CALIBRATION I ASSERU!LY INSTRUCTION PAGE : 1
SECHRIST FOR 1, N o w FLulnIc ! u l w I F o m OF: s
-e!!m'- / ; + 2 - 1~t 4 5 3 EOP7870
911 9/01
PURPOSE:
This procedure is to instruct authorized personnel to replace and calibrate Sechrist ModcI
IV 100B Fluidic Manifold with Non Fluidic Manifold.
I:::ca.
1
--
1
REQUIRED TOOLS:
1.1 Remove four screws holding the lower control panel to the control ]nodule as
illustrated in Fig. 1.
1.2 Remove Ioi~rszrcws iron7 the back panel and disconnect rib1)url c k l c from the
MPU hoard.
FIG. 1
1.3 Disconr~ec~t thc solenoid md all rubirlg ti-on1 the existing Fluidic Manilbld iis
illustrated in big. 2.
1.5 Remove existing Manual Breath Valve and Manual Breath But-ton and hlntiual
Breath Cover as iljustrated in Fig. 2.
FIG. 2
2.1 If unit has 1 ' 16" tubing from mixcr to u~anifold,rcrnokc. reyuldtor asscmbly (Pih'
TV3 18) from vmlilalor. Remove double back tape from hottom of ~ e g ~ i l a t o r .
Remove OIIC existjng 118" NP'L' plug and thc huo fittings al llle in and out 1)~7silio1l.
Secure (3) 10-32x1/8 adaptc~(P/N IV321) and (2) 1 8 tilhc fillings ( P h' 1\'_?11) to
in and out positions, sccure 1 , l b" tube fitting (PiNJiy310) as iIluslr31t.d in Fig.3.
h1sr.k "ill" side of regulator for easy installalion. Keplace double hack tapc on
r c ~ ~ ~ l a tRe~nove
or. 1116" tube fitting from ~rlixerand secure 118" tube fitting (P, N
1V3 1 1) Install regulator back in veiltilator. Rcplace old l i l b" ilLibing\kith P/N
HB333 as illustrated i n Fig.3.
DOCUMENT NO. 140126 IiEk': 4
w&&z-
rl&ll&*
CALIBRATION I ASSERlBLY INSTRUCTION PAGE : 4
SECHRIST
-!!i=ye=i
FOR I V 1 OOBNON- FLUIDIC RIAR'IFOLD OF: 8
EO#7870
7.2. If unit has 1/8" tubing remove one existing li8" NPT plug from the reguIator
assembly as illustrated in Fig.3. Secure one 10-32 x 1IS " adapter ( P j ' N 3 2 1) and
1/16" tube fitting (P/N3 10) to the regulator asscmblj..
2.3 Install new style rna~lualbreath valve with retaining ring and ma!lual breath cover
as illustrated in Fig. 3.
2.4 Carefillly take New Non-Fluidic Manifold (PIN 2 1435) out from the package.
2.5 Align the lnspiratory and Expiratory Control Valves to the appropriate Iloles in
lower control pancl.
2.6 Use retaining pliers to snap the retaining ring back on the Control Valves.
2.7 Install lnspiralory aid Expiratory Control knobs :IS iIlustrated in Fig. 3
hOTE: To install thc lnspiralory and Expiratory przssure knobs, first rohle thc
Inspiratory and Expiratory valve stems clockwise until they seat gcntly. Rotalz
each valve stem countcrclocku~ise00" i 15 "; inslall the knobs 1s it11 knob slop pit1
rcsting against the valve stop pin. Turn both valves fuIly countcrclockwisc.
I:
1
I'
1;
.: ..I.. 1 .
1, ..
DOCUh'lENT NO. 140126 REV: 3
w*s-
&~brr&*
CL4L+1BRATION/ ASSEILIELY IKSTRUCTION PACE : 7
SELHRIST FOR IV 1 OOBNOS- FI.UIDJC MANIYOI.D OF: 8
LOff7870
9/191!Il
3.1 SETUP
3.1.1 Connect 50 psl air and oxygen supply to the air and oxygen fitting as illustraled i n
Fig. 4 to the gas mixer. Connect patient circuif asselnbly, test lrtng, a11J relicf
valve. Turn relief valve all the way in.
'
-
.-- I-[.,
1 3.1.2 Corulcct 0-60 ?si pressure gauge to the gas regulator assembly (i'!N IVS 18).
/'>> ,
' 33-1.3 Open R4 6- RS Yi torn, R6 1 turn, R3 '/; turn and R2 '/, tun, on lhc new Non-
1 -
-1
j Fluidic Manifold. Remove plug and install one IV3 3 0 (1/1G" tubc- fitting) at port
,\ J
E on the Non Fluidic hqanifold as shorvn in Fig. 3. Connect 0-5 psi prcssurc
I Y' gauge to the port labeled "E".
r-- a
~><-q
2.
Note: Rcfererlcc Fig. 4 for the vdlve locntlons.
b ;\,
3.2 Calibrate Pressure Rcguhtor
3.6.1 Slowly Turn the Expiratory pressure control knob to fully clockwise. Adjust the
It4 until the manonlctcr on the ventilator reads 20 k 2.0 cm H . 0 .
' \(J'~
', \ 3.8.1 Turn the ventilator mode to CPAP. Whilc pushing manual breath butran, slo\vly
turn the Inspiralory pressure control knob fully clock\vise. Adjust R2 to 70-75 cm
'I of H20.
7.9.1 Tuln the Mode Switch to Vcnt. Turtl thc Inspi ratory pressure control knob
unti 1 the manometer re:~ds30 L 1 cm of H20. Turn the Expiratory pressure
control knob countcrciockn~isclo clockwise. Verify Illat the manotllctcr
rcnding does not valy by morz than 3 cm of H20. Repeat with Tnspiralory
Pressurc sct bctwctln 70 2nd 75 cnl t 120.Turn the Expiralory prcssurc
cuntl-ol k~lobirntjl the manonicrtet. reads 10 2 1 cm of HzO. Turn the
lnspiratuq prcssurc control k l ~ a bcounterclock~\~ise to clockwise, vcrify
lhal 1112 I I I ; I I I O I I K ~ ~ Treading docs riot vary by morc than 3 till of HzO.
Rcpeal \iilh Expiralory pressurc set between 0 and -2 cm J120.