General: The GC must be fully PC controlled and should operate with a laboratory
temperature from 10 °C to 40 °C and humidity range from 10% to 80%.
Gas Chromatography
Injector:
1. Programmable split/split less capillary injector.
2. Pressure rang 0-150 psi.
3. Spilt vent prevents contamination of split valve and lab air.
4. Automatic control of split vent by split flow or split ratio
5. Programmed flow, programmed pressure, and constant flow mode capabilities.
6. Temperature-programmable inlet.
7. Maximum Temperature limits up to ≥ 450 °C.
8. Temperature programmed split and/or split less operation.
9. Split ratio must be not less than 7500:1
10. Suitable for columns Wide bore (0.53 mm) or narrow bore (0.05-0.32 mm).
11. Operating Pressure of up to ≥ 60 psi.
12. Capability to isolates the column and detector from the effects of high levels of
solvent.
13. Allows the use of different solvents such as methylene chloride with an ECD.
Autosampler
1. Injection speed Normal, fast, slow.
2. Sample positions not less than 100.
3. At least 4 wash vials of 4 ml.
4. Injection into Front Channel or to the other Channel
5. Sample size (1- 250µl for liquid samples).
6. Must be supplied with all accessories may needed for the operation.
ECD Detector:
1. Software flow control of makeup gas.
2. Temperature protect 470 ˚C by software
3. Operating temperature 100 ˚C to 450 ˚C in 1 ˚C increments
4. Linearity > 104
5. MDL < 6 fg lindan per sec.
Columns:
Capillary Columns set dedicated for analysis of the required DBPs (Disinfectant By-
products ) analysis for 4 years will be supported with separate warranty.
Training :
Training on Software, operation, and applications, external training for at least 10
working Days for 2 chemists.
Warranty:
4 Years from installation.
Consumables:
4 years additional consumables (detailed list must be attached).
Operation Requirements:
All required items needed for operation and installation of the system must be included.
The vendor must have at least one experienced, formally trained and approved by
manufacture to do installation, maintenance and validation of the instrument and have at
least one application expert for providing application training.
The system should be top quality, most recent version, and must not refurbish.
Country of Origin: USA, EU.