Laser Instruction

The world leader in serving science Manual

Dec 2008 Ver.2 Bremen


Before start the Laser analysis

 How often need to Cross Calibration?


• Everyday more than 50 samples, it would be nice run the Cross Calibration every
week
• During tuning with laser, if you see low sensitivity compare with last analysis on same
setting, it would need to run the Cross Calibration

 How to run Cross calibration


• With Tune-F (P/N SP1600698) solution on Hot Plasma mode

 How often need to Mass Calibration?


• If you break the vacuum of XSERIES 2, it is necessary run the Mass Calibration
 How to run Mass Calibration-
• Standard way with Tune-A (P/N SP1600633) or Tune-F (P/N SP1600698) solution on
Hot Plasma mode every 3 month
or
• Laser with NIST 612 glass on Hot Plasma See the next slide

2
Mass calibration with NIST 612 on laser-Setting of laser

Total time should be more


than 4 min

3
Mass calibration with NIST 612 on laser-Check the sensitivity

 Check the enough sensitivity each element should be more than


100000cps

4
Mass calibration with NIST 612 on laser- start Mass Calibration

 Start the Mass Calibration

5
System of Laser

Connect to PC (Trigger card)

Check the water line

Go to Torch Additional gas(He)

6
Accessory
Xt Ni Skimmer
Quartz torch cone
bonnet PN 3600811
PN 3601149

Silver
Screen
PN 3601219

Quartz torch
1.5mm
PN 3601145

Ni
Sample cone
PN 3600812

Elbow
connector
PN 3600730

7
Gas Line between Laser with X Series2

Torch
Nebliser Gas(Ar)

Nebliser gas

From Laser
GAS OUT

8
Cable Connection

Behind of PC Cable from


Laser

Video To Laser
Card system

9
Laser Set Up

1. Laser System -Turn on (30min before)


2. Laser Soft Ware – open

3. Starting calibration of Sample stage automatically

10
Spot size -1

Spot size is not necessary to check


every time. this section is just
information how to check or change
the spot size

11
Spot size -2

 Choose the Imaged Aperture

12
Screen calibration-1

 Set the Calibration Glass on the Laser sample cell


• Focus the calibration glass

 Select the Calibration screen

Screen calibration is not necessary to


run every time. this section is just
information how to check the screen
calibration
13
Screen Calibration-2

 Drug the mouth to scale the between line

14
Screen Calibration-3

 This distance should be 100um, if it is not the right distance, you


need to fill in the right number

 Zoom position contains 0%, 25%, 50%, 75%, 100%, each zoom
position need to check the screen calibration on same way
 After check the all of the zoom, please back on 0%

15
Communication test

If laser software
will appear, the
connection test is
success

16
Plasma Turn On -1

1. Check the gas Line of Laser – “On-Line” mode on Laser Software

17
Plasma Turn On -2

2. Add-gas flush out Manuality on PlasmaLab, then turn on the Plasma,


Add-gas keep flush out (only during the gas purge)

18
Plasma Turn On -3

1. Add-gas return to “0” (Tuning torch box) Manually- if you keep flush Add gas,
Plasma can not be turn on-

2. Wait until “Operate”

19
Plasma Turn On -4

3. After “Operate” appeared, Add gas should be increasing slowly

He gas should be increased slowly due to maintain


the plasma up to 200mL/min.

20
Tuning-1

1. Go to the PlasmaLab Software – Check the ACL or Available Accessories-

21
Laser ACL Script

 You can create the ACL script

How to install this ACL script is on next slid

22
How to install ACL script -1

23
How to install ACL script -2

24
Tuning-2

1. Laser door should be closed


1. Laser Warm up (Recommend time is for 15min)

2. Fire click

It would be nice to increase output of laser by low(20%),


middle(50%), high(75%) steps due to warm up the laser

25
How to install ACL script -3

Click the “ save as “ and


fill in the title

After save, it possible to


choose the ACL script
completely

26
Tuning-3

2. Set the Sample on the sample stage


(Laser Gas line is “Bypass” put on
sample lock the stage)
NIST Glass 612(http://www.nist.gov/ )is
best for Tuning standard material

• Set the sample • Lock the sample stage

27
Tuning-4

3. After “Purge” click, it is changed to “On-Line” automatically

28
Tuning-5-a

1. Turn on the light of Laser

2-a. Find out the sample on the monitor – Manually -

To optimize
Focus
To optimize the
sample position

29
Tuning-5-b

2-b. Find out the sample on the monitor – Autofocus -

30
Tuning-6
1. Create the tuning line – Choose the “Raster”- Drug the mouse on the Screen

31
Tuning-7
2. “Properties” click
Tuning laser parameter

Total Time is tuning


time; to get enough
time, you should
create the Raster
Area

32
Tuning-8

1. “Run Scan” click, 2. Highlight the “Enable Laser During Scans”

33
Tuning-9

1. Select the Element –139La, 232Th, 238U, 248ThO, 220BKG - Check the sensitivity
139La > 4000cps/ppm
238U>
10000cps/ppm

2. Oxide Ratio should be below of “1%”

34
Tuning-10

1. Checking the signal on the monitor, gas parameter will adjust Neb gas - Add
gas - Neb gas at first
2. Torch position optimize
3. ExtractionLens,Lens-1,FocusLens optimize

Recommended Parameter Xt cone


Neb gas ; 0.6 to 0.8 L/min
Add Gas (He gas) ; 400 to 700 mL/min
Focus ; 10 to 13 v
L1 ; -1000 to -1300v
Extraction ; -100 to -150V

4. Save the setting parameter

35
How to create the
The world leader in serving science Experiment

This section highlight the how to create


the method and how to put together the
data using Ni alloy sample
Most important thing to create the method

 The duration time on the sample list, it should not be shorter than
total time. please see the Create Experiment-8 with carefully

• Duration time > uptake time + wash out time + laser warm up +
ablation time + extra 5sec

!!If the duration time is not correct time, sample list won’t start!!

37
Create Experiment-1

1. Select the “Profile TRA”

2. Select the “default”

38
Create Experiment-2

1. Type the Uptake and Washout

2. Select the Analyte

39
Create Experiment-3

1. “Dwell time” and “Resolution” select ( Ni Sample )

Dwell time ; 15ms to 50ms


Resolution ; High Resolution is useful to separate from interference
peak and avoid to be aging the detector for counting matrix element

40
Create Experiment-4

1. Internal Standard – correction using matrix Element

Don’t fill in the Internal Standard page now – please Ignore -

41
Create Experiment-5

1. Isotope Ratio – this is also able to be appearing after finish analysis -

42
Create Experiment-6

1. Select the Calibration method

Fully Quant : there is certification standard concentration

Semi-quantified : If there is no certification of


concentration, you can use Semi-quantified method

Addition : Standard addition method is not using for LA-


ICP-MS

Internal Standard and equation elements should be “None”

43
Create Experiment-7
1. Create the ablate line and ablate setting

Sample Line ablation time


1000um/(20um/msec)=50sec

1000um

44
Create Experiment-8

1. Sample List Scan Patterns is same


number with Laser Pattern

2. Duration (Secs) UptakeTime + Laser WarmUp Time + Ablation + wash out Time
LA Warm-Up
Gas Blank 30sec Sample Ablation: Ablated Particles go through the torch 55sec
5sec

45
Create Experiment-9

1. Fully Quantitative Concentration

2. Save the experiment

3. Start the experiment

46
How to combine the Result-1
1. Go to “Acquisition Parameters” and Select the “Per analyte”

47
Compare with Spectra

1. choose the one sample spectra


2. click this icon,
3. other sample double click ,
you can see comparison for each spectra

48
How to remove the noise

Click

High light

Click

49
Combine data -1

Next Slide


Click

Gas blank Region Sample ablation region

50
Combine data -2

Chose the Baseline


Region


High light the samples

③ You can change the parameter and put the data


together

Right click, and


“Copy parameters”

51
Combine data -3

1. Internal Standard – correction using matrix Element

1. Chose the matrix Element of sample

2. Fill in each information

52
Result

Sample is Result

53
How to install laser software and video card

Find the homepage from NEW WAVE http://www.new-wave.com/ and Click “Service” and select “Spares”

54
How to install laser software and video card

55
How to install laser software and video card

56
How to install laser software and video card

57
How to install laser software and video card

58