You are on page 1of 12

DIABETES TECHNOLOGY & THERAPEUTICS

Volume 11, Supplement 1, 2009


Mary Ann Liebert, Inc.
DOI: 10.1089=dia.2009.0004

Clinical Experience with Continuous Glucose


Monitoring in Adults

Kristin Fabiato, M.D.,1 John Buse, M.D., Ph.D.,1 Michelle Duclos, M.P.H.,1
Joseph Largay, P.A.C., C.D.E.,1 Camille Izlar, M.S., R.D., L.D.N., C.D.E.,1 Thomas OConnell, M.D.,1
Jennifer Stallings, R.D., L.D.N., C.D.E.,1 and Kathleen Dungan, M.D.2

Abstract
Despite recent advances in therapy, achieving adequate glycemic control may be difficult for a large number of
patients with diabetes. Real-time (RT)-continuous glucose monitoring (CGM) has the potential to improve
glycemic control through immediate feedback to the properly trained patient. However, limitations exist both in
interpreting the results of published randomized clinical trials on CGM use and in extrapolating the results to the
diabetes population at large. This review summarizes the evidence for use, identifies suitable candidates, de-
scribes optimal implementation, and employs case scenarios in order to emphasize practical aspects of RT-CGM
use in adults. Establishment of expectations and comprehensive education in intensive insulin therapy and RT-
CGM use are necessary for successful implementation. Because the technology has been shown to be most useful
in patients who are actively viewing and responding to RT data, patients should receive explicit instructions for
active self-adjustment of insulin and lifestyle elements. While the technology is improving, false alarms remain a
significant barrier to optimal use. The utility of RT-CGM for patients with severe hypoglycemia or hypoglycemia
unawareness has not been established. Finally, studies are needed to determine the sustainability of improve-
ments in glycemic control, as well as cost-effectiveness and practicality of implementation into busy real-world
practice.

Introduction trials indicate that only frequent intensive use leads to im-
provement in glycemic control or reduction in hypoglyce-
mia in insulin-treated patients.1013 RT-CGM, however, has
D espite recent advances in diabetes therapy, many
patients still fail to achieve the minimal glycosylated
hemoglobin (A1c) target of less than 7.0% set forth by the
greater potential to improve glycemic control through im-
mediate feedback to the properly trained patient and there-
American Diabetes Association.1 Observational studies have fore serves as the focus of the current review. In this review,
shown that glycemic control is linked to frequency of self- we will summarize the evidence for use, identify suitable
monitored blood glucose (SMBG) testing in insulin-treated candidates, describe optimal implementation, and employ
patients.2 The link between frequency of SMBG and glycemic case scenarios in order to emphasize practical aspects of
control in patients with type 2 diabetes not treated with in- RT-CGM use. We will focus on adults with type 1 diabetes
sulin is not so clear.3,4 A recent panel concluded that SMBG mellitus (T1DM), based on the available evidence, unless
should be performed once or twice per day in most non otherwise specified.
insulin-requiring patients.5 It has not been disputed, how-
ever, that even SMBG performed four to six times per day can Evidence
miss important hyperglycemic and hypoglycemic events.6,7
Interpretation of studies
This has, therefore, spurred a growing interest in continuous
glucose monitoring (CGM). Retrospective CGM systems In general there are several limitations in interpreting the
(CGMSs) were approved for use by the Food and Drug Ad- current CGM literature and extrapolating to the general adult
ministration (FDA) in June 1999.8 The first real-time (RT) population. Initial studies were uncontrolled, making it dif-
CGM, the GlucoWatch Biographer (Cygnus Inc., Redwood ficult to interpret whether benefit originated from a true effect
City, CA), was approved by the FDA in 2001.9 Controlled of CGM or resulted from the Hawthorne effectpatients who

1
UNC Highgate Diabetes and Endocrinology Clinic, University of North Carolina at Chapel Hill, Chapel Hill, North Carolina.
2
The Ohio State University, Columbus, Ohio.

S-93
S-94 FABIATO ET AL.

know they are being observed alter their behavior consciously Advanced insulin dosing features of sensor-augmented in-
or subconsciously enough to affect glycemic control. In ad- sulin pumps may provide a more straightforward means of
dition, blinding is difficult because subjects must use RT data applying RT-CGM information to short-term modifications
to make adjustments in insulin therapy. Many of the early in insulin therapy. In a recent 6-month trial of 146 patients with
trials were also conducted in pediatric populations. Further- T1DM, patients were assigned to sensor-augmented pump
more, the patients enrolled in these studies were in general therapy versus pump therapy alone.18 The sensor group
highly motivated and had access to well-trained study teams achieved an A1c reduction similar to that of the control group.
who were specifically funded to take the time and make the However, the sensor group was able to achieve the reduction
effort to optimize the utility of the technology. The patients in in A1c without an increase in the rates of hypoglycemia,
one study,14 for example, had scheduled follow-up visits at whereas the control group had an increased frequency of
weeks 1, 4, 8, 13, 19, and 26 plus an additional phone call in hypoglycemia. Furthermore, subgroup analysis demonstrated
between each visit to review blood glucose values. How CGM that patients who maintained >60% sensor utilization achieved
will translate into real-world practice in which this degree of a significant reduction in A1c relative to the control group. This
intensive intervention cannot generally be provided is thus once again emphasizes the importance of adherence.
unclear. Extensive exclusion criteria, in many studies includ- Finally, a randomized controlled trial enrolled 65 patients
ing severe hypoglycemia or hypoglycemic unawareness, of- with poorly controlled (A1c 810%) type 2 diabetes mellitus
ten limit applicability to patients who would theoretically (T2DM), treated with oral hypoglycemics, insulin, or a com-
stand to gain significant benefit from sensor use. Finally, bination of the two.19 No changes were made to insulin doses
studies are generally short term (3 months or less). As with or oral hypoglycemic agents in patients assigned to either
any intervention requiring consistent patient action, it is group unless hypoglycemia occurred. Patients in the RT-CGM
possible that RT-CGM could lose effectiveness if patients be- group only wore the device for 3 days at a time each month for
come complacent over time. the 3-month study. Each group received lifestyle counseling
monthly based on SMBG or RT-CGM readings. However, the
group assigned to RT-CGM was also instructed to increase
Effect on A1c
their activity levels and eat less in response to hyperglycemia
Several controlled trials have suggested moderate effects alarms and to confirm hypoglycemic alarms with SMBG be-
on A1c in adults with consistent use. fore treating. The results demonstrated a significantly greater
A recent randomized controlled trial stratified 322 subjects reduction in A1c in patients randomized to RT-CGM com-
with T1DM by age group and assigned them to RT-CGM pared to controls (1.1% vs. 0.4% respectively, P 0.004). The
(device varied according to patient preference) or SMBG for findings were associated with significant reductions in caloric
26 weeks.14 Eighty-three percent of the 52 adult subjects over intake, body mass index, and a significant increase in exercise
the age of 25 years had a baseline A1c between 7% and 8%. time over the 3-month study. The results are contrary to those
Subjects in this subgroup who were randomized to RT-CGM of studies involving patients with T1DM in which continuous
achieved a statistically significant 0.5% greater reduction in intensive use is necessary. It is not entirely unexpected, how-
A1c at 26 weeks relative to the control group. In contrast, ever, because many patients with T2DM still have endogenous
subjects between 15 and 24 years of age did not benefit. The insulin production and consequently would be more likely to
difference in outcomes may be attributable to frequency of respond to diet and lifestyle modification. Nevertheless, over
sensor use: 83% of those over 25 years wore the sensor at least half of the patients were receiving insulin, and the duration of
6 days per week compared to only 30% of subjects 1524 years diabetes was over 10 years in both groups, suggesting that
old. This suggests that the technology is only useful with high- endogenous insulin production was poor. It is unclear whe-
intensity use, probably through RT modification of diet, ther patients who performed SMBG measurements only
medication, and lifestyle patterns. Indeed, patients >25 years would have achieved similar reductions in A1c if they were
of age had lower baseline A1cs than the younger patients, also also instructed to respond in similar ways to their SMBG
suggesting that other factors, such as differences in knowl- readings as those with RT-CGM. Further study is needed to
edge base, awareness and=or concern over prevention of determine whether reduction in A1c is dependent upon in-
complications, or frequency of SMBG may have affected sulin use. Cost-benefit analysis and an assessment of the du-
outcomes. This study is unique in that it provides details of rability of the effect over time are necessary to determine
specific standardized instructions for adjustment of therapy whether RT-CGM may be feasible for the millions of patients
in response to RT-CGM.15 with poorly controlled T2DM in the United States alone.
Another controlled study found that a combination of pe-
diatric and adult patients with poorly controlled T1DM
Hypoglycemia
(baseline A1c >9.5%) had a 0.6% greater reduction in A1c at 3
months relative to controls using RT-CGM.16 Once again, Hypoglycemia is regarded by many patients and providers
continuous use was paramount, as those patients who un- as the primary barrier to reaching treatment goals.20 Fear of
derwent intermittent RT-CGM use (3 day periods every 2 hypoglycemia causes many patients to target higher glucose
weeks) did not achieve benefit. Ninety-five percent of the levels than currently recommended and makes them reluctant
patients in the continuous arm reported making changes to to implement adjustments in therapy (see Scenario 4 below).
their insulin dose, diet, or lifestyle based on the results of the Furthermore, hypoglycemia unawareness and serious hypo-
RT-CGM, suggesting that even patients with poor control glycemia can be associated with death and disability.21
(and possibly less meticulous self-care habits) can be moti- CGMS has been utilized as a way to better characterize
vated to succeed. This is also supported by less rigorous, non- hypoglycemia as well as to limit its occurrence.12 However,
randomized data.17 CGMS may overestimate the frequency of hypoglycemia
CLINICAL EXPERIENCE WITH CGM IN ADULTS S-95

when compared to concomitant SMBG values,22 and it has 3. Patients contemplating pregnancy or who are currently
reported hypoglycemia in normal subjects at a similar rate as pregnant33,34 (discussed by Chitayat et al.35 in this
in patients with T1DM.23 Nevertheless, interstitial glucose supplement)
may reflect physiologic glucose levels in tissues such as the 4. Intermittent RT-CGM in patients with T2DM19
brain better than blood glucose and may thus be the more
More data are needed to recommend for or against the use
relevant measure. For example, interstitial glucose more clo-
of RT-CGM for other settings, such as in patients with gas-
sely follows recovery of cognitive function after a hypogly-
troparesis, athletes, or hospitalized patients. In contrast, pa-
cemic event than blood glucose.24
tients who are unwilling to engage in scrupulous self-care
Although there may be greater accuracy with newer gen-
behaviors would not be expected to benefit. That is not to say
eration RT-CGM devices, it is still unclear whether all hypo-
that RT-CGM will not one day be capable of supplanting
glycemic events detected by sensors are clinically relevant.
SMBG, just as SMBG has replaced urinary glucose assess-
Controlled trials have achieved moderate reductions in A1c in
ments, but further refinements in the technology are needed.
specific populations without a concomitant increase in hy-
poglycemia.16,18 This is a substantial achievement given that
Implementation
intensive insulin regimens targeting improvements in glycemic
control almost always increase the risk of hypoglycemia, a Improvements in glycemic control have been most pro-
phenomenon that was most comprehensively documented in nounced when RT-CGM is implemented in conjunction with
the Diabetes Control and Complication Trail and the United comprehensive education and close follow-up. Patients, and
Kingdom Prospective Diabetes Study. A short-term controlled even providers, may become overwhelmed when confronted
study did demonstrate that patients using RT-CGM spent 21% by the stream of information that RT-CGM imparts. There-
less time with glucose values <55 mg=dL than controls.25 fore, RT-CGM instruction lends itself to a similar model as
However, no study has demonstrated a reduction in severe that used for insulin pumps, using well-trained diabetes ed-
hypoglycemic events or hypoglycemia unawareness, outcomes ucators to guide patients through a gradual, stepwise im-
for which these studies were not designed, given that patients plementation process. This type of education may be more
at high risk were not generally recruited. difficult for those physicians outside of a research or large
academic setting or without access to diabetes educators, but
Glycemic variability should not preclude sensor use altogether.
It has been hypothesized that glycemic variability may
Expectations
carry a risk for complications independent of A1c.26 Glycemic
excursions in patients whose A1c is at goal, for example, have Patients may have misconceptions regarding the capability
been linked to macrovascular and possibly microvascular of RT-CGM.36 In a study of patients with T1DM and T2DM,
complications.27,28 Glycemic variability also predicts the risk most agreed that the device was easy to use and understand,
of severe hypoglycemia.29,30 but patients were least satisfied with regards to the calibration
RT-CGM has also been investigated as a tool for modifying interval and the tendency toward error.37 Patients must un-
glycemic excursions. In one controlled study of RT-CGM, derstand that the current technology is only capable of pro-
insulin-treated patients with T1DM and T2DM experienced a viding adjunctive assistance. In our center, we have had
23% reduction in postprandial hyperglycemia and spent 26% several patients who, despite meticulous education, have
more time in the euglycemic range as compared to patients in made serious errors resulting in extreme glycemic excursions
the control group.25 Improvements in measures of glycemic as a result of not confirming RT-CGM with SMBG results
variability were reported elsewhere.31,32 Reduction in gly- prior to making corrections. As with SMBG, RT-CGM is only
cemic variability using RT-CGM, therefore, may foretell useful if patients actually view the data and implement
benefits on complications, but this will need to be examined changes in insulin therapy, diet, and other lifestyle elements.
more specifically. As such, patients must shift to making decisions based upon a
constant flow of glucose trends, rather than upon sporadic
CGM Use glucose points in time. This may be especially challenging for
seasoned patients, who may have difficulty accepting this
Patient selection
paradigm shift in diabetes self-management (see Scenario 5
As the technology improves, a larger percentage of patients below). It may be helpful for patients to attend introductory
stand to benefit from CGM. RT-CGM education sessions or investigate products on-line
Until then, the cost of RT-CGM, as well as its modest effi- before they commit to the technology. Likewise, it has been
cacy in selected patients managed in highly experienced well- our experience that it can take 36 months or more for patients
funded centers, warrants judicious patient selection. The most to become self-empowered, that is, to understand the infor-
ideal candidates appear to be those who are well educated mation well enough to make changes to their regimens on
about their diabetes and highly motivated and yet still unable their own.
to achieve an A1c target of <7%.
Other candidates for whom limited data are available who Baseline knowledge
may benefit include:
We recommend that patients undergo individual or group
1. Patients with severe hypoglycemia, hypoglycemic un- training covering intensive diabetes self-management prior to
awareness, or fear of hypoglycemia that impedes ad- embarking on training for RT-CGM. Because of the com-
justment of therapy plexity of new decisions that will be required, even patients
2. Patients with significant glycemic lability who have received such education in the past would benefit at
S-96 FABIATO ET AL.

least from condensed refresher instruction. Some risk be- with SMBG before action is taken to correct the finding. As the
coming overwhelmed by the plethora of glucose values and technology improves, requirements for reliance upon SMBG
even ignoring the alarms. Other patients may be tempted to confirm RT-CGM values may lessen.
to repeatedly bolus in response to hyperglycemia, placing The capability of RT-CGM to reflect blood glucose is in part
themselves at risk of hypoglycemia, or, conversely, to over- reliant upon appropriate calibration with SMBG. It is crucial
compensate for hypoglycemia with excessive carbohydrate that patients perform SMBG testing correctly, and only on the
intake. Therefore, patients should understand the timeaction fingertips for calibration purposes. Calibration must be per-
profile of insulin prior to starting RT-CGM. For pump pa- formed during steady state, when patients are fasting, or at
tients, this includes the time of active insulin set on their least pre-meal, when glucose levels are changing the least. In
pump, although lockout features can provide protection from particular, it has been shown that a few calibrations during
insulin stacking. Patients must receive explicit instructions steady state are more important for determining RT-CGM
regarding when to re-bolus and how to reduce the insulin accuracy than many calibrations during times of fluctuating
dose to avoid insulin stacking (see RT use section below). blood sugars (i.e., during hypoglycemia, exercise, or post-
Conversely, patients must understand how to appropriately meal).13,41 Trend arrows on some devices as well as automatic
manage hypoglycemia. In one study, insulin stacking was calibration lockouts can assist the patient in identifying ap-
uncommon in patients using RT-CGM, but overzealous propriate times for calibration. With that said, there are oc-
treatment of hypoglycemia was frequent enough to partially casional patients who despite exhaustive education, close
obscure the gains in glycemic control.38 follow-up, and repeated calls to company hotlines, can never
achieve acceptable RT-CGM performance with one or more
Selection of the device devices. Whether this is related to patient factors that are as
yet unrecognized or technical issues is unclear but disturbing
A comparison of the four devices currently approved by
to patients and practitioners alike.
the FDA is beyond the scope of this article. However, features
such as size, duration of initial calibration, ease of subsequent
Targets
calibrations, sensor and battery life, water resistance, accuracy
and density of data, and display features such as trend arrows Targets appear on line graphs of the receiver and on soft-
and graphs, as well as software programs for data review, are ware readouts for patient reference. Targets should be pro-
all considerations in choosing a sensor. Compatibility with grammed based on a patients particular needs, depending on
insulin pumps and blood glucose monitors also differs. Fi- the patients control and comfort level. Targets should gen-
nally, devices differ by cost. The selection should depend erally be set higher for patients with severe hypoglycemia. An
upon patient preference, but educators and=or providers expert consensus suggests either setting standard pre-meal
comfort level may also play a role. and post-meal targets of 80130 mg=dL and 140180 mg=dL,
As stated above, the FDA-approved sensor life varies from respectively, or setting wider targets based on the patients
as little as 3 days with the Medtronic (Northridge, CA) device baseline SMBG profile.42 Setting a wide target initially may
to as many as 7 days with the DexCom (San Diego, CA) help to improve patient compliance because patients whose
sensor. With the significant expense of sensors and the often readouts continually appear outside of the target range may
poor reimbursement, many patients have been forced to ex- become easily frustrated not only by a sense of failure but by
tend the duration of use of each sensor through recalibration. frequent and disruptive alarms. Targets can then be gradually
An article by Garg et al.39 supports 10-day use of the DexCom tightened as the patient becomes accustomed to the device.
sensor, but the safety and accuracy of such off-label use re- Patients should be warned that early in the process it is not
quire further study. unreasonable to expect as many as 50% of the readings to be
It may also be possible that sensors could suit individual outside the target range.
patients differently. In our experience, if a patient does not
respond to one sensor, a different sensor could be tried. In Alarms
practice, however, this may not generally be feasible once a
Alarm thresholds may or may not be set at the actual target
sensor has already been purchased.
values, depending upon the desired sensitivity and specific-
Although an extensive discussion of reimbursement is be-
ity. One study demonstrated that a hypoglycemia alert was
yond the scope of this article, it is important to note here that
able to distinguish a blood sugar of 70 mg=dL or less with 67%
several major insurance carriers have recently approved CGM
sensitivity and 90% specificity; however, the false alert per-
use. As more cost-benefit data become available, reimburse-
centage was high at 47%.38 For hyperglycemia defined as
ment for CGM use may become more widespread.
>250 mg=dL, the sensitivity was 63% with a specificity of 97%
and a false alert percentage of 19%.
Technical considerations
Alarms may be programmed for hypoglycemia and hy-
Based on comparisons with blood glucose meters, patients perglycemia, but also for projected hypoglycemia or hyper-
may become frustrated and get the impression that RT-CGM glycemia on some devices. Projected alarms activate when the
is simply an inaccurate tool.37 However, as discussed by CGM glucose value and the rate of change predict that the
Cengiz and Tamborlane40 in this supplement, interstitial glucose value will cross the actual hyperglycemia or hypo-
glucose imparts fundamentally different information than glycemia alarm threshold at a preset time, such as 1030 min.
blood glucose. Patients and providers must understand that In general, fewer false readings occur at longer intervals.38
interstitial glucose measurements tend to lag behind blood According to the consensus panel, alarms should be in-
glucose, particularly during rapid swings. Thus, hyperglyce- troduced one at a time so that patients can become familiar-
mia and hypoglycemia trends and alarms must be confirmed ized with the appropriate responses required for each without
CLINICAL EXPERIENCE WITH CGM IN ADULTS S-97

becoming overwhelmed.42 An alarm set at 70 mg=dL may


be appropriate in most patients except those with hypogly-
cemia unawareness, where 80 mg=dL or higher may be Perform SMBG
more desirable. However, the current sensor technology is not
as accurate at 70 mg=dL, so the false-positive percentage will
likely be higher at this setting. The consensus panel recom- Before Meals Between Meals
mends an initial hyperglycemia alarm to be set at 250 mg=dL
and be reset to 180200 mg=dL over time.42 This process ne-
cessitates an individualized approach because frequent Calculate insulin dose Calculate insulin dose
(I:CHO plus correction) (Correction dose only)
alarms in the less motivated patient could spur frustration
and complacency, but may not be sensitive enough in patients
with hypoglycemia unawareness. Alarms may also be dis-
abled or set to vibrate in order to minimize patient attrition,
RT-CGM Adjustment of Insulin Bolus
but this may defeat the purpose of the device.

Change in Prandial Insulin


Interpretation of data and adjustment in therapy Glucose Trend Trend Arrows
Dose
Data may be interpreted based on instantaneous readouts Rising > 40 mg/dl ( ) Increase 20%
Rising 20-40 mg/dl ( ) Increase 10%
(in response to glucose trends or alarms) or retrospective use Rising or falling by <20 mg/dl No arrows No change
(in response to cumulative patterns over several days or Glucose falling 20-40 mg/dl ( ) Decrease 10%
weeks). The expert panel recommended no action in the first Glucose falling by > 40 mg/dl ( ) Decrease 20%

37 days to allow a patient to simply acclimate to the device.42


FIG. 1. Algorithm for RT-CGM adjustment of insulin.
I:CHO, insulin:carbohydrate.
RT use
The RT goal for the patient is to be able to identify short-
SMBG measurement in 15 min and discouraging the
term trends in the data and to make changes in nutrition or
practice of interrupting the basal rate on pumps, which
insulin regimen concurrently in order to prevent or treat
can set a patient up for wide swings in glucose.42
glucose values that are out of the target range. For example,
a blood glucose of 70 mg=dL may be treated presumptively While limited data are available in adults, pediatric patients
in a patient whose interstitial glucose is dropping rapidly, and parents using this algorithm conveyed good under-
whereas the same number in a patient whose glucose is standing and initial acceptance, but by 13 weeks, only 59%
climbing could simply be observed. Subjects must also be reported using it at least 50% of the time. This may reflect the
cautioned that the display on the current sensors is such that development of complacency with the device or algorithm,
despite a flat trend arrow, a patients blood sugar could be but the trends seen toward progressively more insulin ad-
changing by as much as 59 mg=dL=h, depending on the par- justments over time suggests that use of the sensor was
ticular device.42 helpful in helping patients recognize subtle intricacies in
One limitation of the current literature is that many studies glucose patterns and subsequently individualize their treat-
generally have not described validated or even standardized ment plans.
algorithms for implementing changes in therapy, despite the
fact that they may have been used. Two exceptions are the Retrospective use
DirecNET and the Juvenile Diabetes Research Foundation
Randomized trials have clearly shown that patients achieve
studies, which used very similar algorithms (Fig. 1).15,43 In the
benefit largely from intensive, daily use of the device, sug-
Juvenile Diabetes Research Foundation study, patients took
gesting that retrospective analysis may only provide mar-
action within two main scenarios:
ginal, if any, benefit.1013 However, optimal use of RT-CGM
1. Upward or downward adjustment of bolus insulin based may also involve more in-depth study of the graphs acces-
upon the rate of change in sensor glucose. For example, sible on the CGM software or website provided by each
patients are instructed to increase their pre-meal bolus manufacturer. Patients should be instructed not to become
by 10% for a 2040 mg=dL=h increase (indicated as one preoccupied with the day-to-day variability in glucose mea-
up arrow on the Medtronic sensor) and increase pre- surements on these readouts, but to make adjustments in
meal bolus by 20% for an upward trend of >40 mg= basal or bolus insulin based upon patterns seen on at least 2
dL=h (indicated as 2 up arrows on the Medtronic sen- out of 3 days.15,43 Patients and providers may also gain insight
sor). Downward arrows were treated in an opposite into the appropriate timing and distribution of insulin boluses
manner. No change in bolus dose required if the change in response to variable meal composition, as well as the re-
was <20 mg=dL=h. sponse to exercise or medications, such as pramlintide (Fig. 2)
2. Therapy to prevent or treat hypoglycemia or hyperglycemia and prednisone (Fig. 3). Finally, as demonstrated with tradi-
based upon projected or actual alarms. For high alerts, tional CGMS, RT-CGM may be useful for identifying recur-
patients are instructed to wait at least 2 h before using rent trends, such as the dawn phenomenon, and guiding
repeated correction dosing. They consumed 15 g of adjustments in basal rates.44,45
carbohydrates for low alerts and 10 g for projected low The published algorithms for RT-CGM provide detailed
alerts. In addition, a consensus panel advised good instructions for self-adjustment of insulin with retrospective
hypoglycemia management skills, including repeating use.15,43 Although these algorithms are complex, they have
S-98 FABIATO ET AL.

FIG. 2. RT-CGM use to evaluate the effect of pramlintide on glucose excursions.

been shown to be feasible in study populations. However, less hyperglycemia and making adjustments in both his basal
sophisticated patients are likely to require simplification or settings and his insulin to carbohydrate ratio, the patient
individualized guidance. A consensus statement has recom- was able to obtain an A1c of 6.5%. While attenuating
mended that patients review and download their data prior to his glycemic excursions and tightening his overall control,
each appointment to identify problem areas on which to focus the sensor alarms helped prevent the increase in hypoglyce-
prior to each appointment.42 This could reduce the amount of mic episodes possible with tightened control. Unfortunately,
time spent in each encounter. Providers, in turn, may find it the patients insurance company elected not to cover the
useful to customize the devices software analysis by identi- sensor so the patient had to discontinue its use. The patients
fying only a few graphs on which they wish to focus during a A1c rose to 7.7% off of the sensor despite continued frequent
visit. Graphs of the weekly blood glucose trend in addition to SMBG.
breakdown by meals and daily blood glucose in instances of
severe highs or lows may be particularly useful. Scenario 2. This is a 39-year-old patient with T1DM for
18 years. The patients A1c had been averaging 6.5% for
Clinical Scenarios several years with use of an insulin pump and SBMG at least
four to six times daily. When sensors became available, the
A few clinical scenarios will be presented in this section to patient purchased one and paid for it out-of-pocket. Analysis
illustrate several of the points discussed thus far in this review of sensor data has taught him how various foods affect his
article. blood sugars. He now administers his pre-meal bolus earlier
and has had an attenuation of postprandial hyperglycemia.
Ideal candidates The patients A1c has improved slightly to 6.2% without an
Scenario 1. This is a 40-year-old patient with T1DM for increase in hypoglycemia. In fact, the patient has actually re-
approximately 20 years, complicated by proliferative reti- gained hypoglycemia awareness and now feels safer even
nopathy. Despite initiation of an insulin pump and frequent when not wearing the sensor.
SMBG (at least four times daily), the patient was unable to
obtain an A1c <7%. The patient received training from a These two scenarios reiterate several key points, in partic-
skilled diabetes educator and purchased the RT-CGM device ular that consistent sensor use with RT adjustments in diet,
that paired with his insulin pump. The patients pre-RT-CGM lifestyle, and therapy leads to the most significant results.
A1c was 7.9%. By identifying trends towards hypo- and Patients who receive adequate education and guidance and
CLINICAL EXPERIENCE WITH CGM IN ADULTS S-99

FIG. 3. A 42-year-old woman with T1DM and multiple sclerosis was treated with a prednisone taper to prevent a multiple
sclerosis flare. With the use of CGM, the patient and her healthcare providers were able to identify a significant increase in
fasting and postprandial glucose: (a) before prednisone and (b) after prednisone. This led to a 50% increase in her basal rate
and an increase in her insulin to carbohydrate ratio from 1:8 to 1:5, which resulted in more stable blood glucose (BG) levels.

feel comfortable making these adjustments will likely respond to the alarms during the second incident. After the
experience the most benefit. The issue of insurance coverage medical staff was made aware of the hypoglycemic episodes,
remains a significant limiting factor, and cost-benefit analyses the patients glycemic goals were briefly loosened, and the
are needed to help determine which patients qualify. Lastly, patients total daily insulin dose was decreased by 25%. Over
the ability of RT-CGM to help patients regain hypoglycemia time, with more education and slow titration, the patients
awareness is one of the key potential benefits of sensor tech- control has again been tightened without an increase in hy-
nology; however, assessment of the true effect has been lim- poglycemia. This is an example of the utility of sensors for
ited by frequent exclusion of these patients from controlled permitting tightening of glycemic control without increasing
trials and the inaccuracy of current sensor technology in the hypoglycemia. However, the lag time between interstitial
hypoglycemic range. glucose and capillary glucose may limit the ability to prevent
hypoglycemia if rapid changes in capillary glucose occur.
Pitfalls Further, the relative inaccuracy of current sensor technology
Scenario 3. This is a 42-year-old woman with T1DM for in the hypoglycemic range and associated frequent false
28 years. The patient was deemed a good candidate for the alarms may contribute to patient disregard and even discon-
sensor because of an A1c of 7.9% despite SMBG four to six tinuation of use.
times daily (Fig. 4a). The patient received similar training as
the patient in Scenario 1, and with adjustment of her insulin Scenario 4. This is a 31-year-old man with T1DM for
regimen, the patient was able to obtain an A1c of 6.3% (Fig. approximately 8 years who had a severe hypoglycemic epi-
4b). However, the patient had two episodes of severe hypo- sode requiring medical assistance soon after his diagnosis. As
glycemia overnight requiring assistance from a family mem- a result, the patient and his wife have developed a magnified
ber (Fig. 4c). In both instances, the alarms on the insulin pump fear of hypoglycemia and attempt to maintain his blood
sounded for several hours before the patient received assis- sugars consistently in the mid-200 mg=dL range (Fig. 5a). The
tance. During the first incident, the patient slept through the patient was considered to be a good candidate for use of the
alarm. This problem was reported previously with the use of sensor as the alarms could be set to detect hypoglycemia,
sensor-augmented pumps. Because the sensor alarm is part of allowing the patient to tighten his overall control. However,
the pump, which remains attached to the body during sleep, despite approximately 25 patient-hours with medical staff
thick bedding may muffle the alarms.46 The patient did not including significant education and multiple adjustments to
S-100 FABIATO ET AL.

FIG. 4. A patient with severe hypoglycemia, a 42-year-old woman with diabetes for 28 years who was able to obtain
improvement in A1c with the sensor. However, the patient experienced an increase in hypoglycemia with the decrease in
A1c: (a) CGM data before initiation of sensor, (b) CGM data after glucose control was briefly loosened and then re-tightened,
and (c) CGM daily data during one of two episodes of severe hypoglycemia requiring assistance (alarm bells) that the patient
did not hear. BG, blood glucose.

his insulin regimen in addition to behavioral counseling, the several visits, providers have recommended raising the hy-
patients A1c remains above 9% despite use of the sensor for poglycemia target and shortening the alarm snooze time in
more than 1 year (Fig. 5b). He maintains that the sensor is order to increase the sensitivity of hypoglycemia warnings.
inaccurate because he reports feeling low frequently despite Recommendations were also made to lower the hypergly-
sensor readings in the 200 mg=dL range. Unfortunately, cemia target and shorten the alarm snooze time from 2 h to
the patients fear of hypoglycemia has been so intense that 1 h to limit hyperglycemia and, therefore, decrease the risk of
he overrides the bolus system and occasionally shuts off diabetic ketoacidosis. The patient, however, maintains his
the pump completely when he sees downward trends. independence in the management of his diabetes and rou-
This example illustrates the limitations of RT-CGM in tinely changes the alarms back to their previsit settings.
overcoming certain psychological barriers in patients with After over a year of continuous use, the patient has been able
diabetes. to obtain a 0.6% decrease in A1c with sensor use, bringing
him almost to goal. Making the additional changes to the
Scenario 5. This is a 67-year-old man with a 20-year low and high targets, however, could lead to a further im-
history of late-onset T1DM. The patient was also thought to provement in glycemic control. This scenario illustrates that
be a good candidate for the sensor given an A1c of 7.6% some seasoned patients may have difficulty accepting pro-
despite frequent SMBG. The patient received adequate ed- vider input and marked changes in the approach to therapy,
ucation about use of the sensor and has become extremely limiting their ability to achieve maximal benefit. A patients
comfortable with making changes to the settings. Over willingness to make both self-directed and provider-initiated
CLINICAL EXPERIENCE WITH CGM IN ADULTS S-101

FIG. 5. Psychological barriers to optimal RT-CGM use. The example patient is a 31-year-old man with a severe hypogly-
cemic episode soon after his diagnosis of T1DM with a subsequent magnified fear of hypoglycemia. This severely limited his
willingness to implement changes suggested by medical staff and contributed to lack of improvement in glycemic control
with use of the sensor for 13 months: (a) CGM data before initiation of sensor and (b) CGM data after 13 months of sensor
use. BG, blood glucose.

changes should be assessed in determining appropriate Amylin, Hoffman-LaRoche, Intekrin, Lilly, Medtronic, Novo
candidates. Nordisk, Novartis, Osiris, Pfizer, National Institutes of
Health=National Heart Lung and Blood Institute, Tolerx,
Conclusions and Transition Therapeutics, honoraria fees from American
Academy of Phyiscan Assistants, Amylin, Novo Nordisk,
Limitations exist in evaluating the current randomized
Pfizer, Sanofi-Aventis, and Smiths Medical, consulting fees
trial data for the effect of RT-CGM on glycemic control. The
from Amylin, Astra-Zeneca, BMS, and Sanofi-Aventis, and
technology is rapidly evolving, but substantial barriers exist
lecture fees from Amylin, Novo Nordisk, and Smiths Medi-
to widespread implementation. Longer-term studies in more
cal. T.O. reports receiving lecturing fees from Amylin, Pfizer,
diverse practices and populations are needed, and neither
Sanofi-Aventis, Takeda, Medtronic, and DexCom. J.B. is a
cost-effectiveness nor impact on quality of life has been ade-
shareholder of Insulet and under contract with the University
quately studied. Steps, such as simplified treatment algo-
of North Carolina; he is an investigator for Amylin and
rithms, are needed to make implementation of the sensor
Medtronic and consultant to Abbott, Amylin, and Johnson
more conducive to an office setting with busy practitioners.
and Johnson. K.D. reports receiving grant support from Novo
However, it appears that with appropriate patient selection,
Nordisk, Diramed, and Tolerx and consulting fees from Eli
comprehensive education, and careful follow-up by expert
Lilly and Diramed and serves on the advisory board of Gly-
practitioners, adult patients who consistently use RT-CGM
comark.
can make changes in their treatment based on RT and retro-
spective trends, improve self-efficacy, and experience signif-
References
icant benefit.
1. Ong KL, Cheung BM, Wong LY, Wat NM, Tan KC, Lam
Author Disclosure Statement KS: Prevalence, treatment, and control of diagnosed dia-
betes in the U.S. National Health and Nutrition Ex-
No competing financial interests exist for K.F., M.D., C.I., amination Survey 19992004. Ann Epidemiol 2008;18:
and J.S. J.L. reports receiving grants or research support from 222229.
S-102 FABIATO ET AL.

2. Haller MJ, Stalvey MS, Silverstein JH: Predictors of control of from a 12-week study. Diabetes Technol Ther 2007;9:203
diabetes: monitoring may be the key. J Pediatr 2004;144:660 210.
661. 18. Hirsch IB, Abelseth J, Bode BW, Fischer JS, Kaufman FR,
3. Kempf K, Neukirchen W, Martin S, Kolb H: Self-monitoring Mastrototaro J, Parkin CG, Wolpert HA, Buckingham BA:
of blood glucose in type 2 diabetes: a new look at published Sensor-augmented insulin pump therapy: results of the first
trials. Diabetologia 2008;4:686688. randomized treat-to-target study. Diabetes Technol Ther
4. Poolsup N, Suksomboon N, Jiamsathit W: Systematic review 2008;10:377383.
of the benefits of self-monitoring of blood glucose on gly- 19. Yoo HJ, An HG, Park SY, Ryu OH, Kim HY, Seo JA, Hong
cemic control in type 2 diabetes patients. Diabetes Technol EG, Shin DH, Kim YH, Kim SG, Choi KM, Park IB, Yu JM,
Ther 2008;10(Suppl 1):S-51S-66. Baik SH: Use of real time continuous glucose monitoring
5. Hirsch IB, Bode BW, Childs BP, Close KL, Fisher WA, Gavin system as a motivational device for poorly controlled type 2
JR, Ginsberg BH, Raine CH, Verderese CA: Self-monitoring diabetes. Diabetes Res Clin Pract 2008;82:7379.
of blood glucose (SMBG) in insulin- and noninsulin-using 20. Irvine AA, Cox D, Gonder-Frederick L: Fear of hypoglyce-
adults with diabetes: consensus recommendations for im- mia: relationship to physical and psychological symptoms in
proving SMBG accuracy, utilization, and research. Diabetes patients with insulin-dependent diabetes mellitus. Health
Technol Ther 2008;10:419439. Psychol 1992;11:135138.
6. Hay LC, Wilmshurst EG, Fulcher G: Unrecognized hypo- 21. Boland E, Monsod T, Delucia M, Brandt CA, Fernando S,
and hyperglycemia in well-controlled patients with type 2 Tamborlane WV: Limitations of conventional methods of
diabetes mellitus: the results of continuous glucose moni- self monitoring of blood glucose: lessons learned from 3
toring. Diabetes Technol Ther 2003;5:1927. days of continuous glucose sensing in pediatric patients with
7. Bode BW, Schwartz S, Stubbs HA, Block JE: Glycemic type 1 diabetes. Diabetes Care 1991;24:18581862.
characteristics in continuously monitored patients with type 22. The Diabetes Research in Children Network (DirecNet)
1 and type 2 diabetes: normative values. Diabetes Care Study Group: Eight-point glucose testing versus the Con-
2005;28:23612366. tinuous Glucose Monitoring System in evaluation of gly-
8. FDA approves new glucose monitoring system for diabetics. cemic control in type 1 diabetes. J Clin Endocrinol Metab
http:==www.fda.gov=bbs=topics=NEWS=NEW00682.html 2005;90:33873391.
(accessed March 2009). 23. Larsen J, Ford T, Lyden E, Colling C, Mack-Shipman L, Lane
9. FDA approves new glucose test for adult diabetics. J: What is hypoglycemia in patients with well-controlled
http:==www.fda.gov=bbs=topics=NEWS=2001=NEW00758 type 1 diabetes treated by subcutaneous insulin pump with
.html (accessed March 2009). use of the continuous glucose monitoring system? Endocr
10. Ludvigsson J, Hanas R: Continuous subcutaneous glucose Pract 2004;10:324329.
monitoring improved metabolic control in pediatric patients 24. Evans ML, Pernet A, Lomas J, Jones J, Amiel SA: Delay in
with type 1 diabetes: a controlled crossover study. Pediatrics onset of awareness of acute hypoglycemia and of restoration
2003;111:933938. of cognitive performance during recovery. Diabetes Care
11. Tanenberg R, Bode B, Lane W, Levetan C, Mestman J, 2000;23:893897.
Harmel AP, Tobian J, Gross T, Mastrototaro J: Use of the 25. Garg S, Zisser H, Schwartz S, Bailey T, Kaplan R, Ellis S,
Continuous Glucose Monitoring System to guide therapy in Jovanovic L: Improvement in glycemic excursions with a
patients with insulin-treated diabetes: a randomized con- transcutaneous, real-time continuous glucose sensor. Dia-
trolled trial. Mayo Clin Proc 2004;79:15211526. betes Care 2006;29:4450.
12. Chico A, Vidal-Ros P, Subira M, Novials A: The continuous 26. Hirsch IB, Brownlee M: Should minimal blood glucose
glucose monitoring system is useful for detecting unrecog- variability become the gold standard of glycemic control?
nized hypoglycemia in patients with type 1 and type 2 di- J Diabetes Complications 2005;19:178181.
abetes but is not better than frequent capillary glucose 27. Ceriello A: Postprandial hyperglycemia and diabetes com-
measurements for improving metabolic control. Diabetes plications: is it time to treat? Diabetes 2005;54:17.
Care 2003;26:11531157. 28. Hirsch IB: Glycemic variability: its not just A1c anymore!
13. Reach G: Continuous glucose monitoring and diabetes Diabetes Technol Ther 2005;7:780783.
health outcomes: a critical appraisal. Diabetes Technol Ther 29. Cox DJ, Gonder-Frederick L, Ritterband L, Clarke W, Ko-
2008;10:6980. vatchev BP: Prediction of severe hypoglycemia. Diabetes
14. The Juvenile Diabetes Research Foundation Continuous Care 2007;30:13701373.
Glucose Monitoring Study Group: Continuous glucose mon- 30. Kilpatrick ES, Rigby AS, Goode K, Atkin SL: Relating mean
itoring and intensive treatment of type 1 diabetes. N Engl J blood glucose and glucose variability to the risk of multiple
Med 2008;359:14641476. episodes of hypoglycemia in type 1 diabetes. Diabetologia
15. JDRF CGM Study Group: JDRF randomized clinical trial to 2007;50:25532561.
assess the efficacy of real-time continuous glucose monitor- 31. Rigla M, Hernando ME, Gomez EJ, Brugues E, Garca-Saez
ing in the management of type 1 diabetes: research design G, Capel I, Pons B, de Leiva A: Real-time continuous glucose
and methods. Diabetes Technol Ther 2008;10:310321. monitoring together with telemedical assistance improves
16. Deiss D, Bolinder J, Riveline JP, Battelino T, Bosi E, Tubiana- glycemic control and glucose stability in pump-treated pa-
Rufi N, Kerr D, Phillip M: Improved glycemic control in tients. Diabetes Technol Ther 2008;10:194199.
poorly controlled patients with type 1 diabetes using real- 32. Kovatchev B, Clarke W: Continuous glucose monitoring
time continuous glucose monitoring. Diabetes Care (CGM) reduces risks for hypo-and hyperglycemia and glu-
2006;29:27302732. cose variability in diabetes [abstract 0086-OR]. http:==
17. Bailey TS, Zisser H, Garg SK: Reduction in hemoglobin professional.diabetes.org=Abstracts_Display.aspx?TYP1&
A1c with real-time continuous glucose monitoring: results CID54075 (accessed March 2009).
CLINICAL EXPERIENCE WITH CGM IN ADULTS S-103

33. Murphy HR, Rayman G, Lewis K, Kelly S, Johal B, Duffield 42. Hirsch IB, Armstrong D, Bergenstal RM, Buckingham B,
K, Fowler D, Campbell PJ, Temple RC: Effectiveness of Childs BP, Clarke WL, Peters A, Wolpert H: Clinical appli-
continuous glucose monitoring in pregnant women with cation of emerging sensor technologies in diabetes man-
diabetes: randomised clinical trial. BMJ 2008;337:907910. agement: consensus guidelines for continuous glucose
34. Byrne EZ, Zisser HC, Jovanovic L: Continuous glucose monitoring. Diabetes Technol Ther 2008;10:232244.
monitoring: is it helpful in pregnancy? Curr Diabetes Rev 43. Diabetes Research in Children Network (DirecNet) Study
2008;4:223226. Group: Use of the DirecNet Applied Treatment Algorithm
35. Chitayat L, Zisser H, Jovanovic L: Continuous glucose (DATA) for diabetes management with a real-time contin-
monitoring during pregnancy. Diabetes Technol Ther uous glucose monitor (the FreeStyle Navigator). Pediatr
2009;11(Suppl 1):S-105S-111. Diabetes 2008;9:142147.
36. Hermanns N, Kulzer B, Gulde C, Eberle H, Haak T: Patient 44. Meneghini E, Grieco G, Camerini M, Mion E, Brambilla MC,
reported outcomes of continuous glucose monitoring (CGM) Pizzi G, Fossati C, Disoteo O, Bonomo M: Determinants of
with real time access to glucose values vs. retrospective morning fasting hyperglycemia in type 1 diabetes mellitus:
analysis; a cross-over study [abstract 1883-P]. http:== analysis of 700 72-hour continuous glucose monitoring
professional.diabetes.org=Abstracts_display.aspx?TYP1& profiles [abstract 403-P]. http:==professional.diabetes.org=
CID69716 (accessed March 2009). Abstracts_Display.aspx?TYP1&CID70378 (accessed March
37. Bode B, Silver M, Weiss R, Martin K: Evaluation of a con- 2009).
tinuous glucose monitoring system for home-use conditions. 45. Zisser HC, Bevier WC, Jovanovic L: Restoring euglycemia in
Managed Care 2008;17:4045. the basal state using continuous glucose monitoring in
38. Bode B, Gross K, Rikalo N, Schwartz S, Wahl T, Page C, subjects with type 1 diabetes mellitus. Diabetes Technol Ther
Gross T, Mastrototaro J: Alarms based on real-time sensor 2007;9:509.
glucose values alert patients to hypo- and hyperglycemia: 46. Buckingham B, Wilson DM, Lecher T, Hanas R, Kaiserman
the Guardian continuous monitoring system. Diabetes K, Cameron F: Duration of nocturnal hypoglycemia prior to
Technol Ther 2004;6:105113. seizures. Diabetes Care 2008;31:21102112.
39. Garg SK, Voelmle MK, Gottlieb P: Feasibility of 10-day use
of a continuous glucose monitoring system in adults with Address reprint requests to:
type 1 diabetes. Diabetes Care 2009;32:436438. John Buse, M.D., Ph.D.
40. Cengiz E, Tamborlane WV: A tale of two compartments: Chief, Division of Endocrinology and Metabolism
interstitial versus blood glucose monitoring. Diabetes Tech- The University of North Carolina Chapel Hill
nol Ther 2009;11(Suppl 1):S-11S-16. 8027 Burnett Womack, CB 7172
41. Diabetes Research in Children Network (DirecNet) Study Chapel Hill, NC 25799
Group: Evaluation of factors affecting CGMS calibration.
Diabetes Technol Ther 2006;8:318325. E-mail: jbuse@med.unc.edu

You might also like