Professional Documents
Culture Documents
a r t i c l e i n f o a b s t r a c t
Article history: A new trapped ion mobility spectrometry (TIMS) apparatus has been developed and evaluated. Building
Received 5 February 2016 on the original prototype design, the new spectrometer includes an upstream “ion accumulation trap”.
Accepted 7 March 2016 Inclusion of the new trap allows for storage of a population of ions during the analysis of a previously
Available online 14 March 2016
stored population of ions. By accumulating and analyzing ions in parallel, rather than in series, a duty
cycle of 100% can be achieved. The new design retains the flexibility of the original design—the ability to
Keywords:
exchange mobility range, resolution, and speed/duty cycle—while in all instances, maintaining superior
TIMS
ion utilization efficiency. When operating at 100% duty cycle, the spectrometer is shown to trap a broad
Ion mobility-mass spectrometry
Duty cycle
range of ions (m/z 622–2722) with an average trapping efficiency of ∼70%. Trapping efficiency was found
Parallel accumulation to be constant over a range of accumulation/analysis times from 20 up to 85 ms where an average resolv-
Charge capacity ing power of 60 was achieved. Compared to transmission mode (MS-only) operation, the combination
of high mobility resolving power, high ion utilization efficiency, and relatively fast scan times achieved
with TIMS is shown to result in an improvement in mass spectral peak intensities of more than an order
of magnitude.
© 2016 Elsevier B.V. All rights reserved.
http://dx.doi.org/10.1016/j.ijms.2016.03.004
1387-3806/© 2016 Elsevier B.V. All rights reserved.
J.A. Silveira et al. / International Journal of Mass Spectrometry 413 (2017) 168–175 169
back of a drift tube has also improved the efficiency of ion injection vacuum stage of a prototype impact ESI-QqTOF (Bruker Daltonics,
as well as the collection of the radially diffuse ion swarm [37]. Billerica, MA) mass spectrometer. As depicted in Fig. 1a, the appa-
A subsequent ion funnel design featured a dedicated trapping ratus is comprised of a set of electrodes that form three regions:
region for enhanced utilization efficiency when used to pulse ions the entrance funnel, the TIMS tunnel, and the exit funnel. Ions and
into a drift tube [38]. Still, drift tube IMS demands that the initial N2 (g) are introduced to the first vacuum stage through the capil-
temporal width of the ion packet be very small with respect to the lary. During operation, ions are deflected into the entrance funnel
drift time (∼1%) to achieve high resolving power (R ≥ 50). For typical while gas may be pumped through a port positioned opposite to
drift times of ∼50 ms, this limits injection times to a few tenths of the capillary exit. Alternatively, during TIMS operation, gas flow
milliseconds and places a major constraint on the instrument duty through the port is partially or completely restricted such that
cycle. Though proof-of-principle temporal multiplexing strategies gas is diverted through the TIMS tunnel and is pumped through
have been shown to increase the ion throughput [39], challenges a secondary port located in the exit funnel region. Specific details
due to peak aliasing and the finite duty cycle limit of ∼50% remain. of the first-generation TIMS instrumentation can be found in the
In addition, temporal multiplexing strategies eliminate the ability literature [45,46].
to perform LC-IMS-MS and LC-IMS-MS/MS due to the need to recon- The second-generation TIMS funnel, featuring a 96 mm-long
struct the IMS-MS dataset over a complete multiplexing sequence. tunnel, was tested on a prototype microTOF (Bruker Daltonics, Bil-
A more recent report described the use of helium (as an alternative lerica, MA) mass spectrometer. In this design, the entrance and exit
to nitrogen) in the trapping region [40]. Owing to the greater mobil- funnel regions are identical to the original design; however, the
ity of ions in a lighter gas, an increase in the charge density of the tunnel length is increased by roughly a factor of two over that of
injected ion packet was observed, resulting in a sensitivity improve- the first-generation prototype (46 mm). Doubling the length of the
ment of more than an order of magnitude. Though more complete tunnel allows for the inclusion of a new “ion accumulation trap”
ejection of the accumulated ions was reported when using helium, upstream of the analysis region. The accumulation trap enables
the IMS duty cycle (1.6–33%) was still far from 100% since the capac- a mode of operation wherein incoming ions from the entrance
ity of the ion funnel/trap was not sufficient to store ions for the funnel can be accumulated, while simultaneously, a previously
entire duration of the 60 ms drift time. accumulated population of ions is mobility-analyzed in the “anal-
Introduced in 2011 by Park and coworkers, trapped ion mobility ysis region”. Inasmuch as ions are continuously accumulated, this
spectrometry (TIMS) was developed as a highly versatile alterna- mode of operation eliminates the need to block ions at the deflec-
tive to conventional drift tube IMS [41–43]. The general principles tion plate and, by definition, allows for a duty cycle of 100%.
of TIMS [44] and a theory [45,46] have been recently reported. In The second-generation TIMS funnel has three potential modes of
TIMS, ions are accumulated and stored according to their mobil- operation: (1) MS-only/transmission mode wherein the DC fields
ity in a manner analogous to the condensed phase electrophoretic and gas flow continuously guide the continuous ion beam to the
technique termed “electric field gradient focusing”. However, TIMS detector, (2) serial TIMS operation wherein ions are sequentially
also contains an elution step wherein ions are released and sepa- accumulated and analyzed, and (3) parallel accumulation TIMS
rated along an electric field gradient (EFG) plateau. The complete operation wherein ion accumulation and analysis occur in parallel.
analysis sequence occurs inside an ion funnel located in the first In MS-only mode, ions are simply transmitted from the entrance
vacuum stage of the instrument. Among other variables, the resolv- funnel to the exit funnel and subsequently, downstream to the
ing power has been shown to be dependent on the scan rate of the mass spectrometer without mobility analysis. In serial TIMS mode,
electric field (ˇ) during the elution step, as well as the drag force act- ions are accumulated and mobility-analyzed sequentially, as previ-
ing on the ions. In agreement with theory, experimental conditions ously described [45]. However, with the second-generation design,
that maximize the work done on ions—longer scans that maxi- one may take advantage of the additional storage capacity afforded
mize the “effective path length” and/or elevated pressures/higher by the longer tunnel [49]. Finally, in parallel accumulation mode,
gas velocities that increase the drag force—have yielded resolving one may achieve a 100% duty cycle, as outlined above. How-
power up to ∼230 for singly charged ions, ∼270 for doubly charged ever, by using the deflection plate to block ions for part of the
peptide ions [47] and ∼300 for multiply charged protein ions [48]. experimental cycle, it is also possible to employ longer analysis
Though exceptional performance has been demonstrated, thus times (while maintaining relatively short accumulation times to
far, TIMS has operated on the basis of an experimental sequence avoid charge capacity issues) when high resolution IMS analysis is
wherein ions are sequentially accumulated and analyzed. desired.
In the present work, we report a new spectrometer and The entrance and exit funnel regions of the second-generation
operational scheme that significantly improves the overall TIMS- spectrometer were pumped using a Roots-type multistage dry vac-
TOF MS instrument performance. The new design features a uum pump (∼600 m3 /h, Alcatel, 601B Roots Blower) backed by an
tunnel—roughly twice the length of the original design—that con- Ebara SA-20 mechanical pump that yielded a pressure of 3.1 and
tains an ion accumulation trap upstream of the analysis region. 0.9 mbar in each respective region. The pressure difference gener-
Spatially decoupling the accumulation from the analysis region ated across the tunnel has been shown to yield gas velocities on
allows the events to occur simultaneously, thus enabling TIMS the order of ∼150 m/s [46]. Ions exiting the TIMS funnel are guided
operation at 100% duty cycle. As demonstrated herein, operating into the TOF accelerator operated at ∼11 Hz. Both instruments were
the new spectrometer in this manner yields high TIMS resolving equipped with an electrospray ionization (ESI) source operated
power, high ion utilization efficiencies, and marked improvements with a 5 L/min N2 (g) flow rate. For all experiments, the shape of
in the TOF signal intensities, making this approach particularly well the EFG rising edge with respect to position followed a square root
suited for high throughput analyses and other applications requir- function that has been previously shown to yield equivalent axial
ing high sensitivity. storage space across the mobility range [49].
To characterize the instrument performance, a 1 pmol/L tryp-
tic digest of bovine serum albumin (BSA, Protea Biosciences,
2. Experimental Morgantown, WV) was directly infused at ∼150 L/min from a
50:50 mixture of water:acetonitrile containing 0.1% formic acid.
A schematic representation of the first-generation spectrome- In addition, low concentration tuning mix (Agilent Technologies,
ter, its operation, and the coordinate system referred to herein is Santa Clara, CA) was used to determine the resolving power by Eq.
shown in Fig. 1a. The TIMS funnel was incorporated into the first (1), efficiency by Eq. (2), and signal enhancement factor by Eq. (3)
170 J.A. Silveira et al. / International Journal of Mass Spectrometry 413 (2017) 168–175
Fig. 1. Comparison of the original 46 mm-long tunnel and serial analysis scheme (a) with the 96 mm-long TIMS tunnel and parallel accumulation scheme (b). In (c), the duty
cycle is plotted as a function of accumulation time for serial (purple line) and parallel accumulation schemes (dotted blue line). In the serial mode of operation, the sum of
the elution and trap steps were assumed to be 50 ms. Note that the transfer step in (b) is very small (<1%) compared to the accumulation/elution step. Furthermore, ions
entering the accumulation trap during the transfer step are immediately transmitted to the analyzer, thus allowing for a duty cycle which is truly 100%.
for TIMS experiments where a broad mass range (m/z 622–2722) funnel/tunnel regions by pulsing the potential of the deflector lens
was analyzed: to an attractive potential. During this time, trapped ions acquire
an equilibrium position along the rising edge and are held for a
R = K/K (1)
user-defined time period, typically, on the order of a few milli-
SignalTIMS 1 seconds. During the elution step, the magnitude of the EFG profile
Efficiency = 100% · · (2) is decreased from an initial value at a fixed scan rate such that
SignalTrans Duty Cycle
ions of progressively higher K elute from the analyzer. Given that
ITIMS scanning the EFG at a slower rate is known to improve the resolv-
Enhancement factor = (3)
ITrans ing power, the ability to vary the time during the accumulation
In Eqs. (1)–(3), R is the resolving power and K is the ion mobility and scanning steps makes it possible to trade speed/duty cycle for
coefficient. ITIMS and ITrans are taken to be the maximum intensity resolving power.
observed for a species during TIMS and the constant intensity of
a species contained in the continuous beam in MS-only experi-
ments, respectively. It is important to recognize that efficiency, as 3.2. Theoretical charge capacity estimation
defined herein, is a measure of the complete system (particularly
with respect to tuning of the entrance funnel, TIMS tunnel, and exit It is, of course, natural that any device used to store ions should
funnel) when operating in parallel accumulation TIMS mode versus have a limited charge capacity. Therefore, part of our effort in
the system as a whole in MS-only mode. SignalTIMS , used for cal- understanding TIMS must be to determine an approximate charge
culating the efficiency, is taken to be the area under an IMS-MS capacity and the consequences of exceeding this finite limit. In
peak whereas SignalTrans is the chromatographic area observed in considering the operation of Paul traps, one may consider the stor-
MS-only mode during the time required for a TIMS experiment. age capacity (i.e., the maximum number of charges that may be
confined in the physical dimensions of the trap) and the analyt-
3. Results and discussion ical capacity (i.e., the maximum number of charges that may be
confined without substantially degrading analytical performance).
3.1. Ion storage in TIMS Furthermore, it is known that in Paul traps, the distribution of
charges with respect to m/z has an influence on relative perfor-
As depicted in Fig. 1, a feature of TIMS is that the trapping posi- mance as a function of mass. That is, if the ion density is relatively
tion for a particular ion is related to the ion mobility coefficient. The high in a narrow m/z range, the performance of the analyzer in
drag force exerted by the gas flow on an ion in the TIMS tunnel is a that particular m/z range will be degraded relative to the per-
simple function of the ion’s mobility, qvg /K. Also, the counteracting formance outside of that range [50]. Similar statements can be
force from the DC electric field is a function of the axial position made with respect to the TIMS device. In the present work, we
(proportional to qz1/2 ) as implied by the plot in Fig. 1. Accordingly, emphasize the analytically useful charge capacity of the TIMS ana-
ions of low K are dragged by the gas to an axial position further into lyzer.
the tunnel where the strength of the electric field on the rising edge If one considers a line of charge trapped in a TIMS analyzer, an
is greater. Conversely, ions of higher K are trapped at a position near associated electric field strength can be estimated based on funda-
the tunnel entrance where the strength of the electric field on the mental principles. Consider that the rising edge of the EFG profile
rising edge is relatively weaker. (where the ions are stored according to their mobility) is ∼23 mm
A serial analysis sequence (see Fig. 1a) is comprised of accu- long and the plateau region (over which the ions must “elute”) is
mulation, blocking, and elution steps. The second blocking step ∼18 mm in length [46]. As an approximation, we therefore assume
involves preventing additional ions from entering the entrance 106 charges, stored with uniform density in a line 23 mm in length,
J.A. Silveira et al. / International Journal of Mass Spectrometry 413 (2017) 168–175 171
Fig. 4. (a) Average resolving power for tuning mix ions (m/z 622–2722) versus accumulation/analysis time for operation of the 96 mm-long tunnel at 100% duty cycle. The
dotted line represents an extrapolation of the separation performance expected based on fourth-root dependence of the data acquired at relatively short accumulation/analysis
times (≤13 ms). The average TIMS peak area versus accumulation/analysis time is shown in (b). In (c) and (d), the duty cycle was varied using a fixed analysis time of 120 ms to
study the signal intensity (c) and the resolving power (d). Panel (e) contains a 2D TIMS-MS plot acquired on the 96 mm-long tunnel using the parallel accumulation/analysis
scheme operating at 100% duty cycle (120 ms accumulation/analysis time).
significant dependence when the experiment was performed at varied timescales indicated that (1) at 85 ms, the total charge capac-
accumulation/analysis times between 20 and 85 ms. Rather, the ity of the trap has not yet been reached and (2) that m/z-dependent
average trapping efficiency was found to be ∼70% for all timescales stability in the trap is not the primary factor that contributes to
investigated. That the efficiency remains constant for each m/z at ion losses. Rather, the m/z dependence observed in Fig. 5a is likely
174 J.A. Silveira et al. / International Journal of Mass Spectrometry 413 (2017) 168–175
4. Conclusions
[7] S. Trimpin, D.E. Clemmer, Ion mobility spectrometry/mass spectrometry snap- [31] F. Meier, S. Beck, N. Grassl, M. Lubeck, M.A. Park, O. Raether, M. Mann, Parallel
shots for assessing the molecular compositions of complex polymeric systems, accumulation-serial fragmentation (PASEF): multiplying sequencing speed and
Anal. Chem. 80 (23) (2008) 9073–9083. sensitivity by synchronized scans in a trapped ion mobility device, J. Proteome
[8] F.A. Fernandez-Lima, C. Becker, K. Gillig, W.K. Russell, M.A.C. Nascimento, D.H. Res. 14 (12) (2015) 5378–5387.
Russell, Experimental and theoretical studies of (CsI)n Cs+ cluster ions pro- [32] P. Benigni, C.J. Thompson, M.E. Ridgeway, M.A. Park, F. Fernandez-Lima, Tar-
duced by 355 nm laser desorption ionization, J. Phys. Chem. A 112 (44) (2008) geted high-resolution ion mobility separation coupled to ultrahigh-resolution
11061–11066. mass spectrometry of endocrine disruptors in complex mixtures, Anal. Chem.
[9] C. Bleiholder, N.F. Dupuis, T. Wyttenbach, M.T. Bowers, Ion mobility-mass 87 (8) (2015) 4321–4325.
spectrometry reveals a conformational conversion from random assembly to [33] S.A. Shaffer, K. Tang, G.A. Anderson, D.C. Prior, H.R. Udseth, R.D. Smith, A
-sheet in amyloid fibril formation, Nat. Chem. 3 (2) (2011) 172–177. novel ion funnel for focusing ions at elevated pressure using electrospray ion-
[10] J.A. Silveira, K.A. Servage, C.M. Gamage, D.H. Russell, Cryogenic ion mobility- ization mass spectrometry, Rapid Commun. Mass Spectrom. 11 (16) (1997)
mass spectrometry captures hydrated ions produced during electrospray 1813–1817.
ionization, J. Phys. Chem. A 117 (5) (2013) 953–961. [34] S.A. Shaffer, D.C. Prior, G.A. Anderson, H.R. Udseth, R.D. Smith, An ion funnel
[11] K.A. Servage, J.A. Silveira, K.L. Fort, D.H. Russell, Evolution of hydrogen bond interface for improved ion focusing and sensitivity using electrospray ioniza-
networks in protonated water clusters H+ (H2 O)n (n = 1 to 120) studied by tion mass spectrometry, Anal. Chem. 70 (19) (1998) 4111–4119.
cryogenic ion mobility-mass spectrometry, J. Phys. Chem. Lett. 11 (2014) [35] J.B. Fenn, M. Mann, C.K. Meng, S.F. Wong, C.M. Whitehouse, Electrospray ion-
1825–1830. ization for mass spectrometry of large biomolecules, Science 246 (1989) 64–71.
[12] K.A. Servage, K.L. Fort, J.A. Silveira, L. Shi, D.E. Clemmer, D.H. Russell, Unfolding [36] M. Karas, F. Hillenkamp, Laser desorption ionization of proteins with molecular
of hydrated alkyl diammonium cations revealed by cryogenic ion mobility- masses exceeding 10,000 Daltons, Anal. Chem. 60 (20) (1988) 2299–2301.
mass spectrometry, J. Am. Chem. Soc. 137 (28) (2015) 8916–8919. [37] K. Tang, A.A. Shvartsburg, H.N. Lee, D.C. Prior, M.A. Buschbach, F. Li, A.V.
[13] T. Wyttenbach, G. von Helden, M.T. Bowers, Gas-phase conformation of biolog- Tolmachev, G.A. Anderson, R.D. Smith, High-sensitivity ion mobility spectrome-
ical molecules: Bradykinin, J. Am. Chem. Soc. 118 (35) (1996) 8355–8364. try/mass spectrometry using electrodynamic ion funnel interfaces, Anal. Chem.
[14] B.T. Ruotolo, J.L.P. Benesch, A.M. Sandercock, S.J. Hyung, C.V. Robinson, Ion 77 (10) (2005) 3330–3339.
mobility-mass spectrometry analysis of large protein complexes, Nat. Protoc. [38] B.H. Clowers, Y.M. Ibrahim, D.C. Prior, W.F. Danielson, M.E. Belov, R.D. Smith,
3 (7) (2008) 1139–1152. Enhanced ion utilization efficiency using an electrodynamic ion funnel trap
[15] M.F. Bush, Z. Hall, K. Giles, J. Hoyes, C.V. Robinson, B.T. Ruotolo, Collision cross as an injection mechanism for ion mobility spectrometry, Anal. Chem. 80 (3)
sections of and their complexes: a calibration framework and database for gas- (2008) 612–623.
phase structural biology, Anal. Chem. 82 (22) (2010) 9557–9565. [39] M.E. Belov, M.A. Buschbach, D.C. Prior, K. Tang, R.D. Smith, Multiplexed ion
[16] N.A. Pierson, L. Chen, S.J. Valentine, D.H. Russell, D.E. Clemmer, Number of mobility spectrometry-orthogonal time-of-flight mass spectrometry, Anal.
solution states of Bradykinin from ion mobility and mass spectrometry mea- Chem. 79 (6) (2007) 2451–2462.
surements, J. Am. Chem. Soc. 133 (35) (2011) 13810–13813. [40] Y.M. Ibrahim, S.V.B. Garimella, A.V. Tolmachev, E.S. Baker, R.D. Smith, Improving
[17] T. Wyttenbach, M.T. Bowers, Structural stability from solution to the gas phase: ion mobility measurement sensitivity by utilizing helium in an ion funnel trap,
native solution structure of ubiquitin survives analysis in a solvent-free ion Anal. Chem. 86 (11) (2014) 5295–5299.
mobility-mass spectrometry environment, J. Phys. Chem. B 115 (42) (2011) [41] M.A. Park, Apparatus and method for parallel flow ion mobility spectrometry
12266–12275. combined with mass spectrometry, US Patent 8,288,717 (2010).
[18] Z. Hall, A. Politis, M.F. Bush, L.J. Smith, C.V. Robinson, Charge-state dependent [42] F. Fernandez-Lima, D.A. Kaplan, M.A. Park, Integration of trapped ion mobil-
compaction and dissociation of protein complexes: insights from ion mobility ity spectrometry with mass spectrometry, Rev. Sci. Instrum. 82 (2011)
and molecular dynamics, J. Am. Chem. Soc. 134 (7) (2012) 3429–3438. 126106.
[19] E.R. Schenk, M.E. Ridgeway, M.A. Park, F. Leng, F. Fernandez-Lima, Isomerization [43] F. Fernandez-Lima, D. Kaplan, J. Suetering, M. Park, Gas-phase separation using
kinetics of AT hook decapeptide solution structures, Anal. Chem. 86 (2) (2013) a trapped ion mobility spectrometer, Int. J. Ion Mobil. Spectrom. 14 (2–3) (2011)
1210–1214. 93–98.
[20] L. Chen, S.H. Chen, D.H. Russell, An experimental study of the solvent- [44] D.R. Hernandez, J.D. DeBord, M.E. Ridgeway, D.A. Kaplan, M.A. Park, F.
dependent self-assembly/disassembly and conformer preferences of grami- Fernandez-Lima, Ion dynamics in a trapped ion mobility spectrometer, Analyst
cidin A, Anal. Chem. 85 (16) (2013) 7826–7833. 139 (2014) 1913–1921.
[21] S. Warnke, C. Baldauf, M.T. Bowers, K. Pagel, G. von Helden, Photodissociation of [45] K. Michelmann, J. Silveira, M. Ridgeway, M. Park, Fundamentals of trapped ion
conformer-selected ubiquitin ions reveals site-specific cis/trans isomerization mobility spectrometry, J. Am. Soc. Mass Spectrom. 26 (1) (2015) 14–24.
of proline peptide bonds, J. Am. Chem. Soc. 136 (29) (2014) 10308–10314. [46] J.A. Silveira, K. Michelmann, M.E. Ridgeway, M. Park, Fundamentals of trapped
[22] K.J. Pacholarz, P.E. Barran, Distinguishing loss of structure from subunit dis- ion mobility spectrometry. Part II. Fluid dynamics, J. Am. Soc. Mass Spectrom.
sociation for protein complexes with variable temperature ion mobility mass (2016), http://dx.doi.org/10.1007/s13361-015-1310-z (in press).
spectrometry, Anal. Chem. 87 (12) (2015) 6271–6279. [47] J.A. Silveira, M.E. Ridgeway, M.A. Park, High resolution trapped ion mobility
[23] J.C. May, J.A. McLean, A uniform field ion mobility study of melittin and impli- spectrometery of peptides, Anal. Chem. 86 (12) (2014) 5624–5627.
cations of low-field mobility for resolving fine cross-sectional detail in peptide [48] M.E. Ridgeway, J.A. Silveira, J.E. Meier, M.A. Park, Microheterogeneity within
and protein experiments, Proteomics 15 (16) (2015) 2862–2871. conformational states of ubiquitin revealed by high resolution trapped ion
[24] W. Cui, H. Zhang, R.E. Blankenship, M.L. Gross, Electron-capture dissociation mobility spectrometry, Analyst 140 (20) (2015) 6964–6972.
and ion mobility mass spectrometry for characterization of the hemoglobin [49] M.A. Park, N. Park, J.A. Meier, M.E. Ridgeway, J.A. Silveira, Equal opportunity ion
protein assembly, Protein Sci. 24 (8) (2015) 1325–1332. storage: electric field optimization for complex mixture analysis by trapped ion
[25] S.J. Allen, K. Giles, T. Gilbert, M.F. Bush, Ion mobility mass spectrometry of pep- mobility spectrometry, in: Proceedings from the 63rd Annual American Society
tide, protein, and protein complex ions using a radio-frequency confining drift for Mass Spectrometry Conference, St. Louis, MO, May 31–June 4, 2015.
cell, Analyst 141 (3) (2016) 884–891. [50] D.A. Kaplan, R. Hartmer, A. Brekenfeld, J. Franzen, M. Schubert, An examination
[26] S. Tenzer, A. Moro, J. Kuharev, A.C. Francis, L. Vidalino, A. Provenzani, P. Macchi, of the physics of the high capacity trap, in: R.E. March, J.F.J. Todd (Eds.), Practical
Proteome-wide characterization of the RNA-binding protein RALY-interactome Aspects of Trapped Ion Mass Spectrometry, CRC Press, Boca Raton, FL, 2010, pp.
using the in vivo-biotinylation-pulldown-quant (iBioPQ) approach, J. Proteome 593–617.
Res. 12 (6) (2013) 2869–2884. [51] D. Halliday, R. Resnick, Physics: Part 2, John Wiley and Sons, New York, NY,
[27] V. Parviainen, S. Joenvaara, N. Tohmola, R. Renkonen, Label-free mass spec- 1960.
trometry proteome quantification of human embryonic kidney cells following [52] A.V. Tolmachev, H.R. Udseth, R.D. Smith, Modeling the ion density distribution
24 hours of sialic acid overproduction, Proteome Sci. 11 (1) (2013) 11–38. in collisional cooling RF multipole ion guides, Int. J. Mass Spectrom. 222 (2003)
[28] R.P. Dator, K.W. Gaston, P.A. Limbach, Multiple enzymatic digestions and ion 155–174.
mobility separation improve quantification of bacterial ribosomal proteins [53] K. Michelmann, J. Bossmeyer, M.E. Ridgeway, J.A. Silveira, M. Mann, M.A. Park,
by data independent acquisition liquid chromatography–mass spectrometry, In improved capacity trapped ion mobility spectrometry, in: Proceedings from
Anal. Chem. 86 (9) (2014) 4264–4270. the 63rd Annual American Society for Mass Spectrometry Conference, St. Louis,
[29] S. Helm, D. Dobritzsch, A. Rodiger, B. Agne, S. Baginsky, Protein identifica- MO, May 31–June 4, 2015.
tion and quantification by data-independent acquisition and multi-parallel [54] J.C. May, C.R. Goodwin, N.M. Lareau, K.L. Leaptrot, C.B. Morris, R.T. Kurulugama,
collision-induced dissociation mass spectrometry (MSE ) in the chloroplast A. Mordehai, C. Klein, W. Barry, E. Darland, G. Overney, K. Imatani, G.C. Stafford,
stroma proteome, J. Proteomics 98 (2014) 79–89. J.C. Fjeldsted, J.A. McLean, Conformational ordering of biomolecules in the gas
[30] D. Helm, J.P.C. Vissers, C.J. Hughes, H. Hahne, B. Ruprecht, F. Pachl, A. Grzyb, K. phase: nitrogen collision cross sections measured on a prototype high reso-
Richardson, J. Wildgoose, S.K. Maier, H. Marx, M. Wilhelm, I. Becher, S. Lemeer, lution drift tube ion mobility-mass spectrometer, Anal. Chem. 86 (4) (2014)
M. Bantscheff, J.I. Langridge, B. Kuster, Ion mobility tandem mass spectrome- 2107–2116.
try enhances performance of bottom-up proteomics, Mol. Cell. Proteomics 13
(2014) 3709–3715.