Anti-TPO
Antibodies to thyroid peroxidase
Elecsys 2010
MODULAR ANALYTICS E170
06368590 190 100 cobas e 411
cobas e 601
cobas e 602
Anti-TPO
Antibodies to thyroid peroxidase
Due to possible evaporation effects, samples, calibrators and controls on ▪ as required: e.g. quality control findings outside the defined limits
the analyzers should be analyzed/measured within 2 hours. Quality control
Materials provided For quality control, use PreciControl ThyroAB.
See “Reagents – working solutions” section for reagents. In addition, other suitable control material can be used.
Materials required (but not provided) Controls for the various concentration ranges should be run individually at
▪ 06472931190, Anti‑TPO CalSet, for 4 x 1.5 mL least once every 24 hours when the test is in use, once per reagent kit, and
following each calibration.
▪ 05042666191, PreciControl ThyroAB, for 2 x 2 mL each of The control intervals and limits should be adapted to each laboratory’s
PreciControl ThyroAB 1 and 2 individual requirements. Values obtained should fall within the defined
▪ 11732277122, Diluent Universal, 2 x 16 mL sample diluent limits. Each laboratory should establish corrective measures to be taken if
▪ 03183971122, Diluent Universal, 2 x 36 mL sample diluent values fall outside the defined limits.
Follow the applicable government regulations and local guidelines for
▪ General laboratory equipment quality control.
▪ Elecsys 2010, MODULAR ANALYTICS E170 or cobas e analyzer
Calculation
Accessories for Elecsys 2010 and cobas e 411 analyzers:
The analyzer automatically calculates the analyte concentration of each
▪ 11662988122, ProCell, 6 x 380 mL system buffer sample (either in IU/mL or kIU/L).
▪ 11662970122, CleanCell, 6 x 380 mL measuring cell cleaning Limitations - interference
solution The assay is unaffected by icterus (bilirubin < 1129 µmol/L or < 66 mg/dL),
▪ 11930346122, Elecsys SysWash, 1 x 500 mL washwater additive hemolysis (Hb < 0.93 mmol/L or < 1.5 g/dL), lipemia (triglycerides
< 23.9 mmol/L or < 2100 mg/dL) and biotin (< 40.9 nmol/L or < 10 ng/mL).
▪ 11933159001, Adapter for SysClean
Criterion: Recovery within ± 10 % of initial value.
▪ 11706802001, Elecsys 2010 AssayCup, 60 x 60 reaction vessels
Samples should not be taken from patients receiving therapy with high
▪ 11706799001, Elecsys 2010 AssayTip, 30 x 120 pipette tips biotin doses (i.e. > 5 mg/day) until at least 8 hours following the last biotin
Accessories for MODULAR ANALYTICS E170, cobas e 601 and administration.
cobas e 602 analyzers: No interference was observed from rheumatoid factors up to a
▪ 04880340190, ProCell M, 2 x 2 L system buffer concentration of 1500 IU/mL.
▪ 04880293190, CleanCell M, 2 x 2 L measuring cell cleaning In vitro tests were performed on 23 commonly used pharmaceuticals. No
solution interference with the assay was found.
▪ 03023141001, PC/CC‑Cups, 12 cups to prewarm ProCell M and In rare cases, interference due to extremely high titers of antibodies to
CleanCell M before use analyte‑specific antibodies, streptavidin or ruthenium can occur. These
effects are minimized by suitable test design.
▪ 03005712190, ProbeWash M, 12 x 70 mL cleaning solution for run For diagnostic purposes, the results should always be assessed in
finalization and rinsing during reagent change conjunction with the patient’s medical history, clinical examination and other
▪ 12102137001, AssayTip/AssayCup Combimagazine M, findings.
48 magazines x 84 reaction vessels or pipette tips, waste bags
Limits and ranges
▪ 03023150001, WasteLiner, waste bags Measuring range
▪ 03027651001, SysClean Adapter M 5.00‑600 IU/mL (defined by the lower detection limit and the maximum of
Accessories for all analyzers: the master curve). Values below the lower detection limit are reported as
< 5.00 IU/mL. Values above the measuring range are reported as
▪ 11298500316, Elecsys SysClean, 5 x 100 mL system cleaning > 600 IU/mL.
solution
Lower limits of measurement
▪ 11298500160, Elecsys SysClean, 5 x 100 mL system cleaning
solution (for USA) Lower detection limit of the test
Assay Lower detection limit: < 5.00 IU/mL
For optimum performance of the assay follow the directions given in this The lower detection limit represents the lowest analyte level that can be
document for the analyzer concerned. Refer to the appropriate operator’s distinguished from zero. It is calculated as the value lying two standard
manual for analyzer‑specific assay instructions. deviations above that of the lowest standard (master calibrator, standard
1 + 2 SD, repeatability study, n = 21).
Resuspension of the microparticles takes place automatically prior to use.
Read in the test‑specific parameters via the reagent barcode. If in Dilution
exceptional cases the barcode cannot be read, enter the 15‑digit sequence Samples with anti‑TPO concentrations above the measuring range can be
of numbers. diluted manually with Diluent Universal. The recommended dilution is 1:5.
Bring the cooled reagents to approximately 20 °C and place on the reagent The concentration of the diluted sample must be > 200 IU/mL. After dilution,
disk (20 °C) of the analyzer. Avoid foam formation. The system multiply the result by the dilution factor.
automatically regulates the temperature of the reagents and the Please note: The autoantibodies are heterogeneous and this gives rise to
opening/closing of the bottles. non-linear dilution phenomena for individual samples.
Calibration Expected values
Traceability: This method has been standardized against the NIBSC Extended studies with the Elecsys Anti‑TPO assay performed on samples
(National Institute for Biological Standards and Control) 66/387 Standard. from 208 healthy test subjects in 3 clinical centers in Austria and Germany
Every Elecsys reagent set has a barcoded label containing specific showed a borderline value of 34 IU/mL for 95 % of the results.
information for calibration of the particular reagent lot. The predefined For detailed information about reference intervals in children, adolescents
master curve is adapted to the analyzer using the relevant CalSet. and pregnant women, refer to the brochure “Reference Intervals for
Calibration frequency: Perform calibration on all analyzers as follows: Children and Adults”, English: 04640292, German: 04625889.
▪ with every reagent kit This booklet also contains results of a detailed study about influencing
factors on thyroid parameters in a well characterized reference group of
Renewed calibration on all analyzers: adults. Different inclusion and exclusion criteria were applied (e.g.
▪ daily: when using the same reagent kit on the analyzers sonographic results (thyroid volume and density) as well as criteria