You are on page 1of 6

See discussions, stats, and author profiles for this publication at: https://www.researchgate.

net/publication/215603742

Monitoring of Circuit Breakers using Ion Mobility Spectrometry to detect SF6-


Decomposition

Article · January 1999

CITATIONS READS

16 177

3 authors, including:

Jörg Ingo Baumbach Evelinda Trindade


Hochschule Reutlingen University of São Paulo
320 PUBLICATIONS   3,779 CITATIONS    17 PUBLICATIONS   58 CITATIONS   

SEE PROFILE SEE PROFILE

Some of the authors of this publication are also working on these related projects:

Computational Methods for Breath Metabolomics in Clinical Diagnostics View project

IUPAC JCAMP-DX View project

All content following this page was uploaded by Jörg Ingo Baumbach on 22 May 2014.

The user has requested enhancement of the downloaded file.


Monitoring of Circuit Breakers using Ion Mobility Spectrometry to
detect SF6-Decomposition

J.I. Baumbach1, P. Pilzecker2, E. Trindade3


1
Institut für Spektrochemie und Angewandte Spektroskopie, Bunsen-Kirchhoff-Str.11,
D-44139 Dortmund, Germany
2
G.A.S. Gesellschaft für Analytische Sensorsysteme mbH, Emil-Figge-Str. 76-80,
D-44227 Dortmund, Germany
3
Laboratório Central de Pesquisa e Desenvolvimento COPEL/UFPR, C.P.318,
CEP 80001-970, Curitiba, PR, Brazil

Dedicated to Prof. Dr. D. Klockow on the occasion of his 65th Birthday

Abstract artificially produced SF6 are discussed10-17. The


In high-voltage systems insulated with SF6 basic physical and chemical properties of SF6,
electrical discharges as partial discharges, its behaviour under various types of discharges
sparks or arcs cause SF6 decomposition and simulations of different types of equipment
leading to the formation of some toxic and used in electric power industry have been
corrosive by-products. There is an urgent need broadly investigated14-16. Due to its efficient
for more information on the origins and infrared absorption and chemical inertness,
quantities of contaminants expected to arise SF6 contributes to the stratospheric ozone
from the use of SF6 filled electrical power depletion. The technical and scientific
equipment. A low resolution ion mobility community and international associations such
spectrometer used as analysis instrument as CIGRÉ (International Conference on Large
delivers a shift of the position of the main peak High Voltage Electric Systems) have discussed
obtained. This shift is correlated to the the need for regulations relating to SF6
concentration of decomposition products handling, the use of new mixtures, as well as
formed in SF6. This paper presents the results recycling or re-use procedures. There is no
of investigations on the fill gas in circuit doubt, that more information is needed on the
breakers in gas insulated substations during origins and the amounts of the contaminants in
operation and of the reclaimed gas after a the SF6 used in electrical power equipment.
recycling procedure. Results from the Purity standards for fresh and reclaimed SF6,
investigation of 36 different circuit breakers in as well as methods to check the gas quality are
an operating substation are presented, which also required. Hence, the establishment of a
can lead to new methods to check the fill gas diagnostic tool, which allows the on-site and
quality. on-line analysis of the insulating gas contained
in the gas insulated switchgear (GIS) is of
Introduction considerable interest.
In the last 30 years the use of SF6 insulated Ion mobility spectrometry is a technique
switchgears in high voltage substations has designed for sensitive environmental
increased considerably. This is due to their monitoring using low cost, rugged and highly
enhanced lifetime, reduced required area and mobile instruments. Using this technique to
volume and longer maintenance cycles4-6. Very supervise the insulating gas in GIS during
high standards in the engineering and operation offers the opportunity of event-based
manufacturing result in the high reliability of maintenance2,18,20-22.
these substations. However, in long term
operation failures still arise which may lead to Ion Mobility Spectrometry
severe damage and additional repair costs1,3,9. Ion mobility spectrometry is a technique which
Also, problems with toxic and corrosive was designed for the detection of trace
decomposition products and the potential compounds within a gas, e.g. gaseous
greenhouse gas effect arising from the pollutants in air7,8,19,23-24. It combines high

Received for review November 10, 1999, Accepted December 9, 1999


Copyright © 1999 by International Society for Ion Mobility Spectrometry
J.I. Baumbach et al.: „Monitoring of Circuit Breakers using ...”, IJIMS 2(1999)1, 35-39, p. 36

Figure 1:
Point to connect the circuit breaker to the ion mobility spectrometer

sensitivity and relatively low technical Faraday plate to the ionization region is used
expenditure with high speed data acquisition. to protect the drift region from chemical
The main advantage over other common reactions between the analyte molecules and
detection methods is the fact, that the to allow only ions of the analyte formed to
instrument can work on-line, continuously and enter the drift region but no uncharged
unsupervised. All these advantages also apply molecules of the analyte. One of the
to the application introduced here. However, preconditions of IMS operation is that
there are some essential differences because (hopefully) no charge transfer reactions occur
of the application in the industry here to the in the drift region of the IMS.
traditional operation of the ion mobility In our case, however, an application in
spectrometer (IMS). This shall be made clear in industrial area is considered. Therefore, no
the following. drift gas is available in the substation
The instrument is based on the drift of ions at investigated. In contrast to the common
ambient pressure under the influence of an operation of the IMS a low resolution IMS not
electric field. The ions undergo a separation protected against analyte molecules entering
process based on various drift velocities due to the drift region was build. Here, the real start
different masses and charges on the way position of the ion swarm is unknown.
towards the Faraday plate. A so called drift gas Normally, a Bradbury-Nielsen shutter opens for
(clean air or nitrogen), which flows from the short time intervals between some µs and
Copyright © 1999 by International Society for Ion Mobility Spectrometry
J.I. Baumbach et al.: „Monitoring of Circuit Breakers using ...”, IJIMS 2(1999)1, 35-39, p. 37

comparable conditions
reference spectra of pure SF6
Peakshift Phase C were taken on-site and they
Peak Position / ms
1,0 Phase C 29.5 showed no significant
Phase B 29.1 difference to the spectra taken
Phase A 28.4
Reference 27.1 in the laboratory. It is notable
that the data acquisition time
Signal / V

(including data processing) was


0,5
only a few minutes under real
Mean Peak Position
(5 Measurements)
conditions which makes this
Phase A (28.4+0.2) ms
Phase B (29.1+0.3) ms method especially interesting to
Phase C (29.5+0.2) ms
the potential users. The flux
0,0
Peak Position
was adjusted to about 2 L/h at
ambient pressure. The total
10 20 30 40 50
measurement time was about 5
Drift Time / ms
min. The filling valve was
connected to the IMS (see
Figure 2: Figure 1) with a DILO® adapter
and a Teflon® line with an inner
Spectra obtained on the three phases of a circuit breaker and diameter of 1 mm. A needle
a reference spectrum of new SF6 valve allowed the adjustment of
a constant small flux through
about 1 ms and only during this time intervals the IMS. All 36 circuit breakers on the site were
ions can enter the drift region. Thus, in our investigated using this procedure. In Figure 2
case a flow of contaminated SF6 will enter the change of the peak positions calculated
ionization and drift region. Therefore, charge from the SF6 spectra of the 3 compartments of
transfer reactions also occur in the drift region. the phases A, B and C (former nomenclature
The mechanism of ion formation in SF6 is R, S and T) of a circuit breaker are clearly
currently under investigation using a high visible. This leads to the conclusion that
resolution IMS. despite the molecular sieves installed in the
Collecting the ions on a Faraday plate delivers compartments, significant amounts of
a time dependent signal corresponding to the by-products are present in the gas of this
mobility of the arriving ions. Normally,
such an ion mobility spectrum
contains information about the nature
of the different trace compounds
present in the sampled gas. In our
case a broad peak occurs. But the 2.5

shift of the peak position could be 2.0


correlated to the total number of
/ ms

decomposition products25. 1.5


Peak Shift

1.0

Results and Discussion 0.5


For investigations on SF6, a table-top
0.0
ion mobility spectrometer (IMS) was
developed and employed in a
er

A
ak

conventional substation equipped


re
it B

B
with SF6 filled circuit breakers under Ph
rcu

ase
Ci

normal operation. The IMS was C

connected to the gas compartment


with a thin polyethylene tube, which Figure 3:
allowed a small flux of SF6 to pass
through the instrument, adjustable Peak shift between the reference spectrum and the
with a needle valve. To ensure spectrum obtained from 36 different circuit breakers
Copyright © 1999 by International Society for Ion Mobility Spectrometry
J.I. Baumbach et al.: „Monitoring of Circuit Breakers using ...”, IJIMS 2(1999)1, 35-39, p. 38

ranges (see Figure 4). However,


since the distance etween the error
source and the gas sampling point is
16
not known, the local concentrations
14
> 500 > 750 could be much higher. The CIGRE
12
ppmv ppmv values for the concentration (total
number of decomposition products)
10
acceptable are between 500 ppmv
Number

8 and 2000 ppmv. As can be seen in


6
Figure 4 some circuit breakers were
over these limits and were this
4 method to be used for event
2 orientated maintenance then these
circuit breakers should be subject to
0
0.5 1.0 1.5 2.0 2.5 maintenance.
Peak Shift / ms The use of IMS, as presented here,
allows an automated monitoring
system to be set up. The structure of
Figure 4: such a system is shown in Figure 5.
Crucial compartments of the
Classification of the peak shift and interpretation in substation should be equipped with
terms of mean decomposition products expected an IMS connected to the gas inside
the compartments. The acquired
specific circuit breaker. The identity and origin spectra could be transferred via a local area
of the by-products could be investigated using network or other telecommunications system to
more selective analytical methods. The the control room where the IMS server is
performance of the monitoring system, placed. Here the spectra can be evaluated by
however, is demonstrated through detection of comparison to reference data, providing a
the peak shifts (see Figure 3). The shift in the basis for decisions on maintenance.
peak position is correlated to the mean
concentration of the decomposition products. Conclusions
Therefore, the number of circuit breakers in A sensitive method was developed for on-site
different classes of peak shift ranges monitoring of SF6 quality which proved to be
corresponds with classes of concentration suitable for the detection of small changes in
composition of the gas.
Experiments carried out
Gas insulated Available network
connection for
on-site at a gas insulated
substation system extension
Power plant substation under real
Gas sampling
and feedback
conditions revealed the
IMS Connection ability of the instrument
employing existing
Control room
network connections to withstand the
IMS
Monitored environmental conditions
compart- IMS Operating outside the laboratory.
ments IMS
company The encouraging results
IMS-server
of the form the basis for a
substation
IMS rugged continuously
IMS Data transfer to control room
operating monitoring
system for circuit
breakers in gas insulated
Figure 5: substations.

Concept of a continuously monitoring system to control on-line and Acknowledgments


on-site the gas quality in different compartments of a gas insulated The financial support of
substation the Bundesministerium
Copyright © 1999 by International Society for Ion Mobility Spectrometry
J.I. Baumbach et al.: „Monitoring of Circuit Breakers using ...”, IJIMS 2(1999)1, 35-39, p. 39

für Bildung, Wissenschaft, Forschung und [14] Vigreux, J.: Application of condition monitoring
techniques in gas insulated substations. - Electra
Technologie and the Ministerium für
(1986) 48-59
Wissenschaft und Forschung des Landes
Nordrhein-Westfalen are gratefully [15] Mauthe, G., Pettersson, K., Gleeson, D., König, D.,
Lewis, J., Molony, ; P.O'Connell; A.Porter;
acknowledged. L.Niemeyer: Handling of SF6 and its decomposition
products in gas insulated switchgear (GIS). - Electra
References 136 (1991) 69-89
[1] Babusci, G., Colombo, E., and Speciali, R., [16] Mauthe, G., et al., Handling of SF6 and its
“Assessment of the Behaviour of Gas-Insulated decomposition products in gas insulated switchgear
Electrical Components in the Presence of an Internal (GIS) -2nd Part- Electra 137 (1991) 81-108
Arc”, CIGRE 1998, 21, 1998.
[17] Mauthe, G., Pyror, B.M, Niemeyer, L., Probst, R.,
[2] Baumbach, J. I., Irmer, A. v., Klockow, D., Alberti Poblotzki, J., SF6 Recycling Guide - Re-use of SF6
Segundo, S. M., Sielemann, S., Soppart, O., and gas in electrical power equipment and final disposal
Trindade, E., “Characterisation of SF6 Decomposition CIGRE 23.10 Task Force01 23.10 (1997) 1-47
Products caused by Discharges in Switchgears using
Ion Mobility Spectrometry”, Proceedings of the fourth [18] Soppart, O., et al., “Ion mobility spectrometry for the
international workshop on ion mobility spectrometry, characterization of SF6-decomposition products
Cambridge, August 1995, 1995. caused by partial discharges”, 9th International
Symposium on High Voltage Engineering, Graz,
[3] Chan, T. M., Heil, F., and Kopejtkova, D., “Report on August 28 - September 1, 1995, 2269-2261 -
the Second International Survey on High Voltage 2269-2264, 1995.
Gas Insulated Substations (GIS) Service
Experience”, CIGRE 1998, 23-102, 1998. [19] Roehl, R. E., “Environmental and process
applications for ion mobility spectrometry”, Appl.
[4] Christophorou, L. G., Olthoff, J. K., and Green, D. S., Spectrosc. Rev., 26, 1-57, 1991.
“Gases for Electrical Insulation and Arc Interruption:
Possible Present and Future Alternatives to Pure [20] Segundo, S. M. A., Soppart, O., Janissek, P. R.,
SF6”, NIST Technical Note, 1425, 1-44, 1997. Baumbach, J. I., Neves, E. F. A., Klockow, D., and
Gutz, I. G. R., “Monitoring of Sulphur Hexafluoride
[5] Mauthe, G. et al., “SF6 and the global atmosphere”, (SF6) - Environmental and Technological Relevance”,
Electra,164, 121-131, 1996. Proceedings of the International Colloquium on
[6] Moritz, G., “25 Jahre Betriebserfahrungen mit Process related Analytical Cemistry in
SF6-isolierten Schaltanlagen”, Elektrizitätswirtschaft, Envirionmental Investgations, Gramado, Brazil, 63,
92, 1191-1194, 1993. 1996.

[7] Eiceman, G.E., Karpas, Z., “Ion mobility [21] Soppart, O., Baumbach, J. I., Alberti, S. M., and
spectrometry”, CRC Press, Boca Raton, Ann Arbor, Klockow, D., “Partial Discharge Ion Mobility
London, Tokyo,1-228, 1994. Spectrometry for Rapid Quality Assessment of SF6
used in High Voltage Substations”, Field Screening
[8] Hill, H. H., Siems, W. F., St.Louis, R. H., and Europe, 1997. In: J. Gottlieb, H. Hötzl, K. Huck and
McMinn, D. G., “Ion mobility spectrometry”, R. Niessner: Field Screening Europe, Proceedings of
Anal.Chem., 62, 1201A-1209A, 1990. the First International Conference on Strategies and
[9] Neumann, C., Balzer, G., and Hudasch, M., Techniques for the Investigation and Monitoring of
“Insulation Coordination Practice of German Utilities Contaminated Sites, Sept. 29 - Oct. 1, 1997,
- Procedures, Experience and Future Trends”, Karlsruhe, Germany, p. 355-358
CIGRE 1998, 33-102, 1998. [22] Soppart, O., Baumbach, J. I., Alberti, S. M., Moraes
[10] Ko, M.K.W., Sze, N.D., Wang, W.-C., Atmospheric e Silva, J. M., and Klockow, D., “Monitoring the
Sulphur Hexafluoride: Sources, Sinks and Quality of Sulphur Hexafluoride (SF6) in Gas
Geenhouse Warming. - Journal of Geophysical Insulated Switchgear sing Ion Mobility
Research 98 (1993) 10-499 - 10.507 Spectrometry”, Proceedings of the V SEMEL, August
1996, Curitiba, PR, Brazil, 2, 562-571, 1996.
[11] Maiss, M., Levin, I., Global Increase of SF6 Observed
in the Atmosphere. - Geophysical Research Letters [23] St.Louis, R. H., and Hill, H. H., “Ion mobility
21 (1994) 569-572 spectrometry in analytical chemistry”, Critical Review
in Analytical Chemistry, 21, 321-355, 1990.
[12] Van Burnt, R.J., Production rates for oxyfluorides
SOF2, SO2F2, and SOF4 in SF6 corona discharges. - [24] Widmer, H. M., and Morrissey, M. A.,
Journal of Research of the National Bureau of “Neochromatographic technologies - ion mobility
Standards 90 (1985) 229-253 spectrometry”, Chimia, 43, 268-277, 1989.

[13] Van Brunt, R.J., Olthoff, J.K., Sauers, I., Morrison, [25] Baumbach, J.I., Klockow, D., Kurrat, M., Segundo,
H.D., Robins, J.R., Chu, F.Y., Detection of S2F10 S.M.A., Soppart, O., Verfahren zur Überwachung des
produced by electrical discharges in SF6 - Proc. 10th Qualitätszustandes des Füllgases
Int. Conf. on GAS DISCHARGES AND THEIR Schwefelhexafluorid in gasgefüllten Anlagen.
APPLICATIONS, Swansea 13-18 September 1992, German Patent DE 195 28 290. - Priority 2.8.1995
p. 418-421

Copyright © 1999 by International Society for Ion Mobility Spectrometry

View publication stats

You might also like