SAMPLING A N D ANALYSIS O F COMMERCIAL FATS A N D OILS ',

AOCS Official Method Ca 3a-46


Replacing Ca 3-25 and Ca 3-46 Reapproved 1997
h

Insoluble impurities"
DEFINITION
This method determines dirt, meal and other foreign substances insoluble in kerosene and petroleum
ether.
SCOPE
Applicable to all normal fats and oils.

APPARATUS 3. Filter through the prepared Gooch crucible with the aid
1. Gooch crucible--prepared with a glass-fiber filter with- of a vacuum. Wash with five 10-mL portions of hot
out organic filler (Reeve Angel 934 AH or WhatmanJ kerosene, allowing each portion to drain before adding
GF/C). Wash the filter with y , ~ x , all;l2b& and e h s the next.
Dry to constant weight at I0 I k 1"C Cool in a desicca- 4. Wash thoroughly with petroleum ether to remove all of
tor to room temperature and weigh. J the kerosene. Dry the crucible and contents to a
2. Filter flask of convenient size ahd Gooch crucible constant weight at 101 k 1°C. cool to room temperature
v in a desiccator and weigh.
(
adapter.
F

REAGENTS (see Notes, Caution) CALCULATIONS


1. Petroleum ether-AOCS Specification H 2-41. 1. Insoluble impurities, % =
2. Kerosene-refined petroleum distillate with a flash gain in mass.of crucible
point not below 23°C (75°F). as determined by the x loo
American Society for Testing and Materials Standard mass of sample taken for moistvre
Method D56, using the tag closed tester. The kerosene
should be filtered through a Gooch crucible, prepared NOTES
as in Apparatus, I, before using. Caution
Petroleum ether and kerosene are flammable solvents. A
PREPARATION O F SAMPLE fume hood should be used at all times when working with
1. Samples must be mixed thoroughly. If necessary, soften these solvents.
with gentle heat (do not melt) and mix thoroughly with
an efficient mixer. NUMBERED NOTES
. I . Samples (e.g., certain feed stocks) with a higher than
PROCEDURE usual percentage of insoluble impurities may be diffi-
1. Use the residue from the moisture and volatile matter cult, if not impossible, to filter. A study among five
determination (AOCS Official Methods Ca 2b-38 or Ca laboratories indicated that, for such samples, a 2-g
2d-25), or a sample prepared in the same manner (see subsample (rather than a 5-g subsample as specified in
Notes, I). AOCS Ofticial Methods Ca 2b-38 and c a 2d-25) can be
2. Add 50 mL of kerosene to the residue.and heat in a used without greatly affecting the precision of the
,,,-
\
.- water bath to dissolve the fat. method.