 EMU1_FROM_SB
OUT
12 R558 E15 IO_E15 IO_P11 P11
18 F13 T13 1% 330 0.0625W
R559 IO_F13 IO_T13
13 R560 D16 IO_D16 IO_M12 M12 2 TI_JTAG_SEL
OUT
2 R561 F14 IO_F14 IO_R13 R13
3 MEM_WE_N R562 D15 IO_D15 IO_N12 N12
4 R563 E12 IO_E12 IO_R14 R14 R578
7 R564 D14 IO_D14 IO_P12 P12 1
17 R565 E13 IO_E13 IO_T15 T15

2
9 C15 R16 150
R566 IO_C15 IO_R16 DS44 SB SAFE
0 MEM_BSY_N C14 P13 0.0625W
IO_C14 IO_P13 HSMY-C190
22 E14 1%
R567 IO_E14 YELLOW
3 R568 D13 IO_D13

1
FLASH_ADDR[24..0] 56 SYM 2 OF 2
IO 0.0625W
1% GND

R579
FLASH_CTL[6..0] 0
IO

2
150
DS45 SB NORMAL
0.0625W
HSMG-C190
1%
GREEN
D D

REV 2 01July04
GND

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:16 2005 cc2_systembist 2 of 3 5138201SCH 2 32 of 40


1 2 3 4 5 6 7 8

Figure 5-42 5138201SCH_s32_r2

Section 3.0 — Corona 2 Board, 5138201 Page 148


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= 512MBIT FLASH MEMORY =======

A A

3V3

R583
10K
0.0625W
1%
MEM_BUSY_N

3V3

C265 C266 C267 U47


10N 10N 10N S29GL512N10FAI01
16V 16V 16V
10% 10% 10%
G5 VCC

B D8 VIO1 B
F1 VIO2
0 GND
1 F2 CE_N RY_BY_N A4
2 G2 OE_N
3 A5 WE_N
4 B4 NWP_ACC
5 B5 RESET_N
6 F7 BYTE_N
FLASH_CTL[6..0]
0 E2 A0 DQ0 E3 R584 0
1 D2 A1 DQ1 H3 R585 1
3V3 3V3 R582 2 C2 A2 DQ2 E4 R586 2
1K 3 A2 A3 DQ3 H4 R587 3
0.0625W 4 B2 A4 DQ4 H5 R588 4
1% 5 D3 A5 DQ5 E5 R589 5
R580 R581 6 C3 A6 DQ6 H6 R590 6
10K 10K 7 A3 A7 DQ7 E6 R591 7
0.0625W 0.0625W 8 B6 A8 DQ8 F3 R592 8
1% 1% 9 A6 A9 DQ9 G3 R593 9
GND 10 C6 A10 DQ10 F4 R594 10
11 D6 A11 DQ11 G4 R595 11
12 B7 A12 DQ12 F5 R596 12
C 13 A7 A13 DQ13 G6 R597 13 C
14 C7 A14 DQ14 F6 R598 14
15 D7 A15 DQ15_A_1 G7 R599 15
16 E7 A16
17 B3 56
A17
18 C4 0.0625W FLASH_DATA[15..0]
A18
19 D5 1%
A19
20 D4 A20
21 C5 A21
22 B8 A22
23 C8 A23
24 F8 A24 GND<2..0>

FLASH_ADDR[24..0]
GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:16 2005 cc2_systembist 3 of 3 5138201SCH 2 33 of 40


1 2 3 4 5 6 7 8

Figure 5-43 5138201SCH_s33_r2

Chapter 5 - System Cabinet Page 149


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= CAN INTERFACE HARDWARE =======

A A

NOTE: THE CORONA2 DOES NOT REQUIRE THE


IN EN_3V3_OFF_N
ISOLATION PROVIDED BY THE GENERIC CAN CORE.
CAN3V3_OFF_N
IN
THUS, THE CIRCUITRY HAS BEEN DELETED.
IN TXA_EN_N THE PORTS ARE LEFT TO ELIMINATE ERRORS.
CAN_TX_TO_XCVR
IN
CAN_BUS<9..1>
IO
IN RXA_EN_N

CAN_3V3TST_N
OUT
B B
CAN_RX_FROM_XCVR
OUT

C C

S2 S3
219-04MST 3V3 219-04MST 3V3 SWITCH FUNCTIONS

NODE_ID(6..1) - SELECT THE CANOPEN NODE ID.


3
7

6
5

0
1

0 NODE_ID(1) 8 O1 I1 1 4 NODE_ID(5) 8 O1 I1 1 NOTE THAT NODE_ID(0) IS 0.

1 NODE_ID(2) 7 O2 I2 2 5 NODE_ID(6) 7 O2 I2 2 BAUD_RATE(1..0) - SELECT THE CAN BAUD RATE.


00 = 125KBPS
R474 2 NODE_ID(3) 6 O3 I3 3 6 BAUD_RATE(0) 6 O3 I3 3 01 = 250KBPS
R467 R469 R471 R473 1K 10 = 500KBPS
1K 1K 1K 1K 0.0625W 3 NODE_ID(4) 5 O4 I4 4 7 BAUD_RATE(1) 5 O4 I4 4 11 = 1MBPS
1% 1% 1% 1% 1%
D D
CAN_SETUP[7..0]
OUT

REV 2 01July04
R468 R470 R472
1K 1K 1K
1% 1% 1%

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

GND cc2_pwr_iso_clk 1 of 3 5138201SCH 2 34 of 40


Fri Sep 02 14:45:13 2005

1 2 3 4 5 6 7 8

Figure 5-44 5138201SCH_s34_r2

Section 3.0 — Corona 2 Board, 5138201 Page 150


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

A A
======= POWER SUPPLIES =======

NOTE: THE 3V3 ISOLATED SUPPLY CONTAINED IN THE


GENERIC CAN CORE 2 HAS BEEN
REMOVED IN THE CORONA IMPLEMENTATION.

B B

3V3

3V3
R476
GND TP 10K
0.0625W
TP11 1%

TIN CC2_RST_N 5 CLK_RST[5..0]


C R475 C255 C
1K +80%-20% 3V3 VR1
0.0625W 0.1U TPS7301QD
1% U41 16V U42 TP12
1V9
TPS3823 74LVC1G00
RESET_N 8 TIN
GND
5 VDD 5 VCC 3 VIN1 VOUT1 5 R481
4 VIN2 VOUT2 6 R478 1K
4 1 1 A 0.0625W
WDI RESET_N 102K
3 Y 4 2 7 1%
MR_N EN_N SENSE_FB
2 2 B
GND
3 1 1%
RST_REQ_N GND GND R477
C256 169K
3.3V 0.1U 1%
C254 S4 16V 1.2V TO 9.75V C257
0.1U KT11B1JM +80%-20% 10U
16V GND 10V
+80%-20% MOUNT OVER CU ISLAND +80%-20%
1 1 3 3 WITH HEAT SINK COMPOUND
SURFACE AREA = 10 SQ. CM.GND
R610
2 2 4 4 1K GND VIAS ON 0.1 CTRS ALONG
R479 R480
0.0625W PERIPHERY CONNECTED TO ALL 1 1
1% DIGITAL GROUND PLANES 0.1W 0.1W
1% 1%
GND MANUAL RESET GND
D D
TMS320 CORE SUPPLY

REV 2 01July04
GND

GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:13 2005 cc2_pwr_iso_clk 2 of 3 5138201SCH 2 35 of 40


1 2 3 4 5 6 7 8

Figure 5-45 5138201SCH_s35_r2

Chapter 5 - System Cabinet Page 151


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= MAX3064 CPLD FOR CLOCK AND WATCHDOGS =======

3V3

A A
R490
TDI_DSP
OUT
C259 C260 C261 C262
56 0.0625W 1% 5
10N 10N 10N 10N
16V 16V 16V 16V
10% 10% 10% 10% U45 CLK_FAIL 2
EPM3064A_44-10
0

VCCIO<1-0> CLK_RST[5..0]
IO
GND VCCINT<1-0>
R493
TDI_CPLD2 1 TDI_IO
IN 56 1%
7 7 TMS_IO IO_18 18
TMS[7..0] 7 26 19 5 0.0625W
IN TCK_IO IO_19
32 TDO_IO IO_20 20 FPGA_CLK
OUT
TCK[7..0] IO_21 21 RST_REQ_N
IN
38 IN_OE1 IO_22 22 6 CONFIGURE RST_REQ_N I/O PIN
ALT_CLK 40 IN_OE2_GCLK2 IO_23 23 7 AS OPEN COLLECTOR
IN
CPLD2_CLK 37 IN_GCLK1 IO_25 25 8
RESET_N 39 IN_GCLR IO_27 27 9
IO_28 28 10
FPGA_WD 2 IO_2 IO_31 31 11
IN
3 IO_3 IO_33 33 12
FPGA_RS_REQ 5 IO_5 IO_34 34 13
IN DS42
B 6 IO_6 IO_35 35 14
HSMG-C190
B
DSP_WD 8 IO_8 IO_42 42 15
IN
0 10 IO_10 IO_43 43 16
DSP_RS_REQ 1 12 IO_12 IO_44 44 2 1
IN
2 13 IO_13
3 14 IO_14 GNDINT<1-0> GREEN R494
R483 R484 4 15 IO_15 GNDIO<3-0> CPLD2 HB 150
1

1K 1K RESET DS41 0.0625W


0.0625W 0.0625W (RED WHEN RESET HSMH-C190 1%
1% 1% IS ACTIVE) RED

CLK_CHK
2

R488 R489
GND GND 150 10K GND GND
0.0625W 0.0625W
1% 1% CPLD2_GPIO[16..0]
IO

GND GND

R485 1V9
56
C 0.0625W C
1%
C263 U43
10N 74AUC1G00
16V
10%
3V3 5 VCC

1 A
R492
GND Y 4 DSP_CLK
R486
2

B OUT
2
30MHZ CLOCK 56
GND 3 56
0.0625W
FL6 0.0625W 1%
BMB0603-A-121 1%
V Y1
S1803B-30.0000 3V3 GND
1

4 VCC VOUT 3 U44


R482 C264 74LVC1G00
1 2 10N
ENABLE GND 16V
1K 10% 5 VCC
C258 0.0625W
30MHZ
16V 1% 1 A
R491
Y
D 0.1U R487 GND
B
4 CLK_RST[0] D
+80%-20% GND 2
3 56
GND
56 0.0625W

REV 2 01July04
0.0625W 1%
1%
GND
GND GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:13 2005 cc2_pwr_iso_clk 3 of 3 5138201SCH 2 36 of 40


1 2 3 4 5 6 7 8

Figure 5-46 5138201SCH_s36_r2

Section 3.0 — Corona 2 Board, 5138201 Page 152


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= CAN CORE 2 BYPASS CAPACITORS =======

A A

BANK 1 BANK 2
3V3 3V3 3V3 3V3 3V3 3V3 3V3 3V3
3V3 3V3 3V3 3V3 3V3 3V3

C214 C217 C220 C223 C227 C230 C233 C236


C195 C197 C200 C202 C205 C207 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U DSP VDDIO PINS 16V 16V 16V 16V 16V 16V 16V 16V
16V 16V 16V 16V 16V 16V

B 3V3 3V3 3V3 3V3 3V3 3V3 3V3 3V3 B


C215 C218 C221 C224 C228 C231 C234 C237
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
1V9 1V9 1V9 1V9 1V9 1V9 1V9 1V9 1V9 1V9 16V 16V 16V 16V 16V 16V 16V 16V

C196 C198 C201 C203 C206 C208 C209 C210 C212 C213
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V 16V 16V 16V 16V 16V BANK 3 VCCIO PINS BANK 4

CPLD BYPASS
DSP VDD PINS 3V3 3V3 3V3 3V3 3V3 3V3 3V3 3V3

C216 C219 C222 C225 C226 C229 C232 C235


C 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
C
16V 16V 16V 16V 16V 16V 16V 16V

3V3 M25P40 3V3 74LVC125A


BYPASS BYPASS VCCINT PINS

C199 C204
0.1U 0.1U
16V 16V
+80%-20% +80%-20%

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:08 2005 cc2_decoupling 1 of 1 5138201SCH 2 37 of 40


1 2 3 4 5 6 7 8

Figure 5-47 5138201SCH_s37_r2

Chapter 5 - System Cabinet Page 153


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

5V0
======= DECOUPLING CAPACITORS =======
C61 C67 C73 C79 C85 C91 C96 C101 C106 C111 C116 C124 C126 C131
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V
+80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20%

A A

GND

3V3

C62 C68 C74 C80 C86 C92 C97 C102 C107 C112 C117 C121 C127 C132 C137 C142 C146 C150 C153
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V
+80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20%

GND

3V3

C63 C69 C75 C81 C87 C93 C98 C103 C108 C113 C118 C122 C128 C133 C138 C143 C147 C151 C154
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V
+80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20%
B B
GND

3V3

C64 C70 C76 C82 C88 C94 C99 C104 C109 C114 C119 C123 C129 C134 C139 C144 C148 C152 C155
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V
+80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20%

GND

C C

1V5

C65 C71 C77 C83 C89 C95 C100 C105 C110 C115 C120 C125 C130 C135 C140
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V
+80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20%

GND

1V5

C66 C72 C78 C84 C90


0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V
+80%-20% +80%-20% +80%-20% +80%-20% +80%-20%
D D
GND

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:05 2005 decoupling 1 of 3 5138201SCH 2 38 of 40


1 2 3 4 5 6 7 8

Figure 5-48 5138201SCH_s38_r2

Section 3.0 — Corona 2 Board, 5138201 Page 154


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= DECOUPLING CAPACITORS =======

A A

5V0

C156 C158 C160


0.1U 0.1U 0.1U
16V 16V 16V
+80%-20% +80%-20% +80%-20%

GND

B B

3V3

C157 C159 C161 C163 C165 C166 C167 C168 C169 C170 C171 C172 C173 C174 C175 C176 C177 C178 C179 C180
0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U 0.1U
16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V 16V
+80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20% +80%-20%

C GND C

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:06 2005 decoupling 2 of 3 5138201SCH 2 39 of 40


1 2 3 4 5 6 7 8

Figure 5-49 5138201SCH_s39_r2

Chapter 5 - System Cabinet Page 155


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

1 2 3 4 5 6 7 8

======= UNUSED DEVICES =======

A A

B B

======= BULK CAPACITANCE =======


N12V0
1V5 1V5 12V0
5V0
C192 C194
2

C186 C188 C189 C190 100U 100U


C181 C183 100U 100U 100U 100U 16V 16V
C 100U 100U 16V 16V 16V 16V 20% 20% C
16V 16V 20% 20% 20% 20%
20% 20%
1

GND
GND GND GND
GND

3V3 12V0

C182 C184 C185 C187


100U 100U 100U 100U C191 C193
16V 16V 16V 16V 0.1U 0.1U
20% 20% 20% 20% 16V 16V
+80%-20% +80%-20%

GND GND

D D

REV 2 01July04
GE Healthcare
Company Restricted

B Size
Block Path Block Sheet Dwg No Rev Sheet
rawing Print for complete definition.
HEET - FOR REFERENCE ONLY.

Fri Sep 02 14:45:06 2005 decoupling 3 of 3 5138201SCH 2 40 of 40


1 2 3 4 5 6 7 8

Figure 5-50 5138201SCH_s40_r2

Section 3.0 — Corona 2 Board, 5138201 Page 156


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 4.0 — Jedi X-ray Generator, 2374870


4.1 Overview

Figure 5-51 2374870SCH_s1_r0

Chapter 5 - System Cabinet Page 157


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.2 Jedi Power Module, 2375932

4.2.1 Generic IF Board, 2295467-3


State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Page 3 of 4

Figure 5-52 2295467-3ADW_s2_r1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 158


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Approved Document - 2295467-3SCH_r1.pdf Page 2 of 9

Figure 5-53 2295467-3SCH_s1_r1


Chapter 5 - System Cabinet Page 159
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Approved Document - 2295467-3SCH_r1.pdf Page 3 of 9

Figure 5-54 2295467-3SCH_s2_r1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 160


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Approved Document - 2295467-3SCH_r1.pdf Page 4 of 9

Figure 5-55 2295467-3SCH_s3_r1

Chapter 5 - System Cabinet Page 161


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Approved Document - 2295467-3SCH_r1.pdf Page 5 of 9

Figure 5-56 2295467-3SCH_s4_r1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 162


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Approved Document - 2295467-3SCH_r1.pdf Page 6 of 9

Figure 5-57 2295467-3SCH_s5_r1

Chapter 5 - System Cabinet Page 163


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Approved Document - 2295467-3SCH_r1.pdf Page 7 of 9

Figure 5-58 2295467-3SCH_s6_r1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 164


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Approved Document - 2295467-3SCH_r1.pdf Page 8 of 9

Figure 5-59 2295467-3SCH_s7_r1

Chapter 5 - System Cabinet Page 165


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
State: RELEASE - Document is released and under formal Change Control. Changes are subject to the ECR/ECO Process.
See the GEMS ePDM System to determine the status of this document.

Approved Document - 2295467-3SCH_r1.pdf Page 9 of 9

Figure 5-60 2295467-3SCH_s8_r1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 166


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.2.2 AEC Board, 2222253-5

Figure 5-61 2222253-5ADW_s1_r1

Chapter 5 - System Cabinet Page 167


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-62 2222253-5SCH_s1_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 168


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-63 2222253-5SCH_s2_r0

Chapter 5 - System Cabinet Page 169


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-64 2222253-5SCH_s3_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 170


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-65 2222253-5SCH_s4_r0

Chapter 5 - System Cabinet Page 171


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-66 2222253-5SCH_s5_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 172


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.2.3 KV Control Board, 2382791-3

Figure 5-67 2382791-3ADW_s1_r0

Chapter 5 - System Cabinet Page 173


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-68 2382791-3ADW_s2_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 174


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-69 2382791-3SCH_s1_r0

Chapter 5 - System Cabinet Page 175


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-70 2382791-3SCH_s2_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 176


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-71 2382791-3SCH_s3_r0

Chapter 5 - System Cabinet Page 177


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-72 2382791-3SCH_s4_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 178


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-73 2382791-3SCH_s5_r0

Chapter 5 - System Cabinet Page 179


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-74 2382791-3SCH_s6_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 180


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-75 2382791-3SCH_s7_r0

Chapter 5 - System Cabinet Page 181


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-76 2382791-3SCH_s8_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 182


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-77 2382791-3SCH_s9_r0

Chapter 5 - System Cabinet Page 183


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-78 2382791-3SCH_s10_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 184


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-79 2382791-3SCH_s11_r0

Chapter 5 - System Cabinet Page 185


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-80 2382791-3SCH_s12_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 186


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-81 2382791-3SCH_s13_r0

Chapter 5 - System Cabinet Page 187


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-82 2382791-3SCH_s14_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 188


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-83 2382791-3SCH_s15_r0

Chapter 5 - System Cabinet Page 189


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-84 2382791-3SCH_s16_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 190


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-85 2382791-3SCH_s17_r0

Chapter 5 - System Cabinet Page 191


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-86 2382791-3SCH_s18_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 192


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-87 2382791-3SCH_s19_r0

Chapter 5 - System Cabinet Page 193


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-88 2382791-3SCH_s20_r0

Section 4.0 — Jedi X-ray Generator, 2374870 Page 194


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-89 2382791-3SCH_s21_r0

Chapter 5 - System Cabinet Page 195


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.2.4 Inverter Power Assembly, 2268972

Figure 5-90 2268972ADW_s1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 196


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

RESONANT CIRCUIT POWER COMPONENTS


(detail area of sheet 2290172SCH and 2290173SCH)

M+
DC Bus TB+
(from AC/DC Board)
Snubber
DC Bus Inductor
Capacitor 0.5 uH
100 uF
Note 2
Note 1

IGBT
Note 3

2256018 (Mid-point Cable) (White) HV Primary 2


(White) HV Primary 4
Resonant Capacitor
3.75 uF Note 1
1:80 Note 1 Jedi STD HV Tank
2186730
HV Primary 1
3.75 uF HV Primary 3
1:200 Note 1
Note 1

BLK YEL
IGBT
Note 3

9 uH 1:40 Note 1
2268974
DC Bus Snubber 2252099 Note 1
Capacitor Inductor BLK BLU
100 uF 0.5 uH
BLK GRN
Note 1 Note 2
I_INV_MEAS
M- To Gate CMD J1
DC Bus TB- ILP_MEAS
(from AC/DC Board)

Inverter Power assembly schematics


NOTES:
1 — Part of Capacitor Assembly part number 2268587 Drawing Number
2 — Part of Snubber Inductor Assembly part number 2272812 2268972 SCH rev 000
3 — IGBTs shown on Quad Snubber Schematic part number 2244697
Creation from
Rev 000, 8 Jun 2001 J. Soto 2290173SCH rev B Sheet 1 of 1

Figure 5-91 2268972SCH_s1_r0

Chapter 5 - System Cabinet Page 197


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.2.4.1 Quad Snubber Board, 2307323

Figure 5-92 2307323ADW_rBf

Section 4.0 — Jedi X-ray Generator, 2374870 Page 198


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-93 2307323SCH_s1

Chapter 5 - System Cabinet Page 199


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-94 2307323SCH_s2

Section 4.0 — Jedi X-ray Generator, 2374870 Page 200


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-95 2307323SCH_s3

Chapter 5 - System Cabinet Page 201


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.2.4.2 Gate Cmd Board, 2209547-3

Figure 5-96 2209547-3ADW_s1_rA

Section 4.0 — Jedi X-ray Generator, 2374870 Page 202


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-97 2209547-3ADW_s2_rA

Chapter 5 - System Cabinet Page 203


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-98 2209547-3SCH_s1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 204


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-99 2209547-3SCH_s2

Chapter 5 - System Cabinet Page 205


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-100 2209547-3SCH_s3

Section 4.0 — Jedi X-ray Generator, 2374870 Page 206


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-101 2209547-3SCH_s4

Chapter 5 - System Cabinet Page 207


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-102 2209547-3SCH_s5

Section 4.0 — Jedi X-ray Generator, 2374870 Page 208


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.3 HV Tank Assembly, 2186730

Figure 5-103 2186730ADW_s1_r2

Chapter 5 - System Cabinet Page 209


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-104 2186730SCH_s1_r2

Section 4.0 — Jedi X-ray Generator, 2374870 Page 210


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.3.1 KV Measure Board, 2184565

Figure 5-105 2184565ADW_s2_r3

Chapter 5 - System Cabinet Page 211


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-106 2184565SCH_s1_r1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 212


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.4 EMC and AC/DC Module, 2372292

Figure 5-107 2372292ADW_r1

Chapter 5 - System Cabinet Page 213


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.4.1 EMC Filter Board, 2209840-2


See related Drawing Print for complete definition.
DRAWING SHEET - FOR REFERENCE ONLY.

Figure 5-108 2209840-2ADW_s1_rA

Section 4.0 — Jedi X-ray Generator, 2374870 Page 214


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
elated Drawing Print for complete definition.

Figure 5-109 2209840-2SCH_s1_rA

Chapter 5 - System Cabinet Page 215


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-110 2209840-2SCH_s2_rA

Section 4.0 — Jedi X-ray Generator, 2374870 Page 216


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.4.2 AC/DC Board, 2300401-2 (schematic not available)

Chapter 5 - System Cabinet Page 217


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.5 Auxiliary Module, 2381787

4.5.1 Heater Board, 2271116

Figure 5-111 2271116SCH_s1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 218


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-112 2271116SCH_s2

Chapter 5 - System Cabinet Page 219


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-113 2271116SCH_s3

Section 4.0 — Jedi X-ray Generator, 2374870 Page 220


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-114 2271116SCH_s4

Chapter 5 - System Cabinet Page 221


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-115 2271116SCH_s5

Section 4.0 — Jedi X-ray Generator, 2374870 Page 222


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-116 2271116SCH_s6

Chapter 5 - System Cabinet Page 223


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-117 2271116SCH_s7

Section 4.0 — Jedi X-ray Generator, 2374870 Page 224


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-118 2271116SCH_s82242438-3ADW_r1.pdf

Chapter 5 - System Cabinet Page 225


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.5.2 LVPS Power Supply, 2212166-2


See related Drawing Print for complete definition.
DRAWING SHEET - FOR REFERENCE ONLY.

2212166-2ADW_s3_rA

Section 4.0 — Jedi X-ray Generator, 2374870 Page 226


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
See related Drawing Print for complete definition.
DRAWING SHEET - FOR REFERENCE ONLY.

Figure 5-119 2212166-2ADW_s2_rA

Chapter 5 - System Cabinet Page 227


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
See related Drawing Print for complete definition.
DRAWING SHEET - FOR REFERENCE ONLY.

Figure 5-120 2212166-2ADW_s3_rA

Section 4.0 — Jedi X-ray Generator, 2374870 Page 228


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-121 2212166-2SCH_s1_rA

Chapter 5 - System Cabinet Page 229


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
p
g

Figure 5-122 2212166-2SCH_s2_rA

Section 4.0 — Jedi X-ray Generator, 2374870 Page 230


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
p
g

Figure 5-123 2212166-2SCH_s3_rA

Chapter 5 - System Cabinet Page 231


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS
p
g

Figure 5-124 2212166-2SCH_s4_rA

Section 4.0 — Jedi X-ray Generator, 2374870 Page 232


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

4.5.3 Rotation Board, 2242438-3

Figure 5-125 2242438-3ADW_s1_r1f

Chapter 5 - System Cabinet Page 233


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-126 2242438-3SCH_s1

Section 4.0 — Jedi X-ray Generator, 2374870 Page 234


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-127 2242438-3SCH_s2

Chapter 5 - System Cabinet Page 235


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-128 2242438-3SCH_s3

Section 4.0 — Jedi X-ray Generator, 2374870 Page 236


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-129 2242438-3SCH_s4

Chapter 5 - System Cabinet Page 237


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-130 2242438-3SCH_s5

Section 4.0 — Jedi X-ray Generator, 2374870 Page 238


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-131 2242438-3SCH_s6

Chapter 5 - System Cabinet Page 239


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-132 2242438-3SCH_s7

Section 4.0 — Jedi X-ray Generator, 2374870 Page 240


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-133 2242438-3SCH_s8

Chapter 5 - System Cabinet Page 241


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-134 2242438-3SCH_s9

Section 4.0 — Jedi X-ray Generator, 2374870 Page 242


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-135 2242438-3SCH_s10

Chapter 5 - System Cabinet Page 243


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-136 2242438-3SCH_s11

Section 4.0 — Jedi X-ray Generator, 2374870 Page 244


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Figure 5-137 2242438-3SCH_s12

Chapter 5 - System Cabinet Page 245


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 5.0 — System Cabinet Bulkhead, 5134189

Figure 5-138 5134189DDW_s2_r2

Section 5.0 — System Cabinet Bulkhead, 5134189 Page 246


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Chapter 6 - Master Interconnect Schematic (MIS) and Charts


Section 1.0
Grounding & Wire Gauge Selection in Master Interconnect Schematics
1.1 Before You Begin

Some critical care areas require unique grounding to maintain patient safety and comply with local regulations. Please understand and comply with all local regulatory requirements before proceeding.

1.2 Equal-potential Grounding “General Rules”

An equal potential grounding system is generally the method of choice for meeting patient safety requirements. There are three rules for establishing an equal potential grounding system:
1.) Only one electrical ground point is permitted near the patient. Under no circumstances are two independent grounds allowed in the sensitive area around the patient. Patient Vicinity is defined as the space within 2.5 meters of the patient location.
2.) Any exposed, conductive surfaces located in the patient vicinity, that are likely to become energized, must be grounded to a reference ground point located at the table base.
3.) All grounding conductors must be #10 AWG minimum (Refer to Section 1.5.4 - Wire Gauge Guidelines For Field Supplied Power Runs) and 0.1 ohm maximum resistance between grounding points.

1.3 equal potential Grounding - “Recommendations”

There are a number of ways to establish an equal potential grounding system. Possibly the most convenient and inexpensive method is the following.
First, establish three grounding points:

• A System Reference, located at the power cabinet.

• A Patient Reference, located in the table base. The recommended Patient Reference is a ground terminal bus bar in the table base.

• A equal potential Terminal, located in the table base. The equal potential Terminal, unlike other ground points, has exposed connectors and is accessible to the operator.
Second, group equipment in the room by function. Establish a ground point in each equipment group:

• Create a protective Earth Terminal, located in a central component of each equipment group.
• Tie components within each equipment group to the Protective Earth Terminal. The Protective Earth Terminal is, in turn, tied to either the System Reference, or the Patient Reference.
• Tie the conductive surfaces of film changers, injectors, and other diagnostic equipment located in the patient vicinity to the Patient Reference, or the equal potential Terminal.
• Link all three major ground points, System Reference, Patient Reference, and equal potential Terminal, with insulated copper conductors whose size is determined by NEC resistance requirements and by the length of the required runs.
For more information on grounding, refer to Direction 46-014505, Electrical Safety - Equipment Grounding. Also check local codes.

1.4 Vendor “Add-on Components” - Ground Requirements

Component must keep separate signal and chassis grounds.

1.5 Grounding using the MIS Map

1.5.1 Purpose
Use the system MIS map to aid in the installation of ground runs for the basic system components and options. After interconnections are complete, verify adequacy of the grounding system by performing electrical leakage current checks per GE Healthcare
Direction 46-014546, Electrical Safety - Leakage Currents.

1.5.2 MIS Map Symbols


Symbols on the map are defined below. Typical requirements are included where appropriate.

Chapter 6 - Master Interconnect Schematic (MIS) and Charts Page 247


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

XX1
EQUIP.
NAME
Equipment - Cabinet, suspension, module, etc.
Protective Earth Terminal - Located in a central component of each equipment group. Terminal is provided for the permanent connection, at system installation, of protective earth conductors (safety ground runs).

Protective Earth Terminal Used as a Reference.


System Reference - Located in the LFX power cabinet. Also,
Patient Reference - Located in the table base.

Equal potential Terminal - Located in the table base. Terminal is provided for the connection of Equal potential grounding conductors to equipment that is temporarily in the patient vicinity.

Functional ground - A connection to the metal frame or enclosure of a component. Typical requirements on functional ground:
• Resistance between functional ground and cabinet power ground is 0.1 ohm max.
• Resistance between ground pin (if cord connected device) and any exposed metal parts on device is 0.1 ohm max.
• Green/yellow insulated wire is used for connections.

S
Signal Ground Terminal - Return ground point for internal component signals. Typical requirements on signal ground:
• Connect to Patient Reference or System Reference with #14 AWG. (Refer to Section 1.5.4 - Wire Gauge Guidelines For Field Supplied Power Runs.)
000
MIS Run (Field Supplied Wire) - Ground link contained within a MIS run. Typical requirements for ground wires within power distribution runs:
• Ground wire is at least as large as the supply conductors. Exception: #10 AWG ground is acceptable with #8 AWG supply conductors.
• Wire has green/yellow insulation.
• Ground wire is installed with, and follow the same path, as the supply conductors.
###
A
TYPE A
Discrete Safety Ground Run Conductor (Field Supplied Wire) - Links Protective Earth Terminals and metal surfaces in order to prevent shock hazards, and minimize leakage currents. Typical requirements:
• #10 AWG copper minimum.
• Wire has green/yellow insulation.
• Resistance of run is 0.1 ohm maximum. Size per Table 6-1.

COPPER WIRE MAXIMUM RUN LENGTH


AWG FEET METER
#10 66 20
#8 96 29
#6 140 43
#4 183 56
#2 243 74
#1/0 297 91
Table 6-1 Safety Grounds - Maximum Length

###
A
TYPE B
Discrete Signal Ground Run Conductor (Field Supplied Wire) - Links internal ground points in order to maintain a common signal reference. Typical requirements:
• #10 AWG copper minimum
• Insulation on wire is not green/yellow
• Resistance of run is 0.1 ohm maximum. Size per table.
###
A
TYPE C
Discrete Ground Run Conductor (Field Supplied Wire) that serves as both Safety and Signal ground - Typical requirements are the same as for Safety ground run conductor:
• #10 AWG copper minimum.
• Wire has green/yellow insulation.

Section 1.0 Grounding & Wire Gauge Selection in Master Interconnect Schematics Page 248
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

• Resistance of run is 0.1 ohm maximum. Size per table.


0000
MIS Run (Cable) - Ground link contained within a factory supplied MIS run.

1.5.3 Power Distribution


This section is intended to define power distribution wiring variations required by particular combinations of selected options.
R&F Positioner power distribution connections depend on the combination of equipment involved. Table 6-2 suggests connections for several possible combinations.

DEVICE SIZE - TYPE LOCATION


XT Suspension 10A - 120V - 1 pole CB8
SG-100 10A - 120V - 1 pole CB9
Spares CB10, CB11
Table 6-2 R&F POSITIONER POWER CONNECTIONS

1.5.4 Wire Gauge Guidelines For Field Supplied Power Runs


The Definium 6000 system is designed to distribute power internally to all modules and devices. In many cases, pre-terminated, multi-conductor cables are provided for this power distribution. Remaining power runs require field supplied wiring.
Wire gauges used for field supplied wiring depend on two considerations:
• Code requirements for minimum wire gauge as a function of circuit breaker size.
• System voltage regulation requirements.
For power runs, go directly to the Voltage Regulation Table. Find the column that matches the circuit breaker size and the rows corresponding to the voltage regulation requirement on the run. The proper wire gauge is indicated based on the length of the run. (Data
from the NEC Minimum Table is built into the Voltage Regulation Table.)
For power ground runs, use the same gauge as used for power runs. Exception: Where #8 AWG wire is used for power runs, #10 AWG is an acceptable ground size.
For Equal potential safety grounds, use #10 AWG minimum. Use larger gauge, or parallel runs to keep each run to 0.1 Ohm maximum. Refer to Section 1.5.

1.5.4.1 NEC Minimum


National Electric Code (NEC) Table 310-16 (75C wire) identifies minimum wire gauges for power wires based on circuit breaker size. Table 250-90 identifies mini mum wire gauges for ground wires based on circuit breaker size. These recommendations may be
modified as required for local code.
Table 6-3 is based on the NEC data noted above. It lists the recommended minimum wire size for discrete power runs and the associated line impedance for a 100 foot run.

Circuit Minimum Line Impedance Minimum Ground


Breaker Conductor Gauge Per 100 Foot Run Gauge (AWG)
Rating (AWG) (Ohms)
10/15 14 0.521 14
20 12 0.330 12
25/30 10 0.208 10
40/50 8 0.133 10
Table 6-3 Recommended Wire Sizes

1.5.4.2 Voltage Regulation


A second consideration, especially for longer runs, is voltage regulation requirements of the system. In general, most systems power runs, fall into one of three categories:
• Inter-cabinet 240 VAC @ 1% regulation.
• Inter-cabinet 120 VAC @ 3% regulation.
• Peripheral device 120 VAC @ 6% regulation.
Voltage regulation requirements for each particular power run are identified on the MIS charts.

Chapter 6 - Master Interconnect Schematic (MIS) and Charts Page 249


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Tables 1-4 through 1-7 indicate the wire size required for various voltage regulation categories. These requirements are a function of current, and run length.

Conductor gauge RUN LENGTH (IN FEET) FOR CIRCUIT BREAKER SIZES
AWG 5A 10A 15A 20A 25A 30A
240 vac @ 1% FEET
14 0-67 0-33 - - - -
12 67-109 33-54 0-36 0-26 - -
10 109-183 54-91 36-60 26-45 0-35 0-29
8 183-280 91-139 60-92 45-68 35-54 29-45
6 - 139-225 92-149 68-111 54-87 45-72
4 - - 149-215 111-161 87-127 72-105
240 vac @ 3% FEET
14 0-203 0-101 0-67 0-50 0-40 0-33
12 - 101-164 67-109 50-81 40-65 33-54
10 - 164-276 109-183 81-137 65-109 54-91
8 - - 183-280 137-209 109-167 91-139
6 - - - - 167-270 139-225
4 - - - - - -
240 vac @ 6% FEET
14 0-407 0-203 0-135 0-101 0-81 0-67
12 - - 135-219 101-164 81-131 67-109
10 - - - 164-276 131-220 109-183
8 - - - - 220-336 183-279
6 - - - - - -
4 - - - - - -
Table 6-4 WIRE GAUGE SIZES FOR 240 VAC LOAD AND REGULATION IN FEET (based on 60 hz at 45 degrees C)

Note: Limit all power runs to 200 ft. maximum. Larger values are shown for reference only.

Conductor gauge RUN LENGTH (IN FEET) FOR CIRCUIT BREAKER SIZES
AWG 5A 10A 15A 20A 25A 30A
120 vac @ 1% FEET
14 0-33 - - - - -
12 33-54 0-26 - - - -
10 54-91 26-45 0-29 - - -
8 91-139 45-68 29-45 0-33 0-26 -
6 139-225 68-110 45-72 33-53 26-42 0-35
4 - 110-160 72-106 53-78 42-62 35-50
120 vac @ 3% FEET
14 0-101 0-50 0-33 0-25 - -
12 101-164 50-81 33-54 25-40 0-32 0-26
10 164-276 81-137 54-91 40-68 32-54 26-45
8 - 137-209 91-139 68-103 54-82 45-68
Table 6-5 WIRE GAUGE SIZE FOR 120 VAC LOAD AND REGULATION IN FEET (based on 60 hz at 45 degrees C)

Section 1.0 Grounding & Wire Gauge Selection in Master Interconnect Schematics Page 250
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Conductor gauge RUN LENGTH (IN FEET) FOR CIRCUIT BREAKER SIZES
AWG 5A 10A 15A 20A 25A 30A
6 - - 139-225 103-168 82-134 68-110
4 - - - 168-245 134-195 110-160
120 vac @ 6% FEET
14 0-203 0-101 0-67 0-50 0-40 0-33
12 - 101-164 67-109 50-81 40-65 33-54
10 - 164-276 109-183 81-137 65-109 54-91
8 - - 183-280 137-209 109-167 91-139
6 - - - - 167-270 139-225
4 - - - - - -
Table 6-5 WIRE GAUGE SIZE FOR 120 VAC LOAD AND REGULATION IN FEET (based on 60 hz at 45 degrees C)

Note: Limit all power runs to 200 ft. maximum. Larger values are shown for reference only.

CONDUCTOR GAUGE RUN LENGTH (IN METERS) FOR CIRCUIT BREAKER SIZES
AWG 5A 10A 15A 20A 25A 30A
240 vac @ 1% METERS
14 0-20.4 0-10.1 - - - -
12 20.4-33.2 10.1-16.5 0-11.0 0-7.9 - -
10 33.2-55.8 16.5-27.7 11.0-18.3 7.9-13.7 0-10.7 0-8.8
8 55.8-85.3 27.7-42.4 18.3-28.0 13.7-20.7 10.7-16.5 8.8-13.7
6 - 42.4-68.6 28.0-45.4 20.7-33.8 16.5-26.5 13.7-21.9
4 - - 45.4-65.5 33.8-49.1 26.5-38.7 21.9-32.0
240 vac @ 3% METERS
14 0-61.9 0-30.8 0-20.4 0-15.2 0-12.2 0-10.1
12 - 30.8-50.0 20.4-33.2 15.2-24.7 12.2-19.8 10.1-16.5
10 - 50.0-84.1 33.2-55.8 24.7-41.8 19.8-33.2 16.5-27.7
8 - - 55.8-85.3 41.8-63.7 33.2-50.9 27.7-42.4
6 - - - - 50.9-82.3 42.4-68.6
4 - - - - - -
240 vac @ 6% METERS
14 0-124.1 0-61.9 0-41.1 0-30.8 0-24.7 0-20.4
12 - - 41.1-66.8 30.8-50.0 24.7-39.9 20.4-33.2
10 - - - 50.0-84.1 39.9-67.1 33.2-55.8
8 - - - - 67.1-102.4 55.8-85.0
6 - - - - - -
4 - - - - - -
Table 6-6 WIRE GAUGE SIZES FOR 240 VAC LOAD AND REGULATION IN METERS (based on 60 hz at 45 degrees C)

Chapter 6 - Master Interconnect Schematic (MIS) and Charts Page 251


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Note: Limit all power runs to 61 meters maximum. Larger values are shown for reference only.

CONDUCTOR GAUGE RUN LENGTH (IN METERS) FOR CIRCUIT BREAKER SIZES
AWG 5A 10A 15A 20A 25A 30A
120 vac @ 1% METERS
14 0-10.1 - - - - -
12 10.1-16.5 0-7.9 - - - -
10 16.5-27.7 7.9-13.7 0-8.8 - - -
8 27.7-42.4 13.7-20.7 8.8-13.7 0-10.1 0-7.9 -
6 42.4-68.6 20.7-33.5 13.7-21.9 10.1-16.2 7.9-12.8 0-10.7
4 - 33.5-48.8 21.9-32.3 16.2-23.8 12.8-18.9 10.7-15.2
120 vac @ 3% METERS
14 0-30.8 0-15.2 0-10.1 0-7.6 - -
12 30.8-50.0 15.2-24.7 10.1-16.5 7.6-12.2 0-9.8 0-7.9
10 50.0-84.1 24.7-41.8 16.5-27.7 12.2-20.7 9.8-16.5 7.9-13.7
8 - 41.8-63.7 27.7-42.4 20.7-31.4 16.5-25.0 13.7-20.7
6 - - 42.4-68.6 31.4-51.2 25.0-40.8 20.7-33.5
4 - - - 51.2-74.7 40.8-59.4 33.5-48.8
120 vac @ 6% METERS
14 0-61.9 0-30.8 0-20.4 0-15.2 0-12.2 0-10.1
12 - 30.8-50.0 20.4-33.2 15.2-24.7 12.2-19.8 10.1-16.5
10 - 50.0-84.1 33.2-55.8 24.7-41.8 19.8-33.2 16.5-27.7
8 - - 55.8-85.3 41.8-63.7 33.2-50.9 27.7-42.4
6 - - - - 50.9-82.3 42.4-68.6
4 - - - - - -
Table 6-7 WIRE GAUGE SIZE FOR 120 VAC LOAD AND REGULATION IN METERS (based on 60 hz at 45 degrees C)

Note: Limit all power runs to 61 meters maximum. Larger values are shown for reference only.

Section 1.0 Grounding & Wire Gauge Selection in Master Interconnect Schematics Page 252
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 2.0
The Master Interconnect Schematic (MIS) Map
2.1 Introduction

2.1.1 Using the MIS Map


The MIS Map has one primary purpose: it shows the interconnection runs required for the core components and all available options. Since all cable connections are indicated, need to refer to MIS charts except for signal tracing or troubleshooting should be
eliminated.

2.1.2 Symbols/Conventions
Identifiers on the MIS maps indicate where connections are normally made to cabinets or major assemblies. MIS runs are identified by lines between connection points, with a numbered circle on the line for each run.

EX1 CABINET A2
INTERNAL CABLE
CABINET A1 (INCLUDED ONLY
MODULE FOR OPTIONS)
J9
A1 A 1
SUBASSEMBLY X
(BOARD)
A1 A 2 A3 J2
J2 J15
MIS RUN
NUMBER MIS RUN
(EXTERNAL CABLE)
xx xxx NUMBER
(EXTERNAL WIRE)

Figure 6-1 EQUIPMENT AND RUN IDENTIFICATION


Runs that end in arrows going to a voltage value indicate connections to the Generator Power Unit supplies. See Figure 6-2. Runs that end in arrows going to a voltage value which also says “Hospital Supply” indicate connections to the power distribution system
of the hospital. See Figure 6-2.

EX1 CABINET

J15
XX
To Power Cabinet
XXX

To Hospital Distribution3

Hospital Power

Figure 6-2 POWER SOURCE IDENTIFICATION

Chapter 6 - Master Interconnect Schematic (MIS) and Charts Page 253


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

2.2 System MIS Maps (Full Sized)

The System MIS Map is also shipped as separate document supplied with your Definium 6000 system.

       

)HL7LDQ 6\VWHP 0,60DS


$ $
0DJLF 3&
6\VWHP &DELQHW
:DOO%R[
+RVSLWDO (WKHUQHW
:DOOVWDQG
$ *E(FDUG $
1HWZRUN
% --HGL&$1 $ $ $ %
-HGL&$1$- :6&$1ದ$- &$1RSHQ
0RXVH 86%

-6\VWHP&$1RSHQ $ $ $


&$10DJLFದ$- :63:5ದ$- :63:5
.H\ERDUG
86% 9$&IURP3'8 $ $ &RQWURO5RRP3ZUದ$- '363:5ದ$-
$ '363:5

& *1'VWXG $ *1'VWXG :6,21&+$0%(5ದ$-


$ :6,21&+$0%(5
&
%$5&2'(
86%
5($'(5
$ $ *1'VWXG
*1'VWXG *1'VWXG *1'VWXG
,PDJH 0RQLWRU
'9,, 9LGHR

' 276 '


$
56 7RXFKVFUHHQ

9*$ $
9*$

$ $
276&$1ದ$- &$1RSHQ
( +DQGVZLWFK
5&,0
(
*1'VWXG
$ *1'VWXG
$  5&,0ದ$-
$
6WDWRU)DQ 3UHVVXUH6ZLWFKದ$- 6WDWRU)DQ 3UHVVXUH6ZLWFK

$
+97UDQVIRUPHU$QRGH $QRGH
+RVSLWDO 
) 9$&  SKDVH
)
 SKDVH SRZHU FDEOH 7R&DELQHW
+97UDQVIRUPHU&DWKRGH
$ &DWKRGH
VXSSO\ &LUFXLW%UHDNHU

*5281',1* $
2763:5ದ$- 2763:5

;5$< 21
/,*+7
* '225 7R3'87HUPLQDOV
*
,17(5/2&.

+ 7,7/(

6\VWHP0,60$3
5(9 +
),567 0$'( )25 )HL7LDQ

6+7&217 21


       

Figure 6-3 System MIS Map

Section 2.0 The Master Interconnect Schematic (MIS) Map Page 254
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

Section 3.0 3.5 MIS CHART 11708A


MIS CHARTS
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11708A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
3.1 MIS CHART 11629A SYSTEM CABINET SKL OTS
(Pins not listed
BULKHEAD A25J5 are not connected) OTS Power In OTS
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11629A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
SYSTEM CABINET SKL OTS
(Pins not listed 3.6 MIS CHART 11710A
BULKHEAD A25J41 are not connected) ALB BOX L5 OTS

INTERCONNECT CHART NUMBER


SHEET 1 OF 1 11710A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
3.2 MIS CHART 11644A SYSTEM CABINET SKL OTS
(Pins not listed
BULKHEAD A25J42 are not connected) Tube Stator£‹Fan & Pressure Switch OTS
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11644A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
SYSTEM CABINET SKL WALLSTAND 3.7 MIS CHART 11711A
(Pins not listed
BULKHEAD A25J206 are not connected) J2 WALLSTAND
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11711A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
3.3 MIS CHART 11690A SYSTEM CABINET SKL OTS
(Pins not listed
BULKHEAD GND are not connected) GND OTS
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11690A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
1 Ground 1
SYSTEM CABINET SKL OTS OVERHEAD TUBE STAND
(Pins not listed
HV TRANSFORMER CATHODE are not connected) CATHODE X-RAY TUBE

S SMALL S 3.8 MIS CHART 11755A

C COMMON C
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11755A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
L LARGE L SYSTEM CABINET SKL WALLSTAND
(Pins not listed
BULKHEAD A25J2 are not connected) J5 WALLSTAND
3.4 MIS CHART 11691A
3.9 MIS CHART 11756A
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11691A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
SYSTEM CABINET SKL
(Pins not listed
OTS OVERHEAD TUBE STAND INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11756A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
HV TRANSFORMER ANODE are not connected) ANODE X-RAY TUBE
SYSTEM CABINET SKL WALLSTAND
(Pins not listed
BULKHEAD A25J1 are not connected) J4 WALLSTAND
S SMALL S

C COMMON C

L LARGE L

Page 255
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

3.10 MIS CHART 11757A 3.13 MIS CHART 11761A

INTERCONNECT CHART NUMBER


SHEET 1 OF 1 11757A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11761A REV 0
APPLICATION: DEFINIUM 8000
SYSTEM CABINET SKL WALLSTAND SYSTEM CABINET SKL WALLBOX
(Pins not listed (Pins not listed
BULKHEAD GND are not connected) GND WALLSTAND are not connected)
BULKHEAD A25J107 CAN WALLBOX

1 RESERVED 1
1 Ground 1
(WHT/BLU)
5

9 3
2 CAN-L 2
Wht 5
Grn 4
Brn (WHT/ORN)
3.11 MIS CHART 11759A 8 4
Wht
Brn
Wht
3
4

7 2
Grn
Wht 3 CAN-GND 3
3
Wht
Orn 2 (WHT/GRN)
Orn
6 1 Wht
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11759A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
Wht
Blu 1
Blu
Wht
2

4 CAN-L-LPBK 4
1 (WHT/BRN)
SYSTEM CABINET SKL WALLSTAND
(Pins not listed
BULKHEAD A25J14 are not connected) J1 WALLSTAND 5 CAN SHIELD 5
Wiring Side (???)

CAN-OPT
3.12 MIS CHART 11760A 1
Blu
1 6
(BLU/WHT)
6
1
6 Wht Wht
Blu 2 2
Orn
7 2 Wht
7 CAN-H 7
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11760A REV 0
APPLICATION: DEFINIUM 8000
8
Wht
Orn
4
3
Grn
Wht
3 (ORN/WHT)
Wht
SYSTEM CABINET SKL WALLBOX Brn 4 4
(Pins not listed Twisted Pin's
9 3
Brn
Wht 8 CAN-H LPBK 8
Pair

BULKHEAD A25J108 are not connected) JEDI CAN WALLBOX


1
2
1
2
6
7
Wht
Grn 5 5 (GRN/WHT)
3 3 8
4 4 9
Postion 5 not used

1 RESERVED 1 9 CAN-V+ 9
(WHT/BLU)
5

9 3
Wht 5 2 CAN-L 2
Grn 4

8 4
Brn
Wht
(WHT/ORN)
Wht 4
Brn 3
Grn
7 2 Wht
3 3 CAN-GND 3
Wht
Orn 2
Orn
(WHT/GRN)
6 1 Wht
Wht 2
Blu 1
Blu
Wht 4 CAN-L-LPBK 4
1
(WHT/BRN)

5 CAN SHIELD 5
Wiring Side (???)

1 1 6 CAN-OPT 6
Blu
1
6 Wht Wht (BLU/WHT)
Blu 2 2
Orn
7 2
Wht
Wht
7 CAN-H 7
Orn 3 3
8 4
Grn
Wht
(ORN/WHT)
Wht
Brn 4 4
Brn
Twisted
Pair
Pin's
9 3
Wht
8 CAN-H LPBK 8
Wht
1

3
2
1
2
3
6
7
8
Grn 5 5 (GRN/WHT)
4 4 9
Postion 5 not used

9 CAN-V+ 9
(BRN/WHT)

Section 3.0 MIS CHARTS Page 256


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

3.14 MIS CHART 11762A 3.15 MIS CHART 11763A

INTERCONNECT CHART NUMBER


SHEET 1 OF 1 11762A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11763A REV 0
APPLICATION: DEFINIUM 8000
SYSTEM CABINET SKL WALLBOX SYSTEM CABINET SKL WALLBOX
(Pins not listed (Pins not listed
BULKHEAD A25J109 are not connected) RCIM WALLBOX
BULKHEAD A25J6 are not connected) POWER WALLBOX
1 RESET SWITCH_P 1 RS485 RESET switch signal
2 AUTO-POS SWITCH_P 2 RS485 AUTO-POSITION BUTTON signal
3 PREP SW_P 3 RS485 Hand switch Prep switch signal
4 EXP SW_P 4 RS485 Hand switch Exposure switch signal 1 120 VAC 1
33
5 TRAIL SW_P 5 RS485 Hand switch Trial switch signal (BROWN)
17 6 X-RAY ON_P 6 RS485 System X-RAY ON signal
7 TUBE OVERHEAT_P 7 RS485 Tube Overheat indicator control signal Neutral
50 8 EXP INHIBIT_P 8 RS485 Exposure Inhibit indicator control signal 2 2
9 ERROR MESSAGE_P 9 RS485 Error Message Indicator control signal
(BLUE)
10 LIMIT INDICATION_P 10 RS485 Limit Indicator control signal
11 24VDC 11 Power 24V Power supply Ground
12 24VDC 12 Power 24V Power Supply 3 3
(GREEN)
13 24VDC 13 Power 24V Power Supply
14 24VDC 14 Power 24V Power Supply
15 Spare_1_P 15 RS485 Spare RT line with Host
34 1 16 Spare_2_P 16 RS485 Spare RT line with Host
18 17 Switch Power 17 Sw Power Keep Alive Power pin for switch (Power ON, E-Stop)
18 Power ON 18 24V Power ON signal, one contact of Power ON switch
19 E-STOP 19 24V E-STOP signal, one contact of E-stop switch
20
21
MIC SIG
SPEAKER SIG
20
21
Analog
Analog
Microphone output signal.
Speaker input signal.
3.16 MIS CHART 11764A
22 Spare_3_P 22 RS485 Spare RT line with Host
23 Spare_4_P 23 RS485 Spare RT line with Host
24 GND 24 GROUND 24V Power Supply Grounding
25 GND 25 GROUND 24V Power Supply Grounding
26
27
RESET SWITCH_N
AUTO-POS SWITCH_N
26
27
RS485
RS485
RESET switch signal
AUTO-POSITION BUTTON signal
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11764A REV 0
APPLICATION: DEFINIUM 8000
28 PREP SW_N 28 RS485 Hand switch Prep switch signal SYSTEM CABINET SKL WALLBOX
29 EXP SW_N 29 RS485 Hand switch Exposure switch signal (Pins not listed
30 TRAIL SW_N 30 RS485 Hand switch Trial switch signal
31 X-RAY ON_N 31 RS485 System X-RAY ON signal GROUND STUD GND are not connected) GND WALLBOX
32 TUBE OVERHEAT_N 32 RS485 Tube Overheat indicator control signal
33 EXP INHIBIT_N 33 RS485 Exposure Inhibit indicator control signal
34 ERROR MESSAGE_N 34 RS485 Error Message Indicator control signal SAFETY GND
33
35 LIMIT INDICATION_N 35 RS485 Limit Indicator control signal
50 36 GND 36 GROUND 24V Power Supply Ground (GRN/YEL)
17 37 GND 37 GROUND 24V Power Supply Ground
38 GND 38 GROUND 24V Power Supply Ground
0.1 OHM Maximum
39 GND 39 GROUND 24V Pow er Supply Ground
40 Spare_1_N 40 RS485 Spare RT line with host
41 Spare_2_N 41 RS485 Spare RT line with host
42 Ground 42 Ground 24V Power Supply Ground
43 Not used 43
34 44 Not used 44
45 MIC SIG RTN 45 Analog Microphone output signal return.
46 SPEAKER SIG RTN 46 Analog Speaker Input signal return.
47 Spare_3_N 47 RS485 Spare RT line with host
1 48 48 RS485 Spare RT line with host
18 Spare_4_N
49 GND 49 GROUND 24V Power Supply Ground
50 GND 50 GROUND 24V Power Supply Ground

Page 257
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

3.17 MIS CHART 11767A 3.18 MIS CHART 11768A

INTERCONNECT CHART NUMBER


SHEET 1 OF 1 11767A REV 0
APPLICATION: DEFINIUM 8000
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11768A REV 0
APPLICATION: DEFINIUM 8000
WALLBOX Workstation/PC Workstation/PC WALLBOX Workstation/PC Workstation/PC
(Pins not listed (Pins not listed
JEDI/Generator CAN are not connected) J2 Workstation/PC Rear Panel CAN are not connected) J1 Workstation/PC Rear Panel

1 RESERVED 1 RESERVED
(WHT/BLU) 1 1
(WHT/BLU)
5 5
9 3
Wht 5 2 CAN-L 2
9 3
5 CAN-L
Grn 4
Brn (WHT/ORN)
Wht
Grn 4 2 2
8 4 Wht 8 4
Brn
Wht
(WHT/ORN)
Wht 4 Wht 4
Brn 3
Brn 3
Grn
7 2 Wht
3 3 CAN-GND 3 7 2
Grn
Wht CAN-GND
Wht
Orn 2
(WHT/GRN) Wht 3 3 3
6 1
Orn
Wht
Orn 2
Orn (WHT/GRN)
6 1 Wht
Wht 2 Wht 2
Blu 1
Blu 1
Blu
Wht 4 CAN-L-LPBK 4 Blu
CAN-L-LPBK
1
(WHT/BRN)
Wht
1 4 4
(WHT/BRN)

5 CAN SHIELD 5 CAN SHIELD


Wiring Side Wiring Side 5 5
(???) (???)

1 1 6 CAN-OPT 6 1 CAN-OPT
Blu
(BLU/WHT)
1
Blu 6 6
1
6 Wht Wht 1
6 Wht Wht (BLU/WHT)
Blu 2 Blu
2 2 2
Orn Orn
7 2
Wht
Wht
7 CAN-H 7 7 2
Wht
Wht
CAN-H
Orn 3
Grn
3
(ORN/WHT) Orn 3 3 7 7
8 4 Wht 8 4
Grn
Wht
(ORN/WHT)
Wht Wht
Brn 4 4 Brn 4
Brn Brn
4
Twisted
Pair
Pin's
9 3
Wht
8 CAN-H LPBK 8
Twisted Pin's
9 3
Wht CAN-H LPBK
1 1 6 Wht
Grn 5 5 (GRN/WHT)
Pair
1 1 6 Wht
5
8 8
3
2 2
3
7
8
3
2 2
3
7
8
Grn 5
(GRN/WHT)
4 4 9 4 4 9
Postion 5 not used Postion 5 not used

9 CAN-V+ 9 CAN-V+
(BRN/WHT) 9 9
(BRN/WHT)

Section 3.0 MIS CHARTS Page 258


GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

3.19 MIS CHART 11769A 3.20 MIS CHART 11770A

INTERCONNECT CHART NUMBER


SHEET 1 OF 1 11769A REV 0
APPLICATION: DEFINIUM 8000
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11770A REV 0
APPLICATION: DEFINIUM 8000
WALLBOX SKL RCIM WALLBOX Workstation/PC Workstation/PC
(Pins not listed (Pins not listed
RCIM are not connected) RCIM
Magic Power 120 - 7A are not connected) PWR Workstation/PC Rear Panel
1 RESET SWITCH_P (WHT/TAN) 1 RS485 RESET switch signal
2 AUTO-POS SWITCH_P (TAN/WHT) 2 RS485 AUTO-POSITION BUTTON signal
3 PREP SW_P (WHT/BRN) 3 RS485 Hand switch Prep switch signal
33 4 EXP SW_P (BRN/WHT) 4 RS485 Hand switch Exposure switch signal 120 VAC
5 TRAIL SW_P (WHT/PINK) 5 RS485 Hand switch Trial switch signal 1 1
17 X-RAY ON_P (PINK/WHT) (BROWN) IEC210 Female
6 6 RS485 System X-RAY ON signal
7 TUBE OVERHEAT_P (WHT/ORN) 7
50
EXP INHIBIT_P (ORN/WHT)
RS485 Tube Overheat indicator control signal
Neutral
1 2
8 8 RS485 Exposure Inhibit indicator control signal
9 ERROR MESSAGE_P (WHT/YEL) 9 RS485 Error Message Indicator control signal
2 2
LIMIT INDICATION_P (YEL/WHT)
(BLUE) 3
10 10 RS485 Limit Indicator control signal
11 24VDC (WHT/GRN) 11 Power 24V Power supply
12 24VDC (GRN/WHT) 12 Power 24V Power Supply Ground
13 24VDC (WHT/BLU) 3 3
24VDC (BLU/WHT)
13 Power 24V Power Supply (GREEN)
14 14 Power 24V Power Supply
1 15 Spare_1_P (WHT/VIO) 15 RS485 Spare RT line with Host
34 Spare_2_P (VIO/WHT)
18 16 16 RS485 Spare RT line with Host
17 Switch Power (WHT/GRY) 17 Sw Power Keep Alive Power pin for switch (Power ON, E-Stop)
18 Power ON (GRY/WHT) 18 24V Power ON signal, one contact of Power ON switch
19 E-STOP (TAN/BRN) 19 24V E-STOP signal, one contact of E-stop switch
20 MIC SIG (BRN/TAN) 20 Analog Microphone output signal
21 SPEAKER SIG (TAN/PINK) 21 Analog Speaker input signal
22
23
Spare_3_P (PINK/TAN)
Spare_4_P (TAN/ORN)
22
23
RS485
RS485
Spare RT line with Host
Spare RT line with Host
3.21 MIS CHART 11774A
24 GND (ORN/TAN) 24 GROUND 24V Power Supply Grounding
25 GND (TAN/YEL) 25 GROUND 24V Power Supply Grounding
26 RESET SWITCH_N (YEL/TAN) 26 RS485 RESET switch signal
AUTO-POS SWITCH_N (TAN/GRN)
27
28
29
PREP SW_N (GRN/TAN)
EXP SW_N (TAN/BLU)
27
28
29
RS485
RS485
RS485
AUTO-POSITION BUTTON signal
Hand switch Prep switch signal
Hand switch Exposure switch signal
INTERCONNECT CHART NUMBER
SHEET 1 OF 1 11774A REV 0
APPLICATION: DEFINIUM 8000
30 TRAIL SW_N (BLU/TAN) 30 RS485
X-RAY ON_N (TAN/VIO)
Hand switch Trial switch signal WALLBOX User Video I/O Panel (left)
31 31 RS485 System X-RAY ON signal
32 TUBE OVERHEAT_N (VIO/TAN) 32
(Pins not listed
RS485 Tube Overheat indicator control signal
33 EXP INHIBIT_N (TAN/GRY) 33 RS485 Exposure Inhibit indicator control signal Monitor 1 120 - 1A are not connected) PWR Left Monitor Rear Panel
34 ERROR MESSAGE_N (GRY/TAN) 34 RS485 Error Message Indicator control signal
35 LIMIT INDICATION (BRN/PINK) 35 RS485 Limit Indicator control signal
36 GND (PINK/BRN) 36 GROUND 24V Power Supply Ground
37 GND (BRN/ORN) 37 GROUND 24V Power Supply Ground 120 VAC (BROWN)
38 GND (ORN/BRN) 38 GROUND 24V Power Supply Ground 1 1
GND (BRN/YEL) IEC210 Female
39 39 GROUND 24V Power Supply Ground
40 Spare_1_N (YEL/BRN) 40 RS485 Spare RT line with host
41 Spare_2_N (BRN/GRN) Neutral (BLUE) 1 2
41 RS485 Spare RT line with host
42 Ground (GRN/BRN) 42 Ground 24V Power Supply Ground
2 2
43 Not used (BRN/BLU) 43
3
44 Not used (BLU/BRN) 44
45 MIC SIG RTN (VIO/BRN) 45 Analog Microphone output signal return
Ground (GREEN)
46 SPEAKER SIG RTN (BRN/VIO) 46 Analog Speaker Input signal return
3 3
47 Spare_3_N (BRN/GRY) 47 RS485 Spare RT line with host
48 Spare_4_N (GRY/BRN) 48 RS485 Spare RT line with host
49 GND (PINK/ORN) 49 GROUND 24V Power Supply Ground
50 GND (ORN/PINK) 50 GROUND 24V Power Supply Ground

3.22 MIS CHART 11776A

INTERCONNECT CHART NUMBER


SHEET 1 OF 1 11776A REV 0
APPLICATION: DEFINIUM 8000
SYSTEM CABINET SKL Workstation/PC
(Pins not listed
GROUND STUD GND are not connected) GND Workstation/PC Rear Panel

SAFETY GND
(GRN/YEL)
0.1 OHM Maximum

Page 259
GE HEALTHCARE
DIRECTION 5135611-100, REVISION 4 DEFINIUM™ 6000 SYSTEM SCHEMATICS AND DRAWINGS

3.23 MIS CHART 20101A

INTERCONNECT CHART NUMBER


SHEET 1 OF 1 20101A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
Magic PC MAGIC OTS
(Pins not listed
BULKHEAD RS232 are not connected) RS232 CONN OTS

1 RXD 1

2 TXD 2

3 12VDC 3

4 GND 4

3.24 MIS CHART 20102A

INTERCONNECT CHART NUMBER


SHEET 1 OF 1 20102A REV 0
APPLICATION: REVOLUTION XR/D - 2X PLATFORM
Magic PC MAGIC OTS
(Pins not listed
BULKHEAD VGA are not connected) Concole A2 J2 OTS

1 RXD 1

2 TXD 2

3 12VDC 3

4 GND 4

Section 3.0 MIS CHARTS Page 260


ay
Xr
262
GE HEALTHCARE
NO.1, YONGCHANG NORTH ROAD
ECONOMIC & TECHNOLOGICAL DEVELOPMENT ZONE
FAX: 86-10-67881850
TELE: 86-10-58068888

BEIJING, P. R. CHINA 100176

You might also like