CO 2 Incubator
HERAcell ® 150 i / 240 i
Trademarks
HERAcell® is a registered trademark of Thermo Scientific.
Thermo Scientific is a brand of Thermo Fisher Scientific Inc.
All trademarks mentioned in the service instructions are the exclusive property of
the respective manufacturers.
Contents
1 General notes 6
1.1 Identification of the device and of the documentation......................................... 7
1.2 Fundamental requirements for service work ........................................................ 7
1.3 Applicability of the instructions ............................................................................. 8
1.4 Warranty ................................................................................................................. 8
1.5 Explanation of safety information and symbols ................................................... 9
1.5.1 Safety information and symbols used in the operating instructions............. 9
1.5.2 Additional symbols for safety informations: .................................................. 9
1.5.3 Symbols on the device .................................................................................. 10
1.6 Overview safety notes ......................................................................................... 11
1.6.1 Safety instructions for repairs ....................................................................... 11
1.6.2 Safety notes on carbon dioxide (CO2) .......................................................... 13
1.6.3 Safety notes on oxygen (O2) ......................................................................... 13
1.6.4 Safety notes on nitrogen (N2) ........................................................................ 14
50117760 / 09.2008 3
®
Service Instructions | HERAcell 150 i / 240 i
Contents
50117760 / 09.2008 4
®
Service Instructions | HERAcell 150 i / 240 i
Figures
50117760 / 09.2008 5
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
50117760 / 09.2008 6
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
Please note that maintenance of the device must be carried out only by
adequately skilled personnel. Prior to repairs, please read these service
instructions carefully.
Skills of service personnel • The replacement of defined spare parts must be performed only by the
Technical Service of Thermo Fisher Scientific or by authorized service
personnel who have been trained by Thermo Fisher Scientific.
• Electrical work must be carried out only by a qualified electrician.
• Work to gas supply lines and to gas supply system components must
be carried out only by trained gas and air conditioning expert person-
nel.
Process procedures for personnel • Service works must be carried out in accordance with the applicable
national regulations.
Process procedures for personnel • For any operation of this device, the service personnel must prepare
clear and concise written instructions in the language of the operating
and cleaning personnel based on the operating instructions, applicable
safety data sheets, plant hygiene guidelines, and technical regulations,
in particular:
– which decontamination measures are to be applied for the CO2
incubator and accessories,
– which protective measures apply when specific agents are used,
– Iwhich measures are to be taken in the case of an accident.
• Repairs to the device must be carried out only by trained and authori-
zed expert personnel.
50117760 / 09.2008 7
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
Original document for translation The contents of the service instructions are subject to change without fur-
ther notice.
• Concerning translations into foreign languages, the German version of
these service instructions is binding.
• Keep these service instructions close to the device so that safety
instructions and important information are always accessible.
• Should you encounter problems that are not detailed adequately in
these service instructions, please contact Thermo Scientific immedia-
tely for your own safety.
1.4 Warranty
Requirements for warranty Thermo Scientific warrants the operational safety and functions of the CO2
incubator only under the condition that:
• the device is operated and serviced exclusively in accordance with its
intended purpose and as described in these operating instructions,
• the device is not modified,
• only original spare parts and accessories that have been approved by
Thermo Fisher Scientific are used,
• inspections and maintenance are performed at the specified intervals.
The warranty is valid from the date of delivery of the device to the operator.
50117760 / 09.2008 8
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
Harmful liquids!
Electric shock!
Hot surfaces!
Fire hazard!
Explosion hazard!
Suffocation hazard!
50117760 / 09.2008 9
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
50117760 / 09.2008 10
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
The general safety instructions point out potential hazards during repairs.
To prevent constant repetitions, the following safety instructions are not
mentioned explicitly in the individual sections; instead, the sections con-
tain only references to the general safety instructions.
Prior to repairs, read the safety instructions carefully.
For any repairs, the listed location requirements apply.
Contamination hazard!
The incubator can be used for the permanent or temporary
storage of toxic or infectious substances or materials. There-
fore, the device or device components may be contaminated.
Always observe the hygiene regulations of the operator!
• Infectious or toxic substances or residues thereof must
always be removed completely from the work space.
• Prior to repairs, make sure that the device has been cleaned
and disinfected to exclude infection hazards!
• The operator must provide a declaration of decontamina-
tion.
Electric shock!
Contact with current-carrying components may cause a lethal
electric shock.
Prior to repairs, disconnect the device from the power supply
system:
• Isolate the device electrically.
• Protect the device from accidental reconnection.
• Make sure the device is de-energized.
• Ground and short the device.
• Isolate adjacent components or components to repair that
are under voltage.
50117760 / 09.2008 11
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
Start-up:
If safety appliances were removed or disabled during repairs,
the device must not be started up before the safety appliances
have been reinstalled and checked for proper operation!
Before the incubator is started up and operated, a test run must
be performed.
Recycling:
All components of the device can be recycled.
If toxic or infectious substances were used in the device, the
components must be decontaminated as required before they
are discarded or shipped. For shipped components, a certificate
of decontamination must be attached.
50117760 / 09.2008 12
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
Suffocation hazard!
Large amounts of CO2 released into the room atmosphere may
cause suffocation.
If CO2 is released, initiate safety measures immediately!
• Leave the room immediately and do not allow others to
enter the room!
• Inform security service or fire department!
Oxygen explosion!
O2 may explode in combination with oils, greases, and
lubricants. If highly compressed oxygen comes in contact with
grease- or oil-containing substances, the mixture may explode!
• For cleaning these device components, use only oil- and
grease-free lubricants.
• Keep all connections and components of the oxygen system
free from substances that contain oil, grease, or lubricant!
Fire hazard!
Released oxygen promotes combustion. Do not use open
flames in the vicinity of oxygen-operated systems!
• Do not smoke in the vicinity of oxygen systems.
50117760 / 09.2008 13
®
Service Instructions | HERAcell 150 i / 240 i
1 General notes
Suffocation hazard!
Large amounts of N2 released into the room atmosphere may
cause suffocation.
If N2 is released, initiate safety measures immediately!
• Leave the room immediately and do not allow others to
enter the room!
• Inform security service or fire department!
50117760 / 09.2008 14
®
Service Instructions | HERAcell 150 i / 240 i
Side cutters
Cable stripper
Blade terminal pliers
Long nose pliers with cutter small
Forceps
Level small
Water pump pliers
Measuring devices:
The measuring devices listed below are recommended measu-
ring devices. You can also use measuring devices with compa-
rable specifications from other manufacturers.
For the use of the measuring devices, the manufacturer's
instructions are binding.
Temperature probe:
Supplier Thermo Fisher Scientific
ID No. 50117403
Specifications:
Sensor Cable resistance thermometer PT100
in 4-wire circuit
Operating temperature -80°C to +260°C
50117760 / 09.2008 15
®
Service Instructions | HERAcell 150 i / 240 i
2.2.3 CO 2 meter
Designation IR-CO2 gas tester
Supplier Thermo Fisher Scientific
ID No. 50060283
Specifications:
Measuring range 0 % to 60 % CO2
Uncertainty in measurement:
from 0% to 1% CO2 ± 0.05 % CO2
from 1% to 25% CO2 ± 5 % of measured value
from 25% to 60% CO2 ± 10 % of measured value
2.2.4 O 2 meter
Designation X-am 7000
Supplier Dräger AG
www.draeger.com
O 2 sensor:
Designation XS O2 1003
Specifications:
Measuring range max. 0 % to 100 Vol. % O2
Linearity error ≤ ±1 Vol. % to 40 Vol. % O2
≤ ± 2 Vol. % larger 40 Vol. % O2
Repeatability ≤ ±1 % of measured value
Digital display resolution 0.5 Vol. %
Ambient conditions:
Temperature, continuous 0 °C to 50 °C
Humidity, continuous 10 % to 90 % r.h.
Humidity, temporary 5 % to 99 % r.h.
Pressure 700 hPa to 1300 hPa
Pressure factor ≤ ±5 % of measured value / hPa
Humidity factor ≤ ±0.01 % of measured value / % r.h.
50117760 / 09.2008 16
®
Service Instructions Main Board | HERAcell 150 i / 240 i
50117760 / 09.2008 17
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 18
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 19
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 20
®
Service Instructions | HERAcell 150 i / 240 i
The wire cage [1] that covers the blower wheel [4] is
attached with two screws [5] to the measuring cell [3].
Each screw has a sealing ring that protects it from
getting lost.
Functional test
Fig. 4: Wire cage installation
After installation, check to see if the blower
wheel can rotate freely.
50117760 / 09.2008 21
®
Service Instructions | HERAcell 150 i / 240 i
Functional test
After installation, check to see of the
blower wheel can rotate freely.
50117760 / 09.2008 22
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 23
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 24
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 25
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 26
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 27
®
Service Instructions | HERAcell 150 i / 240 i
Procedure:
1. Request the desired submenu for device
configuration.
Renew calibrations:
The saving routine will overwrite the exi-
sting calibration values on the main board.
For a start-up of the device:
• Start device using auto-start routine.
• After this renew all calibrations and cor-
responding adjustments for:
- Temperature calibration,
- CO2 calibration,
- if required, O2 calibration.
50117760 / 09.2008 28
®
Service Instructions | HERAcell 150 i / 240 i
A Adjust size.
B Adjust material of interior container.
50117760 / 09.2008 29
®
Service Instructions | HERAcell 150 i / 240 i
C Adjust voltage.
D Switch gas tight screen on/off.
50117760 / 09.2008 30
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 31
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 32
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 33
®
Service Instructions | HERAcell 150 i / 240 i
contra-con setting:
contra-con must be activated (on) by
default. Otherwise the decontamination
routine cannot be started.
50117760 / 09.2008 34
®
Service Instructions | HERAcell 150 i / 240 i
Isothermal container:
Do not use a container filled with water
as isothermal container as the evaporation
of the water will result in a too low
temperature.
Excessive work space temperature:
Excessive temperature in the work space
after the calibration can be reduced
by opening the doors for approx. 30
seconds.
50117760 / 09.2008 35
®
Service Instructions | HERAcell 150 i / 240 i
5. Perform auto-start.
50117760 / 09.2008 36
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 37
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 38
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 39
®
Service Instructions | HERAcell 150 i / 240 i
3. Perform auto-start.
50117760 / 09.2008 40
®
Service Instructions | HERAcell 150 i / 240 i
7.3 O2 configuration
An O2 calibration is required:
• After an O2 sensor has been replaced,
• after change-over of the O2 sensor to another
control range.
7.4 O2 calibration
50117760 / 09.2008 41
®
Service Instructions | HERAcell 150 i / 240 i
Performing an O2 calibration:
For a comparison measurement (O2 calibration), a
calibrated measuring instrument should be used (see
chap. 2.2‚ pg. 15).
4. Perform auto-start.
50117760 / 09.2008 42
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 43
®
Service Instructions | HERAcell 150 i / 240 i
50117760 / 09.2008 44
®
Service Instructions | HERAcell 150 i / 240 i
Renew calibrations:
The saving routine will overwrite the existing calibration values
on the main board. For a start-up of the device renew all cali-
brations and corresponding adjustments.
50117760 / 09.2008 45
®
Service Instructions | HERAcell 150 i / 240 i
2. Calibrate pressure-sensitivity:
In the SETTINGS menu request the dialog box
TOUCH CALIBRATION [2].
Follow the instructions at the display and press
the displayed crosses [1] with a blunt pin
running two sensitivity test cycles.
50117760 / 09.2008 46
www.Thermo.com