You are on page 1of 27

REVIEW

Advanced ICD Troubleshooting: Part II


CHARLES D. SWERDLOW* and PAUL A. FRIEDMAN†
From the *Cedars-Sinai Medical Center, Los Angeles, California, and †Mayo Clinic, Rochester, Minnesota, USA

Failure to Deliver Therapy or Delay of Therapy in whom slow VT may be catastrophic2 (see figure
Absent or delayed therapy may be caused 8 in Part I of this review). The VT detection interval
by programmed values (including human error), should be increased if antiarrhythmic drug ther-

m
ICD system performance, or a combination of the apy is initiated, particularly with amiodarone or a
two. The most common causes are ICD inactiva- sodium-channel blocking (Type 1) drug.6,7 It may
tion, VT slower than the programmed detection in- be prudent to measure the cycle length of induced
terval, SVT-VT discriminators, undersensing, and VT at electrophysiological testing after initiation

co
intradevice software interactions. Pacemaker-ICD of drug therapy.6 However, spontaneous VT often
interactions, once important, are now rare due to is slower than induced VT.8
incorporation of dual- and triple-chamber pace-
makers in ICDs. SVT-VT Discriminators
SVT-VT discriminators may prevent or delay

d.
ICD Inactivation therapy if they misclassify VT or VF as SVT.9–12
If detection is programmed OFF for surgery Discriminators that reevaluate the rhythm diagno-
using electrocautery, reprogramming must be per- sis during an ongoing tachycardia, such as stability
and most dual-chamber algorithms, reduce the risk
formed at the end of the procedure, a fact easily for-
gotten, especially in outpatient surgery. One study
reported an unexplained 11% annual incidence
of transient suspension of detection.1 Medtronic
ic
of underdetection of VT compared with discrimi-
nators that withhold therapy if the rhythm is not
classified correctly by the initial evaluation, such
er
ICDs sound an audible Patient AlertTM 6 hours af- as onset or chamber of origin. The minimum cycle
ter the ICD is programmed “off” to alert the patient length for SVT-VT discrimination should be set to
to the possibility of inadvertent.2 prevent clinically significant delay in detection of
hemodynamically unstable VT.
ac

VT Slower Than the Programmed Detection


Interval Single-Chamber Discriminators
In most ICD patients, VT with cycle lengths Although spontaneous VT often begins with
>400–450 ms are tolerated well, while repeated irregular R-R intervals, stability algorithms clas-
inappropriate therapies are not. However, slow VT sify it as VT as soon as R-R intervals regularize,
.p

may be catastrophic in patients with severe LV dys- even transiently. Thus, they rarely prevent detec-
function or ischemia.2 All SVT-VT discrimination tion of VT (∼0.5%) if they are programmed to nom-
algorithms (except ELAs)3–5 deliver less inappro- inal settings and the patient is not receiving an
priate therapies if the VT detection interval is pro- antiarrhythmic drug that decreases cycle length
w

grammed to a shorter cycle length so that fewer regularity in VT9,13 (see “Antiarrhythmic Drugs”).
SVTs are evaluated. To prevent underdetection Previously, we noted that single-chamber stability
of irregular VT, the VT detection interval should may be programmed in Medtronic dual-chamber
w

be set at least 40–50 ms longer than the slow- ICDs to reject atrial fibrillation during redetection.
est predicted VT for consecutive-interval count- In this case, stability is applied before the dual-
ing and 30–40 ms longer for X-of-Y or interval + chamber algorithm. Because stability responds to
w

interval-average counting.5 It should be a longer irregular rhythms by resetting the VT counter to


cycle length if rapidly conducted SVT is unlikely zero, the dual-chamber algorithm does not eval-
or SVT-VT discrimination is reliable at long cycle uate rhythms rejected by stability. Thus, stability
lengths (ELA ICDs). A long VT detection interval is may delay detection of VT even if the ventricular
important in patients with advanced heart failure, rate is greater than the atrial rate.
Morphology discriminators also reevaluate
rhythm diagnosis during ongoing tachycardias.
Address for reprints: Charles D. Swerdlow, M.D., 8635 W. Third But if they misclassify monomorphic VT initially,
Street, Ste 1190 W, Los Angeles, CA 90048. Fax: (310) 659-8387; the error usually persists and prevents detection
e-mail: swerdlow@ucla.edu for the duration of the tachycardia. The St. Jude
Received February 15, 2005; revised June 1, 2005; accepted morphology algorithm, which analyzes only the
June 6, 2005. rate-sensing electrode, continuously misclassifies


C 2006, The Authors. Journal compilation 
C 2006, Blackwell Publishing, Inc.

70 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

5–10% of monomorphic VTs as SVT. But, when rithms should be increased from the nominal value
it is restricted to tachycardias with ventricular of 30 seconds to reduce inappropriate therapies. In
rate ≤ atrial rate in dual-chamber ICDs, only 2% most patients, a duration of 2–5 minutes is reason-
of VTs are misclassified.11 The Medtronic mor- able.
phology algorithm, which usually analyzes high-
voltage electrograms, misclassifies 1–2% of VTs as Undersensing
SVTs.14 In contrast, the onset discriminator uni- VF may be undersensed due to combinations
formly misclassifies VT if it either accelerates grad- of programming (sensitivity, rate, or duration),
ually across the sinus-VT zone boundary or occurs low-amplitude electrograms, rapidly varying elec-
during SVT with cycle length in the VT zone. trogram amplitude, drug effects, or post-shock tis-
sue changes. Clinically significant undersensing of

m
Dual-Chamber Discriminators VF is rare in modern ICD systems if the baseline R-
If the atrial lead dislodges to the ventricle, the wave amplitude is ≥5–7 mV.16 Post-shock under-
atrial channel records a ventricular electrogram. sensing was an important clinical problem in older
Any VT will appear to the ICD as a tachycardia ICD systems using integrated-bipolar leads with

co
with 1:1 AV relationship and nearly simultaneous closely spaced electrodes.17 Currently, the most
atrial and ventricular activation. Some discrimina- common causes of VF undersensing are drug or hy-
tors designed to withhold therapy from SVT with perkalemic effects that slow VF into the VT zone,
1:1 AV conduction may then withhold appropri- ischemia, and rapidly varying electrogram ampli-
ate VT therapy. These include the Medtronic 1:1 tude.4,18 ICDs that adjust dynamic range based on

d.
SVT rule and the St. Jude Rate Branch Algorithm the amplitude of the sensed R-wave (Guidant) may
without additional discriminators, which should be most vulnerable to the latter, extremely rare
not be programmed until the atrial lead is stable. problem.18 Prolonged ischemia from sustained VT
VT with 1:1 VA conduction is a difficult diag- slower than the VT detection interval may cause
nosis for dual-chamber algorithms. Guidant’s older
Atrial ViewTM misclassifies it as SVT unless the on-
set of VT is abrupt. Medtronic’s PR LogicTM mis-
ic deterioration of signal quality, resulting in under-
sensing of VF. Lead, connector, or generator prob-
lems may also present as undersensing.
er
classifies it as SVT in the rare cases in which the
VA interval is both long and constant. Guidant’s Pacemaker-ICD Interactions
Rhythm IDTM and St. Jude’s BranchTM misclassify Interactions between ICDs and separate pace-
it if a morphology match occurs.11,15 makers have become rare since ICDs incorpo-
ac

rated dual-chamber bradycardia pacing in the late


Duration-Based “Safety-Net” Features to Override 1990s. Nevertheless, a few combined systems have
Discriminators not been revised due to vascular access problems
These programmable features deliver therapy or other reasons. The multiple potential interac-
if an arrhythmia satisfies the ventricular rate cri- tions have been reviewed and testing protocols to
.p

terion for a sufficiently long duration even if detect them have been developed.19–21 The princi-
discriminators indicate SVT (Guidant Sustained- pal interaction that may delay or prevent ICD ther-
Duration OverrideTM (SRD), Medtronic High Rate apy is oversensing of high-amplitude pacemaker
TimeoutTM , and St. Jude Maximum Time to stimulus artifacts. If this occurs during VF, repet-
w

DiagnosisTM (MTD)). The premise is that VT will itive automatic adjustment of sensing threshold
continue to satisfy the rate criterion for the pro- and/or gain may prevent detection of VF.
grammed duration while the ventricular rate dur-
w

ing transient sinus tachycardia or atrial fibrillation Intradevice Interactions


will decrease below the VT rate boundary. The lim- Today, “intradevice interactions”—in which
itation is delivery of inappropriate therapy when bradycardia pacing features of dual-chamber ICDs
w

SVT exceeds the programmed duration, which oc- interact with and impair detection of VT or VF—
curs in ∼10% of SVTs at 1 minute and 3% of SVTs pose a greater challenge than pacemaker-ICD inter-
at 3 minutes.12,14 The decision to use a discrimi- actions.22 During high-rate, atrial or dual-chamber
nator override should be based on clinical factors pacing, sensing may be restricted to short periods
including the probability that discriminators will of the cardiac cycle because of the combined ef-
prevent detection of VT, the likely consequences of fects of ventricular blanking after ventricular pac-
failure to detect VT, and the likelihood of sustained ing and cross-chamber ventricular blanking after
SVT in the VT rate zone. For example, override fea- atrial pacing, which is needed to avoid cross talk. If
tures should be considered whenever a morphol- a sufficient fraction of the cardiac cycle is blanked,
ogy algorithm is programmed without inducing VT systematic undersensing of VT or VF may occur.
at electrophysiologic study. The programmed du- When pacing and blanking events occur at in-
ration in Guidant Atrial ViewTM and St. Jude algo- tervals that are multiples of a VT cycle length,

PACE, Vol. 29 January 2006 71


SWERDLOW AND FRIEDMAN

m
co
d.
Figure 1. Failure to detect VT due to an intradevice interaction. The rate-smoothing algorithm in-
troduced atrial and ventricular pacing complexes with associated blanking periods that prevented
detection of VT during post-implant testing. An external rescue shock was required. Shown from
top to bottom are surface ECG, atrial electrogram, ventricular electrogram, and event markers.
ic
At top VT is induced by programmed electrical stimulation with drive cycle length 350 ms and
premature stimuli at 270, 250, and 230 ms (intervals labeled next to event markers). The first
sensed ventricular event occurs 448 ms after the pacing drive (“PVC 448”). The rate smoothing
er
algorithm drives pacing to prevent a pause after the “premature ventricular complex” (PVC),
labeled AP↓1638. A ventricular-paced event does not follow the first AP↓ because a ventricular
event is sensed (VT 415). Subsequent rate smoothing generated atrial and ventricular pacing
pulses (indicated by AP↓ and VP↓ markers, respectively). The resultant post-pacing blanking
ac

periods are shown in the figure as horizontal bars. PAB denotes cross-chamber (post-atrial-pace)
ventricular blanking period. VBP denotes same-chamber (post-ventricular-pace) blanking period.
Together, they prevent approximately 4 of every 6 VT complexes from being sensed. Since the VT
counter must accumulate 8 out of 10 consecutive complexes in the VT zone for detection of VT
to occur, VT is not detected.
.p

ventricular complexes are repeatedly under- VT/VF are complex and difficult to predict, but
sensed, delaying or preventing detection (see they usually elicit a programmer warning. Gen-
Fig. 1).22–24 erally, aggressive rate-smoothing (a small allow-
w

Although uncommon, intradevice interac- able percentage change in R-R intervals), a high
tions have been reported most frequently with the upper pacing rate, a long and fixed AV interval
use of the Rate SmoothingTM algorithm in Guidant favor undersensing and should be avoided. If rate-
w

ICDs.22–24 This algorithm is intended to prevent smoothing is required, parameter settings that re-
VT/VF initiated by sudden changes in ventricular sult in warnings should be avoided by program-
rate.25 It prevents sudden changes in ventricular ming a dynamic AV delay, limiting the upper rate
w

rate by pacing both the atrium and ventricle at in- to less than 125 beats/min, and shortening ventric-
tervals based on the preceding (baseline) R-R inter- ular blanking after atrial-paced events to maximize
val. As an unintended consequence, it may prevent the sensing window for VT/VF. This programming
sensing of VT/VF in some patients because it intro- reduces, but does not eliminate, the risk of under-
duces repetitive post-pace blanking periods. The sensing.22–24,26
algorithm applies rate-smoothing to baseline inter-
vals independent of their cycle length, including Unsuccessful Therapy
intervals in the VT or VF zones. Intradevice inter- Because defibrillation success is probabilistic,
actions that result in delayed or absent detection occasional shocks fail, but failure of more than two
of VT/VF are most common and most danger- maximum-output shocks is rare if the safety mar-
ous when VT is fast. The parameter interrelation- gin is adequate.27 If an ICD classifies a shock as un-
ships that result in delayed or absent detection of successful, stored electrograms must be reviewed

72 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

to determine both if the shock was delivered for A few patient-related causes that would oth-
true VT/VF and if the shock actually failed to ter- erwise require operative revision may be resolved
minate VT/VF. Shocks from chronic ICD systems by programming shock pathway and waveform
that defibrillated reliably at implant may fail to ter- parameters. For ICDs with fixed-tilt waveforms,
minate true VT or VF because of patient-related or waveform duration depends on output capaci-
ICD system-related reasons. The problem of high- tance and pathway resistance.28 ICDs with pro-
implant DFTs has been reviewed.3,4 In chronically grammable waveform duration or tilt (St. Jude)
implanted systems, most patient-related causes of permit optimization of waveform parameters in-
unsuccessful shocks can be reversed, but most dependent of pathway resistance. Shortening the
system-related causes require operative interven- entire waveform might reduce DFT.29 Shortening
tion. phase-two might reduce DFT after amiodarone

m
therapy is started.30 Migration of an active-can
Misclassified Therapy pulse generator low on the chest wall can increase
ICDs misclassify effective therapy as ineffec- DFT by altering the shock vector. This may be re-
tive if VT/VF recurs before the ICD determines the solved by excluding the pulse generator from the

co
VT/VF episode as terminated and reclassifies the shock circuit, which may be done noninvasively
post-therapy rhythm as sinus. Decreasing the du- in Medtronic ICDs.
ration for redetection of sinus rhythm (St. Jude)
might correct this classification error. However, ICD System-Related Reasons
this type of misclassification usually does not con- ICD-related causes include insufficient pro-

d.
stitute a clinical problem, while post-shock detec- grammed shock strength, battery depletion, gen-
tion of nonsustained VT does. An exception oc- erator component failure, lead failure, device-
curs when one iteration of antitachycardia pacing lead connection failures, and lead dislodgment.
is programmed for the first therapy and a shock for
the second. If antitachycardia pacing terminates ic
VT, but it recurs before the rhythm is classified as
sinus, the recurrent VT will receive a shock in-
In evaluating data from the ICD interrogation,
attention should be paid to arrhythmia dura-
tion, electrograms from the shocking electrodes,
charge time, battery voltage, the relationship be-
er
stead of potentially effective antitachycardia pac- tween programmed and delivered shock strength,
ing. Programming multiple iterations of antitachy- impedance of the high-voltage lead at the time of
cardia pacing can mitigate this problem. However, shock, and trend plots of lead impedance. Pro-
the “Smart Mode” in Medtronic ICDs (nominally longed episodes may increase the shock strength
ac

“on”) disables antitachycardia pacing if it is classi- required to convert VT or VF. They may be
fied as unsuccessful on four consecutive episodes. caused by delayed detection and/or prolonged
Thus, if successful antitachycardia pacing is mis- charge times. Programmer software identifies some
classified repeatedly, Smart Mode will disable it. system-related problems and displays messages to
In this setting, Smart Mode should be programmed alert the clinician.
.p

off.
ICDs may also misclassify shocks as ineffec- Problems Identified at Routine Follow-Up
tive if the post-shock rhythm is SVT in the VT Battery Depletion
rate zone (catecholamine-induced sinus tachycar-
w

dia or shock-induced atrial fibrillation). Solutions Battery Discharge Curves—Relationship to


include applying SVT-VT discriminators to rede- Elective-Replacement and End-of-Service Indicators
tection of VT, increasing shock strength to prevent ICDs use batteries composed of lithium an-
w

shock-induced atrial fibrillation, or administering odes and silver vanadium oxide (Ag 2 V 4 O 11 ) cath-
β-blockers to slow post-shock SVT. odes31 in which reduction at the cathode occurs in
multiple steps. The corresponding discharge char-
Patient-Related Factors
w

acteristics and battery behavior during capacitor


Factors that raise DFTs reversibly at the cel- charging have been a source of confusion among
lular level include hyperkalemia, antiarrhythmic clinicians familiar with the slow, constant voltage
drugs, and ischemia. Pleural or pericardial effu- decline typical of lithium-iodine pacemaker bat-
sions raise the DFT by forming parallel intratho- teries. Initially, the cathodal contribution to bat-
racic current paths that shunt current away from tery voltage is due to two simultaneous reactions
the heart. Progressive cardiac enlargement or new corresponding to reduction of V+5 (V+5 + e− →
myocardial infarction may also raise the DFT and V+4 ) and Ag+1 (Ag+1 + e− → Ag+0 ). Unloaded cell
are less easily reversed. Usually, defibrillation test- voltage falls from about 3.2 to 3.1 V at about 30%
ing should be performed to confirm reliable de- battery depletion. In most ICD batteries, voltage
fibrillation after the suspected cause has been re- then falls rapidly to about 2.6 V, where it reaches a
versed. plateau until the battery is 80–90% depleted. This

PACE, Vol. 29 January 2006 73


SWERDLOW AND FRIEDMAN

voltage decrease is caused by a combination of fac- lodgment to the ventricle might also result in inap-
tors, including build-up of inert material at the propriate shocks caused by ventricular oversens-
cathode and depletion of the V+5 valence state of ing of noise generated by contact between the leads
vanadium so that the cathodal half-cell voltage re- (Fig. 3).
flects reduction of V+4 (V+4 + e− → V+3 ). The 2.6-V
plateau is nominal performance for most (but not Diagnosis of Lead Failure
all) ICD batteries. End-of-service indicators corre- A systematic approach to the patient with sus-
spond to either an unequivocal reduction in un- pected lead failure requires evaluation of elec-
loaded voltage about 0.1 V below this plateau or trograms from all electrodes on the lead, pacing
an excessive increase in charge time. threshold, pacing impedance, and painless high-
voltage electrode impedance; system radiography;

m
Premature Battery Depletion stored episode electrograms, data logs, and flash-
The most common cause is excessive exter- back memory (extended recording of R-R intervals
nal power drain due to pacing or capacitor charg- in Medtronic ICDs).36 The yield of these tests is
ing. Pacing problems include unnecessary ventric- shown in Table I. The Sensing Integrity CounterTM

co
ular pacing, high pacing outputs, and lead insula- (Medtronic) displays the number of nonphysio-
tion failures. The most common cause of asymp- logic short intervals detected in the 120–140-ms
tomatic premature battery depletion related to ca- range. With intact, true-bipolar leads, the count
pacitor charging is repeated aborted shocks due to remains at zero. With intact, integrated-bipolar
repetitive nonsustained VT or oversensing due to leads, the large proximal sensing electrode (the

d.
lead-connector problems. Repeated shocks due to distal ICD coil) permits occasional oversensing of
VT storm might also rapidly deplete the battery. diaphragmatic myopotentials, but a count >300
Some diagnostic features have high power con- over 3 months suggests a lead problem.39
sumption. In Medtronic ICDs, “prestorage” of elec-
trograms prior to each VT/VF episode results in
continuous amplification of the unfiltered electro-
gram, substantially reducing longevity. In contrast,
ic
Pacing Impedance
The normal value for pacing impedance is a
function of lead design; but for any lead (except
er
prestorage in Guidant and St. Jude ICDs, which Y-adapted lead combinations), impedance <200
store only filtered electrograms, does not cause sig-  indicates an insulation defect and impedance
nificant battery depletion. ICD component failure >2,000  suggests a conductor failure if a loose
is a rare cause of battery depletion. set screw or faulty adapter are excluded.40,41 Mod-
ac

ern ICDs perform serial impedance measurements


Lead Problems automatically. The range of variability of pacing
Ventricular lead problems are the most com- lead impedances has been studied for coaxial pace-
mon cause of ICD system malfunction.32–35 They maker leads,42 but corresponding data for coaxial
usually present either as oversensing, resulting and multi-lumen defibrillation electrodes are lim-
.p

in inappropriate shocks, or as abnormal diagnos- ited. Generally, conductor failures do not present
tic measurements at routine follow-up. Presen- with moderate changes in impedance. Either the
tation as undersensing during VF or unsuccess- impedance exceeds 2,000  (continuously or in-
ful therapy is uncommon but often fatal. Clin- termittently), or it is normal, and conductor failure
w

ical and lead-design factors identify patients at is diagnosed by identification of a characteristic


high risk for ventricular lead problems and merit pattern of oversensing. Pacing lead impedance is
intensified follow-up. Noninvasive assessment is insensitive to inner insulation failure of Medtronic
w

usually diagnostic.36 The introduction of remote 6936 leads.41 Isolated pacing insulation failure of
ICD monitoring,37,38 which permits transmission multi-lumen leads is sufficiently rare that the sen-
of complete ICD interrogation via telephone or the sitivity of pacing lead impedance is unknown.
w

internet, may enhance noninvasive identification Inner insulation failure in Medtronic 6936 or
of subclinical lead failure (Fig. 2). Patients with 6966 leads may be detected by a fall in ring-coil
Medtronic ICDs may present with audible “Patient impedance with normal pacing impedance.41 Re-
Alert” warnings (see below). view of ring-coil impedance may require sending
In ventricular ICDs, atrial lead malfunctions the programmer’s “save-to-disk” file to the manu-
may present as failure to discriminate VT from facturer for analysis by a specialized software.
SVT. Atrial lead dislodgment to the ventricle re-
sults in sensing of ventricular electrograms on the High-Voltage Impedance
atrial channel, causing detection of one “atrial” Shocking-pathway impedance normally is
event for each ventricular event. This may cause 25–75  in transvenous systems. Values out-
dual-chamber algorithms misclassify of VT as SVT side this range suggest a conductor defect (high
and withhold appropriate therapy. Atrial lead dis- impedance) or insulation failure/short circuit

74 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

m
co
d.
ic
er
Figure 2. Asymptomatic malfunction in coaxial defibrillation lead (Medtronic model 6936) detected by internet-based,
remote patient monitoring (Medtronic CareLinkTM ). The Sensing Integrity Counter provides a cumulative count of
ac

nonphysiologic short intervals caused by oversensing. Multiple episodes of nonsustained VT with cycle lengths 140–
210 ms also suggest oversensing. In the absence of exposure to electromagnetic interference, this is highly suggestive of
an inner-insulation failure. The pacing and defibrillation lead impedances are insensitive to early insulation failure.
Analysis of ring-coil impedance from the programmer “save-to-disk” file is often diagnostic. See text. Courtesy of
Physician Assistant Sheila Reynolds and Dr. Dwight Reynolds.
.p

(low impedance). Periodic assessment of the a high-energy shock. If a proximal coil is present,
impedance of the high-voltage electrodes using its integrity is not assessed. Medtronic MarquisTM
w

weak test pulses may detect a loss of lead integrity ICDs and newer models report independent prox-
before inappropriate or ineffective therapies occur. imal and distal coil impedances that closely ap-
In Medtronic and Guidant ICDs, painless pulses proximate those of high-voltage shocks. When a
are delivered automatically at periodic intervals lead defect is suspected, a full-output shock may
w

to create trend plots. In St. Jude ICDs, high-voltage be necessary to evaluate lead integrity. Partial insu-
lead impedance is assessed by delivering a 12-V lation defects may not be identified by low-energy
test pulse using the programmer command. Pa- pulses that deliver insufficient current to activate
w

tients feel this pulse and may perceive it to be un- the shorted high-output protection feature (see
comfortable. below).
High-voltage impedance measured by weak
pulses correlates well with values measured Electrogram Amplitude (R-Wave, P-Wave) and Pacing
by high-energy shocks.43,44 In Guidant ICDs, Threshold
impedance values recorded painlessly are compa- The amplitude of the R-wave during sinus
rable to those recorded during high-voltage shock. rhythm should be at least 5–7 mV for assurance
In older Medtronic ICDs, the pulse is delivered be- that the low-amplitude electrograms during VF
tween the lead tip and the distal coil, resulting will be sensed reliably.16 Causes of reduced elec-
in nominal low-voltage impedance values of 11– trogram amplitude include antiarrhythmic drugs,
20 , significantly lower than the impedance of myocardial infarction, shocks, fibrosis at the lead

PACE, Vol. 29 January 2006 75


SWERDLOW AND FRIEDMAN

m
co
d.
ic
Figure 3. Atrial lead dislodgment causing inappropriate shock. Top right box shows the episode summary, indicating
an episode detected as VF (via the combined count), for which the first VF therapy was delivered. The interval
er
summary plot (middle panel) demonstrates a regular “atrial” rate (boxes) of 640 ms, with a highly variable ventricular
cycle length (black circles), leading to detection and delivery of a 34.6-J shock. The bottom panel shows the atrial
electrogram (Atip to Aring), near-field ventricular electrogram (Vtip to Vring), calculated A-A intervals, event markers,
and calculated V-V intervals. Each ventricular electrogram is associated with an atrial-channel electrogram that has
ac

large amplitude and slope. The sensing of ventricular events on the atrial channel in the absence of sensed atrial events
suggests atrial lead dislodgment to the ventricle. The presence of smaller deflections on the ventricular electrogram
(the first is sensed as “VS 550”) is consistent with oversensing of “noise” signals caused by contract with the atrial
lead. This oversensing results in detection of VF and delivery of shocks during normal rhythm. AR = atrial refractory
sense; VS = ventricular sense; TS = tachycardia sense (in VT zone); VF = fib sense (in VF zone).
.p

tip, new conduction defects, and lead dislodg- resynchronization ICDs dislodge proximally into
ment. These same factors may also elevate the the coronary sinus or right atrium.
w

pacing threshold. The introduction of automatic Subclavian crush occurs when a lead is com-
capture assessment45 in conjunction with remote pressed in the narrowly confined space between
monitoring systems that permit review of device the first rib and clavicle;47,48 this common site of
measurements and trend plots between clinic vis- fracture requires careful examination if lead failure
w

its37,38 may further facilitate early diagnosis of lead is suspected (Fig. 4). Radiography may also detect
problems. conductor defects, inadequate pin insertion into
the header, abandoned or incompletely removed
w

leads, or torsion due to Twiddler’s syndrome. How-


Radiography ever, some intact electrical connections in leads
and adapters appear radiographically to have con-
Abnormal radiographic features are identified ductor discontinuity (“pseudo fracture,” Fig. 5).
in less than half of patients with ICD lead malfunc- Fluoroscopy of an unopened lead or adapter (if
tions.36,46 Lead dislodgment is diagnosed when the available) can be used to determine the expected
lead tip is in a different position than in the post- radiographic appearance.
operative radiograph, but variability due to car-
diac and respiratory motion may prevent diagnosis
of minor dislodgments. Typically, RV apical leads Real-Time Telemetry
dislodge toward the tricuspid valve; atrial leads Real-time display of intracardiac electrograms
dislodge into the RV; and coronary venous leads of is used to assess the effect of body position and

76 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

TABLE I.
Detection of Malfunction in Transvenous ICD Leads

Test Sensitivity*
Asymptomatic Patients with All Patients with
Patients Inappropriate Shocks Lead Failures
(n = 11) (n = 7) (n = 18)

Individual tests

m
Stored electrograms 27% 100% 56%
X-ray 27% 29% 28%
Pacing threshold 36% 14% 28%
R-wave amplitude 55% 0% 6%

co
Pacing impedance 9% 0% 6%
High-voltage lead impedance 29% 14% 43%
Defibrillation threshold 0% 0% 0%
Combination of tests
All pacing/sensing tests 82% 14% 78%

d.
Pacing/sensing, stored electrograms and x-ray 100% 100% 100%
Sensing and detection of induced VF 29% 14% 43%

Modified from Luria et al.36

maneuvers on recorded signals. The appearance of


ic electrogram (between shocking electrodes), and
er
noise during Valsava or inspiration may be caused sensing/pacing markers may permit localization
by myopotential oversensing49 or a structural lead of the malfunction to a specific component of the
defect. Abnormalities of pacing function, R-wave lead. Medtronic ICDs permit use of special teleme-
amplitude, pacing and shocking impedance, and try Holter monitors for troubleshooting.50 They
stored evidence of oversensing39 usually indicate
ac

provide a simultaneous, 24-hour recording of a sur-


lead defect. Simultaneous surface ECG recording face ECG lead with two intracardiac electrograms
with telemetry of the sensing signal, the far-field and an extended Marker ChannelTM (Fig. 5).
.p
w
w
w

Figure 4. Subclavian crush injury to defibrillation electrode. (A) Note the sharp bend on the lead
at the site where it passes between the subclavian vein and the first rib (arrow). (B) Lead extracted
from a patient with subclavian crush. Note the disrupted insulation and the disarray of the filers
(the circumferentially coiled conductor) at the site of crush. Reprinted with permission from Lloyd
et al.46

PACE, Vol. 29 January 2006 77


SWERDLOW AND FRIEDMAN

m
co
d.
ic
er
Figure 5. Oversensing identified by telemetry Holter monitor. Telemetered ECG and ventricular
ac

integrated-bipolar and high-voltage electrograms are recorded with extended marker channel
showing usual markers (top line) and multiple other device parameters. The sixth line shows the
VF counter which increments from 0 to 3 during oversensing. Saturation of the sensing electrogram
and intermittent oversensing are characteristic of electrode problems, in this case intermittent
conductor failure. The Holter was performed because the Short Interval Counter was abnormal.
.p

In this case, the lead impedance was normal.

Stored Episodes shocks delivered per episode, or all therapies in a


zone delivered. Most serious problems identified
w

Lead, adapter, or set screw malfunction often


presents as oversensing, resulting in inappropri- by alerts are lead-related40 (Fig. 6). Unfortunately,
ate detection of VF and aborted or inappropriate ICD recipients may not hear alert tones or may not
shocks. This results from make-break contact noise recognize their significance. Alerts are a useful ad-
w

or oversensing of extracardiac signals due to insu- junct to, but not a substitute for, periodic device
lation defects. The pattern of oversensing often in- interrogation and follow-up.
cludes extremely short nonphysiological intervals
w

(<140 ms, Fig. 6). In Medtronic ICDs, the combi- Risk Factors for Lead Failure
nation of a high Sensing Integrity CountTM of non-
physiological intervals, nonsustained tachycardia Clinical factors predictive of lead failure in-
episodes with nonphysiological intervals, and ab- clude epicardial leads,34,51 abdominal pulse gener-
normal lead impedance triggers a lead alert warn- ator location,36,47 coaxial defibrillation leads,36,41
ing on interrogation.39 subclavian venous access,48,52 and dual-chamber
(as opposed to single-chamber) ICD systems.52 In
epicardial systems, sensing and defibrillation are
Patient Alerts performed by different leads, eliminating over-
Medtronic ICDs emit patient-alert tones for sensing as an early warning for failure of defibrilla-
pacing or high-voltage lead impedances out of tion leads. Thus, periodic assessment of shocking
range, low battery voltage, long charge time, three lead impedance is essential. Currently, epicardial

78 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

m
co
d.
ic
Figure 6. Conductor failure identified by lead alert. Top panel shows marked, abrupt increases
in all measured impedance values triggering a lead alert on July 28, 2 days after the patient was
er
seen in clinic. The patient did not hear the alert tones, probably because the ICD was implanted
submuscularly. Middle panel shows episode summaries from inappropriately detected episodes
of VF during intense stationary bicycle exercise in “spinning” class (August 1). Bottom panel
shows the corresponding stored electrogram from the true-bipolar sensing lead. Signals are prob-
ably caused by mechanical motion at the fracture site. Saturation of rate sensing electrogram is
ac

characteristic of conductor failure; but persistence of oversensing throughout the cardiac cycle is
not and is probably caused by intense exercise.

systems are used only in patients with a univen- layer.36,57 A drop in the ring-to-coil impedance
.p

tricular heart or mechanical tricuspid valve. Sub- is an early marker of this failure,41 which results
cutaneous patches, which have similar construc- in oversensing of nonphysiological signals. Use of
tion to epicardial patches and have similar high ring-to-coil impedance, the nonsustained episode
failure rates, also warrant careful observation.34,53 log, and Sensing Integrity Counter may permit
w

Abdominal pulse generator placement, indepen- identification of early failure before oversensing
dent of epicardial or transvenous approach, may causes inappropriate shocks (Fig. 2). A “signa-
cause lead stress associated with tunneling and ture” presentation of this type of lead failure is
w

mechanical friction of the pulse generator against oversensing of electrical noise following shocks
the lead.36,54 Cardiac resynchronization leads have (Fig. 7).41
a higher dislodgment rate than other electrodes.55
Adapters increase the number of mechanical Protecting ICDs from High-Voltage, Short Circuits
w

connections and stress points and may increase Low impedance on the shock electrodes in-
the risk of system malfunction. A loose set screw dicates an insulation failure that diverts shock
(in the header or an adapter) can mimic lead frac- current from the heart. It may produce sufficient
ture, with make-break contact noise and elevated peak current that the ICD’s output circuit fails by
impedance. “electrical overstress.”58 For example, an insula-
Coaxial leads have higher failure rates than tion failure in the pocket may produce a short cir-
multi-lumen leads, in which conductors are longi- cuit from the high-voltage coil to the can with a
tudinally arrayed along the length of the lead.56 resistance of 8 . A maximum-output (∼800 V)
Medtronic TransveneTM coaxial leads (models shock would result in a peak current of 100 A in
6936 and 6966) are prone to failure caused the high-voltage output circuit, causing it to fail
by metal ion oxidation of a middle insulation catastrophically.59 Shorting between two active

PACE, Vol. 29 January 2006 79


SWERDLOW AND FRIEDMAN

m
co
d.
ic
er
Figure 7. Post-shock oversensing. Panels show recordings of true-bipolar sensing and marker
channel during VF induced at replacement of an ICD pulse generator attached to a chronically
ac

implanted Medtronic model 6936 TransveneTM , coaxial defibrillation electrode. Top and middle
panels are continuous. At left of top panel, T-wave shock (CD) induces VF, which is reliably sensed,
detected, and terminated by a programmed 20-J shock (CD 19.6 J) in the middle of the middle
panel. Post-shock intermittent oversensing saturates electrogram signals, resulting in failure to
identify sinus rhythm and repetitive inappropriate redetection of VF, with multiple shocks aborted
.p

using the programmer. The lower panel shows one aborted shock with detection of VF (FD). The
short charge time between FD and CE markers is caused by high residual voltage output capacitor
after previous aborted shocks. Post-shock oversensing is the typical early sign of inner insulation
failure in this coaxial defibrillation electrode.
w

intracardiac electrodes of opposite polarity can with no effective means to rescue the patient60 (see
cause the same effect. To prevent electrical- Fig. 8).
w

overstress component failure, modern ICDs protect


the output circuit by aborting the shock pulse im-
mediately if the current in the defibrillation path- Pulse Generator Failure
w

way is excessive. Depending on the manufacturer, Implantable ICD pulse generator failure is
this corresponds to a series lead impedance of 15– rare, although the incidence is largely unknown
20 . The presumption is that the short circuit due to limitations in current reporting methods.61
diverts current from the heart, preventing shocks Pulse generator failure might result from primary
from terminating VT/VF. In addition to preserv- semiconductor component failure caused by ex-
ing the generator, this feature has a safety ben- posure to extreme voltages or currents (electri-
efit. Simultaneous insulation failure of the rate- cal overstress). It can occur due to inappropri-
sensing and high-voltage components of RV leads ate internal arcing, commonly associated with the
may present as oversensing. Diverting the resultant hybrid circuit.58 In older ICDs without shorted
inappropriate shock prevents unnecessary pain. high-output protection, pulse generators fail if
Further, it may prevent fatal proarrhythmia from shocks are delivered into a short circuit resulting
a weak shock delivered in the vulnerable period from lead insulation failure.

80 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

m
co
d.
ic
Figure 8. Protection for shorted output of high-voltage circuit. Text therapy summary indicates sequence of therapies
er
after induction of VF by a 0.6 J T-wave shock (T-shock) during replacement of an ICD pulse generator attached to a
chronically implanted Guidant Model 00125 defibrillation electrode. The measured high-voltage lead impedance is
<20 . Despite programmed shocks strengths of 15, 20, and 30 J, delivered energies are 0.2, 0.3, and 0.3 J, respectively,
because of shorted output protection feature. Short charge times for second and third shocks correspond to high
ac

residual energy on capacitors at end of previous shock. Top left panel shows interval plot. Top right panel shows third
unsuccessful shock (followed by minimum distortion of post-shock, high-voltage electrogram due to low voltage) and
external rescue. An insulation failure in the pocket shorted the high-voltage lead to the generator can.

Pulse generator failure may manifest as ab- periods. These interlocks limit combinations
.p

sence of telemetry, emission of device tones, inap- of the upper rate limit and AV delay based on
propriate shock, premature battery depletion, dis- the VT detection interval. Adaptive, rate-related
play of fault codes upon interrogation, inability to shortening of the post-pacing ventricular blanking
interrogate or program, failure to charge the capac- period permits higher pacing rates with less
w

itors or retain a charge, early battery depletion, or ventricular blanking, but might increase the risk
failure to deliver therapy. Failed pulse generators of T-wave oversensing.62
should be explanted and returned to the manu- In patients with LV dysfunction, RV apical
facturer for analysis, and the incident should be pacing promotes ventricular dyssynchrony, atrial
w

reported to the Food and Drug Administration’s fibrillation, and heart failure.63,64 Dual-chamber
Manufacturer and User Facility Device Experience ICDs should be programmed to minimize ventric-
database (http://www.fda.gov/cdrh/maude.html). ular pacing in this population. One approach is
w

Rigorous reporting and analysis may distinguish to use a nontracking pacing mode (DDI) and a
random component failure from a systemic failure, long AV interval. However, this may introduce pa-
correctable by design or manufacturing modifica- rameter interlocks or—in conjunction with rate-
tions before many patients are affected.61 smoothing—intradevice interactions resulting in
undersensing of VT/VF. Specific algorithms that
Considerations for Bradycardia Pacing in ICDs promote intrinsic conduction periodically deter-
To prevent intradevice interactions (see mine whether intrinsic conduction is present
“Pacemaker-ICD Interactions”), dual-chamber during dual-chamber pacing. They either prolong
ICDs have interlocks between pacing and VT/VF the AV interval (Search HysteresisTM in Guidant
detection parameters that limit the fraction of the ICDs; AutoIntrinsic Conduction SearchTM in St.
cardiac cycle affected by pacing-related blanking Jude) or change the pacing mode to a modified

PACE, Vol. 29 January 2006 81


SWERDLOW AND FRIEDMAN

m
co
d.
Figure 9. Algorithm for maximizing intrinsic AV conduction. (A) Hospital telemetry strip. The
ic
first four complexes demonstrate atrial pacing with intrinsic ventricular conduction in a patient
with MVPTM Mode programmed in a Medtronic ICD. The fifth atrial-paced beat is not conducted,
followed by AV pacing on the sixth beat. This algorithm paces in the AAI mode with ventricular
er
surveillance, to minimize ventricular pacing. Ventricular pacing occurs only if a nonrefractory
sensed or paced atrial event is not conducted (allowing for a maximum pause of two times
the lower rate limit plus 80 ms). Nominal performance of this algorithm may be confused with
pacemaker malfunction. (B) From top to bottom: surface ECG, electrogram markers, and near-field
ac

ventricular electrograms. The first three complexes depict DDDR pacing mode with MVPTM Mode
programmed on. The algorithm periodically checks for the return of AV conduction to minimize
ventricular pacing (seen in the fourth complex, AS-VS). AS = atrial sense; VS = ventricular sense;
VP = ventricular pace.
.p

AAI mode with ongoing assessment of AV conduc- ing ICDs.65–67 The two widely spaced ventricular
tion (AAI+ or AAIR+ mode, Medtronic MVPTM ). leads in resynchronization ICDs introduce novel
The AAIR+ mode eliminates interlocks between intradevice interactions between resynchroniza-
the pacing upper rate limit and VT zone, permit- tion pacing and detection of VT/VF. We address
w

ting paced rates in the VT zone. In the AAIR+ their unique troubleshooting implications. Opti-
mode, early ventricular events reset the VA inter- mizing pacing for hemodynamic benefit68 is be-
val and ventricular pacing does not occur at a fixed yond the scope of this review.
w

or dynamic AV delay. The AAIR+ mode thus elim-


inates interlocks to permit rate-responsive pac- Ventricular Sensing Problems
ing in the VT zone without the risk of intrade- Early, resynchronization systems utilized
vice interactions that may prevent detection of
w

adapters that have multiple modes of pacing and


VT by competitive bradycardia pacing22–24 (see sensing malfunction.68,69 The first dedicated sys-
“Pacemaker-ICD Interactions”). Because the PR in- tem (Guidant Contak CDTM ) sensed between the
terval is not limited, a single nonconducted P- LV and RV electrodes to determine R-R inter-
wave may occur (see Fig. 9). A potential risk of this vals. This “extended bipolar” sensing configura-
pacing mode is ventricular proarrhythmia caused tion “merges” events sensed in the LV or RV
by pause-dependent VT. into a single recording channel. During pacing,
RV and LV depolarizations are synchronized and
Troubleshooting Cardiac Resynchronization refractory periods prolonged, preventing double-
Therapy in ICDs counting. But during conducted rhythms or VT, R-
The vast majority of cardiac resynchroniza- wave double-counting occurs whenever the inter-
tion pacing in the United States is performed us- val between local electrograms of conducted beats

82 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

m
co
d.
Figure 10. Multiple shocks in a patient with a cardiac resynchronization ICD with extended
bipolar sensing (Contak CTM ) due to double counting initiated by a pacemaker-mediated tachy-
cardia prevention algorithm. Electrograms from top to bottom are atrial, ventricular near-field,
ic
and ventricular far-field. The initial rhythm is sinus tachycardia, with P-synchronous pacing at
the maximum tracking rate (“VP-MT”). This algorithm periodically extends the post-ventricular
atrial refractory period during upper-rate limit pacing to abort a pacemaker-mediated tachycardia
er
(indicated by “PMT-B”). The following atrial-sensed event falls in the extended post ventricular
atrial refractory period (indicated by the parenthesis, “(AS)”) and is therefore not tracked. Cessa-
tion of ventricular pacing permits double counting of the intrinsic-wide QRS complex by extended
bipolar sensing, leading to inappropriate detection of VF. VS = ventricular sense in sinus zone.
ac

Other abbreviations as in Figure 11.

at the RV and LV electrodes is greater than the ven- promote inappropriate shocks by temporarily sus-
tricular blanking period, leading to double sens- pending ventricular pacing and permitting double-
ing of a single ventricular depolarization.70 Any counting (Fig. 10).
.p

event that inhibits ventricular pacing even tran- Extended bipolar ventricular sensing also in-
siently and thus permits AV conduction (e.g., pre- creases the risk of oversensing P-waves, most
mature complex, conducted sinus tachycardia, or commonly when the LV lead dislodges into the
SVT) may cause R-wave double-counting and in- coronary sinus. This may inhibit pacing, prevent-
w

appropriate therapy (Fig. 10 [and fig. 4 in Part I of ing resynchronization therapy for heart failure, or
this review]). cause ventricular asystole in patients with com-
R-wave double-counting may result in shocks plete heart block. Oversensing of atrial arrhyth-
w

for SVT slower than the programmed VT detection mias may cause inappropriate shocks despite a
interval because alternate device-detected inter- slow ventricular rate.69 Oversensed events may
vals correspond to (1) the interval between the two have complex and unanticipated interactions with
w

sensed components of the ventricular electrogram detection enhancements (Fig. 11).


and (2) the difference between the true ventricu-
lar cycle length and the double-counting interval.
See Figure 6 for details. Because alternate “R-R” LV Threshold and RV Anodal Capture
intervals in any conducted rhythm increment the Even if they use RV sensing, most ICDs
VF counter, all detected VT or SVT episodes are use dual-cathodal (“extended bipolar”) pacing in
classified as VF and treated with shocks, regard- which the LV and RV tip electrodes serve as the
less of cycle length. Because there are two detected combined cathode. The common anode may be ei-
intervals for each ventricular electrogram, tran- ther the RV coil (integrated-bipolar pacing config-
sient nonsustained tachycardias may satisfy the uration) or the RV ring (true-bipolar pacing con-
VF detection criterion, resulting in aborted shocks. figuration). Some ICDs permit simultaneous (or
Pacemaker algorithms—such as those designed to temporally offset) true-bipolar pacing in both the
terminate pacemaker-mediated tachycardia—may LV and RV (Fig. 11).

PACE, Vol. 29 January 2006 83


SWERDLOW AND FRIEDMAN

m
co
d.
ic
er
ac

Figure 11. Extended and true bipolar left LV pacing configurations (available in the Contak Renewal 3TM ICD). The top
two panels depict extended bipolar pacing, in which pacing occurs between LV and RV electrodes. This configuration
has the advantage of requiring only a single LV electrode, but the disadvantage of permitting anodal capture (Fig. 13).
The ability to select either the distal or proximal LV electrode as cathode for pacing functionally permits “moving” the
.p

LV lead via noninvasive reprogramming. This can be advantageous postoperatively if the pacing threshold increases
or phrenic-nerve stimulation occurs. The bottom two panels depict true-bipolar LV pacing, which eliminates the
possibility of anodal RV capture, but requires a bipolar LV lead.
w

Determining whether the LV lead is captur- the pacing pulse may be supra-threshold in one
ing can be difficult. At least seven possible states chamber and subthreshold in the other, leading to
of ventricular pacing can occur in cardiac resyn- loss of resynchronization. In these ICDs, a change
w

chronization systems that use an extended bipolar of ventricular-electrogram morphology during a


pacing configuration. These are no capture (intrin- threshold test indicates loss of capture at one elec-
sic conduction) and capture from six specific pac- trode; the corresponding change in the surface
ing configurations: RV cathodal + LV cathodal (the ECG usually indicates which pacing configura-
w

intended state); LV (cathodal) only; RV cathodal tion is capturing (see below). Current systems per-
only; RV anodal only; RV anodal + LV cathodal; mit independent programming of RV and LV out-
and RV (cathodal + anodal) + LV (“triple site” pac- puts, facilitating determination of the RV and LV
ing). Most of these can be distinguished by anal- threshold.
ysis of intracardiac electrograms combined with a ECG changes associated with biventricular
12-lead ECG, which is invaluable for follow-up of pacing have been reviewed.68 During determina-
resynchronization pacing systems. tion of the LV pacing threshold, the QRS complex
In early cardiac resynchronization systems, usually widens with loss of LV capture. It becomes
dual-cathodal pacing applies the same voltage to less negative in lead I and more negative in lead
the commonly wired LV and RV cathodes. Since III as ventricular excitation originates from the in-
the local LV and RV thresholds usually differ, ferior RV as opposed to the lateral LV (Fig. 12).

84 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

m
co
d.
ic
er
Figure 12. QRS morphology during biventricular pacing. The morphology of the paced QRS
complex depends on lead positions and the distribution and properties of diseased myocardium.
ac

Left panel: LV capture is suggested by a markedly negative QRS in lead I. Middle panel: Loss of
LV capture is associated with a positive QRS compels in lead I and a more negative deflection in
lead III. Right panel: With loss of capture, intrinsic conduction returns. In this case, the intrinsic
QRS is narrow.
.p

During dual-cathodal pacing, a ventricular electro- fibrillation coil (integrated-bipolar pacing), which
gram that follows the pacing stimulus by 50–200 has a large surface area (low current density). Thus,
ms may assist in determining the site of capture. It anodal capture is extremely rare in Medtronic and
w

occurs when one ventricular electrode senses the Guidant ICDs, which use integrated-bipolar pac-
wavefront propagating from the site of capture in ing. St. Jude ICDs use true-bipolar pacing if the LV
other ventricle, giving rise to the second event. lead is unipolar and the RV lead is true-bipolar.
“Anodal capture” occurs when the delivered Anodal capture is more common in this pacing
w

voltage captures at the RV anode either in isola- configuration.


tion or in conjunction with the LV cathode. When
anodal capture results in “triple-site” pacing (LV Ensuring Delivery of Effective Resynchronization
w

cathode, RV cathode, and RV anode), the QRS du- In order to provide effective cardiac resyn-
ration may decrease by 10–20 ms and QRS ampli- chronization, a high frequency of biventricular
tude of leads I and aVL may increase.71 If anodal pacing must be delivered. A practical goal is to
RV capture is mistaken for LV capture, the pac- resynchronize 90% of R-R intervals. Figure 15
ing output may be programmed to a subthreshold shows a systematic approach to confirming resyn-
value in the LV, resulting in loss of cardiac resyn- chronization. Loss of resynchronization occurs if
chronization (Figs. 13 and 14). It is much more LV capture fails. Kay provides a comprehensive re-
common if the RV anode is a ring electrode (true- view of the differential diagnosis of loss of resyn-
bipolar pacing), which has a small surface area chronization for reasons other than loss of LV
(high current density) and may be in direct contact capture.72 The most common causes are intrin-
with endocardium than if the RV anode is a de- sic AV conduction due to a long programmed AV

PACE, Vol. 29 January 2006 85


SWERDLOW AND FRIEDMAN

dersensing and ventricular oversensing may simi-


larly minimize ventricular pacing and limit resyn-
chronization.
Resynchronization ICDs have features de-
signed to maintain LV stimulation. To prevent LV
T-wave oversensing, Guidant RenewalTM ICDs
incorporate an LV refractory period, which may
inhibit LV pacing. Additionally, they include an
LV protection period after a sensed or paced event
during which pacing will not occur. Although de-
signed to prevent pacing in the LV vulnerable pe-

m
riod, this parameter reduces the maximum LV pac-
ing rate and may inhibit cardiac resynchronization
(Fig. 16).
Resynchronization ICDs also employ algo-

co
rithms designed to maximize ventricular pac-
ing during potentially disruptive events such as
premature ventricular complexes or rapidly con-
ducted atrial arrhythmias. In Medtronic resyn-
chronization ICDs, the Ventricular Sense Res-

d.
Figure 13. Possible modes of anodal capture during
biventricular pacing. In all examples, an extended ponseTM feature triggers pacing in one or both
bipole paces between LV and RV leads. The top left panel ventricles after each RV-sensed event.74 Both
demonstrates biventricular pacing using an integrated- Medtronic and Guidant systems include al-
gorithms that temporarily shorten the post-
bipolar RV lead, in which the common anode is the dis-
tal defibrillation coil. Since the RV coil has a large sur-
face area and thus a low current density, anodal capture
is rare. The remaining panels depict biventricular pac-
ic
ventricular atrial refractory period to regain atrial
tracking and restore resynchronization after pre-
mature ventricular complexes or during sinus
er
ing between a LV electrode and a true-bipolar RV lead, in tachycardia faster than the nominal upper rate
which the common anode is the RV ring electrode. The limit. Medtronic’s Conducted AF ResponseTM
top right panel depicts normal function, with capture resynchronizes conducted beats in atrial fibrilla-
only at the LV tip during LV-only pacing. The bottom tion up to a minimum R-R interval without increas-
ac

left panel depicts capture only at the anode. This oc- ing ventricular rate. Guidant’s Ventricular Rate
curs if the anodal RV-ring threshold is lower than the LV RegularizationTM algorithm is intended to restore
threshold and the pacing output is between the RV an- resynchronization and ventricular regularity by
odal threshold and LV cathodal thresholds. The bottom pacing the ventricle during irregular conduction
right panel depicts anodal and cathodal capture. This of atrial fibrillation. Concealed conduction into the
.p

may confound the determination of LV pacing threshold. AV node from the paced events may slow AV nodal
See Figure 14. conduction, thereby limiting the pacing-induced
increase in ventricular rate. These and other algo-
rithms designed to maximize cardiac resynchro-
w

delay and loss of atrial tracking at fast atrial rates.


Others include frequent premature ventricular nization therapy may result in pacing after QRS
complexes, sensing malfunction (see above), pro- onset on surface ECGs and pacing at “unexpected”
gramming considerations and/or specific pacing times (Fig. 17).
w

algorithms (see below). Interruption of cardiac Phrenic-Nerve Stimulation


resynchronization pacing is common, but it can
usually be restored by noninvasive or invasive in- Left phrenic-nerve stimulation by an LV elec-
w

tervention.73 trode may make cardiac resynchronization intol-


erable. It may become manifest only after im-
ICD Programming and Resynchronization Algorithms plantation due to postural changes or minor lead
Any parameter setting that minimizes ventric- migration. In Guidant ICDs, daily impedance
ular pacing or permits ventricular fusion will ad- measurements performed to assess lead integrity
versely affect cardiac resynchronization. Such set- temporarily increase the pacing amplitude. They
tings include a long AV delay, a prolonged post- may need to be programmed “off” in patients who
ventricular atrial refractory period or extension of have phrenic stimulation at the higher output. De-
the post-ventricular atrial refractory period after creasing the pacing output noninvasively elimi-
a premature ventricular complex, a low maximal- nates phrenic stimulation if there is a sufficient
tracking rate, AV search hysteresis, rate-smoothing safety margin between the LV and phrenic-nerve
up or down, or use of a DDI pacing mode. Atrial un- thresholds. Data are insufficient to determine how

86 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

m
co
d.
ic
er
Figure 14. Anodal capture during testing of LV capture threshold. LV pacing occurs between a LV cathode and a
RV ring anode. (A) Upper left panel: From top to bottom: surface ECG, electrogram markers and intervals, atrial
electrogram, and ventricular near-field electrogram. At left, pacing output is 2.5 V and anodal capture is present,
with capture occurring at both the LV electrode and the RV ring. No electrograms are visible on the LV channel
ac

due to the simultaneous RV and LV capture. With a decrement in pacing output to 2.25 V, the QRS morphology
changes on the ECG (circled complex); and electrograms appear on the ventricular channel, representing local RV
depolarization. Inset at lower right shows that the time between the pacing stimulus and the RV electrogram represents
the interventricular conduction time during LV pacing. (B) With further decrement in LV pacing output from 0.75 to
.p

0.5 V, true loss of LV capture occurs, with widening of the surface ECG due to left bundle branch block. The local RV
depolarizations are similar during LV pacing (in A) and intrinsic AV conduction (panel B).

often phrenic-nerve stimulation can be eliminated Sensing, Detection, and Pacing


w

by reprogramming from unipolar to bipolar LV pac- Drug effects on the amplitude and slew rate
ing or changing the LV pacing vector on a multi- of local electrograms may impair sensing. This
polar LV lead (Guidant EZTRAK 2†TM ). The alter- may delay or prevent detection of VT/VF. Antiar-
w

native is repositioning the LV lead. rhythmic drugs that slow the rate of VT below
the programmed VT rate cutoff can prevent ini-
tial detection of VT2 or divert appropriate ther-
Troubleshooting Interactions with ICDs
apy during reconfirmation (Fig. 18). Class IC
w

Antiarrhythmic Drugs drugs bind sodium channels in the open state,


Antiarrhythmic drugs are prescribed com- resulting in greater drug effect during tachy-
monly to ICD patients.75–78 Serious drug-ICD inter- cardias. They may alter ventricular-electrogram
actions associated predominantly with the use of morphology sufficiently in SVT to invalidate
sodium- or potassium-channel blocking (Vaughn- a morphology-algorithm template acquired at a
Williams class IA, IC, or III) drugs have been re- normal sinus rate. Because VT displays increased
viewed7,76,79 and are summarized in Table II. These cycle length variability in patients taking Class
interactions must be considered when antiarrhyth- IC drugs,9,13 interval-stability detection enhance-
mic drug treatment is initiated or the dose is al- ments may misclassify VT as atrial fibrillation.
tered.7 The effect of Class IC drugs to increase pacing

PACE, Vol. 29 January 2006 87


SWERDLOW AND FRIEDMAN

m
co
d.
Figure 15. Systematic approach to insuring delivery of cardiac resynchronization therapy. CRT =
ic
cardiac resynchronization therapy; PVC = premature ventricular complex; VTns = nonsustained
VT; AVN = AV node.
er
ac
.p
w
w
w

Figure 16. Inhibition of LV pacing by the programmable LV protection period in a Contak Re-
newal 3TM . Shown from top to bottom are ECG lead II, RV electrogram, and LV electrogram. A
programmer command for LV pacing was issued during atrial fibrillation. Since the LV protec-
tion period prevented pacing at the selected rate, markers indicating inhibition of pacing appear
(“inh-LVP”). Programming the protection period off permitted assessment of the LV threshold.
RVS = right ventricular sense.

88 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

m
co
d.
ic
er
Figure 17. Algorithms designed to maintain a high percent of cardiac resynchronization pacing during atrial fibrilla-
tion may be confused with inappropriate pacing. (A) Hospital telemetry shows ventricular pacing occurring after the
QRS onset due to the Medtronic Ventricular Sense ResponseTM algorithm, which introduces a triggered biventricular
pacing pulse after each RV-sensed event. (B) Surface ECG lead II, event markers, and the RV tip to LV tip electrogram
ac

are displayed during atrial fibrillation. The first three complexes represent biventricular pacing (“BV”). The last com-
plex represents a conducted complex that was sensed (“VS”). The split marker associated with the “VS” indicates
that a triggered pace pulse resulted. (C) Hospital telemetry shows pacing during rapidly conducted atrial fibrillation
due to an algorithm designed to maximize cardiac resynchronization pacing. This pacing, with the patient at rest, is
distinct from rapid, rate-responsive pacing that is triggered by a sensor.
.p

thresholds is greatest at rapid rates, so that an- Defibrillation


titachycardia pacing could be subthreshold even Antiarrhythmic drugs may cause potentially
if bradycardia pacing is supra-threshold at slower life-threatening rises in DFTs. Drug-defibrillation
w

rates.80 This rarely presents a clinical problem be- interactions are complex. Studies are confounded
cause the pulse amplitude for antitachycardia pac- by anesthetic effects, heterogeneity in ICD leads
ing nominally is higher than that for bradycardia and waveforms, and inter-species variability (e.g.,
w

pacing. When Class IA, IC, or III antiarrhythmic human vs canine vs porcine). Drugs that block the
drugs are started in an ICD patient, the slowest fast inward sodium current (such as lidocaine) el-
VT zone detection interval should be increased by evate the DFT, whereas agents those that block re-
≥40 ms, SVT-VT discriminators should be active
w

polarizing potassium currents (such as sotalol and


to prevent inappropriate therapy of SVT with rate dofetilide) lower the DFT slightly.7 Improved de-
overlapping that of drug-slowed VT, the interval- fibrillation safety margins have mitigated these ad-
stability discriminator should be programmed to verse pharmacologic effects, and potent sodium-
a less specific value, a new morphology tem- channel blocking drugs are rarely prescribed for
plate should be acquired during atrial pacing at chronic oral therapy in ICD patients. However,
a rapid rate, and a sustained-duration override chronic oral amiodarone increase DFTs to a clin-
should be considered. Noninvasive programmed ically relevant degree in some patients.81 When
stimulation may be appropriate to determine if treatment is initiated with amiodarone or another
VT/VF is detected accurately and terminated drug that often elevates the DFT, the defibrillation
promptly. safety margin should be confirmed if it is both

PACE, Vol. 29 January 2006 89


SWERDLOW AND FRIEDMAN

TABLE II.
Adverse Interactions Between Antiarrhythmic and ICDs

ICD Function Effect Drugs

Undersensing (rare) Diminished electrogram slew rate/amplitude 1A, 1C


Underdetection of VT VT slowed to rate < detection rate IA, IC, III
Increase in cycle length variability of VT results in
misclassification of VT as SVT by stability enhancement
Inappropriate detection of SVT Organization of atrial fibrillation into atrial flutter with 1:1 1C

m
AV conduction
Use-dependent altered morphology of ventricular 1C
electrogram causes misclassification of SVT as VT

co
Increase in pacing threshold Baseline Amiodarone, 1C
Rapid pacing rates for antitachycardia 1C
Bradyarrhythmias Battery depletion from more bradycardia pacing Amiodarone, β-blockers,
calcium antagonists
Pacing-induced dyssynchrony

d.
Increase in defibrillation threshold Unsuccessful shocks/death Amiodarone, 1B
More aborted or delivered shocks Proarrhythmia IA, IC, III

marginal (e.g., safety margin <10 J) and if an in-


sufficient safety margin would result in a change
in therapy (revision of the defibrillation system or
discontinuation of the drug). Similar considera-
ic may contribute to repetitive episodes of VT (“VT
storm”).85
er
tions apply to substantial increases in amiodarone Electromagnetic Interference
dose. Ubiquitous environmental electromagnetic
energy can generate electrical potentials on ICD
sensing electrodes. Such electromagnetic interfer-
Metabolic Effects
ac

ence (EMI) can result in inappropriate detection


Hyperkalemia is the metabolic abnormality of VT/VF, failure to sense VT/VF, and inappro-
with the greatest clinical effect on ICD func- priate pacing or inhibition of pacing. A static
tion. In addition to increase in pacing thresh- magnetic field causes reversion to magnet mode,
old,7,82,83 it can present as T-wave oversensing,83,84 which usually disables detection of VT/VF but
.p

R-wave double-counting, or failure to detect VT or does not alter pacing. Electromagnetic interfer-
VF (Fig. 19). Hypokalemia and hypomagnesemia ence with frequencies less than ∼2 Hz may be
w
w
w

Figure 18. Underdetection of VF caused by antiarrhythmic drugs (amiodarone and mexiletine).


Displayed from top to bottom are the atrial, ventricular near-field and ventricular far-field elec-
trograms, and event markers. VT degenerates to VF. The left panel shows detection of VF. Despite
persistence of VF (seen in far-field ventricular electrograms), slowing in local near-field electro-
gram results in four events at cycle length 378–388 ms, slower than the VT detection interval
(“VS”) resulting in diversion of therapy (“Dvrt Chrg”). Detection occurred eventually, and the
resultant ICD shock terminated VF. Atrial asystole occurs during VT/VF.

90 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

m
co
d.
ic
er
Figure 19. Underdetection of VF caused by hyperkalemia (potassium 6.7) in the setting of chronic
ac

amiodarone therapy. (A) Near-field and far-field ventricular electrograms and a Marker Channel of
a Guidant Prizm VRTM ICD are shown. The top panel shows a sine-wave VT in the far-field channel,
which is detected as VF because local electrograms on the near-field channel are double-counted.
At right of upper panel, four intervals greater than the programmed VF detection interval of 316
.p

ms (190 bpm) result in an aborted shock. Lower panel shows persistence of VF after the shock is
aborted. Too few intervals are sensed in the VF zone to permit detection of VF. The patient was
resuscitated by external shock. (B) ICD system testing after correction of hyperkalemia. Induced
VF is sensed reliably and terminated by an ICD shock. The near-field electrogram is narrow,
w

indicating that the double electrograms present during the clinical arrhythmia were caused by
functional conduction block. Courtesy of Dr. Felix Schnoell.
w

detected as R-waves and prevent automatic ad- Electromagnetic interference is a lesser prob-
justment of sensitivity from sensing VT/VF. Over- lem for true-bipolar sensing than for integrated-
sensing of higher frequency electromagnetic in- bipolar sensing. It is diagnosed based on history of
terference may cause inappropriate detection of exposure and characteristic high-frequency, non-
w

VT/VF. Pacemaker function may respond as track- physiological signals that are larger in far-field
ing (atrial channel) or inhibition (ventricular chan- than near-field electrograms (fig. 3 in part I of this
nel). These effects have been reviewed.86,87 With review). Diminishing programmed sensitivity may
the exception of industrial and military sources, mitigate interference. VT/VF detection may be dis-
clinically significant interference by nonmedical abled during planned, unavoidable exposure (e.g.,
sources is rare.88,89 Interference from sources in surgical electrocautery) if an external defibrillator
the medical environment has been studied specif- and trained personnel are present. Field strength
ically, including magnetic resonance imaging90,91 in the patient’s environment may be measured if
radiofrequency catheter ablation92 and surgical external electromagnetic interference is suspected.
electrocautery.93 We address limited issues rele- Static magnetic fields >5 G should be avoided. ICD
vant to troubleshooting. detection of time-varying, external magnetic fields

PACE, Vol. 29 January 2006 91


SWERDLOW AND FRIEDMAN

depend on their frequency and orientation as well two distinct rates of VT and antitachycardia pac-
as the size and number of inductive loops in the ing for monomorphic VT that overlaps in rate with
lead. Reasonable field intensity limits are 6000 polymorphic VT. Other experts recommend two
V/m for 60 Hz alternating current, 70 V/m for rate zones with the slowest at cycle length 340–320
high-frequency fields (>150 kHz), 80 Gauss for ms for patients whose only spontaneous arrhyth-
modulated magnetic fields with frequency <10 mia is VF. This approach lowers the risk of inap-
MHz, and 1 G for those with frequency >10MHz.94 propriate therapy but increases the risk of not treat-
Given the variable response to time-varying fields, ing VT. In the largest primary-prevention trial,99
one approach to assessing electromagnetic safety ICDs were programmed to a single detection zone
of an environment is to place an identical and at cycle length 320 ms, without SVT-VT discrim-
identically programmed ICD in a saline bath or inators. Approximately half of shocks were inap-

m
conductive gel and expose it to the environment. propriate; the incidence of sudden death from un-
The ICD should be connected to the same lead detected VT has not been reported.
with several loops to assess the effect of induced The sinus-VT rate boundary should be slow
currents. Electrograms and telemetered markers enough to ensure detection of all hemodynami-

co
should be recorded. Alternatively, the patient’s cally compromising VT. The boundary between
ICD electrogram and marker channels may be the two VT zones should be based on the cy-
recorded while the patient is operating equip- cle length at which different types or fewer tri-
ment, preferably with the ICD in a “monitor- als of antitachycardia pacing are preferred. The
only” mode and a contingency plan for prompt VT-VF rate boundary is based on the cycle length

d.
defibrillation. below which antitachycardia pacing should not
be delivered. In Medtronic ICDs, which use
consecutive-interval counting above the VF Inter-
Post-Implant Programming to Optimize Therapy
val and X-of-Y counting below it,62 this bound-
The medieval jurist Henry de Bracton ap-
plied the phrase “An ounce of prevention is worth
a pound of cure” to philosophy of law. But it
ic
ary should be set to prevent underdetection of ir-
regular, polymorphic VT by consecutive-interval
counting.
er
is equally applicable to ICDs: Appropriate pro-
gramming minimizes the need for troubleshoot-
ing. In clinical studies, better programming of Duration for Detection and Redetection
SVT-VT discriminators would have prevented ap- Detection duration prior to antitachycardia
ac

proximately one-quarter of inappropriate therapy pacing should not be decreased from nominal
for SVT.11,95 In practice, this fraction probably is values because therapy is immediate after detec-
higher. Further, reports continue to include fatal tion and undersensing of monomorphic VT is
proarrhythmia and sudden deaths due to ICD pro- rare. It should be increased in patients who have
gramming errors.4,60,96 long episodes of nonsustained VT. Substantial in-
.p

creases in duration for detection of VT proba-


Sensing
bly are safe in St. Jude and Guidant ICDs, which
If the amplitude of the sinus-rhythm R-wave at use counting methods that are insensitive to occa-
implant is ≥7 mV, nominal sensitivity near 0.3 mV sional long ventricular intervals or undersensing:
w

provides adequate sensing of VF.16 More sensitive Guidant ICDs use X-of-Y counting, and St. Jude
settings may be selected if the R-wave is ≤3 mV. ICDs count based on the interval and average of the
Less sensitive settings decrease the risk of T-wave previous three intervals. In contrast, consecutive-
w

oversensing in Medtronic ICD, but may increase interval counting used by Medtronic ICDs in the
the risk of undersensing VF. VT zone may underdetect VT if occasional long
ventricular intervals or undersensing occur.9 Un-
Detection Zones and Zone Boundaries
w

less VT is known to be highly regular, the num-


Typical values for these rate boundaries are ber of intervals to detect VT usually should not
360–500 ms for slower VT, 340–300 ms for faster be more than 50% greater than the nominal value
VT, and 280–240 ms for VF. Some experts (includ- in Medtronic ICDs. Duration for detection of VF
ing the authors) recommend three-zone program- may be reduced from nominal only if there is
ming in most patients implanted for secondary no alternative method to ensure reliable detec-
prevention—even those whose only clinical ar- tion. Redetection time for VT should be increased
rhythmia is VF—because most spontaneous VF from nominal if VT is well-tolerated, but anti-
begins with rapid VT and most rapid VT can be tachycardia pacing is not. Redetection time for VF
terminated by antitachycardia pacing.97,98 Three should be decreased if post-shock undersensing
zones permit different antitachycardia pacing for occurs.

92 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

SVT-VT Discrimination A potential risk of simplifying the user inter-


Some SVT discriminators should be pro- face is loss of programming flexibility and trou-
grammed at implant in all patients with intact AV bleshooting tools. ICDs are required to perform
conduction. See Section III D. multiple functions based on input signals and clin-
ical conditions that vary widely from patient to
patient. Features optimized for one condition do
Programmed Therapy not perform optimally for another condition. Con-
Antitachycardia pacing should be pro- sider familiar clinical trade-offs: The preferred so-
grammed in all patients unless it is known to lution to rejection of far-field R-waves depends on
be ineffective or proarrhythmic. See Section IV the relative sizes and timing of the P-wave and far-
C The first shock strength for VF should be set field R-wave as well as the likelihood that the pa-

m
either in relation to an established implant safety tient will develop atrial fibrillation or flutter. The
margin or to maximum.16,27 In modern ICDs with preferred solution to discriminating SVT from VT
short charge times, we recommend that shocks for with a 1:1 AV relationship depends on various fea-
hemodynamically stable VT be programmed to tures such as the reliability of morphology discrim-

co
the same strength as those for VF, rather than to a ination with available electrograms, the difference
low energy. This maximizes the likelihood that a between VA and AV times during VT and SVT, and
ventricular shock will terminate (inappropriately the specific responses of VT and SVT to antitachy-
detected) rapidly conducted, atrial fibrillation. It cardia pacing. Prophylactic implantation of ICDs
also minimizes both the risk that a ventricular without prior electrophysiological evaluation in-

d.
shock will accelerate VT and the risk that it creases the conflicting requirements for reliable
will be weaker than the atrial upper limit of detection of all VTs while rejecting unanticipated
vulnerability (and thus initiate atrial fibrillation if SVTs.
ICD manufacturers are working to provide a
But we recognize that practice differs with regard
to this issue.
ic
it is delivered during the atrial vulnerable period).
simplified user interface that provides sufficient
options for common clinical situations. At the
same time, they seek to maintain and increase
er
the wide range of flexible options that permit op-
Evolving Trends timal programming for outlier patients or clini-
The ICD has expanded from a single-purpose cal conditions. Manufacturers must also provide
“shock box” to a multi-purpose platform for elec- physicians with adequate technical references for
ac

trical cardiac therapies; ICDs under development troubleshooting. Reference manuals are no longer
include sensors for monitoring heart failure and packaged with ICDs, although some manufacturers
other comorbidities. This increasing functionality provide online manuals.
and complexity is likely to expand the domain of Present ICDs provide troubleshooting alerts to
troubleshooting and to result in yet unknown, un- the patient and physician for specific risk condi-
.p

intended consequences, including novel intrade- tions.40 Most relate simply to a parameter “out-of-
vice interactions. range,” but at least one lead-integrity alert that in-
Present ICDs have approximately 500 pro- tegrates multiple parameters has been verified and
grammable features. Despite simplified user in- released.39,102 The capability to download ICD data
w

terfaces that might reduce operator errors, both remotely over the Internet37,38 permits develop-
nonlethal100,101 and lethal4,96 programming errors ment of automated troubleshooting algorithms that
occur. The automatic, self-checking, and self- require computing power beyond that available to
w

adjusting functions in today’s ICDs are a wel- implanted ICDs or even ICD programmers. One ap-
come addition, given the time pressures of present plication would be high-level, automated analysis
clinical practice and the fact that troubleshooting of stored arrhythmia episodes to evaluate the ac-
w

is both time consuming and poorly reimbursed. curacy of SVT-VT discrimination and recommend
However, automated features should be disabled programming changes. Physician-initiated or au-
in some conditions. Examples discussed in this tomated ICD reprogramming via internet could
review include the need to disable automatic up- both simplify troubleshooting and introduce novel
dating of morphology-algorithm templates in pa- problems.
tients with rate-related aberrancy and the need to
disable Medtronic Smart Mode when VT recurs
rapidly after effective antitachycardia pacing. Fu-
Acknowledgments: We wish to thank Jim Gilkerson,
ture automated features may also introduce un- Ph.D. (Guidant), Walter Olson, Ph.D., and Jeff Gillberg, M.S.
foreseen consequences that require novel forms of (Medtronic), and Tami Shipman, B.S. (St. Jude Medical) for
troubleshooting. their diligent reviews of this article.

PACE, Vol. 29 January 2006 93


SWERDLOW AND FRIEDMAN

References 22. Shivkumar K, Feliciano Z, Boyle NG, et al. Intradevice interaction


in a dual chamber implantable cardioverter defibrillator preventing
1. Kolb C, Deisenhofer I, Weyerbrock S, et al. Incidence of anti-
ventricular tachyarrhythmia detection. J Cardiovasc Electrophysiol
tachycardia therapy suspension due to magnet reversion in im-
2000; 11:1285–1288.
plantable cardioverter defibrillators. Pacing Clin Electrophysiol
23. Glikson M, Beeman AL, Luria DM, et al. Impaired detection of ven-
2004; 27:221–223.
tricular tachyarrhythmias by a rate-smoothing algorithm in dual-
2. Bansch D, Castrucci M, Bocker D, et al. Ventricular tachycardias
chamber implantable defibrillators: Intradevice interactions. J Car-
above the initially programmed tachycardia detection interval in
diovasc Electrophysiol 2002; 13:312–318.
patients with implantable cardioverter-defibrillators: Incidence,
24. Cooper JM, Sauer W, Verdino R. Absent ventricular tachycardia de-
prediction and significance. J Am Coll Cardiol 2000; 36:557–565.
tection in a biventricular implantable cardioverter-defibrillator due
3. Shukla HH, Flaker GC, Jayam V, et al. High defibrillation thresholds
to intradevice interaction with a rate smoothing pacing algorithm.
in transvenous biphasic implantable defibrillators: Clinical predic-
Heart Rhythm 2004; 1:728–731.
tors and prognostic implications. Pacing Clin Electrophysiol 2003;
25. Wietholt D, Kuehlkamp V, Meisel E, et al. Prevention of sus-
26:44–48.
tained ventricular tachyarrhythmias in patients with implantable
4. Swerdlow C, Shivkumar K. Implantable cardioverter defibrillators:
cardioverter-defibrillators-the PREVENT study. J Interv Card Elec-

m
Clinical aspects. In: Zipes DP, Jalife J (eds.): Cardiac Electrophysi-
trophysiol 2003; 9:383–389.
ology: From Cell to Bedside, 4th Ed. Philadephia, W.B. Saunders,
26. Shalaby AA. Delayed detection of ventricular tachycardia in a dual
2004, pp. 980–993.
chamber rate adaptive pacing implantable cardioverter defibrilla-
5. Bansch D, Steffgen F, Gronefeld G, et al. The 1 + 1 trial: A prospec-
tor: A case of intradevice interaction. Pacing Clin Electrophysiol
tive trial of a dual- versus a single-chamber implantable defibril-
2004; 27:1164–1166.
lator in patients with slow ventricular tachycardias. Circulation

co
27. Gold MR, Higgins S, Klein R, et al. Efficacy and temporal stability
2004; 110:1022–1029.
of reduced safety margins for ventricular defibrillation: Primary
6. Brode SE, Schwartzman D, Callans DJ, et al. ICD-antiarrhythmic
results from the Low Energy Safety Study (LESS). Circulation 2002;
drug and ICD-pacemaker interactions. J Cardiovasc Electrophysiol
105:2043–2048.
1997; 8:830–842.
28. Block M, Breithardt G. Optimizing defibrillation through improved
7. Goldschlager N, Epstein A, Friedman P, et al. Environmental and
waveforms. Pacing Clin Electrophysiol 1995; 18:536–538.
drug effects on patients with pacemakers and implantable car-
29. Mouchawar G, Kroll M, Val-Mejias JE, et al. ICD waveform op-

d.
dioverter/defibrillators: A practical guide to patient treatment.
timization: A randomized, prospective, pair-sampled multicenter
Arch Intern Med 2001; 161:649–655.
study. Pacing Clin Electrophysiol 2000; 23:1992–1995.
8. Monahan KM, Hadjis T, Hallett N, et al. Relation of induced to
30. Merkely B, Lubinski A, Kiss O, et al. Shortening the second phase
spontaneous ventricular tachycardia from analysis of stored far-
duration of biphasic shocks: Effects of class III antiarrhythmic
field implantable defibrillator electrograms. Am J Cardiol 1999;
drugs on defibrillation efficacy in humans. J Cardiovasc Electro-
83:349–353.
9. Swerdlow CD, Ahern T, Chen PS, et al. Underdetection of ventric-
ular tachycardia by algorithms to enhance specificity in a tiered-
therapy cardioverter-defibrillator. J Am Coll Cardiol 1994; 24:416–
424.
ic physiol 2001; 12:824–827.
31. Takeuchi ES, Quattrini PJ, Greatbatch W. Lithium/silver vanadium
oxide batteries for implantable defibrillators. Pacing Clin Electro-
physiol 1988; 11:2035–2039.
32. Nunain SO, Roelke M, Trouton T, et al. Limitations and late com-
er
10. Weber M, Bocker D, Bansch D, et al. Efficacy and safety of
plications of third-generation automatic cardioverter-defibrillators.
the initial use of stability and onset criteria in implantable car-
Circulation 1995; 91:2204–2213.
dioverter defibrillators. J Cardiovasc Electrophysiol 1999; 10:145–
33. Mattke S, Muller D, Markewitz A, et al. Failures of epicardial and
153.
transvenous leads for implantable cardioverter defibrillators. Am
11. Glickson M, Swerdlow CD, Daoud EG, et al. Optimal combination
Heart J 1995; 130:1040–1044.
of discriminators for differentiating ventricular from supraventric-
ac

34. Glikson M, Luria D, Friedman PA, et al. Are routine arrhythmia


ular tachycardia by dual-chambers defibrillators. J Cardiovasc Elec-
inductions necessary in patients with pectoral implantable car-
trophysiol 2005; 16:732–739.
dioverter defibrillators? J Cardiovasc Electrophysiol 2000; 11:127–
12. Brugada J, Mont L, Figueiredo M, et al. Enhanced detection crite-
135.
ria in implantable defibrillators. J Cardiovasc Electrophysiol 1998;
35. Dorwarth U, Frey B, Dugas M, et al. Transvenous defibrillation
9:261–268.
leads: High incidence of failure during long-term follow-up. J Car-
13. Frang P, Kus T, Vinet A, et al. Underdetection of ventricular tachy-
diovasc Electrophysiol 2003; 14:38–43.
.p

cardia using 40 ms stability criterion: Effect of antiarrhythmic ther-


36. Luria D, Glikson M, Brady PA, et al. Predictors and mode of detec-
apy. Pacing Clin Electrophysiol 1997; 20:2882–2892.
tion of transvenous lead malfunction in implantable defibrillators.
14. Klein G, Manolis A, Viskin S, et al. Clinical performance of
Am J Cardiol 2001; 87:901–904.
waveletTM morphology discrimination algorithm in a worldwide
37. Joseph GK, Wilkoff BL, Dresing T, et al. Remote interrogation and
single chamber ICD population. (abstract) Circulation 2003; 110:III-
monitoring of implantable cardioverter defibrillators. J Interv Card
345.
w

Electrophysiol 2004; 11:161–166.


15. Gold MR, Shorofsky SR, Thompson JA, et al. Advanced rhythm dis-
38. Schoenfeld MH, Compton SJ, Mead RH, et al. Remote monitoring
crimination for implantable cardioverter defibrillators using elec-
of implantable cardioverter defibrillators: A prospective analysis.
trogram vector timing and correlation. J Cardiovasc Electrophysiol
Pacing Clin Electrophysiol 2004; 27:757–763.
2002; 13:1092–1097.
39. Gunderson B, Patel A, Bounds C. Automatic identification of im-
w

16. Swerdlow CD. Implantation of cardioverter defibrillators without


plantable cardioverter-defibrillator lead problems using intracar-
induction of ventricular fibrillation. Circulation 2001; 103:2159–
diac electrograms. Comput Cardiol 2002; 29:121–124.
2164.
40. Becker R, Ruf-Richter J, Senges-Becker JC, et al. Patient alert in im-
17. Panotopoulos P, Krum D, Axtell K, et al. Ventricular fibrillation
plantable cardioverter defibrillators: Toy or tool? J Am Coll Cardiol
sensing and detection by implantable defibrillators: Is one better
w

2004; 44:95–98.
than the others? A prospective, comparative study. J Cardiovasc
41. Ellenbogen KA, Wood MA, Shepard RK, et al. Detection and man-
Electrophysiol 2001; 12:445–452.
agement of an implantable cardioverter defibrillator lead failure:
18. Dekker LR, Schrama TA, Steinmetz FH, et al. Undersensing of VF
Incidence and clinical implications. J Am Coll Cardiol 2003; 41:73–
in a patient with optimal R wave sensing during sinus rhythm.
80.
Pacing Clin Electrophysiol 2004; 27:833–834.
42. Danilovic D, Ohm OJ. Pacing impedance variability in tined steroid
19. Cohen AI, Wish MH, Fletcher RD, et al. The use and interaction of
eluting leads. Pacing Clin Electrophysiol 1998; 21:1356–1363.
permanent pacemakers and the automatic implantable cardioverter
43. Vollmann D, Luethje L, Zenker D, et al. Subthreshold test pulses
defibrillator. Pacing Clin Electrophysiol 1988; 11:704–711.
versus low energy shock delivery to estimate high energy lead
20. Geiger MJ, O’Neill P, Sharma A, et al. Interactions between transve-
impedance in implanted cardioverter defibrillator patients. Pacing
nous nonthoracotomy cardioverter defibrillator systems and per-
Clin Electrophysiol 2003; 26:457–460.
manent transvenous endocardial pacemakers. Pacing Clin Electro-
44. Mehdirad AA, Stohr EC, Love CJ, et al. Implantable defibrillators
physiol 1997; 20:624–630.
impedance measurement using pacing pulses versus shock deliv-
21. Glikson M, Trusty JM, Grice SK, et al. A stepwise testing protocol for
ery with intact and modified high voltage lead system. Pacing Clin
modern implantable cardioverter-defibrillator systems to prevent
Electrophysiol 1999; 22:437–441.
pacemaker-implantable cardioverter-defibrillator interactions. Am
45. Sperzel J, Pitschner HF, Schwarz T, et al. Automatic capture
J Cardiol 1999; 83:360–366.

94 January 2006 PACE, Vol. 29


ADVANCED ICD TROUBLESHOOTING

verification in ICD lead systems using intracardiac ventricular 68. Asirvatham S, Hayes DL. Optimization of biventricular devices. In:
evoked response and reduced coupling capacitance. Europace Hayes DL, Wang PJ, Sackner-Bernstein J, et al. (eds.): Resynchro-
2003; 5:83–89. nization and Defibrillation in Heart Failure: A Practical Approach.
46. Lloyd M, Hayes D, Friedman P. Pacemaker and ICD Radiography. Oxford, UK, Blackwell, 2004, pp. 139–162.
Cardiac Pacing and Defibrillation: A Clinical Approach. Armonk, 69. Lipchenca I, Garrigue S, Glikson M, et al. Inhibition of biventricular
NY, Futura, 2000, pp. 485–519. pacemakers by oversensing of far-field atrial depolarization. Pacing
47. Kron J, Herre J, Renfroe EG, et al. Lead- and device-related com- Clin Electrophysiol 2002; 25:365–367.
plications in the antiarrhythmics versus implantable defibrillators 70. Garcia-Moran E, Mont L, Brugada J. Inappropriate tachycardia de-
trial. Am Heart J 2001; 141:92–98. tection by a biventricular implantable cardioverter defibrillator.
48. Gallik DM, Ben-Zur UM, Gross JN, et al. Lead fracture in cephalic Pacing Clin Electrophysiol 2002; 25:123–124.
versus subclavian approach with transvenous implantable car- 71. Bulava A, Ansalone G, Ricci R, et al. Triple-site pacing in patients
dioverter defibrillator systems. Pacing Clin Electrophysiol 1996; with biventricular device-incidence of the phenomenon and car-
19:1089–1094. diac resynchronization benefit. J Interv Card Electrophysiol 2004;
49. Babuty D, Fauchier L, Cosnay P. Inappropriate shocks delivered by 10:37–45.
implantable cardiac defibrillators during oversensing of activity of 72. Kay G. Troubleshooting and reprogramming of cardiac resynchro-

m
diaphagmatic muscle. Heart 1999; 81:94–96. nization therapy. In: Ellenbogen KA, Kay GN, Wilkoff B (eds.): De-
50. Feldman CL, Olson WH, Hubbelbank M, et al. Identification of an vice Therapy for Congestive Heart Failure. Philadephia, Elsevier,
implantable defibrillator lead fracture with a new Holter system. 2004, pp. 294–310.
Pacing Clin Electrophysiol 1993; 16:1342–1344. 73. Knight BP, Desai A, Coman J, et al. Long-term retention of cardiac
51. Brady PA, Friedman PA, Trusty JM, et al. High failure rate for an epi- resynchronization therapy. J Am Coll Cardiol 2004; 44:72–77.

co
cardial implantable cardioverter-defibrillator lead: Implications for 74. Tse HF, Newman D, Ellenbogen KA, et al. Effects of ventricular rate
long-term follow-up of patients with an implantable cardioverter- regularization pacing on quality of life and symptoms in patients
defibrillator. J Am Coll Cardiol 1998; 31:616–622. with atrial fibrillation (atrial fibrillation symptoms mediated by
52. Kron J. Clinical significance of device-related complications in pacing to mean rates [AF SYMPTOMS study]). Am J Cardiol 2004;
clinical trials and implications for future trials: Insights from the 94:938–941.
Antiarrhytmics Versus Implantable Defibrillators (AVID) trial. Card 75. Pacifico A, Hohnloser SH, Williams JH, et al. Prevention of
Electrophysiol Rev 2003; 7:473–478. implantable-defibrillator shocks by treatment with sotalol. d,l-

d.
53. Kuhlkamp V. Initial experience with an implantable cardioverter- Sotalol Implantable Cardioverter-Defibrillator Study Group. N Engl
defibrillator incorporating cardiac resynchronization therapy. J Am J Med 1999; 340:1855–1862.
Coll Cardiol 2002; 39:790–797. 76. Rajawat YS, Dias D, Gerstenfeld EP, et al. Interactions of antiar-
54. Mehta D, Nayak HM, Singson M, et al. Late complications in pa- rhythmic drugs and implantable devices in controlling ventric-
tients with pectoral defibrillator implants with transvenous defib- ular tachycardia and fibrillation. Curr Cardiol Rep 2002; 4:434–

55.
ing Clin Electrophysiol 1998; 21:1893–1900.
Daoud EG, Kalbfleisch SJ, Hummel JD, et al. Implantation tech-
niques and chronic lead parameters of biventricular pacing dual-
ic
rillator lead systems: High incidence of insulation breakdown. Pac-

chamber defibrillators. J Cardiovasc Electrophysiol 2002; 13:964–


440.
77. Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an
implanted defibrillator in patients with coronary disease at high
risk for ventricular arrhythmia. Multicenter automatic defibrilla-
tor implantation trial investigators. N Engl J Med 1996; 335:1933–
er
970. 1940.
56. Guidant. CardiacRhythm Management Product Performance Re- 78. AVID Investigators. A comparison of antiarrhythmic-drug ther-
port. St. Paul, MN, Guidant Corporation, 2004. apy with implantable defibrillators in patients resuscitated from
57. Hauser RG, Cannom D, Hayes DL, et al. Long-term structural fail- near-fatal ventricular arrhythmias. N Engl J Med 1997; 337:1576–
ure of coaxial polyurethane implantable cardioverter defibrillator 1583.
ac

leads. Pacing Clin Electrophysiol 2002; 25:879–882. 79. Rajawat YS, Patel VV, Gerstenfeld EP, et al. Advantages and pitfalls
58. Hauser RG, Hayes DL, Almquist AK, et al. Unexpected ICD pulse of combining device-based and pharmacologic therapies for the
generator failure due to electronic circuit damage caused by elec- treatment of ventricular arrhythmias. Pacing Clin Electrophysiol
trical overstress. Pacing Clin Electrophysiol 2001; 24:1046–1054. 2004; 27:1670–1681.
59. Sweeney MO. Exploding implantable cardioverter defibrillator. J 80. Hook BG, Perlman RL, Callans DJ, et al. Acute and chronic cycle
Cardiovasc Electrophysiol 2001; 12:1422–1424. length dependent increase in ventricular pacing threshold. Pacing
60. Vollmann D, Luthje L, Vonhof S, et al. Inappropriate therapy and Clin Electrophysiol 1992; 15:1437–1444.
.p

fatal proarrhythmia by an implantable cardioverter defibrillator. 81. Pelosi F Jr, Oral H, Kim MH, et al. Effect of chronic amiodarone
Heart Rhythm 2005; 2:307–309. therapy on defibrillation energy requirements in humans. J Cardio-
61. Swerdlow CD, Cannom DS. ICD product reliability: Lessons from vasc Electrophysiol 2000; 11:736–740.
electrical overstress failures. Pacing Clin Electrophysiol 2001; 82. Dohrmann ML, Goldschlager NF. Myocardial stimulation thresh-
24:1043–1045. old in patients with cardiac pacemakers: Effect of physiologic
w

62. Wood M, Swerdlow C, Olson WH. Sensing and arrhythmia detec- variables, pharmacologic agents, and lead electrodes. Cardiol Clin
tion by implantable devices. In: Ellenbogen K, Kay GN, Wilkoff BL 1985; 3:527–537.
(eds.): Clinical Cardiac Pacing and Defibrillation, 2nd Ed. Philade- 83. Koul AK, Keller S, Clancy JF, et al. Hyperkalemia induced T wave
phia, W.B. Saunders, 2000, pp. 68–126. oversensing leading to loss of biventricular pacing and inappropri-
63. Sweeney MO, Hellkamp AS, Ellenbogen KA, et al. Adverse effect ate ICD shocks. Pacing Clin Electrophysiol 2004; 27:681–683.
w

of ventricular pacing on heart failure and atrial fibrillation among 84. Arthur W, Kaye GC. Hyperkalemia diagnosed by implantable car-
patients with normal baseline QRS duration in a clinical trial of dioverter defibrillator T wave sensing. Pacing Clin Electrophysiol
pacemaker therapy for sinus node dysfunction. Circulation 2003; 2001; 24:908–909.
107:2932–2937. 85. Moulik PK, Attar MN, Rose EL, et al. Successful resuscitation of a
w

64. Wilkoff BL, Cook JR, Epstein AE, et al. Dual-chamber pacing or ven- patient with electrical storm. Emerg Med J 2003; 20:E4.
tricular backup pacing in patients with an implantable defibrillator: 86. Pinski SL, Trohman RG. Interference in implanted cardiac devices,
The Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial part I. Pacing Clin Electrophysiol 2002; 25:1367–1381.
(see comment). JAMA 2002; 288:3115–3123. 87. Pinski SL, Trohman RG. Interference in implanted cardiac devices,
65. Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization part II. Pacing Clin Electrophysiol 2002; 25:1496–1509.
therapy with or without an implantable defibrillator in advanced 88. Groh WJ, Boschee SA, Engelstein ED, et al. Interactions be-
chronic heart failure (see comment). N Engl J Med 2004; 350:2140– tween electronic article surveillance systems and implantable
2150. cardioverter-defibrillators. Circulation 1999; 100:387–392.
66. Young JB, Abraham WT, Smith AL, et al. Combined cardiac resyn- 89. McIvor ME, Reddinger J, Floden E, et al. Study of pacemaker and
chronization and implantable cardioversion defibrillation in ad- implantable cardioverter defibrillator triggering by electronic arti-
vanced chronic heart failure: The MIRACLE ICD Trial. JAMA 2003; cle surveillance devices (SPICED TEAS). Pacing Clin Electrophys-
289:2685–2694. iol 1998; 21:1847–1861.
67. Bradley DJ, Bradley EA, Baughman KL, et al. Cardiac resyn- 90. Anfinsen OG, Berntsen RF, Aass H, et al. Implantable cardioverter
chronization and death from progressive heart failure: A meta- defibrillator dysfunction during and after magnetic resonance
analysis of randomized controlled trials (see comment). JAMA imaging. Pacing Clin Electrophysiol 2002; 25:1400–1402.
2003; 289:730–740. 91. Roguin A, Zviman MM, Meininger GR, et al. Modern pacemaker

PACE, Vol. 29 January 2006 95


SWERDLOW AND FRIEDMAN

and implantable cardioverter/defibrillator systems can be magnetic sus tested antitachycardia pacing in implantable cardioverter de-
resonance imaging safe: In vitro and in vivo assessment of safety fibrillators: A prospective study including 200 patients. Circulation
and function at 1.5 T. Circulation 2004; 110:475–482. 1998; 97:66–74.
92. Burke MC, Kopp DE, Alberts M, et al. Effect of radiofrequency cur- 98. Wathen MS, Sweeney MO, DeGroot PJ, et al. Shock reduction using
rent on previously implanted pacemaker and defibrillator ventric- antitachycardia pacing for spontaneous rapid ventricular tachy-
ular lead systems. J Electrocardiol 2001; 34(Suppl.):143–148. cardia in patients with coronary artery disease. Circulation 2001;
93. Casavant D, Haffajee C, Stevens S, et al. Aborted implantable car- 104:796–801.
dioverter defibrillator shock during facial electrosurgery. Pacing 99. Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable
Clin Electrophysiol 1998; 21:1325–1326. cardioverter-defibrillator for congestive heart failure. N Engl J Med
94. Medtronic Inc. Electromagnetic interference and the work site. 2005; 352:225–237.
2003:3. 100. Ozahowski TP, Greenberg ML, Mock P, et al. Implantable car-
95. Wilkoff BL, Kuhlkamp V, Volosin K, et al. Critical analysis dioverter defibrillator clinic casualties: Inadvertent reprogramming
of dual-chamber implantable cardioverter-defibrillator arrhythmia during routine implantable cardioverter defibrillator follow-up.
detection: Results and technical considerations. Circulation 2001; Pacing Clin Electrophysiol 1996; 19:1524–1525.
103:381–386. 101. Swerdlow C. Optimal programming of sensing and detection in

m
96. Mitchell LB, Pineda EA, Titus JL, et al. Sudden death in patients single chamber ventricular ICDs. Card Electrophysiol Rev 2001;
with implantable cardioverter defibrillators: The importance of 5:85–90.
post-shock electromechanical dissociation. J Am Coll Cardiol 2002; 102. Gunderson BD, Patel AS, Bounds CA, et al. An algorithm to pre-
39:1323–1328. dict implantable cardioverter-defibrillator lead failure. J Am Coll
97. Schaumann A, von zur Muhlen F, Herse B, et al. Empirical ver- Cardiol 2004; 44:1898–1902.

co
d.
ic
er
ac
.p
w
w
w

96 January 2006 PACE, Vol. 29

You might also like