Short Operation Instruction


eraflash & eraflash LT – Flash Point Testers
Turning on the analyzer ASTM D 6450
Position the analyzer properly on the bench. After This standard method requires the smaller sample cup
connecting the power cable to the back panel of the where only 1 mL must be filled. Fill the sample cup with a
analyzer, press the main power switch. pipette to the 1 mL mark and add a stirring magnet.
The analyzer will load the firmware. During this process,
the LED on the power button will flash. The loading takes Starting the measurement
approximately one minute. After loading, the analyzer will Put the sample cup into the opening of the front panel and
automatically display the measuring menu. close the door. Press the button to start the
measurement.
Measurement settings As soon as the measurement was started, the screen
Choose “Method”, “Profile”, “Sample ID”, “Operator” name changes to the measure screen which provides
and enter either the expected flash point or the start and information about the status of the measurement. The
final temperature. entire measurement is fully automated. A measurement
Measurement “Profiles” can be entered and modified in can be aborted by pressing the button
the “Settings” menu. Profiles include preset measurement
temperatures, appropriate sample names and if required
an offset for correlation to other instruments.
For unknown samples a scanning measurement (SCAN)
can be used, applying a wide measuring range entering
Tinitial and Tfinal (session switch grey).
For routine samples with approximately known flash point
simply enter the expected flash point (Texpected; session
switch red / white). Tinitial and Tfinal are set automatically as
required by the selected standard method.

When the measurement is finished, the result is displayed.


At the same time it is also saved in the result file. Press
to return to the measurement screen. Clean the
electrodes and the sample cup to be ready for the next
measurement.

Results menu
The results are stored directly in the “Results” menu and
can be viewed and compared there.
The results are shown for each method separately.
Change the method to view the corresponding results.

Printing of results & LIMS transfer


Select the results to be printed / transferred by a clicking
and holding, then press the “PRINT” or “LIMS” button.
The selected results are printed or sent to LIMS. (For
printer & LIMS configuration see the “System” menu).
ASTM D7094, ASTM D93 c and ASTM D56 c, SCAN Export results on USB memory stick
If you choose these standard methods then the 2 mL To export all results connect a USB thumb drive and wait
sample cup must be used. Use a pipette to pour the a few seconds. The memory stick is automatically
sample into the sample cup up to the 2 mL mark. Put a mounted. Pressing the “EXPORT” button will store all
stirring magnet into the cup for stirring the sample during results from all standard methods (including the log-file).
the test.

Short Instruction Manual ERAFLASH & ERAFLASH LT eralytics GmbH


Version v81001_2017-03 office@eralytics.com
Page 1 of 1 www.eralytics.com

You might also like