You are on page 1of 8

Analytica Chimica Acta 503 (2004) 187194

Improving the lower detection limit of potentiometric sensors


by covalently binding the ionophore to a polymer backbone
Martin Pntener, Tams Vigassy, Ellen Baier, Alan Ceresa, Ern Pretsch
Laboratorium fr Organische Chemie, ETH Hnggerberg, CH-8093 Zrich, Switzerland

Received 2 July 2003; received in revised form 30 September 2003; accepted 14 October 2003

Abstract

The lower detection limit of polymeric membrane ion-selective electrodes (ISEs) is impaired by zero-current ion fluxes through the
organic phase. This adverse effect is largely eliminated by covalently attaching the ionophore to a polymer backbone. To this purpose, the
Pb2+ -selective ligand, 4-tert-butylcalix[4] arene-tetrakis(N,N -dimethylthioacetamide) is substituted on its upper rim by a diol derivative
which is subsequently copolymerized with poly(tetrahydrofuran)diol and 2,2,4-trimethylhexamethylene diisocyanate to the corresponding
polyurethane. By measurements on sandwich membranes, it is shown that through binding the ionophore to the polymer, the mobility of Pb2+
in the ISE membrane is strongly reduced. As a consequence, the response range of such an ISE is extended by several orders of magnitude.
This is the case even when using an internal electrolyte that with an ISE based on a mobile ionophore leads to strong deviations from the linear
response because of ion uptake from the sample into the membrane or ion release from the membrane into the sample. With a conventional
inner filling solution of 101 M Pb(NO3 )2 , a lower detection limit of 1.7 109 M Pb2+ has been achieved in the presence of 104 M Na+ .
2003 Elsevier B.V. All rights reserved.

Keywords: Potentiometric sensor; Polymeric ionophore; Detection limit; Ion fluxes

1. Introduction sample [810]. The lower detection limits and selectivities


have been found to improve by many orders of magnitude
During recent years, it has been well established that the when compensating such gradients through careful adjust-
lower detection limit and selectivity behavior of traditional ment of the composition of the internal solution [9,1114]
polymer membrane ion-selective electrodes (ISEs) [1,2] or by applying a weak electric current [1517]. However, as
have been adversely affected by zero-current transmembrane revealed by a detailed theoretical analysis [9,10,14,18], per-
fluxes of the measuring (primary) ions [3]. Already very fect compensation of transmembrane ion fluxes is only pos-
small changes of <1% in the total ionic concentration of the sible for individual sample compositions and any changes
ISE membrane are sufficient to induce leaking of primary in them inevitably leads to concentration gradients in one
ions into the sample and thus cause the concentration in the or the other direction, and, consequently, to a biased ISE
contacting aqueous layer to be ca. 106 M, even if the bulk response. For trace analysis of environmental samples [19],
sample virtually did not contain primary ions. Consequently, ISEs have been optimized by a twofold strategy: (1) by
the lower detection limit was found to be around 106 M adjusting the composition of the inner solution so as to
and the best selectivity coefficients were around 103 to induce an ion exchange similar to that caused by a typical
104 in practically all reported cases [4,5]. Usually, the ori- sample to be measured, and (2) by reducing transmembrane
gin of concentration gradients in the ISE membrane is the ion fluxes through optimizing the membrane composition
coextraction of primary ions together with their counterions [2022]. The topic of this contribution is to investigate
from the internal solution [6,7] and/or the exchange of pri- the influence on the response behavior of ISEs when the
mary ions from the membrane by interfering ions from the ionophore is covalently bound to a polymer backbone.
Covalent bonding of electrically charged [2325] and
neutral [26] ionophores to polymer backbones has already
Corresponding author. Tel.: +41-1-632-29-26; fax: +41-1-632-11-64. been reported many years ago. While the primary aim of
E-mail address: pretsch@org.chem.ethz.ch (E. Pretsch). those contributions was to improve the lifetime of ISEs,

0003-2670/$ see front matter 2003 Elsevier B.V. All rights reserved.
doi:10.1016/j.aca.2003.10.030
188 M. Pntener et al. / Analytica Chimica Acta 503 (2004) 187194

the extension of the linear response range by immobilizing a 5 h Soxhlet extraction (4 molecular sieve), discarding
the ionophore was also noticed [24,26]. However, since the the first portion (30 ml) of the distillate. After the water
lipophilicity of the parent mobile ionophores was rather content was determined to be lower than 5 ppm, the reac-
poor, the reason for this improvement was not necessarily tion mixture was cooled to 70 C. Freshly distilled TMDI
due to the reduction in transmembrane ion fluxes [27]. Sub- (0.20 g, 0.94 mmol, 5 equiv.) was added and the reaction
sequently, different authors [2839] have covalently bound mixture was copolymerized at 70 C for 7 days. Once
ionophores but again in order to improve the lifetime of the GPC analysis showed no further increase in polymer chain
corresponding ISEs or to inhibit the leakage of components length after adding three drops of TMDI to the reaction
into the sample during in vivo measurements [25,32], but not mixture, abs. MeOH (0.2 ml) was added and the mixture
to attain better lower detection limits. More recently, we have was stirred for 1 h. The product was then slowly precipi-
shown that the mobility of ionophores and transmembrane tated with hexane cooled to 0 C (300 ml) under vigorous
ion fluxes can be drastically reduced by covalent bonding of stirring and stirring was continued for 2 h. The precipi-
a H+ -selective ionophore to polyurethane [40]. In the present tate was redissolved in CH2 Cl2 (50 ml) and precipitated
work, we describe the immobilization of the ionophore, twice with cooled hexane (0 C, 300 ml) obtaining the
4 - tert - butylcalix[4]arene - tetrakis(N,N -dimethylthioacet- ionophore 3 (0.72 g, 37% yield) after drying for 1 week in
amide), introduced in 1994 [41], which shows an extremely a drying oven at 40 C. Viscosimetric analysis of 3 gave a
high Pb2+ selectivity [42]. weight-averaged molecular mass, Mw , of 28 850 g mol1
and Mw /Mn = 6.31, Mn being the number-averaged
molecular mass.
2. Experimental

2.1. Reagents 2.3. ISE membranes and electrodes

All solutions were prepared with freshly deionized water In general, the membrane components (typically 200 mg
(18 M cm) and acids and salts of puriss. p.a. or MicroSelect of total mass) were dissolved in freshly distilled THF
quality (Fluka AG, CH-9471 Buchs, Switzerland). The plas- (2.0 ml) during ca. 2 h and poured into a glass ring (24
ticizers, bis(2-ethylhexyl) sebacate (DOS) and o-nitrophenyl or 26 mm i.d.) fixed on a Teflon support. After overnight
octyl ether (o-NPOE), poly(vinyl chloride) (PVC), the evaporation of the solvent at RT, 5 mm disks were punched
lipophilic salts, sodium tetrakis[3,5-bis(trifluoromethyl)phe- from the mother membrane (thickness, ca. 200 m) and
nyl]borate (NaTFPB), potassium tetrakis[3,5-bis(trifluoro- glued with THF to a plasticized PVC tubing, mechanically
methyl)phenyl]borate (KTFPB), and tetradodecylammo- fixed to a 1000 l pipette tip. On the basis of the mem-
nium tetrakis(4-chlorophenyl)borate (ETH 500), as well branes containing DOS (ca. 75 mg), PVC (37 mg), KTFPB
as the ionophore, 4-tert-butylcalix[4]arene-tetrakis(N,N -di- (5 mmol kg1 ), and ionophore 3 (2.3422.21 mg), the ef-
methylthioacetamide), were Selectophore products (Fluka). fective concentration of 3 was calculated. The compositions
The ion buffers, disodium ethylenediamine tetraacetic of all membranes are shown in Table 1. Inner filling so-
acid (Na2 EDTA) and nitrilotriacetic acid (NTA), were lutions were: for conventional ISEs, 101 M Pb(NO3 )2 ;
also from Fluka. Chemicals and solvents for the synthe- for ISE membranes showing strong ion uptake, 101 M
ses were obtained from Aldrich, Fluka, Merck, Riedelde Na2 EDTA, 101 M NaOH, pH 7.5, 105 M Pb(NO3 )2 ; for
Han, or Sigma in puriss. p.a. quality and were used ISEs with optimized lower detection limits [19], 103 M
without further purification if not otherwise mentioned. NTA, 2.1 101 M NaOH, pH 7.0, 104 M Pb(NO3 )2 . A
Poly(tetrahydrofuran) (PTHF) and 2,2,4-trimethylhexa- second inner filling solution, in contact with the first one
methylene diisocyanate (TMDI) were from BASF AG through a diaphragma, consisted of 101 M NaCl (conven-
(D-67056 Ludwigshafen, Germany) and Hls Chemie tional ISEs and ISE membranes with strong ion uptake) or
Schweiz AG (CH-8700 Ksnacht, Switzerland), respec- 2103 M NaCl (with optimized lower detection limit). The
tively. The diol derivative 2 was obtained in our laboratory ISEs were conditioned in 5 109 M Pb(NO3 )2 containing
by a synthesis of seven steps, which will be described 103 M CaCl2 and 104 M HNO3 (Fig. 4) or 5 1010 M
elsewhere [43]. Pb(NO3 )2 with 104 M NaCl and 104 M HNO3 (Fig. 5)
for 48 h.
2.2. Synthesis of ionophore 3
2.4. EMF measurements
A solution of dry 2 (0.25 g, 0.19 mmol, 1 equiv.) in dry
1,2-dichloroethane (80 ml) was stirred at RT under Ar for Measurements were performed with a 16-channel elec-
10 min. The oligomeric diol PTHF (1.50 g, 0.75 mmol, trode monitor (Lawson Labs Inc., Malvern, PA 19355)
4 equiv.) was added and the yellowish solution was heated in magnetically stirred solutions at ambient tempera-
to 70 C until PTHF was fully dissolved. The clear solu- ture (2022 C). Activity coefficients were obtained from
tion was then heated to 110 C and freed from water by the DebyeHckel approximation and EMF values were
M. Pntener et al. / Analytica Chimica Acta 503 (2004) 187194 189

Table 1
Composition of ISE membranes
Membrane Ionophore, mg (mmol kg1 , wt.%) Anionic sites, mg (mmol kg1 ) Plasticizer, mg Polymer, mg (wt.%) 102 M Pb
no. (NO3 )2 , l
1 3 NaTFPB KTFPB o-NPOE DOS PVC PU

Iaa 2.11 (9.85) 0.36 (1.96) 134.8 66.4 (32.5)


Iba 0.37 (2.05) 133.6 66.4 (33.1)
IIaa 21.53 (7.87, 9.61) 0.63 (3.12) 135.7 66.2 (29.6)
IIba 0.57 (3.10) 135.8 66.4 (32.8)
IIIb 2.13 (9.36) 0.94 (4.86) 144.2 68.2 (31.7)
IVb 1.95 (8.03) 0.97 (4.68) 139.2 67.0 (29.1) 21.4 (9.27)
Vb 22.80 (8.18, 9.98) 0.99 (4.81) 137.5 67.2 (29.4)
VIb,c 1.96 (8.02) 0.98 (4.69) 140.0 67.4 (29.2) 20.9 (9.05)
VIIb,c 22.21 (8.17, 9.96) 1.04 (5.17) 133.1 66.6 (29.9)
VIIId 5.01 (10.38) 2.01 (4.95) 299.1 151.2 (33.1) 113.0
IXd 29.67 (10.42, 12.71) 1.03 (5.00) 134.6 66.1 (28.6) 57.5
a For double membranes.
b For selectivity measurements.
c For Fig. 4.
d For Fig. 5.

corrected for liquid-junction potentials with the Hender- 3. Results and discussion
son equation. Sample pH values were determined with a
Metrohm glass electrode (No. 6.0133.100, Metrohm AG, In the presence of zero-current transmembrane ion fluxes,
CH-9010 Herisau, Switzerland). The reference electrode the primary ion activity near the membrane surface, aI ,
was a Metrohm double junction Ag/AgCl reference elec- differs from that in the bulk of the sample, aI (bulk). At
trode (No. 6.0729.100) with 3 M KCl as reference elec- steady state, this difference given by the second term on the
trolyte and 1 M KCl as bridge electrolyte. right-hand side of Eq. (1) depends on the difference between
the concentration of the primary ion complex on the inner
2.5. Selectivity measurements ([IL] ) and outer ([IL]) side of the membrane, the diffusion
coefficients, D, of ionic species in the membrane (m) and
Prior to selectivity measurements, the ISEs were condi- the aqueous sample (aq), and the thicknesses, d, of the dif-
tioned in 102 M NaCl for 2 days. The sequence of the fusion layers of both phases:
investigated ions was: Na+ , H+ , K+ , Mg2+ , Ca2+ , Cd2+ ,
Dm daq
Pb2+ , Cu2+ , and Ag+ . For each of these ions, the mem- aI = aI (bulk) + ([IL] [IL]) (1)
brane was conditioned in the corresponding 102 M ni- Daq dm
trate solution (during 0.5 h), after which calibration curves
In this approximate equation, derived for the steady state
were taken using a Liquino 711 (Metrohm) for automatic
[8,14,18], activities are set equal to concentrations. At
dilution. Even with this procedure, sub-Nernstian slopes
submicromolar sample activities of Iz i, already very small
were obtained for some highly discriminated cations. In
changes (below 1%) in the concentration of primary ion
these cases, the selectivity values represent upper limits
complexes, ILz i, on either side of the membrane induce
only.
relevant deviations between aI and aI (bulk). If the concen-
tration of ILz i in the membrane decreases toward the sample
2.6. Sandwich membrane measurements phase, the lower detection limit, aI (DL), is determined by
the primary ions leaking from the membrane [14]:
Membranes were individually conditioned in 102 M
 
Pb(NO3 )2 for 12 h. The two membranes (without and with 1 z I + zJ
ionophore 1 or 3) for the sandwich membranes were se- log aI (DL) = +
2 6
lected if their EMF values agreed within 1 mV. Then,   
one membrane disk ( 6 mm) without and another with Dm daq RT zI /zJ  pot zI /zJ
log KIJ aJ (2)
ionophore (1 or 3) were stacked together and mounted in a Daq dm zI
Philips IS 561 electrode body (Mller Glasblserei, Zrich,
Switzerland), with the ionophore-loaded membrane on the Here, zI and zJ are the charges of primary (I) and interfering
pot
outer side. It took <1 min between assembling the sand- (J) ions, KIJ is the potentiometric selectivity coefficient,
wich membrane and starting the potential measurements. and RT is the concentration of ion exchanger sites in the
Again, 102 M Pb(NO3 )2 served as internal electrolyte and membrane. It is assumed that [IL] = RT /zI . On the other
sample solution. hand, if [IL] = 0 so that [IL] strongly decreases toward the
190 M. Pntener et al. / Analytica Chimica Acta 503 (2004) 187194

inner solution, a sudden step-like potential drop is observed tension of that study, we have modified the Pb2+ -selective
at aI (step) [18]: ionophore, 4-tert-butylcalix[4]-tetrakis(N,N -dimethylthio-
Dm daq RT acetamide) [41,42] by introducing on its upper rim a diol
aI (step) = (3) derivative (2) and subsequently copolymerizing it with
Daq dm zI
poly(tetrahydrofuran)diol and 2,2,4-trimethylhexamethylene
To construct ISEs for submicromolar activity measurements, diisocyanate to the corresponding polyurethane (see Fig. 1).
the composition of the internal solution has to be designed Although synthetically more demanding, the modification
so as to minimize the transmembrane ion gradients [44]. of the upper rim was chosen in order not to interfere with
Since, however, any change in the sample composition in- the coordination sphere. A similar strategy was applied re-
evitably induces some concentration gradients, as a second cently for modifying a related Na+ -selective calix[4]arene
measure, the value of the right-hand side of Eq. (3) (the same [39]. The synthesis of the diol precursor 2 required seven
term also occurs in Eq. (2)) must be minimized. The benefi- steps and will be described elsewhere.
cial influence of various measures including the increase in The copolymerization reaction yielded a mixture of
the polymer content or the thickness of the membrane [20], polymers with a weight mean molecular mass of Mw =
the decrease in RT [19,20], strong stirring, application of a 28.850 g mol1 (see Section 2). As shown by the ratio
flow-through cell or rotating electrodes [4547] as well as Mw /Mn = 6.31 (with Mn as the numerically weighted
the introduction of lipophilic particles into the membrane molecular mass), the polymer mixture has a rather high dis-
matrix [22] have been documented. As a further possibil- persity. The effective molar concentration of the complexing
ity, the present work investigates the influence of covalently calixarene units has been determined potentiometrically by
pot
binding the ionophore to a polymer backbone on the re- evaluating the selectivity coefficient, KPbNa , as a function of
sponse behavior of the corresponding ISE membranes. Co- the concentration of ionophore 3 relative to that of the cation
valent attachment of an ionophore is also a requirement for exchanger sites (TFPB ), kept constant at 5 mmol kg1
developing rugged miniaturized ISEs [33,34]. (Fig. 2). As long as the molar ratio of binding sites to ion
Recently, we have shown that covalent attachment of a exchanger, [L]/[TFPB ] is <0.5, about the same potentio-
H+ -selective ionophore to a polyurethane backbone can metric selectivity is expected as for a ligand-free membrane.
eliminate the ion-flux-induced sudden potential drop caused Since the concentration of the uncomplexed Pb2+ ions in
by a strong transmembrane pH gradient [40]. As an ex- the membrane sharply decreases at [L]/[TFPB ] = 0.5, a

O
O O + HO O
O OH OH
9 27
OH PTHF
CH2 CH2
O O O +

N S N S N S
4 3 O C N
N C O

TMDI
2
1
1,2-dichloroethane,
o
70 C, 7 d, 37%

O
O O polymer
O O
O
CH2 polymer
O O

N S N S
3

3
Fig. 1. Structure of the mobile ionophore 1 and scheme of the synthesis of the covalently immobilized ionophore 3. The number of poly(tetrahydrofuran)
(PTHF) linear-chain polymer units is an average.
M. Pntener et al. / Analytica Chimica Acta 503 (2004) 187194 191

ionophore per 100 mg of total membrane is only marginal,


4 further experiments were made with an excess of ionophore
3 (10 mg/100 mg of total membrane) in order to avoid selec-
2 tivity changes due to small variations of the ionophore con-
centration.
log KPpbotNa

1
Next, we determined the effective transmembrane ion
0
mobilities of Pb2+ in o-NPOE membranes by using the
-1 sandwich membrane method [4850]. For this experiment,
-2 two membranes, one with the ionophore (1 or 3) and the
-3 other without, were prepared and conditioned separately.
The EMF of the cell was measured using the same solu-
0 2 4 6 8 10 tion (102 M Pb(NO3 )2 ), also used for conditioning, on both
Ionophore 3 [mg/100 mg membrane] sides of the membrane. Since in this symmetrical setup, the
pot
Fig. 2. Potentiometric selectivity coefficient, log KPbNa , as a function of transmembrane potential ideally is zero, the measured value
the concentration of ionophore 3 (cf. Fig. 1 and Table 1) relative to a of ca. 300 mV (cf. Fig. 3) for the individual membranes is
constant concentration of TFPB (5 mmol kg1 ). mainly due to different internal filling solutions. Then, the
two kinds of membranes were joined together and mounted
strong selectivity gain is expected at this concentration. In in an electrode body, with the ionophore-loaded membrane
the range of 0.5 < [L]/[TFPB ] < 1, the ligand concentra- on the sample side. The initial potential reflecting the com-
tion is not sufficient to form a 1:1 complex with the monova- plex formation constant [48,50] (see below) was stable for
lent interfering ion. Therefore, the Pb2+ selectivity is better 10 and 30 min with 1 and 3, respectively, and then started to
in this range and decreases somewhat because of the forma- drift toward the value obtained with the single membranes
tion of Na+ complexes at higher ionophore concentrations. (Fig. 3). The apparent large difference between the time re-
The experimental trend (Fig. 2) nicely fits these expecta- sponses found with ionophores 1 and 3 reflects the differ-
tions except that the change in selectivity at [L]/[TFPB ] = ence in their mean mobility. With the mobile ionophore 1,
0.5 is less sharp than presumed for a strongly complexing the equilibrium value is approached after 4 h, while with
ionophore (see below). This might be due to the fact that the polymeric ionophore 3, it is still not reached even af-
the ionophore is a mixture of polymers, with possibly some- ter 11 h. Since the latter corresponds to a mixture of rather
what varying complexation behavior. Nevertheless, from this high dispersity, no attempt has been made to fit a diffusion
study, the effective ligand concentration can be determined coefficient to the observed curves.
as 8.21 105 mmol mg1 , which corresponds to the in- The initial EMF difference across double membranes
corporation of 84% of the initially used diol into the poly- is directly related to the effective ion complex formation
mer backbone. Since the selectivity gain around 6 mg of constants of the ionophore in the membrane [48,50]. The

800

700

3
EMF / mV

600

500
1

400

300
0 1 2 3 4 5 6 7 8 9 10 11 12
Time / h

Fig. 3. Time response of ISEs with sandwich membranes (see Section 2). At t = 0, an ionophore-free ion exchanger membrane (inner side) was joined
to an analogous membrane additionally containing ionophore 1 or 3 (cf. Fig. 1 and Table 1).
192 M. Pntener et al. / Analytica Chimica Acta 503 (2004) 187194

Table 2
pot
Potentiometric selectivity coefficients, log KPbJ , obtained with the separate solution method for ionophores 1 and 3 (cf. Fig. 1) on polymer membranes
plasticized with o-NPOE or DOSa
Ion J 1, o-NPOE 1, o-NPOE with PU 1, DOS 1, DOS with PU 3, o-NPOE 3, DOS (membrane VII)
(membrane III) (membrane IV) [19] (membrane VI) (membrane V)
H+ 5.3 0.3 (60.2) 2.9 0.3 (53.4) 7.5 0.2 2.3 0.1 (56.1) 1.3 0.2 (49.3) 1.7 0.6 (50.7)
Na+ 7.2 0.2 (56.6) 4.3 0.6 (53.6) 7.5 0.2 3.6 0.1 (56.4) 3.5 0.2 (41.0) <4.4 0.5 (42.6)
K+ 7.8 0.1 (54.8) 5.3 0.2 (49.7) 6.9 0.2 4.4 0.1 (55.1) 4.1 0.2 (38.0) <4.3 0.4 (44.8)
Ag+ >13.9 2.9 (25.4)b >12.9 0.7 (28.3)b 14.7 0.2 (55.1)b >5.9 2.1 (11.7) >6.5 0.9 (12.9)
Mg2+ <14.3 0.2 (11.4) 9.4 0.2 (22.7) 13.9 0.2 8.4 0.1 (25.8) <6.6 0.3 (17.2) <6.6 0.6 (19.5)
Ca2+ 12.9 0.1 (23.4) 8.0 0.1 (26.2) 13.1 0.2 7.5 0.1 (26.9) 6.1 0.2 (23.7) 6.4 0.6 (25.0)
Cu2+ 6.1 0.1 (30.7) 5.2 0.04 (30.8) 4.0 0.2 5.0 0.1 (30.1) <2.3 0.2 (14.7) <2.3 0.6 (15.6)
Cd2+ 4.1 0.1 (30.9) 3.9 0.1 (27.3) 6.3 0.1 3.6 0.1 (23.8) 4.5 0.2 (24.3) 5.5 0.6 (24.3)
a Standard deviations obtained with three ISEs. In parentheses: response slopes (mV) for the measuring range of 102 to 104 M.
b Slopes for 105.2 to 106 M Ag+ .

values obtained with Pb2+ according to the formalism used (A) consisted of 101 M Pb(NO3 )2 , which induces a rather
in Ref. [48] are log = 18.4 0.5 and 17.3 0.3 for 1 strong leaking of Pb2+ to the sample due to its coextraction
and 3, respectively, in a PVC/o-NPOE membrane. Earlier, on the inner side of the membrane. Indeed, with this internal
this complex formation constant of 1 in a PVC/DOS and a solution, the lower detection limit of the ISE membrane with
PVC/o-NPOE matrix was determined as log = 15.9 0.2 1 was only ca. 106.5 M Pb2+ . In contrast, with the poly-
[42] and 21.3 0.3 [27], respectively. Although the attach- meric ionophore 3, the lower detection limit was improved
ment of the coordinating calix[4]arene unit to the polymer by about one logarithmic unit (Fig. 4, two top curves). The
backbone leads to some loss of complex stability, this re- second internal solution (B) consisted of 105 M Pb(NO3 )2
duction is not substantial and the polymeric ionophore 3 still and 101 M Na2 EDTA, its pH being adjusted to 7.5 with
belongs to the compounds forming the strongest 1:1 com- NaOH. The ISE with the mobile ionophore 1 showed the ex-
plexes in membranes found so far [49]. pected apparently super-Nernstian response between 106
pot
The potentiometric selectivity coefficients, log KPbJ , were and 107 Pb2+ (Fig. 4, bottom curve). Also this bias was
determined for ionophores 1 and 3 using o-NPOE or DOS as eliminated when using the polymeric ionophore 3. It must
plasticizer (Table 2). With the mobile ionophore (1), the se- be noted that neither of these internal solutions of extreme
lectivities obtained with o-NPOE were somewhat better than compositions would be applied to practical issues because
those reported earlier with DOS. Such differences have been
observed repeatedly, and were ascribed to the coordinating
ability of DOS [27,51]. Interestingly, the same trend was
not found with the polymeric ionophore (3). This ionophore
gave a significant loss of selectivity with respect to all inter-
fering ions. As shown by the similar value of the complex
formation constant with Pb2+ (see above), the selectivity
loss is a consequence of stronger interactions with the inter-
EMF / mV

fering ions and not due to weaker interactions with Pb2+ .


Obviously, the other functional groups of the polyurethane
1
are rather good coordinating sites for the interfering alkali A
and alkaline earth metal cations. As a further evidence, the 3
selectivities were also determined with membranes based
on the mobile ionophore 1, which contained about the same
amount of the polyurethane (PU) backbone as membranes
based on 3 (Table 2, columns 3 and 5). The strong loss of 50 mV
selectivity due to the presence of this PU further confirms 3
the above interpretation. It must be emphasized that the se- B
lectivities of the ISE membranes with 3 (Table 2, columns 1
6 and 7) are still very good and sufficient for obtaining a
-9 -8 -7 -6 -5 -4 -3
Pb2+ response down to submicromolar concentrations.
The influence of the reduced mobility of 3 on the response log aPb
to dilute Pb2+ samples was tested with two internal solu-
Fig. 4. Potentiometric response curves of ISEs based on ionophore 1
tions of extreme compositions. In order to ensure the same or 3 (cf. Fig. 1 and Table 1) and having as internal electrolyte either
viscosity, the membrane with 1 contained the same amount A (101 M Pb(NO3 )2 , conventional) or B (105 M Pb(NO3 )2 , 101 M
of PU as that with 3 (see Table 1). The first internal solution Na2 EDTA, 101 M NaOH, pH 7.5).
M. Pntener et al. / Analytica Chimica Acta 503 (2004) 187194 193

ing the selectivity behavior, it must be emphasized that the


selectivity coefficients of <106 for Ca2+ and Mg2+ are
1
still much better than those achieved prior to recent devel-
100 mV
3 opments of the past 5 years. Also, with the lower detection
limit of 109 M Pb2+ in the presence of 104 M NaCl at
pH 4.0, the new sensor currently belongs to the best ones.
EMF / mV

1 In spite of this, there is room for improvements by using


bg I other polymers instead of polyurethanes.
3

4. Conclusions

bg II It is shown that the covalent binding of an ionophore to


a polymer backbone significantly reduces the adverse influ-
ence of zero-current ion fluxes in an ISE membrane. With the
new immobilized Pb2+ -selective ionophore, the lower detec-
-12 -11 -10 -9 -8 -7 -6 -5 -4 -3
tion limit for ISEs having a conventional inner electrolyte has
became comparable to that obtained with the corresponding
log aPb
mobile ionophore and an optimized inner solution. Although
Fig. 5. Potentiometric response curves of ISEs based on the ionophore 1 the use of a polyurethane backbone strongly reduced the se-
and with an internal solution for optimized lower detection limit (105 M lectivities, the general strategy of reducing transmembrane
Pb(NO3 )2 , 103 M NTA, 2.1 101 M NaOH, pH 7.0) in comparison ion fluxes by covalent attachment of ionophores to polymers
with those of ISEs based on ionophore 3 and having a conventional inner looks very promising on the basis of the present results.
electrolyte (101 M Pb(NO3 )2 ). Sample background: bg I, 103 M CaCl2 ,
104 M HNO3 ; bg II, 104 M NaCl, 104 M HNO3 .

Acknowledgements
much smaller ionic gradients are sufficient to avoid biases
This work was financially supported by the Swiss National
caused by transmembrane ion fluxes.
Science Foundation, an internal research grant of the ETH
As a further test, the response of an ISE based on
Zrich, Orion Research Inc. (Beverly, MA), and the National
ionophore 1 and an optimized inner solution [19] was
Institutes of Health (grant R01-GM59716). We thank Dr. D.
compared with that of an ISE with ionophore 3 and a
Wegmann for careful reading of the manuscript.
conventional internal electrolyte solution (Fig. 5). With a
103 M CaCl2 solution as sample background, this latter
internal solution with the ISE membrane based on the mo- References
bile ionophore 1 gave a lower detection limit of 106.2 M
Pb2+ (Fig. 4). On the other hand, when it was combined [1] W.E. Morf, The Principles of Ion-Selective Electrodes and of Mem-
with the immobilized ionophore 3 using the detection limit brane Transport, Elsevier, New York, 1981.
was significantly improved but still less good by about one [2] E. Bakker, P. Bhlmann, E. Pretsch, Chem. Rev. 97 (1997) 3083
order of magnitude than that obtained with ionophore 1 3132.
and the optimized inner solution (Fig. 5, top two curves). [3] E. Bakker, E. Pretsch, Anal. Chem. 74 (2002) 420A426A.
[4] P. Bhlmann, E. Pretsch, E. Bakker, Chem. Rev. 98 (1998) 1593
As indicated by the selectivity coefficients (Table 1), this 1687.
is due to the combined interference of H+ (KPbH (aH )2
pot
[5] E. Bakker, P. Bhlmann, E. Pretsch, Electroanalysis 11 (1999) 915
= 5 1010 ) and Ca2+ (kPbCa aCa = 5.9 1010 ). If, how-
pot 933.
[6] S. Mathison, E. Bakker, Anal. Chem. 70 (1998) 303309.
ever, a 104 M NaCl solution is used as sample background [7] R.E. Gyurcsnyi, E. Pergel, R. Nagy, I. Kapui, B.T.T. Lan, K. Tth,
(KPbNa (aNa )2 = 2.6 1012 ), the response of the ISE with
pot
I. Bitter, E. Lindner, Anal. Chem. 73 (2001) 21042111.
3 and the conventional internal electrolyte is as good as that [8] T. Sokalski, T. Zwickl, E. Bakker, E. Pretsch, Anal. Chem. 71 (1999)
of 1 with the optimized inner solution (Fig. 5, two bottom 12041209.
[9] T. Sokalski, A. Ceresa, M. Fibbioli, T. Zwickl, E. Bakker, E. Pretsch,
curves). This result indicates that covalent attachment of Anal. Chem. 71 (1999) 12101214.
ionophores to a polymer has a high potential in constructing [10] T. Zwickl, T. Sokalski, E. Pretsch, Electroanalysis 11 (1999) 673
rugged electrodes since they can efficiently help to reduce 680.
transmembrane ion fluxes. From the various methods for [11] T. Sokalski, A. Ceresa, T. Zwickl, E. Pretsch, J. Am. Chem. Soc.
improving the lower detection limits [3,11,14,20], cova- 119 (1997) 1134711348.
[12] W. Qin, T. Zwickl, E. Pretsch, Anal. Chem. 72 (2000) 32363240.
lent binding to a polymer is especially relevant in view of [13] A.C. Ion, E. Bakker, E. Pretsch, Anal. Chim. Acta 440 (2001) 7179.
constructing miniaturized sensors having long lifetimes. [14] A. Ceresa, A. Radu, S. Peper, E. Bakker, E. Pretsch, Anal. Chem.
Although the polyurethane applied is not optimal regard- 74 (2002) 40274036.
194 M. Pntener et al. / Analytica Chimica Acta 503 (2004) 187194

[15] E. Lindner, R.E. Gyurcsnyi, R.P. Buck, Electroanalysis 11 (1999) [33] R.P. Kusy, J.Q. Whitley, R.P. Buck, V.V. Cosofret, E. Lindner, Poly-
695702. mer 35 (1994) 21412147.
[16] E. Pergel, R.E. Gyurcsnyi, K. Tth, E. Lindner, Anal. Chem. 73 [34] J.A.J. Brunink, R.J.W. Lugtenberg, Z. Brzzka, J.F.J. Engbersen,
(2001) 42494253. D.N. Reinhoudt, J. Electroanal. Chem. 378 (1994) 185200.
[17] W.E. Morf, M. Badertscher, T. Zwickl, N.F. de Rooij, E. Pretsch, J. [35] D.N. Reinhoudt, J.F.J. Engbersen, Z. Brzzka, H.H. van den Vlekkert,
Electroanal. Chem. 526 (2002) 1928. G.W.N. Honig, H.A.J. Holterman, U.H. Verkerk, Anal. Chem. 66
[18] W.E. Morf, M. Badertscher, T. Zwickl, N.F. de Rooij, E. Pretsch, J. (1994) 36183623.
Phys. Chem. B 103 (1999) 1134611356. [36] R.J.W. Lugtenberg, R.J.M. Egberink, A. van den Berg, J.F.J.
[19] A. Ceresa, E. Bakker, D. Gnther, B. Hattendorf, E. Pretsch, Anal. Engbersen, D.N. Reinhoudt, J. Electroanal. Chem. 452 (1998)
Chem. 73 (2001) 343351. 6986.
[20] A. Ceresa, T. Sokalski, E. Pretsch, J. Electroanal. Chem. 501 (2001) [37] L.Y. Heng, E.A.H. Hall, Anal. Chem. 72 (2000) 4251.
7076. [38] L.Y. Heng, E.A.H. Hall, Electroanalysis 12 (2000) 178186.
[21] T. Sokalski, I. Bedlechowicz, M. Maj-Zurawska, A. Hulanicki, Fre- [39] E. Malinowska, L. Gawart, P. Parzuchowski, G. Rokicki, Z. Brzzka,
senius J. Anal. Chem. 370 (2001) 367370. Anal. Chim. Acta 421 (2000) 93101.
[22] T. Vigassy, R.E. Gyurcsnyi, E. Pretsch, Electroanalysis 15 (2003) [40] M. Pntener, M. Fibbioli, E. Bakker, E. Pretsch, Electroanalysis 14
375382. (2002) 13291338.
[23] L. Ebdon, A.T. Ellis, G.C. Corfield, Analyst 104 (1979) 730738. [41] E. Malinowska, Z. Brzzka, K. Kasiura, R.J.M. Egberink, D.N.
[24] L. Ebdon, A.T. Ellis, G.C. Corfield, Analyst 107 (1982) 288294. Reinhoudt, Anal. Chim. Acta 298 (1994) 253258.
[25] S.C. Ma, N.A. Chaniotakis, M.E. Meyerhoff, Anal. Chem. 60 (1988) [42] A. Ceresa, E. Pretsch, Anal. Chim. Acta 395 (1999) 4152.
22932299. [43] M. Pntener, E. Pretsch, Anal. Chim. Acta, in preparation.
[26] S. Daunert, L.G. Bachas, Anal. Chem. 62 (1990) 14281431. [44] E. Bakker, E. Pretsch, Trends Anal. Chem. 20 (2001) 1119.
[27] M. Telting-Diaz, E. Bakker, Anal. Chem. 73 (2001) 55825589. [45] Q. Ye, M.E. Meyerhoff, Anal. Chem. 73 (2001) 332336.
[28] K. Kimura, T. Matsuba, Y. Tsujimura, M. Yokoyama, Anal. Chem. [46] A. Zirino, R. De Marco, I. Rivera, B. Pejcic, Electroanalysis 14
64 (1992) 25082511. (2002) 493498.
[29] Y. Tsujimura, M. Yokoyama, K. Kimura, Anal. Chem. 67 (1995) [47] T. Vigassy, R.E. Gyurcsnyi, E. Pretsch, Electroanalysis 15 (2003)
24012404. 12701275.
[30] Y. Tsujimura, M. Yokoyama, K. Kimura, Electroanalysis 5 (1993) [48] Y. Mi, E. Bakker, Anal. Chem. 71 (1999) 52795287.
803807. [49] Y. Qin, E. Bakker, Anal. Chim. Acta 421 (2000) 207220.
[31] Y. Tsujimura, T. Sunagawa, M. Yokoyama, K. Kimura, Analyst 121 [50] M.M. Shultz, O.K. Stefanova, S.B. Mokrov, K.N. Mikhelson, Anal.
(1996) 17051709. Chem. 74 (2002) 510517.
[32] V.V. Cosofret, E. Lindner, R.P. Buck, R.P. Kusy, J.Q. Whitley, Elec- [51] E. Bakker, A. Xu, E. Pretsch, Anal. Chim. Acta 295 (1994) 253
troanalysis 5 (1993) 725730. 262.

You might also like