Professional Documents
Culture Documents
gemedical.com
Technical
Publication
Direction 2406876-100
Revision 7
GE Medical Systems
Precision THUNIS-800+ R&F System
Service Manual
Copyright © 2004~2008
by General Electric Company, Inc.
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
Page 2
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
LEGAL NOTES
TRADEMARKS
Adobe, the Adobe logo, Acrobat, the Acrobat logo, Exchange, and PostScript are trademarks of
Adobe Systems Incorporated or its subsidiaries and may be registered in certain jurisdictions.
Microsoft is a registered trademark and Windows is a trademark of Microsoft Corporation.
All other products and their name brands are trademarks of their respective holders.
COPYRIGHTS
All Material, Copyright © 2004~2008 by General Electric Company, Inc. All rights reserved.
IMPORTANT PRECAUTIONS
LANGUAGE
警告 • 本维修手册仅提供英文版本。
(ZH-CN) • 如果维修服务提供商需要非英文版本,客户需自行提供翻译服务。
• 未详细阅读和完全理解本维修手册之前,不得进行维修。
• 忽略本警告可能对维修人员,操作员或患者造成触电、机械伤害或其他形式的伤害。
• このサービスマニュアルには英語版しかありません。
(JA) • サービスを担当される業者が英語以外の言語を要求される場合、翻訳作業はその業
者の責任で行うものとさせていただきます。
• このサービスマニュアルを熟読し理解せずに、装置のサービスを行わないでくださ
い。
• この警告に従わない場合、サービスを担当される方、操作員あるいは患者さんが、
感電や機械的又はその他の危険により負傷する可能性があります。
(KO) • 고객의 서비스 제공자가 영어 이외의 언어를 요구할 경우, 번역 서비스를 제공하는 것
은 고객의 책임입니다.
IMPORTANT...X-RAY PROTECTION
X-ray equipment if not properly used may cause injury. Accordingly, the instructions herein
contained should be thoroughly read and understood by everyone who will use the equipment
before you attempt to place this equipment in operation. The General Electric Company, Medical
Systems Group, will be glad to assist and cooperate in placing this equipment in use.
Although this apparatus incorporates a high degree of protection against x-radiation other than the
useful beam, no practical design of equipment can provide complete protection. Nor can any
practical design compel the operator to take adequate precautions to prevent the possibility of any
persons carelessly exposing themselves or others to radiation.
It is important that anyone having anything to do with x-radiation be properly trained and fully
acquainted with the recommendations of the National Council on Radiation Protection and
Measurements as published in NCRP Reports available from NCRP Publications, 7910 Woodmont
Avenue, Room 1016, Bethesda, Maryland 20814, and of the International Commission on
Radiation Protection, and of any other local authorities, and take adequate steps to protect against
injury.
The equipment is sold with the understanding that the General Electric Company, Medical Systems
Group, its agents, and representatives have no responsibility for injury or damage which may result
from improper use of the equipment.
Various protective materials and devices are available. It is urged that such materials or devices be used.
Revision History
Revision Date Reason for change
1 Dec. 15, 2004 Frist release
2 May 10, 2005 Remove corresponding contents for Analog Systems
3 Apr. 10, 2006 Optimize Installation and calibration, add System checks, and
update MIS charts
4 Aug. 4, 2006 Seperate Class A and Class M Service Manual
5 Nov. 30, 2006 Update Schematics, FRU, Table I/O Board Information
6 Jun. 20,2007 Introduce in new touch screen
7 Jan. 24,2008 Add watchdog installation, CPI touch screen website, DICOM/
DSA/watchdog license labels attachment, collimator mounting
screws PM, and update FRU list
Preface
Publication Conventions
Standardized conventions for representing information is a uniform way of communicating
information to a reader in a consistent manner. Conventions are used so that the reader can easily
recognize the actions or decisions that must be made. There are a number of character and
paragraph styles used in this publication to accomplish this task. Please become familiar with them
before proceeding forward.
It’s important that you read and understand hazard statements, and not just ignore them.
Section 1.0
Safety & Hazard Information
Proper product safety labeling allows a person to safely use or service a product. The format and
style for safety communications reflected in this publication represents the harmonization of IEC/
ISO 3864 and ANSI Z535 standards.
Within this publication, different paragraph and character styles are used to indicated potential
hazards. Paragraph prefixes, such as hazard, caution, danger and warning, are used to identify
important safety information. Text (Hazard) styles are applied to the paragraph contents that are
applicable to each specific safety statement.
Preface Page 17
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
NOTICE - Indicates information or a company policy that relates directly or indirectly to the
safety of personnel or protection of property. This signal word is associated directly with a
hazard or hazardous situation and is used in place of 'DANGER,' 'WARNING,' or 'CAUTION.'
It can include:
• Destruction of a disk drive
• Potential for internal mechanical damage, such as to a X-ray tube
Page 18 Preface
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
Section 2.0
Publication Conventions
2.1 General Paragraph and Character Styles
Prefixes are used to highlight important non-safety related information. Paragraph prefixes (such
as Purpose, Example, Comment or Note) are used to identify important but non-safety related
information. Text styles are also applied to text within each paragraph modified by the specific
prefix.
The current section and its title The current chapter and its title
are always shown in the footer of are always shown in the footer of
the left (even) page. the right (odd) page.
An exclamation point in a triangle is used Paragraphs preceeded by a symbol
to indicate important information to the user. (e.g. bullets) contain information that
Paragraphs preceeded by Alphanumeric has no specific order.
characters (e.g. numbers) contain infor-
mation that must be followed in a specific order.
Headers and footers in this publication are designed to allow you to quickly identify your location.
The document part number and revision number appears in every header on every page. Odd
numbered page footers indicate the current chapter, its title and current page number. Even page
footers show the current section and its title, as well as the current page number.
Preface Page 19
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
2.3 Computer Screen Output/Input Text Character Styles
Within this publication, mono-spaced character styles (fonts) are used to indicate computer text
that’s either screen input and output. Mono-spaced fonts, such courier, are used to indicated text
direction. When you type at your keyboard, you are generating computer input. Occasionally you
will see the math operator “greater-than” and “less-than” symbols used to indicate the start and
finish of variable output. When reading text generated by the computer, you are reading it as
computer generated output. In addition to direction, characters are italicized (e.g. italics) to indicate
information specific to your system or site.
Example: This paragraph’s font represents computer generated screen “fixed” output.
Fixed Output Its output is fixed from the sense that it does not vary from application
to application. It’s the most commonly used style used to indicate
filenames, paths and text that do not change from system to system. The
character style used is a fixed width such as courier.
Example: This paragraph’s font represents computer screen output that is
Variable Output “variable”. It’s used to represent output that varies from application to
application or system to system. Variable output is sometimes found placed
between greater-than and less-than operators for clarification. For
example: <variable_ouput> or <3.45.120.3>. In both cases, the < and >
operators are not part of the actual input.
Example: This paragraph’s font represents fixed input. It’s computer input that is
Fixed Input typed-in via the keyboard. Typed input that does not vary from application
to application or system to system. Fixed text the user is required to
supply as input. For example: cd /usr/3p
Example: This paragraph’s font represents computer input that can vary from
Variable Input application to application or system to system. With variable text, the
user is required to supply system dependent input or information. Variable
input sometimes is placed between greater-than and less-than operators.
For example: <variable_input>. In these cases, the (<>) operators would
be dropped prior to input. For example: ypcat hosts | grep <3.45.120.3>
would be typed into the computer as
ypcat hosts | grep 3.45.120.3
without the greater-than and less-than operators.
2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys)
Different character styles are used to indicate actions requiring the reader to press either a hard or
soft button, switch or key. Physical hardware, such as buttons and switches, are called hard keys
because they are hard wired or mechanical in nature. A keyboard or on/off switch would be a hard
key. Software or computer generated buttons are called soft keys because they are software
generated. Software driven menu buttons are an example of such keys. Soft and hard keys are
represented differently in this publication.
Example: A power switch ON/OFF or a keyboard key like ENTER is indicated by applying a character style
Hard Keys that uses both over and under-lined bold text that is bold. This is a hard key.
Example: Whereas the computer MENU button that you would click with your mouse or touch with your hand
Soft Keys uses over and under-lined regular text. This is a soft key.
Page 20 Preface
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
ACROYMN DEFINITION
ABC / (ABS) Automatic brightness control/system. Regulates brightness
ABD Automatic Brightness Detection
ACTOR A Software Module that performs some function or action
ADC or A/D Analog to Digital Converter
ADCM Active Distortion Correction Module
ADCS Active Distortion Correction System
AEC Automatic exposure control. Technique used to control brightness signal to cut exposure.
ATHENA Digital System SW, which runs on the Windows 95 OS and on the Saturn Hardware
ATLAS CORE Operating system of the X-Ray products
CALYPSO GEMS’ engine ring name for the System control Software application
CAN Controller Area Network - A serial communication method in use on Definium 6000.
CANopen A protocol for a CAN network. Used interchangeably with CAN
CAT Collimator Alignment Tool
CCD Charge Coupled Devices
CCU Camera Control Unit
CDS Correlated Double Sampling
COTS Commercial Off The Shelf Tool
CORONA Engineering name for I/O Circuit Board used in RAD system cabinet
CPU Central Processing Unit
CPU Control processor unit. Microprocessor and peripherals which run the software/firmware
DAC Digital to Analog Converter
DSC Digital Servo Controller
DSC2 Digital Servo Controller, Version 2
EFC Earth Field Compensation
FKP Function key processor. Processor located in RCIM
FOV Field of View
FPGA Field programmable gate array. It is programmed by the CPU core after the reset and
handles all the exposure control logic including the system interface real-time lines
FRU Field Replaceable Unit (Spare Part)
G1, G2, G3 Grids 1, 2 and 3
GUI Graphical User Interface
HALO Engineering name for Relay Circuit Board used in RAD system cabinet
HOTKEY Keyboard Shortcut or key strokes used to execute a software command
HV RIPPLE High voltage variations due to inverter current pulses. Typically measured in percent.
HVPS High Voltage Power Supply
IGBT Insulated gate bipolar transistor power switch.
Ilp HV power inverter parallel resonant current. Current in the parallel inductor
Glossary Page 21
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
ACROYMN DEFINITION
Ilr HV power inverter serial resonant current. Current in the serial inductor
IO or I/O Input/Output
IUI Integrated User Interface
JEDI High Voltage Generator for X-Ray tube
LAT Lateral
LFC Load from Cold
LONG Longitudinal
LOTO LockOut/Tag Out
LP Line Pair
LSL Lower Specification Limit
LUT Look Up Table
MAINS Hospital Supplied AC Power
MAGIC System Control Computer / Workstation PC
MITRA Software supplier for DICOM application SW
MOS Metal oxyde semiconductor. Power switch
NDF Neutral Density Filter
NVRAM Non Volatile Random Access Memory
OEM Original Equipment Manufacture; Actual manufacturer that makes the device for GEHC.
OLC On-Line Center
OS Operating System
OTS Overhead Tube Suspension
PA Power Assist
PCI Pherperial Control Interface
PDF Portable Document Format
POS Positioner
RAD Radiography
R&F Radiography & Fluoroscopy
RCIM RAD Control Interface Module.
RED HAT ® SW version of Linux. Red Hat is an integrator and supplier of the Linux operating system.
ROI Region of Interest
POT Potentiometer
RPM RedHat (Linux) Package Manager software
RS422 A Serial Communications Hardware Connection Type
RT Real Time
PWA Printed Wire Board is synonymous with printed circuit board (PCB).
SABER Applications SW for positioning
SAE Society of Automotive Engineers - International organization that sets standards for
materials and fasteners in both English and Metric units, See www.sae.org
SBC Single Board Computer
SCAT System Compatibility Assurance Testing performed by GE prior to delivery.
Page 22 Glossary
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
ACROYMN DEFINITION
SID Source to Image Distance - The distance from the X-ray tube focal spot to the image plane
(i.e. detector, image intensifier)
STATE MACHINE Software or hardware functions that handle the state of a computer and authorizes it to go
to the next state upon reception of specific events.
SUIF Service User Interface
SW Abbreviation for the word Software
SYSTEM CONTROLLER See MAGIC
SYSTEM RESET A system reset is accomplished by pressing the green button located on the IUI. Pressing
it causes all of the computer to shutdown and restart. Thus re-booting the operating
systems and system applications.
TTL Transistor to Transistor Logic
UPS Un-interruptible Power Supply
U/T Abbreviation for Under Table
USL Upper Specification Limit
XT X-Ray Tube
Glossary Page 23
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
This page is intentionally left blank.
Page 24 Glossary
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
Table of Contents
Preface
Publication Conventions ...................................................................................... 17
Section 1.0
Safety & Hazard Information ........................................................................... 17
1.1 Hazard Messages............................................................................................................ 17
1.2 Text Format of Signal Words ........................................................................................... 17
1.3 Symbols and Pictorials Used ........................................................................................... 18
Section 2.0
Publication Conventions ................................................................................. 19
2.1 General Paragraph and Character Styles........................................................................ 19
2.2 Page Layout..................................................................................................................... 19
2.3 Computer Screen Output/Input Text Character Styles .................................................... 20
2.4 Buttons, Switches and Keyboard Inputs (Hard & Soft Keys) ........................................... 20
Section 1.0
Introduction
The goal of this document is to guide the service representative through an entire X-ray room
installation.
The following On-site Requirements must be accounted for prior to beginning installation of the
system.
Section 2.0
Objective and Scope of this Manual
This direction will lead you in a correct sequence through the manuals needed to install the
equipment. Following the suggested sequence will result in an orderly installation. However,
extenuating circumstances may dictate a deviation from the recommended sequence. Feel free to
deviate, but keep track of your progress by checking off the tasks in the flowchart as you complete
them.
Check equipment according to the packing list for your system against the main catalog numbers
in the FDO (Field Delivery Order). Also, verify that all the manuals are present in the binders
according to the Table of Contents.
Before starting to install the system, identify all the cables and check their length to verify that they
are long enough. If you find a short cable, contact us immediately so we can send the required
length cable to you and thereby, not delay the installation.
2.3 Interconnection
Interconnection of products by furnished cables or contractor supplied wiring is indicated by
numbered MIS runs in the installation sections. Run interconnect charts are available, which
describe the connection and function of each wire.
2.4 Presentation
The procedures in this steering manual are presented in jobcard forms.
Section 3.0
On-Site Requirements
3.1 EMC Requirements
If electromagnetic compliance is mandatory for the local country where this system is installed, this
equipment is allowed to be installed only in X-ray protected rooms, which provide an attenuation of
at least 12 dB for radio disturbances from 30 MHz to 1 GHz.
3.3 Documentation
Directions are necessary for correct installation of the Pecision THUNIS-800+ R&F System.
Procedures from each document will be used and reflect the latest information available at this time.
Not all procedures in this manual apply to all sites. Before you begin installation, cross out the
procedures that do not apply to your site.
Section 4.0
Description of System
5 4
7
8
Section 5.0
Installation Process
See System Installation Roadmap & Applied Time Standards on the next page.
Unpack & Position Demount Lock Screw Table Tilt Motor Digital Spot Device Wiring Harness Tube Stand Tilt Table to Vertical
Table Bracket Position
Refer to Page 51 and Refer to Page 55 Refer to Page 55 Refer to Page 58 Refer to Page 60 Refer to Page 61 Refer to Page 63
Page 53
2 Person 1 Hour 1 Person 0.25 Hour 2 Person 0.5 Hour 2 Person 0.5 Hour 2 Person 0.25 Hour 2 Person 0.5 Hour 1 Person 0.25 Hour
Refer to Page 64 Refer to Page 71 Refer to Page 75 Refer to Page 80 Refer to Page 82 To Be Continued
2 Person 1 Hour 2 Person 0.5 Hour 2 Person 0.5 Hour 2 Person 1 Hour 2 Person 2 Hours
DAY 2
DAY 2 DAY 2 DAY 2 DAY 2 DAY 2 DAY 2 DAY 2
Unpack PDU Install PDU Unpack Integrated Install Integrated Electrical Connec- Accessories Fix Table
Console Console tions
Refer to Page 89 Refer to Page 90 Refer to Page 92 Refer to Page 93 Refer to Page 99 Refer to Page 101 Refer to Page 109
2 People 1 Hour 2 Person 1 Hour 2 Person 1 Hour 2 Person 2 Hour 1 Person 0.5 Hour 1 Person 0.5 Hour 2 Person 1 Hour
Chapter 4 - System
Complete MI, Checks
Cleanup Site Equipment and Per- Power Up Refer to Page 123 To Be Continued
sonnel Readiness
1 Person 4 Hour
DAY 3
DAY 3 DAY 3 DAY 3 DAY 3 DAY 3
Chapter 4 - System Chapter 5 - Configu- Final Cover Installa- Labels and Rating Preliminary
Checks (Cont.) raton & Calibration tion Plates Application Training
Refer to Page 123 Refer to PS800+ Turn Over to Customer
Refer to Page 129 Refer to Page 110 Refer to Page 113 Operator Manual
1 Person 4 Hour 1 Person 2 Hour 1 Person 0.5 Hour 1 Person 0.5 Hour (2406878-100)
Page 41
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
Page 42
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
Section 1.0
System Installation Guide
1.1 Mechanical Installation
6 Attach Warning Labels and Rating Plates People Required: 1 Hours to Complete: 0.5
Attach all the warning labels and detent markers.
Section 2.0
Installation Handoff Checklist
Pre-Installation Checklist
Delivery Date: Sales Person:
Customer: FDO No.: Room #
Equipment:
2.) Floor is strong enough for intended equipment and mounting methods
approved – seismic regulatory codes considered?
3.) Delivery route accommodates all intended equipment?
8.) Support structures installed for floor and wall mounted equipment?
Interconnections Completed
1.) Signal cable, power and grounding plans produced?
4.) Flexible, stranded wire provided for System input power connection?
5.) System “feeder” power cables pulled and sufficient length available at
disconnect box for connections?
General Completed
1.) Walls and floor clear of all obstructions?
Comments:
Inspection Date(s):
11.) List of any installation drawing deviations (all deviations should be signed by customer
and reviewed by the installation specialist).
12.) Packing materials disposed of.
13.) Dollies and crates returned:
Positioner table dolly
____________________________________ ____________________________________
Mechanical Installer Signature Field Engineer Signature
____________________________________ ____________________________________
Mechanical Installer Signature Field Engineer Signature
Section 3.0
Cables
3.1 Routing
• System MIS cables should be pulled to/from the System Cabinet before the System Cabinet
is moved into the final mounting position.
• Some system cables are pre-connected to a sub-system and should not be disconnected in
order to pull the cable(s) through the conduits or wall ducts. Therefore, the following cables
should be pulled in the directions listed below:
- Wall Stand cables should be pulled from the Wall Stand to the System Cabinet.
- Magic PC cables should be pulled from the Wall Stand to the Magic PC.
• If excess cables must be coiled and stored, they must be coiled in a “figure 8" configuration,
NOT a single coil stack.
• Power cables must be kept separate from signal cables as much as possible. In wall ducts or
raceway, power cables should be routed in a separate channel from signal cables. Inside
system cabinets, power cables should not be bundled together with signal cables.
Section 4.0
Positioner Table
4.1 Unpack
This section will describe the basic steps required to un-crate the table.
• Do not discard any packing materials such as envelopes, boxes, bags until all parts are
accounted for against the packing list.
• Check the mechanical conditions and external appearances of all parts for possible damages
or missing items.
• The distribution center or factory must be notified immediately of any damage or shortage of
parts.
1.) Remove the top cover and four top cross members from the table shipping crate. See Figure 2-
3.
2.) Remove any additional cross members from the inside of the crate.
3.) Remove the sides of the shipping container. See Figure 2-4.
5.) Take out all the boxes out from the crate.
2.) Pull the table down the ramp and to the field, and confirm the location of the table with operator.
DO NOT put your feet under the table base, for the table is so heavy and may hurt you.
3.) Adjust caster bolts to upper the casters above the table base level.
Table Base
Tilt Motor
Motor Base
2.) Set the motor belt between pulleys of motor and gear box.
3.) Loosen the four motor mount nuts to adjust the motor belt tension.
Belt
Motor
cable
Motor
mount
nuts
Belt Tensioner
4.) Tighten the motor belt tensioner, so that the clearance between the support plate and the
tensioner plate is approximate 0~1 mm, and align the screw and the edge of the motor base.
7.) Insert table tilt motor cable (2409826) between the two counters, connect the cable, and fix the
U V W PE
1 2 3 4
1 0 0 0
Note: Set S1=”0001” and Jumper J8 to Pin1-Pin2 connection to support power input before
installation, after installaion, set them back to the default (S1=”0000” and Jumper J8 to Pin2-
Pin3 connection.
Note: If the country DO NOT have AC220V, please tilt table manually.
3.) Insert the digital spot device, and install it with the four screws.
Note: DO NOT need demount table top.
Note: Adjust the postion of digital spot device, place along marked guidelines, and then tighten the
four screws.
2.) Take the system wiring harness out from the back of the table base.
4.) Divide the cables into two groups for installation (one group is for I.I., the other is for other
subsystems behind the tube stand).
Tube Stand
Back Cover
5.) Connect the cables on the back of tube stand, and ty-rap the cables.
Tilt the table LESS THAN 90 degrees to avoid the table’s overturn.
3.) Power off.
4.3.7 I.I.
1.) Demount the plate for protection.
Front
Cover
Bottom
Cover
Cable
Slot
Note: Follow the process below during installing the 20 m cable (5122285) on I.I. side.
At I.I. side, the shielded cable should make 4 paths in the ferrite (3 loops).
a.) Insert the 20 m cable through the cable slot without ferrite.
b.) Open the ferrite and place the cable inside. The length of cable between the ferrite and
the connector should be 30 cm.
h.) Fix the cable to the Image Head using the cable clamp. Place the ferrite close to the cable
clamp before tightening.
Cable
Clamp
7.) Connect the I.I. power cable (2409831) and ground cable.
2409831
9.) Put the rubbery protection belt into the gap of I.I. and I.I. cover.
Screws
Screws
Release
Lock
Lock
Trunion
7.) Clean HV cable connectors, and then apply a thin layer of silicon compound.
8.) Connect the anode and cathode HV cables to the X-ray tube.
Note: Check the polarity of HV cables carefully, and DO NOT connect the two HV cables reversedly.
9.) Adjust tube to the position easy for operation.
10.) Demount the two screws for tube fan.
2.) Take out the collimator flange, demount the handle, and remount it onto the other side.
Take out
Note: In order to securely mount the collimator, please add Loctite 242 to the screws on the stop
plates, also need to add Loctite 242 to the 4 mounting screws on the collimator flange while
installing the collimator.
3.) Demount the four mounting screws on the tube.
4.) Put the blue handle between the collimator flange and the tube, and mount both the handle
and collimator flange using four screws. See Figure 2-47.
Collimator Flange
Blue Handle
5.) Mount collimator onto the collimator flange using the two stop plates.
9.) Connect collimator cables, and fix them with cable clamps.
12.) Connect the tube control cable (2409507) and grounding cable (2409508).
Note: Ground cable must be connected properly.
Collimator ground cable should be connected to tube E.
A196 RAD14
14.) Route and ty-rap the cables, and insert them into the cable hose.
Note: Suggest wear gloves when insert cables into cable hose.
15.) Mount the three hose brackets, and mount the cable hose into the hose brackets.
Hose
Hose Bracket
Section 5.0
Generator Cabinet
5.1 Unpack
• Do not discard any packing materials such as envelopes, boxes, bags until all parts are
accounted for against the packing list.
• Check the mechanical conditions and external appearances of all parts for possible damages
or missing items.
• The distribution center or factory must be notified immediately of any damage or shortage of
parts.
1.) Remove the top and side covers of the shipping container and cross braces.
6.) Position generator cabinet to the location that does not prevent the table’s movement.
2.) Demount generator side cover with six screws and washers.
Demount
6.) Clean HV cable connectors, and then apply a thin layer of silicon compound.
7.) Connect the anode and cathode HV cables to the generator oil tank.
HV Tank HV Cable
+ +
- -
Note: Check the polarity of HV cables carefully, and do not connect the two HV cables reversedly.
8.) Connect tube control cable (2409507) and tube ground cable (2409508).
2409508
2409507
To prevent possible body or equipment damage, please avoid inappropriate power up.
11.) Remount the protective cover.
12.) Connect CCU generator I/F cable (2408888) and touch screen console communication cable
(5123315) on generator I/F board.
b.) Connect touch screen grounding cable (5124835) from touch screen to generator room I/
F board.
c.) Connect generator room I/F cable (2408887).
d.) Connect door lock (Pin 4, 5) and room light (Pin 11, 12) on TB4.
AEC Channel 1
AEC Channel 2
16.) Ty-rap the cables and remount all the generator cable protective covers.
Section 6.0
PDU Cabinet
6.1 Unpack
• Do not discard any packing materials such as envelopes, boxes, bags until all parts are
accounted for against the packing list.
• Check the mechanical conditions and external appearances of all parts for possible damages
or missing items.
• The distribution center or factory must be notified immediately of any damage or shortage of
parts.
1.) Remove the top and side covers of the shipping container and cross braces.
Section 7.0
Integrated Console
7.1 Unpack
• Do not discard any packing materials such as envelopes, boxes, bags until all parts are
accounted for against the packing list.
• Check the mechanical conditions and external appearances of all parts for possible damages
or missing items.
• The distribution center or factory must be notified immediately of any damage or shortage of
parts.
1.) Remove the top and side covers of the shipping container and cross braces.
6.) Position console cabinet beside the integrated console, either left or right.
Demount
Demount
PS1
PS2
System Console
Power cable 2409500
Note: Follow the process below during installing the 20 m cable (5122285) on PCU side.
At PCU side, the shielded cable should make 3 paths in the ferrite (2 loops).
a.) Open the ferrite and place the cable inside. The length of cable between the ferrite and
the connector shall be about 7 cm.
5.) Take out the watchdog (5121184) from Accessory Box, install it to "Rockey" as below, and
connect workstation cables.
Power
4
4
6
6
2
2
Key 3
Key 3
1
1
5
5
COM1
Rockey
Keyboard
Mouse
Hospital Network
8.) Install in-room monitor (optional) on the cart, put them in-room, and connect in-room monitor
cables.
Note: Connect in-room monitor if customer orders it.
Section 8.0
Electrical Connections
1.) Configure S1 back to "0000" on table logic I/O board (2408348-1), and configure Jumper J8
back to Pin2-Pin3 connection.
1 2 3 4
0 0 0 0
4.) Connect other cables on table bulkhead, and ty-rap the cables.
J431
J442
J531
J542
J432
Section 9.0
Accessories
9.1 Install Compressor Force
1.) Take out the compressor with a dowel and bolt.
2.) Insert the dowel into the pole of the compressor.
Note: Align the holes in the dowel and pole.
Latter fix-
ing pin
1.) Insert the two former fixing pins into both side rails.
2.) Insert the two latter fixing pins into the slots in both side rails.
Note: There are three slots in each side rail.
Install the footrest by cutomer’s requirements.
Latter fix-
ing pin
Side rail
Lock
Lever
Screw down
Latter fix-
ing pin
Former
fixing pin
Screw down
2.) Install the outboard compression band pole (the pole with compression band fixed in), adjust
the compression band pole to a proper position, and then, tighten the bolt.
3.) Install the inboard compression band pole (the pole without compression band fixed on), adjust
the compression band pole to a proper position, and then, tighten the bolt.
Section 10.0
Fix Table
Note: Confirm the location of table with customers before fixing the table.
1.) Drill four 14 mm diameter holes 100 mm deep with masonry drill.
Section 11.0
Footstep Installation (Optional)
1.) Demount the footstep cover with 5 screws.
2.) Install the footstep clinging to table base with two bolts.
3.) Re-mount the footstep cover.
Section 12.0
Final Cover Installation
Note: Install the subsystem covers after system checks and configuration/calibration.
2.) Install generator cabinet top cover with two ground cables.
2.) Install the console cabinet back and side covers, and then install the rubbery caps in a plastic
poke (2239428).
Note: Leave at least 20 cm between console cabinet and wall, or the high temperature in console
cabinet will damage the server.
Section 13.0
Labels and Rating Plates
Attach English/multilingual labels to overlay the English/Chinese labels attached in factory.
13.1 System
13.2 Table
5123129
5123128
13.3 Generator
13.4 PDU
Watchdog
DICOM
DSA
Section 14.0
MI ==> FE Handoff Checklist
Purpose: Mechanical Installer to Field Engineer handoff checklist to confirm that all critical
installation steps have been completed.
3.) Cable shield clamps installed and tightened (how many, where).
4.) All mounting bolts installed and tightened (floor and wall).
Section 15.0
FE==> Apps Handoff Checklist
Purpose: Field Engineer to Application Specialist handoff checklist to confirm that all critical
installation steps have been completed.
16.) Are PACS configured and SMPTE and/or test images pushed to
system for successful display (be sure burn/no burn is properly
set for PACS)?
Section 16.0
Final Steps
16.1 Database Backup
After calibration, perform complete database backup. Refer to Backup procedure in the Utilities
appendix.
4)
II.) DSA
Section 1.0
Hardware
1.1 Table
Note: If the customer select In room foot pedal (Option). There is a configuration for customer to select:
1.) S2= "0000" -(Default - Remote configuration): Console foot switch and In-room foot switch are
controlled by "Remote" button. They can't be activated at the same time during fluoroscopy
operation.
2.) S2="0001" -(Customer - Classical configuration): Console foot switch and In-room foot switch
are not controlled by "Remote" button. They can be activated at the same time during
fluoroscopy operation.
A196 Tube
SW1: 1~8 OFF
RAD14 Tube
SW1: 3 ON
1, 2, 4~8 OFF
Note: 1.) There are many ABC inputs on generator I/F board; TH800+ shall use J8, which is a BNC
connector.
2.) "Out" in table above means the jumper shall be jumped out (not connected).
3.) The "*" in table above means "don't care".
4.) For some jumpers, such as JW11, JW19 and JW21, there are 3 pins named 1,2,3. So PINs2-
3 means pin 2 and 3 shall be connected.
Jumper
Jumper
Jumpers locate
on this board
Note: 1.) Some jumpers are simple jumpers, such as JW2 and JW3...etc. There are only 2 pins for these
jumpers. ON means connect, OFF means not connected.
2.) Some jumpers are a little bit complex, for example, JW44. 45, 46. There are only 4 pins for 3
jumpers. The figure below shows how to set these jumpers.
JW45 OFF
JW44 ON JW45 ON JW44 OFF JW46 ON JW44 OFF
JW46 OFF
JW45 OFF
JW46 OFF
Pin1 Pin2 Pin3 Pin4 Pin1 Pin2 Pin3 Pin4 Pin1 Pin2 Pin3 Pin4
3.) The figures above are just illustration to show the usage of this kind of jumpers. The marks on
the board have to be carefully read to ensure correct jumper setting. JW49, 50, 54,55,59,60
are nearly the same.
1.3 PDU
Factory configured voltage is 380V. You can configure the voltage by jumpers on the transformer.
1.5 Collimator
1.) Make the collimator controller of positioner console on max position.
2.) Superpose the adjustment tool on the collimator. This adjustment tool may be a transparent
material with the figure below, such as a white paper or card. The line’s width on the
adjustment tool shall be 1 mm. Push the lamp button of collimator to light it.
Spot/Fluoro
1/1
51mm
51mm
S1 S2
0011 0000
Note: S1, S2 configure is from high to low. Below diagram is an example for S1 configure.
1 2 3 4
1 1 0 0
4.) Adjust Potentiometer of X-ray Control board (2408350) to control the collimator. The
adjustment specifications are shown below. Visually check the collimator blade light vision
edge on the line inside. So it shall meet the requirement <+/-0.5mm.
Configure S1=”1011”, S2=”0000”. Adjust VR2 and VR4 for Spot 1/1 position.
Configure S1=”0101”, S2=”0000”. Adjust VR6 and VR8 for Fluoro 1/1 position.
Section 2.0
Softwares
Note: All the installation drivers are stored in E:\Setup\. The file E:\Setup\Recovery.gho cannot be moved.
All the factory configured parameters are stored in D:\Service Tools\. If you do any modifications on
the parameters, please also store in this path as another file name. DO NOT replace the factory
configured file.
All the softwares are factory installed.
Since the recovery disk includes all the softwares (Operating System, Service Package, Drivers,
GenWare, PCU, and Image Brightness Measure installation softwares), suggest recovering system
with the recovery disk if necessary, and then, install the DSA software.
Below software versions are for examples. Please confirm the current software versions in your
system by referring to the CDs shipped with system.
Note: The touch screen membrane can be easily scratched by sharp objects. To avoid possible
damage, items such as pens, pointers, tools, and etc should not be brought near the touch screen
membrane.
Do not attemp to install unauthorized software onto the console. The generator must be
programmed not to exceed the x-ray tube maximum operating limits shown in the Operator
Manual (Part Number 2406878-100). Intended life and reliability will not be obtained unless
generators are operated within published specifications.
The viewing angle adjustment screws at the rear of the touch screen console should only be
removed by qualified persennel. Removing these screws may create instability, causing possible
equipment damage or personal injury. The screws may be loosened sufficiently by the user
to adjust the viewing angle of the display, but must be retightened immediately.
The X-ray generator on / off buttons are located on the left-bottom of the touch screen frame.
The touch screen console will power up when the generator is switched on. When the console has
completed its startup routine, the main console menu will appear on the touch screen console. Four
menu selections are available as described below.
The software on the USB will damage a desktop PC. The software must be installed on a CPI Touch
Screen computer only.
After a brief delay, the system utilities menu will be displayed. The message Access
Denied indicates that an incorrect password was used.
The factory-default password may be changed by a service engineer as described
later. If this was done, the password defined above will not allow access to the
system utilities menu.
* APR Editor Allows the APR to be edited. Changes may be saved to memory.
* APR Backup / Allows the APR data to be backed up, and backed-up APR data to
Restore be restored. The factory-default APR data is available in several
languages.
* Date / Time Allows the date and time to be set or changed.
Setup
* Receptor Allows predefined receptor symbols to be assigned to each image
Symbols receptor button.
* Touch Screen • Allows for the setting of specific console operating
Setup parameters.
• Sets up the serial communication ports on the touch screen
console.
• Allows adjustment of the sound volume.
• Enables / disables compatible equipement (i.e. infimed digital
interface).
• Allows the operator and service passwords to be changed.
* Touch Screen Allows for electrical alignment of the touch sensitive membrane
Calibration with the “buttons” displayed on the touch screen.
* Data Link Prepares the console for communication with an external
computer. This is a service tool intended for use by authorized
service personnel only.
- Main Menu Press to return to the main console menu.
Note: It is strongly suggested that you review the operator manual (2406878-100) before proceeding. A
good understanding of the terminology and APR menu structures is needed to make APR changes.
1.) From the system utilities menu, press APR Editor. A screen similiar in appearance to the
normal operating screen will be displayed as below. However, as a reminder that you are in
2.) Refer to the applicable subsections for the procedures to change parameters and technique
for existing APR items, and to edit, add, or delete APR, procedural, or menu items.
Note: When a region of interest (i.e. SKULL) is selected, a pop-up menu will appear similar to that in
normal APR mode. In addition to “new” ADD, EDIT, and DELETE buttons, a MOVE UP and
MOVE DOWN button will be displayed.
The MOVE UP and MOVE DOWN buttons allow the items in a menu or submenu to be
rearranged. To do this, select the item to be moved. This will highlight the item. Press MOVE
UP to move the selected item up in the list, and MOVE DOWN to move the selected item down
in the list.
Note: When editing or adding an APR item, menu item, or procedural item, an English keyboard will
pop up at the bottom of the screen. A partial keyboard with international symbols may be
displayed at the top of the screen. To display the desired international characters, select the
language of your choice from the list near the top right side of the screen. Characters may then
be entered via either keyboard.
Be aware of the generator and x-ray tube limits when setting the parameters. Failure to do so
may cause a limit error when the edited procedure is chosen on the touch screen.
2.) Holding the mouse over any “Item“ will show a pop-up window containing the same information
as in the “Properties“ panel but in a more concise way as shown below.
3.) Use the APR Data panel on the right hand side of the APR Editor to insert new Menus ( ),
Procedures ( ) and / or Items ( ).
Menu = - can contain other Menus (i.e. SKULL, SINUSES, FACIAL BONES), Procedures
(i.e. SKULL PROC) or Items (i.e. SKULL TOWNES, SKULL LAT).
Procedure = - contains a list of items. Upon selecting the first Item in a Procedure and
completing an exposure the Touch Screen will automatically select the next Item in the list and
so on until the entire list of items under the selected Procedure is exhausted.
Item = - anything contained within Menus and / or Procedures that is not a Menu or a
Procedure (i.e. SKULL TOWNES, SKULL LAT, SKULL BASAL, SKULL ZYGO.ARCH)
4.) To edit an item in the Table of APR click the selected item once, after the appearance of the
pop up window containing all parameters values for the selected item, click again, the item
should be available for editing. (see below)
b.) After the pop up info show up click the item again.
5.) Use the Properties panel on the left hand side of the APR Editor to modify the parameters for
each item and /or procedure.
6.) Select the parameter value to modify on the Properties panel and double click it. A drop down
box will be displayed containing a range of values available for the selected parameter. Notice
that only the items parameters values can be modified. Menus and procedures do not contain
any parameters.
7.) Changing from small to large Focus, for the pediatric size patient of the Skull Townes item.
a.) Select the Skull Townes item.
b.) Double click the parameter to be modified (Focus in our case) and select the desired
value.
1 2 3
4 5 6
The current image receptor symbols may be replaced with predefined symbols chosen from the
receptor symbol library.
Each image receptor button will always select a predefined image receptor (i.e. table Bucky, wall
Bucky, DR, etc.). Before changing the image receptor symbols, it must be clearly understood which
image receptors are selected by each image receptor button. Each image receptor button should
then have a logical and intuitive symbol assigned to that position.
1.) From the system utilities menu, press Receptor Symbols.
2.) A pop-up window will display as above showing the image receptor buttons with the currently
assigned symbols near the right side of the receptor symbols window, and the library of
available receptor symbols near the left side of the window.
3.) Select the image receptor button for which the symbol is to be changed. The receptor numbers
in the figure above correspond to the receptor numbers.
4.) Select an appropriate symbol for the selected receptor from the symbols library..
5.) Press to assign the selected symbol to the selected image receptor button.
6.) Repeat steps 3 and 5 for each image receptor whose symbol is to be changed.
7.) Press OK to continue or CANCEL to return to the system utilities menu without making any
changes. If OK was pressed, a pop-up window will display asking if you wish to save the
changes. Press YES to save the changes. Press NO to return to the system utilities menu;
CANCEL cancels the changes.
8.) To delete unused symbols from the receptor symbols library, select the symbol to be deleted.
The selected symbol will be highlighted.
- Press .
- A pop-up window will display asking if you wish to delete the selected item. Press YES to
delete the item; NO cancels the deletion.
- DO NOT DELETE SYMBOLS YOU MAY WANT TO USE IN THE FUTURE.
Settings
1.) Program the functions defined in the table above by checking or unchecking the applicable
item.
2.) Program the Digital Port to COM1. This selects the communication port for the optional digital
interface i.e. the InfiMed imaging system.
3.) Program the DAP Port. Use default.
4.) Program the Data Link Port to COM2.
5.) Program the Languages. This selects the language for operator and error messages, and for
text on the buttons, etc. Voice messages may only be availabe in English.
6.) Select the Generator Type as Indico 100.
7.) Set the Screen Saver Interval as OFF. This sets the time from the last activity on the touch
screen until the screen saver is activated.
8.) Set the Sound Volume by dragging the slider to the left (lower) or right (louder). This sets the
Features
1.) From the touch screen setup window, select FEATURES. Available digital interfaces will be
shown on the screen.
2.) Check GE Hualun Digital Interface to enable the function.
3.) Press OK to continue or CANCEL to return to the system utilities menu without making any
changes. If OK is pressed, a pop-up window will display asking if you wish to save the
changes. Press YES to save the changes. Press NO to return to the system utilities menu;
CANCEL cancels the changes.
Passwords
Be sure to record the new service password before changing the factory-default password. If
the new password is subsequently lost, consult the factory or re-install the touch screen
software in order to restore the factory-default service password.
6.) To change the service password, press the ... button to the right of the top service password
line. A password window will pop up. Enter the new password.
- Press Clear to cancel an incorrect entry.
- Press Cancel to close the password pop-up window.
- Press Accept to accept the new password.
7.) Repeat the previous step to enter the new password on the second service password line.
8.) Press OK to continue; CANCEL cancels the changes.
9.) A pop-up window will display asking if you wish to save the changes. Press YES to save the
changes. Press NO to return to the system utilities menu; CANCEL cancels the changes.
Error Logs
The error logs function records console errors only. This screen exists for CPI debugging purposes.
The information recorded will generally be of no value to field-service personnel.
- If the calibration was done properly, you will be taken to the main menu.
2.2 Recovery
If necessary, do recovery with the recovery disk which includes Operating System, Service
Package, Drivers, GenWare, PCU, and Image Brightness Measure installation softwares.
Suggest recovering system with the recovery disk if necessary.
If you want to intall the softwares respectively, please refer to Precision THUNIS-800+ Restricted
Service Manual (2406879-100).
1.) Insert the recovery disk into the CD-ROM, and close the CD-ROM and reboot the workstation.
2.) When a time counter appears on the left-top corner, press “enter“ key to start the recovery
process. At the same time, there will be a click sound from the PC speaker. If the site uses a
CRT monitor, there will be nothing on the screen, press “enter“ after the click sound.
3.) Waiting, after recovery finishes, system will reboot automatically. If the site uses a CRT
monitor, and cannot see anything in the screen, when the click sound appears for the second
time, do nothing until the sound disappears, and boot from the harddisk.
5.) Enter “Control Panel” in the path “Start -> Settings -> Control Panel“.
6.) Select the third tab in “Regional and Language Options“, and select the compatible language
in the “Select a language to match the language version of the non-Unicode program you want
to use“.
2.3 DSA
After system recovery, please install DSA software.
2.3.2 Uninstallation
English/Chinese version of DSA software is factory installed. You should uninstall the English/
Chinese version, and install the English/Local Language version.
6.) After remove DSA from control panel, go to regedit to remove DSA completely.
7.) Run “regedit“ from “Start -> Run“.
2.3.3 Installation
English/Chinese version of DSA software is factory installed.
2.) Accept the terms in the license agreement", and click "Next".
3.) Install, the setup wizard will register the software to OS automatically.
2.) Input the Password in the popup window, and click "OK".
Note: Please refer to Precision THUNIS-800+ Restricted Service Manual (2406879-100) to get the
password.
4.) Input report title in the "DSA Config Tool" interface, and setup password as customer
requirements.
5.) Select one modality. Usually use default. If there is any other products use the modality name
as default in hospital, please change it to another one.
6.) Select "Apply DSA" checkbox to activate DSA if ROCKEY has not been activated yet, input
the Serial Number.
Note: If customer order DSA, please get the letter from customer, open it, input the serial number,
and activate DSA.
If DSA is activated before, the serial number of DSA has already stored in the ROCKEY, so
DO NOT click “Apply DSA“ during re-installation.
7.) Select "Apply DICOM" checkbox to activate DICOM if ROCKEY has not been activated yet,
input the Serial Number..
If DICOM is activated before, the serial number of DICOM has already stored in the ROCKEY,
so DO NOT click “Apply DICOM“ during re-installation.
8.) Click "Reversion" to restore the patient database, select the version customer needs and press
"Reversion" in the popup "Database Reversion" dialog, and then, the computer will restart.
2.) Input the Password in the popup window, and click "OK".
Note: Please refer to Precision THUNIS-800+ Restricted Service Manual (2406879-100) to get the
password.
4.) Click "OK" to close the error dialog, click "Start". Step the right foot switch to get the frame, and
confirm there is an image.
8.) Re-start the computer. If DSA software is auto log on, the configuration is completed.
Chapter 4 - Configuraton & Calibration Page 181
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
2.3.6 Configure DSA / DICOM Function Environment Parameters
1.) Run dsa.exe in D:\dsa. The main interface will be full of the screen. Supposing that there is a
patient and image series, click "DCM Store".
2.) Input all the correct information of the PACS server supplied by hospital in the "Dcm Store
SCU" popup window, and then, click "Verify" to test the network connection.
3.) "Verification Success" dialog will popup in several seconds if the information is correct, That
means PACS server information has been configured now. Click "OK" to close it.
6.) Input the service password in the popup dialog as you set before.
7.) The "Settings" window will popup. Set the parameters as operators requirements in “Print
Scheme Setup“ tab.
8.) Input all the correct information of the DICOM Printer supplied by hospital in the "Server Setup"
11.) Input all the correct information of the DICOM Printer supplied by hospital in "WorkList SCU",
and then, click "Verify". If the information is correct, the connection successful information
b.) Select the location you prefer to save the backups, give a name (such as “Generator
Parameters“), and then, click “Save“. Sugget to save the backup file to D:\Service Tools\.
b.) Restore the factory configured backed-up file from D:\Service Tools\.
Note: Check the settings after downloading the parameters as below.
A196
RAD14
6.) To upload an APR file from a USB flash disk to the touch screen select the desired file (ex.
PS800+English-V01.apr) from the USB Device window and press . Once the file is
copied successfully, it will appear in the Touch Screen window of the USB File Transfer Utility.
7.) Using this way, upload all APRs to touch screen.
8.) When finished, press Close to exit to window "APR Backup and Restore".
9.) Delete the original APRs of touch screen. Such as "ENGLISH.APR"
10.) Restore "PS800+English.apr" as current APR.
11.) When finished, press Close to exit and go to main screen.
Section 3.0
Image
3.1 Tube Conditioning / Seasoning
Tube conditioning or “seasoning” is particularly important for new tubes or tubes that have not been
used for several days. This should be performed on each X-ray tube before attempting auto
calibration, as an unseasoned tube may not operate properly at higher kV values without arcing.
Note: Close the collimator blades and put a folded Lead Apron on the table-top to protect I.I tube.
Select “Receptor 3“ or “Receptor 4“ before tube conditioning / seasoning.
3.1.1 Overview
The generator does X-ray tube auto calibration at 50 kV, 60 kV, 70 kV, 80 kV, 100 kV and 120 kV.
The tube normally needs to be seasoned before it can be operated at the higher voltages
encountered during auto calibration.
Tube seasoning is started by auto calibrating the kV stations up to and including part of the 70 kV
station. The tube is then seasoned at 70 kV. Progressively higher kV stations are then auto
calibrated and seasoned. Finally the entire kV and mA range is auto calibrated, then the tube is
seasoned at the remaining high kV values.
Manually releasing the exposure button during auto calibration of a particular kV station in the
following procedure prevents the generator from attempting operation beyond that kV/mA value.
Note: Low speed only exposures are recommended for the seasoning exposures , to prevent excessive
hear build-up in the housing from the stator windings or the rotor bearings.
3.1.2 Procedure
Note: X-ray tube seasoning should be done on LARGE focus in order to minimize tube wear. And then,
SMALL focus.
1.) Run Genware Generator Utilities software, and open Setup -> Auto Tube Calibration.
Note: The procedure below is intended for seasoning an X-ray tube prior to attempting tube auto
calibration. To season a tube that does not need to be calibrated, simply follow steps b, d, f, h,
and i.
a.) Start the tube auto calibration sequence , and manually terminate the exposure at 70
kV and 250 mA.
b.) Season the tube at 70 kV by taking approximately 10 exposures of 200 mA and 100
ms. These exposures should be taken at the rate of approximately one every 15 seconds.
c.) Restart the auto calibration sequence and manually terminate the exposure at 100 kV
and 250 mA.
d.) Season the tube at 100 kV by taking approximately 5 exposures of 200 mA and 100 ms.
These exposures should be taken at the rate of approximately one every 15 seconds
e.) Restart the auto calibration sequence and manually terminate the exposure at 120 kV and
160 mA.
f.) Season the tube at 120 kV by taking approximately 5 exposures of 160 mA and 100 ms.
These exposures should be taken at the rate of approximately one every 15 seconds.
g.) Restart the auto calibration sequence and allow the auto calibration sequence to
complete.
h.) Season the tube at 130 kV by taking approximately 5 exposures of 100 mA and 50 ms.
These exposures should be taken at the rate of approximately one every 15 seconds.
i.) Repeat step 8 at 140 kV, and then at 145 kV.
3.) Do the calibration of SMALL focus by repeating the above steps.
3.2 ABC
Note: For the details, refer to in Class M Service Manual (2406879-100).
PCU
Generator
J8
Gen I/F bd
J4 Co-axis cable
AEC Channel 1
AEC Channel 2
3.) Place 15cm Plexiglas on positioner table, make sure it cover the FOV.
4.) Power up system, and ready for digital spot.
Note: If the digital spot or DSA exposure technique shall be adjusted, a simpler way is to adjust the
density value on touch screen.
The density value can be set to: +5, +4, +3, +2, +1, 0, -1, -2, -3, -4, -5.
Set density to a value > 0 will increase exposure time in AEC mode.
Set density to a value < 0 will decrease exposure time in AEC mode.
Section 4.0
Network
4.1 DICOM Configuration
Please refer to step 7 in Setup and Activation on page 173.
Section 1.0
Positioner Table
1.1 Main Function
1.2 Architecture
Refer to Digital System Table Schematics on page 289.
Trigger Signal
Logic board
Logic Calculation
Trigger signal, such as limit switch, console joystick, table side button, footswitch and so on, input
to Logic Board. After logic calculation by Logic Board, signal will output to X-ray Control Board and
Motor Relay Board.
X-ray Control Board will achieve collimator blade close/open and communicate with TDSP Board.
Motor Relay Board will drive four movements. Tilting movement and Tube arm up&down movement
are drived by Frequency Inverter (3 phase motor); Table top left&right movement and Compressor
up&down movement are drived by 1 phase motor.
Section 3.0
PDU
3.1 Fuction
Power supply to all subsystems include integrated console, positioner table, tube fun, and
generator.
• 220VAC for console;
• 220VAC and 110VAC for positioner table;
• 220VAC for tube fun;
• 380VAC/400VAC for generator.
3.2 Architecture
Refer to Digital System PDU Schematics on page 294.
Console
220VAC
Positioner table
Tube fun
380VAC/400VAC Isolation transformer
Console
110VAC
Positioner table
Tube fun
Generator
Section 4.0
Integrated Console
4.1 Function
4.1.2 Intercom
- Console Mic to table speaker;
- Table Mic to Console speaker.
4.2 Architecture
Refer to Digital System Console Schematics on page 296.
TDSP board
WW/WL adjustment interface
Touchscreen parameter
techspot X-ray on
signal
Console interface
PCU realtime
Digital box
control
realtime interface
I.I
Generator room interface
Section 5.0
Image Chain
5.1 Main Feature
• Video output.
• Real time image processing with flexible setting.
• Various CCD operating modes up to 1000x1000 pixels/30 frames per second.
• Internal or external synchronization.
• Defective pixel correction.
• Motorized iris and neutral density filter.
• Light measurement path for Automatic Exposure Control.
XRII:TH9428 HP2
Optics
Compact Camera control
lens head board
X-rays TH 8740-020
Light
measurement
HVPS: TH 7195-3 path
Processing
and Control
Unit
EQUIPMENT
PDB
OCB
LVDS / 1Kx1Kx12bit LVDSÆCamera link
I.I
Optics
Tube PCI/ Frame Grabber
Grid
Console IF board PCI/ 1750 card
Digital BOX
FootSW and Spot signal
Monitor
PRECISION THUNIS-800+ SERVICE MANUAL
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
Section 6.0
Digital Box
The Precision THUNIS-800+ digital box mainly consists of digital Imaging workstation, a control
interface board, and image frame grabber.
An analog electrical signal is obtained when the X-ray penetrates through human tissue and passes
by the image intensifier and TV camera. Such a signal is output to D.I.W where it is amplified, and
then digitized at a high speed by image frame grabber. In order to real-time display and acquire
images, the system is required to configue a high-performance and high-speed PC system,
including digital image processing unit, analog video-frequency logarithm transformation, and
image frame grabber.
The control interface board functions to guarantee the synchronization between X-ray machine’s
exposure and image acquisition so as to complete the digital spot.
The digital fluoroscopy and digital abstraction function of the system can easily be used to guide
the direction of cathetherization in the process of intervention.
The digital subtraction means to subtract normal-structure images of angiography and only display
blood vessel images. Therefore, besides the after-angiography images, the normal-structure
images before angiography are also needed.
Image output modes include image printing, image recording, and laser photography etc. Printing
is mainly used for preparing a clinic report. The printed image can only be used for reference owing
to the restriction of paper and printer. Disc recording may be used for long-term image storage and
resource sharing. The output interface of laser camera is in accordance with DICOM3 standard,
which is convenient for connection to PACS or HIS.
Chapter 6 - Troubleshooting
GE MEDICAL SYSTEMS
Table Top Move >2, or 1->3 is 220 VAC Y Check Motor Y Change Motor or
Function Disable When Joystick or Table Cable capacitor
Table Movement
Side Button
Function Disable
N
N
N N
Re-connect Cable or
Check Console Joystick or Table Re-assemble
Check Joystick or Table
N Side Button Input “BXXX” or Mechanical Part
Side Button Connection
“CXXX”. Is it charged when
of Change
Joystick or Table Side Button is
Corresponding Ones
Activated?
Change Inverter
PRECISION THUNIS-800+ SERVICE MANUAL
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
1.1.2 Tube Fan Not Work
Cable N
Change Cable
Connected?
END
Press Exposure
Restore GenWare
Factory
Configuration
No Lights on either
N Change the Lamp
Console Side or Remote
Switch
Control?
N
N
Check PDU
Change Transformer
Calibrate Collimator by
Refer to Collimator Y
Calibration, No FOV
Indication Error?
Change Collimator
1.2.1 Process
1.) In “Remote“ console control state, press “Film clear“, “ID No. Set+“ and “Film No. Set+“ button
together.
2.) Check the 4 Remain indicator state. Normal--”OFF”, Diagnose--”ON”.
3.) Press “ID No. Set-“ or “ID No. Set+“ button to switch IO monitor contents.
4.) Press “Film clear“, “ID No. Set+“ and “Film No. Set+“ buttons together again. Turn back to
diagnostic program.
SEG2:A
bit3:1 bit2:0 bit1:1 bit0:0
limit_sw(11) limit_sw(10) limit_sw(9) limit_sw(8)
Compress_limit_sw UD_range_sw Down_limit_sw Up_limit_sw
“1“@>Compressor up “0“@<Tube UD range “1“@Tube down limit “0“@Tube up not limit
limit
SEG3:0
bit3:0 bit2:0 bit1:0 bit0:0
limit_sw(7) limit_sw(6) limit_sw(5) limit_sw(4)
Right_limit_sw Left_limit_sw Reverse_degree_sw Vertical_over_sw
“0“@Top Right not limit “0“@Top Left not limit “0“@>-18 degree “0“@<90 detent
SEG4:2
bit3:0 bit2:0 bit1:1 bit0:0
limit_sw(3) limit_sw(2) limit_sw(1) limit_sw(0)
Incline_over_sw Incline_stop_sw Level_stop_sw Vertical_stop_sw
“0“@>-30 detent “0“@>-30 detent “1“@<=0 detent “0“@<90 detent
Result:
1.) It shows limit switch states.
2.) Compressor is at the Up limit position. And the Tube arm is at the Down limit position.
3.) Tabletop isn’t at limit position, and it is at >-18 degree.
4.) Tabletop is at <=0 degree detent position.
1.3.1 BOP
Start Pressing the button starts the converter. This button is dis-
converter abled by default.
Activate the button: P0700 = 1 or P0719 = 10 ... 15
Stop OFF1 Pressing the button causes the motor to come to a
converter standstill at the selected ramp down rate. This button is
disabled by default.
Activate the button: P0700 = 1 or P0719 = 10 ... 15
OFF2 Pressing the button twice (or once long) causes the
motor to coast to a standstill.
This function is always enabled.
Change Press this button to change the direction of rotation of the
direction motor. Reverse is indicated by a minus (-) sign or a flashing
decimal point. Disabled by default.
Activate the button: P0700 = 1 or P0719 = 10 ... 15.
Jog motor In the "Ready to power-on" state, when this key is pressed,
the motor starts and rotates with the pre-set jog frequency.
The motor stops when the button is released. Pressing this
button when the motor is running has no effect.
2
Press until P0003 is displayed
3
Press to access the parameter value level
4
Press or to the required value (example:
3)
5
Press to confirm and store the value
Note: The following important restrictions should be considered when using the Cloning procedure:
- Only the current dataset is uploaded to the BOP.
- Once the cloning procedure has started, it cannot be interrupted.
- It is possible to copy data from inverters of different power and voltage ratings.
- During download, if the data is not compatible with the inverter, the default
- values for the parameter will be written to the inverter.
- During the cloning process any data already held by the BOP is overwritten.
- If the download or upload of data fails, the inverter will not function correctly.
Fault Significance
F0001 Overcurrent
F0002 Overvoltage
F0003 Undervoltage
F0004 Inverter Overtemperature
F0005 Inverter I2t
F0011 Motor Overtemperature I2
F0051 Parameter EEPROM Fault
F0052 Powerstack Fault
F0060 Asic Timeout
F0072 No Data from USS (RS485 link) during
Telegram Off Time
F0085 External Fault
Alarms Significance
A0501 Current Limit
A0502 Overvoltage limit
A0503 Undervoltage Limit
A0505 Inverter I2
A0511 Motor Overtemperature I2
A0910 Vdc-max controller de-activated
A0911 Vdc-max controller active
A0920 ADC parameters not set proerly
PDU Output N
Refer to PDU Check
Normally?
Generator 3 Main N
Change Fuses
Fuses Work
N Check AC Mains
3 Phase Power
Power Supply from
Works Normally
Hospital
Refer to Generator
Indication LEDs in
Generator Service
Manual
Section 3.0
PDU
Whether have N
Power Up again
Main Power Input
N
CB1 Closed? Close CB1
N
CB2 Closed? Close CB2
Fuse F1 Works N
Change Fuse F1
Normally?
Change Indication
Lamp
END
turn up volume,
still no voice?
console
Mic cable and N
Re-connect or change cable
table speaker
cable Ok?
footswitch N
Change footswitch
Ok?
Change intercom
board
turn up volume,
still no voice?
table Mic
cable and table N
Re-connect or change cable
speaker cable
Ok?
Change intercom
board
Power-Up N N
Console Fuse
Indication LED Change
Works?
Light
Y Y
PDU Output N
Circular LED
Breaker CB6 Close
Flash
Closed?
END
Page 230
GE MEDICAL SYSTEMS
Console Light is Abnormal
Y Y Y
N
Change TDSP Bd Display “0000” Change TDSP Board
OK
PRECISION THUNIS-800+ SERVICE MANUAL
4.4
GE MEDICAL SYSTEMS
Emergency Fault
Y N Change Button or
Cable
Chapter 6 - Troubleshooting
N
Check A2-RY1 Relay is OK? Change Relay
Check Cable
PRECISION THUNIS-800+ SERVICE MANUAL
Page 231
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
4.5 Remote Function Fault
Remote Function
Fault
N N
Reset DIP Swtich
N
Y
Chapter 6 - Troubleshooting
Change Touch
Screen
PRECISION THUNIS-800+ SERVICE MANUAL
Page 233
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
Section 5.0
Image Chain (I.C.)
Trouble Action
No fluoro image displayed in 1.Check if monitor is powered correctly.
monitor on Analog System 2. If so, check is there are characters displayed on bottom left of the moni-
tor.
3. If so, the monitor and PCU are correct, chech if generator work status.
4. If there are no characters displayed on monitor, check if the cable con-
nect CCD camera and PCU connected correctly.
5. If so, connected PCU to COM port of PC, run TH8740.exe, go to display
page, check if “Analog video standard“ is set to item 3: progressive. If not,
change it to item 3.
6. If PCU can’t be connected to control software, the PCU is defective,
replace a new one.
The image is not as good as 1. Check PCU setting according to TST file.
usual 2. Check if exposure technique, kV mA, mAs are resonable.
3. If not, check PCU setting according t oTST file and check generator set-
ting according to TST file. Set PCU and generator parameters again and
redo AEC, ABC calibration when necessary.
Fluoro ABC fail, fluoro kV, mA 1. Check if ABC cable is connected correctly from PCU to generator.
reached highest value 2. If so, check if ABC jumper on generator interface board are set cor-
rectly.
3. Check PCU setting according to TST file.
AEC exposure fail, exposure is 1. Check if AEC cable is connected correctly from PDB board to generator.
cut off by back up time AEC channel (top one) shall be used for Analog and Digital System. AEC
channel 2 shall be used for DSA of Digita System.
2. Check that J2 of PDB shall be used for Digital System, and J3 of PDB
shall be used for Analog System.
3. If so, check if AEC jumper on generator interface board are set cor-
rectly.
4. Connect to Generator by Genware, check if receptor is set according to
TST file.
5. If all above arecorrect, disconnect AEC cable from generator side, take
a fixed exposure and measure the output of Thales PDB board by oscillo-
scope, there shall be DC voltage during exposure. If not, the PDB or optics
are defective.
6. Replace a new PDB and do test above again, if pass, the PDB is defec-
tive, if not, the optics are defective.
The orientation of image is not Press PCU OSD button to enter PCU’s OSD menu, change vertical&hori-
correct zantal inversion state of PCU until the orientation of image is correct.
Display error on 20" monitor in Check the input selection switch on the back of the monitor. There are
Digital System both analog video input and PC VGA signal input on the back of the
monitor. The switch must be correctl set.
Y
OK? OK
Change Image
Convert Bd
Download GenWare Y
Factory Parameters (*.gen) OK
Change Generator
Interface Board
No AEC Function
Download GenWare Y
Factory Parameters (*.gen) OK
is OK?
AEC Cable Y
Connection is OK
OK?
Change Photodiode
Board of I.I.
No Image Acquisition
with X-ray Exposure
DICOM Print UI Y
Wait or Close Print UI
Running
Y Restart DSA, Y
END
Whether OK?
Y N
N N
Change Frame
Grabber Card and
Image Converter
Board
Y Y
Have X-ray Collimator Open?
No Image during Fluoro
Y Perform as Error
Have Error Codes
Codes
Chapter 6 - Troubleshooting
Room I/F Mis
DSA Progressing Y N Re-connect or
Cable
Bar Works? Change Cable
Connected?
N Y
Y N
Page 239
GE MEDICAL SYSTEMS
DIRECTION 2406876-100, REVISION 7 PRECISION THUNIS-800+ SERVICE MANUAL
5.6 No Spot Function
No Spot Function
Refer to
Have X-ray Exposure Y troubleshooting “No
When Joystick Spot? Image Acquisition
with X-ray Exposure”
Y Check Generator
Anode Works?
Cable Connections
Y
Y
Y
Y
Re-connect Cable
from Console I/F
Board to Generator OK
Room I/F Board
Type Code
Error 0x01
Warning 0x10
Information 0x11
Location Code
Patient management 0x01
Image acquisition 0x02
Image processing 0x03
Patient report 0x04
Recording CD 0x05
Dicom 0x06
Power on system
Y
See (2)
N Enter Digital main screen
Y
N
Digital software is damaged; Operation of patient history management,
reinstall it. eg. add, store, or delete.
Y Digital software is
Connected. During fluoro, Digital damaged; reinstall it.
N monitor displays real-time fluoro
images.
Y
See (3) Connected. During fluoro, press down the
fluoro acquisition key, and acquisition is N
normal.
Y See (4)
N
Connected. Exposure
and acquisition images
are normal.
Y
See (5) N
Image printing is normal.
Y
See (6)
Normal image writing
N
Y See (7)
OK
(2)
N
Characters displayed
Y N N
Enter
Main-board
CMOS
Reconnect power
cord, reconnect Y
signal lines, or
N Video card Replace
Y replace the monitor.
damaged; replace main-board
Caused by abnormal it. End.
power-off, quit it.
N
Y
End Replace hard disk and
reinstall system software.
End.
(3)
Y
Confirm all set values
of GRABSETUP are Operate PCITEST and End
correct. press the fluoro footswitch;
PCITEST displays.
Y
N
Normal operation of
Press the fluoro footswitch, and the
EsayGrab
control box will give a sound instruction.
Y
N
Y
Data acquisition cable Incorrect CAM file storage
with bad contact or path (correct path: N
damaged D:\DSA\a201b.cam)
(4)
(5)
N
Press down the hand brake,
Press down the hand brake, and EXP and X-ray machine is exposed.
N
lamp of the interface board shines.
N
Y
Y
Exposure time: 33ms
Interface board Y
damaged; replace it
Readjust AEC
parameters
N
Acquisition is normal,
Interface board End
but images are tool
damaged; replace it.
bright or dark.
Faulty CCD
operating mode
N End
Y
(6)
Normal printing
N
Y
N
N
Operate dsasetup, and
reset the printer curve.
(7)
N
Insert a new writable
CD
Can CD writing work be
OK
ended?
Section 1.0
General
To use this equipment with an optimal situation at any time, the periodic maintenance must be
required.
Section 2.0
Preventive Maintenance Action List
Operation Location Description Done
1.) General
General cleaning and painting Chapter 7 - Section 3 Preventive Maintenance - 3.1.1
Visual inspection Chapter 7 - Section 3 Preventive Maintenance - 3.1.2
2.) Positioner Table
Equipment appearance check Chapter 3 - Section 2 Functional Check - Step 1
Room doorlight check Chapter 3 - Section 2 Functional Check - Step 2
Switch check Chapter 3 - Section 2 Functional Check - Step 2
Indicator lamp check Chapter 3 - Section 2 Functional Check - Step 2
Lever Operation Chapter 3 - Section 2 Functional Check - Step 2
Indicator check (Digital Display) Chapter 3 - Section 2 Functional Check - Step 2
Insulated resistance check Chapter 7 - Section 3 Preventive Maintenance - 3.2.1
Fan Chapter 3 - Section 2 Functional Check - Step 5
Rotor and collimator cable check Chapter 3 - Section 2 Functional Check - Step 5
Collimator beam limiting adjustment Chapter 4 - Section 1 Configuration & Calibration - 1.5
Collimator field lamp illumination Chapter 7 - Section 2 Preventive Maintenance - 3.2.2
Collimator field lamp and timer Chapter 3 - Section 2 Functional Check - Step 3
Collimator blade movement Chapter 3 - Section 3 Functional Check - Step 3
Collimator mounting screws check 1. Check the flange
2. If the collimator mounting screws
are loose, apply LOCTITE onto the
screws.
Tilting stop check Chapter 3 - Section 2 Functional Check - Step 4
Tilting limit switch check Chapter 3 - Section 2 Functional Check - Step 4
Tilting transmission oil leak check Chapter 3 - Section 2 Functional Check - Step 4
Tilting motor V-belt check Chapter 7 - Section 3 Preventive Maintenance - 3.2.4
Tilting guide bearing check Chapter 7 - Section 3 Preventive Maintenance - 3.2.5
Table lateral movement check Chapter 3 - Section 2 Functional Check - Step 4
Table lateral movement gear check Chapter 7 - Section 3 Preventive Maintenance - 3.2.6
Compression force check Chapter 7 - Section 3 Preventive Maintenance - 3.2.7
Compression cone limit check Chapter 3 - Section 2 Functional Check - Step 4
Compression cone roller check Chapter 7 - Section 3 Preventive Maintenance - 3.2.8
Intercom foot switch Chapter 3 - Section 2 Functional Check - Step 2
Microphone Chapter 3 - Section 2 Functional Check - Step 2
Intercom volume Chapter 3 - Section 2 Functional Check - Step 2
Foot stand Chapter 3 - Section 2 Functional Check - Step 14
Shoulder Rest Chapter 3 - Section 2 Functional Check - Step 14
Hand Grip - -
3.) Generator
Section 3.0
Preventive Maintenance
Maintenance Is To Be Performed Only By Competent, Trained Personnel Who Are Familiar With
The Potential Hazards Associated With This Equipment.
Maintenance Activities In The Table Below Are To Be Performed By Authorized Service Personnel
Only. However The Owner Or Operator Of The Equipment Must Schedule The Suggested Mainte-
nance Activity When Required By Contacting Their Local Service Agency.
Note: Maintenance schedule frequency may be determined by certain regulatory requirements of the
country or state in which the installation is located. Always check the local codes and regulations
when determining a maintenance schedule.
Always Switch Off Mains Power To The Generator And Wait A Minimum Of 5 Minutes For Capaci-
tors To Discharge Before Beginning Any Preventative Maintenance, Including Cleaning.
Observe Esd Precautions. Keep All Static - Sensitive Components And Circuit Boards In Their
Static - Shielding Packaging Until Ready To Install. Ensure That You Are Grounded At All Times
When Handling Static - Sensitive Components And Circuit Boards.
3.1 General
3.1.1 General Cleaning and Painting
Keep all surfaces clean and touch-up paint chips.
Open System
Breaker
Voltage between
Power input and
Ground
3.) Remove lid at the back of X-ray tube stand. Using 152R band, adjust compression force
4.) Use only fresh oil, type Shell DIALA AX or equivalent. It is critical that air is not added when
topping up the oil. The following procedure is strongly recommended when adding oil.
- Use a new clean syringe to remove oil from the container. A 60 cc catheter tip syringe is
recommended. Approximately 60 cc of oil is required to raise the oil level by one
millimeter.
- Turn the syringe upright and expel any trapped air.
- Place the tip of the syringe through the oil-fill plug and into the oil, ensuring that it is below
the surface of the oil.
- Gently eject the oil from the syringe into the HT tank, while making sure that the tip of the
syringe remains below the surface of the oil until all of the oil is emptied from the syringe.
- Repeat the previous steps until the required amount of oil has been added.
5.) Replace the oil fill plug. Once the plug is installed and the screw properly seated, tighten the
screw 4 turns. This will secure the oil fill plug. Wipe up any oil spills. Dispose of soiled absorber
in compliance with government requirements and ensure conformity to local disposal
regulations. THE OIL DOES NOT CONTAIN PCBs.
3.3.3 Cleaning
Note: Refer to General Cleaning in section 3.1.1.
Do Not Disassemble, Incinerate, Or Short-circuit The Battery(S) In This Product. Do Not Put It
In Trash That Is Disposed Of In Landfills; Dispose Of It As Required By Local Ordinances.
The Fluorescent Lamp In The Lcd Display Contains Mercury. Do Not Put It In Trash That Is
Disposed Of In Landfills; Dispose Of It As Required By Local Ordinances.
The Lcd Is Made Of Glass. If The Lcd Breaks Due To Rough Handling Or Dropping, And The
Internal Fluid Gets In Your Eyes Or On Your Hands, Immediately Wash The Affected Areas
With Water For At Least 15 Minutes. Seek Medical Attention If Any Symptoms Are Present
After Washing.
The Aec And Abs Verification Procedures Require The Production Of X-rays. Observe Correct
Operating Procedures, And Take Appropriate Precautions Against X-radiation.
3.4.1 AEC
The following procedure may be used to verify that the AEC circuits are functioning on generators
equipped with AEC (Automatic Exposure Control).
1.) Switch the console and generator ON, and select an appropriate radiographic image
receptor.
2.) Align the X-ray tube and the selected image receptor such that the central ray is directly over
the center field of the AEC pickup device. Set the focal spot to film plane distance to 40 in. (1
m).
3.) Select AEC mode of operation. Select center field, large focus.
MINIMUM EXPOSURE TIME:
3.4.2 ABC
The following procedure may be used to verify that the ABS circuits are functioning on generators
equipped with ABS (automatic brightness stabilization).
1.) Switch the console and generator ON, and select an appropriate fluoroscopic image
receptor.
2.) Select ABS mode.
3.) Open the collimator or beam limiting device shutters to maximum. With no object in the
image field, press the fluoro footswitch. Confirm that the collimator or beam limiting device is
fully open by observing the monitor.
4.) While continuing to press the fluoro footswitch, note the fluoro kV display at the control
console or remote fluoro control. This should be less than 55 kV.
5.) Close the collimator or beam limiting device and press the fluoro footswitch. With no object in
the image field, ensure that the radiation is fully blocked by observing the monitor. If necessary,
cover the image intensifier with lead aprons to block all radiation.
6.) While continuing to press the fluoro footswitch, note the fluoro kV display at the control
console or remote fluoro control. This should be the maximum available fluoro kV, 110 kV or
125 kV (depending on whether the upper fluoro kV limit has been set to 110 kV or 125 kV).
7.) Place an absorber (20 cm of water, or 1.5 in. ( 3.8 cm ) of pure aluminum, or equivalent) in the
center of the image field.
8.) Press the fluoro footswitch. While viewing the absorber on the monitor, open the collimator or
beam limiting device sufficiently to cover as much of the absorber as possible. Ensure that the
X-ray beam does not extend beyond the sides of the absorber as stray radiation will adversely
affect the kVp reading in the next step.
9.) While continuing to press the fluoro footswitch, note the fluoro kV display at the control
console or remote fluoro control. This should stabilize between 70 and 80 kV.
Section 1.0
Positioner Table
F1
F2
F3
F4
Item # 32
Item # 5 Item # 22
Item # 26 Item # 24
Item # 18
Item # 7 Item # 17
Item # 48 Item # 45
Item # 41
Item # 44
Item # 39 Item # 40
Item # 48
Item # 27 Item # 28
Item # 21
Item # 37
Item # 20
Item # 46
Item # 8 Item # 6
Item # 49 Item # 15
Section 2.0
Generator
Item Part No. ePDM No. Name
1 5123270 5123270 Generator CPU Board
2 5123273 5123273 Generator I/F Board
3 5123275 5123275 Room Interface Board
4 5123276 5123276 Control Board
5 5123277 5123277 Filament Board (Small Focus)
6 5123278 5123278 Filament Board (Large Focus)
7 5123279 5123279 Auxiliary Board
8 5123280 5123280 Inverter Board
9 5123281 5123281 Power Input Board
10 5123282 5123282 Resonant Board
11 5123283 5123283 Universal AEC Board
12 5123287 5123287 DSS PWBA
13 5123290 5123290 DC BUS Capacitor
14 5123292 5123292 Line Contactor
15 5123293 5123293 Mains Diode
16 5123294 5123294 Fuse, A70QS10-14F
17 5123295 5123295 Fuse, FNQ-10
18 5123296 5123296 Fuse, FNQ-2
19 5123297 5123297 Fuse, GDC-1.6
20 5123298 5123298 Fuse, GDC-2
21 5123299 5123299 Fuse, GDC-2.5
22 5123300 5123300 Fuse, GDC-5
23 5123301 5123301 Fuse, MDA-2
24 5123302 5123302 Fuse, MDA-7
25 5123303 5123303 Fuse, MDA-12
26 5123304 5123304 Fuse, MDL-4
27 5123305 5123305 Fuse, OTS-60
28 5123306 5123306 Relay
29 5123307 5123307 Connector
30 5123308 5123308 Axial Fan
31 5123309 5123309 Auxiliary Transformer
32 5123310 5123310 Room I/F Transformer
33 5123311 5123311 HV Tank
34 5123312-1 5123312-1 Touch Screen Console
35 5123289 5123289 Digital Input Board
36 5123312-2 5123312-2 Touch Screen Upgrade Kit
Note: If it is the old touch screen (5123312), please order touch screen kit (5123312-2) for upgrade. If it
is the new touch screen (5123312-1), just order the new one for replacement.
Item # 4
Item # 26
Item # 5 Item # 7
Item # 6
Item # 12
Item # 25 Item # 24
Item # 16 Item # 28
Item # 3
Item # 32
Item # 15
Item # 10
Item # 30
Item # 13
Item # 14
Item # 13
Item # 9 Item # 17
Item # 27
Item # 31
Item # 18
Item # 23
Item # 8
Item # 33
Item # 1
Item # 35
Item # 11 Item # 2
Item # 21
Item # 21
Item # 29
Item # 22
Item # 21
Item # 19
Section 3.0
PDU
Item Part No. ePDM No. Name
1 5121163 5121163 Fuse Holder
2 5117119 5117119 Relay (Up Delay 110VAC Coil)
3 5117046 5117046 Contactor (30A)
4 5117120 5117120 Relay (110VAC Coil)
5 5117122 5117122 Relay (24VDC Coil)
6 5120198 5120198 Relay Holder
7 5117136 5117136 LED
8 5117139 5117139 Resistance, R1
9 5117140 5117140 Fuse, F1
Item # 2
Item # 7 Item # 6
Item # 4, 5
Item # 3
Item # 1 Item # 9
Item # 8
Item # 9
Item # 8
Item # 7
Item # 1
Item # 5 Item # 4
Item # 16
Item # 14
Item # 10
Item # 15
Item # 6
Item # 13
Item # 12
Item # 11
Item # 19
Item # 20
Item # 21
Section 5.0
Image Chain
Item Part No. ePDM No. Name
1 5122279 5122279 Optics for Digital RIU
2 5122280 5122280 Photodiode Board
3 5122278 5122278 CCD Camera Head
4 5122282 5122282 TH7195 Power Supply
5 5122284 5122284 PCU
6 5122277 5122277 XRII with Housing
7 5122281 5122281 Optics Control Board
Item # 1 Item # 3
Item # 7
Item # 2
Item # 5
Item # 6
Item # 4
Item # 7
Item # 4, 7
Item # 1 Item # 2, 3
Item # 5 Item # 6
b1
b2
a2
X-ray Field
10.) If pass, it’s ok. If fail, please adjust collimator lamp, and redo alignment of visually defined x-
ray fields until pass.
Section 7.0
Miscable
Item Part No. ePDM No. Name
1 2409055 2409055 TDSP CN331 MIS Cable
2 2409056 2409056 TDSP CN332 MIS Cable
3 2409498 2409498 I.I. Power and Control Cable
4 2408887 2408887 Generator Room I/F Cable
5 2408888 2408888 CCU Generator I/F Cable
6 2409500 2409500 System Console Power Cable
7 2409502 2409502 Table Power Cable
8 5123220 5123220 Image Cnvtr Board Power Cable
10 2409507 2409507 Tube Control Cable
11 2409637 2409637 PDU Power-On Cable
12 2408886 2408886 Cabinet TDSP CN333 Signal Cable
13 2408889 2408889 CPI RS232 Communication Cable
14 2408924 2408924 Console MIC Audio Cable
15 2408925 2408925 Console Speaker Audio Cable
16 2409497 2409497 Speaker Volume Adjust Cable
17 2409496 2409496 Mic Volulme Adjust Cable
19 5122632 5122632 HV Cable
20 2409811 2409811 Table Board Interface Cable 1
21 2409812 2409812 Table Board Interface Cable 2
22 2409813 2409813 Table Tilt Over Switch Cable
23 2409814 2409814 Table Tilt Stop Switch Cable
24 2409815 2409815 Table Limit Switch Cable
25 2409816 2409816 Table I/O Board Power Cable
26 2409817 2409817 Cassette Signal Cable
27 2409818 2409818 Table Relay Control Cable
29 2409820 2409820 Collimator Control Cable
30 2409821 2409821 Table Top Motor Cable
31 2409822 2409822 Table Relay Board Power Cable
32 2409823 2409823 Tilt Inverter Control Cable
33 2409824 2409824 Tube Inverter Control Cable
34 2409825 2409825 Swtich Power Input Cable
35 2409828 2409828 Console Move Control Input Cable
36 2409829 2409829 Console Spot Control Input Cable
38 2409831 2409831 I.I. Power Cable
39 2409832 2409832 Compressor Motor Control Cable
40 2409833 2409833 Collimator Power Cable
41 2409835 2409835 Collimator Control Cable 2
Section 2.0
Symbols/Conventions
The equipment outlines on the map identify where connections are made. These outlines are usually cabinets or major assemblies. Internal divisions
show modules/circuit boards. These components are identified using the component location codes defined for each product. (If you are unfamiliar
with these codes refer to the schematics direction for any major product. It is worth your effort to learn these codes.)
MIS runs are identified by lines between connection points, with the MIS number in an oval at each end of the line.
Section 3.0
Digital System MIS Map
A hardcopy version of system MIS map, 2409796-2SCH, is shipped with system. This schematic shows all of MIS cables and their termination points.
A version appropriate for viewing electronically is provided here.
1 2 3 4 5 6 7 8
C Patient
Remote R&F Table
Table Electronics
Grid 10:1
Film Cassette
D
ROOM Interface
H Image Intensifier
Generator Interface V
P
S Optics Imager
AEC Board CCD
E CCU
AEC Feedback (Photodiode) Camera Digital
Digital Interface ABS Feedback (Proportional DC) Subsystem
CPI
F Touchscreen RS232
XRay Console Console
Thales Inteface
To Gen Room IF &
Hundicmed Digital IF Boards
Varian
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
Integrated Console
65/50KW R&F Generator
A P/N 5123374 Digital Box A
Generator I/F Boards (Workstation)
LCD Monitor
Analog Video_3 Video
In-Room Monitor
X-Ray ON CDR-W
Analog Video_4 MIS 5162261
Warning TB4 @ Room Interface 1750 I/O Card
Light
DICOM
Frame Grabber
Keyboard
B Door
Video Convert Board PC Mouse
/ B
Interlock TB4 @ Room Interface /
PC
Switch /
PWR
COM1 COM2 1
1
Digital Box 1
MIS 5162380
J2 @ Digital IF board MIS 2409362 (
(
MIS 2408888 (
J8 @Gen IF board J4 (ABC Feedback Signal)
II Image Chain
MIS 5122285
{ AEC Channel#2 for DSA Camera Head LVDS J1 (To Camera Head)
Pin7-, Pin12+
AEC Channel#1 for spot MIS 2409797 AEC Feedback Signal
J432
Xray APR Console
II PWR Imager PCU COM2
(Touchscreen LCD)
MIS 2409637 24VDC @Room Interface
D J4 @generator IF board MIS 5123315 Gen
D
GND COM1
Stud MIS 5124835
Xray Tube J4
RS232
HV Tank Anode+ MIS 5122632 To Tube Anode +
E U
Bright Adjustment
E
V HV Tank Cathode- To Tube Cathode -
W
MIS 2409498 J5 (II Power andControl)
MIS 2409492 GND
Tube Stator MIS 2409507 To Stator Contrast Adjustment
GND
MIS 2409493 GND MIS 2409508 Stud Tube Grounding Stud
Console Interface Board
MIS 2409494 HV Tank MIS 5162733 Collimator Lamp SW
MIS 2409495
To Fan PWR 220VAC TDSP Board Fluoro Foot switch
GND Fluoro Switch
J431
Stud
F TB2
U
Cabinet GND J331 MIS 2409055 CN331
Mic Vol Ctrl F
J442
V J5 MIS 2409505-6 J332 MIS 2409056 CN332
W Speaker Vol CTRL
GND In-room speake MIS 2409700-1 J2 (Speaker_2)
J3 MIS 2409502 TM1 Control Room Speaker
In-room MIC J5 (MIC_2)
GND GND
J1 ( ON/OFF Signal) Stud Cabinet GND MIS 2409503 Stud
J531
In-Room Control room MIC
J341 DC PWR:
Footswitch
Hospital TB1 Table Electrical P/N 5121496-1
5VDC, +/-12VDC, 24VDC Intercom FootSwitch
G 360-440VAC U Intercom G
3 Phase V J4 MIS 2409500 220VAC Power GND
Supply W Stud
GND
GND Stud Cabinet GND MIS 2409501 Console GND
Console Table
Power Distribution
Console Cabinet
System PDU P/N 5119248
P/N 5122648
H Connected With Bulkhead
REV MADE BY DATE
REVISIONS
PCN NO. DESCRIPTION
MADE
APPROVALS YY/MM/DD TITLE
1 2 3 4 5 6 7 8
Chapter 10 - Schematics
Section 1.0
Positioner Table
1.1 Digital System Table Schematics
Tilt
Logic IO Board Motor Relay Board Frequency Inverter 750W R/F/I
J4 IO 5120415 U 3Phase Motor
A 2408348-1 2409218 GND V A4 A
1 1 220Vac,4P,750W
Tilt Stop LS Assy J3 A1 B1 A1 B2 2
Tilt_inverter_DO
2
A2 FC1 W M1 5120417-1
A5 G24 Tilt_inverter_DI0
1 3 3
1LS Vertical_stop_sw Tilt_inverter_DI1 P
4 4 4
(Vertical) G24 Tilt_inverter_DI2 E
2 5 5 P 2409826
Level_stop_sw Tilt_inverter_+24V
2LS (Level) 5 6 6 E
G24 Tilt_inverter_G24
3 7 7 L
3LS Incline_stop_sw
6 N
(Incline) IDC 10p 2409823
2409814 Molex 6p Tube
Frequency Inverter 370W U/D
B Tilt Over LS Assy
A4 G24
J5 J5
GND
IO 5120414
U
V A3 3Phase Motor B
1 1 1 A2 FC2 220Vac,4P,370W
4LS Incline_over_sw Tube_inverter_DO W M2
4 2 2 5120416-1
(Incline) G24
2 J12 J12 3
Tube_inverter_DI0
3
5LS Vertical_over_sw Board Interface Cable2 Tube_inverter_DI1 P
5 4 4
(Vertical) G24 2409812 Tube_inverter_DI2 E
3 5 5 P 2409827
6LS Reverse_degree_sw IDC 20p IDC 20p Tube_inverter_+24V
6 6 6 E
(-18 ~ -30) Tube_inverter_G24
2409813 Molex 6p
7 7 L
CPLD J N
IDC 10p 2409824
2 5119273-4
TM11 J4 MAX3256A 1
C 21LS
5119274
1
G24
Left_limit_sw
1
J2 2 J1 5119273-3 C
(Left) 2 2 3 4
G24 JTA 0Vac
22LS 3 3 3
Right_limit_sw G MNL 3p 110Vac 5166053
(Right) 4 4 2
G24 IDC 10p 220Vac
11LS 5 5 1
Up_limit_sw
(Up) 6 6
5119276-1 7
G24
7
J3 MNL 4p 2409822 A2
12LS
(Down) 8
Down_limit_sw
8
J7 A2 RY1 A2 Fuse
5119276-2 G24 +24Vdc TM1
13LS 9 9 1
(U/D) UD_limit_sw Power_Relay- F1 10A 220Vac
10 10 2 7 1 1 Table Power Cable
4
5119273-2
D 5119276-3 2409815 IDC 10p Molex 3p
J1
2409818
5
8
F2 6A 220Vac
2 2 to PDU -2409502 D
Main Frame Assy 351C J351 +24Vdc CN52 Switch Power 80W F3 2A
1 1 9 3 3
A3 1N 1
+24Vdc 1 2
24G
3
5120419 6
110Va
Lamp_side_sw 2 A1 P1 F4 2A
2 2 A B c 4 4
3 3
G24 3
JST 4p
CN
Rise_side_sw 5G 5119273-1
4 4 4 3 3 1 5 5
G24 5 +5Vdc L
5 5 4 1 1
Fall_side_sw 6
GND N
6 6 5 6 7 8 2 3 6 6
FallInv_side_sw MNL 4p 2409816 PE
7 7 7 4 5
G24 JST 8p
5119990-
8 8 8
E 9 9
Up_side_sw
Down_side_sw
9
CN51
JST 5p 2409825
Tower A6
2 E
10 10 10 5119987
G24 Main A3
11 11 11 5119988
+5Vdc
Left_side_sw PWA A1
5G
12 12 12 5119989
+15vdc
Right_side_sw Base A4
-15vdc
13 1 Bulkhead 5119990-1
15G
15G
13 13 Table Ground Cable
Console_SM_SW
14 14 14 15W, 5Ohm 2
Table_EM_SW A2 to PDU -2408503
15 15 15
+24Vdc A1 R1 Molex 4p
16 16 16
17 17 17
18 18
+24V
18 J1 MNL 4p E
19 19
Emergency_CPLD
19
J332 J442
A2
Console Spot Control Input Cable TDSP 332 MIS Cable
F 20 20 20
DB50
F
IDC 20p 1 2 3 4 2409829 to Console -2409056
5119278 IDC 20p J11 J11 IDC 50p
TOP Board Interface Cable1
5119986 TM12 220Vac Area Designations
L/R 220 1 2409811
N1
Top_motor1 2 IDC 50p IDC 50p Xray Control Board
A3 N2
Top_motor2 3 J6 J82 2408350 A1 - Table Base PWAs assembly
M3 -
2409821 J41 J342 A1 B3 A2 - Table Power in Assmbly
-
Gear Motor J341 J331 J431 A3 - Table Main Frame Assembly
- Console Move Control Input Cable TDSP 331 MIS Cable
220Vac, 25W - J352
DB50 A4 - Table Base Motor Assembly
G 5120418-1 -
IDC 50p
2409828 to Console -2409055 A5 - Table Front Assembly G
- NOTES A6 - Table Tower Assembly
-
Table 1. All sheets of this schematic will be at the same revision. The A7 - Table Tube and Collimator
5155234 Top revision of this sheet will be the revision of the entire document. A8 - Table Case and Image Chain
5119277
2. = Indicates signals are part of a cable
11P 12P 21P 22P 13P
SIDE SW 3PS 2PS 1PS
10
12
13
11
14
15
19
18
7
1
2
S S S S S REVISIONS
A5 REV MADE BY DATE ECR NO. DESCRIPTION
11 OCT 2004
H Front cover To Main A3
5121493
1
2
Xu Yong
Xu Yong 15 DEC 2004 2010239
Initial Design
Update Some Cables Part No
MADE
APPROVALS YY/MM/DD TITLE
Precision THUNIS-800+ Table Schematics
REV H
3 Xu Yong 06 MAR 2006 Update For -Rainstorm
CHECK
FIRST MADE FOR: Storm 5
Inv_Fall Fall Rise Up Down Left Right Emergency 4 Jia Yingjie 03 JUL 2006
2030734 Update For Rainstorm APPR
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
Bulkhead
A2
Motor Relay Board Image Chain J432
2409218 5122283 I.I Power Cable I.I Power and Control MIS Cable
A J3
DB15
A
A1 B2 A8 2409831 to Console -2409498
Compreesor 110Vac
Compress_motor1 1 CCD Camera to PCU MIS Cable
A6 342ZN
30
Compress_motor2 2 To Main A3 to Console -5122285
0
30
A6
R
1
3 J1,J2,J4 5121494
30 AEC Feedback Coaxial Cable
110Vac,
M4 151C 1
30
MNL 3p J5,J12
Torque Motor 2
30 to CPI -2409797
5123381-1 151R 3
30
342ZN
4
30
R Xray HV Cable
152 4
30 44CN Logic IO Board Xray Tube
B LS
R 7
30 3 3
2408348-1 5121886
to CPI -5122632 B
8
30 2 2
(UP)
J41 Tube Anode Cable
9 1 1 A1 B1 A7
5121655 Compress_limit1_sw to CPI -2409507
4 4 8
Compress_limit_sw
5 5 9 Tube Ground Cable
Tower SW G24
MCN 6 6
G24
10
2 5 to CPI -2409508
A6 11P
1 1 9 9
Up_side_sw
2 2 10 10 4 Tube Fan Power Cable
Up S Down_side_sw
3 3 11 11 3
4 4 4 4
Lamp_Tower_sw
1
FAN to PDU -2409505-6
3 5 5 5 5
+24Vdc
2
J352 A7
SFD interface Cable
C Dow 12P 6 6 12 12
Spare_Tower_sw
7 C
n S MNL 6p 5119984 MNL 12p 2409832
1 IDC 10p IDC 34p
5121657
4
J341 J531
5 In-Room Console Move Control Input Cable DB50 In-Room Foot SW
5121497 5121496-1
Cassette CN83 J6 IDC 50p
5119985
1 Mic+ MIC
MIC 3 Mic-
1 1 REVISIONS
2 2
A6 2
3 3
REV MADE BY DATE ECR NO. DESCRIPTION
11 OCT 2004 Initial Design
H 4 4
5119983 Molex 4p
1
2
Xu Yong
Xu Yong 15 DEC 2004 2010239 Update Some Cables Part No
MADE
APPROVALS YY/MM/DD TITLE
Precision THUNIS-800+ Table Schematics
REV H
3 Xu Yong 06 MAR 2006 Update For -Rainstorm
CHECK
FIRST MADE FOR: Storm 5
APPR
4 Jia Yingjie 03 JUL 2006
2030734 Update For Rainstorm
Jia Yingjie OTHER
GEHL MEDICAL SYSTEMS 5117026SCH SHT/CONT ON
5 13 DEC 2006 Update For Rainstorm 2 /2
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7 8
Tube Fan Power Cable -2409505-6 3'8
D D
/6/6/6
%XONKHDG
/6/6/6
E $5 E
7DEOH7RS
)UDPH
70 70
/6
/6 F1 Table Power Cable -2409502
/6 F2
/6
/6 F3
F &1 F4 F
Ground Cable
2409503
)HUULWH
7RS/5 70
)HUULWH
A3 )HUULWH
M3
A4
M1
G G
7LOW5),
)HUULWH
)HUULWH )HUULWH
A3
7XEH8'
M2
H REVISIONS APPROVALS YY/MM/DD TITLE
REV H
REV MADE BY DATE ECR NO. DESCRIPTION
MADE
PS800+ Table Cable Routing
3636363636363636
1 Xu Yong 11 JAN 2006 Initial Design
CHECK
Precision 800+ Table
FIRST MADE FOR: 1
APPR
GEHL MEDICAL 5122468-1ADW SHT/CONT ON
- OTHER SYSTEMS 1/1
1 2 3 4 5 6 7 8
Section 2.0
Generator
Please refer to the Generator Schematics in CPI Service Manual.
Section 3.0
PDU
3.1 Digital System PDU Schematics
p g
Figure 10-1 5121161SCH_s1_r1
Section 4.0
Integrated Console
4.1 Digital System Console Schematics
Section 1.0
Scope
This section provides detailed information about each MIS run in the form of individual MIS cable
charts.
Each of the following charts defines a single interconnection in the system. The charts are arranged
in the order, according to MIS numbers. In some cases, a particular MIS number is repeated, except
for a letter added at the end. Such a set of charts use the same cable, and carry essentially the
same signals, however the applications for each chart is different - the point of connection is
different (at one or both ends) for each chart in the set.
Section 2.0
How to Use
Each MIS cable shown on the MIS map in 2409796-2(Digital System), has its own MIS chart. Each
chart identifies all signals present on the cable, and the specific pins associated with each signal.
See Figure 11-1 for an example.
Section 3.0
MIS Charts
3.1 System Main Input Power Cable
Note: The four cables should be provided by hospital, three are power supply cables, and the other one
is ground cable.
TB1
24G
1 CN51-4 (Brown)
+24V
6 CN51-2 (Red)
1 5G CN52-4 (Blue)
+5V
4 CN52-2 (Purple)
12G
2 CN51-6 (Black)
+12V
5 CN51-5 (Orange)
-12V
3 CN51-8 (Yellow)
VCC
1 1
VCC
2 2
VCC
3 3
VCC
4 4
VCC
5 5
VCC
6 6
GND
7 7
GND
8 8
GND
9 9
GND
10 10
GND
11 11
GND
12 12
F_BIT
13 13
R_BIT
14 14
F_ON
15 15
R_ON
16 16
EMERGENCY (No use)
17 17
IDNO+ (No use)
18 18
/Y5
19 19
/Y6
20 20
12V
21 21
12V
22 22
TECH_SPOT
23 23
XRAY_ON
25 25
ABNORMAL (No use)
26 26
GND
39 39
GND
40 40
GND
41 41
GND
42 42
GND
43 43
GND
44 44
12V
45 45
12V
46 46
12V
47 47
12V
48 48
12V
49 49
12V
50 50
REMOTE PREP+
1 TB6-7
REMOTE PREP-
14 TB6-8
REMOTE FLUORO+
2 TB6-5
REMOTE FLUORO-
15 TB6-6
REMOTE EXP+
3 TB6-9
REMOTE EXP-
16 TB6-10
MULTISPOT+
4 TB5-11
MULTISPOT-
17 TB5-12
XRAY ON+
6 TB6-2
XRAY ON-
19 TB6-1
RS_TXD
TX 2 2 TX
RX 3 RS_RXD 3 RX
GND
GND 5 5 GND
GND GND
FG
1 5
L
2 1
N
3 3
L
2 2 (White)
3 N 3 (Black)
1 FG 4 (Green)
GND GND
GND GND
24G
1 3 (Black)
24V
2 2 (Red)
FG
1 1
L
2 2
N
3 3
GND
1 1
CHANNLE_B
2 2
VCC
3 3
CHANNLE_A
4 4
GND
5 5
GND
1 1
CHANNLE_B
2 2
VCC
3 3
CHANNLE_A
4 4
GND
5 5
GND
1 1
+12V
2 2
5V
3 3
-12V
4 4
GND
5 5
MIC-
2 2
MIC+
1 1
SP+
1 1
SP-
2 2
From Signal To
1 Not used 1
2 Not used 2
3 24V 3
4 RISE_CONSOLE_SW 4
5 SPARE_CONSOLE_SW 5
6 FULL_CONSOLE_SW 6
7 FULLINV_CONSOLE_SW 7
8 24V 8
9 DOWN_CONSOLE_SW 9
10 UP_CONSOLE_SW 10
11 24V 11
Table 11-1 Console Move Control Signals
From Signal To
1 Not used 1
2 Not used 2
4 REC_DA<0> 4
3 REC_DA<1> 3
6 REC_DA<2> 6
5 REC_DA<3> 5
8 REC_DA<4> 8
7 REC_DA<5> 7
10 REC_DA<6> 10
9 REC_DA<7> 9
11 TRA_DA<0> 11
12 TRA_DA<1> 12
13 TRA_DA<2> 13
14 TRA_DA<3> 14
15 TRA_DA<4> 15
16 TRA_DA<5> 16
17 TRA_DA<6> 17
18 TRA_DA<7> 18
19 TRA_DA<8> 19
20 TRA_DA<9> 20
21 TRA_DA<10> 21
22 TRA_DA<11> 22
23 TRA_DA<12> 23
24 TRA_DA<13> 24
25 TRA_DA<14> 25
26 TRA_DA<15> 26
27 TRA_DA<16> 27
Table 11-2 Console Spot Control Signals
Coaxial
SYNC_SIGNAL IN SYNC_SIGNAL OUT
RAD_STATUS
1 1
FLUORO_STATUS
20 2
SINGLE_SPOT
2 3
GATE_TRIGGER
21 4
GATE_SEL
3 5
DSA_SEL
22 6
PATH_SEL (not use)
4 7
REPLAY_SEL (not use)
23 8
SINGLE_FRM_FORWARD (not use)
5 14
SINGLE_FRM_BACKWARD (not use)
24 15
WW_INCRES
6 16
WW_DECRES
25 17
WL_DECRES
7 18
WL_INCRES
26 19
(not use)
8 20
(not use)
27 21
FLUORO ON +
1 1
FLUORO ON -
14 2
RAD PREP +
2 3
RAD PREP -
15 4
RAD READY+
3 5
RAD READY-
16 6
MODE SELECT+
4 7
MODE SELECT-
17 8
II MODE 1+
5 9
II MODE 1-
18 10
II MODE 2+
6 11
II MODE 2-
19 12
RAD EXP +
7 13
RAD EXP -
20 14
II MODE 3+
8 15
II MODE 3-
21 16
PULSE FLUORO ON+
10 17
PULSE FLUORO ON-
23 18
XRAY ON +
11 19
XRAY ON -
24 20
PULSE FLUORO SYNC IN+
12 21
PULSE FLUORO SYNC IN-
25 22
Workstation
20'' Monitor VGA Digital Box
RED
1 1
GREEN
2 2
BLUE
3 3
RED SHIELD
6 6
GREEN SHIELD
7 7
BLUE SHIELD
8 8
SYNC GND
10 10
H_SYNC
13 13
V_SYNC
14 14
PART No.
1 of 2
2409365
VID0+
1 23
VID0-
26 4
VID1+
2 3
VID1-
27 22
VID2+
3 21
VID2-
28 2
VID3+
4 1
VID3-
29 20
VID4+
5 28
VID4-
30 9
VID5+
6 8
VID5-
31 27
VID6+
7 26
VID6-
32 7
VID7+
8 5
VID7-
33 25
VID8+
9 33
VID8-
34 14
PART No.
2 of 2
2409365
GND
1 1
GND
14 14
X0-
2 2
X0+
15 15
X1-
3 3
X1+
16 16
X2-
4 4
X2+
17 17
XCLK-
5 5
XCLK+
18 18
X3-
6 6
X3+
19 19
SERTC+
7 7
SERTC-
20 20
SERTFG-
8 8
SERTFG+
21 21
CC1-
9 9
CC1+
22 22
TB2 U A HV Tank
TB2 V B HV Tank
TB2 W C HV Tank
HV Tank
GND GND
1 1 SIG_IN
2 2 SIG_OUT
3 3 GND
1 1 SIG_IN
2 2 SIG_OUT
3 3 GND
GND
1 1
GND
4 4
GND
5 5
II_CTRL<1>
6 6
GND
7 7
GND
8 8
II_CTRL<2>
13 13
II_CTRL<0>
14 14
II_POWER(24V)
15 15
J4 Power Strip
0VAC 0VAC
J3 TM1
GND GND
J5 Fan
PE PE
C 1
M 2
S 3
T1 5
T2 6
Shield
GND GND
Ground
PE PE
Room Interface
J1 Boards
SPEAKER +
1 1
SPEAKER -
2 2
MIC +
4 4
MIC -
5 5
MIC - Shield
6 6
From Signal To
1 DI_CPLD<0> 1
2 DI_CPLD<1> 2
3 DI_CPLD<2> 3
4 DI_CPLD<3> 4
5 DI_CPLD<4> 5
6 DI_CPLD<5> 6
7 DI_CPLD<6> 7
8 DI_CPLD<7> 8
9 DO_CPLD<0> 9
10 DO_CPLD<1> 10
11 DO_CPLD<2> 11
12 DO_CPLD<3> 12
13 DO_CPLD<4> 13
14 DO_CPLD<5> 14
15 DO_CPLD<6> 15
16 DO_CPLD<7> 16
17 DO_CPLD<8> 17
18 DO_CPLD<9> 18
19 DO_CPLD<10> 19
20 DO_CPLD<11> 20
21 DO_CPLD<12> 21
22 DO_CPLD<13> 22
23 DO_CPLD<14> 23
24 DO_CPLD<15> 24
25 COL_CPLD<0> 25
26 COL_CPLD<1> 26
27 COL_CPLD<2> 27
28 COL_CPLD<3> 28
Table 11-3 Xray Control Signals
+24V
1 1 (Brown)
+24V
2 2 (Yellow)
+24V
3 3 (Blue)
INCLINE-OVER
4 4 (Gray)
VERTICAL-OVER
5 5 (White)
REVERSE-DEGREE
6 6 (Red)
+24V
1 1 (Brown)
+24V
2 2 (Yellow)
+24V
3 3 (Blue)
4 VERTICAL-STOP 4 (Gray)
5 LEVEL-STOP 5 (White)
6 INCLINE-STOP 6 (Red)
IO Board J4 Limit SW
24V
1 1 (Black)
LEFT LIMIT SW
2 2 (Brown)
24V
3 3 (Red)
RIGHT LIMIT SW
4 4 (Orange)
24V
5 5 (Yellow)
UP LIMIT SW
6 6 (Green)
24V
7 7 (Blue)
DOWN LIMIT SW
8 8 (Purple)
24V
9 9 (Gray)
RANGE SW
10 10 (White)
+24V
IO Board J1-1 CN52-1
24G
IO Board J1-2 CN52-3
5G
IO Board J1-3 CN51-3
+5V
IO Board J1-4 CN51-1
+15V
Xray Board J1-1 CN51-5
15G
Xray Board J1-2 CN51-6
15G
Xray Board J1-3 CN51-7
-15V
Xray Board J1-4 CN51-8
+5V
SFD-1 CN51-2 (Red)
5G
SFD-2 CN51-4 (Black)
+5V
R1-1 CN51-2 (Red)
5G
R1-2 CN51-4 (Black)
+24Vdc
1 1
Cassette in SW
2 2
+24V
1 1 (Red)
POWER-RELAY
2 2 (Black)
+24Vdc
1 1
Cassette in SW
2 2
Col RES-
7 1
Col RES+
10 2
Col FED UD
1 3
Col FED LR
4 4
Col LR
5 5
Col UD
8 6
Mot LR+
9 7
Mot LR-
12 8
Mot UD+
3 9
Mot UD-
6 10
GND
11 11
2 12
220VAC
1 1 (Red)
TOP_MOT1
2 2 (Black)
TOP_MOT2
3 3 (White)
220VAC
1 A2 RY1 Pin6
110VAC
2 A2 Fuse F4
0VAC
3 TM1 Pin4
PE
4 Braid
Frequency Inverter
Relay Board J4 IO FC1
GND
1 1
TILT INVERTER D0
2 2
TILT INVERTER DI0
3 3
TILT INVERTER DI1
4 4
TILT INVERTER DI2
5 5
TILT INVERTER +24V
6 6
TILT INVERTER 24G
7 7
NC1
8 8
NC2
9 9
PE
10 Braid
Frequency Inverter
Relay Board J5 IO FC2
GND
1 1
TUBE INVERTER D0
2 2
TUBE INVERTER DI0
3 3
TUBE INVERTER DI1
4 4
TUBE INVERTER DI2
5 5
TUBE INVERTER +24V
6 6
TUBE INVERTER 24G
7 7
NC1
8 8
NC2
9 9
PE
10 Braid
Positioner Table
Positioner Table
Frequency Inverter
FC1 UVW Motor1
U
1 1
V
2 2
W
3 3
PE
4 4
Frequency Inverter
FC2 UVW Motor2
U
1 1
V
2 2
W
3 3
PE
4 4
110Vac
1 1
0Vac
2 2
PE
3 3
GND
1 1
GND
4 4
GND
5 5
II_CTRL<1>
6 6
GND
7 7
GND
8 8
II_CTRL<2>
13 13
II_CTRL<0>
14 14
II_POWER(24V)
15 15
COM_MTR1
2 2
110VAC
3 1
PART No.
2 of 2
2409832
TOWER_SW_LAMP
4 1
+24V
5 2
COM_LS1
6 8
COM_LS
7 9
+24V
8 10
+24V
9 5
UP_SIDE_SW
10 4
DOWN_SIDE_SW
11 3
SPARE_TOWER_SW
12 7
110Vac
1 1
0Vac
2 2
PE
3 3
COL_RES_
1 3
COL_RES+
2 12
COL_FED_VD
3 9
COL_FED_LR
4 4
COL_LR
5 5
COL_VD
6 10
MOT_LR+
7 1
MOT_LR_
8 6
MOT_VD+
9 7
MOT_VD_
10 2
GND
11 11
12 8
GND GND
Table Power in
Fuse Assy RY1
220VAC
1 1
220VAC
2 2
220VAC
3 2
220VAC
A2 RY1-4 1
220VAC
A2 RY1-5 2
0VAC
1 1
0VAC
2 2
1 COM COM 1
2 N.O. N.O. 2
Table Main Frm TM11 21LS / 22LS Table Main Frm Assy
Assy
+24VAC
1 21LS-1
LEFT-LIMIT-SW
2 21LS-2
+24VAC
3 22LS-3
RIGHT-LIMIT-SW
4 22LS-4
COM
1 1
NC
2 2
COM
1 1
NC
2 2
COM
1 1
NC
2 2
Mic+
1 1
Mic-
2 2
3 3
COM_MTR2
1 301
COM_MTR1
2 302
110VAC
3 300
TOWER_SW_LAMP
4 4
+24V
5 5
COM_LS1
6 307
COM_LS
7 308
+24V
8 309
+24V
9 1
UP_SIDE_SW
10 2
DOWN_SIDE_SW
11 3
SPARE_TOWER_SW
12 6
spk+
1 1
spk-
2 2
Tower A6
E E
Positioner Table
E E
Positioner Table
E E
Positioner Table
E E
Positioner Table
Table Power in
Assy TM1 GND
E E
TableCase and
Table Main Frm Assy GND Image Chain
E E
E E
E E
J531 Bulkhead
+24V
SW_COM 18
FOOT_IRC_SW
SW_NO 17
PE
SW_PE SHIELD
+24V
18 18
FOOT_IRC_SW
17 17
+24V
1 1
COM_LS
2 2
12Vac
L
0Vac
N
+24V
1 1,6
UP_SIDE_SW
2 2
DOWN_SIDE_SW
3 3
TOWER_SW_LAMP
4 4,7
+24V
5 5,8
Camera Head
see table 4 see table 4
GND GND
3.99 5122518, Digital Box KeyBoard Bracket Right Side Grounding Cable
GND GND
HV Tank Tube
IF board J4 GEN
GND GND
GND GND
PCU Workstation
+5
1 1
RXD
2 2
TXD
3 3
NC
4 4
GND
5 5
NC
6 6
NC
7 7
NC
8 8
NC
9 9
Workstation
LCD Monitor VGA2 Digital Box
RED
1 1
GREEN
2 2
BLUE
3 3
RED SHIELD
6 6
GREEN SHIELD
7 7
BLUE SHIELD
8 8
SYNC GND
10 10
H_SYNC
13 13
V_SYNC
14 14
220VAC
L, N, FG Power In
+5
1 1
RXD
3 2
TXD
2 3
NC
4 4
GND
5 5
NC
6 6
NC
7 7
NC
8 8
NC
9 9
SW - COM
1 L3
SW - NO
2 L4