You are on page 1of 6

Electrophoresis 2014, 00, 1–6 1

Tomas Drevinskas1 Research Article


Audrius Maruška1
Vitalis Briedis2
Capacitance-to-digital: The upgrades of
1 Faculty of Natural Sciences,
Department of Biology,
single chip detector
Vytautas Magnus University,
Kaunas, Lithuania The capacitance-to-digital single chip detector was upgraded. The paper discusses hard-
2 Faculty of Pharmacy, Lithuanian
ware issues and benefits of the designed/upgraded detector. The device can be operated
University of Health Sciences,
Kaunas, Lithuania from rechargeable lithium-ion battery as stand-alone, portable system and is capable of
transmitting real-time data wirelessly. The detector and additional modules (battery, battery
holder, microcontroller board, wireless module) weight is less than 85 g. Electrophoretic
Received July 4, 2014 separation in low conductivity 20 mM MES/L-His buffer, pH 6.1, was performed in order
Revised September 11, 2014
to evaluate detection parameters. The system is capable of quantification of potassium ions
Accepted September 30, 2014
down to 0.31 ␮M. Investigation of differential signal acquisition configuration showed im-
proved performance regarding external noise and temperature fluctuations. The system
can be a solution for stand-alone, field-portable capillary format separation detector.

Keywords:
Capacitance-to-digital converter / Capillary electrophoresis / Miniaturization
DOI 10.1002/elps.201400320

 Additional supporting information may be found in the online version of this


article at the publisher’s web-site

1 Introduction systems [7, 8]. Not only separation devices require modern
detection systems, but miniaturized mass analyzers [9], fluo-
Since the introduction of chromatography in 1903 by Tsvett [1] rescence detectors [10], spectrophotometers [11], impedance
and electrophoresis in 1930s by Tiselius [2], separation meth- analyzers [12], capacitive sensors [13], gas sensors [14],
ods were upgraded/developed into sophisticated analytical Fourier transform infrared spectrometers [15] as well. The
tools for complex mixtures analysis [3]. Separation systems wireless data transmission implementation into modern
are coupled to a detector that detects migrating analytes detectors is presented in the scientific literature [16].
through the detection window. Such systems are commonly Integration of modern detection systems or coupling
purchased as benchtop devices. Numerous suppliers can be them with separation systems is a challenge. Special elec-
found providing different bench-top size analytical systems trode optimization [17], additional coupling interface [18] etc.
worldwide. Contrary, field-portable, stand-alone, miniatur- are needed. Frequently, after coupling the device or system
ized separation/detection systems, or modules are described has bigger dimensions [17], or such systems need additional
more often in scientific literature [4]. Demanding industrial units for operation (laboratory power supply, alternating
guidelines [5], ambitiously growing scientific community [6] current source, connectors etc.) [19].
require automated, miniaturized, wireless, smart systems Single chip techniques provide numerous advantages:
for chemical analysis, environmental monitoring, process low noise, low power consumption, small-size of the de-
control etc. vice [20]. Therefore capacitive sensor AD7745 integrated cir-
Numerous scientific articles describe miniaturized cuit (IC) was selected as a candidate designing the single chip
analytical tools and detectors integrated into separation detector [21]. Manufacturers state, that this IC can be powered
from a low voltage source of 3.3 V and consumes 700 ␮A cur-
rent. Such characteristics provide a possibility for low power
Correspondence: Professor Audrius S. Maruška, Faculty of Natu- consumption applications.
ral Sciences, Department of Biology, Vytautas Magnus University, Often designers/engineers face additional problems,
Vileikos 8–206, LT44404 Kaunas, Lithuania when designing miniaturized systems. Printed circuit board
E-mail: a.maruska@gmf.vdu.lt
traces routing can be a challenge due to cross-talk and other
Fax: +37-037-327908
parasitic effects [22].
Capacitance-to-digital (CDC) technology allows measure-
Abbreviations: CDC, capacitance-to-digital converter; GND,
ground; IC, integrated circuit; LDO, low dropout; PCB, printed
ment of capacitance and conversion of it to digital code [23].
circuit board There are several publications with this technology applied


C 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
2 T. Drevinskas et al. Electrophoresis 2014, 00, 1–6

for liquid holdup measurement [24], detection of volatile com- capillary electrophoresis system (Agilent Technologies, Wald-
pounds [14], structural health monitoring [16] etc. This tech- bron, Germany). The data acquisition and peak integration
nology has already been applied in separation science [25,26]. software was programmed in house using open-source soft-
In our previous work the single chip (AD7745) was used ware Processing 2 (USA). Other calculations were performed
as whole detector. The device operation required laboratory using Microsoft Excel (USA).
power supply, additional electronics prototyping board, and
USB cable for data transmission [27]. Only one single-ended
electrode configuration was investigated in the former set- 2.3 Sample preparation and electrophoretic
up. Differential signal acquisition concept can be applied for separation
detection. Such technique provides higher resistance to ex-
ternal noise [28]. In addition, Wheatstone bridge is a widely Stock solutions of analytes (NaCl, KH2 PO4 , TRIS, histamine,
used measurement technique in the modern devices (digi- melittin, and bee venom) and 20 mM MES/L-His buffer/BGE,
tal scales, pressure sensors etc.). This measurement imple- pH 6.1 were prepared by dissolving adequate mass of sub-
ments four wires and four resistances (impedances). Usually stance in bidistilled water and frozen at −20°C to pre-
three resistances are equal and known and one is used as a vent microbial growth. BGE of 1 M acetic acid (pH 2.4)
sensor [29]. This method allows a simple proportional calcu- was prepared by diluting adequate volume of acetic acid with
lation of unknown resistive value of the sensor. Inequalities bidistilled water. Before use, the stock solutions were de-
due to the presence of different length wires are neglected, frosted and diluted to required concentration using bidistilled
therefore only the change of sensor resistivity is recorded water. Buffer solution was degassed in vacuum for 20 s before
[29, 30]. analysis.
The aim of this work was to design a stand-alone, Electrophoretic separations were performed using 50 ␮m
battery-powered, wireless, capacitance-to-digital single chip id, 365 ␮m od fused silica capillary. Analysis temperature was
detector for capillary electrophoresis and the improvement set at 25°C. A 14 and 17 kV voltages were used for detection
of detection parameters using different signal acquisition evaluation and real-sample analysis correspondingly. Sam-
configurations. ples were injected hydrodynamically at 50 mbar for 5 s. Sepa-
ration in MES/L-His buffer was performed in less than 5 min
and separation in 1 M acetic acid was performed in 12 min.
2 Materials and methods For evaluation of detection parameters 20 mM MES/L-His
buffer was selected as BGE. Metal cations were separated
2.1 Chemicals using TRIS buffer [31]. Capillary was flushed for 2 min with
0.1 M NaOH, bidistilled water, and BGE prior to each analysis.
L-Histidine (L-His) (99.0%), monohydrate MES (98.0%) were Each analysis was repeated three times. LOQ were calculated
purchased from Alfa Aesar (Germany), sodium hydrox- in accordance with valid pharmaceutical analytical method
ide (NaOH) (99.0%) was purchased from Reachem (Slo- validation guidelines [32].
vakia), potassium dihydro phosphate (NaH2 PO4 ) (99.3%) was
from Fisher Scientific (USA). 2-Amino-2-(hydroxymethyl)-
1,3-propanediol (TRIS) (99.8%) was purchased from Merck 3 Results and discussion
(Germany), sodium chloride (NaCl) was purchased from
Lachema (Czech Republic), acetic acid (99.8%) was purchased 3.1 Design of battery powered detector
from Lach-ner (Czech Republic). Histamine (⬎97.0%) and
melittin (91.8%) were from Sigma-Aldrich (Germany). Bee Battery holder, holder clips, and detection cell case were
venom was kindly provided by the local beekeeper. Bidistilled 3D-printed using a Printrbot 3D printer (Printrbot, USA).
water (produced in a laboratory) was used for preparation of Microcontroller development board and wireless data trans-
the solutions. mission module was glued on top of the battery holder us-
ing hot glue. AD7745 chip was soldered on 2-sided printed
circuit board (PCB) that was carefully routed in accordance
2.2 Instrumentation with technical documentation [23] and mixed signal lay-
out guidelines [33]. The AD7745 contains digital and ana-
AD7745 chip was purchased from Analog Devices (USA), log signal conversion parts, therefore IC placement affects
LT1761 3.3 V low dropout (LDO) voltage regulator was pur- the designed device performance. The device was designed
chased from Linear Technology (USA). Other electronic parts of several modules (prototyping board, wireless transmis-
and wires were purchased from the local store, APC220 wire- sion module, and AD7745 capacitance detector), therefore
less data transmission module was from DFROBOT (China). solid/unsplit grounding cannot be applied [33].
Sparkfun pro micro prototype development board was pur- In our previous applications the single chip detector
chased from SparkFun (USA). Li-ion (3.7 V, 1800 mAh) cell was powered from laboratory power supply [27]. All ground
phone battery from Kruger & Matz (Poland) was used to (GND) wires were connected to the single point of power
power the device. Analyses were performed using HP 3D CE plug grounded wire. The battery powered, stand-alone devices


C 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
Electrophoresis 2014, 00, 1–6 Microfluidics and Miniaturization 3

Figure 1. Schematic diagram of the


detector. (A) Single-ended configu-
ration. (B) Differential configuration.
(C) Wheatstone bridge configuration.
Drawn using Cadsoft Eagle (USA).

Figure 2. Photographs of different con-


figuration detectors. (A) Single-ended
configuration, (B) differential configura-
tion, (C) shielded device with copper foil.

cannot be connected to a common grounded wire, therefore of wire/trace inequality and different parasitic values for each
other solutions of this problem should be made. input pin of AD7745. Software code was adjusted so that the
High resolution analog-to-digital converters require a capacitive offset calibration register [22] would have 15 fF
very low noise power supply [34]. Consequently, low noise, offset at start (Supporting Information Vid. 1).
micro-power, LDO voltage regulator LT1761 was used for on-
board voltage stabilization. Stabilized voltage of 3.3 V pow-
3.2 Performance evaluation
ered the AD7745 chip. Decoupling capacitors were selected
as suggested in the datasheet [35]. Power supply of AD7745
First of all the previously reported version of single-ended
was bypassed using 100 nF ceramic, 1 and 10 ␮F tantalum
single chip detector [27] was reproduced (Fig. 2A). Several
capacitors (Fig. 1).
changes were implemented, namely: (i) PCB was etched with
Electrode geometry of the device was similar to geom-
GND layer under all electrode area; (ii) AD7745 chip and
etry used in our previous work [27]. Electrodes were made
passive electronic parts were completely shielded. Unshielded
from stainless steel sleeves (syringe needles) (0.4 mm id,
detector with single-ended configuration resulted in higher
0.6 mm od, 19 mm length) and soldered onto the PCB leav-
noise floor than unshielded detector with differential signal
ing a gap of 0.2 mm between actuator and sensor electrodes
acquisition configuration (Fig. 2B). Resolution parameters
(Supporting Information Fig. 1). Shielded wires were used
were calculated (Table 1).
for power supply from the battery and for digital data con-
nections. Data were collected at 9 Hz rate via inter-integrated
circuit interface and instantly sent to the computer via a wire- Table 1. Comparison of performance of different configuration
less system. The total mass of the designed detector was 82.5 g detectors
(Supporting Information Fig. 2). No. Shield Effective resolution Wheatstone bridge
Three different electrode configurations were tested: (i) (peak-to-peak bits) (Differential)
one electrode couple in single-ended mode (Fig. 1A); (ii) two
electrode couples in differential mode (Fig. 1B); (iii) four Single-ended Differential
electrode couples—Wheatstone bridge (Fig. 1C) (differential
1 Shielded 16.6 17.3 16.7
mode). In differential operation mode offset calibration was
2 Unshielded 13.4 16.7 16.3
performed. This procedure was necessary for compensation


C 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
4 T. Drevinskas et al. Electrophoresis 2014, 00, 1–6

Table 2. Detector performance regarding capillary wash Table 3. Recorded values of capacitance for different electrode
procedure configurations and different flush solutions

Effective resolution (peak-to-peak bits) No. Solution Configuration Value at Value at Difference
15°C (fF) 30°C (fF) (fF)
No. Flushing solution Active electrode Active and reference
couple washed electrode couples 1 Bidistilled water Single-ended 145.2 155.7 10.5
washed 2 Bidistilled water Differential 19.2 20.0 0.8
3 20 mM MES/L-His Single-ended 287.1 315.5 28.4
1 Bidistilled water 17.3 17.3 buffer
2 MES/L-His buffer 17.3 17.3 4 20 mM MES/L-His Differential 9.6 13.5 3.9
3 0.1 M NaOH 12 17 buffer

The results obtained show, that single-ended electrode ing temperature fluctuations test the capillaries inside mea-
configuration is highly susceptible to noise, therefore the de- surement electrode couple (for single-ended configuration)
vice must be completely shielded (Fig. 2C). The unshielded and measurement/reference electrode couple (for differen-
device differential electrode configuration offered similar per- tial electrode configuration) were injected with the test liquid
formance to the shielded device. (water and 20 mM MES/L-His buffer). Temperature shifts
During detector performance evaluation it was noticed, from 15 to 30°C and vice versa were set. During temperature
that the noise floor of a detector differs when different so- change period in capillary cassette the values of capacitance
lutions are flushed through the capillary. The sensor elec- were recorded. Lower baseline change during the tempera-
trode connected to Cin(±) and the high conductivity liquid ture shifting was observed for differential electrode configura-
(0.1 M NaOH) inside the capillary form a capacitively cou- tion than for single-ended electrode configuration (Table 3).
pled antenna which degrades signal-to-noise ratio/effective The differential signal acquisition mode showed increased
resolution. However, undesirable background noise can be performance during temperature fluctuations.
reduced by injecting the same liquid into reference electrode One of the main problems of contactless conductivity de-
couple. Consequently, two antennas are formed providing tection (or similar complex impedance measurement tech-
same pattern of noise into Cin+ and Cin– of AD7745 chip niques) is caused by fluctuating temperature [36, 37]. In
inputs resulting in subtracted noise and higher performance this case, temperature fluctuations affect both: active and
(Table 2). Similar tendencies showing reduced performance reference electrode couples, therefore, undesirable baseline
of capacitively coupled contactless conductivity detector (C4 D) shifting is negligible or reduced (Fig. 3).
detectors and higher background noise are presented by other
researchers [36]. The latter issue is the main cause of degraded
performance for single-ended C4 D detectors when high con- 3.3 Separation and detection evaluation
ductivity buffer is used for separation. In another work [30]
slight degradation of effective resolution from 16.8 bits (low In our previous work [27] it was shown that CDC detector
conductivity BGE) to 16.0 bits (high conductivity BGE) bits is capable of detecting of inorganic and organic cations and
was reported. anions in high and low conductivity buffers. Detection capa-
Another valuable feature of the detector was noticed dur- bilities in different BGE solutions were investigated. In this
ing the capillary cassette temperature stabilization period. paper, a low conductivity buffer was tested using model ana-
Low baseline shifting was observed when the temperature of lytes in order to evaluate performance of the designed device.
capillary cassette is increased or decreased. The effect of tem- Separation of three cationic analytes was performed in 20 mM
perature fluctuations was evaluated and compared between MES/L-His pH 6.1 buffer solution. Baseline separation was
differential and single-ended electrode configurations. Dur- achieved in less than 5 min (Fig. 4).

Figure 3. Temperature fluctuations


influence on CDC detector reading.
Temperature shifted down from 30°C
to 15°C and vice versa. Gray line—
single-ended electrode configuration
reading, black line—differential elec-
trode configuration reading. Condi-
tions: 50 ␮m id, 365 ␮m od fused silica
capillary (A) 20 mM MES/L-His buffer
flushed, (B) bidistilled water flushed.
Detection: 3.3 V, 32 kHz.


C 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
Electrophoresis 2014, 00, 1–6 Microfluidics and Miniaturization 5

can be stated, that the measurement is performed only in the


liquid present inside the capillary.
The detector was designed manually (all electronic parts
and electrodes were soldered by hand). The use of sophis-
ticated pick-and-place machines for electrode soldering is
expected to improve the symmetry that should increase
the performance of the detector. Such techniques are ap-
plied in the difference or instrumentation amplifiers, when
symmetry plays an important role on device performance
[28].

3.4 Real-sample separation

Figure 4. Separation of 1–K+ (44 ␮M), 2–Na+ (52 ␮M) ions and
The results obtained by other authors show that bee venom
3–TRIS (28 ␮M). Conditions: 50 ␮m id fused silica capillary,
effective length (Leff )–37.5 cm, total length (Ltot )–50.0 cm, 14 kV can be successfully separated in acidic BGE [38, 39]. In the
voltage potential, 25°C separation temperature, injection at present work, separation of bee venom was performed in the
50 mbar * 5 s. BGE–20 mM MES/L-His buffer, pH 6.1. Detection: BGE of 1 M acetic acid, pH 2.4. Conditions were not optimized
3.3 V, 32 kHz. for separation, since the intention was only to provide data
showing, that separation and detection of complex mixture is
Table 4. Calculated LOQ for different electrode configuration
possible using the detection setup of designed configuration.
detectors
Separation was performed in 12 min (Fig. 5). Analytes were
No. Range (␮M) LOQ (␮M) detected using a CDC detector with differential electrode con-
figuration. The reference electrode couple contained empty
Single-ended Differential Wheatstone fused silica capillary (50 ␮m id, 365 ␮m od). The results ob-
electrodes bridge
tained showed excellent linearity (R2 ⬎ 0.99) of the histamine
K+ 11–27500 0.54 0.31 0.56 calibration curve. LOQ for histamine was 0.4 ␮M. Perfect
Na+ 14–3500 0.49 0.25 0.45 migration time repeatability (RSD ⬍ 1.0%) and good peak
TRIS 7–1875 0.83 0.55 0.83 area repeatability (RSD ⬍ 3.6%) were achieved. Histamine
and melittin peaks were identified in the electropherogram
of separated bee venom using external standard method.
Obtained results showed excellent linearity (R2 ⬎ 0.99) of Determined concentration of histamine in the sample was
the calibration curves for the selected analytes and electrode 40 ␮M and histamine part in bee venom was found to be
configurations tested. High migration time repeatability 1.0%.
(RSD ⬍ 1.2%) and good peak area repeatability (RSD ⬍ 2.2%)
were achieved. LOQs were calculated for all the analytes and
different electrode configurations used. LOQ for K+ were
0.54 using single-ended setup and 0.31 ␮M using differen-
tial electrode setup (Table 4). In the previous work, LOD
(LOD is assumed three times lower than LOQ) of K+ was
1.6 ␮M using the same conditions, which shows an improve-
ment of more than eight times for similar electrode con-
figurations. The improvement of sensitivity is caused by an
implementation of on-board voltage regulator. The total im-
provement of sensitivity of upgraded detector was more than
15 times.
Better performance was expected from Wheatstone
bridge electrode configuration, however the results were very
similar to that obtained using single-ended configuration.
This can be explained by the fact that in Wheatstone bridge
electrode configuration the detection cell was separated from
other electronic parts. Although shielded wires were used Figure 5. Separation of bee venom sample (0.4 mg/mL) dissolved
for signal transmittance, separated modules usually decrease in water. Conditions: 50 ␮m id, 365 ␮m od fused silica capillary,
Leff —37.5 cm, Ltot —50.0 cm, 17 kV voltage potential, 25°C sepa-
performance of the device [33]. ration temperature, injection at 50 mbar * 5 s. BGE—1 M acetic
What is most important, the differential electrode config- acid, pH 2.4. Detection: 3.3 V, 32 kHz. Peaks: 1–40 ␮M histamine,
uration provides a subtraction of parasitic effects, therefore it 2–melittin.


C 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com
6 T. Drevinskas et al. Electrophoresis 2014, 00, 1–6

4 Concluding remarks [15] Wallrabe, U., Solf, C., Mohr, J., Korvink, J. G., Sensor.
Actuat. B 2004, 123–124, 459–467.
Stand-alone battery-powered capacitance-to-digital detector [16] Mascarenas, D. D. L., Flynn, E. B., Todd, M. D., Overly, T.
for capillary electrophoresis, capable of transmitting real-time G., Farinholt, K. M., Park, G., Farrar, C. R., J. Sound Vib.
data wirelessly, was designed. Different signal acquisition 2010, 329, 2410–2420.
electrode configurations were evaluated, showing approxi- [17] Francisco, K. J. M., do Lago, C. L., Electrophoresis 2009,
mately 15 times higher LOQs comparing it to the previous 30, 3458–3464.
work. The effect of external noise and thermal fluctuations [18] Maruška, A., Stankevičius, M., Stanius, Ž., Procedia
on detector performance was reduced. Developed system was Chem. 2010, 2, 54–58.
applied for the real-sample analysis. [19] Lichtenberg, J., de Rooij, N. F., Verpoorte, E., Elec-
trophoresis 2002, 23, 3769–3780.
This research was financed by the Research Fund of Vytautas [20] Leach, W. M. Jr., Proc. IEEE 1994, 82, 1515–1538.
Magnus University, grant No. P-BF-13-07. Instrumental sup- [21] Analog Devices Inc., 24-Bit Capacitance-to-Digital Con-
port from Agilet Technologies (donation of HP 3D CE capillary verter with Temperature Sensor. AD7745AD7746. Prod-
electrophoresis system) is highly acknowledged. uct datasheet, Norwood 2005.
[22] Nazarian, S., Pedram, M., Int. J. Electron 2008, 95–99,
The authors have declared no conflicts of interest. 903–937.
[23] Pratt, S., Analog Dialogue 2006, 40, 1–4.
5 References [24] An, Z., Ningde, J., Lusheng, Z., Zhongke, G., Measure-
ment 2014, 49, 153–163.
[1] Tswett, M. S., Proceedings of the Warsaw Society of Nat- [25] Kiplagat, I. K., Kubáň, P., Pelcová, P., Kubáň, V., J. Chro-
uralists [i.e., natural scientists], Biology Section, 1905, matogr. A 2010, 1217, 5116–5123.
14–6, 20–39.
[26] Takeuchi, M., Li, Q., Yang, B., Dasgupta, P. K., Wilde, V.
[2] Tiselius, A., The Moving-Boundary Method of Studying E., Talanta 2008, 76, 617–620.
the Electrophoresis of Proteins, Nova Acta Regiae Soci-
etatis Scientiarum Upsaliensis, Uppsala 1930. [27] Drevinskas, T., Kaljurand, M., Maruška, A., Electrophore-
sis 2014, 35, 2401–2407.
[3] Hjerten, S., Free Zone Electrophoresis, Almquist and
Wiksells Boktryckeri AB, Uppsala 1967. [28] Mancini, R., Op Amps for Everyone, Texas Instruments
Inc., Dallas 2002.
[4] Rı́os, Á., Zougagh, M., Avila, M., Anal. Chim. Acta 2012,
740, 1–11. [29] Hoffmann, K., Applying the Wheatstone Bridge Cir-
cuit, Hottinger Baldwin Messtechnik GmbH, Darmstadt
[5] International Conference on Harmonisation of Techni- 1986.
cal Requirements for Registration of Pharmaceuticals for
Human Use. The Value and Benefits of ICH to Drug Regu- [30] Thoren, M., Linear Technol. Appl. Note 2005, 96, 1–20.
latory Authorities—Advancing Harmonization for Better [31] Kubáň, P., Kobrin, E.-G., Kaljurand, M., J. Chromatogr. A
Health. Geneva 2010. 2012, 1267, 239–245.
[6] Moed, H. F., Glänzel, W., Schmoch, U., Handbook of [32] International Conference on Harmonisation of Technical
Quantitative Science and Technology Research, Kluwer Requirements for Registration of Pharmaceuticals for
Academic Publishers, Dordrecht 2004. Human Use. Validation of Analytical Procedures: Text
[7] Guijt, R. M., Armstrong, J. P., Candish, E., Lefleur, V., And Methodology Q2(R1), 2005.
Percey, W. J., Shabala, S., Hauser, P. C., Breadmore, M. [33] Ott, H. W., Printed Circuit Design, 2004 (June), 8–11.
C., Sensor. Actuat. B 2011, 159, 307–313. [34] Reeder, R., Designing Power Supplies for High Speed
[8] Stachowiak, T. B., Svec, F., Fréchet, J. M. J., J. Chro- ADC, Technical Article, Analog Devices Inc., Norwood
matogr. A 2004, 1044, 97–111. 2012.
[9] Badman, E. R., Cooks, R. G., J. Mass Spectrom. 2000, 35, [35] Linear Technology, LT1761 Series. 100 mA, Low Noise,
659–671. LDO Micropower Regulators in TSOT-23. Product
[10] Geng, X., Wu, D., Guan, Y., Talanta 2012, 88, 463–467. datasheet, Milpitas 2005.

[11] Schmidt, W., Biochem. Biophys. Methods 2004, 58, [36] Brito-Neto, J. G. A., Fracassi da Silva, J. A., Blanes L., do
125–137. Lago, C. L., Electroanalysis 2005, 17, 1198–1206.

[12] Chen, C. J., Liu, J. T., Chang, S.-J., Lee, M.-W., Tsai, J.-Z., [37] Brito-Neto, J. G. A., Fracassi da Silva, J. A., Blanes L., do
J. Taiwan Inst. Chem. Eng. 2012, 43, 678–684. Lago, C. L., Electroanalysis 2005, 17, 1207–1214.

[13] Yurish, S. Y., Procedia Chem. 2009, 1, 441–444. [38] Pacáková, V., Stulı́k, K., J. AOAC Int. 2000, 83, 549–
554.
[14] Botsialas, A., Oikonomou, P., Goustouridis, D., Ganetsos,
Th., Raptis, I., Sanopoulou, M., Sensor. Actuat. B 2013, [39] Kokot, Z. J., Matysiak, J., Urbaniak, B., Dereziński, P.,
177, 776–784. Anal. Bioanal. Chem. 2011, 399, 2487–2494.


C 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim www.electrophoresis-journal.com

You might also like