This workbook contains flow check forms for high volume particulate sampler for PM10 or TSP.
This workbook will be used by WDNR for audits of these types of samplers
There are three types of flow controls used in these samplers. A separate check form has been created for each type.
Instructions
Fill in the Orifice Qa Calibration worksheet with the calibration data points for the orifice being used. Alternatively, the Orifice
ID, Calibration Date, Slope and Intercpet may be entered manually in the appropriate fields of the check form.
Based on the type of instrument flow control being used, select the apprpriate check sheet.
Volumetrically controlled (Choked flow venturi) - Use the Volumetric worksheet
Mass Flow Controlled (MFC, or Rheostat) - use the MFC worksheet
Computer Controlled (Plus type samplers) - use the Computer worksheet
Protection
The worksheets are protected and only cetain cells have been unlocked. Some unlocked cells may contain formulas or
references which may be desirable to overwrite with manual entries.
If the sheets need to be edited beyond the unlocked fields, the protection may be turned off in the file menu. Protection is not
password locked.
Hi-Vol (TSP or PM10)
Particulate Sampler Audit/Verification
Volumetric Flow Controller
Calculations
Pf (in H2O) = stagnation pressure Pf (mm Hg) = Pf (in H2O) * 25.4 / 13.61
Audit: Check 1 point at the Pf of a clean filter with an orifice other than the one normally used to verfiy by site opearator
Hi-Vol (TSP or PM10)
Particulate Sampler Audit /Verification
Mass Flow Controller
Calibration Data Results Calibration Date: 4/29/2013 Orifice Used: 111 Orifice Cal Date: 1/15/2013
Point/Plate 1 Point/Plate 2 Point/Plate 3 Point/Plate 4 Point/Plate 5 Slope: 19.08848
Qa (m3/min): 1.841 1.611 1.430 1.317 1.158 Intercept: 1.31322
(Ic) Corrected Dickson 36.0 32.5 29.0 26.5 23.0 Corr Coef: 0.99646
Comments: Performed an audit on indiustrial sampler at MFC Site - chekcs looked good.
Calculations
Point Orifice Orifice Actual Orifice Standard Instrument Act. Instrument Std Actual Q Δ% Standard Q Δ%
Number inches H2O m3/min m3/min Q m3/min Q m3/min m3/min m3/min
1 2.7 1.074 1.062 1.133 1.114 5.47 4.93
2
3
4
5
Calculations
Audit: Check 1 point at the operational flow rate with an orifice other than the one normally used to verfiy by site opearator
HiVol (TSP or PM10)
Particulate Sampler Calibration
Volumetric Flow Controller
Site Calibration Orifice
Location: Serial:
Airs ID Cal Date:
Date: Qa Slope:
Sampler Serial #: Qa Int.:
Calibrated By:
Temp (deg F): Pf (Clean Filter):
Ta (deg C): Pa (mm Hg):
Ta (deg K): Reason for Cal:
PreMaintenance
1
2
3
4
5
Calculations
Calibrator Flow (Qa) = 1/Slope*(SQRT(H20*(Ta/Pa))Intercept)
Pf in mm Hg = in H2O * 25.4/13.61
Pressure Ratio (Po/Pa) = 1Pf/Pa
% Difference = (Look Up FlowCalibrator Flow)/Calibrator Flow*100
PostMaintenance
1
2
3
4
5
Initial Certification check 5 points at, above and below Pf of a clean filter
Field Flow Check: Check 1 ponit at the Pf of a clean filter
Orifice Calibration Form for terms of Qa Fill in the gray squares
Date 10/15/2012
Orifice ID 102
Ta (C) 21.9
Ta (K) 294.9
Pa (mm Hg) 737.2
Orif.
Press
Known Roots Drop
Volume Press. H
Vm Time Time Time P (inches
Point (M3) (minutes) (seconds) minutes - decimal (mm Hg) H2O)
1 3.00 1.00 23.09 1.3848 28.3 10.38
2 3.00 1.00 28.96 1.4827 24.9 9.13
3 3.00 1.00 36.34 1.6057 21.1 7.82
4 3.00 1.00 44.77 1.7462 18.1 6.64
5 3.00 1.00 55.54 1.9257 15.1 5.5
6 3.00 2.00 8.20 2.1367 12.4 4.49
7 3.00 2.00 25.56 2.4260 9.7 3.49
8 3.00 2.00 49.62 2.8270 7.2 2.56
9 3.00 4.00 53.08 4.8847 2.5 0.83
Va Qa y axis
2.8848 2.0832 2.0377
2.8987 1.9550 1.9111
2.9141 1.8149 1.7687
2.9263 1.6759 1.6298
2.9386 1.5260 1.4833
2.9495 1.3804 1.3402
2.9605 1.2203 1.1816
2.9707 1.0508 1.0120
Use Qa slope2.9898 0.6121 0.5762
and intercept
for
Volumetircally
controlled
samplers
Qa slope (m) 0.99264
Qa intercept (b) -0.03112
coefficient (r ) 0.9999952