Professional Documents
Culture Documents
whether the ultimate quantity being measured is mass, volume at base conditions,
energy or volume at flowing conditions. Appropriate conversions, relative to the gas
physical properties and process conditions, must be applied to accurately obtain the
desired quantity. The equation for flow rate at base conditions, including flow pressure
Fp
effect compensation factor ( ), is such that,
Fp Fp
Qb Q mUncompensa ted Q mUncmpensat ed
G Eq. (8.3)
b r b ( Air )
The relationship for the calculation of energy, including flow pressure effect
Fp
compensation factor ( ), is such that,
8.4 MAINTENANCE
Users should follow the manufacturer’s recommendations for maintenance. Monitoring
diagnostics, performing the periodic meter verification procedures discussed in Section 9.1 and
possibly the trending/long-term monitoring of performance indicators, can identify if abnormal
conditions develop.
The decision to perform periodic flow test is left to the user or defined by contract. Periodic flow
test can be calendar-based or, as a minimum, condition-based. A number of companies have
30
flow-tested Coriolis meters that have been in service for significant (in excess of 15 years)
periods of time. These test results have shown that meters, free from flow tube corrosion, erosion
or mechanical damage, perform as originally flow-calibrated, within the uncertainty of the flow lab.
This testing has also shown that coating of the flow tubes by debris or product residue may
change flow tube balance and cause a subsequent zero shift, but if the coating is stable, left as-is
and the meter is re-zeroed, the meter performs as originally calibrated. See Section 9.2, “Flow
Performance Testing” for further discussion.
9 Meter Verification and Flow Performance Testing
The meter manufacturer should provide the meter operator with written field meter verification test
procedures that will allow the Coriolis meter, as a component of the measuring system, to be verified
as operating properly and performing within the measurement uncertainty limits required by the
designer/operator. Further discussion of these checks can also be found in Appendix C, Section 3.4.1.
9.1 FIELD METER VERIFICATION
The field verification of a Coriolis metering system consists of monitoring and evaluating metering
conditions, diagnostic indicators output by the transmitter and/or ancillary devices of the metering
system designed to identify possible change in the system’s performance and the cause. The
evaluation of these indicators will guide the operator in determining the need to re-zero the
Coriolis meter, execute a flow performance test (in-situ or laboratory), adjust maintenance
intervals and implement design improvements, if necessary, to the metering system.
The operator should follow design-specific meter verification procedures recommended by the
manufacturer and as a minimum the following general meter verification procedures should be
performed.
Meter Transmitter Verification – The meter transmitter verification should coincide with
the meter zero check. It should include the following procedures:
Meter Zero Verification – A change in the meter zero value can be used as an indicator
of change in the metering conditions. This can be caused by contamination and coating of
31