1. Product Introduction
HPM-300A is a consumption/standby power analyzer and a practical and economical handheld-type wattmeter. It enables
the measurement of the power consumption (power standby) of home appliances and business equipment with 0.1 W to 4
KW power consumption, in a Plug & Play way {with an AC 220V clamp sensor (1000A), it can measure up to 220 KW
power consumption}, thus offering simple and easy operation even by users at production plants and industrial sites. It can
also measure AC, DC, and combined electric power consumption.
※ There is an extra charge for the purchase of a clamp sensor (KRW 90,000/ea).
Clamp sensor models:
* CP-09 - 200A (AC use only); ±2% accuracy
* CP-11 - 400A (DC use only); ±5% accuracy
* CP-12 - 1000A (AC use only); ±3% accuracy
Warning
- Extra caution against a high voltage should be taken when using this equipment.
- The inappropriate use of the equipment may cause casualties or electric shock.
- Observe all the safety rules stated in the accompanying manual.
- Ensure the complete use of the provided power connection wire (load wire).
⊙ Safety Instructions
This equipment has been designed in accordance with the CE CAT II & FCC safety standards. Users, however, must
see to it that they have a thorough knowledge of the guidelines for the use of the equipment, as shown below, and
must use it with full attention, to ensure its safe operation.
- Do not load general batteries, only rechargeable batteries, and do not use an AC adaptor while loading general
batteries (the Li-ion batteries are built-in).
- When using the selected function, do not go over the voltage or current limit, which had been set as the maximum.
Do not attempt to operate the equipment if you cannot close its back cover.
- 3 -
Warning
If the blinking “ ”signappearsonthescreen,rechargethebatteriesimmediatelyoruseanadaptor.
For battery recharging, use a 12V 500mA adaptor.
- Recharge the batteries for about five hours, or connect them to an adaptor for five minutes, before using the
measuring device for the first time.
⊙ General Characteristics
- The use of a wide LCD enables the simultaneous display of the measured voltage, electric current, electric energy,
and frequency values, facilitating prompt and easy reading.
- ±0.4% electric power accuracy
- ±0.3% voltage and electric current accuracy
- Voltage measurement range: AC 1~500V/DC 1~990V (±0.3%, ±3 digit)
- Electric current measurement range (direct measuring): AC 1mA~15A/DC 1mA~15A (±0.3%, ±3 digit)
- Electric energy measurement range (direct measuring): 0.1W~ 4kW (±0.4%, ±3 digit)
- Power measurement factor: ±0.001~1.000 (±0.5%, ±3 digit)
- Auto/manual voltage and electric current range
- Operating temperature: 0~50°C
- Storing temperature: -10~60°C
- Operating power source: 7.4V lithium ion (use after recharging for five hours)/12V (300mA) DC adaptor
- Temperature coefficient: 0.1 X (stated accuracy)/1°C (no greater than 18°C or greater than 28°C)
- Altitude: 2000 m (6,561.7 feet)
- Size: 89 mm (W) x 188 mm (H) x 47 mm (D)
- Weight: 600 g (inclusive of the batteries)
MODE AC/DC measurement mode conversion (for “Energy On,” no conversion is done)
Electric power (W), electric energy (Wh), and power factor (PF) mode conversion
Power
button
CURRENT Conversion mode with an electric current unit (manual or auto range)
In the kilowatt hour electric energy (Wh) mode, when the energy is off, press a button
ENERGY
CLEAR
for 3 sec to reset it.
- 4 -
LCD backlight on/off
Symbol Meaning
Earth (grounding)
- 5 -
2. AC Electric Power Measurement (When Using a Load Wire, No Greater
than 15A Power)
Caution
To prevent an accident during the measurement, make sure that the location of HPM-300A is in an appropriate mode
before connecting the test lead in to the instrument.
LOAD
<LOAD WIRE - Using the AC electric power measurement mode (direct measuring)>
<When testing electric power consumption no greater than 15A>
- 6 -
<When using a probe, AC electric power measurement mode (direct measuring)>
<When measuring electric power consumption no greater than 15A>
MODE
1. Press the button to select the AC mode.
2. Each time you press the Power button, it is converted to electric power → electirc energy → power factor.
3. The electric power range has been set to "Auto Range."
4. Connect the power source accurately. Do not interconnect the electric current port and the voltage port in a reverse
way.
5. Load wire connection method.
* Red load wire: “A” terminal → Connect to the 15A terminal of HPM-300A.
* Black load wire: “±” terminal → Connect to the COM terminal of HPM-300A.
* Yellow load wire: “V” terminal → Connect to the V terminal of HPM-300A.
6. When the preparations have been completed, connect the measurement load to the load wire’s plug socket
and plug it into a power source.
! Warning: To prevent electric shock, apply the power source to the load wire’s plug socket only after connecting the
terminal.
7. The voltage, electric current, and frequency will be displayed on the LCD.
9. When the accumulation pauses, press again and the accumulation will stop.
- 7 -
3. AC Electric Power Clamp-Sensor-used Measurement (Clamp Sensors
CP-09, CP-11, and CP-12)
Warning
To prevent an accident during the measurement, make sure that the location of HPM-300A is in an appropriate mode
before connecting the test lead into the instrument.
MODE
1. Press and select the AC mode.
2. When using a current sensor, conversion to the CT mode is required.
AUTO RANGE
| Press RANGE
(00.00mA) RANGE
| Press RANGE
(000.0mA) RANGE
- 8 -
| Press RANGE
(0.000A) RANGE
| Press RANGE
(00.00A) RANGE
| Press RANGE
Warning
To prevent an accident during the measurement, make sure that the location of HPM-300A
is inanappropriate mode before connecting the test lead into the instrument.
- 9 -
< Interconnection (direct measuring) when measuring the DC voltage and electric current >
※ When there is a residual voltage or electric current value in the DC mode, offset before measuring.
- Voltage offset: Press “V-RST” after short-circuiting the voltage terminals and the ground (using a wire or pincet).
hevoltagevaluewillbeoffsetafteryouhearthebeepingsound.
- Electric current offset: Select a conversion unit mode first.
① Press “A-RST” in 00.00mA and it will be offset after you hear the beeping sound.
② Press “A-RST” after conversion to 0.000A by pressing “RANGE.” It will be offset after you hear the beepi ng
sound.
③ Press “RANGE” to convert to the CT mode, then short-circuit the CT terminal and press “A-RST.” It will be
offset after you hear the beeping sound.
- 10 -
5. DC-Electric Power Clamp-Sensor-used Measurement (Electric Current
Sensor Using CP-11 with Not Greater than 400A Power)
- 11 -
AUTO RANGE
| Press RANGE
(00.00mA) RANGE
| Press RANGE
(000.0mA) RANGE
| Press RANGE
(0.000A) RANGE
| Press RANGE
(00.00A) RANGE
| Press RANGE
8. Before measuring, press “DC ZERO” on the electric current sensor until the electric current value of HPM-300A on
the LCD reaches “0".
9. Apply a voltage to the V (+) and COM (-) terminals of HPM-300A.
10. When the electric current sensor is connected to a load, the electric current value will be displayed on the LCD.
11. Thevoltage, electric current, and electric power usage will be displayed on the LCD.
- 12 -
7. Range Modification
8. PC Communication Method
(1) Communication port: Connect to a created port in the case of a serial communication port or a USB cable.
(2) Transmission speed: 9600
(3) Parity: none
(4) Stop bit: 1
(5) Flow control: none
(6) When the English capital letter “S” is transmitted from a PC or Micom, or when “RS-232” is pressed, data will
be consecutively received and “COMM” will be displayed on the product’s screen. When the English capital letter
“E” is transmitted, the communication will be cancelled.
(7) “S” on the communication screen means the start of a data set, while “E” means the end.
- 13 -
UNIT (data at the decimal point)
PF 0.000PF - - - - - -
Press “HOLD” to temporarily pause the HPM-300Aoperation. The operation will be suspended until “HOLD” is pressed
again. Pressthe button to operate the LCD backlight.
An electric current sensor can be used for measuring electric power consumption without blocking or connecting a circuit.
This sensor is equipped with a fireproof and electric-shock-free case for hand protection in compliance with the safety
standards.
An electric current sensor may be separately purchased. You may choose from three different types (KRW 90,000).
※ Types of electric current sensor: CP-9 (AC use only, with a not-greater-than-200A power), CP-11 (combined AC/DC
receiver, with a not-greater-than-400A power), CP-12 (AC use only, with a not-greater-than-500A power)
- 14 -
(using an electric current sensor). The inappropriate operation of the equipment can cause casualties.
3. Using the equipment with a cable with damaged coating is very dangerous, and careless contact with the equipment
can cause electric shock. Make sure to wear rubber boots when using the equipment, and use it on a rubber floor
with verified insulated products. Normal operation in a grounded state must also be maintained to ensure safety.
4. Do not open or repair the equipment’s case while operating the equipment.
④ General Characteristics
- Operating conditions: 0~50°C, humidity not greater than 70%
- Storing temperature: -20~60°C, 0~80% humidity
- Temperature coefficient: 0.1X (stated accuracy)/1°C (not greater than 18°C; 28°C or above)
- Altitude: 2000 m (6,561.7 feet)
- Size: CP-09 - 48 mm (W) x 124 mm (H) x 30 mm (D)
- CP-11 - 63 mm (W) x 90 mm (H) x 40 mm (D)
- CP-12 - 82 mm (W) x 230 mm (H) x 40 mm (D)
- Output: Coil cable inclusive of banana plug
- Electronic characteristic: 23±5°C, 70% humidity
- Output: 0~1V RMS (1Meg ohm or above the input impedance)
- Clamp sensor accuracy: CP-09 - 1~200A, 50~60Hz, ±2%, AC electric current measurement use only, transmission rate:
- 1mV/1A
- CP-11 - 1~400A, 50~60Hz, ±5%, DC electric current measurement availability, transmission rate: 10mV/40A;
1mV/400A
- CP-12 - 1~500A, 50~60Hz, ±2%, AC electric current measurement use only, transmission rate: 1mV/1A
- 15 -
11. Others
Maintenance
Warning
Take note of the following to prevent electric shock or damage:
1. Do not separate the case when the power source is applied.
2. When replacing a fuse damaged by overload, make sure to replace with an identically rated (250V, 15A) fuse.
3. Do not open the case except when replacing the fuse.
4. When other maintenance is required, consult with the company’s A/S team.
5. The company shall not be held accountable for damages caused by equipment repairs executed by the users
themselves or by non-company personnel, other than replacing the fuse.
Calibration
It is recommended that calibration be carried out in accordance with the equipment specification (once a year).
Accessories/Components
Load wire (1 ea), AC adaptor (1 ea), test probe (3 ea), additional crocodile-shaped clip (2 ea) (upon the purchase of an
electric current sensor)
: Cable supply for data communication use (charges will be applied)
- 16 -
ADAPTOR
TEST
PROBE
LOAD WIRE
- 17 -
12. Appendix
WT = VI*CONɵ
WTOTAL = W1 + W2
- 18 -
(3) Three-phase four-wire electric power measurement (3∅, 4-wire)