PRINCIPLE OF OPERATION THE Principle of operation of the quoted source shall be on phantom

AND CONTROL load source. Independent control variation shall been provided for all
OPERATION three phases(namely R-Y-B phases) for variation of parameters viz.,
voltage, current& Phase angle by using coarse and fine adjustments
variacs
Output voltage 3x300Vac(or better),50HzL-N
specifications Variable from 0 to 300v(by using both coarse and fine controls) for all
three phases, independently.
System shall be mounted with three numbers of 4.5 digit digital
voltmeter (fitted on the panel)to measure the output voltage by phase
wise. The accuracy of these voltmeter shall be 0.5%
Burden: 3x1000VA from 1A to 120A and 300VA from 0 to 1A for current
circuit
Predefined current settings : 1A,5A,10A,50A and 120A
Import &export mode Individual import /export selector switch for individual phase shall be
provided
Power factor range and Range: 0to 360deg(ZPF lag to UPF to ZPF lead) for each phase
adjustment controls Power factor adjustment shall be possible by coarse and fine control.
Predefined power factor settings: UPF 0.866 lag/lead,0.5lag/lead and
ZPF lag/lead.
Power factor System shall be mounted with three numbers of 4.5 digit digital power
measurement factor meter (fitted on the front panel), to measure the power factor by
phase wise, the accuracy of these power factor meters shall be 0.5%.
Frequency measurement System shall be mounted with three numbers of 4.5 digit digital
frequency meter (fitted on the front panel), to measure the power
factor by phase wise, the accuracy of these frequency meters shall be
0.5%.
Isolation current Suitable isolation current transformer ICT)with interconnecting cables
transformer shall be provided for all the three phases
Input supply requirements Three phase 4 wire 415V10%,48-52 Hz power supply,
--mains ON/OFF switch shall be provided.
--push to ON & push to OFF, illuminated push button switches for
voltage circuit.
--push to ON& push to OFF, illuminated push button switch for current
circuit
--emergency stop push button