You are on page 1of 19

6.610.

Trends in Materials for Spine Surgery


M Marcolongo, S Sarkar, and N Ganesh, Drexel University, Philadelphia, PA, USA
2011 Published by Elsevier Ltd.

6.610.1. Introduction 128


6.610.2. Anatomy and Physiology of the Spine 128
6.610.2.1. The Vertebral Column 128
6.610.2.2. The Intervertebral Disc 129
6.610.2.3. The Nucleus Pulposus 129
6.610.2.4. The Annulus Fibrosus 129
6.610.2.5. Intervertebral Disc Degeneration 130
6.610.3. Spinal Fusion 131
6.610.3.1. Metals for Spinal Cages and Interbody Spacers 132
6.610.3.2. Polymers and Composites for Fusion Cages 132
6.610.3.3. Wires, Tapes, and Cables in Spinal Fusion 133
6.610.3.4. Pedicle Screws, Plates, Bands, and Rods for Fusion and Dynamic Stabilization 134
6.610.3.5. Bioresorbable Materials for Spinal Fusion Implants 134
6.610.4. Total Disc Arthroplasty 135
6.610.4.1. Material Requirements 136
6.610.4.1.1. Implant endplate 136
6.610.4.1.2. Bone implant interface 136
6.610.4.1.3. Bearing surfaces 136
6.610.4.1.4. General material specifications 136
6.610.4.2. Discontinued Artificial Discs 137
6.610.4.2.1. Acroflex 137
6.610.4.3. Currently Available Devices 137
6.610.4.3.1. Charite artificial disc 137
6.610.4.3.2. ProDisc 138
6.610.4.3.3. Maverick 139
6.610.4.3.4. Flexicore 140
6.610.4.4. Future Trends 140
6.610.5. Annulus Repair 140
6.610.5.1. Procedure 140
6.610.5.2. Mechanical Role 140
6.610.5.3. Reported Treatment Methods 141
6.610.5.3.1. Suture techniques 141
6.610.5.3.2. Sealant and bulking techniques 141
6.610.5.4. Current Technologies 141
6.610.5.4.1. Xclose tissue repair system 141
6.610.5.5. Barrier Technologies 142
6.610.5.5.1. Inclose surgical mesh system 142
6.610.5.5.2. BarricaidW prosthesis 143
6.610.5.5.3. SpineJet XLW 143
6.610.5.6. Material Requirements 143
6.610.5.7. Future Trends 144
6.610.6. Concluding Remarks 144
References 144

Glossary induced under load and in the presence of repeated relative


Discogenic back pain Discogenic back pain is also known surface motion and wear.
as lumbar disc pain. The most frequent symptoms of Hemangiomas An abnormal proliferation of blood vessels
discogenic back pain are lower back pain and spasms. that may occur in any vascularized tissue.
Fretting corrosion Fretting corrosion refers to corrosion Laminectomy Spine surgery that removea the portion of the
damage at the asperities of contact surfaces. This damage is bone arch on the dorsal surface of a vertebra.

127
128 Orthopedic Surgery Spinal Treatment

Osteopenia A condition in which bone mineral density is Vertebroplasty and kyphoplasty Medical spinal procedures
lower than normal, but not low enough to be classified as in which a material is injected through a small hole in the
osteoporosis. skin (percutaneously) into a fractured vertebra and hardens;
Pseudo elasticity A reversible response to an applied stress, the goal of the procedure is to relieve the pain caused by
caused by a phase transformation between the austenitic and osteoporotic compression fractures.
martensitic phases of a crystal.

Abbreviations NP Nucleus pulposus


AF Annulus fibrosus PG Proteoglycan
GAG Glycosaminoglycan TDA Total disc arthroplasty
IVD Intervertebral disc

6.610.1. Introduction visualization of spinal structures, innovations of novel surgical


procedures, and a better understanding of the biology and
The spine has two main purposes: (1) it must provide a strong structure of the native tissue. Early materials were adapted
mobile central axis to support the skeleton and (2) it must also from procedures designed for the repair of bones and joints
provide protection for the delicate nerves that travel from the in the extremities; however, as more is understood about the
brain to the periphery. A balance between stability and flexi- unique structure and mechanical needs of the spine, novel
bility are essential to achieve this dual purpose. Aging, injury, materials are being developed for use in spine specific proce-
disease, and genetic conditions can disrupt this balance result- dures. Tissue engineering and drug delivery have also moti-
ing in back pain with more than 60% of the population experi- vated the development of new materials with porous and
encing back pain throughout a lifetime.1 Low back pain alone regenerative surfaces. This chapter outlines the current materi-
accounts for 2% of all US physician office visits.2 Pain can be als utilized in the major surgical interventions (spinal fusion,
generated in the spine from damage to several anatomical total disc replacement, and annulus repair) and spine pathol-
elements including the vertebral bodies, intervertebral discs ogies as well as the future direction of material development.
(IVDs), facet joints, dura of the nerve roots, ligaments, facia,
and muscles.3 Stabilization of the spine has been the major
focus of surgical interventions to alleviate back pain.4 6.610.2. Anatomy and Physiology of the Spine
Interventional techniques have evolved from the earliest use
6.610.2.1. The Vertebral Column
of metal screws and plates to correct scoliosis adapted from
peripheral extremity repair in the late 1950s.5 Internal fixation The human vertebral column or spine functions to transfer
of the spine using pedical screws was first introduced in 1959 by loads and bending moments of the head, trunk, and any
Boucher in Canada.6 In the 1960s, the advent of bony fusion external loads to the pelvis, allows sufficient physiological
techniques greatly increased the success of these procedures and movement and flexibility of the upper body, and protects the
currently, advancements are still being made on materials choice spinal cord from dangers due to motion and trauma.14 It also
and structural design of spinal internal fixation devices.7 The serves as protection to other vital internal organs and as a base
IVD has also been a major focus in spinal procedures with the of attachment for upper-body ligaments, tendons, and mus-
first surgeries on disc protrusion reported in 1934 by Mixter and cles. These functions help explain its form and structure.
Barr.6 Current techniques in treating discogenic back pain The spine consists of 24 individual vertebrae, the sacrum,
include removal of disc material and replacement with pros- and the coccyx, and spans from the skull to the pelvis. The
thetic devices8; however, advances in knowledge of the IVD vertebrae are stacked on top of each other and are grouped into
ultrastructure have led to recent investigations into the repair five distinct regions there are 7 cervical vertebrae, 12 thoracic,
of individual IVD components, that is, nucleus replacements9 5 lumbar, 5 fused sacral, and 3 fused vertebrae that make up
and annulus repair technologies (see Chapter 6.613, Nucleus the coccyx (Figure 1). From posterior perspective, the spine
Replacement).10 Recent advances in minimally invasive surgery appears to be straight, but lateral view shows four normal
have also made a large impact on spine surgery and the potential curves that mechanically serve to give the spinal column
for novel therapies. increased flexibility and shock-absorption ability.
Most notable is the development of the vertebroplasty and The vertebral body is the primary weight bearing area, and
kyphoplasty procedures developed in the late 1980s for the adjacent vertebrae are connected by an IVD (Figure 2). The ver-
percutaneous treatment of vertebral fractures (see Chapter tebral foramen is the large hole in the center of the vertebra that
6.611, Injectable Bone Cements for Spinal Column Augmen- is surrounded by lamina, forming the spinal canal. The spinal
tation: Materials for Kyphoplasty/Vertebroplasty).1113 The cord runs through this opening and is protected by it. The
development of materials for spinal procedures has been spinous process protrudes from the central posterior region of
greatly influenced by advances in technology that permit the the vertebra and acts as a connection point for ligaments. There
Trends in Materials for Spine Surgery 129

Lateral (side) Posterior (back)


spinal column spinal column

Cervical Cervical

Thoracic Thoracic

Lumbar Lumbar
Figure 3 Vertebral compression fracture. Reproduced from
Stryker-Interventional. 2008. Compression Fractures of the Spine.
Sacrum Sacrum http://www.helpingbacks.com/spinalfractures.php.

Coccyx Coccyx
largest avascular tissue in the human body requiring nutrients
Figure 1 Regions of the human spine with respect to the vertebrae and to enter and waste to be removed primarily through diffusion
corresponding intervertebral discs (http://healthbase.wordpress.com/ and fluid flow.14 The disc structure is quite complex combining
2007/06/13/spinal-fusion-surgery/). three distinct regions: the nucleus pulposus (NP), an inner
gelatinous core; the annulus fibrosus (AF), a surrounding
ordered fibrous structure; and the bony endplates that are
Lumbar vertebrae partly cartilaginous and partly bony and serve as an interface
between the disc and the vertebral bodies (Figure 4). The
Transverse cellular content of the disc is very low, less than 0.25% of
process Vertebral
tissue volume.16 The extracellular disc matrix plays the main
body role in determining the mechanical properties of the disc.
Spinous
process
6.610.2.3. The Nucleus Pulposus

Lamina The central NP region of the IVD is a gelatinous structure that is


mostly water (77% of wet weight) but also contains a large
Vertebral constituent of proteoglycans (PGs) (14% of wet weight) which
Spinous body are composed of charged glycosaminoglycan (GAG) chains
Pedicle
process covalently attached to a protein core. Randomly organized
Axial Lateral collagen fibers are also present in the NP (about 4% of wet
(overhead (side weight) and associate with the PGs to form a loose network.17
view) view) Collagen fibers within the NP provide tensile strength while
PGs create a large osmotic swelling pressure and draw water
Figure 2 The lumbar vertebrae. Reproduced from Toida, T.;
into the tissue (Figure 5). Charged groups on the GAG chains
Kakinuma, N.; Toyoda, H.; Imanari, T. Anal. Sci. 1994, 10, 537541.
of the PGs carry with them mobile counter ions such as Na
which draw water into the tissue because of the osmotic imbal-
are pairs of transverse processes which are orthogonal to the ance generating a hydrostatic pressure within the NP.18
spinous process and provide attachment for the back muscles.
There are also four facet joints on each vertebra that are found in
6.610.2.4. The Annulus Fibrosus
pairs (one superior and one inferior), interlocking with adjacent
vertebrae to provide stability to the spine. When osteoporosis The AF of the IVD is a structure that serves to contain the
occurs in the spine, vertebral collapse, or compression fractures, gelatinous NP. The chemical composition of the AF comprises
may occur (Figure 3.). Vertebral compression fractures may also 6590% wet weight of water, 5070% dry weight of collagen,
be caused by hemangiomas, multiple myeloma, and osteolytic 1020% dry weight of PGs, and noncollagenous proteins such
metastases.15 Multiple fractures lead to a stooped or hunched as elastin.19 The structure of the AF is laminate in nature, con-
posture, chronic pain, and loss of height. sisting of a minimum of 15 (posterior) and 25 (lateral) concen-
tric layers. The layers are made up of type I collagen fibers which
alternate in angles from 30 at the peripheral AF to 44 at the
6.610.2.2. The Intervertebral Disc
central AF with reference to the transverse plane of the disc
The IVD separates adjacent vertebrae and acts as a shock (Figure 6). The concentric rings of the AF are referred to as
absorber, dissipating energy under loading. The disc is the lamellae, and the spaces between them as interlamellar septae.19
130 Orthopedic Surgery Spinal Treatment

Annulus
Nucleus
fibrosus
pulposus
4 cm

Nerve Vertebral endplate


root

710
mm

Lamellae of
annulus fibrosus
Cartilaginous
Vertebral end-plate Nucleus pulposus
body Annulus fibrosus
Figure 4 The intervertebral disc (IVD) location between adjacent vertebrae and the individual structures and average dimensions of the IVD.
Reproduced from Raj, P. P.; Fipp, A. Pain Practice 2008, 8, 18.

Proteoglycan
Link protein swelling P
pressure H2O T
Aggrecan

20
Hyaluronan a
Collagen
tension
40
Collagen
fibril

Figure 6 Load transfer in the intervertebral disc (IVD) and lamellar


structure of the annulus fibrosus. When a compressive load (C) is placed
on the IVD, a hydrostatic pressure (P) is generated in the NP and tensile
stress (T) on the annulus. Lamella of the annulus has oblique collagen
fibers in alternating directions. On average, there are 20 collagen lamellae
Figure 5 Constituents of the nucleus pulposus and subsequent in the annulus with approximately 40 collagen bundles per lamella.
development of hydrostatic pressure. Reproduced from Kiani, C.; Liwen, C.; Reproduced from Adams, M.; Roughley, P. Spine 2006, 31, 2151.
Yao Jiong, W. U.; Yee, A. J.; Burton, B. Y. Cell Res. 2002, 12, 1932.

Spinal compression generates a hydrostatic pressure in the induced accelerated disc degeneration.21 The exact mechanism
NP that places the annulus into tension thereby resisting the of IVD degeneration is not well understood; however, possible
applied load (Figure 6).20 The AF has a very highly organized degenerative cascades have been proposed which lead to even-
structure which results in complex anisotropic behavior with tual changes in disc mechanical properties (Figure 7). Initial
loading. The tensile, compressive, and shear properties of the changes in the IVD are thought to be brought about by changes
AF differ in the axial, radial, and circumferential directions. in the nutrient flow to and waste product flow out of the
The AF can be further divided into the inner and outer AF on nucleus. Calcification of the cartilaginous endplates, the
the basis of the structural differences, the inner AF being sub- major route of nutrient supply to the disc, leads to harsh
jected to higher hydrostatic pressures of the NP than the outer AF cellular environments, subsequently reducing matrix synthesis
which is subject to tensile forces. and increasing enzyme production further reducing PG and
collagen content of the disc, in particular in the NP.21,23,24 The
loss of PGs in the NP has major effects on the load-bearing
6.610.2.5. Intervertebral Disc Degeneration
behavior of the disc, causing the osmotic pressure of the disc to
The IVD of the spine is subject to degenerative changes induced fall and reducing the disc water content disrupting the NPs
by normal aging as well as injury, loading, or genetically ability to behave hydrostatically under load. Loading can then
Trends in Materials for Spine Surgery 131

Stress

Load

Young
Endplatee sccal pillaries
Vertebra
Stress
erosis A P
matrix damage Endplat
rs
epto
Nocic
Nociceptors

Cytokines
Annulus
s
Fissure Lactate
es Matrix
as Old
ote degradation
Oxygen Pr
Nutrients
Nucleus
t
or

Disc cell
sp
an

Stre
Endplate sclerosis
Tr

ss
Endplate capillaries
Degenerated
Load
Distance
22
Figure 7 Cascade of events in intervertebral disc (IVD) degeneration and changes in mechanical properties of the IVD with degeneration
(Normal disc to degenerated disc from top to bottom). Reproduced from Adams, M. A.; Dolan, P.; McNally, D. S. Matrix Biol. 2009, 28, 384389.

Herniation of the lumbar disc Signal changes in vertebral end plates

L5 L5

(a)

L5 L5
Degeneration of the lumbar disc
(c)

Annular fissures with high-intensity signals

L5 L5

L5 L5

(b) (d)

Figure 8 MRI and schematic images of intervertebral disc disruptions. (a) Herniation of the lumbar disc with extrusion of disc material found in 118%
of asymptomatic subjects. (b) Degeneration of the lumbar disc increases with age and is found in 2570% of asymptomatic subjects. (c) Signal
changes in the vertebral endplates, found in 10% of asymptomatic subjects. (d) Disc with annular fissures seen in 1433% of asymptomatic
subjects. Despite high prevalence of these damage modes in healthy persons, these findings are often associated with serious back pain and are
often treated with spinal fusion or disc replacement strategies. Adapted from Carragee, E. J. N. Engl. J. Med. 2005, 352, 18911898.

lead to inappropriate stress concentrations along the endplate 6.610.3. Spinal Fusion
and in the annulus with progressive changes as disc degenera-
tion proceeds (Figure 7). Spinal fusion and stabilization utilizes a wide range of devices
Changes in stress concentration can lead to damage of such as screws, wires, artificial ligaments, and vertebral cages.26
the AF as well as the endplates with stress concentrations The instrumentation used in fusion surgery is designed to
being associated with discogenic pain.17 Several manifesta- stabilize the spine during fusion and promote osseous
tions of disc degeneration and damage can be visualized fusion.7 Material choice for these devices is therefore tar-
with MRI and are represented in Figure 8. Although these geted toward the promotion of bone ingrowth, and imme-
alterations in the IVD are prevalent in the asymptomatic diate mechanical stabilization of the spine. Postoperative
population, disc disruptions have been associated with imaging and analysis of surgical success is also of impor-
back pain and are targets for surgical interventions such as tance in material choice for spinal fusion devices.27 Early
spinal fusion, total disc replacement, nucleus replacement, spinal fusion devices were based on metals commonly uti-
and annulus repair.25 lized in other orthopedic surgeries such as in hip implants.
132 Orthopedic Surgery Spinal Treatment

Metals are still widely used in spine fusion implants today; bilaterally between the vertebral bodies.28,29 The ends of this
however, advances have been made in implant design and device are capped with ultrahigh molecular weight polyethyl-
material development in order to address the unique needs ene (UHMWPE) plugs in order to contain the bone grafts and
of the spine including maintenance of spine flexibility and to minimize adhesion to surrounding nerves and blood ves-
to avoid the occurrence of adjacent area damage. sels. The strength of the titanium material allows for the load
bearing function of the spine and protection of the graft mate-
rial while the implant design which incoroporates threading
6.610.3.1. Metals for Spinal Cages and Interbody Spacers
to stabilize the implant, and a hollow center to allow for bone
The removal of the IVD in order to address discogenic back delivery, achieves the other functions of this device. Metals,
pain and disc herniation will lead to the loss of disc height and however, are subject to fatigue failure if the graft does not
lead to spinal instability. Cage structures have been developed fuse; also the high modulus of metals can lead to stress shield-
to assist in implanting allo- and autologous bone graft inter- ing and eventual failure of the device to integrate into its
body spacers. These cages enhance stabilization of the spinal surroundings.26 See Table 1 for some common metals used
segment and prevent postoperative collapse. in spine implants and their material properties.
The BAK interbody fusion system is a cage structure made Cobalt chrome has also been developed into a porous
of porous Ti6Al4V shell with a cavity for the placement of metal sponge for use in spinal surgery.32 The porous structure
bone grafts (Figure 9).28 This device is implanted centrally or was designed to match the structure of cancellous bone and the
implant could be loaded axially up to 28.5 MPa. Problems
arose, however, with metal ion release because of the high
surface area imparted by the porous structure. Released metal
ions may increase pain at the implant site and cause loosening
of the implant and local bone resorption at the implant
site.33,34 Decreased pH levels at the implant site can be brought
about by the inflammatory response and infection and may
also have a deleterious effect on metallic implants causing
fretting corrosion.33,34

6.610.3.2. Polymers and Composites for Fusion Cages


In order to address some of the drawbacks of metal fusion
cages, composite materials have been developed and utilized
in spinal cages. A cage made from a composite of poly ether
ketone ether ketone ketone (PEKEKK) with pyrolytic carbon
fibers interspersed within the polymer matrix was developed.35
The material was molded to incorporate teeth located superi-
orly and inferiorly in order to resist expulsion (Figure 10).
Material properties were modulated to better mimic those of
Figure 9 The BAK cervical spinal fixation device. A threaded hollow bone in order to better support physiological loads as the
titanium implant that allows bone grafts to be implanted within the shell graft material slowly integrates into the surgical site and in
to facilitate fusion. Ends are capped with UHMWPE to minimize adhesion. case the graft does not fuse. The cage was designed to have a
Reproduced from Martz, E. O.; Goel, V. K.; Pope, M. H.; Park, J. B. Youngs modulus of 17 GPa, close to that of cortical bone.35
J. Biomed. Mater. Res. 1997, 38, 267288. This matched modulus decreases the likelihood of stress

Table 1 Material properties of common metals in spinal implants

Property 316L SS (cold worked) Wrought CoNiCrMo Ti6Al4V Pure Ti (Grade 4)

ASTM designationa F 138 (grade 2) F 562 (cold worked and aged) F 136 F 67
Density (g cm3)b 7.9 9.2 4.5 4.51
Elastic modulus (GPa)a 193 220234b 105 100
Yield strength (MPa)c 690 1586 795 483655
Tensile strength (MPa)c 860 1793 860 550
Elongation (%)c 12 8 10 15
Fatigue strength (MPa) 310 500 520 240
(107 cycles)a (F55 cold worked) (cold worked) (forged, annealed) (as fabricated)
a
From Black.30
b
From Park and Lakes.31
c
ASTM F 138, ASTM F 562, ASTM F 136, and ASTM F 67.
Source: ASTM. Annual Book of ASTM Standards; ASTM: Philadelphia, PA, 1992; Park, J. B.; Lakes, R. S. Biomaterials: An Introduction; Springer: Berlin, 2007; Black, J. Orthopaedic
Biomaterials in Research and Practice; Churchill Livingstone: New York, 1988; Martz, E. O.; Goel, V. K.; Pope, M. H.; Park, J. B. J. Biomed. Mater. Res. 1997, 38, 267288.
Trends in Materials for Spine Surgery 133

shielding. An additional benefit of the nonmetallic implant is fusion could be identified because of the radiolucency of the
the ability to visualize the postoperative implant area because PAEK materials which continued to be a major reason for the
of the radiolucent property of the material. use of this material (Figure 11).36,40
Polyaryletherketones (PAEKs) are in the family of high In addition to carbon reinforced PEEK, neat PEEK is also
temperature thermoplastic polymers and have been widely utilized in the development of spinal cages and implants.4143
used for spinal cage implants after their introduction in the Studies have also been conducted to investigate the improve-
1990s as an improvement over metal devices by Acromed ment of PEEK cages to incorporate accelerated fusion perfor-
(Cleveland, OH, now DePuy Spine, Raynham, MA).36 PAEKs mance by adding bioactive materials such as hydroxylapatite or
consist of an aromatic backbone molecular chain intercon- rhBMP-2.41,43,44 Wear and fracture can be potential concerns
nected by ketone and ether functional groups. Specifically, poly with the use of PEEK as a load-bearing implant material; how-
(aryl-ether-ether-ketone), PEEK, and poly(aryl-ether-ketone- ever, these concerns are not unique to PEEK. Although debris
ether-ketone-ketone), PEKEKK, have been used in orthopedic from wear has been noted in biopsies from carbon fiber PEEK
and spinal implants. The chemical structure of PAEKs allows cages, there has been no evidence of an inflammatory reaction
stability at high temperatures, exceeding 300  C,37 resistance to making PAEK type materials promising in this application.45
chemical and radiation damage,38 compatibility with reinfor- PEEK materials have also been investigated for other spine
cing agents such as glass and carbon fibers, and greater strength stabilization devices such as posterior dynamic stabilization
than even metals.36 Mechanical and material properties for some devices46 and total disc replacements.36
PEEK formulations in comparison to other commonly used
polymer materials are given in Table 2.36 These materials are
6.610.3.3. Wires, Tapes, and Cables in Spinal Fusion
also radiolucent making them a good alternative to metallic
spine biomaterials. Fusion of single or multiple spinal segments is also accom-
PEEK and PEKEKK were first used in the development of the plished through the use of wires, tapes, and cables to secure
Brantigan cage with 68% by weight carbon fiber reinforcement adjacent bone segments. The most commonly used wire mate-
in a 2-year clinical trial beginning in 1989.39 Clinical results rial is 316L stainless steel (Table 1).26 Stainless steel is rela-
were good with 100% interbody fusion identified. Interbody tively strong, ductile, and relatively inexpensive. Stainless steel
316L is especially preferred because of its resistance to corro-
sion as a result of low carbon content (0.03% maximum)
which prevents sensitization of the stainless steel. Nevertheless,
even stainless steel 316L is susceptible to localized corrosion
associated with a loss of the protective chromium oxide surface
layer.47 A concern in the use of metal wires for fixation is the
possibility of neurological damage as well as concerns that the
wire may cut through the bone if the bone is soft, which is
especially the case in children.26 As an alternative, polymer
Figure 10 Carbon fiber composite cage made from Ultrapek. tapes have been introduced such as nylon and woven strands
Reproduced from Martz, E. O.; Goel, V. K.; Pope, M. H.; Park, J. B. of polyethylene (Mersiline) for use in procedures on scoliotic
J. Biomed. Mater. Res. 1997, 38, 267288. children48,49

Table 2 Typical average physical properties of PEEK and CFR-PEEK structural composite biomaterials, compared with UHMWPE and polymethyl
methacrylate (PMMA)

Property (ISO) Selected lavibio PEEK biomaterials (OPTIMALTI) UHMWPE PMMA

Unfilled 30% (w/w) chopped carbon fiber 68% (v/v) continuous carbon fiber
(OPTIMALTI) reinforced (LTICA30) reinforced (Endolign)

Polymer type Semicrystalline Semicrystalline Amorphous


Molecular weight 0.080.12 0.080.12 0.080.12 26 0.10.8
(106 g mol1)
Poissons ratio 0.36 0.40 0.38 0.46 0.35
Specific gravity 1.3 1.4 1.6 0.9320.945 1.1801.246
Flexural modulus 4 20 135 0.81.6 1.54.1
(GPa)
Tensile strength 93 170 >2000 3948 2449
(MPa)
Tensile elongation 3040 12 1 350525 12
(%)
Degree of 3035 3035 3035 3975 Noncrystalline
crystallinity (%)

Testing conducted at 23  C.
134 Orthopedic Surgery Spinal Treatment

6.610.3.4. Pedicle Screws, Plates, Bands, and Rods for better integration. Stainless steel fatigue strength is enhanced
Fusion and Dynamic Stabilization by the implantation of nitrogen ions by inducing the forma-
tion of extremely hard nitrides at the surface of the material
Screws, plates, and rods are often used to augment spinal
while maintaining the bulk properties of the material.26 Tita-
fusion. Screws placed in the pedicles or vertebral bodies act as
nium can be modified by the introduction of bioactive titanate
anchor points for bands, rods and plates working together to
layers of, for example, sodium titanate, calcium titanate, or
restore spinal stability. The screwbone interface is especially
titanium oxide which induce apatite formation on the implant
influential in the success of fixations. Surface treatments of
surfaces and have been demonstrated to enhance osteoinduc-
metal implants of stainless steel50 or titanium51 have been
tivity as well as osteoconductivity for improved implant
utilized to increase fatigue life of the implants and promote
integration.51
Dynamic stabilization may be an alternative to fusion in
some patients and may provide an improvement over fusion
by altering load bearing and controlling abnormal motion while
maintaining some flexibility. Dynamic stabilization devices
incorporate a flexible ligament like component or other spring
or mobile joint to allow motion in the spine while controlling
movement such as flexion and extension. The diverse function-
ality of these devices often requires the use of both flexible
polymeric materials and more rigid materials such as metals
(Table 3).7

6.610.3.5. Bioresorbable Materials for Spinal Fusion


(a) (b) Implants

Figure 11 (a) Brantigan spine fusion cage and (b) lateral radiograph of Another key element in the successful stabilization of the spine
a Brantigan cage with a solid fusion. (Note that the Brantigan cage has in fusion surgeries is the restoration of the normal curvature of
tantalum microspheres for visualization on radiographs.) Reproduced the spine. If the material system is too rigid, it becomes difficult
from Kurtz, S. M.; Devine, J. N. Biomaterials 2007, 28, 48454869. to bend and twist to provide the appropriate curvature.

Table 3 Dynamic stabilization devices

Category and trade name Key features

Pedicle screws and artificial ligaments


Dynesys (Dynamic Neutralization System for the Semirigid artificial ligament system composed of titanium alloy pedicle screws and
Spine); Zimmer Spine, Warsaw, Ind (18) polycarbonaturethane spacers connected by polyster cords (artificial ligaments) that are
placed under tension
Graf ligament; Surgicraft, Redditch, England (19) Nonelastic polyester ligament looped around pedicle screws and placed under tension to
prevent rotation while allowing some flexion; if tension is too high, hyperlordosis,
foraminal narrowing, and nerve root impingement may result
IsoBar; Scientx USA, Maitland, FL Includes a mobile joint within metal rods
M-Brace; Applied Spine Technologies, New Haven, Implanted by using a minimally invasive technique
Conn
Stabilimax NZ; Applied Spine Technologies Utilizes a dual spring mechanism
Dynamic Soft Stabilization System (20) Elliptical metal coil connecting adjacent pedicle screws
Inter-spinour process decompression system
Wallis ligament; Abbott Spine, Austin, Tex (21) Interspinous process distraction device composed of a spacer held between spinous
processes with Dacron tape
X STOP; St Francis Medical Technologies, Concord, Titanium device in two pieces, placed between adjacent spinous processes to hold spine in
NH (22) flexion; can be inserted by using a minimally invasive approach with a local anesthetic and
is thus useful in elderly patients with degenerative spinal stenosis
Diam; Medtronic So famor Danek, Memphis, Tenn H-shaped silicone device held in place by a mesh band and suture
Coflex; Paradigm Spine, New York, NY Posterior U-shaped device that controls flexion and extension
element replacement systems
Total Facet Arthroplasty System; Archus Orthopedics, Consists of a sphere that slides along a curved plate anchored by pegs passing into the
Redmond, Wash vertebral body
Total Posterior System; Impliant, Princeton, NJ Posterior elements are removed and a plastic device is implanted and anchored with devices
similar to pedicle screws

Note: Some of the devices listed may not yet have been approved by the FDA for clinical use.
Source: Zucherman, J. F.; Hsu, K. Y.; Hartjen, C. A.; et al. Spine 2005, 30, 1351; Senegas, J.; Etchevers, J. P.; Vital, J. M.; Baulny, D.; Grenier, F. Rev. Chir. Orthop. Reparatrice Appar.
Mot. 1988, 74, 15; Sengupta, D. K. The Lumbar Spine; Lippincott, Williams & Wilkins: Philadelphia, PA, 2004; Vol. 1, pp 373383; Graf, H. Rachis 1992, 4, 123137; Stoll, T. M.;
Dubois, G.; Schwarzenbach, O. Eur. Spine J. 2002, 11, 170178; Senegas, J. Eur. Spine J. 2002, 11, 164169.
Trends in Materials for Spine Surgery 135

Inducing an abnormal curvature may lead to stress induced technique; however, this technique can also induce damage
osteopenia of the vertebral bodies and spinal degeneration in in the polymers causing changes in biocompatibility and bio-
adjacent spinal levels.26 This is, however, in contrast to the mechanical characteristics. Treatment with ethylene oxide or
needs of a rigid system in order to promote the solid fusion plasma sterilization provides better alternatives and shows
process. Therefore, the use of a system whose rigidity decreases limited effect on molecular weight and tensile strength of
with time is desirable. Bioresorbable materials provide promise PLA-based materials.53,54 Biocompatibility of PLA and related
for such a system. Bioresorbable cages have stiffnesses compa- polymers has generally been demonstrated to be good with
rable to those of bone, are radiolucent, and resorb over time.52 the absence of significant toxicity.52 A local reduction in pH
These cages provide the additional benefits of eliminating risks imparted by the degradation products, however, may be
associated with permanent implants such as implant removal responsible for some adverse tissue reactions such as osteolysis
on revision surgery. In addition, degradation of the polymer and bone resorption around the implant.55
may allow for the incorporation of a drug release system into Bioresorbable PLA plates have performed well in compari-
the implanted device. son to titanium plates in controlled clinical studies. In an
The most commonly used biodegradable materials are original study by Nabhan et al., bioresorbable plates made of
polyesters derived from poly(a-hydroxy acids) such as poly L-lactic and D,L-lactic acid in the ratio of 80:20 in the INION S-1
(lactic acid) (PLA) and poly(glycolic acid) (PGA), and their Biodegradable Anterior Cervical Fusion System were investi-
isomers and copolymers. The chemical backbones of these mate- gated in a prospective randomized controlled study with 40
rials are hydrolytically unstable, and the chains will degrade patients.56 The bioresorbable plates in this study were designed
when placed in aqueous environments such as the body. PLA to hydrolyze over the course of 2 years. Plates maintained 90%
and PLGA degrade to lactic acid and glycolic acid respec- of their initial strength after 6 months, then continued
tively and are natural, biocompatible metabolic compounds. to decrease slowly to 70% after 9 months, and continued to
The strength and degradation rate of these polymers can be absorb after 2 years. The fusion incidence and rates of fusion
modified by the fabrication of copolymer systems. PGA degrades were found to be comparable to those of titanium plates but
within months and is therefore used only in small amounts in with several added benefits associated with the use of a biode-
copolymer systems, while PLA is more widely used in spinal gradable polymer. Advantages of the PLA plates included
implants. Chemical and physical properties of some commonly the radiolucency of the plates which allowed visualization
used PLA and PLA isomers are summarized in Table 4. of the graft material giving reassurance that the plates had not
The polymer degradation kinetics and mechanical proper- migrated. The plates were also somewhat flexible and were able
ties can be influenced by the polymer system utilized; however, to be bent to match the vertebral contours of the patients.
the implant design and geometry itself has a large influence on Gradual resportion of the plates may allow the fusion to grad-
the rates and types of degradation experienced by the implant. ually assume the role of structural support and enhance fusion
PLA devices degrade faster when they are less porous, less rates while reducing stress shielding.56
permeable, and have a bulkier design and wall thicknesses.52
Sterilization of the implant can also have drastic effects on the
physical and mechanical properties of the polymer. Many con-
6.610.4. Total Disc Arthroplasty
ventional sterilization techniques such as steam sterilization
can cause melting of the polymer system and therefore cannot
The first mode of treatment proposed for the treatment of the
be utilized. Ionizing radiation is an alternative sterilization
majority of patients who suffer from disc degeneration is non-
operative care. On the basis of carefully considered guidelines,
Table 4 Chemical and physical characteristics of PLLA and PLDLLA surgical interventions are recommended in the event that
conservative remedies do not address the clinical needs of
PLLA PLLA (own 70/30 70/30
the patient. Arthrodesis or spinal fusion is the traditionally
(literature) use) PLDLLA PLDLLA
accepted surgical intervention technique advocated in cases
(literature) (own use)
of isolated disc degeneration in the cervical or lumbar spine.
Mn 240 500 140 250 The procedure involves the removal of a motion segment
(g mol1) through the use of bone grafts and sometimes through internal
Mw 395 500 253 300 fixation. Indications for the procedure are instability of a
(g mol1) motion segment caused by traumatic injury or degeneration.57
D (Mw/Mn) 1.64 1.81 A successful fusion procedure eliminates motion that causes
Intr. visc. 2.68 1.57 pain and offers the ability to restore intervertebral height and
(dl g1)
alignment.58
Inh. visc. 2.42 1.45
(dl g1)
However, spinal fusion has its limitations, some of which
Crystallinity 13.0 Amorphous Amorphous include the following41:
(%)
Tg ( C) 6065 6467 5560 6062
No significant improvement seen in the success rate of
fusion procedures despite the advances made in terms
Tm ( C) 173178 180 Amorphous Amorphous
Degr. time >24 3648 1218 >12
of surgical techniques and instruments;
(min) Morbidities associated with bone graft harvests during
the procedure;
Reproduced from Wuisman, P.; Smit, T. H. Eur. Spine J. 2006, 15, 133148. Accelerated adjacent level disc degeneration;
136 Orthopedic Surgery Spinal Treatment

Long postoperative recuperation times; 6.610.4.1.2. Bone implant interface


Increased costs of fusion procedures. Attachment of an implant to the endplate can be achieved
through the use of screws, serrations, or fins. These structures
Disc arthroplasty is a surgical procedure that has been getting
help insert the device, prevent early displacement, and facili-
a lot of attention in the spine community. The procedure aims to
tate long term stability. Long term attachment of the device can
potentially eliminate a painful disc while preserving/restoring
be facilitated by the presence of microstructured features on the
motion.58 The goals of arthroplasty include successful pain relief
surface of the implant in order to encourage bony ingrowth.
and functional recovery, acceptable levels of morbidity asso-
Other options for inducing bony ingrowth include titanium
ciated with the procedure (equal to or less than those seen in
spraying or meshes, calcium sulfate coatings, and hydroxyapa-
fusion), shorter postoperative recuperation times, and easily
tite coatings.58
implantable surgical techniques.41 The biomechanical objectives
of disc arthroplasty are the restoration of disc function to relieve
abnormal stress or strain caused by degeneration. More impor- 6.610.4.1.3. Bearing surfaces
tantly, it is essential that the procedure does not cause any Special attention is to be paid while selecting materials for the
detrimental biomechanical effects on surrounding structures bearing surfaces of the implant. The bearing surface usually deter-
within the motion segment and on adjacent levels.41 mines the long term stability of a well-fixed TDA implant. The
articulation of these surfaces must allow low friction motion,
resist permanent deformation, and show good wear resistance
6.610.4.1. Material Requirements
(see Chapter 6.614, Wear: Total Intervertebral Disc Prostheses).
In order to understand the material requirements, it is neces- Typically, surgical intervention to correct disc degeneration
sary to understand the biomechanics of disc arthroplasty. is performed on patients between the ages of 35 and 50 years.
The following are some of the biomechanical parameters that In order to reduce incidence of revision surgeries, the estimated
might be considered while designing a total disc replacement life of a prosthesis is proposed to be over 40 years. Taking into
implant59: consideration the number and amplitude of load cycles a
lumbar disc experiences in a year, it can be assumed that the
Restoration of mobility while avoiding segmental instability;
average adult bends 125 000 (extension and flexion move-
Restoration of correct spinal alignment;
ments) annually and takes about 2 million steps (gait cycles),
Protection of adjacent structures from overloading and
the device is expected to endure roughly 85 million cycles of
resultant accelerated degeneration;
loading during its lifetime without significant degeneration.60
Stability and wear properties of the device.
It is recommended that the device be tested for a 100 million
The material requirements for a TDA prosthesis are dis- cycles of loading. This hypothesis is made after taking into
cussed according to each component of the artificial disc to account the implants capacity to resist fatigue, and immuno-
be designed, along with some examples where possible. logic response to wear particles including their toxicity.
UHMWPE is used as a bearing material in several implants
owing to its low creep properties. Even though polyethylene
6.610.4.1.1. Implant endplate
shows wear debris in its use with other joints, animal studies of
The vertebral endplate is subject to repetitive physiological
metal-on-polyethylene disc implants have not shown signifi-
loads, and therefore, a material chosen for this component
cant evidence of wear particles.61,62
needs to be durable enough to withstand such types of
loads. Other aspects to be considered include the reactivity
of the material, its modulus of elasticity, ultimate tensile 6.610.4.1.4. General material specifications
strength, imaging characteristics, ease of manufacture, and Metals are considered to be good candidates for a total disc
cost.57 Commonly used materials for endplates are stainless replacement device owing to their inherent high fatigue
steel, cobaltchromium alloys, and titanium alloys.58 strength. Also, metal alloys have been shown to possess good
Stainless steel implants are inexpensive to fabricate, show biocompatibility.57 On the other hand, nonmetallic materials
relatively low corrosion, and have high moduli of elasticity. such as polymers and elastomers are also good candidates
Cobalt chromium alloys have properties in between those of because of the similarities between their mechanical properties
steel and titanium alloys. Depending on the alloy, processing and those of native disc tissue. Over a short period of time, the
of the end product can be tuned to a wide range of strengths lower elastic moduli of nonmetals help replicate disc behavior
and ductility, thus making these alloys versatile in their appli- adequately. However, in terms of a long term device, such
cability as implants. However, such implants are more ex- materials tend to pose a challenge to develop. Overall, the
pensive to manufacture.58 Currently used prostheses have need of the hour is to develop materials that demonstrate
endplates made out of these types of alloys. both biomechanical applicability and biocompatibility while
Titanium alloys are attractive as implant materials owing to being user friendly in a surgical environment.63
their high biocompatibility. They are also more corrosion In order to meet the needs for an artificial disc, combinations
resistant than the other two types of alloys. Further, they of both metallic and nonmetallic components have been
yield fewer debris particles as detected by imaging with mag- researched. In most cases, these take the form of metal
netic resonance. Their mechanical properties are closer to those polymermetal sandwiches, wherein the metal portion is used
of cancellous bone, but they have higher susceptibility to wear to improve fixation using screws/spikes or porous coatings to
debris generation. Pure titanium implants have not been stud- promote tissue ingrowth. The polymer component of the fixed
ied clinically, but titanium coatings are being used currently. implant can provide the flexibility seen in native discs.57
Trends in Materials for Spine Surgery 137

6.610.4.2. Discontinued Artificial Discs 6.610.4.2.1.2. Second generation


There is very little information available in the literature about
6.610.4.2.1. Acroflex
the second generation of the Acroflex device (Figure 12). In
The Acroflex artificial disc was developed in the 1970s by
this iteration of the design, the polymeric core was composed
Arthur Stefee. The device represents a leap in the concepts
of a silicone elastomer and clinical trials were performed in
and design of artificial discs.39,62 The design of the disc com-
eight patients. Data for this generation of studies are not avail-
prises two porous metal end plates coated with titanium and
able because of the controversies surrounding the use of the
fused to a hyperplastic polymer disc which is made up of a
silicone gel in breast implants. Class action lawsuits had been
hexane based polyolefin rubber.62,64 The design of the device is
filed against the manufacturer of the silicone (Dow Corning),
such that the vertebral bone is intended to grow into the
making the developers of the disc hesitate to publish their
titanium plates to ensure long term fixation, while the polymer
findings.62
disc is intended for restoration of flexibility and motion to the
functional spinal unit.39
6.610.4.2.1.3. Third generation
The Acroflex device had significant biomechanical advan-
The third generation of the Acroflex disc once again used a
tages. It helped in the preservation of disc height and allowed
hexane based polyolefin rubber core. The endplate design was
semiconstrained motion in all modes including compression.58
modified to reduce the profile to facilitate easier implantation.
Over the years, there have been three generations of the
The endplates were made from a Ti6Al4V ELI alloy, coated
Acroflex disc developed. Each generation of the device success-
with porous beads made of pure Ti.10 Mechanical testing was
fully passed the mechanical testing protocols to which they
performed on the device and concluded to be satisfactory.
were subjected, only to fail in the short term when implanted
Debris testing was not undertaken as the disc had no articulat-
in a series of limited human clinical trials.10,39
ing surfaces.65
Animal studies conducted on 20 baboons showed signifi-
6.610.4.2.1.1. First generation cant evidence of osteointegration into the porous coating after
The first generation of Acroflex was composed of CoCr alloy 6 months. There was one case of failure and histological reac-
and polyethylene as the polymeric core cemented together and tion because of the presence of wear particles in one animal.66
implanted through an anterior approach.39 The disc was tested Pilot human trials were conducted with two groups of
in cadaveric spines; however, concerns were raised regarding patients, with subtle design differences in the endplate of the
the bonecement interface, and the device idea was subse- device. Both sets of trials did not yield significant results to
quently abandoned.62 conclude the success of the device.10,66
A cementless design was developed in the 1980s using Eventually, even after three decades of changes to the design
titanium end plates with a double layer of sintered 250 mm and materials used in the Acroflex artificial disc, the device did
beads to enhance long term fixation. In this version of the disc, not achieve significant clinical success. The design concept of
the polymeric core was made out of a polyolefin rubber known using a one piece nonarticulating artificial disc is very attrac-
as Hexsyn, manufactured by Goodyear for fatigue resistant tive, as it would mimic natural disc function very well. How-
biomedical applications.39 ever, this is hindered by the complex loading environment
The device was subjected to preliminary mechanical testing, seen in the spine. The device was successful in terms of achiev-
and despite some incidences of local material failure, the tests ing rapid osteointegration. The biggest hurdle to the success of
were judged to be successful. The next step was animal testing; the Acroflex artificial disc though was the designers inability to
however, the company was never able to obtain permission to find a suitable polymeric core material to meet the required
carry out animal studies. Instead, on the basis of in vitro studies needs for an artificial disc.62
to study biocompatibility, a decision was made to proceed to
human trials directly.39
6.610.4.3. Currently Available Devices
The device was implanted in six patients and follow-
up studies were conducted. However, in 1990 the FDA 6.610.4.3.1. Charite artificial disc
raised concerns of the use of Hexsyn, which contained The Charite Artificial Disc (marketed by DePuy Spine), unlike
2-mercaptobenzothiazole, a potential carcinogenic.57 The disc the Acroflex was designed with the goal of replicating the
was eventually removed from the market. kinematics of the disc as opposed to the motion. The device

Figure 12 Second generation Acroflex implant with radiographic images post implantation. Reproduced from Melrose, J.; Smith, S. M.; Little, C. B.;
Moore, R. J.; Vernon-Roberts, B.; Fraser, R. D. Eur. Spine J. 2008, 17, 11311148.
138 Orthopedic Surgery Spinal Treatment

was designed by Kurt Shellnack M.D., and Karin Buttner- by 10% during testing. They claimed that there was minimal
Janz M.D., Ph.D, who were affiliated at the time to the wear and debris formation posttesting67; however, it has been
Charite Center for Musculoskeletal Surgery at the University seen that their claims are consistent with testing in a single
of Berlin.67 The basic design of the device incorporates two direction, which produces far less wear particles than testing in
metallic end plates made of cobalt-chromium alloy57 which multiple directions.
are attached to adjacent vertebral bodies, and articulate against
a central polymeric core made up of UHMWPE.67 6.610.4.3.1.2. SB Charite III
The UHMWPE core has two domed (convex) surfaces Waldemar Link GmbH and Co. was the company that com-
that help to articulate against the concave metallic endplates. mercialized the Charite artificial disc, resulting in the third
The hypothesis is that in conditions of full flexion or tension, generation SB Charite III (Figure 14). Major changes were
the core is free to translate or rotate when the polyethylene rim made to the endplate material, which was now constructed
of the core contacts the metal rims of the endplates. The device out of a proprietary CoCr alloy called VACUCAST (0.25% C,
is implanted through an anterior approach.67 The device pro- 2820% Cr, 5.56.5% Mo, max 0.5% Ni, max. 0.5% Fe,
vides unconstrained motion during rotation, and semi con- 0.41% Si, 1% Mn, and the remaining 57.169% Co). The
strained motion during flexion, extension, lateral bending, and back of the alloy was satin finished by corundum blasting,
translation, to some extent.58 and the device was fitted with six sharp teeth to achieve fixa-
The Charite Aritficial disc has undergone four iterations in tion. The UHMWPE core design remained unchanged.62
design since its inception and offers the largest and longest There have been a number of clinical studies undertaken to
clinical trial of any existing artificial disc.57,62 The rationale for document the performance of the SB Charite III. In summary,
choosing an UHMWPE core was because of the outstanding most of these studies have demonstrated that the Charite arti-
friction and fatigue properties of the material. ficial disc has the potential to preserve motion and survive long
term implantation in the body with good to excellent out-
6.610.4.3.1.1. SB Charite I and II comes. However, cohort studies undertaken have shown varia-
Only retrospective clinical data are available for the SB Charite bility in the complication and success rates of the device, which
I and II, the first two generations of the device. The first two raises concerns regarding the repeatability and reproducibility
generations were made up of stainless steel endplates, which of the procedure.
had protruding teeth for better fixation.62 In the SB I disc, the
end plates were circular and shaped like bottle caps, while in 6.610.4.3.2. ProDisc
the SB II they were expanded by providing lateral wings to The ProDisc (Synthes Spine, Paoli, PA) also consists of metallic
enhance contact with the vertebral end plates.62 The SB I and endplate with an UHMWPE core. Unlike the Charite, the Pro-
II discs were implanted in 15 and 22 patients respectively and Disc is attached firmly to the inferior end plate through a
follow-up data obtained after a period of 17.5 years. Follow-up locking mechanism. The convex surface of the polymer core
studies revealed a success rate of 87% (13/15) for the SB I disc, articulates against the superior end plate. There have been two
and 68% (15/22) for the SB II discs (Figure 13). The success was generations of the ProDisc, ProDisc I and II, but only the
defined in terms of the absence of long term complications.68 ProDisc II has been commercialized.69
The inventors performed mechanical testing on the devices The ProDisc has undergone the second largest clinical trial
and found that the height of the UHMWPE core was reduced of any artificial disc after the Charite. The ProDisc I had metal-
lic endplates made of Ti alloy, whose back surfaces were
plasma sprayed with a titanium coating to promote bone
growth.70 Follow-up reports for ProDisc I showed some proce-
dure related complications. Minimal wear of the UHMWPE
core was detected.70
Clinical studies on the ProDisc II have only been short term
(1731 months), with reports of positive outcomes. It is to be
noted that the studies for ProDisc are more scientifically
SB I SB II
sound than the studies for the Chartite disc, which has only
Figure 13 SB Charite I and II. been retrospectively studied in Europe. Studies on the ProDisc

SB III

Figure 14 SB Charite III, and radiographic image of the implanted device.


Trends in Materials for Spine Surgery 139

Figure 15 Components of the ProDisc. From left to right: superior endplate, polyethylene inlay, and inferior endplate. Source: Synthes
Product Literature.

Figure 16 Front and side views of the ProDisc-L artificial disc. Source: Synthes Product Literature.

Figure 17 The Maverick artificial disc. Reproduced from Mathews, H.; Lehuec, J.; Friesem, T.; Zdeblick, T.; Eisermann, L. Spine J. 2004, 4, 268275.

II however, are prospective and employ validated outcome The endplates have been treated with hydroxyapatite to pro-
measures such as the Oswestry Disability Index, which can mote bone growth.62
be compared with alternative spine procedures (Figures 15 Clinical trials of the Maverick disc began in 2002. Two-year
and 16).62 follow-up results have been reported so far. Clinical success
was reported in 75% of the patients. No revisions or device
6.610.4.3.3. Maverick related complications were reported in this study. Mechanical
The Maverick total disc replacement device was developed in testing on the disc showed good shock loading, and showed no
2001 by Medtronic Sofamor Danek (Figure 17). Initial proto- mechanical damage after 10 million loading cycles. However,
types were designed using alumina ceramic, but the commer- wear debris was produced, but the test methods were not
cially available device now features a two piece metal-on-metal published in the literature. On the basis of this, the life span
design, which comprises superior and inferior endplates of the device was predicted to be around 31.5 years in vivo, but
made of CoCr alloy. Bearing surfaces are made of CoCr alloy this claim cannot be validated until further studies are
based on ultra low wear rates obtained in hip replacements. conducted.71,72
140 Orthopedic Surgery Spinal Treatment

To sum, the Maverick disc is a good alternative to metal- the implant. The all metal implants show excellent wear resis-
polymer-metal disc replacements that are currently available. tance; however, the generation of wear debris is an issue of
Two-year follow-up studies and biomechanical testing have concern. The consequences of these wear particles and their
been reported. Metal-on-metal articulation produces less wear corrosion solubility products are unknown at present. There
than metal-on-UHMWPE prostheses; however, it also produces might also be concerns regarding the maintenance of bony
metallic wear debris. This can be a cause for concern. Only attachment with the degradation of bone tissue with age and
further studies can tell if the Maverick disc will be clinically osteoporosis with respect to these devices.
successful. Nevertheless, the future holds a lot of promise for this
field, and as more data are published on the available and
6.610.4.3.4. Flexicore developing implants, changes in design and materials can be
The Flexicore disc manufactured by Stryker Spine is the most conceptualized and implemented to meet the objective of
constrained of lumbar discs and consists of four CoCr achieving a successful implant that will meet all the needs
components: a superior and an inferior baseplate, a spherical of patients who require total disc replacement.
head that attaches to the superior endplate, and a shield
(Figure 18).73
The device is capable of supporting tensile loading because 6.610.5. Annulus Repair
of the presence of the spherical head. It also contains mechan-
6.610.5.1. Procedure
ical stops in order to limit flexion/extension and lateral bend-
ing. The device endplates are coated with titanium plasma The repair of the AF has seemed to be an intuitive idea for
spray coating to promote bone growth. Mechanical testing many researchers, but given the limitations posed by the cur-
has been conducted in tension, compression, and shear. The rent microscopic and endoscopic surgical techniques, this has
studies showed that the device is mechanically sound and has not been attempted much. Advanced technologies are being
strength limits excessive to the load limits for the IVD. Wear developed to close annulus defects. These could also enhance
testing showed that minimal or no wear damage was seen.73 the use of intradiscal prostheses which are contained by the AF.
The disc is undergoing a clinical trial at present, but the Efforts are also underway to develop devices that will support
results are not expected for several years. The success of the the annulus after a discectomy procedure.
Flexicore disc can only be determined when more clinical data
become available.
6.610.5.2. Mechanical Role
The AF of the IVD is a structure that serves to contain the
6.610.4.4. Future Trends
gelatinous central portion of the disc, the NP. The chemical
The field of total disc replacement is a rapidly evolving area of composition of the AF comprises 6590% water, 5070% dry
new technologies and techniques, and great strides have been weight of collagen, 1020% dry weight of PGs, and noncolla-
made in the development of spinal implants that are able to genous proteins such as elastin.19
replicate and preserve motion in the IVD. The arena of total disc The structure of the AF is laminate in nature, consisting of a
replacement has seen many leaps in terms of achieving short minimum of 15 (posterior) and 25 (lateral) concentric layers.
term benefits such as patient satisfaction and speed of recovery. The layers are made up of type I collagen fibers which alternate
New implants are being designed, developed, and tested. in angles from 28 at the peripheral AF to 44 at the central AF
The near future might hold answers regarding the long term with reference to the transverse plane of the disc. The concen-
benefits of using total disc replacements, especially the goal of tric rings of the AF are referred to as lamellae, and the spaces
reducing the incidence of adjacent segment degeneration. between them as interlamellar septae.19
More comprehensive follow-up studies will reveal the future The AF has a very highly organized structure which results
for these implants. in complex anisotropic behavior. The tensile, compressive, and
From a materials standpoint, there have been different shear properties of the AF differ in the axial, radial, and circum-
design concepts explored. The use of all-metal implants and ferential directions. The AF can be further divided into the
that of metalpolymer implants have been investigated. At this inner and outer AF on the basis of the structural differences,
juncture, it is hard to say which system works better. Each type the inner AF being subjected to higher hydrostatic pressures of
of design has its own inherent advantages and disadvantages; the NP than the outer AF which is subject to tensile forces.
for example, the main issue with metalpolymer implants is Age related degradation of annulus tissue manifests in the
the wear of the polymer core, which can cause early failure of form of annular tears and fissures, and can ultimately lead to

Figure 18 Components of the Flexicore artificial disc. Reproduced from Valdevit, A.; Errico, T. Spine J. 2004, 4, 276288.
Trends in Materials for Spine Surgery 141

disc herniation.74 Circumferential delamination and the for- for the same. He tested his method on 105 patients, and
mation of radial fissures can cause the extrusion of the NP recorded no reherniations, neurological impairment, or post-
material which is contained by the annulus.13 This is referred operative radiculopathy in them.83
to as disc herniation. Herniation can be defined as the loca- Lehmann et al. have reported suture related techniques to
lized displacement of disc material beyond the limits of the close annular defects by using 4-0 sutures to close the flaps of
intervertebral disc space.75 When this happens from the poste- the posterior longitudinal ligament, peridural membrane, and
rior region of the annulus, the release of chemical inflamma- outer fibers of the annulus. In a study conducted in 152
tory signals can result in severe pain in the back.13,76 patients, a larger percentage of patients who received sutures
Degeneration of the annulus results in a transfer of load had less postsurgery pain compared to those who did not
from the NP to the posterior portion of the annulus. receive them; however, the result was not statistically signifi-
To compensate for the added stress, the annulus tends to cant. The investigators did not report on the incidence of
broaden, which often causes significant protrusion of the reoperation or reherniation.84
disc. This manifests in the form of circumferential disc bulging Cauthen et al. have studied suturing of the annulus with a
or focal herniation. Tissue remodeling in the posterior annulus focus on reducing the rate of reoperations. He studied 254
is often accompanied by calcification and other changes that patients and reported that there was a 21% chance of recurrent
pose a challenge to the treatment of this issue.77 disc herniation when the patient did not receive sutures at
In the past, surgical intervention to reverse disc degenera- 2 years. The use of one suture to perform microsurgical sutur-
tion has focused on re-establishing disc height and geometry, ing showed a decreased recurrence rate of 10% while the use
while minimizing the flexibility of the disc. But recently, tech- of more than one suture brought it down to as low as 5%.85
niques that seek to retain or restore disc motion are being
investigated in a bid to avoid the elimination of spinal segmen- 6.610.5.3.2. Sealant and bulking techniques
tal motion. Some of these include the replacement of the total Biocompatible tissue adhesives have been used in many areas of
disc (i.e., the NP and a significant portion of the annulus) with the body containing soft tissue, for example, fascial hernias or
a prosthetic implant. An alternative to this approach is to cardiovascular procedures.86,87 Such procedures, however, have
replace the NP while retaining a major part of the annulus.78 not been extended to the AF because of the undesirable effects of
Studies have shown that an incision through the annulus adhesion that might occur with neural tissue in the vicinity.
can alter the mechanics of the involved disc. Hampton et al. Instead, sealing biomaterials, which are polymeric solutions
showed that incisions made in the posterior annulus showed that cure in situ, have been injected into the annulus.88 In
limited healing potential and the presence of a persistent defect concept, these bulking hydrogels are intended to form a sealant
could form a pathway for the leakage of nuclear fluid into which will take on the complex, irregular shape of the AF defect,
surrounding perineural tissue.76 This view was reinforced by and may bond strongly to the neighboring tissue around the
other researchers who investigated annulotomy techniques in defect. There have been no clinical reports of such a technique
animal models. They found that a box type or window annu- being applied to the annulus; instead, these types of biomater-
lotomy technique produces a significantly weaker healing ials seem more applicable to nucleus replacement.89,90
response than a slit or cruciate incision.79,80
On the basis of the results reported, it can be inferred that 6.610.5.4. Current Technologies
the AF has a limited capacity for regeneration after an annu-
lotomy. On the basis of the type of annulotomy performed, the 6.610.5.4.1. Xclose tissue repair system
healing process leads to the formation of a thin layer of inferior The Xclose Tissue Repair System marketed by Anulex Tech-
fibrous tissue. The poor regenerative capacity may be due to the nologies, Inc. is an FDA approved technology, and is recom-
fact that the external repairs do not meet the needs of fiber mended for reapproximation of soft tissue such as the AF after
recruitment to withstand the tensile forces applied to the annu- a discectomy procedure (Figure 19). The system consists of two
lus.81,82 In order for the annulus to shift between axial loading tension band devices and one knot pusher. The device works
and circumferential tension, it is necessary that the nucleus by placing tension bands at the annulotomy site and securing
volume remains elastic, deformable, and contained. When them with tension bands.78
the annulus suffers damage from an annulotomy or degrada- 6.610.5.4.1.1. Material
tion, its ability to retain the nucleus is lowered.19 In light of All implanted components of the device are made up of poly-
the above discussion, we can say that preservation of as much ethylene terephthalate.78,91
annular tissue as possible should be the goal during surgical
interventions to treat disc degeneration. 6.610.5.4.1.2. Procedure
All the components of the kit are provided in sterile conditions
and preloaded on disposable delivery instruments. A tension
6.610.5.3. Reported Treatment Methods
band consists of a pair of nonabsorbable suture loops each of
6.610.5.3.1. Suture techniques which is attached to a T-anchor. A third loop of suture with a
There have been reports of closing annular defects with sutures pretied knot is then used to connect the two loops together,
after a microdiscectomy. The first instance of suture usage to while facilitating the tightening of the tension band, which
close annular defects was reported by Yasargil in 1977, wherein helps to draw the construct together and hence closes the defect
he used 7-0 sutures to close a defect made to remove nucleus in the tissue.78
material. He reported that this method could help prevent Cadaveric experiments have been performed to study the use
adhesions in the annulus, but did not suggest any mechanism of the Xclose system. The results were presented at the annual
142 Orthopedic Surgery Spinal Treatment

meeting of the Spine Arthroplasty Society in Berlin. The ability of not detrimentally affect the repaired tissue, and the system has
the system to remain in place after cyclic loading of the repaired shown stability upon cyclic loading. So far, the Xclose system
tissue was analyzed. A partial laminectomy was performed to has been implanted in over 6000 patients in the United States.93
make a posterior annular incision. A limited amount of nucleus The Xclose system is undergoing a Phase IV randomized
was removed (0.2 cm3), and the incision was repaired using clinical trial at present, but no clinical data are currently avail-
the Xclose system. Nine spinal units were tested in this study.92 able. In 2009, the company also launched a Post Market Clinical
Cyclic fatigue loading was performed, consisting of 20 K study in which 750 patients from 34 facilities around the United
cycles of flexionextension, 15 K cycles of lateral bending, States were enrolled.93 On the basis of data from the clinical
and 5 K cycles of axial rotation. Calibrated digital macropho- trials, more can be said about the materials that are used to
tography was used to assess the systems ability to keep the construct the device and their role in the body post implantation.
repaired tissue in place, and the knot slippage was measured to Until data from more long term studies in vivo are made avail-
check for loosening of the suture knots.78 able, the final understanding of this approach cannot be
The study showed that all annulotomy defects remained formulated.
closed and intact after 40 K cycles of loading. Post cyclic load-
ing, the loosening in the system was minimal and knot slip- 6.610.5.5. Barrier Technologies
page was measured to be 0.8  0.7 mm. Also, the system did
not migrate out of the annulus or pull out of the tissue.92 6.610.5.5.1. Inclose surgical mesh system
On the basis of the results of the study, we can infer that the The Inclose Surgical Mesh System is marketed by Anulex
Xclose system is a simple and effective method to close annu- Technologies (Minnetonka, Minnesota) (Figure 20). The
lar defects incurred during a discectomy. The materials used do device is designed to act as a barrier and provide a scaffold
for the repair of soft tissue such as the AF.94 The surgical mesh
implant is made up of a monofilament polymeric braid which
is preloaded on a disposable delivery device.

6.610.5.5.1.1. Materials
The braided mesh is biocompatible, expandable, and com-
posed of polyethylene terephthalate,91 a commonly used poly-
mer in medical devices.75

6.610.5.5.1.2. Procedure
The mesh is a low profile device which enables it to be inserted
into the annulus in a minimally invasive manner. The device is
inserted using a preloaded delivery device supplied with the
kit. The system is secured in place using nonabsorbable surgi-
cal suture anchors.
The cylindrical device is mounted on a delivery instrument
which allows it to extend in a circumferential direction to form
a barrier by latching the cylinder ends together. The mesh
device can be inserted into any tissue aperture, such as an
Figure 19 Schematic of the Xclose Tissue Repair System. annular defect, and once deployed into position, it can be
Reproduced from http://www.anulex.com/anulex_technology/xclose.asp. tethered using sutures.75

Figure 20 Inclose Surgical Mesh System and the configuration of the device post implantation (http://anulex.com/anulex_technology/
inclose_us.asp).
Trends in Materials for Spine Surgery 143

6.610.5.5.1.3. Studies on the device anchors the annulus and reinforces it while closing the
In vitro biomechanics studies were conducted using the device, defect.104 The frame functions as a shape memory alloy, and
and demonstrated that implantation of the Inclose system did exhibits pseudo elasticity (Intrinsic Therapeutics).
not alter segmental spinal biomechanics. The study also showed
that the implant position remained intact despite application of 6.610.5.5.2.2. Procedure
complex cyclic loading.95 In theory, a stable spinal motion seg- The Barricaid Prosthesis is implanted into the annulus
ment may be able to reduce risks of herniation in an adjacent disc through a small incision (5  10 mm defect) made on the
by avoiding stress redistribution. Also, this study suggests that the posterior side, as part of a standard limited discectomy, just
repair of the annulus using such a device might indirectly avoid prior to wound closure (Intrinsic Therapeutics). The device
the need of nucleus material removal which is often a solution covers a major portion of the medialateral distance in the
employed by surgeons to mitigate the effects of disc herniations. posterior AF, and expands cephaladcaudad postinsertion.75
Animal studies on intradiscal implants are faced with com-
plications of downsizing the device in order to accommodate 6.610.5.5.2.3. Studies on the device
smaller tissue areas. In general, no animal models exist that can In vitro biomechanical studies have been performed to evaluate
accurately represent the exact nature of disc space characteristics the Barricaid device, and they have demonstrated that the rate
as seen in humans and correlate to human clinical use. Despite of recurrent herniations is decreased.99 The ability of the device
these limitations, animal models were designed to study the to withstand intradiscal pressure was demonstrated in a study by
surgical implantation of the Inclose device and histological subjecting the device to complex applied loads, and measuring
response to the PET mesh. the intradiscal pressures.100 The clinical use of the device was
Peppelman et al. studied Inclose implantation in a Nubian also tested, and the study demonstrated that compared to
goat model and showed that a lateral approach could be used to patients who were not implanted with the prosthesis, patients
place the implant safely in the proximity of the annulus. When who received the device after a standard discectomy were seen to
positioned and affixed properly, the implant remained intact experience restoration in disc height lost because of the proce-
over the course of 12 weeks, as seen using radiography. Also, the dure.101 However, the evaluation of the long term effects of the
device was found to be incorporated histologically without any device in a clinical setting will require longer follow-up studies.
detrimental effects on the surrounding tissues.96
The device was studied after implantation in cadaveric and 6.610.5.5.3. SpineJet XLW
animal models. On the basis of a report submitted at the Spine The SpineJet HydroSurgery system is a novel technology that
Arthroplasty Society in Berlin in 2007, the studies demonstrated harnesses the power of water in medical applications such as
that the presence of wear particulates was minimal. Also, motion surgery (Figure 21). The system comprises a controlled hair-
of the implant was negligible and it had no detrimental effects on thin supersonic stream of water that can be used as an effective
spinal flexibility. Early findings showed that there were no recur- cutting, ablation, and collection device. The advantage of such
rent herniations when the device was implanted.97 a system is that it has the power of laser technology and radio-
Early clinical use of the Inclose mesh device has shown frequency devices, but avoids the damage to tissue that such
promising results. Around 10 patients were implanted with the technologies cause.102 The SpineJet XL is a percutaneous
device after an endoscopic L4-L5 or L5-S1 discectomy. The hydrodiscectomy device that can be used in the minimally
implant was inserted with minimal disruption to the annulus, invasive surgical intervention of intervertebral disc degenera-
with access openings between 3 and 8 mm. There was inade- tion. The device aspirates tissue as it is being used, and can be
quate length of follow up, because of which long term effects of used to remove cartilage and nucleus material.
the implant cannot be predicted yet. However, it was noted The device is designed with four cutting surfaces. Annular
that the best technical results were obtained in patients with thinning is accomplished using a windshield wiper motion to
focal, protruded, and/or extruded disc herniations.98 scrape the heel and toe of the device along the inner surface of
the annulus. Lateral cutting sides are used to scrape cartilage
off the endplates to increase the surface area for interbody
6.610.5.5.2. BarricaidW prosthesis
lumbar fusion.102
The Barricaid Prosthesis device is currently not clinically
The device is mainly used for the removal of NP, in order to
available in the United States. It is a device aimed at recon-
perform spinal fusion. The systems use of cold fluid prevents
struction of the annulus in the region of disc herniation. The
thermal damage to tissue. The design of the device is such that
Barricaid implant works by forming a strong and flexible wall
it enables the removal of precise amounts of nucleus tissue
between the AF and NP, where it is implanted.
without annular puncture or endplate damage.102
The salient features of the prosthesis include its ability to
The device underwent a trial on 13 patients in 20082009.103
create a mechanical barrier that closes the annulus defect, while
Initial results suggested that percutaneous hydrodiscectomy
allowing more nucleus material to be retained in the disc.
could be a good alternative for the treatment of lumbar degener-
According to the manufacturers, the device can be implanted
ative disc disease. However, more clinical data are required to
quickly, simply, and securely (Intrinsic Therapeutics).104
properly assess the device before more widespread use.

6.610.5.5.2.1. Material
6.610.5.6. Material Requirements
The Barricaid Prosthesis consists of a woven mesh made of
polytetrafluoroethylene (PTFE) supported by a nitinol frame. On the basis of the discussion of current materials/technolo-
The frame is made up of a nickeltitanium alloy base, which gies, we can say that closure of an annulus defect using suture
144 Orthopedic Surgery Spinal Treatment

Saline
Venturi
suction
effect

In flow
tube

Saline
stream
Evacuation
Waste
tube

Figure 21 Schematic of the SpineJet XL hydrodiscectomy system and the complete setup.

approximations would require materials that are nonabsorb- development of novel materials and techniques in surgical
able, given the nature of an annular defect. With respect to the spine interventions. Composite materials and polymers have
barrier type of technologies that have been developed so far, been utilized in spinal implants as an improvement over metals
the material in question should be able to support the annulus which do not have matching mechanical properties to the
appropriately on the basis of the biomechanical environments spinal anatomy and are subject to wear. A better understanding
it is subject to, without the device getting damaged or causing of spine anatomy, physiology, and pathophysiology has also
damage to the annulus; that is, the material has to have led to the development of new materials with several materials
mechanical properties matching those of the annulus. Any being developed specifically for application in the spine. Failure
motion of the device due to functioning of the annulus should of earlier devices has especially expanded the breadth of mate-
not generate any wear particles that might bring about adverse rials utilized in spinal interventions. Adjacent level disc degen-
effects in the tissue. Further, considering that such types of eration with spinal fusion, expulsion of implant materials
techniques are limited by the method of implantation of the through surgical sites, and subsidence of implants into the
device, it is also required that the material in question has endplates have driven the development of new surgical techni-
properties that enable molding it in a manner that facilitates ques and provided the impetus for the development of novel
easy and efficient implantation into the AF. spine materials. This is especially illustrated by the develop-
ment of in situ curing hydrogel and nonhydrogel nucleus repla-
cements where materials were designed such that they could be
6.610.5.7. Future Trends
delivered with minimal damage to the AF, a major way for
The field of annulus repair is still in its infancy, with data still nucleus replacement as well as the development of sealants
being collected on implantation of devices that are currently for the AF itself. Another drive in materials for spine implants
available in the market. However, preserving the integrity of is the utilization of tissue engineering approaches. Tissue scaf-
the annulus is an important consideration while repairing the folds that support nucleus and AF cell growth are under investi-
intervertebral disc, as it is the only means of gaining access to gation; however, major limitations in nutrient supply to the
the NP. On the basis of the discussion presented above, it is implant after implantation must be addressed before tissue
intuitive to try to preserve as much of the annulus as possible as engineering approaches for the IVD can be realized.
the annulus is responsible for the mechanical properties of the
disc. As more follow-up data become available, investigators
can look to modifying the currently available technologies to References
refine their applicability, as well as lay the foundation for the
design and development of newer devices that can be 1. Atlas, S. J.; Deyo, R. A. J. Gen. Intern. Med. 2001, 16, 120131.
2. Martin, B. I.; Deyo, R. A.; Mirza, S. K.; et al. JAMA 2008, 299, 656664.
implanted minimally invasively, which will help retain the 3. Kuslich, S. D.; Ulstrom, C. L.; Michael, C. J. Orthop. Clin. North Am. 1991, 22,
properties of the annulus while allowing access to the NP. 181187.
4. Benzel, E. C. Biomechanics of Spine Stabilization. American Association of
Neurological Surgeons: Rolling Meadows, IL, 2001.
6.610.6. Concluding Remarks 5. Knoeller, S. M.; Seifried, C. Spine (Phila Pa 1976) 2000, 25, 28382843.
6. Klenerman, L. The Evolution of Orthopaedic Surgery. RSM Press: Oxon, UK, 2002.
7. Rutherford, E.; Tarplett, L.; Davies, E.; Harley, J.; King, L. Radiographics 2007,
Early spinal implants were based on metals because of the 27, 1737.
utility of metals in other orthopedic devices. The complex load- 8. Sakalkale, D. P.; Bhagia, S. A.; Slipman, C. W. Pain Physician 2003, 6, 195198.
ing nature and function of the spine however necessitated the 9. Coric, D.; Mummaneni, P. V. J. Neurosurg. Spine 2008, 8, 115120.
Trends in Materials for Spine Surgery 145

10. Melrose, J.; Smith, S. M.; Little, C. B.; Moore, R. J.; Vernon-Roberts, B.; 63. Edeland, H. J. Biomed. Eng. 1985, 7, 57.
Fraser, R. D. Eur. Spine J. 2008, 17, 11311148. 64. Szpalski, M.; Gunzburg, R.; Mayer, M. Eur. Spine J. 2002, 11, 6584.
11. Deramond, H.; Depriester, C.; Galibert, P.; Le Gars, D. Radiol. Clin. North Am. 65. Serhan, H. Personal Communication, 2005; DePuy Spine, Raynham, MA.
1998, 36, 533546. 66. Cunningham, B.; Lowery, G.; Serhan, H.; et al. Eur. Spine J. 2002, 11, 115123.
12. Galibert, P.; Deramond, H.; Rosat, P.; Le Gars, D. Neurochirurgie 1987, 33, 166. 67. Buttner-Janz, K.; Hochschuler, S.; McAfee, P. The Artificial Disc. Springer: Berlin,
13. Thongtrangan, I.; Le, H.; Park, J.; Kim, D. H. Neurosurg. Focus 2004, 16, E13. 2003.
14. White, A. A.; Panjabi, M. M. Clinical Biomechanics of the Spine; Lippincott: 68. Putzier, M.; Funk, J.; Schneider, S.; et al. Eur. Spine J. 2006, 15, 183195.
Philadelphia, PA, 1990. 69. Delamarter, R.; Bae, H.; Pradhan, B. Orthop. Clin. North Am. 2005, 36, 301313.
15. Lieberman, I. H.; Togawa, D.; Kayanja, M. M. Spine J. 2005, 5, 305S316S. 70. Tropiano, P.; Huang, R.; Girardi, F.; Cammisa, F., Jr.; Marnay, T. J. Bone Joint
16. Errington, R. J.; Puustjarvi, K.; White, I. R. F.; Roberts, S.; Urban, J. P. G. J. Anat. Surg. 2005, 87, 490.
1998, 192, 369378. 71. Lehuec, J.; Kiaer, T.; Friesem, T.; Mathews, H.; Liu, M.; Eisermann, L. Spine
17. Raj, P. P.; Fipp, A. Pain Pract. 2008, 8, 18. 2003, 28, 346.
18. Kiani, C.; Liwen, C.; Yao Jiong, W. U.; Yee, A. J.; Burton, B. Y. Cell Res. 2002, 72. Mathews, H.; Lehuec, J.; Friesem, T.; Zdeblick, T.; Eisermann, L. Spine J. 2004,
12, 1932. 4, 268275.
19. Bron, J.; Helder, M.; Meisel, H.-J.; Van Royen, B.; Smit, T. Eur. Spine J. 2009, 73. Valdevit, A.; Errico, T. Spine J. 2004, 4, 276288.
18, 301313. 74. Hickey, D.; Hukins, D. Spine 1980, 5, 106.
20. Adams, M.; Roughley, P. Spine 2006, 31, 2151. 75. Kenneth Yonemura, J. S.; Peppelman, W., Jr.; Griffith, S.; Davis, R,.;
21. Roughley, P. J. Spine (Phila Pa 1976) 2004, 29, 26912699. Cauthen, J. C., III. In Spinal Reconstruction: Clinical Examples of
22. Masuda, K.; Lotz, J. C. Tissue Eng. B Rev. 2010, 16, 147158. Applied Basic Science, Biomechanics and Engineering, 1st edn.; Kai-Uwe
23. Le Maitre, C.; Pockert, A.; Buttle, D.; Freemont, A.; Hoyland, J. Biochem. Soc. Lewandrowski, M. J. Y. E., Kalfas, I., Park, P., Mclain, R. F., Trantolo, D. J., Eds.;
Trans. 2007, 35, 652655. Informa Healthcare: New York, 2007.
24. Urban, J. P. G.; Smith, S.; Fairbank, J. C. T. Spine 2004, 29, 2700. 76. Hampton, D.; Laros, G.; Mccarron, R.; Franks, D. Spine 1989, 14, 398401.
25. Carragee, E. J. N. Engl. J. Med. 2005, 352, 18911898. 77. Adams, M.; Mcnally, D.; Dolan, P. J. Bone Joint Surg. Br. 1996, 78,
26. Martz, E. O.; Goel, V. K.; Pope, M. H.; Park, J. B. J. Biomed. Mater. Res. 1997, 38, 965972.
267288. 78. Griffith, S., Davis, R., Hutton, W., Eds.; In Repair of the Anulus Fibrosus of the
27. Vaccaro, A. R.; Singh, K.; Haid, R.; et al. Spine J. 2003, 3, 227237. Lumbar Disc; Raymedica, LLC: USA, 2007.
28. Kuslich, S. D.; Bagby, G. The BAK interbody fusion system: Early clinical results 79. Ahlgren, B.; Lui, W.; Herkowitz, H.; Panjabi, M. Spine 2000, 25, 2165.
of treatment for chronic low back pain. In 8th NASS Annual Meeting. San Diego, 80. Ethier, D.; Cain, J.; Yaszemski, M.; et al. Spine 1994, 19, 2071.
USA, 1993; pp 175176. 81. Guerin, H.; Elliott, D. J. Orthop. Res. 2007, 25, 508516.
29. Matge, G. Acta Neurochir. 1998, 140, 18. 82. Iatridis, J. C. P.; Weidenbaum, M. M. D.; Setton, L. A. P.; Mow, V. C. P.
30. Black, J. Orthopaedic Biomaterials in Research and Practice. Churchill Spine 1996, 21(10), 11741184.
Livingstone: New York, 1988. 83. Yasargil, M. Adv. Neurosurg. 1977, 4, 81.
31. Park, J. B.; Lakes, R. S. Biomaterials: An Introduction; Springer: Berlin, 2007. 84. Lehmann, T. R.; Titus, M. K. Refinements in technique for open lumbar
32. Waisbrod, H. Arch. Orthop. Trauma Surg. 1988, 107, 222225. discectomy. In Proceedings of the International Society for the Study of the
33. Merritt, K.; Brown, S. A. Biological Effects of Corrosion Products from Metals. Lumbar Spine (ISSLS), June 1997.
ASTM International: Philadelphia, PA, 1985; p 195. 85. Cauthen, J. Spinal Arthoplasty: A New Era in Spine Care; Quality Medical:
34. Merritt, K.; Brown, S. A. Clin. Orthop. Relat. Res. 1996, S233S243. St. Louis, MO, 2005; pp 157177.
35. McMillin, C. R.; Brantigan, J. W. Soc. Adv. Mater. Process Eng. 1993, 582590. 86. Basu, S.; Marini, C.; Bauman, F.; et al. Ann. Thorac. Surg. 1995, 60, 1255.
36. Kurtz, S. M.; Devine, J. N. Biomaterials 2007, 28, 48454869. 87. Miyano, G.; Yamataka, A.; Kato, Y.; et al. J. Pediatr. Surg. 2004, 39,
37. Hay, J. N.; Kemmish, D. J. Polymer 1987, 28, 20472051. 18671870.
38. Katoozian, H.; Davy, D. T.; Arshi, A.; Saadati, U. Med. Eng. Phys. 2001, 23, 505511. 88. Haldimann, D. System for repairing inter-vertebral discs. US Patent App.
39. Brantigan, J. W.; Steffee, A. D. Spine (Phila Pa 1976) 1993, 18, 21062107. 10/207,285, 2002.
40. McMillin, C. R. Soc. Adv. Mater. Process Eng. 1993, 591598. 89. Bao, Q.; Yuan, H. Spine 2002, 27, 1245.
41. Cho, D.-Y.; Lee, W.-Y.; Sheu, P.-C. Surg. Neurol. 2004, 62, 378385. 90. Carl, A.; Ledet, E.; Yuan, H.; Sharan, A. Spine J. 2004, 4, 325329.
42. Ferguson, S. J.; Visser, J. M. A.; Polikeit, A. Eur. Spine J. 2006, 15, 149156. 91. Wilke, H.-J. P.; Neef, P. M. D.; Caimi, M. M. D.; Hoogland, T. M. D.;
43. Toth, J. M.; Wang, M.; Estes, B. T.; Scifert, J. L.; Seim, H. B., III; Turner, A. S. Claes, L. E. P. Spine 1999, 24, 755762.
Biomaterials 2006, 27, 324334. 92. Bourgeault, C.; Beaubien, B. P.; Freeman, A. L.; Bentley, I.; Houfburg, R.;
44. Mastronardi, L.; Ducati, A.; Ferrante, L. Acta Neurochir. 2006, 148, 307312. Griffith, S. L. In Biomechanical Assessment of Anulus Fibrosus Repair with
45. Togawa, D.; Bauer, T. W.; Brantigan, J. W.; Lowery, G. L. Spine 2001, 26, 2744. Suturetethered Tissue Anchors. Proceedings of the 2007 Annual Meeting of the
46. Senegas, J. Eur. Spine J. 2002, 11, 164169. Spine Arthroplasty Society (SAS7), Berlin, Germany, May 2007.
47. Lemons, J. E.; Lucas, L. C. J. Arthroplasty 1986, 1, 143147. 93. Anulex Technologies, Inc. Completes Patient Enrollment Phase of Post Market
48. Gaines, R. W., Jr.; Abernathie, D. L. Spine 1986, 11, 907. Study and Announces the Treatment of over 6000 Patients with the Xclose Tissue
49. OBrien, J. P.; Stephens, M. M.; Prickett, C. F.; Wilcox, A.; Evans, J. H. Clin. Repair System, 2009. http://www.anulex.com/about_anulex/news_events.asp.
Orthop. Relat. Res. 1986, 203, 168. 94. Anulex Technologies, Inc. Anulex Technologies Receives FDA 510(k) Clearance
50. Liu, Y.; Njus, G. O.; Bahr, P. A.; Geng, P. O. Spine 1990, 15, 311. for the Inclose Surgical Mesh System. 2005 [Online].
51. Kokubo, T.; Yamaguchi, S. Materials 2009, 3, 4863. 95. Cunningham, B. C.; Hu, N. B.; Beatson, H.; McAfee, P. In Fifth Global Symposium
52. Wuisman, P.; Smit, T. H. Eur. Spine J. 2006, 15, 133148. on Motion Preservation Technology; Author, C. D., Ed.; Spine Arthroplasty
53. Gogolewski, S.; Mainil-Varlet, P.; Dillon, J. G. J. Biomed. Mater. Res. 1996, 32, Society: New York, 2005.
227235. 96. Peppelman, W. C.; Kraus, D. R.; Sherman, J.; Yonemura, K.; Griffith, S. L.;
54. Nuutinen, J.-P.; Clerc, C.; Virta, T.; Tulamo, R. M.; Tormala, P. J. Biomater. Sci. Cauthen, J. In Feasibility Results of a Novel Annular Repair Device in a Goat
Polym. Ed. 2002, 13, 13251336. Model. World Spine III, Rio de Janeiro, Brazil, 2005.
55. Jun, S.; Harold, A. J. Appl. Biomater. 1993, 4, 1327. 97. Guillermo Bajares, A. P.; Diaz, Manuel; et al. Spine Arthroplasty Soc. 2007, 7.
56. Nabhan, A. M. D.; Ishak, B. C. M.; Steimer, O. M. D.; et al. J. Spinal Disord. Tech. 98. Bajaras, G.; Perez-Olivia, A. In Fifth Global Symposium on Motion Preservation
2009, 22, 155161. Technology; Spine Arthroplasty Society (SAS): New York, 2005.
57. Bao, Q.; Mccullen, G.; Higham, P.; Dumbleton, J.; Yuan, H. Biomaterials 1996, 99. Gorensek, M.; Vilendecic, M.; Eustacchio, S.; et al. Spine J. 2006, 6, 144.
17, 11571167. 100. Yeh, O.; Chow, S.; Small, M.; Einhorn, J.; Lambrecht, G. Spine J. 2006, 6, 146147.
58. White, A.; Lawrence, J.; Grauer, J. In Spinal Reconstruction: Clinical Examples of 101. Emir Kamaric, O. Y.; Velagic, A.; Einhorn, J.; Osman, G.; Vilandecic, M.;
Applied Basic Science, Biomechanics and Engineering, Informa Healthcare: Lambrecht, G. Spine J. 2005, 5, S178S179.
New York, 2007; p 263. 102. HYDROCISION, The SpineJet System. 2010. http://www.hydrocision.com.
59. Galbusera, F.; Bellini, C.; Zweig, T.; et al. Eur. Spine J. 2008, 17, 16351650. 103. Han, H. J.; Kim, W. K.; Park, C. K.; Oh, S. H.; Lee, D. G.; Doh, E. S. Kor. J. Spine
60. Hedman, T. P.; Kostuik, J. P.; Fernie, G. R.; Heller, W. G. Spine 1991, 16, S256. 2009, 6, 187191.
61. Cunningham, B.; Dmitriev, A.; Hu, N.; McAfee, P. Spine 2003, 28, S118. 104. Intrinsic Therapeutics, Inc. The BarricaidW Prosthesis: Optimizing Outcomes for
62. Kurtz, S. M.; Edidin, A. A. Spine Technology Handbook. Academic Press: Discectomy Patients. Available at http://www.intrinsic-therapeutics.com/
New York, 2006. barricaid_ard.shtml

You might also like