You are on page 1of 4

General Model RAQN

Plastic Rotameter ®
Specifications

This type of Rotameter is designed for measurement of liquids


and gases. The conical plastic metering tube has a free rotating
float.
The mounting position is vertical with flow direction upwards.
The flow is indicated by the top position of the float and can be
read from the standard scale on the metering tube or from a
connected scale.

FEATURES
● Breakage proof
● Easy to handle
● Highly corrosion-resistant
● Low pressure loss
● Visible monitoring of the measured medium
● Local indication without resource
● Selectable scale

TECHNICAL SPECIFICATIONS OF OPTIONS


STANDARD SPECIFICATIONS
Measurable flow rates
- Water (20 °C) : 10 l/h to 10 m3/h LIMIT SWITCH (OPTION CODE: /GM1 AND /GM2)
- Air (20 °C)(1 bar abs.) : 160 l/h to 250 m3/h (for floats with magnet only and flow > 16 l/h water or 630 l/h air)
Measuring range : 10:1 Type : Reed-contact with bistable switching
Accuracy : According to VDI 3513, Max. switching voltage : 230 V
Class 4 (2,5) Max. switching current : 0.6 A
Max. pressure : 10 bar Max switching capacity : 12 VA or 12 W
Max. temperature Temperature range : -10 ºC to +70 ºC
- Polyamide metering tube : -10 to + 60 °C Protection : IP 65
- Polysulphone metering tube : -10 to + 100 °C Internal capacity : 0 nF
Material process connection Internal inductivity : 0 mH
- Inner thread : Steel Electrical connection : LIYY 2 x 0.34, 1 m long
- Glue socket : PVC Weight : 35 g
Material metering tube : Polyamide or Polysulphone
Design (valve) : With or without optional valve POWER SUPPLY FOR LIMIT SWITCHES
Length : Approx. 400 mm (OPTION CODE: /WXX)
Weight : Dependent on design Type : Transistor relay acc. DIN 19234
0.4 to 2.1 kg (NAMUR)
Power supply : 230 V AC + 10%/-15%,
(see “Dimensions” on page 4)
50/60 Hz (/W2A,/W2B)
24 V DC ± 20% (/W4A,/W4B)
Relay output : 1 potential-free changeover
contact (/W2A, /W4A)
2 potential-free changeover
contacts (/W2B, /W4B)
Switching capacity : Max. 250 V AC; max. 2 A
Control circuit : Intrinsically safe [EEx ia] IIC
Entity parameters : acc. certificate of conformity
PTB Nr. Ex-94.C2086

YOKOGAWA GS 1R1B4-E-H
3rd Edition
ROTA YOKOGAWA
2

MODEL SPECIFICATIONS
Diameter process Model Process conn. Material Material Design Metering tube
connection Inner Glue Process Rotameter (valve) length/
Without valve thread socket connection diameter
Inner thread Glue socket Code Code Code Code Code Code Code
Rp 3/8 inch RAQN42 R1 - ST; PC PA; PS NNN Q1
Rp 1/2 inch RAQN01 R1 - ST; PC PA; PS NNN Q1
Rp 3/4 inch RAQN23 R1 - ST; PC PA; PS NNN Q1
Rp 1 inch RAQN02 R1 - ST; PC PA; PS NNN Q1
RAQN02 R1 - ST PA; PS NNN Q2
32 mm RAQN02 - K1 PC PA; PS NNN Q1
Rp 1 1/4 inch RAQN03 R1 - ST, PC PA; PS NNN Q2
40 mm RAQN03 - K1 PC PA; PS NNN Q2
Rp 1 1/2 inch RAQN04 R1 - ST; PC PA; PS NNN Q4
Rp 2 inch RAQN05 R1 - ST; PC PA; PS NNN Q4
63 mm RAQN05 - K1 PC PA; PS NNN Q4
With valve
Rp 1 inch RAQN02 R1 - ST1) PA; PS RAE Q1
Rp 1 1/4 inch RAQN03 R1 - ST1) PA; PS RAE Q2
Rp 2 inch RAQN05 R1 - ST1) PA; PS RAE Q4
Process connnection: Inner thread Rp ......... -R1
Glue socket (only in PVC) .......... -K1
Material process connection: Steel .................. ST1)
PVC ................... PC
Material Rotameter: Polyamide ............................... -PA
Polysulphone .......................... -PS
Design: Without valve .................................................. NNN
Valve in input .................................................. RAE
The suffix code of the combination metering tube - float can be read from the flow tables on page 3 ............. -kkkkk-kkkkk

Style ................................................................................................................................................................................................. *A
Options (see table below) ......................................................................................................................................................................... /kk
1)
The connection of the Rotameter with valve is made of brass.

OPTIONS
Options Option code Description
Mark /B1 Stainless steel tag plate
/BK Customer-spec. notes (please specify)
Limit Contact Function Application
switch /GM1 MlN-contact mag. Switch contact float with magnet only
/GM2 MAX-contact (> 16 l/h water or 630 l/h air)
If two limit switches are selected both option codes should be specified
Test/ /H1 Certificate of degrease
Certificates /PT With flow table for recalculation
/PP With inspection certificate for pressurised air
Additional /W2A Transmitter relay KFA6-SR2-Ex1.W / 230 V AC, 1 channel, changeover contact
Instruments /W2B Transmitter relay KFA6-SR2-Ex2.W / 230 V AC, 2 channel, changeover contacts
/W4A Transmitter relay KFD2-SR2-Ex1.W / 24 V DC, 1 channel, changeover contact
/W4B Transmitter relay KFD2-SR2-Ex2.W / 24 V DC, 2 channel, changeover contacts

RECOMMENDATIONS FOR DETERMINING THE MODEL CODE


Please specify in the following order
- Measuring range for water/liquid or air/gas
- With or without optional limit switch
- With or without optional valve

First select the required measuring range from the flow table (last column) and specify the float insertion for the optional limit switch. Then
the suffix code of the combination metering tube - float can be fixed.
The suffix code of the measuring tube length/diameter with or without the optional valve completes the model code specification.

GS 1R1B4-E-H
3

FLOW TABLE WITH THE METERING TUBE - FLOAT COMBINATION


Flow table Suffix code of the metering tube - float combination
Water/Liquids 20°C Metering tube Float
Recommended comb. Altemative comb. -k k kk k -kk k k k
Row A Row B
Max. Pressure Max. Pressure Length Diameter Metering Scale Material Dia- Flow Insertion
flow drop*) flow drop*) tube cone meter mark
[l/h] [mbar] [l/h] [mbar] code code code code code code code code
- - 100 2,5 Q 1 53 PP; PC 1 N
250 15 160 6 Q 1 53
400 16 250 7 Q 1 54 Row A=SS 1 Row A=3 M2); N
630 18 400 8 Q 1 57 Row B=PP;PC Row B=2
1 000 26 630 10 Q 1 61 A;
- - 1 000 6 Q 2 63 P PP; PC 1 N
2 500 30 - - Q 2 63 Row A=SS 2 Row A=3 M2); N
- - 1 600 15 Q 2 63 Row B=PP;PC Row B=2
- - 4 000 13 Q 4 71 PP;PC 1
10 000 50 - - Q 4 71 Row A=SS 4 Row A=3 M2); N
- - 6 300 25 Q 4 71 Row B=PP;PC Row B=2

Flow table Suffix code of the metering tube - float combination


Air/Gas 20°C; 1 bar absolute Metering tube Float
Recommended comb. Altemative comb. -k k kk k -kk k k k
Row A Row B
Max. Pressure Max. Pressure Length Diameter Metering Scale Material Dia- Flow Insertion
flow drop*) flow drop*) tube cone meter mark
[l/h] [mbar] [l/h] [mbar] code code code code code code code code
- - 1 600 1 Q 1 53 4 N
2 500 2 6 300 11 Q 1 53 Row A
4 000 2 10 000 14 Q 1 54 PP; PC 1 Row A=5 N
6 300 3 16 000 16 Q 1 57 Row B=7 Row B
10 000 3 25 000 19 Q 1 61 A; M2); N
- - 40 000 12 Q 2 63 P 6 N
25 000 5 - - Q 2 63 PP; PC 2 Row A=5 N
- - 63 000 26 Q 2 63 Row B=7 M2); N
- - 63 000 3 Q 4 71 4 N
- - 160 000 16 Q 4 71 6 M2); N
100 000 7 - - Q 4 71 PP;PC 4 Row A=5 N
- - 250 000 34 Q 4 71 Row B=7 M2); N

Description
Length metering tube 350 mm .................... Q
Diameter metering tube R 1 1/2 inch to R2 3/4 inch. ....... k
Cone metering tube See flow table .............................................. kk
Medium scala Connection scale .......................................................... A;
metering tube Metering tube with % indication only ............................ P
Material float 1.4571 ............................................................................................. SS
Polypropylene .................................................................................. PP
PVC ................................................................................................. PC
Diameter metering tube 21,6 mm to 54,1 mm ................................................................................................ k
Flow mark Factor 0,4 to 1 (Water) (code 1 to 3)
Factor 6,3 to 25 (Air) (code 4 to 7) .......................................................................................... k
Float insertion Without magnet ......................................................................................................................................... N
With magnet (for limit switches /GM1 and /GM2) ...................................................................................... M2)
2)
For optional limit switches /GM1 and /GM2
*)
The indicated flow pressure is a target value and may deviate based on the type of Rotameter.

Other measuring tube-float combinations with other float materials and a variety of measuring ranges are available on request.

GS 1R1B4-E-H
4

DIMENSIONS

Metering tube without Plastic insertion parts Steel insertion parts RAQN with
connecting parts etc.

øG

350 A 356 350 A A


350
C B

200 350

øG

ød øg

WEIGHTS AND DIMENSIONS


Metering Unit (mm) Weight
tube G D A d B C g (kg)
Q1 G11/2” 34 400 32 366 400 Rp1” 0,4
Q2 G2” 48 408 40 372 410 Rp11/4” 0,7
Q4 G23/4” 81 432 63 370 418 Rp2” 2,1

GERMANY AUSTRALIA UNITED STATES OF AMERICA


YOKOGAWA Yokogawa Deutschland GmbH
Berliner Strasse 101-103
D-40880 RATINGEN
Yokogawa Australia Pty Ltd.
Private mail bag 24
Centre Court D3
Yokogawa Industrial Automation
4 Dart Road
NEWNAN, GA 30265-1040
Tel. +49-2102-4983 0 25-27 Paul Street North Tel. +1-770-254 0400
AUSTRIA Fax +49-2102-4983 22 NORTH RYDE, N.S.W. 2113 Fax +1-770-251 6427
EUROPEAN HEADQUARTERS Tel. +61-2-805 0699
Yokogawa Europe B.V. Yokogawa Austria Ges.m.b.H. ISO 9001
Franzensbrückenstrasse 26 ITALY Fax +61-2-888 1844
Vanadiumweg 11, Yokogawa Italia S.r.l.
3812 PX AMERSFOORT A-1021 WIEN
Tel. +43-1-216 50 43 Vicolo D. Pantaleoni, 4 SINGAPORE
The Netherlands 20161 MILANO Yokogawa Engineering Asia Pte. Ltd.
Tel. +31-33-4641 611 Fax +43-1-216 50 43 33
Tel. +39-2-66 24 11 11, Tampines Street 92
Fax +31-33-4641 610 Fax +39-2-645 57 02 SINGAPORE, 528872
BELGIUM CERTIFICATED
E-mail: info@yokogawa.nl Tel. +65-783 9537 FIRM
Yokogawa Belgium N.V./S.A.
Minervastraat 16 SPAIN Fax +65-786 2606
THE NETHERLANDS Yokogawa España S.A.
Yokogawa Nederland B.V. 1930 ZAVENTEM
Tel. +32-2-719 55 11 C/Francisco Remiro, N°2, Edif. H Manufactured by:
Computerweg 1, 28028 MADRID GERMANY
3606 AV MAARSSEN Fax +32-2-725 34 99 CENTRAL/EAST REGION
Tel. +34-1-724 20 80 Rota Yokogawa GmbH & Co. KG
Tel. +31-346-59 12 11 Fax +34-1-355 31 40 Rheinstrasse 8 Via Yokogawa Austria: Czechia,
Fax +31-346-56 61 96 FRANCE Slovakia, Hungary, Poland,
Yokogawa Contrôle Bailey S.A. D-79664 WEHR
UNITED KINGDOM Tel. +49-7761-567 0 Croatia, Slovenia, Jugoslavia,
Vélizy Valley Bulgaria, Romania, Macedonia,
18-20 Rue Grange Dame Rose Yokogawa United Kingdom Ltd. Fax +49-7761-567 126
Stuart Road, Manor Park, Bosnia & Herzegovina
78140 VELIZY VILLACOUBLAY
Tel. +33-1-39 26 10 00 RUNCORN SOUTH AFRICA
Cheshire WA7 1TR Yokogawa South Africa (Pty) ltd. Distributors in:
Fax +33-1-39 26 10 30 Denmark, Finland, Greece,
Tel. +44-1-928 597100 67 Port Road, Robertsham
Fax +44-1-928 597101 Southdale 2135, JOHANNESBURG Norway, Portugal, Russian
Tel. +27-11-680-5420 Federation, Sweden,
Fax +27-11-680-2922 Switzerland and Turkey.

Block 03,01-98

GS 1R1B4-E-H Printed in The Netherlands


Subject to change without notice 03-802(A) Q
Copyrightf