You are on page 1of 274

P O I N T O F C A R E

T E S T I N G

Roche OMNI S

Service Manual

Ident.Nr.: =057620922 =
Dr uckvor lage: =---=
Änderungsstand: =---=
Bezeichnung: Titelblatt OMNIS
Service Manual

Druckvorlagenherstellung: fa. pupo


Roche Diagnostics GmbH
D-68298 Mannheim / Germany
www.roche.com

Copyright © 2004 Roche Diagnostics GmbH, all rights reserved


The contents of this document may not be reproduced in any form or communicated to any third party without the prior writ-
ten consent of Roche Diagnostics. Every effort is made to ensure its correctness. Subject to change without notice.
REF/No. 03309614001
Rev. 4.0, December 2004

First edition: May 2003


– Important information! – Always follow it! –

This Service Manual contains vital warning and safety information.

This instrument is only intended for one area of application which is described in the instructions. The most
important prerequisites for use, operation, and safety are explained to ensure smooth operation. No warranty
or liability claims will be covered if the instrument is used in ways other than those described or if the neces-
sary prerequisites and safety measures are not observed.

The instrument may be operated only by persons whose qualifications enable them to comply with the safety
measures that are necessary during operation of the instrument.
To avoid direct contact with biological materials, wear lab clothing, protective gloves, protec-
tive glasses and, if necessary, mouth cover and apply disinfection and sterilization procedures.
Adjustments and maintenance performed with removed covers and connected power may be attempted only
by a qualified technician who is aware of the associated dangers.
Instrument repairs are only to be performed by the manufacturer or qualified service personnel.

Only accessories and supplies either delivered by or approved by Roche are to be used with the instrument.
This ensures the performance and measuring accuracy of the unit. These items are manufactured especially
for use with this instrument and meet the highest quality requirements.

Operation of the instrument with solutions whose composition is not consistent with that of the original
solutions' can negatively affect, above all, the long-term measurement accuracy. Deviations in the composi-
tion of the solutions can also decrease the service life of the electrodes.

The quality control requirements must be completed at least once daily for safety reasons. Because the mea-
surement results delivered by the instrument depend not only on the instrument's proper operation but also
on a variety of external influences (for example: preanalytics), the results produced by this instrument should
be appraised by an expert before additional measures are taken based on the results.

Explanation:

Meaning: "Important, see Instructions for Use."

– Important information! – Always follow it! –


– Operating safety information –

The instrument has been constructed and tested according to the protective measures stipulat-
ed by IEC 1010-1 / EN 61010-1: 2001-02 for electrical measurement, control and laboratory
instruments, as well as EN 61010-2: 2001-12 and EN 61010-2: 2002-01, special requirements for
IVD products and was delivered from the factory in flawless condition with regards to safety
features. In order to preserve this condition and to ensure safe operation, the operator of the
instrument must respect the notices and safety warnings contained in this Service manual.

• This instrument is classified under the protection class I according to


IEC 1010-1 / EN 61010-1.

• The instrument meets the conditions for overvoltage category II.

• The instrument meets the conditions for contamination level 2.

• Do not operate the instrument in an explosive environment or in the vicinity of explosive


anesthetic mixtures containing oxygen or nitrous oxide.

• If an object or fluid ingresses into the interior of the unit, disconnect the unit from the
power supply and give the unit to a trained specialist for examination before you continue
using it.

• The instrument is suitable for long-term operation indoors.

CAUTION:

• The power cord may be plugged only into a grounded socket. When using an extension
cord, make sure it is properly grounded.

• Any rupture of the ground lead inside or outside the instrument or a loose ground connec-
tion can render hazardous operation of the instrument. Intentional disconnection of the
grounding is not permitted.

• The instrument is not suitable for operation with a direct current power supply.
Use only the original mains plug delivered with the Roche OMNI S.

– Operating safety information –


Contents

Contents
1 Introduction
1.1 Symbols ............................................................................................................................. 1-1

1.2 Important notes and warnings ..................................................................................... 1-2

1.3 Disinfectant ...................................................................................................................... 1-2


1.3.1 Deproteinizer...................................................................................................................................................1-3

1.3.2 Other disinfectants........................................................................................................................................1-3

1.4 Decontamination ............................................................................................................. 1-3


1.4.1 Decontaminating the surfaces .................................................................................................................1-4
1.4.2 Screen................................................................................................................................................................1-4

1.4.3 Barcode scanner............................................................................................................................................1-4

1.4.4 Input unit ..........................................................................................................................................................1-4

1.4.5 Cleaning of the measuring chambers BG, ISE and MSS................................................................1-4

1.4.6 Tubes..................................................................................................................................................................1-4

1.5 ESD protective measures .............................................................................................. 1-5


1.5.1 Explanation of the phenomenon .............................................................................................................1-5

1.5.2 Influence of electrostatic charges on components ..........................................................................1-5

1.5.3 Why is ESD protection so important today?........................................................................................1-6

1.5.4 How can ESD protection be guaranteed..............................................................................................1-6

1.5.5 Conclusion .......................................................................................................................................................1-6

2 Revisions
2.1 Service manual ................................................................................................................ 2-1
2.1.1 Revision 1.0......................................................................................................................................................2-1

2.1.2 Revision 2.0......................................................................................................................................................2-1

2.1.3 Revision 3.0......................................................................................................................................................2-3

2.1.4 Revision 4.0......................................................................................................................................................2-4

3 Roche OMNI S – Components

Service Manual, Roche OMNI S, Version 4.0, December 2004 1


Contents

3.1 Important notes ................................................................................................................3-1

3.2 Shutdown ...........................................................................................................................3-1


3.2.1 The OMNI S remains switched off for more than 24 hours ......................................................... 3-1

3.3 Removing the rear panel ................................................................................................3-1

3.4 Folding up the central measuring unit ........................................................................3-2

3.5 T&D system (Turn & Dock) ............................................................................................3-2


3.5.1 Replacing the T&D module....................................................................................................................... 3-2

3.5.2 Replacing the fill port.................................................................................................................................. 3-4

3.5.3 Replacing the fill port holder.................................................................................................................... 3-4

3.5.4 Replacing the needle .................................................................................................................................. 3-4

3.5.5 Replacing the sample inlet path ............................................................................................................. 3-5

3.5.6 Replacing the bypass nipple .................................................................................................................... 3-5

3.5.7 Replacing the optical sensor SS6........................................................................................................... 3-5


3.5.8 Replacing the T&D sensor cover ............................................................................................................ 3-6

3.5.9 Replacing the T&D disk.............................................................................................................................. 3-6

3.5.10 Replacing the T&D tubing set.................................................................................................................. 3-7

3.5.11 Replacing the plug control........................................................................................................................ 3-7

3.5.12 Replacing the T&D actuator board ........................................................................................................ 3-9

3.5.13 Replacing the stepper motor.................................................................................................................... 3-9


3.5.14 Replacing the T&D linear actuators .................................................................................................... 3-10

3.6 Fluid mixing system (FMS) ......................................................................................... 3-11


3.6.1 Replacing the FMS..................................................................................................................................... 3-11

3.6.2 Replacing the FMS control board ........................................................................................................ 3-13

3.6.3 Replacing the FMS tubing set ............................................................................................................... 3-13

3.7 Sample distributor (SD) .............................................................................................. 3-13


3.7.1 Replacing the sample distributor cartridge ...................................................................................... 3-13
3.7.2 Replacing the sample distributor.......................................................................................................... 3-15

3.7.3 Replacing the SS1, SS2, SS3 and SS4 optical sensors ................................................................. 3-17

3.8 BG measuring chamber ............................................................................................... 3-18

2 Service Manual, Roche OMNI S, Version 4.0, December 2004


Contents

3.8.1 Replacing the BG measuring chamber .............................................................................................. 3-18

3.8.2 Replacing the fan protector (BG module)......................................................................................... 3-19


3.8.3 Replacing the electrode locking lever (BG module) ..................................................................... 3-19

3.8.4 Replacing the center seal (BG module)............................................................................................. 3-19

3.9 ISE measuring chamber (OMNI S 3-6 only) ........................................................... 3-20


3.9.1 Replacing the ISE measuring chamber.............................................................................................. 3-20

3.9.2 Replacing the fan protector (ISE module) ........................................................................................ 3-20

3.9.3 Replacing the electrode locking lever (ISE module)..................................................................... 3-20


3.9.4 Replacing the center seal (ISE module) ............................................................................................ 3-20

3.10 MSS measuring chamber (OMNI S 5 and 6 only) ................................................. 3-21


3.10.1 Replacing the MSS measuring chamber........................................................................................... 3-21

3.10.2 Replacing the MSS MC cover................................................................................................................ 3-22

3.10.3 Replacing the MSS clip............................................................................................................................ 3-22

3.10.4 Replacing the fan protector (MSS module) ..................................................................................... 3-23

3.10.5 Replacing the electrode locking lever (MSS module).................................................................. 3-23

3.10.6 Replacing the center seal (MSS module) ......................................................................................... 3-23

3.10.7 Replacing the MSS measuring chamber amplifier........................................................................ 3-23

3.11 Measuring chamber cartridge ................................................................................... 3-25


3.11.1 Replacing the measuring chamber cartridge .................................................................................. 3-25

3.11.2 Replacing the MC Actor Control Board............................................................................................. 3-26

3.12 COOX module (OMNI S 2, 4 and 6 only) ................................................................. 3-27


3.12.1 Overview ........................................................................................................................................................ 3-27

3.12.2 Replacing the lamp unit bilirubin ......................................................................................................... 3-28

3.12.3 Replacing the PolyOX-KX control board ........................................................................................... 3-29

3.12.4 Replacing the Polychromator Bilirubin............................................................................................... 3-30

3.12.5 Replacing the primary light guide........................................................................................................ 3-32

3.12.6 Replacing the cuvette holder................................................................................................................. 3-33

3.12.7 Replacing the cuvette ............................................................................................................................... 3-34

3.12.8 Replacing the cuvette seals.................................................................................................................... 3-34

3.12.9 Replacing the tubing set tHb/PolyOX................................................................................................. 3-34

Service Manual, Roche OMNI S, Version 4.0, December 2004 3


Contents

3.12.10 Replacing the hemolyzer ......................................................................................................................... 3-34

3.12.11 Replacing the Hemolyzer tube .............................................................................................................. 3-36


3.12.12 Replacing the HSIN board ...................................................................................................................... 3-36

3.13 tHb/SO2 module (OMNI S 1, 3 and 5 only) ............................................................ 3-38


3.13.1 Replacing the tHb/SO2 module ............................................................................................................ 3-38

3.13.2 Replacing the tubing set tHb/PolyOX ................................................................................................. 3-38

3.13.3 Replacing the tHb/SO2 heating unit ................................................................................................... 3-38

3.13.4 Replacing the Hb cartridge..................................................................................................................... 3-39

3.14 Peristaltic pumps (PP) ................................................................................................. 3-40


3.14.1 Replacing a peristaltic pump ................................................................................................................. 3-40

3.14.2 Replacing the pump head....................................................................................................................... 3-41

3.14.3 Replacing the pump tube........................................................................................................................ 3-42

3.15 Bottle compartment ..................................................................................................... 3-42


3.15.1 Replacing the bottle compartment cover .......................................................................................... 3-42

3.15.2 Replacing the gas dampers.................................................................................................................... 3-42


3.15.3 Replacing the waste separator.............................................................................................................. 3-43

3.15.4 Replacing the waste docking mechanism ........................................................................................ 3-43

3.15.5 Replacing the docking mechanism S1 (rinse solution) ............................................................... 3-44

3.15.6 Replacing the docking mechanisms S2 and S3 (fluid packs) ................................................... 3-44

3.15.7 Replacing the micro switch module .................................................................................................... 3-45

3.15.8 Replacing the bacteria filter ................................................................................................................... 3-45

3.15.9 Replacing the Transponder Control (waste, S1, S2, S3) .............................................................. 3-46

3.16 Vacuum system ............................................................................................................. 3-46


3.16.1 Replacing the vacuum pump completely .......................................................................................... 3-46

3.16.2 Replacing the vacuum pump ................................................................................................................. 3-48

3.16.3 Replacing the vacuum pump head...................................................................................................... 3-49

3.16.4 Replacing the moisture trap ................................................................................................................... 3-50

3.16.5 Replacing the waste tube........................................................................................................................ 3-50

3.16.6 Replacing the vacuum pump protector ............................................................................................. 3-51

3.16.7 Replacing the VP tube set....................................................................................................................... 3-51

4 Service Manual, Roche OMNI S, Version 4.0, December 2004


Contents

3.16.8 Replacing the vacuum control board ................................................................................................. 3-52

3.16.9 Replacing the Actuator Board V4/V6.................................................................................................. 3-53

3.17 Valves .............................................................................................................................. 3-54


3.17.1 Replacing the valves of the measuring chamber cartridge........................................................ 3-54

3.17.2 Replacing the valves from sample distributor cartridge and Hb cartridge .......................... 3-55

3.17.3 Replacing the waste valves V13/V18 .................................................................................................. 3-56

3.17.4 Replacing the waste valve V6 ................................................................................................................ 3-57

3.17.5 Replacing the air valve V4....................................................................................................................... 3-59

3.17.6 Replacing the FMS valve (MIX) ............................................................................................................ 3-59

3.18 Mainboard unit .............................................................................................................. 3-59


3.18.1 Replacing the Mainboard unit............................................................................................................... 3-59

3.18.2 Replacing the sample sensor board.................................................................................................... 3-60

3.19 Interface unit .................................................................................................................. 3-61


3.19.1 Replacing the Interface unit................................................................................................................... 3-61

3.19.2 Replacing the AutoQC/COOX cable ................................................................................................... 3-63

3.20 Screen/PC unit .............................................................................................................. 3-64


3.20.1 Replacing the Screen/PC Unit............................................................................................................... 3-64
3.20.2 Replacing the hard disk........................................................................................................................... 3-68

3.20.3 Replacing the floppy disk drive............................................................................................................. 3-69

3.20.4 Replacing the battery and charger ...................................................................................................... 3-70

3.20.5 Replacing the TFT backlight................................................................................................................... 3-71

3.20.6 Replacing the screen cable .................................................................................................................... 3-72

3.20.7 Software update.......................................................................................................................................... 3-73

3.21 Printer .............................................................................................................................. 3-75


3.21.1 Replacing the Printer ................................................................................................................................ 3-75
3.21.2 Replacing the winder unit....................................................................................................................... 3-76

3.21.3 Replacing the paper lid............................................................................................................................ 3-76

3.21.4 Replacing the printer-cutter lid............................................................................................................. 3-76

3.22 Fan unit ............................................................................................................................ 3-76

Service Manual, Roche OMNI S, Version 4.0, December 2004 5


Contents

3.22.1 Replacing the cooling fan unit .............................................................................................................. 3-76

3.23 Replacing the valve bus cable ................................................................................... 3-77

3.24 Replacing the actuator bus main controller cable ................................................ 3-78

4 AutoQC module
4.1 Initial installation of the AutoQC module ...................................................................4-1

4.2 Replacing the AutoQC module .....................................................................................4-2

4.3 Replacing the AQC snap lock .......................................................................................4-3

4.4 Replacing the AQC magnetic valve .............................................................................4-4

4.5 Replacing the AQC board ..............................................................................................4-5

4.6 Replacing the YZ distributor board .............................................................................4-6

4.7 Replacing the Z distributor board ................................................................................4-7

4.8 Replacing the flex cable (short) ...................................................................................4-8

4.9 Replacing the flex cable (long) ....................................................................................4-9

4.10 AQC sample tube .............................................................................................................4-9


4.10.1 Replacing the AQC sample tube complete......................................................................................... 4-9
4.10.2 Replacing the AQC sample tube short............................................................................................... 4-11

4.10.3 Replacing the AQC sample tube long ................................................................................................ 4-11

4.11 Replacing the X motor ................................................................................................. 4-12

4.12 Replacing the Y motor ................................................................................................. 4-12

4.13 Replacing the Z motor ................................................................................................. 4-13

4.14 Replacing the AutoQC steel tube .............................................................................. 4-14

4.15 Replacing the AQC temperature sensor .................................................................. 4-15

4.16 Replacing the wash port ............................................................................................. 4-16

4.17 Replacing the toothed belt (short) ........................................................................... 4-17

4.18 Replacing the toothed belt (long) ............................................................................. 4-18

5 Fluid actions

6 Service Manual, Roche OMNI S, Version 4.0, December 2004


Contents

5.1 Overview ........................................................................................................................... 5-1


5.1.1 OMNI S 5 and 6 .............................................................................................................................................5-1

5.1.2 OMNI S 3 and 4 .............................................................................................................................................5-2

5.1.3 OMNI S 1 and 2 .............................................................................................................................................5-3

5.2 Pin assignment of Packs S2 and S3 ........................................................................... 5-4


5.2.1 Pack S2..............................................................................................................................................................5-4

5.2.2 Pack S3 (OMNI S 5 and 6 only) ...............................................................................................................5-4

5.3 T&D positions ................................................................................................................... 5-5

5.4 Measurement ................................................................................................................... 5-6


5.4.1 Sample input ...................................................................................................................................................5-6

5.4.2 Sample distribution.................................................................................................................................... 5-10


5.4.3 Aspirating MSS standby solution and starting FMS ..................................................................... 5-12

5.4.4 Washing the measuring chamber vicinity......................................................................................... 5-14

5.4.5 Washing tHb/SO2 or COOX and preparing Mix1........................................................................... 5-16

5.4.6 Recalibrating BG/ISE and drying tHb/SO2 or COOX .................................................................... 5-18

5.5 Calibration ...................................................................................................................... 5-20


5.5.1 BG/ISE cleaning .......................................................................................................................................... 5-20

5.5.2 Calibrating the conductivity for calibrations solution A or B ..................................................... 5-22

5.5.3 O2 zero point calibration ......................................................................................................................... 5-24

5.5.4 Na conditioning........................................................................................................................................... 5-25

5.5.5 Mix1/Mix2 calibration, aspirating BG/ISE reference solution.................................................... 5-26

5.5.6 O2-air calibration........................................................................................................................................ 5-28


5.5.7 Aspirating MSS reference solution, calibration with standby solution .................................. 5-29

5.5.8 Cal1 Calibration........................................................................................................................................... 5-30

5.5.9 Cal2/Cal3/Cal4 Calibration ..................................................................................................................... 5-31

5.5.10 COOX calibration........................................................................................................................................ 5-32

6 Protected software functions


6.1 Protected setup ............................................................................................................... 6-1
6.1.1 Sample counter ..............................................................................................................................................6-1

Service Manual, Roche OMNI S, Version 4.0, December 2004 7


Contents

6.1.2 Baro settings .................................................................................................................................................. 6-1

6.1.3 Miscellaneous settings ............................................................................................................................... 6-1


6.1.4 Serial number................................................................................................................................................. 6-2

6.1.5 Instrument type ............................................................................................................................................. 6-2

6.1.6 Polarization voltage...................................................................................................................................... 6-2

6.1.7 COOX module (OMNI S 2, 4, 6 only)..................................................................................................... 6-3

6.1.8 AutoQC module............................................................................................................................................. 6-3

6.1.9 Log monitor .................................................................................................................................................... 6-3

6.2 Protected system functions ...........................................................................................6-4


6.2.1 AQC adjustment............................................................................................................................................ 6-4

6.3 Protected infos .................................................................................................................6-4


6.3.1 Service log file export.................................................................................................................................. 6-4

6.4 Protected DB functions ..................................................................................................6-4


6.4.1 Delete all databases .................................................................................................................................... 6-4

6.4.2 Delete databases & archives.................................................................................................................... 6-4

7 Maintenance
7.1 OMNI S 1, 3, 5 (tHb/SO2) ..............................................................................................7-1
7.1.1 Yearly replacement ...................................................................................................................................... 7-1

7.1.2 Replace every three years ......................................................................................................................... 7-2

7.2 OMNI S 2, 4, 6 (COOX) ...................................................................................................7-4


7.2.1 Yearly replacement ...................................................................................................................................... 7-4

7.2.2 Replace every three years ......................................................................................................................... 7-5

7.3 AutoQC module ................................................................................................................7-7


7.3.1 Yearly replacement ...................................................................................................................................... 7-7

8 Troubleshooting
8.1 Important Notes ...............................................................................................................8-1

8.2 System stops .....................................................................................................................8-1

8.3 Module stops ................................................................................................................. 8-15

8 Service Manual, Roche OMNI S, Version 4.0, December 2004


Contents

8.4 System warnings ........................................................................................................... 8-24

8.5 Value flags ...................................................................................................................... 8-27

8.6 Important test routines ................................................................................................ 8-53


8.6.1 Contact path test ........................................................................................................................................ 8-53

8.6.2 Aggregates ................................................................................................................................................... 8-55


8.6.3 Control sensors ........................................................................................................................................... 8-57

8.6.4 Meas. sensors.............................................................................................................................................. 8-58

8.6.5 General fluidics test................................................................................................................................... 8-63

8.6.6 Stability monitor .......................................................................................................................................... 8-75


8.6.7 General hardware test .............................................................................................................................. 8-78

8.7 Sensor limits (Sensor report) ..................................................................................... 8-82

9 Index

Service Manual, Roche OMNI S, Version 4.0, December 2004 9


Contents

10 Service Manual, Roche OMNI S, Version 4.0, December 2004


1 Introduction

1 Introduction
1.1 Symbols ............................................................................................................................. 1-1

1.2 Important notes and warnings ..................................................................................... 1-2

1.3 Disinfectant ...................................................................................................................... 1-2


1.3.1 Deproteinizer...................................................................................................................................................1-3

1.3.2 Other disinfectants........................................................................................................................................1-3

1.4 Decontamination ............................................................................................................. 1-3


1.4.1 Decontaminating the surfaces .................................................................................................................1-4

1.4.2 Screen................................................................................................................................................................1-4

1.4.3 Barcode scanner............................................................................................................................................1-4


1.4.4 Input unit ..........................................................................................................................................................1-4

1.4.5 Cleaning of the measuring chambers BG, ISE and MSS................................................................1-4

1.4.6 Tubes..................................................................................................................................................................1-4

1.5 ESD protective measures .............................................................................................. 1-5


1.5.1 Explanation of the phenomenon .............................................................................................................1-5
1.5.2 Influence of electrostatic charges on components ..........................................................................1-5

1.5.3 Why is ESD protection so important today?........................................................................................1-6

1.5.4 How can ESD protection be guaranteed..............................................................................................1-6

1.5.5 Conclusion .......................................................................................................................................................1-6

Service Manual, Roche OMNI S, Version 4.0, December 2004 1-I


1 Introduction

1-II Service Manual, Roche OMNI S, Version 4.0, December 2004


1 Introduction

1 Introduction
1.1 Symbols

Sections marked with this symbol (see Instructions for Use!) contain information
that must be observed to avoid potential injuries (to patients, users and third
parties).

Risk of infection!

All sections or passages that are marked with this symbol describe procedures
and/or indicate conditions or dangers that could damage or lead to a malfunction
in the Roche OMNI S, and which therefore should never be attempted.

Passages that are marked with this symbol warn of an immediate danger in
connection with electrical wiring or components.

All sections or passages that are marked with this symbol warn of specific dangers
in connection with static discharge. Packages that are marked with this symbol
must only be opened by trained technical staff (see section 1.5: "ESD protective
measures" on Page 1-5).

NOTE: All sections or text locations marked with "NOTE" describe safe procedures that are
intended to provide the user with additional help.

Service Manual, Roche OMNI S, Version 4.0, December 2004 1-1


1 Introduction

1.2 Important notes and warnings

• Never operate the analyzer in the vicinity of volatile or explosive gases


(e.g. anesthetic gases, etc.)!
• The device must be connected to a 2-pin, grounded socket.
• The power cable and the plug must be undamaged. Damaged power cables and
plugs must be replaced immediately.
• Before opening the back panel switch off the device and disconnect the power
cable.
• Replace damaged fuses with the specified type of fuse.

• Service and repair work must be done only as specified in this manual. Unqual-
ified service or repair work may result in warranty claims not being granted.
• Use only suitable tools and test instruments for service and repair work.
• Do not allow fluids to enter the interior of the device, because this may damage
the electronics.
• Use only slightly moistened cloths or cotton sticks to clean the device.

1.3 Disinfectant

Use only liquid disinfectant such as protein remover (Roche deproteinizer) or an


alcohol-based (about 70%) surface disinfectant.
Do not spray disinfectant directly onto the instrument because this could cause
malfunctions in the electronics.
Do not use any type of bleaching agent. Exception: Roche Deproteinizer.

Do not attempt to clean/decontaminate any part of the instrument before shutting


it down and unplugging it from the power source.
Before plugging in the device again and switching it on always wait for 15 minutes
to allow the disinfectant to evaporate.

1-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


1 Introduction

1.3.1 Deproteinizer

Composition

Aqueous NaOCI solution with active chlorine (< 2%)

Hazards identification

Due to the basic and oxidizing character of the reagent ("Deproteinizer") local irritations
after contact with eyes, skin or mucous membranes cannot be excluded.

First Aid measures

After inhalation: breath fresh air, drink large amounts of water


After skin contact: wash with generous amounts of water, remove contaminated
clothing
After eye contract: rinse eyes with generous amounts of water, contact an eye doctor
After drinking: drink large amounts of water, avoid vomiting, contact a doctor

1.3.2 Other disinfectants


Use standard alcohol-based (70%) disinfectants to clean the surface of the device. Read the
product information first.

Never use standard disinfectant to clean the tubes and tubing paths under any
circumstances!
Do not use any type of bleaching agent. Exception: Roche Deproteinizer.

1.4 Decontamination
The following procedures are intended to prevent the risk of infection (including hepatitis
and HIV) when replacing parts that were in contact with blood or other biological fluids.

Always wear gloves!

See section 1.3: "Disinfectant" on Page 1-2.

Service Manual, Roche OMNI S, Version 4.0, December 2004 1-3


1 Introduction

1.4.1 Decontaminating the surfaces


Decontaminate all exterior surfaces of the device with cloth moistened with disinfectant.
Removable panels (such as printer cover, device cover, T&D cover) must be removed for
cleaning.

1.4.2 Screen
1. Press [Analyzer] > [System] > [Wash & Clean] > [Clean screen]. The touch function
of the screen is now deactivated for 30 seconds.
2. Clean the surface of the screen with a cloth moistened with disinfectant.
3. After 30 seconds the software automatically returns to the previous menu.

1.4.3 Barcode scanner


Decontaminate the exterior surfaces of the barcode scanner with a cloth moistened with
disinfectant.

1.4.4 Input unit


Press [Analyzer] > [System] > [Wash & Clean] > [Clean input unit] and follow the
instructions on the screen.

1.4.5 Cleaning of the measuring chambers BG, I S E and M S S


1. Remove the Roche OMNI S cover.
2. Open the appropriate measuring chamber.
3. Open the electrode clamping lever in the measuring chamber and remove the
electrodes.
4. Clean the measuring chamber with a cloth moistened with disinfectant.
5. Plug the electrodes back into place and close the electrode clamping lever.
6. Close the measuring chamber and replace the cover.

1.4.6 Tubes

Use only deproteinizer to clean the tubes!

Press [Analyzer] > [System] > [Wash & Clean] > [Decontaminate tubing] and follow the
instructions on the screen.
For a detailed description of this function see the OMNI S Reference manual.

1-4 Service Manual, Roche OMNI S, Version 4.0, December 2004


1 Introduction

1.5 ES D protective measures

ESD = Electrostatic Sensitive Device

1.5.1 Explanation of the phenomenon


The most frequent cause of electrostatic discharge is friction of various materials such as
plastic, synthetic fiber, hard rubber or paper.
It can also be caused by bending or applying pressure to a material.
Discharges that are dangerous to components occur when "charged" bodies or components
do not have a discharge connection (ground) to discharge the electrostatic charge.
The discharges normally do not have a high capacity, but there are often voltage differences
in the range of several thousand volts. The discharges are perceived as small shocks or visible
sparks.
Examples:
• Shoes with rubber soles:
Friction is caused by walking. A person develops an electrical charge different from the
ground. A discharge occurs when an object (e.g. door handle) is contacted.
• Synthetic fiber clothing:
Discharge is audible and is visible in darkness.

The dryer the air the greater the risk of electrical discharge caused by friction.
Electrostatic charges are less likely to build up in humid air, particularly when
saturated with water vapor.
The probability of ESD phenomena is therefore particularly great in a northern
hemisphere winter in centrally heated rooms where the humidity is low.

1.5.2 Influence of electrostatic charges on components


If a person with an electrostatic charge touches a component, a discharge over a connection
of an IC or semiconductor element may occur. The resulting voltages may damage the com-
ponent.
Critical situations can occur while repairing or testing components if they are lying on a
more or less conductive surface (e.g. table top) and a person with an electrostatic charge
touches them. A discharge through a critical component connection can also occur in this
case.

Service Manual, Roche OMNI S, Version 4.0, December 2004 1-5


1 Introduction

1.5.3 Why is ES D protection so important today?


Formerly, current controlling semiconductors were mostly used (TTL, normal transistors,
etc.).
Today the principle of current logic is virtually the only technology in use in MOS and
CMOS components.
Currents generated by an electrostatic discharge (as much as several kV!) destroy the sensi-
tive component inputs or cause hidden damage. This damage is generally not immediately
detectable.
Another effect is that the clearances inside the ICs continuously become smaller. The inter-
nal wires become thinner and thinner, the allowable maximum input voltages become small-
er and the effects of any discharges become more and more critical.

1.5.4 How can ES D protection be guaranteed


A continuous discharge is required when working on an electronic assembly. Do this as fol-
lows:
• Use ESD wrist bands (special wrist bands connected to a protective ground).
• Repairs and tests on assemblies must only be conducted on tables with ESD mats 1 con-
nected to grounds or ESD wrist bands.
• Always pick up components at the edge (e.g. like a photo).
• Components must always be transported in ESD packages 1 or appropriate storage or
shipping containers 1 (use original packaging!).
• Shoes with rubber soles and/or clothing of synthetic fibers must not be worn in work-
shops where electronic components are repaired.
• Use a humidifier if necessary to maintain optimum humidity in the work area.
• Do not touch assemblies or components with the hand after testing.
• Always transport and ship assemblies or components for repair in ESD protective pack-
aging only. This will prevent further damage that may result in misinterpretation of the
original cause of the fault.

1.5.5 Conclusion
Of course, not all circuit boards and electronic assemblies require such careful handling. An
electrical board that only carries simple plug connectors does not require ESD packaging.
In cases of doubt always use ESD packaging.

1. The ESD mat consists of materials that have a very low, defined conductivity (1012 ohms). These materials prevent
charges caused by friction from building up and protect the component from damage.

1-6 Service Manual, Roche OMNI S, Version 4.0, December 2004


2 Revisions

2 Revisions
2.1 Service manual ................................................................................................................ 2-1
2.1.1 Revision 1.0......................................................................................................................................................2-1

2.1.2 Revision 2.0......................................................................................................................................................2-1

2.1.3 Revision 3.0......................................................................................................................................................2-3

2.1.4 Revision 4.0......................................................................................................................................................2-4

Service Manual, Roche OMNI S, Version 4.0, December 2004 2-I


2 Revisions

2-II Service Manual, Roche OMNI S, Version 4.0, December 2004


2 Revisions

2 Revisions
2.1 Service manual

2.1.1 Revision 1.0

Applicable from
Revision Nr. Date of Release Modified Chapters
Software Version
1.0 Mai 2003 First edition 1.0

2.1.2 Revision 2.0

Applicable from
Revision Nr. Date of Release Modified Chapters
Software Version
2.0 March 2004 2, 3, 4, 5, 6, 7, 8, 9 2.0

Chapter-Page Modification / Addition


Cover Page Revision Nr. and Date
Table of contents Update
2-1 Revision list
3-I - 3-VI Table of contents
3-11 Replacing the T&D linear actuators
3-12 Replacing the FMS
3-13 Replacing the FMS tubing set
3-15 Replacing the sample distributor
3-22 Replacing the MSS clip
3-25 Replacing the measuring chamber cartridge
3-28 Replacing the lamp unit bilirubin
3-30 Replacing the Polychromator Bilirubin
3-40 Replacing the pump head
3-42 Replacing the waste separator
3-45 Replacing the Transponder Control
3-46 Replacing the vacuum pump completely
3-50 Replacing the VP tube set
3-53 - 3-58 Valves
3-60 Replacing the Interface unit
3-63 - 3-69 Screen/PC unit

Service Manual, Roche OMNI S, Version 4.0, December 2004 2-1


2 Revisions

Chapter-Page Modification / Addition


3-70 Replacing the Printer
3-72 Replacing the valve bus cable
4-I Table of contents
4-1 Initial installation of the AutoQC module
4-2 Replacing the AutoQC module
4-4 Replacing the AQC magnetic valve
4-13 Replacing the AutoQC steel tube
4-14 Replacing the AQC temperature sensor
4-17 Replacing the toothed belt (long)
5-I Table of contents
5-1 - 5-34 Fluid actions
6-I Table of contents
6-1 Baro settings; Miscellaneous settings
6-3 COOX module
6-4 T&D needle calibration
6-5 Delete all databases
7-I Table of contents
7-1 Maintenance
7-4 Maintenance
8-I Table of contents
8-41 Troubleshooting
8-46 Troubleshooting
8-48 - 8-53 Important test routines
8-54 - 8-64 General fluidics test
9-1 to 9-4 Index

2-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


2 Revisions

2.1.3 Revision 3.0

Applicable from
Revision Nr. Date of Release Modified Chapters
Software Version
3.0 June 2004 2, 3, 4, 6, 8, 9 3.01

Chapter-Page Modification / Addition


Cover Page Revision Nr. and Date
Table of contents Update
2-1 Revision list
3-37 Replacing the tHb/SO 2 module
3-64 Replacing the Screen/PC Unit
3-66 Replacing the hard disk
6-1 Sample counter
6-2 Instrument type
6-3 COOX module
6-4 AVR-tools; Initialization; Shutdown PC
6-5 Simulation mode; AQC adjustment; Logmonitor-connection information
8-36 Value flag 2024 deleted
8-53 COOX (OMNI S 2, 4, 6 only)
8-54 General fluidics test
8-68 General hardware test
9-1 to 9-4 Index

Service Manual, Roche OMNI S, Version 4.0, December 2004 2-3


2 Revisions

2.1.4 Revision 4.0

Applicable from
Revision Nr. Date of Release Modified Chapters
Software Version
4.0 December 2004 1, 2, 3, 6, 7, 8, 9 4.00

Chapter-Page Modification / Addition


Cover Page Text, Revision Nr. and Date
Table of contents Update
1-2, 1-3 Disinfectants
2-4 Revision list
3-1 Note updated
3-37 tHb/SO2 module replacement
3-38 Note inserted
3-41 Replacing the pump tube
3-64 to 3-66 Screen/PC unit
3-73 Software update with FTP
6-1 Baro settings
6-4 Protected system functions, protected infos, AQC adjustment
7-1, 7-4 Yearly replacement
8 System stops 10002, 10029, 10039, 10155, 10470 new
8 System stops 10005, 10028 modified
8 Module stops 20081, 20083 new
8 System warnings 30008-30011, 30028-30034 new
8 Value flags 3022, 3033, 4003, 4008, 4024, 8083, 8084, 9001, 9002, 9003,
9004, 9005, 9006, 9008, 9009, 9010 new
8 Value flags 1020-1024, 1032, 1036, 1038, 1050, 1051, 2009, 2011, 2012,
2016, 2021, 2022, 2028, 2031, 2040, 2041, 2043, 2044, 3000, 3004, 3013,
3014, 5010-5012, 6026, 8036, 8070, 8076 modified
8-54 Aggregates: T&D module
8-57 to 8-59 Measurement sensors: BG, ISE, MSS
9-1 to 9-4 Index

2-4 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3 Roche OMNI S – Components


3.1 Important notes ............................................................................................................... 3-1

3.2 Shutdown .......................................................................................................................... 3-1


3.2.1 The OMNI S remains switched off for more than 24 hours ..........................................................3-1

3.3 Removing the rear panel ............................................................................................... 3-1

3.4 Folding up the central measuring unit ....................................................................... 3-2

3.5 T&D system (Turn & Dock) ........................................................................................... 3-2


3.5.1 Replacing the T&D module........................................................................................................................3-2

3.5.2 Replacing the fill port...................................................................................................................................3-4


3.5.3 Replacing the fill port holder ....................................................................................................................3-4

3.5.4 Replacing the needle ...................................................................................................................................3-4

3.5.5 Replacing the sample inlet path ..............................................................................................................3-5

3.5.6 Replacing the bypass nipple .....................................................................................................................3-5

3.5.7 Replacing the optical sensor SS6............................................................................................................3-5

3.5.8 Replacing the T&D sensor cover .............................................................................................................3-6

3.5.9 Replacing the T&D disk ..............................................................................................................................3-6

3.5.10 Replacing the T&D tubing set ..................................................................................................................3-7

3.5.11 Replacing the plug control ........................................................................................................................3-7


3.5.12 Replacing the T&D actuator board.........................................................................................................3-9

3.5.13 Replacing the stepper motor ....................................................................................................................3-9

3.5.14 Replacing the T&D linear actuators .................................................................................................... 3-10

3.6 Fluid mixing system (FMS) ......................................................................................... 3-11


3.6.1 Replacing the FMS..................................................................................................................................... 3-11
3.6.2 Replacing the FMS control board ........................................................................................................ 3-13

3.6.3 Replacing the FMS tubing set ............................................................................................................... 3-13

3.7 Sample distributor (SD) .............................................................................................. 3-13


3.7.1 Replacing the sample distributor cartridge...................................................................................... 3-13

3.7.2 Replacing the sample distributor ......................................................................................................... 3-15

3.7.3 Replacing the SS1, SS2, SS3 and SS4 optical sensors................................................................. 3-17

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-I


3 Roche OMNI S – Components

3.8 BG measuring chamber ............................................................................................... 3-18


3.8.1 Replacing the BG measuring chamber .............................................................................................. 3-18

3.8.2 Replacing the fan protector (BG module)......................................................................................... 3-19

3.8.3 Replacing the electrode locking lever (BG module) ..................................................................... 3-19
3.8.4 Replacing the center seal (BG module)............................................................................................. 3-19

3.9 ISE measuring chamber (OMNI S 3-6 only) ........................................................... 3-20


3.9.1 Replacing the ISE measuring chamber .............................................................................................. 3-20

3.9.2 Replacing the fan protector (ISE module)......................................................................................... 3-20

3.9.3 Replacing the electrode locking lever (ISE module) ..................................................................... 3-20

3.9.4 Replacing the center seal (ISE module)............................................................................................. 3-20

3.10 MSS measuring chamber (OMNI S 5 and 6 only) ................................................. 3-21


3.10.1 Replacing the MSS measuring chamber........................................................................................... 3-21
3.10.2 Replacing the MSS MC cover................................................................................................................ 3-22

3.10.3 Replacing the MSS clip ............................................................................................................................ 3-22

3.10.4 Replacing the fan protector (MSS module)...................................................................................... 3-23

3.10.5 Replacing the electrode locking lever (MSS module) .................................................................. 3-23

3.10.6 Replacing the center seal (MSS module).......................................................................................... 3-23

3.10.7 Replacing the MSS measuring chamber amplifier ........................................................................ 3-23

3.11 Measuring chamber cartridge ................................................................................... 3-25


3.11.1 Replacing the measuring chamber cartridge .................................................................................. 3-25

3.11.2 Replacing the MC Actor Control Board ............................................................................................. 3-26

3.12 COOX module (OMNI S 2, 4 and 6 only) ................................................................. 3-27


3.12.1 Overview ........................................................................................................................................................ 3-27

3.12.2 Replacing the lamp unit bilirubin ......................................................................................................... 3-28

3.12.3 Replacing the PolyOX-KX control board ........................................................................................... 3-29

3.12.4 Replacing the Polychromator Bilirubin............................................................................................... 3-30

3.12.5 Replacing the primary light guide........................................................................................................ 3-32

3.12.6 Replacing the cuvette holder ................................................................................................................. 3-33


3.12.7 Replacing the cuvette ............................................................................................................................... 3-34

3.12.8 Replacing the cuvette seals.................................................................................................................... 3-34

3-II Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.12.9 Replacing the tubing set tHb/PolyOX................................................................................................. 3-34

3.12.10 Replacing the hemolyzer ......................................................................................................................... 3-34


3.12.11 Replacing the Hemolyzer tube .............................................................................................................. 3-36

3.12.12 Replacing the HSIN board ...................................................................................................................... 3-36

3.13 tHb/SO2 module (OMNI S 1, 3 and 5 only) ............................................................. 3-38


3.13.1 Replacing the tHb/SO2 module ............................................................................................................ 3-38

3.13.2 Replacing the tubing set tHb/PolyOX................................................................................................. 3-38

3.13.3 Replacing the tHb/SO2 heating unit................................................................................................... 3-38


3.13.4 Replacing the Hb cartridge..................................................................................................................... 3-39

3.14 Peristaltic pumps (PP) ................................................................................................. 3-40


3.14.1 Replacing a peristaltic pump ................................................................................................................. 3-40

3.14.2 Replacing the pump head....................................................................................................................... 3-41

3.14.3 Replacing the pump tube........................................................................................................................ 3-42

3.15 Bottle compartment ...................................................................................................... 3-42


3.15.1 Replacing the bottle compartment cover.......................................................................................... 3-42
3.15.2 Replacing the gas dampers.................................................................................................................... 3-42

3.15.3 Replacing the waste separator.............................................................................................................. 3-43

3.15.4 Replacing the waste docking mechanism ........................................................................................ 3-43

3.15.5 Replacing the docking mechanism S1 (rinse solution) ............................................................... 3-44

3.15.6 Replacing the docking mechanisms S2 and S3 (fluid packs) ................................................... 3-44

3.15.7 Replacing the micro switch module.................................................................................................... 3-45


3.15.8 Replacing the bacteria filter ................................................................................................................... 3-45

3.15.9 Replacing the Transponder Control (waste, S1, S2, S3) .............................................................. 3-46

3.16 Vacuum system ............................................................................................................. 3-46


3.16.1 Replacing the vacuum pump completely.......................................................................................... 3-46

3.16.2 Replacing the vacuum pump................................................................................................................. 3-48

3.16.3 Replacing the vacuum pump head...................................................................................................... 3-49

3.16.4 Replacing the moisture trap................................................................................................................... 3-50

3.16.5 Replacing the waste tube........................................................................................................................ 3-50

3.16.6 Replacing the vacuum pump protector ............................................................................................. 3-51

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-III


3 Roche OMNI S – Components

3.16.7 Replacing the VP tube set....................................................................................................................... 3-51

3.16.8 Replacing the vacuum control board.................................................................................................. 3-52


3.16.9 Replacing the Actuator Board V4/V6.................................................................................................. 3-53

3.17 Valves .............................................................................................................................. 3-54


3.17.1 Replacing the valves of the measuring chamber cartridge........................................................ 3-54

3.17.2 Replacing the valves from sample distributor cartridge and Hb cartridge........................... 3-55

3.17.3 Replacing the waste valves V13/V18 .................................................................................................. 3-56

3.17.4 Replacing the waste valve V6 ................................................................................................................ 3-57

3.17.5 Replacing the air valve V4....................................................................................................................... 3-59

3.17.6 Replacing the FMS valve (MIX)............................................................................................................. 3-59

3.18 Mainboard unit .............................................................................................................. 3-59


3.18.1 Replacing the Mainboard unit............................................................................................................... 3-59

3.18.2 Replacing the sample sensor board.................................................................................................... 3-60

3.19 Interface unit ................................................................................................................. 3-61


3.19.1 Replacing the Interface unit ................................................................................................................... 3-61
3.19.2 Replacing the AutoQC/COOX cable.................................................................................................... 3-63

3.20 Screen/PC unit .............................................................................................................. 3-64


3.20.1 Replacing the Screen/PC Unit ............................................................................................................... 3-64

3.20.2 Replacing the hard disk ........................................................................................................................... 3-68

3.20.3 Replacing the floppy disk drive............................................................................................................. 3-69

3.20.4 Replacing the battery and charger ...................................................................................................... 3-70

3.20.5 Replacing the TFT backlight................................................................................................................... 3-71

3.20.6 Replacing the screen cable .................................................................................................................... 3-72


3.20.7 Software update.......................................................................................................................................... 3-73

3.21 Printer .............................................................................................................................. 3-75


3.21.1 Replacing the Printer................................................................................................................................. 3-75

3.21.2 Replacing the winder unit ....................................................................................................................... 3-76

3.21.3 Replacing the paper lid ............................................................................................................................ 3-76

3.21.4 Replacing the printer-cutter lid............................................................................................................. 3-76

3-IV Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.22 Fan unit ............................................................................................................................ 3-76


3.22.1 Replacing the cooling fan unit .............................................................................................................. 3-76

3.23 Replacing the valve bus cable ................................................................................... 3-77

3.24 Replacing the actuator bus main controller cable ................................................ 3-78

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-V


3 Roche OMNI S – Components

3-VI Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3 Roche OM N I S – Components
3.1 Important notes
Various components of the OMNI S, such as T&D module, tubes, waste container,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.

After components have been replaced the previous calibration parameters may have
changed and it may be necessary to input new values.
For this reason after modifications the system must be calibrated and then QC mea-
surements must be conducted to adapt the calibration parameters to the new con-
ditions.
Always use screwdrivers of suitable size.

3.2 Shutdown

Unless otherwise specified, always switch off the OMNI S before you start
modifying it!

3.2.1 The OM N I S remains switched off for more than 24 hours


Press [Analyzer] > [System] > [Utilities] > [Put out of operation]
and follow the instructions on the screen.

Use only deproteinizers to decontaminate the tubing paths!

3.3 Removing the rear panel


1. Switch off the OMNI S (note section 3.2 "Shutdown" on page 3-1).
2. Unscrew the screw holding the power supply in place and pull the power supply towards
the back.
3. Remove device cover and printer cover.
4. Unscrew the five rear-panel screws and remove the rear panel.
Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-1


3 Roche OMNI S – Components

3.4 Folding up the central measuring unit


1. Remove the device cover, printer cover and T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module").
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit",1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

1 2
1 A 2

1 2

B 3 C
4

3
Fig. 1: Central measuring unit

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, because otherwise it will be
irreparably damaged by raising the central measuring unit.

3.5 T&D system (Turn & Dock)

3.5.1 Replacing the T&D module


1. Switch off the OMNI S (note section 3.2 "Shutdown" on page 3-1).
2. Remove the device cover and T&D cover of the OMNI S.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module").
4. Carefully pull the sample inlet path out to the right and remove it.

3-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

5. Turn the T&D lock back and pull the tubing strip on the right side of the T&D module
out of its retainer.

2 AutoQC tube

1
2
Tubing strip

T&D lock
Sample inlet path

Fig. 2: T&D module

6. Unplug the AutoQC tube from the tube connection.

AutoQC tube

Fig. 3: T&D module - AutoQC tube

7. Unscrew the three screws fastening the T&D module (see Fig. 2: "T&D module", 1, 2, 3).

Do not remove the fastening screws completely, because they are retained in
place by plastic rings to prevent loss.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-3


3 Roche OMNI S – Components

8. Pull the T&D module forward and unplug the actuator bus cable.

Actuator bus cable

Fig. 4: T&D module - actuator bus cable

9. Remove T&D module.


Use the reverse order for assembly.

3.5.2 Replacing the fill port


1. Remove T&D cover.
2. Rotate fill port holder to the right and remove it.
Use the reverse order for assembly.

3.5.3 Replacing the fill port holder


1. Remove T&D cover.
2. Rotate fill port holder to the right and remove it.
3. Remove the fill port from the fill port holder.
Use the reverse order for assembly.

3.5.4 Replacing the needle


1. Remove T&D cover.
2. Rotate fill port holder to the right and remove it.
3. Rotate needle to the right and remove it.
Use the reverse order for assembly.

After replacing the needle start a T&D needle calibration (Component test > Aggre-
gates > T&D module).

3-4 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.5.5 Replacing the sample inlet path


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Remove the device cover and T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
Use the reverse order for assembly.

After replacing the sample inlet path start a T&D needle calibration (Component
test > Aggregates > T&D module).

3.5.6 Replacing the bypass nipple


1. Remove the OMNI S cover.
2. Slide the bypass holder down (see Fig. 18: "Bypass nipple" on page 3-14).
3. Pull the bypass nipple up and disconnect it (see Fig. 18: "Bypass nipple" on page 3-14).
Use the reverse order for assembly.

3.5.7 Replacing the optical sensor S S6


1. Disassemble T&D module as directed in chapter 3.5.1 "Replacing the T&D module"
on page 3-2.
2. Remove the sensor cover on the left side of the T&D module.
3. Hold down the locking button of the sensor plug and pull the sensor plug out.

Sensor plug Sensor SS6 Sensor cover


Fig. 5: Sensor SS6

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-5


3 Roche OMNI S – Components

4. Slide the optical sensor out and remove it.


Use the reverse order for assembly.

Note the position of the optical sensor during installation: there are three possible
positions and it must be replaced in the one from which it was removed.

3.5.8 Replacing the T&D sensor cover


1. Remove the sample inlet path as directed in chapter 3.5.5 "Replacing the sample inlet
path" on page 3-5.
2. Remove the sensor cover (see Fig. 5: "Sensor SS6" on page 3-5).
Use the reverse order for assembly.

3.5.9 Replacing the T&D disk


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.

Activate the function program


[Analyzer] > [System] > [Component test] > [T&D module]
and press [P1] to put the T&D disk in the position needed for replacement.

1. Remove T&D cover.


2. Rotate fill port holder to the right and remove it.

Fill port

fill port holder

Fig. 6: fill port holder

3. Now use the fill port holder as a tool:


Insert the fill port holder with the flat side into the slot in the T&D disk and turn it 90
degrees to the right or left. Hold the T&D disk in place during this process.
4. Remove T&D disk.
Use the reverse order for assembly.

3-6 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.5.10 Replacing the T&D tubing set


1. Remove the T&D disk as directed in chapter 3.5.9 "Replacing the T&D disk".
2. Pull the T&D moisture trap forward.

Moisture trap 2

Tubing set

T&D disk

Tubing strip

Fig. 7: T&D tubing set

3. Pull the tubing strip on the right side of the T&D module upwards from its holder.
4. Unplug the AutoQC tube from the tube connection.
5. Pull the T&D tubing set forward and remove it.
Use the reverse order for assembly.

Make sure that the tubes are correctly positioned in the holder guides when install-
ing the new T&D tubing set.

3.5.11 Replacing the plug control


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Disassemble T&D module as directed in chapter 3.5.1 "Replacing the T&D module" on
page 3-2.
2. Remove T&D disk as directed in 3.5.9 "Replacing the T&D disk" on page 3-6.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-7


3 Roche OMNI S – Components

3. Rotate needle to the right with the needle holder and remove it.

needle

needle holder

Fig. 8: needle

4. Unplug the SCD (Sample Container Detection) cable on the actuator board and slide it
out of the holder.

SCD cable

Fig. 9: T&D module - SCD cable

5. Unscrew the three screws holding the housing cover at the bottom of the T&D module
and remove the cover.

1 2 Housing cover

Fig. 10: T&D module - housing cover on the bottom

3-8 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

6. Unscrew the four screws on the back of the plug control.

2
1 2

3
4

Fig. 11: T&D module - plug control fixing screws

7. Pull the plug control out to the front.


Use the reverse order for assembly.

3.5.12 Replacing the T&D actuator board

Electrostatically sensitive components!

1. Disassemble T&D module as directed in chapter 3.5.1 "Replacing the T&D module" on
page 3-2.
2. Disconnect all cables and the optical sensor from the actuator board.
3. Unscrew the three screws holding the actuator boards and remove the actuator board.
Use the reverse order for assembly.

Note the correct positions of the various plug connectors on the cables during
installation.

3.5.13 Replacing the stepper motor

Electrostatically sensitive components!

1. Disassemble T&D module as directed in chapter 3.5.1 "Replacing the T&D module" on
page 3-2.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-9


3 Roche OMNI S – Components

2. Disconnect the two plugs (labeled POS and TURN) from the actuator board.

Linear actuator connectors


2

"POS" and "TURN" connectors

Fig. 12: T&D module - Actuator board

3. Unscrew the screws (stainless steel) holding the stepper motor and remove the stepper
motor (see Fig. 13: "T&D stepper motor", 1, 2).

2
T&D stepper motor
1

Fig. 13: T&D stepper motor

Use the reverse order for assembly.

3.5.14 Replacing the T&D linear actuators

Electrostatically sensitive components!

1. Disassemble T&D module as directed in chapter 3.5.1 "Replacing the T&D module" on
page 3-2.
2. Remove T&D disk as directed in chapter 3.5.9 "Replacing the T&D disk".
3. Disconnect the linear actuator plug on the actuator board.
4. Unscrew the three screws holding the housing cover at the bottom of the T&D module
and remove the cover.
5. Unscrew the two screws holding the linear actuator.

3-10 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

6. Unscrew the spindle nut of the linear actuator with a 4.5 mm wrench and remove it
together with the washer.

Spindle nuts

Linear actuators
Fig. 14: T&D module - linear actuators

7. Pull the linear actuator back and remove it.


8. Remove the spindle cover and support plate from the linear actuator and fasten to the
new one.
Use the reverse order for assembly.

Before assembling the new linear actuator, unscrew the spindle 1 cm on the motor
side to ensure that the front end of the spindle does not press against the tube of the
sample distributor cartridge during installation.

3.6 Fluid mixing system (FM S)

3.6.1 Replacing the FM S

Electrostatically sensitive components!

1. Remove the rear panel of the device as directed in chapter 3.3 "Removing the rear panel"
on page 3-1.
2. Remove T&D cover.
3. Unscrew the four screws holding the right side panel and remove the side panel.

Do not remove the fastening screws completely, because they are secured against
loss.

4. If an AutoQC module (optional) is installed in the OMNI S, remove it to allow free


access to the FMS (from the back of the device).

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-11


3 Roche OMNI S – Components

5. Unscrew the two screws holding the plug protective plate of the control board and
remove the protective plate (see Fig. 15: "FMS control board", 1, 2).

1 Cable connectors
- red/green/blue
- yellow/red

Fig. 15: FMS control board

6. Disconnect the two FMS cable connectors on the board


(see Fig. 15: "FMS control board").
7. Disconnect the two cable connectors (ground) from the FMS actuators.
8. Open the bottle compartment cover and remove all bottles or packs from the OMNI S.
9. Disconnect all tube connections exiting from the FMS (see Fig. 16: "FMS unit", A, B, C).

1 2

B 3 C
4
Fig. 16: FMS unit

10. Unscrew the four screws (see Fig. 16: "FMS unit", 1, 2, 3, 4) holding the FMS and pull
the FMS forward.

Do not remove the fastening screws completely, because they are secured against
loss.

Use the reverse order for assembly.

Test the functioning of the FMS valves (V19 and VM) under
[Analyzer] > [System] > [Component test] > [Valves] > [FM S].

3-12 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.6.2 Replacing the FM S control board

Electrostatically sensitive components!

1. Remove the rear panel as directed in chapter 3.3 "Removing the rear panel" on page 3-1.
2. If an AutoQC module (optional) is installed in the OMNI S, remove it to allow free
access to the FMS (from the back of the device).
3. Unscrew the two screws holding the plug protection plate of the actuator board and
remove the plug protection plate (see Fig. 15: "FMS control board" on page 3-12).
4. Pull out all plug connectors on the FMS control board.
5. Unscrew the two screws holding the FMS control board and remove the board.
Use the reverse order for assembly.

3.6.3 Replacing the FM S tubing set

Activate the function program


[Analyzer] > [System] > [Utilities] > [Maintenance] > [Change FM S tubing]
and follow the instructions on the screen.

3.7 Sample distributor (SD)

3.7.1 Replacing the sample distributor cartridge

Activate the function program


[Analyzer] > [System] > [Utilities] > [Maintenance]
to close all valve actuators.

1. Remove the device cover and T&D cover of the OMNI S.


2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-13


3 Roche OMNI S – Components

4. Disconnect the two tubes that run from the sample distributor cartridge to the
measuring chamber cartridge.

Linear clamp

Tubes to measuring chamber cartridge

5
Bypass nipple
6

Sample distributor liquid connection

Fig. 17: Sample distributor cartridge

5. Open the locking lever (Plexiglas cover) of the MSS input pump and slide the linear clip
(white plastic piece) upwards.
6. Slide the bypass holder down and raise the bypass nipple.

1 2
Bypass nipple

B Bypass holder
C
3
4

Fig. 18: Bypass nipple

3-14 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

7. Unscrew the six screws (see Fig. 17: "Sample distributor cartridge", 1, 2, 3, 4, 5, 6) hold-
ing the sample distributor cartridge and pull the sample distributor cartridge up and to
the right.

Do not remove the fastening screws completely, because they are secured against
loss.

Use the reverse order for assembly.

Make sure that there is no contamination between heating plate and sample dis-
tributor cartridge.

3.7.2 Replacing the sample distributor

Electrostatically sensitive components!

1. Disassemble the sample distributor cartridge as directed in section 3.7.1 "Replacing the
sample distributor cartridge" on page 3-13.
2. Remove printer cover.
3. Unscrew the three screws holding the central measuring unit (see Fig. 1: "Central mea-
suring unit",1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

4. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, because otherwise it will be
irreparably damaged by raising the central measuring unit.

5. Disassemble the OMNI S fan unit as directed in 3.22.1 "Replacing the cooling fan unit"
on page 3-76.
6. Unscrew the two screws holding the mainboard unit (see Fig. 57: "Mainboard unit" on
page 3-60).
7. Fold the mainboard unit down.
8. Disconnect the two MSS measuring chamber connector cables on the mainboard and
remove the MSS measuring chamber as directed in section 3.10.1 "Replacing the MSS
measuring chamber" on page 3-21.
9. Remove the ISE measuring chamber as directed in section 3.9.1 "Replacing the ISE mea-
suring chamber" on page 3-20.
10. Remove the BG measuring chamber as directed in section 3.8.1 "Replacing the BG mea-
suring chamber" on page 3-18.
11. Remove the Hb cartridge as directed in section 3.13.4 "Replacing the Hb cartridge" on
page 3-39.
12. Disconnect the two connector cables of the sample distributor (name: PELT-QK, VB)
on the mainboard.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-15


3 Roche OMNI S – Components

13. Disconnect the two capacitor mainboard connecting wires on the sample distributor.
14. Disconnect the sample detection board from the mainboard.

1 A 2
Sample detection board
1 2

B 3 C
4

Fig. 19: Sample detection board

15. Pull the liquid connector of the sample distributor down


(see Fig. 17: "Sample distributor cartridge" on page 3-14).
16. Unscrew the four screws holding the sample distributor
(Fig. 20: "Fastening screws of the sample distributor", 1, 2, 3, 4).

1 1
A 2

1 2

B 3 C
4

3
4

Fig. 20: Fastening screws of the sample distributor

17. Pull the sample distributor unit forward and remove it.
Use the reverse order for assembly.

Make sure that there is no contamination between heating plate and sample
distributor cartridge.
When installing the BG, ISE and MSS modules, make sure not to confuse the
modules and plug-in connectors!
During assembly, ensure the correct installation of the fluid seal between lateral
channel and pump frame!

3-16 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.7.3 Replacing the S S1, S S2, S S3 and S S4 optical sensors

Electrostatically sensitive components!

1. Remove the device cover, printer cover and T&D cover of the OMNI S.
2. SS1, SS2, SS4: disassemble the sample distributor cartridge as directed in section 3.7.1
"Replacing the sample distributor cartridge" on page 3-13.
SS3: disassemble the Hb cartridge as directed in section 3.13.4 "Replacing the Hb car-
tridge" on page 3-39.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit",1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

7. Disassemble the OMNI S fan unit as directed in 3.22.1 "Replacing the cooling fan unit"
on page 3-76.
8. Unscrew the two screws holding the mainboard unit (see Fig. 57: "Mainboard unit" on
page 3-60).
9. Fold the mainboard unit down.
10. Disconnect the sensor on the sample detection board.
11. Optical sensor SS1: Disassemble the mainboard unit according to section 3.18.1
"Replacing the Mainboard unit" on page 3-59.
Unscrew the two screws holding the VBI valve (see Fig. 53:
"Sample distributor valves and Hb cartridge valves" on
page 3-55) and remove the valve. Now slide the sensor out and
remove it.
optical sensor SS2: Disassemble the mainboard unit according to section 3.18.1
"Replacing the Mainboard unit" on page 3-59.
Unscrew the two screws holding the V22 valve (see Fig. 53:
"Sample distributor valves and Hb cartridge valves" on
page 3-55) and remove the valve. Now slide the sensor out and
remove it.
optical sensor SS3: Disassemble the mainboard unit according to section 3.18.1
"Replacing the Mainboard unit" on page 3-59. Unscrew the
two screws holding the V2 valve (see Fig. 53: "Sample distrib-
utor valves and Hb cartridge valves" on page 3-55) and remove
the valve. Now slide the sensor out and remove it.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-17


3 Roche OMNI S – Components

optical sensor SS4: Unscrew the two screws holding the VSI valve (see Fig. 53:
"Sample distributor valves and Hb cartridge valves" on
page 3-55) and remove the valve. Now slide the sensor out and
remove it.
Use the reverse order for assembly.

3.8 BG measuring chamber

3.8.1 Replacing the BG measuring chamber


1. Disassemble the sample distributor cartridge as directed in section 3.7.1 "Replacing the
sample distributor cartridge" on page 3-13.
2. Open the BG measuring chamber, remove the electrodes and the electrode locking lever
and pull out the tube to the measuring chamber cartridge on the left side.
3. Close the measuring chamber and unscrew the three screws holding the BG module
(see Fig. 21: "Screws holding the BG and ISE modules", 1, 2, 3).

1 1 2
A 2
6
1 2

4
5
B 3 C
4

1
2

Fig. 21: Screws holding the BG and ISE modules

4. Unscrew the three screws holding the ISE module (see Fig. 21: "Screws holding the BG
and ISE modules", 4, 5, 6).
5. Lift the ISE module and pull the BG module forward.
6. Disconnect the mainboard plugs from the BG module and remove the BG module.
Use the reverse order for assembly.

Make sure that there is no contamination between heating plate and sample
distributor cartridge.

3-18 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.8.2 Replacing the fan protector (BG module)

1 1 2
A 2
Fan protector

1 2

B 3 C
4

Fig. 22: Fan protector

1. Disassemble the BG module as directed in 3.8.1 "Replacing the BG measuring chamber"


on page 3-18.
2. Release the four hooks at the end of the fan protector.
3. Lever the head of the NTC cable out of the fan protector with a screwdriver and slide
the cable out.
4. Remove the fan protector.
Use the reverse order for assembly.

3.8.3 Replacing the electrode locking lever (BG module)


1. Remove the OMNI S cover, open the BG measuring chamber and remove the electrodes.
2. Half open the electrode locking lever, hold it in place with both hands and pull it up.
3. Disconnect the tube and remove the electrode locking lever.
Use the reverse order for assembly.

3.8.4 Replacing the center seal (BG module)


1. Remove the OMNI S cover and open the BG measuring chamber.
2. Open the electrode locking lever.
3. Remove the electrodes.
4. Remove the center seal (blue plastic ring on the right interior panel) with tweezers.
Use the reverse order for assembly.

Do not use tools to install the new center seal, otherwise it may be damaged.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-19


3 Roche OMNI S – Components

3.9 I SE measuring chamber (OM N I S 3-6 only)

3.9.1 Replacing the I SE measuring chamber


1. Disassemble the sample distributor cartridge as directed in 3.7.1 "Replacing the sample
distributor cartridge" on Page 3-13.
2. Open the ISE measuring chamber, remove the electrodes and the electrode locking lever
and pull out the tube to the measuring chamber cartridge on the left side.
3. Close measuring chamber and unscrew the three screws holding the ISE module
(see Fig. 21: "Screws holding the BG and ISE modules" on page 3-18, 4, 5, 6).
4. Disconnect the tube adapter to the MSS outlet pump.
5. Unscrew the five screws holding the MSS module.
6. Lift the MSS module and pull the ISE module forward.
7. Remove the board plug from the ISE module and remove the ISE module.
Use the reverse order for assembly.

Make sure that there is no contamination between heating plate and sample dis-
tributor cartridge.

3.9.2 Replacing the fan protector (I S E module)


1. Disassemble the ISE module as directed in 3.9.1 "Replacing the ISE measuring chamber"
on page 3-20.
2. Loosen the four hooks at the ends of the fan protector
(see Fig. 22: "Fan protector" on page 3-19).
3. Release the head of the NTC cable from the fan protector with a screwdriver and slide it
out.
4. Remove fan protector.
Use the reverse order for assembly.

3.9.3 Replacing the electrode locking lever (I S E module)


1. Remove the OMNI S cover and open the ISE measuring chamber.
2. Half open the electrode locking lever, hold it in place with both hands and pull it up.
3. Disconnect the tube and remove the electrode locking lever.
Use the reverse order for assembly.

3.9.4 Replacing the center seal (I S E module)


1. Remove the OMNI S cover and open the ISE measuring chamber.
2. Open the electrode locking lever.
3. Remove the electrodes.

3-20 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

4. Remove the center seal (blue plastic ring on the right interior panel) with tweezers.
Use the reverse order for assembly.

Do not use tools to install the new center seal, otherwise it may be damaged.

3.10 M SS measuring chamber (OM N I S 5 and 6 only)

3.10.1 Replacing the M S S measuring chamber

Electrostatically sensitive components!

1. Disassemble the sample distributor cartridge as directed in 3.7.1 "Replacing the sample
distributor cartridge" on Page 3-13.
2. Open the locking lever (Plexiglas cover) of the MSS outlet pump and slide the linear
clamp (white plastic part) up (see Fig. 39: "Peristaltic pump" on page 3-40).
3. Pull the tubes out of the tube guide and remove the tube adapter from the MSS module.

MSS outlet pump

Linear clamp

Tube adapter

Tube guide

Fig. 23: Tube adapter

4. Open the MSS measuring chamber, remove the electrodes and the MSS cassette and pull
out the tube on the left side.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-21


3 Roche OMNI S – Components

7. Disconnect the two MSS connecting cables on the mainboard.


8. Unscrew the five screws (internal hex) holding the MSS module and remove the module
(see Fig. 24: "MSS module screws", 1, 2, 3, 4, 5).

1
2

B C
A

4
3
5

Fig. 24: MSS module screws

Use the reverse order for assembly.

Make sure that there is no contamination between heating plate and sample
distributor cartridge.

3.10.2 Replacing the M S S MC cover


1. Remove the OMNI S cover and open the MSS measuring chamber.
2. Pull the MC cover upward.
Use the reverse order for assembly.

3.10.3 Replacing the M S S clip

Electrostatically sensitive components!

1. Remove the OMNI S cover and open the MSS measuring chamber.
2. Unscrew the screws marked 1, 2, A, B, and C of the MSS module and open the module
(see Fig. 24: "MSS module screws" on page 3-22).

Do not remove the fastening screws completely, because they are secured against
loss.

3. Remove the MSS clip.


Use the reverse order for assembly.

3-22 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.10.4 Replacing the fan protector (M S S module)


1. Disassemble the MSS module as directed in section 3.10.1 "Replacing the MSS measur-
ing chamber" on page 3-21.
2. Loosen the four hooks at the ends of the fan protector (see Fig. 22: "Fan protector" on
page 3-19).
3. Release the head of the NTC cable from the fan protector with a screwdriver and slide it
out.
4. Remove the fan protector.
Use the reverse order for assembly.

3.10.5 Replacing the electrode locking lever (M S S module)


1. Remove the OMNI S cover and open the MSS measuring chamber.
2. Half open the electrode locking lever, hold it in place with both hands and pull it up.
3. Disconnect the tube and remove the electrode locking lever.
Use the reverse order for assembly.

3.10.6 Replacing the center seal (M S S module)


1. Remove the OMNI S cover and open the MSS measuring chamber.
2. Open the electrode locking lever.
3. Remove electrodes and MSS-cartridge.
4. Remove the center seal (blue plastic ring on the right interior panel) with tweezers.
Use the reverse order for assembly.

Do not use tools to install the new center seal, otherwise it may be damaged.

3.10.7 Replacing the M S S measuring chamber amplifier

Electrostatically sensitive components!

1. Remove the device cover and the T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-23


3 Roche OMNI S – Components

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disassemble the fan shield as directed in 3.22.1 "Replacing the cooling fan unit" on page
3-76.
7. Disconnect the MSS connector cable on the mainboard.
8. Unscrew the screws marked 1, 2, A, B, and C of the MSS module and open the module
(see Fig. 24: "MSS module screws" on page 3-22).
9. Remove the rocker switch, the MSS cover and the amplifier board of the MSS measuring
chamber from above and pull the amplifier out to the back.
Use the reverse order for assembly.

3-24 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.11 Measuring chamber cartridge

3.11.1 Replacing the measuring chamber cartridge

Activate the function program


[Analyzer] > [System] > [Service] > [Replacing the MC cartridge]
to pull in all valve actuators.

1. Remove device cover.


2. Disconnect all tubes on the measuring chamber cartridge
(see Fig. 25: "Measuring chamber cartridge", A, B, C).

Tension lever

Linear clamp

1
2
A
MSS outlet pump
B

3 D
Main pump

4 C
E

Fig. 25: Measuring chamber cartridge

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-25


3 Roche OMNI S – Components

3. Disconnect tube adapter to the bottle compartment, the connector tube to the waste
separator and the connector tube to V4 on the T-piece (see Fig. 26: "MC tube adapter").

Tube adapter

Connector tubes

T-piece

Fig. 26: MC tube adapter

4. Open the BG measuring chamber and disconnect the measuring chamber outlet tube
(see Fig. 25: "Measuring chamber cartridge", E).
5. Open the ISE measuring chamber and disconnect the measuring chamber outlet tube
(see Fig. 25: "Measuring chamber cartridge", D).
6. Open the locking lever (Plexiglas cover) of the main pump and the MSS outlet pump
and slide the linear clamp (white plastic part) up (see Fig. 39: "Peristaltic pump" on
page 3-40).
7. Remove the tube adapter from the MSS module.
8. Disconnect the two tubes that run from the sample distributor cartridge to the measur-
ing chamber cartridge (see Fig. 17: "Sample distributor cartridge" on page 3-14).
9. Unscrew the five screws holding the measuring chamber cartridge
(see Fig. 25: "Measuring chamber cartridge", 1, 2, 3, 4, 5).

Do not remove the fastening screws completely, because they are secured against
loss.

10. Pull the measuring chamber cartridge up and remove it.


Use the reverse order for assembly.

3.11.2 Replacing the MC Actor Control Board

Electrostatically sensitive components!

1. Remove the device cover and the T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.

3-26 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

4. Unscrew the three screws holding the central measuring unit


(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disconnect all valves of the measuring cartridge and the valve bus cable on the MC
Actor Control Board.
7. Lever the MC Actor Control Board from the snap fastener and remove it.
Use the reverse order for assembly.

3.12 COOX module (OM N I S 2, 4 and 6 only)

3.12.1 Overview
1 A 2
Lamp unit Cuvette holder
1 2
Halogen lamp black
Filters
B C Waste
3
Lens Apertures
4 grey
Beamsplitter

Prism
Reference lamp Light guides Cuvette
Photodiode

Holographic
grating
Results Hemolyzer
tHb, SO2, O2Hb, HHb, Sample
COHb, MetHb, Bili

Diode array (CCD) Flat cable3


PolyOX-KX-
Polychromator control 9700 LY3

Fig. 27: COOX module: Overview

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-27


3 Roche OMNI S – Components

3.12.2 Replacing the lamp unit bilirubin

Electrostatically sensitive components!

1. Remove the rear panel of the device as directed in section 3.3 "Removing the rear panel"
on page 3-1.
2. Remove the T&D cover of the OMNI S.
3. Unscrew the four screws holding the right side panel and remove the side panel.

Do not remove the fastening screws completely, because they are secured against
loss.

4. Loosen the two cable clamp screws on the PolyOX-KX control board.

1 1 2
A 2

1 2
Cable clamp
B 3 C
4

Fig. 28: PolyOX-KX control board - cable clamp

5. Disconnect the power supply cable (identification: J7) and the two lamp plugs on the
lamp unit (identification: J1, J2) on the board.
6. Open the bayonet closure of the primary light guide and pull it off
(see Fig. 29: "Lamp unit").

Bayonet closure

Holding screw

Fig. 29: Lamp unit

3-28 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

7. Unscrew the screw holding the lamp unit (see Fig. 29: "Lamp unit").
8. Release the cable from the flat cable clip.
9. Remove the lamp unit.
Use the reverse order for assembly.

3.12.3 Replacing the PolyOX-KX control board

Electrostatically sensitive components!

1. Remove the rear panel of the device as directed in section 3.3 "Removing the rear panel"
on page 3-1.
2. Remove the T&D cover of the OMNI S.
3. Unscrew the four screws holding the right side panel and remove the side panel.

Do not remove the fastening screws completely, because they are secured against
loss.

4. Loosen the two cable clamp screws.


5. Remove all plug connectors (identification: J1, J2, J3, J7, J16) and the ground from the
PolyOX-KX control board.
6. Unscrew the ten screws holding the PolyOX-KX control board.

1 2
3

10 4

9 5

8 7 6
Fig. 30: PolyOX-KX Control Board

7. Pull the PolyOX-KX control board out from the side and remove it.
Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-29


3 Roche OMNI S – Components

3.12.4 Replacing the Polychromator Bilirubin

After replacing the polychromator bilirubin, a code must be entered in the protec-
tive settings (see chapter 6 "Protected software functions", section 6.1.7 "COOX
module (OMNI S 2, 4, 6 only)")! For this purpose, record the 16 digit code located
on the polychromator bilirubin!

1. Remove the rear panel of the device as directed in section 3.3 "Removing the rear panel"
on page 3-1.
2. Unscrew the four screws holding the right side panel and remove the side panel.

Do not remove the fastening screws completely, because they are secured against
loss.

3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).
6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

7. Pull the two light guides out of the clamp.

Light guide clamp

Fig. 31: Light guide clamp

8. Disconnect the two tubes from the cuvette holder (see Fig. 32: "Cuvette holder", A, B)
and open the cuvette holder.

3-30 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

9. Unscrew the two screws holding the cuvette holder and pull out the cuvette holder to
the front (see Fig. 32: "Cuvette holder", 1, 2).

1 2

A B

Fig. 32: Cuvette holder

10. Disconnect the light guide (gray) leading to the polychromator from the cuvette holder
and slide it out to the back.
11. Pull the light guide leading to the polychromator out of the clamp beside the polychro-
mator and slide it out to the right.

Light guide clamp

Fig. 33: Clamp beside polychromator

12. Open the bayonet closure of the primary light guide on the lamp unit and remove it
(see Fig. 29: "Lamp unit" on page 3-28).
13. Unscrew the two screws holding the light guide conduit and pull the light guide conduit
out to the back (see Fig. 35: "Light guide conduit", 1, 2).
14. Remove the plugs from the power supply (from the PolyOX-KX control board, name:
J16) and ground (from frame).
15. Unscrew the three screws holding the COOX module and remove the COOX module.
16. Disconnect the flat cable from the polychromator bilirubin.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-31


3 Roche OMNI S – Components

17. Unscrew the two screws at the bottom of the polychromator bilirubin and remove it.

Polychromator Bilirubin

Fig. 34: Polychromator Bilirubin

Use the reverse order for assembly.

Make sure that the bayonet connector is properly locked when reconnecting the
light guide!

3.12.5 Replacing the primary light guide


1. Remove the rear panel of the device as directed in section 3.3 "Removing the rear panel"
on page 3-1.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Pull the primary light guide and the light guide for bilirubin out of the clamp
(see Fig. 31: "Light guide clamp" on page 3-30).
7. Disconnect the two tubes from the cuvette holder (see Fig. 32: "Cuvette holder", A, B)
and open the cuvette holder.
8. Unscrew the two screws holding the cuvette holder and pull out the cuvette holder to
the front (see Fig. 32: "Cuvette holder", 1, 2).

3-32 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

9. Disconnect the primary light guide and the light guide for bilirubin from the cuvette
holder and slide them out to the back.

Never bend the primary light guide, because this may damage it.

10. Open the bayonet closure of the primary light guide plug and remove it.
11. Unscrew the two screws holding the light guide conduit and pull the light guide conduit
out to the back (see Fig. 35: "Light guide conduit", 1, 2).

Primary light guide

Light guide conduit

1 2

A B
1 2

Fig. 35: Light guide conduit

Use the reverse order for assembly.

3.12.6 Replacing the cuvette holder


1. Remove the device cover, the printer cover and the T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Remove the two light guides from the clamp (see Fig. 31: "Light guide clamp"
on page 3-30).
7. Disconnect the two tubes (A and B) from the cuvette holder (see Fig. 32: "Cuvette
holder" on page 3-31) and open the cuvette holder.
8. Unscrew the two screws (1 and 2) holding the cuvette holder and pull the cuvette
holder out to the front (see Fig. 32: "Cuvette holder" on page 3-31).
9. Disconnect the two light guides and the flat cable from the cuvette holder and slide it
out to the back.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-33


3 Roche OMNI S – Components

10. Disconnect the cuvette holder ground cable from the frame.
11. Remove the cuvette holder.
Use the reverse order for assembly.

3.12.7 Replacing the cuvette


1. Remove the OMNI S cover and open the cuvette input.
2. Open the retainer and remove the cuvette.
Use the reverse order for assembly.
After replacing the cuvette, perform a COOX calibration using tHb calibrator or a blood
sample whose tHb values are exactly known
(see Instructions for Use section 6.4.3 "COOX calibration"):
3. Perform a COOX calibration and enter "Yes" at "Cuvette replaced".
4. Apply the value „tHb (measured) ---“ by pressing „Accept“.
5. Perform another COOX calibration and enter "No" at "Cuvette replaced".
The COOX-Modul will be ready for measurement after the second calibration.

3.12.8 Replacing the cuvette seals


1. Remove the OMNI S cover and open the cuvette input.
2. Open the retainer and remove the cuvette.
3. Remove the two cuvette seals with tweezers.
Use the reverse order for assembly.

3.12.9 Replacing the tubing set tHb/ PolyOX


1. Remove the OMNI S cover.
2. Disconnect the left and right tubes from the hemolyzer.
Use the reverse order for assembly.

3.12.10 Replacing the hemolyzer


1. Remove the device cover, the printer cover and the T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

3-34 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disconnect the hemolyzer plug from the HSIN board.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-35


3 Roche OMNI S – Components

7. Unscrew the two screws (internal hex) holding the hemolyzer head and remove it.

2 B
Hemolyzer
1
A

Fig. 36: Hemolyzer

8. Unscrew the two screws (A, B) holding the hemolyzer and pull it out to the back.
Use the reverse order for assembly.

When installing the hemolyzer head tighten the two screws alternately to prevent
distortion!
Do not reverse the two tubes on the hemolyzer when reconnecting them!

3.12.11 Replacing the Hemolyzer tube


1. Remove the OMNI S cover.
2. Disconnect the tube on the hemolyzer (out) and at the cuvette input and remove it.
Use the reverse order for assembly.

3.12.12 Replacing the H S I N board

Electrostatically sensitive components!

1. Remove the device cover, the printer cover and the T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

3-36 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disconnect the wide plug and the hemolyzer plug from the HSIN board.
7. Disconnect the flat cable from the HSIN board.

Actuator bus cable plug

1 2

HSIN board
Fig. 37: Hemolyzer - HSIN board

8. Remove the plug from the actuator board and unscrew the two mating component
screws (see Fig. 37: "Hemolyzer - HSIN board", 1, 2).
9. Pull the HSIN board out to the back and remove it.
Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-37


3 Roche OMNI S – Components

3.13 tHb/ SO 2 module (OM NI S 1, 3 and 5 only)

3.13.1 Replacing the tHb/ SO 2 module

Don’t use a tHb/SO 2 module with revision 02 with a software version < 3.00 . For
software version < 3.00 use a tHb/SO 2 module with revision 00 or revision 01.
These revisions are also compatible with sotware versions 3.00 !

1. Remove the device cover, T&D cover and printer cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disconnect the tube connector from the tHb/SO 2 module.


7. Unscrew the two screws of the left tube connecting piece and keep them for later
installation of the new tHb/SO 2 module.
8. Disconnect the connector cable to the mainboard on the tHb/SO 2 module.
9. Unscrew the two screws holding the tHb/SO 2 module.
10. Pull the tHb/SO 2 module forward and remove it.
Use the reverse order for assembly.

3.13.2 Replacing the tubing set tHb/ PolyOX


1. Remove the OMNI S cover.
2. Disconnect the left and right tubes from the tHb/SO2 module.
Use the reverse order for assembly.

3.13.3 Replacing the tHb/ SO 2 heating unit


1. Disconnect the tHb/SO 2 module as directed in section 3.13.1 "Replacing the tHb/SO2
module".
2. Disconnect and remove the heating unit.
Use the reverse order for assembly.

3-38 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

When assembling the tHb/SO 2 heating unit, make sure it is placed correctly in order
to avoid possible damage.

3.13.4 Replacing the Hb cartridge

Activate the function program


[Analyzer] > [System] > [Service] > [Replacing the Hb cartridge]
to position all valve actuators correctly.

1. Remove the device cover and T&D cover of the OMNI S.


2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Slide the bypass holder down and fold the bypass nipple up
(see Fig. 18: "Bypass nipple" on page 3-14).
4. Disconnect the bypass nipple, the connector tube to the tHb/SO 2 module
(or hemolyzer) and the waste tube (on the waste separator).
5. Unscrew the two screws holding the Hb cartridge and remove the cartridge
(see Fig. 38: "Hb cartridge", 1, 2).

Fig. 38: Hb cartridge

Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-39


3 Roche OMNI S – Components

3.14 Peristaltic pumps (PP)

3.14.1 Replacing a peristaltic pump


1. Switch off the OMNI S (see section 3.2 "Shutdown" on page 3-1.)
2. Remove the device cover and printer cover of the OMNI S.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

7. Disassemble the fan shield as directed in 3.22.1 "Replacing the cooling fan unit"
on page 3-76.
8. Remove the plug on the back of the pump.
9. Open the locking lever (Plexiglas cover) of the peristaltic pump and slide the linear
clamp (white plastic part) up (see Fig. 39: "Peristaltic pump").

Locking lever
1 2

Linear clamp

3 4
Fig. 39: Peristaltic pump

10. Remove the pump tube.


11. Unscrew the four screws holding the peristaltic pump, slide it up and pull it out for-
wards (see Fig. 39: "Peristaltic pump", 1, 2, 3, 4).
Use the reverse order for assembly.

When positioning the pump tube around the pump head, make sure that the tubes
are not crossed.

3-40 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

Test the function of the peristaltic pump with


[System] > [Component test] > [Aggregates] > [Peristaltic pumps].

3.14.2 Replacing the pump head


1. Remove the OMNI S cover.
2. Open the locking lever (Plexiglas cover) of the peristaltic pump and slide the linear
clamp (white plastic part) up (see Fig. 39: "Peristaltic pump").
3. Remove the pump tube.
4. Loosen the setscrew on the pump head with a 1.5 mm Allen wrench.

Setscrew

Fig. 40: Pump head

5. Pull the pump head up.


6. Connect the new pump head and fasten it with the setscrew. Note that the setscrew must
grip the flat side of the shaft (remove blue cap).

Flat side of shaft

Fig. 41: Pump head shaft

7. Position the pump tube around the pump head. Note that the tubes must not be
crossed.
The remaining assembly is done in reverse order.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-41


3 Roche OMNI S – Components

3.14.3 Replacing the pump tube


1. Remove the OMNI S cover.
2. Open the locking lever (Plexiglas cover) of the peristaltic pump and slide the linear
clamp (white plastic part) up (see Fig. 39: "Peristaltic pump").
3. Remove the old pump tube.
4. Position the new pump tube around the pump head in the correct way. Note that the
tubes must not be crossed.
The remaining assembly is done in reverse order.

3.15 Bottle compartment

3.15.1 Replacing the bottle compartment cover


1. Open the bottle compartment cover.
2. Disconnect the two cables (left and right).
3. Lift the two front fork heads of the gas dampers (left and right) from the retaining pins
of the bottle compartment cover with a screwdriver (see Fig. 42: "Gas damper").

Back fork head

Front fork head

Fig. 42: Gas damper

4. Remove the bottle compartment cover.


Use the reverse order for assembly.

3.15.2 Replacing the gas dampers


1. Open the bottle compartment cover.
2. Left gas damper: unscrew the five screws holding the left side panel of the OMNI S and
remove the side panel.

Do not remove the fastening screws completely, because they are secured against
loss.

Right gas damper: unscrew the four screws holding the right side panel and remove the
side panel.

3-42 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

Do not remove the fastening screws completely, because they are secured against
loss.

3. Lift the front fork head of the respective gas damper from the retaining pin of the bottle
compartment cover with a screwdriver (see Fig. 42: "Gas damper").
4. Lift the back fork head of the respective gas damper from the retaining pin of the
OMNI S device frame with a screwdriver (see Fig. 42: "Gas damper").
5. Remove the respective gas damper.
Use the reverse order for assembly.

3.15.3 Replacing the waste separator


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Remove the OMNI S cover.
2. Disconnect all tubes from the waste separator.
3. Rotate the waste separator about a quarter turn counterclockwise and remove it.
4. Remove the waste container sensor from the waste separator.

When removing the sensor make sure that the light pipe is not bent.

Use the reverse order for assembly.

3.15.4 Replacing the waste docking mechanism


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Remove the OMNI S cover.
2. Open the bottle compartment cover and the waste docking mechanism.
3. Pull the waste bottle out to the front.
4. Pull the waste tube out of the waste separator (see Fig. 43: "Waste docking mecha-
nism").

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-43


3 Roche OMNI S – Components

5. Unscrew the screw holding the waste docking mechanism (see Fig. 43: "Waste docking
mechanism").

waste separator

waste tube

Holding screw

Fig. 43: Waste docking mechanism

6. Pull the waste docking mechanism 1 cm forward and then pull it down.
7. Pull the waste tube out of the waste docking mechanism.
Use the reverse order for assembly.

3.15.5 Replacing the docking mechanism S1 (rinse solution)


1. Remove the OMNI S cover.
2. Open the bottle compartment cover and the S1 docking mechanism.
3. Pull S1 rinse solution forward.
4. Unscrew the screw holding the S1 docking mechanism.
5. Pull the S1 docking mechanism 1 cm forward and then pull it down.
6. Pull the rinse tube off the nipple and remove the S1 docking mechanism.
Use the reverse order for assembly.

3.15.6 Replacing the docking mechanisms S2 and S3 (fluid packs)


1. Remove the device cover and T&D cover of the OMNI S.
2. Open bottle compartment cover and S2 (or S3) docking mechanism.
3. Pull S2 (or S3) fluid pack forward.
4. Unscrew the screw holding the S2 (or S3) docking mechanism.

Holding screw

Fig. 44: S2/S3 docking mechanism

3-44 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

5. Open the hook inside the top cover of the S2 (or S3) docking mechanism and pull it
out.
6. Disconnect the tube connector and remove the S2 (or S3) docking mechanism.
Use the reverse order for assembly.

Make sure that the tube connector is in the correct position during installation!

3.15.7 Replacing the micro switch module


1. Remove the OMNI S cover.
2. Open the bottle compartment cover.
3. Unscrew the screw holding the micro switch module.

Holding screw

Microswitch

Fig. 45: Microswitch

4. Pull micro switch module forward.


5. Disconnect the two plug connectors and remove the micro switch module.
Use the reverse order for assembly.

3.15.8 Replacing the bacteria filter


1. Open the bottle compartment cover.
2. Slide the bacteria filter left or right and pull it forward.

Bacteria filter

Fig. 46: Bacteria filter

Use the reverse order for assembly.

Install the bacteria filter with the "Inlet" label pointing to the left.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-45


3 Roche OMNI S – Components

3.15.9 Replacing the Transponder Control (waste, S1, S2, S3)


1. Remove the rear panel of the device as directed in section 3.3 "Removing the rear panel"
on page 3-1.
2. Pull the AutoQC drawer out and remove the ampoule block.
3. For S1 rinse only: close AutoQC drawer and unscrew the two screws holding the
transponder control cover plate.
4. For S3 fluid pack transponder control only: pull the waste exhaust tube from the guide;
unscrew the two screws holding the plug shield of the FMS control board and remove
the plug shield (see Fig. 15: "FMS control board", 1, 2); unscrew the two screws holding
the FMS control board and slide the board to the side.
5. Disconnect the cable from the transponder control.
6. Unscrew the screw holding the transponder control and remove the transponder
control.
Use the reverse order for assembly.

3.16 Vacuum system

3.16.1 Replacing the vacuum pump completely


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Switch off Roche OMNI S and open bottle compartment cover.
2. Remove the device cover and T&D cover of the OMNI S.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

3-46 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

7. Unscrew the five screws holding the left side panel of the OMNI S and remove the side
panel.

Do not remove the fastening screws completely, because they are secured against
loss.

8. Disconnect the actuator bus cable from the vacuum pump.

Cable clip

Actuator bus cable

Fig. 47: Vacuum pump - actuator bus cable

9. Hold down the lock button of the waste sensor plug and pull it out.
10. Disconnect the exhaust tube.
11. Open the cable clip and disconnect the waste tube at the T-piece and slide it out.

Waste tube

T-piece

Fig. 48: waste tube

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-47


3 Roche OMNI S – Components

12. Unscrew the three screws holding the vacuum pump (see Fig. 49: "Vacuum pump", 1, 2,
3) and remove the vacuum pump.

Exhaust tube Moisture trap


Pump motor cable VPP sensor

Fig. 49: Vacuum pump

Use the reverse order for assembly.

3.16.2 Replacing the vacuum pump


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Remove the vacuum pump completely as directed in section 3.16.1 "Replacing the vac-
uum pump completely" on page 3-46.
2. Disconnect the waste tube on the inlet side of the pump unit
(see Fig. 49: "Vacuum pump").
3. Disconnect the pump motor cable from the vacuum control board
(see Fig. 49: "Vacuum pump").

3-48 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

4. Unscrew the screw holding the vacuum control board and slide it to the side.

Vacuum control board

Holding screw

Fig. 50: Vacuum control board

5. Unscrew the three screws holding the pump unit and remove it.
6. Unscrew the three low-frequency bearings of the pump unit and fasten them to the new
pump unit.
The remaining assembly is done in reverse order.

3.16.3 Replacing the vacuum pump head


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Remove the vacuum pump completely as directed in section 3.16.1 "Replacing the vac-
uum pump completely" on page 3-46.
2. Disconnect the waste tube on the inlet side of the pump unit.
3. Unscrew the four screws holding the pump head unit with a torx screwdriver and
remove it.
Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-49


3 Roche OMNI S – Components

3.16.4 Replacing the moisture trap


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Unscrew the five screws holding the left side panel of the OMNI S and remove the side
panel.

Do not remove the fastening screws completely, because they are secured against
loss.

2. Pull the moisture trap from the holder and disconnect the two tubes
(see Fig. 49: "Vacuum pump" to page 3-48).
3. Remove the moisture trap.
Use the reverse order for assembly.

3.16.5 Replacing the waste tube


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Remove the waste docking mechanism as directed in chapter 3.15.4 "Replacing the waste
docking mechanism" on page 3-43.
2. Pull the waste tube from the waste docking mechanism and remove it.
Use the reverse order for assembly.

3-50 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.16.6 Replacing the vacuum pump protector


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Unscrew the five screws holding the left side panel of the OMNI S and remove the side
panel.

Do not remove the fastening screws completely, because they are secured against
loss.

2. Disconnect the two tubes and the sensor on the vacuum pump protector
(see Fig. 49: "Vacuum pump" on page 3-48).
3. Loosen the two screws holding the vacuum pump protector by unscrewing them out
about 1 cm.
4. Pull the vacuum pump protector up and remove it.
Use the reverse order for assembly.

3.16.7 Replacing the VP tube set


Various components of the OMNI S, such as T&D module, tubes, waste bottle,
waste separator etc., contain liquid biological residues after use that may cause
infection.
Handle these components with care and avoid contact with skin. Always wear
gloves when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.
1. Switch off Roche OMNI S and open bottle compartment cover.
2. Remove the device cover and T&D cover of the OMNI S.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-51


3 Roche OMNI S – Components

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

7. Unscrew the five screws holding the left side panel of the OMNI S and remove the side
panel.

Do not remove the fastening screws completely, because they are secured against
loss.

8. For this purpose, record the barcode located on the polychromator bilirubin!
(see Fig. 43: "Waste docking mechanism" on page 3-44).
9. Open the cable clamp, disconnect the waste tube at the T piece, and unthread it
(see Fig. 48: "waste tube" on page 3-47).
10. Disconnect the waste tube at the vacuum pump protector and remove it.

Waste Tube

Air tube

Vakuum pump protector

Fig. 51: VP tube set

11. Disconnect the air tube at the vacuum pump protector and at the vacuum pump and
remove it.
Use the reverse order for assembly.

3.16.8 Replacing the vacuum control board

Electrostatically sensitive components!

1. Remove the device cover and T&D cover of the OMNI S.


2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.

3-52 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

4. Unscrew the three screws holding the central measuring unit


(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Remove all connectors from the vacuum control board.


7. Unscrew the screw holding the vacuum control board and remove it
(see Fig. 50: "Vacuum control board").
Use the reverse order for assembly.

3.16.9 Replacing the Actuator Board V4/V6


1. Remove device cover and T&D cover.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disconnect the connectors from the V4/V6 actuator board.


7. Unscrew the screw holding the V4/V6 actuator board and remove the board.
Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-53


3 Roche OMNI S – Components

3.17 Valves

3.17.1 Replacing the valves of the measuring chamber cartridge


These instructions apply to the following valves: V3, V5, V8, V9, V12, V14, V16,
V17, VIO, VBO

V12 V17

V8 V16

V14
V9

VIO

V5

VBO V3

Fig. 52: Valves of the measuring chamber cartridge

1. Switch off the OMNI S (note section 3.2 "Shutdown" on page 3-1).
2. Remove the device cover and T&D cover of the OMNI S.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

3-54 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

7. Disconnect the valve cable from the slot.


8. Unscrew the two screws holding the respective valve.
9. Pull the valve back and remove it.
Use the reverse order for assembly.

3.17.2 Replacing the valves from sample distributor cartridge and Hb cartridge
These instructions apply to the following valves: V10, V15, V20, V21, V22, VBI, VII, VSI,
V2, V23, V24

V10 V15

V20

VSI

VII

V21

VBI

V22
V24
V23
V2

Fig. 53: Sample distributor valves and Hb cartridge valves

1. Switch off the OMNI S (note section 3.2 "Shutdown" on page 3-1).
2. Remove the OMNI S cover.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-55


3 Roche OMNI S – Components

3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

7. Unscrew the two screws holding the mainboard retainer


(see Fig. 57: "Mainboard unit", 1, 2).
8. Fold the mainboard retainer down.
9. Disassemble the fan shield as directed in 3.22.1 "Replacing the cooling fan unit"
on page 3-76.
10. Disconnect the valve bus cable of the respective valve from the slot.
11. Unscrew the two screws holding the respective valve.
12. Pull the valve back and remove it.
13. Insert the valve spindle of the new valve manually.
Use the reverse order for assembly.

3.17.3 Replacing the waste valves V13/V18

Attention: V13 and V18 are different parts:


V13 = valve undamped
V18 = VP waste valve

V18 (VP waste valve)

V13 (valve undamped)

Fig. 54: Waste valves V13 and V18

1. Remove the vacuum pump completely as directed in section 3.16.1 "Replacing the vac-
uum pump completely" on page 3-46.

3-56 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

2. Disconnect the two tubes on the vacuum pump protector (see Fig. 49: "Vacuum pump"
on page 3-48).
3. Loosen the two screws holding the vacuum pump protector by unscrewing them out
about 1 cm.
4. Pull the vacuum pump protector up and remove it.
5. Pull the tube out of the waste valve V13 or V18 (see Fig. 54: "Waste valves V13 and
V18").
6. Disconnect the power supply cable of the valve V13 or V18 on the board.
7. Unscrew the two screws holding the waste valve V13 or V18 and remove the valve.
8. Remove the MC valve centering device and fasten it to the new valve.
Use the reverse order for assembly.

3.17.4 Replacing the waste valve V6


1. Switch off the OMNI S (note section 3.2 "Shutdown" on page 3-1).
2. Disassemble the waste docking mechanism as directed in section 3.15.4 "Replacing the
waste docking mechanism" on page 3-43.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

7. Unscrew the five screws holding the left side panel of the OMNI S and remove the side
panel.

Do not remove the fastening screws completely, because they are secured against
loss.

8. Disconnect the connectors from the V4/V6 actuator board.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-57


3 Roche OMNI S – Components

9. Unscrew the screw holding the V4/V6 actuator board and slide the board to the side.

Relief clamp

Actuator Board V4/V6

Holding screw

Fig. 55: Actuator board

10. Disconnect the waste tube from the waste docking mechanism and pull it out of the
valve (see Fig. 56: "waste valve V6").
11. Unscrew the two screws holding the V6 waste valve (see Fig. 56: "waste valve V6", 1, 2).

waste tubing
1 2 V6

Fig. 56: waste valve V6

12. Pull the V6 waste valve back and remove it.


Use the reverse order for assembly.

3-58 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.17.5 Replacing the air valve V4

Attention: V4 is a separate valve called "valve OMNI S", do not use the valve called
"valve unit", it is different!

1. Switch off the OMNI S (note section 3.2 "Shutdown" on page 3-1).
2. Remove device cover and T&D cover.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.
5. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

7. Disconnect cable of V4 from the V4/V6 actuator board.


8. Loosen the relief clamp and remove the tubes.
9. Carefully pry out valve V4 by using a flat-head screw driver.
Use the reverse order for assembly.

3.17.6 Replacing the FM S valve (M IX)


1. Disassemble the FMS unit as directed section 3.6.1 "Replacing the FMS" on page 3-11.
2. Unscrew the two screws holding the MIX valve and remove the valve.
Use the reverse order for assembly.

3.18 Mainboard unit

3.18.1 Replacing the Mainboard unit

Electrostatically sensitive components!

1. Switch off the OMNI S (note section 3.2 "Shutdown" on page 3-1).
2. Remove the device cover and T&D cover of the OMNI S.
3. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
4. Carefully pull the sample inlet path out to the right and remove it.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-59


3 Roche OMNI S – Components

5. Unscrew the three screws holding the central measuring unit


(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

6. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

7. Unscrew the two screws holding the mainboard unit


(see Fig. 57: "Mainboard unit", 1, 2).

1
2

Fig. 57: Mainboard unit

8. Fold the mainboard unit down.


9. Remove all plug connectors on the mainboard.
10. Disconnect all cable connectors fixed to the mainboard with plugs on the modules.

The BG and ISE module plug connectors on the mainboard are permanently
attached: do not pull them off!

11. Press the mainboard unit to the left or right to release it from the retainer pins.
12. Remove the mainboard unit.
Use the reverse order for assembly.

3.18.2 Replacing the sample sensor board


See section 3.18.1 "Replacing the Mainboard unit" number 1 to 8.
Use the reverse order for assembly.

3-60 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.19 Interface unit

3.19.1 Replacing the Interface unit

Electrostatically sensitive components!

1. Remove the rear panel of the device as directed in section 3.3 "Removing the rear panel"
on page 3-1.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disconnect the connector cable of the AutoQC module and the COOX module at the
interface board.

AutoQC module
connector cable

COOX module
connector cable

Fig. 58: Interface connections

7. Disconnect the screen cable on the left interior of the interface unit
(see Fig. 70: "Fixing axis of the swivel arm" on page 3-72).

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-61


3 Roche OMNI S – Components

8. Disconnect the interface unit connector cable on the mainboard.

Interface unit
connector cable

Fig. 59: Interface unit connector cable

9. Disassemble the printer as directed in section 3.21.1 "Replacing the Printer"


on page 3-75.
10. Unscrew the four screws holding the interface unit (see Fig. 60: "Interface unit").

2
3
4
1
4

Fig. 60: Interface unit

Grasp the interface unit while unscrewing the mounting screws to prevent it
from sliding off!

11. Pull the interface unit to the back and remove it.
Use the reverse order for assembly.

During installation, ensure that the cable connections and the orientation of the
printer connection cable (see Fig. 72: "Printer connector cable") are correct!

3-62 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.19.2 Replacing the AutoQC /COOX cable


1. Remove the rear panel of the device as directed in section 3.3 "Removing the rear panel"
on page 3-1.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disconnect the respective cable on the interface unit


(see Fig. 58: "Interface connections" on page 3-61).
7. For AutoQC cable only: remove the AutoQC drawer; open the bayonet closure of the
primary light guide connector and remove it; unscrew the two screws holding the light
guide conduit and pull the light guide conduit out through the back (see Fig. 35: "Light
guide conduit" on page 3-33, 1, 2).
8. Disconnect the cables on the COOX and AutoQC control board, pull them out and
remove them.
Use the reverse order for assembly.

During installation, ensure that the cable connections are correct!

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-63


3 Roche OMNI S – Components

3.20 Screen/PC unit

3.20.1 Replacing the Screen/ PC Unit

Before replacing the screen/PC unit, all data must be backed up!
The settings data can be imported and exported.

After replacing the screen/PC unit, a code must be entered in the protective settings
(OMNI 2, 4, 6 only)(see chapter 6 "Protected software functions", section 6.1.7
"COOX module (OMNI S 2, 4, 6 only)")! For this purpose, record the 16 digit code
located on the polychromator bilirubin (see Fig. 62: "Polychromator bilirubin" on
page 3-67)!

1. Export the Setup data: [Setup] > [Load / save setup] > [Export].
2. Print the Status report:
[Info] > [Miscellaneous reports] > [Status report] > [Print status report].
3. Backup all databases:
[Data manager] > [Utilities] > [Backup / Restore] > [Backup now]
(see Reference Manual Section 3.2.7 "Utilities").
4. Mark the last entry in the Backup/Restore list and copy it onto floppy disk
(see Reference Manual Section 3.2.7 "Utilities").
5. Switch off the OMNI S (note section 3.2 "Shutdown" on page 3-1).
6. Disconnect the connector cable on the screen and slide it out.
7. Unscrew the nut fixing the screen at the bottom of the swivel arm with a 13 mm wrench.

Fixing nut
2
3

Brake packet
4

Fig. 61: Fixing the screen

8. Remove the brake packet.


9. Remove the screen.
Use the reverse order for assembly.

The following settings must be performed after replacing the screen/PC unit:

3-64 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

For OM N I S 1, 3, 5

• Check the software version ([Info] > [Software versions]) and perform a software
update, if necessary ([Analyzer] > [System] > [Utilities] > [Communication] >
[SW upgrade]) (see section "Software update with FTP" on page 3-73, section "Soft-
ware update with floppy disks" on page 3-73 and Reference Manual chapter 3 "Software
modes", section "SW upgrade").
• Check date and time.
• Assign the instrument type and load the default setup (see chapter 6 "Protected
software functions", section 6.1.5 "Instrument type"):
- In the menu [Setup] > [Protected Setup] > [Instrument type],
select the instrument type.
- Press the [Load default setup] key to load the default settings of the instrument
stored on the harddisk.
- Press the [Save and exit] key.
• Import the previously exported Setup data: [Setup] > [Load / save setup] > [Import].
• Print the Status report:
[Info] > [Miscellaneous reports] > [Status report] > [Print status report],
compare it to the previously printed report and adjust settings if necessary.
• Check the following settings and adjust if necessary:
- [Setup] > [Interfaces]: Network, COM 1 and ASTM.
- [Setup] > [Security]: Instrument functions, User management and
Profile management.
- [Setup] > [Parameters] > [Misc. Settings]: Activated for calibrations, Activated for
measurements, QC lock, QC warning, Multirules.
- [Setup] > [Display & Reports] > [Calibration] > [Cal. report]: Sensor report,
Parameter report.
- [Setup] > [Protected setup] > [Miscellaneous settings] > [Cal. intervals U.S.].
• Restore the backup data from the floppy disk:
[Data manager] > [Utilities] > [Backup / Restore] > [Copy from disk]
(see Reference Manual Section 3.2.7 "Utilities").
• Restore the databases:
[Data manager] > [Utilities] > [Backup / Restore] > [Restore archive]
(see Reference Manual Section 3.2.7 "Utilities").
• Reinsert the wastewater bottle, the S1 Rinse Solution bottle, the S2 Fluid Pack and the
S3 Fluid Pack (OMNI S 5 only).
• Insert a new MSS sensor (OMNI S 5 only).
• Reread the barcodes of the electrodes or insert new electrodes.
• Check the barometer settings (see chapter 6 "Protected software functions",
section 6.1.2 "Baro settings").
• Check the polarization voltage for MSS (for OMNI S 5 only) and P O 2 and re-enter the
default values, if necessary (see chapter 6 "Protected software functions", section 6.1.6
"Polarization voltage").
• Enter the serial number of the OMNI S (see chapter 6 "Protected software functions",
section 6.1.4 "Serial number") - it is located on the rear panel of the OMNI S.
• Perform the QC material setup: [Setup] > [QC material].

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-65


3 Roche OMNI S – Components

For OM N I S 2, 4, 6

• Check the software version ([Info] > [Software versions]) and perform a software
update, if necessary ([Analyzer] > [System] > [Utilities] > [Communication] >
[SW upgrade]) (see section "Software update with FTP" on page 3-73, section "Soft-
ware update with floppy disks" on page 3-73 and Reference Manual chapter 3 "Software
modes", section "SW upgrade").
• Check date and time.
• Assign the instrument type and load the default setup (see chapter 6 "Protected
software functions", section 6.1.5 "Instrument type"):
- In the menu [Setup] > [Protected Setup] > [Instrument type],
select the instrument type.
- Press the [Load default setup] key to load the default settings of the instrument
stored on the harddisk.
- Press the [Save and exit] key.
• Import the previously exported Setup data: [Setup] > [Load / save setup] > [Import].
• Print the Status report:
[Info] > [Miscellaneous reports] > [Status report] > [Print status report],
compare it to the previously printed report and adjust settings if necessary.
• Check the following settings and adjust if necessary:
- [Setup] > [Interfaces]: Network, COM 1 and ASTM.
- [Setup] > [Security]: Instrument functions, User management and
Profile management.
- [Setup] > [Parameters] > [Misc. Settings]: Activated for calibrations, Activated for
measurements, QC lock, QC warning, Multirules.
- [Setup] > [Display & Reports] > [Calibration] > [Cal. report]: Sensor report,
Parameter report.
- [Setup] > [Protected setup] > [Miscellaneous settings] > [Cal. intervals U.S.].
• Restore the backup data from the floppy disk:
[Data manager] > [Utilities] > [Backup / Restore] > [Copy from disk]
(see Reference Manual Section 3.2.7 "Utilities").
• Restore the databases:
[Data manager] > [Utilities] > [Backup / Restore] > [Restore archive]
(see Reference Manual Section 3.2.7 "Utilities").
• Reinsert the wastewater bottle, the S1 Rinse Solution bottle, the S2 Fluid Pack and the
S3 Fluid Pack (OMNI S 6 only).
• Insert a new MSS sensor (OMNI S 6 only).
• Reread the barcodes of the electrodes or insert new electrodes.
• Check the barometer settings (see chapter 6 "Protected software functions",
section 6.1.2 "Baro settings").
• Check the polarization voltage for MSS (for OMNI S 6 only) and P O 2 and re-enter the
default values, if necessary (see chapter 6 "Protected software functions",
section 6.1.6 "Polarization voltage").
• Perform the COOX adjustments again:
- Remove the printer cover and the T&D cover.
- Open the container compartment.
- Unscrew the 4 mounting screws of the right side wall and remove the side wall.

3-66 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

- Read and record the 16 digit code located on the polychromator bilirubin
(see Fig. 62: "Polychromator bilirubin").

16 digit code

Fig. 62: Polychromator bilirubin

- In the menu [Setup] > [Protected Setup] > [COOX module] >
[COOX adjustments], enter the 16 digit code by means of the keyboard symbol
(see chapter 6 "Protected software functions", section 6.1.7 "COOX module
(OMNI S 2, 4, 6 only)").
• Perform a calibration of the polychromator ([Analyzer] > [System] > [Calibrations]).
• Perform a COOX calibration ([Analyzer] > [System] > [Calibrations]).
• Enter the serial number of the OMNI S (see chapter 6 "Protected software functions",
section 6.1.4 "Serial number") - it is located on the rear panel of the OMNI S.
• Perform the QC material setup: [Setup] > [QC material].

Tip: Select a measurement report after loading the default setup and starting the PC. As
default, the measurement report "example" will be selected, if available. If this is not
available, create and select a new measurement report.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-67


3 Roche OMNI S – Components

3.20.2 Replacing the hard disk

Electrostatically sensitive components!

Don’t build a hard disk from a screen/PC unit with the order number 03313476001
("old" version 2.x for instruments with a serial number < 1500) into a screen/PC
unit with the order number 04364015001 ("new" version 3.x for instruments with
a serial number > 1500)! These hard disks are not compatibly!

After replacing the hard disk, a code must be entered in the protective settings
(OMNI S 2, 4, 6 only) (see chapter 6 "Protected software functions", section 6.1.7
"COOX module (OMNI S 2, 4, 6 only)")! For this purpose, record the 16 digit code
located on the polychromator bilirubin (see Fig. 62: "Polychromator bilirubin" on
page 3-67)!

1. Disassemble the screen as directed in section 3.20.1 "Replacing the Screen/PC Unit"
on page 3-64.
2. Push the screen joint to the vertical position.

Screen joint

Fig. 63: Screen joint

3. Unscrew the eight screws on the back of the screen.


4. Remove the rear panel of the screen case and disconnect the speaker cable
(see Fig. 68: "Screen" on page 3-71).

3-68 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

5. Unscrew the two screws holding the hard disk.

1
2
3
Hard disk

4
Connector cable

Fig. 64: Hard disk

6. Disconnect the hard disk connector cable and remove the hard disk.
Use the reverse order for assembly.

After replacing the hard disk, perform the same settings as described in section
3.20.1 "Replacing the Screen/PC Unit" on page 3-64!

3.20.3 Replacing the floppy disk drive

Electrostatically sensitive components!

1. Disassemble the screen as directed in section 3.20.1 "Replacing the Screen/PC Unit".
2. Push the screen joint to the vertical position
(see Fig. 63: "Screen joint" on page 3-68).
3. Unscrew the eight screws on the back of the screen.
4. Remove the rear panel of the screen case and disconnect the speaker cable
(see Fig. 68: "Screen" on page 3-71).
5. Unscrew the two screws holding the floppy disk drive.

1 Floppy disk drive


2
3

Connector cable
4

2
Fig. 65: Floppy disk drive

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-69


3 Roche OMNI S – Components

6. Disconnect the floppy disk drive connector cable and remove the floppy disk drive.
Use the reverse order for assembly.

3.20.4 Replacing the battery and charger

Electrostatically sensitive components!

1. Disassemble the screen as directed in section 3.20.1 "Replacing the Screen/PC Unit".
2. Push the screen joint to the vertical position
(see Fig. 63: "Screen joint" on page 3-68).
3. Unscrew the eight screws on the back of the screen.
4. Remove the rear panel of the screen case and disconnect the speaker cable
(see Fig. 68: "Screen" on page 3-71).
5. Unscrew the four screws holding the battery and charger (be careful of the plastic
washers).

4
1
2
3
Battery and charger

2 3 4

Fig. 66: Battery and charger

6. Remove battery and charger.


Use the reverse order for assembly.

Be careful of the plastic washers during installation!

When installing the battery and charger circuit in a screen/PC unit with a serial
number < 1500, it is necessary to affix an aluminum block (thermal cooling inter-
face) to the housing!
• Peel of the film of the aluminum block by using tweezers (the thicker side of the
SILPAD) and position it as shown in Fig. 67: "Alu Block" (the thicker side of the
SILPAD toward the metal plate).

Peel off the foil on one side of the aluminum block only!

3-70 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

Alu Block
Fig. 67: Alu Block

3.20.5 Replacing the TFT backlight


1. Disassemble the screen as directed in section 3.20.1 "Replacing the Screen/PC Unit".
2. Push the screen joint to the vertical position
(see Fig. 63: "Screen joint" on page 3-68).
3. Unscrew the eight screws on the back of the screen.
4. Remove the rear panel of the screen case and disconnect the speaker cable
(see Fig. 68: "Screen" on page 3-71).
5. Disconnect light connector and remove it (see Fig. 68: "Screen" on page 3-71).
6. Unlock the touch plug and disconnect it (see Fig. 68: "Screen" on page 3-71).
7. Unscrew the six screws holding the shield.

2 3 4
Speaker cable

Touch plug

Light plug
1 6 5
Fig. 68: Screen

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-71


3 Roche OMNI S – Components

8. Lift the display unit out of the plastic frame.


9. Disconnect the light unit cable.
10. Release the left and right locks on the screen (push inwards) and remove the light unit.

2 Locks

Light unit Light unit


Fig. 69: Light unit

Use the reverse order for assembly.

The touch plug must be pushed in sufficiently far during reinstallation and must
not be bent!

3.20.6 Replacing the screen cable


1. Remove the rear panel as directed in chapter 3.3 "Removing the rear panel" on page 3-1.
2. Disconnect the screen cable at the interface unit and at the screen.
3. Unscrew the screw holding the swivel arm with an Allen wrench (see Fig. 70: "Fixing
axis of the swivel arm" on page 3-72) and remove the swivel arm-screen unit.

Fixing axis of the swivel arm

Screen cable

Fig. 70: Fixing axis of the swivel arm

4. Unscrew the two screws holding the screen cable on the bottom of the swivel arm.
5. Open the two clips and remove the screen cable.
6. Fasten the two clips to the new screen cable.
Use the reverse order for assembly.

3-72 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.20.7 Software update

Software update with floppy disks

For the software update with floppy disks, the four system disks OMNI S are necessary.
1. Choose [Analyzer] > [System] > [Utilities] > [Shutdown PC].
2. Turn of the instrument after shutting down and wait at least 3 minutes.
3. Insert disk no. 1 in the disk drive and switch on the instrument.
4. The software update will be started. Follow the instructions on the screen−insert the
next floppy disk when needed and touch the screen.
5. Remove floppy disk 4 after the update was finished. The Roche OMNI S automatically
warms up, performs a calibration and is ready again.

Software update with FTP

The software update needs a FTP server software that can be purchased separately
in specialized trade. For this software, Roche cannot grant any support.

Apply the network settings on the computer: On Roche CODEX COE 3.1-PCs, they are
located in the Start Menu, (Settings >) Control Panel > Network Connections:
1. Select the network connection that will be used for connecting the computer and the
OMNI S.
2. Select Properties in the context menu (right-click with the mouse).
3. In the network protocol area, select "Internet Protocol (TCP/IP)" and click to "Proper-
ties".
Tip: Note the current settings to restore them after the software update.
4. Enter the following settings:
IP address: 192.168.0.100 Subnet mask: 255.255.255.0
No further settings are required.
5. Install the FTP server software on the computer (standard PC oder Laptop with ether-
net network connection) according to the instructions of the manufacturer.
Apply the network settings on the OMNI S:
6. Select [Settings] > [Interfaces] > [Network].
7. Enter the following settings:
IP address: 192.168.0.101 Net mask: 255.255.255.0
No further settings are required.
Press the "Back" button to save this settings.
Connect the computer and the OMNI S:
8. Connect the computer and the OMNI S using a crossover RJ45 network cable (available
in specialized trade).
Tip: Direct network connections between PC and OMNI S are only possible by using a
crossover network cable.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-73


3 Roche OMNI S – Components

Tip: Verify proper connection using the command PING on the computer.
Therefore, select Start - Run and enter CMD (or COMMAND).
Enter PING 192.168.0.101 in the command prompt-a proper connection will respond
with "Reply from 192.168.0.101 ...". When "Request timed out." is shown, inspect the
network settings and the cabling.
9. Run the FTP server software on the computer.
The software update package must be placed into the FTP folder (root folder of the
server) or a subfolder.
Tip: Details on the settings of the FTP server can be found in the documentation of the
manufacturer.
Start the software-upgrade on the OMNI S:
10. Select [Analyzer] > [System] > [Utilities] > [Communication] > [SW-Upgrade].
11. Apply the settings according to the settings of the computer:
Server: 192.168.0.100 User: anonymous
Path: / Password: update
When using the anonymous access, the serial number of the OMNI S will be automati-
cally used as password. Possibly the anonymous access of the FTP server is deactivated
and protected with user name and password. When the software package is not in the
root folder of the ftp server, the path has to be adjusted accordingly.
12. Press [Upgrade]. The software-upgrade will be started.
13. After successful completion of the software-upgrade, a message will be shown. Restart
the PC-software by pressing [Yes] in order to activate the new software version.
If the software-upgrade does not complete, the status messages can give conclusion on the
error source. Under normal conditions, "Connected to 192.168.0.100" is shown in the first
line. This indicates a proper connection to the FTP server. However if "Not connected"
appears, verify the network connection and settings. "550 File does not exits." indicates that
the installation package is in another path or not present on the server.

Update of video sequences

The update of the video sequences is available for Screen/PC-units "new" version
3.x or later for instruments with a serial number > 1500 or higher!
The playback of the videos is possible with software version 4.00 or later only!

Never perform the update of video sequences together with a software update!
The software update package has to be removed from the FTP server or moved to a
different FTP folder.
An update of video sequences is similar to a software update with FTP, the playback of the
videos is possible with software version 4.00 or later. This requires the "Roche OMNI S
Video Update CD".
1. Perform a software update if necessary and restart the OMNI S.
2. Prepare the PC and the Roche OMNI S same as for the software update with FTP until
step 9 (see section "Software update with FTP" on page 3-73).
3. Instead of the software update package, copy the video sequences update package from
the Video Update CD into a new subfolder (e.g. "/video") or into an empty FTP folder.

3-74 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

4. Continue with step 10 of the software update with FTP. As the path for the installation,
select the new folder (e.g. "/video") on the OMNI S.
Due to the amount of data, the update may need several time to finish, depending on the
transfer rate.

3.21 Printer

3.21.1 Replacing the Printer


1. Remove the device cover, printer cover and T&D cover of the OMNI S.
2. Remove the printer paper and press the printer lever on the right side back.
3. Pull the paper winder out to the right.
4. Unscrew the two screws holding the printer unit
(see Fig. 71: "OMNI S printer unit", 1, 2).

Fig. 71: OMNI S printer unit

5. Lift the printer and disconnect the connector cable from the printer unit.
6. Remove the printer unit.

Fig. 72: Printer connector cable

Use the reverse order for assembly.

During assembly, observe the position of the printer connector cable


(see Fig. 72: "Printer connector cable")!

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-75


3 Roche OMNI S – Components

3.21.2 Replacing the winder unit


1. Disassemble the printer unit as directed in section 3.21.1 "Replacing the Printer"
on page 3-75.
2. Unscrew the screw holding the winder unit on the bottom of the printer unit.
3. Disconnect the cable connector and remove the winder unit.
Use the reverse order for assembly.

During initial installation of the winder unit the plastic bar on the winder unit
must be broken off to allow the unit to be attached.

3.21.3 Replacing the paper lid


1. Remove the paper cover.
2. Press the paper cover to the left and pull it up.
Use the reverse order for assembly.

3.21.4 Replacing the printer-cutter lid


1. Remove the printer lid and open the printer-cutter lid.
2. Press the printer-cutter lid to the left or right and pull it up.
Use the reverse order for assembly.

3.22 Fan unit

3.22.1 Replacing the cooling fan unit


1. Remove the device cover, printer cover and T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Unscrew the two screws holding the fan unit and lift it up.
7. Disconnect the two connector cables of the fan unit at the mainboard and remove the
fan unit.
Use the reverse order for assembly.

3-76 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

3.23 Replacing the valve bus cable


1. Remove the device cover, printer cover and T&D cover of the OMNI S.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Disassemble the fan shield as directed in 3.22.1 "Replacing the cooling fan unit" on page
3-76.
7. Unscrew the two screws holding the mainboard retainer
(see Fig. 57: "Mainboard unit" on page 3-60).
8. Fold the mainboard retainer down.
9. Release the valve bus cable plug connector on the sample distributor
(see Fig. 73: "Valve bus cable (Sample distributor/mainboard)").

Plug connector on the


sample distributor

Plug connector on
the mainboard

Fig. 73: Valve bus cable (Sample distributor/mainboard)

10. Release valve bus cable connector on the mainboard


(see Fig. 73: "Valve bus cable (Sample distributor/mainboard)").

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-77


3 Roche OMNI S – Components

11. Release the valve bus cable plug connector on the measuring chamber cartridge valve
unit (see Fig. 74: "Valve bus cable").

Plug connector on the measuring


chamber cartridge valve unit

Plug connector on the


V4/V6 actuator board

Fig. 74: Valve bus cable

12. Release the valve bus connector cable on the V4/V6 actuator board and remove the
cable (see Fig. 74: "Valve bus cable").
Use the reverse order for assembly.

3.24 Replacing the actuator bus main controller cable


1. Remove the rear panel as directed in chapter 3.3 "Removing the rear panel" on page 3-1.
2. Rotate the T&D lock on the T&D module 90 degrees to the right
(see Fig. 2: "T&D module" on page 3-3).
3. Carefully pull the sample inlet path out to the right and remove it.
4. Unscrew the three screws holding the central measuring unit
(see Fig. 1: "Central measuring unit" on page 3-2, 1, 2, 3).

Do not remove the fastening screws completely, because they are secured against
loss.

5. Raise the central measuring unit forward until it clicks audibly into place.

The sample inlet path must be removed at this time, otherwise it will be
destroyed by folding the central measuring unit out.

6. Remove the light guide conduit with the COOX light guide as directed in 3.12.5
"Replacing the primary light guide" on page 3-32.
7. Remove the two screws holding the actuator bus plug.
8. Release the controller cable connector on the hemolyzer.
9. Release the controller cable connector on the vacuum control board.

3-78 Service Manual, Roche OMNI S, Version 4.0, December 2004


3 Roche OMNI S – Components

10. Release the controller cable connector on the FMS control board.
11. Release the controller cable connector on the T&D module.
12. Pull the controller cable out of the cable guide and remove it.
Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 3-79


3 Roche OMNI S – Components

3-80 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4 AutoQC module
4.1 Initial installation of the AutoQC module .................................................................. 4-1

4.2 Replacing the AutoQC module ..................................................................................... 4-2

4.3 Replacing the AQC snap lock ....................................................................................... 4-3

4.4 Replacing the AQC magnetic valve ............................................................................ 4-4

4.5 Replacing the AQC board .............................................................................................. 4-5

4.6 Replacing the YZ distributor board ............................................................................. 4-6

4.7 Replacing the Z distributor board ............................................................................... 4-7

4.8 Replacing the flex cable (short) .................................................................................. 4-8

4.9 Replacing the flex cable (long) .................................................................................... 4-9

4.10 AQC sample tube ............................................................................................................ 4-9


4.10.1 Replacing the AQC sample tube complete .........................................................................................4-9

4.10.2 Replacing the AQC sample tube short............................................................................................... 4-11


4.10.3 Replacing the AQC sample tube long ................................................................................................ 4-11

4.11 Replacing the X motor ................................................................................................. 4-12

4.12 Replacing the Y motor ................................................................................................. 4-12

4.13 Replacing the Z motor ................................................................................................. 4-13

4.14 Replacing the AutoQC steel tube .............................................................................. 4-14

4.15 Replacing the AQC temperature sensor .................................................................. 4-15

4.16 Replacing the wash port .............................................................................................. 4-16

4.17 Replacing the toothed belt (short) ............................................................................ 4-17

4.18 Replacing the toothed belt (long) ............................................................................. 4-18

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-I


4 AutoQC module

4-II Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4 AutoQC module
4.1 Initial installation of the AutoQC module
1. Switch off the OMNI S.
2. Remove the back panel of the device as directed in section 3.3 "Removing the rear
panel" on page 3-1.
3. Unscrew the two shipping screws wrapped with a red band on the AutoQC module.
4. Remove the red shipping valve cap from the AQC magnetic valve.
5. Unscrew the two screws holding the AutoQC drawer cover on the bottom of the OMNI S
and fasten the drawer cover to the AutoQC module.
6. Remove the AutoQC snap lock from the module and plug it into the matching opening
in the OMNI S (see Fig. 5: "snap lock (b)" on page 4-3).
7. Slide the AutoQC module into the opening and fasten it to the bottom of the OMNI S
with the two fixing screws.

The AutoQC module rails must fit flush to the OMNI S cover frame so the two
fixing screws can grip.

8. Remove the cylinder pin from the AQC rinse tube and connect the rinse tube to the
angled nipple (see Fig. 2: "Auto QC module - cable and tube connections" on page 4-2).
9. Connect the AQC sample tube to the angle unit (see Fig. 1: "AutoQC module: rinse
tube, sample tube, connector cable").
10. Push the connector cable through the AQC board cover (see Fig. 1: "AutoQC module:
rinse tube, sample tube, connector cable") and connect it to the AutoQC control board
(see Fig. 2: "Auto QC module - cable and tube connections").

Connector cable

Rinse tube

Sample tube

Fig. 1: AutoQC module: rinse tube, sample tube, connector cable

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-1


4 AutoQC module

4.2 Replacing the AutoQC module

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the OMNI S.


2. Remove the back panel of the device as directed in section 3.3 "Removing the rear
panel" on page 3-1.
3. Pull the AutoQC module out to the left.
4. Disconnect the AQC sample tube at the angle unit and pull it out
(see Fig. 2: "Auto QC module - cable and tube connections").

Connector cable AQC sample tube Angle unit Angled nipple

Wash tube
Fig. 2: Auto QC module - cable and tube connections

5. Disconnect the cable (at the board) and the wash tube (at the angled nipple)
(see Fig. 2: "Auto QC module - cable and tube connections").
6. Unscrew the two screws holding the AutoQC module on the bottom of the OMNI S
(see Fig. 3: "AutoQC fixing screws", 1, 2).

1 2

Fig. 3: AutoQC fixing screws

7. Pull the AutoQC module out to the left and remove it.
Use the reverse order for assembly.

4-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4.3 Replacing the AQC snap lock


The snap lock of the OMNI S AutoQC module consists of two parts. Part a is fixed to the
AutoQC module; part b is fixed to the right frame of the OMNI S.

Replacing the AQC snap lock (a)


1. Remove the back panel of the device as directed in section 3.3 "Removing the rear
panel" on page 3-1.
2. Pull the AutoQC module out to the left.
3. Unscrew the two screws holding the snap lock on the back of the AutoQC module and
remove them (see Fig. 4: "snap lock (a)", 1, 2).

1 2
1
2

Fig. 4: snap lock (a)

Use the reverse order for assembly.

Replacing the snap lock (b)


1. Remove the back panel of the device as directed in section 3.3 "Removing the rear
panel" on page 3-1.
2. Unscrew the four screws holding the right side panel and remove the side panel.

Do not remove the fastening screws completely, because they are secured against
loss.

3. Pull the AutoQC module out to the left.


4. Press the brackets on the back of the snap lock and slide it through the opening towards
the AutoQC module.

1 2
1
2

Fig. 5: snap lock (b)

Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-3


4 AutoQC module

4.4 Replacing the AQC magnetic valve

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the OMNI S.


2. Remove the back panel of the device as directed in section 3.3 "Removing the rear
panel" on page 3-1.
3. Pull the AutoQC module out to the left.
4. Remove the cap from the magnetic valve (see Fig. 6: "AutoQC magnetic valve").

Magnetic valve

1 2

Cap

Fig. 6: AutoQC magnetic valve

5. Unscrew the two screws holding the magnetic valve


(see Fig. 6: "AutoQC magnetic valve", 1, 2).
6. Pull the connector cable from the board and remove the magnetic valve.
Use the reverse order for assembly.

Activate the function program


[Analyzer] > [System] > [Component test] > [Valves] > [AQC module]
to test the functioning of the magnetic valve.

4-4 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4.5 Replacing the AQC board

Electrostatically sensitive component!

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Disassemble the AutoQC module as directed in 4.2 "Replacing the AutoQC module"
on page 4-2.
3. Unscrew the two screws holding the AutoQC control board cover and remove it
(see Fig. 7: "AutoQC board cover", 1, 2).

Fig. 7: AutoQC board cover

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-5


4 AutoQC module

4. Pull all cables (names: J8, J9, J11, J14) from the AutoQC control board as well as the
temperature sensor cable out.

First unlock the flex cable connector by lifting the ends of the connector then pull
the cable from the board.

XY motor cable (J9)


1
1 A

Temperature sensor cable

Magnetic valve cable (J11)

Flex cable (long) (J8)

2 B

Fig. 8: AutoQC Control Board

5. Unscrew the two screws (1, 2) holding the AutoQC control board, remove the two
spacer pins (A, B) (see Fig. 7: "AutoQC board cover") and remove the AutoQC control
board.
Use the reverse order for assembly.

Activate the function program


[Analyzer] > [System] > [Protected functions] > [AQC compensation]
to run a position compensation in the x and y direction.

4.6 Replacing the YZ distributor board

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Loosen the two screws holding the AutoQC cover by turning them about one
revolution.

4-6 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4. Pull the AutoQC cover back and then left.


5. Loosen the two screws holding the cross-slide cover by turning them about one
revolution.
6. Pull the cross-slide cover up.
7. Unscrew the cable guide fixing screw on top of the YZ distributor board and pull the
cable guide to the right (see Fig. 9: "YZ distributor board").

Cable guide

Flex cable (short)

Spacer pin

Motor cable Flex cable (long)

Fixing screw
Fig. 9: YZ distributor board

8. Pull the motor cable and the two flex cables (short and long) from the YZ distributor
board (see Fig. 9: "YZ distributor board").

First unlock the flex cable connector by lifting the ends of the connector then pull
the cable from the board.

9. Unscrew the screw holding the YZ distributor board, disassemble the spacer pin and
remove the YZ distributor board (see Fig. 9: "YZ distributor board").
Use the reverse order for assembly.

4.7 Replacing the Z distributor board

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Loosen the two screws holding the AutoQC cover by turning them about one
revolution.
4. Pull the AutoQC cover back and then left.

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-7


4 AutoQC module

5. Loosen the two screws holding the cross-slide cover by turning them about one
revolution.
6. Pull the cross-slide cover up.
7. Unscrew the screw holding the Z distributor board.

AQC sample tube

Motor cable

Flex cable (short)

Fixing screw

Setscrew

Fig. 10: Z distributor board

8. Disconnect the AQC sample tube, the motor cable and the flex cable (short) from the Z
distributor board (see Fig. 10: "Z distributor board").

First unlock the flex cable connector by lifting the ends of the connector then pull
the cable from the board.

9. Unscrew the cross-slide axis setscrew and remove it upwards (see Fig. 10: "Z distributor
board").
10. Remove the cross-slide unit and the Z distributor board.
Use the reverse order for assembly.

4.8 Replacing the flex cable (short)

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Loosen the two screws holding the AutoQC cover by turning them about one
revolution.
4. Pull the AutoQC cover back and then left.
5. Loosen the two screws holding the cross-slide cover by turning them about one
revolution.
6. Pull the cross-slide cover up.
7. Unscrew the screw holding the cable guide on top of the YZ distributor board and
remove the cable guide to the right (see Fig. 9: "YZ distributor board" on page 4-7).

4-8 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

8. Remove the flex cable (short) on the Z distributor board (see Fig. 10: "Z distributor
board" on page 4-8) and on the YZ distributor board (see Fig. 9: "YZ distributor board"
on page 4-7).

First unlock the flex cable connector by lifting the ends of the connector then pull
the cable from the board.

Use the reverse order for assembly.

4.9 Replacing the flex cable (long)

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Unscrew the screw holding the cable guide on top of the YZ distributor board and move
the cable guide to the right (see Fig. 9: "YZ distributor board" on page 4-7).
4. Remove the flex cable (long) on the YZ distributor board (see Fig. 9: "YZ distributor
board" on page 4-7) and on the control board (see Fig. 8: "AutoQC Control Board" on
page 4-6).

First unlock the flex cable connector by lifting the ends of the connector then pull
the cable from the board.

Use the reverse order for assembly.

4.10 AQC sample tube

4.10.1 Replacing the AQC sample tube complete


The AQC sample tube complete consists of: AQC sample tube short
AQC sample tube long
AutoQC safety spring
AutoQC angle unit

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-9


4 AutoQC module

3. Remove the back panel of the device as directed in section 3.3 "Removing the rear
panel" on page 3-1.
4. Unscrew the four screws holding the right side panel and remove the side panel.

Do not remove the fastening screws completely, because they are secured against
loss.

5. Remove the AQC sample tube from the cross-slide (see Fig. 10: "Z distributor board" on
page 4-8).
6. Remove the safety spring of the AQC sample tube on the back of the T&D module and
disconnect the AQC sample tube and remove it (see Fig. 11: "AutoQC AQC sample tube
and safety spring").

Safety spring

AQC sample tube

Fig. 11: AutoQC AQC sample tube and safety spring

7. Unscrew the screw (1) holding the angle unit (see Fig. 2: "Auto QC module - cable and
tube connections" on page 4-2) and remove the angle piece with the AQC sample tube.

Do not remove the fastening screws completely, because they are secured against
loss.

Use the reverse order for assembly.

4-10 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4.10.2 Replacing the AQC sample tube short

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Remove the AQC sample tube (short) from the cross-slide (see Fig. 10: "Z distributor
board" on page 4-8).
4. Disconnect the AQC sample tube (short) from the angle unit and remove it.
Use the reverse order for assembly.

4.10.3 Replacing the AQC sample tube long

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Remove the back panel of the device as directed in section 3.3 "Removing the rear
panel" on page 3-1.
4. Unscrew the four screws holding the right side panel and remove the side panel.

Do not remove the fastening screws completely, because they are secured against
loss.

5. Remove the AQC sample tube from the cross-slide (see Fig. 10: "Z distributor board" on
page 4-8).
6. Remove the safety spring of the AQC sample tube (long) on the back of the T&D mod-
ule and disconnect the AQC sample tube and remove it (see Fig. 11: "AutoQC AQC sam-
ple tube and safety spring").
7. Disconnect the AQC sample tube (long) from the angle unit and remove it.
Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-11


4 AutoQC module

4.11 Replacing the X motor

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Disassemble the AutoQC module as directed in 4.2 "Replacing the AutoQC module" on
page 4-2.
3. Unscrew the two screws holding the AutoQC control board cover and remove it
(see Fig. 7: "AutoQC board cover" on page 4-6).
4. Disconnect the X motor cable on the control board (see Fig. 8: "AutoQC Control Board"
on page 4-6).
5. Unscrew the two screws holding the X motor retainer (see Fig. 12: "X motor retainer")
and remove the retainer.

Fixing screws
of the X motor retainer

Fig. 12: X motor retainer

6. Unscrew the two X motor fixing screws and remove the motor.
7. Remove the gear wheel from the motor and attach it to the new motor.
Use the reverse order for the further assembly.

4.12 Replacing the Y motor

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Loosen the two screws holding the AutoQC cover by turning them about one
revolution.

4-12 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4. Pull the AutoQC cover back and then left.


5. Loosen the two screws holding the cross-slide cover by turning them about one
revolution.
6. Pull the cross-slide cover up.
7. Unscrew the screw holding the cable guide on top of the YZ distributor board and move
the cable guide to the right (see Fig. 9: "YZ distributor board" on page 4-7).
8. Disconnect the motor cable from the YZ distributor board
(see Fig. 9: "YZ distributor board" on page 4-7).
9. Unscrew the two screws holding the motor and remove the Y motor.
10. Remove the gear wheel from the motor and attach it to the new motor.
Use the reverse order for the further assembly.

4.13 Replacing the Z motor

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Loosen the two screws holding the AutoQC cover by turning them about one
revolution.
4. Pull the AutoQC cover back and then left.
5. Loosen the two screws holding the cross-slide cover by turning them about one
revolution.
6. Pull the cross-slide cover up.
7. Disconnect the motor cable from the Z distributor board (see Fig. 10: "Z distributor
board" on page 4-8).
8. Unscrew the cross-slide axis setscrew and pull it up (see Fig. 10: "Z distributor board"
on page 4-8).
9. Unscrew the two screws holding the motor and remove the Z motor.
10. Remove the gear wheel and the guide sleeve from the motor and attach them to the new
motor.
Use the reverse order for the further assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-13


4 AutoQC module

4.14 Replacing the AutoQC steel tube

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Remove the ampule mats, ampule block and the AutoQC cup.
4. Loosen the two screws holding the AutoQC cover by turning them about one
revolution.
5. Pull the AutoQC cover back and then left.
6. Loosen the two screws holding the cross-slide cover by turning them about one
revolution.
7. Pull the cross-slide cover up.
8. Disconnect tube (A) on the steel tube from the AQC needle.
9. Unscrew the setscrew of the steel tube (see Fig. 13: "AutoQC steel tube").

A
Setscrew

B
steel tube

Fig. 13: AutoQC steel tube

10. Release tube (B) from the steel tube and pull the steel tube down and out.
Use the reverse order for assembly.

Make sure that the steel tube connector tube for the silicone tube points forward!

4-14 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4.15 Replacing the AQC temperature sensor

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.

The OMNI S must always be switched off, because the new AutoQC temperature
sensor must be configured. This is done after the boot process.
The AutoQC temperature sensor must not be connected or disconnected with the
power on.

2. Pull the AutoQC module out to the left.


3. Remove the ampule mats, the ampule block and the AutoQC cup.
4. Disconnect the cable from the AutoQC temperature sensor
(see Fig. 8: "AutoQC Control Board" on page 4-6).
5. Unscrew the two screws holding the longitudinal slide cover and remove it
(see Fig. 14: "Longitudinal slide cover", 1, 2).

Temperature sensor

Fig. 14: Longitudinal slide cover

6. Remove the AutoQC temperature sensor and clean glue residues from the longitudinal
slide cover.
7. Glue the new AutoQC temperature sensor to the longitudinal slide cover.
Use the reverse order for assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-15


4 AutoQC module

4.16 Replacing the wash port

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Unscrew the screw holding the wash port and remove it (see Fig. 15: "Wash port").

Wash port

Fixing screw

Fig. 15: Wash port

4. Disassemble the wash port retainer by pressing the two snap locks together and remove
the wash port.
Use the reverse order for the further assembly.

4-16 Service Manual, Roche OMNI S, Version 4.0, December 2004


4 AutoQC module

4.17 Replacing the toothed belt (short)

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Loosen the two screws holding the cross-slide cover by turning them about one
revolution.
4. Pull the cross-slide cover up.
5. Unscrew the pulley screw (see Fig. 16: "Toothed belt (short)") and remove the toothed
belt.

Driver pin

Toothed belt (short)

Pulley

Fig. 16: Toothed belt (short)

6. Tension the new toothed belt over the toothed disk (see Fig. 16: "Toothed belt (short)")
and over the pulley.
7. Fasten the toothed belt to the pulley with the screw.
8. Position the toothed belt on the driver pin of the cross-slide
(see Fig. 16: "Toothed belt (short)").
Use the reverse order for the further assembly.

Service Manual, Roche OMNI S, Version 4.0, December 2004 4-17


4 AutoQC module

4.18 Replacing the toothed belt (long)

Activate the function program


[Analyzer] > [System] > [Component test] > [Aggregates] >
[AQC motors] > [AQC motors] > [Service Position]
to bring the motors to the service position.

1. Switch off the Roche OMNI S.


2. Pull the AutoQC module out to the left.
3. Remove the ampule mats, ampule block and the AutoQC cup.
4. Loosen the two screws holding the AutoQC cover by turning them about one
revolution.
5. Pull the AutoQC cover back and then left.
6. Loosen the two screws holding the cross-slide cover by turning them about one
revolution.
7. Pull the cross-slide cover up.
8. Unscrew the two screws holding the AutoQC control board cover and remove it
(see Fig. 7: "AutoQC board cover"page 4-5).
9. Disconnect the cable from the AutoQC temperature sensor
(see Fig. 8: "AutoQC Control Board" on page 4-6).
10. Unscrew the two screws holding the longitudinal slide and remove the shield
(see Fig. 14: "Longitudinal slide cover").
11. Unscrew the pulley screw and remove the toothed belt.
12. Tension the new toothed belt over the toothed disk and over the pulley.
13. Fasten the toothed belt to the pulley with the screw.
14. Position the toothed belt on the driver pin of the longitudinal slide.
Use the reverse order for the further assembly.

4-18 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

5 Fluid actions
5.1 Overview ........................................................................................................................... 5-1
5.1.1 OMNI S 5 and 6 .............................................................................................................................................5-1

5.1.2 OMNI S 3 and 4 .............................................................................................................................................5-2

5.1.3 OMNI S 1 and 2 .............................................................................................................................................5-3

5.2 Pin assignment of Packs S2 and S3 ........................................................................... 5-4


5.2.1 Pack S2..............................................................................................................................................................5-4

5.2.2 Pack S3 (OMNI S 5 and 6 only) ...............................................................................................................5-4

5.3 T&D positions ................................................................................................................... 5-5

5.4 Measurement ................................................................................................................... 5-6


5.4.1 Sample input ...................................................................................................................................................5-6

5.4.2 Sample distribution.................................................................................................................................... 5-10

5.4.3 Aspirating MSS standby solution and starting FMS ..................................................................... 5-12

5.4.4 Washing the measuring chamber vicinity......................................................................................... 5-14

5.4.5 Washing tHb/SO2 or COOX and preparing Mix1........................................................................... 5-16

5.4.6 Recalibrating BG/ISE and drying tHb/SO2 or COOX .................................................................... 5-18

5.5 Calibration ...................................................................................................................... 5-20


5.5.1 BG/ISE cleaning .......................................................................................................................................... 5-20
5.5.2 Calibrating the conductivity for calibrations solution A or B ..................................................... 5-22

5.5.3 O2 zero point calibration ......................................................................................................................... 5-24

5.5.4 Na conditioning........................................................................................................................................... 5-25

5.5.5 Mix1/Mix2 calibration, aspirating BG/ISE reference solution.................................................... 5-26

5.5.6 O2-air calibration........................................................................................................................................ 5-28

5.5.7 Aspirating MSS reference solution, calibration with standby solution .................................. 5-29

5.5.8 Cal1 Calibration........................................................................................................................................... 5-30

5.5.9 Cal2/Cal3/Cal4 Calibration ..................................................................................................................... 5-31

5.5.10 COOX calibration........................................................................................................................................ 5-32

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-I


5 Fluid actions

5-II Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

5 Fluid actions
5.1 Overview

5.1.1 OMN I S 5 and 6

The OMNI S 5 contains a tHb/SO 2 module instead of the COOX module!

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-1


5 Fluid actions

5.1.2 OMN I S 3 and 4

The OMNI S 3 contains a tHb/SO 2 module instead of the COOX module!

Main pump

Vacuum
pump
30°C
Barometer 15
17 12

16

13
14

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3

M
19

FMS
W
Waste S1 S2
container Rinse solution

5-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

5.1.3 OMN I S 1 and 2

The OMNI S 1 contains a tHb/SO 2 module instead of the COOX module!

Main pump

Vacuum
pump
30°C
Barometer 15
17 12

16

13
14

Moisture
trap Bacteria
filter 37°C
Vacuum MC
pump shield
18 cartridge T&D
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3

M
19

FMS
W
Waste S1 S2
container Rinse solution

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-3


5 Fluid actions

5.2 Pin assignment of Packs S2 and S3

5.2.1 Pack S2

Reference solution BG Na Conditioner

6 7
Reference solution ISE BG/ISE cleaning solution

5 8

4 1 O2 zero point solution


empty
3 2

Calibration solution A Calibration solution B

5.2.2 Pack S3 (OM N I S 5 and 6 only)

Reference solution MSS Calibration solution Cal4

6 7
Calibration solution Cal3
empty
5 8

4 1 Calibration solution Cal2


empty
3 2

Calibration solution Cal1 Standby solution

5-4 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

5.3 T&D positions


• Position 1: releases the fill port for inserting a sample
• Position 2: add the S1 rinse solution
• Position 3: position for "aspiration from the syringe“
• Position 4: dock position
• Position 5: for O 2 zero point solution from S2 Fluid Pack
• Position 6: calibration solution 2 from S3 Fluid Pack A (OMNI S 5 and 6 only)
• Position 7: Na conditioning solution from S2 Fluid Pack
• Position 8: calibration solution 3 from S3 Fluid Pack A (OMNI S 5 and 6 only)
• Position 9: cleaning solution from S2 Fluid Pack
• Position 10: AutoQC connection
• Position 11: calibration solution 4 from S3 Fluid Pack A (OMNI S 5 and 6 only)

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-5


5 Fluid actions

5.4 Measurement

5.4.1 Sample input

Injecting sample

V24

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9
10
11

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Waste separator Fill port


V23 SS3 SS2 SS6

5-6 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

The sample container is detected by a light barrier on the T&D module.


Sample path: fill port >
sample inlet path >
SS2
When the sample has reached SS2 the user is prompted (by beeping and
prompt window) to remove the sample container.
Excess sample material:
V24 (open) >
V23 (open) >
Bypass >
Waste separator

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-7


5 Fluid actions

Aspirate sample

Main pump

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Fill port
SS2

5-8 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

The sample container is detected by a light barrier on the T&D module. The sample is aspi-
rated through the cross channel with the main pump.
Sample path: fill port >
sample inlet path >
SS2
When the sample has reached SS2 the user is prompted (by beeping and
prompt window) to remove the sample container.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-9


5 Fluid actions

5.4.2 Sample distribution

Main pump MSS output pump

10
17 12 15

20

16 8
MCO CCL 7 CCR MCI

13
14 VSI

9
5
6 VIO
MCO MCC 4 VII

MCI

18 T&D
5 21

4
Waste MCO MCC 3 VBI 6 7
3 VBO 5
8

separator MCI
4
9

10
3

1|3
22
2
11
T&D tubing set

4
8 23
24 2
3
4
5
6
7 8 9 11
10

2
2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

S1 Rinse
solution

1 - separate BG
2 - separate tHb/SO 2 or COOX

3 - position BG, separate ISE


4 - position ISE

5 - separate MSS

6 - wash rail
7 - position MSS

8 - position tHb/SO 2 or COOX

5-10 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

When the sample is detected by SS2 the sample distribution is started. The sequence of the
individual aspiration phases depends on the existing or requested modules.
Separating BG:
The start of the sample is aspirated from SS2 to SS1. The volume SS2-SS1 is specified for
filling the BG measuring chamber.
Separating tHb/SO 2 or COOX:
One sample segment is aspirated from SS2 to SS3. The volume SS2-SS3 is specified for
filling the tHb/SO 2 or COOX module.The user can specify in the setup whether
tHb/SO 2 / COOX or ISE is separated first.
Positioning BG and separating ISE:
This aspiration process assumes that the sample column tip is at SS1. It is aspirated via VBI
and VBO with the main pump. The start of the sample movement is detected by the conduc-
tivity (breaking of the calibration solution column in the measuring chamber). If this does
not occur at the right time, an aspiration error is displayed. The positioning of the sample
is controlled by the MCI-MCO contact path. If a timeout occurs because an excessively low
signal or a non-reproducible conductivity signal, an aspiration error is displayed. During
the aspiration the sample is "observed" with SS1; if the volume of the sample is too small,
this interferes with the aspiration process.
After closing the VBI the ratio of the conductivities of MCI-MCO to MCI-MCC is checked.
If there are air bubbles in the MCI-MCO area, this can result is the aspiration process
restarting and the VBI opening again. When the corresponding conductivity level is
reached, the positioning of the sample is complete.
During the BG positioning ISE is separated between SS2 and SS1.
Positioning ISE:
This aspiration process is similar to positioning BG. If an ISE measurement is not required,
the remainder of the sample is aspirated to VII through the sample distributor to establish
the required initial state for a subsequent MSS positioning.
Separating MSS and washing rail:
A sample segment is aspirated to the SS4.
Positioning MSS:
The sample is aspirated by turning the MSS output pump and opening the VSI. The tearing
of the standby solution and the start of the sample is detected by the conductivity at the MSS
input. Then it must also be possible to detect the tearing of the standby solution and the
sample start, at which the MSS output pump is stopped. The completed sample positioning
is checked again for freedom from bubbles via the conductivity distances.
Positioning tHb/SO 2 or COOX:
The sample segment for measurement of tHb/SO 2 or COOX is positioned last. The valves
V2, V4, V17 and V16 are opened and aspirated with the main pump. The sample is still
hemolyzed in the event of a COOX measurement.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-11


5 Fluid actions

5.4.3 Aspirating M S S standby solution and starting FM S

MSS output pump MSS input pump

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI 1
9

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Calibration solution B Calibration solution A

MSS standby solution

1 - aspirating MSS standby solution

2 - starting FMS

5-12 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

The MSS is positioned in four subsections:


Preconditioning by aspiration of standby solution (1)
This process is done before aspiration of the sample. A little reference solution is aspirated
via V8 and the MSS reference electrode (see section 5.5.7 "Aspirating MSS reference solu-
tion, calibration with standby solution" on page 5-29) and is checked with the conductivity
section at the MSS output. Then the measuring chamber is filled with standby solution.
Aspirate sample
The sample is aspirated by turning the MSS output pump and opening the VSI (see section
5.4.2 "Sample distribution" on page 5-10). The tearing of the standby solution and the start
of the sample is detected by the conductivity at the MSS input. Then it must also be possible
to detect the tearing of the standby solution and the sample start, at which the MSS output
pump is stopped. The completed sample positioning is checked again for freedom from bub-
bles via the conductivity distances.
Aspiration of standby solution (2) to determine glu and lac
The determination of the MSS measurement values is always based on the reference poten-
tial of the standby solution; for this reason the MSS cassette is filled with standby solution
(by stopping the MSS input pump) after washing out the sample (with a mixture of standby
solution and air by rotating the two MSS pumps). The quality of the aspiration is assessed
by the conductivity distance at the MSS output.
The MSS input pump is used to fill the tube in the lateral channel (with standby solution),
for defined aspiration (aspirating the standby solution) of the MSS measuring chamber cas-
sette and for washing valve V10.
Aspiration of standby solution (3) to determine urea
The MSS cassette is filled with standby solution again. The quality of the aspiration and the
stroke (after repeated aspiration of standby solution) is assessed by the conductivity dis-
tance at the MSS output.

The sample segment for measurement of tHb/SO 2 or COOX is positioned last. The valves
V2, V4, V17 and V16 are opened and aspirated with the main pump (see section 5.4.2 "Sam-
ple distribution" on page 5-10). The sample is still hemolyzed in the event of a COOX mea-
surement.
At the same time the FMS is started for these procedures (for the pending recalibration of
the BG and ISE modules) and mixed from the A and B Mix1 solutions.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-13


5 Fluid actions

5.4.4 Washing the measuring chamber vicinity

Vacuum pump T&D module (position 2)

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

S1 Rinse
solution

waste separator

5-14 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

This section is used to clean sample residues from the aspiration path, the bypass and the
sample distributor.
The T&D disk docks at position 2 (S1 Rinse solution); then a rinse solution-air mix is aspi-
rated into the waste separator over the aspiration path and the bypass or the sample distrib-
utor (with V15 and V16 open) by creating a low pressure with the vacuum pump. The cor-
rect rinse solution to air ratio is assessed with the SS2 sensor.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-15


5 Fluid actions

5.4.5 Washing tHb/ SO 2 or COOX and preparing Mix1

Vacuum pump Main pump

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10

2 3

2
11
T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
AQC 1 2

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

S1 Rinse
solution

Waste separator FMS

1 - washing tHb/SO 2 or COOX

2 - preparing Mix1

5-16 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

When the measuring chamber area is clean, the tHb/SO 2 or COOX module is washed. The
vacuum pump runs rinse solution "in the reverse direction" from V3 through the measuring
module to the Hb cartridge; air packets are mixed in through V4 to increase the cleaning
effect.
Mix1 is aspirated from the FMS to the FMS tube simultaneously.

Then the measuring chamber area is dried. During this procedure the T&D disk is undocked
with the vacuum pump running until the measuring chamber area is dry. The valve positions
are the same as when washing the measuring chamber area
(see 5.4.4 "Washing the measuring chamber vicinity" on page 5-14).

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-17


5 Fluid actions

5.4.6 Recalibrating BG /I S E and drying tHb/ SO 2 or COOX

Vacuum pump Main pump

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 7

V4 3 VBO
MCI
5
6
8

9
4

10

1 3

2
11
T&D tubing set

2 24
22
3
4
5
6
7 8 9 11
2 10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Waste separator FMS

1 - recalibrating BG/ISE

2 - drying tHb/SO 2 or COOX

5-18 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

To recalibrate BG and ISE the main pump first aspirates a mixture of air (from a reservoir
between sample distributor and measuring chamber cartridge) and Mix1 through the mea-
suring chambers. When this has sufficiently washed them, the measuring chambers are filled
with Mix1 by shutting off the air feed.
This filling is assessed by conductivity measurement: with liquid solution a stable and plau-
sible value must be measured between MCI and MCO.
Then the inlet valve is closed and reference solution is aspirated through valves V5 or V9
(see 5.5.5 "Mix1/Mix2 calibration, aspirating BG/ISE reference solution" on page 5-26) until
a clear increase in conductivity between MCI and MCO can be detected. Finally the outlet
and reference valves are closed and the conductivity is checked again.

After each washing of the BG or ISE, the rail is washed and emptied to counteract electrical
couplings to the pump elastomer.

Along with the recalibration the tHb/SO 2 or COOX module is dried. Air is aspirated into the
waste separator with the vacuum pump through V4, measuring module, V2 and V23.

Before the instrument is released for the next measurement, Mix1 residue is aspirated into
the waste separator with the main pump with valves V22, V15 open.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-19


5 Fluid actions

5.5 Calibration

5.5.1 BG / I S E cleaning

Main pump T&D module (position 9)

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Waste separator Cleaning solution

5-20 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

This procedure removes deposits in the BG or ISE measuring chamber. The required clean-
ing solution is in the S2 Fluid Pack and is aspirated into the measuring chambers through
the T&D disk (position 9), the aspiration channel and the sample distributor.
The calibration of the conductivity is performed next.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-21


5 Fluid actions

5.5.2 Calibrating the conductivity for calibrations solution A or B

Main pump

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS

Calibration solution B

Calibration solution A

5-22 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

FMS adjustment:
This procedure starts by aspirating the FMS through V19 (air) until it is empty. After a short
pressure equalization V19 is closed and calibration solution B is aspirated into the BG mea-
suring chamber (see 5.5.2 "Calibrating the conductivity for calibrations solution A or B" on
page 5-22).
Solution B must be detected in the BG measuring chamber by the conductivity measuring
distance within a set period (after 9.64 sec at the earliest and after 12.34 sec at the latest).

Calibrating the conductivity for calibration solution B


In this section the conductivity of calibration solution B is precisely measured in the BG and
ISE measuring chamber. The precision of this measurement is very significant for Mix1 and
Mix2, because they are based on this measurement.
Solution B is already in the sample distributor at V21. The BG measuring chamber is cleaned
with a mixture of air and calibration solution B after the FMS adjustment. Then the BG mea-
suring chamber is filled with calibration solution B with no air bubbles and the conductivity
is determined.
Then calibration solution B is aspirated into the ISE measuring chamber and also measured
there.

Calibrating the conductivity for calibration solution A


This procedure is similar to the calibration of the conductivity for calibration solution B,
with the only difference being that a FMS adjustment is not made first.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-23


5 Fluid actions

5.5.3 O 2 zero point calibration

Main pump O 2 electrode T&D module (position 5)

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Waste separator O 2 zero point solution

This procedure determines the second calibration point of the O 2 electrode. The O 2 zero
point solution is in the S2 Fluid Pack and is aspirated into the BG measuring chamber with
the vacuum pump and the main pump through the T&D disk (position 5) and the aspiration
channel through SS2 and SS1.
After completing this procedure the measuring chamber area is briefly washed
(see 5.4.4 "Washing the measuring chamber vicinity" on page 5-14).

5-24 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

5.5.4 Na conditioning

Vakuum pump Main pump Na electrode T&D module (position 7)

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Waste separator Na conditioner

This procedure is necessary to retain the sensitivity of the Na electrode. The Na conditioner
is in the S2 Fluid Pack and is aspirated into the ISE measuring chamber with the vacuum
pump and the main pump through the T&D disk (position 7) and the sample inlet path
through SS2 and SS1.
After this procedure the measuring chamber area is washed (see 5.4.4 "Washing the measur-
ing chamber vicinity" on page 5-14), and the tHb/SO 2 or COOX module is also washed and
Mix1 is prepared (see 5.4.5 "Washing tHb/SO2 or COOX and preparing Mix1" on
page 5-16).
In the case of the tHb/SO 2 module a calibration is conducted during the washing procedure
by measuring the rinse solution values.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-25


5 Fluid actions

5.5.5 Mix1/ Mix2 calibration, aspirating BG / I S E reference solution

Main pump

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

3
9

MCO MCC VII


VIO
MCI

18
2 5 21
1 T&D

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

BG Reference solution FMS

Calibration solution B
ISE Reference solution
Calibration solution A

1 - Mix1/Mix2 Calibration

2 - aspirating BG reference solution

3 - aspirating ISE reference solution

5-26 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

Mix1 Calibration
This section refers to the BG and ISE measuring chamber only. The Mix1 mixture is pre-
pared similarly to a recalibration in the sample distributor after a measurement (see 5.4.6
"Recalibrating BG/ISE and drying tHb/SO2 or COOX" on page 5-18) and is ready for entry
to the measuring chamber at V21.
The main pump first aspirates a mixture of air (from a reservoir between sample distributor
and measuring chamber cartridge) and Mix1 through the measuring chambers. When this
has sufficiently washed them, the measuring chambers are filled with Mix1 by shutting off
the air feed.
This filling is assessed by conductivity measurement: with liquid solution a stable and plau-
sible value must be measured between MCI and MCO.
Then the inlet valves are closed and reference solution is aspirated through valves V5 or V9
until a clear increase in conductivity between MCI and MCO can be detected. Finally the
outlet and reference valves are closed and the conductivity is checked again.

Mix2 Calibration
The fluid procedure corresponds exactly to the Mix1 calibration.
The only difference is the mixture ratio of calibration solutions A and B:
Mix1: 33.3% calibration solution A
66.6% calibration solution B
Mix2: 60.0% calibration solution A
40.0% calibration solution B
The Mix2 calibration determines the slopes of the BG electrodes CO 2 and pH along with all
ISE electrodes.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-27


5 Fluid actions

5.5.6 O 2 -air calibration

Main pump O 2 electrode T&D module (position 4)

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

This procedure determines the first calibration point of the O 2 electrode. Ambient air is
slowly aspirated from the T&D disk (position 4) into the BG measuring chamber over the
aspiration channel and sample distributor.
Then the electrodes are conditioned with Mix1 (see 5.5.5 "Mix1/Mix2 calibration, aspirating
BG/ISE reference solution" on page 5-26).

5-28 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

5.5.7 Aspirating M S S reference solution, calibration with standby solution

MSS output pump

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI


2 6 7
3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6
1

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Waste separator MSS reference solution

MSS Standby solution

1 - aspirating MSS reference solution

2 - calibration with standby solution

The MSS cassette is preconditioned by first aspirating standby solution through a metal pipe
in the sample distributor and valves V10 and VSI. The filling is checked at the entrance to
the measuring chamber (MCI - CCR) and at the output to the measuring chamber
(CCL - MCO) by conductivity distance measurements.
Then the reference solution is aspirated to the reference electrode through V8. Next the MSS
cassette is rinsed with standby solution again.
Next the calibration solutions Cal1 to Cal4 (see section 5.5.8 "Cal1 Calibration" and 5.5.9
"Cal2/Cal3/Cal4 Calibration") are aspirated and after a rinse with standby solution a refer-
ence measurement for the glu and lac parameters is made. The reference measurement for
urea is made after repeated rinsing with standby solution.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-29


5 Fluid actions

5.5.8 Cal1 Calibration

MSS output pump V20

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Waste separator Calibration solution Cal1

This procedure determines the sensitivity of the sensors. Cal1 is aspirated into the MSS
measuring chamber from the S3 Fluid Pack through a metal pipe in the sample distributor
and valves V20 and VSI and measured. The positioning is checked by conductivity measure-
ments.

5-30 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

5.5.9 Cal2/Cal3/Cal4 Calibration

MSS output pump T&D module (Cal2=Pos. 6)


(Cal3=Pos. 8)
(Cal4=Pos.11)

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI

18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

11
2 T&D tubing set
22

24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Cal2
Waste separator Cal3
Cal4
Cal2 Calibration
This procedure determines the linearity of the sensors. Cal2 is aspirated from the S3 Fluid
Pack via the T&D disk (position 6), sample inlet path and sample distributor into the MSS
measuring chamber and measured.
Cal3 Calibration
This procedure determines the interference sensitivity of the sensors. Cal3 is aspirated from
the S3 Fluid Pack via the T&D disk (position 8), sample inlet path and sample distributor
into the MSS measuring chamber and measured.
Cal4 Calibration
This procedure determines the interference sensitivity of the sensors. Cal4 is aspirated from
the S3 Fluid Pack via the T&D disk (position 11), sample inlet path and sample distributor
into the MSS measuring chamber and measured.

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-31


5 Fluid actions

5.5.10 COOX calibration

Main pump

10
17 12 15

20

16 8
MCO CCL CCR MCI

13
14 VSI

MCO MCC VII


VIO
MCI
3
18 T&D
5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
9
4

10
3

2 2
11
T&D tubing set
22
1
24 2
3
4
5
6
7 8 9 11
10

4 23
AQC
2
AQC
4
6

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

Waste separator SS2 fill port

1 - calibrator input

2 - separator COOX

3 - aspirate excess out

4 - position calibrator

5-32 Service Manual, Roche OMNI S, Version 4.0, December 2004


5 Fluid actions

The calibration does not begin automatically but must be initiated by the user. The fill port
of the T&D module is released for input from an external tHb calibrator.
Once the SS2 sensor has detected the solution and the sample container has been removed,
the positioning begins: a separation at the SS3 sensor is made, then the excess solution is
aspirated out through the sample distributor. The calibration solution is now aspirated
through V2 and the hemolyzer through the cuvette with the main pump. The optical mea-
surement and storage of the cuvette layer thickness is made here.

Finally the COOX module is washed and dried (see section 5.4.5 "Washing tHb/SO2 or
COOX and preparing Mix1" or 5.4.6 "Recalibrating BG/ISE and drying tHb/SO2 or COOX").

Service Manual, Roche OMNI S, Version 4.0, December 2004 5-33


5 Fluid actions

5-34 Service Manual, Roche OMNI S, Version 4.0, December 2004


6 Protected software functions

6 Protected software functions


6.1 Protected setup ............................................................................................................... 6-1
6.1.1 Sample counter ..............................................................................................................................................6-1

6.1.2 Baro settings ...................................................................................................................................................6-1

6.1.3 Miscellaneous settings................................................................................................................................6-1

6.1.4 Serial number..................................................................................................................................................6-2

6.1.5 Instrument type ..............................................................................................................................................6-2

6.1.6 Polarization voltage.......................................................................................................................................6-2

6.1.7 COOX module (OMNI S 2, 4, 6 only)......................................................................................................6-3

6.1.8 AutoQC module .............................................................................................................................................6-3

6.1.9 Log monitor .....................................................................................................................................................6-3

6.2 Protected system functions .......................................................................................... 6-4


6.2.1 AQC adjustment.............................................................................................................................................6-4

6.3 Protected infos ................................................................................................................ 6-4


6.3.1 Service log file export ..................................................................................................................................6-4

6.4 Protected DB functions .................................................................................................. 6-4


6.4.1 Delete all databases .....................................................................................................................................6-4

6.4.2 Delete databases & archives.....................................................................................................................6-4

Service Manual, Roche OMNI S, Version 4.0, December 2004 6-I


6 Protected software functions

6-II Service Manual, Roche OMNI S, Version 4.0, December 2004


6 Protected software functions

6 Protected software functions


6.1 Protected setup

6.1.1 Sample counter


This function allows manual editing of the internal sample counter. See Samples for the
number of samples input (can be edited by the keyboard symbol), QC samples for the num-
ber of QC samples input (can be edited by the keyboard symbol) and Measurements for the
number of measurements made (can be edited by the keyboard symbol).

6.1.2 Baro settings


This function allows various barometer settings to be manually edited.
The following options are available:
Baro [mmHg] input the baro value with the keyboard symbol.
[Update E E PROM] this key allows the data of the vacuum EEPROM to be updated
by the OMNI S software. The data on the hard disk are written
to the EEPROM.

P_atm, SENS and ZERO are default settings which must not be changed!

IMPORTANT:
From approx. 2000 m above sea level or air pressure < 570 mmHg (75.98 kPa), the
specifications for parameter PO 2 are no longer fulfilled and the parameter must no
longer be used for evaluation of the clinical decisions.
After successful installation, the parameter must be permanently deactivated!

6.1.3 Miscellaneous settings


This function is used to activate various auxiliary programs and service actions.
The following options are available:
[Demo mode] This function simulates the OMNI S with all measurement values for
demonstration purposes. The measurement values are not calibrated;
instead they are spread around a mean value.
[Cal. intervals U.S.] This key reduces the 2P calibration interval of PO 2 from 72 hours to the
2P calibration interval of the other parameters to correspond to specif-
ic legal requirements in the USA.
[Bili] This key activates (green) or deactivates (red) the parameter bilirubin
(Roche OMNI S 2, 4 and 6 only).

Service Manual, Roche OMNI S, Version 4.0, December 2004 6-1


6 Protected software functions

[Service mode] This key activates an external keyboard connected to the OMNI S.
[Hotline info] This key edits the information hotline (that appears under Customer
Information).

6.1.4 Serial number


This function assigns a serial number after pressing the keyboard symbol.

6.1.5 Instrument type


This function is used to define the instrument type that is valid for the respective
instrument. Marking the desired instrument types and pressing the key
[Save and exit] assigns the new type to the OMNI S software.
The key [Load default setup] resets any changes made to the instrument type and loads the
default settings of the instrument stored on the hard disk.

After replacing the screen/PC unit or the hard disk, assign the valid instrument type
and load the default settings. When system stop 10005 occurs after replacing the
screen/PC unit or the hard disk and booting the PC, assign the valid instrument
type and load the default settings.

6.1.6 Polarization voltage

Upol O2
The keyboard symbol under Upol O 2 can be used to change the polarization voltage of the
oxygen electrode (default value = 750 mV). The [Adjust] key is used to confirm and import
the newly defined value.

Upol MSS (OMNI S 5 and 6 only)


The keyboard symbol under Upol MSS can be used to change the polarization voltage of the
MSS sensors (default value -350 mV). The [Adjust] key is used to confirm and import the
newly defined value. The polarization voltage can only be changed with the MSS cassette
inserted.

6-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


6 Protected software functions

6.1.7 COOX module (OM N I S 2, 4, 6 only)


Setup
The number of samples during a measurement is shown under Scans
(1, 2, 4, 8 and 16 samples are possible).
The duration of sampling during a measurement is shown under Integration time
(12 ms to 12000 ms are possible).

COOX switches

This setting is intended for internal purposes only and must be changed only when
specifically directed!

The [Suppress negative values] key activates (green) or deactivates (red) the suppression
of the output of negative values during a measurement. The measurement values are set
to 0 in the event of activation and a matching tHb compensation is made.

COOX adjustments
The offset values for the polychromator bilirubin are stored here. After replacing the poly-
chromator bilirubin, the screen/PC unit or the hard disk, they must be re-entered by means
of entering the code. The code is located on the polychromator and must be entered using
the keyboard symbol under Set.
The [Set default value] key loads the default settings of the polychromator bilirubin stored
on the harddisk.

6.1.8 AutoQC module


The [Temperature correction] key activates (green) or deactivates (red) the temperature
correction of the pH, PO 2 and PCO 2 measurement values during a QC measurement to
receive standardized measurement results.

6.1.9 Log monitor


The [Log monitor printout] key activates (green) or deactivates (red) the output of all log
monitor data at the internal thermal printer.
The keyboard symbol can also be used to specify the number of generations and the number
of log file entries in the printout.

Deactivate this function to prevent permanent stress on the thermal printer!

Service Manual, Roche OMNI S, Version 4.0, December 2004 6-3


6 Protected software functions

6.2 Protected system functions

6.2.1 AQC adjustment


• The start [Start] key initializes a position compensation of the AutoQC motors to ensure
that ampoules are pierced at the correct position after replacement of the AutoQC
board. This should be performed he course of AQC maintenance.

To avoid damage, the AQC ampoule mats must be removed from the AQC ampoule
block before starting the adjustment!

Be aware of possible injuries due to moving parts!

6.3 Protected infos

6.3.1 Service log file export


This function exports log files to a floppy disk.
The naming of the logfile archive is as follows:
logfile_xxxx_yyyyyy_zzzz.pck
x...serial number
y...date
z...time

When there is too little free space on the floppy disk, data can be deleted. If this is
confirmed with „Yes“, all existing data on the floppy disk will be deleted.

6.4 Protected DB functions

6.4.1 Delete all databases


This function is used to delete all patient, measurement, QC measurement, calibration and
instrument databases.

The instrument is restarted!

6.4.2 Delete databases & archives


This function deletes all databases and archives (corresponds to backups).

6-4 Service Manual, Roche OMNI S, Version 4.0, December 2004


7 Maintenance

7 Maintenance
7.1 OMNI S 1, 3, 5 (tHb/SO2) .............................................................................................. 7-1
7.1.1 Yearly replacement.......................................................................................................................................7-1

7.1.2 Replace every three years ..........................................................................................................................7-2

7.2 OMNI S 2, 4, 6 (COOX) .................................................................................................. 7-4


7.2.1 Yearly replacement.......................................................................................................................................7-4

7.2.2 Replace every three years ..........................................................................................................................7-5

7.3 AutoQC module ............................................................................................................... 7-7


7.3.1 Yearly replacement.......................................................................................................................................7-7

Service Manual, Roche OMNI S, Version 4.0, December 2004 7-I


7 Maintenance

7-II Service Manual, Roche OMNI S, Version 4.0, December 2004


7 Maintenance

7 Maintenance
7.1 OM NI S 1, 3, 5 (tHb/ SO 2 )

7.1.1 Yearly replacement


Various components of the OMNI S, such as T&D module, tubes, waste container,
etc., contain liquid biological residues after use that may cause infection.
Handle these components with care and avoid contact with skin.Always wear gloves
when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.

Sample tube
See section 3.5.4 "Replacing the needle" on page 3-4.

Fill port
See section 3.5.2 "Replacing the fill port" on page 3-4.

T&D tubing set


See section 3.5.10 "Replacing the T&D tubing set" on page 3-7.

Pump head
See section 3.14.2 "Replacing the pump head" on page 3-41.

PP pump tube
See section 3.14.3 "Replacing the pump tube" on page 3-42.

FMS tubing set


See section 3.6.3 "Replacing the FMS tubing set" on page 3-13.

Waste separator
See section 3.15.3 "Replacing the waste separator" on page 3-43.

Bypass nipple
See section 3.5.6 "Replacing the bypass nipple" on page 3-5.

Service Manual, Roche OMNI S, Version 4.0, December 2004 7-1


7 Maintenance

Bacteria filter
See section 3.15.8 "Replacing the bacteria filter" on page 3-45.

T&D disk
See section 3.5.9 "Replacing the T&D disk" on page 3-6.

Tubing set tHb/PolyOX


See section 3.13.2 "Replacing the tubing set tHb/PolyOX" on page 3-38.

The tubing paths must also be disinfected annually and the baro value must be
checked.

7.1.2 Replace every three years


Various components of the OMNI S, such as T&D module, tubes, waste container,
etc., contain liquid biological residues after use that may cause infection.
Handle these components with care and avoid contact with skin.Always wear gloves
when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.

sample tube
See section 3.5.4 "Replacing the needle" on page 3-4.

Fill port
See section 3.5.2 "Replacing the fill port" on page 3-4.

T&D tubing set


See section 3.5.10 "Replacing the T&D tubing set" on page 3-7.

Pump head
See section 3.14.2 "Replacing the pump head" on page 3-41.

PP pump tube
See section 3.14.3 "Replacing the pump tube" on page 3-42.

FMS tubing set


See section 3.6.3 "Replacing the FMS tubing set" on page 3-13.

7-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


7 Maintenance

Waste separator
See section 3.15.3 "Replacing the waste separator" on page 3-43.

Bypass nipple
See section 3.5.6 "Replacing the bypass nipple" on page 3-5.

Bacteria filter
See section 3.15.8 "Replacing the bacteria filter" on page 3-45.

T&D disk
See section 3.5.9 "Replacing the T&D disk" on page 3-6.

Tubing set tHb/PolyOX


See section 3.13.2 "Replacing the tubing set tHb/PolyOX" on page 3-38.

Sample distributor cartridge


See section 3.7.1 "Replacing the sample distributor cartridge" on page 3-13.

Measuring chamber cartridge


See section 3.11.1 "Replacing the measuring chamber cartridge" on page 3-25.

Hb cartridge
See section 3.13.4 "Replacing the Hb cartridge" on page 3-39.

Waste tubing
3.16.5 "Replacing the waste tube" on page 3-50.

Vacuum pump head


See section 3.16.3 "Replacing the vacuum pump head" on page 3-49.

VP tubing set
See section 3.16.7 "Replacing the VP tube set" on page 3-51.

The tubing paths must also be disinfected annually and the baro value must be
checked.

Service Manual, Roche OMNI S, Version 4.0, December 2004 7-3


7 Maintenance

7.2 OM N I S 2, 4, 6 (COOX)

7.2.1 Yearly replacement


Various components of the OMNI S, such as T&D module, tubes, waste container,
etc., contain liquid biological residues after use that may cause infection.
Handle these components with care and avoid contact with skin.Always wear gloves
when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.

Needle
See section 3.5.4 "Replacing the needle" on page 3-4.

Fill port
See section 3.5.2 "Replacing the fill port" on page 3-4.

T&D tubing set


See section 3.5.10 "Replacing the T&D tubing set" on page 3-7.

Pump head
See section 3.14.2 "Replacing the pump head" on page 3-41.

PP pump tube
See section 3.14.3 "Replacing the pump tube" on page 3-42.

Cuvette
See section 3.12.7 "Replacing the cuvette" on page 3-34.

Cuvette seals
See section 3.12.8 "Replacing the cuvette seals" on page 3-34.

Hemolyzer tube
See section 3.12.11 "Replacing the Hemolyzer tube" on page 3-36.

FMS tubing set


See section 3.6.3 "Replacing the FMS tubing set" on page 3-13.

7-4 Service Manual, Roche OMNI S, Version 4.0, December 2004


7 Maintenance

Waste separator
See section 3.15.3 "Replacing the waste separator" on page 3-43.

Bypass nipple
See section 3.5.6 "Replacing the bypass nipple" on page 3-5.

Bacteria filter
See section 3.15.8 "Replacing the bacteria filter" on page 3-45.

T&D disk
See section 3.5.9 "Replacing the T&D disk" on page 3-6.

Tubing set tHb/PolyOX


See section 3.13.2 "Replacing the tubing set tHb/PolyOX" on page 3-38.

The tubing paths must also be disinfected annually and the baro value must be
checked.

7.2.2 Replace every three years


Various components of the OMNI S, such as T&D module, tubes, waste container,
etc., contain liquid biological residues after use that may cause infection.
Handle these components with care and avoid contact with skin.Always wear gloves
when working with the device!
The tubes may drip a little after being disconnected. Remove excess fluids with a
clean, absorbent cloth.

Needle
See section 3.5.4 "Replacing the needle" on page 3-4.

Fill port
See section 3.5.2 "Replacing the fill port" on page 3-4.

T&D tubing set


See section 3.5.10 "Replacing the T&D tubing set" on page 3-7.

Service Manual, Roche OMNI S, Version 4.0, December 2004 7-5


7 Maintenance

Pump head
See section 3.14.2 "Replacing the pump head" on page 3-41.

PP pump tube
See section 3.14.3 "Replacing the pump tube" on page 3-42.

Cuvette
See section 3.12.7 "Replacing the cuvette" on page 3-34.

Cuvette seals
See section 3.12.8 "Replacing the cuvette seals" on page 3-34.

Hemolyzer tube
See section 3.12.11 "Replacing the Hemolyzer tube" on page 3-36.

FMS tubing set


See section 3.6.3 "Replacing the FMS tubing set" on page 3-13.

Waste separator
See section 3.15.3 "Replacing the waste separator" on page 3-43.

Bypass nipple
See section 3.5.6 "Replacing the bypass nipple" on page 3-5.

Bacteria filter
See section 3.15.8 "Replacing the bacteria filter" on page 3-45.

T&D disk
See section 3.5.9 "Replacing the T&D disk" on page 3-6.

Tubing set tHb/PolyOX


See section 3.13.2 "Replacing the tubing set tHb/PolyOX" on page 3-38.

Sample distributor cartridge


See section 3.7.1 "Replacing the sample distributor cartridge" on page 3-13.

7-6 Service Manual, Roche OMNI S, Version 4.0, December 2004


7 Maintenance

Measuring chamber cartridge


See section 3.11.1 "Replacing the measuring chamber cartridge" on page 3-25.

Hb cartridge
See section 3.13.4 "Replacing the Hb cartridge" on page 3-39.

Waste tubing
3.16.5 "Replacing the waste tube" on page 3-50.

Vacuum pump head


See section 3.16.3 "Replacing the vacuum pump head" on page 3-49.

VP tubing set
See section 3.16.7 "Replacing the VP tube set" on page 3-51.

The tubing paths must also be disinfected annually and the baro value must be
checked.

7.3 AutoQC module

7.3.1 Yearly replacement

AQC maintenance kit


The AQC maintenance kit consists of the following components
• Tube for magnetic valve
Replacement: remove the tube at the angle nipples and replace it with a new one.
• Sample tube complete
Replacement: See section 4.10.1 "Replacing the AQC sample tube complete" on
page 4-9.
• Wasch port
Replacement: See section 4.16 "Replacing the wash port" on page 4-16.

Service Manual, Roche OMNI S, Version 4.0, December 2004 7-7


7 Maintenance

7-8 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

8 Troubleshooting
8.1 Important Notes .............................................................................................................. 8-1

8.2 System stops .................................................................................................................... 8-1

8.3 Module stops ................................................................................................................. 8-15

8.4 System warnings ........................................................................................................... 8-24

8.5 Value flags ...................................................................................................................... 8-27

8.6 Important test routines ................................................................................................ 8-53


8.6.1 Contact path test ........................................................................................................................................ 8-53

8.6.2 Aggregates ................................................................................................................................................... 8-55

8.6.3 Control sensors ........................................................................................................................................... 8-57

8.6.4 Meas. sensors.............................................................................................................................................. 8-58

8.6.5 General fluidics test................................................................................................................................... 8-63

8.6.6 Stability monitor .......................................................................................................................................... 8-75

8.6.7 General hardware test .............................................................................................................................. 8-78

8.7 Sensor limits (Sensor report) ..................................................................................... 8-82

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-I


8 Troubleshooting

8-II Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

8 Troubleshooting
8.1 Important Notes

Always check all cable and tube connections for the relevant component before
starting disassembly! Serious error messages are frequently the result of loose or
incorrectly attached cable and tube connections.

Risk of infection! Always wear gloves to work on the device! If a manual inspection
of tubes is required, always wear eye protection, because fluids may be sprayed at
high pressure form tube connections!

8.2 System stops


Error 10001

"Flap W"
• Check docking mechanism for waste again. Open and close several times.
• Replace docking mechanism for waste
(see section 3.15.4 "Replacing the waste docking mechanism" on page 3-43).

Fehler 10002

"Out of operation"
• Perform installation procedures (see "Instructions for Use" Chapter 1, "Introduction",
section "Installation").

Error 10005

"Comm. error PC-µC"


• Communications between PC and microcontroller is interrupted. Turn the instrument
off and on again.
• After replacing the Screen/PC-unit or the hard disk, assign the valid instrument type
and load the default settings (see 6.1.5 "Instrument type" on page 6-2).

Error 10006

"Download error"
• The download process to the microcontroller failed. Turn the instrument off, wait at
least 3 minutes, and turn it back on.
• Reinstall software.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-1


8 Troubleshooting

• Replace the mainboard unit


(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).
• Check AQC/PolyOX cable for damage and replace if necessary.

Error 10008

• Waste container is missing. Insert container.

Errors 10009, 10010

"Current overload"
• A component has a short circuit.

Error 10012

"µC memory overrun"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.

Error 10016

"Bottle compartment cover open"


• Check functioning of microswitch module and replace if necessary
(see section 3.15.7 "Replacing the micro switch module" on page 3-45).

Errors 10017, 10018

10017: "Bypass wash error 1"


10018: "SD wash error 1"
• Call up "Wash" again.
• Clean wash water nozzle.
• Replace the T&D tubing set
(see section 3.5.10 "Replacing the T&D tubing set" on page 3-7).

Error 10019

"Waste container full"


• Insert empty waste container or enter correct fill level.

8-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Error 10020

"Rinse bottle empty"


• The customer may have refilled the wash water container. Insert new wash water con-
tainer.

Error 10021

"Flap S1"
• Check docking mechanism for wash water again. Open and close several times.
• Replace docking mechanism for wash water
(see section 3.15.5 "Replacing the docking mechanism S1 (rinse solution)"
on page 3-44).

Error 10022

"µC Reset"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.

Error 10024

"No wash-water container"


• The wash water container transponder is not installed, is damaged or not labeled. Insert
new wash water container.

Error 10025

"Check date and time"


• Input current date.

Errors 10026, 10027

10026: "Bypass wash error 2"


10027: "SD wash error 2"
• Call up "Wash" again.
• Check waste tube and replace if necessary
(see section 3.16.5 "Replacing the waste tube" on page 3-50).
• Check sample inlet path and replace if necessary
(see section 3.5.5 "Replacing the sample inlet path" on page 3-5).
• Replace the T&D tubing set
(see section 3.5.10 "Replacing the T&D tubing set" on page 3-7).
• Insert new wash-water container.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-3


8 Troubleshooting

Error 10028

• Insert new S1 rinse solution bottle.

Error 10029

"S1 expired"
• Insert new S1 rinse solution bottle.

Error 10032

"Microcontroller communications error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.

Error 10033

"COOX communications error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the PolyOX KX control board
(see section 3.12.3 "Replacing the PolyOX-KX control board" on page 3-29).

Error 10034

"AQC communications error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the AQC board
(see section 4.5 "Replacing the AQC board" on page 4-5).

Error 10039

"Vacuum reduction not ok"


• Change valve V13
(see section 3.17.3 "Replacing the waste valves V13/V18" on page 3-56).
• Start vacuum pump test: if the pump output is more than 20% over or under the refer-
ence value, the entire vacuum pump must be replaced
(see section 3.16.1 "Replacing the vacuum pump completely" on page 3-46).

8-4 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Error 10040

"Vacuum error"
• Start general fluidics test: if there is a leak between the T&D module and the vacuum
pump, it will be detected here. The relevant component must be replaced.
• Start vacuum pump test: if the pump output is more than 20% over or under the refer-
ence value, the entire vacuum pump must be replaced
(see section 3.16.1 "Replacing the vacuum pump completely" on page 3-46).

Error 10041

"Waste error 1"


• Insert new waste container.
• The waste docking mechanism is contaminated or damaged. It must be cleaned or
replaced
(see 3.15.4 "Replacing the waste docking mechanism" on page 3-43).

Error 10042

"Waste error 2"


• The waste ventilation path is blocked. Insert new waste container.

Error 10043

"Sample sep. sensor error"


• In the sensor test check the functioning of the waste container sensor and if necessary
replace it.
• Replace the vacuum control board
(see section 3.16.8 "Replacing the vacuum control board" on page 3-52).

Errors 10044

"Sample sep. error"


• Insert new waste container.
Waste separator empty:
• In the sensor test check the functioning of the waste container sensor and if necessary
replace the waste container sensor.
Waste separator full:
• Check waste system (valve, waste tube, waste container) for blockages and replace the
relevant component.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-5


8 Troubleshooting

Error 10045

"Check waste"
• Check waste fill level and enter it correctly.

Error 10046

"Waste line blocked"


• Insert new waste container.
Waste separator empty:
• In the sensor test check the functioning of the waste container sensor and if necessary
replace the waste container sensor.
Waste separator full:
• Check waste system (valve, waste tube, waste container) for blockages and replace the
relevant component.

Error 10047

"VPS error"
Vacuum pump protector full:
• Replace vacuum pump protector
(see section 3.16.6 "Replacing the vacuum pump protector" on page 3-51).
Vacuum pump protector empty:
• In the sensor test check the functioning of the VPS sensor and if necessary replace it
(see section 3.16.6 "Replacing the vacuum pump protector" on page 3-51).
• Replace the vacuum control board
(see section 3.16.8 "Replacing the vacuum control board" on page 3-52).

Errors 10048...10051

"Errors SS1...SS4"
• Call up "Wash".
• Check the corresponding sample sensor in the sensor test and if necessary calibrate by
calling up "Wash".
• Visually inspect the corresponding sample sensor for liquid residue.
Fluid present:
a) Test corresponding valves and if necessary replace them
(see section 3.17.2 "Replacing the valves from sample distributor cartridge and Hb car-
tridge" on page 3-55).
b) Check tubing system for blockages and replace the relevant components.
No fluid present:
a) Replace corresponding sample sensor
(see section 3.7.3 "Replacing the SS1, SS2, SS3 and SS4 optical sensors" on page 3-17).
b) Replace sample detection board
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

8-6 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Error 10052

"Error SS6"
• Check installation of sample inlet path.
• Call up "Wash".
• Check sample sensor in the sensor test and if necessary calibrate by calling up "Wash".
• Visually inspect the sample sensor for liquid residue.
Fluid present:
Check tubing system for blockages and if necessary replace T&D tubing set
(see section 3.5.10 "Replacing the T&D tubing set" on page 3-7).
No fluid present:
a) Replace sample sensor
(see section 3.5.7 "Replacing the optical sensor SS6" on page 3-5).
b) Replace the T&D actuator board
(see section 3.5.12 "Replacing the T&D actuator board" on page 3-9).

Error 10054

"Baro sensor error"


• Update Baro value.

Errors 10056...10097

• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check AQC/PolyOX cable for damage and replace if necessary.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).
• Check connector cable between mainboard and interface unit.
• Replace the interface unit
(see section 3.19.1 "Replacing the Interface unit" on page 3-61).

Error 10098

"Timing error COOX measurement."


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check AQC/PolyOX cable for damage and replace if necessary.
• Replace the PolyOX KX control board
(see section 3.12.3 "Replacing the PolyOX-KX control board" on page 3-29).
• Check connector cable between mainboard and interface unit.
• Replace the interface unit.
(see section 3.19.1 "Replacing the Interface unit" on page 3-61).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-7


8 Troubleshooting

Errors 10099...10106

"Timing error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check AQC/PolyOX cable for damage and replace if necessary.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).
• Check connector cable between mainboard and interface unit.
• Replace the interface unit
(see section 3.19.1 "Replacing the Interface unit" on page 3-61).

Error 10107

"Timing error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check AQC/PolyOX cable for damage and replace if necessary.
• Replace the PolyOX KX control board
(see section 3.12.3 "Replacing the PolyOX-KX control board" on page 3-29).
• Check connector cable between mainboard and interface unit.
• Replace the interface unit
(see section 3.19.1 "Replacing the Interface unit" on page 3-61).

Errors 10108...10128

• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check AQC/PolyOX cable for damage and replace if necessary.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).
• Check connector cable between mainboard and interface unit.
• Replace the interface unit
(see section 3.19.1 "Replacing the Interface unit" on page 3-61).

Error 10130

"Timing error COOX"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check AQC/PolyOX cable for damage and replace if necessary.
• Replace the PolyOX KX control board
(see section 3.12.3 "Replacing the PolyOX-KX control board" on page 3-29).
• Check connector cable between mainboard and interface unit.
• Replace the interface unit
(see section 3.19.1 "Replacing the Interface unit" on page 3-61).

8-8 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Errors 10152...10154

"BG/ISE/MSS Measuring chamber open"


• Check all cable plug connectors on the relevant board.
• Replace affected measuring chamber.

Error 10155

"AQC cover open"


• Check all plug connectors on the AQC board.
• Check the magnet and the magnetic switch on the AQC cover.

Errors 10160...10261

"File error"
• Reinstall software.
• Replace the hard disk (see section 3.20.2 "Replacing the hard disk" on page 3-68).

Errors 10288...10294

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Errors 10299, 10300

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the sample distributor (see section 3.7.2 "Replacing the sample distributor" on
page 3-15).

Errors 10301...10310

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.

Errors 10318...10327

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the sample distributor (see section 3.7.2 "Replacing the sample distributor" on
page 3-15).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-9


8 Troubleshooting

Error 10340

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the T&D actuator board
(see 3.5.12 "Replacing the T&D actuator board" on page 3-9).

Errors 10341, 10342

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 10343

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the vacuum control board
(see section 3.16.8 "Replacing the vacuum control board" on page 3-52).

Error 10346

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the HSIN board
(see section 3.12.12 "Replacing the HSIN board" on page 3-36).

Error 10347

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 10348

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the BG measuring chamber
(see section 3.8.1 "Replacing the BG measuring chamber" on page 3-18).

8-10 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Error 10349

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the ISE measuring chamber
(see section 3.9.1 "Replacing the ISE measuring chamber" on page 3-20).

Error 10350

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the MSS measuring chamber
(see section 3.10.1 "Replacing the MSS measuring chamber" on page 3-21).

Error 10351

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the sample distributor (see section 3.7.2 "Replacing the sample distributor" on
page 3-15).

Error 10352

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the FMS control board
(see section 3.6.2 "Replacing the FMS control board" on page 3-13).

Errors 10353, 10388

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace Screen/PC unit
(see section 3.20.1 "Replacing the Screen/PC Unit" on page 3-64).

Error 10389

"Program error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the printer (see section 3.21.1 "Replacing the Printer" on page 3-75).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-11


8 Troubleshooting

Errors 10416...10426

"Hardware conflict / Error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 10427

"Hardware error"
• Replace the T&D actuator board
(see section 3.5.12 "Replacing the T&D actuator board" on page 3-9).

Errors 10428, 10429

"Hardware error"
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 10430

"Hardware error"
• Replace the vacuum control board
(see section 3.16.8 "Replacing the vacuum control board" on page 3-52).

Error 10431

"Hardware error"
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 10433

"Hardware error"
• Replace the HSIN board
(see section 3.12.12 "Replacing the HSIN board" on page 3-36).

Error 10434

"Hardware error"
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

8-12 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Error 10435

"Hardware error"
• Replace the BG measuring chamber
(see section 3.8.1 "Replacing the BG measuring chamber" on page 3-18).

Error 10436

"Hardware error"
• Replace the ISE measuring chamber
(see section 3.9.1 "Replacing the ISE measuring chamber" on page 3-20).

Error 10437

"Hardware error"
• Replace the MSS measuring chamber
(see section 3.10.1 "Replacing the MSS measuring chamber" on page 3-21).

Error 10438

"Hardware error"
• Replace the sample distributor (see section 3.7.2 "Replacing the sample distributor" on
page 3-15).

Error 10439

"Hardware error"
• Replace the FMS control board
(see section 3.6.2 "Replacing the FMS control board" on page 3-13).

Error 10440

"Hardware error"
• Replace the Screen/PC unit
(see section 3.20.1 "Replacing the Screen/PC Unit" on page 3-64).

Errors 10441...10452

"Hardware error"
• Replace the sample distributor (see section 3.7.2 "Replacing the sample distributor" on
page 3-15).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-13


8 Troubleshooting

Errors 10456...10467

"EEPROM error"
• Replace the sample distributor (see section 3.7.2 "Replacing the sample distributor" on
page 3-15).

Error 10468

"EEPROM error"
• Replace the FMS control board
(see section 3.6.2 "Replacing the FMS control board" on page 3-13).

Error 10469

"EEPROM error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the vacuum control board
(see section 3.16.8 "Replacing the vacuum control board" on page 3-52).

Error 10470

"Transponder error"
• Remove and replace waste bottle, S1 rinse solution, S2 and S3 fluid pack.
• Check all cable plug connectors of the transponder antennas.
• Replace the Transponder Control waste, S1, S2 or S3 (see section 3.15.9 "Replacing the
Transponder Control (waste, S1, S2, S3)" on page 3-46).
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 10471

"T&D error"
• Press "Initialization".
• T&D disk, fill port or plug control are very contaminated or damaged. Clean the rele-
vant components or if necessary replace.
• Replace the T&D module (see section 3.5.1 "Replacing the T&D module" on page 3-2).

Error 10472

"Transponder error S1"


• Pull out waste container and put it back in.
• Insert new waste container.

8-14 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

• Replace transponder control S1 (see section 3.15.9 "Replacing the Transponder Control
(waste, S1, S2, S3)" on page 3-46).

Error 10488

"HW test error"


• Call up General hardware test and replace damaged parts.

Error 10489

"Remote lock"
Instrument locked by OMNILINK.
• To remove this error press "Continue". Log on as a user with the privileg "Remote
unlock". More information on user management, see Reference Manual Chapter 3,
„Software modes“, section „Security“.

8.3 Module stops


Error 20000

"SD temp. error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Call up General hardware test and check the sample distributor voltage.
Voltage = 5V:
Replace sample distributor
(see section 3.7.2 "Replacing the sample distributor" on page 3-15).
Voltage < 5V:
Replace power supply and sample distributor
(see section 3.7.2 "Replacing the sample distributor" on page 3-15).

Error 20016

"FMS volume error"


• Aspirate solution B and call up the general fluidics test. After removing possible faults,
call up the mixture calibration.
• Replace S2 fluid pack.
• Replace pump tube of main pump
(see Chapter 7: 3.14.3 "Replacing the pump tube" on page 3-42).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-15


8 Troubleshooting

• Test functioning of BG measuring chamber by electrode simulator:


Simulator test OK:
a) Replace SCon.
b) Replace fluid mixing system
(see section 3.6.1 "Replacing the FMS" on page 3-11).
Simulator test nOK:
see next item.
• Test functioning of the ISE measuring chamber by electrode simulator:
Simulator test OK:
Replace BG measuring chamber
(see section 3.8.1 "Replacing the BG measuring chamber" on page 3-18).
Simulator test nOK:
Replace mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20017

"FMS error"
• Aspirate solution B and solution A and call up the general fluidics test. Install a new
S2 fluid pack after correcting any errors.
• Use fluid actions to aspirate solution A and B.
Evaluation OK:
a) Replace FMS control board
(see section 3.6.2 "Replacing the FMS control board" on page 3-13).
b) Replace the fluid mixing system
(see section 3.6.1 "Replacing the FMS" on page 3-11).
If these actions cannot correct the errors, see the next item.
Evaluation nOK:
Check tubing paths for solutions A and B for blockages and replace the relevant
component.

Error 20018

"Pack S2 empty"
• Install a new S2 fluid pack.
• Download log file and send instrument for repair.

Error 20019

"Flap error S2"


• Replace S2 docking mechanism
(see section 3.15.6 "Replacing the docking mechanisms S2 and S3 (fluid packs)"
on page 3-44).

8-16 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Error 20020

"No pack S2"


• Check plug connectors of the transponder antenna cable.
• Replace transponder antenna.
• If "Transponder Error" (no. 20022, 20094) occurs simultaneously with this error,
replace mainboard unit (see section 3.18.1 "Replacing the Mainboard unit" on page 3-
59).

Error 20021

"ADC error cond BG/ISE"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20022

"Transponder error S2"


• Remove and replace S2 fluid pack.
• Check plug connectors of the transponder antenna cable.
• Replace transponder antenna.
• If multiple "Transponder Errors" (no. 20022, 20094) occur, replace mainboard unit (see
section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20023

"S2 on board time expired"


• Insert new S2 fluid pack.

Error 20024

"HW test error FMS"


• Call up General hardware test and replace damaged parts.

Errors 20032...20034

"BG temp. error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Call up general hardware test and check the BG measuring chamber temperature.
nok:
Replace power supply and BG measuring chamber
(see section 3.8.1 "Replacing the BG measuring chamber" on page 3-18).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-17


8 Troubleshooting

• Check cable press fit of the BG connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20035

"BG temp. error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check function of fan unit and if necessary replace it
(see section 3.22.1 "Replacing the cooling fan unit" on page 3-76).
• Replace the BG measuring chamber
(see section 3.8.1 "Replacing the BG measuring chamber" on page 3-18).
• Check cable press fit of the BG connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20037

"ADC error BG"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the BG measuring chamber
(see section 3.8.1 "Replacing the BG measuring chamber" on page 3-18).
• Check cable press fit of the BG connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Errors 20048...20050

"ISE temp. error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Call up general hardware test and check the ISE measuring chamber temperature.
nok:
Replace power supply and ISE measuring chamber
(see section 3.9.1 "Replacing the ISE measuring chamber" on page 3-20).
• Check cable press fit of the BG connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20051

"ISE temp. error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check function of fan unit and if necessary replace unit
(see section 3.22.1 "Replacing the cooling fan unit" on page 3-76).

8-18 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

• Replace the ISE measuring chamber


(see section 3.9.1 "Replacing the ISE measuring chamber" on page 3-20).
• Check cable press fit of the BG connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20053

"ADC error ISE"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the ISE measuring chamber
(see section 3.9.1 "Replacing the ISE measuring chamber" on page 3-20).
• Check cable press fit of the BG connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20054

"HW test error ISE"


• Call up "General hardware test" and replace damaged parts.

Errors 20065...20067

"COOX lamp error"


• Call up polychromator calibration.
• Replace the lamp unit Bilirubin
(see section 3.12.2 "Replacing the lamp unit bilirubin" on page 3-28).
• Replace the PolyOX KX control board
(see section 3.12.3 "Replacing the PolyOX-KX control board" on page 3-29).

Errors 20068, 20069

"COOX lamp error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the PolyOX KX control board
(see section 3.12.3 "Replacing the PolyOX-KX control board" on page 3-29).

Error 20071

"COOX HW error"
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the Polychromator Bilirubin
(see section 3.12.4 "Replacing the Polychromator Bilirubin" on page 3-30).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-19


8 Troubleshooting

• Replace the PolyOX KX control board


(see section 3.12.3 "Replacing the PolyOX-KX control board" on page 3-29).

Error 20072

"COOX temp error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the cuvette holder
(see section 3.12.6 "Replacing the cuvette holder" on page 3-33).
• Replace the HSIN board
(see section 3.12.12 "Replacing the HSIN board" on page 3-36).

Fehler 20073

„HW test error Hemolyzer“


• Call up "General hardware test" and replace damaged parts.

Error 20080

"PP error"
• To correct errors replace tubing at the MSS outlet pump and call up MSS system cali-
bration.
• Replace pump head of MSS outlet pump
(see section 3.14.2 "Replacing the pump head" on page 3-41).
• Replace MSS outlet pump
(see section 3.14.1 "Replacing a peristaltic pump" on page 3-40).
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20081

"MSS pol. running"


• The phase of the MSS polarization is not yet completed. With the automatic performed
system calibration the sensor will become "Ready".

Error 20082

"MSS pol. not ok"


• Repeat MSS polarization.
• Replace the MSS measuring chamber
(see section 3.10.1 "Replacing the MSS measuring chamber" on page 3-21).

8-20 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

• Check SS1 to SS4 sample sensors in the sensor test and if necessary calibrate by calling
up "Wash".
• Check SS1 to SS4 sample sensors visually for fluid residues.
Fluid present:
a) Test corresponding valves and if necessary replace them
(see section 3.17.2 "Replacing the valves from sample distributor cartridge and Hb car-
tridge" on page 3-55).
b) Check tubing system for blockages and replace the relevant components.
No fluid present:
a) Replace corresponding sample sensor
(see section 3.7.3 "Replacing the SS1, SS2, SS3 and SS4 optical sensors" on page 3-17).
b) Replace sample detection board
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20083

"MSS pol. not ok"


• Repeat MSS polarization.

Error 20084

"HW test error MSS"


• Call up General hardware test and replace damaged parts.

Error 20085

"Pack S3 empty"
• Install a new S3 fluid pack.
• Download log file and send instrument for repair.

Error 20086

"Flap error S3"


• Replace S3 docking mechanism
(see section 3.15.6 "Replacing the docking mechanisms S2 and S3 (fluid packs)"
on page 3-44).

Errors 20087...20089

"MSS temp. error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Call up general hardware test and check the MSS measuring chamber voltage.
Voltage = 5V:
Replace BG measuring chamber
(see section 3.10.1 "Replacing the MSS measuring chamber" on page 3-21).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-21


8 Troubleshooting

Voltage < 5V:


Replace power supply and BG measuring chamber
(see section 3.10.1 "Replacing the MSS measuring chamber" on page 3-21).
• Check cable press fit of the BG connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20091

"No pack S3"


• Check plug connectors of the transponder antenna cable.
• Replace transponder antenna.
• If "Transponder Error" (no. 20022, 20094) occurs simultaneously with this error,
replace mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Errors 20092, 20093

"ADC error MSS"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the MSS measuring chamber
(see section 3.10.1 "Replacing the MSS measuring chamber" on page 3-21).
• Check cable press fit of the BG connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20094

"Transponder error S3"


• Remove and replace S3 fluid pack.
• Check plug connectors of the transponder antenna cable.
• Replace transponder antenna.
• If multiple "Transponder Errors" (no. 20022, 20094) occur, replace mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20095

• Insert new S3 fluid pack.

Error 20096

"tHb/SO 2 temperature error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.

8-22 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

• Call up general hardware test and check the temperature of the tHb/SO 2 module.
nok:
Replace power supply and tHb/SO 2 module
(see section 3.13.1 "Replacing the tHb/SO2 module" on page 3-38).
• Check cable press fit of the tHb/SO 2 connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20097

"ADC error tHb"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace the tHb/SO 2 module
(see section 3.13.1 "Replacing the tHb/SO2 module" on page 3-38).
• Check cable press fit of the tHb/SO 2 connector cable on the mainboard for damage.
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Error 20112

"SS5 error"
• Call up "Wash AutoQC".
• Check SS5 sample sensor in the sensor test and if necessary calibrate by calling up
"Wash".
• Visually inspect the sample sensor for fluid residue.
Fluid present:
Check tubing system for blockage and replace relevant components.
No fluid present:
Replace Z-distributor board
(see section 4.7 "Replacing the Z distributor board" on page 4-7).
• Replace the mainboard unit
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Errors 20113...20118

"module error AQC"


• Call up "Wash AutoQC".
• Check all plug connectors on the AQC board.
• Replace the AQC board (see section 4.5 "Replacing the AQC board" on page 4-5).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-23


8 Troubleshooting

Error 20120

"AQC wash error"


• Call "wash AQC" and visually check whether a water-air mixture is being aspirated
through the wash tube.
Water only: wash nozzle is blocked; clean or replace nozzle.
Air only: check function of AQC motors and if necessary replace them.
• Check wash port for damage and if necessary replace it
(see section 4.16 "Replacing the wash port" on page 4-16).
• Check AQC needle for damage or blockage and if necessary replace it
(see section 4.14 "Replacing the AutoQC steel tube" on page 4-14).
• Replace the T&D disk (see section 3.5.9 "Replacing the T&D disk" on page 3-6).

Errors 20122...20141

"AQC pos. error"


• Call up "Wash AutoQC".
• Check function of AQC motors and if necessary replace them.

Error 20142

"AQC temp. error"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check room temperature.
• Replace temperature sensor
(see section 4.15 "Replacing the AQC temperature sensor" on page 4-15).
• Replace the AQC board (see section 4.5 "Replacing the AQC board" on page 4-5).

8.4 System warnings


Warning 30005

"Cl electrode is contaminated (defective)"


• Call up "Start internal cleaning" for the ISE measuring chamber.
• Call up "Start external cleaning" for the ISE measuring chamber.
• Replace Cl electrode.

Warning 30008

"Pack S2 level is low"


• If necessary, insert a new S2 fluid pack (see Instructions for Use Chapter 6 "Mainte-
nance", section "Exchange of solutions and packs").

8-24 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Warning 30009

"Pack S3 level is low"


• If necessary, insert a new S3 fluid pack (see Instructions for Use Chapter 6 "Mainte-
nance", section "Exchange of solutions and packs").

Warning 30010

"Rinse level is low"


• If necessary, insert a new S1 rinse solution (see Instructions for Use Chapter 6 "Mainte-
nance", section "Exchange of solutions and packs").

Warning 30011

"Waste is nearly full"


• If necessary, replace and/or empty the waste container (see Instructions for Use Chap-
ter 6 "Maintenance", section "Exchange of solutions and packs").

Warning 30012...30025

"Preparation nOK"
• Call up fill routine for the corresponding solution.
• Depending on the solution insert new S1 rinse solution or S2/S3 fluid pack.
• Check tubing system with the general fluidics test or manually for damage or blockage
according the corresponding diagrams in Chapter 5 and replace the relevant compo-
nent.

Warning 30028

"Check AQC material"


• If necessary, insert a new mat (see Instructions for Use Chapter 6 "Maintenance", sec-
tion "Changing of AutoQC mats").

Warning 30029

"Check parameter status"


• To correct the error perform "Calibration for Ready":
System > Calibrations.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-25


8 Troubleshooting

Warning 30030

"Perform man. QC measurement"


• Perform QC measurement.

Warning 30031

"Perform maintenance"
• Perform the maintenance (see Instructions for Use Chapter 6 "Maintenance").

Warning 30032

"Printer error - Lever open"


• Close printer lever (see Instructions for Use Chapter 6 "Maintenance", section "Repla-
cing printer paper").
• Check printer connector cable.
• Replace printer (see section 3.21.1 "Replacing the Printer" on page 3-75).

Warning 30033

"Printer error - Paper empty"


• Insert printer paper (see Instructions for Use Chapter 6 "Maintenance", section "Repla-
cing printer paper").
• Check printer connector cable.
• Replace printer (see section 3.21.1 "Replacing the Printer" on page 3-75).

Warning 30034

"HW test error fan"


• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check function of fan unit and if necessary replace unit
(see section 3.22.1 "Replacing the cooling fan unit" on page 3-76).

8-26 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

8.5 Value flags


Flag no. 1000

"Sample pos. error (1)"


No constant conductivity over the distance MCI-MCO when positioning in the measuring
chamber.
Possible causes:
- Blockage or leakage
- Leaky electrodes
- Solution is fragmented
- Solution is unhomogenous
- Separation of electrode membrane
Troubleshooting:
• If discoloration can be detected by sample material in the internal electrolyte of the
electrode, it must be replaced.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Call up "Contact path test" and aspirate solution B (see contact path test)
• Insert electrode simulator in the relevant module and act accordingly.

Flag no. 1001

"Sample pos. error (2)"


Conductivity over the distance MCI-MCO too low when positioning in the measuring
chamber.
Possible causes:
- Blockage or leaks
- Difficulty in aspiration of solution B because of FMS problems
Troubleshooting:
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Replace S2 fluid pack.
• Check/replace S2 docking mechanism (see section 3.15.6 "Replacing the docking mech-
anisms S2 and S3 (fluid packs)" on page 3-44).
• Replace the FMS unit (see section 3.6.1 "Replacing the FMS" on page 3-11).
• Replace the FMS board
(see section 3.6.2 "Replacing the FMS control board" on page 3-13).

Flag no. 1002

"Sample sep. error (1)"


No air packet was detected between calibration solution and sample.
Possible causes:
- Deposits or blockage in the measuring chamber

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-27


8 Troubleshooting

Troubleshooting:
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Test/replace V4.
• Check/replace air tube at V4.

Flags no. 1003, 1004

"Ref. sol. asp. error"


"Ref. sol. pos. error"
No reference solution was detected.
Possible causes:
- Blockage or leaks in the reference system
Troubleshooting:
• Call up "Fill routines" and then "Fill reference electrode" for the relevant module.
• Replace S2/S3 fluid pack.
• Check plug connector of the reference tube.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Replace reference electrode.

Flag no. 1005

"End of sample detected"


SS1 detects the end of the sample, but the measuring chamber is not yet filled:
Possible causes:
- Irregular sample
- Insufficient sample
- Improper sample
Troubleshooting:
• Repeat measurement/calibration, ensure sufficient and homogenous sample input.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Test/replace V4.
• Check/replace air tube at V4.

Flag no. 1006

"Irregular sample (1)"


The measuring chamber is not homogeneously filled.
Possible causes:
- Irregular sample
- Leakage in measuring channel
Troubleshooting:
• Repeat measurement/calibration, ensure sufficient and homogenous sample input.

8-28 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

• If discoloration can be detected by sample material in the internal electrolyte of the


electrode, it must be replaced.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Call up "Contact path test" and aspirate solution B (see contact path test)
• Insert electrode simulator in the relevant module and act accordingly.

Flag no. 1007

"Cuvette not empty"


A sample detection was not possible:
Possible causes:
- Deposits in the cuvette
- Light level too low
- Pump delivery volume too high
Troubleshooting:
• Run "Polychromator calibration".
• Run "Internal cleaning".
• Replace PP tube (main pump).

Flag no. 1008

"Irregular sample (2)"


The conductivity changed after closing the valve at the measuring chamber input.
Possible causes:
- Insufficient wetting in measuring channel
- Separation of electrode membrane
- Leakage in electrode
Troubleshooting:
• Perform "Calibration for Ready".
• Call up "Wetting routine" for the relevant module.
• Check electrodes for leakage and if necessary replace.
• Check sample distributor-measuring chamber connection for leakage.
• Test/replace measuring chamber inlet valve
(see section 3.17.2 "Replacing the valves from sample distributor cartridge and Hb car-
tridge" on page 3-55).

Flag no. 1009

"Sample pos. error (3)"


The measuring signal was not constant or the measuring channel was not filled homoge-
neously.
Possible causes:
- Blockage or leakage

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-29


8 Troubleshooting

- Leaky electrodes
- Sample is fragmented
- Sample is unhomogenous
- Separation of the electrode membrane
Troubleshooting:
• If discoloration can be detected by sample material in the internal electrolyte of the
electrode, it must be replaced.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Call up "Contact path test" and aspirate solution B (see contact path test)
• Insert electrode simulator in the relevant module and act accordingly.

Flag no. 1010

"Sample pos. error (4)"


The conductivity was too high.
Possible causes:
- Standby solution aspiration problems
- PP tube (MSS outlet) defective
Troubleshooting:
• Check/replace PP tube (MSS outlet).
• Call up fill routines "Aspirate standby solution" and "Aspirate reference solution".
• Call up "Contact path test" and aspirate standby solution (see contact path test).
• Replace S3 fluid pack.

Flag no. 1011

"Sample pos. error (5)"


No constant conductivity at sensor input was detected.
Possible causes:
- Blockage or leakage
- PP tube (MSS outlet) defective
Troubleshooting:
• Check/replace PP tube (MSS outlet).
• Call up fill routines "Aspirate standby solution" and "Aspirate reference solution", then
run "Calibration for ready".
• Run "MSS polarization".
• Call up "wetting routine" for MSS.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Call up "Contact path test" and aspirate standby solution (see contact path test).
• Insert electrode simulator in the relevant module and act accordingly.
• Replace MSS cassette.

8-30 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Flag no. 1012

"Sample pos. error (6)"


Conductivity at sensor input was too low.
Possible causes:
- Blockage
- PP tube (MSS outlet) defective
- No fluid in the measuring chamber
Troubleshooting:
• Check/replace PP tube (MSS outlet).
• Call up fill routines "Aspirate standby solution" and "Aspirate reference solution", then
run "Calibration for ready".
• Run "MSS polarization".
• Call up "Wetting routine" for MSS.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Replace MSS cassette.

Flag no. 1013

"Sample sep. error (2)"


No air packet was detected between calibration solution and sample. The sample was not
aspirated or aspirated too late.
Possible causes:
- Deposits or blockage in measuring path.
- PP tube (MSS outlet) defective
Troubleshooting:
• Check/replace PP tube (MSS outlet).
• Call up fill routines "Aspirate standby solution" and "Aspirate reference solution", then
run "Calibration for ready".
• Run "MSS polarization".
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Replace MSS cassette.

Flag no. 1014...1017

"Sample pos. error (7...10)"


Conductivity at sensor input was too low after positioning the sample /calibration solution
in the measuring chamber:
Possible causes:
- Unhomogenous sample/calibration solution
- Insufficient sample/calibration solution detected
- Measuring path insufficiently wetted

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-31


8 Troubleshooting

Troubleshooting:
• Ensure that sample input is sufficient and homogenous.
• Run "MSS polarization".
• Call up "Wetting routine" for MSS.

Flag no. 1020

"Sample distribution error (1)"


SS1 did not detect a solution.
Possible causes:
- Blockage or leakage during positioning.
Troubleshooting:
• Call up "Wash".
• Check that bypass nipple is correctly seated.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Call up "Component test > Control sensors > Sample sensors" and "Calibrate"
(see Reference Manual section 3.3.1 "Component test").
• Replace the SD cartridge
(see section 3.7.1 "Replacing the sample distributor cartridge" on page 3-13).
• Replace sample detection board
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Flag no. 1021

"Sample distr. error (2)"


No solution detected at SS3 although SS2 detected a solution.
Possible causes:
- Blockage or leakage
Troubleshooting:
• Call up "Wash".
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Call up "Component test > Control sensors > Sample sensors" and "Calibrate"
(see Reference Manual section 3.3.1 "Component test").
• Replace the Hb cartridge (see section 3.13.4 "Replacing the Hb cartridge" on page 3-39).
• Replace cuvette.

Flag no. 1022

"Sample distr. error (3)"


Sample excess could not be extracted via sample distributor.
Possible causes:
- Blockage

8-32 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Troubleshooting:
• Call up "Wash".
• Check that bypass nipple is correctly seated.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Call up "Component test > Control sensors > Sample sensors" and "Calibrate" (SS2)
(see Reference Manual section 3.3.1 "Component test").
• Replace sample detection board
(see section 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Flag no. 1023

"Sample distr. error (4)"


SS4 did not detect a solution.
Possible causes:
- Blockage or leakage
- Insufficient sample
Troubleshooting:
• Ensure that there is sufficient sample.
• Call up "Wash".
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Call up "Component test > Control sensors > Sample sensors" and "Calibrate" (SS4)
(see Reference Manual section 3.3.1 "Component test").
• Replace the SD cartridge
(see section 3.7.1 "Replacing the sample distributor cartridge" on page 3-13).

Flags no. 1025, 1026

"No sample detected (2)"


"No sample detected (3)"
No constant conductivity of the sample was detected:
Possible causes:
- Fragmented sample in sample distributor
Troubleshooting:
• Ensure sample entry is free of air bubbles.

Flag no. 1027

"No sample detected (4)"


No sample detected in COOX module.
Possible causes:
- Sample path leaks or is blocked
- Cuvette fill leaks
- Main pump tube defective

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-33


8 Troubleshooting

Troubleshooting:
• Call up "Wash".
• Check/replace main pump PP tube.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.

Flag no. 1028...1031

"Insufficient sample (1...4)"


Insufficient sample for module was detected:
Possible causes:
- Unhomogenous sample
- Insufficient sample
Troubleshooting:
• Ensure that sample input is sufficient and homogenous.

Flag no. 1032

"Sample distr. error (5)"


No calibration solution 1 ( CAL 1) detected:
Possible causes:
- Blockage or leakage
- Main pump PP tube defective
Troubleshooting:
• Call up "Fill routines" and run "Aspirate CAL 1".
• Check/replace main pump PP tube.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Check/replace S3 docking mechanism (see section 3.15.6 "Replacing the docking mech-
anisms S2 and S3 (fluid packs)" on page 3-44).
• Call up "Component test > Control sensors > Sample sensors" and "Calibrate" (SS4)
(see Reference Manual section 3.3.1 "Component test").

Flag no. 1036

"Insufficient sample (5)"


Insufficient sample for MSS module was detected.
Possible causes:
- Unhomogenous sample
- Insufficient sample
Troubleshooting:
• Ensure that sample input is sufficient and homogenous.
• Call up filling routines "Aspirate CAL 2, CAL 3, CAL 4".
• Call up "General fluidics test" and check that tubing paths are open and without leaks.

8-34 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

• Check/replace S3 docking mechanism (see section 3.15.6 "Replacing the docking mech-
anisms S2 and S3 (fluid packs)" on page 3-44).
• Call up "Component test > Control sensors > Sample sensors" and "Calibrate" (SS4)
(see Reference Manual section 3.3.1 "Component test").

Flag no. 1037

"No sample detected (8)"


No sample detected in hemolyzer.
Possible causes:
- Aspiration path leaks or is blocked
Troubleshooting:
• Call up "Hemolyzer test".
• Check/replace main pump PP tube.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Check COOX module cuvette for leaks and if necessary replace.

Flag no. 1038

"No sample detected (9)"


Wash water is not detected by the tHb/SO 2 module (calibration).
Possible causes:
- Aspiration path leaks or is blocked
Troubleshooting:
• Call up "Wash".
• Check/replace main pump PP tube.
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Check COOX module cuvette for leaks and if necessary replace.
• Call up "Component test > Control sensors > Sample sensors" and "Calibrate" (SS3)
(see Reference Manual section 3.3.1 "Component test").

Flag no. 1050

"No sample in SIP"


No sample was recognized at SS2.
Possible causes:
- Sample path leaks or is blocked
- No sample was input
Troubleshooting:
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Check seating of T&D lock.
• Check/replace bypass nipple.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-35


8 Troubleshooting

• Prevent direct sunlight on the plug control (false input).


• Call up "Component test > Control sensors > Sample sensors" and "Calibrate" (SS2)
(see Reference Manual section 3.3.1 "Component test").

Flag no. 1051

"No AQC sample detected "


No sample was recognized at SS5:
Possible causes:
- Sample path leaks or is blocked
Troubleshooting:
• Call up "Wash AutoQC".
• Check AQC ampoule assembly (incorrect programming).
• Check/replace AQC steel tube
(see section 4.14 "Replacing the AutoQC steel tube" on page 4-14).
• Call up "Component test > Control sensors > Sample sensors" and "Calibrate" (SS5)
(see Reference Manual section 3.3.1 "Component test").

Flag no. 1052

"No vacuum"
No regulated low pressure could be build up during aspiration process.
Possible causes:
- Defective vacuum system
- Incorrect Baro setup
Troubleshooting:
• Check Baro settings.
• Call up "Vacuum system" in "Control sensors" test (see vacuum test).

Flag no. 2004

"Repro. not ok"


Sensor signal cannot be reproduced.
Possible causes:
- Invalid sample type used
- Measuring path contaminated
- Bubbles in internal electrolyte of the electrodes on the membrane
- Interference signals in measuring path during measurement value capture
- Interference signals over the reference electrodes during measurement value capture
- Loss of sensitivity of an electrode

8-36 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Troubleshooting:
• Perform "Calibration for Ready".
• Call up "Wetting routine" for the relevant module.
• Remove the electrode and tap the body of the electrode with a finger nail to detach any
air bubbles from the electrode membrane.
• Call up "Stability monitor" and check "Details" of the "Electrode slope". If the elec-
trode/cartridge is defective, it is displayed. Change the corresponding electrode or car-
tridge.
• If the sample channel is visibly contaminated, start "Internal cleaning" of the module
(caution: MSS excepted).
• If the error affect potentiometric electrodes/sensors (pH, urea, all ISE electrodes), the
reference electrode must be replaced.

Flag no. 2009

"Recal. not OK"


Signal difference occurring during recalibration of PCO 2 was too large.
Possible causes:
- Invalid sample type used
Troubleshooting:
• Perform "Calibration for Ready".
• Call up "Stability monitor" and check "Details" of the "Electrode slope". If the PCO 2
electrode is faulty, it will be displayed and must be replaced.

Flag no. 2011

"Sensor signal instable"


Sensor signal cannot be reproduced.
Possible causes:
- Invalid sample type used
- Bubbles in internal electrolyte of the electrodes on the membrane
- Interference signals in measuring path during measurement value capture
- Interference signals over the reference electrodes during measurement value capture
Troubleshooting:
• Check seating of unit cover.
• Perform "Calibration for Ready".
• If this interference concerns the entire module, the reference electrode must be
replaced.
• Remove the electrode and tap the body of the electrode with a finger nail to detach any
air bubbles from the electrode membrane.
• Call up "Stability monitor" and check "Details" of the "Electrode slope". If the elec-
trode/cartridge is defective, it is displayed. Change the corresponding electrode or car-
tridge.
• Check ground cable plug connector on container compartment door.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-37


8 Troubleshooting

• Check seating of pump gasket of the MSS outlet pump.


• Insert electrode simulator and act accordingly.
• Replace affected measuring chamber.

Flag no. 2012

"Sensor signal is drifting"


The sensor signal is drifting.
Possible causes:
- Invalid sample type used
- Wash water contaminated
- Bubbles in internal electrolyte of the electrodes on the membrane
- Interference signals in measuring path during measurement value capture
- Wash water aspiration problems
Troubleshooting:
• Perform "Calibration for Ready".
• Remove the electrode and tap the body of the electrode with a finger nail to detach any
air bubbles from the electrode membrane.
• Call up "Stability monitor" and check "Details" of the "Electrode slope". If the elec-
trode/cartridge is defective, it is displayed. Change the corresponding electrode or car-
tridge.
• Call up fill routine "Aspirate rinse".
• Replace S1 bottle.
• Replace measuring chamber cartridge.
• Insert electrode simulator and act accordingly.
• Test/replace V4.
• Check/replace air tube at V4.

Flag no. 2016

"Signal too high"


The 1st sampling value of the O 2 zero point calibr. is > 60 mV
Possible causes:
- An air bubble has falsified the calibration
- Defective electrode
- Unhomogenous O 2 zero point solution
Troubleshooting:
• Perform "Calibration for Ready".
• Call up "Wetting routine" for BG.
• In the fluid actions "Aspirate O 2 zero point solution".
• Replace S2 fluid pack.
• Change electrode PO 2 .
• Insert electrode simulator and act accordingly.

8-38 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Flag no. 2021

"Drift alarm"
A drifting of the sensor signal was detected in the "Ready" state.
Possible causes:
- The electrode drifts exceed the default limit values
- Invalid sample type used
- Wetting problems
Troubleshooting:
• Perform "Calibration for Ready".
• Call up "Wetting routine" for BG.
• Call up "Stability monitor", start and check "Details" of the "Electrode slope". If the
electrode/cartridge is defective, it is displayed. Change the corresponding electrode or
cartridge.
• If the flag refers to the PO 2 electrode, call up "Aspirate MIX 1" in the fluid actions for
BG. The PO 2 sensor signal must be no greater than 50 mV within two minutes. If this is
the case, call up "Internal cleaning" for BG. In case the error was resolved, set the inter-
val for "BG cleaning" to 3 days. Perform an "External cleaning" with Deproteinizer if
necessary.
• Insert electrode simulator and act accordingly.

Flag no. 2022

"Ref. drift alarm"


A drifting of the sensor signal was detected at the reference electrode in the "Ready" state.
Possible causes:
- Bubbles interfere with reference system
- S2 Fluid Pack docking mechanism contaminated
Troubleshooting:
• In the fill routines call up "Fill reference electrode" and then run "Calibration for
ready".
• Check S2 docking mechanism for contamination and if necessary clean it.
• Replace reference electrode.
• Call up "General fluidics test" and check tubing paths (particularly the seal of the refer-
ence valves).
• Insert electrode simulator and act accordingly.

Flag no. 2023

"1 point sensitivity drifts"


The sensor has not completed the startup phase and the 1-point sensitivity is still strongly
increasing.
Troubleshooting:
• Perform "Calibration for Ready".

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-39


8 Troubleshooting

Flag no. 2024

"Air bubble detected"


Before washing out the sample, the conductivity value was not detected as pressure-stable.
Possible causes:
- A blood gas electrode has not filled bubble-free
- Leaky VBI
- Leaky sample path in BG.
Troubleshooting:
• Call up "General fluidics test" and check that tubing paths are open and without leaks.

Flag no. 2028

"Mean not OK"


Measurement value of calibration is outside the expected range.
Possible causes:
- Defective electrode
- Incorrect Baro value
Troubleshooting:
• Call up a "System Calibration" for the relevant module.
• If only one electrode is affected, replace that electrode.
• If more than one electrode/sensor is affected, replace S2 or S3 fluid pack depending on
the module.
• Check Baro settings.
• Insert electrode simulator and act accordingly.

Flags no. 2029, 2030

2029: "ADC >>>>>"


2030: "ADC <<<<<"
Sensor signal is above or below the measuring range.
Possible causes:
- Reference system not filled
- Leaky/empty electrode
Troubleshooting:
• In the fill routines call up "Fill reference electrode" and then run "Calibration for
ready".
• Correct electrode seating.
• Replace relevant electrode/cartridge.
• Insert electrode simulator and act accordingly.

8-40 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Flag no. 2031

"Conditioning not ok"


The Na electrode was not conditioned.
Possible causes:
- Docking mechanism of S2 fluid pack leaky
- T&D leaky when aspirating conditioner
• Call up „Aspirate Na cond solution“. Then call up "Conditioning".
• Replace S2 fluid pack.
• Check S2 docking mechanism for contamination and if necessary clean it.
• Check tubing system with the general fluidics test or manually for damage or blockage
according the corresponding diagrams in Chapter 5 and replace the relevant compo-
nent.

Flags no. 2035, 2036

2035: "Linearity not OK"


2036: "Interference sensitivity"
The sensitivity of the sensor is too low.
Possible causes:
- Sensor worn out
Troubleshooting:
• Replace MSS cassette.
• Replace S3 fluid pack.

Flag no. 2040

"Cuvette not OK"


Calibrated layer thickness outside the expected range.
Possible causes:
- Sontamination of the cuvette
Troubleshooting:
• Replace cuvette.
• Call up "Polychromator calibration".
• Call up "COOX calibration".

Flag no. 2041

"Ref. point not OK"


Reference point is outside the expected range.
Possible causes:
- Sample feed blocked

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-41


8 Troubleshooting

Troubleshooting:
• Call up "Wash".
• Call up "General fluidics test" and check that tubing paths are open and without leaks.
• Replace cuvette.
• Perform "Polychromator calibration".
• Perform "COOX calibration".

Flag no. 2042

"Wavelength not OK"


A specific wavelength is outside the expected range.
Troubleshooting:
• Call up "Polychromator calibration".
• Replace polychromator
(see section 3.12.4 "Replacing the Polychromator Bilirubin" on page 3-30).
• Replace lamp unit (see section 3.12.2 "Replacing the lamp unit bilirubin" on page 3-28).

Flags no. 2043, 2044

"Hemolysis not OK (1, 2)"


The hemolyzer is not ready for operation; the hemolyzer output is outside the expected
range.
Troubleshooting:
• Visually inspect that hemolyzer is dry.
• Call up hemolyzer test and depending on the result replace hemolyzer unit or HSIN
board.

Flag no. 2070

"Sensor signal unstable"


Sampling criterion for tHb measurement value calculation violated.
Possible causes:
- Irregular sample
- Sample feed leaks or is blocked
Troubleshooting:
• For the tHb/SO 2 module call up "Start internal cleaning".
• Call up "General fluidics test" and check tubing paths.
• Test vacuum pump and take action according to the result.
• Replace S1 bottle.

8-42 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Flag no. 2071

"Measuring path dirty"


The absolute limits of water calibration are outside the specified range.
Possible causes:
- Sensor dirty
Troubleshooting:
• For the tHb/SO 2 module call up "Start internal cleaning". Repeat as necessary.
• Replace S1 bottle.
• Replace tHb/SO 2 module.

Flag no. 3000

"D2 / D3 not performed"


MSS standby solution was not aspirated.
Possible causes:
- Aspirate standby solution not OK
- Aspirate reference solution not OK
Troubleshooting:
• Call up fill routines "Aspirate standby solution" and "Aspirate reference solution", then
run "Calibration for ready".

Flag no. 3001

"FMS volume error"


This flag corresponds to module stop no. 20016
(see section 8.3 "Module stops" on page 8-15).

Flag no. 3002

"Temperature error"
This flag corresponds to the temperature module stop for the relevant module
(see section 8.3 "Module stops" on page 8-17).

Flag no. 3003

"FMS error"
This flag corresponds to module stop no. 20017
(see section 8.3 "Module stops" on page 8-16).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-43


8 Troubleshooting

Flag no. 3004

"AQC module error"


This flag corresponds to AQC module stop
(see section 8.3 "Module stops" on page 8-23).

Flags no. 3013, 3014

"Ref. point not ok"


Problem with precalibration; problem with calibration with MSS standby solution.
Possible causes:
- Aspirate standby solution not OK
- Aspirate reference solution not OK
Troubleshooting:
• In the fill routines call up "Aspirate standby solution" and "Fill ref. electrode MSS".
Subsequently perform "Calibration for Ready".
• In the wetting routines call up "Wetting MSS".
• In the contact path test call up "MSS standby solution" and take action according to the
result.
• Replace MSS cassette.

Flag no. 3015

„1P not OK“


Problem at calibration point 1;
Possible causes:
- Second point is not calibrated since an error occurred at the first point.
Troubleshooting:
• Perform "Calibration for Ready".

Flag no. 3022

"Conductivity not OK"


Conductivity calibration was not ended properly
Troubleshooting:
• Print sensor report, check the error messages for conductivity for CAL B (B Pot) and
CAL A (A Pot).
• Call up "General fluidics test" and check tubing paths.
• Replace S2 fluid pack.

8-44 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Flag no. 3033

"0P not OK"


O 2 zero calibration was not ended properly.
Possible causes:
- Aspiration of O 2 zero point solution not OK
- Soiling or leaky point
Troubleshooting:
• Call up fluid action "Aspirate O2 zero point solution".
• check fill port and T&D disk for soiling (see Instructions for Use Chapter 6 "Mainte-
nance").
• Call up "General fluidics test" and check tubing paths.
• Replace S2 fluid pack.

Flag no. 4003

"Calibration pending (1)"


New sensor was inserted, no calibration values present.
Troubleshooting:
• Perform "Calibration for Ready".

Flag no. 4008

"Calibration pending (2)"


Calibration is missing or was canceled.
Troubleshooting:
• Perform "Calibration for Ready".

Flag no. 4024

"Calibration pending (3)"


Sensor is or was deactivated.
Troubleshooting:
• Perform "Calibration for Ready".

Flag no. 5006

"Calculation error (1)"


µC calculation error.
Possible causes:
- An illegal calculation operation has occurred

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-45


8 Troubleshooting

Troubleshooting:
• Remove and replace containers/packs.
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Check Baro values.
• Reinstall software.

Flag no. 5007

"Calculation error (2)"


µC calculation error.
Possible causes:
- An illegal calculation operation has occurred
Troubleshooting:
• Remove and replace relevant electrode.
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Reinstall software.

Flag no. 5010...5012

"Calculation error (3...5)"


µC calculation error with COOX.
Possible causes:
- An illegal calculation operation has occurred
Troubleshooting:
• Check cuvette for soiling.
• Call up "Polychromator calibration".
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Reinstall software.

Flag no. 6026

"ADC error"
Problem with ADC on amplifier board
Troubleshooting:
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Replace relevant module.
• Check cable press fit of the module connector cable on the mainboard for damage.
• If the conductivity is not OK, replace mainboard
(see 3.18.1 "Replacing the Mainboard unit" on page 3-59).

8-46 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Flag no. 6030

"COOX HW error"
This flag corresponds to module stop no. 20071
(see section 8.3 "Module stops" on page 8-19).

Flags no. 6031, 6032

6031: "Neon lamp not OK"


6032: "Halogen lamp not OK"
The light output of the neon lamp/halogen lamp is too low.
Possible causes:
- Cuvette contaminated
- Light guide damaged
- Polychromator electronics defective
Troubleshooting:
• Call up "Internal cleaning" and then call up "Polychromator calibration".
• Replace cuvette.
• Check/replace cuvette holder
(see section 3.12.6 "Replacing the cuvette holder" on page 3-33).
• Check primary light guide for damage and if necessary replace it
(see section 3.12.5 "Replacing the primary light guide" on page 3-32).
• Replace lamp unit
(see section 3.12.2 "Replacing the lamp unit bilirubin" on page 3-28).
• Replace polychromator unit
(see section 3.12.4 "Replacing the Polychromator Bilirubin" on page 3-30).

Flag no. 7034

"Solution exhausted"
The pH value of the O 2 zero solution is < 5.
Possible causes:
- Solution used up
- Contamination of pack
Troubleshooting:
• Replace S2 fluid pack.

Flag no. 8036

"IfS not OK"


Ratio of BSA/Glu or BSA/Lac for CAL 3 out of range.
Possible causes:
- Interference correction not possible

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-47


8 Troubleshooting

Troubleshooting:
• Perform "Calibration for Ready".
• Replace MSS cassette.
• Replace S3 fluid pack.
• Check tubing for solutions CAL 1 and CAL 3 (see Chapter 5).

Flag no. 8061...8063

"Interferences (1...3)"
Invalid calculated values.
Possible causes:
- Interferences
- Unsuitable Sample type used
Troubleshooting:
• Check sample material.
• Perform "Polychromator calibration".
• Replace cuvette.
• Call up hemolyzer test and depending on the result replace hemolyzer unit or board.

Flag no. 8070

"IfS repro not OK"


Non-reproducible sensor signal at interference sensor.
Possible causes:
- Unsuitable sample type used
Troubleshooting:
• Call up "Automatic wetting routine" for MSS.
• Call up "MSS polarization".
• Perform "Calibration for Ready".
• Replace MSS cassette.

Flags no. 8071, 8072

"Calculation error (1, 2)"


UC calculation error on interference sensor.
Possible causes:
- An illegal calculation operation has occurred
Troubleshooting:
• Remove and replace MSS cassette.
• Remove and replace S3 fluid pack.
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Reinstall software.

8-48 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Flag no. 8073

"IfS sensor signal unstable"


Interference sensor signal cannot be reproduced.
Possible causes:
- Unsuitable sample type used
- Measuring path contaminated
- Interference signals in measuring path during measurement value capture
- Interference signals over reference electrodes during measurement value capture
Troubleshooting:
• If this interference applies to the complete MSS module, the reference electrode must
be changed.
• Perform "Calibration for Ready".
• Call up "Stability monitor", start and check "Details" of the "Electrode slope". If the
MSS cassette is faulty, this will be displayed and the MSS cassette must be replaced.
• Insert electrode simulator and act accordingly.

Flag no. 8074

"IfS ref. point not ok"


Problem at reference point of standby sol. at IfS.
Possible causes:
- Aspirate standby solution not OK
- Aspirate reference solution not OK
Troubleshooting:
• In the fill routines call up "Aspirate standby solution" and "Fill MSS reference elec-
trode". Subsequently perform "Calibration for Ready".
• In the wetting routines call up "Wetting MSS".
• In the contact path test call up "MSS standby solution" and take action according to the
result.
• Replace MSS cassette.

Flag no. 8075

"IfS ADC error"


Problem with ADC on the amplifier board of a Ifs.
Troubleshooting:
• Replace MSS module.
• If the conductivity is not OK, replace mainboard
(see 3.18.1 "Replacing the Mainboard unit" on page 3-59).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-49


8 Troubleshooting

Flag no. 8076

"IfS mean not OK"


Final calibration value of a Interference Sensor (IfS) is out of range.
Possible causes:
- Transponder error on S3 fluid pack
- Defective electrode
- Incorrect Baro value
Troubleshooting:
• Call up "System Calibration" for MSS.
• Replace MSS cassette.
• Replace S3 fluid pack.
• Check Baro settings.
• Insert electrode simulator and act accordingly.

Flags no. 8077, 8078

"IfS ADC >>>>>"


"IfS ADC <<<<<"
Interference sensor signal is above/below ADC range
Possible causes:
- Reference system not filled
- Leaky/empty electrode
Troubleshooting:
• In the fill routines call up "Fill reference electrode" and then run "Calibration for
ready".
• Check electrode seating.
• Replace MSS cassette.
• Insert electrode simulator and act accordingly.

Flag no. 8080...8082

"Interferences (4...6)"
Possible causes:
- Unsuitable sample type
- For Bilirubin: tHb value is less than 3 g/dl
Troubleshooting:
• Check sample material: sample too old, partially hemolyzed, unhomogenous, incorrect
anticoagulant?

8-50 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Flag no. 8083

"Sample type conflict"


Possible causes:
- Unsuitable sample type used
Troubleshooting:
• Check sample material.

Flag no. 8084

"Invalid sample type"


All calculated values that are based on measurement values of arterial blood cannot be dis-
played.
Possible causes:
- Instead of arterial blood a different blood type was used
Troubleshooting:
• Use arterial blood as sample material and select the proper sample type.

Flag no. 9000

"Calculation error (3)"


PC calculation error.
Possible causes:
- An illegal calculation operation has occurred
Troubleshooting:
• Remove and replace packs.
• Turn the instrument off, wait at least 3 minutes, and turn it back on.
• Reinstall software.

Flag no. 9001

„Not activated“
Parameter is not activated for measurement (gray).

Flag no. 9002

„Not calibrated“
Parameter is not calibrated (red cross).
Troubleshooting:
• Perform "Calibration for ready".

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-51


8 Troubleshooting

Flag no. 9003

„Measurement running“
Will be displayed instead of a measurement result as long as the measurement was not
finished.

Flag no. 9004

„Out of range (-)“


Measurement value is outside (below) the normal range.

Flag no. 9005

„Out of range (+)“


Measurement value is outside (above) the normal range.

Flag no. 9006

„tHb not OK“


Indicates at the parameter SO 2 , that the tHb value is outside the normal range.

Flag no. 9008

„Missing data“
The calculation of this parameter is not possible because of missing base values.

Flag no. 9009

„Measurement aborted“
The measuremant was aborted by the user.

Flag no. 9010

„System stop occured“


A system stop occured during the measurement.

8-52 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

8.6 Important test routines

8.6.1 Contact path test

BG / I S E

Aspirate CAL B:
Positioning: MCI-MCO contact paths
Status NOT OK:
This error pattern corresponds to value flag no. 1001.
VI closed: MCO measurement value
Status NOT OK:
This error pattern corresponds to value flag no. 1008.
Cond. ratio: MCC-MCO contact path
Status NOT OK:
This error pattern corresponds to value flag no. 1006.
Ref. aspiration: MCO measurement value
Status NOT OK:
This error pattern corresponds to value flag no. 1003.
VO closed MCO measurement value
Status NOT OK:
This error pattern corresponds to value flag no. 1004.
External sample:
Separation: MCC measurement value
Status NOT OK:
This error pattern corresponds to value flag no. 1002.
Positioning: MCO measurement value
Status NOT OK:
This error pattern corresponds to value flag no. 1000.
Cond. ratio: MCC-MCO contact path
Status NOT OK:
This error pattern corresponds to value flag no. 1006.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-53


8 Troubleshooting

M S S (Standby solution) (OM N I S 5 and 6 only)

Positioning MCO: MCO measurement value


Status NOT OK:
- Call up "Aspirate MSS standby solution"
in the aspiration routines.
- Measurement value near zero: check seating of MSS cassette.
- Measurement value slightly lower: replace MSS cassette.
- Insert electrode simulator and act accordingly.
Ref. aspiration: MCO measurement value
Status NOT OK:
- Call up "Aspirate MSS reference solution"
in the aspiration routines.
- Replace S3 fluid pack.
- Call up "General fluidics test" and check tubing paths.
Amp. Lac/Glu are measurement values that have no meaning yet.
Check MCI: MCI measurement value
Status NOT OK:
- Call up this test several times. This value can improve greatly
as a result.
- Call up MSS wetting routine.
- Call up MSS internal cleaning and then insert
new MSS cassette.

M S S pos. sample (OM N I S 5 and 6 only)

This test function is only significant when using the electrode simulator.

M S S external sample (OM N I S 5 and 6 only)

Separation MCI: MCI measurement value


Status NOT OK:
This error pattern corresponds to value flag no. 1002.

8-54 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Positioning MCI: MCI measurement value


Status NOT OK:
This error pattern corresponds to value flag no. 1000.
Separation MCO: MCI measurement value
Status NOT OK:
This error pattern corresponds to value flag no. 1002.
Positioning MCO: MCI measurement value
Status NOT OK:
This error pattern corresponds to value flag no. 1000.
Amp. Lac/Glu are measurement values that have no meaning yet.
Check MCI: MCI measurement value
Status NOT OK:
- Sample is unhomogenous.

8.6.2 Aggregates

T&D module

Initialization: The T&D disc performs an automatic initialization.


Adjustment of disk: This function enables manual adjustment of an extremely
twisted T&D disc: loosen retaining arm behind the T&D disc
with an Allen wrench, T&D disc in pos.1 and press "Adjustment
of disk. Then tighten retaining arm and call up "Initialization".
T&D needle calibration: This function is used for adjusting the position of the T&D disk
to the SIP tip. It must be performed after replacing the SIP or the
tubing set.
Place syringe: This function is used to check the needle position.
It should be centered in the retainer.
If the needle projects too far, it is more difficult to attach a
sample container. If the needle is too far back, there is
a danger that the sample container will fall off.
If the needle is not centered in the container, a "T&D needle

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-55


8 Troubleshooting

calibration" must be performed. If this is not successful,


perform the following actions:
- Replace needle.
- Replace T&D disk.
- Replace inlet needle.
- Replace T&D module.
Aspirate from syringe: This function is used for visual inspection of the needle
(bent, broken).

Peristaltic pumps

This test function tests the function of the three peristaltic pumps (the instrument variants
OMNI S 1-4 only have one peristaltic pump). The following features must be noted here:
Noise check: If a pump is too loud, this indicates that the pump head
may be eccentric - The pump head of the relevant pump
must be replaced.
In addition, an excessively noisy pump can indicate that a
pump tube is too tightly seated - The pump tube of the
relvant pump must be replaced.
Direction of rotation: After replace a peristaltic pump the
correct direction of rotation must be checked with the test -
If the direction of rotation is reversed, the pump is incorrectly
connected.
Tubing paths: This test function can often be useful as a support when testing
the tubing, because it allows precise control of the pumps.

Hemolyzer (OM N I S 2, 4, 6 only)

U PC : Phase-angle adjustment voltage


Status NOT OK:
- The hemolyzer is not empty: call up "Wash".
- The hemolyzer is damaged: replace unit.

8-56 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

I N1 : Current setting
Status NOT OK:
- The hemolyzer is damaged: replace unit.
I N2 : Measure setting
Status NOT OK:
- The HSIN board is defective: replace board.
IH: Hemolyzer current
Status NOT OK:
- The hemolyzer is damaged: replace unit.
- The HSIN board is defective: replace board.
U AI Voltage at air and idle
Status NOT OK:
- The hemolyzer is not empty: call up "Wash".
- The hemolyzer is damaged: replace unit.
R AP : Resistance with air and power
Status NOT OK:
- The hemolyzer is not empty: call up "Wash".
- The hemolyzer is damaged: replace unit.
I SP : Current with sample and power
Status NOT OK:
- The hemolyzer is damaged: replace unit.
- The HSIN board is defective: replace board.
U SP Voltage with sample and power
Status NOT OK:
- The hemolyzer is damaged: replace unit.
- The HSIN board is defective: replace board.

8.6.3 Control sensors

Vacuum System

There are four effective test options for the vacuum system that can be used for error anal-
ysis:
• Set 200 mBar as target value and switch on pump: the underpressure should be between
170 and 230 mBar after 2 to 3 seconds. If this cannot be reached, valve V13 or the con-
necting tube must be checked and if necessary replaced.
• Set 400 mBar as target value and switch on pump: the underpressure should be between
360 and 440 mBar after 10 seconds. If this cannot be reached, valve V13 or the connect-
ing tube must be checked and if necessary replaced.
• Set MAX as target value and switch on pump: the underpressure should reach at least
75% of the local Barometric Pressure as sensed on Control sensors/Baro sensor. If this
is not reached, check the pump head and the VPS sensor and if necessary replace.
• Set MAX as target value, switch on pump and then off after 10 seconds: the underpres-
sure should not be less than 15 mBar after one minute. If this is not reached, there is a
leak in the vacuum system: check moisture trap, VPS, V13, V18, V6, V23, main pump,
waste separator and if necessary replace.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-57


8 Troubleshooting

Baro sensor

The value of the Baro sensor must match the actually measured Baro value (with a calibrated
barometer). If the deviation is greater than 4 mBar, the Baro value must be input again into
the protected setup.

IMPORTANT:
From approx. 2000 m above sea level or air pressure < 570 mmHg (75.98 kPa), the
specifications for parameter PO 2 are no longer fulfilled and the parameter must no
longer be used for evaluation of the clinical decisions.
After successful installation, the parameter must be permanently deactivated!

8.6.4 Meas. sensors

BG

Use this function to monitor the BG module.

The current sensor values will be displayed in the column "Signal". Standard deviation in
column "Std. dev." must not exceed the follwing values for optimal aspiration of Mix 1:
Sensor Signal from Signal to Std. dev.
[mV] [mV] [mV]
P CO 2 -1300 2000 1.5
PO2 -- -- 1.5
pH -2100 1600 3

Aspirate Mix 1: Mix 1 will be positioned into the measurement chamber.


External sample: An external sample will be positioned into the measurement chamber.

The boundaries provided for standard deviation are valid for Mix 1 only!

8-58 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

ISE

Use this function to monitor the ISE module.

The current sensor values will be displayed in the column "Signal". Standard deviation in
column "Std. dev." must not exceed the follwing values for optimal aspiration of Mix 1:
Sensor Signal from Signal to Std. dev.
[mV] [mV] [mV]
Ca 2+ -2000 2200 1.2
K+ -1700 2100 1.2
Na + -1700 2300 1.2
Cl - -2300 1700 1.2

Aspirate Mix 1: Mix 1 will be positioned into the measurement chamber.


External sample: An external sample will be positioned into the measurement chamber.

The boundaries provided for standard deviation are valid for Mix 1 only!

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-59


8 Troubleshooting

MSS

Use this function to monitor the MSS module.

The current sensor values will be displayed in the column "Signal". Standard deviation in
column "Std. dev." must not exceed the follwing values for optimal aspiration of MSS stand-
by solution:
Sensor Signal from Signal to Std. dev.
BSA -10 nA 100 nA 0.8 nA
Lac -10 nA 100 nA 0.4 nA
Glu -10 nA 100 nA 0.4 nA
NH 4 -100 mV 100 mV 0.2 mV
K -100 mV 100 mV 0.2 mV
Urea -100 mV 100 mV 0.2 mV

Aspirate MSS MSS standby solution will be positioned into the measurement
standby solution: chamber.
External sample: An external sample will be positioned into the measurement chamber.

The boundaries provided for standard deviation are valid for MSS standby solution
only!

8-60 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

tHb/ SO 2 (OM N I S 1, 3, 5 only)

Use this function to monitor the tHb/SO 2 module.


Position: Wash water is positioned in the module.
Scans: This button starts the monitoring process which determines the values
for transmissionlight and light scatter (each for wavelengths 1 to 4) and
the characteristic value for the scan.
Dry: The tHb/SO 2 module is dried.
The test result (ok- green checkmark / not ok- red X) is displayed under Status.

If a large negative error value is displayed for one of the diodes (1 to 4) for transmission
light (T) as well as scattered light (S), it indicates that a laser diode is defective (this type of
error corresponds to value flag no. 2071).

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-61


8 Troubleshooting

COOX (OM N I S 2, 4, 6 only)

Use this function to monitor the oximeter module.


Lamp test Activate Halogen lamp (green) - When the test is complete, the values
and the status (ok - green checkmark / not ok - red x) are displayed in
the table.

Status NOT OK:


If an error value is displayed on the halogen lamp x-peak, it indicates a
grid shift (photodiode array movement). This error pattern
corresponds to value flag no. 6032 (see page 8-47).
If necessary replace the polychromator unit.

Activate Neon lamp (green) - When the test is complete, the values and
the status (ok - green checkmark / not ok - red x) are displayed in the
table.

Status NOT OK:


The light output of the neon lamp is too low.
If an error value is displayed on the neon lamp x-peak, it indicates a
grid shift (photodiode array movement). This error pattern
corresponds to value flag no.6031 (see page 8-47). If necessary replace
the polychromator unit.

8-62 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

8.6.5 General fluidics test


Note: The general fluidics test described below applies to the device variants OMNI S 5
and 6! For device variants OMNI S 1 to 4, the number of areas in which fluidics
tests can be performed is reduced accordingly.

The general fluidics test divides the OMNI S into 11 ranges and perform fluid tests in these
ranges. The test results are displayed on the screen. The possible test results are as follows:
• OK - The test for that section could not find any errors.
• Leak or blocked - The range that has a leak or is blocked is shown exactly; the error can
be found in this range.
• Unknown - An error was detected that is not directly in this range but only affects this
range.
• Error - The range could not be tested because of errors in other ranges.
If errors are found, the relevant range must be tested for leakage or blockage from the fol-
lowing diagrams. The 11 ranges of the general fluidics test are classified as followed:

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-63


8 Troubleshooting

Separator main pump range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer
17 12 15

20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na
+
K
+
Ca
2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

8-64 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Sample distributor range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-65


8 Troubleshooting

FM S range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

8-66 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

M S S measuring chamber range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-67


8 Troubleshooting

Sample input range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

8-68 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

S S2 - S S3 range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-69


8 Troubleshooting

Converging line range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

8-70 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

BG measuring chamber range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-71


8 Troubleshooting

I S E measuring chamber range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

8-72 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Hb module range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-73


8 Troubleshooting

Bypass range

Main pump MSS output pump MSS input pump

Vacuum
pump
30°C
10
Barometer 15
17 12
20

16 8
MCO CCL CCR MCI

13
VSI
14
MSS SS4
30°C
Ref. Dummy Urea / Lac / Glu Thermoblock

Moisture
trap Bacteria 9
filter 37°C
MCO MCC VII
Vacuum MC VIO
pump shield
cartridge ISE MCI
T&D
18 Ref. SCon Cl – Na+ K+ Ca 2+
37°C
Exhaust 5 21

MCO MCC VBI 6 7


3 VBO 5
8

MCI
BG Turn &
9

37°C
Ref. SCon pH O2 CO2 SS1 4

Waste Dock 10

Hb 3

separator cartridge 2
11
T&D tubing set
22
Hemoly- 24 2
3
4
5
6
7 8 9 11
10

4
COOX zer 23
SS6
2
SS3 SS2 AQC
AQC

6 5 1 7 8 2 3 6 2 1 8 7 3

M
19

FMS
W
Waste S1 S2 S3
container Rinse solution

8-74 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

8.6.6 Stability monitor


The stability monitor can be used to check the device stability of the OMNI S over a defined
period. The following limit values are used for the assessment:
Aspiration status
Medium Limit value
Mix1 < 5%
Mix2 < 20 %
O2 zero < 30 %
Standby solution < 5%
Solution C1 < 5%
Solution C2 < 10 %
Solution C3, C4 < 30 %
Na Cond < 30 %
Clean < 30 %

Electrode status
Sensor Limit value
all < 10 %

Electrode voltage scanning


Sensor Limit value
all BG 2 mV
all ISE 2 mV
Urea, K, NH4 0.4 mV
BSA 1.6 nA
Glu, Lac 0.8 nA

Standard deviation
Sensor Limit value
pH 3 mV
pO2 1.5 mV
pCO2 1.5 mV
Cl 1.2 mV
Na 1.2 mV
K 1.2 mV
Ca 1.2 mV
Urea 0.2 mV
Glu 0.4 nA
Lac 0.4 nA
BSA 0.8 nA

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-75


8 Troubleshooting

Standard deviation spec.


Module Limits
ISE Ca*0.55 < K < Ca*0.95
ISE Ca*0.30 < Na < Ca*0.70
ISE Ca*0.05 < Cl < Ca*0.45

Electrode slope
Sensor Limit value
pH > -300 mV
pCO2 > -130 mV
Na < 125 mV
Cl < -100 mV
Ca < -75 mV
K < 125 mV
Glu C2/C1 < 3.8
Lac C2/C1 < 3.8
Urea C2/C1 < 1.6

Conductivity drift
Module Limits
BG 99 % < CondDrift < 101 %
ISE 99 % < CondDrift < 101 %

FMS stability
Abs (default - cond) > 4 %

Air pressure
570 mmHg < BaroVal < 800 mmHg

IMPORTANT:
From approx. 2000 m above sea level or air pressure < 570 mmHg (75.98 kPa), the
specifications for parameter PO 2 are no longer fulfilled and the parameter must no
longer be used for evaluation of the clinical decisions.
After successful installation, the parameter must be permanently deactivated!

8-76 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Temperature drift
Module Sensor Limits Designation
BG / ISE Temp_Cov 37° +/-0.15 Cover temperature
BG / ISE Temp_MC 37° +/-0.15 Temperature of SCon
BG / ISE Temp_MCI 37° +/-0.15 Peltier temp. meas. chamber trough In
BG / ISE Temp_MCO 37° +/-0.15 Peltier temp. meas. chamber trough Out
MSS Temp_MC 30° +/-4 Temperature of rocker
MSS Temp_MCI 30° +/-0.15 Peltier temp. meas. chamber trough In
MSS Temp_MCO 30° +/-0.15 Peltier temp. meas. chamber trough Out
SD BG_MSS_IN 37° +/-0.15 Temperature at BG In
SD ISE_D_IN 37° +/-0.15 Temperature at ISE In
SD BG_MSS_IN 30° +/-0.15 Temperature at MSS In
SD ISE_D_IN 30° +/-0.15 Temperature at standby solution In

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-77


8 Troubleshooting

8.6.7 General hardware test


This test is used to check the hardware components of the OMNI S. This test allows any
faults in components to be detected and limited. The following components are checked:

Power supplies

24V power supply


Error message: 24V power supply nOK
Remedy: - Check 24V cable (+ plug) from power supply to mainboard.
- Check actuator bus cable and plug.
12V power supply
Error message: 12V power supply nOK
Remedy: - Check 12V cable (+ plug) from power supply to mainboard.
2.5V (MB)
Error message: 2.5V power supply nOK
Remedy: - Check mainboard and if necessary replace.

Heatings

Heating BG
Error message: Peltier BG input nOK
Peltier BG output nOK
Remedy: - Check cable plug from mainboard to BG module.
- Replace BG module.
Error message: Peltier fan BG input nOK
Peltier fan BG output nOK
Remedy: - Check cable plug from mainboard to BG module.
- Check BG fan for mechanical blockage.
- Replace BG module.
Heating ISE
Error message: Peltier ISE input nOK
Peltier ISE output nOK
Remedy: - Check cable plug from mainboard to ISE module.
- Replace ISE module.
Error message: Peltier fan ISE input nOK
Peltier fan ISE output nOK
Remedy: - Check cable plug from mainboard to ISE module.
- Check ISE fan for mechanical blockage.
- Replace ISE module.

8-78 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

Heating MSS
Error message: Peltier MSS input nOK
Peltier MSS output nOK
Remedy: - Check cable plug from mainboard to MSS module.
- Replace MSS module.
Error message: Peltier fan MSS input nOK
Peltier fan MSS output nOK
Remedy: - Check cable plug from mainboard to MSS module.
- Check MSS fan for mechanical blockage.
- Replace MSS module.
Heating SD
Error message: Peltier SD 37°C-BG nOK
Peltier SD 37°C-ISE nOK
Peltier SD 37°C-MSS1 nOK
Peltier SD 37°C-MSS2 nOK
Remedy: - Check cable plug from mainboard to sample distributor.
- Replace sample distributor.
Error message: Peltier fan SD 37°C at BG input nOK
Peltier fan SD 37°C at ISE input nOK
Peltier fan SD 37°C at MSS input nOK
Peltier fan SD 37°C above MSS input nOK
Remedy: - Check cable plug from mainboard to sample distributor.
- Check SD fan for mechanical blockage.
- Replace sample distributor.

Valves on valve bus

Valve + No.
Error message: Valve + No. - nOK
Remedy: - Check plug and cable of relevant valve.
- Replace valve.
- Replace valve board.

Turn & Dock

Stepper motor turn


Error message: T&D stepper motor turn nOK
Remedy: - Check plug and cable of stepper motor.
- Replace stepper motor.
Stepper motor fillport
Error message: T&D stepper motor fillport nOK
Remedy: - Check plug and cable of stepper motor.
- Replace stepper motor.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-79


8 Troubleshooting

Stepper motor dock


Error message: T&D stepper motor dock nOK
Remedy: - Check plug and cable of stepper motor.
- Replace stepper motor.

FM S

V19
Error message: Valve 19 nOK
Remedy: - Check plug and cable of valve.
- Replace valve.
- Replace FMS board.
VMIX
Error message: Valve MIX nOK
Remedy: - Check plug and cable of valve.
- Replace valve.
- Replace FMS board.

PP Main

Stepper motor main pump


Error message: Main pump nOK
Remedy: - Check plug and cable of main pump.
- Replace main pump.
- Replace mainboard.

PP M S S

Stepper motor PP_MSS1 (In)


Error message: MSS pump 1 nOK
Remedy: - Check plug and cable of pump.
- Replace pump.
- Replace mainboard.
Stepper motor PP_MSS2 (Out)
Error message: MSS pump 2 nOK
Remedy: - Check plug and cable of pump.
- Replace pump.
- Replace mainboard.

Vacuum pump

V13
Error message: Valve 13 nOK
Remedy: - Check plug and cable of valve.
- Replace valve.
- Replace vacuum control board.

8-80 Service Manual, Roche OMNI S, Version 4.0, December 2004


8 Troubleshooting

V18
Error message: Valve 18 nOK
Remedy: - Check plug and cable of valve.
- Replace valve.
- Replace vacuum control board.
Vacuum pump
Error message: Power supply of the vacuum pump nOK
Remedy: - Check plug and cable of vacuum pump.
- Replace pump unit.
- Replace vacuum control board.

Main fan

Fan 1
Error message: Main fan 1 nOK
Remedy: - Check plug and cable of fan unit.
- Check fan unit for mechanical blockage.
- Replace fan unit.
- Replace mainboard.
Fan 2
Error message: Main fan 2 nOK
Remedy: - Check plug and cable of fan unit.
- Check fan unit for mechanical blockage.
- Replace fan unit.
- Replace mainboard.

Hemolyzer

Error byte Hemolyzer


Error message: Function of hemolyzer nOK
Remedy: - Unscrew head of hemolyzer, check hemolyzer for
moisture and if necessary dry; then repeat test.
Remedy: - Check plug and cable of hemolyzer.
- Replace hemolyzer.
- Replace HSIN board.

Service Manual, Roche OMNI S, Version 4.0, December 2004 8-81


8 Troubleshooting

8.7 Sensor limits (Sensor report)


Conductivity BG: min typ max
A Pot.: 55mV 75mV 105mV
B Pot.: 1320mV 1500mV 1710mV
PCO 2 : min typ max
1P Pot.: -1300mV -1000mV 2000mV
Slp Pot.: -160mV -143mV -110mV
PO 2 : min typ max
1P Pot.: 250mV 330mV 550mV
Slp Pot.: -10mV 2mV 15mV
pH: min typ max
1P Pot.: -2100mV 500mV 1600mV
Slp Pot.: -340mV -305mV -250mV
Conductivity ISE: min typ max
A Pot.: 14mV 50mV 70mV
B Pot.: 930mV 1030mV 1250mV
Ca 2+ : min typ max
1P Pot.: -2000mV 500mV 2200mV
Slp Pot.: -95mV -87mV -65mV
K+: min typ max
1P Pot.: -1700mV 550mV 2100mV
Slp Pot.: 110mV 130mV 140mV
Na + : min typ max
1P Pot.: -1700mV 950mV 2300mV
Slp Pot.: 110mV 130mV 140mV
Cl - : min typ max
1P Pot.: -2300mV 700mV 1700mV
Slp Pot.: -120mV -108mV -90mV
Glu: min typ max
1P Sens.: 10nA 50nA 270nA
Lin.: 3 4,5 5,25
3P Sens.: 0nA 10nA 100nA
Lac: min typ max
1P Sens.: 10nA 50nA 315nA
Lin.: 3 4,5 5,25
3P Sens.: 0nA 10nA 100nA
BSA: min typ max
1P Sens.: -10nA 0nA 10nA
3P Sens.: 0nA 10nA 100nA
COOX: min typ max
d: 60µm 100µm 130µm

8-82 Service Manual, Roche OMNI S, Version 4.0, December 2004


9 Index

9 Index Cal. intervals U.S. ............................................................. 6-1


Central measuring unit .................................................. 3-2
Contact path test ............................................................ 8-53
A Control sensors ............................................................... 8-57
Cooling fan
AutoQC
Replacing the cooling fan unit .......................... 3-76
Initial installation of the AutoQC module .... 4-1
COOX module
Maintenance kit ....................................................... 7-7
Overview .................................................................... 3-27
Replacing the AutoQC module ......................... 4-2
Replacing the cuvette ........................................... 3-34
Replacing the AutoQC steel tube .................... 4-14
Replacing the cuvette holder ............................. 3-33
Replacing the click-in closure ............................ 4-3
Replacing the cuvette seals ................................. 3-34
Replacing the Control board .............................. 4-5
Replacing the lamp unit ...................................... 3-28
Replacing the flex cable (long) ........................... 4-9
Replacing the Polychromator Bilirubin ........ 3-30
Replacing the flex cable (short) ......................... 4-8
Replacing the PolyOX-KX control board .... 3-29
Replacing the magnetic valve ............................. 4-4
Replacing the primary light guide ................... 3-32
Replacing the sample tube ................................... 4-9
Replacing the tubing set tHb/PolyOx ............ 3-34
Replacing the temperature sensor .................. 4-15
Cuvette .........................................................................7-4, 7-6
Replacing the toothed belt (long) ................... 4-18
Cuvette seals ...............................................................7-4, 7-6
Replacing the toothed belt (short) ................. 4-17
Replacing the wash port ..................................... 4-16 D
Replacing the X motor ........................................ 4-12
Replacing the Y motor ......................................... 4-12 Decontamination ............................................................. 1-3
Replacing the YZ distributor board ................. 4-6 Demo mode ........................................................................ 6-1
Replacing the Z distributor board .................... 4-7 Disinfectant ........................................................................ 1-2
Replacing the Z motor ........................................ 4-13 E
AutoQC/COOX cable .................................................. 3-63
ESD protection .................................................................. 1-5
B
F
Baro sensor ....................................................................... 8-58
BG measuring chamber Fill port ........................................................ 7-1, 7-2, 7-4, 7-5
Replacing the BG measuring chamber ......... 3-18 Floppy disk drive
Replacing the center seal .................................... 3-19 Replacing the floppy disk drive ........................ 3-69
Replacing the electrode locking lever ............ 3-19 Fluid actions ....................................................................... 5-1
Replacing the fan protector ............................... 3-19 Aspirate sample ........................................................ 5-8
Bili .......................................................................................... 6-1 Calibration ................................................................ 5-20
Bottle compartment COOX calibration ................................................. 5-32
Replacing the bacteria filter ............................... 3-45 Na conditioning ..................................................... 5-25
Replacing the bottle compartment cover .... 3-42 O2 zero point calibration ................................... 5-24
Replacing the docking mechanism S1 .......... 3-44 Overview ...................................................................... 5-1
Replacing the docking mechanisms S2 and S3 3- Sample distribution .............................................. 5-10
44 Sample input .............................................................. 5-6
Replacing the gas dampers ................................ 3-42 FMS
Replacing the microswitch module ................ 3-45 Replacing the FMS ................................................ 3-11
Replacing the waste docking mechanism .... 3-43 Replacing the FMS control board ................... 3-13
Bypass holder ............................................ 7-1, 7-3, 7-5, 7-6 Replacing the FMS tubing set ........................... 3-13
Bypass nipple ..................................................................... 3-5 Replacing the FMS valve (MIX) ....................... 3-59
FMS tubing set ......................................... 7-1, 7-2, 7-4, 7-6
C
G
Cable connectors
Replacing the main controller cable .............. 3-78 General fluidics test ....................................................... 8-63
Replacing the valve bus cable ........................... 3-77 General hardware test ................................................... 8-78

Service Manual, Roche OMNI S, Version 4.0, December 2004 9-1


9 Index

H O
Hard disk OMNI S 1 ............................................................................ 5-3
Replacing the hard disk .......................................3-68 OMNI S 2 ............................................................................ 5-3
Hb cartridge OMNI S 3 ............................................................................ 5-2
Replacing the Hb cartridge ................................3-39 OMNI S 4 ............................................................................ 5-2
Hemolyzer OMNI S 5 ............................................................................ 5-1
Replacing the hemolyzer .....................................3-34 OMNI S 6 ............................................................................ 5-1
Replacing the Hemolyzer tube .........................3-36
Replacing the HSIN board .................................3-36 P
Test ..............................................................................8-56 PC unit ............................................................................... 3-64
Hemolyzer tube .........................................................7-4, 7-6 Peristaltic pumps
Hotline info ........................................................................6-2 Replacing a peristaltic pump ............................ 3-40
Replacing the pump head .................................. 3-41
I
Replacing the pump tube ................................... 3-42
Insert needle ...............................................................7-1, 7-2 Test .............................................................................. 8-56
Interface unit Pin assignment of Packs S2 and S3 ........................... 5-4
Replacing the Interface unit ...............................3-61 Printer
ISE measuring chamber Replacing the cutter lid ....................................... 3-76
Replacing the center seal .....................................3-20 Replacing the paper cover .................................. 3-76
Replacing the electrode locking lever .............3-20 Replacing the Printer ........................................... 3-75
Replacing the fan protector ...............................3-20 Replacing the winder unit .................................. 3-76
Replacing the ISE measuring chamber ..........3-20 Protected software functions ...................................... 6-1

M R
Mainboard unit Rear panel
Replacing the Mainboard unit ..........................3-59 Disassembly ............................................................... 3-1
Replacing the sample sensor board ................3-60 Revisions ............................................................................. 2-1
Maintenance .......................................................................7-1
OMNI S 1 ....................................................................7-1 S
OMNI S 2 ....................................................................7-4 Sample distributor
OMNI S 3 ....................................................................7-1 Replacing the sample distributor .................... 3-15
OMNI S 4 ....................................................................7-4 Replacing the sample distributor cartridge . 3-13
OMNI S 5 ....................................................................7-1 Replacing the SS1, SS3 and SS4 sensors ....... 3-17
OMNI S 6 ....................................................................7-4 Sample inlet path
Meas. sensors Replacing the sample inlet path ......................... 3-5
BG ................................................................................8-58 Screen
ISE ................................................................................8-59 Replacing the battery and charger .................. 3-70
MSS ..............................................................................8-60 Replacing the Screen ............................................ 3-64
Measuring chamber cartridge Replacing the screen cable ................................. 3-72
Replacing the MC Actor Control Board .......3-26 Replacing the TFT backlight ............................. 3-71
Replacing the measuring chamber cartridge 3-25 Screen/PC unit ................................................................ 3-64
Module stops ....................................................................8-15 Sensor limits .................................................................... 8-82
MSS measuring chamber Sensor report ................................................................... 8-82
Replacing the center seal .....................................3-23 Service mode ...................................................................... 6-2
Replacing the electrode locking lever .............3-23 Shutdown ............................................................................ 3-1
Replacing the fan protector ...............................3-23 Software update
Replacing the measuring chamber amplifier 3-23 video sequences ...................................................... 3-74
Replacing the MSS clip ........................................3-22 with floppy disks .................................................... 3-73
Replacing the MSS MC cover ...........................3-22 with FTP .................................................................... 3-73
Replacing the MSS measuring chamber .......3-21 Stability monitor ............................................................ 8-75

9-2 Service Manual, Roche OMNI S, Version 4.0, December 2004


9 Index

Symbols ................................................................................ 1-1 V


System stops ....................................................................... 8-1
System warnings ............................................................. 8-24 Vacuum system
Replacing the Actuator Board V4/V6 ............ 3-53
T Replacing the moisture trap .............................. 3-50
Replacing the pump head ................................... 3-49
T&D module Replacing the vacuum control board ............ 3-52
Replacing the actuator board ............................. 3-9 Replacing the vacuum pump ............................ 3-48
Replacing the fill port holder .............................. 3-4 Replacing the vacuum pump completely ..... 3-46
Replacing the linear actuators .......................... 3-10 Replacing the VP waste water stop ................. 3-51
Replacing the needle .............................................. 3-4 Replacing the waste valve V6 ............................ 3-57
Replacing the optical sensor ................................ 3-5 Replacing the waste valves V13/V18 .............. 3-56
Replacing the plug control ................................... 3-7 Test .............................................................................. 8-57
Replacing the stepper motor ............................... 3-9 VP tube set ................................................................ 3-51
Replacing the T&D disk ........................................ 3-6 waste tube ................................................................. 3-50
Replacing the T&D module ................................ 3-2 Value flags ......................................................................... 8-27
Replacing the T&D sensor cover ....................... 3-6 Valves .................................................................................. 3-54
Replacing the T&D tubing set ............................ 3-7 Replacing the valves from MC cartridge / Hb cart-
T&D positions .......................................................... 5-5 ridge ............................................................................. 3-54
Test .............................................................................. 8-55 Replacing the valves of the sample distributor 3-
T&D positions ................................................................... 5-5 55
tHb/SO2 module
Replacing the tHb/SO2 heating unit ............. 3-38 W
Replacing the tHb/SO2 module ...................... 3-38
Replacing the tubing set tHb/PolyOx ........... 3-38 Waste water separator ........................... 7-1, 7-3, 7-5, 7-6
Transponder Control Waste water tubing ..................................................7-3, 7-7
Replacing the Transponder Control .............. 3-46

Service Manual, Roche OMNI S, Version 4.0, December 2004 9-3


9 Index

9-4 Service Manual, Roche OMNI S, Version 4.0, December 2004

You might also like