SCAD 500L

DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL

No 632 ; 22nd Main


4th “T” Block ; Jayanagar
Bangalore – 560 041

# TEL:- 001-91-80 – 2245 4993 # FAX:- 266 55 333

E-MAIL:- pyrodynamics@bsnl.in
WEB:- www.pyrodynamics-india.com
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Table of Contents:-

1) General

2) Specifications

3) Input power

4) Front panel controls & wiring diagrams

5) Gain adjustment

6) Calibration

7) RS232 Interface

8) Flow chart.

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

1) General:-

SCAD500L is a two Channel LVDT signal conditioner.

SCAD 500L uses a 128*64 pixel Graphics LCD display


On Line Data can be displayed in the following formats:-
- ASCII Format of 2 channels simultaneously.
- Graphical Display of user selectable any one channel.
- Peak Hold – Max / Min of 2 channels.

The data can be viewed on line in ASCII or Graphical Format during the experiment /
test. On completion of the test the user can go into the PEAK HOLD Menu and view Max
/ Min data of 2 channels simultaneously.

Voltages can be converted into user defined engineering units via the ax2+bx+c scaling
option.

All the menu settings and the Bridge Balance is stored in a built in NVRAM.
Menu settings can be password protected.

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

2) Specifications:-
EXCITATION 5Vrms,5kHz
VOLTAGE
GAIN variable using external potentiometer
FULL SCALE +/- 10 VDC
ANALOG OUTPUT
DISPLAY 128x64 Pixel Graphics LCD
Simultaneous Display of 2 channels in mm or user
Defined engineering units and Current Date and
Time.
ENGINEERING Each Channel can be set for user defined
UNITS engineering units.
DISPLAY UPDATE 1 sample per second per channel.
RATE
INPUT POWER +/- 15VDC & +5VDC. 110 – 230 VAC, 50 /60 Hz
Adapter supplied with the system.

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

3) Input Power:-
Input Power: - 100 – 240VAC; 47-63 Hz; Use only the supplied AC Adapter

AC INPUT

Green Light on the


AC Adapter
indicates that the
AC Adapter is
Switched on

** Ensure that there is


Proper Ground.
** Always ensure that the
SCAD 500L is switched off
Prior to connecting the
AC Adapter.

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

4) Front panel controls & wiring diagrams:-


SCAD 500L uses a 128*64 pixel Graphics LCD display. The Display can be programmed
via Front Panel Key Board.

Press the SHIFT key to access the Menu’s and


Sub Menu’s.

Press the ENTER key to move to the next menu


in the main or the Sub menu or to the next line
in the sub menu.

Press the EXIT key to come out of a particular


menu.

Press the UP key for the cursor to move up in


The main and sub menu.

Press the DOWN key for the cursor to move


Down in the main and Sub menu.

Press the RIGHT key for the cursor to shift to the next position to the right.

DECIMAL, AUTOZERO, CH ON/OFF, PEAK HOLD, GRAPH, and MENU are shortcut keys
use by the SHIFT key.

Two external potentiometers for gain settings of CH A and CH B.

LVDT Connector Type A:-


PIN CONFIGURATION
1 Excitation input 1
2 Ground(Center tap)
3 Excitation input 2
4 Secondary output 1
5 Secondary output 2
Female

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

LVDT Connector Type B:-


PIN CONFIGURATION
1 Excitation input 1
2 Secondary output 1
3 Ground(Center tap)
4 Secondary output 2
5 Excitation input 2
Female

Analog output Pin configuration:-


PIN CONFIGURATION COLOUR
1 Vout1(LVDT type A) RED
2 Vout2(LVDT type B) GREEN
3 Ground BLACK

Female

5) Gain adjustment:-
Set the LVDT position to maximum linear range (+/-2.5mm) and vary the GAIN
pot till the required range (+/-10v etc.,).

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

6) Calibration:-

There are two methods of calibration:-

1. Manual Calibration:
Fix the LVDT to be calibrated in the Calibration Test Rig as shown below

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Connect the LVDT to SCAD 500 L Displacement Measurement System and


Power on the system

* SCAD 500 L *
Initializing….……………
NML Ver 2010

Ch L1 +00008.842 mm

Ch L2 +00005.714 mm

24 / 02 / 10 01 : 36 : 58

Press SHIFT Key

SHIFT

Ch L1 +00008.842 mm PASS WORD


CHANGE PW
Ch L2 +00005.714 mm RE-INITIALIZE
CLOCK

24 / 02 /10 01 : 36 : 58

ENTER FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

PASS WORD
CHANGE PW Enter Password
RE-INITIALIZE 00000.0000
CLOCK

EXCITATION CAL SETTINGS


CALIBRATION AUTO CALIBRATE
PEAK HOLD & GRAPH +y = ax^2 + bx + c
SETTINGS
- y = ax^2 + bx + c

+y = ax^2 + bx + c

a Are you sure you want to


b Clear Previous
c a, b & c values Y/N ? N
CLEAR a, b & c

Press UP button

Are you sure you want to


Clear Previous
a, b & c values Y/N ? Y
Then ENTER , EXIT,EXIT…data menu.

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Now set the zero position of LVDT by rotating the micrometer head, the analog output
can be monitored on LCD or using Analog output terminals.

Move the LVDT by 2.5 mm –ve side and adjust the gain (vary GAIN A/B pot till the
output voltage becomes -9.999V.)

Take the analog output readings of corresponding displacements by rotating micro meter
as shown below and plot the graph in excel sheet.

There are two plots here: - 1. –ve values (-2.5mm to 0)


2. +ve values (0 to +2.5mm)
nd
Add trend line and get 2 order polynomial equations….

The equations are in the form of Y=ax^2+bx+c.

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Now these equations can directly enter in to SCAD500L.

a L1 +00000.0014
a L2 +00000.0000

b L1 +00000.2635
b L2 +00001.0000

-y = ax^2 + bx + c

a
b
c
CLEAR a, b & c

c L1 -00000.0012
Press UP & DOWN Keys to change the Value, c L2 +00000.0000
LEFT & RIGHT Keys to shift the Position

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

a L1 -00000.0011
a L2 +00000.0000

b L1 +00000.2639
b L2 +00001.0000

+y = ax^2 + bx + c

a
b
c
CLEAR a, b & c

c L1 -00000.0010
Press UP & DOWN Keys to change the Value,
LEFT & RIGHT Keys to shift the Position
c L2 +00000.0000

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

2. Auto Calibration:

Fix the LVDT to be calibrated in the Calibration Test Rig as shown below

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Connect the LVDT to SCAD 500 L Displacement Measurement System and


Power on the system

* SCAD 500 L *
Initializing….……………
NML Ver 2010

Ch L1 +00008.842 mm

Ch L2 +00005.714 mm

24 / 02 / 10 01 : 36 : 58

Press SHIFT Key

SHIFT

Ch L1 +00008.842 mm PASS WORD


CHANGE PW
Ch L2 +00005.714 mm RE-INITIALIZE
CLOCK

24 / 02 /10 01 : 36 : 58

ENTER FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

PASS WORD
CHANGE PW Enter Password
RE-INITIALIZE 34567.0000
CLOCK

Set the Excitation Amplitude and


Frequency
to the LVDT to the desired values
EXCITATION
CALIBRATION
PEAK HOLD & GRAPH AMP FREQ
SETTINGS Ex L1 5 V AC 5 KHz
Ex L2 5 V AC 5 KHz

PAGE 2

CAL SETTINGS
AUTO CALIBRATE
+y = ax^2 + bx + c
- y = ax^2 + bx + c

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

ENTER FLOW

CALL SETTINGS
1) Go to Cal Settings Menu and Enter the No of cal point desired
2) Enter the Negative Displacement Range and Positive Displacement Ranges

Press UP & DOWN Keys to change


the Value, LEFT & RIGHT Keys to
shift the Position and ENTER Key for
next Ch Selection
CAL SETTINGS No of Cal Points L1 11
AUTO CALIBRATE - DISP L1 -00002.5000 mm
+y = ax^2 + bx + c +DISP L1 +00002.5000 mm
- y = ax^2 + bx + c No of Cal Points L2 11
- DISP L2 -00002.5000 mm
+DISP L2 +00002.5000 mm

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

AUTO CALIBRATE

Press UP & DOWN Keys for Select ion

CAL SELECTION
----------------------------------------------
CAL SETTINGS Ch L1
AUTO CALIBRATE Ch L2
+y = ax^2 + bx + c
- y = ax^2 + bx + c

Select the Particular Channel to


Calibrate

CAL POINTS COUNT 00


SET POINT -00002.500 mm At this stage find LVDT zero
by rotating micro meter, then
Ch L1 -00009.954 V
Rotate the micro meter –ve
side exactly -2.5mm from the
initial point (zero position)

CAL POINTS COUNT 01 Rotate micro meter anti-clockwise


SET POINT -00002.000 mm by 0.5mm, now our
Ch L1 -00007.935 V SET POINT is -2.000 mm
&
CAL POINT is 01

CAL POINTS COUNT 02 Rotate micro meter anti-clockwise


SET POINT -00001.500 mm
by 0.5mm, now our
Ch L1 -00005.873 V SET POINT is -1.500 mm
&
CAL POINT is 02

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Next Page
CAL POINTS COUNT 03
SET POINT -00001.000 mm Rotate micro meter anti-clockwise
Ch L1 -00003.947 V
by 0.5mm, now our
SET POINT is -1.000 mm
&
CAL POINT is 03

CAL POINTS COUNT 04


SET POINT -00000.500 mm Rotate micro meter anti-clockwise
by 0.5mm, now our
Ch L1 -00001.764 V
SET POINT is -0.500 mm
&
CAL POINT is 04

CAL POINTS COUNT 05


SET POINT +00000.000 mm
Rotate micro meter anti-clockwise
by 0.5mm, now our
Ch L1 +00000.000 V SET POINT is -0.000 mm
&
CAL POINT is 05

CAL POINTS COUNT 06 Rotate micro meter anti-clockwise


SET POINT +00000.500 mm
by 0.5mm, now our
Ch L1 +00001.934 V SET POINT is +0.500 mm
&
CAL POINT is 06

CAL POINTS COUNT 07


SET POINT +00001.000 mm Rotate micro meter anti-clockwise
by 0.5mm, now our
Ch L1 +00003.967 V
SET POINT is +1.000 mm
&
CAL POINT is 07

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Next Page

CAL POINTS COUNT 08


SET POINT +00001.500 mm Rotate micro meter anti-clockwise
by 0.5mm, now our
Ch L1 +00005.789 V
SET POINT is +1.500 mm
&
CAL POINT is 08

CAL POINTS COUNT 09


SET POINT +00002.000 mm Rotate micro meter anti-clockwise
by 0.5mm, now our
Ch L1 +00007.934 V
SET POINT is +2.000 mm
&
CAL POINT is 09

CAL POINTS COUNT 10 Rotate micro meter anti-clockwise


SET POINT +00002.500 mm by 0.5mm, now our
Ch L1 +00009.946 V
SET POINT is +2.500 mm
&
CAL POINT is 10

CAL SELECTION
---------------------------------------------- Now Select second LVDT
Ch L1 connected to other channel
Ch L2 for Calibration

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

The table below shows will give the idea of CAL POINT & SET POINT

CAL POINT SET POINT Ideal LVDT


(micro meter LVDT Readings
movement in mm) readings (Values given
below is only
for depiction)
00 -2.5 -10.000 -9.954
01 -2.0 -8.000 -7.935
02 -1.5 -6.000 -5.873
03 -1.0 -4.000 -3.947
04 -0.5 -2.000 -1.764
05 0 0.000 0.000
06 +0.5 +2.000 +1.934
07 +1.0 +4.000 +3.967
08 +1.5 +6.000 +5.789
09 +2 +8.000 +7.934
10 +2.5 +10.000 +9.946

After this the values b & c of +Y or –Y will generate automatically by SCAD500L

Note: Auto calibration is limited to 1st order polynomial (mx+b), to get better curve fit
use manual calibration i.e, 2nd order polynomial (ax^2+bx+c).

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

7) RS 232 Interface:-

The data can be interfaced to the host PC via the RS 232 Interface.

PLEASE NOTE THAT FOR STREAMING OF DATA, THE FRONT PANEL LCD SHOULD
BE IN DATA MODE AND THE DATA DISPLAYED ON THE LCD.

In case your PC does not have a RS 232 Port, install the USB to RS 232 Converter. The drivers
for the converter are available in the CD ROM.
To know the Port Settings, MY COMPUTER – CONTROL PANEL – SYSTEM – HARDWARE
–DEVICE MANAGER – HARDWARE – PORTS.

The Port should have the following settings:-


- Select the correct Com Port.

BAUD RATE 57600


DATA BITS 8
PARITY NONE
STOP BITS 1
FLOW CONTROL NONE

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

8) Flow chart:-

* SCAD 500 L *
Initializing….……………

Ch L1 +00008.842 mm

Ch L2 +00005.714 mm

24 / 02 / 10 01 : 36 : 58

Press SHIFT Key

SHIFT

Ch L1 +00008.842 mm PASS WORD


CHANGE PW
Ch L2 +00005.714 mm RE-INITIALIZE
CLOCK
24 / 02 /10 01 : 36 : 58

Press UP & DOWN keys to move the menu cursor -


Press ENTER and EXIT keys to navigate through menu’s

ENTER FLOW

SHIFT FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
Enter Password Web: www.pyrodynamics-india.com
E-mail: pyrodynamics@bsnl.in Enter Password
34567.0000 34578.0000
Invalid Entry

EXCITATION
CALIBRATION Press UP & DOWN Keys to change the Value,
PEAK HOLD & GRAPH LEFT & RIGHT Keys to shift the Position
SETTINGS

Enter old PW Enter old PW


34578.0000 34578.0000

Invalid Entry
Unauthorized User

PASS WORD
CHANGE PW
RE-INITIALIZE
CLOCK Enter old PW
34567.0000
Enter New PW
23456.0000

Ch L1 +00008.842 mm
* SCAD 500 L *
Ch L2 +00005.714 mm
Initializing….……………
24 / 02 / 10 01 : 36 : 58

Wed, Feb, 24, 2010


* SCAD 500 L *
01 : 36 : 58 PM
ENTER FLOW Initializing….……………
EXIT FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

AMP FREQ
Ex L1 5 V AC 5 KHz
Ex L2 5 V AC 5 KHz

CAL SETTINGS
AUTO CALIBRATE
EXCITATION +y = ax^2 + bx + c
CALIBRATION - y = ax^2 + bx + c
PEAK HOLD & GRAPH
SETTINGS

CLEAR P_H
PEAK HOLD
Y/T GRAPH
X/Y GRAPH

Ch L1 +00008.842 mm

Ch L2 +00005.714 mm

24 / 02 / 10 01 : 36 : 58

UNIT
DECIMAL DIGITS
CHANNEL ON/OFF
SET TIME

ENTER FLOW

EXIT FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Press UP & DOWN Keys for change the values,


LEFT & RIGHT Keys to shift the Position and
ENTER Key for next Ch selection
No of Cal Points L1 11
- DISP L1 -00002.5000 mm
+DISP L1 +00002.5000 mm

No of Cal Points L2 11
- DISP L2 -00002.5000 mm
+DISP L2 +00002.5000 mm

Press UP & DOWN Keys for Selection

CAL SELECTION
----------------------------------------------
Ch L1
CAL SETTINGS Ch L2
AUTO CALIBRATE
+y = ax^2 + bx + c
- y = ax^2 + bx + c

Press UP & DOWN Keys for Selection

+y = ax^2 + bx + c

a
b
c
CLEAR a, b & c

Press UP & DOWN Keys for Selection

-y = ax^2 + bx + c

a
b
c
CLEAR a, b & c

ENTER FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

AUTO CALIBRATE

Press UP & DOWN Keys for Select ion

CAL SELECTION
----------------------------------------------
CAL SETTINGS Ch L1
AUTO CALIBRATE Ch L2
+y = ax^2 + bx + c
- y = ax^2 + bx + c

CAL POINTS COUNT 00


SET POINT -00002.500 mm

Ch L1 -00009.954 V

CAL POINTS COUNT 01


SET POINT -00002.000 mm

Ch L1 -00007.935 V

CAL POINTS COUNT 02


SET POINT -00001.500 mm

Ch L1 -00005.873 V

Next Page

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

CAL POINTS COUNT 03 CAL POINTS COUNT 08


SET POINT -00001.000 mm SET POINT +00001.500 mm

Ch L1 -00003.947 V Ch L1 +00005.789 V

CAL POINTS COUNT 04 CAL POINTS COUNT 09


SET POINT -00000.500 mm SET POINT +00002.000 mm

Ch L1 -00001.764 V Ch L1 +00007.934 V

CAL POINTS COUNT 05 CAL POINTS COUNT 10


SET POINT +00000.000 mm SET POINT +00002.500 mm

Ch L1 +00000.000 V Ch L1 +00009.946 V

CAL POINTS COUNT 06 CAL SELECTION


SET POINT +00000.500 mm ----------------------------------------------
Ch L1
Ch L1 +00001.934 V Ch L2

CAL POINTS COUNT 07


SET POINT +00001.000 mm

Ch L1 +00003.967 V

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

a L1 +00000.0000
a L2 +00000.0000

b L1 +00001.0000
b L2 +00001.0000

+y = ax^2 + bx + c

a
b
c
CLEAR a, b & c

c L1 +00000.0000
c L2 +00000.0000

Press UP & DOWN Keys to change the Value,


LEFT & RIGHT Keys to shift the Position

Are you sure you want to


Clear Previous
a, b & c values Y/N ? N

ENTER FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

a L1 +00000.0000
a L2 +00000.0000

b L1 +00001.0000
b L2 +00001.0000

- y = ax^2 + bx + c

a
b
c
CLEAR a, b & c

c L1 +00000.0000
c L2 +00000.0000

Press UP & DOWN Keys to change the Value,


LEFT & RIGHT Keys to shift the Position

Are you sure you want to


Clear Previous
a, b & c values Y/N ? N

ENTER FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Press UP & DOWN Keys to Select YES or NO

Are you sure you want to


Clear Peak_Hold Y/N? N

Ch L1
PMAX +00008.953
P MIN - 00006.741
Ch L2
PMAX +00007.592
CLEAR P_H P MIN - 00005.356
0
PEAK HOLD
Y/T GRAPH
X/Y GRAPH

GRAPH SELECTION
----------------------------------------------
Ch L1
Ch L2

GRAPH SELECTION
-------------- Select X-AXIS -----------
Ch L1
Ch L2

GRAPH SELECTION
-------------- Select Y-AXIS -----------
Ch L1
Ch L2

ENTER FLOW

EXIT FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Press UP & DOWN Keys to change Y scale, LEFT & RIGHT Keys to change Time scale
Press SHIFT key to monitor PMAX & PMIN values
Press EXIT Key to Pause and ENTER Key to resume the Graph

+01024

Ch L1
+00003
Time
+00021
- 01024

+01024

Ch L2
+00003
Time
+00021
- 01024

GRAPH SELECTION
----------------------------------------------
Ch L1
Ch L2

ENTER FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Press UP & DOWN Keys to change Y scale,


LEFT & RIGHT Keys to change X scale
+00128

+00001
L1 / L2
+00004
T 0021
- 00128
- 00128

+00128
GRAPH SELECTION
-------------- Select X-AXIS ----------- +00001
Ch L1 L2 / L1
Ch L2 +00004
T 0021
- 00128
- 00128

GRAPH SELECTION
-------------- Select Y-AXIS -----------
Ch L1
Ch L2

ENTER FLOW

EXIT FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

Press UP & DOWN Keys to change the Value,


LEFT & RIGHT Keys to shift the Position and
ENTER Key to to move to next Selection

Select unit L1 mm
Select unit L2 mm

Press UP & DOWN Keys to set and


ENTER key to move to next setting

No of Decimals L1 3
No of Decimals L2 3

UNIT
DECIMAL DIGITS
CHANNEL ON/OFF
SET TIME

Press UP & DOWN Keys for selection

Select Ch L1 Enable
Select Ch L2 Enable

Press UP & DOWN Keys to set and


ENTER key to move to next setting

Day, Mon, DD, YYYY


Wed, Feb, 24, 2010
Hr : Mn : Se **
01 : 36 : 58 PM

ENTER FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL
PYRODYNAMICS * Ph: 91-80-2 245 4993; Fax: 91-80-2 66 55 333
E-mail: pyrodynamics@bsnl.in Web: www.pyrodynamics-india.com

SHIFT SHORT KEYS

SHIFT

Ch L1 * +00000.000 mm

Ch L2 * +00000.000 mm

24 / 02 / 10 01 : 36 : 58

Ch L1
PMAX +00008.953
P MIN - 00006.741
Ch L2
PMAX +00007.592
Press UP key P MIN - 00005.356
SHIFT
Press RIGHT Key Press RIGHT key to Clear
Ch L1 +00008.842 mm
Press DOWN Key
Ch L2 +00005.714 mm +00032

Press LEFT Key


24 / 02 / 10 01 : 36 : 58
Ch L1
Press EXIT Key +00003
Time
+00021
- 00032

Press UP & DOWN Keys for Selection

Select Ch L1 Enable
Select Ch L2 Enable

Press UP & DOWN Keys for Selection

No of Decimals L1 3
No of Decimals L2 3

ENTER FLOW

UP or DOWN or LEFT or RIGHT FLOW

SCAD500L DISPLACEMENT MEASUREMENT SYSTEM


OPERATIONAL MANUAL