Report Generator Lite

(NS9200LT)

User’s Manual

NEC Avio Infrared Technologies Co., Ltd.


Table of Contents

1.1 Introduction ..............................................................................................................................................................................2


2.1 Contents ......................................................................................................................................................................................3
3.1 Product Overview ..............................................................................................................................................................3
3.2 Cautions .......................................................................................................................................................................................3
4.1 How to Install..........................................................................................................................................................................4
5.1 Program uninstalling ..................................................................................................................................................... 10
6.1 Startup of Report Generator Lite (from the Start Menu) .................................................... 12
6.2 Report Generator Lite Startup (from the Shortcut Icon)..................................................... 12
7.1 Menu Structure ................................................................................................................................................................. 13
8.1 Names and Functions ................................................................................................................................................. 15
8.1.1 Detail of Controller .............................................................................................................................................. 24
8.1.2 Saving thermal image data ........................................................................................................................... 31
8.1.3 POINT ................................................................................................................................................................................ 33
8.1.4 BOX ...................................................................................................................................................................................... 38
8.1.5 Delete all objects ................................................................................................................................................... 42
8.6 Create Graph ....................................................................................................................................................................... 43
8.6.3 Trend................................................................................................................................................................................... 43
9.1 Menu Operations.............................................................................................................................................................. 45
9.1.1 Thermal Image Menu .......................................................................................................................................... 45
9.1.2 Edit Menu ....................................................................................................................................................................... 55
9.1.3 View Menu ..................................................................................................................................................................... 80
9.1.4 Option menu ................................................................................................................................................................ 83
9.1.5 Help menu ...................................................................................................................................................................... 89
10.1 Specifications................................................................................................................................................................... 90
10.2 Operation Environment .......................................................................................................................................... 92

-2- No.95691-2464-0000
1.1 Introduction

Thank you for purchasing our Report Generator Program (NS9200LT). Please read this manual before
operating the program. This manual provides the information necessary to operate the Report Generator
Lite. If you encounter any problems in the manual, please contact our sales representative.

For safe operation


This program is manufactured considering safe operation. Mishandling and mistake can cause
malfunctioning. Carefully read and understand the contents of this manual.

Note
● The contents of this manual are subject to change without notice.
● No part of this manual may be reproduced or transcribed without written permission.
●* Please inform us of any points that are not clear or are missing from this manual
● Irrespective of item * above We do not assume any responsibility for results due to user operations.

-2- No.95691-2464-0000
2.1 Contents

1)Install disk(CD-ROM) x 1
It is installation CD-ROM to install this program.

3) THE SOFTWARE LICENSE AGREEMENT x1


Please be sure to read before the using.

3.1 Product Overview


This program is formed with the template, which compiles thermal image playback and report generator
and report, which analyze output layout. As for the report generator which possesses the operability,
which is similar to the Windows explorer, display function and the figure where the thermal image, the
visible picture, and synthesis indication are possible (the designated POINT, BOX) it is possible to output
the temperature data of the specific part which is used. Furthermore, it is possible in the database to
make the time, which searches the past file with register of the thermal image and search function short.
At the report generator, the thermal image and the figure data, which are set, are output by the
designated template, they are possible to draw up the user individual report instantaneously.

3.2 Cautions

1. To operate the Report Generator Lite, observe the operating conditions of the PC under the proper
operating environment.
2. The master disk of the Report Generator Lite should be stored with care. Avoid improper storage
environment where exposed to harsh temperature/humidity, direct sunlight, and dust. (0-40degC,
< 85%)
3. Whatever reason, no part of Report Generator Lite may be reproduced, copied, transferred, and
sold.
4. For purchased devices other than the Report Generator Lite, follow the warnings described in their
respective devices. For the products manufactured or sold other than NEC Avio Infrared
Technologies Co., Ltd., follow the description and conditions of their respective manufacturers and
sales agencies.
5. NEC Avio Infrared Technologies Co., Ltd. does not assume any responsibility for results due to
user operations of Viewer Program.

Company and product names in this manual are their respective trademarks or registered trademarks.

-3- No.95691-2464-0000
4.1 How to Install

Note [Attention]
When setting up, Administrator authority with a certain
user name login, or, Administrator from obtaining
permission, please execute setup.


Operation

(1) Start up the computer and Windows.


The Windows desktop screen appears.

My computer
This icon denotes this computer. Double-click this icon to
see the hard disk or REMOVABLE contents.

Start button
Buttons to be used for application software.

-4- No.95691-2464-0000
(2) CD of “NS9200LT Report Generator Lite” is inserted in CD-ROM drive.

(3) CD-ROM drive is opened, and "NS9200LT" is double-clicked.

(4) A mouse cursor is united with “Setup.exe” of REPORTGEN and it double-clicks there.

-5- No.95691-2464-0000
Since a message as shown in the following figure is displayed when, as for the cases other than
the authority of Administrator, the authority of the user at the time of login double-clicks
Note "SETUP.EXE", be careful to be installed by the authority of Administrator.

The login name and password of


Administrator are entered.

Please right-click “setup.exe", and select "Run as administrator" when you


use WindowsVista. Moreover, please select "Allow" when the
Note confirmation dialog is continuously displayed by Windows.

WindowsVista

-6- No.95691-2464-0000
Reference When checking the present login user's authority, "a user and a password" or a "user account"
of a "control panel" is chosen.

Authority of user at time of login, the authority of Administrator at other than in case of use,
the copy of the file (download it includes), there are times when error message is indicated
Note
with operation such as retention and deletion. This OS (Windows) with because it is the
phenomenon which it occurs with the functional restriction, the copy (retention) ahead,
becomes operational by the fact that it is modified to "desktop" of the login user or the "my
document" etc..

(6) Start the setup program. The “Welcome” screen appears. Then, click the [Next (N)] button.

-7- No.95691-2464-0000
(7) The [Select Installation Folder] screen appears for specifying the folder for installing the program.
The target drive for installation is the Windows drive as the default. To change the drive, click
[Browse…] button. You can change the folder. If changes are unnecessary, click the [Next (N)]
button.

If the name of the folder for installation includes two-byte characters, faulty
operations may result. It is, therefore, strongly recommended to install the
Note program under the default setting or using only one-byte characters for the
folder name.

(8) If the following "next" button is pushed, since the dialog which checks the start of installation of the
following figure will be displayed, a "Next" button is pushed when you may start.

-8- No.95691-2464-0000
(9) After the installation of this program completed, the Setup completion screen appears. Then, click the
[Finish] button to finish the setup program.

- Perform setup while other application programs do not operating. If an error


message appears, reset the computer and reinstall the program according to
Note the setup steps.
- Never remove the master program disk during the setup.

-9- No.95691-2464-0000
5.1 Program uninstalling
This section explains how to uninstall this program.


Operation

(1) Click the Start to display the menu.

(2) Place the mouse on Settings and move rightward (in Windows2000).
(3) Select and click Control Panel.

- 10 - No.95691-2464-0000
(4) The Control Panel screen appears. Double-click Add/Remove Programs

(4) The Add/Remove Programs screen appears. Find “Report Generator LT” using scroll bar and click
this item.

(5) Click the [Remove] button.

- 11 - No.95691-2464-0000
(6) A screen for confirmation appears. Click the [Yes] button in this screen.

6. Report Generator Lite Startup

6.1 Startup of Report Generator Lite (from the Start Menu)

Place the mouse pointer on [Program] in the Start menu and click [Report
Generator Lite] in [NEC_Avio], then the Report Generator Lite starts.

NS9200 Report Generator

6.2 Report Generator Lite Startup (from the Shortcut Icon)


Operation
Double-clicking the shortcut icon on the desktop start up the Report Generator Lite.

- 12 - No.95691-2464-0000
7.1 Menu Structure
The Report Generator Lite menu bar structure and functions are as follows.

1.Thermal Image
Open: Replays one thermal image
Open Files: Replays more than one thermal image
Save: Saves by overwriting the current state
Save as: Saves the current state with a different name
Print: The present picture by which it is indicated by the preview
window is printed.
Print setting: It sets the printer which it prints.
Send E-mail: Send E-mail
Exit: Exits the Report Generator Lite

2.Edit
cut: Cut the selected file.
copy: Copy the selected file.
paste: Paste the copied file.

Image copy: A display image is copied.(Thermal only, Thermal with color bar,
Thermal with measurement condition)
Save image file: Saving in BMP or JPEG format (one picture by a operation)
Save thermal image: Saves only thermal image in BMP or JPEG format
Save thermal image with color bar: Saves the image in BMP or JPEG format
with a color bar.
Save thermal image with measurement condition: Saves the image in BMP or
JPEG format with measurement condition.

Save AVI file


Save thermal image as AVI: Saves only thermal image portion in AVI format
Save thermal image with color bar as AVI:Saves the screen to AVI format
Save thermal image with measurement condition as AVI: Saves the image in AVI
Format with condition bar and measurement condition.
Generate Report: Data is outputted to the specified template and
a report is created.

3.View
Tool Bar: Switches Show or Hide of tool bar
Status Bar: Switches Show or Hide of status bar
Update Folder: Updates to the folder information to the newest state
New Folder: Creates a new folder in the focused directory/folder

- 13 - No.95691-2464-0000
5.Object
Add Object:
POINT: Add the One POINT.
BOX: Add the One BOX.
Delete Object:
All Objects: Delete All Objects.
All Points: Delete All Points.
All Boxes: Delete All Boxes.

6.Graph
Trend: Display the time series data of the selection object.
(The time series data is made in order of the page of
the thermal image that is opening now.)

7.Option
Save continuously: Saves several images in BMP or JPEG file all at once
Level/Sense hold: Holds current Level and Sense setting
Environment: Sets temporary folder for saving thermal image files
The contents of a trend graphical representation are set up.

8.Help
About Report Generator Lite

- 14 - No.95691-2464-0000
8.1 Names and Functions
Functional name and that function of the report generator become as follows.

(5) Tool bar (6) Controller

(2) Folder selection

(7) Information Tab

(1) Main image


view window

(4) Status bar (3) Image preview


Function
(1) Main image display window
This window displays a thermal image and a visual image appended to the thermal image.
The image is displayed when a thermal image is opened from the Thermal Image menu or
double-clicking is made at the Image preview window.

(2) Folder selection


Select a folder including the image files you wish to view. Operate the same method as for
Windows Explorer. To update to the newest state, select [Update Folder] in the [View] menu.

(3) Image preview window


Thermal files, BMP files, and JPEG files in the selected folder are displayed as preview. By
double-clicking the image, thermal image appears on the main image screen. The preview
image can be scrolled with the scroll bar.

- 15 - No.95691-2464-0000
(4) Status bar
Menu bar status is indicated. “Hide” or “Show” in Status bar in the [View] menu can be switched.
* A check mark is indicated while the status is in “Show.”

With status bar Without status bar

Moreover, when moving a mouse cursor on a thermal image, the coordinates value and temperature value
of a position of a mouse cursor are displayed on the program lower right.

On a status bar, X coordinates,


Y coordinates, and a temperature value
are shown from the left side.

When a temperature value is not displayed on a status bar, it is clicking a thermal


Note image top once, and is displayed.

- 16 - No.95691-2464-0000
(5) Tool bar
As for the tool bar, the icon of the following figure is one group.

[Explanation about functions]


"Open" …Open the "Open File" dialog to play back a thermal image.
(jpg,sit,six, tmp,svt and svx)
"Open Files" …Open the "Open Files" dialog to play back multiple thermal images. (jpg,sit,six and
tmp)
"Save " …Save the file.
"Print" …Print the displayed thermal image out.
"Cut" …Cut the selected file or a setting object is cut off.
"Copy" …Copy the selected file or a setting object is copied.
"Paste" …Paste the copied file or a setting object is paste.
"Rescale Thermal Image"…Rescale a thermal image. (100% to 5000%)

Open

Save
Zoom
Cut Select the zoom tool or factor

Paste

Copy

Print

Open Files

- 17 - No.95691-2464-0000
Details display

Thumbnail display

Preview display
Set a BOX

Set a POINT

Select a analysis objects

Object setting

[Explanation about functions]


"Set a POINT" …Set a POINT.
"Set a BOX" …Set a BOX (rectangle).
"Select a analysis object"…Alter a Object set on a thermal image.
“Object setting” …Setting in the selective object is modified.

"Preview display" …Preview the main image display window (display an image).
"Thumbnail display" …Display image files in the selected folder as thumbnail (reduced) display.
"Details display” …Display detailed information about image files in the selected folder.

- 18 - No.95691-2464-0000
(6) Controller
By the controller, a setup of Level & Span and a Sense and the change of a page, voice replay, and
visible / fusion display are set up to the thermal image at the time of a preview display or a thumbnail
display.

Level span upper value


(10s step, 1s step, 0.1s step)

Level value
(10s step, 1s step, 0.1s step)

Level span lower value


(10s step, 1s step, 0.1s step)

Sensitivity setting
(When the level span is set, the sensitivity value is
automatically calculated.)

Multi Sense
(The resolution of a display color is changed.)

Ambient Temperature
(Used to set the ambient temperature in for emissivity
correction)

Emissivity
Set Emissivity of Whole image.

Color mode
Change the color palette
ISO setting
Only the specification range applies a color palette.
Color Step
8,16,32,64,128,256 colors
Page
Replay, Stop, Span, Previous page or Next page

Voice replay
(Replay, Pause, Stop)

Display switching
(Thermal, Visible, Fusion)

Next / Back Button


(The next or back image can be viewed in
the image preview.)

Editing can be made for the contents in editing boxes on the screen. To
Reference
validate the settings that have entered, press the [Enter] key when the focus
is on the edit box.

- 19 - No.95691-2464-0000
(7) Image Info tab
The Image info tab displays the information on the previewed thermal image file, output results of the set
Analysis objects, graph display, and other information. To switch between the tabs, press the button on
the right. ( )

Image info
The Image Information tab displays the file information on the previewed thermal image.


File name: Displays the full pathname of the previewed thermal image file.

Meas.Date: Displays the measured day of a thermal image file.

Meas.Time: Displays the measured time of a thermal image file.

Comment: Displays the comment of a thermal image file
(It is indicated from the first to 80 characters)


Function
It is retained in the thermal image file, comment it is possible to input and to compile and.


Operation
(1) Item “comment” of the image information tub is clicked.

(2) When it clicks, because comment dialogue is indicated, comment is input.


When OK” button is pushed, information of the comment which is input is retained in
the thermal image file.

The frequency of input, in case of 1 byte word input is to 638 words.


Comment, in thumbnail indication and the detailed screen, is indicated from the first to 23
Note characters. In addition, in the preview screen it is indicated from the first to 79 characters.
When printing is printed from the first to 29 characters.

- 20 - No.95691-2464-0000
 Condition
The Condition tab displays the measurement condition information on the previewed thermal image
file.


Camera Model: Displays the used model (model of a thermo-tracer) at measurement.

Range: Displays the range (range of measurement temperature) at
measurement.

Image mode: Displays the setting information on the image mode or the average
at measurement.
(OFF: Sigma OFF, OFFx2: Sigma OFFx2, OFFx4: Sigma OFFx4,
S2: Sigma 2, S2x2: Sigma 2x2, S2x4: Sigma 2x4,
S8: Sigma 8, S8x2: Sigma 8x2, S8x4: Sigma 8x4,
S16: Sigma 16, S16x2: Sigma 16x2, S16x4: Sigma 16x4,
S32: Sigma 32, S32x2: Sigma 32x2, S32x4: Sigma 32x4,
S64: Sigma 64, S64x2: Sigma 64x2, S64x4: Sigma 64x4,


External lens: Displays the lens type when a lens other than the standard lens
was used.

Emissivity: Displays the Emissivity at measurement.

Amb Temp: Displays the environmental temperature value when performing
environmental reflective compensation in a thermo-tracer
at measurement.

POINT
The POINT tab displays the POINT information set at the preview display.


Number: Displays the number of the POINT. (POINT a to POINT t )

Temperature: Displays the temperature value indicated by the POINT.

Emissivity: Displays the emissivity set to the POINT.

Coordinates: Displays the coordinates of the POINT.

- 21 - No.95691-2464-0000
BOX
The BOX tab displays the BOX information set in the preview display.


Number: Displays the BOX number. (A to T)

Max: Displays the maximum temperature value in the BOX.

Min: Displays the minimum temperature value in the BOX.

Ave: Displays the average temperature value in the BOX.

Emissivity: Displays the emissivity set to the BOX.

Coordinates: Displays the coordinates of the upper-left and the lower-right corners
of the BOX.

Max coordinates: Displays the coordinates of the maximum temperature value
in the BOX.

Min coordinates: Displays the coordinates of the minimum temperature value
in the BOX.

- 22 - No.95691-2464-0000
Trend
The Trend tab displays a trend graph using the horizontal axis for pages and the vertical axis for
temperature while multiple thermal images (svt files are included) are played back.
The target is the temperature value of the POINT. Other than that, for the BOX, POLYGON, LINE
and CIRCLE, the maximum, minimum, and average temperature values are adopted.

[For the temperature value of the POINT]


Set the display temperature range of the
vertical axis. It is possible to use the spin
button or direct key entry. Cursor is indicated in the
position of the page, which
presently it is indicatory.
Cursor cannot be dragged
with a mouse.

Set the number of display


partitions of the trend
graph for the number of
partitions.

The object (object name), which indicates


the trend graph, is selected.
[Indicate list] Temperature value of
POINT: POINT cursor is indicated.
BOX: BOX (A,B,C…T)

[For the temperature value of Objects other than the POINT]

Maximum temperature value

Set the display starting


Average temperature value
position of the trend
graph for the starting
point.
Minimum temperature value

Set the display range


(pages) of the horizontal
axis for DIV.

- 23 - No.95691-2464-0000
8.1.1 Detail of Controller

(1) Level/Span setting

(3) Sensitivity setting

(2) Level setting


(5) Emissivity

(6) Color mode


(4) Multi Sense
(8) Color step

(9) Page (7) ISO setting


(10) Voice replay

(11) Visual/Fusion display switching


Function
(1) Level Span setting
Set the upper and lower values for the level for the main view image using the spin
buttons or through direct value input. The span width can be set within the range at
measuring (upper and lower values).

(2) Level setting


Set the level (the middle value) for the main view image using the spin buttons or
through direct value input.


Level span upper value: 10s step
Level span upper value: 1s step
Operation
Level span upper value: 0.1s step

Level value: 10s step


Level value: 1s step
Level value: 0.1s step
Click the spin
buttons or enter
a value directly.
Level span Lower value: 10s step
Level span Lower value: 1s step
Level span Lower value: 0.1s step

- 24 - No.95691-2464-0000

Function
(3) Sensitivity setting
Directly enter or use the spin buttons to input the sensitivity value for the main screen image.
When the level span is set, a check mark is disappears in the check box ( ). In this case,
the sensitivity value is automatically calculated. By adding a check mark in the check box
( ) on the left, a sensitivity value equivalent to that of the Thermo Tracer can be set.


Operation Click the spin
buttons or enter
a value directly.

When the level span is set, a check mark is disappears in the


check box ( ). In this case, the sensitivity value is automatically
calculated. By adding a check mark in the check box ( ) on the
left, a sensitivity value equivalent to that of the Thermo Tracer can
be set.


Function
(4) Multi Sense
Color separation ability (OFF, Middle, High/Low) is changed to a Main image view window.
A push on a "multi-sense" button displays a "multi-sense setting" dialog. The target color
separation ability is chosen and the "O.K." button is pushed.


Operation
When this button is clicked,
multiple sense setting dialogue
is indicated.

OFF: Used to release the multi-sense display


Middle: Used to view the details of the central
temperature portion
High/Low: Used to view the details of the hot
and cold portions

- 25 - No.95691-2464-0000

Function
(5) Emissivity
Display image, emissivity of the whole main revision is executed.
It can set emissivity value in the range of the 0.1~1.0.
In addition, as for environmental temperature, when measuring when executing “Reflection
calibration”, environmental temperature is indicated.
When measuring when “Reflection calibration” is not executed, “20.0 degC” is set.

Operation

By the fact that this spin button is


clicked, environmental
By the fact that this spin button temperature can be modified
is clicked, emissivity can
be modified


Function
(6) Color mode
Change the color mode for the main screen image using the pull-down menu
(Fine/Shine/Rainbow/Brightness/Hot iron/Medical/Monochrome positive/negative)


Operation Pressing this pull-down button
permits of color mode
selection from among the color
pallet menu.


Function
(7) ISO setting
ISO it sets to the main display image, (be able to become colored only the designated range,).
(8 pattern). Initializing is indicated, ISO number equivalent of 3 being set.


Operation
3. When 2 or more it is necessary, ISO
number is clicked.

4.It decides with “the OK”


button.

1. Click the ISO number. 5.Setting of the ISO number which


it appoints is eliminated.

6. The ISO setting which it sets


is cancelled.
2. With the slider bar (upper limit
and the lower limit) it can
become colored and appoints
the range.

- 26 - No.95691-2464-0000

Function
(8) Color step
Color gradation for the main screen image can be changed using the spin buttons (8, 16, 32,
64, 128, 256 colors).


Operation

You can change the


color gradation with


Function
this spin button.

(9) Page
If any thermal images or the svt files are replayed, image rendering of the previous, next, or
specified page is possible. In addition in case of plural playbacks, appointing playback interval,
it is possible to play back automatically.
* This item can be selected only when several images or the svt file are replayed.


Operation Click the spin
buttons or enter
a value directly.

Replay Stop 100ms, 200ms and 500ms, 1 seconds,


2 seconds, 5 seconds, 10 seconds,
20 seconds, 30 seconds, 60 seconds,
it can select automatic playback interval,
from midst of the highest speed.

Such as when emissivity revision refresh rate is not in time and graphical representation and
when load of the personal computer is large, because (scan disk etc), there are times when,
Note please note.

 (10) Voice replay


Function If a thermal image file has voice comments, voice replay, temporary halt, stop, and mute is
available.
* Without voice comments, this function does not operate.


Operation Replay button
Pause button

Slider bar
Mute button
(Voice ON/OFF) Stop button

- 27 - No.95691-2464-0000

Function
(11) Switching between thermal, visible, and fusion display
You can switch between a thermal image, visible image, and fusion display.
* When no visible image is attached, any visible image is not displayed.


Operation To display a visible image, press the "Visible" button. To combine and display a thermal image
and a visible image, press the "Fusion" button.

To switch between a Thermal image,


a Visible image, and Fusion display, use
these buttons by setting them ON or OFF.

When a thermal image is displayed When a visible image is displayed

When fusion image is displayed

[Setting fusion display]


1. When you press the Fusion Display button, the following dialog is displayed.

When it turned with rotary mode, angle is


indicated. The indicatory range becomes
Expansion and contraction -5° from +5°.
mode/Rotary mode is changed. With the spin button, it can modify with
By the fact that it clicks, it becomes 1° step.
rotary mode.
When you reset to expansion and
contraction mode, it clicks for
the second time.

You can adjust the color of


the thermal image using
the slide bar within the You can change the
range of level and span transparency of the thermal
image using the slide bar.
setting.
(Only at adjustment)

The application temperature


range of the slide bar is
indicated.

- 28 - No.95691-2464-0000
[Expansion and contraction mode]
2. To move a thermal image displayed on the visible image, use the mouse. Alter the image in the
vertical, horizontal, and skew directions by using the mouse to set the image in the optimum position.
If you use the "Transparency" slide in the dialog to adjust the transparency of the thermal image at
setting, the setting becomes easier.

3. If you display an infrared image of only the specific temperature range in the level and span setting
range, use the slide bar to adjust the upper and lower limits in the dialog.

Only for the specific


temperature range, a
thermal image can be
displayed.

4. When setting is completed, click a part of the visible image other than the thermal image.
The combination adjust mode (the dialog is closed) is canceled.

Because when changing the setting and the cancellation of fusion setting mode,
Note image position is formalized, there are times when the infrared image slips a little.

- 29 - No.95691-2464-0000
[Rotary mode]
2.When it modifies in rotary mode, moves mouse cursor to the four corners of the infrared image, the
mouse cursor idea contest it is indicated as the rotary idea contest.
In addition, with the spin button of synthetic setting dialogue, at a time 1° step angle can be
modified.
With the spin button, it can modify with 1° step.

3.Being the place where it becomes the rotary icon, when you push the spin button in angle of rotation,
it is possible to turn the center of the infrared image as a fulcrum.

Turning angle is
up to +5°.

Center

Turning angle is
up to -5°.

- 30 - No.95691-2464-0000
8.1.2 Saving thermal image data
1. Use drag & drop to copy thermal image data to the specified folder.

A) Click the thermal image to be saved on the image preview window to select.
(It is also possible to select multiple images by using the Ctrl key or the Shift key at the
click.)

B) Confirm that the destination folder of the thermal image data is displayed in the folder
selection window. If not displayed, click + on the folder selection window to display
the folder.

C) When you drag the selected thermal image by the mouse cursor, the file is moved
(copied).

Move the file by drag


& drop.

Selected thermal image

If you click a folder and the like (other than + ) on the folder selection window,
when you try to display the destination folder, the image preview window is
Note changed to the selected folder.

2. Save thermal image data in the specified folder of a personal computer by using
the right-click menu.

A) Right-click the image to be saved on the image preview window. When the right-click
menu is displayed, select "Save".
Menu when you right-click on
the image preview window.

- 31 - No.95691-2464-0000
B) When the "Reference Folders" window is displayed, select the saved location and click
the [OK] button. To cancel the processing, click the [Cancel] button.

Select the saving


destination folder.

Cancel button

OK button
Button to make a new folder

C) If duplicated files exist at saving, the following confirmation window is displayed.


To overwrite the data, click the [Yes] button. To cancel the processing, click [No].

No button
Yes button

- 32 - No.95691-2464-0000
8.1.3 POINT
The POINT is the function to display the temperature value of the POINT on the image for the displayed
thermal image.
Add a POINT to the location where the temperature value is to be displayed. You can set the added
POINT. You can set the font color and the size of the temperature value, the emissivity of the POINT,
and the coordinates. The methods of setting are:
#The method the tool bar ( ) of pushing the icon, clicking on the thermal image.
#Select [Add Object]-[POINT] of the right click menu.
#Select [Object]- [Add Object]-[POINT] of the menu bar. And at the designated point setting window it
sets.

8.1.3.1 Adding a POINT


Function
You can add a POINT to the location where the temperature value is to be displayed. Up to
20 points can be added. Additional method is as follows:
#A right-click menu is opened in a place (point) to add on a thermal image,
and it is "Add Object" - "POINT" is selected.
#Push the POINT icon( ) of the tool bar, click in the optional place.
#"Object"- "Add Object"- "POINT" of a menu bar is selected, and it sets up on a POINT
setting window.


Operation To add a POINT to a location, open the right-click menu at the location where you want to add
a point on the thermal image and select "Add Object" - "POINT", or click the POINT icon
( ) of the toolbar and click on the thermal image. The alphabetic letters a to t are
assigned to the added POINTs.

A POINT is added, and the point name


(the alphabetic letter subsequent the
existing letter is assigned) and the
temperature value are displayed.
Example) If "Add object" - "POINT" is
executed when a to d have been set, the
POINT e is added.

Right-click somewhere other


than POINTS, and select
"Add Object" - "POINT".

- 33 - No.95691-2464-0000
8.1.3.2 Moving a POINT


Function
You can move the coordinates of the displayed POINT to different coordinates.
(The displayed temperature value is changed to the value of the destination coordinates.)


Operation Point the mouse cursor to the POINT to be moved, and move the POINT by drag & drop. (To
check the coordinates, select and click the POINT to display the coordinates of the POINT as
a pop hint.)

To display the coordinates of the POINT as


a pop hint, select and click the POINT.

Drag & drop

You can also set the coordinates by setting coordinates in the POINT setting
See
window. For the operation method, see "8.1.3.3 Setting a POINT".

- 34 - No.95691-2464-0000
8.1.3.3 Setting a POINT


Function
You can set a POINT. You can set the following six setting items; "emissivity setting",
"display of emissivity", "X coordinate", "Y coordinate", "display style (size and color)", and
"deletion".


Operation
1. When mouse cursor is adjusted to the POINT of the setting object, the right click menu is
opened on the POINT and “Set Object” is selected, “Point setting" window starts.
In addition, because in when the object other than the POINT is not selected, the “Point
selective” picture is indicated by the fact that the button( ) on the tool bar is pushed, it
selects the object, it is possible also to start “Point setting" window.

Right-click on the POINT


The “Object setting” button on
and select "Set Object".
the tool bar is clicked, the object
of POINT is selected.

[At the time of the right click menu selection] [At the time of object setting button selection]

2. When the "Point setting" window is started, set each of items. To update the settings, click the
[OK] button. To cancel the setting processing, click the [Cancel] button. If you click the
[Delete] button, the POINT is deleted.
When you click the [Style] button, the change setting window is started.

The name of the POINT is displayed. Emissivity setting


Click the spin button or enter a value in the
text box directly.
Display Emissivity (The range is between 0.10 and 1.00. You
setting can set a value in increments of 0.01.)
Set whether
the emissivity X coordinate setting
is displayed. Click the spin button, or enter a value in the
text box directly.
(Emissivity indication is
reflected on all the Y coordinate setting
POINT.) Click the spin button, or enter a value in the
text box directly.

OK button
Style button
Cancel button

The "Font Setting"


Delete button
window is started.

- 35 - No.95691-2464-0000
Change setting window

You can set a font color and a display size in the "Font Setting" window.
To update the settings, click the [OK] button. To cancel the update, click the [Cancel] button.

The currently selected font color is displayed.

To select a font size,


click the pull-down
button.

Cancel button
OK button
Select the font color to be set from
the color palette and click it.

- 36 - No.95691-2464-0000
8.1.3.4 Deleting a POINT


Function
You can delete a POINT.


Operation To delete a POINT, point the mouse cursor to the POINT to be deleted, open the right-click
menu on the POINT, and select [Delete Object]. You can also delete a POINT from the Point
Setting dialog. In addition, [Delete] of the Point setting window you push the button, or you
can delete even by the fact that the Cut button ( ) on the tool bar is pushed.

Right-click on the POINT, In the state where the object is not


and select "Delete Object". selected a push on a ( ) button
displays this dialog.

Operation
To delete all the set POINTs, right-click on somewhere other than a POINT, and select "Delete
Object " - "All Points". Moreover, all can be deleted if "Object"-"Delete Object"- "All Points" is
selected.

[At the time of the right click menu selection] [At the time of object setting button selection]

- 37 - No.95691-2464-0000
8.1.4 BOX
The BOX function displays the maximum temperature, the minimum temperature, and the average
temperature of the area (rectangle) on the image for the displayed thermal image. Add a BOX to the
location where you want to display the temperature value. You can set and edit the added BOX and set
a font color and a size of the display object and the emissivity of the BOX.
The methods of setting are:
#The method the tool bar ( ) of pushing the icon, clicking on the thermal image.
#Select [Add Object]-[BOX] of the right click menu.
#Select [Object]- [Add Object]-[BOX] of the menu bar. And at the designated BOX setting window it
sets.

8.1.4.1 Adding a BOX


Function
You can add a BOX to the location where you want to display the temperature value. Up to
20 boxes can be added.


Operation
At each method below, addition of the BOX is possible.
The BOX which is added is allotted to the name to of `A' ~ `T' of alphabet.
Additional method is as follows:

#A right-click menu is opened in a place (point) to add on a thermal image,


and it is "Add Object" - "BOX" is selected.
#Push the designated BOX icon( ) of the tool bar, click in the optional place.
#"Object"- "Add Object"- "BOX" of a menu bar is selected, and it sets up on a BOX setting
window.

Right-click on somewhere other


than a Object, and select
"Add Object" - "BOX".

- 38 - No.95691-2464-0000
8.1.4.2 Moving a BOX


Function You can move the coordinates of the displayed BOX to different coordinates.
(The displayed temperature value is changed to the value of the destination coordinates.)


Operation
Point the mouse cursor to the BOX to be moved and move the BOX name (A to T) by drag &
drop.

1. Clicking a BOX displays the


selected frame.

2. You can move a thermal


image by drag & drop.

You can also move a BOX by setting coordinates in the BOX setting window.
See
For the operation method, see "8.1.4.3 Setting a BOX".

- 39 - No.95691-2464-0000
8.1.4.3 Setting a BOX

Function
You can set a BOX. You can set the following five setting items; "emissivity setting",
"upper-left X coordinate", "upper-left Y coordinate", "Height", "Width", "Style (size and color)",
and "deletion".


Operation
1.When mouse cursor is adjusted to the BOX of the setting object, the right click menu is
opened on the BOX and “Set Object” is selected, “BOX setting" window starts.
Moreover, when the BOX object is selected, BOX setting window can also be started by pushing
the button( ) on a tool bar.

Right-click on a BOX and To alter a Object, click the


select "Set Object". BOX and drag four
corners and the center of
each line.

(2) When the "BOX setting" window is started, set items. To update the settings, click the [OK]
button. To cancel the setting,, click the [Cancel] button. To delete a BOX, click the [Delete]
button.
When you click the [Style] button, the change setting window is started. The setting
method of changing the style is the same as that of a POINT.
X coordinate of the upper
left corner of the box Y coordinate of the upper-left corner of the box
Click the spin button or Click the spin button or enter a value in the text box
enter a value in the directly.
text box directly.
Box height
Set the box length in the vertical direction.
Update of the
emissivity image
Mark this to update Box width
the emissivity correction Set the box length in the horizontal direction.
results to the image
display.
(Emissivity indication is Emissivity setting
reflected on all the BOX.) Set the emissivity.

OK button Style button


Cancel button
The "Font Setting"
Delete button window is started.

- 40 - No.95691-2464-0000
8.1.4.4 Deleting a BOX


Function You can delete a BOX.


Operation
To delete a BOX, point the mouse cursor to the BOX to be deleted, open the right-click menu
on the BOX, and select "Delete Object". You can also delete a BOX from the BOX Setting
dialog. In addition, [Delete] of the BOX setting window you push the button, or you can delete
even by the fact that the Cut button( ) on the tool bar is pushed.

Right-click on the box and


select "Delete Object".


Operation
To delete all the set boxes, right-click on somewhere other than boxes and select "Delete
Object" - "All Boxes". Moreover, all can be deleted if "Object"-"Delete Object"- "All Boxes" is
selected.

[At the time of the right click menu selection] [At the time of object setting button selection]

- 41 - No.95691-2464-0000
8.1.5 Delete all objects


Function
You can delete all the objects currently set.


Operation To delete all objects, open the right-click menu on a thermal image other than the set objects,
and select "Delete Object"-“All Objects". Moreover, all can be deleted if "Object"-"Delete
Object"- "All Objects" is selected.
Right-click on a thermal image other
than the set Objects, and select
"Delete Object".- "All Objects".

[At the time of the right click menu selection] [At the time of object setting button selection]

- 42 - No.95691-2464-0000
8.6 Create Graph


Function
You can create a graph for the target image or the target Object.
Three types of graphs; Trend, can be created.

8.6.3 Trend


Function You can create a trend graph by plotting temperature values of a set Object in the time series
while multiple thermal images (sit or tmp) are played back or an svt file is played back.
In this trend graph, the horizontal axis indicates pages and the vertical axis indicates
temperature values.
Trend graphs are intended for all Objects, including POINT and BOX.
For a BOX other than a POINT, the maximum temperature value, the minimum temperature
value, and the average temperature value are displayed in a trend graph.
To create a trend graph, select "Trend" from the right-click menu on a thermal image, or select
"Graph" - "Trend" from the menu bar.


Operation

1.While a graph is selected (on a POINT, if a POINT is the target), open the right-click menu and
select "Trend", or select "Graph" - "Trend" from the menu bar.

Right-click on the selected Object, and


select "Trend".

2.Creating a trend graph is started. During creation, the following dialog is displayed. To cancel
processing during creation, press the "Cancel" button.

To cancel creation of a trend graph


during processing, press the
"Cancel" button.

- 43 - No.95691-2464-0000
3.A profile is created in the information tab.
[For the POINT temperature value]
For a POINT, the trend of all set POINTs is displayed.

It indicates by cursor, using as a standard


the position applicable to the page
displayed currently.
(Trend cursor cannot be dragged)

The object (object name), which


displays trend graph, is selected.
[List of data]
POINT: POINT
BOX : BOX(A,B,C…T) Cursor Temperature

[For the temperature value of Objects other than POINT]


For Objects (BOX) other than POINT, the maximum temperature value, the minimum temperature
value, and the average temperature value of the selected Object are displayed.

Maximum temperature value

Average temperature value

Minimum temperature value

Other than POINT object,


Maximum (max) and
Minimum (min) and
Average (ave) is indicated.

*1
The graph it outputs which temperature value, at configuration it can set.
Reference
As for the trend graph, as for automatic operation renewal it is not done by the modification of
movement and the like in the setting object. The case where the data is renewed, please
execute the compilation in the trend graph for the second time.

The trend graph is drawn up from page of the image, which presently is indicated.
Because of this, presently it is indicated in before the trend graph drawing up,
Note
verification of page number is recommended. (Data of 1st data = present page of trend)

When the trend graph is drawn up in the plural objects, it is necessary beforehand to
select the plural objects.
As for plural object selective methods, please refer to the chapter of 8.6.2 Line Profile.

- 44 - No.95691-2464-0000
9.1 Menu Operations
This section explains menu bar functions and operations of the Report Generator.
For menu structure, see [7.1 Menu Structure].

9.1.1 Thermal Image Menu

[Open]


Function This command opens a thermal image file from the menu bar (opens one image).


Operation
(1) From the menu bar, select [Thermal Image], then [Open].

(2) The [Open] screen appears. Select an image and click the [Open] button to display th
e image in the Main Display Window. Clicking the [Cancel] button closes the
[Open] screen.

Thermal Image Selection Screen Change View Format


Create New Folder ・ Large Icon
Move Up One Folder Level ・ Small Icon
・ List
Move to Last Folder Displayed ・ Details
・ Thumbnail
Search File Location

Close
Window

Preview of
Selected Image

Select file format.


(jpg,sit,six, tmp,svt, or svx)
File Selection Window
Toggle infrared and visible
image previews ON/OFF.

- 45 - No.95691-2464-0000
[Open Files]


Function This command opens thermal image files from the menu bar (opens several).


Operation
(1) From the menu bar, select [Thermal Image], then [Open Files]. The [Open Files] screen appears.
(2) From [File location], select the folder and directory where a thermal image is saved.
(3) To limit display of thermal image files by file type, select [File of type].
(4) From the file list, select several thermal images and click the [Add] button to add them to the
page list. To delete an entry from the page list, select any files from the page list and click the
[Delete] button.
(5) Click the [Open (O)] button to display the file in the Main Display Window. At this time, the
thermal image opens as a page function. Click the [Cancel] button to close the [Open Files]
screen.

Thermal Image Selection Screen

File List
Click the file you want to display while holding down the
Ctrl Key and Shift key. If selected, the background color
changes to blue. Search File Location

Select File Type jpg,sit,six, or tmp

Add/Delete Buttons

Register/Read List
You can register several files as a page list (.sfl).
-While file(s) are displayed in the page list, click the [Register
Page List
List] button. (See 1. Register List for operations thereafter)
You can import page lists that have been registered before.
-Click the [Import List] button.
Click the file you want to display while holding down the Ctrl
Key and Shift key. If selected, the background color changes
to blue. (See 2. File List Load for operations thereafter)

- 46 - No.95691-2464-0000
Reference As shown in the following figure, if two or more picture reproduction of
the thermal image file is carried out, a picture preview window will be displayed.

Ex) In case of the thermal image file of 1st page


1 Page TH780006.SIT
The name of thermal image

Page number

Reference
When opening two or more thermal image files, it can reproduce also by
the following methods.

1. Two or more thermal image files currently displayed on the image preview window are
selected in a mouse cursor.

Select two or more files.

2. Next, if "Thermal Image" - "Open Files" is selected from a menu bar or


the button( ) on a tool bar is pushed, the "Open Files" dialog will be displayed, a
nd the thermal image files selected as the Page list will be added.

The selected file is


automatically added to a page list.

3.If the button "Open (O) " is clicked, it will be displayed on a main display window.

- 47 - No.95691-2464-0000
1. Register List
(1) From the [Open Files] screen, ensure file(s) are selected in the page list, and then click the
[Register List] button.

(2) The [Register List] screen appears. Enter a file name and click the [Save] button t
o save the file in the [Save In] location in .sfl format. Clicking the [Cancel] button c
ancels registration.

Create New Folder


Change View Format
>Large Icon
Move Up One Folder Level >Small Icon
Move to Last Folder Displayed >List
>Details
Search File Location
>Thumbnail

File Display
Window

Save button

Cancel button
Enter a file name.

- 48 - No.95691-2464-0000
2. File List Load
(1) From the [Thermal Image Selection] screen, click the [File List Load] button, and the
[File List Load] screen appears.

(2) The [File List Load] screen appears. Select a .sfl file and click the [Open (O)]
button to import the file whose location is displayed in [Look in]. Clicking the
[Cancel] button closes the [File List Load] screen.

Create New File Change View Format


>Large Icon
Move Up One Folder Level >Small Icon
>List
Move to Last Folder Displayed
>Details
>Thumbnail
Search File Location

File Display
Window

Open button

Select a file from the Cancel button


File Display Window

- 49 - No.95691-2464-0000
[Save]


Function Overwrites and saves a thermal image data that is being displayed/edited.


Operation
(1) From the menu bar, select [Thermal Image], then [Save].

(2) The Confirmation screen appears. Click the [Yes] button to overwrite and save. Click the [No]
button to cancel.

No button
Yes button

- 50 - No.95691-2464-0000
[Save As]


Function This command saves a thermal image data that is being displayed/edited as a new file.


Operation
(1) From the menu bar, select [Thermal Image], then [Save As].

(2) The [Save As] screen appears. Click on the [Save] button to crete a new thermal image file
with the designated name. Click the [Cansel] button to cancel.
Change View Format
>Large Icon
Create New File >Small Icon
Move Up One Folder Level >List
>Details
Move to Last Folder Displayed >Thumbnail

Search File Location

Close
Window

Insert a file name.

Save button
Cancel button

Reference If a file name you entered (selected) is already in use, the [Save] screen
appears. Click the [Yes] button from the Confirmation screen to overwrite
and save.

- 51 - No.95691-2464-0000
[Print]


Function
The thermal image which the preview it is indicatory is printed.


Operation
A) From the menu bar, select [Thermal Image], then [Print].

B) The printing verification dialogue which is set with setting of the printer is indicated.
[OK] or when [Cancel] and so on pushes the printing execution button, printing starts.

Reference Because printing verification dialogue differs depending upon the setting of the
personal computer of use, please note.

- 52 - No.95691-2464-0000
[Print setting]


Function
It sets the printer.


Operation
A) From the menu bar, select [Thermal Image], then [Print setting].

B) Setting dialogue of the printer is indicated. It sets the printer.

Reference Because setting dialogue of the printer differs depending upon the setting of the
personal computer of use, please note.

- 53 - No.95691-2464-0000
[Exit]


Function This function exits the Report Generator.


Operation From the menu bar, select [Thermal Image], then [Exit] to exit the Report Generator.

- 54 - No.95691-2464-0000
9.1.2 Edit Menu
[cut]


Function The image file which is selected in thumbnail display or details display is cut off.
In addition, when the object which sets is selected, the object is cut off.


Operation
[In case of file cut off]
A) In thumbnail display or details display, the image file which we would like to cut off is
selected.
B) From the menu bar, select [Edit], then [cut].

C) Cut the selected image file.

[In case of object cut off]


Please refer to the section (deletion of 8.1.3.4 Deleting a POINT, deletion of
8.1.4.4 Deleting a BOX, deletion of 8.1.5.4 Deleting a LINE, deletion of 8.1.6.4 Deleting
a POLYGON and deletion of 8.1.7.4 Deleting a CIRCLE) of deletion in each object.

Even when two or more objects are selected, the number of the objects deleted is one.
Note In that case, a priority serves as a BOX > LINE > POLYGON > CIRCLE. .
Moreover, priority is given to the object of a large number in the same object.

[copy]
- 55 - No.95691-2464-0000

Function The image or Image info tab which is indicated in the main image view window is
copied in the clipboard. In addition, when the object which sets is selected,
the isomorphic object is copied.


Operation
1. An object is selected and copied.
[In case of main image copy]

A) The free place on the main image view window is clicked.


B) From the menu bar, select [Edit], then [copy] or
click the button( ) on the tool bar.

[In case of image info tab copy]

A) Click the Image info tab.

Click here

B) From the menu bar, select [Edit], then [copy] or click the button( ) on the tool bar.
The right it clicks inside the Image info tab, selects the menu “copy”.

C) If applications, such as MS-Excel(MS-Word), are started and it performs [Paste],


a image will paste.

- 56 - No.95691-2464-0000
Even when two or more objects are selected, the number of the objects copied is one.
Note In that case, a priority serves as a BOX > LINE > POLYGON > CIRCLE. .
Moreover, priority is given to the object of a large number in the same object.

[paste]


Function At thumbnail display or details display, it copies the file, it is possible to paste to the
specified folder. In addition, when the object was copied, the same object pastes on t
he main image window.


Operation

A) From the menu bar, select [Edit], then [copy]. Or a object is selected and copied.
B) From the menu bar, select [Edit], then [paste].
In case of the object copy, “Edit”- “Paste”of the menu bar selects.

C) In the specified folder, the file which was copied is duplicated.


Or the same object pastes on the main image window.

- 57 - No.95691-2464-0000
[Image copy]

 The image which is indicated in the main image window (1 image), “thermal only”
Function “thermal with color bar” or “thermal with measurement condition” with copies in the
clipboard.


Operation
1. From the menu bar “Edit”-“Image copy”-“Copy thermal image”
Or “Save thermal image with color bar” and selects which.

2.If applications, such as MS-Excel(MS-Word), are started and it performs [Paste],


a image will paste.

- 58 - No.95691-2464-0000
[Image Save]


Function This command saves an image displayed in the main image window in the Bitmap or a JPEG
format(Compression ratio is fixed). Choose one of two menus to save images: [thermal
image], [thermal image with color bar]

1. Save thermal image


Function Save a displayed image in the Bitmap or JPEG format.


Operation
(1) Choose [Edit] from the menu bar, and then click [Save image file] and select [Save thermal
image].

(2) [Save thermal image] screen appears. Indicate the location, filename and file format (Bitmap or
JPEG), and then click the [Save] button. The image is saved in the specified location. Click the
[Cancel] button to stop processing

Save button
Cancel button
Select Bitmap or JPEG from
the pull-down menu.

- 59 - No.95691-2464-0000
2. Thermal image with color bar


Function
Save a displayed image with color bar in the Bitmap or JPEG format.


Operation
(1) Choose [Edit] from the menu bar, and then click [Save image file] and select [Save thermal
image with color bar].

(2) [Save thermal image with color bar] screen appears. Specify the location, filename, and file
format (Bitmap or JPEG), and then click the [Save] button. The image is saved in the specified
location. Click the [Cancel] button to stop processing

Save button

Cancel button
Select Bitmap or JPEG from
the pull-down menu

- 60 - No.95691-2464-0000
3. Thermal image with measurement condition


Function
Save a displayed image with color bar and measurement conditionin the Bitmap or JPEG
format.


Operation
(1) Choose [Edit] from the menu bar, and then click [Save image file] and select [Save thermal
image with measurement condition].

(2) [Save thermal image with measurement condition] screen appears. Specify the location,
filename, and file format (Bitmap or JPEG), and then click the [Save] button. The image is
saved in the specified location. Click the [Cancel] button to stop processing

Save button

Select Bitmap or JPEG from Cancel button


the pull-down menu

Example

- 61 - No.95691-2464-0000
[AVI Save]


Function This command saves an image displayed in the main image window as an AVI format.
Choose one of two menus to save AVI: [Save thermal image as AVI] and [Save screen as
AVI] (Enable for replaying of two or more thermal image files)

1. Save thermal image as AVI


Function
Save a thermal image only in the AVI format.


Operation
(1) Choose [Edit] from the menu bar, and then click [Save AVI file] and select [Save thermal
image as AVI].

(2) The [AVI transform] screen appears. Specify the first page, the last page and a frame rate, and
then click the [OK] button. The AVI is saved in the specified location. Click the [Cancel] button
to stop processing.

Page Setting
Enter the page range
directly or by using the
spin buttons. OK button

Cancel button

For 60 - 1fps

For 2 - 3600spf

60 - 1fps
Available from the pulldown menu:
2 - 3600spf 60fps, 30fps, 15fps, 5fps,1fps
Select a value among 2 - 3600spf by the spin button

- 62 - No.95691-2464-0000
(3) The [Save As] screen appears. Indicate the filename and click the [Save] button. Click the
[Cancel] button to stop processing

AVI file only

Save button

Cancel button

(4) When you click the [Save] button, the [Comment] box appears and the video is saved as an
AVI format. Click the [Cancel] button to stop saving at the page.

Cancel button

2. Save thermal image with color bar as AVI


Function Save a displayed image with color bar in the AVI format.


Operation
(1) Choose [Edit] from the menu bar, then click [Save AVI file] and select [Save screen as AVI].

(2) Perform the same steps from (2) to (4) of “1. Save thermal image as AVI”, hereafter.

Only the thermal images with the same type format can be saved as AVI.
Note Example: sit file 10 images : OK
tmp file 10 images; OK
svt file: OK
sit file 5 images + tmp file 5 images; NG
BMP (or JPG) file 10 images; NG

- 63 - No.95691-2464-0000
During AVI converting memory capacity of the personal computer decreases,
Note when continuation becomes impossible, message below is indicated.

Memory capacity is insufficient.


Please end other application programs
and vacate memory capacity

3. Save thermal image with measurement condition as AVI


Function Save a displayed image with color bar and measurement condition in the AVI format.


Operation
(1) Choose [Edit] from the menu bar, then click [Save AVI file] and select [Save thermal image with
measurement condition as AVI].

(2) Perform the same steps from (2) to (4) of “1. Save thermal image as AVI”, hereafter.

Example

- 64 - No.95691-2464-0000
[Generate Report]
Before outputting a report, check the security setting of MS-Word (Excel).

Note [Notes on security]


While you use this product, either of the following messages might be
reported from Microsoft Word. That is to check whether a program built into
Microsoft Word (Excel) is a secured program.
These messages are generated 1. when a report is output 2. when a template
layout is modified 3. when a template is added 4. when a file is opened after
creation, and in other cases.
You can delete these messages by selecting "Tool" - "Macro" - "Security" from
Word and setting the security level of "Security" to [Low]. But deletion
makes the program susceptible to viruses. Therefore, deletion is not
recommended.

Example 1) Message displayed when the security level is set to "High" or "Very High"

Example 2) Message generated by Microsoft Word (Displayed when the security level is "Middle".)

When this message is displayed, click "Enables Macro (E)".

- 65 - No.95691-2464-0000
See Checking and changing the security level of a macro
<<MS-Word>>
A) Start MS-Word by selecting [Start] - [Programs].
B) Select [Tools (T)] - [Macros (M)] - [Security (S)] from the menu bar as shown below.

C) After selection, when the following dialog is displayed, mark a radio button other than "Very
High" or "High" and click the "OK" button. At this time, if you select "Low", a warning message is
not displayed at execution of the program as described in the window below. Note that, however,
a malicious program is also executed without any warning message.

- 66 - No.95691-2464-0000
【For use of this software with Office2007】
In case for using this software with Office2007, the following settings are required.

1. Setting of Office2007 (MS-Word, MS-Excel)


In case for outputting data to Word2007( or Excel2007) by “ Generate Report “ function, the
security setting of “ ActiveX Settings ” and “ Macro Settings ” needs to be changed properly.

1-1 The previous setting of execution (the setting of ActiveX)


A) The first to do is right click on the ribbon and display menu. Then “Customize Quick
Access Toolbar...” is selected.

Fig 1. “Customize Quick Access Toolbar...” is selected in Word2007.

B) After “Word Options” dialogue is displayed, “Trust Center” is selected and then
“Trust Center Settings...” button is pressed

Fig 2. “Word Options” dialogue

- 67 - No.95691-2464-0000
C) After “ Trust Center” dialogue is displayed,
“Enable all controls without...” is selected in “ActiveX Settings”

Fig 3. ”Trust Center”-”The setting of ActiveX”

1-2. The previous setting of outputting report(Security setting in Macro)


”Enable all macros” is configured in Macro setting.
A) After “Trust Center” is displayed,
“Enable all macros...” is selected in “Macro Settings”

Fig 4. ”Trust Center”-”Macro Settings”

- 68 - No.95691-2464-0000
B) This is all to do for setting. Now the output report function is available for Word2007and
NS9200LT Report Generator Lite Software.

Fig 5. output report image

- 69 - No.95691-2464-0000
[Generate Report]


Function To create an analysis result report about thermal images, you can output analysis results to a
report template and create a report file.
The created report file is saved in the database folder set by selecting "Options" -
"Environment” from the menu bar.


Operation A) Select "Edit" - "Generate Report" from the menu bar.

B) The Report Wizard is displayed. First of all click the template of the output destination on
the "Installed templates" list.
C) Then, specify the thermal image to be output.
Leave only the images to be output in the list by using the Add or Delete button and the
"Ctrl" key or the "Shift" key together.
For svt files, you can specify the Start page and the End page.
1) Specify a template.
* For changing and adding a template
layout, see [Add] button and
[Change layout] button.

3) Specify the thermal image to be


output.
(If you mark "Thumbnail display",
Preview of a template you can check images.)

4)For SVT files, you can specify the


Start page and the End page. 5) Specify the method of
outputting the report.

- 70 - No.95691-2464-0000
D) Select the image display format from a) Infrared image only, b) Infrared image + visible
image, or c) Fusion image.
E) If necessary, enter a title, company name, creator, and comment, and press the "Send"
button.
F) When the send button is pressed, data is output to the specified template.
The template creates a report file by using the sent data and saves the file.

7) Press the "Send" button.

6) Specify the display format of an output image.


a) Infrared image only
b) Infrared image + visible image
c) Fusion image

* If you select an infrared image only or a


combined image, a visible image is not
output.

* When no window to display each image exists


in the target template, the image to be output
is not displayed.
It is recommended that you should check the
template layout before output (sending).

G) The created report file is saved in the same location as the database folder
set by selecting [Options] – [Environment ].
If multiple images are output, multiple report files are saved
in the database folder.

Example) Database folder output path: C:¥report-output


The report files are saved in "C:¥report-output¥200607201228¥Report".

(Creation YearMonthDayHourMinutes)

- 71 - No.95691-2464-0000
[Changing a template layout]
Change the layout of the template to be output.


Operation
(1) Specify the template of which layout is to be changed from "Installed templates" in the Report
Wizard.

(2) Press the "Change Layout" button.

(1) The specified template is started in the template file format.

Image display window

- 72 - No.95691-2464-0000
(4) Use the toolbar for creating a template, located at the button of the started template file, to
move and add a image display window, and to add image data and Object data.
H) Button to allow you to move a display window

G) Button to save template formats


F) Adding data codes

E) Adding trend graph display windows


D) Adding histogram display windows
C) Adding line profile display windows

B) Adding visible image display windows

A) Adding thermal image display windows

Reference When "Add-in" menu is selected, the template creation toolbar


is displayed for Word2007.

- 73 - No.95691-2464-0000
[Explanation about adding data codes]
Set the image data and Object data output from a report creation program to a template additionally.
Set data in the order of "1. Image" to "4. Parameter" as shown in the following dialog.
After completion of setting, press the "Insert" button to insert the setting codes.

0. Report Info 1. Image


Set Title, Company, Creator, Comment. Select the target images of the thermal
The characters entered image display window. Up to 10 thermal
in the Report Wizard are displayed. images can be set.

2. Object
Set which type of data is output.
Information; File information on thermal images
POINT; Information on specified points
BOX; BOX information

3. Target
1. If you set a Object in 2. Object, set a Object name.
(a,b… or A,B…)
2. If you set Information in 2. Object, set the following list.
FileName; File name
MEMO; Memo information saved by the memo function
DATE; Measured date
Cancel button TYPE; Used thermo-tracer type
Close the Upper Range; Upper limit of the measured range
Data_Code dialog. Lower Range; Lower limit of the measured range
Emissivity; Emissivity at measurement
MaxTemp;The maximum temperature value in the image
Insert button MinTemp; The minimum temperature value in the image
When you press the Insert button, the AveTemp; The average temperature value in the image
setting codes selected in 0. to 3.
are inserted to the template.

- 74 - No.95691-2464-0000
Reference When Information is selected, [3.Target] after [Insert]
the selecting pushing the button, it is completion.
(4.There is no selection of Parameter)

There is no selection of Parameter.

[When the data of the graphics file is selected]

4. Parameter
2. Object, 3.The data that the object,
which is selected with Target, makes
output is appointed.

[When the object data is selected]

Ex) When is set to the thermal image (No.1), sets the temperature value of POINT
name b, as description below.

Thermal image No.1

POINT data

POINT b

Temperature

- 75 - No.95691-2464-0000
Reference
The output enable data in each object becomes as follows.

[Select POINT object]


1. Temp: POINT Temperature
2. AddressX: X coordinate of POINT
3. AddressY: Y coordinate of POINT
4. Emissivity: Emissivity of POINT

[Select BOX object]


1. MaxTemp: Maximum temperature inside BOX
2. MinTemp: Minimum temperature inside BOX
3. AveTemp: Average temperature inside BOX
4. Emissivity: Emissivity of BOX
5. UpperLeftX: X coordinate with respect to the left of BOX
6. UpperLeftY: Y coordinate with respect to the left of BOX
7. LowerRightX: X coordinate under the right of BOX
8. LowerRightY: Y coordinate under the right of BOX
9. Area: Area inside BOX

[Select LINE object]


1. MaxTemp: Maximum temperature on LINE
2. MinTemp: Minimum temperature on LINE
3. AveTemp: Average temperature on LINE
4. Emissivity: Emissivity of LINE
5. Length: Length of LINE

[Select POLYGON object]


1. MaxTemp: Maximum temperature inside POLYGON
2. MinTemp: Minimum temperature inside POLYGON
3. AveTemp: Average temperature inside POLYGON
4. Emissivity: Emissivity of POLYGON
5. Area: Area inside POLYGON

[Select CIRCLE object]


1. MaxTemp: Maximum temperature inside CIRCLE
2. MinTemp: Minimum temperature inside CIRCLE
3. AveTemp: Average temperature inside CIRCLE
4. Emissivity: Emissivity of CIRCLE
5. CenterX: Central X coordinate of circle
6. CenterY: Central Y coordinate of circle
7. Area: Area inside CIRCLE

- 76 - No.95691-2464-0000
[Explanation about the button to save template formats]
Save the created and edited template with a new name.


Operation
(1) When you press the [SaveAs] button on the template creation toolbar, the "Save as Template"
dialog is displayed.

(2) Enter a file name and press the "Save" button.

Enter a file name.

To save the template,


click the Save button.

*.dot

See In the initial setting, the saving destination of the "Save as Template" dialog is the
"Template" folder to which the report creation program is installed.
If you change only a file name and save the file, you do not need to press the "Add"
button in the report wizard.

To save a file in the template format, you can also use "File" - "Save As" of
See
MS-Word. At this time, be sure to select "Document template (*.dot)" as a
type of the file and pay attention to the saving destination.

- 77 - No.95691-2464-0000
[Explanation about the button to allow you to move a display window]
To move and alter the image display window that is added or has been already set, unlock the image
display window.


Operation
A) Specify the template of which layout you want to change from the "Installed templates" in the
Report Wizard.
B) Press the [Change Layout] button and open a template.
C) Press the "Move" button on the template creation toolbar. To move and alter a window, select
a image display window by clicking it, and drag the entire window or each line by the mouse.

This indicates that


the plane is selected
by the mouse.

Design mode button

D) When the image display window has been moved and altered, press the design mode button.
Alternatively, you can also unlock the window by selecting "View" - "Toolbar" - "Control Tool
Box" of the WORD menu, and press the design mode button.

Design mode button

After the window has been moved or altered, if you do not unlock the design
mode, you cannot add an image display window or a data code. Therefore,
Note be sure to unlock the design mode.

- 78 - No.95691-2464-0000
[Adding a template]
Add a template to the "Installed templates" list.


Operation
(1) Press the "Add" button of "Installed templates" in the Report Wizard.

(2) When the "Open File" dialog is displayed, select the template to be added and press the
"Open" button.

To add a template to
the list, click the
"Open" button.

Reference Only templates in the MS-Word format are intended.


The extension of the template format of MS-Word is "*.dot".
The extension of the template format of MS-Excel is "*.xlt".

- 79 - No.95691-2464-0000
9.1.3 View Menu

[Tool Bar]


Function Switches Show/Hide of the tool bar.

 Choose [View] from the menu bar, and select [Tool Bar]. Tool Bar appears, when it has a
Operation check mark beside it. Select the status bar again to hide it.

[Status Bar]


Function Switches Show/Hide of the status bar.

 Choose [View] from the menu bar, and select [Status Bar]. Status Bar appears, when it has a
Operation check mark beside it. Select the status bar again to hide it.

Reference The status bar shows the contents of the menu bar at the cursor position if
the mouse-cursor is placed on the menu bar.

- 80 - No.95691-2464-0000
[Update Folder]


Function Updates the folder select window to the latest status.


Operation Choose [View] from the menu bar and select [Update Folder]. The folder selection window is
updated to the latest status.

A new added folder appears

- 81 - No.95691-2464-0000
[New Folder]


Function Creates a new folder at the place you specify it in the folder select window.


Operation
A) In the folder select window, click the location you wish to create a new folder.

Click the location you wish.

B) Choose [View] from the menu bar, and select [New Folder].

C) A “new folder" is created at the place specified at step 1. The folder can be renamed. Type a
new holder name when you change it. Otherwise, the folder name is "New folder” as it is.

A created holder

Reference When you click [New Folder] at the location where a “New folder” already
exists, new folders are numbered in a serial order such as “New folder (1)”,
“New folder (2) ...

- 82 - No.95691-2464-0000
9.1.4 Option menu
[Save continuously]


Function Save all thermal images displayed in the main image window as a Bitmap format or a JPEG
format(Compression ratio is fixed) at a time.
(Enable for replaying of two or more thermal image files)


Operation
A) Choose [Option] from the menu bar, and then click [Save continuously] and select [BMP] or
[JPG] (file format you wish to convert to).

B) [Auto Convert] screen appears. Indicate the first and last page, and then click the [OK] button.
Click the [Cancel] button to stop processing

OK button

Cancel button

The First Page & the Last Page Settings


Enter the page range directly or by using
the spin buttons.

C) Each file is saved with the same name in the same folder as source file.

Source Thermal Image


files

Converted
BMP(JPG)
files

- 83 - No.95691-2464-0000
- When the file of the changed file, a same name, and this extension exists
Note in the same folder as a changing agency file, it is overwritten
compulsorily.
- The Saves continuously is effect on the Thermal Image Only (BMP, JPG
and other formats cannot be converted.)
- Name showed after conversion:
Thermal Image (sit, tmp) : Thermal Image File Name.BMP (or JPG)
Thermal Image (svt) : Thermal Image File Name + S
+Page Number.BMP (or JPG)
Visible Image : Replaying Thermal Image File Name V.BMP
(or JPG)
- The compression ratio of a JPG file is fixed.

[Level/Sense hold]


Function The current settings of level and sensitivity are held for all files displayed in the main
image window. (Enable for replaying of two or more thermal image files)


Operation
(1) Choose [Option] from the menu bar, and select [Level/Sense hold].

(2) A check mark appears at the top of [Level/Sense hold] when it is selected. Select [Lev
el/Sense hold] again to hide it. (The check mark does not appear.)

- 84 - No.95691-2464-0000
[Environment]


Function You can set environments, such as Object settings (color, character size, and Object emissivity),
Excel output settings (setting title, coordinates display, and decimal places), Temperature unit
settings (changing a date format, and changing a temperature format), and Folder setting (ahead
database outputting the folder, the template folder, the template folder and the initializing file
archive folder for Data Capture on Excel), Trend (trend graphical representation temperature
appointment) each environment is set.

A) Objects Setting (setting a color, a character size, and object emissivity)


Set the character size and display color of Objects, such as a POINT, BOX, LINE, POLYGON, and
CIRCLE.
This operation is the same as the function by selecting "Set Object" - "Style" from the right-click
menu on the set Object.


Operation
1. Select [Option] – [Environment].
2. The "Environment Setting" dialog is displayed. Then, click the Object setting tab.

Press the button of


the target Object.

A figure different
each page can be Mark this to adjust the
set by checking emissivity of the target Object
here. to the emissivity of the image file
automatically.

3. When you press the button of the target Object in the window above, the following dialog
for the corresponding Object is displayed

4. When you press the "Change" button, the "Font Setting" dialog is displayed and you can
specify the display character size and the display color. To enable the settings, press
the "OK" button.

- 85 - No.95691-2464-0000
B) Setting a temperature unit


Operation
1. Select [Option] – [Environment].
2. The "Environment Setting" dialog is displayed. Then, click the Temperature Unit tab.

Change the display format of


date.

Change the temperature display


format to C (centigrade) or
F (Fahrenheit).

Click the OK button to update the settings.

- 86 - No.95691-2464-0000
C) Setting of folder
The output place of a database, the preservation place of the template which the user created,
the preservation place of the template for captures which the user created, and the preservation
place of the initial-setting file of a report generation program are specified.
In addition, the image data by which the report output was carried out, object data, and a report file are
saved at the output place of a database.

[Database Folder]

 1. Select [Option] – [Environment].


Operation 2. The "Environment Setting" dialog is displayed. Then, click the “Folder setting” tab.
3. Consequently “Browse” button of the “Database folder” is pushed.

Specify the destination to which


the database folder is built.

The full path of the destination to which


the database folder is built is displayed.

Click the OK button to update the settings.


# A database folder specifies the place (point), which manages collectively the image data
by which the report output was carried out, figure data, a report file, and a database file.


Operation
[Report Template Folder]

1. Select [Option] – [Environment].


2. The "Environment Setting" dialog is displayed. Then, click the “Folder setting” tab.
3. Consequently “Browse” button of the “Report template folder” is pushed.

Specify the destination to which


the Template folder is built.

The full path of the destination to which


the Template folder is built is displayed.

Click the OK button to update the settings.


# A Report template folder specifies the place (point) where the created template is saved,
when there are new creation of a template and change of a layout by the user.

- 87 - No.95691-2464-0000

Operation
[Configuration File Folder]

1. Select [Option] – [Environment].


2. The "Environment Setting" dialog is displayed. Then, click the “Folder setting” tab.
3. Consequently “Browse” button of the “Configuration file folder” is pushed.

The full path of the destination to which


the Configuration file folder is built Specify the destination to which
is displayed. the Template folder is built.

Click the OK button to update the settings.


# A configuration file is a file which saves the configuration (object setting information etc.)
of a report generation program.

D) Trend
The graph it outputs which temperature value in the trend graph, it sets.


Operation 1. Select [Option] – [Environment].
2. The "Environment Setting" dialog is displayed. Then, click the “Trend” tab.
3. It makes which temperature value in the trend graph indicate, it selects.
When it checks in check box of each temperature value, that temperature value is
indicated in the trend graph.
(Maximum Temperature, Minimum Temperature, Average Temperature)

It becomes effective with the


OK button.

- 88 - No.95691-2464-0000
9.1.5 Help menu
[Version of Report Generator Lite]


Function Opens the Version of Report Generator Lite screen.


Operation
Choose [Help] from the menu bar and select [About Report Generator Lite], and then the
[About Report Generator Lite] screen opens.

Version of Report Generator Lite

Click the [OK] button to close the screen.

OK button

Program version is shown

Program name is shown

- 89 - No.95691-2464-0000
10. General Specifications
10.1 Specifications
1. Overview
This program is intended for creating a report based on our Viewer program. This program can
report mass data in a short time and create a report easily by using the template function.
In addition to the functions of the Viewer program, this program has the database function that manages
image data by Windows Explorer-like operations.

2. General specifications
● Target models
F30,TH9260,TH7800 series, TH7700 series, TH9100 series, TH7102 series, TS9100, TS7302,
TH5100 series,TS9230,TS9260
* Custom ranges are not supported.
● Target data formats
F30:*.jpg TH9260,TS9260: .six,.svx
TS9230,TH7800 series, TH7700 series, TH9100 series, TH7102 series, TS9100, TS7302: .sit, .svt
TH5100 series: .tmp
● Product components (NS9200LT)
 Product instruction manual 1

3. Processing specifications
The NS9200LT Report Generator Lite has the following functions.
A) File management function
1) Thumbnail display (infrared, visible, fusion display)
2) Condition search (date, file name, and comment (MEMO))
B) Thermal image display function (sit, tmp, and svt)
1) Thumbnail display of thermal image files (Windows Explorer)
2) Playing back thermal image files
3) Previewing images
4) Save …Data is saved (overwritten) with the same file name in the currently displayed
condition.
5) Save as …Save data with a new file name in the currently displayed condition.
C) Display change function
1) Setting a temperature level (level) (The range to be set conforms to the specifications of
the measured model.)
2) Setting sensitivity (sense) (The range to be set conforms to the specifications of the
measured model.)
3) Color scale …Select one from 8, 16, 64, 128, and 256 colors.
4) Changing a level span
5) Playing back voice …Playback, stop, pause, and mute
6) Switching between an infrared image, a visible image, and a combined image
7) Color/monochrome display (You can select color/monochrome display from 14 types.)
…Fine color positive/Fine color negative, color positive/color
negative, monochrome positive/monochrome negative, SHINE
color positive/SHINE color negative, brightness color
positive/brightness color negative, Hot Iron color positive/Hot
Iron color negative, Medical color positive/Medical color
negative
8) Multi-sense …Changes color resolution. (Middle: Center temperature part, High/Low:
Upper limit and lower limit temperature part)
9) ISO setting …Set isothermal band display (a total of 8).
10) Page …When multiple files are played back or an svt file is played back, a
display image is updated to the next or previous page.
11) Creating a new folder…Create a new folder at the location specified in the folder
selection window.

- 90 - No.95691-2464-0000
D) Edit function
1) Saving images (The BMP and JPEG file formats of the currently displayed image are
created.) …With the color bar, and thermal images only
2) Saving AVI (Images are saved in the AVI file format.)
…With the color bar, and thermal images only
* You can select the AVI saving frame rate from 60 fps, 30 fps, 15fps, 10 fps, 5 fps, 1 fps,
and 2 to 60 spf (1 spf step). Fps…frame per second, Spf…second per frame
3) Selecting a folder (search for a thermal image file)
4) Read zoom (100% to 5000%)

E) Pattern setting function


1) Setting a POINT: Set a POINT on a thermal image (up to 20 points)
2) Setting a BOX: Set a BOX (frame) on a thermal image (up to 20)
3) Correcting Object emissivity: Change the emissivity of the set Object.
(ε = 0.10 to 1.00: 0.01 step)
4) Setting ambient temperature at correction of emissivity: Set the ambient temperature at
data measurement for
correction of emissivity.

F) Analysis, processing, and edit functions


1) Sending data to external Excel: Transfers the temperature data of the specified pattern
to Excel sheets.
Targets: IR image, POINT, POINT trend, BOX, BOX trend
Copy: Copy the selected file or object.
2) Image copy: Copies preview display images in the BMP format.
3) Correcting emissivity (IR image and Objects)
4) Correcting lens distortion
5) Ambient compensation
6) Subtraction (whole image, area, similar object)

G) Graph creation function


1) Trend
Temperature trend : POINT
Maximum, minimum, and average trends: BOX

H) Report output function


1) Outputting reports (including templates) (one or multiple sheets): MS-Word format
MS-Excel format
2) Printing reports

- 91 - No.95691-2464-0000
I) Others
1) Setting options
Save continuously (display screen)…Executes file conversion of the current
playback images at a time to the BMP format
or the JPEG format
Maintaining level and sense…Maintains the current level and sense value.
Environment setting…Setting Objects (setting a color, a character size, and
emissivity), setting Excel output (setting a title, coordinates
display, decimal places), setting Temperature unit
(changing a date format, and changing a temperature
format), Folder setting(Database, Report template, Capture
template, configuration file),Trend setting,Fusion setting

10.2 Operation Environment


The following system environment is necessary to use the NS9200 Report Generator Program:

<Hardware>
 Operating system
Computer compatible with Microsoft® Windows® 2000 professional (Service Pack 4 or higher)
Computer compatible with Microsoft® Windows® XP professional (Service Pack 2 or higher)
Windows Vista Ultimate(32bit)
(NS9200 Report Generator operates with English OS environment.)
 Computer
Personal computer with the processor of Pentium 3 1.4 GHz or faster
 Memory
512 MB or more
 Display
Display with a resolution of XGA (1024 x 768) or more and a high color of 65,536 pixels
(16bits) or more

 Peripherals
HDD: 30 MB required for installing this program
USB port: Although the program is operated, 1 necessity
<Software>
 Operating system required for operating the NS9200.
Windows 2000 Professional (Service Pack 4 or higher),
Windows XP Professional (Service Pack 2 or higher)
Windows Vista Ultimate(32bit)

 Commercially available application required for operating NS9200.


Microsoft Word 2000, 2002, 2003 or 2007
Microsoft Excel 2000, 2002, 2003 or 2007

Microsoft, Windows, Word, and Excel are Microsoft's registered trademarks in the Unites States and
other countries. Product names mentioned above are a trademark or registered trademark of each
relevant company.

- 92 - No.95691-2464-0000
11. Others
1. This software program is a copyrighted work of NEC Avio Infrared Technologies Co., Ltd. It is
prohibited by law to copy, reproduce, and modify a part or all of this software without permission. Since
this software, however, is protected (usage is restricted) by the master USB key, you can install this
software to multiple personal computers.

2. Details of this software may be modified for improvement with no advance notice in the future.

3. If there are any problems or questions, contact our branch office nearby.