RETURN TO THIS MANUAL'S TABLE OF CONTENTS, RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) NM3 5.3 Oxygen Ratio Monitor/ Controller Adjustment 53.1 53.7 53.8 53.9) 53.14 NOTE: Remove the rear cover of the vapor box. Refer to replacement procedure 4.10. Connect a calibrated oxygen monitor to the Freshgas Outlet Disconnect the pipeline hoses. Open the oxygen and nitrous oxide cylinder valves. Set the System Power switch to ON. Close the O, flow control valve. Open the N,O flow control val c Lo its counter-clockwise stop If there is no nitrous oxide flow, proceed to Step 5.3.10. If there is nitrous oxide flow, loosen setscrew E and move the cam mounting bracket to the right until the nitrous oxide flow stops. Tighten setscrew E. Slowly open the O; flow control valve. If the nitrous oxide starts to flow when the oxygen flow is between 200 and 300 ml per minute, proceed to Step 5.3.14. If the nitrous oxide does not start to flow when the oxygen flow is between 200 and 300 ml per minute, loosen setscrew E and move the cam mounting bracket to the left until nitrous oxide flow starts Tighten setscrew E. Repeat steps 5.3.6 thru 5.3.12 until no further adjustment of the cam. mounting bracket is needed. Adjust the oxygen flow to a point where the nitrous oxide flowmeter indicates 10 liters per minute. Steps marked with an asterisk (*) do not apply to later model s that do not have the O,/N,O ratio lamp on the alarm channel, 58 RETURN TO THIS MANUAL'S TABLE OF CONTENTS, RETURN TO CD-ROM TABLE OF CONTENTS NM3: ADJUSTMENT AND CALIBRATION PROCEDURES (continued) NARKOMED 3 REAR VIEW (02 TUBING. MPL SWITCH ADJUSTMENT N20 (TURN CLOCKWISE \ 4 TO DECREASE \ * SETPOINT) \ A MONTTORCONTROLLEA “vou / NountING BRACKET e swiron Moun SWITCH ADJUSTMENT CaM (URN clocaWise . TO CLOSE CONTACTS) Figure 5-3: OXYGEN RATIO MONITOR/CONTROLLER ADJUSTMENTS 59 RETURN TO THIS MANUAL'S TABLE OF CONTENTS, RETURN TO CD-ROM TABLE OF CONTENTS ADJUSTMENT AND CALIBRATION PROCEDURES (continued) NM 5.3.15 The oxygen concentration should be between 21% and 29% oxygen. * 5.3.16 Ifthe yellow O,/N,O FLOW RATIO lamp on the alarm panel is lighted, proceed to step 5.3.20. * 5.3.17 If the yellow O,/N,O FLOW RATIO lamp on the alarm panel is not lighted, loosen setscrew H and turn the switch adjustment cam until it is pointing downward * 5.3.18 Loosen setscrew A and move the switch mount until it is nearly touching the switch adjustment cam. Tighten setserew A. * 5.3.19 Turn the switch adjustment cam clockwise until the contacts close and the O,/N,O FLOW RATIO lamp is lighted. Tighten setscrew H * 5.3.20 Close the N,0 flow control valve. * 5.3.21 Adjust the oxygen flow to 1 liter per minute * 5.3.22 Slowly open the NO flow control valve. * 5.3.23. ‘The yellow OJN,O FLOW RATIO lamp on the alarm panel should light as soon as the ORMC limits the flow of nitrous oxide * 5.3.24 Adjust the oxygen flow to 1.5 liters per minute. * 5.3.25. Slowly open the NO flow control valve. * 5.3.26 The yellow OJN,O FLOW RATIO lamp on the alarm panel should light as soon as the ORMC limits the flow of nitrous oxide 5.3.27 Adjust the oxygen flow to 2 liters per minute. 5.3.28 Slowly open the NO flow control valve. * 5.3.29 The yellow OJN,O FLOW RATIO lamp on the alarm panel should light as soon as the ORMC limits the flow of nitrous oxide 5.3.30 Close the oxygen flow control valve. The nitrous oxide flow should decrease proportionally, and the oxygen concentration should remain between 21% and 29% oxygen. 5-10 RETURN TO THIS MANUAL'S TABLE OF CONTENTS, RETURN TO CD-ROM TABLE OF CONTENTS NM3: ADJUSTMENT AND CALIBRATION PROCEDURES (continued) 5.3.32 5.3.33 5.3.38 The nitrous oxide flow should stop when the flow of oxygen is between 200 and 300 ml per minute. If the 0/N,O FLOW RATIO lamp turns off when the nitrous oxide flow drops to between 100 and 200 ml per minute, proceed to Step 5.3.34. If the O,IN,O FLOW RATIO lamp does not turn off when the nitrous oxide flow drops to between 100 and 200 ml per minute, adjust the MPL switch until the OJN,O FLOW RATIO lamp turns off, Close the ©, eylinder valve and the O, flow control valve. Press the 0, Flush valve to drain oxygen pressure from the system. Close the N,O flow control valve. Reconnect the pipeline hoses. Replace the rear cover of the vapor box. Perform the PMS Procedure given in Section 6 541

You might also like