You are on page 1of 7

Article

pubs.acs.org/ac

Photoacoustic Spectroscopy for the Quantification of N2O in the Off-


Gas of Wastewater Treatment Plants
Klemens M. Thaler,† Christoph Berger,† Carmen Leix,‡ Jörg Drewes,‡ Reinhard Niessner,†
and Christoph Haisch*,†

Chair of Analytical Chemistry, Institute of Hydrochemistry, Technical University of Munich, Marchioninistrasse 17, 81377 Munich,
Germany

Chair of Urban Water Systems Engineering, Technical University of Munich, Am Coulombwall 3, 85748 Garching, Germany
*
S Supporting Information

ABSTRACT: Different configurations of photoacoustic (PA)


setups for the online-measurement of gaseous N2O, employing
semiconductor lasers at 2.9 and 4.5 μm, were developed and tested.
Their performance was assessed with respect to the analysis of N2O
emissions from wastewater treatment plants. For this purpose, the
local N2O emissions of a wastewater treatment bioreactor was
sampled by a dedicated mobile sampling device, and the total N2O
emissions were analyzed in the gastight headspace of the bioreactor.
We found that the use of a quantum-cascade laser emitting at about
4.53 μm, operated in a wavelength modulation mode, in
combination with a conventional longitudinal PA cell yielded the
highest sensitivity (<100 ppbv). However, we also observed a strong cross-sensitivity to humidity, which can be explained by
increased V−T relaxation. This observation in combination with the limited dynamic range (max conc. ∼ 3000 ppmv) led us to
the use of the less-sensitive but spectroscopically more robust 2.9 μm laser. A detection limit below 1 ppmv, a dynamic range of
more than 4 orders of magnitude, no influence of humidity or any other substance relevant to the off-gas analysis, as well as a
comparable low price of the laser source made it the ideal tool for N2O analyses of the off-gas of a wastewater treatment plant.
Such a system was implemented successfully in a full-scale wastewater treatment plant. The results regarding the comparison of
different PA setups can be transferred to other systems, and the optimum performance can be selected according to the specific
demands.

N itrous oxide (N2O) is a strong greenhouse gas with a


global warming potential 298 times higher than that of
CO2 over a 100-year time horizon.1 N2O is also an ozone-
in the mid-infrared (MIR) region to quantify N2O. A limit of
detection (LOD) of 1 ppbv and a precision of 0.1 ppbv can be
achieved.7,8 Instruments are commercially available and can be
depleting substance (ODS) with its annual emissions of 17.7 applied to field measurements. However, the commercial
teragrams of nitrogen being the largest ODS emitted in the 21st instruments are hardly suitable for routine online analysis due
century.2 About 30% of these N2O emissions are related to to the rather high costs.9 Lasers in the MIR region are also used
human activities.3 This leads to an increase in the atmospheric for multielement absorption analyses.10,11 In this case, only one
N2O concentration of about 0.2−0.3% year−1 or an increase of absorption line is selected, corresponding to the laser
almost 20% since 1750.4 Anthropogenic emissions come from wavelength. Photoacoustic (PA) spectroscopy is a highly
agricultural activities, fossil fuel combustion, and wastewater sensitive method to measure optical absorption and can be
treatment. For the calculation of present and future N2O realized in small and robust setups, thus being particularly
emissions and the goal of reducing global warming, reliable data suitable for field measurements.
on N2O production and emission are required, which PA Detection of N2O. The PA effect is based on the
necessitates robust sampling and measuring strategies. Gas conversion of absorbed optical energy into heat, which again
chromatography equipped with electron capture detection is leads to thermal expansion.12 For gas-phase PA spectroscopy,
the method of choice at World Meteorological Organization’s usually a narrowband light source is intensity-modulated with a
Global Atmosphere Watch (GAW) stations, and it is also the frequency in the acoustic range (i.e., some kHz), which
most widely used analytical method.5 However, it has some generates a modulated acoustic expansion (i.e., an acoustic
disadvantages: frequent calibration is required and reference wave). Alternatively, wavelength scanning over a specific
gases have to be injected frequently to avoid signal drift. The
major disadvantage is its inability to measure continuously.6 By Received: February 9, 2017
contrast, optical techniques allow rapid and continuous Accepted: February 24, 2017
measurement. Fourier-transform infrared spectroscopy is used Published: February 24, 2017

© XXXX American Chemical Society A DOI: 10.1021/acs.analchem.7b00491


Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

absorption line of the target substance is possible. PA


spectroscopy has been applied in a surprisingly large variety
of configurations. In 1993, Yang et al. detected N2O in the
near-infrared region at 775.25 nm, generated by a titanium−
sapphire laser.13 Costopoulos et al. used a pulsed optical
parametric oscillator emitting at 2.86 μm, yielding an
impressive LOD of 60 ppbv. Others used a mechanically
modulated continuous-wave CO2-laser emitting at 10.532
μm,14 a semiconductor laser emitting around the N2O
absorption band at 2.292 μm,15 or a quantum cascade laser
(QCL).16,17
Lasers with a rapid wavelength tuning capability, such as
QCLs, allow for PA spectroscopy by wavelength modulation, as
first proposed by Castleden et al.,18 instead of the common
intensity modulation. Modulation of the wavelength over a
small range around a narrow absorption peak, typically in the
range of less than the fwhm of the absorption line, results in a
PA signal with twice the modulation frequency, allowing for
second-order detection. This signal is essentially baseline-free as
it is influenced neither by incoherent noise (acoustic noise from Figure 1. Fundamental experimental PAS setup.
outside the PA cell) nor by coherent noise, as caused by
wideband window absorption.19,20 Detection methods with
QCLs and multi pass sample cells have been developed.21 Only 134 mW. The wavelength was chosen to fit the absorption peak
a few examples on the combination of QCLs with photo- of N2O in the IR. Wavelength tuning of the QCL is performed
acoustic spectroscopy (PAS) are described in the litera- by changing the laser temperature by means of a temperature
ture.16,17,22,23 control unit (TEC-1089, Meerstetter Engineering GmbH,
In this paper, we present two new PA instruments for the Switzerland). The laser-diode driver (LDD; LDD-1124,
detection of N2O emissions from wastewater treatment plants Meerstetter Engineering GmbH, Switzerland) is connected to
(WWTPs). The instruments were combined with a dedicated an Arduino Uno (Arduino.cc, Italy), which triggers the internal
sampling system to collect gaseous emissions from WWTP modulation of the LDD. This is necessary because the LDD has
reactors. The intended use of this system is a routine no reference output signal as required for lock-in detection.
application in WWTPs and similar applications to monitor The laser is modulated by a sinusoidal current (Imax = 320 mA)
N2O emissions for a better process understanding and with a frequency of 2 kHz (modulation depth = 30 mA).
eventually for the development of N2O mitigation strategies. Temperature of the QCL is set to 18.3 °C, corresponding to an
Hence, the price and robustness are key parameters for the emission wavelength at the absorption peak of N2O. The laser
assessment of such a system. Consequently, we tested two current modulation results in a laser wavelength modulation
different PA systems, one employing a QCL source at 4.5 μm around the center of the N2O absorption peak, which in turn
excitation wavelength and one using a distributed feedback leads to a 2f PA signal. Thus, the first longitudinal acoustic
(DFB) laser source at 2.9 μm. At 4.5 μm, a higher sensitivity is mode of the resonant PA cell at 4 kHz is excited. The QCL
expected because N2O has its strongest optical absorption band light beam, collimated by a lens directly built into the laser unit
at this wavelength (see Figure S1, Supporting Information and then guided through two identical PA cells by two gold
(SI)), whereas the DFB laser source is significantly more cost- mirrors. All optical components were purchased from Thorlabs
efficient; a QCL system is typically 3 to 5 times costlier than a (Dachau, Germany). The second PA cell serves as reference
conventional diode laser of a comparable emission power. cell, which is sealed and filled with 1000 ppmv N2O. This cell is

■ EXPERIMENTAL SETUP
The aim of the presented work was the optimization of the
built identically to the first PA cell and refilled every six months.
The PA signal of this cell is demodulated at 3f, resulting in a
zero PA signal at the absorption maximum, which again allows
analytical performance of a PA system with a clear focus on the for a precise identification of the maximum. This information is
routine application in the rough environment of WWTPs with used to control the QCL wavelength via the QCL temperature
respect to sensitivity and selectivity as well as robustness and and the in-house made control software written in LabView
costs. The complete analytical system for the quantification of 2013 (National Instruments).
N2O emissions from WWTPs comprises a gas sampling hood, a The second laser system consists of a 2.9 μm DFB laser
PA system, and a pump providing a continuous gas flow from diode (Nanoplus, Germany) with an optical power of 5.5 mW.
the hood to the detector. We compared two different laser An integrated thermoelectric cooling device and a negative-
systems, operating at two different emission wavelengths, temperature-coefficient thermistor was attached to the laser
different PA cell designs, and different signal detection and diode (LD). Temperature and current were controlled by a
demodulation systems. LDC1000 (Profile, U.S.A.), leading to a variable emission from
Laser Systems. The PA setup basically comprises laser 2893 to 2900 nm. The laser is modulated with a simple
source including the corresponding driver and thermostatic transistor-to-transistor logic signal at 4 kHz. The laser beam is
control and a PA cell with an acoustic resonator (see Figure 1). focused into the PA cell described above by an aspherical lens
One laser source is a QCL (AdTechoptics, U.S.A.) purchased ( f = 25.4 mm). For this setup, a second PA cell was not
from EQ Photonics (Eching, Germany) with an emission band necessary. Temperature regulation was precise enough to
ranging from 4.53 to 4.54 μm and a maximum output power of ensure a high stability and therefore no wavelengths drift.
B DOI: 10.1021/acs.analchem.7b00491
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

PA Cells. The PA cells are conventional PA cells featuring valve is connected to the top of the floating hood to prevent air
longitudinal acoustic resonances in a central optical resonator from entering.
formed by a cylindrical glass tube (length 40 mm, inner To prevent foam, which can cover the reactor’s liquid surface
diameter 4 mm) equipped in its middle with a condenser as high as 20 cm, from entering the sampling line, a sieve cloth
microphone (EK-23029, Knowles, U.S.A.) for the PA signal (60 μm mesh) is placed between the gas-outlet and the tubing.
detection and a suitable preamplifier, as described elsewhere.24 A pump (VDE0530, KNF Neuberger GmbH, Germany)
In earlier experiments, we successfully used a Helmholtz type provides a continuous gas-flow of 700 mL/min through the
resonator as the PA detection cell in combination with a whole system of the sampling hood and PA analyzer. During
QCL.23 This cell type is particularly suitable for laser sources the aerated phases of the process, excess air can escape from the
with limited beam quality. The system essentially consists of ball check valve, which is connected to a mass flow meter (red-
two coaxial cylinders (diameter 22 mm, length 2 mm), y, Vögtlin Instruments AG, Switzerland). The concentration
combined by a connecting duct (diameter 4 mm, length 2 measured by our PA device, in combination with the total
mm). The laser source is positioned directly at the center of emitted volume, which is the sum of the flow through our
one of the cell inlets without any focusing element. Aluminum instrument and the volume emitted through the ball check
was chosen as the cell material due to its high reflectivity for the valve, allows for a precise calculation of the total N2O
laser light; additionally, the inner surface of the cell was emissions. During unaerated phases, ambient air is aspirated
polished. A gas inlet is positioned laterally into the same cell, through the gas-inlet for keeping the system at ambient
while a microphone sits in the other cell right opposite to the pressure. The low flow rate in the sampling hood allows for
connecting tube. Finite element simulation (Comsol Multi- equilibration between the liquid-phase and the gas phase N2O
physics, Germany) was performed to estimate the eigenfre- concentration. The sampling hood is held in position by an
quency of the cell (see Figure S2, SI). aluminum bar of 4 m length, connected to the hood by a hinge
As an alternative to the microphone described above, which joint enabling us to follow the water surface even during
was used in longitudinal PA cells, a microelectro-mechanical turbulent aeration phases. Simultaneous determination of the
dissolved N2O concentration is performed using an electro-
system (MEMS) microphone from Knowles (SPM0408LE5H-
chemical N2O sensor with a detection limit of 0.5 μM, a
TB, Knowles, U.S.A.) was tested. The advantage of this
response time below 45 s and a working range up to 300 μM
microphone is its high sensitivity of 126 mV/Pa and an output
(Unisense Environment A/S, Denmark), fixed in the center
impedance of 200 Ω, which makes higher amplification
part of the hood approximately 20 cm below the liquid surface.
possible. As an all solid-state device, it is particularly robust, As an alternative to the measurement using the sampling hood,
chemically as well as mechanically. the total N2O emission of the reactor could be sampled via the
Signal Recording and Demodulation. Three different headspace, as this reactor is covered by a nearly gastight hatch.
approaches for signal recording and demodulation were An exhaust tube on its top allows for a time-resolved, spatially
compared. The most conventional system uses a commercial integrating analysis of the gas emission.
lock-in amplifier (SR 830, Stanford Research, U.S.A.) to detect Calibration Gases. For the calibration of the PA-systems,
the preamplified PA signal. In the second approach, the signals commercial test gases of different concentrations were used:
are recorded using an analogue to digital converter (ADC) (NI- 98%, 1%, 200 ppmv and 20 ppmv N2O in nitrogen (Westfalen
PCI 4474 dynamic signal acquisition card, National Instru- AG, Germany). The test gases were diluted with pure nitrogen
ments, U.S.A.) recording four channels at 102.4 kS/S and a 24- (99.99% N2) using rotameters. To assess the influence of
bit resolution. The signal lock-in is performed by an in-house humidity on the PA signal, water-saturated test gases were
software, also programmed in LabView 2013. The third achieved by bubbling the gases through a washing bottle and
approach employs a simple PC soundcard (24-bit resolution, then diluted to the required humidity.


Xonar DG, ASUS, Taiwan) instead of the ADC for the signal
recording, but the same software-based demodulation. Temper- RESULTS AND DISCUSSION
ature and humidity of the gas are measured by analogue sensors Comparison of the Recording and Demodulation
(HIH4010−004, Honeywell, U.S.A. and TMP36GZ, Analog Systems. A comparison of the three data recording systems
Devices, Inc. U.S.A.) connected to the measurement computer was performed using the 2.9 μm DFB laser source. It was found
via the Arduino. The whole setup is integrated in a mobile that the Stanford Research SR830 lock-in amplifier yields an
housing, allowing field-measurements. approximately 10% lower standard deviation, resulting in a
Gas Sampling System for WWTPs. For the collection of slightly lower LOD than the combination of the ADC with the
the emitted off-gas at a specific spot in a full-scale wastewater software demodulator (see Figure S4, SI).
treatment plant (more specifically in the nitritation step of a The comparison of the ADC with the soundcard yielded the
two-stage deammonification process being intermittently surprising result that there was no significant difference
aerated by aeration membranes installed at the bottom of the between the two with respect to the LOD and the relative
reactor with an aeration duration of 45 min followed by an standard deviation of the signals. A major limitation of the
unaerated phase of 15 min), a floating sampling hood was soundcard approach is that it has only two channels, one for the
placed in the reactor (see Figure S3, SI). The square-shaped PA signal and one for the modulation reference required for the
hood is made up of stainless steel (V2A) with floating bodies lock-in, meaning that the detection of two independent PA
on each side and a cross-sectional surface of 0.5 m2. A gas-inlet channels is not possible (i.e., the signal from two PA cells
and a gas-outlet are located on two opposite corners of the cannot be recorded simultaneously). The SR830 has two input
hood. The gas-outlet is connected via gastight tubing to the PA- channels plus a separate channel for the modulation reference,
detection system, which actively sucks in the sample air. Short but these two channels can only be used in a differential way,
distances between the hood and the PAS system were not for independent measurements. Only the ADC has four
emphasized to enable a quick response time. A ball check independent channels, of which one is required for the
C DOI: 10.1021/acs.analchem.7b00491
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

modulation reference. As mentioned above, for the 4.5 μm


QCL laser source a second PA cell with a corresponding
detection channel is required. Hence, we used the combination
of the ADC with a software lock-in throughout the following
experiments.
Comparison of the Laser Sources. The combination of
the intensity modulated (1f) 2.9 μm DFB laser system with the
conventional longitudinal PA cell and the ADC detection
system demonstrated to be a good compromise between
sensitivity, robustness, and cost efficiency. Figure 2 depicts a

Figure 3. Correlation of the PA signal excited at 2.9 μm to the N2O


concentration (T = 295 K, cwater = 6.83 g/m3).

laser system was tested also for long-term stability. After a


continuous use of 24 h/day for 3 months, a repeated calibration
Figure 2. Simulated spectrum of 70 ppmv N2O + 1% H2O in air, on
was within the standard deviation of the previous signals.
the basis of absorption data from the HITRAN database (red line), With the use of the MEMS-microphone instead of the
compared with the measured spectrum with a DFB laser (blue line). electret microphone, a significantly higher sensitivity was
expected. Indeed, experimental results show signals that are
doubled. However, the noise also more than doubled.
scan over the wavelength range accessible with this laser source, Additionally, the MEMS microphone we used features a nearly
measured in air containing 70 ppmv N2O in comparison to the constant frequency response down to 100 Hz, making the
theoretical absorption values of N2 containing N2O and H2O as system very sensitive to the ubiquitous 1/f background noise.
extracted from the HITRAN database.25 While the overall Thus, continuous flow measurements were impossible. The
correlation between theory and measured data is satisfying, the LOD of this configuration was 3.85 ppmv. The approach of
divergence between the two increases at the high wavelength combining two of these microphones, one in each of the two
range of the covered spectrum. The reason is the lower laser cells of the Helmholtz resonator, operated in a differential
emission intensity of the laser at higher temperatures. A mode, was not successful.
temperature ramp is required to tune the laser emission The N2O absorption at 4533.7 nm is the highest of the
wavelength. For further measurements, 2894.7 nm was chosen complete IR spectrum and it is not cross-sensitive to any other
as the signal wavelength. The H2O influence should be relevant component (CH4, NH3, CO2, CH3SH). Consequently,
negligible in this region of the spectrum. For an application in the highest sensitivity of the PA analysis is to be expected at this
WWTPs cross-sensitivity to other gases was assessed. The wavelength. The selected QCL features its central wavelength
following substances were reported to be emitted from typical at exactly this wavelength. Figure S6 (SI) displays a calibration
reactors: methane, ammonia, carbon dioxide, hydrogen sulfide, of the PA system at this wavelength, produced with 2f
and methyl mercaptan.26,27 Absorption coefficients of the wavelength-modulation schema and the conventional longi-
mentioned gases are significantly lower than the absorption tudinal PA cell. The LOD here was found to be 77 ppbv, thus
coefficient of N2O in the two wavelength ranges of the lasers as well below the natural atmospheric background concentration
it can be appreciated in S5 (SI).28 The absorption spectra and more than sufficient for our application. However, the
presented there are calculated with the concentration listed in dynamic range of the system when using the QCL is limited to
the reference by using the HITRAN data.25 This low spectral a maximum of about 3000 ppmv; at higher concentrations, the
overlap in combination with the wavelength tuning approach PA signal saturates, and above 4000 ppmv it even decreases.
rules out the risk of spectral interferences. We assume that this behavior is due to the high optical
A calibration with this setting and the resonant PA cell was absorption at the selected wavelength, which leads to a
performed (see Figure 3). From this data, an LOD of 2.0 ppmv significant loss of optical energy along the optical axis before
can be calculated (3·sb-criterion). Further improvement to an reaching the center of the PA cell. This effect is not observed
LOD of 0.82 ppmv is possible by switching to 2f modulation; at for the 2.9 μm PA setup because of the lower absorption
the same time, the calibration is linear up to a concentration of coefficient of N2O at 2.9 μm.
1%. This dynamic range of more than 4 orders of magnitude is Another important feature is shown in Figure 4, where the
certainly one of the most impressive figures of merit of this PA PA signal at different constant N2O concentrations is plotted
detection system. In industrial process monitoring like the against the humidity. A spectral interference of the H2O
intended use at WWTPs, such a dynamic range is essential, as absorption with the N2O measurement can be excluded. For
the N2O concentrations can vary greatly depending on diverse the 2.9 μm DFB laser, no cross-sensitivity to H2O is observed.
influencing factors (e.g., the aeration regime). The 2.9 μm DFB The QCL PA signals of N2O scale linearly with the humidity.
D DOI: 10.1021/acs.analchem.7b00491
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

software lock-in amplifier and OEM-laser drivers makes it


possible to integrate the whole setup in a small housing of 30
cm × 40 cm × 22 cm (see Figure S7, SI), making the system
highly mobile. A disadvantage of this system is the high
dependency of the measured signal on the humidity. Even
though calibrations for different humidities have been
performed, a complete correction, particularly for varying
humidities, is difficult and reduces the precision. Possible
humidification of the incoming sampling gas may result in
higher signals and controlled conditions for quantification. A
desiccation of the sampling gas may result in analyte loss.
Additionally, the lower signals are disadvantageous.
The LOD of the 2.9 μm PA setup is more than 1 order of
magnitude lower than that of the QCL PA system. The reason
is the lower power of the laser diode and the lower absorption
coefficient of N2O at 2.9 μm. The dynamic range of this system
is better than that of the QCL PA setup. A major advantage of
this setup is the humidity independence. Table 1 compares the
Figure 4. QCL PA signal at different humidities and concentrations tested PA systems.
(no. of repeated measurements: 5; no. of data points: 9).
Table 1. Comparison of the Different PA Systems
At a temperature of 22 °C and a relative humidity of 100%, the Investigated
PA signals of N2O is about 7 times higher than at 0% humidity
QCL system
for all N2O concentrations in the linear calibration range. A (4.5 μm) DFB system (2.9 μm)
possible reason for this behavior is vibrational−translational longitud. PA Helmholtz PA
(V−T) relaxation of N2O molecules. Briefly, collisions with longitud. PA cell cell cell
H2O molecules lead to a reduced lifetime of the state excited by sensitivity 77 ppbv 0.82 ppmv 1.98 ppmv
the absorption, which again lead to a more intense, in-phase PA dynamic range 104 104 n.d.
signal.29−32 H2O influence strong − −
Obviously, for a practical PA analysis at this wavelength of mech. stability + +++ ++
4.5 μm, either the humidity should be kept constant or the mobility ++ + ++
influence of humidity should be eliminated. We chose the latter
option: the humidity sensor mentioned above is used to
measure simultaneously temperature and humidity of the gas In summary, we decided to use the 2.9 μm DFB laser system
flowing through the PA cell. Our LabView control program with 2f modulation and ADC signal recording for the field
implements a routine that corrects the N2O concentration applications described below. It provides the best compromise
using the measured values of the sensor and the slope of the between sensitivity on the one hand and robustness,
linear regressions for constant N2O concentrations at different spectroscopical as well as mechanical, on the other hand.
humidities measured in the laboratory. A disadvantage of this These stability considerations include the long-time stability of
procedure is the spatial separation of humidity determination the calibration, which did not feature any significant changes
and PA measurement, leading to a temporal offset that has to throughout our field measurements (i.e., for at least 3 months).
be corrected for. For our specific application at WWTPs, we The limited dynamic range and the cross sensitivity of the 4.5
found that the relative humidity (100%) as well as the μm QCL system to water plus the significantly higher price
temperature (22 °C) were constant throughout all our make this approach less suitable for the specific application,
experiments. particularly as the sensitivity of the 2.9 μm DFB system is more
Alternative Cell Design and Setups. Finite element than sufficient for this task.
simulations with COMSOL Multiphysics 5.0 (Comsol A.B., Field Application on a WWTP. As an example application
Sweden) determined the resonance frequency of the Helmholtz we tested the 2.9 μm PA system for its practical applicability.
resonator to be 3506 Hz, while experimentally, we found a Two sampling approaches were tested: using the gas sampling
resonance maximum at 3640 Hz, which is a satisfying result hood designed for spatially resolved sampling and total
considering the limitations of such static simulation. The emission analysis over the entire reactor headspace. The
determined resonance frequency fits into the frequency range temporal behavior of the N2O concentration in the liquid-phase
of maximum sensitivity for both microphones. and in the PA-measured gas-phase over three consecutive
The use of intensity modulation (on/off) was not possible aeration cycles is shown in Figure 5. It can be appreciated that
for this cell type because absorption of the laser at the cell wall the liquid-phase concentration is increasing over a period of 15
results in a very high background signal. Therefore, 2f min. Then aeration starts, and the liquid-phase concentration
frequency modulation was used. Measurements with this cell decreases quickly throughout the full aeration cycle of 45 min,
type show a high standard deviation for concentrations below 1 as N2O is stripped out of the liquid. With the stripping gas,
ppmv. The detection limit was determined to be 1.98 ppmv, dissolved N2O is transferred into the gas phase, as the PA data
which is 2 times higher than that for the longitudinal PA cell at reveal. When the aeration ends, the liquid-phase concentration
2f modulation. starts to build up again. The gas-phase concentration increases
Comparison of the Two PA Setups. The highest at the beginning of the aeration, reaches a maximum after about
sensitivity is achieved for the QCL PA system. The use of 15 min, and then decreases constantly. The reason is that the
E DOI: 10.1021/acs.analchem.7b00491
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

■ AUTHOR INFORMATION
Corresponding Author
*E-mail: Haisch@tum.de. Tel. +49 89 2180-78242. Fax -99
78242.
Author Contributions
The manuscript was written through contributions of all
authors. All authors have given approval to the final version of
the manuscript.
Notes
The authors declare no competing financial interest.

Figure 5. N2O concentration in the gas phase (blue) measured by our


PA system and in the liquid phase (red), monitored at a WWTP.
■ ACKNOWLEDGMENTS
The authors thank S. Wiesemann and R. Hoppe of the
institute’s mechanical workshop for the construction of the
liquid-phase concentration is already significantly reduced by mobile PA-system and the development of the gas sampling
the stripping, and thus, only a small fraction of N2O can be device. We thank also Prof. Frank K. Tittel for his support.
transferred into the gas phase. Financial support of the project and particularly of K.T., C.B.,
The data shown in Figure 5 reveal that the PA setup has a and C.L. by the International Graduate School of Science and
higher temporal resolution than the electrochemical sensor. Engineering IGSSE of the Technical University of Munich is
The short-term increases in the decreasing gas-phase gratefully acknowledged.


concentration (e.g., at 12.10 h) is due to local fluctuations.
The bioreactor content is not homogeneous and has strong REFERENCES
spatial variations. A mixing device installed at medium height of (1) Myhre, G., Shindell, D.; Bréon, F.-M., Collins, W.; Fuglestvedt, J.;
the reactor leads to local fluctuations. The total emission Huang, J.; Koch, D.; Lamarque, J.-F.; Lee, D.; Mendoza, B.; Nakajima,
sampling at the top of the reactor cover avoids this problem. An T.; Stephens, G.; Takemura, T.; Zhang, H.. In Climate Change 2013:
overnight analysis is plotted in Figure S8 (SI). Qualitatively, the The Physical Science Basis. Contribution of Working Group I to the Fifth
behaviors of both, gas-phase and liquid-phase N2O concen- Assessment Report of the Intergovernmental Panel on Climate Change;
trations are identical. The variations over the whole analysis Stocker, T. F., Qin, D., Plattner, G.-K., Tignor, M., Allen, S. K.,
time can be attributed to spatially variable conditions in the Boschung, J., Nauels, A., Xia, Y., Bex, V., Midgley, P. M., Eds.;
reactor, such as different aeration intensities causing not only Cambridge University Press: Cambridge, U.K., 2013.
differential stripping effects, but also varying dissolved oxygen (2) Ravishankara, A. R.; Daniel, J. S.; Portmann, R. W. Science 2009,
326, 123−125.
concentration affecting the biological N2O production.


(3) Wuebbles, D. J. Science 2009, 326, 56−57.
(4) Machida, T.; Nakazawa, T.; Fujii, Y.; Aoki, S.; Watanabe, O.
CONCLUSIONS Geophys. Res. Lett. 1995, 22, 2921−2924.
For the detection and quantification of gaseous N2O (5) World Meteorological Organization (WMO). GAW Report No.
concentration, different PA setups were tested and compared, 185. Guidelines for the Measurement of Methane and Nitrous Oxide and
namely, a QCL system at 4.5 μm and a DFB system at 2.9 μm. their Quality Assurance; WMO: Geneva, Switzerland, 2009.
The LOD was improved by the use of 2f modulation. The costs (6) Hamilton, J. F.; Lewis, A. C. In Analytical Techniques for
of the setups were reduced using a PC soundcard as an analog- Atmospheric Measurement; Dwayne, H., Ed.; Blackwell Publishing:
to-digital converter and a software lock-in amplifier instead of a Oxford, 2006.
(7) Hammer, S.; Griffith, D. W. T.; Konrad, G.; Vardag, S.; Caldow,
much larger instrumental lock-in amplifier. This also allows for C.; Levin, I. Atmos. Meas. Tech. 2013, 6, 1153−1170.
the development of a robust and highly mobile setup, which (8) Griffith, D. W. T.; Deutscher, N. M.; Caldow, C.; Kettlewell, G.;
was tested at a WWTP. The results regarding the comparison Riggenbach, M.; Hammer, S. Atmos. Meas. Tech. 2012, 5, 2481−2498.
of different PA setups is valid for other analyst accessible in the (9) Rapson, T. D.; Dacres, H. TrAC, Trends Anal. Chem. 2014, 54,
IR range and can thus be transferred to other applications, such 65−74.
as automotive exhaust gas monitoring or industrial process (10) Yu, Y.; Sanchez, N. P.; Griffin, R. J.; Tittel, F. K. Opt. Express
monitoring. The specific optimum performance can be selected 2016, 24, 10391−10401.
according to the specific demands. (11) Dong, L.; Tittel, F. K.; Li, C. G.; Sanchez, N. P.; Wu, H. P.;


Zheng, C. T.; Yu, Y. J.; Sampaolo, A.; Griffin, R. J. Opt. Express 2016,
ASSOCIATED CONTENT 24, A528−A535.
(12) Haisch, C. Meas. Sci. Technol. 2012, 23, Article No. 012001.
*
S Supporting Information
(13) Yang, X. K.; Petrillo, C. J.; Noda, C. Chem. Phys. Lett. 1993, 214,
The Supporting Information is available free of charge on the 536−540.
ACS Publications website at DOI: 10.1021/acs.anal- (14) Schramm, D. U.; Sthel, M. S.; da Silva, M. G.; Carneiro, L. O.;
chem.7b00491. Souza, A. P.; Vargas, H. Rev. Sci. Instrum. 2003, 74, 513−515.
Additional data including absorption coefficients of N2O, (15) Grossel, A.; Zeninari, V.; Parvitte, B.; Joly, L.; Courtois, D. Appl.
finite element simulation of acoustic pressure and sound Phys. B: Lasers Opt. 2007, 88, 483−492.
(16) Rocha, A. M.; Sthel, M. S.; de Castro, M. P. P.; Mothe, G. A.;
pressure, image of gas sampling hood, relative standard Silva, W. C.; Perez, V. H.; da Silva, M. G.; Miklos, A.; Vargas, H. Energy
deviation of a commercial lock-in amplifier compared to Fuels 2014, 28, 4028−4032.
a software lock-in amplifier, simulated absorption spectra (17) Lima, J. P.; Vargas, H.; Miklos, A.; Angelmahr, M.; Hess, P.
of the relevant biogas components, calibration of the Appl. Phys. B: Lasers Opt. 2006, 85, 279−284.
QCL-PA-system, image of PA prototype instrument, and (18) Castleden, S. L.; Kirkbright, G. F.; Spillane, D. E. M. Anal. Chem.
N2O concentration in WWTP (PDF) 1981, 53, 2228−2231.

F DOI: 10.1021/acs.analchem.7b00491
Anal. Chem. XXXX, XXX, XXX−XXX
Analytical Chemistry Article

(19) Peng, Y.; Zhang, W.; Li, L.; Yu, Q. X. Spectrochim. Acta, Part A
2009, 74, 924−927.
(20) Schilt, S.; Kosterev, A. A.; Tittel, F. K. Appl. Phys. B: Lasers Opt.
2009, 95, 813−824.
(21) Santoni, G. W.; Daube, B. C.; Kort, E. A.; Jiménez, R.; Park, S.;
Pittman, J. V.; Gottlieb, E.; Xiang, B.; Zahniser, M. S.; Nelson, D. D.;
McManus, J. B.; Peischl, J.; Ryerson, T. B.; Holloway, J. S.; Andrews,
A. E.; Sweeney, C.; Hall, B.; Hintsa, E. J.; Moore, F. L.; Elkins, J. W.;
Hurst, D. F.; Stephens, B. B.; Bent, J.; Wofsy, S. C. Atmos. Meas. Tech.
2014, 7, 1509−1526.
(22) Couto, F. M.; Sthel, M. S.; Castro, M. P. P.; da Silva, M. G.;
Rocha, M. V.; Tavares, J. R.; Veiga, C. F. M.; Vargas, H. Appl. Phys. B:
Lasers Opt. 2014, 117, 897−903.
(23) Lendl, B.; Ritter, W.; Harsek, M.; Niessner, R.; Haisch, C.
Proceedings of IEEE Sensors 2006, EXCO; Daegu, Korea, October 22−
25, 2006; pp 338−341.
(24) Beck, H. A.; Niessner, R.; Haisch, C. Anal. Bioanal. Chem. 2003,
375, 1136−1143.
(25) The HITRAN Database. http://www.cfa.harvard.edu/
HITRAN/.
(26) Frison, N.; Chiumenti, A.; Katsou, E.; Malamis, S.; Bolzonella,
D.; Fatone, F. J. Cleaner Prod. 2015, 93, 126−133.
(27) Lackey, J. C.; Peppley, B.; Champagne, P.; Maier, A. Waste
Manage. Res. 2015, 33, 767−771.
(28) Du, Z.; Zhen, W.; Zhang, Z.; Li, J.; Gao, N. Appl. Phys. B: Lasers
Opt. 2016, 122, 100.
(29) Rooth, R. A.; Verhage, A. J. L.; Wouters, L. W. Appl. Opt. 1990,
29, 3643−3653.
(30) Kosterev, A. A.; Mosely, T. S.; Tittel, F. K. Appl. Phys. B: Lasers
Opt. 2006, 85, 295−300.
(31) Avramides, E.; Hunter, T. F. J. Chem. Soc., Faraday Trans. 2
1979, 75, 515−527.
(32) Avramides, E.; Hunter, T. F. Chem. Phys. 1981, 57, 441−451.

G DOI: 10.1021/acs.analchem.7b00491
Anal. Chem. XXXX, XXX, XXX−XXX

You might also like