Professional Documents
Culture Documents
AGPL-QP-RFT-03
General Procedure
GENERAL PROCEDURE
Remote Field Testing (RFT)
of Ferromagnetic Tubes
AGPL-QP-RFT-03
Revision History:
Rev No Doc Release Dates Prepared by Reviewed by Authorized by
00 11-Feb-2013 Sabari Selavan Sumit Lamba Tawfik Mohamed
01 12-Jan-2015 Sabari Selavan Sumit Lamba Tawfik Mohamed
02 15-August 2017 Sabari Selavan Sumit Lamba Tawfik Mohamed
03 22-June-2018 Sabari Selavan Sumit Lamba Tawfik Mohamed
1
Ref No.
AGPL-QP-RFT-03
General Procedure
1. Purpose
The puspose of this procedure is to establish the general requirements for RFT
inspection on ferromagnetic tubes and also includes - tube preparation and cleaning,
probes selection, calibration process, equipment set up process, reporting, etc.
2. Scope
This procedure is applicable to Remote Field Testing (RFT) of ferromagnetic tubes
with an internal diameter of between 8mm and 120mm.
3. References
3.1 ASME BPVC Section V -2015 ASME Boiler & Pressure Vessel Code
3.2 ASME BPVC Section V -2015 Remote Field Testing (RFT) Examination
Article 17 Method
3.3 ASTM E 2096 – 08 In-Situ Examination of Ferromagnetic
Heat Exchanger Tubes Using Remote Field
4. Personnel
Non-destructive Testing personnel conducting inspections, analyzing data and
reporting results shall be qualified & certified ASNT NDT ECT level 2 with AGPL’s
Written Practice established in compliance with SNT-TC-1A (2015) or equivalent
(BS EN or ISO or AINDT Level II).
5. Equipment
• The equipment selected shall as a minimum be capable of displaying results on
impedance plane readout, and be capable of discriminating phase and amplitude.
• Within this procedure the following tube inspection equipment is considered:
Eddyfi Ectane 2, Olympus MS5800 or equivalent.
• Inspection system shall be capable of operating in single or multi frequency, in
both absolute and differential modes.
• Probes selected shall be suitable for the minimum required fill factor of not less
than 70%.
• The probes suitable for RFT inspection of heat exchangers and boilers are
specified in Attachement 1.
2
Ref No.
AGPL-QP-RFT-03
General Procedure
• Probe must be selected in accordance with tube diameter, however the gap 1-
1.5mm between probe and tube internal diameter must be considered for
allowing probe to go through.
• Probes shall be capable of detecting the wastage in the calibration tubes at the
required frequencies selected for the inspection.
3
Ref No.
AGPL-QP-RFT-03
General Procedure
Example #1
Example # 2
4
Ref No.
AGPL-QP-RFT-03
General Procedure
8. System Set Up
The test equipment shall be set up in accordance with this procedure and the onboard
Wizard on the equipment. The following information shall be contained within the
instruction:
• Description of equipment being inspected including (client, item and item number).
• Specification of tubing being inspected including (tube material, tube diameter,
tube wall thickness, tube length).
• Test probe specification including (probe diameter, fill factor, test mode –
differential / absolute).
• In case of a suitable calibration tube not available, alternative procedure could be
followed as provided by Level III after client’s acceptance.
• Select a frequency such that the signal from the support plate lies horizontal on the
voltage plane.
Data analysis shall be carried out by using the attenuation spiral in the Multiview or
Magnifi Software.
Calibration shall be carried out immediately prior to the start of the inspection.
• The test equipment shall be calibrated in accordance with this procedure detailing
signal settings and amplitudes as well as data capture settings.
• The equipment shall be balanced with the probe in the calibration tube on an area
with no defects.
• All defects in the calibration tube shall be calibrated as per the process channels of
the software.
• Any change in probe, extension cable, test instrument or any other part of the
examination system shall be cause for a re-calibration.
• The test equipment response shall be checked and verified by use of the calibration
tube standards as follows:
At least once every four hours.
At each change of operator or any probe, cable or inspection equipment.
Whenever malfunctioning is suspected.
5
Ref No.
AGPL-QP-RFT-03
General Procedure
When a new log file is opened, and the last tube in the previous log file shall
be re-inspected to check that no significant change in differential or absolute
response has taken place.
If, during any check, it is determined by the data analyst that the test equipment is
out of calibration the equipment shall be re-calibrated and a new data file opened.
The data analyst shall determine tubes, if any, are to be re-examined.
12. Calibration
Calibration of the test equipment and relevant modules shall be undertaken in
accordance with manufacturer’s procedures or an authorized Agent of the equipment
being utilized. The period between calibrations shall not exceed 1 year.
6
Ref No.
AGPL-QP-RFT-03
General Procedure
13.Reporting
The following information is required for the report as a minimum:
• Customer, location, equipment tested and serial numbers.
• Size, material type and grade as well as the configuration of tubes being inspected.
• Tube numbering system.
• Extent of examination, location of tubes (to inspect drawing) and length.
• Personnel and qualifications.
• Equipment used and accessories (in the technique sheet).
• Calibration standards used (in the technique sheet).
• Outline of the technique used.
A copy of all the inspection data and reports shall be handed to the operations
inspection division of the client.
7
Ref No.
AGPL-QP-RFT-03
General Procedure
ATTACHEMENT 1
mm 6.05 5.59 5.16 4.57 4.19 3.76 3.40 3.05 2.77 2.41 2.11 1.83 1.65 1.47 1.24
in 0.238 0.220 0.206 0.180 0.165 0.148 0.135 0.120 0.109 0.095 0.083 0.072 0.065 0.058 0.049
22.22 mm 0.875 in 090 100 100 110 120 130 130 140 140 150 160 160 160 170 170
25.40 mm 1.000 in 120 120 130 140 150 150 160 170 170 180 180 190 190 190 200
31.75 mm 1.250 in 180 180 180 200 200 200 220 220 220 240 240 240 240 240 260
38.10 mm 1.500 in 220 240 240 260 260 260 280 280 300 300 300 300 300 300 300
50.80 mm 2.000 in 340 360 360 380 380 380 400 400 400 420 420 420 420 420 440
mm 7.62 7.21 6.58 6.05 5.59 5.16 4.57 4.19 3.76 3.40 3.05 2.77 2.41 2.11
in 0.300 0.284 0.259 0.238 0.220 0.206 0.180 0.165 0.148 0.135 0.120 0.109 0.095 0.083
50.80 mm 2.0 in 260 320 320 320 320 320 360 360 360 360 360 360 360 400
63.50 mm 2.5 in 400 400 400 400 450 450 450 450 450 450 450 500 500 500
76.20 mm 3.0 in 500 500 500 550 550 550 550 550 550 550 600 600 600 600
Note: Frequency of probe (LF, MF, and HF) will depend on combination of the
material and thickness of the tubes.