Professional Documents
Culture Documents
17073
REVISION 1
VRT TABLE
SERVICE MANUAL
MEDICAL
SYSTEMS
GENERAL @ ELECTRIC
GENERAL@ E L E C T RI C
GENERAL ELECTRIC COMPANY MEDICAL SYSTEMS GROUP
MILWAUKEE WISCONSIN 53201 U S A TELEX 269679
COPENHAGEN FRANKFURT MADRID TORONTO LONDON
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
TABLE OF CONTENTS
J
CHAPTER TITLE
3 INSTALLATION/ADJUSTMENTS
4 FUNCTIONAL CHECKS
5 TROUBLESHOOTING INFORMATION
6 ADJUSTMENTS
8 NOT INTENDED
9 RENEWAL PARTS
10 PERIODIC MAINTENANCE
REVISION HISTORY
REV DATE REASON FOR CHANGE
TabCh 3 - TabCh 7 -
3-1 1 7- i 0
3-1I 1 7-1 thru 7-22 0
3-1 thru 3-2 1 7-23 thru 7-24 1
3-3 thru 3-9 0 7-24A 1
3- 10 thru 3-26 1 7-25 1
3-27 0 7-26 0
3-28 thru 2-29 1 7-27 thru 7-28 1
3-30 thru 3-33 0 7-29 thru 7-49 0
3-34 thru 3-35 1 5-62 thru 5-71 1
3-36 0 5-7 1A 1
5-718 1
5-72 thru 5-90 1
I
I
Il111 NO POSTAGE
NECESSARY
IF MAILED
IN THE
UNITED STATES
I
I'
BUSINESS REPLY MAIL
FIRST CLASS PERMIT NO 6657 MILWAUKEE WISCONSIN
I
I'
I
I
I
I'
I
~~
E l
E l
4-
XI
= I
I
I
I
I
I
I
IC
[r
Ilb:a
:1
It
FOLD FOLD
I
I
I
I
I
I
I
I
I
I
I
I
I
SUGGESTION SUGGESTION SUGGESTION SUGGESTION SUGGESTION SUGGESTION I
I
I
I
I
I
NOIlS399nS NOIlS399~S NOIlS399flS NOIlS399nS NOIlS399nS NOllS399nS I
I
IF ADDITIONAL PAGES ARE INCLUDED, FOLD AND STAPLE TO LOWER PART OF THIS
SHEET BEFORE MAILING
SERVICE REPRESENTATIVE
(Please Pnnt)
CITY
DIAL COMM NO
I
.
I
I
MAIL CODE
I
DIRECTION NO 13861K CODE September 6 1985 I
I
I
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
CHAPTER 3 - INSTALLATION
TABLE OF CONTENTS
I
L
CHAPTER 3 - INSTALLATION
SECTION 1 -INTRODUCTION
1-1 How To Use This Chapter Read the "Damage in Transportation ' form before
unpacking the equipment Unpacking except that
This chapter provides the sequence needed t o necessary t o inspect for possible shipping damage,
'
install the equipment listed in the COMPATIBILITY jhould not be done until equipment is moved into
LISTING in Direction 17119, VRT Table Planning - I
the x-ray room Shortage claims on small parts can
-- be avoided if packing material, envelopes, pack-
Check equipment according t o the' packing-list, I
agk, etc , are carefully examined Do not discard
(Product Delivery Instruction) Check each system any material until all items have been located
against the main catalog numbers In the System r9
-- - - - --
I
I
ct,
-I
Listing Chart
" _ -- - 1-4- Materials Required For Installation
Collect and verify that all manuals are present 1
i
Interconnection of products by furnished cables or Installation"kit catalog number B0578AA i s
contractor supplied wiring are indicated by num- required t o install the VRT table
bered Master Interconnect System (MIS) runs in ;
the installation sections that follow Interc%n6ectlt Table base anchor: are provided in the installation
charts are available which describe the connection , kit Reference Direction 171 19 for anchoring in-
cables, except power and qround, and are fur- structions -
- -
I
r i ,
I
' I
4
INSTALL AND INTERCONNECT
Generator SARC II
Phototimer/Quantamat
4
INSTALL AND INTERCONNECT
Tableflube Column Mamo Interlock
Sentry 111
+
BUCKY INSTALLATION
Changex Vertical Cassette Holder
Katum Head Stand Multi Size Cassette Holder
Vertical Bucky Stand Horizontal SID
1X Ray Tube
Ion Chamber
INSTALL
Counterweights
Grid
VRT ADJUSTMENTS
SID Pot Motor Speed
Tom0 Plane Height Time After Exposure
Tube Stand Sentry 111
0 Section 1, Unpacking, Setup, and Furnished 0 Section 2, Unpacking, Setup and Furnished
Parts parts
2 Mount tom9 relay assembly 46-21692261 on 2 Mount relay assembly 46-21692261 on the
the generator 56 panel' Refer t o Illustration generator 20 panel Refer t o Illustration 3-1
6 I >
3- 1
0 If the system does not have a SARC II Rotor
Controller, perform the following
i
(marked 44 and 49, in series with the VRT tom0 1 Refer t o MSI 850-11/1250 IV Service Manual
interlock relay Reference MIS 836,. (VRT Table A0701B/A0702A, Chapter 3 and perform the
J4 t o DXD 350-11 56TB1 panel) following sections
6 Tape the end of the normally open contact lead 0 Section 1, Before You Begin
marked 72 Tape the end of the normally open
contact lead marked B12 These leads are not 0 Section 2, Unpacking Setup, and Furnished
used for this application Parts
0 3-1 Line Safety Switch to Power Unit 0 3-1 Control and High Voltage Transform-
er
0 3-2 X-ray Control to Power Unit 50/60 Hz
0 3-2 High Voltage Transformer t o Power
0 3-3 X-ray Control to High Voltage Trans- Unit
former 50/60 Hz
0 Section 4, Interconnection Auxiliary Items,
0 3-4 Power Unit t o High Voltage Trans-
former 50/60 Hz 0 4-1 Handswitch
0 3-12 Installation of X-ray Tube Casing Inter- 0 4-3 Connection of Input Power
lock
0 4-5 System with only one RAD tube
0 4-5 Bucky connections
0 4-6 Connection of X Ray Tube Casing
2 Mount relay assembly 46-21692261 on the
generator 20 panel Refer t o Illustration 3-2 0 Section 5-10 032 Rotor Panel
0 Section 2, Unpacking, Setup, and Furnished Section,6, Quantamat:lI Power Supply (A90
Parts " I ' ", I
t
Panel) t r 3
1 1 I t" 8 * t)
i 1 1 I tf-
0 Section 5,5-8, 032 Rotor Panel Section 8, Connect A90 Panel t o Three
Phase Generator
2 Mount relay assembly 46-21692261 on the
generator 20 panel Refer t o Illustration 3 2 0 Section 9, Connect A90 Panel t o Single
Phase Generator
0 If the installation does not have a SARC II Rotor
Controller perform the following 0 Section 11, A90 Panel t o Ion Chamber
L
I . . Detectors
0 Connect relay coil lead marked A2 t o ' 0 SectionL12, Film Area Compensation With
D32TS1 6 (schematic location 160, A) and
the lead marked 46 t o 32TB1-G or 32TB1-F - - -Sentry-ll and Sentry-Ill- - - Automatic Col-
limator
(schematic location 162, C)
I
, I
0 Connect -relays normally open contacts - --SECTION 4 SARC-11 ROTOR CONTROLLER INSTAL- -
(marked 72 and 812) in series with VRT- - - - - - - - i- V T l 0N
tom0 interlock relay at 2OTBl-Blf-(sche- - -+-
1-
NOTE
matic location 53, G) , ' I L
9
0 Tape the end of the normally open contact: '- -Skip this section if a SARC 11 1s not t o be In-
lead marked 44 Tape the end of the nor- stalled 1 r
i
mally open contact lead marked 45 These
leads are not used for this application - _t - Refr$ t o SARC-II Service Manual SM A8105D
I - -_ -- Chaptei3, and perform the following sections
I
Skip this section if a Quantamat II Phototimer 2 Remove the lead from TSl-H6 and tape the
is not t o be installed end The other end of this wire will be removed
in the generator Connect a jumper between
Refer t o Quantamat-ll Phototimer Service Manual TSl-H6 and TSl-A1 Schematic location
SM A8400B Chapter 3, and perform the following RClASA1 (46-193550-5) 3, E
' I $ I i
-
SECTION 5 VRT TABLE INSTALLATION 2 Carefulby mark each mounting hole location
required t o mount both the Rail Stand and
5-1 Table Site Preparation Table Base See note
NOTE
NOTE
If Rail Stand and Table Base mounting holes
Reference Illustration 3-3, 3-4 and 3-5 on are not accurately located, table will not
pages 3-6 through 3-8 regarding information operate properly
on mounting hole layout, before beginning
the following procedure 3 Accurately drill each (112" diameter) mounting
hole required for both the Rail Stand and Table
Base
1 Remove the floor covering (if any) from around
mounting points of Rail Stand and Table Base 4 Remove all dust from the mounting holes
b
2526
opening for Raceway 232
I
L 3818
970
J
NO ACCESS
62] L5 MAX HEIGHT
63 CONSTRICTED BY CABLE EXIT
TO TUBE STAND
0IDEALENTRANCE
\ /
,F - - -
\
\
\ [-Y'
\ I
I
I
! I
I
I
I
D
D
st
I
I
@ I
5 Reference Illustration 3-6 Locate the supplied 6 Place your 3' or 4' level over mounting points
shims required to make up the difference and verify that tolerance in both directions
between the removed (if any) floor covering (longitudinal and lateral) Is less than 030" if
and the original floor covering using a 4' level, or 020'' if using a 3' level
TUBE COLUMN
PALLET
SPACER USED TO
MOUNT M x 75
I
COVER CAPS
37593 DO1
SHIM IDENTIFICATION
ILLUSTRATION3 6
5-2 Rail Stand Installation 3 Reference Illustration 3-8 Remove cover caps
from rail stand Loosen screws and remove both
1 Position rail stand over i t s mounting holes rail stand covers
I I COVER CAPS
d 37582 DO1
36999 D10
RAIL
SCREW TAND
LOCATION UNTING
RAIL LOCATION
..-
STAND
MIOUNTING
LOCATION
-
I
TOM0 BELT COVER
S C REW
LOCATION bAIL
STAND
SCREW
LOCATION
MOUNTING
LOCATION
37583 DO1
RAIL STAND PREPARATION
ILLUSTRATION 3 9
MOUNTING
PLATE
36998 D10
SHIPPING DOLLY REMOVAL
ILLUSTRATION3 12
r J 1
I G;3 TABLEBASESHIM
AND PLACEMENT
37585 DO1
'1 MOUNTING
PLATE
f
37587 DO1
LONGITUDINAL DIRECTION
tLLUSTRATlON 3 14
CAUTION
This section is very important, if you do not
take the time to level your table, as described 37584 DO1
in the following procedure, it will not
function properly1 HEIGHT ALIGNMENT
ILLUSTRATION 3 16
2 Reference Illustration 3-16 Use supplied shims 6 Reference Illustration 3-17 Using supplied
as needed (notice shim style and placement washers (behind head of anchor bolts), install,
under rail stand) to meet the required 4rnrn but do not tighten (4) table base mounting
height difference bo1t s
NOTE
The (4) shorter bolts are used on the top
portion of table and (2) longer bolts are used
on the bottom portion of table
n t
3 - LOCATIONS
37585 DO1
REQUIRED
WASHER TABLE BASE MOUNTING
ILLUSTRATION 3 17
-
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
NOTE
Use the top bearing rail on rear of rail stand as
a reference for the longitudinal leveling
NOTE
37583 DO1
0 The use of a feeler gauge i s needed to veify Reference Section 5-3 step 6 Verify that 3 mm
tolerances spacer is in place
Reference Illustration 3-19 Use supplied shims Check Sections 5-4 through 5-8 Adjust if
as needed (notice shim style and placement necessary
between bottom of rail stand and table base
floor mounting bolts) to meet tolerance as Reference Sections 5-2 and 5-3 SECURELY
stated in step 2 TIGHTEN all rail stand and table base mounting
bolts Torque t o 20-30 ft -Ibs
CAUTION
Never place any shims between top portion Check Sections 5-4 through 5-8 Adjust if
of rail stand and table necessary
SECTION 6 -TUBE COLUMN INSTALLATION 3 Reference Illustration 3-21 Remove covers from
the vertical carriage
6-1 Shipping Brace Preparation
4 Reference Illustration 3-21 Place support arm
1 Reference Illustration 3-21 Before installing 28" from the center of the trunnion
trunnion, decide whether the column will be
installed from the head or foot end of rail stand NOTE
and which side of column the trunnion's long
To release safety brake on the tube support
arm will be mounted
arm, do the following
NOTE
0 Reference Illustration 3-22
As seen in Illustration 3-21 the trunnion has a
long and short arm The long arm will pass Place your screw driver on spring loaded
over the rail stand when installing the column safety brake lever and press down t o release
HORIZONTAL ARM
HORIZONTAL \ \
-< REMOVE*%?
COVERS
PLUNGER
ILLUSTRATION3 22
6-2 Tube Column Preparation 7 To install the column base on rail stand, the
column base must be rotated toward the
1 Remove shipping braces from support cradles longer trunnion arm
and the two red shipping braces from column
base 8 Reference Illustration 3-24 Remove (2)
mounting screws from tom0 coupling plate,
NOTE and remove
When trunnion is bolted t o the tube column
in another area and moved to the room, the
door must be a t least 49 ' wide
CAUTION
Watch out for overhead obstructions when
raising the column t o its vertical position
SECTION 7- SENTRY Ill AUTO COLLIMATOR IN- There must be no interference between the
STALLATION upper component tray and the cable drape
Redo tie wrap point i f necessary
Detach and discard the connector cover from
Verify that bucky wheels are always
the electronics package See Illustration 3-26 supporting the bucky in i t s entire travel
range If not, check longitudinal alignment
Remove and discard the two hex standoffs See with a level
Illustration 3-26
If adjustment is necessary, remove both side
Detach and discard the shields from the trim covers from the upper table base
electronics package See Illustration 3-26
Adjust bucky guide bar height by loosening
the guide bar support screws to move the bar
Mount the electronics package t o the bottom up or down as necessary
tray of the table base with hardware from the
installation kit ((4) 1/4-20 x 1/2" Hex head Tighten hardware
screws and (4) 1/4 lockwashers) Connect P2-P9
connectors Remove (2) mounting screws from the upper
REMOVE HEX right side of the table frame
REMOVE SHIELD STANDOFFS REMOVE CONNECTOR
AND DISCARD AND DISCARD COVER AND DISCARD
Route the ribbon cable and ground wire of the
1
keyboard through the slot
-
SECTION 8 BUCKY INSTALLATION
NOTE
Connection 06 goes t o upper component tray ELECTRICAL
I/F 06, cable 2a goes t o lower component tray CABLES
0 6 AND Za
I/F P2a Reference Illustration 3-27
37771 DO1
4 Insert cassette tray, and move the bucky by BUCKY CARRIAGE
hand over the entire bucky travel range and ILLUSTRATION 3 27
check the cable drape of cable 06 and 2a
0 Install raceway t o Changex as directed in 0 Section 6-5-1, Trim and Cover Installation
Direction 14232
0 Section 6-5-2, Pointer Indication Check
0 Section 2, Preparation for Installation
0 Section 6-5-3, Bucky Table Adjustment
0 Section 3, Locating the Baseplate
[7 Appendix A Supplement for General Electric
0 Section 4, Installing the Changer Stand Interfaces
0 Section 5 Preparing the Changer for Mount- 8-4 Vertical Bucky Stand Installation And
ing interconnect
0 Section 14 Installing Optional Patient Com- Section 10, Trim Covers and Finalize Cable
pression Device Drape
8-3 Katum Head Stand Installation And Inter- 0 Section 11, Mount Front Cover
connect
0 Section 12, Lock Adjustment
Refer t o Katum Region X40 Head Unit Service
Manual and perform the following sections Section 13, Counterbalancing
0 Install raceway interface t o the counter- 1 The power t o the table must be off
balanced vertical cassette holder as directed In
Direction 14236 2 The table top must be off
0 Section 3-1, Capacitor Modification 3 Remove the ion chamber parts from shipping
box Reference Illustration 3-28
0 Section 3-2, Detector Mounting
NOTE
0 Connect MIS 222
The Channel Assembly and t h e Cable
Retermination K i t s are used for the MPX
0 Section 3, Counterbalancing
application only
0 Section 5, Film-in-place Switch Connection
4 Install Ion Chamber power cable into the table,
and secure t o the existing Bucky cables Ref-
0 Section 6, Vertical Positioning Check
erence Illustration 3-29
0 Connect MIS 128
5 Remove pre amplifier bracket, and pre-
amplifier, from the ion chamber bracket Ref-
0 Connect MIS 594
erence Illustration 3-30
1
ION CHAMBER
TY RAP POWER CABLE TYRAP
37233 P10
6 Remove Ion Chamber bracket from the Ion 7 Remove the Pre-Amp cover from the pre-
Chamber Reference Illustration 3-31 amplifier housing Reference lllustratlon 3-32
SCREW ION
CHAMBER
i
I
i
1
i
; *
t
ION CHAMBER
37224 P10
BRACKET
ION CHAMBER WITH BRACKET
ILLUSTRATION 3 31 ,.
PRE AMPLIFIER
P10
PRE AMPLIFIER WITH COVER
ILLUSTRATION 3 32
8 Remove the (2) 6-32 hex nuts from pre- 9 Loosen the (2) hex head bolts on the table
amplifier insulating plate Reference Illustra- which secure the tables pre-amplifier bracket
tion 3-33 Reference Illustration 3-34
PRE AMPLIFIER PRE AMPLIFIER
I
/
NUT
INSULATING
PLATE
37227 P10
10
TABLE PRE AMPLIFIER MOUNTING BRACKET
ILLUSTRATION 3 34
ILATING
rE
NUT
MOUNTING SCREWS
37229 P10
PRE AMPLIFIER INSTALLATION
ILLUSTRATION3 35
10 Install the pre-amplifier insulating plate and 11 Install the cover and four screws on the pre-
pre-amplifier (with the cable end facing the amplifier housing
mounting side of the bracket) onto the pre-
amplifier bracket Secure with the (2) 6-32 x 12 Interchange the wires marked 4 and 6 on the
3/8" screws from the ton Chamber installation terminal strip, then switch the tags on the
kit and (2) 6-32 hex nuts removed in step 8 wires Reference Illustration 3-36
Reference Illustration 3-35
13 Remove the (4) screws from the table preamp-
Iifier bracket Reference II Iustration 3-35
D
W
a
36991 D10
TERMINAL STRIP IDENTIFICATION
ILLUSTRATION 3 36
14 Install the terminal strip marker from the 17 Install the (2) 6-32 x 1/4" screws from the
termination kit (with letter facing up), onto installation kit, through the (2) small Ion
the table pre-amplifier bracket and secure Chamber holding brackets and into the side of
with the four screws removed in step 13 the Ion Chamber Reference Illustration 3-37
37231 P10
ION CHAMBER INSTALLATION
ILLUSTRATION 3 37
-a SECTION 10 -GRID INSTALLATION 3 Place support bracket on pivot pins (head end
side of bucky)
CAUTION
4 Loosely mount grid on springs, square the grid
Free grid of i t s wrappings with care, the and tighten all screws Reference Illustration
riveted tabs and brackets must not bend' 3 39
Engage the pivot PIns Into mounting bracket
5 Check grid freedom of motion by manually
Reference Illustration 3-38
pushing against grid and releasing it suddenly
2 Mount support bracket to grid Reference Illus- Oscillation should continue for a t least 20
tration 3-39 seconds
PIVOT
PIN
37769 DO1
PIVOT PINS
ILLUSTRATION 3 38
SUPPORT
BRACKET
SPRINGS
\ d /
37770 DO1
SUPPORT BRACKET
ILLIJSTRAT ON 3 39
-
SECTION 11 X RAY TUBE UNIT MOUNTING 7 Install collimator interface plate and align t o
Direction SM D5303C, Sentry Three Automatic
Move the support arm down t o position shown Collimator, Section 4-2
in Illustration 3-40
8 Install collimator t o aid in balancing the
Remove red round bar from the column, while system
holding vertical carriage in place Move the
carriage down a little bit more, so that the 9 Loosen x ray tube flange mounting plate
counterweight carriage clears the holes in the Rotate tube and collimator assembly so that
column collimator points down Retighten mounting
plate
Insert the red bar through the column holes 10 Install the new collimator SID scale sticker
until counterweight carriage rest on bar supplied with the VRT table
Loosen x-ray tube flange mounting plate Remove the two front cover mounting
Reference Illustration 3-40 screws and pull the front face of the
collimator forward
Rotate the x-ray tube flange mounting plate
180" from the position shown in Illustration 0 Rotate the SID drum t o the 44 cm scale
3-41, and move the tube flange mounting plate 0 Apply the new SID sticker over the existing
forward about 25 ' scale by rotating the drum
FLA
37774 DO1
TUBE MOUNTING
ILLUSTRATION3 40
*
REV 1 DIRECTION 17073
For MX 100 tube insert 24 pc 2 4 kg and 19 3 Install cable drape saddle (this saddle locks into
pc 24kg place with brackets from inside the column)
For MX 75 tube insert 25 pc 2 4 kg
1 TRIM WEIGHT
37773 DO1
COUNTERWEIGHT INSTALLATION
ILLUSTRATION 3 41
If the system has a MST 625-11 generator 2 Route HV cables, collimator/stator, SID, and
vertical carriage cables over saddle, into cable
Refer t o MST 625-11 Service Manual A0651B/BE, take-up and then into table base Reference
Chapter 3 and perform the following section Illustration 3-42
FULCRUM
SADDLE ASS EM8 LY
8 RACK ET
CABLE
TAKE UP
0 HV cables must be on the outer edges of Lower component tray I/F connections P2a
cable takeup P3a, PSa, P6a, P7al P8a, P1 Oa
0 Collimator electronic harness connections P2,
3 The following cables must be routed under
P3, P4 P5, P6, P8, P9 P7
lower component tray
4 Connect incoming power to KL4, with the use
0 Upper component tray I F connections 01, of ring terminals Reference Illustration 3-43
02,03,04,05,06,08
CAUTION
NOTE Push on terminals will permit an unsafe
condition when the lower componet tray IS
Cable 08 is emergency off pulled out
WIRE X
37605 D10
LeadLabel 2 3 4 5 6 7 8 9 10 11 12
I
Nom
volts 0 120 192 200 207 214 220 225 230 235 240
r
37605 010
POWER CONNECTIONS
ILLUSTRATION 3 43
5 Set wire ' X ' (KL3 Relay) to incoming power line 6 Ty-rap excess length of cable t o fulcrum assem
voltage Reference illustration 3-43 and bly bracket t o allow easy access t o fulcrum
schematic on page371 of Chapter 7 assembly
0 220 VAC is required from the autotrans- 0 Verify that there is no interference between
former cables and table mechanics
To turn on table power rotate the red Mount belt connector to horizontal carriage
emergency stop button clockwise Reference Illustration 3-44
Check DC power supply voltages T R l , TR2, Mount cable take-up connector to horizontal
andTR3 (voltage should be 22 2 1 lVDC carriage with (4) Allen Head screws
37777 DO1
BELT CONNECTOR
ILLUSTRATION 3 44
'e -
SECTION 14 MOUNTING TABLE TOP 3 Slide table in over plastic roller, press and hold
foot treadle (all the way down)
4 Carefully move table top over brake and
1 Place table top on table base, and remove the
continue to move until the plastic roller (on
four end pieces from table top Reference Illus-
foot end of table) is cleared
tration 3-45
5 Push table top down and slide it over plastic
2 Move table top to foot end until you clear the roller Center table top, release foot treadle,
plastic roller Reference Illustration 3-46 and remount the four end pieces on table
frame
END PIECES
II END PIECES
37775 DO1
END PIECES
ILLUSTRATION 3 45
37776 DO1
TABLE TOP
ILLUSTRATION 3 46
-
SECTION 1S TOMO RELAY INSTALLATION NOTE
Installation of the tom0 relay prevents the torno This precedure assumes that the generator
sweep from starting until the rotor is up to had been installed previously If this is a
speed,causing improper exposures completely new installation, only install the
b wire from 20TB1-9 t o H5 in the SARC II
0 Mount the tom0 relay a t the top of the
20TB1 panel
-
DXS 650 I I APPLICATION (WITH SARC-II)
b
2
In the generator, move the wire on 20TB1-612
t o 20TB1-12 (12 should be a unused terminal)
r JUMPER
- - - -
rL
marked 72 t o 20TB1-12 and the lead marked I Remove
812 t020TBl-Bl2 I andtape
I both ends
0 This contact will prevent a tom0 sweep
before the rotor is up to speed
b
3 Move the wire on 20TB1-116 (coming from H5
in the SARC II ) t o 20TB1-9 (9 should be an I
20781 9 2OTBl A2
I
unused terminal) I
b I
4 Remove and tape both ends of the wire I
I
connecting 20TB1-79 and terminal H6 in the
I Move from
SARC II I 2OTBl 116
b I to9
I
5 Connect the normally open contact lead I f
TOMO RELAY
marked 44 to 20TB1-79 and the lead marked 45 I
2OTB1
t o 2OTB1-116 I 20161
I
'i 1-
B12
I -812
0 This contact replaces the rotor controller I 116
exposure interlock contacts I
T
I 0 CONTACT
b
6 Connect the relay coil lead marked A2 t o
2OTB 1-A2 and the lead marked 46 to 20TB 1-9
I
L - 79
2\, ROTOR EXPOSURE
- 12
0 - - -
MSI 850 WMSI 1250 IV 81MST 625WMST 1050-11
APPLICATIONS (WITH SARC-II)
NOTE
-
SECTlON 16 FlNAL STEPS After all checks tests, and data records have been
completed perform the following
Refer t o the following manuals for completing the
installation 0 Select test films t o be retained
0 MST 625-11 Generator Service Manual SM 0 Complete and return Product Locator cards
A0651B/ BE, Section 31
TABLE OF CONTENTS
1 TABLE TOP 4- 1
2 FOOT TREADLE 4- 2
3 BUCKY CARRIAGE 42
4 VERTICAL CARRIAGE 43
8 TOMOGRAPH IC 4-4
LONGITUDINAL
LATERAL
SCREWKOUNTERNUT
BRAKE
ADJUSTMENT SCREW
37850 DO1
LATERAL
BRAKE
ADJUSTMENT SCREW
SECTION 4 - VERTICAL CARRIAGE When necessary, adjust the lock by tightening the
nut /counterlock Reference Illustration 4-3
1 Depress vertical lack switch, then move tube
unit through full vertical range of travel 0 Repeat procedure as necessary to achieve
moving efforts of 8 pounds maximum and lock
0 Movement must be easy and quiet holding efforts of 22 pounds
0 Must have cushioned end stops a t the top SECTION 5 -TUBE STAND DRIVE
and bottom
Rotate x-ray tube unit to chest stand location,
Coupling bar must slide easily through
(drive the tube stand with the collimator tube
vertical carriage
stand button)
2 Connect spring scale to the collimator handle 0 Column must stop automatically at 40" SID
and measure the lock holding force in the
up/down movement 0 Column must stop automatically a t 72' SID
0 Chest stand LED must be ON a t the control
0 Lock holding force, must be (22 pounds) panel
minimum Repeat step 1 using control panel tube stand
buttons
-
SCREWKOUNTE RLOCK
I-
3 Rotate tube unit off table chest radiography in SECTION 7 - X-RAY TUBE ALIGNMENT
both directions and then return to 0 degrees
(beam down) position 1 Lower tube arm t o within 5" above table top
0 Tube unit will easily rotate with a positive 2 Draw a large X on a sheet of paper and place it
detent a t the f 90" position on table top
0 Detent does release when angulation switch 3 Turn collimator field light on and collimate t o a
i s pressed small field size
0 Tube unit has detent a t 0" position 4 Center the X in the field light and tape in place
SECTION 6 -TUBE STAND ROTATION /DRIVE 5 Raise tube arm t o i t s upper limit of travel
1 Couple unit into RAD MODE 0 Projected center of light field must not shift
more than 1/4" in any direction from the
2 Press column rotation switch at the collimator, center position established a t the lower
and rotate column 90" in each direction and height
return t o 0" (beam down) position
0 Column stops automatically at the end of the 0 Coupling bar moves freely
travel
0 No cable interference while unit is in motion
There should be no "rasping" sound
0 Bucky hits end stop a t 40" (215 mm) plane
Clearance between bucky and horizontal height
carriage is okay
0 Cassette tray strikes the control panel when
4 Repeat the above procedure with column loadinghnloading, prior to a Prep and
rotated - 90" position TOMO DEMO
-
SECTION 9 CON MAG SECTION 10 -CON SID
1 Couple unit into CON MAG MODE and drive the 1 Couple unit into CON SID MODE and drive the
tube stand (with the tube stand button a t the tube stand (with the tube stand buttons a t the
control panel) I control panel)
TABLE OF CONTENTS
-
SECTION 1 BUCKY CENTERING LATCHES - Couple into RAD MODE and check latch
function
0 When centering bucky latches interfere with
TOM0 coupling bar, latch distance can be too - Latches should move in and out without
tight Reference Illustration 5-1 interference with the coupling bar
- To adjust, remove latch cover (remove
screws and pull down/forward the latch NOTE
cover) Reference Illustration 5-1
- Open either lock screw on the left or right
If the plunger is not centered to the coupling
bar longitudinally, shift bucky left or right t o
side and set latch distance (turn adjustment
achieve centering
screws in or out) Reference Illustration 5-1
37910 DO1
BUCKY SWITCHES
ILLUSTRATION5 1
SECTION 2 -COUPLING BAR STRIKES BUCKY SECTION 3 - NO BUCKY LIGHT ON THE CONTROL
PANEL
0 When coupling bar strikes the bucky or the
plunger does not engage deep enough 0 Microswitch may not be in contact with
cassette tray Reference Illustration 5-2
- Bucky strikes bar - remove (4) screws and
turn back adjusting screws Reference Cassette tray does not go in all the way
Illustration 5-2
- Check connections, and verify that they
- Adjust bucky coupling unit the required are in line with each other Reference
distance and tighten all screws Reference illustration 5-2
illustration 5-2
ALIGNMENT
CONNECTlONS
NT SCREWS
COUPLING BAR
ILLUSTRATON 5 2
SECTION 4 - BUCKY LIGHT IS FLASHING SECTION 6 -CANNOT COUPLE INTO CON SID MODE
100 CM SID
MICROSWITCH
MICROSWITCH
SOLENOID
PLUNGER
37959 DO1
MICROSWITCHLOCATION
ILLUSTRATION 5 3
SECTION 7 -COUPLING PROBLEMS - loosen lock screws, and turn socket head
screws in or out Reference Illustration 5-4
0 Are caused, (while coupling into TOM0 or
CON MAG) when upper and/or lower NOTE
coupling plate jaws are set too tight
The engaging cam to the coupling plate jaws
- Adjust coupling jaws on fulcrum tower should have a snap fit
Reference Illustration 5-4
LOCK
37908 DO1
COUPLING JAWS
ILLUSTRATION 5 4
p BRAKE
TRACK
37907 DO1
BRAKE GAP
ILLUSTRATION 5 5
SECTION 9 - LOCKED OUT BETWEEN MODES - Tube i s down and microswitch is actuated
Reference lllustratlon 5-6
0 Mode button lights on control panel are flash-
ing, check coupling bar microswitches Ref- - Select TOM0 (tube and guide pin are down
erence Illustration 5-6 and switch is actuated) Reference Illus-
tration 5-6
- erence
Rotate column t o position as shown
Illustration 5-6
Ref-
NOTE
COUPLING BAR
MICROSWITCHES
37906 DO1
COUPLING BAR
ILLUSTRATION 5 6
--+-'+-----( 44
43 35 30
I I l l 1
I I
ON OFF
OVERRIDE 2
NORMAL
OVERRIDE 1
\\ Sentry Three I
--
It --I\
36990 D10
SENTRY THREE AUTOMATIC COLLIMATOR FOR THE VRT RADIOGRAPHIC SYSTEM
ILLUSTRATION 5 7
The switches that have changed are the following Switch marked ROT - Push and hold in
Switch marked FOOT - Push and hold the button t o release the locks Man-
in the button The column, with col- ually move the column, with colli-
limator and tube, will automatically mator and tube, to the desired
move to the right (when facing the ROTATIONAL setting and then release
column) until the button is released the button When the button i s re-
When the button is released the leased, the locks will return t o the
column will stop moving locked position
d n X TI
r
- -
L ;f
0
nL
z$ 0
0 P
56 P
N
__i ---
I
I;
.i
P
P
-
- P 0
1
d
f
4
F
c
X I I n 1 !Oh ! 1 0 I 8 I
r
L -
30
I
?J 3
P '6
+ 5
111: I
f;
m-
0
I*
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
-
SECTION 11 VRT FLOW DIAGRAMS
11-1 Power-Up
I INITALIZE SYSTEM I
DISPLAY E l 5 NO
4 PRE RAMTEST OK?
STOP SYSTEM
I INlfALlZE LED5 I
FLASHING T O M 0
RAD, CONSID
CONMAG
BRAKES, SOLENOIDS
~ ~
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
~
11-2 Selecting CON SID Mode of Operation - Move the bucky a little t o the left and
t o the right
0 Press/hold CON SID button until the column - When you can feel resistance in both
automatically stops a t table center directions, the bucky i s properly
coupled
0 Couple the Bucky manually i f not already 0 Tube stand can now be moved with tube
coupled stand buttons to the left or the right
I CONSIDPROGRAM I
I EXITMAINPROGRAM I I SETlNDEXlOOSID I
I YES
KEYBOARD LOOP NO
SID 100 ?
CONSID ?
NO
Cp[+]
KEYBOARD LOOP
v
.1 No PUSH OUT PLUNGER
I EXITMAINPROGRAM I RELEASE COUPLING BAR
REDRAW CATCH BUCKY
OPEN VERTICAL BRAKE
BUCKY BRAKE ON
CLOSE K7
OPEN K2
RELEASE EXPOSE
I
NO
KEYBOARD LOOP
KEY PRESSED7
-
I I
c $.
OTHER MAIN KEY DRIVE COLUMN TO THE DRIVE COLUMN TO THE
EXIT MAIN PROGRAM
11-3 Selecting CON MAG Mode of Operation - Move the bucky a little to the left and
to the right
a
0 Presdhold CON MAG button until column
automatically stops at table center When you can feel resistance in both
directions, the bucky is properly
0 Couple bucky manually if not already coupled
coupled
--
GENERAL ELECTRIC MEDICAL SYSTEMS
I
KEYBOARD LOOP
CONMAG 7
YES
KEYBOARD LOOP
CONMAG? 14
NU
KEY PRESSED
CENTER COLUMN
SID 100 ?
I
TUBE BRAKE ON
BUCKY ROUTINE
VRT TABLE SERVICE MANUAL
DIRECTION 17073
1
a KEYBOARD LOOP No I
c
BUCKYCENTERED?
NO
KEYBOARD LOOP
KEY PRESSED ?
'I
11-4 Selecting RAD Mode Operation 0 When the RAD led i s on, this indicates that
the bucky is coupled t o the tube stand
0 Presdhold RAD button a t control panel,
until column automatically stops at table
0 When the RAD led is flashing, this indicates
center that the bucky is not coupled t o tube stand
RAD PROGRAM I
KEY PRESSED
CENTER COLUMN
Cp!q*I
KEYBOARD LOOP
RELEASE COLUMN
ROTATION
BUCKY ROUTINE
YES
NO
1
PUSH OUT PLUNGER
EXIT MAIN PROGRAM FIX COUPLING BAR
REDRAW CATCH BUCKV
OPEN VERTICAL BRAKE
RELEASE SID UNIT
CLOSE K7
OPEN K2
RELEASE EXPOSE
LED RAD IS ON
I
KEY PRESSED?
1 I
11-6 Selecting Tom0 Mode of Operation Select speed and angle (press the desired
speed/angle buttons)
0 Presdhold Tom0 button until column auto-
matically stops at table center @ Presdhold PREP button until column
reaches PREP position
0 Couple bucky manually if not already 0 Select PLANE HEIGHT (using the up or down
coupled button)
- Move the bucky a little t o the left and To make a TOMO DEMO SWEEP press/hold
t o the right TOMO DEMO button until sweep has been
completed
- When you can feel resistance in both
directions, the bucky i s properly - When in TOMO, 40°, and SLOW SPEED
coupled the LED's will continue t o flash, select a
new angle
TOMO PROGRAM
9
KEYBOARD LOOP KEY PRESSED
.1
I CENTERCOLUMN I
f=3<Z*J KEYBOARD LOOP
YES
NO
SID 100
1 TOMO 7 I
I TOMO? I c
PUSH OUT PLUNGER
RELEASE COUPLING BAR
REDRAW CATCH BUCKY
OPEN VERTICAL BRAKE
COUPLE SID UNIT
CLOSE K7
CLOSE K2
CLOSE K5
NO SPEED + ANGLE
SELECTED?
F FLASHING PREP
I
PREP POSITION
PREP 7
T
EXIT TO MAIN PROGRAM
1
I
PREP LED ON I
NO
KEYBOARD LOOP
KEY PRESSED ?
I
-1-
OTHER MAIN KEY
EXIT MAIN PROGRAM
I I
11 II
b
DRIVE COLUMN TO THE
RIGHT
-
I 1 DRIVE COLUMN TO THE
LEFT I
T O M 0 FLOW CHART
ILLUSTRATION 5 12 (Continued)
BUCKY
COMPONENT STATUS
vc IM4 solenoid
(coupling bar) t ----- ENERGIZED
I BU IS5 switch latches out
I sensor out ENERGIZED
1
vc IS3 switch ----- CLOSED
BU I M3 solenoid plunger
in coupling bar ----- ENERGIZED
vc IS4 switch
(coupling bar is locked) t ----- OPEN
1 BU IS4 switch plunger
BU I M 1solenoid plunger
CLOSED
i s lifted ENERGIZED
1
BU IM 5 selenoid latches
are in OFF
FULCRUM TOWER
, I BU 1% Switch latch in I
sensor in OFF I
VC/M5 bottom fulcrum ENERGIZED ENERGIZED
solenoid is coupled
1c
VUS6 bottom fulcrum CLOSED BUIS3 top fulcrum switch i s coupled CLOSED
is switch coupled A
FULCRUM COUPLED
I T0MOLED"ON" I
Several tests are performed and on the LED Display Micro Processor Board
an error code i s displayed which hopefully will E01 - not implemented
show the serviceman the part of the hardware
E02 Keyboard selection error Replace 025
which is defective Replace D26
E03 LED s driver error Replace D29
The diagnostics reside i n the eprom and i s Replace D30
therefore always available without the necessity of
LED s driver error Replace 035
other software tools
E04
. 7 segment driver error
Replace 036
Replace D32
Replace D33
12-1-2 General Guidelines
E06 Chest stand out DO D4Time for Replace 034
MPX Replace D37
When troubleshooting errors are encountered i n
power-up software, it is strongly recommended
I E07 I - not imdemented ~
Power Board
that the system be "reset" to verify if any errors
reoccur before beginning t o troubleshoot the E08 Tube stand control motor speed Replace D13
control Replace D12
system Reset the system by powering down or
using the 'reset" switch on the micro processor EO9 Solenoids Replace D15
Replace D l 4
board This will cause the power-up diagnostics t o
re-execute This will determine if the error was due E10 Solenoids Replace D17
Replace D16
t o a momentary fault or i f additional/similar errors
are also present E l1 Relay Control Replace D19
Replace D18
After troubleshooting all errors, re-execute power- NOTE If diagnostics loops on addresses (meaning it counts
up diagnostics by resetting the system in order t o from E08 to E l 1) then there are problems with the data bus
DE2 The lines may be shorted together or shorted t o GND or
verify if any errors till exist vcc
- ~~
ERROR
-
CODE
REASON
Micro-Processor Board
WHAT TO
DO NR
TEST
000
ICPU Stack Test
TEST ERROR CODE
E12
E12 CPU stack error Replace D10 000 CPU General Test E13
and D13 001 Eprom Test Checksum E14
(Rams) 002 Ram Test E15
Replace D7 003 VIA 6522 Timer Test E16
(CPU) 004 Interrupt Test 1 E17
004 Interrupt Test 2 E18
E13 CPU error Replace 07 004 Interrupt Test 3 E19
(CPU) 005 N D Converter Test E20
E14 Eprom Test checksum error Replace D17 006 InterfacelLatch Test E02 Error code
(EDrom) 006 InterfaceRatch Test E03 isthe
E15 Ram error Replace D10 006 InterfacelLatch Test E04 address of
and D13 006 InterfacelLatch Test E05 the defec
(Rams) 006 InterfacelLatch Test E06 tive latch
006 InterfacelLatch Test E08
E16 Via timer error Replace 0 5 006 InterfacelLatch Test EO9
(VIA) 006 InterfacelLatch Test E10
E17 Interrupt error Replace 0 5 006 InterfaceRatch Test E l1
I
E18 Interrupt error Replace 0 5 007 LED Test - all LED s should be on No error code
E19 Interrupt error Replace DS must be visually
controlled
E20 N D Converter test error
In this test the 5V power supply of NOTE If an error occurs the error code is displayed on panel
the system is measured on channel 7 and the system hangs in a death loop To restart the system
of D22 If the value of the voltage is the "RESET BUTTON (51) has t o be depressed or a power
not within 4 8V and 5 2 V error E20 is offlpower on cycle has to be done
displayed
1 Check SV power supply 12-3-2 Entering Board Level Diagnostics
2 Check 5V at D22n Replace D22
3 Check address of MUX D22 Replace D28 Diagnostics operation, including power up diag-
should be HHH Replace 0 3 1
Replace D16
nostics, is selected if SW2 (Diagnostic SW) on the
4 CheckSVatD16l14 Replace 024 processor board is in the ON position
E21 - n o t implemented
It is necessary to press the RESET (51) button on a
E29
running unit to enter the diagnostics mode
E30- Power supply 1 voltage below 22V Check power
-
53s SUDDIV
OTE If agnostics loopson addresses (meaning it counts
Nr 1 The tests are the same as in power-up diagnostics,
ut It I S p' ,sibleto select the followinq options
from E08 t o E06) then there are problems with the data bus
DE1 The lines may be shorted together or shorted t o GND or CONSOLE
FUNCTION
VCC It may be the same problem if the display shows SWITCH
unimplemented error codes The display hangsjust on this
"DE1 I FAST SELECTS STOP LOOP ON TEST MODE
All the tests are executed again and again the
test number is shown on the display
An error code is shown on the display too This
test mode does not stop at an error it continues
12-3 Operator - Interface testing
MED SELECTS STOP LOOP ON TEST ERROR
12-3-1 Entering Power-Up Diagnostics DOWN SELECTS NEXT TEST EACH TIME DEPRESSED
This test loops on test mode like FAST but in
Normal table operation, including power up diag- steps
nostics, is selected if SW2 (Diagnostic SW) on the UP SELECTS REPEATLASTTEST
processor board i s in the OFF position During The test which was executed last will be
power up diagnostics, the following tests are repeated
Test number and error is shown on the display
executed in this sequence
MEMORY CELL TEST - The data pattern 55H i s A 'Walking 0" is written and read back from each
written t o all memory locations and then verified register
The data pattern i s then inverted t o AAH and
memory test again This test verifies that each An error will cause EXX error message on the SID
memory cell can be set and reset display XX is the corresponding chip select address
from the defective register
Any RAM error will cause an E l 5 error message on
the SID display
A simple 8 bit checksum IS calculated and compared The LEDs on the panel and the SID display are
with the checksum in memory location FFF9H switched on for about 1 second The LEDs must be
visualIy control led
An error will cause an E14 error message on the SID
display No error message will be displayed
-
CHAPTER 6 ADJUSTMENTS
TABLE OF CONTENTS
1 GENERATOR ADJUSTMENTS 6- 1
1-1 DXD 350-11 6- 1
1-2 DXS 650-11 6- 1
1-3 MSI 850-11/1250-IV 6- 1
1-4 MST 625-11 6- 2
1-5 MST 1050-11 6-3
1-6 Phototimer 6-3
1-7 VRT Cassette Tray Adjustment 6-4
1-8 SENTRY-Ill Auto Collimator Adjustment 6-4
1-9 SENRTY-Ill Adj Multi-Size Cassette Holder 6-6
1-10 Multi-Sized Cassette Holder 6-6
CHAPTER 6 -ADJUSTMENTS
0 Section 10, X-ray Tube Filament Selection 0 Section 8-3, Connecting the High Voltage
Divider
0 Section 11, High Voltage Cables
0 Section 8-4, High Voltage Output and
0 Section 12, High Voltage Divider Connec- Primary Current Check
tion
0 Section 9, Meter Accuracy and MAS Adjust-
0 Section 13, Line Regulation Adjustment ment
0 Section 14, Radiographic MA Adjustment 0 Section 10, Line Regulation Test and
Preliminary No Load Voltage Adjustments
0 Section 15, Space Charge Compensation
0 Section 12, Tube Protector-Tube Rating Set-
0 Section 16, KVP Calibration UP
1 Refer t o SM A0650B, DXS 650-11 X-Ray Unit 4 Refer t o SM D1154A or SM D1176A, X-Ray Tube
Service Manual, Chapter 3 and perform the Unit Part 7, Run-in Procedure
following sections
1-3 MSI 850-ll/1250-IV Generator Adjustments
CAUTION
1 Refer t o SM A0701B/A0702A, MSI 850-11/1250-1V
Before starting disconnect the High Voltage X-Ray Unit Service Manual, Chapter 3, and
Cables from the X-Ray Tube perform the following sections
0 Section 5, Adjustment considerations 0 Section 5, Adjustment Considerations
0 Section 6, Power Supply Tests and Adjust- 0 Section 7, Power Supply Tests and Adjust-
ments
ments
0 Section 8, Regulator and Voltage Adjust- 0 Section 23, Precontacting and Millisecond
ments Timing
0 Section 13, Line Regulator Test 3 Refer t o SM D1154A or SM D1176A, X-Ray Tube
Unit Part 7, Run-In Procedure
0 Section 14, Radiographic Primary No-Load
1-4 MST 625-11 Generator Adjustments
0 Section 15, Timer
1 Refer t o SM A0651B/BE, MST 625-11 X-Ray Unit
0 Section 16, mAs Compensation Service Manual, Chapter 3 and perform the
following sections
0 Section 17, Tube Protector kVp x mA Signal
CAUTION
0 Section 18 Tube Protector-Tube Rating
Setup Before starting, disconnect the High Voltage
Cables from the X-Ray Tube
0 18-1, Determining Needs for Rating
Cards 0 Section 5-4, Anode Rotation, Maxiray 75
Tube Unit
0 18-2, Offset Adjustment 70R56
0 Section 5-6, Mammography Requirements
0 18-3, Rating Card Setup
0 Section 6, Power Supply Tests-No Load
0 Section 19, mA Overload Adjustment
0 Section 7, Tube Selection and Rotation
0 Section 20, Radiographic mA Adjustments
0 Section 8, KVP Servo
0 20-1, mAAdjustments
2 Refer t o Service Manual SM A8105D, SARC-II
0 20-3, Filament Adjustments Service Manual, Chapter 3 and perform the
procedures described in Section 6-1 through
0 Section 21, Space Charge and kVp 6-9, and the following sections
Adjustments Perform 21-1 , Conventional
Tube, Space Charge, and kVp 0 Section 10, Preliminary Checks
0 Section 22 Filament and Standby Currents 0 Section 1 1,Panel Meter mA-mAs Check
0 Section 16, Tube Protector Adjustment 0 Section 15, Two Radiographic Tubes,
Emission Test
0 Section 17, mA Overload Adjustment
0 Section 16, Tube Protector
0 Section 18, Radiographic mA Adjustment
0 Section 17, mA Overload Adjustment
0 Section 19, Radiographic kVp Check
0 Section 18, Radiographic mA Adjustments
0 Section 20, Millisecond Timing Final Adjust-
ment (73R9) 0 Section 19, kVp Waveform, Precontacting
Adjustment
0 Section 21, Radiation mAs Final Adjustment
(76R27) Section 20, Millisecond Timing Final
Adj ustment (73R9)
0 Section 22, kVp Derating Check, Long
Cables 0 Section 21, Radiation mAs Final Adjustment
(76R27)
0 Section 23, Space Charge Compensation
Check 0 Section 22, kVp Derating Check, Long
Cables
3 Refer t o SM D1154A or SM D1176A, X-Ray Tube
Unit Part 6, Filiment and High Voltage testing 0 Section 23, Space Charge and kVp Adjust-
ment
4 Refer t o SM D1154A or SM D1176A, X-Ray Tube
Unit Part 7, Run-in Procedure 0 23-1, Conventional RAD Tubes
0 Section 17, Ion Chamber Detector Sensi- 5 With 6 84 f 02 VDC applied to pin 3 of CT and
tivity Adjustment com pin 4 insert 17" cassette within the
longitudinal jaws of the cassette tray
0 Section 17-2, The VRT Table i s not included
in the 17-2 section title However, the in- 0 Adjust CT/R4 for 6 43 f 26 VDC a t pin 8
structions provided are applicable t o the
VRT Table 6 Insert 8" cassette within the longitudinal jaws
Insert 8"cassette within the lateral jaws 0 Section 14, X-Ray Tube Unit and Collimator
Leveling Also refer t o tube stand leveling
0 Adjust CT/R2 for 1 92 f 08 VDC instructions in the VRT service manual
Repeat steps 2 and 3 until voltages are within 0 Section 15, Optional Scale Label Attach-
tolerances ment
U Section 16, Alignment of Bucky-Cassette 0 If necessary adjust R91 (A101 until light
Tray Mating Parts field i s within 1/4" of the 10-1/4"
pattern (preferably slightly undersize)
0 Section 18-2, Adjustment for Chest Unit
Whose Image Receptor can be Positioned K Remove the 14x17 cassette and insert the
Horizontaly 8x10 cassette with the 8" dimension along
the longitudinal axis
0 Section 19, Source-to-Image Distance (SID)
Proportional Voltage Adjustment L Insert the tray into the bucky (collimator
should be in the automatic mode)
Section 20, Damping Adjustment
M Move the DVM positive lead to TP1 on the
Section 21, Motor Stall Adjustment A10 card
1 Using the following procedure, adjust the N Adjust R112 (on All Lat table card) to
collimator field size +
obtain + 1 90 02VDC
A Align the tube unit and collimator t o the 0 Remove the 8x10 cassette and insert a
table bucky 14x17 cassette with the 17" dimension
along the longitudinal axis
B Set focal spot t o 40" above the bucky film
plane (use the collimator measuring tape) P Adjust R97 (on Lat table card) t o
obtain + 4 25 f 02VDC
C Place a 8x10 cassette in tray and position
the 8" dimension along the transverse
(lateral) axis
Q With the field light on, measure the
longitudinal dimension of the light field
D Insert the tray into the bucky (collimator projected onto the RAT pattern
should be in the automatic mode)
0 If necessary adjust R97 (A10) until light
E Connect a DVM to the following field is within 1/4" of the 10-1/4"
connections pattern (preferably slightly undersize)
C Repeat step B using a 10x12 and then 1-9 SENTRY-Ill Adjustments For Multi-size Cas-
a 14x 17 cassette sette Holder
D Select 70 kVp, 1005, 07 sec, and RAD Refer t o Sentry-Ill Automatic Collimator Service
mode Manual, SM D5305C, Chapter 3 and perform the
following sections
E Insert a 8x10 cassette into bucky
0 Section 22-2, Adjustment for Chest Unit
F Rotate collimator 45" and make an Cassette Size Sensing System
exposure
0 Section 22-3, Adjustment for 14x17' Image
G Measure each square and verify that the Receptor Size Chest Unit
outside dimensions of the square are 114"
smaller than the film edge
1-10 Multi-sized Cassette Holder Adjustment
H Repeat steps F,and G using a 14x17 cassette
Refer t o SM B3515A, Multi-sized Cassette Holder,
0 Section 23, Adjustment of Quantamat Film Chapter 3, and perform the following sections
Area Switching
0 3-8, Field Size Potentiometers
0 Section 24, Large Field Size Check
0 3-9, Field Size Calibration
0 Section 25, Tube and Cutoff Check and
Adjustment 0 3-10, Counterbalancing
a -
SECTION 2 VRT TABLE ADJUSTMENTS 2 Loosen Allen head screws on pulley plate
ALLEN HEAD
SCREWS
' ADJUSTMENT
SCREW
37749 DO1
BELT ADJUSTMENT
ILLUSTRATION 6 1
n
67 Copyr ghte 1985 by General Elect I C C 1
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
1 Engage column rotation lock and vertical travel 1 Remove motor assembly trim cover Reference
lock Illustration 6-2
\
MOTOR ASSEMBLY
TRIM COVER
37750 DO1
TRIM COVER REMOVAL
ILLUSTRATION 6 2
0 Drive tube stand t o head end of table and 0 Drive tube stand t o foot end of table and
place the bucky to foot end of table place the bucky t o head end of table
SET SCREWS
MOTOR
ASSEMBLY
POT
37751 DO1
MOTOR ASSEMBLY
ILLUSTRATION 6 3
a 2-4 Adjustment of Bucky Plunger t o Coupling Bar 0 Engage plunger and retract by going from
TOMO into RAD MODE
0 When plunger does not engage to coupling bar
after opening lock screw, adjust height differ- 0 Verify that plunger does not hit the bucky
ence by turning the height screw in or out frame Reference Illustration 6-5
(raising or lowering) Reference Illustration 6-5
0 Verify that the micro switches are opening/
0 To adjust 2 mm lift for TOMO, CONSID or CON closing when the plunger is engaging with the
MAG Mode of Operation, adjust tom0 bar lift coupling bar, or when plunger retracts from
screw See Illustration 6-5 coupling bar Reference Illustration 6-5
NOTE
NOTE
If plunger i s not centered to the coupling bar
This adjustment can only be done while going longitudinely, refer to Chapter 4, Bucky Cen-
into one of these modes tering Latches
MICRO SWITCH
MICRO SWITCH
LIFT SCREW
BUCKY FRAME
HEIGHTSCREW
PLUNGER
37752 DO1
PLUNGER ASSEMBLY
ILLUSTRATION 6 5
2-5 Tube Stand Drive Speed Adjustment 4 Reset counter t o zero by pushing in the bottom,
on counter
Fast Speed Adjustment
5 One person will presdhold the TOMO DEMO
1 Select TOMO MODE, 40" and fast speed a t the button a t the control panel and a second per-
control panel son will push/ hold in the microswitch (towards
the now rotating coupling) and keep time for
2 PREP unit and presdhold the TOMO DEMO (1) minute and then release Reference Illus-
button, until motion stops tration 6-6
3 Loosen (4) motor mounting bracket nuts and 0 If the counter does not show 132 f 1 RPM,
push motor assembly down adjust pot (P4) located on the power board
(upper component tray) Reference Illustra-
Reference Illustration 6-6 tion 6-7
YI R
MICRO SWITCH P1
P2
P3
MOUNTING
BRACKET NUTS PU
P5
co
/
I I 1
37753 DO1
DRIVE MOTOR
ILLUSTRATION 6 6 POWER BOARD
ILLUSTRATION6 7
37754 DO1
a
CopyrighP 1985 by General Electric Co 6-12
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
8
Presslhold TOMO MODE button until the
TOMO light is on and couple into TOMO
MODE
14 Manually adjust the left and right thumb- 1 Select a 4 second exposure time at the gen-
wheels on the collimator until both sets of erator control console
collimator blades are closed
2 Presdhold MED speed button until the MED
To verify that both sets of collimator blades speed light is on
are closed, momentarily press the FIELD
LIGHT button t o see if any field light is 3 Makeatomoexposure
visible on the table
4 Compare oscilloscope waveforms on channel 1
When light is visible on the table, adjust the , (exposure time) and channel 2 (tomo sweep
thumbwheels until there i s no light time) Reference Illustration 6-8
-
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
2-8 Centering of X-Ray Tube t o Film 5 Again use collimator field light t o center the x-
ray field to the cassette b y rotating tube/
1 Couple the column t o the bucky in RAD MODE collimator assembly in and out t o obtain lateral
and set tube t o 100 cm SID and longitudinal alignment
2 Place a cassette into the cassette tray 0 If you do not have proper adjustment place
a level, vertically on column (both side and
3 Use the collimator field light t o adjust the x ray rear) and make adjustments using the
field t o the cassette following procedure
/* I
SOCKET HEAD
a
37763 DO1 37764 DO1
TUBE SUPPORT ARM TUBE COLUMN ALIGNMENT
ILLUSTRATION 6 10 ILLUSTRATION 6 1 1
a 2-9 Tom0 Phantom Plane Height Adjustment 7 Presdhold tomography PREP button until the
PREP light is on
With the power on, perform the following
8 Set x-ray controls t o 45 kVp 50 mA and 2
Load film in a cassette
seconds
/
Place cassette in the tray and insert into the
Bucky NOTE
Manually move the Bucky t o the center For the MPX application only, check the
position exposure times a t the generator control
console to insure that they agree with the
Presghold TOMO button until the TOMO light settings on the VRT control panel, for each
is on (coupled into the TOMO MODE) selectable sweep and angle
ADJUSTMENT
SCREW
COUNTER NUT
MOUNTING
SCREW
SOCKET
MOUNTING HEAD SCREW
SCREW
37756 D l 0
COUPLING PLATE
ILLUSTRATION 6 12
%--
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
~~ ~ ~
10 Press/hold REDUCTION SET button on the colli- 2-10 Tomo Plane Height Adjustment
mator This will allow the collimator blades t o
be operated manually 1 Select TOMO MODE on control panel (LED
display must be on)
11 Hold in the FIELD LIGHT button and center the 2 Drive plane height t o 000 by pressing the
collimator light with the circular pattern of DOWN button at control panel
the light field centered on the inclined surface
of the phantom 3 Select RAD MODE
4 Check (with a metric tape measure), the dis-
12 Adjust the left and right thumbwheels until tance between the horizontal bar and the
the size of the light field i s slightly larger than pointer Reference Illustrations 6-13,6-14
the phantom and release the FIELD LIGHT
button NOTE
13 Take a tom0 exposure at 40 mm, t 1 mm, Must uncouple t o get into this position
plane height
NOTE
NOTE
When moving the fulcrum assembly, watch
If a tom0 exposure is terminated before the for the service cable, push down the upper
selected exposure time is completed, press the coupling jaws while you are moving the ful-
tomography PREP button before starting a crum assembly
new tom0 exposure
- Must go out of TOMO MODE
14 Remove the cassette and develope the film - Distance for 0 mm plane height = 63 mm
t 1 mm, adjust if necessary
15 If the film indicates that the plane height
needs t o be adjusted, refer t o the TOMO - Distance for 2 15 mm plane height = 170 5
PLANE HEIGHT ADJUSTMENT mm 2 1 mm, adjust i f necessary
37757 DO1
MEASURING THE DISTANCE THROUGH BUCKY OPENING
ILLUSTRATION6 13
I n- HORIZONTAL BAR
ILLUSTRATION 6 14
5 To adjust the distance at the fulcrum tower, 7 Make adjustments by moving the triangle up or
pull out the fulcrum assembly by removing (2) down (adjust the toothed belt) Reference Illus-
fulcrum mounting screws Reference Illustra- tration 6-15
tion 6-9
TRIANGLE
/
TOOTHED BELT
37759 DO1
FULCRUM ASSEMBLY
ILLUSTRATION 6 15
8 Using the following procedure, readjust the 0 Turn power off and remove connector O2A
end stop microswitches (only if necessary), so
whenever the table is at 63 mm or 170 5 mm 0 Hook up ohmmeter t o the 10 turn 10 KSZ pot,
the upper or lower microswitch is actuated between pins 2 and 3 Reference Illustration
Reference Illustration 6-16 6-17
37760 DO1
ILLUSTRATION 6 16
0
1 OK*POT
37761 DO1
POT LOCATION
ILLUSTRATION6 1 7
0 Drive the triangle manually t o i t s upper 0 Reconnect 02A and turn power on
position (63mm)
0 Testing the Drive
0 Verify that you have a OQ reading
Select TOM0 MODE a t control panel
0 When adjustment is necessary, remove the
pot by loosening (2) set screws and pull Couple into the mode and pull out upper
down this assembly Reference Illustration component tray
6-15
Drive plane height up/down and check the
Turn shaft until you have a reading of OQ reading on the digital display (control
panel)
Remount the pot and tighten set screws
Check with a tape measure the plane
Drive the triangle manually down (to height, by inserting the tape through the
170 5 mm) bucky slot onto the triangle
0 Verify that you have a 10KQ f3 percent
reading there is a difference between the digital
display and the measured distance between
0 Pull back the service cable 02A t o foot end the triangle and the marker, make adjust-
side of the table ments to the microprocessor, using the fol-
lowing procedure
0 Reinstall fulcrum assembly by moving in the
tower and pushing down the upper coupling
jaws Reference Illustration 6-13
- Drive fulcrum plane height down to the
0 MM position (using the Down button on
control panel)
0 Verify that when the assembly IS in place, the
fulcrum frame is all the way against the two
tapered end stops, located at the foot end
- Turn on service switch located on upper
interface Reference Illustration 6-18
side of the table base
SERVICE SWITCH
37762 DO1
UPPER COMPONENT TRAY
ILLUSTRATION 6 18
- Press and hold (Down) button while 2-11 Setting Chest Stand Distance
adjusting R29 for 000 mm
1 Center tube stand, and angulate tube 90"
- Release (Down) button and turn off the toward wall bucky location Reference
service switch Illustration 6-19
- Drive fulcrum plane height up t o the 2 Measure the distance between the center of
215mm position (using the Up button on tu be/colIimator (use colIimator tape measure)
the control panel) and the film at the wall bucky (in mm)
- Turn on the service switch
3 Subtract 1046 mm from the measured distance
- Press/hold (Up) button and adjust R30 for 4 Divide the result by 2 04 mm
215mm
- Release (Up) button and turn off the The end result is the number of bits t o be
service switch turned on at the DIP switch (53 on processor
board)
WALL
37765 001
37766 DO1
PROCESSOR BOARD
ILLUSTRATION 6 20
LSB M SB
0 ON
Value
37767 DO1
53 DIP SWITCH
ILLUSTRATION 6 21
TABLE OF CONTENTS
~
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 0 DIRECTION 17073
I
major options Option interconnect maps show the
cables and additional connection changes that are
required for adding the indicated option Tube
interconnect maps show the connections for
various X-Ray tube applications
D6
STATOR
SID POT -
X- RAY TUBE
ANODE
lo7
CATHODE
to7
PRESSURE SWITCH
SENTRY
COLLIUATOR
4P2 4P1
SENTRY
P6a P7a P k ELEC-
TRONICS
ZfWNAL PZQ
846 1p2
PlOa
07 848 P4
J4
844 P9
I I
I
VERTICAL
CASETTE
HOLDER
CONNECTIONS BETWEEN SENTRY
ELECTRONICS AND VR7 TABLE
ARE IN TABLE HARNESS
VERTICAL
BUCKY
-
CHANCEX
T O M 0 KIT J6 36953 D l l
DXD 350 -1
STATOR
ANODE
I
HIGH VOLTAGE
TRANSFORMER
28TP 28TS1
56182
56781
a- VRT TABf Pqa
ISENT RY LEC
BUCKY
Y
X-RAY
SWITCH
36955-010
DXS 650 - T
1L
1
INDEPENDENT
SUPPLY
192 T O 2 4 0 VAC
SINGLE PHASE
I I
{GI SARC
ROTOR
CONTROLLER
-n 3
- STATOR
ANODE
CATHODE
50-A SO-C
HIGH VOLTAGE
TRANSFORMER
YfT TABLE
ION CHAMBER
TABLE-WALL
BUCKY
Y
1- ‘I ,
SINGLE PHASE
LINE SAFETY
DISCONNECT II
,I
X - R A Y HAND
SWITCH 1 36958-010
MSI 850 -1
OR 1250 - 14
ANODE
CATHODE
PRESSURE
SWITCH
I SO-A SO-C
TS3 POWER
UNIT
HIGH VOLTAGE
TRANSFORMER
18TB1
I
THREE
I I
PHASE
SAFETY
DISCON-
NECT
-
VRT TABLE J4
20781
VRT TABK&qa
P
(SENTRY
(SENTRY
VRT TABl#fPl(k
WE?
I
MSI X- RAY OUANTAMAT-E TABLE-WALL
IONCHAMBER
UN17
BUCKY
W ARNlNG
LIGHT X-RAY HAND
76956-010
INDEPENDENT
SUPPLY
192 TO
240 VAC
SINGLE
PHASE
sARR$ToRn
CONTROLLER *-I STATOR
ANODE
A5TS1 A5TS2
CATHODE
@- PRESSURE
SWITCH
-
l8TSl & l8TS2
SO-A SO-C
POWER
UNI 1
HIGH VOLTAGE
TRANSFORMER
18TB1
I
I 1
1 I
I
I I
20TB1
rqg NOTE1
I
I
I I I
WARNING
LIGHT
I
I
I I1
WALL BUCKY
a 1-1 MIS Chart Overview Only those charts which apply to VRT/MSI/MST/DXD
based systems are provided
This section provides detailed information about
each MIS cable in the form of individual MIS cable
charts All of the charts are in numerical order Any MIS
.-.I 109 n
I T S l C C COWCl I O D b L I
on Lono C C Y T C ~
8NC I L f I I C N C 1 aMOUND
sic n r l 1IOLL
scr.: t nmNU4L
CWOC 7 \ "E"01C 2
I
II m, II I'
rn
P ADDITIONAL
CONNECTION PATHS
BLU
0 VhC-(l3f-
1 1 3 vFlc< st-
RED I
BLK
1 f.
192
1 4TP- 1
h
V (et.
A 1 4 - 6 K2
.*-=e WHT I
I
- - - -
r >> 1 ANODE
I I
X-RAY TUBE
CATHODE
I
CABLES 4 6 - 1 2 5 3 8 8
45 OR 60 FEET
M P X M A S T E R C O N T R O L 1/0
I
B U C K Y PWR
2 (GEN 1 1 S V A C )
250 V A
( OVAC 1
X- RAY
COMND
LOGIC
24VDC
-
351 J58
BUCKY PWR
(GEN iisuAcl
250 UA
STABILIZER
M=-t-HF
- -
BE RTN CelJAC)
- -
EXPOSURE 4
I
LOGIC
RELAYS
z
ROlOR X-RAY
HkNO SW - -
FURNISHED CABLE
46-382712G1 6@FT(18.3M)
.
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 0 DIRECTION 17073
QUANTAMATIC I O N CHAMBER
300 VOLTS
I I
NOTE I 11 I12
RETERMINATE I N
F I E L D AS SHOWN. 24 VOLT R T N
I I
61 I10
AREA 3
I I
71 16
SINT
I I
el I 3
AREA 2
I
i
1
I
ION CHAMBER DETECTOR
91 I 5 AND PREAMP
AREA 1
I I
101 I 4
+ 1 5 VOLTS DC
I I
111 I 1
- 15 VOLTS DC
I I
121 12
15 VOLT R T N
NOT USED I I
131 I8
I C OUTPUT
I 1
141 17
SH IELD
MPX MASTER C O N T R O L
I
OIL 5- TERMINAL
PRESSURE MODU S T Y L E P L U G
SWITCH
A ) C O N N E C T E A C H R A D T U B E INTERLOCK T O J 3 9 OF T H E
A P P R O P I A T E 1/01
REF, CH, 8 MULt
TUBE 1 M G 2 A8 J 3 9 t 2 4 71
TUBE 2 A4 J 3 9 [2-c7I
TUBE 3 All J39 [2+7 1
TUBE 4 A12 J 3 9 t 2-F7 I
3, VASCULAR S Y S T E M S
A ) C O N N E C T MODU P L U G AT M G 2 A B J 3 9 F O R F L U O R 0 TUBE,
B ) C O N N E C T MODU P L U G A T M G 2 A 8 J 4 0 F O R AUX, R A D TUBE,
CABLE 46-17559461 1 2 0 F T
B A S I C 110 46-184896
OIL 5- TERMINAL
PRESSURE MODU S T Y L E PLUG
SWITCH
TUBE 1 MG2A8 J 1 5 A
TUBE 2 MG2A8 J 1 5 B
TUBE 3 MG2A8 J 1 5 C
TUBE 4 MG2A8 315D
3, VASCULAR SYSTEMS
A ) CONNECT MODU PLUG AT M G 2 A 8 J 1 5 B FOR FLUOR0 TUBE
5 TERMINAL
OIL MODU PLUG
PRESSURE
(REMOVED FOR
SWITCH INSTALLATION)
DISCONNECT LEAD
BENERATOR COMING F R O M 1 AT SCHM. REF.
I 525
CABLE 6 6 - 1 7 5 5 9 4 6 1 60FT
TO TABLE ION 7
DETECTOR COWARATOR 4
- n
W 7
CKT ION CHAmER
I
-
PRE-AMP BOARD
10 SEE
OUANTA)*(F\T
S M A6400B
TO FILM AREA
3 BRN 6
n
SELECTOR SWITCH W 6
START INTEG
S I GNAL
TO FILM AREA
SELECTOR SWITCH
TO FILM AREA
SELECTOR SWITCH
-<< 2
1
<<
-a
15
GRY
YEL
PUR
3
n
W
5
n
W
4
n
W
3
+15 VDC
4? BLU 2
n
W 2
I +/- 15 VDC
RETURN
BLK 8
n
8
--t
4IEL
-
l0
<
FURNISHED CABLE
46-156314-Gt 50 FT
USEABLE LENGTH 40 FT (12 2 m)
I A90P10 TSl 1
QUANTAMAT ELECTRONICS
I
6
300 VDC
++ ION DETECTOR
-
POWER PRE AMP
SUPPLY
SEE
+24 VDC 10 11
n
RETURN v
TO FILM AREA
SELECTOR SWITCH
START INTEG
SIGNAL
-+
3
+
++
14
15
n
v
n
v
6
3
10
TOFILMAREA
SELECTOR SWITCH
START AREA
SELECTOR W I T C H
-
1
L
n
v
n
W
5
4
9
1
+15 VDC
++ 8
n
v
1
++ 10
r-
2 3
2 8 4
h
RETURN __(+v
SHIELD
1 1 n
10 I 11
I
- -- I
1
-
-
"3
3
- I
11s VAC- - -
18TS1-24'
18TS1-21
I
8LU
9
-
o vac
-t--
RED
18TS3-S ELK
-
18TS3-3
____o___I
18TS3-4 UHT
GND GND
,&ECABLES
?-
STATOR CABLE
4615532Xil
OFT (183m)
A2 BLU
llSVAC w
I
4 RED
0
1
FAN SO.
0 VAC 4 1 C
28K3-0
r------ 1
I
r PHASED 0v 1
1 8
1
I I so.
I
0 I
I
I
ROTOR
CONTROL
CIRCUIT
I
I
STATOR
I I 7 09
I COMMON ,+C-l} w
-I
ELU
RE 0 0 FAN SO
I
66n11
~PHASEDOV n
Y
I I
I ROTOR I SO
STATOR
-- -
181s-24- BLU- BLK
115VAC I
DIFFERENT COLORS
m s i-7
*7
I
__o__
- 18TS12
STATOR
a Y---
181s16'
WHT I
0 BVAC A16
n
W
I I + STABILIZER
0VAC
1 BL 4 J P 18
n I I / / X-RAY
W
\ AB COMND.
K15 HD.SW,
BUCKY
MOT1 ON BL RTN 1s
HD.SW. IIO
C H A S S I S GND
R E M O V E JUMPER B E T W E E N
12TS1-2&3
NOTE1 BUCKY MAY BE I N I T I A T E D BY
COMPLETING A C I R C U I T B E T W E E N
T E R M I N A L S TS3- 586 ( I S O L A T E D )
OR TS3-3&B ( N O N ISOLFITED).
FURN, CABLEi46-223110G1
5 0 FT ( 1 5 . 2 M )
X
@-1,
w
\
-
ROTOR
115VAC
X- RAY
-
A2
115VAC
RELA f ADDED
TC 20 PANEL
20K2
-
@ 0%
W
A2
<
<<Z
L,
A2J4-8
JUMPER
ADDED
OPTO
T
Q/
ISOLATOR
EXPOSURE
INTERLOC K
A5K1
LINE
CONTACTO?
SF
[RANSFER A2J4-5 1
-
1 <&-
OPTO ISOLATOR
15
1P4-15 1P4-5 FLO A 2 J4 -6 41
115VAC
- 1 4 GAUGE
- 18 GAUCE HlFE
X UNUSED TERMINAL WIRE FIELD SUPPLIED
7 23 Copyright@1985 by General Electric Co
GENERAL ELECTRIC M E D I C A L S Y S T E M S VRT TABLE SERVICE MANUAL
~ ~~~
FUNCTION
SELECT
SWITCH
GRlD CO!!
~.u 7
A __
I SWITCH I RELEASE
TOMO
7 07- J X30-3
} START<
115VAC
07-TI X30-1
CLOSED FOR
w e-( M
07-T X22- 4 MAMMO
EL> FILTER IN
PLACE
X UNUSED TERMINAL
m<
V5VAC HIGH 1P6-A lS1
A
< I< IF I I
115VAC LOW
I 1P6-C
I F2
I 1P6-E
61TB3 DXS
JUMPER FOR NON
I MPX GENERATOR
I 2OTBl MSI/MST
I
I
FUSE HI I
r -- 1
I -
I I SEE MIS
841
FOR K10 CONTACTS
I I
T-
2 4 0 VOLTS LOW
I L -- J
CABLE 46-240202G1 30 FT
t
7 25 Copyrtghts 1985 by General Electric Co
Dx S/MSI/M ST/G ENER A T O R WALL BUCKY
~ZOTBI
A1 OVAC 8
\
\
A 2 115VAC 4 \
JUMPER
FUNCTION
SELECT
SWITCH
CLOSED
DURING
MOTION
-
A2J4-13
K209
\ R E L A Y ADDED
7 0 20 PANEL
/
@ 0% <
I WPER
ADDED
w
dzL1 A5K1
m 117VAC
A2 A2
W
X-RAY HAND SWITCH
n
ROTOR 0 A2J4-0 4/
117VAC O P T 0 ISOLATOR
<e A2J4-1
SF
TRANSFER A2J4-5 4/
- <&- I
-7-
OPTO ISOLATOR
A 2 J4 -6
-- 14 GAUGE
-- 18 GAUGE WIRE
X UNUSED TERMINAL
WIRE FIELD SUPPLIED
7-27 Copyright@1985 by General Electric Co
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
A2K21B
MEDIUM FILM AREA 7
JI
TO OUANTAMAT
< <- >-<
9OP17-14
1P20-14 92
>-< <1PB-p
L f ILM
<
AREA
. COMPENSATION
+24VDC FROM
2 A2K19A
W A N T AMAT
< > <- -
’ 9OP17-15 < ><- 1P20-15 91
<L<
Klo
-c
CLOSED FOR
07-T
H <<L
<-< X22-4 MAMMO
EL> FILTER IN
PLACE
X UNUSED TfRMINAL
X-RAY TOMO
X-RAY RAD.
;1 7914TCONMAG
OR
CONSID
BULKY
I ~
N IN-OUT
OVAC
_-_-------- LED IS ON
L,. - - - - J 88 v
BUCKY W I
a
1 'WALL
.. .-
.-BUCKY
rc
SELECT OR
--i
SWITCH I
I
I SEKKTE I
r- CONSOLE I
I
INO 1
L-
0
NO 2 i
--
I
-
4 SPLICE 1 2
WBUCKY
CONTACTS
I
I
BUCKY UP
CLOSED I
OURING
TO SPEED OPERATION 1
9 07-F X29-1
BUCKY 1
<t--(H<
BUCKY
BUCKY K6 UP TO
UP TO
:1 < < ",'",-'T
07- H SPEED
K8
GRID COCK
AND
RELEASE
7 07-J X30-3
ADDED TO TOMO
H H <-
START
ANGLE
' 1
K7
RAD
EXPOSURE
56 PANAL EXPOSURE 6
H H
07-k
REACHED
J
X30- 4 KQ
START
* A2 12 07- M X30-1
I 115VAC
<*H< -- ,
NOTE
NOTE
HEAVY LINES
ARE ADDED WIRES
A2K265
EXPOSURE INHIBIT INTERLOCK EXPOSUFE
X-RAY TU6E ENERGIZED SIDE INHIBIT
PRESSURE SWITCH RELAY
ROTOR X-RAY
42
T
TO OUANTAMAT
C
AREA
IT T-II
COMPENSATION
A 9OP 17-14 <-@-( < ( < 1p8-p FILM
AREA
COMPENSATION
A2K19A
+24VDC FROM
W A N T AMAT
1P8-R
A90P17-15 <-@-< <L
K V ? LOCKOUT
07-P X22-1
@-<
K V P LOCKOUT
< <-
tf--(
07-T
J
X22- 4
CLOSED FOR
MAMMO
K1o
FILTER IN
PLACE
- - SENTRY - - -
115VAC HIGH r ELECTRONICS
@-< < A
<I<
1P6-A 1s1
IF I
I
I IT1
I
115VAC L O W
@-< < C < 1P6-C
I F2
OFF SWITCH
F
FUSE KI I
r -- 1
I I
I I
--
SEE MIS 841 I -- I
FOR K10 CONTACTS
2 4 0 VOLTS LOW
L -- _I
CABLE 46-240202G2 50 FT
Copyrighe 1985 by General Electric Co 7-32
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 0 DIRECTION 17073
COMMON P3
All-13 1J16-17 1P3-A
DlFF COMMON
All -77 2
IP C
B 4 K1-4B
SMD5303
CHAPTER F +6 8VDC CASSETTE SIZE SUPPLY
C 5 K1-6C
E 6
FILM IN PLACE
F 9
0
240 VOLTS
HIGH
-- T+, IK I O 0
I I VRT
I 0 AUTO
TRANSFORMER
I
I
240 VOLTS I I
LOW A l I l K l 0
"1
GROUND
POT
$)s>&& All- 10
TO LOGIC
-$><>A>-+ All- 13
OVDC
24
>- Ali- 21
ov3c
TO LOGIC
9, PNK
SHIELD DO N
CONNE H U
-
IN PLACE^+>
CASSETTE
-L
6
6
W
3,
v
BLU
$>A>+>-+ 1J15-12 A3- 25 -
+15VDC
7
F 25 A4- 23
))--$>+>-+ T 3 LOGIC
N
ROTATE R R 23 A4- 27
S2-1B
IY
AI
1,!
~ 2 - 1/WHT/
~
WHT
>->+;) TO LOGIC
13 BLK
-y:O LOGIC
BUSKY
HO IZ
S5A
LOWER
+lo VOLTS DC
t
UP
ccw
FOCAL SPOT TO
FLOOR DISTANCE
DIFFERENTIAL
COMMON
c 1 1 CURRENT CAR-
I I RYING COMMON
I I
I I
v
SHIELD )>-<Go GROUND STUD
VOLTS
SOURCE TO FLOOR
DISTANCE
x7
40 INCH SID C 8 I
x7
7 2 INCH SID C 3
4
SHIELD J
1P1-9 1P6-K
X32-3 0 7 - Y
+24 VOLTS )- ) )L> )-
< <F
TUBE VERTICAL
TUBE X26-2 07-2 LOCK
VERTICAL
LOCK
> >- >L> >-{ <L*L
452
COLUMN
COLUMN X27-2 07-W ROTATION
> >- >&> >-
<
,
ROTATION < L O L
SWITCH 453
TUBE
ANGULATION
07-3 05-3 07- V
H H b % - +4sL
1
M> >-
< i L< GROUND
- STUD
INCREASE
TRANS CASSETTE
SIZE SIGNAL
A11-25 ) LATERAL
I I SIZE
CURRENT CARRYING
I I
>
I .
COMMON I I
I I
A11-13
[TP-11) >*< L
I I
I 1
DIFFERENTIAL I I
A11-35 COMMON
+
29 I I
LONG CASSETTE
SIZE
A11-39 > >HI I
LONGITUDINAL
SIZE
+6 84DC
CSS POTS
All-11
UP-2) ) A >>------( I I
I I
SHIELD
GROUND 0
STUD
1 1
I I
> 07-5 B
+24V I I
>H <
p
1 1
X32-3
11
I
v:
CASSETTE TRAY
IN PLACE < E
x4- 9 > 07- r
>H
U
H & C
29
10 17
CASSETTE SIZE IN INCHES
+lSVDC/CATH TABLE
TOP INTERLOCK
P20-M
A3- 30 > >-e
TABLE IN VERTICAL AND
CASSETTE IN PLACE
P20-F
A4-38 > > H e
CATH T A a L E TOP DOWN
P2a-N
A2-6 > > H e
TABLE TRENDELENBURG AND
CASSETTE IN PLACE
P2a-P
A4-35 > >H+
+15VDC
a
7-4 1 CopyrightQ 1985 by General Electric Co
768-746z
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
Al- 5
<-
07- X X32-2
i +e+ e CLOSE5
C<L +24V3C Wl-IN
Buck I
K15 K6 UD 7 3
07-U X32-4 T SPEED
A1-61
-<e9
11
07-5 X22- 2
r<e <z<
K16
A1-62
K161 L
e-<
07- V X22- 5
@ BElWEEN tm3DES
AN3 TOMO
LED O N
07- X 31- 1
WHEN
HAND SW IO 07-C X31-2 K7 START
< A1- 35
OE
2 8 ) >RTN2< (28
07-2
<<
X32- 3
( ++ 2 4 V D C
07-Q X32- 6
OPT0
ISOLATOR
RDY 2 TOMG START
A 1- 16 07-M X32- 7
+24VDC OPT 0
~ p 4 - I GEN RDY ISOLATOR
A1- 20 CNTCT 07-2 X32- 4 ROTO2 UP
TO SPEED
CON1 ACTOR
J7-3
TABLE
K11 POWER
24 V RTN ON
A1-60 ACC 2D
BIT COM 07- AA x3-5
19
>-----<e+<-
A1-66 ACC 2D 1
19
A1-69
ACC 2D
BIT 1
< 21
-e+<-< 07- H X3-2
A1-73 ACC 2D
22
JD
AI- 77 ACC 20
23 BIT 3 07-K x3-4
-tis--<<-<<
A1-81 ACC 20
07-N
< 24
We+(-<<
BIT 4 x3- 3
I 1J15-5 I
I
I
t------JI)'"-" "))
I I
I I )N>WHT 0
I I n
v
I I GROUND
4TB1-5
I I
i I X64>->07-D> I
I I I I
I
I > I1
d;>1p7-M4TP1-7
4T6'-6
M I=l >>'Y EL
I 1J16-50 T I
1P7-X X
' 01Pll-2
I - I T O POWER AMP
z
BLADE
DRIVE MOTOR
(LENGTH DIM 1
II
I 1J16-48 I
I~ 7 - e 8
I ->
I BLADE
I 1P11-3 I TO POWER AMP DRIVE MOTOR
RED (WIDTH DIM 1
I - - - - - - - J
I I
*HH*
+loo' +80'
I
I A4R131 J J 4P4- 1 4510 4513 4P4-3
n - - +4 M
-
* ) 10' 10'
I 07-F
I
I
I
I X6-3
I
>- > TUBE TbrrARD HEAD
) I )
1
I- m *
A4R133 D I
> -
A
D
>HHt-++b
4TB2-8 451 I 4516 4P4- 2
I I -80' -100'
I
I +15VDC>->I>-
B I
I
B
> B
REDUCTION
-
7
I I
I
I
I
I
" 4P3-A
I A2R5 Y 4P3-D
I
>-HI- 4515 KEY SWITCH
I C 4P3-c
Q
>*Ht-
I
< A 4PS-2 4DS3
>M> 4'
i I EXP
HOLD
I I LAMP
I A2R75 F 4P8-1
4DS2
I I
> > N ~ A ULAMP
TOMATIC
CABLE
46-178793G5
7 45 Copyright@1985 by General Electric Co
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
- - SENTRY - - -
115VAC HIGHr 1P6-A
ELECTRONICS
1s1
-3 H MI< IF I I
I I
I
1DS1
CB7- 10
I I
I
115VAC LOW I
1P6-C
-+ >-< / c ' I F2 w o
I
I 1P6-E
CB1-6
A -+ H <
240 VOLTS LOW
7 46
SEE MIS 841
FOR K10 CONTACTS
I
CABLE 46-240202G1 30 FT
I
GENERAL ELECTRIC M E D I C A L S Y S T E M S VRT TABLE SERVICE MANUAL
IP83
, TUSEX INHIBIT
K I IO AI-76 EN INTLKS
1-13 <-y-i-l
K20
@JP20BE < ewe-( RELAY
1 A2K 21B
KEDX
J18
-<2> ><- <L AREA
COMPENSATION
A2K19A
AREA SEL COM
(+24VDC) 1P8-R
J1-16<2> < >- <L
, +24VDC 07-P X-22-1
-L
K la
MAMO
CLOSED FOR
INTLK x
A5-4
J 4 - 2 4 H
MAMO INTLK
&-a <A><<L
<-< JP1-A
A5- 6
>-(
07- T X22-4 MAMMO
FILTER IN
PLACE
XA = TUBE 1
8 = TUBE 2
C = TUBE 3
D = TUBE 4
+15
r----
I CHANGE X
1 VOLTS r- SENTRY ELECTRONICS
I
I K 71 11P3-E
I
<-1- I 1J15-12
H A 3 - 2 5
I J6-18
I I I
I I I
1 FILM
IN
I
I
I
I
T
PLACE I I
I FILM
IN
I
I PLACE
I
L
I I
I 1P3-F
<-1- I 1J15-25
H A - 2 3
iTI
MG2A8
J522 + 24!!!.?N++-
a - 2
- - - - L -
YSI/MST-A3
DXD 350-11
59 MPX PMlA3
> 240
CB1-6
I
I
I
- -- -I
// D \ VRT AUTO
\\
\ TRANSFORMER
STUD K10
I I
USI/MST/DXD I
HDEPEhOENT
SUPPLY
YPX-PM1 A3
I I
I I K10
LOW STUD
-+I
STUD 10 AMP CIRCUIT BREAKER
36957-DlO
0 -
SECTION 2 VRT SCHEMATICS
Due to page number references located on all schematics, the following pages will be numbered
accordingly The wiring diagram will start this section on page 5-62
MIS 845
1J P6/n
7
P6aiK Field L a m r bP2 /K
M 07/31 KL213
X2612 -
) Page 75
.40 051 3
C Page69
R
cR
1
.41 X3215 -
)Page 76
H Tubestand Drive HeadEnd H
4
X3418 -
)B Tubestand Drive FootEnd B
4
I
X3417 -
Page76
To lJP4je
t15VDC Torno Interlocks
-42 X 3312 -
Page77
Sentry
IPL
M I S 848
,D Hori z Tom0 On Siqnal
- 43 ’ x33n
page77
-
5-62
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
C C
>-
D 0
)
E E
-1
H H
4- To Chest Unit MIS 840 842 or 853
J J
) - - ,
D D
I-::
- I ,
To o, Differential D, sw
, I
Sentry
Common
' I
MIS 0L3 C, Currentcurrying C. ns I
Common T
lJp9?
+1SVDC 0711 x712 -
I
)Page 0L
)c LOlnch SID ,2 X7/1 -
To Wage a&
Sentry
E 721nch SID
.
,3 X713
'Page 8~
-
MIS 0CL
Shield
KLGI+ KL4/+ 58 I
c T - - K ?'' -
5-63
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
MIS 850
1JP7h P7a Ix
>-
F F
C
I
- C
To
Sentry
Electronic
IP7
V
> I I 'v
H ~i
> I ! >
P P
, >
I. 1. From
K 1 1. K Collimor
>
'VI Drive
Button
-
,
- 5 TubeStand Dri/e
Head End
X612
'Pop 05
5-64
~~~ -
1 ’‘F Exposurehhibflap
Load Side
c Exposurelnhibit
X- Ray
To LorseFilmAreo H
, C ont r o I
Sentrv
Electronics MIS 837
R, MediumFilmArea MPX 852
IP8
,p + 2 & V D C from ,p
Quontarnat
E 07/13 KVP lock out x2211-
X-Roy 1
CI T C
-
1 Lateral
Size
Lon’gitudinol
Size
To
Sentry B CassetteTroy
GE
Electronio
IP2
MIS 8L6
+ 2LV
‘Page71+UV Trt
KL211
’ XL19-
Page 83
E, +15VM: Table Anaulotim Interlocks <A6 -1
x3313-
Page77
TableHorizontal end Css inRoce j5 x331L-
) Page77
,F Tablevertical and
Css mPlace
Z S V O C TableT~~
Interlock
,N TableToDDown ,
N
,P css in Place a n L , P
Table Trendelenourg
5-65
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
07/46 x315 -
J4{19 Acc2D BitCom _____< )Page 01
,20 Acc 2 D B i t 0
0 X311 -
- C 'Page 8~
21 kc 1 X312 -
> 2D B i t 1 )Pap&
-
,22 AG 2D B i t 2 . 10 X316
)Page&
x314 -
,23 A c c 20 Bit3
9
' page81
24 kc2-
0 Bit4 12
' x313 -
Page
29 RDY (Intrt
.
I
11 X 3217 -
)Page 76
30 RW2 rtn
.14 x32/4-
Page 76
MPX
Generator
+21V DC .-15 x2212 -
) Page 73
,I7 Tomo M o d e 2 .
,18
)
X2215 -
Pagen
,3 BE2 P
17 X32I9 -
) Page 76
,13 BE r t n 2 . 20 X3212-
Page 76
. 21 x32/3-
-
,B OE rtn 2 Page 76
>
31 d ~ l o X R ~ , . j
-
, 25 X 3112
) Page77
26 X2913-
2 OVAC \ ) Page77
l1 X - R a v F w l l r e 27 X2914-
\ 1 Page77
-
>
9
BuJty 1 . 28 X29ll
Page 77
M S I lMST
DXSIDXD ,
lo Bucky UP to Sgeed .29 X2912-
Page 77
Generator 30 x3013-
Fv Pagen
12 I 32 x3011-
VACi !L.A
J2 ) Page 77
On
5-66
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
~~~
---
5 <7 or 7, +24V T r l KL211
c CPage71
VCIS3
6 vi0 6, Coupiina BarClamDs Open x3511
'Page76
2 2 ge X26ll
vc IM4 KL2lb
1
- rt
- Page 71
-1
FOlSl .2 End Swirch Down
'X2513
Page 7~
p I:
?Ll/RI Columnposition x214
1: (3
C Return --
'Page 82
'XZI'
Page 8 2
I
SW ,6 Tocho Generator
' Tr4lS
page71
5-67
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
-
BUlMl
w2
O6aA bl
ge
06;: +Tirf
Clamps
xPage
2 6 17L
1
BU It4 3
VI0 9. Plunaer
' x2812
Page E
KUlS
'Page71
' x2713
Page E
5-68
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
E
- VClM6
-6
5 bl
SY
.5
-6
t24V rtn T r l
xnn
'page75
rt -1
I
C o b 0 Degrees XSl2 as
)POW
7-'
br .2 B L V Tr 1 KL2ll
2- *
4
gn ge -12 KL119
+ -12 ,
ge -7
>
SwitchFukumQ&d Bottom XPage
U 1 6 76
\
gr .lo C o l w r a k e X2216
) Page7L
+05d2
-
OSbI205al3 br
.
0513 T , , h o m * OIILO
'Page 6 2
VClMl
6 - 5 sw _5 +2LV rtn T r l KL26
Page 71
VCIM2
>- 2 ge .-2 Pluager CON SID lOOcm x2714
) Poge 75
"di, vcls' :
7
gr '7 t24V Trl
- KL213
Page 71
Tr SI-isv X116
)
[ >
Y119
Page 71
{ qsv:+l!iV
)Y115
I
m
h
Voltage depends on 3
Q,
cn
Pot value set on
Page71 > X2ll2 2
Page 73 x2111
l M )X211L
PlOaIc
C
-
A1
-
A2 071~3 0811
K 1C
2u) VAC Remove Jumper For Emergency
MPX Generator Installation
F6 01154 0012
1A KLGlS Kl, KLLlb
----L - d KR +UV b e 7 0
-226v Un+ZLV!%ge62
22v
2ov
Tr 2
+ KL21L +2&V
'page70
ov
3 2A
- KL217
Page 67-70
II ll ll
3 2A
X21lG Page73
f
2 ,k21/1 Pap73
I I 3 ,x2112 hgen
L ,03110 Page67
5 ,0316 Page67
6 ,X2313 Page73
7 ,X2311 Page73
Tr 5
12A
KI
0
CB
INCOMING
I
KL3 l$AC
LINE
02-N
KLW+ KLU+ 3
+O O C A ]
NOTE Set ' X " t o incoming power line voltage Reference Chapter3, Section 13-1
5-7 1
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
Component Designations
I
~
5 71A
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
K L4 Terminal strip
L1 'Line ON" lamp
VUMl Solenoid tube angulation
VUM2 Solenoid SID plunger
VUM3 Solenoid vertical brake
VUM4 Solenoid coudina bar free
1 VUMS I Solenoid bottom Tom0 coupling
T V U M 6 I Solenoid column rotation
I VUM7
vus 1
1 Solenoid column lock
Switch SID plunger engaged
vus2 Switch 100 cm SID
vus3- Switch coupling bar in place
vus4 Switch couolina bar locked
I
-~
VUSS I Switch column 0 degrees
VUS6 Switch fulcrum coupled bottom
VUSID SID potentiometer
a
5-7 1B
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
tsv
+sv
14 LSl4
5 72
2 E
F
r
m
r
+
N ColumnMotorDrive 3a
2
2 x2?L n
03/1
Paae 67
d
X
ul 6-
0
-?Column or Drive
- 0 , !531*
c 0312
Page 67
X
W x23/3
4-
z
w
J
Page71
x21 I 4
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
a
X
17
I I
P q e 72 Page 72
FIl
Page 72 Page 72
5-74
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL,
----- --- --
h
4
016
L S 2cc
1 Page 72 Page 72 I
Page72
5-75
\ GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV1- __
/-------
- -
.
.
-1-
/
\-
-------- / \
-\
-
-. DIRECTION 17073
I--
Fulcrum 0611C
coJpled up Page68
Fulcrum 7
coupled betta Page69
+2
5-76
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
~
I K5
&I
I
t I
I
1
-
I
1
I
I
Page 76
+2LV
0
I I I
p$
d
I
10 9
A I 1,
6
Page76
- 1
+sv I1I
cL 6 (&
Page72
6
Page72
J
g
Page 72
5-77
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
+ sv
r
M
07
A1
A2
A3
IOIN A4
AS
A6
AI
A0
A9
ai0
+ 5v + 5v A1 1
A12
A13
A14
A15
CPU I
6502
01
02
c3
03
04
05
Ob
07
02
+sv
0
I R /k
+ 5v
r
4
'$ 02
4
5
555
3
I
C4 2
Page 79
5-78
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
PA0
05 PA 1
PA i
PA 3
PA 1
IS
IS1
PA 5
PA 6
PA '
3
1 Inhibit
012
L S 138
m-ts
Page78
tS2
a531
VI A
0522
3
30
31
32
33
Page78 I"
05
D6
07 PB( a
PB * 011
2 PB
7 91w L S 2L5
PB
02
PBi
IRa PB' Page80
I Reset PB'
9
PB
Oir Page 80
11 Direct1on
Power Boord-Selett
1
10211
r t-s Page 80
5-79
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
I I10 Select I
+ 5v
? Cx f
Power Board
Page72
@
Page
a@
70
5-80
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
b
+5v
I t L
00
c1
2764 O2
Page 78 03
E 000
-FFFk
04
0:
D(
0;
OR
27128
coo0
-FtFF
+ sv
+5v +15v -15v
0 T
PI TI TI
I Z Y 8V I Z Y 18V I
5-8 1
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
100v
lOOK
IU M
027 LS 2LL
I
5-82
E8-S
O’lZ 51 BE 0
AS +
I
,
4
WZ
OS(
X X
f f
b
ELOLL N01133tlla
1VflNVH 331AWS 318Vl U A SW31SAS lV31a3W 31113313lM3N39
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
4
Page 80
Page 79
3z-
Page 79
5-84
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
Chest-Stond- Distnnce
""i
Page 80
LS 2u
DL1
LS Z L
1 1
Page 79 I
5-85
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
,-,
F7g 18 3 5 16 7 12 5 t L
-' LS Z L
O 39
I
To Page90
5-86
GENERAL ELECTRIC MEDICAL SYSTEMS VRT fABLE SERVICE MANUAL
REV 1 DIRECTION 17073
Topa e90
@ -1
LS 2LI
Slow
Med
Fast
50
100
20°
LOO
Con Sid
91121lSl 61161 21191 SI looQ
*5v
L
To Page
5-87
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
Page 79
Page 80
?a e 8 0
Display Port
5-88
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
XlB132
x i m
XlBllL
7 1 XlBll6
- I
I2
1
1
X1B 125
To Display Port
11
IT, &"F Page 00
%
+5v
!--4
5+sv
E3+5
+SL
IEl
25
26 R2
X1B /GO
-,x'B134
+ 5v
.T'T* X1BI1
5-89 ,
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
LED Port
Page 07
'age 00
C X*l l 113
> .
I 1 1 T
I 1 I Tons'o. f
13 12
)
To Switch
Decode Port
Page 0 6
I t3
. 5-90
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
~ ~~
TABLE OF CONTENTS
-
CHAPTER 9 RENEWAL PARTS
-
SECTION 1 BOARD LAYOUTSKOMPONENTS
- -
POWER BOARD LAYOUT 660583 46-1 79908P371
I I -
?[
a
9- 1
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
=
0
- I
.
a
N
X
-
-
Cx
mQ
0
9-2
c
e
I
* u
VI
0
-
@
-
X
~~
COMPONENT LAYOUT
A12 12 2K I I I 4 * 2 SIL
A13 147K I I 14+ 1 SIL
A14 150 4 + 1 SIL
A15 2 2K 4 2 SIL
A I7 4 7K 4 + 1 SIL
A18 2 2K 8 + 1 SIL
A19 2 2K 8 + 1 SIL
A20 I 2 2K I I 1 4 + 1 SIL
A2 1 I150 I 4 + 1 SIL
A22 4 7K 4 + 1 SIL
A2 3 2 2K 4 * 2 SIL
A24 4 7K 4 + 1 SIL
Resistor-arrays Processor
I A1 14 7K 15% I Several I 5-SIL 1
1 A2 147K
I
15% I Severat I 5-SIL I
A3 2 2K 5% Several 9-SIL
A4 2 2K 5% Several 9-SIL
IA5 12 2K 15% I Several I 5-SIL I
A6 1OK 5% Several 5-SIL
A7 4 7K 5% Several 5-SIL
IA8 I150 15% I Several I 9-SIL
A9 4 7K 5% Several 9-SIL
J
A10 1OK 5% Several 8-SIL
Nr TY Pe Manufacturer Remark
D26 74LS244 TI 46-2032OOP148
027 74LS244 TI 46-2032OOP148
028 7407 TI 46-2032OOP63
029 74LS373 TI 46-2032OOP147
D30 74LS244 TI 46-2032OOP148
03 1 74LS373 TI 46-2032OOP147
032 74LS373 TI 46-2032OOP147
033 74LS244 TI 46-2032OOP148
D34 74LS373 TI 46-2032OOP147
035 74LS373 TI 46-2032OOPl47
D36 74LS244 TI 46-2032OOP148
037
-
74LS244 TI 46-2032OOP148
D38 74LS240 TI 46-2032OOP194
D39 74LS244 TI 46-2032OOP148
0
040 74LS244 TI 46-2032OOPl48
D41 74LS244 TI 46-2032OOP148
CY 100pF116V 1 Pc Electrolyt
cx 1OOpFl63V 10 Pc Ceramic
c1 1nFl63V Ceramic
c2 1nFl63V Ceramic
c3 1nFl63V Ceramic
c4 1nFl63V Ceramic
P1 2 5K Pi her PTlO
P2 2 5K Pi her PTlO
P3 2 5K Piher PTlO
IP4 I5K I Piher I PT10
IP5 125K I Pther I PTlO
Rl-R3 2 2K 5%
R4 150 5%
R7 2 2K 5%
I Solenoid 1OU H
I
1878108-T1103-K
1
1 1
:able ~
X1BIC
3417-6600
XlNC AMP
3417-6640 3M
X2NC AMP
342 1-6600 3M
X2BIC AMP
1 3421-6600 3M
1 FIat-cab1e 40-DIL 1 6 m Length
1 Falt-cable 20-DIL 5 m Lenath
1N4004
Diodes Pr
IlN4148 146-136314P1
Connectors Proce
AMP Flat-cable 40
3595-6002 3M
I
IL x1
3592-6002
350828- 1
AMP
3M
AMP Mate-N-Lock
I 9 POL
2o
Zonnectors Power I
Switches J
51 7101 C + K Reset-button
52 8631 C + K Service-switch
53 10-DIPSwitch Chest-stand
U42 4N37
u43 4N37
u44 4N37
u45 4N37
U46 4N37
u47 4N37
U48 I4N37 I I
I4N37
I
I
I
I
I
U50 I4N37 I I
US1 I4N37 I I
U52 I4N37 I I
u 53
~~
I4N37
I
I
I
I
I
u54 I4N37 I
u55 I4N37 I I
-
U56 I4N37 I I
u 57 I4N37
1 -
I
I
I
I
'US8 I ON37 I I
I4N37 I I '
U60 4N37
U61 4N37
U- 62
-- I
4N37
- I I
37854 Doi
13-
11,12-
’9
37699 Dol
9-10
~~~~ ~~ -
m
18
30 31
29 -27
28
37857 Do1
9-1 1
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
-43
-62
41
-58
59
60
\ 37856 DO1
65,66,67,68,69,70,71
37857 001
9-12
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
49
46
132 83 84 82 64 81 85 87 88 89 86 93 80
9-13
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
\ 117 IT7
9-14
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
198’
200/ .I75
201/
176
a 37859 DO1
9-15
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
204
\ 219
37861 DO1
211’
222 223
210’
261 ’
9-16
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
/ \
37863 DO1
e 235 236
238 25
233
24L
252
257
24 3
251
9-17
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
264
242
i -I
37865 DO1
239
2.41 w3
e
-268
,277
.270
9-18
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
297
,291
312
-290
-293
-295
-299
-300
/ \
303 304305
298
-322
-318
316
319
-317
320
321
37866-DO 1
9-19
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
‘325
37868 DO1 e
,327
,326
321
37869 DO1
9-20
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
351 348 352 353 354.365 349 347348 364 361 362
3w.
377-
38 3 384. -391392
393
382 -390
- 389
375
376
\I
378
37870 DO1
9-21
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
411
- 478
41 2
488
,495 496
497L98
427,
429 1
,426
424
,422
428 421
' 418
.419
37072 DO1
9-22
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
423
444
'442
37875 DOI
,438
436
440
494
e
9-23
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
459
,456
452 - 469
458 - 457
454 -482
453
461 ,462 463,
4a
466
467
I
37874 DO1
9-24
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
-
SECTION 3 -TABLE BASE PART NAMESlORDERlNG NUMBERS
Failed parts may be replaced only with original parts as listed below Always order by the GE part number If
no GE number i s listed order Hans Pausch number with prefix of HN' All parts t o be ordered from GE
32 Not used
33 Not used
34 Not used
35 Not used
36 Not used
37 Not used
38 Not used
39 Not used
Name
47 1 [ Notused 1
473 Not used
474 Boom compl 660 721 1
37
35
19
38
40
10
44.
~ 3 0
21 -
H43
22-
-31 32
3334
1 8 d 36
‘39
41 42
9-39
~~ ~~
-72
- 73
-69
-70 71
’56
37878 001
. 55
5L-
53 -
’15
52
37879 001
9-40
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
74.
-68
r,
67-
i
L
66-
24
9-41
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
~~ ~
89 86 77 9293 108
I I I I
110
76
80 84
81
94
8283
84
85
a
62 l& 99 103 102
79
78
a
9-42
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
56
57
59
60
58
.
133
132
9-43
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
.126
121
121
122 123
122 123
9-44
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
a 150-
152-
153- 151
183 185- 178179
180
140-
187’ 15L
18L/ l70
167-
.168 186
-155
1‘64
9-45
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
177
176
175
191
192
196
9-46
1
206
22i
9-47
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
197 198
23’ 5
233
9-48
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
252/
250/
249
253 /
308
314
257
9-49
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
260
258
,270
313
287,
-303
-2%
- 289
-290
311
-288
301
-
312
29 1
300
298.
299
9-50
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
283
9-51
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
318
319 320-
-310
4 60
9-52
~~
-328
-333
33 6
-335
,342
340
3 41
9-53
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
L59
9-54
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
409
-391
*4\
-390
-382
410-
-385
LOO -
419 - -399
, 406- -386
LOB’ -398
407’
‘413
41’7 ‘414 LIS
41 6
9-55
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
432
L 29
L28
9-56
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
4 41
L45
-456
457
9-57
GENERAL ELECTRIC MEDICAL SYSTEMS VRT TABLE SERVICE MANUAL
REV 1 DIRECTION 17073
VIEW A - B VIEW C- D
PI I
LL8
L Li7 Li9 L 53 i6
LLL6 L 50 L\51
L 51 L'5 2
L52
I
*T +==------
"f
-qB --- D
VIEW E - F
LSL
./-
9-58
GENERAL ELECTRIC MEDICAL SYSTEMS VRT?ABLE SERVICE MANUAL
REV 1 DIRECTION 17073
Failed parts may be replaced only with original parts as listed below Always order by the GE part number If
no GE number i s listed, order Hans Pausch number with prefix of HN' All parts t o be ordered from GE
-
CHAPTER 10 PERIODIC MAINTENANCE
Surface conditions
0 Lateral play
0 Condition and adjustment of plastic rollers
0 Bumper condition
0 Guide bearing condition
Bearing tracks and rails (clean and wipe with light machine oil)
~~
COMPONENT
VERIFY THAT LISTED COMPONENTS ARE IN WORKING ORDER
DESCRIPTION
I 12 Coupling bar
band con-
0
0
0
General condition
Latch condition
Verify operation
dition
14 Lateral cassette holder 0 Verify operation
condition
0 Verifv oDeration